SIST EN ISO 15587-1:2003
(Main)Water quality - Digestion for the determination of selected elements in water - Part 1: Aqua regia digestion (ISO 15587-1:2002)
Water quality - Digestion for the determination of selected elements in water - Part 1: Aqua regia digestion (ISO 15587-1:2002)
This part of ISO 15587 specifies a method for extracting trace elements from a water sample using aqua regia as a digestion agent. The method is applicable to all types of waters with a suspended solids mass concentration of less than 20 g/l and a mass concentration of total organic carbon (TOC) expressed as carbon of less than 5 g/l.
The aqua regia digestion method is empirical and it might not release elements completely. However, for most environmental applications, the result is fit for purpose.
Aqua regia digestion is suitable for the release of: Ag, Al, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, P, Pb, Sb, Se, Sn, Sr, Tl, V, Zn. It is not suitable for the digestion of refractory compounds such as SiO 2, TiO2 and Al2O3. The presence of chloride in the digestion solution may limit the application of analytical techniques.
The method is generic and may be implemented using a wide variety of equipment provided the digestion composition is unchanged, the digestion temperature is known, and the digestion duration is in accordance with this temperature.
Wasserbeschaffenheit - Aufschluss für die Bestimmung ausgewählter Elemente in Wasser - Teil 1:Königswasser-Aufschluss (ISO 15587-1:2002)
Dieser Teil der ISO 15587 legt ein Verfahren zur Königswasser-Extraktion von Spurenelementen aus einer Wasserprobe fest. Das Verfahren ist auf alle Wässer mit einer Massenkonzentration an suspendierten Stoffen kleiner 20 g/l und einer Massenkonzentration an gesamtem organisch gebundenem Kohlenstoff (TOC) kleiner 5 g/l anwendbar.
Der Königswasser-Aufschluss ist eine empirische Festlegung und es werden nicht unbedingt alle Elemente vollständig aufgeschlossen. Jedoch ist das Ergebnis für die meisten Fragestellungen im Bereich der Umweltanalyse geeignet.
Der Königswasser-Aufschluss ist geeignet für die Freisetzung von: Ag, Al, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, P, Pb, Sb, Se, Sn, Sr, Tl, V, Zn. Der Aufschluss ist nicht geeignet für refraktäre Verbindungen wie SiO2, TiO2 und Al2O3. Die Anwesenheit von Chlorid in der Aufschlusslösung kann die Anwendbarkeit von bestimmten Analysentechniken einschränken.
Das Verfahren ist universell anwendbar und kann unter Verwendung verschiedener Ausrüstungen durchgeführt werden, vorausgesetzt
3 die Zusammensetzung der Aufschlusslösung wird nicht verändert,
3 die Aufschlusstemperatur ist bekannt, und
3 die Aufschlussdauer und Temperatureinstellung sind angepasst.
die Aufschlussdauer und Temperatureinstellung sind angepasst.
Qualité de l'eau - Digestion pour la détermination de certains éléments dans l'eau - Partie 1: Digestion a l'eau régale (ISO 15587-1:2002)
L'ISO 15587-1 spécifie une méthode d'extraction d'éléments sous forme de traces dans un échantillon d'eau, en utilisant de l'eau régale comme agent de digestion. La méthode est applicable à tous les types d'eaux dont la concentration massique en solides en suspension est inférieure à 20 g/l et la concentration massique en carbone organique total (COT), exprimée en carbone, est inférieure à 5 g/l.
La méthode de digestion à l'eau régale est empirique et elle n'extrait pas toujours en totalité les éléments. Toutefois, pour la plupart des applications environnementales, le résultat répond aux besoins.
La digestion à l'eau régale permet la libération des éléments suivants: Ag, Al, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, P, Pb, Sb, Se, Sn, Sr, Tl, V, Zn. La méthode ne convient pas à la digestion de composés réfractaires comme SiO2, TiO2 et Al2O3. La présence de chlorures dans la solution de digestion peut limiter l'application des techniques analytiques.
