Methods for the calibration of vibration and shock transducers — Part 44: Calibration of field vibration calibrators

This document specifies the instrumentation and procedure to be used for performing calibration of field vibration calibrators (FVCs). It is not applicable to FVCs used for the calibration of transducers. These are covered by ISO 16063‑21. Procedures and requirements of in situ calibration by FVC are beyond the scope of this document. Annex B provides more information on the application of FVC.

Méthodes pour l'étalonnage des transducteurs de vibrations et de chocs — Partie 44: Étalonnage des calibreurs de vibrations pour usage in situ

General Information

Status
Published
Publication Date
01-Nov-2018
Current Stage
9093 - International Standard confirmed
Start Date
08-Mar-2024
Completion Date
13-Dec-2025
Ref Project
Standard
ISO 16063-44:2018 - Methods for the calibration of vibration and shock transducers — Part 44: Calibration of field vibration calibrators Released:11/2/2018
English language
10 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


INTERNATIONAL ISO
STANDARD 16063-44
First edition
2018-11
Methods for the calibration of
vibration and shock transducers —
Part 44:
Calibration of field vibration
calibrators
Méthodes pour l'étalonnage des transducteurs de vibrations et de
chocs —
Partie 44: Étalonnage des calibreurs de vibrations pour usage in
situ
Reference number
©
ISO 2018
© ISO 2018
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2018 – All rights reserved

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Requirements for apparatus and other conditions . 1
4.1 General . 1
4.2 Reference transducer and amplifier . 2
4.3 Voltage measuring instrument for true RMS transducer chain output . 2
4.4 Frequency measuring instrument . 3
4.5 Distortion measuring instrument . 3
4.6 Dummy mass . 3
4.7 Environmental conditions . 3
5 Calibration procedure . 3
5.1 Preparation of calibration . 3
5.2 Calibration . 4
5.3 Other evaluation (optional) . 4
6 Calculation and expression of the result . 4
6.1 Calculation of the result . 4
6.2 Expression of the result . 5
Annex A (normative) Expression of uncertainty of measurement in calibration .6
Annex B (informative) Configuration and application of field vibration calibrators .8
Bibliography .10
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www .iso
.org/iso/foreword .html.
This document was prepared by Technical Committee ISO/TC 108, Mechanical vibration, shock and
condition monitoring.
A list of all the parts in the ISO 16063 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/members .html.
iv © ISO 2018 – All rights reserved

Introduction
ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and
shock transducers.
This document focuses on field vibration calibrators (FVCs). In this context, FVCs are mainly used for
in situ checks of vibration and shock transducers providing sine wave vibration at known frequency
and magnitude under field conditions. The FVC acts as a calibrated vibration source for in situ checks of
transducer sensitivity that is specified or requested, for example, in ISO 8041-1 or in ISO 8042.
Operational frequency and acceleration RMS value of FVCs are usually 160 Hz or 159,2 Hz, and 3,16 m/
2 2 2
s , 9,81 m/s or 10 m/s , respectively, which are widely accepted as reference conditions. There are
also FVCs with selectable acceleration magnitudes and frequencies. In comparison with stationary
calibration systems, FVCs have limitations in shaker power and inertial mass. Therefore, they can be
unsuitable for heavy test objects, high acceleration magnitudes and large displacements.
Using the calibration procedure described by this document, the acceleration generated by an FVC
can be traceable, through chain of calibration, to a primary or national standard as defined by ISO/
IEC Guide 99 (“the VIM”) with associated uncertainty defined by ISO/IEC Guide 98-3 (“the GUM”).
INTERNATIONAL STANDARD ISO 16063-44:2018(E)
Methods for the calibration of vibration and shock
transducers —
Part 44:
Calibration of field vibration calibrators
1 Scope
This document specifies the instrumentation and procedure to be used for performing calibration of
field vibration calibrators (FVCs).
It is not applicable to FVCs used for the calibration of transducers. These are covered by ISO 16063-21.
Procedures and requirements of in situ calibration by FVC are beyond the scope of this document.
Annex B provides more information on the application of FVC.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 2041, Mechanical vibration, shock and condition monitoring — Vocabulary
ISO 16063-21, Methods for the calibration of vibration and shock transducers — Part 21: Vibration
calibration by comparison to a reference transducer
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 2041 apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https: //www .iso .org/obp
— IEC Electropedia: available at http: //www .electropedia .org/
4 Requirements for apparatus and other conditions
4.1 General
Figure 1 shows the apparatus for the calibration. In some cases, reference transducer and amplifier
may be a single device as transducer chain.
Key
1 FVC
2 dummy mass
3 reference transducer
4 amplifier
5 distortion meter
6 voltmeter
7 frequency counter
8 fixing attachment (only for hand-held FVCs)
9 base
Figure 1 — Calibration apparatus
4.2 Reference transducer and amplifier
The reference transducer (preferably together with amplifier) shall be calibrated in accordance with
the comparison method given in ISO 16063-21 or other known methods with expanded uncertainty
of less than 2 % (magnitude). The uncertainty is the expanded uncertainty using a coverage factor of
2. When the FVC allows selectable acceleration magnitudes and frequencies, the reference transducer
shall be calibrated at each magnitude and frequency in advance.
The reference transducer shall have a low transverse sensitivity of preferably less than 2 % to minimize
the influence of the FVC's transverse vibration. The mass of the reference transducer shall be equal to,
or smaller than the capability of the FVC to be tested. Additional mass (dummy mass) may be employed
for evaluation of acceleration, distortion and other performances of the FVC.
4.3 Voltage measuring instrument for true RMS transducer chain output
A voltmeter measuring true RMS at the reference transducer chain output is used. Required expanded
uncertainty of voltmeter is 0,3 % at the selected reference frequency (coverage factor of 2).
2 © ISO 2018 – All rights reserved

4.4 Frequency measuring instrument
A frequency counter at the reference transducer chain output is used. Required expanded uncertainty
is 0,05 % at the selected reference frequency (coverage factor of 2).
4.5 Distortion measuring instrument
th
A distortion meter or an FFT analyser for total distortion
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...