Cement — Test methods — Part 2: Chemical analysis by X-ray fluorescence

ISO 29581-2:2010 describes a performance-based method for the chemical analysis of cement for SiO2, Al2O3, Fe2O3, CaO, MgO, SO3, K2O, Na2O, TiO2, P2O5, Mn2O3, SrO, Cl and Br using X-ray fluorescence (XRF). It can be applied to other relevant elements when adequate calibrations have been established. ISO 29581-2:2010 describes an alternative method for analyses of cement for conformity and information purposes, based on beads of fused sample and analytical validation using certified reference materials, together with performance criteria. A method based on pressed pellets of unfused sample can be considered as equivalent, providing that the analytical performance satisfies the same criteria. The use of fused beads generally improves the accuracy of analysis for non-volatile elements since it eliminates variability arising from differences in mineralogical forms or oxidation states. Pressed pellets generally improve the accuracy of analysis for volatile elements and can give adequate accuracy for the routine analysis of non-volatile elements. The presence of sulfide in a sample also leads to restrictions on the scope of the analysis that can be undertaken using the XRF technique based upon fused beads. In particular, sulfate (SO3) cannot be determined directly from such a fused bead because of the contribution to the analysis from the unknown amount of sulfide. In addition, sulfide cannot be determined directly (or accurately, indirectly) because of the contribution of the unknown amount of sulfate to the analysis and from the possibility that some sulfide can be lost by volatilization during fusion. Consequently, the method of ISO 29581-1, included as Annex D of ISO 29581-2:2010, is the reference method for determining the sulfate content of samples containing sulfide species. Other methods can be used, provided they are calibrated, either against the reference method or against internationally accepted reference materials, in order to demonstrate their equivalence. In the case of dispute, unless otherwise agreed by all parties, only the reference method in ISO 29581-1 can be used. ISO 29581-2:2010 describes methods that apply principally to cements, but which can also be applied to their constituent materials and to other materials, the standards for which call up these methods.

Ciments — Méthodes d'essais — Partie 2: Analyse chimique par spectrométrie de fluorescence X

General Information

Status
Published
Publication Date
17-Feb-2010
Technical Committee
Drafting Committee
Current Stage
9093 - International Standard confirmed
Completion Date
01-Dec-2020
Ref Project

Buy Standard

Standard
ISO 29581-2:2010 - Cement -- Test methods
English language
30 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 29581-2
First edition
2010-03-01


Cement — Test methods —
Part 2:
Chemical analysis by X-ray fluorescence
Ciments — Méthodes d'essais —
Partie 2: Analyse chimique par spectrométrie de fluorescence X





Reference number
ISO 29581-2:2010(E)
©
ISO 2010

---------------------- Page: 1 ----------------------
ISO 29581-2:2010(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


COPYRIGHT PROTECTED DOCUMENT


©  ISO 2010
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2010 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 29581-2:2010(E)
Contents Page
Foreword .iv
Introduction.v
1 Scope.1
2 Normative references.2
3 Terms and definitions .2
4 General requirements for testing.3
5 Reagents and reference materials.4
6 Apparatus.5
7 Preparation of a test sample of cement .6
8 Flux .7
9 Determination of loss on ignition and the change in mass on fusion of the cement .8
10 Factoring test results and correcting total analyses for presence of sulfides and halides.10
11 Preparation of fused beads and pressed pellets .12
12 Calibration and validation.14
13 Calculation and expression of results .23
14 Performance criteria (repeatability, accuracy and reproducibility limits) .24
Annex A (informative) Examples of fluxes .25
Annex B (informative) Sources of certified reference materials.26
Annex C (informative) Examples of calibration standards and monitor beads and pellets.27
Annex D (informative) Determination of the sulfate content of samples containing sulfide species.28
Bibliography.30

