EN 60601-2-4:2011/A2:2026
(Amendment)Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators
- Status
- Published
- Publication Date
- 12-Nov-2026
- Technical Committee
- CLC/TC 62 - Electrical equipment in medical practice
- Current Stage
- 6060 - Document made available - Publishing
- Start Date
- 13-Nov-2026
- Due Date
- 01-Sep-2026
- Completion Date
- 13-Nov-2026
- Directive
- 93/42/EEC - Medical devices
EN 60601-2-4:2011/A2:2026 is a CENELEC amendment to the particular requirements for the basic safety and essential performance of cardiac defibrillators. It updates requirements for manual defibrillators and automated external defibrillators (AED), with a focus on charging time, battery capacity, alarm behavior, rhythm recognition, and electromagnetic compatibility (EMC). Manufacturers, test laboratories, and compliance teams use it to check whether a defibrillator design and its verification evidence match the amended requirements.
What does EN 60601-2-4:2011/A2:2026 specify?
EN 60601-2-4:2011/A2:2026 specifies targeted changes to the cardiac defibrillator particular standard rather than a full rewrite of the base document. The amendment updates selected clauses on safety, essential performance, labeling, charging performance, battery endurance, rhythm analysis, and EMC.
The document is organized as replacement or added text in clauses 201.1, 201.2, 201.3, 201.7.2.103, 201.8.5.3, 201.8.5.5.1, 201.12.3, 201.101, 201.102, 201.107, and 202, with supporting notes in Annex AA and Annex BB.
| Annex | What it covers |
|---|---|
| Annex AA | Explanatory notes for alarm systems, charging-time tests, and rhythm recognition detector performance |
| Annex BB | Updated clause mapping and defined-term index entries for amended requirements |
What are the key requirements of EN 60601-2-4:2011/A2:2026?
Alarm systems are driven by risk management
EN 60601-2-4:2011/A2:2026 adds a new clause on alarm systems. It says risk management determines when alarms are needed to indicate unsatisfactory patient states or equipment states, and IEC 60601-1-8 applies when alarms are required.
In practice, the defibrillator manufacturer must justify alarm choices for the intended use environment and the level of operator training. Any deviation from the alarm standard has to be validated, for example through usability evidence.
Charging time limits are tightened for manual defibrillators and AEDs
EN 60601-2-4:2011/A2:2026 sets ready-to-shock timing limits for frequent-use and infrequent-use manual defibrillators, and for frequent-use and infrequent-use AEDs. The timing is measured under defined worst-case conditions, including operation at 90 % of rated mains voltage and battery-depleted conditions.
For example, the standard measures the interval from discharge or disarm to ready-for-discharge, and for AEDs also from rhythm recognition detector (RRD) activation. This matters because the device must be ready quickly enough after a shockable rhythm is identified or after power-up.
Battery endurance is tested at 0 °C
EN 60601-2-4:2011/A2:2026 requires a new, fully charged battery to support at least 20 defibrillation discharges at 0 °C at the maximum delivered energy. Separate cycling patterns are given for manual defibrillators and AEDs.
For AEDs, the requirement is tied to the programmed defibrillation sequence and the maximum selectable energy setting if the operator can change it. This means battery claims need to be supported by test evidence under cold conditions and realistic discharge spacing.
Rhythm analysis with artifacts must be described and justified
EN 60601-2-4:2011/A2:2026 adds a requirement for manufacturers who specify an RRD with artifacts. They must state sensitivity and specificity, identify the artifact type, and explain how it is measured.
The overall RRD performance still has to meet the clause requirements, and risk management must identify hazards linked to rhythm detection in the presence of artifacts. In practice, this pushes manufacturers to document how the algorithm behaves during motion or CPR-like interference.
EMC requirements are expanded and made more specific
EN 60601-2-4:2011/A2:2026 adds EMC test-plan requirements and pass/fail criteria for all immunity test levels. The defibrillator must not deliver inappropriate therapy, change state unintentionally, lose data, or lose basic safety or essential performance.
The amendment also sets emissions and immunity expectations for all operating modes, including radiated RF fields, electrostatic discharge (ESD), surges, bursts, conducted disturbances, and proximity fields. It also treats trains and railroad stations as a special environment with harsh disturbance conditions, which matters for public-access AEDs used near rail systems.
What terms does EN 60601-2-4:2011/A2:2026 define?
- ALARM SYSTEM - A set of technical alarm functions used to warn the operator about unsatisfactory patient or equipment states.
- AUTOMATED EXTERNAL DEFIBRILLATOR (AED) - A defibrillator that analyzes rhythm and guides or controls shock delivery for external use.
- HOME HEALTHCARE ENVIRONMENT - The use environment addressed for equipment used outside a professional healthcare facility, including relevant EMC conditions.
- ELECTROMAGNETIC ENVIRONMENT (EM ENVIRONMENT) - The electromagnetic conditions around the equipment that can affect immunity and emissions performance.
- RISK MANAGEMENT - The process used to identify hazards and decide the control measures and acceptance criteria for the device.