La méthode est générique et peut être mise en oeuvre à l'aide d'une grande variété de matériels sous réserve que la composition de la digestion ne soit pas modifiée, que la température de digestion soit connue, et que la durée de digestion corresponde à cette température.
Kakovost vode - Razklop za določevanje izbranih elementov v vodi - 1. del: Razklop z zlatotopko (ISO 15587-1.2002)
General Information
Standards Content (Sample)
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Wasserbeschaffenheit - Aufschluss für die Bestimmung ausgewählter Elemente in Wasser - Teil 1:Königswasser-Aufschluss (ISO 15587-1:2002)Qualité de l'eau - Digestion pour la détermination de certains éléments dans l'eau - Partie 1: Digestion a l'eau régale (ISO 15587-1:2002)Water quality - Digestion for the determination of selected elements in water - Part 1: Aqua regia digestion (ISO 15587-1:2002)13.060.50VQRYLExamination of water for chemical substancesICS:Ta slovenski standard je istoveten z:EN ISO 15587-1:2002SIST EN ISO 15587-1:2003en01-maj-2003SIST EN ISO 15587-1:2003SLOVENSKI
STANDARD
SIST EN ISO 15587-1:2003
EUROPEAN STANDARDNORME EUROPÉENNEEUROPÄISCHE NORMEN ISO 15587-1March 2002ICS 13.060.50English versionWater quality - Digestion for the determination of selectedelements in water - Part 1: Aqua regia digestion (ISO 15587-1:2002)Qualité de l'eau - Digestion pour la détermination decertains éléments dans l'eau - Partie 1: Digestion à l'eaurégale (ISO 15587-1:2002)Wasserbeschaffenheit - Aufschluss für die Bestimmungausgewählter Elemente in Wasser - Teil 1:Königswasser-Aufschluss (ISO 15587-1:2002)This European Standard was approved by CEN on 1 March 2002.CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this EuropeanStandard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such nationalstandards may be obtained on application to the Management Centre or to any CEN member.This European Standard exists in three official versions (English, French, German). A version in any other language made by translationunder the responsibility of a CEN member into its own language and notified to the Management Centre has the same status as the officialversions.CEN members are the national standards bodies of Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece,Iceland, Ireland, Italy, Luxembourg, Malta, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and United Kingdom.EUROPEAN COMMITTEE FOR STANDARDIZATIONCOMITÉ EUROPÉEN DE NORMALISATIONEUROPÄISCHES KOMITEE FÜR NORMUNGManagement Centre: rue de Stassart, 36
B-1050 Brussels© 2002 CENAll rights of exploitation in any form and by any means reservedworldwide for CEN national Members.Ref. No. EN ISO 15587-1:2002 ESIST EN ISO 15587-1:2003
EN ISO 15587-1:2002 (E)2CORRECTED
2002-04-17ForewordThis document (ISO 15587-1:2002) has been prepared by Technical Committee ISO/TC 147"Water quality" in collaboration with Technical Committee CEN/TC 230 "Water analysis", thesecretariat of which is held by DIN.This European Standard shall be given the status of a national standard, either by publicationof an identical text or by endorsement, at the latest by September 2002, and conflictingnational standards shall be withdrawn at the latest by September 2002.According to the CEN/CENELEC Internal Regulations, the national standards organizations ofthe following countries are bound to implement this European Standard: Austria, Belgium,Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, Ireland, Italy,Luxembourg, Malta, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland and theUnited Kingdom.Endorsement noticeThe text of the International Standard ISO 15587-1:2002 has been approved by CEN as aEuropean Standard without any modifications.NOTE
Normative references to International Standards are listed in annex ZA (normative).SIST EN ISO 15587-1:2003
EN ISO 15587-1:2002 (E)3Annex ZA(normative)Normative references to International Publicationswith their relevantEuropean publicationsThis European Standard incorporates by dated or undated reference, provisions from otherpublications. These normative references are cited at the appropriate places in the text, and thepublications are listed hereafter. For dated references, subsequent amendments to or revisionsof any of these publications apply to this European Standard only when incorporated in it byamendment or revision. For undated references the latest edition of the publication referred toapplies (including amendments).NOTE