© ISO 2010 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 29581-2:2010(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 29581-2 was prepared by Technical Committee ISO/TC 74, Cement and lime.
ISO 29581 consists of the following parts, under the general title Cement — Test methods:
⎯ Part 1: Analysis by wet chemistry
⎯ Part 2: Chemical analysis by X-ray fluorescence
iv © ISO 2010 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 29581-2:2010(E)
Introduction
This part of ISO 29581 incorporates the following technical principles based on comments received by the
Secretariat.
a) It provides an analytical method based on X-ray fluorescence (XRF) for use as the alternative method for
the analysis of cement. When correctly calibrated according to the specified procedures and reference
materials, it provides a method of suitable precision for conformity and information purposes.
b) It introduces a reference method for TiO , P O , SrO and Br analysis.
2 2 5
c) Traceability of the method relies upon reference materials and “pure” chemicals so that the ultimate
traceability to basic international chemical standards relies upon classical analytical methods that are
outside of the scope of this part of ISO 29581.
XRF and other instrumental methods, such as differential thermal analysis for determination of carbon dioxide,
atomic absorption spectroscopy, etc., can be used as alternative methods, provided they are calibrated
against the reference methods, or against internationally accepted reference materials.

© ISO 2010 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 29581-2:2010(E)

Cement — Test methods —
Part 2:
Chemical analysis by X-ray fluorescence
1 Scope
This part of ISO 29581 describes a performance-based method for the chemical analysis of cement for SiO ,
2
Al O , Fe O , CaO, MgO, SO , K O, Na O, TiO , P O , Mn O , SrO, Cl and Br using X-ray fluorescence
2
2 3 2 3 3 2 2 2 5 2 3
(XRF). It can be applied to other relevant elements when adequate calibrations have been established.
This part of ISO 29581 describes an alternative method for analyses of cement for conformity and information
purposes, based on beads of fused sample and analytical validation using certified reference materials,
together with performance criteria.
A method based on pressed pellets of unfused sample can be considered as equivalent, providing that the
analytical performance satisfies the same criteria.
NOTE 1 The use of fused beads generally improves the accuracy of analysis for non-volatile elements, since it
eliminates variability arising from differences in mineralogical forms or oxidation states. Pressed pellets generally improve
the accuracy of analysis for volatile elements and can give adequate accuracy for the routine analysis of non-volatile
elements.
NOTE 2 The presence of sulfide in a sample also leads to restrictions on the scope of the analysis that can be
undertaken using the XRF technique based upon fused beads. In particular, sulfate (SO ) cannot be determined directly
3
from such a fused bead because of the contribution to the analysis from the unknown amount of sulfide. In addition, sulfide
cannot be determined directly (or accurately, indirectly) because of the contribution of the unknown amount of sulfate to
the analysis and because of the possibility that some sulfide can be lost by volatilization during fusion. Consequently, the
method of ISO 29581-1, included as Annex D to this part of ISO 29581, is the reference method for determining the sulfate
content of samples containing sulfide species.
Other methods can be used, provided they are calibrated, either against the reference method or against
internationally accepted reference materials, in order to demonstrate their equivalence.
In the case of dispute, unless otherwise agreed by all parties, only the reference method in ISO 29581-1 can
be used.
This part of ISO 29581 describes methods that apply principally to cements, but which can also be applied to
their constituent materials and to other materials, the standards for which call up these methods.
International Standard specifications state which methods can be used.
© ISO 2010 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 29581-2:2010(E)
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO Guide 30, Terms and definitions used in connection with reference materials
ISO Guide 31, Reference materials — Contents of certificates and labels
ISO 29581-1, Cement — Test methods — Part 1: Analysis by wet chemistry
EN 196-7, Methods of testing cement — Part 7: Methods of taking and preparing samples of cement
EN 197-1, Cement — Part 1: Composition, specifications and conformity criteria for common cements
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
beads
glassy discs of fused sample for analysis by X-rays in the spectrometer
3.2
pellets
compressed discs of finely ground sample for analysis by X-rays in the spectrometer
3.3
calibration beads or pellets
beads or pellets used for establishing the calibration equation
3.4
analysis beads or pellets
beads or pellets containing the sample being analysed
3.5
accuracy
closeness of agreement between a test result and the certified value for a reference material
3.6
repeatability
closeness of agreement among independent test results obtained with the same method on identical test
items in the same laboratory by the same operator using the same equipment within short intervals of time
3.7
reproducibility
closeness of agreement between independent test results obtained with the same method on identical test
items in different laboratories with different operators using different equipment
3.8
expert laboratory
laboratory capable of consistently meeting the expert performance criteria set out in Clause 14
3.9
normal laboratory
laboratory capable of consistently meeting the normal performance criteria set out in Clause 14
2 © ISO 2010 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 29581-2:2010(E)
4 General requirements for testing
4.1 Number of tests
Analysis of a cement can require the determination of a number of its chemical elements. For each
determination, one or more tests shall be carried out in which the number of measurements taken shall be as
specified in the relevant clause of this part of ISO 29581.
Where the analysis is one of a series subject to statistical control, the determination of each chemical element
by a single test shall be the minimum required.
Where the analysis method (including preparation and measurement) is checked at least once a week, as in
accordance with 12.5.1, a determination of each chemical element by a single test shall be the minimum
required. In the other cases, the number of tests for the determination of each chemical element shall be two;
see also Clause 13.
4.2 Accuracy and precision limits
4.2.1 Accuracy limit
The accuracy performance criterion in this part of ISO 29581 is measured as a limit on the closeness of
agreement between a test result and an accepted reference value for a certified reference material. The limits
for accuracy, expressed in percent absolute, are set out in Table 2; one set is appropriate to the performance
it is expected that an “expert” laboratory can achieve, whereas the other is appropriate for a “normal”
laboratory.
4.2.2 Repeatability limit
The repeatability performance criterion in this part of ISO 29581 is measured as a limit on the repeatability