- RHYTHM RECOGNITION DETECTOR (RRD) - The function that analyzes the patient rhythm and supports shock decision-making in an AED.
Who uses EN 60601-2-4:2011/A2:2026?
Manufacturers of cardiac defibrillators use EN 60601-2-4:2011/A2:2026 to design the device and prepare verification evidence. Test laboratories use it for charge-time tests, battery endurance checks, and EMC immunity testing.
Quality and regulatory teams use it to compare claims, labels, and test reports against the amended requirements. Hospitals, home healthcare organizations, and buyers use it when specifying equipment that must perform safely in rescue situations.
What changed in EN 60601-2-4:2011/A2:2026 from the previous edition?
EN 60601-2-4:2011/A2:2026 is Amendment 2 to IEC 60601-2-4:2010 and it aligns the defibrillator standard with IEC 60601-1:2005+A1:2012+A2:2020 and IEC 60601-1-2:2014+A1:2020.
The amendment updates or adds requirements for:
- scope text and normative references
- disposable defibrillator electrode labeling
- maximum mains voltage handling
- defibrillation protection
- alarm systems
- charging time for manual defibrillators and AEDs
- battery capacity and discharge endurance
- rhythm recognition detector performance with artifacts
- EMC emissions, immunity, and special railway-environment considerations
- Annex AA notes and Annex BB mappings
Which standards are used with EN 60601-2-4:2011/A2:2026?
- IEC 60601-1:2005+A1:2012+A2:2020 - The general standard for basic safety and essential performance.
- IEC 60601-1-2:2014+A1:2020 - The collateral EMC standard, including test methods and EMC performance requirements.
- IEC 60601-1-11:2015+A1:2020 - Requirements for medical electrical equipment used in the home healthcare environment.
- IEC 60601-1-8:2006+A1:2012+A2:2020 - Requirements for alarm systems.
- ISO 15223-1:2021 - Symbols for medical device labels and information supplied with the device.
- CISPR 11 and CISPR 16-2-3 - Emissions classification and related test guidance.
- IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-8 - EMC immunity test methods used in the amended requirements.
What does the EN 60601-2-4:2011/A2:2026 document contain?
The amendment contains replacement paragraphs, added clauses, and revised test instructions rather than a complete reissue of the cardiac defibrillator standard. It gives measurement-based methods for charging time, battery endurance, and EMC verification, together with pass/fail criteria for therapy delivery and recovery behavior.
It also includes explanatory notes in Annex AA on why the charging-time and rhythm-analysis requirements are written as they are. Annex BB updates clause references and defined-term entries tied to the amended sections.
For a design or verification team, the document is mainly a set of targeted performance and test requirements that must be read together with the base cardiac defibrillator standard and the referenced IEC and ISO standards.
Relations
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 17-Nov-2026
- Effective Date
- 17-Nov-2026
- Effective Date
- 17-Nov-2026
- Effective Date
- 07-Nov-2023
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN 60601-2-4:2011/A2:2026 is a amendment published by CLC. Its full title is "Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators". This standard covers: Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators
EN 60601-2-4:2011/A2:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
EN 60601-2-4:2011/A2:2026 has the following relationships with other standards: It is inter standard links to EN 61000-4-5:2014, EN 60601-1:2006, EN 61000-4-8:2010, EN 60601-1:2006/A2:2021, IEC 60601-1-8:2006, EN 60601-1:2006/A1:2013, EN 60601-1-2:2015/A1:2021, EN 60601-1-2:2015, EN 61000-4-4:2012, EN IEC 61000-4-3:2020, EN 60601-1-8:2007/A1:2013, EN 1789:2007+A2:2014, EN 1789:2007+A1:2010, EN 13718-1:2008, EN 60601-2-4:2011. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN 60601-2-4:2011/A2:2026 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/295. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
EN 60601-2-4:2011/A2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-januar-2027
Medicinska električna oprema - 2-4. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti srčnih defibrilatorjev - Dopolnilo A2 (IEC 60601-2-
4:2010/AMD2:2026)
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and
essential performance of cardiac defibrillators (IEC 60601-2-4:2010/AMD2:2026)
Medizinische elektrische Geräte - Teil 2-4: Besondere Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Defibrillatoren (IEC 60601-2-
4:2010/AMD2:2026)
Appareils électromédicaux - Partie 2-4: Exigences particulières pour la sécurité de base
et les performances essentielles des défibrillateurs cardiaques (IEC 60601-2-
4:2010/AMD2:2026)
Ta slovenski standard je istoveten z: EN 60601-2-4:2011/A2:2026
ICS:
11.040.10 Anestezijska, respiratorna in Anaesthetic, respiratory and
reanimacijska oprema reanimation equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN 60601-2-4:2011/A2
NORME EUROPÉENNE
EUROPÄISCHE NORM November 2026
ICS 11.040.10
English Version
Medical electrical equipment - Part 2-4: Particular requirements
for the basic safety and essential performance of cardiac
defibrillators
(IEC 60601-2-4:2010/AMD2:2026)
Appareils électromédicaux - Partie 2-4: Exigences Medizinische elektrische Geräte - Teil 2-4: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des défibrillateurs cardiaques wesentlichen Leistungsmerkmale von Defibrillatoren
(IEC 60601-2-4:2010/AMD2:2026) (IEC 60601-2-4:2010/AMD2:2026)
This amendment A2 modifies the European Standard EN 60601-2-4:2011; it was approved by CENELEC on 2026-11-02. CENELEC
members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this amendment the
status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Türkiye and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-2-4:2011/A2:2026 E
European foreword
The text of document 62D/2317/FDIS, future edition 3 of IEC 60601-2-4/AMD2, prepared by SC 62D
"Particular medical equipment, software, and systems" of IEC/TC 62 "Medical equipment, software,
and systems" was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2027-11-30
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2029-11-30
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
Endorsement notice
The text of the International Standard IEC 60601-2-4:2010/AMD2:2026 was approved by CENELEC
as a European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standard indicated:
IEC 60601-1-11:2015 NOTE Approved as EN 60601-1-11:2015 (not modified)
CISPR 16-2-3 NOTE Approved as EN 55016-2-3
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1 Where an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available
here: www.cencenelec.eu.
