IEC 60601-2-37:2024
(Main)Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
IEC 60601-2-37:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.7.2.13. This document does not cover ultrasonic therapeutic equipment. Equipment used for the imaging or diagnosis of body structures by ultrasound in conjunction with other medical procedures is covered. IEC 60601-2-37:2024 cancels and replaces the second edition published in 2007 and Amendment 1:2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) technical and editorial changes resulting from the amended general standard IEC 60601 1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and its collateral standards IEC 60601-1-xx,
b) technical and editorial changes as a result of maintenance to normative references;
c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics standards. In particular, Clause 201.11 about protection against excessive temperatures and other hazards has been fully revised.
Appareils électromédicaux - Partie 2-37: Exigences particulières pour la sécurité de base et les performances essentielles des appareils de diagnostic et de surveillance médicaux à ultrasons
L'IEC 60601-2-37:2024 s’applique à la SECURITE DE BASE et aux PERFORMANCES ESSENTIELLES des APPAREILS DE DIAGNOSTIC A ULTRASONS comme cela est défini en 201.3.217, désignés ci-après sous le terme APPAREILS EM. Si un article ou un paragraphe est destiné spécifiquement à être applicable uniquement aux APPAREILS EM, ou uniquement aux SYSTEMES EM, le titre et le contenu de cet article ou de ce paragraphe l’indiquent. Si cela n’est pas le cas, l’article ou le paragraphe s’applique à la fois aux APPAREILS EM et aux SYSTEMES EM, selon le cas. Les DANGERS inhérents à la fonction physiologique prévue de l'APPAREIL EM ou du SYSTEME EM dans le cadre du domaine d'application du présent document ne sont pas couverts par des exigences spécifiques contenues dans le présent document, à l'exception du 201.7.2.13. Le présent document ne couvre pas les appareils thérapeutiques à ultrasons. Les appareils utilisés pour réaliser l'imagerie ou le diagnostic de structures du corps par ultrasons, en association avec une autre procédure médicale, sont couverts. L'IEC 60601-2-37:2024 annule et remplace la deuxième édition parue en 2007, et l’Amendement 1:2015. Cette édition constitue une révision technique.
Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
a) modifications techniques et rédactionnelles, qui résultent de la norme générale amendée IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 et IEC 60601-1:2005/AMD2:2020 et de ses normes collatérales IEC 60601-1-xx ;
b) modifications techniques et rédactionnelles qui résultent de la maintenance des références normatives ;
c) modifications techniques et rédactionnelles qui résultent des évolutions correspondantes des normes du CE 87 Ultrasons. En particulier, l’Article 201.11 concernant la protection contre les températures excessives et les autres dangers a été entièrement révisé.
General Information
Relations
Overview
IEC 60601-2-37:2024 is the third edition (Edition 3.0, 2024) of the particular standard for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. It applies to ultrasonic diagnostic ME EQUIPMENT and ME SYSTEMS (as defined in the standard) and supersedes IEC 60601-2-37:2007 and Amendment 1:2015. The document focuses on safety and performance requirements for diagnostic imaging and monitoring by ultrasound, explicitly excluding ultrasonic therapeutic equipment. It also covers equipment used for imaging in conjunction with other medical procedures.
Key topics and requirements
- Scope and applicability - clarifies when clauses apply to ME EQUIPMENT, ME SYSTEMS, or both; excludes physiological-function hazards except where noted.
- Basic safety & essential performance - requirements supplementing the general IEC 60601-1 series to address ultrasonic-specific risks.
- Protection against hazards - measures for electrical, mechanical, thermal and radiation hazards (notably Clause 201.11 on protection against excessive temperatures and other hazards has been fully revised).
- Acoustic output and reporting - requirements and an acoustic output reporting table for consistent third‑party reporting and communication of ultrasound exposure indices.
- Testing and measurement - test methods for surface temperature of transducers, acoustic and thermal properties, and distributed essential performance tests.
- Electromagnetic compatibility (EMC) - EMC requirements and disturbance tests for safe operation in clinical environments.
- Programmable systems (PEMS) - guidance and requirements for programmable electrical medical systems used in diagnostic ultrasound.
- Identification, marking and documentation - operator information (TI/MI), labeling, and technical documentation requirements to support safe clinical use.
- Annexes with guidance - informative annexes (AA, BB, CC, DD, EE) providing rationale, CISPR 11 classification guidance, operator information interpretation, test setups, and example output tables.
Applications and who uses this standard
- Manufacturers of diagnostic ultrasound systems and transducer assemblies - for design, risk management, and conformity.
- Test laboratories and certification bodies - to develop test procedures and assess compliance.
- Regulatory authorities and notified bodies - to evaluate safety/performance claims during market approval.
- Clinical engineers and procurement teams - to interpret device labeling, acoustic output, and safe‑use guidance.
- R&D and quality teams - to align product development with updated IEC 60601-1 amendments and TC 87 Ultrasonics developments.
Related standards
- IEC 60601-1 (general medical electrical safety) and its amendments (AMD1:2012, AMD2:2020)
- Collateral IEC 60601-1-xx standards and relevant TC 87 Ultrasonics documents
Keywords: IEC 60601-2-37:2024, ultrasonic medical diagnostic equipment, medical electrical equipment, acoustic output, transducer temperature, EMC, PEMS, basic safety, essential performance.
Standards Content (Sample)
IEC 60601-2-37 ®
Edition 3.0 2024-07
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-37: Particular requirements for the basic safety and essential performance
of ultrasonic medical diagnostic and monitoring equipment
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews, graphical symbols and the glossary.
committee, …). It also gives information on projects, replaced With a subscription you will always have access to up to date
and withdrawn publications. content tailored to your needs.
IEC Just Published - webstore.iec.ch/justpublished
Electropedia - www.electropedia.org
Stay up to date on all new IEC publications. Just Published
The world's leading online dictionary on electrotechnology,
details all new publications released. Available online and once
containing more than 22 500 terminological entries in English
a month by email.
and French, with equivalent terms in 25 additional languages.
Also known as the International Electrotechnical Vocabulary
IEC Customer Service Centre - webstore.iec.ch/csc
(IEV) online.
If you wish to give us your feedback on this publication or need
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
IEC 60601-2-37 ®
Edition 3.0 2024-07
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-37: Particular requirements for the basic safety and essential performance
of ultrasonic medical diagnostic and monitoring equipment
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.55; 17.140.50 ISBN 978-2-8322-9372-0
– 2 – IEC 60601-2-37:2024 RLV © IEC 2024
CONTENTS
FOREWORD . 4
INTRODUCTION . 2
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terms and definitions . 10
201.4 General requirements . 17
201.5 General requirements for testing ME EQUIPMENT . 18
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 18
201.7 ME EQUIPMENT identification, marking and documents . 18
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 23
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 23
201.10 Protection against unwanted and excessive radiation HAZARDS . 23
201.11 Protection against excessive temperatures and other HAZARDS . 24
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 30
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 32
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 32
201.15 Construction of ME EQUIPMENT . 32
201.16 ME SYSTEMS . 32
201.17 *Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 32
202 ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 33
212 Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment
(EMS) . 36
Annexes . 37
Annex AA (informative) Particular guidance and rationale . 38
Annex BB (informative) Guidance in classification according to CISPR 11 . 47
Annex CC (informative) Guidance to the MANUFACTURER on the interpretation of TI and
MI to be used to inform the OPERATOR . 48
Annex DD (informative) Example set-up to measure surface temperature of externally
applied TRANSDUCER ASSEMBLIES . 52
Annex EE (informative) Acoustic output table intended for third parties . 55
Bibliography . 58
Index of defined terms . 62
Figure AA.1 – Method a) for an external probe . 41
Figure AA.2 – Method b) for an external probe . 42
Figure AA.3 – Method b) for an external probe . 42
Figure DD.1 – Set-up of an example test object to measure the surface temperature of
externally applied transducers . 54
Table 201.101 – List of symbols . 16
Table 201.102 – Distributed essential performance requirements . 18
Table 201.103 – Acoustic output reporting table . 22
Table 201.104 – Overview of the tests noted under 201.11.1.3 . 29
Table CC.1 – Relative importance of maintaining low exposure indices in various
scanning situations . 50
Table DD.1 – Acoustic and thermal properties of tissues and materials . 52
Table DD.2 – Weight % pure components . 53
Table EE.1 – Example of acoustic output table for third parties . 56
– 4 – IEC 60601-2-37:2024 RLV © IEC 2024
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-37: Particular requirements for the basic safety
and essential performance of ultrasonic medical
diagnostic and monitoring equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC Publication(s)"). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
This redline version of the official IEC Standard allows the user to identify the changes
made to the previous edition IEC 60601-2-37:2007+AMD1:2015 CSV. A vertical bar
appears in the margin wherever a change has been made. Additions are in green text,
deletions are in strikethrough red text.
IEC 60601-2-37 has been prepared by subcommittee 62B: Medical imaging equipment,
software, and systems, of IEC technical committee 62: Medical equipment, software, and
systems. It is an International Standard.
This third edition cancels and replaces the second edition published in 2007 and
Amendment 1:2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) technical and editorial changes resulting from the amended general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and
its collateral standards IEC 60601-1-xx,
b) technical and editorial changes as a result of maintenance to normative references;
c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics
standards. In particular, Clause 201.11 about protection against excessive temperatures
and other hazards has been fully revised.
The text of this International Standard is based on the following documents:
Draft Report on voting
62B/1318/CDV 62B/1348/RVC
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
In this document, the following print types are used:
– requirements and definitions: roman type.
– test specifications: italic type.
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020, IN THIS DOCUMENT OR AS NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by the
clause number. References to subclauses within this document are by number only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
– 6 – IEC 60601-2-37:2024 RLV © IEC 2024
The verbal forms used in this document conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance with
this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
• reconfirmed,
• withdrawn, or
• revised.
IMPORTANT – The "colour inside" logo on the cover page of this document indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this document using a colour printer.
INTRODUCTION
In this document, safety requirements additional to those in the general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, and IEC 60601-1:2005/AMD2:2020 are
specified for ULTRASONIC DIAGNOSTIC EQUIPMENT.