where independent test results are obtained with the same method on identical test items (material) in the
same laboratory by the same operator using the same equipment within a short interval of time. The limits for
repeatability, expressed in percent absolute, are set out in Table 1; one set is appropriate to the performance
it is expected that an “expert” laboratory can achieve, whereas the other is appropriate to a “normal”
laboratory.
4.2.3 Reproducibility limit
The reproducibility performance criterion in this part of ISO 29581 is measured as a limit on the reproducibility
where test results are obtained with the same method on identical test items (material) in different laboratories
with different operators using different equipment. The limits for reproducibility, expressed in percent absolute,
are set out in Table 3; one set is appropriate to the performance it is expected that an “expert” laboratory can
achieve, whereas the other is appropriate to a “normal” laboratory.
4.2.4 Laboratory competence
The laboratory shall demonstrate that it can achieve the required performance in accordance with 12.3.3 and
12.3.4.
4.3 Expression of mass
Express mass in grams to the nearest 0,000 5.
4.4 Other methods
Other methods may be used, provided they are calibrated, either against the reference method or against
internationally accepted reference materials, in order to demonstrate their equivalence.
© ISO 2010 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 29581-2:2010(E)
5 Reagents and reference materials
5.1 Pure reagents
Reagents shall be of analytical quality and, wherever possible, pure oxides or carbonates, except for the
calibration of such elements as sulfur, chlorine, bromine or phosphorus, which do not form stable oxides or
carbonates, where some guarantee of stoichiometry is required.
Reagents shall be free of (or corrected for) the presence of water (and, in the case of oxides, carbon dioxide)
when weighed out for fusion. Also, the reagents shall be in a known oxidation state. The specified procedure
ensures that the correct oxidation state is obtained.
The reagents used to prepare the standard beads for cations shall be pure oxides or carbonates of at least
99,95 % purity (excluding moisture or CO ).
2
Reagents shall be used in a known stoichiometry in terms of content. In order to achieve this, they can be
treated before use as follows.
a) Determine the loss on ignition for silicon dioxide (SiO ), aluminium oxide (Al O ) and magnesium oxide
2 2 3
(MgO) as follows.
1) Ignite the reagent at, for example, (1 175 ± 25) °C and maintain at this temperature for 30 min.
2) Cool in a desiccator to room temperature and reweigh.
3) After allowing for this loss, weigh the appropriate amount of the unignited material to prepare the
bead.
b) Dry manganese oxide (Mn O ) and titanium(IV) oxide (TiO ) as follows.
2 3 2
1) Ignite the reagent at, for example, (1 000 ± 25) °C and maintain at this temperature for 30 min.
2) Cool in a desiccator to room temperature before use.
c) Dry iron (III) oxide (Fe O ) as follows.
2 3
1) Ignite the reagent at, for example, (700 ± 25) °C and maintain at this temperature for 30 min.
2) Cool in a desiccator to room temperature before use.
d) Dry calcium carbonate (CaCO ), strontium carbonate (SrCO ), potassium carbonate (K CO ) and sodium
3 3 2 3
carbonate (Na CO ).
2 3
1) Heat the reagent at, for example, (250 ± 10) °C and maintain at this temperature for 2 h.
2) Cool in a desiccator to room temperature before use.
5.2 Reference materials
5.2.1 Certified reference materials
Certified reference materials (CRMs) are materials, e.g. cement, supplied by an organization conforming to
the requirements for the competence of reference material producers in accordance with ISO Guide 30.
CRMs shall be supplied with a certificate of analysis giving information on the average value and standard
deviation in accordance with ISO Guide 31.
4 © ISO 2010 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 29581-2:2010(E)
5.2.2 Industrial reference materials
Industrial reference materials (IRMs) are materials, e.g. cement, prepared and homogenized by a laboratory.
The reference analysis of an IRM shall be the average result from inter-laboratory co-operative testing
involving at least four laboratories able to meet the performance criteria given in 12.3.
5.3 Calibration standards
Calibration standards are prepared in the laboratory from pure, analytical-grade reagents, IRMs, CRMs or a
combination of these. They shall be formulated to provide a series of calibration standards covering the range
of maximum to minimum values for each element being analysed and shall be evenly distributed between
those limits. The variation in concentrations of the elements shall be independent of each other. There shall
be a minimum of seven calibration standards in a series.
5.4 Binding agent
A binding agent, e.g. wax, whose influence on the elements being analysed has been determined, is used in
the grinding of samples during the preparation of pressed pellets. Carry out a pellet-preparation monitoring
check (see 12.5) whenever the batch of binding agent is changed.
6 Apparatus
6.1 Balance, capable of weighing to an accuracy of ± 0,000 5 g.
6.2 Fusion vessels and casting moulds, of a non-wetted platinum alloy, such as Pt/5 % Au or Pt/Rh.
Vessels that serve both as a fusion vessel and as a casting mould (i.e. a combined fusion mould) may be
used. If moulds become distorted in use, then they shall be reshaped by pressing in a suitable former. If the
bottom (flat) surface of the bead is used for analysis, it is necessary that the internal base of the mould also be
kept flat and free from blemishes.
NOTE Cleanliness of fusion vessels is important in achieving accurate analyses. This can be achieved, for example,
by boiling in dilute hydrochloric acid, 1:10 by volume or citric acid, 100 g/l.
6.3 Lids, optional, of a platinum alloy (not necessarily non-wetted).
6.4 Furnace, e.g. an electric resistance, muffle or high-frequency induction furnace, capable of operating at
(250 ± 10) °C, (700 ± 25) °C, (950 ± 25) °C, (1 000 ± 25) °C and (1 175 ± 25) °C.
6.5 Automatic fusion apparatus, for use in automatic bead preparation (see 11.4).
An automatic fusion apparatus may be used, provided that the performance criteria in 12.3 can be met.
6.6 Cooling apparatus, consisting of any device, such as a narrow jet of air that can be directed to the
centre of the base of the casting mould (for example, by the base of a bunsen burner without a barrel) or a
water-cooled metal plate.
NOTE Normally, cooling in air is sufficient but some difficult samples can require a cooling apparatus in order to cool
the melt rapidly. This is necessary to obtain a homogeneous bead and to free the melt from the casting mould.
6.7 Heat reservoir, for the casting mould, required in special circumstances when using moulds of small
sizes, so that the mould does not cool too rapidly when removed from the furnace.
6.8 Spectrometer, X-ray fluorescence, capable of meeting the performance criteria given in 12.3.
NOTE It is required to set appropriate measuring conditions to satisfy the performance criteria based on the type of
samples, the type of apparatus, elements being analysed and their content, etc.
© ISO 2010 – All rights reserved 5