Where an International Publication has been modified by common modifications,indicated by (mod.), the relevant EN/HD applies.PublicationYearTitleEN/HDYearISO 36961987Water for analytical laboratory use —Specifications and test methodsEN ISO 36961995ISO 5667-31994Water quality – Sampling – Part 3: Guidanceon the preservation and handling of samplesEN ISO 5667-31995SIST EN ISO 15587-1:2003
SIST EN ISO 15587-1:2003
Reference numberISO 15587-1:2002(E)© ISO 2002
INTERNATIONAL STANDARD ISO15587-1First edition2002-03-01Water quality — Digestion for the determination of selected elements in water — Part 1: Aqua regia digestion Qualité de l'eau — Digestion pour la détermination de certains éléments dans l'eau — Partie 1: Digestion à l'eau régale
SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) PDF disclaimer This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat accepts no liability in this area. Adobe is a trademark of Adobe Systems Incorporated. Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
©
ISO 2002 All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO's member body in the country of the requester. ISO copyright office Case postale 56 • CH-1211 Geneva 20 Tel.
+ 41 22 749 01 11 Fax
+ 41 22 749 09 47 E-mail
copyright@iso.ch Web
www.iso.ch Printed in Switzerland
ii © ISO 2002 – All rights reserved
SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) © ISO 2002 – All rights reserved iii Contents Page Foreword.iv 1 Scope.1 2 Normative references.1 3 Terms and definitions.2 4 Principle.2 5 Reagents.3 6 Apparatus.3 7 Sampling.4 8 Procedure.4 9 Test report.6 Annex A (informative)
Digestion in an open system using electrical heating.7 Annex B (informative)
Digestion in an open system using microwave-assisted heating.9 Annex C (informative)
Digestion in a closed system using microwave-assisted heating.11 Annex D (informative)
Digestion in a closed system using an autoclave.15 Annex E (informative)
Performance checks.17 Bibliography.18
SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) iv © ISO 2002 – All rights reserved
Foreword ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 3. The main task of technical committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to the member bodies for voting. Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote. Attention is drawn to the possibility that some of the elements of this part of ISO 15587 may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent rights. ISO 15587-1 was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 2, Physical, chemical and biochemical method. ISO 15587 consists of the following parts, under the general title Water quality — Digestion for the determination of selected elements in water: Part 1: Aqua regia digestion Part 2: Nitric acid digestion Annexes A to E of this part of ISO 15587 are for information only.
SIST EN ISO 15587-1:2003
INTERNATIONAL STANDARD ISO 15587-1:2002(E) © ISO 2002 – All rights reserved 1 Water quality — Digestion for the determination of selected elements in water — Part 1: Aqua regia digestion WARNING — Persons using this part of ISO 15587 should be familiar with normal laboratory practice. This part of ISO 15587 does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions. 1 Scope This part of ISO 15587 specifies a method for extracting trace elements from a water sample using aqua regia as a digestion agent. The method is applicable to all types of waters with a suspended solids mass concentration of less than 20 g/l and a mass concentration of total organic carbon (TOC) expressed as carbon of less than 5 g/l. The aqua regia digestion method is empirical and it might not release elements completely. However, for most environmental applications, the result is fit for purpose. Aqua regia digestion is suitable for the release of: Ag, Al, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Hg, K, Mg, Mn, Mo, Na, Ni, P, Pb, Sb, Se, Sn, Sr, Tl, V, Zn. It is not suitable for the digestion of refractory compounds such as SiO2, TiO2 and Al2O3. The presence of chloride in the digestion solution may limit the application of analytical techniques. The method is generic and may be implemented using a wide variety of equipment provided the digestion composition is unchanged, the digestion temperature is known, and the digestion duration is in accordance with this temperature. 