A general guidance and rationale for the requirements of this document are given in Annex AA.
Knowledge of the reasons for these requirements will not only facilitate the proper application
of this document but will, in due course, expedite any revision necessitated by changes in
clinical practice or as a result of developments in technology.
The approach and philosophy used in drafting this document for safety of ULTRASONIC
DIAGNOSTIC EQUIPMENT are consistent with those in standards of the IEC 60601-2 series that
apply to other diagnostic modalities, such as X-ray equipment and magnetic resonance systems.
In each case, the safety standard is intended to require increasing sophistication of output
display indicators and controls with increasing energy levels in the interrogating field of
diagnosis. Thus, for all such diagnostic modalities, it is the responsibility of the OPERATOR to
understand the risk of the output of the ULTRASONIC DIAGNOSTIC EQUIPMENT, and to act
appropriately in order to obtain the needed diagnostic information with the minimum risk to the
PATIENT.
INTRODUCTION TO AMENDMENT 1
The second edition of IEC 60601-2-37 was published in 2007. Since that publication, the parent
standard, IEC 60601-1:2005, entered maintenance, under which an amendment (IEC 60601-
1:2005/AMD1:2012) and a consolidated edition 3.1 (IEC 60601-1:2005 and IEC 60601-
1:2005/AMD1:2012) were published. This amendment to IEC 60601-2-37:2007 addresses
three issues:
1) technical changes proposed by National Committees as a result of 4 years of practical
usage,
2) technical and editorial changes resulting from the amended general standard
IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012 and its collateral standards
IEC 60601-1-xx, and
3) technical changes as a result of maintenance to normative references.
– 8 – IEC 60601-2-37:2024 RLV © IEC 2024
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-37: Particular requirements for the basic safety
and essential performance of ultrasonic medical
diagnostic and monitoring equipment
The clauses and subclauses of the general standard apply except as follows:
201.1 Scope, object and related standards
Clause 1 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, and
IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.1.1 *Scope
Replacement:
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC
DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
ME EQUIPMENT and to ME SYSTEMS, as relevant.
case, the clause or subclause applies both to
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within
the scope of this document are not covered by specific requirements in this document except
in 7.2.13 and 8.4.1 of this standard 201.7.2.13.
NOTE See also subclause 4.2 of this standard.
This document does not cover ultrasonic therapeutic equipment. Equipment used for the
imaging or diagnosis of body structures by ultrasound in conjunction with other medical
procedures is covered.
201.1.2 Object
Replacement:
The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217.
201.1.3 Collateral standards
Addition:
This document refers to those applicable collateral standards that are listed in Clause 2 of
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and
Clause 201.2 of this document.
IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020, IEC 60601-1-12:2014 and
IEC 60601-1-12:2014/AMD1:2020 apply as modified in Clause 202 and Clause 212 respectively.
All other published collateral standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 as appropriate for the particular ME EQUIPMENT under
consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard IEC 60601-
1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of sections, clauses and subclauses of this document corresponds to that of the
general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-
1:2005/AMD2:2020 with the prefix "201" (e.g. 201.1 in this document addresses the content of
Clause 1 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020) or applicable collateral standard with the prefix "20x" where x
is the final digit(s) of the collateral standard document number (e.g. 202.6 in this document
addresses the content of Clause 6 of the IEC 60601-1-2 collateral standard, 203.4 in this
document addresses the content of Clause 4 of the IEC 60601-1-3 collateral standard, etc.).
The changes to the text of the general standard IEC 60601-1:2005, IEC 60601-
1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are specified by the use of the following
words:
"Replacement" means that the clause or subclause of the general standard IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is replaced completely by this document.
"Addition" means that the text of this document is additional to the requirements of the general
standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-
1:2005/AMD2:2020 or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is amended as indicated by the text of this document.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.154, additional definitions in this document are numbered
beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items
aa), bb), etc.
Subclauses or figures which are additional to those of a collateral standard are numbered
starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for IEC 60601-1-2,
203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, any applicable collateral
standards and this document taken together.
Where there is no corresponding section, clause or subclause in this document, the section,
clause or subclause of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012
and IEC 60601-1:2005/AMD2:2020 or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or
– 10 – IEC 60601-2-37:2024 RLV © IEC 2024
applicable collateral standard, although possibly relevant, is not to be applied, a statement to
that effect is given in this document.
201.2 Normative references
Clause 2 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 applies except as follows:
Addition:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1:2005/AMD2:2020
IEC 60601-1-12:2014, Medical electrical equipment – Part 1-12: General requirements for basic
safety and essential performance – Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems intended for use in the emergency medical services
environment
IEC 60601-1-12:2014/AMD1:2020
IEC 60601-2-18:2009, Medical electrical equipment – Part 2-18: Particular requirements for the
basic safety and essential performance of endoscopic equipment
IEC 62127-1:20072022, Ultrasonics – Hydrophones – Part 1: Measurement and
characterization of medical ultrasonic fields up to 40 MHz
IEC 62127-1:2007/AMD1:2013
IEC 62359:2010, Ultrasonics – Field characterization – Test methods for the determination of
thermal and mechanical indices related to medical diagnostic ultrasonic fields
IEC 62359:2010/AMD1:2017
CISPR 11:2024, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurement
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in the general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and in
IEC 62359 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following
addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp
NOTE 1 An index of defined terms is given after the Bibliography.
NOTE 2 A list of symbols used in this document is found in Table 201.101.
___________
There exists a consolidated edition (3.1) including IEC 60601-1:2005 and its Amendment 1 (2012).
There exists a consolidated edition (1.1) including IEC 62127-1:2007 and its Amendment 1 (2013).
Addition:
201.3.201
BONE THERMAL INDEX
TIB
THERMAL INDEX for applications such as foetal (second and third trimester), in which the
ultrasound beam passes through soft tissue and a focal region is in the immediate vicinity of
bone
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.17, modified – The definition
no longer refers to neonatal cephalic applications, and The original notes have been deleted.]
201.3.202
COMBINED-OPERATING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that combines more than one DISCRETE-
OPERATING MODE
201.3.203
CRANIAL-BONE THERMAL INDEX
TIC
THERMAL INDEX for applications, in which the ultrasound beam passes through bone near the
beam entrance into the body, such as paediatric and adult cranial or neonatal cephalic
applications
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.21, modified – The definition
now includes a reference to neonatal cephalic applications, and The original notes have been
deleted.]
201.3.204
DEFAULT SETTING
specific state of control the ULTRASONIC DIAGNOSTIC EQUIPMENT will enter upon power-up, new
PATIENT select, or change from non-foetal to foetal applications
201.3.205
DISCRETE-OPERATING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT in which the purpose of the excitation
of the ULTRASONIC TRANSDUCER or ULTRASONIC TRANSDUCER element group is to utilise only one
diagnostic methodology
201.3.206
FULL SOFTWARE CONTROL OF ACOUSTIC OUTPUT
means by which the ULTRASONIC DIAGNOSTIC EQUIPMENT manages the acoustic output
independent of direct OPERATOR control
201.3.207
INVASIVE TRANSDUCER ASSEMBLY
transducer which, in whole or in part, penetrates inside the body, either through a body orifice
or through the surface of the body
– 12 – IEC 60601-2-37:2024 RLV © IEC 2024
201.3.208
MECHANICAL INDEX
the displayed parameter representing potential cavitation bioeffects
Indicator of the risk for bioeffects due to mechanical or nonthermal mechanisms, such as
cavitation
Symbol: MI
Unit: None
Note 1 to entry: See IEC 62359 for methods of determining the MECHANICAL INDEX.
201.3.209
MULTI-PURPOSE ULTRASONIC DIAGNOSTIC EQUIPMENT
ULTRASONIC DIAGNOSTIC EQUIPMENT that is intended for more than one clinical application
201.3.210
NON-SCANNING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that involves a sequence of ultrasonic
pulses that give rise to ultrasonic scan lines that follow the same acoustic path
201.3.211
PRUDENT USE STATEMENT
affirmation of the principle that only necessary clinical information should be acquired and that
high exposure levels and long exposure times should be avoided
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.40, modified – The definition
has been reworded.]
201.3.212
SCANNING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that involves a sequence of ultrasonic
pulses that give rise to scan lines that do not follow the same acoustic path
201.3.213
SOFT TISSUE THERMAL INDEX
TIS
THERMAL INDEX related to soft tissues
Unit: None
[SOURCE: IEC 62359:2010, 3.52, modified – The original notes have been deleted.]
201.3.214
THERMAL INDEX
TI
indicator of the risk of bioeffect due to thermal mechanisms expressed as the ratio of
ATTENUATED OUTPUT POWER at a specified point to the ATTENUATED OUTPUT POWER required to
raise the temperature at that point in a specific tissue model by 1 °C
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.56, modified – The term
"ATTENUATED ACOUSTIC POWER" has been replaced twice by the term "ATTENUATED OUTPUT
POWER" Addition of “indicator of the risk of bioeffect due to thermal mechanisms expressed as
the”, and the original note has been deleted.]
201.3.215
TRANSDUCER ASSEMBLY
those parts of ULTRASONIC DIAGNOSTIC EQUIPMENT comprising either the ULTRASONIC TRANSDUCER
and/or the ULTRASONIC TRANSDUCER ELEMENT GROUP, or both, together with any integral
components, such as an acoustic lens or integral stand-off
Note 1 to entry: The TRANSDUCER ASSEMBLY is usually separable from the ultrasound instrument console.
[SOURCE: IEC 62127-1:2007, 3.69 IEC 62359:2010 and IEC 62359:2010AMD1:2017, 3.57,
modified – the original term "medical diagnostic ultrasound equipment" has been replaced by
"ULTRASONIC DIAGNOSTIC EQUIPMENT" in the definition.]
