IEC 62366-1
Edition 1.0 2020-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
AMENDMENT 1
AMENDEMENT 1
Medical devices –
Part 1: Application of usability engineering to medical devices

Dispositifs médicaux –
Partie 1: Application de l'ingénierie de l'aptitude à l'utilisation aux dispositifs
médicaux
IEC 62366-1:2015-02/AMD1:2020-06(en-fr)

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.

IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform Electropedia - www.electropedia.org
The advanced search enables to find IEC publications by a The world's leading online dictionary on electrotechnology,
variety of criteria (reference number, text, technical containing more than 22 000 terminological entries in English
committee,…). It also gives information on projects, replaced and French, with equivalent terms in 16 additional languages.
and withdrawn publications. Also known as the International Electrotechnical Vocabulary

(IEV) online.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Glossary - std.iec.ch/glossary
details all new publications released. Available online and 67 000 electrotechnical terminology entries in English and
once a month by email. French extracted from the Terms and definitions clause of
IEC publications issued between 2002 and 2015. Some
IEC Customer Service Centre - webstore.iec.ch/csc entries have been collected from earlier publications of IEC
If you wish to give us your feedback on this publication or TC 37, 77, 86 and CISPR.

need further assistance, please contact the Customer Service

Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.

Recherche de publications IEC - Electropedia - www.electropedia.org
webstore.iec.ch/advsearchform Le premier dictionnaire d'électrotechnologie en ligne au
La recherche avancée permet de trouver des publications IEC monde, avec plus de 22 000 articles terminologiques en
en utilisant différents critères (numéro de référence, texte, anglais et en français, ainsi que les termes équivalents dans
comité d’études,…). Elle donne aussi des informations sur les 16 langues additionnelles. Egalement appelé Vocabulaire
projets et les publications remplacées ou retirées. Electrotechnique International (IEV) en ligne.

IEC Just Published - webstore.iec.ch/justpublished Glossaire IEC - std.iec.ch/glossary
Restez informé sur les nouvelles publications IEC. Just 67 000 entrées terminologiques électrotechniques, en anglais
Published détaille les nouvelles publications parues. et en français, extraites des articles Termes et définitions des
Disponible en ligne et une fois par mois par email. publications IEC parues entre 2002 et 2015. Plus certaines
entrées antérieures extraites des publications des CE 37, 77,
Service Clients - webstore.iec.ch/csc 86 et CISPR de l'IEC.

Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
IEC 62366-1
Edition 1.0 2020-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
A MENDMENT 1
AM ENDEMENT 1
Medical devices –
Part 1: Application of usability engineering to medical devices

Dispositifs médicaux –
Partie 1: Application de l'ingénierie de l'aptitude à l'utilisation aux dispositifs

médicaux
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.01 ISBN 978-2-8322-8384-4

– 2 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
FOREWORD
This amendment has been prepared by subcommittee 62A: Common aspects of electrical
equipment used in medical practice, of IEC technical committee 62: Electrical equipment in
medical practice, and ISO technical committee 210: Quality management and corresponding
general aspects for medical devices.
The text of this amendment is based on the following documents:
FDIS Report on voting
62A/1386/FDIS 62A/1397/RVD
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.
The committee has decided that the contents of this amendment and the base publication will
remain unchanged until the stability date indicated on the IEC web site under
"http://webstore.iec.ch" in the data related to the specific publication. At this date, the
publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
_____________
FOREWORD
In the fourth paragraph, replace “ISO 14971:2007” with “ISO 14971:2019”, format “medical
device user interfaces” in small caps and replace “manufactures” with “MANUFACTURERS” to
correct the spelling and the format.

INTRODUCTION to Amendment 1
The first edition of IEC 62366-1 was published in 2015. Since its publication, experts working
in the field have identified several inaccuracies that warrant correction. In total, 22 issues
were identified and presented to the National Committee members of IEC/SC 62A and to the
Member Bodies of ISO/TC 210. A majority of the members of both committees that stated a
position supported developing this amendment to address the identified issues while making
no fundamental changes to the USABILITY ENGINEERING PROCESS as originally conceived in
IEC 62366-1:2015.
To assist the USER to implement the USABILITY ENGINEERING PROCESS, the technical report
MANUFACTURERS in
IEC TR 62366-2 is available, which contains tutorial information to assist
complying with this document, as well as more generally to design MEDICAL DEVICES that goes
beyond SAFETY-related aspects of USER INTERFACES and offers more detailed descriptions of
USABILITY ENGINEERING methods that can be applied.