2 Normative references The following normative documents contain provisions which, through reference in this text, constitute provisions of this part of ISO 15587. For dated references, subsequent amendments to, or revisions of, any of these publications do not apply. However, parties to agreements based on this part of ISO 15587 are encouraged to investigate the possibility of applying the most recent editions of the normative documents indicated below. For undated references, the latest edition of the normative document referred to applies. Members of ISO and IEC maintain registers of currently valid International Standards. ISO 3696:1987, Water for analytical laboratory use — Specification and test methods ISO 5667-3:1994, Water quality — Sampling — Part 3: Guidance on the preservation and handling of samples SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) 2 © ISO 2002 – All rights reserved
3 Terms and definitions For the purposes of this part of ISO 15587, the following terms and definitions apply. 3.1 aqua regia solution obtained by mixing one volume of nitric acid with three volumes of hydrochloric acid 3.2 wet digestion treatment with reactive liquid chemicals with the aim of bringing analytes into solution so as to be suitable for subsequent analysis NOTE The degree to which analytes are brought into solution suitable for subsequent analysis may depend on the principle of the subsequent analysis. Some analytical methods require the element in solution to be independent of its chemical state, others require a specific valence or a specific ionic species, for instance hydride atomic spectrometry, photometry and electrochemical methods. For the latter, special attention should be paid to the presence of chemically stable organometals [5]. 4 Principle The test portion is digested with aqua regia at a defined temperature. The temperature is at least the boiling point (103 °C at 101,3 kPa) and at most 175 °C. At the boiling point (101,3 kPa), the minimum duration to release the digestible fraction is 120 min. The maximum duration is set at four times the minimum duration at that temperature. By convention, both the required minimum and maximum duration is assumed to halve with every 15 °C increase in temperature above the boiling point (see Figure 1). After digestion the digest is clarified, if necessary.
Figure 1 — Range of permitted digestion temperature and duration SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) © ISO 2002 – All rights reserved 3 5 Reagents During analysis, use only reagents of recognized analytical grade that meet the purity requirements of the subsequent analysis. Verify their purity by performing a blank test. 5.1 Water, complying with grade 1 of ISO 3696. Grade 2 may be used if it meets the purity requirements for the subsequent analysis. 5.2 Hydrochloric acid, c(HCl) = 12,0 mol/l, ρ = 1,2 kg/l, approximately 37 % by mass. 5.3 Nitric acid, c(HNO3) = 15,8 mol/l, ρ = 1,4 kg/l. Nitric acid is available both as ρ(HNO3) = 1,40 kg/l, approximately 65 % by mass and ρ(HNO3) = 1,42 kg/l, approximately 69 % by mass. Both are suitable. 5.4 Antifoaming agent, for instance n-dodecane (C12H26). 6 Apparatus 6.1 Digestion vessel, temperature- and pressure-resistant and capable of containing the mixture of sample and digest solution. If a closed vessel is applied, its inner wall shall be inert and shall not release substances to the digest in excess of the purity requirements of the subsequent analysis. 6.2 Vapour recovery system, capable of recovering the evaporated digest during open digestion (e.g. reflux condenser) and designed to minimize losses. Materials in contact with the vapour shall be inert and shall meet the purity requirements of the subsequent analysis. 6.3 Volatile species trap, in an open digestion system capable of trapping one or more volatile measurement species which may pass through the vapour recovery system (6.2). This trap may comprise condensation equipment or equipment to absorb these species in nitric acid or aqua regia. Materials in contact with the vapour shall be inert and shall meet the purity requirements of the subsequent analysis. 6.4 Heating device, capable of maintaining the target temperature for the required time period. 