201.3.216
TRANSMIT PATTERN
combination of a specific set of transducer beam-forming characteristics (determined by the
transmit aperture size, apodisation shape, and relative timing/phase delay pattern across the
aperture, resulting in a specific focal length and direction), and an electrical drive waveform of
a specific fixed shape but variable amplitude
[SOURCE: IEC 62359:2010, 3.58]
201.3.217
ULTRASONIC DIAGNOSTIC EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT that is intended for ultrasonic medical examination
201.3.218
ULTRASONIC TRANSDUCER
device capable of converting electrical energy to mechanical energy within the ultrasonic
frequency range and/or reciprocally of converting mechanical energy to electrical energy
[SOURCE: IEC 62127-1:2007/AMD1:2013, 3.73 IEC 62127-1:2022, 3.88]
201.3.219
ATTENUATED PULSE-AVERAGE INTENSITY
I
pa,α
value of the acoustic PULSE-AVERAGE INTENSITY after attenuation and at a specified point, and
given by
(−αzf /10 dB)
awf
I = I (z)10
pa,α pa
(−αzf /10dB)
awf
I zI= z 10 (1)
( ) ( )
pa,α pa
where
α is the ACOUSTIC ATTENUATION COEFFICIENT as defined in IEC 62359:2010, definition 3.1;
z is the distance from the EXTERNAL TRANSDUCER APERTURE to the point of interest;
f is the ACOUSTIC WORKING FREQUENCY as defined in IEC 62359:2010 and
awf
IEC 62359:2010/AMD1:2017, definition 3.4;
I (z) is the PULSE-AVERAGE INTENSITY measured in water as defined in IEC 62127-1:2007
pa
and IEC 62127-1:2007/AMD1:2013, definition 3.47 IEC 62127-1:2022, 3.53.
–2
Unit: W m
– 14 – IEC 60601-2-37:2024 RLV © IEC 2024
201.3.220
NUMBER OF PULSES PER ULTRASONIC SCAN LINE
n
pps
number of acoustic pulses travelling along a particular ULTRASONIC SCAN LINE
Note 1 to entry: Here ULTRASONIC SCAN LINE refers to the path of acoustic pulses on a particular BEAM AXIS in
SCANNING and NON-SCANNING MODES.
Note 2 to entry: This number can be used in the calculation of any ultrasound temporal average value from
HYDROPHONE measurements.
Note 3 to entry: The following shows an example of the NUMBER OF PULSES PER ULTRASONIC SCAN LINE and the
NUMBER OF ULTRASONIC SCAN LINES (";" indicates the end of a frame):
1 2 3 4; 1 2 3 4; 1 2 3 4… n = 1; n = 4
pps sl
1 1 2 2 3 3 4 4; 1 1 2 2 3 3 4 4; … n = 2; n = 4
pps sl
1 1 1 1 2 2 2 2 3 3 3 3 4 4 4 4; 1 1 1 1 2 2 2 2 3 3 3 3 4 4 4 4; … n = 4; n = 4
pps sl
1 1 2 2 3 3 4 4 1 1 1 2 2 2 3 3 3 4 4 4; 1 1 2 2 3 3 4 4 1 1 1 2 2 2 3 3 3 4 4 4; … n = 5; n = 4 (within one frame
pps sl
the pulses down each line may not occur contiguously).
Within one frame, all scan lines may not have the same n value. An example is: 1 2 2 3 3 4; 1 2 2 3 3 4; … avg
pps
n = 1,5; max n = 2; n = 4
pps pps sl
[SOURCE: IEC 61157:2007/AMD1:2013, 3.45, modified – The fourth example in the Note 3 to
entry has been corrected.]
201.3.221
ULTRASOUND ENDOSCOPE
ENDOSCOPE with built-in ULTRASOUND TRANSDUCERS
201.3.222
ENDOSCOPE
medical instrument having viewing means, with or without optics, introduced into a body cavity
through a natural or surgically created body opening for examination, diagnosis or therapy
Note 1 to entry: ENDOSCOPES may be of rigid, flexible or capsule type, each of which may have different image
pick-up systems (e.g. via lenses or electronic/ultrasonic sensors) and different image transmission systems (e.g.
optical (via lenses or fibre bundles), or electrical/electronic).
Note 2 to entry: Note 1 to entry differs from NOTE 1 of definition 3.1 in ISO 8600-1 in order to include 'capsule'
endoscopes.
[SOURCE: IEC 60601-2-18:2009, 201.3.203]
201.3.223
DEPTH FOR PEAK PULSE-INTENSITY INTEGRAL
z
pii
position of maximum SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY for NON-SCANNING MODE
components, determined beyond the BREAK-POINT DEPTH, z , on the BEAM-AXIS
bp
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z from the external transducer
bp
aperture to the plane of maximum PULSE-INTENSITY INTEGRAL (pii) as approximated by the PULSE-
PRESSURE-SQUARED INTEGRAL (ppsi)
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.24, modified ─ Notes to entry
1, 2 and 3 deleted.]
201.3.224
DEPTH FOR PEAK ATTENUATED PULSE-INTENSITY INTEGRAL
z
pii, α
position of maximum ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY for NON-SCANNING
MODE components, determined beyond the BREAK-POINT DEPTH, z , on the BEAM-AXIS
bp
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak ATTENUATED PULSE-
bp
INTENSITY INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.71, modified ─ Main term
” replaced with “depth for peak attenuated pulse-intensity integral”,
“depth for maximum
pii, α
Notes to entry 1, 2 and 3 deleted and addition of a new Note to entry. ]
201.3.225
DEPTH FOR PEAK SUM OF PULSE-INTENSITY INTEGRALS
z
sii
position of maximum SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY for SCANNING MODE
components, determined beyond THE BREAK-POINT DEPTH, z , on the BEAM-AXIS
bp
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak SCAN-INTENSITY
bp
INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
Note 2 to entry: The subscript 'sii' indicates the scan intensity integral (sii). The sii for SCANNING MODE components
at a particular point is determined from the sum over a complete scan frame of the PULSE-INTENSITY INTEGRALS of the
ULTRASONIC SCAN LINES that make up the scanning components of a combined mode. Non-scanned components are
excluded from the sum. See IEC 62359 and IEC 62127-1 for more details.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.74, modified ─ Main term
“depth for maximum sii” replaced with “depth for peak sum of pulse-intensity integrals”, Notes
to entry 1, 2 and 3 deleted and addition of new Notes 1 and 2 to entry.]
201.3.226
DEPTH FOR PEAK SUM OF ATTENUATED PULSE-INTENSITY INTEGRALS
z
sii,α
position of maximum ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY for SCANNING
MODE components, determined beyond the BREAK-POINT DEPTH, z , on THE BEAM-AXIS
bp
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak ATTENUATED SCAN-
bp
INTENSITY INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
Note 2 to entry: The subscript "sii" indicates the "scan intensity integral" that is the sum at a particular point of the
PULSE-INTENSITY INTEGRALS of the ULTRASONIC SCAN LINES comprising a SCANNING MODE component. See IEC 62359
and IEC 62127-1 for additional details.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.75, modified ─ Main term
“depth for maximum sii ” replaced with “depth for peak sum of attenuated pulse-intensity
α
integrals”, addition of “peak” in the definition, Notes to entry 1, 2 and 3 deleted and addition of
new Notes 1 and 2 to entry.]
– 16 – IEC 60601-2-37:2024 RLV © IEC 2024
201.3.227
DEPTH FOR MECHANICAL INDEX
z
MI
depth on the BEAM-AXIS from the EXTERNAL TRANSDUCER APERTURE to the plane of maximum
ATTENUATED PULSE INTENSITY INTEGRAL (pii ) ATTENUATED PULSE-PRESSURE-SQUARED-INTEGRAL
α
(ppsi )
α
Unit: m
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.23]
201.3.228
THERMAL OFFSET
∆T
offset
difference between a) the temperature of the APPLIED PART of the TRANSDUCER ASSEMBLY at
steady-state in the measurement setting before transmitting begins and b) the steady-state
temperature at the same location in the measurement setting when the TRANSDUCER ASSEMBLY
was not present
Note 1 to entry: The value of the THERMAL OFFSET can be positive, negative or zero.
201.3.229
ULTRASOUND
acoustic oscillation whose frequency is above the high-frequency limit of audible sound (about
20 kHz)
[SOURCE: IEC 60050-802:2011, 802-01-01]
Table 201.101 – List of symbols
Symbol Term Reference
A = –12 dB OUTPUT BEAM AREA IEC 62359
aprt
d = EQUIVALENT BEAM DIAMETER IEC 62359
eq
f = ACOUSTIC WORKING FREQUENCY IEC 62359
awf
I = ATTENUATED PULSE-AVERAGE INTENSITY
pa,α
pii = PULSE-INTENSITY INTEGRAL IEC 62359
pii = ATTENUATED PULSE-INTENSITY INTEGRAL IEC 62359
α
I = ATTENUATED SPATIAL-PEAK PULSE-AVERAGE INTENSITY IEC 62359
sppa,α
I = SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY IEC 62359
spta
I = ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY IEC 62359
spta,α
I (z) = ATTENUATED TEMPORAL-AVERAGE INTENSITY IEC 62359
ta,α
MI = MECHANICAL INDEX IEC 62359
n = NUMBER OF PULSES PER ULTRASONIC SCAN LINE IEC 61157
pps
P = OUTPUT POWER IEC 62359
P = BOUNDED-SQUARE OUTPUT POWER IEC 62359
1x1
P = ATTENUATED OUTPUT POWER IEC 62359
α
p = ATTENUATED PEAK-RAREFACTIONAL ACOUSTIC PRESSURE IEC 62359
r,α
p = PEAK-RAREFACTIONAL ACOUSTIC PRESSURE IEC 62359
r
prr = PULSE REPETITION RATE IEC 62359
Symbol Term Reference
srr = SCAN REPETITION RATE IEC 62127-1
TI = THERMAL INDEX IEC 62359
TIB = BONE THERMAL INDEX IEC 62359
TIC = CRANIAL-BONE THERMAL INDEX IEC 62359
TIS = SOFT-TISSUE THERMAL INDEX IEC 62359
t = PULSE DURATION IEC 62359
d
X,Y = –12 dB OUTPUT BEAM DIMENSIONS IEC 62359
z = DEPTH FOR BONE THERMAL INDEX IEC 62359
b
z = BREAK-POINT DEPTH IEC 62359
bp
z = DEPTH FOR PEAK PULSE-INTENSITY INTEGRAL IEC 62359
pii
z = DEPTH FOR MECHANICAL INDEX IEC 62359
MI
z = DEPTH FOR PEAK ATTENUATED PULSE INTENSITY INTEGRAL IEC 62359
pii,α
z = DEPTH FOR PEAK SUM OF PULSE INTENSITY INTEGRALS IEC 62359
sii
z = DEPTH FOR PEAK SUM OF ATTENUATED PULSE INTENSITY INTEGRALS IEC 62359
sii,α
z = DEPTH FOR TIS IEC 62359
s
= THERMAL OFFSET
∆T
offset
∆T = temperature rise caused by the TRANSDUCER ASSEMBLY transmitting
tx
201.4 General requirements
Clause 4 of the general standard IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.4.1 Conditions for application to ME EQUIPMENT or ME SYSTEMS
Addition:
An ULTRASOUND ENDOSCOPE where the imaging means is limited to ultrasound shall be
considered an ULTRASOUND TRANSDUCER and shall meet the requirements of this document.