---------------------- Page: 10 ----------------------
ISO 29581-2:2010(E)
6.9 Flow gas, maintained at as constant a room temperature as possible.
The temperature of the flow gas cylinder and of the connecting pipework is critical in order to prevent drift in
sensitivity of the flow proportional counters. Pipework shall be as short as practical and run, whenever
possible, within the temperature-controlled room housing the spectrometer. Where this is not possible, the
cylinder shall be kept in a temperature-controlled cabinet (room temperature ± 2 °C) or otherwise maintained
at a constant room temperature. For the same reason, new cylinders shall be allowed to equilibrate for at least
2 h to room temperature before use.
NOTE 1 The flow gas is used in the gas flow proportional counter of the XRF spectrometer.
NOTE 2 The composition of gas can change as the cylinder becomes exhausted. Cylinders should be replaced before
they become completely empty.
6.10 Grinding equipment, capable of grinding the sample, with binding agent if necessary, to a suitable
fineness.
6.11 Pellet press, capable of applying a pressure suitable for production of pellets with a consistent,
consolidated surface to meet the performance requirements given in 12.3.
6.12 Mould, usually of steel, of suitable strength to withstand the press without distortion and of suitable size
to produce a pellet to fit the spectrometer.
7 Preparation of a test sample of cement
Before chemical analysis, treat the laboratory sample, taken in accordance with EN 196-7, as follows to obtain
a homogeneous test sample.
a) Take approximately 100 g of the laboratory sample by means of a sample divider or by quartering.
b) Sieve this portion on a 150 µm or 125 µm sieve until the residue remains constant.
c) Remove metallic iron from the material retained on the sieve by means of a magnet (see Note 1).
d) Grind the iron-free fraction of the retained material so that it completely passes the 150 µm or 125 µm
sieve.
e) Transfer the sample to a clean, dry container with an airtight closure and shake vigorously to mix it
thoroughly.
f) Carry out all operations as quickly as possible to ensure that the test sample is exposed to ambient air for
only the minimum time.
NOTE 1 Where the analysis is one of a series subject to statistical control and the level of the metallic iron content has
been shown to be insignificant in relation to the chemical properties being determined, it is not necessary to remove
metallic iron. Where the level of metallic iron is significant, it is required to record and report the amount in the results.
NOTE 2 Where the sample contains quartz, it can be necessary to grind the sample to pass a 90 µm sieve in order to
obtain a satisfactory fusion (see Clause 11). The time and temperature required to obtain a satisfactory fusion is affected
by the fineness of the sample.
NOTE 3 Where pressed pellets are used, accuracy can be improved by grinding the sample more finely.
6 © ISO 2010 – All rights reserved

