IEC 60601-2-16:2008
(Main)Medical electrical equipment - Part 2-16: Particular requirements for basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment
Medical electrical equipment - Part 2-16: Particular requirements for basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment
IEC 60601-2-16:2008 applies to the basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment. IEC 60601-2-16:2008 does not take into consideration the dialysing fluid control system of Haemodialysis equipment using regeneration of dialysing fluid and central delivery systems. It does however take into consideration the specific safety requirements of such Haemodialysis equipment concerning electrical safety and patient safety. IEC 60601-2-16:2008 specifies the minimum safety requirements for haemodialysis equipment. These devices are intended for use either by medical staff or for use by the patient or other trained personnel under the supervision of medical expertise. IEC 60601-2-16:2008 includes all medical electrical equipment that is intended to deliver a haemodialysis, haemodiafiltration and haemofiltration treatment to a patient suffering from kidney failure. This third edition cancels and replaces the second edition of IEC 60601-2-16, published in 1998. This edition constitutes a technical revision. Changes since the previous edition include, among others, a summary of additional essential performance requirements, revision of terms and definitions, classification of applied parts and limits as numbers for protection against hazardous outputs, in the annex. The contents of the corrigendum of October 2008 have been included in this copy.
Appareils électromédicaux - Partie 2-16: Exigences particulières pour la sécurité de base et les performances essentielles des appareils d'hémodialyse, d'hémodiafiltration et d'hémofiltration
La CEI 60601-2-16:2008 s'applique à la sécurité de base et aux performances essentielles des appareils d'hémodialyse, d'hémodiafiltration et d'hémofiltration, désignés ci-après sous le terme d'appareils d'hémodialyse. La CEI 60601-2-16:2008 ne prend pas en compte le système de contrôle du liquide de dialyse de l'appareil d'hémodialyse, utilisant la régénération du liquide de dialyse et les systèmes de transmission centralisés. Toutefois, elle prend en compte les exigences de sécurité spécifiques de l'appareil d'hémodialyse concernant la sécurité électrique et celle du patient. La CEI 60601-2-16:2008 spécifie les exigences minimales de sécurité relatives aux appareils d'hémodialyse. Ces appareils sont destinés àêtre utilisés soit par le personnel médical, soit par le patient, soit par d'autres personnes formées, sous le contrôle d'un personnel ayant une bonne compétence médicale. La CEI 60601-2-16:2008 s'applique à tout appareil em destiné à fournir un traitement d'hémodialyse, d'hémodiafiltration et d'hémofiltration à un patient souffrant d'insuffisance rénale. Cette troisième édition annule et remplace la deuxième édition de la CEI 60601-2-16 publiée en 1998. La présente édition constitue une révision technique. Les évolutions par rapport à l'édition précédente comprennent, entre autres, un résumé des exigences supplémentaires relatives aux performances essentielles, la révision des termes et des définitions, la classification des parties appliquées et des limites chiffrées relatives à la protection contre les caractéristiques de sortie dangereuses, en Annexe. Le contenu du corrigendum d'octobre 2008 a été pris en considération dans cet exemplaire.
General Information
Relations
Overview
IEC 60601-2-16:2008 (Edition 3.0) is a particular standard in the IEC 60601 series that specifies basic safety and essential performance requirements for medical electrical equipment used for haemodialysis, haemodiafiltration and haemofiltration. This third edition (2008) updates and replaces the 1998 edition, incorporates the October 2008 corrigendum, and provides technical revisions including additional essential performance requirements, revised terms and definitions, and classification of applied parts.
Key scope points:
- Applies to haemodialysis equipment intended to deliver dialysis therapies to patients with kidney failure.
- Focuses on electrical safety, patient safety and minimum safety requirements for equipment used by clinical staff or trained personnel (including supervised patient use).
- Excludes some items from detailed coverage (e.g., extracorporeal circuits, dialysers, dialysing fluid concentrates, water treatment equipment, and peritoneal dialysis devices).
- Does not fully address dialysing-fluid control systems that use regeneration or central delivery but does consider their electrical and patient safety aspects.
Key Topics and Requirements
- Basic safety and essential performance definitions and mandatory requirements aligned with IEC 60601‑1 (general standard).
- Classification of applied parts and limits for protection against hazardous outputs (see annexes and tables such as the essential performance table 201.101).
- Electrical protection against shocks and hazardous outputs; specific requirements for equipment construction and circuitry.
- Risk and fault condition management, including handling hazardous situations and specified fault responses.
- Alarms, usability and EMC - requirements for alarm systems, usability testing and electromagnetic compatibility (referenced collateral clauses).
- Informative guidance and rationale in Annex AA to support interpretation of particular requirements.
Practical Applications
- Design and development of haemodialysis, haemodiafiltration and haemofiltration machines to meet regulatory safety expectations.
- Risk management, verification and validation testing for manufacturers and test laboratories.
- Conformity assessment and technical documentation for regulatory submissions and CE/market approvals.
- Clinical procurement, biomedical engineering assessment, and safe deployment in dialysis clinics and hospitals.
Who Uses This Standard
- Medical device manufacturers and design engineers (product development and compliance teams)
- Regulatory and quality assurance professionals (technical files, conformity assessment)
- Test laboratories and certification bodies (type testing and inspections)
- Clinical engineers, nephrology clinicians and procurement officers evaluating device safety
Related Standards
- IEC 60601-1 - General standard for basic safety and essential performance (this particular standard takes precedence where modified)
- IEC 60601 collateral standards (e.g., IEC 60601-1-2 for EMC; note IEC 60601-1-3 does not apply here)
- IEC 60601-2-39 - Peritoneal dialysis equipment (different dialysis modality)
Keywords: IEC 60601-2-16, haemodialysis equipment, haemodiafiltration, haemofiltration, medical electrical equipment, basic safety, essential performance, patient safety, electrical safety, applied parts, dialysis device compliance.
Frequently Asked Questions
IEC 60601-2-16:2008 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Medical electrical equipment - Part 2-16: Particular requirements for basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment". This standard covers: IEC 60601-2-16:2008 applies to the basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment. IEC 60601-2-16:2008 does not take into consideration the dialysing fluid control system of Haemodialysis equipment using regeneration of dialysing fluid and central delivery systems. It does however take into consideration the specific safety requirements of such Haemodialysis equipment concerning electrical safety and patient safety. IEC 60601-2-16:2008 specifies the minimum safety requirements for haemodialysis equipment. These devices are intended for use either by medical staff or for use by the patient or other trained personnel under the supervision of medical expertise. IEC 60601-2-16:2008 includes all medical electrical equipment that is intended to deliver a haemodialysis, haemodiafiltration and haemofiltration treatment to a patient suffering from kidney failure. This third edition cancels and replaces the second edition of IEC 60601-2-16, published in 1998. This edition constitutes a technical revision. Changes since the previous edition include, among others, a summary of additional essential performance requirements, revision of terms and definitions, classification of applied parts and limits as numbers for protection against hazardous outputs, in the annex. The contents of the corrigendum of October 2008 have been included in this copy.
IEC 60601-2-16:2008 applies to the basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment. IEC 60601-2-16:2008 does not take into consideration the dialysing fluid control system of Haemodialysis equipment using regeneration of dialysing fluid and central delivery systems. It does however take into consideration the specific safety requirements of such Haemodialysis equipment concerning electrical safety and patient safety. IEC 60601-2-16:2008 specifies the minimum safety requirements for haemodialysis equipment. These devices are intended for use either by medical staff or for use by the patient or other trained personnel under the supervision of medical expertise. IEC 60601-2-16:2008 includes all medical electrical equipment that is intended to deliver a haemodialysis, haemodiafiltration and haemofiltration treatment to a patient suffering from kidney failure. This third edition cancels and replaces the second edition of IEC 60601-2-16, published in 1998. This edition constitutes a technical revision. Changes since the previous edition include, among others, a summary of additional essential performance requirements, revision of terms and definitions, classification of applied parts and limits as numbers for protection against hazardous outputs, in the annex. The contents of the corrigendum of October 2008 have been included in this copy.
IEC 60601-2-16:2008 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 60601-2-16:2008 has the following relationships with other standards: It is inter standard links to IEC 60601-2-16:2008/COR1:2008, IEC 60601-2-16:1998, IEC 60601-2-16:2012. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
You can purchase IEC 60601-2-16:2008 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of IEC standards.
