Abstract

ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute.
ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases:
oxygen;
nitrous oxide;
medical air;
helium;
carbon dioxide;
xenon;
specified mixtures of the gases listed above;
oxygen-enriched air;
air for driving surgical tools;
nitrogen for driving surgical tools;
use with vacuum.

Status
Not Published
Public Enquiry End Date
29-Nov-2026
Technical Committee
VAZ - Healthcare
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
06-Oct-2026
Due Date
23-Feb-2027

Buy Documents

Draft

oSIST prEN ISO 18082:2026

English language (15 pages)
Preview
Preview
e-Library read for
×1 day

oSIST prEN ISO 18082:2026 is a SIST standard for anaesthetic and respiratory equipment. It specifies the dimensions, allocation and marking of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases and vacuum, so connector interfaces stay gas-specific. It is aimed at manufacturers, designers, test labs and others who fit, inspect or verify medical devices and gas supply components.

What does oSIST prEN ISO 18082:2026 specify?

oSIST prEN ISO 18082:2026 covers NIST connectors used at operating pressures up to 1 400 kPa and for vacuum systems at pressures not greater than 60 kPa absolute. It applies to connectors for oxygen, nitrous oxide, medical air, helium, carbon dioxide, xenon, specified mixtures of those gases, oxygen 90+ (O 90+), air and nitrogen for driving surgical tools, and vacuum.

The document focuses on the connector interface itself. It leaves out manufacturer-supplied information and points users to the manufacturer of each medical device for operating information about permanently fitted connectors. The introduction also separates NIST from other connector systems such as DISS, SIS and terminal-unit quick connectors.

The standard is organized into Scope, Normative references, Terms and definitions, Dimensions and allocation of NIST connectors, Marking, Test for durability of markings, and a Bibliography.

What are the key requirements of oSIST prEN ISO 18082:2026?

oSIST prEN ISO 18082:2026 is mainly a dimensional and identification standard. The practical job is to make sure the wrong gas service cannot be connected by mistake, and that the connector can still be identified after cleaning and use.

Pressure range and gas service

Clause 1 limits the connector family to low-pressure gas services up to 1 400 kPa and vacuum up to 60 kPa absolute. Clause 1 also names the gases and mixtures covered. In practice, users must select the correct connector family for each gas service and not apply it outside those pressure conditions.

Allocation and geometry

Clause 4 assigns NIST connectors to gas services in Table 1 and states that left-hand threads have not been allocated. The body, nipple and nut dimensions are defined by Figures 1 to 4 and Tables 2 to 4. In practice, this is what keeps different gas services physically incompatible and gives manufacturers the dimensions to design, machine and inspect.

Verification method

Clause 4 says compliance is verified by measurement and visual inspection. That means dimensional checking is not optional and the connector identification must also be checked by eye. This is the basis for incoming inspection, production control and final release testing.

Marking

Clause 5 requires durable, legible marking with the gas symbol from Table 5, and it allows the gas name to be added. The marking must be readable to a person with corrected visual acuity of 1 from 0.5 m away at 215 lux. In practice, the marking has to remain visible under normal lighting and handling, not just when new.

Durability test for markings

Clause 6 gives a wipe test for marking durability: the markings are rubbed by hand with a cloth soaked in distilled water, then methylated spirit, then isopropyl alcohol, each for 15 s at ambient temperature. The marking is then checked for legibility. This gives manufacturers a simple way to show that identification survives routine cleaning.

Sealing parts

Table 4 gives the dimensions of the “O” rings used with the connector ranges. The figures also identify the position of the marking symbols and the assembly arrangement. For users, this means the sealing elements are part of the connector system and must match the defined sizes.

What terms does oSIST prEN ISO 18082:2026 define?

  • gas-specific - Designed so that connectors for different gas services or vacuum services cannot be connected together.
  • medical gas - A gas or gas mixture used for patient administration or as a source of pneumatic power for medical or surgical tools.
  • non-interchangeable screw-threaded connector (NIST connector) - A threaded connector family that uses different diameters and left- or right-hand threads to preserve gas-specificity.
  • NIST - The abbreviation used throughout the document for non-interchangeable screw-threaded connectors.

Who uses oSIST prEN ISO 18082:2026?

oSIST prEN ISO 18082:2026 is used by manufacturers of anaesthetic and respiratory equipment, by makers of medical devices with permanently fitted NIST connectors, and by engineers who design medical gas interfaces. It is also relevant to quality control and inspection teams that measure dimensions, check allocation and verify marking durability.

In practice, it applies to devices and systems such as pressure regulators, hose assemblies, flow-metering devices, lung ventilators and anaesthetic workstations, as well as supplies for surgical tools and vacuum services.

What changed in oSIST prEN ISO 18082:2026 from the previous edition?

oSIST prEN ISO 18082:2026 is the first edition to contain the requirements for NIST connectors for medical gases that were previously included in ISO 5359:2008 and ISO 5359:2008/Amd.1:2011. For users, that means the connector dimensions, allocation and marking are gathered in one dedicated document.

Which standards are used with oSIST prEN ISO 18082:2026?

The document cites and references several related standards and companion documents:

  • ISO 5359 - low-pressure hose assemblies for medical gases and vacuum.
  • ISO 4135:2022 - vocabulary for anaesthetic and respiratory equipment, including the definition of medical gas.
  • ISO 4135:2022/DAmd 1.2 - will add a definition for oxygen 90+.
  • ISO 7396-1:2016+Amd1:2017 - medical gas pipeline systems, including the gas-specific concept.
  • ISO 9170-1:2017 - terminal units for medical gas pipeline systems, a different connector system named for comparison.
  • ISO 48-2:2018 - hardness testing for rubber, used for the “O” ring notes.
  • BS 4518:1982+A2:2014 - metric dimensions for toroidal sealing rings (“O” rings) and housings.
  • CGA V-5:2019 and AS 2896:2021 - other non-interchangeable connector systems mentioned in the introduction.