© IEC 2020
INTRODUCTION
Replace, in the second paragraph, “Figure A.4” with “Figure A.5”.
Replace, in the NOTE, “functionality” with “performance”.
Replace, in the last paragraph, “benefits” with “advantages”.
Replace the existing footnote 1 with the following:
IEC TR 62366-2:2016, Medical devices – Part 2: Guidance on the application of usability engineering to medical
devices.
1 * Scope
In the second sentence, replace “with CORRECT USE and USE ERRORS, i.e., NORMAL USE” with
“with NORMAL USE, i.e., CORRECT USE and USE ERROR”.
Replace NOTE 1 with the following:
NOTE 1 SAFETY is freedom from unacceptable RISK. Unacceptable RISK can arise from USE ERROR, which can lead
to exposure to HAZARDS including loss or degradation of clinical performance.
Replace the existing footnote 2 with the following:
IEC TR 62366-2:2016, Medical devices – Part 2: Guidance on the application of usability engineering to medical
devices.
2 Normative references
Replace “ISO 14971:2007” with “ISO 14971:2019”.
3 Terms and definitions
Replace, in the first paragraph, “ISO 14971:2007” with “ISO 14971:2019”.
3.1
ABNORMAL USE
*
Replace, in the existing definition and its example, “intentional” with “deliberate” in 3 places.
3.2
ACCOMPANYING DOCUMENTATION
Replace the existing definition, notes to entry and source with the following:
3.2
ACCOMPANYING DOCUMENTATION
materials accompanying a MEDICAL DEVICE and containing information for the USER or those
accountable for the installation, use, maintenance, decommissioning and disposal of the
MEDICAL DEVICE, particularly regarding safe use
Note 1 to entry: The ACCOMPANYING DOCUMENTATION can consist of the instructions for use, technical description,
installation manual, quick reference guide, etc.
Note 2 to entry: ACCOMPANYING DOCUMENTATION is not necessarily a written or printed document but could involve
auditory, visual, or tactile materials and multiple media types.

– 4 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
Note 3 to entry: MEDICAL DEVICES that can be used safely without instructions for use are exempted from having
instructions for use by some authorities with jurisdiction.
[SOURCE: ISO 14971:2019, 3.1, modified – Note 3 to entry has been added.]
3.20
USE ENVIRONMENT
Add, in Note 1 to entry, the following second sentence:
Social attributes such as team versus individual, chaotic versus calm, stress level and length of shift can also play
a role.
3.23
* USE SPECIFICATION
Replace, in Note 3 to entry, “ISO 14971:2007” with “ISO 14971:2019”.
3.25
USER GROUP
Replace the definition with the following:
subset of USERS who are differentiated from other USERS by factors that are likely to influence
their interactions with the MEDICAL DEVICE
NOTE 1 to entry: Attributes of USER GROUPS can include age, culture, expertise.
3.29
USER PROFILE
Replace the existing definition with the following:
summary of the mental, physical and demographic traits of a USER GROUP, as well as
characteristics, such as knowledge, skills and abilities, which can have a bearing on design
decisions
4.1.1 * USABILITY ENGINEERING PROCESS
Replace, in the third paragraph, “ISO 13485:2003” with “ISO 13485:2016”.
Replace, in NOTE 1, “ISO 13485:2003” with “ISO 13485:2016”.
Replace, in the fourth paragraph, “ISO 14971:2007” with “ISO 14971:2019” and “Figure A.4”
with “Figure A.5”.
Replace, in the fifth paragraph, “Figure A.4” with “Figure A.5” and “carried out” with “carried
out iteratively or”.
4.1.2 * RISK CONTROL as it relates to USER INTERFACE design
Replace the first paragraph and list items a) to c) with the following:
To reduce use-related RISK, the MANUFACTURER shall use one or more of the following options,
in the priority listed (as required by ISO 14971:2019, 7.1):
a) inherently safe design and manufacture;
b) protective measures in the MEDICAL DEVICE itself or in the manufacturing PROCESS; and
c) information for SAFETY and, where appropriate, training to USERS.

© IEC 2020
4.1.3 Information for SAFETY as it relates to USABILITY
Replace, in the second paragraph, “intentional” with “deliberate” in 2 places.
4.3 Tailoring of the USABILITY ENGINEERING effort
Delete the compliance check.
5.1 * Prepare USE SPECIFICATION
Replace the fifth dash with:
– * intended USE ENVIRONMENT; and
5.2 * Identify USER INTERFACE characteristics related to SAFETY and potential USE ERRORS
Replace, in the first paragraph, “ISO 14971:2007, 4.2” with “ISO 14971:2019, 5.3”.
Replace the last sentence of the first paragraph with the following:
This identification shall include consideration of the PRIMARY OPERATING FUNCTIONS if they are
MEDICAL DEVICE SAFETY standards.
provided in applicable product-specific
Replace, in NOTE 1, “ISO 14971:2007, C.2.29 to C.2.34” with “ISO/TR 24971:— , A.2.31 to
A.2.37”.
Insert the following footnote:
Under preparation. Stage at the time of circulation: ISO /TR APUB 24971:2020.
Replace the paragraph preceding the compliance check with the following:
The results of this identification of characteristics related to SAFETY and potential USE ERRORS
shall be stored in the USABILITY ENGINEERING FILE.
5.3 * Identify known or foreseeable HAZARDS and HAZARDOUS SITUATIONS
Replace the last sentence of the first paragraph with the following:
This identification shall be conducted as part of a RISK ANALYSIS performed according to
ISO 14971:2019, 5.4.
5.5 * Select the HAZARD-RELATED USE SCENARIOS for SUMMATIVE EVALUATION
Replace the text of the subclause with the following:
The MANUFACTURER shall select the HAZARD-RELATED USE SCENARIOS to be included in the
SUMMATIVE EVALUATION.
The MANUFACTURER shall select:
– all HAZARD-RELATED USE SCENARIOS;