NOTE 1 Examples of such ULTRASOUND TRANSDUCERS include transvaginal, transesophageal (TEE), rectal,
laparoscopic and other similar intra-cavity probes.
An ULTRASOUND ENDOSCOPE having imaging means in addition to ultrasound shall also meet the
requirements of 201.11.6.5 of IEC 60601-2-18:2009.
NOTE 2 Examples of such additional imaging means include optical and CCD.
201.4.3 ESSENTIAL PERFORMANCE
Additional subclause:
201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
Table 201.102 lists the potential sources of unacceptable risk identified to characterize the
ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT and the subclauses in which the
requirements are found.
– 18 – IEC 60601-2-37:2024 RLV © IEC 2024
Table 201.102 – Distributed essential performance requirements
Requirement Subclause
Free from noise on a waveform or artefacts or distortion in an image or error of a
displayed numerical value which cannot be attributed to a physiological effect and which 202.6.2.1.10202.8.1
may can alter the diagnosis.
Free from the display of incorrect numerical values associated with the diagnosis to be
202.6.2.1.10202.8.1
a
performed .
201.12.4.2
a
Free from the display of incorrect safety-related indications.
202.6.2.1.10202.8.1
201.10.101
Free from the production of unintended or excessive ultrasound output.
202.6.2.1.10202.8.1
Free from the production of unintended or excessive TRANSDUCER ASSEMBLY surface
202.6.2.1.10202.8.1
temperature.
Free from the production of unintended or uncontrolled motion of TRANSDUCER
202.6.2.1.10202.8.1
ASSEMBLIES intended for intra-corporeal use.
a
"incorrect" in the sense that the displayed value differs from what is calculated (having been altered during
data transfer), or the calculation itself is not correct.
NOTE In some circumstances the need f
...
IEC 60601-2-37 ®
Edition 3.0 2024-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-37: Particular requirements for the basic safety and essential performance
of ultrasonic medical diagnostic and monitoring equipment
Appareils électromédicaux –
Partie 2-37: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils de diagnostic et de surveillance médicaux à ultrasons
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et
les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews, graphical symbols and the glossary.
committee, …). It also gives information on projects, replaced With a subscription you will always have access to up to date
and withdrawn publications. content tailored to your needs.
IEC Just Published - webstore.iec.ch/justpublished
Electropedia - www.electropedia.org
Stay up to date on all new IEC publications. Just Published
The world's leading online dictionary on electrotechnology,
details all new publications released. Available online and once
containing more than 22 500 terminological entries in English
a month by email.
and French, with equivalent terms in 25 additional languages.
Also known as the International Electrotechnical Vocabulary
IEC Customer Service Centre - webstore.iec.ch/csc
(IEV) online.
If you wish to give us your feedback on this publication or need
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Recherche de publications IEC - IEC Products & Services Portal - products.iec.ch
webstore.iec.ch/advsearchform Découvrez notre puissant moteur de recherche et consultez
La recherche avancée permet de trouver des publications IEC gratuitement tous les aperçus des publications, symboles
en utilisant différents critères (numéro de référence, texte, graphiques et le glossaire. Avec un abonnement, vous aurez
comité d’études, …). Elle donne aussi des informations sur les toujours accès à un contenu à jour adapté à vos besoins.
projets et les publications remplacées ou retirées.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
Le premier dictionnaire d'électrotechnologie en ligne au monde,
Restez informé sur les nouvelles publications IEC. Just
avec plus de 22 500 articles terminologiques en anglais et en
Published détaille les nouvelles publications parues.
français, ainsi que les termes équivalents dans 25 langues
Disponible en ligne et une fois par mois par email.
additionnelles. Egalement appelé Vocabulaire
Electrotechnique International (IEV) en ligne.
Service Clients - webstore.iec.ch/csc
Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
IEC 60601-2-37 ®
Edition 3.0 2024-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-37: Particular requirements for the basic safety and essential performance
of ultrasonic medical diagnostic and monitoring equipment
Appareils électromédicaux –
Partie 2-37: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils de diagnostic et de surveillance médicaux à ultrasons
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.55, 17.140.50 ISBN 978-2-8322-8920-4
– 2 – IEC 60601-2-37:2024 © IEC 2024
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 10
201.3 Terms and definitions . 10
201.4 General requirements . 17
201.5 General requirements for testing ME EQUIPMENT . 18
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 18
201.7 ME EQUIPMENT identification, marking and documents . 18
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 23
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 23
201.10 Protection against unwanted and excessive radiation HAZARDS . 23
201.11 Protection against excessive temperatures and other HAZARDS . 24
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 29
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 31
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 31
201.15 Construction of ME EQUIPMENT . 31
201.16 ME SYSTEMS . 31
201.17 *Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 31
202 ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 31
212 Requirements for medical electrical equipment and medical electrical
systems intended for use in the emergency medical services environment
(EMS) . 33
Annexes . 34
Annex AA (informative) Particular guidance and rationale . 35
Annex BB (informative) Guidance in classification according to CISPR 11 . 43
Annex CC (informative) Guidance to the MANUFACTURER on the interpretation of TI and
MI to be used to inform the OPERATOR . 44
Annex DD (informative) Example set-up to measure surface temperature of externally
applied TRANSDUCER ASSEMBLIES . 48
Annex EE (informative) Acoustic output table intended for third parties . 51
Bibliography . 54
Index of defined terms . 58
Figure AA.1 – Method a) for an external probe . 38
Figure AA.2 – Method b) for an external probe . 39
Figure AA.3 – Method b) for an external probe . 39
Figure DD.1 – Set-up of an example test object to measure the surface temperature of
externally applied transducers . 50
Table 201.101 – List of symbols . 16
Table 201.102 – Distributed essential performance requirements . 18
Table 201.103 – Acoustic output reporting table . 22
Table 201.104 – Overview of the tests noted under 201.11.1.3 . 28
Table CC.1 – Relative importance of maintaining low exposure indices in various
scanning situations . 46
Table DD.1 – Acoustic and thermal properties of tissues and materials . 48
Table DD.2 – Weight % pure components . 49
Table EE.1 – Example of acoustic output table for third parties . 52
– 4 – IEC 60601-2-37:2024 © IEC 2024
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-37: Particular requirements for the basic safety
and essential performance of ultrasonic medical
diagnostic and monitoring equipment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC Publication(s)"). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
IEC 60601-2-37 has been prepared by subcommittee 62B: Medical imaging equipment,
software, and systems, of IEC technical committee 62: Medical equipment, software, and
systems. It is an International Standard.
This third edition cancels and replaces the second edition published in 2007 and
Amendment 1:2015. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) technical and editorial changes resulting from the amended general standard
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and
its collateral standards IEC 60601-1-xx,
b) technical and editorial changes as a result of maintenance to normative references;
c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics
standards. In particular, Clause 201.11 about protection against excessive temperatures
and other hazards has been fully revised.
The text of this International Standard is based on the following documents:
Draft Report on voting
62B/1318/CDV 62B/1348/RVC
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
In this document, the following print types are used:
– requirements and definitions: roman type.
– test specifications: italic type.
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020, IN THIS DOCUMENT OR AS NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by the
clause number. References to subclauses within this document are by number only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
The verbal forms used in this document conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance with
this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website
– 6 – IEC 60601-2-37:2024 © IEC 2024
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
• reconfirmed,
• withdrawn, or
• revised.
IMPORTANT – The "colour inside" logo on the cover page of this document indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this document using a colour printer.
INTRODUCTION
In this document, safety requirements additional to those in IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012, and IEC 60601-1:2005/AMD2:2020 are specified for
ULTRASONIC DIAGNOSTIC EQUIPMENT.
A general guidance and rationale for the requirements of this document are given in Annex AA.
Knowledge of the reasons for these requirements will not only facilitate the proper application
of this document but will, in due course, expedite any revision necessitated by changes in
clinical practice or as a result of developments in technology.
The approach and philosophy used in drafting this document for safety of ULTRASONIC
DIAGNOSTIC EQUIPMENT are consistent with those in standards of the IEC 60601-2 series that
apply to other diagnostic modalities, such as X-ray equipment and magnetic resonance systems.
In each case, the safety standard is intended to require increasing sophistication of output
display indicators and controls with increasing energy levels in the interrogating field of
diagnosis. Thus, for all such diagnostic modalities, it is the responsibility of the OPERATOR to
understand the risk of the output of the ULTRASONIC DIAGNOSTIC EQUIPMENT, and to act
appropriately in order to obtain the needed diagnostic information with the minimum risk to the
PATIENT.
– 8 – IEC 60601-2-37:2024 © IEC 2024
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-37: Particular requirements for the basic safety
and essential performance of ultrasonic medical
diagnostic and monitoring equipment
201.1 Scope, object and related standards
Clause 1 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, and
IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.1.1 *Scope
Replacement:
This document applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC
DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within
the scope of this document are not covered by specific requirements in this document except
in 201.7.2.13.
This document does not cover ultrasonic therapeutic equipment. Equipment used for the
imaging or diagnosis of body structures by ultrasound in conjunction with other medical
procedures is covered.