Standards Content (Sample)
IEC 60601-2-16
Edition 3.0 2008-04
INTERNATIONAL
STANDARD
Medical electrical equipment –
Part 2-16: Particular requirements for basic safety and essential performance of
haemodialysis, haemodiafiltration and haemofiltration equipment
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
IEC Central Office
3, rue de Varembé
CH-1211 Geneva 20
Switzerland
Email: inmail@iec.ch
Web: www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
ƒ Catalogue of IEC publications: www.iec.ch/searchpub
The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,…).
It also gives information on projects, withdrawn and replaced publications.
ƒ IEC Just Published: www.iec.ch/online_news/justpub
Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available
on-line and also by email.
ƒ Electropedia: www.electropedia.org
The world's leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions
in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical
Vocabulary online.
ƒ Customer Service Centre: www.iec.ch/webstore/custserv
If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service
Centre FAQ or contact us:
Email: csc@iec.ch
Tel.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 60601-2-16
Edition 3.0 2008-04
INTERNATIONAL
STANDARD
Medical electrical equipment –
Part 2-16: Particular requirements for basic safety and essential performance of
haemodialysis, haemodiafiltration and haemofiltration equipment
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
PRICE CODE
XA
ICS 11.040.20 ISBN 2-8318-9720-3
– 2 – 60601-2-16 © IEC:2008(E)
CONTENTS
FOREWORD.3
INTRODUCTION.6
201.1 Scope, object and related standards .7
201.2 Normative references.9
201.3 Terms and definitions.9
201.4 General requirements.12
201.5 General requirements for testing of ME EQUIPMENT.15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS .15
201.7 ME EQUIPMENT identification, marking and documents.15
201.8 Protection against electrical HAZARDS from ME EQUIPMENT.19
201.9 Protection against mechanical hazards of me equipment and me systems .21
201.10 Protection against unwanted and excessive radiation HAZARDS.21
201.11 Protection against excessive temperatures and other HAZARDS.21
201.12 * Accuracy of controls and instruments and protection against hazardous
outputs.22
201.13 HAZARDOUS SITUATIONS and fault conditions .31
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS).31
201.15 Construction of ME EQUIPMENT .31
201.16 * ME SYSTEMS .32
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS .33
202 Electromagnetic compatibility – Requirements and tests .33
203 General requirements for radiation protection in diagnostic X-ray equipment.33
206 Usability.33
208 General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems.33
209 Requirements for the reduction of environmental impacts.34
210 Process requirements for the development of physiologic closed-loop
controllers .34
Annexes .35
Annex AA (informative) Particular guidance and rationale.36
Bibliography.52
Index of defined terms used in this particular standard.53
Figure 201.101 – Continuous air infusion test set-up .28
Figure AA.1 – Example of the HAEMODIALYSIS ME SYSTEM.50
Table 201.101 – ESSENTIAL PERFORMANCE requirements .12
60601-2-16 © IEC:2008(E) – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-16: Particular requirements for basic safety and
essential performance of haemodialysis, haemodiafiltration
and haemofiltration equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-16 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This third edition cancels and replaces the second edition of IEC 60601-2-16, published in
1998. This edition constitutes a technical revision. Changes since the previous edition
include, among others, a summary of additional essential performance requirements, revision
of terms and definitions, classification of applied parts and limits as numbers for protection
against hazardous outputs, in the annex.
– 4 – 60601-2-16 © IEC:2008(E)
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/671/FDIS 62D/681/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website.
60601-2-16 © IEC:2008(E) – 5 –
The committee has decided that the contents of this particular standard will remain
unchanged until the maintenance result date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
A bilingual version of this publication may be issued at a later date.
The contents of the corrigendum of October 2008 have been included in this copy.
– 6 – 60601-2-16 © IEC:2008(E)
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HAEMODIALYSIS,
HAEMODIAFILTRATION and HAEMOFILTRATION EQUIPMENT.
60601-2-16 © IEC:2008(E) – 7 –
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-16: Particular requirements for basic safety and
essential performance of haemodialysis, haemodiafiltration
and haemofiltration equipment
201.1 Scope, object and related standards
Clause 1 of IEC 60601-1:2005 applies, except as follows:
201.1.1 Scope
Addition:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
HAEMODIALYSIS, HAEMODIAFILTRATION and HAEMOFILTRATION EQUIPMENT, hereafter referred to as
HAEMODIALYSIS EQUIPMENT.
This International standard does not take into consideration the DIALYSING FLUID control
system of HAEMODIALYSIS EQUIPMENT using regeneration of DIALYSING FLUID and CENTRAL
DELIVERY SYSTEMS. It does however take into consideration the specific safety requirements of
such HAEMODIALYSIS EQUIPMENT concerning electrical safety and PATIENT safety.
This International standard specifies the minimum safety requirements for HAEMODIALYSIS
EQUIPMENT. These devices are intended for use either by medical staff or for use by the
PATIENT or other trained personnel under the supervision of medical expertise.
This International standard includes all ME EQUIPMENT that is intended to deliver a
HAEMODIALYSIS, HAEMODIAFILTRATION and HAEMOFILTRATION treatment to a PATIENT suffering
from kidney failure.
The particular requirements in this International standard do not apply to:
– EXTRACORPOREAL CIRCUITS;
– DIALYSERS;
– DIALYSING FLUID CONCENTRATES;
– water treatment equipment;
– equipment used to perform PERITONEAL DIALYSIS (see IEC 60601-2-39).
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of IEC 60601-1.
NOTE See also 4.2 of IEC 60601-1:2005.
201.1.2 Object
Replacement:
– 8 – 60601-2-16 © IEC:2008(E)
The object of this particular standard is to establish BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for HAEMODIALYSIS EQUIPMENT
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of IEC 60601-1, and Clause 201.2 of this International Standard. (See also Clauses
202, 203, 206, 208, 209 and 210)
IEC 60601-1-3 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in IEC 60601-1, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance, as appropriate for the particular ME EQUIPMENT under
consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements.
A requirement of a particular standard takes priority over IEC 60601-1.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses or figures which are additional to those of the general standard are numbered
starting from 201.101. However due to the fact that definitions in the general standard are
numbered 3.1 through 3.139, additional definitions in this standard are numbered beginning
from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items aa), bb),
etc.
Subclauses or figures which are additional to those of a collateral standard are numbered
starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for IEC 60601-
1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
60601-2-16 © IEC:2008(E) – 9 –
Where there is no corresponding, clause or subclause in this particular standard, the, clause
or subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 52.
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60601-1-9:2007, Medical electrical equipment – Part 1-9: General requirements for basic
safety and essential performance – Collateral Standard: Requirements for environmentally
conscious design
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
IEC 61672-1, Electroacoustics – Sound level meters – Part 1: Specifications
ISO 594-2, Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other
medical equipment – Part 2: Lock fittings
ISO 3744, Acoustics – Determination of sound power levels of noise sources using sound
pressure – Engineering method in an essentially free field over a reflecting plane
ISO 8638, Cardiovascular implants and artificial organs – Extracorporeal blood circuit for
haemodialysers, haemodiafilters and haemofilters
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005, and
IEC 60601-1-10:2007 apply, except as follows:
NOTE An index of defined terms is found beginning on page 53.
201.3.8
*APPLIED PART
Replacement:
EXTRACORPOREAL CIRCUIT and all parts permanently and conductively connected to it (e.g.
DIALYSING FLUID circuit)
NOTE See Figure AA.1 in Annex AA.
201.3.78
PATIENT CONNECTION
Addition:
NOTE The PATIENT blood lines connectors are the individual points on the APPLIED PART through which current
can flow between the PATIENT and the HAEMODIALYSIS EQUIPMENT in NORMAL CONDITION or SINGLE
FAULT CONDITION.
– 10 – 60601-2-16 © IEC:2008(E)
Additions:
201.3.201
* HAEMODIALYSIS EQUIPMENT
ME EQUIPMENT or ME SYSTEM used to perform HAEMODIALYSIS, HAEMODIAFILTRATION and/or
HAEMOFILTRATION
NOTE When the term ME EQUIPMENT is used in headings it is equivalent to HAEMODIALYSIS EQUIPMENT. When the
term ME EQUIPMENT is used in the text it is referring to a general ME EQUIPMENT.