– 6 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
– a subset of the HAZARD-RELATED USE SCENARIOS based on the SEVERITY of the potential
HARM that could be caused by USE ERROR (e.g. for which medical intervention would be
needed); or
NOTE 1 The SEVERITY of HARM is determined in ISO 14971:2019, 5.5.
– a subset of the HAZARD-RELATED USE SCENARIOS based on the SEVERITY of the potential
HARM and based on other circumstances specific to the MEDICAL DEVICE and the
MANUFACTURER.
NOTE 2 Examples of selection schemes are given in Annex A, 5.5, and IEC TR 62366-2.
A summary of any selection scheme, the rationale for its use and the results of applying it shall
be stored in the USABILITY ENGINEERING FILE.
Compliance is checked by inspection of the USABILITY ENGINEERING FILE.
5.7.1 General
Delete, in the first sentence, “SPECIFICATION”.
Replace the first dash under list item b), including Examples 1 and 2, with the following:
– document which USER GROUPS are intended to be included in the test;
EXAMPLE 1 In a FORMATIVE EVALUATION, clinical personnel who are employees of the MANUFACTURER are
used to represent a nurse-USER GROUP.
EXAMPLE 2 In a SUMMATIVE EVALUATION, a panel of practicing intensive care nurses is used to represent
a critical care nursing USER GROUP.
Multiple USER PROFILES may be combined into a USER GROUP for the purposes of a
USABILITY TEST;
Add, in the penultimate dash under list item b) “and” after the semicolon.
Replace, in NOTE 3, “scaling” with “tailoring”.
5.7.3 * SUMMATIVE EVALUATION planning
Replace list item e) with the following:
e) * for a USABILITY TEST,
– how the characteristics of the test participants are representative of the intended USER
PROFILES;
– justifying how the test participants are grouped into distinct USER GROUPS for the
purpose of determining the number of test participants;
– the test environment and conditions of use and a rationale for how they are adequately
representative of the intended USE ENVIRONMENT;
– the definition of CORRECT USE for each HAZARD-RELATED USE SCENARIO; and
– the method of collecting data during the USABILITY TEST for the subsequent analysis of
observed USE ERRORS and use difficulties.
Delete the existing NOTE 5.
5.8 * Perform USER INTERFACE design, implementation and FORMATIVE EVALUATION
Replace, in NOTE 1, “ISO 14971:2007, Subclause 6.6” with “ISO 14971:2019, 7.5”.

© IEC 2020
5.9 * Perform SUMMATIVE EVALUATION of the USABILITY of the USER INTERFACE
Replace the second paragraph with the following:
The MANUFACTURER shall analyse the data of the SUMMATIVE EVALUATION and shall identify all
USE ERRORS and use difficulties that occurred. If a USE ERROR or use difficulty can lead to a
HAZARDOUS SITUATION, the root cause of any such USE ERROR or use difficulty shall be
determined. The root causes should be determined based on methods including observations
of USER performance as well as subjective comments from the USER.
NOTE 1 A use difficulty where a USER almost commits a USE ERROR while performing a TASK, but recovers in time
to avoid making the USE ERROR is sometimes called a “close call”.
Renumber NOTES 1 to 5 as 2 to 6.
Replace item i), without modifying the note, with the following:
i) document why improvement is not necessary or not practicable;
Replace, in NOTE 3, “ISO 14971:2007, 6.2” with “ISO 14971:2019, 7.1 and
ISO/TR 24971:— , Annex C”.
Insert the following footnote:
Under preparation. Stage at the time of circulation: ISO/TR APUB 24971:2020.
Replace item iii) with the following:
iii) evaluate the RESIDUAL RISK according to ISO 14971:2019, 7.3.
Replace, in NOTE 4, “ISO 14971:2007, Subclause 6.6” with “ISO 14971:2019, 7.5”.
Replace, in NOTE 5, “ISO 14971:2007, Clause 7” with “ISO 14971:2019, Clause 8”.
Replace, in the compliance check, "ISO 14971:2007, 6.4” with “ISO 14971:2019, 7.3”.
A.2 Rationale for requirements in particular clauses and subclauses
Clause 1 – Scope
In the sixth paragraph, set benefits in small caps in 2 places.
Replace, in the last paragraph, “benefits” with “advantages”.
Definition 3.1 – ABNORMAL USE
Replace, in the third dash, “RISK benefit” with “BENEFIT-RISK”.
Replace, in Example 2, “RISK/benefit” with “BENEFIT and RISK”.
Replace, in the first sentence of the last paragraph, “inaction” with “lack of USER action”.
Definition 3.11 – PRIMARY OPERATING FUNCTION
Replace the entire paragraph with the following:
For the purposes of this document, a PRIMARY OPERATING FUNCTION is a function that is directly
related to the SAFETY of the MEDICAL DEVICE. PRIMARY OPERATING FUNCTIONS are identified in
some product-specific MEDICAL DEVICE SAFETY standards. These standards require those