201.1.2 Object
Replacement:
The object of this document is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE
requirements for ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217.
201.1.3 Collateral standards
Addition:
This document refers to those applicable collateral standards that are listed in Clause 2 of
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and
Clause 201.2 of this document.
IEC 60601-1-2:2014, IEC 60601-1-2:2014/AMD1:2020, IEC 60601-1-12:2014 and
IEC 60601-1-12:2014/AMD1:2020 apply as modified in Clause 202 and Clause 212 respectively.
All other published collateral standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 as appropriate for the particular ME EQUIPMENT under
consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE requirements.
A requirement of a particular standard takes priority over IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020.
The numbering of sections, clauses and subclauses of this document corresponds to that of
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 with the
prefix "201" (e.g. 201.1 in this document addresses the content of Clause 1 of
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020) or
applicable collateral standard with the prefix "20x" where x is the final digit(s) of the collateral
standard document number (e.g. 202.6 in this document addresses the content of Clause 6 of
the IEC 60601-1-2 collateral standard, 203.4 in this document addresses the content of Clause
4 of the IEC 60601-1-3 collateral standard, etc.). The changes to the text of IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 are specified by the use of
the following words:
"Replacement" means that the clause or subclause of IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is replaced completely by this document.
"Addition" means that the text of this document is additional to the requirements of
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or
applicable collateral standard.
"Amendment" means that the clause or subclause of IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard is amended as indicated by the text of this document.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.154, additional definitions in this document are numbered
beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and additional items
aa), bb), etc.
Subclauses or figures which are additional to those of a collateral standard are numbered
starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for IEC 60601-1-2,
203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, any applicable collateral
standards and this document taken together.
Where there is no corresponding section, clause or subclause in this document, the section,
clause or subclause of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 or applicable collateral standard, although possibly not relevant,
applies without modification; where it is intended that any part of IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 or applicable collateral
standard, although possibly relevant, is not to be applied, a statement to that effect is given in
this document.
– 10 – IEC 60601-2-37:2024 © IEC 2024
201.2 Normative references
Clause 2 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 applies except as follows:
Addition:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1:2005/AMD2:2020
IEC 60601-1-12:2014, Medical electrical equipment – Part 1-12: General requirements for basic
safety and essential performance – Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems intended for use in the emergency medical services
environment
IEC 60601-1-12:2014/AMD1:2020
IEC 60601-2-18:2009, Medical electrical equipment – Part 2-18: Particular requirements for the
basic safety and essential performance of endoscopic equipment
IEC 62127-1:2022, Ultrasonics – Hydrophones – Part 1: Measurement and characterization of
medical ultrasonic fields
IEC 62359:2010, Ultrasonics – Field characterization – Test methods for the determination of
thermal and mechanical indices related to medical diagnostic ultrasonic fields
IEC 62359:2010/AMD1:2017
CISPR 11:2024, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurement
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and in IEC 62359 and the
following apply.
ISO and IEC maintain terminology databases for use in standardization at the following
addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp
NOTE 1 An index of defined terms is given after the Bibliography.
NOTE 2 A list of symbols used in this document is found in Table 201.101.
Addition:
201.3.201
BONE THERMAL INDEX
TIB
THERMAL INDEX for applications such as foetal (second and third trimester), in which the
ultrasound beam passes through soft tissue and a focal region is in the immediate vicinity of
bone
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.17, modified – The original
notes have been deleted.]
201.3.202
COMBINED-OPERATING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that combines more than one DISCRETE-
OPERATING MODE
201.3.203
CRANIAL-BONE THERMAL INDEX
TIC
THERMAL INDEX for applications, in which the ultrasound beam passes through bone near the
beam entrance into the body, such as paediatric and adult cranial or neonatal cephalic
applications
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.21, modified – The original
notes have been deleted.]
201.3.204
DEFAULT SETTING
specific state of control the ULTRASONIC DIAGNOSTIC EQUIPMENT will enter upon power-up, new
PATIENT select, or change from non-foetal to foetal applications
201.3.205
DISCRETE-OPERATING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT in which the purpose of the excitation
of the ULTRASONIC TRANSDUCER or ULTRASONIC TRANSDUCER element group is to utilise only one
diagnostic methodology
201.3.206
FULL SOFTWARE CONTROL OF ACOUSTIC OUTPUT
means by which the ULTRASONIC DIAGNOSTIC EQUIPMENT manages the acoustic output
independent of direct OPERATOR control
201.3.207
INVASIVE TRANSDUCER ASSEMBLY
transducer which, in whole or in part, penetrates inside the body, either through a body orifice
or through the surface of the body
– 12 – IEC 60601-2-37:2024 © IEC 2024
201.3.208
MECHANICAL INDEX
Indicator of the risk for bioeffects due to mechanical or nonthermal mechanisms, such as
cavitation
Symbol: MI
Unit: None
Note 1 to entry: See IEC 62359 for methods of determining the MECHANICAL INDEX.
201.3.209
MULTI-PURPOSE ULTRASONIC DIAGNOSTIC EQUIPMENT
ULTRASONIC DIAGNOSTIC EQUIPMENT that is intended for more than one clinical application
201.3.210
NON-SCANNING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that involves a sequence of ultrasonic
pulses that give rise to ultrasonic scan lines that follow the same acoustic path
201.3.211
PRUDENT USE STATEMENT
affirmation of the principle that only necessary clinical information should be acquired and that
high exposure levels and long exposure times should be avoided
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.40]
201.3.212
SCANNING MODE
mode of operation of ULTRASONIC DIAGNOSTIC EQUIPMENT that involves a sequence of ultrasonic
pulses that give rise to scan lines that do not follow the same acoustic path
201.3.213
SOFT TISSUE THERMAL INDEX
TIS
THERMAL INDEX related to soft tissues
Unit: None
[SOURCE: IEC 62359:2010, 3.52, modified – The original notes have been deleted.]
201.3.214
THERMAL INDEX
TI
indicator of the risk of bioeffect due to thermal mechanisms expressed as the ratio of
ATTENUATED OUTPUT POWER at a specified point to the ATTENUATED OUTPUT POWER required to
raise the temperature at that point in a specific tissue model by 1 °C
Unit: None
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.56, modified – Addition of
“indicator of the risk of bioeffect due to thermal mechanisms expressed as the”, and the original
note has been deleted.]
201.3.215
TRANSDUCER ASSEMBLY
those parts of ULTRASONIC DIAGNOSTIC EQUIPMENT comprising either the ULTRASONIC TRANSDUCER
and the ULTRASONIC TRANSDUCER ELEMENT GROUP, or both, together with any integral
components, such as an acoustic lens or integral stand-off
Note 1 to entry: The TRANSDUCER ASSEMBLY is usually separable from the ultrasound instrument console.
[SOURCE: IEC 62359:2010 and IEC 62359:2010AMD1:2017, 3.57, modified – "medical
diagnostic ultrasound equipment" has been replaced by "ULTRASONIC DIAGNOSTIC EQUIPMENT" in
the definition.]
201.3.216
TRANSMIT PATTERN
combination of a specific set of transducer beam-forming characteristics (determined by the
transmit aperture size, apodisation shape, and relative timing/phase delay pattern across the
aperture, resulting in a specific focal length and direction), and an electrical drive waveform of
a specific fixed shape but variable amplitude
[SOURCE: IEC 62359:2010, 3.58]
201.3.217
ULTRASONIC DIAGNOSTIC EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT that is intended for ultrasonic medical examination
201.3.218
ULTRASONIC TRANSDUCER
device capable of converting electrical energy to mechanical energy within the ultrasonic
frequency range and reciprocally of converting mechanical energy to electrical energy
[SOURCE: IEC 62127-1:2022, 3.88]
201.3.219
ATTENUATED PULSE-AVERAGE INTENSITY
I
pa,α
value of the acoustic PULSE-AVERAGE INTENSITY after attenuation and at a specified point, and
given by
−αzf /10dB
( )
awf
(1)
I zI= z 10
( ) ( )
pa,α pa
where
α is the ACOUSTIC ATTENUATION COEFFICIENT as defined in IEC 62359:2010, definition 3.1;
z is the distance from the EXTERNAL TRANSDUCER APERTURE to the point of interest;
f is the ACOUSTIC WORKING FREQUENCY as defined in IEC 62359:2010 and
awf
IEC 62359:2010/AMD1:2017, definition 3.4;
I (z) is the PULSE-AVERAGE INTENSITY measured in water as defined in IEC 62127-1:2022,
pa
3.53.
–2
Unit: W m
– 14 – IEC 60601-2-37:2024 © IEC 2024
201.3.220
NUMBER OF PULSES PER ULTRASONIC SCAN LINE
n
pps
number of acoustic pulses travelling along a particular ULTRASONIC SCAN LINE
Note 1 to entry: Here ULTRASONIC SCAN LINE refers to the path of acoustic pulses on a particular BEAM AXIS in
SCANNING and NON-SCANNING MODES.
Note 2 to entry: This number can be used in the calculation of any ultrasound temporal average value from
HYDROPHONE measurements.
Note 3 to entry: The following shows an example of the NUMBER OF PULSES PER ULTRASONIC SCAN LINE and the
NUMBER OF ULTRASONIC SCAN LINES (";" indicates the end of a frame):
1 2 3 4; 1 2 3 4; 1 2 3 4… n = 1; n = 4
pps sl
1 1 2 2 3 3 4 4; 1 1 2 2 3 3 4 4; … n = 2; n = 4
pps sl
1 1 1 1 2 2 2 2 3 3 3 3 4 4 4 4; 1 1 1 1 2 2 2 2 3 3 3 3 4 4 4 4; … n = 4; n = 4
pps sl
1 1 2 2 3 3 4 4 1 1 1 2 2 2 3 3 3 4 4 4; 1 1 2 2 3 3 4 4 1 1 1 2 2 2 3 3 3 4 4 4; … n = 5; n = 4 (within one frame
pps sl
the pulses down each line may not occur contiguously).
Within one frame, all scan lines may not have the same n value. An example is: 1 2 2 3 3 4; 1 2 2 3 3 4; … avg
pps
n = 1,5; max n = 2; n = 4
pps pps sl
[SOURCE: IEC 61157:2007/AMD1:2013, 3.45, modified – The fourth example in the Note 3 to
entry has been corrected.]