---------------------- Page: 11 ----------------------
ISO 29581-2:2010(E)
8 Flux
8.1 Choice of flux
8.1.1 General
One of the advantages of the XRF fused cast bead method is that a wide variety of fluxes may be chosen. For
a given calibration, the same flux shall be used throughout. The conditions given in 8.1.2 to 8.1.4 shall be met
for any flux used.
NOTE 1 Fluxes used with success in the analysis of cement are given in Annex A. Pre-fused fluxes have the
advantage of a lower moisture content.
NOTE 2 Reducing the particle size of the flux has been shown to improve fusion at a given temperature.
8.1.2 Dissolution
Under the conditions of preparation used, the sample shall be totally dissolved by the flux and shall not come
out of solution in the casting procedure.
8.1.3 Heavy-element absorber
A heavy-element absorber, such as lanthanum or vanadium oxide, may be incorporated into the flux, provided
⎯ it does not reduce sensitivities to the point that the performance criteria given in 12.3 cannot be met;
⎯ the heavy element does not have a line overlap with any of the elements being determined.
NOTE Lanthanum oxide assists the formation and stabilization of glass but reduces the intensity of the emitted
X-rays.
WARNING — There are restrictions on the use of heavy-metal chemicals in some countries. Care
should be taken in the handling of these and national safety rules observed.
8.1.4 Flux purity
The flux shall be pure with respect to the elements being determined.
Most reagents sold as “flux” grade quality by reputable manufacturers meet this requirement, but an analysis
shall be obtained for each batch of flux supplied. Carry out a bead-preparation monitoring check (see 12.5)
when a batch of flux is changed.
8.2 Moisture in flux
Pre-molten fluxes having a loss on ignition not exceeding 0,50 % mass fraction are preferred. However, if the
flux contains moisture, it shall be dried at a suitable temperature.
8.3 Flux-to-sample ratio
The flux-to-sample ratio selected shall be such that the performance criteria given in 12.3 are met. The mass
fraction ratio, R, of flux to sample used for the calibration shall be used for subsequent analyses.
NOTE As the flux-to-sample mass fraction ratio is greater than one, impurities in the flux can have a negative
influence on the measured result. The greater the ratio of the flux to sample, the greater the influence.
The total mass of sample and flux shall be chosen for the particular casting mould type used, and this mass
shall always remain the same.
© ISO 2010 – All rights reserved 7