6.5 Filter paper, cellulose-based ashless type, with a median pore size adapted to the apparatus used for analysis. It shall not be able to release substances to the digest in excess of the purity requirements of the subsequent analysis and shall not absorb substances from the digest beyond the level of negligible impact on trueness of the analytical result. 6.6 Anti-bumping granules or roughened glass beads, having a diameter of 2 mm to 3 mm and acid-washed [for instance with warm nitric acid (5.3) diluted to 10 % by volume with water (5.1)]. 6.7 Volumetric flask, usually of nominal capacity of 50 ml or 100 ml. 6.8 Graduated pipettes or dispensers, of appropriate volume for the task. SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) 4 © ISO 2002 – All rights reserved
7 Sampling Handle and preserve the water sample in accordance with ISO 5667-3. The digestion procedure (see clause 8) is based on a test portion of 25,0 ml ± 0,1 ml taken from a well-shaken homogeneous water sample. Any test portion larger than 20 ml is allowed, provided that the volume of aqua regia is modified proportionally. If suspended solids make it impossible to deliver a representative test portion of a known volume, deliver the test portion otherwise, for instance by weighing. Calculate the volume from the determination of its mass, possibly corrected for the mass and density of solids. Report the uncertainty of the volume if it is higher than 0,1 ml for a volume of 25 ml. To lower the limit of detection, the preserved sample may be concentrated by evaporation. Evaporation is carried out after addition of nitric acid (5.3) and before addition of hydrochloric acid (5.2). However, the concentrated water sample shall meet the requirements of clause 1. Evaporation is not allowed if volatile analyte species might be present. 8 Procedure 8.1 Preparation Carefully acid-wash digestion equipment in contact with the digestion solution and volumetric ware [for instance with warm nitric acid (5.3) diluted to 10 % by volume with water (5.1)] and rinse with water (5.1). Add a test portion of 25,0 ml ± 0,1 ml to the digestion vessel (6.1). Add 6,0 ml ± 0,1 ml of hydrochloric acid (5.2) followed by 2,0 ml ± 0,1 ml of nitric acid (5.3). The boiling point of this mixture is equal to 103 °C. The volume ratio sample to aqua regia is equal to approximately three. A volume ratio up to four is allowed. If excessive foaming occurs, add one or more drops of antifoaming agent (5.4). Swirl and allow the mixture to stand until any visible reaction has stopped. If the digestion vessel is of the open type, proceed with 8.2. If not, proceed with 8.3. 8.2 Digestion in an open system Connect the digestion vessel (6.1) to the vapour recovery system (6.2). Connect the volatile species trap (6.3) so as to prevent volatile measurement species from passing through the vapour recovery system. Place the digestion vessel in the heating device (6.4). Increase the temperature to the boiling point. Maintain boiling for at least 120 min. As the boiling point decreases with altitude, add 20 min to the minimum duration for every 1 000 m above sea level. The maximum allowed duration equals four times the minimum duration. Once the digestion period is completed, allow the digestion vessel to cool. If the volatile species trap (6.3) is used, add the contents to the digestion vessel. SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) © ISO 2002 – All rights reserved 5 Rinse the trap and/or the condenser with water (5.1) and decant the washings. Collect the water in the digestion vessel. Dismount the digestion vessel from the digestion system. Proceed with 8.4. Examples of digestion in an open system are given in annexes A and B. 8.3 Digestion in a closed system Cap the digestion vessel (6.1) and weigh it. Connect the digestion vessel to the microwave equipment or place it onto the heating device (6.4). Increase the temperature of the digestion mixture slowly to temperature Td between the boiling point (103 °C at 101,3 kPa) and 175 °C. Maintain digestion for a period ∆t of at least 10 min and at most 480 min (101,3 kPa). Given the digestion period ∆t, expressed in minutes, the digestion temperature Td, expressed in degrees Celsius, shall meet the following condition: 206,6 − 21,64 × ln(∆t) < Td < 236,6 − 21,64 × ln(∆t) (1) NOTE 1 Boundary conditions in equation (1) reflect the assumption of clause 4 on release kinetics and duration time. The permitted domain of digestion temperature and duration is shown in Figure 1. Once the digestion period is completed, allow the digestion vessel to cool. Weigh the digestion vessel and accept the digest if the mass loss is consistent with the mass loss of a well documented reference material known to lead to acceptable results. Uncap and vent the digestion vessel in a fume hood. Examples of digestion in a closed system are given in annexes C and D. NOTE 2 Several microwave units control power rather than temperature. A procedure to establish the relationship between microwave power and temperature is given in C.6. 