201.3.202
HAEMODIALYSIS
HD
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by bidirectional diffusive
transport and ULTRAFILTRATION across a semi-permeable membrane separating the blood from
the DIALYSING FLUID
NOTE This process normally includes fluid removal by filtration. This process is usually also accompanied by
diffusion of substances from the DIALYSING FLUID into the blood.
201.3.203
HAEMOFILTRATION
HF
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by unidirectional convective
transport via ULTRAFILTRATION across a semi-permeable membrane separating the blood from
the DIALYSING FLUID. Ultrafiltrate is simultaneously replaced by an approximately isoosmolar
SUBSTITUTION FLUID at a rate such that the difference between the ULTRAFILTRATION rate and
the rate of SUBSTITUTION FLUID addition will lead to removal of the excess fluid over the course
of the treatment
201.3.204
HAEMODIAFILTRATION
HDF
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by a simultaneous
combination of HD and HF
201.3.205
DIALYSER
HD, HDF or HF
a device containing a semi-permeable membrane that is used to perform
201.3.206
DIALYSING FLUID
solution intended to exchange solutes and/or water with blood during HD or HDF
NOTE The words "dialysate" and "dialysis fluid" are commonly used as synonyms of DIALYSING FLUID.
201.3.207
DIALYSING FLUID CONCENTRATE
substances which, when appropriately diluted or dissolved with purified water, produce the
DIALYSING FLUID
201.3.208
SUBSTITUTION FLUID
a fluid administered to the PATIENT via the EXTRACORPOREAL CIRCUIT during HF or HDF
201.3.209
*ULTRAFILTRATION
process of fluid removal from the PATIENT'S blood across the DIALYSER
60601-2-16 © IEC:2008(E) – 11 –
201.3.210
EXTRACORPOREAL CIRCUIT
blood lines and any integral ACCESSORY thereof
201.3.211
TRANSMEMBRANE PRESSURE
TMP
fluid pressure difference exerted across a semi-permeable membrane
NOTE Generally the mean TMP is used. In practice, the displayed TRANSMEMBRANE PRESSURE is usually estimated
from the measured EXTRACORPOREAL CIRCUIT pressure and the measured DIALYSING FLUID pressure, each obtained
at a single point.
201.3.212
*BLOOD LEAK
leakage of blood from the blood compartment to the DIALYSING FLUID compartment of the
DIALYSER
NOTE When performing an HF process, this involves the filtration fluid section.
201.3.213
ARTERIAL PRESSURE
pressure measured in the blood withdrawal line of the EXTRACORPOREAL CIRCUIT
NOTE 1 A difference can be made between the pre-pump pressure, which is upstream of the blood pump, and
post pump pressure, which is downstream of the blood pump .
201.3.214
VENOUS PRESSURE
pressure measured in the blood return line of the EXTRACORPOREAL CIRCUIT
201.3.215
*PROTECTIVE SYSTEM
automatic system, or a constructional feature, specifically designed to protect the PATIENT
against HAZARDS which can arise
201.3.216
*ONLINE HDF
HAEMODIAFILTRATION procedure where the HAEMODIALYSIS EQUIPMENT, based on the DIALYSING
FLUID, produces the SUBSTITUTION FLUID for the HDF treatment, suitable for injection
201.3.217
*ONLINE HF
HAEMOFILTRATION procedure where the HAEMODIALYSIS EQUIPMENT, based on the DIALYSING
FLUID, produces the SUBSTITUTION FLUID for the HF treatment, suitable for injection
201.3.218
CENTRAL DELIVERY SYSTEM
ME SYSTEM which proportions DIALYSING FLUID CONCENTRATE and water for distribution
Part of a
as DIALYSING FLUID to the HAEMODIALYSIS EQUIPMENT or distributes DIALYSING FLUID
CONCENTRATE
201.3.219
NET FLUID REMOVAL
fluid loss from the PATIENT
NOTE Historically this term was “weight loss”.
– 12 – 60601-2-16 © IEC:2008(E)
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 * ESSENTIAL PERFORMANCE
Additional subclauses:
201.4.3.101 * Additional ESSENTIAL PERFORMANCE requirements
ESSENTIAL PERFORMANCE of HAEMODIALYSIS EQUIPMENT includes, but is not limited to the functions
found in the subclauses listed in Table 201.101, which must be met within the tolerances specified
by the MANUFACTURER, if applicable:
Table 201.101 – ESSENTIAL PERFORMANCE requirements
Requirement Subclause
Blood flow 201.4.3.102
DIALYSING FLUID flow 201.4.3.103
NET FLUID REMOVAL 201.4.3.104
SUBSTITUTION FLUID flow 201.4.3.105
Dialysis time 201.4.3.106
DIALYSING FLUID composition 201.4.3.107
DIALYSING FLUID temperature 201.4.3.108
SUBSTITUTION FLUID temperature 201.4.3.109
NOTE Some ESSENTIAL PERFORMANCES listed in Table 201.101 are dependent on the characteristics of the
disposable used (e.g. blood flow is dependent upon the pump segment inner diameter in rotary peristaltic pumps).
201.4.3.102 Blood flow
The blood flow for the HAEMODIALYSIS EQUIPMENT shall be as specified by the MANUFACTURER.
NOTE 1 Only a blood flow lower than the set value is considered as negative for the treatment. Therefore the goal
of testing is to find the highest negative blood flow error.
Compliance is checked under the following test conditions for typical peristaltic pumps.
– Apply a pump segment to the HAEMODIALYSIS EQUIPMENT and let it run for at least 30 min.
– Apply a fluid (e.g. water) with a temperature of 37 °C in the EXTRACORPOREAL CIRCUIT.
– Set the blood flow of the HAEMODIALYSIS EQUIPMENT to 400 ml/min or – if not possible - to
the highest possible blood flow.
– Set the ARTERIAL PRESSURE to -200 mmHg.
– Measure the blood flow.
The values of the measured blood flow shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
NOTE 2 Pump segment fatigue can reduce the blood flow rate.
NOTE 3 The blood flow rate in peristaltic pumps can be affected by negative input pressures
201.4.3.103 DIALYSING FLUID flow
The DIALYSING FLUID flow for the DIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
60601-2-16 © IEC:2008(E) – 13 –
NOTE Only a DIALYSING FLUID flow lower than the set value is considered as negative for the treatment.
Compliance is checked under the following test conditions.:
– Set the HAEMODIALYSIS EQUIPMENT to the HAEMODIALYSIS mode as specified by the
MANUFACTURER.
– Set the HAEMODIALYSIS EQUIPMENT to maximum DIALYSING FLUID flow.
Measure the DIALYSING FLUID flow during 30 min.
– Set the HAEMODIALYSIS EQUIPMENT to minimum DIALYSING FLUID flow.
– Measure the DIALYSING FLUID flow during 30 min.
The values of the DIALYSING FLUID flow shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
201.4.3.104 NET FLUID REMOVAL
The NET FLUID REMOVAL for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
Compliance is checked under the following test conditions.
Test 1 for the balancing part of the HAEMODIALYSIS EQUIPMENT only:
– Set the HAEMODIALYSIS EQUIPMENT in the HAEMODIALYSIS mode, if applicable, with a
DIALYSER according to the MANUFACTURER’s recommendation.
– Apply fluid (e.g. water) in THE EXTRACORPOREAL CIRCUIT.
– Set the highest DIALYSING FLUID flow, if applicable.
– Set the DIALYSING FLUID temperature to 37 °C, if applicable.
– Set the NET FLUID REMOVAL rate to 0 ml/h or the lowest adjustable value.
– Create a blood outlet pressure of 50 mmHg below the highest specified pressure.
– Measure the NET FLUID REMOVAL during an appropriate time interval.
Continue with test 2:
– Set the NET FLUID REMOVAL rate to the maximum value.
– Measure the net fluid removal during an appropriate time interval.
Continue with test 3:
– Create a blood outlet pressure 20 mmHg above the lowest specified pressure.
– Measure the NET FLUID REMOVAL during an appropriate time interval.
The values of the NET FLUID REMOVAL shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
201.4.3.105 SUBSTITUTION FLUID flow
HAEMOFILTRATION and HAEMODIAFILTRATION equipment only:
For
The SUBSTITUTION FLUID flow for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
NOTE Only a SUBSTITUTION FLUID flow lower than the set value is considered as negative for the treatment.
