EN ISO 80601-2-13:2022/A1:2026
(Amendment)Medical electrical equipment - Part 2-13: Particular requirements for basic safety and essential performance of an anaesthetic workstation - Amendment 1 (ISO 80601-2-13:2022/Amd 1:2026)
- Status
- Published
- Publication Date
- 18-Aug-2026
- Technical Committee
- CEN/TC 215 - Respiratory and anaesthetic equipment
- Drafting Committee
- CEN/TC 215/WG 1 - Anaesthetic machines and related equipment
- Current Stage
- 6060 - Definitive text made available (DAV) - Publishing
- Start Date
- 19-Aug-2026
- Due Date
- 19-Aug-2026
- Completion Date
- 19-Aug-2026
EN ISO 80601-2-13:2022/A1:2026 is CEN’s European Amendment 1 to the anaesthetic workstation standard EN ISO 80601-2-13:2022. It updates the particular requirements for basic safety and essential performance of anaesthetic workstations, with new or revised provisions for Oxygen 90+, system recovery, cybersecurity, alarms, connectors, AGSS interfaces and manufacturer information. It is written for manufacturers, test laboratories, technical assessors and buyers who need to know what has changed in the equipment requirements.
What does EN ISO 80601-2-13:2022/A1:2026 specify?
EN ISO 80601-2-13:2022/A1:2026 specifies amendments to the base anaesthetic workstation standard rather than a full standalone text. It changes selected clauses on safety, marking, information for use, monitoring, software, alarms, interfaces and annexes so that the equipment can be assessed against the updated requirements.
The document structure shown here includes the European foreword, amended clauses in the main body, Annex C on marking and information tables, Annex D on symbols, Annex AA on rationale, Annex CC on terms, and an updated bibliography.
| Annex | What it covers |
|---|---|
| Annex C | Marking and information tables linked to the amended clauses |
| Annex D | Symbols added for absorber on/off and reservoir bag connection port |
| Annex AA | Rationale for selected clauses, especially system recovery and interchangeable AGSS |
| Annex CC | List of added terms and abbreviations |
What are the key requirements of EN ISO 80601-2-13:2022/A1:2026?
EN ISO 80601-2-13:2022/A1:2026 mainly adds requirements for Oxygen 90+, system recovery, cybersecurity, alarm handling and interface safety. In practice, these changes affect how the workstation is marked, what the instructions for use must say, what is tested, and what is documented in the risk management file.
Manufacturer information, marking and instructions for use
Clause 201.7 ties the manufacturer’s information, marking, labelling and instructions for use to ISO 20417. That matters because the updated document expects the workstation’s user information to be structured and complete, not just technically correct.
The amendment adds a full set of requirements for Oxygen 90+ compatibility in clauses such as 201.5.101.4, 201.7.2.107, 201.7.9.2.101 and 201.101.7. A workstation and its components must be identified as compatible, conditionally compatible or incompatible with Oxygen 90+, and the instructions for use must state the gas type, any use conditions, the gas composition and the consequences of using Oxygen 90+. For a buyer or hospital engineer, that means Oxygen 90+ use is not an informal option - it must be declared and controlled.
System recovery after malfunction
Clause 201.4.3.101 introduces system recovery for anaesthetic workstations. After a malfunction, the workstation should try to restore essential performance, and if ventilation or gas delivery stops it must allow spontaneous breathing, make manual ventilation available within the stated time, and signal the recovery state by alarm.
This is important because the standard does not treat every fault as a full shutdown. It allows recovery logic where patient risk can be reduced by restarting the affected function, but the recovery duration, alarm type and any change in settings must be documented.
Protection against hazardous user actions
Clause 201.12.4.101 requires protection against accidental adjustment of controls that could create a hazardous situation. The listed cases include permanent deactivation of fresh gas flow, permanent deactivation of vapour delivery, standby activation and accidental switch-off of the workstation or its parts.
The protection method must be single fault safe and analysed in the risk management file. If the workstation has a manual configuration for Oxygen 90+, that configuration must also be protected from unintended use. In practice, manufacturers need a deliberate confirmation, lockout or similar control, and they need to justify it.
Pressure limitation, connectors and breathing system interfaces
Clauses 201.102 and 201.105 cover the patient connection port, reservoir bag connection port, access ports, sampling gas return ports and breathing tubes. The amendment tightens connector compatibility rules and markings, and it adds explicit anti-misconnection measures for operator-detachable parts.