– 8 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
identified PRIMARY OPERATING FUNCTIONS to be an input to the USABILITY ENGINEERING PROCESS.
This document does not require any further identification of PRIMARY OPERATING FUNCTIONS
because this document requires the identification, description and evaluation of HAZARD-
RELATED USE SCENARIOS. However, TASKS to be performed within a HAZARD-RELATED USE
SCENARIO might be related to a function that has been identified as a PRIMARY OPERATING
.
FUNCTION
Definition 3.17 – USABILITY ENGINEERING or HUMAN FACTORS ENGINEERING
Delete the second sentence of the first paragraph and combine the first and second
paragraphs.
USE ERROR
Definition 3.21 –
Replace the first and second sentences of the first paragraph with:
USE ERRORS often can be an indication of USER INTERFACE design flaws that affect the
interaction of a USER with a MEDICAL DEVICE.
Definition 3.22 – USE SCENARIO
Add the following sentence to the end of the second paragraph:
A TASK in a HAZARD-RELATED USE SCENARIO, in which a USE ERROR can lead to significant HARM,
can be thought of as a 'critical task' [48] [49].
Clause 5 – USABILITY ENGINEERING PROCESS
Replace the entire rationale for Clause 5 including Figure A.4 with the following:
The purpose of the USABILITY ENGINEERING PROCESS, as described in this document, is to
provide use-related SAFETY of the MEDICAL DEVICE for the PATIENT, USER and others. To
USABILITY ENGINEERING PROCESS mitigates RISK caused by USER
achieve this purpose, the
interaction problems associated with NORMAL USE, such as USE ERROR.
Figure A.4 illustrates the types of use as described in this document and their relationship to
the concept of “reasonably foreseeable misuse” in ISO 14971.

© IEC 2020
Figure A.4 – Types of use as described in this document and their relationship
to the concept of “reasonably foreseeable misuse” in ISO 14971
OBJECTIVE EVIDENCE to support the determination that use-related RESIDUAL RISK has been
reduced to acceptable levels is generated by conducting SUMMATIVE EVALUATION of the
USABILITY of the USER INTERFACE.
To establish criteria for the acceptability of RESIDUAL RISKS related to USABILITY, the
MANUFACTURER considers relevant available data (e.g., the state of technology, experience
with similar MEDICAL DEVICES, POST-PRODUCTION surveillance reports). The MANUFACTURER can
RESIDUAL RISK
apply these criteria according to ISO 14971, which additionally considers
relative to the BENEFIT of the MEDICAL DEVICE.
A comprehensive RISK MANAGEMENT PROCESS, such as that defined in ISO 14971, requires that
a MANUFACTURER establish, implement, document and maintain a PROCESS for identifying
HAZARDS and HAZARDOUS SITUATIONS associated with a MEDICAL DEVICE, estimating and
evaluating the associated RISKS, controlling those RISKS, and monitoring how effective those
LIFE CYCLE. Such a PROCESS includes the following elements:
controls are throughout the
– RISK ANALYSIS;
– RISK EVALUATION;
– RISK CONTROL; and
– production and POST-PRODUCTION activities.
When applying a comprehensive RISK MANAGEMENT PROCESS to the USER INTERFACE, estimating
the RISK associated with each USE ERROR can be problematic, particularly because no
validated techniques are known to exist to predict, in advance, the likelihood of a person
USE ERROR. However, this document provides a PROCESS that a MANUFACTURER
committing a
can use to analyse, specify, design and evaluate the USABILITY of a MEDICAL DEVICE.
Implementing this PROCESS permits the MANUFACTURER to address the unpredictability of a