Compliance is checked under the following test conditions:
– 14 – 60601-2-16 © IEC:2008(E)
Test 1 for the balancing part of the HAEMODIALYSIS EQUIPMENT and of the therapeutic relevant
SUBSTITUTION FLUID flow:
– Set the HAEMODIALYSIS EQUIPMENT to the HDF or HF mode with a DIALYSER according to the
MANUFACTURER’s recommendation.
– Apply fluid (e.g. water) in the EXTRAXCORPOREAL CIRCUIT.
– Set the NET FLUID REMOVAL flow to 0 ml/h, or – if not possible – to the minimum.
– Set the maximum SUBSTITUTION FLUID flow.
– Set the temperature of the SUBSTITUTION FLUID to 37 °C, if applicable.
– Measure the SUBSTITUTION FLUID flow and the NET FLUID REMOVAL.
Continue with test 2:
– Set the minimum SUBSTITUTION FLUID flow.
– Measure the SUBSTITUTION FLUID flow and the NET FLUID REMOVAL.
The values of SUBSTITUTION FLUID flow and NET FLUID REMOVAL shall be within the tolerances
specified by the MANUFACTURER in the instructions for use.
201.4.3.106 Dialysis time
The accuracy of the Dialysis Time for the HAEMODIALYSIS EQUIPMENT shall be as specified by
the MANUFACTURER.
Compliance is checked by functional measurements relevant for the definition of Dialysis Time
specified by the MANUFACTURER.
201.4.3.107 * DIALYSING FLUID composition
Test method specified by the MANUFACTURER.
201.4.3.108 DIALYSING FLUID temperature
The DIALYSING FLUID temperature for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
NOTE This test applies only to HAEMODIALYSIS EQUIPMENT having a heater for the DIALYSING FLUID.
Compliance is checked under the following test conditions:
– Let the HAEMODIALYSIS EQUIPMENT run until it is in a thermally stable condition.
– The environmental temperature shall be within 20 °C to 25 °C.
– Set the DIALYSING FLUID temperature to 37 °C, if applicable.
– Set the highest DIALYSING FLUID flow.
– Measure the temperature at the DIALYSER inlet.
– Record the temperature during a period of 30 min.
– Set the lowest DIALYSING FLUID flow.
– Measure the temperature at the DIALYSER inlet.
– Record the temperature during a period of 30 min.
The values of the DIALYSING FLUID temperature shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
60601-2-16 © IEC:2008(E) – 15 –
201.4.3.109 SUBSTITUTION FLUID temperature
The tolerances of the SUBSTITUTION FLUID temperature for the HAEMODIALYSIS EQUIPMENT shall
be as specified by the MANUFACTURER.
NOTE This test applies only to HAEMODIALYSIS EQUIPMENT having a heater for the SUBSTITUTION FLUID.
Compliance is checked under the following test conditions.
– Let the HAEMODIALYSIS EQUIPMENT run until it is in a thermally stable condition.
– The environmental temperature is within 20-25 °C.
– Set the SUBSTITUTION FLUID temperature to 37 °C, if applicable.
– Set the highest SUBSTITUTION FLUID flow.
– Measure the temperature of the SUBSTITUTION FLUID at the connection point of the
SUBSTITUTION FLUID line to the blood line.
– Record the temperature over a period of 30 min.
– Set the lowest SUBSTITUTION FLUID flow.
– Measure the temperature of the SUBSTITUTION FLUID at the connection point of the
SUBSTITUTION FLUID line to the blood line.
– Record the temperature over a period of 30 min.
The values of the SUBSTITUTION FLUID temperature shall be within the tolerances specified by
the MANUFACTURER in the instructions for use.
201.4.7 SINGLE FAULT CONDITION for ME EQUIPMENT
Addition:
An example of SINGLE FAULT CONDITION is a failure of a PROTECTIVE SYSTEM (see
201.12.4.4.101, 201.12.4.4.102, 201.12.4.4.103, 201.12.4.4.104, 201.12.4.4.105);
NOTE If air is permanently present in the EXTRACORPOREAL CIRCUIT when the HAEMODIALYSIS EQUIPMENT is used
as intended by the MANUFACTURER, the air is not regarded as a SINGLE FAULT CONDITION, but as NORMAL CONDITION.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.4.3 Units of measure
Addition:
mmHg may be used for measurement of pressures in any part of the HAEMODIALYSIS
EQUIPMENT.
– 16 – 60601-2-16 © IEC:2008(E)
201.7.8.1 *Colours of indicator lights
Addition:
NOTE A red ALARM SIGNAL activated in case of extracorporeal blood loss to the environment (see
201.12.4.4.104.1) is one example of an alarm that requires immediate response by the OPERATOR. If the blood flow
is stopped for an extended period of time (201.12.4.4.104.3), this is another example for a red ALARM SIGNAL. In
most other ALARM CONDITIONS the PROTECTIVE SYSTEM puts the HAEMODIALYSIS EQUIPMENT in a state, which is safe
for the PATIENT, at least temporarily, and therefore such ALARM SIGNAL is indicated by a yellow ALARM SIGNAL. Other
ALARM SIGNALS shall be determined by the MANUFACTURER’S RISK MANAGEMENT PROCESS.
It is acceptable to implement other colour coding schemes in addition to 7.8.1 and 7.8.2 of the
general standard.
If additional colour coding schemes are implemented,
a) for all alarms needing immediate response by the OPERATOR a red alarm signal shall be
activated;
b) the scheme defined by this clause of the general standard shall be the factory default;
c) only the RESPONSIBLE ORGANIZATION shall be able to change the colour coding scheme.
Compliance is checked by functional tests
201.7.9.2 Instructions for use
201.7.9.2.2 Warning and safety notices
Addition:
The instructions for use shall additionally include the following, if applicable:
– a statement which draws the OPERATOR's attention to the precautions necessary to
prevent any cross-infection between PATIENTS;
– a statement which draws the OPERATOR's attention to the HAZARDs associated with
connection and disconnection of the PATIENT;
– a statement which draws the OPERATOR's attention to the potential HAZARDs including any
HAZARDOUS SITUATIONs arising from improper connections of the EXTRACORPOREAL CIRCUIT;
– a statement on the HAZARDS related to incorrect choice of DIALYSING FLUID
CONCENTRATE(S);
– a quantitative description of the possible deviation of each component of the DIALYSING
FLUID in SINGLE FAULT CONDITION depending on the alarm limits of the PROTECTIVE SYSTEM;
– * a statement on the potential HAZARDS related to a possible transport of undesired
DIALYSING FLUID compartment to the blood compartment of the
substances from the
DIALYSER;
– for the PROTECTIVE SYSTEM employed according to 201.12.4.4.104 1a) Note 2:
• a warning stating that this PROTECTIVE SYSTEM reduces the HAZARD in part only and an
explanation of the remaining HAZARDs;
• a description of further measures to reduce the RISK;
– an explanation of the adequate OPERATOR action upon an alarm and potential HAZARDs, if
the alarm is repeatedly cleared without solving the underlying problem.;
– * a warning specifying that any narrow passages in the EXTRACORPOREAL CIRCUIT (such as
kinks in the blood line or cannula that are too thin) may cause haemolysis and that this
HAZARDOUS SITUATION may not be detected by the PROTECTIVE SYSTEMS;
60601-2-16 © IEC:2008(E) – 17 –
– if a PROTECTIVE SYSTEM, according to 201.12.4.4.105 a) Note 1, is applied: a warning
stating that improper functioning of an ultrasonic air detector may be caused by a
coagulum or the application of ultrasound gel;
– a warning stating that air may enter into the EXTRACORPOREAL CIRCUIT at connection points
downstream of the air detector, if pressures are negative; this can occur in cases such as
single needle applications or central venous catheter applications;
– for ONLINE HDF and ONLINE HF:
• a warning stating that only the disinfection procedures defined and validated by the
MANUFACTURER may be used for ONLINE HDF and ONLINE HF;
• information on the required quality of the incoming water and of the DIALYSING FLUID
CONCENTRATES used;
• intervals at which wearing parts (e.g. filter) should be exchanged;
– a warning that the blood flow and thus the treatment efficacy may be reduced when the
pre-pump ARTERIAL PRESSURE is extremely negative; and the range and accuracy of the
flow of such pump(s) and the inlet and outlet pressure range over which this accuracy is
maintained.