This matters because anaesthetic workstations depend on correct gas-path connections. The document requires connectors that do not engage with the wrong ports, specific markings for the bag and exhaust-related interfaces, and breathing tubes that either cannot be wrongly assembled or use a fixed arrangement that makes misconnection detectable.
AGSS, transfer systems and receiving systems
Clause 201.103 and its annex updates cover AGSS, including interchangeable AGSS. The amendment defines an interchangeable AGSS, sets inlet pressure limits, adds marking and information duties, and gives connector rules for transfer systems, receiving systems and disposal hoses.
For users, the practical point is that the exhaust-gas route is now treated as a defined part of the workstation interface. If the workstation can work with an interchangeable AGSS, the interfaces, labels and instructions must make that clear, and the system must stay within the pressure limits given in the clause.
Monitoring, alarms, software and cybersecurity
Clause 201.12.4.104 requires anaesthetic ventilators to have a means to monitor exhaled volume. Annex C also adds rows for carbon dioxide monitoring equipment and oxygen monitoring equipment, and one clause requires airway pressure monitoring equipment to be fitted or provided for before service.
The amendment also tightens alarm handling. Clause 208.6.8.1 addresses how inactive alarm conditions are shown, and clause 208.6.12 requires operator alarm system logs for at least high and medium priority alarm conditions. That helps service teams and investigators track what happened and why an alarm was suppressed or inactive.
Clause 201.14.101 links software life cycle documentation to IEC 62304 for software safety class C, and clause 201.14.102 adds cybersecurity capabilities. It refers to risk controls from IEC/TR 60601-4-5 and sets expected Target Security Levels (SL-T) for user interfaces, remote clinical setting changes and secure updates. In practical terms, remote access must be authenticated, while local operator use is handled differently.
What terms does EN ISO 80601-2-13:2022/A1:2026 define?
- Essential function - a function or capability needed to maintain basic safety, essential performance, minimum clinical functionality and operational availability of the medical device.
- Firecall - an emergency access method for a secured medical device, also known as a break glass feature.
- Interchangeable AGSS - an anaesthetic gas scavenging system intended by design to be used with different models and manufacturers’ anaesthetic workstations.
- Oxygen 90+ (O 90+) - a gas mixture with at least 90% oxygen and specified maximum limits for argon, nitrogen, carbon dioxide, carbon monoxide, nitrogen oxides, sulfur dioxide, water and oil.
- Security level - the required set of countermeasures and intrinsic cybersecurity properties for a zone or conduit based on risk.
- Target Security Levels (SL-T) - the desired security level for a system, set through risk assessment.
- System recovery - automatic restart of a programmable electronic subsystem for part or all of the medical electrical equipment.
- Waste volatile anaesthetic agent capture system - a system that collects and stores waste volatile anaesthetic agents from the exhaust port of a breathing system.
Who uses EN ISO 80601-2-13:2022/A1:2026?
EN ISO 80601-2-13:2022/A1:2026 is used by anaesthetic workstation manufacturers, component suppliers and design engineers who must build equipment to the amended safety requirements. It is also used by test laboratories, certification teams and regulatory or quality managers who verify marking, alarms, connector compatibility, software evidence and risk management files.
Hospitals, biomedical engineers and procurement teams use it when they need to check whether a workstation supports Oxygen 90+, compatible AGSS arrangements, monitoring options and the required user information. Software and cybersecurity specialists also use it because the amendment links the workstation to software life cycle and cybersecurity expectations.
What changed in EN ISO 80601-2-13:2022/A1:2026 from the previous edition?
This amendment adds the Oxygen 90+ concept and its marking, compatibility statements and use instructions. It also adds system recovery requirements, new cybersecurity provisions, software life cycle guidance, alarm logging requirements and several new or revised interface rules for breathing systems, AGSS and connector types.
It updates the document to use ISO 20417 for manufacturer information and adds or revises references to standards used for breathing tubes, connectors, monitoring equipment and alarm-related functions. Annexes C, D, AA and CC are updated to match these changes.
Which standards are used with EN ISO 80601-2-13:2022/A1:2026?
- ISO 20417:2026 - sets the framework for information supplied by the manufacturer.
- IEC 62304:2006+A1:2015 - supports the software life cycle evidence expected for software safety class C.
- IEC/TR 60601-4-5:2021 - provides cybersecurity concepts used for essential function, firecall and SL-T.