– 10 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
USER and minimize USE ERROR. This PROCESS helps the MANUFACTURER accomplish these
objectives by:
a) discovering HAZARDS and HAZARDOUS SITUATIONS related to the USER INTERFACE;
b) designing and implementing measures to control the RISKS related to the USER INTERFACE;
and
c) evaluating that these RISK CONTROL measures are effective in reducing RISK.
Other advantages of the USABILITY ENGINEERING PROCESS can include improved customer
satisfaction, but these aspects are beyond the scope of this document.
Figure A.5 provides an overview of relationship and interactions between the RISK
in ISO 14971 and the USABILITY ENGINEERING PROCESS of this document.
MANAGEMENT PROCESS
RISK MANAGEMENT is a decision-making PROCESS for determining acceptable RISK whereas
USABILITY ENGINEERING is a design and development PROCESS for the USER INTERFACE to
reduce the possibility of USE ERRORS that could result in HARM.
When the MANUFACTURER is identifying the characteristics related to SAFETY of the MEDICAL
DEVICE in accordance with the requirements of ISO 14971:2019, 5.3, the USABILITY
PROCESS can provide the detail necessary (see 5.2) to accomplish this step for
ENGINEERING
the USER INTERFACE of the MEDICAL DEVICE.
Further, when the MANUFACTURER is compiling a list of known or foreseeable HAZARDS and
HAZARDOUS SITUATIONS associated with the MEDICAL DEVICE in accordance with the
requirements of ISO 14971:2019, 5.4, the USABILITY ENGINEERING PROCESS provides a list of
items that are required to be considered (see 5.3) in order to accomplish this step for the
USER INTERFACE of the MEDICAL DEVICE.
ISO 14971 requires that RISKS associated with each of the identified HAZARDOUS SITUATIONS be
estimated (ISO 14971:2019, 5.5), and evaluated (ISO 14971:2019, Clause 6). If a RISK is not
acceptable according to the MANUFACTURER'S RISK acceptability criteria, the MANUFACTURER is
required to identify RISK CONTROL measure(s) that are appropriate for reducing the RISK(S) to
an acceptable level (ISO 14971:2019, 7.1). The MANUFACTURER is then required to implement
RISK CONTROL measures and verify that they are effective in reducing the RISK to
the identified
an acceptable level (ISO 14971:2019, 7.2).

© IEC 2020
A, B, C, D, E and F represent information flow between the two PROCESSES. The heavy solid lines (B, C, D and F)
represent information flow required by this document. New problems identified should be interpreted to mean new
HAZARDS, HAZARDOUS SITUATIONS or HAZARD-RELATED USE SCENARIOS discovered or implemented RISK CONTROL is
ineffective.
a
New problems identified should be interpreted to mean new USE ERRORS, HAZARDS, HAZARDOUS SITUATIONS, or
HAZARD-RELATED USE SCENARIOS have been identified.

– 12 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
Key
A USE SPECIFICATION is an input to ISO 14971:2019, 5.2
B Identified USER INTERFACE characteristics related to SAFETY (see 5.2)
C Identified foreseeable HAZARD and HAZARDOUS SITUATIONS (see 5.3).
D Identified sequences of events leading to HAZARDOUS SITUATIONS from ISO 14971:2019, 5.4 are an input to
determining HAZARD-RELATED USE SCENARIOS (see 5.4).
E The SEVERITY of HARM as determined in ISO 14971:2019, 5.5 is an input to identifying and describing HAZARD-
RELATED USE SCENARIOS.
F The SUMMATIVE EVALUATION (see 5.9) produces OBJECTIVE EVIDENCE and data for determining and evaluating
RESIDUAL RISK related to use in ISO 14971:2019, 7.3.
Figure A.5 – The relationship between the RISK MANAGEMENT PROCESS (ISO 14971:2019)
and the USABILITY ENGINEERING PROCESS (IEC 62366-1)
The USABILITY ENGINEERING PROCESS requires that all known or foreseeable HAZARD-RELATED
USE SCENARIOS (see 5.4) are identified and described prior to selecting the HAZARD-RELATED
USE SCENARIOS (see 5.5) that are used in preparing the USER INTERFACE EVALUATION plan. In
RISK CONTROL options related to use are identified during the development of
this document,
the USER INTERFACE SPECIFICATION with testable requirements (see 5.6). The HAZARD-RELATED
USE SCENARIOS, the RISK CONTROL measures and the USER INTERFACE EVALUATION plan are
iteratively updated based on the results from FORMATIVE EVALUATIONS and from the other parts
of product realization PROCESS, as appropriate.
Both the FORMATIVE EVALUATION and the SUMMATIVE EVALUATION of the implemented USER
are planned in the USER INTERFACE EVALUATION plan (see 5.7). FORMATIVE
INTERFACE
EVALUATION is carried out during USER INTERFACE design and implementation (see 5.8) to
explore the USER INTERFACE, identify the need for improvement or to confirm adequacy of the
USER INTERFACE. For each selected HAZARD-RELATED USE SCENARIO, the implemented USER
INTERFACE is subject to SUMMATIVE EVALUATION (see 5.9) to produce OBJECTIVE EVIDENCE that
use-related RESIDUAL RISK has been reduced to acceptable levels. These steps achieve the
same objective as 5.5 through 7.2 of ISO 14971:2019.
Subclause 5.1 – Prepare USE SPECIFICATION
Replace, in the first sentence, “standard” with “document”.
Replace, in the USE ENVIRONMENT subheading, “USE ENVIRONMENT” with “intended USE
ENVIRONMENT”.
Subclause 5.2 – Identify USER INTERFACE characteristics related to SAFETY and potential
USE ERRORS
Replace, in the first paragraph, “ISO 14971:2007, C.2.1” with “ISO/TR 24971:— , A.2.1”.
Replace, in list item a), “ISO 14971:2007, C.2.9” with “ISO/TR 24971:—, A.2.9”.
Replace, in list item b), “ISO 14971:2007, C.2.12” with “ISO/TR 24971:—, A.2.12”.
Replace, in list item c), “ISO 14971:2007, C.2.26” with “ISO/TR 24971:—, A.2.28” and
“ISO 14971:2007, C.2.27” with “ISO/TR 24971:2019, A.2.29”.
Replace, in list item d), replace “ISO 14971:2007, C.2.29” with “ISO/TR 24971:—, A.2.31”.
Insert the following footnote:
Under preparation. Stage at the time of circulation: ISO/TR APUB 24971:2020.

