prEN ISO 18082
(Main)Anaesthetic and respiratory equipment - Dimensions of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases (ISO/DIS 18082:2026)
- Abstract
ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute.
ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases:
oxygen;
nitrous oxide;
medical air;
helium;
carbon dioxide;
xenon;
specified mixtures of the gases listed above;
oxygen-enriched air;
air for driving surgical tools;
nitrogen for driving surgical tools;
use with vacuum.
- Status
- Not Published
- Publication Date
- 16-Mar-2028
- Technical Committee
- CEN/TC 215 - Respiratory and anaesthetic equipment
- Drafting Committee
- CEN/TC 215/WG 1 - Anaesthetic machines and related equipment
- Current Stage
- 4020 - Submission to enquiry - Enquiry
- Start Date
- 24-Sep-2026
- Due Date
- 24-Sep-2026
- Completion Date
- 24-Sep-2026
prEN ISO 18082 is a draft CEN/ISO standard for anaesthetic and respiratory equipment. It specifies the dimensions, allocation and marking of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases and vacuum services, so manufacturers and system specifiers can match the right connector to the right gas service. It applies to connectors used at operating pressures up to 1 400 kPa and to vacuum systems at pressures not greater than 60 kPa absolute.
What does prEN ISO 18082 specify?
prEN ISO 18082 covers the geometry, gas-service allocation, marking and marking durability for NIST connectors. The document is arranged into six clauses: Scope, Normative references, Terms and definitions, Dimensions and allocation of NIST connectors, Marking, and Test for durability of markings, followed by a bibliography.
The scope ties the document to medical gases and to vacuum use in medical equipment. It also points readers to ISO 5359 for low-pressure hose assemblies used with medical gases and vacuum.
What are the key requirements of prEN ISO 18082?
prEN ISO 18082 is mainly about preventing misconnection by controlling the connector form, the assigned gas service and the identification marking.
- Pressure and service range - Clause 1 limits use to low-pressure medical gas connections up to 1 400 kPa and vacuum at no more than 60 kPa absolute. In practice, this keeps the document focused on the connector types used between cylinders, pipeline services and medical devices.
- Gas-specific allocation - Clause 4 and Table 1 assign each NIST connector range to a particular gas or service, such as oxygen, nitrous oxide, medical air, helium, carbon dioxide, xenon, specified mixtures, oxygen 90+ (O 90+), air or nitrogen for driving surgical tools, and vacuum. The allocation is what helps avoid cross-connection between incompatible services.
- No left-hand thread allocation - Table 1 states that left-hand threads have not been allocated. This matters because the thread direction is part of the gas-specific design and the standard keeps the system deliberately incompatible across services.
- Dimensional conformity - Clause 4 requires the NIST body, nipple and nut to comply with Figures 1 to 4 and Tables 2 to 4. Compliance is checked by measurement and visual inspection, so manufacturers need controlled dimensions and inspection methods, not just general fit.
- Sealing components - Clause 4 includes “O” ring dimensions in Table 4 and shows how they relate to the nipple design in Figure 2. The notes explain that the seal is chosen and compressed so the connector stays gas-tight and operates smoothly.
- Marking - Clause 5 requires durable, legible marking with the gas symbol from Table 5, and allows the gas name to be added. The marking must be readable by a person with corrected visual acuity from 0.5 m under 215 lux, which gives a practical readability target for production and inspection.
- Marking durability test - Clause 6 checks the marking by rubbing it with water, methylated spirit and isopropyl alcohol. In practice, this is a simple production or incoming-inspection test to confirm that the identification will survive normal handling and cleaning.
What terms does prEN ISO 18082 define?
- gas-specific - Having characteristics that prevent connections between different gas services or vacuum services.
- medical gas - A gas or gas mixture intended for patient administration or as a source of pneumatic power for medical or surgical tools.
- non-interchangeable screw-threaded connector (NIST connector) - A threaded connector system that keeps services separate by using different diameters and left- or right-hand threads for each gas or service.
- oxygen 90+ (O 90+) - A named oxygen service listed in the scope, allocation table and marking table.
- LFT - The symbol used in Table 5 for medical air/helium/carbon monoxide, identified in the table as lung function test.
Who uses prEN ISO 18082?
Manufacturers of anaesthetic and respiratory equipment use prEN ISO 18082 when designing and marking connectors fitted to medical devices. Test laboratories and quality teams use it to check dimensions, allocation and marking durability. Hospital engineering, procurement and regulatory staff use it when specifying or assessing medical gas equipment where cross-connection prevention matters.
What changed in prEN ISO 18082 from the previous edition?
This document is presented as the first edition covering NIST connectors for medical gases as a dedicated standard. The foreword says the requirements were previously included in ISO 5359:2008 and ISO 5359:2008/Amd.1:2011.
What does the prEN ISO 18082 document contain?
The document contains dimensional figures for the NIST body, nipple, nut and assembled connector. It also contains allocation tables for gas-specific connector ranges, tables for indexing diameters and “O” ring dimensions, and a marking table that links each gas or service to its symbol.
Clause 5 gives the marking rules, and Clause 6 gives the alcohol-and-water rub test used to confirm durability. The bibliography points to related documents for low-pressure hose assemblies, vocabulary, pipeline systems, terminal units and sealing-ring dimensions.