- ISO 5356-1:2015 - supports the breathing system connector requirements.
- ISO 5367:2023 - supports breathing tubes and breathing sets.
- ISO 80369-7:2021 - supports Luer connectors for access ports.
- ISO 80601-2-55:2018 and ISO 80601-2-55:2018/Amd 1:2023 - provide related monitoring requirements for carbon dioxide and oxygen.
- ISO 9170-1:2017 and ISO 9170-2:2008 - support terminal unit and disposal hose connector requirements in transfer and receiving systems.
What does the EN ISO 80601-2-13:2022/A1:2026 document contain?
The amendment contains updated clause text, inspection checks and functional tests for oxygen compatibility, pressure limiting, alarms, connectors, software and cybersecurity. It also includes new or replaced figures for absorber bypass markings, connector arrangements, reservoir bag marking and AGSS inlet flow patterns.
Annex C gives the marking and information tables tied to the amended clauses, Annex D lists the added symbols, and Annex AA explains the technical background for selected changes such as system recovery and interchangeable AGSS. Annex CC adds the new terms used in the amendment, and the bibliography updates related suction equipment references.
Relations
- Effective Date
- 25-Aug-2026
- Effective Date
- 25-Aug-2026
- Effective Date
- 12-Feb-2026
- Effective Date
- 13-Mar-2024
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN ISO 80601-2-13:2022/A1:2026 is a amendment published by the European Committee for Standardization (CEN). Its full title is "Medical electrical equipment - Part 2-13: Particular requirements for basic safety and essential performance of an anaesthetic workstation - Amendment 1 (ISO 80601-2-13:2022/Amd 1:2026)". This standard covers: Medical electrical equipment - Part 2-13: Particular requirements for basic safety and essential performance of an anaesthetic workstation - Amendment 1 (ISO 80601-2-13:2022/Amd 1:2026)
Medical electrical equipment - Part 2-13: Particular requirements for basic safety and essential performance of an anaesthetic workstation - Amendment 1 (ISO 80601-2-13:2022/Amd 1:2026)
EN ISO 80601-2-13:2022/A1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
EN ISO 80601-2-13:2022/A1:2026 has the following relationships with other standards: It is inter standard links to EN 13718-1:2008, EN 1789:2007+A1:2010, ISO 80601-2-13:2022/Amd 1:2026, EN ISO 80601-2-13:2022. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN ISO 80601-2-13:2022/A1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- European foreword
- 1 Patient interface with a 15/22 mm coaxial connector
- 2 Reservoir bag [22 mm socket]
- 4 Reservoir bag port (22 mm socket)
- 5 Circle absorber assembly
- 6 Flexible bag arm or bag swivel mount with a 22 mm cone (direct connection to the reservoir
- 2 Inspiratory breathing tubes with 22 mm sockets on both ends
- 3 Expiratory breathing tubes with 22 mm sockets on both ends
- 4 Anaesthetic workstation
- Annex CC
- Bibliography
European foreword
This document (EN ISO 80601-2-13:2022/A1:2026) has been prepared by Technical Committee ISO/TC 121 "Anaesthetic and respiratory equipment" in collaboration with Technical Committee CEN/TC 215 “Respiratory and anaesthetic equipment” the secretariat of which is held by BSI. This Amendment to the European Standard EN ISO 80601-2-13:2022 shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by February 2027, and conflicting national standards shall be withdrawn at the latest by February 2027. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN shall not be held responsible for identifying any or all such patent rights. Any feedback and questions on this document should be directed to the users’ national standards body/national committee. A complete listing of these bodies can be found on the CEN website. According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and the United Kingdom.
Endorsement notice The text of ISO 80601-2-13:2022/Amd 1:2026 has been approved by CEN as EN ISO 80601-2- 13:2022/A1:2026 without any modification.
SIST EN ISO 80601-2-13:2022/A1:2026
SIST EN ISO 80601-2-13:2022/A1:2026 International Standard ISO 80601-2-13 Second edition Medical electrical equipment — 2022-04 Part 2-13:
AMENDMENT 1 Particular requirements for basic 2026-08 safety and essential performance of an anaesthetic workstation AMENDMENT 1 Appareils électromédicaux — Partie 2-13: Exigences particulières de sécurité de base et de performances essentielles pour les postes de travail d'anesthésie AMENDEMENT 1 Reference number
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd.1:2026(en)
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.