201.3.221
ULTRASOUND ENDOSCOPE
ENDOSCOPE with built-in ULTRASOUND TRANSDUCERS
201.3.222
ENDOSCOPE
medical instrument having viewing means, with or without optics, introduced into a body cavity
through a natural or surgically created body opening for examination, diagnosis or therapy
Note 1 to entry: ENDOSCOPES may be of rigid, flexible or capsule type, each of which may have different image
pick-up systems (e.g. via lenses or electronic/ultrasonic sensors) and different image transmission systems (e.g.
optical (via lenses or fibre bundles), or electrical/electronic).
Note 2 to entry: Note 1 to entry differs from NOTE 1 of definition 3.1 in ISO 8600-1 in order to include 'capsule'
endoscopes.
[SOURCE: IEC 60601-2-18:2009, 201.3.203]
201.3.223
DEPTH FOR PEAK PULSE-INTENSITY INTEGRAL
z
pii
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z from the external transducer
bp
aperture to the plane of maximum PULSE-INTENSITY INTEGRAL (pii) as approximated by the PULSE-
PRESSURE-SQUARED INTEGRAL (ppsi)
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.24, modified ─ Notes to entry
1, 2 and 3 deleted.]
201.3.224
DEPTH FOR PEAK ATTENUATED PULSE-INTENSITY INTEGRAL
z
pii, α
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak ATTENUATED PULSE-
bp
INTENSITY INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.71, modified ─ Main term
“depth for maximum ” replaced with “depth for peak attenuated pulse-intensity integral”,
pii, α
Notes to entry 1, 2 and 3 deleted and addition of a new Note to entry. ]
201.3.225
DEPTH FOR PEAK SUM OF PULSE-INTENSITY INTEGRALS
z
sii
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak SCAN-INTENSITY
bp
INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
Note 2 to entry: The subscript 'sii' indicates the scan intensity integral (sii). The sii for SCANNING MODE components
at a particular point is determined from the sum over a complete scan frame of the PULSE-INTENSITY INTEGRALS of the
ULTRASONIC SCAN LINES that make up the scanning components of a combined mode. Non-scanned components are
excluded from the sum. See IEC 62359 and IEC 62127-1 for more details.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.74, modified ─ Main term
“depth for maximum sii” replaced with “depth for peak sum of pulse-intensity integrals”, Notes
to entry 1, 2 and 3 deleted and addition of new Notes 1 and 2 to entry.]
201.3.226
DEPTH FOR PEAK SUM OF ATTENUATED PULSE-INTENSITY INTEGRALS
z
sii,α
depth z on the BEAM AXIS and beyond the BREAK-POINT DEPTH z of peak ATTENUATED SCAN-
bp
INTENSITY INTEGRAL
Unit: m
Note 1 to entry: BEAM-AXIS and BREAK-POINT DEPTH are defined in IEC 62359.
Note 2 to entry: The subscript "sii" indicates the "scan intensity integral" that is the sum at a particular point of the
PULSE-INTENSITY INTEGRALS of the ULTRASONIC SCAN LINES comprising a SCANNING MODE component. See IEC 62359
and IEC 62127-1 for additional details.
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.75, modified ─ Main term
“depth for maximum sii ” replaced with “depth for peak sum of attenuated pulse-intensity
α
integrals”, addition of “peak” in the definition, Notes to entry 1, 2 and 3 deleted and addition of
new Notes 1 and 2 to entry.]
– 16 – IEC 60601-2-37:2024 © IEC 2024
201.3.227
DEPTH FOR MECHANICAL INDEX
z
MI
depth on the BEAM-AXIS from the EXTERNAL TRANSDUCER APERTURE to the plane of maximum
ATTENUATED PULSE-PRESSURE-SQUARED-INTEGRAL (ppsi )
α
Unit: m
[SOURCE: IEC 62359:2010 and IEC 62359:2010/AMD1:2017, 3.23]
201.3.228
THERMAL OFFSET
∆T
offset
difference between a) the temperature of the APPLIED PART of the TRANSDUCER ASSEMBLY at
steady-state in the measurement setting before transmitting begins and b) the steady-state
TRANSDUCER ASSEMBLY
temperature at the same location in the measurement setting when the
was not present
Note 1 to entry: The value of the THERMAL OFFSET can be positive, negative or zero.
201.3.229
ULTRASOUND
acoustic oscillation whose frequency is above the high-frequency limit of audible sound (about
20 kHz)
[SOURCE: IEC 60050-802:2011, 802-01-01]
Table 201.101 – List of symbols
Symbol Term Reference
A = –12 dB OUTPUT BEAM AREA IEC 62359
aprt
d = EQUIVALENT BEAM DIAMETER IEC 62359
eq
f = ACOUSTIC WORKING FREQUENCY IEC 62359
awf
I = ATTENUATED PULSE-AVERAGE INTENSITY
pa,α
pii = PULSE-INTENSITY INTEGRAL IEC 62359
pii = ATTENUATED PULSE-INTENSITY INTEGRAL IEC 62359
α
I = ATTENUATED SPATIAL-PEAK PULSE-AVERAGE INTENSITY IEC 62359
sppa,α
I = SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY IEC 62359
spta
I = ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY IEC 62359
spta,α
I (z) = ATTENUATED TEMPORAL-AVERAGE INTENSITY IEC 62359
ta,α
MI = MECHANICAL INDEX IEC 62359
n = NUMBER OF PULSES PER ULTRASONIC SCAN LINE IEC 61157
pps
P = OUTPUT POWER IEC 62359
P = BOUNDED-SQUARE OUTPUT POWER IEC 62359
1x1
P = ATTENUATED OUTPUT POWER IEC 62359
α
p = ATTENUATED PEAK-RAREFACTIONAL ACOUSTIC PRESSURE IEC 62359
r,α
p = PEAK-RAREFACTIONAL ACOUSTIC PRESSURE IEC 62359
r
prr = PULSE REPETITION RATE IEC 62359
srr = SCAN REPETITION RATE IEC 62127-1
Symbol Term Reference
TI = THERMAL INDEX IEC 62359
TIB = BONE THERMAL INDEX IEC 62359
TIC = CRANIAL-BONE THERMAL INDEX IEC 62359
TIS = SOFT-TISSUE THERMAL INDEX IEC 62359
t = PULSE DURATION IEC 62359
d
X,Y = –12 dB OUTPUT BEAM DIMENSIONS IEC 62359
z = DEPTH FOR BONE THERMAL INDEX IEC 62359
b
z = BREAK-POINT DEPTH IEC 62359
bp
z = DEPTH FOR PEAK PULSE-INTENSITY INTEGRAL IEC 62359
pii
z = DEPTH FOR MECHANICAL INDEX IEC 62359
MI
z = DEPTH FOR PEAK ATTENUATED PULSE INTENSITY INTEGRAL IEC 62359
pii,α
z = DEPTH FOR PEAK SUM OF PULSE INTENSITY INTEGRALS IEC 62359
sii
z = DEPTH FOR PEAK SUM OF ATTENUATED PULSE INTENSITY INTEGRALS IEC 62359
sii,α
z = DEPTH FOR TIS IEC 62359
s
∆T = THERMAL OFFSET
offset
= temperature rise caused by the TRANSDUCER ASSEMBLY transmitting
∆T
tx
201.4 General requirements
Clause 4 of IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 applies, except as follows:
201.4.1 Conditions for application to ME EQUIPMENT or ME SYSTEMS
Addition:
An ULTRASOUND ENDOSCOPE where the imaging means is limited to ultrasound shall be
considered an ULTRASOUND TRANSDUCER and shall meet the requirements of this document.
NOTE 1 Examples of such ULTRASOUND TRANSDUCERS include transvaginal, transesophageal (TEE), rectal,
laparoscopic and other similar intra-cavity probes.
An ULTRASOUND ENDOSCOPE having imaging means in addition to ultrasound shall also meet the
requirements of 201.11.6.5 of IEC 60601-2-18:2009.
NOTE 2 Examples of such additional imaging means include optical and CCD.
201.4.3 ESSENTIAL PERFORMANCE
Additional subclause:
201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
Table 201.102 lists the potential sources of unacceptable risk identified to characterize the
ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT and the subclauses in which the
requirements are found.
– 18 – IEC 60601-2-37:2024 © IEC 2024
Table 201.102 – Distributed essential performance requirements
Requirement Subclause
Free from noise on a waveform or artefacts or distortion in an image or error of a displayed
numerical value which cannot be attributed to a physiological effect and which can alter the 202.8.1
diagnosis.
Free from the display of incorrect numerical values associated with the diagnosis to be
202.8.1
a
performed .
201.12.4.2
a
Free from the display of incorrect safety-related indications.
202.8.1
201.10.101
Free from the production of unintended or excessive ultrasound output.
202.8.1
Free from the production of unintended or excessive TRANSDUCER ASSEMBLY surface
202.8.1
temperature.
Free from the production of unintended or uncontrolled motion of TRANSDUCER ASSEMBLIES
202.8.1
intended for intra-corporeal use.
a
"incorrect" in the sense that the displayed value differs from what is calculated (having been altered during
data transfer), or the calculation itself is not correct.
In some circumstances the need for the repetition of an ultrasound examination should be
evaluated as a HAZARDOUS SITUATION, for example, intra-corporeal investigation and stress
testing for cardiopathic PATIENTS.
201.5 General requirements for testing ME EQUIPMENT
Clause 5 of IEC 60601-1:2005, IEC
...
Frequently Asked Questions
IEC 60601-2-37:2024 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment". This standard covers: IEC 60601-2-37:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.7.2.13. This document does not cover ultrasonic therapeutic equipment. Equipment used for the imaging or diagnosis of body structures by ultrasound in conjunction with other medical procedures is covered. IEC 60601-2-37:2024 cancels and replaces the second edition published in 2007 and Amendment 1:2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) technical and editorial changes resulting from the amended general standard IEC 60601 1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and its collateral standards IEC 60601-1-xx, b) technical and editorial changes as a result of maintenance to normative references; c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics standards. In particular, Clause 201.11 about protection against excessive temperatures and other hazards has been fully revised.