© IEC 2020
Subclause 5.3 – Identify known or foreseeable HAZARDS and HAZARDOUS SITUATIONS
Replace, in the third sentence, “Figure E.1 from ISO 14971:2007” with “Figure C.1 from
ISO 14971:2019”.
Subclause 5.6 – Establish USER INTERFACE SPECIFICATION
Replace the entire text with the following:
The detailed and testable design requirements for the USER INTERFACE contained in the USER
INTERFACE SPECIFICATION are generated based on the information collected in the preceding
PROCESS steps. This information includes the USE SPECIFICATION as well as the identified USE
ERRORS and the HAZARD-RELATED USE SCENARIOS. While the USER INTERFACE is evaluated, the
USER INTERFACE SPECIFICATION is updated, as needed.
Subclause 5.7.2 – FORMATIVE EVALUATION planning
Replace the entire paragraph with the following:
The purpose of FORMATIVE EVALUATION of the USER INTERFACE, which could include USABILITY
TESTS, is to assess portions of or the entire USER INTERFACE to ensure it achieves a specified
quality level and to increase the likelihood that the final SUMMATIVE EVALUATION of the
USABILITY of the USER INTERFACE can be conducted successfully. The results of each
FORMATIVE EVALUATION can be used to guide iterations of the design of the USER INTERFACE.
USER INTERFACE design is based on the quality level being
The decision to stop iterating the
measured during the later stages of FORMATIVE EVALUATIONS. No further iterations are required
when the quality level has been achieved that gives the MANUFACTURER the confidence that
the SUMMATIVE EVALUATION conducted at the end of the iterative design cycle can generate
sufficient OBJECTIVE EVIDENCE that use-related RESIDUAL RISK is acceptable.
Subclause 5.7.3 – SUMMATIVE EVALUATION planning
d) (availability of the ACCOMPANYING DOCUMENTATION and provision of training)
Replace the existing note with the following new note:
NOTE ‘Effectiveness as a RISK CONTROL measure’ relates to ISO 14971:2019, 7.2, and not to the defined term,
EFFECTIVENESS.
e) (USABILITY TEST)
Replace the first paragraph with the following paragraph:
This document requires assessment of whether USE ERRORS occurred, the USERS had use
difficulties or the USERS successfully completed the TASKS associated with the HAZARD-
RELATED USE SCENARIOS (i.e. CORRECT USE) that the MANUFACTURER selected for inclusion in
the SUMMATIVE EVALUATION. A use difficulty is a difficulty, or struggle, encountered during use,
which is typically momentary and overcome by the USER. When the use difficulty progresses,
such as, progresses to failure to complete an action or mismatch in mental models, the use
difficulty has progressed to a USE ERROR. A use difficulty where a USER almost commits a USE
ERROR while performing a TASK, but recovers in time to avoid making the USE ERROR is
sometimes called a “close call”.
Subclause 5.9 – Perform SUMMATIVE EVALUATION of the USABILITY of the USER INTERFACE
In the fourth paragraph, replace “USE ERRORS” with “USE ERRORS or use difficulties” and add
the following as the last sentence of the paragraph:
See rationale to 5.7.3 e) for additional information regarding use difficulties.