201.7.9.2.5 ME EQUIPMENT description
Addition:
The instructions for use shall additionally include the following, if applicable:
– a definition of TRANSMEMBRANE PRESSURE if the MANUFACTURER makes use of one different
from that stated in 201.3.211;
– an explanation of the coloured markings on the DIALYSING FLUID CONCENTRATE connectors;
– information on the effective delivered blood flow in single-needle treatments;
– information on the recirculation of blood in the EXTRACORPOREAL CIRCUIT in single-needle
treatments;
– the delay time after which an audible alarm is activated after interruption of the power supply
– for PHYSIOLOGIC CLOSED-LOOP CONTROLLER functions:
a) the technical working principle;
b) the PATIENT parameters which are measured and the physiological parameters which
are controlled;
c) the methods by which these feedback control modes have been evaluated including
beneficial and adverse effects recorded during clinical testing;
see also the collateral standard IEC 60601-1-10.
– * for any data that is displayed or indicated by the HAEMODIALYSIS EQUIPMENT and that may
be used for adjusting the treatment or measuring or confirming the treatment efficacy:
a) a description of the technical working principle;
b) if the measurement is indirect: a statement to the accuracy and possible influencing
factors;
c) the method by which the technical working principle has been evaluated relative to
standard medical care
– for HAEMODIALYSIS EQUIPMENT with APPLIED PARTS other than TYPE CF APPLIED PARTS a
statement if this HAEMODIALYSIS EQUIPMENT can be used together with central venous
catheters. If the HAEMODIALYSIS EQUIPMENT is not suitable for central venous catheters
possible HAZARDs shall be listed.
201.7.9.2.6 Installation
Addition:
– 18 – 60601-2-16 © IEC:2008(E)
The instructions for use shall additionally include the following, if applicable:
– a statement that it is essential for the HAEMODIALYSIS EQUIPMENT to be installed and used
in compliance with appropriate regulations/recommendations on quality of water and other
relevant fluids;
– for CLASS I HAEMODIALYSIS EQUIPMENT, a statement of the importance of the quality of the
protective earth in the electrical installation;
– a statement of the applications in which a POTENTIAL EQUALIZATION CONDUCTOR should be
used;
– the acceptable range of temperature, flow and pressure for inlet water and any CENTRAL
DELIVERY SYSTEM;
– a note emphasizing the importance of compliance with all local regulations regarding the
separation of the HAEMODIALYSIS EQUIPMENT from the water supply, the prevention of back
flow to the potable water source, and prevention of contamination via the drain connection
of the HAEMODIALYSIS EQUIPMENT from any sewer connection;
– if different schemes for colour coding of visual alarms can be configured, a statement that
the RESPONSIBLE ORGANIZATION should select the colour coding scheme which minimizes
the RISK of confusion in their environment;
– if settings of operating parameters or PROTECTIVE SYSTEMs can be configured, a statement
that the RESPONSIBLE ORGANIZATION should select the configuration(s) or explicitly confirm
the default configuration.
201.7.9.2.12 Cleaning, disinfection and sterilization
Addition:
The instructions for use shall additionally include the following, if applicable:
– a description of the method(s) by which sanitization or disinfection is achieved;
– * a statement that the test procedure by which the effectiveness of sanitization or
disinfection has been verified is available on request;
– a warning stating to follow the MANUFACTURER’S instructions to disinfect the HAEMODIALYSIS
EQUIPMENT. If other procedures are used it is the responsibility of the RESPONSIBLE
ORGANIZATION to validate the disinfection procedure for efficacy and safety. This warning
shall specifically list HAZARDs including the failure mode that may result from other
procedures;
– a warning that the RESPONSIBLE ORGANIZATION is responsible for the hygienic quality of
any delivery system(s), e.g. central water supply system, CENTRAL DELIVERY SYSTEMS,
HAEMODIALYSIS EQUIPMENT connecting devices, including the fluid lines from connection
points to the HAEMODIALYSIS EQUIPMENT.
201.7.9.2.14 ACCESSORIES, supplementary equipment, used material
Addition:
The instructions for use shall additionally include the following, if applicable:
– information on DIALYSING FLUID CONCENTRATES, DIALYSERS and blood lines intended to be
used together with the HAEMODIALYSIS EQUIPMENT.
Compliance is checked by inspection of the instructions for use.
201.7.9.3 Technical description
201.7.9.3.1 General
Addition:
60601-2-16 © IEC:2008(E) – 19 –
The technical description shall additionally include the following, if applicable:
• Installation:
– a description of the particular measures or conditions to be observed when installing the
HAEMODIALYSIS EQUIPMENT or bringing it into use. These shall include guidance on the type
and number of tests to be carried out;
– the maximum temperature which can occur at the drain of the HAEMODIALYSIS EQUIPMENT:
– * information about energy consumption, energy delivery to the environment and energy
delivery to the dr
...
IEC 60601-2-16 ®
Edition 3.0 2008-04
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-16: Particular requirements for basic safety and essential performance of
haemodialysis, haemodiafiltration and haemofiltration equipment
Appareils électromédicaux –
Partie 2-16: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d’hémodialyse, d’hémodiafiltration et d’hémofiltration
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by
any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or
IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de la CEI ou du Comité national de la CEI du pays du demandeur.
Si vous avez des questions sur le copyright de la CEI ou si vous désirez obtenir des droits supplémentaires sur cette
publication, utilisez les coordonnées ci-après ou contactez le Comité national de la CEI de votre pays de résidence.
IEC Central Office
3, rue de Varembé
CH-1211 Geneva 20
Switzerland
Email: inmail@iec.ch
Web: www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
Catalogue of IEC publications: www.iec.ch/searchpub
The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,…).
It also gives information on projects, withdrawn and replaced publications.
IEC Just Published: www.iec.ch/online_news/justpub
Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available
on-line and also by email.
Electropedia: www.electropedia.org
The world's leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions
in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical
Vocabulary online.
Customer Service Centre: www.iec.ch/webstore/custserv
If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service
Centre FAQ or contact us:
Email: csc@iec.ch
Tel.: +41 22 919 02 11
Fax: +41 22 919 03 00
A propos de la CEI
La Commission Electrotechnique Internationale (CEI) est la première organisation mondiale qui élabore et publie des
normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications CEI
Le contenu technique des publications de la CEI est constamment revu. Veuillez vous assurer que vous possédez
l’édition la plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue des publications de la CEI: www.iec.ch/searchpub/cur_fut-f.htm
Le Catalogue en-ligne de la CEI vous permet d’effectuer des recherches en utilisant différents critères (numéro de référence,
texte, comité d’études,…). Il donne aussi des informations sur les projets et les publications retirées ou remplacées.
Just Published CEI: www.iec.ch/online_news/justpub
Restez informé sur les nouvelles publications de la CEI. Just Published détaille deux fois par mois les nouvelles
publications parues. Disponible en-ligne et aussi par email.
Electropedia: www.electropedia.org
Le premier dictionnaire en ligne au monde de termes électroniques et électriques. Il contient plus de 20 000 termes et
définitions en anglais et en français, ainsi que les termes équivalents dans les langues additionnelles. Egalement appelé
Vocabulaire Electrotechnique International en ligne.
Service Clients: www.iec.ch/webstore/custserv/custserv_entry-f.htm
Si vous désirez nous donner des commentaires sur cette publication ou si vous avez des questions, visitez le FAQ du
Service clients ou contactez-nous:
Email: csc@iec.ch
Tél.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 60601-2-16 ®
Edition 3.0 2008-04
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-16: Particular requirements for basic safety and essential performance of
haemodialysis, haemodiafiltration and haemofiltration equipment
Appareils électromédicaux –
Partie 2-16: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d’hémodialyse, d’hémodiafiltration et d’hémofiltration
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX XA
ICS 11.040.20 ISBN 978-2-88912-333-9
– 2 – 60601-2-16 IEC:2008
CONTENTS
FOREWORD . 3
INTRODUCTION . 6
201.1 Scope, object and related standards . 7
201.2 Normative references . 9
201.3 Terms and definitions . 9
201.4 General requirements . 12
201.5 General requirements for testing of ME EQUIPMENT . 15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 15
201.7 ME EQUIPMENT identification, marking and documents. 15
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 19
201.9 Protection against mechanical hazards of me equipment and me systems . 21
201.10 Protection against unwanted and excessive radiation HAZARDS . 21
201.11 Protection against excessive temperatures and other HAZARDS . 21
201.12 * Accuracy of controls and instruments and protection against hazardous
outputs . 22
201.13 HAZARDOUS SITUATIONS and fault conditions . 31
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 31
201.15 Construction of ME EQUIPMENT . 31
201.16 * ME SYSTEMS . 32
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 33
202 Electromagnetic compatibility – Requirements and tests . 33
203 General requirements for radiation protection in diagnostic X-ray equipment . 33
206 Usability . 33
208 General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems . 33
209 Requirements for the reduction of environmental impacts. 34
210 Process requirements for the development of physiologic closed-loop
controllers . 34
Annexes . 35
Annex AA (informative) Particular guidance and rationale . 36
Bibliography . 52
Index of defined terms used in this particular standard. 53
Figure 201.101 – Continuous air infusion test set-up . 28
Figure AA.1 – Example of the HAEMODIALYSIS ME SYSTEM . 50
Table 201.101 – ESSENTIAL PERFORMANCE requirements . 12
60601-2-16 IEC:2008 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-16: Particular requirements for basic safety and
essential performance of haemodialysis, haemodiafiltration
and haemofiltration equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-16 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This third edition cancels and replaces the second edition of IEC 60601-2-16, published in
1998. This edition constitutes a technical revision. Changes since the previous edition
include, among others, a summary of additional essential performance requirements, revision
of terms and definitions, classification of applied parts and limits as numbers for protection
against hazardous outputs, in the annex.