IEC 60601-2-37:2024 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of ULTRASONIC DIAGNOSTIC EQUIPMENT as defined in 201.3.217, hereinafter referred to as ME EQUIPMENT. If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this document are not covered by specific requirements in this document except in 201.7.2.13. This document does not cover ultrasonic therapeutic equipment. Equipment used for the imaging or diagnosis of body structures by ultrasound in conjunction with other medical procedures is covered. IEC 60601-2-37:2024 cancels and replaces the second edition published in 2007 and Amendment 1:2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: a) technical and editorial changes resulting from the amended general standard IEC 60601 1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020 and its collateral standards IEC 60601-1-xx, b) technical and editorial changes as a result of maintenance to normative references; c) technical and editorial changes resulting from relevant developments in TC 87 Ultrasonics standards. In particular, Clause 201.11 about protection against excessive temperatures and other hazards has been fully revised.
IEC 60601-2-37:2024 is classified under the following ICS (International Classification for Standards) categories: 11.040.55 - Diagnostic equipment; 17.140.50 - Electroacoustics. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 60601-2-37:2024 has the following relationships with other standards: It is inter standard links to IEC 60601-2-37:2007, IEC 60601-2-37:2007/AMD1:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
You can purchase IEC 60601-2-37:2024 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of IEC standards.
IEC 60601-2-37:2024は、超音波医療診断およびモニタリング機器の基本的な安全性と重要な性能に関する特定の要件を定めたものであり、ME機器の安全性と性能に特化した重要な標準です。この文書は、超音波診断機器に関する技術的および編集的な変更が盛り込まれており、以前の版と比較してより明瞭な規定を提供しています。 この標準の大きな強みは、ME機器とMEシステムの安全性を確保するための具体的な指針が含まれている点です。特に、IEC 60601-1:2005の改訂版からの技術的変更を取り入れることで、現行の技術水準に沿った形での文書の整備がなされています。また、TC 87超音波標準の関連発展に基づく更新も行われており、これにより基準の透明性と一貫性が向上しています。 さらに、201.11条項では過剰温度やその他の危険に対する保護措置が全面的に見直されており、より安全な使用が求められる状況においても対応可能な構造となっています。このように、IEC 60601-2-37:2024は、超音波医療機器の使用に関する重要な規範であり、医療機器の設計、製造、運用におけるニーズに対応した適切な基準を提供します。 この標準は、超音波診断機器の安全性を確保し、医療現場における信頼性を高める役割を果たすため、今後も多くの関係者にとって重要なリファレンスとして機能することが期待されます。医療機器の品質向上を図りながら、患者の安全性を守るためには、この標準の遵守が不可欠です。
IEC 60601-2-37:2024 represents a comprehensive update to the existing standards pertaining to ultrasonic medical diagnostic and monitoring equipment. This standard specifically addresses the BASIC SAFETY and ESSENTIAL PERFORMANCE criteria for ULTRASONIC DIAGNOSTIC EQUIPMENT, as well as clarifying its scope with respect to ME EQUIPMENT and ME SYSTEMS. The delineation between these categories is crucial for manufacturers and healthcare providers, ensuring compliance with safety and performance standards inherent to modern medical devices. One of the strengths of IEC 60601-2-37:2024 lies in its emphasis on identifying HAZARDS associated with the physiological functions of the ME EQUIPMENT. Although the document explicitly states that such inherent hazards are not addressed by specific requirements, it nonetheless provides clear guidelines on the operational safety and performance expected from enumerated equipment. This clarity aids manufacturers in developing devices that adhere to specified safety protocols, effectively minimizing risk to patients and operators alike. The technical revisions made in this edition are particularly notable, as they reflect a thorough approach to incorporating recent advancements in the field of ultrasonics. The adjustments stemming from updates in the general standard IEC 60601 1, along with revisions of normative references, ensure that the standard remains relevant in an ever-evolving medical landscape. The attention to detail in Clause 201.11, especially the comprehensive revision concerning protection against excessive temperatures and other hazards, exemplifies the standard's commitment to addressing modern challenges in ultrasonic diagnostics. Moreover, the standard's exclusion of ultrasonic therapeutic equipment maintains its focused relevance, allowing stakeholders to concentrate specifically on diagnostic applications and their associated requirements. This facilitates a more tailored approach in meeting regulatory expectations and advancing technology within this vital aspect of medical imaging. In summary, IEC 60601-2-37:2024 stands out for its thorough and updated approach to the essential performance and safety of ultrasonic medical diagnostic equipment. Its rigorous framework not only aligns with current technological innovations but also positions itself as a critical reference for ensuring the safety and efficacy of medical devices in the field of ultrasonics.
La norme IEC 60601-2-37:2024 traite des exigences particulières relatives à la sécurité de base et à la performance essentielle des équipements médicaux ultrasonores destinés au diagnostic et à la surveillance. Elle s'applique spécifiquement à la sécurité de base et à la performance essentielle de l'équipement de diagnostic ultrasonore, tel que défini dans la section 201.3.217, désigné ici comme ÉQUIPEMENT MÉDICAL (ME). L’étendue de cette norme est particulièrement pertinente compte tenu de l’évolution des technologies ultrasonores et des exigences croissantes en matière de sécurité dans le domaine médical. En précisant que les risques inhérents à la fonction physiologique prévue de l'ÉQUIPEMENT MÉDICAL ou des SYSTÈMES MÉDICAUX ne sont pas couverts sauf mention explicite, la norme s'assure que les utilisateurs soient conscients des limites de sa portée. Cela élargit son importance pour les fabricants et les professionnels de la santé, qui doivent se conformer à des exigences spécifiques. Parmi les points forts de la norme IEC 60601-2-37:2024, on note des modifications techniques significatives par rapport à l'édition précédente. Cela inclut des changements techniques et éditoriaux suite à la mise à jour de la norme générale IEC 60601-1:2005 et de ses amendements, ainsi que des ajustements en fonction des références normatives pertinentes. Ces révisions permettent de garantir que les exigences de sécurité restent à la pointe de l’innovation. De plus, la révision complète de la sous-clause 201.11 sur la protection contre les températures excessives et d'autres dangers témoigne d'un engagement envers la sécurité des patients et des opérateurs. Cette norme ne couvre pas les équipements thérapeutiques ultrasonores, se concentrant plutôt sur ceux utilisés pour l’imagerie et le diagnostic, ce qui est crucial pour assurer une utilisation appropriée et sûre dans les établissements médicaux. En somme, la norme IEC 60601-2-37:2024 se révèle non seulement actualisée mais également essentielle pour le secteur médical. Elle établit un cadre robuste pour garantir la sécurité et la performance des équipements médicaux ultrasonores, renforçant ainsi la confiance des utilisateurs et des patients dans ces technologies critiques.
Die IEC 60601-2-37:2024 ist ein maßgebliches Dokument, das sich mit den grundlegenden Sicherheitsanforderungen und der wesentlichen Leistung von Ultraschall-Diagnose- und Überwachungsgeräten im medizinischen Bereich befasst. Der Anwendungsbereich dieser Norm konzentriert sich ausdrücklich auf die Grundsicherheit und essentielle Leistung von medizinischen elektrischen Geräten (ME EQUIPMENT) zur Ultraschalldiagnose, wie in 201.3.217 definiert. Durch die klare Abgrenzung, dass spezifische Anforderungen nur für ME EQUIPMENT oder ME SYSTEMS gelten, wird die Norm sowohl für Hersteller als auch für Anwender von ME EQUIPMENT besonders relevant. Ein bemerkenswerter Stärke der IEC 60601-2-37:2024 ist die Berücksichtigung von technischen und redaktionellen Änderungen, die aus der Überarbeitung der allgemeinen Norm IEC 60601-1 resultieren. Diese Anpassungen reflektieren die neuesten Entwicklungen in der Technik und der Sicherheitskontrolle, was zu einer erhöhten Sicherheit und Leistungsfähigkeit der Ultraschallgeräte führt. Besonders hervorzuheben ist die umfassende Überarbeitung von Abschnitt 201.11, der sich mit dem Schutz vor übermäßigen Temperaturen und anderen Gefahren befasst, was die Relevanz der Norm in der heutigen technologischen Landschaft unterstreicht. Zusätzlich wird die Norm durch die Beachtung relevanter Entwicklungen im Bereich der Ultraschall-Standards des Technischen Ausschusses 87 (TC 87) gestärkt. Somit bietet die IEC 60601-2-37:2024 ein aktualisiertes und umfassendes Regelwerk, das sicherstellt, dass Ultraschall-Diagnosegeräte nicht nur effektiv, sondern auch sicher im klinischen Umfeld eingesetzt werden können. Durch die Abdeckung von Anforderungen für Geräte, die zur Bildgebung oder Diagnose von Körperstrukturen durch Ultraschall in Verbindung mit anderen medizinischen Verfahren verwendet werden, bleibt die Norm anpassungsfähig und relevant für die breite Palette an medizinischen Anwendungen. Die Nichtabdeckung von therapeutischen Ultraschallgeräten ist ebenfalls sinnvoll, da sie den spezifischen Fokus der Norm wahrt. Insgesamt ist die IEC 60601-2-37:2024 eine bedeutende Errungenschaft im Bereich der medizinischen Normen, die sowohl die Sicherheit als auch die Leistung von Ultraschall-Diagnosetechnologien betont und somit entscheidend zur Qualitätssicherung im Gesundheitswesen beiträgt.