– 14 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
Table B.1 – Glossary of relevant RISK MANAGEMENT terms
Replace, in the heading, “ISO 14971:2007” with “ISO 14971:2019”.
Delete, in the meaning of HARM, “physical”.
Replace the meaning of HAZARDOUS SITUATION with the following:
circumstance in which people, property, or the environment is/are exposed to one or more
HAZARDS
Table B.2 – Examples of HARM due to RISK caused by USE ERROR(S) or poor USABILITY
Correct the Table title to “Examples of HARM caused by USE ERROR(S) or poor USABILITY”.
Replace, in the heading of the first column of this table, "HAZARD" with “HAZARD or HAZARDOUS
" (one correction on each page of the table).
SITUATION
C.1 General
Replace, in the first paragraph, “scaled” with “tailored”.
Replace, in the third paragraph, “IEC 62366-1:—' with “IEC 62366-1:2015”.
Replace, in Example 1, “IEC 62366-1:—“ with “IEC 62366-1:2015”.
Replace, in Example 2, “IEC 62366-1:—“ with “IEC 62366-1:2015”.
Replace, in Example 3, “IEC 62366-1:—“ with “IEC 62366-1:2015”.
Replace, in Example 4, “IEC 62366-1:—“ with “IEC 62366-1:2015”.
C.2.5 RESIDUAL RISK evaluation
Replace, in the first paragraph, “the overall RESIDUAL RISK according to ISO 14971:2007, 6.4”
with “the RESIDUAL RISK according to ISO 14971:2019, 7.3”.
Annex D – Types of MEDICAL DEVICE use, with examples
Replace, in the first paragraph, “inactions” with “lack of USER actions”. (2 occurrences)
Replace, in the third paragraph, the second and third sentences with the following:
Alternately, NORMAL USE could involve a USE ERROR or the use could involve deliberate
conduct that is beyond any additional means of USER INTERFACE-related RISK CONTROL by the
MANUFACTURER, (i.e., ABNORMAL USE). This does not necessarily mean that ABNORMAL USE
results in a poor outcome for the PATIENT.
Figure D.1 – Interrelationships between the different types of MEDICAL DEVICE use, with
examples
Replace Figure D.1 with the following:

© IEC 2020
Figure D.1 – Interrelationships between the different
types of MEDICAL DEVICE use, with examples

– 16 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
Annex E – Reference to the essential principles
Replace the entire annex with the following:
Annex E
(informative)
Reference to the essential principles
E.1 Non-IVD MEDICAL DEVICES
This document has been prepared to support the essential principles of safety and
performance of non-IVD MEDICAL DEVICES as MEDICAL DEVICES according to
ISO 16142-1:2016 [12]. This document is intended to be acceptable for conformity
assessment purposes.
Compliance with this document provides one means of demonstrating conformance with the
specific essential principles of ISO 16142-1:2016 [12]. Other means are possible. Table E.1
maps the clauses and subclauses of this document with the essential principles of
ISO 16142-1:2016.
Table E.1 – Correspondence between this document and the essential principles
Essential principle of Corresponding clause(s)/ Qualifying remarks
ISO 16142-1:2016, Annex B [12] sub-clause(s) of this document /notes
1 All The part relating to manufacturing is not
addressed.
a) All
b) All
6 All The part related to known or foreseeable
RISKS as they relate to use is addressed.
12.2 —
a) All The part relating to manufacturing is not
addressed.
b) All The part relating to manufacturing is not
addressed.
13.3 All
13.4 All The part related to understood by the
USERS is addressed.
17.4 All Addressed as it relates to safe use. The
part relating to manufacturing is not
addressed
19.1 All Addressed as it relates to safe use.
19.2 All
20.1 All The part relating to manufacturing is not
addressed.
20.2 All The part relating to manufacturing is not
addressed.
E.2 IVD MEDICAL DEVICES
This document has been prepared to support the essential principles of safety and
performance of IVD MEDICAL DEVICES as MEDICAL DEVICES according to ISO 16142-2:2017 [47].
This document is intended to be acceptable for conformity assessment purposes.