Relations
- Effective Date
- 08-Oct-2025
- Effective Date
- 08-Oct-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
prEN ISO 18082 is a draft published by the European Committee for Standardization (CEN). Its full title is "Anaesthetic and respiratory equipment - Dimensions of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases (ISO/DIS 18082:2026)". This standard covers: ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases: oxygen; nitrous oxide; medical air; helium; carbon dioxide; xenon; specified mixtures of the gases listed above; oxygen-enriched air; air for driving surgical tools; nitrogen for driving surgical tools; use with vacuum.
ISO 18082:2014 specifies the dimensions, the allocation and marking of non-interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. ISO 18082:2014 specifies NIST connectors intended for use with the following medical gases: oxygen; nitrous oxide; medical air; helium; carbon dioxide; xenon; specified mixtures of the gases listed above; oxygen-enriched air; air for driving surgical tools; nitrogen for driving surgical tools; use with vacuum.
prEN ISO 18082 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
prEN ISO 18082 has the following relationships with other standards: It is inter standard links to EN ISO 18082:2014/A1:2017, EN ISO 18082:2014. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
prEN ISO 18082 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 1.1 This International Standard specifies the dimensions, the allocation and marking of non-
- 2 Normative references
- 3 Terms and definitions
- 3.1 gas-specific, noun
- 3.2 medical gas, noun
- 3.3 non-interchangeable screw-threaded connector, noun
- 4 Dimensions and allocation of NIST connectors
- 1 NIST nut (see Figure 3)
- 4 NIST nipple (see Figure 2)
- 5 NIST body (see Figure 1)
- 5 Marking
- 6 Test for durability of markings
- Bibliography
Introduction
This document has been prepared in response to the need for a safe method of connecting medical equipment intended to administer medical gases to patients or power medical devices. Medical gases are stored in cylinders or cryogenic vessels, or can be produced on site; several medical devices (e.g. pressure regulators, hose assemblies, flow-metering devices, lung ventilators, anaesthetic workstations) can be fitted between the source of supply and the medical device. At each interface gas-specific connectors are needed to ensure that the intended medical gas is administered to the patient.
While the desirability of achieving agreement on a single International Standard for screw-threaded connectors has never been in doubt, the present pattern of usage has made such agreement impossible. Nevertheless, fears that proliferation of individual national standards or practices will eventually result in potentially dangerous cross-connection between components for different gases have led to the choice of several different connector systems, all of which are intended to be incompatible with each other. This document specifies the dimensions and the allocation of non-interchangeable screw threaded (NIST) connectors for use with medical gases.
This document does not specify the dimensions and the allocation of [1] - diameter index safety system (DISS) connectors specified in CGA V-5:2019 , [2] - sleeve indexed system (SIS) connectors specified in AS 2896:2021 , and [3] - quick connectors designed for terminal units specified in ISO 9170-1:2017 .
oSIST prEN ISO 18082:2026
oSIST prEN ISO 18082:2026 DRAFT International Standard ISO/DIS 18082:2026(en) Anaesthetic and respiratory equipment — Dimensions of non-interchangeable screw-threaded (NIST) low-pressure connectors for medical gases
1 Scope
1.1 This International Standard specifies the dimensions, the allocation and marking of non-
interchangeable screw-threaded (NIST) connectors intended to be used at operating pressures up to 1 400 kPa, and for vacuum systems at pressures not greater than 60 kPa absolute. 1.2 This International Standard specifies NIST connectors intended for use with the following medical gases:
- oxygen;
- nitrous oxide;
- medical air;
- helium;
- carbon dioxide;
- xenon;
- specified mixtures of the gases listed above;
- oxygen 90+ (O 90+);
- air for driving surgical tools;
- nitrogen for driving surgical tools; and for use with vacuum.
[4] NOTE 1 Low-pressure hose assemblies for medical gases and vacuum are specified in ISO 5359 . [5] NOTE 2 ISO 4135:2022/DAmd 1.2 will incorporate a definition for oxygen 90+. 1.3 The information to be supplied by the manufacturer is excluded from the scope of this International Standard because information about the use of NIST connectors is supplied by the manufacturer of each medical device to which the connectors are permanently fitted. NOTE Environmental aspects are dealt with in each International Standard concerning medical devices fitted with NIST connectors.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
oSIST prEN ISO 18082:2026 ISO/DIS 18082:2026(en)
3.1 gas-specific, noun
having characteristics which prevent connections between different gas services or vacuum services [6] [SOURCE: ISO 7396-1:2016+Amd1:2 017 , 3.17]
3.2 medical gas, noun
gas or mixture of gases intended for administration to patients for anaesthetic, therapeutic, diagnostic or prophylactic purposes, or as a source of pneumatic power for medical or surgical tools [7] [SOURCE: ISO 4135:2022 , 3.1.1.20]
3.3 non-interchangeable screw-threaded connector, noun
NIST connector, noun range of threaded connectors intended to maintain gas-specificity by allocation of a set of different diameters and a left- or right-hand screw thread to mating components for each particular gas or service [7] [SOURCE: ISO 4135:2022 , 3.2.3.5]
4 Dimensions and allocation of NIST connectors
The dimensions of the NIST body, nipple and nut shall comply with Figure 1, Figure 2, Figure 3 and Figure 4 and Table 2, Table 3 and Table 4.
...