ISO copyright office CP 401 • Ch. de Blandonnet 8 CH-1214 Vernier, Geneva Phone: +41 22 749 01 11 Email: copyright@iso.org Website: www.iso.org Published in Switzerland
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) Foreword ISO (the International Organization for Standardization) and IEC (the International Electrotechnical Commission) form the specialized system for worldwide standardization. National bodies that are members of ISO or IEC participate in the development of International Standards through technical committees established by the respective organization to deal with particular fields of technical activity. ISO and IEC technical committees collaborate in fields of mutual interest. Other international organizations, governmental and non-governmental, in liaison with ISO and IEC, also take part in the work.
The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of document should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives or www.iec.ch/members_experts/refdocs).
ISO and IEC draw attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO and IEC take no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO and IEC had not received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents and https://patents.iec.ch. ISO and IEC shall not be held responsible for identifying any or all such patent rights. Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO’s adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www.iso.org/iso/foreword.html. In the IEC, see www.iec.ch/understanding-standards. This document was prepared jointly by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment, Subcommittee SC 1, Breathing attachments and anaesthetic machines, and Technical Committee IEC/TC 62, Medical equipment, software, and systems, Subcommittee SC 62D, Particular medical equipment, software, and systems, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/TC 215, Respiratory and anaesthetic equipment, in accordance with the Agreement on technical cooperation between ISO and CEN (Vienna Agreement). A list of all parts in the ISO 80601 and the IEC 80601 series can be found on the ISO and IEC websites. Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html and www.iec.ch/nationalcommittees.
SIST EN ISO 80601-2-13:2022/A1:2026
SIST EN ISO 80601-2-13:2022/A1:2026 AMENDMENT ISO 80601-2-13:2022/Amd 1:2026(en)
Medical electrical equipment — Part 2-13:
Particular requirements for basic safety and essential performance of an anaesthetic workstation AMENDMENT 1
Foreword Replace the text of the third dash by the following:
- consideration of anaesthetic workstations using Oxygen 90+;
201.2 Add the following reference:
ISO 20417:2026, Medical devices — Information to be supplied by the manufacturer 201.3 Add the following terms and definitions:
201.3.239 essential function function or capability that is required to maintain basic safety, essential performance, a minimum of clinical functionality as specified by the manufacturer, and operational availability for the medical device Note 1 to entry: Essential functions include, but are not limited to, the safety instrumented function (basic safety and essential performance), the control function and the availability of urgently needed functions and such allowing the operator to view and manipulate the medical device safely with the most urgently needed performance (operational availability). The loss of essential function is commonly termed loss of protection, loss of control and loss of view respectively.
Note 2 to entry: The term is derived from IEC 62443-4-2:2019, 3.1.20, and has been refined for the purpose and scope of this document.
[SOURCE: IEC/TR 60601-4-5:2021, 3.10]
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) 201.3.240 firecall method established to provide emergency access to a secure medical device Note 1 to entry: In an emergency situation, unprivileged users can gain access to key systems to correct the problem. When a firecall is used, there is usually a review process to ensure that the access was used properly to correct a problem. These methods generally either provide a one-time use user identifier (ID) or one-time password or other suitable measures.
Note 2 to entry: Also referred to as “break glass” feature.