– 4 – 60601-2-16 IEC:2008
This bilingual version, published in 2011-02, corresponds to the English version.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/671/FDIS 62D/681/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
The French version of this standard has not been voted upon.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website.
60601-2-16 IEC:2008 – 5 –
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
The contents of the corrigendum of October 2008 have been included in this copy.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
– 6 – 60601-2-16 IEC:2008
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HAEMODIALYSIS,
HAEMODIAFILTRATION and HAEMOFILTRATION EQUIPMENT.
60601-2-16 IEC:2008 – 7 –
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-16: Particular requirements for basic safety and
essential performance of haemodialysis, haemodiafiltration
and haemofiltration equipment
201.1 Scope, object and related standards
Clause 1 of IEC 60601-1:2005 applies, except as follows:
201.1.1 Scope
Addition:
BASIC SAFETY and ESSENTIAL PERFORMANCE of
This International Standard applies to the
HAEMODIALYSIS, HAEMODIAFILTRATION and HAEMOFILTRATION EQUIPMENT, hereafter referred to as
HAEMODIALYSIS EQUIPMENT.
This International standard does not take into consideration the DIALYSING FLUID control
system of HAEMODIALYSIS EQUIPMENT using regeneration of DIALYSING FLUID and CENTRAL
DELIVERY SYSTEMS. It does however take into consideration the specific safety requirements of
AEMODIALYSIS EQUIPMENT concerning electrical safety and PATIENT safety.
such H
This International standard specifies the minimum safety requirements for HAEMODIALYSIS
EQUIPMENT. These devices are intended for use either by medical staff or for use by the
PATIENT or other trained personnel under the supervision of medical expertise.
This International standard includes all ME EQUIPMENT that is intended to deliver a
HAEMODIALYSIS, HAEMODIAFILTRATION and HAEMOFILTRATION treatment to a PATIENT suffering
from kidney failure.
The particular requirements in this International standard do not apply to:
– EXTRACORPOREAL CIRCUITS;
– DIALYSERS;
– DIALYSING FLUID CONCENTRATES;
– water treatment equipment;
– equipment used to perform PERITONEAL DIALYSIS (see IEC 60601-2-39).
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of IEC 60601-1.
NOTE See also 4.2 of IEC 60601-1:2005.
201.1.2 Object
Replacement:
– 8 – 60601-2-16 IEC:2008
The object of this particular standard is to establish BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for HAEMODIALYSIS EQUIPMENT
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of IEC 60601-1, and Clause 201.2 of this International Standard. (See also Clauses
202, 203, 206, 208, 209 and 210)
IEC 60601-1-3 does not apply.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in IEC 60601-1, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance, as appropriate for the particular ME EQUIPMENT under
consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements.
A requirement of a particular standard takes priority over IEC 60601-1.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the 60601-1-3 collateral
standard, etc.). The changes to the text of the general standard are specified by the use of
the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses or figures which are additional to those of the general standard are numbered
starting from 201.101. However due to the fact that definitions in the general standard are
numbered 3.1 through 3.139, additional definitions in this standard are numbered beginning
from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items aa), bb),
etc.
Subclauses or figures which are additional to those of a collateral standard are numbered
starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for IEC 60601-
1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
60601-2-16 IEC:2008 – 9 –
Where there is no corresponding, clause or subclause in this particular standard, the clause
or subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 52.
Clause 2 of the general standard applies, except as follows:
Addition:
IEC 60601-1-9:2007, Medical electrical equipment – Part 1-9: General requirements for basic
safety and essential performance – Collateral Standard: Requirements for environmentally
conscious design
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
IEC 61672-1, Electroacoustics – Sound level meters – Part 1: Specifications
ISO 594-2, Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other
medical equipment – Part 2: Lock fittings
ISO 3744, Acoustics – Determination of sound power levels of noise sources using sound
pressure – Engineering method in an essentially free field over a reflecting plane
ISO 8638, Cardiovascular implants and artificial organs – Extracorporeal blood circuit for
haemodialysers, haemodiafilters and haemofilters
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005, and
IEC 60601-1-10:2007 apply, except as follows:
NOTE An index of defined terms is found beginning on page 53.
201.3.8
*APPLIED PART
Replacement:
EXTRACORPOREAL CIRCUIT and all parts permanently and conductively connected to it (e.g.
DIALYSING FLUID circuit)
NOTE See Figure AA.1 in Annex AA.
201.3.78
PATIENT CONNECTION
Addition:
NOTE The PATIENT blood lines connectors are the individual points on the APPLIED PART through which current
can flow between the PATIENT and the HAEMODIALYSIS EQUIPMENT in NORMAL CONDITION or SINGLE
FAULT CONDITION.
– 10 – 60601-2-16 IEC:2008
Additions:
201.3.201
* HAEMODIALYSIS EQUIPMENT
ME EQUIPMENT or ME SYSTEM used to perform HAEMODIALYSIS, HAEMODIAFILTRATION and/or
HAEMOFILTRATION
NOTE When the term ME EQUIPMENT is used in headings it is equivalent to HAEMODIALYSIS EQUIPMENT. When the
term ME EQUIPMENT is used in the text it is referring to a general ME EQUIPMENT.
201.3.202
HAEMODIALYSIS
HD
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by bidirectional diffusive
transport and ULTRAFILTRATION across a semi-permeable membrane separating the blood from
the DIALYSING FLUID
NOTE This process normally includes fluid removal by filtration. This process is usually also accompanied by
diffusion of substances from the DIALYSING FLUID into the blood.
201.3.203
HAEMOFILTRATION
HF
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by unidirectional convective
transport via ULTRAFILTRATION across a semi-permeable membrane separating the blood from
the DIALYSING FLUID. Ultrafiltrate is simultaneously replaced by an approximately isoosmolar
SUBSTITUTION FLUID at a rate such that the difference between the ULTRAFILTRATION rate and
the rate of SUBSTITUTION FLUID addition will lead to removal of the excess fluid over the course
of the treatment
201.3.204
HAEMODIAFILTRATION
HDF
process whereby concentrations of water-soluble substances in a PATIENT's blood and an
excess of fluid of a PATIENT with renal insufficiency are corrected by a simultaneous
combination of HD and HF
201.3.205
DIALYSER
a device containing a semi-permeable membrane that is used to perform HD, HDF or HF
201.3.206
DIALYSING FLUID
solution intended to exchange solutes and/or water with blood during HD or HDF
NOTE The words "dialysate" and "dialysis fluid" are commonly used as synonyms of DIALYSING FLUID.
201.3.207
DIALYSING FLUID CONCENTRATE
substances which, when appropriately diluted or dissolved with purified water, produce the
DIALYSING FLUID
201.3.208
SUBSTITUTION FLUID
a fluid administered to the PATIENT via the EXTRACORPOREAL CIRCUIT during HF or HDF
201.3.209
*ULTRAFILTRATION
process of fluid removal from the PATIENT'S blood across the DIALYSER
60601-2-16 IEC:2008 – 11 –
201.3.210
EXTRACORPOREAL CIRCUIT
blood lines and any integral ACCESSORY thereof
201.3.211
TRANSMEMBRANE PRESSURE
TMP
fluid pressure difference exerted across a semi-permeable membrane
NOTE Generally the mean TMP is used. In practice, the displayed TRANSMEMBRANE PRESSURE is usually estimated
from the measured EXTRACORPOREAL CIRCUIT pressure and the measured DIALYSING FLUID pressure, each obtained
at a single point.