8.4 Sample transfer Decant the sample into a suitable acid-cleaned volumetric flask (6.7). In order to transfer the analytes quantitatively, rinse the digestion vessel (6.1) with water (5.1) and decant the washings. Collect the water in the same volumetric flask. If the transfer of particles that might interfere with the subsequent analysis cannot be avoided by simple decantation, decant the digest through a filter paper (6.5), collecting the filtrate in a suitable acid-cleaned volumetric flask (6.7). Wash the digestion vessel with water (5.1) and decant the washings through the filter paper (6.5). Collect the filtrate in the same volumetric flask. Fill the volumetric flask to the mark after addition of reagents required for subsequent sample handling and analysis. The extract is ready for determination. Examples of performance checks are given in annex E. SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) 6 © ISO 2002 – All rights reserved
9 Test report The test report can be prepared separately or in conjunction with the test report of the subsequent analytical method. Either of these shall include the following information: a) a reference to this part of ISO 15587, i.e. ISO 15587-1; b) the identity of the water sample; c) the volume of the test portion and its uncertainty if higher than 0,5 %; d) the conditions of digestion; e) any deviation from this method.
SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) © ISO 2002 – All rights reserved 7 Annex A (informative)
Digestion in an open system using electrical heating A.1 Principle A test portion of a water sample, which may contain up to 20 g/l suspended solids and up to 5 g/l organic carbon, is digested by boiling the test portion under reflux with aqua regia during a period between 120 min and 480 min. A.2 Reagents Use the reagents as specified in clause 5. A.3 Apparatus Usually, borosilicate glass is suitable for parts in contact with the digestion solution. If low concentrations of leachable elements are to be analysed, other construction materials may be used, for instance quartz (high temperatures) and polyethene or polypropene (low temperatures). Examples of leachable elements are B, Na, K and Al. A.3.1 Digestion vessel, made of borosilicate glass, and having a nominal volume of 100 ml. A.3.2 Reflux condenser, a straight-through type, with conical ground-glass joints and made of borosilicate glass. Water-cooled condensers with a minimum effective length of at least 200 mm have been found suitable. A.3.3 Roughened glass beads, having a diameter of 2 mm to 3 mm and acid-washed [for instance with warm nitric acid (5.3) diluted to 10 % by volume with grade 1 water (5.1)]. A.3.4 Temperature-controlled heating apparatus, capable of heating the contents of the digestion vessel to reflux temperature. A.3.5 Volumetric flask, made of borosilicate glass, and having a nominal volume of 100 ml. A.3.6 Graduated pipettes or dispensers. A.4 Sampling Preserve and handle water samples in accordance with ISO 5667-3. Take a test portion of 50,0 ml ± 0,2 ml from a well-shaken homogeneous water sample. If suspended solids make it impossible to deliver a representative test portion of a known volume, the test portion may be delivered otherwise, for instance by weighing. Calculate the volume from the determination of its mass, possibly corrected for the mass and density of solids. Report the uncertainty of the volume if it is higher than 0,2 ml. SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) 8 © ISO 2002 – All rights reserved