[SOURCE: IEC/TR 60601-4-5:2021, 3.11] *201.3.241 interchangeable AGSS anaesthetic gas scavenging system that by design is intended to be used with different models and manufacturers’ anaesthetic workstations Note 1 to entry: Connections of interchangeable AGSSs can be non-operator detachable. 201.3.242 Oxygen 90+ (O 90+)
gas mixture that meets the following criteria for composition:
- Oxygen (O₂): volume fraction of no less than 90%
- Argon (Ar): volume fraction of no more than 5%
- Nitrogen (N₂): volume fraction of no more than 10%
- Carbon dioxide (CO₂): volume fraction of no more than 300 ppm
- Carbon monoxide (CO): volume fraction of no more than 10 ppm
- Nitrogen oxides (NOx): volume fraction of no more than 2 ppm
- Sulfur dioxide (SO₂): volume fraction of no more than 1 ppm
- Water (H₂O): volume fraction of no more than 67 ppm
- Oils: no more than 0.1 mg/m Note 1 to entry: Regional or national regulations can specify the colour coding for Oxygen 90+. Note 2 to entry: It is expected that the definition of Oxygen 90+ is added to ISO 4135. 201.3.243 security level level corresponding to the required set of countermeasures and inherent cybersecurity properties of devices and systems for a zone or conduit based on assessment of risk for the zone or conduit [SOURCE: IEC/TR 60601-4-5:2021, 3.23]
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) 201.3.244 Target Security Levels SL-T Target Security Levels are the desired level of security for a particular system. This is usually determined by performing a risk assessment on a system and determining that it needs a particular level of security to ensure its correct operation [SOURCE: IEC 62443-3-3:2013, 3.3] 201.3.245 system recovery method for fault handling via an automatic restart of programmable electronic subsystem (PESS) for parts of the ME equipment or for the complete ME equipment NOTE 1 to entry: There is guidance or rationale for this definition contained in Clause AA. [SOURCE: ISO 80601-2-12:2023, 201.3.298] 201.3.246 waste volatile anaesthetic agent capture system system that collects and stores waste volatile anaesthetic agents from the exhaust port of a breathing system 201.4 Add the following subclause:
201.4.3.101 System recovery NOTE 1 There is guidance or rationale for this subclause contained in Clause AA. a) Following a malfunction, the anaesthetic workstation should perform a system recovery to attempt to restore essential performance of the anaesthetic workstation. NOTE 2 An anaesthetic workstation or a subassembly function can become disturbed by a malfunction that can jeopardize the essential performance of the anaesthetic workstation. Without system recovery, the patient would have to be disconnected and connected to an alternative means of ventilation or gas delivery. Ventilation or gas delivery would thus be interrupted until an alternative means of ventilation or gas delivery is connected. System recovery, as specified in this subclause, attempts to automatically re-establish ventilation or gas delivery in a shorter period. If necessary, the anaesthetic workstation can be replaced later when it has fewer consequences for the patient´s therapy. b) System recovery can result in the temporary:
1) cessation in the ventilation or gas delivery leading to risk of harm; or 2) reduction in the function of anaesthetic workstation subassemblies without cessation of ventilation or gas delivery.
EXAMPLE The temporary blanking of the display.
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) c) During a system recovery with a cessation of ventilation or gas delivery that can lead to a hazardous situation:
1) the anaesthetic workstation shall allow spontaneous patient breathing. Manual ventilation shall be available within 15 s in accordance with 201.102.12; and NOTE 3 During a system recovery, there can be a short period where manual ventilation is not possible because the pneumatic system may have to be restarted by the anaesthetic workstation. 2) the anaesthetic workstation shall be equipped with an alarm system to indicate system recovery with a cessation of ventilation or gas delivery.
i) The alarm condition for a system recovery with a cessation of ventilation or gas delivery shall be high priority or may be an alternative alarm signal. The type of alternative alarm signal or alarm condition delay and alarm signal generation delay for the alarm signal shall be documented in the risk management file.
d) During a system recovery without a cessation of ventilation or gas delivery, the anaesthetic workstation shall be equipped with an alarm system to indicate system recovery without a cessation of ventilation or gas delivery.
1) The alarm condition for system recovery without a cessation of ventilation or gas delivery shall be at least low priority with an auditory alarm signal or may be an alternative alarm signal. The type of alternative alarm signal or alarm condition delay and alarm signal generation delay for the alarm signal shall be documented in the risk management file. e) Following a system recovery without operator intervention, the anaesthetic workstation shall attempt to operate with the same system configuration settings, ventilation or gas delivery settings and alarm settings as before the system recovery.
1) If ventilation settings, gas delivery settings or alarm settings are different after the system recovery, the anaesthetic workstation alarm system shall indicate any change in settings. 2) The alarm generated shall be at least medium priority.
f) The duration of a system recovery with a cessation of ventilation or gas delivery should be as short as practicable to avoid an unacceptable risk to the patient and shall be documented in the risk management file.
g) The maximum duration of a system recovery of ventilation or gas delivery shall be disclosed in the instructions for use.
Check conformance by inspection of the instructions for use, inspection of the risk management file and functional testing.
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) 201.4 Add the following subclause:
201.4.5 Alternative risk control measures or test methods for ME equipment or ME system Amendment (add prior to the compliance check):
aa) Subsequent revisions of dated references (new editions or amendments) may be used in substitution of a referenced document provided the manufacturer can demonstrate the hazard or hazardous situation addressed in the dated normative reference is adequately resolved in the subsequent revision.