201.3.212
*BLOOD LEAK
leakage of blood from the blood compartment to the DIALYSING FLUID compartment of the
DIALYSER
NOTE When performing an HF process, this involves the filtration fluid section.
201.3.213
ARTERIAL PRESSURE
pressure measured in the blood withdrawal line of the EXTRACORPOREAL CIRCUIT
NOTE A difference can be made between the pre-pump pressure, which is upstream of the blood pump, and post
pump pressure, which is downstream of the blood pump.
201.3.214
VENOUS PRESSURE
pressure measured in the blood return line of the EXTRACORPOREAL CIRCUIT
201.3.215
*PROTECTIVE SYSTEM
automatic system, or a constructional feature, specifically designed to protect the PATIENT
against HAZARDS which can arise
201.3.216
*ONLINE HDF
HAEMODIAFILTRATION procedure where the HAEMODIALYSIS EQUIPMENT, based on the DIALYSING
FLUID, produces the SUBSTITUTION FLUID for the HDF treatment, suitable for injection
201.3.217
*ONLINE HF
HAEMOFILTRATION procedure where the HAEMODIALYSIS EQUIPMENT, based on the DIALYSING
FLUID, produces the SUBSTITUTION FLUID for the HF treatment, suitable for injection
201.3.218
CENTRAL DELIVERY SYSTEM
part of a ME SYSTEM which proportions DIALYSING FLUID CONCENTRATE and water for distribution
as DIALYSING FLUID to the HAEMODIALYSIS EQUIPMENT or distributes DIALYSING FLUID
CONCENTRATE
201.3.219
NET FLUID REMOVAL
fluid loss from the PATIENT
NOTE Historically this term was “weight loss”.
– 12 – 60601-2-16 IEC:2008
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 * ESSENTIAL PERFORMANCE
Additional subclauses:
201.4.3.101 * Additional ESSENTIAL PERFORMANCE requirements
ESSENTIAL PERFORMANCE of HAEMODIALYSIS EQUIPMENT includes, but is not limited to the functions
found in the subclauses listed in Table 201.101, which must be met within the tolerances specified
MANUFACTURER, if applicable:
by the
Table 201.101 – ESSENTIAL PERFORMANCE requirements
Requirement Subclause
Blood flow 201.4.3.102
DIALYSING FLUID flow 201.4.3.103
NET FLUID REMOVAL 201.4.3.104
SUBSTITUTION FLUID flow 201.4.3.105
Dialysis time 201.4.3.106
DIALYSING FLUID composition 201.4.3.107
DIALYSING FLUID temperature 201.4.3.108
SUBSTITUTION FLUID temperature 201.4.3.109
NOTE Some ESSENTIAL PERFORMANCES listed in Table 201.101 are dependent on the characteristics of the
disposable used (e.g. blood flow is dependent upon the pump segment inner diameter in rotary peristaltic pumps).
201.4.3.102 Blood flow
The blood flow for the HAEMODIALYSIS EQUIPMENT shall be as specified by the MANUFACTURER.
NOTE 1 Only a blood flow lower than the set value is considered as negative for the treatment. Therefore the goal
of testing is to find the highest negative blood flow error.
Compliance is checked under the following test conditions for typical peristaltic pumps.
– Apply a pump segment to the HAEMODIALYSIS EQUIPMENT and let it run for at least 30 min.
– Apply a fluid (e.g. water) with a temperature of 37 °C in the EXTRACORPOREAL CIRCUIT.
– Set the blood flow of the HAEMODIALYSIS EQUIPMENT to 400 ml/min or – if not possible - to
the highest possible blood flow.
– Set the ARTERIAL PRESSURE to -200 mmHg.
– Measure the blood flow.
The values of the measured blood flow shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
NOTE 2 Pump segment fatigue can reduce the blood flow rate.
NOTE 3 The blood flow rate in peristaltic pumps can be affected by negative input pressures
201.4.3.103 DIALYSING FLUID flow
The DIALYSING FLUID flow for the DIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
60601-2-16 IEC:2008 – 13 –
NOTE Only a DIALYSING FLUID flow lower than the set value is considered as negative for the treatment.
Compliance is checked under the following test conditions.:
– Set the HAEMODIALYSIS EQUIPMENT to the HAEMODIALYSIS mode as specified by the
MANUFACTURER.
– Set the HAEMODIALYSIS EQUIPMENT to maximum DIALYSING FLUID flow.
Measure the DIALYSING FLUID flow during 30 min.
– Set the HAEMODIALYSIS EQUIPMENT to minimum DIALYSING FLUID flow.
– Measure the DIALYSING FLUID flow during 30 min.
The values of the DIALYSING FLUID flow shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
NET FLUID REMOVAL
201.4.3.104
The NET FLUID REMOVAL for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
Compliance is checked under the following test conditions.
Test 1 for the balancing part of the HAEMODIALYSIS EQUIPMENT only:
– Set the HAEMODIALYSIS EQUIPMENT in the HAEMODIALYSIS mode, if applicable, with a
DIALYSER according to the MANUFACTURER’s recommendation.
– Apply fluid (e.g. water) in THE EXTRACORPOREAL CIRCUIT.
– Set the highest DIALYSING FLUID flow, if applicable.
– Set the DIALYSING FLUID temperature to 37 °C, if applicable.
– Set the NET FLUID REMOVAL rate to 0 ml/h or the lowest adjustable value.
– Create a blood outlet pressure of 50 mmHg below the highest specified pressure.
– Measure the NET FLUID REMOVAL during an appropriate time interval.
Continue with test 2:
– Set the NET FLUID REMOVAL rate to the maximum value.
– Measure the net fluid removal during an appropriate time interval.
Continue with test 3:
– Create a blood outlet pressure 20 mmHg above the lowest specified pressure.
– Measure the NET FLUID REMOVAL during an appropriate time interval.
The values of the NET FLUID REMOVAL shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
201.4.3.105 SUBSTITUTION FLUID flow
For HAEMOFILTRATION and HAEMODIAFILTRATION equipment only:
The SUBSTITUTION FLUID flow for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
NOTE Only a SUBSTITUTION FLUID flow lower than the set value is considered as negative for the treatment.
Compliance is checked under the following test conditions:
– 14 – 60601-2-16 IEC:2008
Test 1 for the balancing part of the HAEMODIALYSIS EQUIPMENT and of the therapeutic relevant
SUBSTITUTION FLUID flow:
– Set the HAEMODIALYSIS EQUIPMENT to the HDF or HF mode with a DIALYSER according to the
MANUFACTURER’s recommendation.
– Apply fluid (e.g. water) in the EXTRAXCORPOREAL CIRCUIT.
– Set the NET FLUID REMOVAL flow to 0 ml/h, or – if not possible – to the minimum.
– Set the maximum SUBSTITUTION FLUID flow.
– Set the temperature of the SUBSTITUTION FLUID to 37 °C, if applicable.
– Measure the SUBSTITUTION FLUID flow and the NET FLUID REMOVAL.
Continue with test 2:
– Set the minimum SUBSTITUTION FLUID flow.
– Measure the SUBSTITUTION FLUID flow and the NET FLUID REMOVAL.
The values of SUBSTITUTION FLUID flow and NET FLUID REMOVAL shall be within the tolerances
specified by the MANUFACTURER in the instructions for use.
201.4.3.106 Dialysis time
The accuracy of the Dialysis Time for the HAEMODIALYSIS EQUIPMENT shall be as specified by
the MANUFACTURER.
Compliance is checked by functional measurements relevant for the definition of Dialysis Time
specified by the MANUFACTURER.
201.4.3.107 * DIALYSING FLUID composition
Test method specified by the MANUFACTURER.
201.4.3.108 DIALYSING FLUID temperature
The DIALYSING FLUID temperature for the HAEMODIALYSIS EQUIPMENT shall be as specified by the
MANUFACTURER.
NOTE This test applies only to HAEMODIALYSIS EQUIPMENT having a heater for the DIALYSING FLUID.
Compliance is checked under the following test conditions:
– Let the HAEMODIALYSIS EQUIPMENT run until it is in a thermally stable condition.