---------------------- Page: 12 ----------------------
ISO 29581-2:2010(E)
8.4 Anti-wetting agent
A small amount of anti-wetting agent may be used, if necessary. An anti-wetting agent, such as lithium
bromide, ammonium bromide, lithium iodide, lithium iodate or ammonium iodide, may be added to the melt to
assist in preventing the cracking of the fused beads on cooling and to aid in the release from the mould.
Where an anti-wetting agent is used, all beads shall be prepared using the same anti-wetting agent added in
the same quantity and at the same stage of bead preparation.
NOTE Bromine or iodine in the anti-wetting agent can remain in the bead under some fusing conditions. It is required
to check for residual bromine or iodine, since these elements can cause line overlapping, such as Br Lα on Al Kα or I Lβ
2
on Ti Kα.
9 Determination of loss on ignition and the change in mass on fusion of the cement
9.1 Principle
In order to be able to total (to 100 % mass fraction) any oxide analysis of cement, a loss on ignition, i.e. the
amount of any combined water and carbon dioxide, is required. In addition, in order to be able to convert an
oxide analysis obtained on the fused-basis using fused-bead XRF, to an oxide analysis on the as-received
basis, an “observed” loss on ignition is also required.
NOTE Where the sample contains no oxidizable species, the loss on ignition and the “observed” loss on ignition are
the same.
The “observed” loss on ignition (see 9.3.1) is a very close approximation to the “change in mass on fusion”
that occurs when a sample is prepared as a fused bead for analysis by XRF. This “observed” loss on ignition
is used in this method to calculate a factor, f, (see Clause 10) to convert test results obtained on the fused
basis to the as-received basis.
The traditional loss-on-ignition determination carried out in an oxidizing atmosphere by igniting in air can be
used to determine both the loss on ignition and the “observed” loss on ignition. Where any oxidizable species
are present, in particular sulfide or sulfur-containing species, a correction can be applied to the “observed”
loss in order to derive a “corrected” loss on ignition for use in totalling the oxide analysis. However, any error
resulting from the oxidation of any metallic iron, bivalent iron or bivalent manganese is usually considered to
be negligible and only the correction for the extent of oxidation of sulfides is applied in the correction.
An alternative method, e.g. automatic equipment, may be used, provided that it can be demonstrated that the
performance criteria given in 12.3 are satisfied.
9.2 Procedure
Weigh, to ± 0,000 5 g, (1,00 ± 0,05) g of cement into
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.