IEC 60601-2-37:2024는 초음파 의료 진단 장비의 기본 안전성과 필수 성능에 대한 특정 요구사항을 규명하는 표준으로, 의료 전기 장비의 안전성을 보장하는 데 있어 중요한 역할을 한다. 이 표준의 범위는 초음파 진단 장비(ME EQUIPMENT)와 관련된 안전 및 성능 요구 사항에 초점을 맞추며, 이를 통해 의료 현장에서의 안전성을 높이고 환자의 보호를 강화하는 데 기여한다. IEC 60601-2-37:2024의 강점은 주요 기술 변경 사항이 반영되어 있다는 점이다. 이전 표준과 비교하여, IEC 60601-1:2005 및 관련 개정판에서의 기술적 및 편집적 변화가 포함되어 있으며, 이는 표준의 현대화를 통해 보다 안전하고 효율적인 의료 기기 사용을 가능하게 한다. 특히, 201.11 조항의 개정은 과도한 온도 및 기타 위험 요소로부터의 보호를 강화하였으며, 이는 초음파 진단 장비의 안전성을 크게 향상시킨다. 또한, 이 문서는 초음파 치료 장비는 제외하고 있지만, 신체 구조의 영상화나 진단에 관련된 장비는 포괄적으로 다루고 있어 의료 환경에서 매우 유용하다. IEC 60601-2-37:2024는 2007년에 발표된 두 번째 판 및 2015년의 수정판을 대체하며, 표준의 기술적 개정은 최신 기술 동향을 반영하고 있어 그 필요성이 매우 적절하다. 결론적으로, IEC 60601-2-37:2024는 초음파 의료 진단 장비의 안전과 성능을 보장하는 데 있어 필수적인 표준으로, 의학적 안전 기준을 충족시키기 위한 평가 기준으로서의 중요성을 갖는다.
IEC 60601-2-37:2024 표준은 초음파 의료 진단 및 모니터링 장비의 기본 안전 및 필수 성능에 관한 특별 요구 사항을 제시합니다. 이 문서는 의료 전기 장비(ME EQUIPMENT)의 기본 안전성과 필수 성능을 보장하기 위해 설정된 기준으로, 특히 초음파 진단 장비에 중점을 두고 있습니다. 문서의 적용 범위는 ME EQUIPMENT와 ME 시스템 모두에 해당하며, 특정 조항이 ME EQUIPMENT 또는 ME 시스템에만 적용될 경우 이를 명시하고 있습니다. 이 표준은 이전의 2007년에 발행된 두 번째 판과 2015년 개정판을 대체하며, 기술적 개정을 포함하고 있습니다. 특히, IEC 60601-1:2005의 개정 표준 및 관련 물질에 따른 기술적 및 편집적 변화가 반영되어 있으며, 이는 최신 기술 발전을 수용하기 위한 노력입니다. 특히 조항 201.11에서 과도한 온도 및 기타 위험으로부터의 보호에 대한 요구 사항이 전면적으로 수정되었다는 점이 주목할 만합니다. IEC 60601-2-37:2024의 강점은 초음파 진단 장비의 안전성과 성능을 명확히 규정하여, 의료 현장에서 사용되는 장비들이 환자의 안전을 보장하고, 효율적인 진단 기능을 수행할 수 있도록 돕는 데 있습니다. 또한, 초음파 치료 장비는 이 문서의 적용 범위에 포함되지 않음을 명확히 하여, 초음파 진단 장비의 특수성을 강조합니다. 결론적으로, IEC 60601-2-37:2024 표준은 초음파 의료 장비의 안전성과 성능에 대한 필수 요구 사항을 제공하여 의료 분야에서의 정확한 진단 및 환자 안전을 보장하는 데 매우 중요합니다.
La norme IEC 60601-2-37:2024 traite des exigences particulières concernant la sécurité de base et la performance essentielle des équipements de diagnostic et de surveillance médicale ultrasonique. Son champ d'application est clairement défini, s'appliquant uniquement aux appareils de diagnostic ultrasonique, excluant les équipements thérapeutiques ultrasoniques. Cela en fait un document crucial pour les fabricants d'équipements médicaux qui souhaitent garantir la conformité de leur matériel aux exigences de sécurité spécifiques. Parmi les forces de cette norme, on note son alignement sur les révisions techniques récentes des normes générales IEC 60601-1, apportant des changements techniques et éditoriaux qui améliorent la clarté et la cohérence. La mise à jour de la clause 201.11, qui concerne la protection contre les températures excessives et d'autres dangers, souligne l'engagement de la norme à fournir des directives sûres et actuelles. De plus, les modifications issues des développements au sein des standards d'ultrason TC 87 enrichissent la pertinence de cette norme dans le contexte clinique moderne. En intégrant des exigences qui abordent à la fois les dangers inhérents à la fonction physiologique des équipements médicaux et les systèmes médicaux, la norme IEC 60601-2-37:2024 se positionne comme un document fondamental, assurant ainsi une sécurité accrue et une performance essentielle fiables dans les environnements médicaux. Cette norme revêt une grande pertinence, car elle prend en compte les évolutions technologiques et les défis contemporains rencontrés dans le domaine du diagnostic par ultrasound.
IEC 60601-2-37:2024 is a significant standard that delineates the particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. The scope of this standard is comprehensive, specifically focusing on the basic safety and essential performance criteria for ultrasonic diagnostic equipment as defined in the standard's terminology. This targeted application ensures that only relevant ultrasonic diagnostic and monitoring equipment is assessed under its provisions, thereby enhancing the clarity and precision of compliance efforts. One of the core strengths of IEC 60601-2-37:2024 is its rigorous approach to safety and performance. It effectively establishes the baseline criteria necessary to mitigate risks associated with ultrasonic medical equipment, which is vital for ensuring patient safety and efficacy in medical diagnostics. By explicitly detailing the applicability of clauses to either ME EQUIPMENT or ME SYSTEMS, the standard provides clear guidance, reducing ambiguity for manufacturers and healthcare providers alike. The updated edition introduces significant technical changes, reflecting ongoing developments in technology and industry practices. These updates are not only essential for alignment with the amended general standard IEC 60601-1 but also reinforce the document's relevance in the fast-evolving field of medical ultrasound. The inclusion of technical and editorial changes stemming from normative references ensures that the standard remains current and comprehensive, addressing the latest advancements and challenges in ultrasonic medical technology. Furthermore, the fully revised Clause 201.11, which focuses on protection against excessive temperatures and other hazards, demonstrates a proactive response to emerging safety concerns. This provision is particularly pertinent given the nature of ultrasonic equipment and the potential risks associated with prolonged use or improper functioning. By canceling and replacing the previous edition, IEC 60601-2-37:2024 signifies a commitment to fostering continuous improvement and enhancing the reliability and effectiveness of ultrasonic medical diagnostic equipment. The technical revisions and updates amplify the standard's value, making it a crucial resource for manufacturers aiming to achieve compliance while ensuring the safety and effectiveness of their products. In conclusion, the IEC 60601-2-37:2024 standard stands out for its comprehensive scope, robust safety criteria, and timely updates that reflect the latest industry developments, positioning it as an indispensable guide for ultrasonic medical diagnostic and monitoring equipment.
Die Norm IEC 60601-2-37:2024 behandelt die grundlegende Sicherheit und die wesentliche Leistung von Ultraschalldiagnosegeräten im medizinischen Bereich. Sie ist speziell auf die Anforderungen und Risiken ausgelegt, die mit der Anwendung von medizinischem elektrischen Geräten (ME EQUIPMENT) verbunden sind, und definiert präzise, welche Aspekte für die Sicherheit und die funktionelle Integrität dieser Geräte berücksichtigt werden müssen. Ein zentraler Stärke der Norm ist die klare Trennung zwischen Anforderungen an ME EQUIPMENT und ME SYSTEMS, was die Anwendung der Norm erleichtert und die Relevanz der Sicherheitsbestimmungen für verschiedene Nutzungskontexte betont. Ein wesentlicher Vorteil ist die Berücksichtigung technischer und redaktioneller Änderungen, die sich aus den aktualisierten allgemeinen Standards (IEC 60601-1), sowie aus den Entwicklungen im Bereich der Ultraschalltechnik ergeben haben. Insbesondere die vollständige Überarbeitung von Abschnitt 201.11 zeigt das Engagement der Norm, sich den Herausforderungen durch übermäßige Temperaturen und anderen Gefahren zu stellen. Dies erhöht nicht nur die Patientensicherheit, sondern auch die Zuverlässigkeit der Ultraschalldiagnosetechnologie. Die Norm ist besonders relevant für Hersteller und Anwender von Ultraschallgeräten, da sie Anforderungen an die Bildgebung und Diagnose von Körperstrukturen durch Ultraschall definiert, während therapeutische Geräte ausgeschlossen sind. Somit stellt die IEC 60601-2-37:2024 einen bedeutenden Schritt in der Standardisierung dar, indem sie sowohl technische Präzision als auch sich entwickelnde klinische Bedürfnisse in der Medizintechnik integriert. Die Möglichkeit der Norm zur Anpassung an zukünftige Entwicklungen und die klare Definition von Anforderungen macht sie zu einem unverzichtbaren Dokument für die Gewährleistung der grundlegenden Sicherheit und der wesentlichen Leistung von medizinischen Ultraschallgeräten.
IEC 60601-2-37:2024は、超音波医療診断機器の基本的な安全性および本質的性能に関する特定の要件を規定しており、医療機器(ME機器)における適用範囲が明確に定義されています。この標準は、ME機器とMEシステムに対する共通の基準を提供し、超音波診断機器の安全性を確保するための重要な文書です。 この標準の強みは、技術的および編集上の改定を行い、最新の国際基準に適応している点です。IEC 60601-1:2005やその改正版に基づく技術的変更を含むことで、より現代的で安全性に裏打ちされた内容に更新されています。特に、201.11項の改定により、過剰温度やその他の危険要因に対する保護に関連する情報が強化されています。 また、IEC 60601-2-37:2024は、超音波治療機器を除外し、超音波を使用した身体構造の画像診断または診断に関連する医療手続きに適用される機器を明示的にカバーしているため、医療現場での使用における適切なガイドラインを提供します。この文書は、医療機器が患者に与える影響を最小限に抑えるための危険性評価を行う上での不可欠な基準となるでしょう。 このように、IEC 60601-2-37:2024は、超音波診断機器のしい本質的性能を保証するための重要な役割を果たすと同時に、医療機器の安全性向上に貢献する信頼性の高い基準と言えます。標準の適用範囲における明確さと、技術的な精度がこの文書の重要な特徴であり、十分な信頼を持って医療現場で利用することができるでしょう。










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...