© IEC 2020
Compliance with this document provides one means of demonstrating conformance with the
specific essential principles of ISO 16142-2:2017 [47]. Other means are possible. Table E.2
maps the clauses and subclauses of this document with the essential principles of
ISO 16142-2:2017[47].
Table E.2 – Correspondence between this document and the essential principles
Essential principle of Corresponding clause(s)/ Qualifying remarks
ISO 16142-2:2017, Annex B [47] sub-clause(s) of this document /notes
1 All The part relating to manufacturing is
not addressed.
a) All
b) All
6 All The part related to known or
foreseeable RISKS as they relate to use
is addressed.
9.1 — The part relating to manufacturing is
not addressed
a) All
b) All
c) All
11.2 —
a) All The part relating to manufacturing is
not addressed.
The part relating to manufacturing is
b) All
not addressed.
c) All The part relating to manufacturing is
not addressed.
h) All The part relating to manufacturing is
not addressed.
11.4 All The part relating to manufacturing is
not addressed.
12.3 All The part related to understood by the
USERS is addressed.
16.6 All Addressed as it relates to safe use. The
part relating to manufacturing is not
addressed.
17.1 All Addressed as it relates to safe use. The
part relating to manufacturing is not
addressed.
17.2 All Addressed as it relates to safe use. The
part relating to manufacturing is not
addressed.
18.1 All Addressed as it relates to safe use.
a) All Addressed as it relates to safe use.
18.4 All The part relating to manufacturing is
not addressed.
b) All
– 18 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
Bibliography
Add to [1], IEC 60601-1:2005/AMD2:—
Add to [2], IEC 60601-1-6:2010/AMD2:—
Add to [3], IEC 60601-1-8:2006/AMD2:—
Replace [4] with the following:
[4] IEC 60601-1-11:2015, Medical electrical equipment – Part 1-11: General requirements
for basic safety and essential performance – Collateral standard: Requirements for
medical electrical equipment and medical electrical systems used in the home healthcare
environment
IEC 60601-1-11:2015/AMD1:—
Replace [6] with the following:
[6] ISO/IEC Guide 63:2019, Guide to the development and inclusion of aspects of safety
in International Standards for medical devices
Replace, in reference [8], “2005” with “2015”.
Replace “2008”, in reference [9], with “2015”.
Replace, in the existing reference [11], "2003" with "2016".
Replace the existing reference [12] with the following:
[12] ISO 16142-1:2016, Medical devices – Recognized essential principles of safety and
performance of medical devices – Part 1: General essential principles and additional
specific essential principles for all non-IVD medical devices and guidance on the
selection of standards
Replace the existing reference [13] with the following:
[13] ISO/TR 24971:— , Medical devices – Guidance on the application of ISO 14971
Replace footnotes 3, 4 and 5 with:
There exists a consolidated edition 3.1(2012) including IEC 60601-1:2005 and Amendment 1:2012.
Under preparation. Stage at the time of circulation of this document: IEC/AFDIS 60601-1:2005/AMD2:2019.
There exists a consolidated edition 3.1(2013) including IEC 60601-1-6:2010 and Amendment 1:2013.
Under preparation. Stage at the time of circulation of this document: IEC/AFDIS 60601-1-6:2010/AMD2:2019
There exists a consolidated edition 2.1(2012) including IEC 60601-1-8:2006 and Amendment 1:2012.
Under preparation. Stage at the time of circulation of this document: IEC/AFDIS 60601-1-8:2006/AMD2:2019
Under preparation. Stage at the time of circulation of this document: IEC/AFDIS 60601-1-11:2015/AMD1:2019
Under preparation. Stage at the time of circulation: ISO/TR APUB 24971:2020.
Add, after the last existing reference [46], the new references as follows:
[47] ISO 16142-2:2017, Medical devices – Recognized essential principles of safety and
performance of medical devices – Part 2: General essential principles and additional
specific essential principles for all IVD medical devices and guidance on the selection
of standards
© IEC 2020
[48] US FDA, (2016), Guidance for Industry and Food and Drug Administration Staff:
Applying Human Factors and Usability Engineering to Medical Devices, February 3,
[49] US FDA, (2016), Draft Guidance for Industry and FDA Staff: Combination Product
Design and Development, February 2016

Index of defined terms
Update the following terms:
HARM . ISO 14971:2019, 3.3
HAZARD . ISO 14971:2019, 3.4
HAZARDOUS SITUATION . ISO 14971:2019, 3.5
INTENDED USE . ISO 14971:2019, 3.6
LIFE CYCLE . ISO 14971:2019, 3.8
MANUFACTURER . ISO 14971:2019, 3.9
MEDICAL DEVICE . ISO 14971:2019, 3.10
OBJECTIVE EVIDENCE . ISO 14971:2019, 3.11
POST-PRODUCTION . ISO 14971:2019, 3.12
PROCEDURE . ISO 14971:2019, 3.13
PROCESS . ISO 14971:2019, 3.14
RECORD . ISO 14971:2019, 3.16
RESIDUAL RISK . ISO 14971:2019, 3.17
RISK . ISO 14971:2019, 3.18
RISK ANALYSIS . ISO 14971:2019, 3.19
RISK ASSESSMENT . ISO 14971:2019, 3.20
RISK CONTROL . ISO 14971:2019, 3.21
RISK EVALUATION . ISO 14971:2019, 3.23
RISK MANAGEMENT . ISO 14971:2019, 3.24
RISK MANAGEMENT FILE . ISO 14971:2019, 3.25
SAFETY . ISO 14971:2019, 3.26
SEVERITY . ISO 14971:2019, 3.27
Add the following term:
BENEFIT . ISO 14971:2019, 3.2

___________
– 20 – IEC 62366-1:2015/AMD1:2020
© IEC 2020
AVANT-PROPOS
Le présent amendement a été établi par le sous-comité 62A: Aspects généraux des
équipements électriques utilisés en pratique médicale, du comité d’études 62 de l’IEC:
Équipements électriques dans la pratique médicale, et du comité technique 210 de l'ISO:
Management de la qualité et aspects généraux correspondants des dispositifs médicaux.
Le texte de cet amendement est issu des documents suivants:
FDIS Rapport de vote
62A/1386/FDIS 62A/1397/RVD
Le rapport de vote indiqué dans le tableau ci-dessus
...