What does the oSIST prEN ISO 18082:2026 document contain?

oSIST prEN ISO 18082:2026 is built around drawings, tables and a simple marking test. Figures 1 to 4 show the body, nipple, nut and assembled connector. Tables 1 to 3 allocate gas services and give the indexed connector dimensions, while Table 4 gives the “O” ring sizes and Table 5 gives the marking symbols.

Clause 6 provides the durability check for the markings, using water, methylated spirit and isopropyl alcohol. For readers, the document is mainly a design and inspection reference for connector geometry, identification and basic marking durability.

Relations

Effective Date
08-Oct-2025
Effective Date
08-Oct-2025

Buy Documents

Draft

oSIST prEN ISO 18082:2026

English language (15 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

oSIST prEN ISO 18082:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Anaesthetic and respiratory equipment - Dimensions of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases (ISO/DIS 18082:2026)". This standard covers: ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases: oxygen; nitrous oxide; medical air; helium; carbon dioxide; xenon; specified mixtures of the gases listed above; oxygen-enriched air; air for driving surgical tools; nitrogen for driving surgical tools; use with vacuum.

ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases: oxygen; nitrous oxide; medical air; helium; carbon dioxide; xenon; specified mixtures of the gases listed above; oxygen-enriched air; air for driving surgical tools; nitrogen for driving surgical tools; use with vacuum.

oSIST prEN ISO 18082:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

oSIST prEN ISO 18082:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 18082:2014, SIST EN ISO 18082:2014/A1:2018. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

oSIST prEN ISO 18082:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 1.1 This International Standard specifies the dimensions, the allocation and marking of non-
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 gas-specific, noun
  • 3.2 medical gas, noun
  • 3.3 non-interchangeable screw-threaded connector, noun
  • 4 Dimensions and allocation of NIST connectors
  • 1 NIST nut (see Figure 3)
  • 4 NIST nipple (see Figure 2)
  • 5 NIST body (see Figure 1)
  • 5 Marking
  • 6 Test for durability of markings
  • Bibliography

Introduction

This document has been prepared in response to the need for a safe method of connecting medical equipment intended to administer medical gases to patients or power medical devices. Medical gases are stored in cylinders or cryogenic vessels, or can be produced on site; several medical devices (e.g. pressure regulators, hose assemblies, flow-metering devices, lung ventilators, anaesthetic workstations) can be fitted between the source of supply and the medical device. At each interface gas-specific connectors are needed to ensure that the intended medical gas is administered to the patient.

While the desirability of achieving agreement on a single International Standard for screw-threaded connectors has never been in doubt, the present pattern of usage has made such agreement impossible. Nevertheless, fears that proliferation of individual national standards or practices will eventually result in potentially dangerous cross-connection between components for different gases have led to the choice of several different connector systems, all of which are intended to be incompatible with each other. This document specifies the dimensions and the allocation of non-interchangeable screw threaded (NIST) connectors for use with medical gases.

This document does not specify the dimensions and the allocation of [1] - diameter index safety system (DISS) connectors specified in CGA V-5:2019 , [2] - sleeve indexed system (SIS) connectors specified in AS 2896:2021 , and [3] - quick connectors designed for terminal units specified in ISO 9170-1:2017 .

oSIST prEN ISO 18082:2026

oSIST prEN ISO 18082:2026 DRAFT International Standard ISO/DIS 18082:2026(en) Anaesthetic and respiratory equipment — Dimensions of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases

1 Scope

1.1 This International Standard specifies the dimensions, the allocation and marking of non-

interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. 1.2 This International Standard specifies NIST connectors intended for use with the following medical gases:

  • oxygen;
  • nitrous oxide;
  • medical air;
  • helium;
  • carbon dioxide;
  • xenon;
  • specified mixtures of the gases listed above;
  • oxygen 90+ (O 90+);
  • air for driving surgical tools;
  • nitrogen for driving surgical tools; and for use with vacuum.

[4] NOTE 1 Low-pressure hose assemblies for medical gases and vacuum are specified in ISO 5359 . [5] NOTE 2 ISO 4135:2022/DAmd 1.2 will incorporate a definition for oxygen 90+. 1.3 The information to be supplied by the manufacturer is excluded from the scope of this International Standard because information about the use of NIST connectors is supplied by the manufacturer of each medical device to which the connectors are permanently fitted. NOTE Environmental aspects are dealt with in each International Standard concerning medical devices fitted with NIST connectors.

2 Normative references

There are no normative references in this document.

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

oSIST prEN ISO 18082:2026 ISO/DIS 18082:2026(en)

3.1 gas-specific, noun

having characteristics which prevent connections between different gas services or vacuum services [6] [SOURCE: ISO 7396-1:2016+Amd1:2 017 , 3.17]

3.2 medical gas, noun

gas or mixture of gases intended for administration to patients for anaesthetic, therapeutic, diagnostic or prophylactic purposes, or as a source of pneumatic power for medical or surgical tools [7] [SOURCE: ISO 4135:2022 , 3.1.1.20]

3.3 non-interchangeable screw-threaded connector, noun

NIST connector, noun range of threaded connectors intended to maintain gas-specificity by allocation of a set of different diameters and a left- or right-hand screw thread to mating components for each particular gas or service [7] [SOURCE: ISO 4135:2022 , 3.2.3.5]

4 Dimensions and allocation of NIST connectors

The dimensions of the NIST body, nipple and nut shall comply with Figure 1, Figure 2, Figure 3 and Figure 4 and Table 2, Table 3 and Table 4.

...