A.5 Procedure Carefully acid-wash the digestion equipment in contact with the digestion solution and volumetric ware [for instance with warm nitric acid (5.3) diluted to 10 % by volume with grade 1 water (5.1)]. Take a test portion of 50,0 ml ± 0,2 ml and transfer it to the digestion vessel (A.3.1). Add the roughened glass beads (A.3.3). Add 12,0 ml ± 0,2 ml of hydrochloric acid (5.2) followed by 4,0 ml ± 0,2 ml of nitric acid (5.3). If excessive foaming occurs, add one or more drops of antifoaming agent (5.4). Swirl and allow the mixture to stand until any visible reaction has stopped. Connect the digestion vessel to the reflux condenser (A.3.2). Place the digestion vessel in the temperature-controlled heating apparatus (A.3.4). Increase the temperature slowly until reflux conditions are reached, ensuring that the condensation zone is lower than one third of the height of the condenser. Maintain the boiling temperature for a period between 120 min and 480 min. As the boiling point (103 °C at 101,3 kPa) decreases with altitude, add 20 min to the minimum of 120 min for every 1 000 m above sea level. Once the digestion period is completed, allow the digestion vessel to cool. Rinse the reflux condenser with grade 1 water (5.1) and decant the washings. Collect the water in the digestion vessel. Dismount the digestion vessel from the digestion system. Decant the sample into the acid-cleaned volumetric flask (A.3.5). In order to transfer the analytes quantitatively, wash the digestion vessel with grade 1 water (5.1) and decant the washings. Collect the water in the same volumetric flask. Fill the volumetric flask to the mark after addition of reagents, which are required for subsequent sample handling and analysis. The extract is ready for determination. A.6 Test report This test report can be prepared separately or in conjunction with the test report of the subsequent analytical method. Either of these shall include the following information: a) a reference to this part of ISO 15587, i.e. ISO 15587-1; b) the identity of the water sample; c) the volume of the test portion and its uncertainty if higher than 0,5 %; d) the conditions of digestion; e) any deviation from this method.
SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) © ISO 2002 – All rights reserved 9 Annex B (informative)
Digestion in an open system using microwave-assisted heating B.1 Principle A test portion of a water sample, which may contain up to 20 g/l suspended solids and up to 5 g/l organic carbon, is digested by boiling the test portion under reflux in a flask heated by microwave unit with aqua regia during a period between 120 min and 480 min. B.2 Reagents Use the reagents specified in clause 5. Verify their purity by performing a blank test. B.3 Apparatus Usually, borosilicate glass is suitable for parts in contact with the digestion solution. If low concentrations of leachable elements are to be analysed, other construction materials may be used, for instance quartz (high temperatures) and polyethene or polypropene (low temperatures). Examples of leachable elements are B, Na, K and Al. B.3.1 Digestion vessel, made of borosilicate glass, and having a nominal volume of 50 ml. B.3.2 Reflux condenser, made of borosilicate glass. B.3.3 Roughened glass beads, having a diameter of 2 mm to 3 mm (or anti-bumping granules), acid-washed [for instance with warm nitric acid (5.3) diluted to 10 % by volume with grade 1 water (5.1)]. B.3.4 Microwave unit, consisting of a programmer and microwave module capable of heating the digestion vessel to the boiling point. B.3.5 Volumetric flask, made of borosilicate glass, and having a nominal capacity of 50 ml. B.3.6 Graduated pipettes or dispensers. B.4 Sampling Preserve and handle water samples in accordance with ISO 5667-3. Take a test portion of 25,0 ml ± 0,1 ml from a well-shaken homogeneous water sample. If suspended solids make it impossible to deliver a representative test portion of a known volume, the test portion may be delivered otherwise, for instance by weighing. Calculate the volume from the determination of its mass, possibly corrected for the mass and density of solids. Report the uncertainty of the volume if it is higher than 0,1 ml. SIST EN ISO 15587-1:2003
ISO 15587-1:2002(E) 10 © ISO 2002 – All rights reserved
B.5 Procedure Carefully acid-wash the digestion equipment in contact with the digestion solution and volumetric ware [for instance with warm nitric acid (5.3) diluted to 10 % by volume with grade 1 water (5.1)]. Take a test portion of 25,0 ml ± 0,1 ml and transfer it to the digestion vessel (B.3.1). Add the roughened glass beads (B.3.3). Add 6,0 ml ± 0,1
...