– The environmental temperature shall be within 20 °C to 25 °C.
– Set the DIALYSING FLUID temperature to 37 °C, if applicable.
– Set the highest DIALYSING FLUID flow.
– Measure the temperature at the DIALYSER inlet.
– Record the temperature during a period of 30 min.
– Set the lowest DIALYSING FLUID flow.
– Measure the temperature at the DIALYSER inlet.
– Record the temperature during a period of 30 min.
The values of the DIALYSING FLUID temperature shall be within the tolerances specified by the
MANUFACTURER in the instructions for use.
60601-2-16 IEC:2008 – 15 –
201.4.3.109 SUBSTITUTION FLUID temperature
The tolerances of the SUBSTITUTION FLUID temperature for the HAEMODIALYSIS EQUIPMENT shall
be as specified by the MANUFACTURER.
NOTE This test applies only to HAEMODIALYSIS EQUIPMENT having a heater for the SUBSTITUTION FLUID.
Compliance is checked under the following test conditions.
– Let the HAEMODIALYSIS EQUIPMENT run until it is in a thermally stable condition.
– The environmental temperature is within 20-25 °C.
– Set the SUBSTITUTION FLUID temperature to 37 °C, if applicable.
– Set the highest SUBSTITUTION FLUID flow.
– Measure the temperature of the SUBSTITUTION FLUID at the connection point of the
SUBSTITUTION FLUID line to the blood line.
– Record the temperature over a period of 30 min.
– Set the lowest SUBSTITUTION FLUID flow.
– Measure the temperature of the SUBSTITUTION FLUID at the connection point of the
SUBSTITUTION FLUID line to the blood line.
– Record the temperature over a period of 30 min.
SUBSTITUTION FLUID temperature shall be within the tolerances specified by
The values of the
the MANUFACTURER in the instructions for use.
201.4.7 SINGLE FAULT CONDITION for ME EQUIPMENT
Addition:
An example of SINGLE FAULT CONDITION is a failure of a PROTECTIVE SYSTEM (see
201.12.4.4.101, 201.12.4.4.102, 201.12.4.4.103, 201.12.4.4.104, 201.12.4.4.105);
NOTE If air is permanently present in the EXTRACORPOREAL CIRCUIT when the HAEMODIALYSIS EQUIPMENT is used
as intended by the MANUFACTURER, the air is not regarded as a SINGLE FAULT CONDITION, but as NORMAL CONDITION.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.4.3 Units of measure
Addition:
mmHg may be used for measurement of pressures in any part of the HAEMODIALYSIS
EQUIPMENT.
– 16 – 60601-2-16 IEC:2008
201.7.8.1 *Colours of indicator lights
Addition:
NOTE A red ALARM SIGNAL activated in case of extracorporeal blood loss to the environment (see
201.12.4.4.104.1) is one example of an alarm that requires immediate response by the OPERATOR. If the blood flow
is stopped for an extended period of time (201.12.4.4.104.3), this is another example for a red ALARM SIGNAL. In
most other ALARM CONDITIONS the PROTECTIVE SYSTEM puts the HAEMODIALYSIS EQUIPMENT in a state, which is safe
for the PATIENT, at least temporarily, and therefore such ALARM SIGNAL is indicated by a yellow ALARM SIGNAL. Other
ALARM SIGNALS shall be determined by the MANUFACTURER’S RISK MANAGEMENT PROCESS.
It is acceptable to implement other colour coding schemes in addition to 7.8.1 and 7.8.2 of the
general standard.
If additional colour coding schemes are implemented,
a) for all alarms needing immediate response by the OPERATOR a red alarm signal shall be
activated;
b) the scheme defined by this clause of the general standard shall be the factory default;
c) only the RESPONSIBLE ORGANIZATION shall be able to change the colour coding scheme.
Compliance is checked by functional tests
201.7.9.2 Instructions for use
Warning and safety notices
201.7.9.2.2
Addition:
The instructions for use shall additionally include the following, if applicable:
– a statement which draws the OPERATOR's attention to the precautions necessary to
prevent any cross-infection between PATIENTS;
– a statement which draws the OPERATOR's attention to the HAZARDs associated with
PATIENT;
connection and disconnection of the
– a statement which draws the OPERATOR's attention to the potential HAZARDs including any
HAZARDOUS SITUATIONS arising from improper connections of the EXTRACORPOREAL CIRCUIT;
– a statement on the HAZARDS related to incorrect choice of DIALYSING FLUID
CONCENTRATE(S);
– a quantitative description of the possible deviation of each component of the DIALYSING
in SINGLE FAULT CONDITION depending on the alarm limits of the PROTECTIVE SYSTEM;
FLUID
– * a statement on the potential HAZARDS related to a possible transport of undesired
substances from the DIALYSING FLUID compartment to the blood compartment of the
DIALYSER;
– for the PROTECTIVE SYSTEM employed according to 201.12.4.4.104 1a) Note 2:
• a warning stating that this PROTECTIVE SYSTEM reduces the HAZARD in part only and an
explanation of the remaining HAZARDs;
• a description of further measures to reduce the RISK;
– an explanation of the adequate OPERATOR action upon an alarm and potential HAZARDs, if
the alarm is repeatedly cleared without solving the underlying problem;
– * a warning specifying that any narrow passages in the EXTRACORPOREAL CIRCUIT (such as
kinks in the blood line or cannula that are too thin) may cause haemolysis and that this
HAZARDOUS SITUATION may not be detected by the PROTECTIVE SYSTEMS;
60601-2-16 IEC:2008 – 17 –
– if a PROTECTIVE SYSTEM, according to 201.12.4.4.105 a) Note 1, is applied: a warning
stating that improper functioning of an ultrasonic air detector may be caused by a
coagulum or the application of ultrasound gel;
– a warning stating that air may enter into the EXTRACORPOREAL CIRCUIT at connection points
downstream of the air detector, if pressures are negative; this can occur in cases such as
single needle applications or central venous catheter applications;
– for ONLINE HDF and ONLINE HF:
• a warning stating that only the disinfection procedures defined and validated by the
MANUFACTURER may be used for ONLINE HDF and ONLINE HF;
• information on the required quality of the incoming water and of the DIALYSING FLUID
CONCENTRATES used;
• intervals at which wearing parts (e.g. filter) should be exchanged;
– a warning that the blood flow and thus the treatment efficacy may be reduced when the
pre-pump ARTERIAL PRESSURE is extremely negative; and the range and accuracy of the
flow of such pump(s) and the inlet and outlet pressure range over which this accuracy is
maintained.
201.7.9.2.5 ME EQUIPMENT description
Addition:
The instructions for use shall additionally include the following, if applicable:
– a definition of TRANSMEMBRANE PRESSURE if the MANUFACTURER makes use of one different
from that stated in 201.3.211;
– an explanation of the coloured markings on the DIALYSING FLUID CONCENTRATE connectors;
– information on the effective delivered blood flow in single-needle treatments;
– information on the recirculation of blood in the EXTRACORPOREAL CIRCUIT in single-needle
treatments;
– the delay time after which an audible alarm is activated after interruption of the power supply;
– for PHYSIOLOGICAL CLOSED-LOOP CONTROLLER functions:
a) the technical working principle;
b) the PATIENT parameters which are measured and the physiological parameters which
are controlled;
c) the methods by which these feedback control modes have been evaluated including
beneficial and adverse effects recorded during clinical testing;
see also the collateral standard IEC 60601-1-10.
– * for any data that is displayed or indicated by the HAEMODIALYSIS EQUIPMENT and that may
be used for adjusting the treatment or measuring or confirming the treatment efficacy:
a) a description of the technical working principle;
b) if the measurement is indirect: a statement to the accuracy and possible influencing
factors;
c) the method by which the technical working principle has been evaluated relative to
standard medical care;
– for HAEMODIALYSIS EQUIPMENT with APPLIED PARTS other than TYPE CF APPLIED PARTS a
statement if this HAEMODIALYSIS EQUIPMENT can be used together with central venous
catheters. If the HAEMODIALYSIS EQUIPMENT is not suitable for central venous catheters
possible HAZARDs shall be listed.
201.7.9.2.6 Installation
Addition:
– 18 – 60601-2-16 IEC:2008
The instructions for use shall additionally include the following, if applicable:
– a statement that it is essential for the HAEMODIALYSIS EQUIPME
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...