201.4.10.101* Add the following new NOTE below NOTE 2:
NOTE 3 Anaesthetic workstations can provide the supply gas to drive other devices such as high flow therapy, oxygen flow meters or suction equipment. See also 201.7.9.2.101 qq). 201.5.101.4 Replace the text of the second dash:
with Oxygen 90+ if the anaesthetic workstation is rated compatible or conditionally compatible with Oxygen 90+.
Replace the third dash by a NOTE:
NOTE The concentration of oxygen and balance gases can vary over time in a gas supply with Oxygen 90+. 201.7 Add the following subclause:
201.7.1.101 Information to be supplied by the manufacturer The information supplied by the manufacturer including marking, labelling and instructions for use shall conform with ISO 20417.
Check conformance by application of ISO 20417.
201.7.2.21* Replace the existing text by:
This subclause does not apply.
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) 201.7.2.101 Delete this subclause completely.
201.7.2.104 Delete this subclause completely.
201.7.2 Add the following subclause:
201.7.2.107 Oxygen 90+ The anaesthetic workstation and its components shall be marked whether it is compatible, conditionally compatible or incompatible with Oxygen 90+. A workstation that can be used with Oxygen 90+ with restrictions shall be rated conditionally compatible with Oxygen 90+, see 201.7.9.2.101, ii).
An Oxygen 90+ compatibility marking shall be placed on the anaesthetic workstation, next to the gas supply connectors. A single marking may be sufficient to cover all components that are described in the instructions for use of the anaesthetic workstation. Additional markings can be required by risk management, e.g. to inform the operator.
NOTE Symbols for the marking of compatibility with Oxygen 90+ have been proposed for inclusion in ISO 7000.
201.7.9.1 Replace the existing text by:
This subclause does not apply.
201.7.9.2.101 Replace the text in list item ii) with the following:
ii) - the type of oxygen the anaesthetic workstation or its individual components are compatible with (oxygen or Oxygen 90+) and; - if the anaesthetic workstation or its components is rated conditionally compatible with Oxygen 90+, a description of the conditions under which it can be used with Oxygen 90+ and; - if the anaesthetic workstation or its components is rated conditionally compatible or compatible with Oxygen 90+, the gas composition of Oxygen 90+ as defined in 201.3.242; Replace the text in list item jj) with the following:
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) jj) if the anaesthetic workstation or its individual components is rated conditionally compatible or compatible for use with Oxygen 90+, the consequences of the use of Oxygen 90+, e.g. possible argon accumulation and varying oxygen concentrations; Add the following list items and NOTE below oo)*:
pp) list of any component or function that is pneumatically powered, the type of gas needed and its consumption of gas in l/min; qq) a description of those components that are pneumatically powered that cannot be used in parallel; NOTE If too many devices are used in parallel, the pressure supply to the anaesthetic workstation can be overloaded.
201.9.2.103* Delete this subclause completely.
201.9.2.104* Delete this subclause completely.
201.12.4.101* Replace the text in list item a) by the following:
a) A means of protection against accidental adjustment of operating controls that can create a hazardous situation shall be provided. This includes:
deactivating fresh gas flow to a patient permanently (i.e. not reactivating without user interaction); deactivating the anaesthetic vapour delivery system permanently (i.e. not reactivating without user interaction); activating the standby mode of an anaesthetic workstation; accidental turning off of the anaesthetic workstation or its individual components. The means for preventing an accidental adjustment shall be single fault safe. The means preventing the accidental adjustment as listed above shall be analysed and documented in the risk management file.
EXAMPLE 1 Requiring a confirmation step can be one method. The same confirmation step as for any other adjustment will possibly not be appropriate.
NOTE 1 Logging of each single operator adjustment leading to permanent deactivation of fresh gas flow or vapour delivery, activation of the standby mode or turning off the anaesthetic workstation or its individual components can help investigating adverse events.
SIST EN ISO 80601-2-13:2022/A1:2026 ISO 80601-2-13:2022/Amd 1:2026(en) NOTE 2 Detecting patient breathing activity during standby or when no fresh gas is being delivered to the patient and announcing an alarm signal (e.g. optical and acoustical with a high volume level) can help to prevent adverse outcomes in such situations.
Add list item d) before the conformance check:
d) If an anaesthetic workstation is equipped with a manual configuration for use with Oxygen 90+, the configuration shall be protected from unintended use.
...