Replace IEC 60601-1-2 and its amendment with:
Publication Year Title EN/HD Year
IEC 60601-1-2 2014 Medical electrical equipment - Part 1-2: EN 60601-1-2 2015
General requirements for basic safety and
essential performance - Collateral
Standard: Electromagnetic disturbances -
Requirements and tests
+ A1 2020 + A1 2021
- - + prAA
Replace IEC 60601-1 and its amendment with:
Publication Year Title EN/HD Year
IEC 60601-1 2005 Medical electrical equipment - Part 1: EN 60601-1 2006
General requirements for basic safety and
essential performance
- - + corrigendum Mar. 2010
+ A1 2012 + A1 2013
- - + A12 2014
+ A2 2020 + A2 2021
- - + A13 2024
Replace ISO 15223-1 and its amendment with:
Publication Year Title EN/HD Year
ISO 15223-1 2021 Medical devices - Symbols to be used with EN ISO 15223-1 2021
information to be supplied by the
manufacturer - Part 1: General
requirements
Add:
Publication Year Title EN/HD Year
IEC 60601-1-8 2006 Medical electrical equipment – Part 1-8: - -
General requirements for basic safety and
essential performance - Collateral
Standard: General requirements, tests and
guidance for alarm systems in medical
electrical equipment and medical electrical
systems
+ A1 2012 + A1 2013
- - + AC 2014
+ A2 2020 - -
IEC/TR 61000-2-5 - Electromagnetic compatibility (EMC) - Part - -
2-5: Environment - Description and
classification of electromagnetic
environments
IEC 61000-4-3 - Electromagnetic compatibility (EMC) - Part EN IEC 61000-4-3 -
4-3 : Testing and measurement techniques
- Radiated, radio-frequency,
electromagnetic field immunity test
IEC 61000-4-4 - Electromagnetic compatibility (EMC) - Part EN 61000-4-4 -
4-4: Testing and measurement techniques
- Electrical fast transient/burst immunity
test
IEC 61000-4-5 2014 Electromagnetic compatibility (EMC) - Part EN 61000-4-5 2014
4-5: Testing and measurement techniques
- Surge immunity test
IEC 61000-4-8 - Electromagnetic compatibility (EMC) - Part EN 61000-4-8 -
4-8: Testing and measurement techniques
- Power frequency magnetic field immunity
test
IEC 60601-2-4 ®
Edition 3.0 2026-09
INTERNATIONAL
STANDARD
AMENDMENT 2
Medical electrical equipment -
Part 2-4: Particular requirements for the basic safety and essential performance
of cardiac defibrillators
ICS 11.040.10 ISBN 978-2-8327-1502-4
IEC 60601-2-4:2010-12/AMD2:2026-09(en)
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
Medical electrical equipment -
Part 2-4: Particular requirements for the basic safety and essential
performance of cardiac defibrillators
AMENDMENT 2
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
Amendment 2 to IEC 60601-2-4:2010 has been prepared by subcommittee 62D: Particular
medical equipment, software, and systems, of IEC technical committee 62: Medical equipment,
software, and systems.
The text of this Amendment is based on the following documents:
Draft Report on voting
62D/2317/FDIS 62D/2327/RVD
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
The language used for the development of this Amendment is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
– reconfirmed,
– withdrawn, or
– revised.
___________
INTRODUCTION to Amendment 2
Amendment 2 to the International standard IEC 60601-2-4 has been prepared by IEC
subcommittee 62D: Particular medical equipment, software, and systems, of IEC technical
committee 62: Medical equipment, software, and systems in medical practice. Amendment 2
changes is to align to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and
IEC 60601-1:2005/AMD2:2020 and IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020.
201.1 Scope, object and related standards
Replace the existing footnote 1 with the following new text:
The general standard is IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020,
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
201.2 Normative references
Replace the existing reference to IEC 60601-1-2, modified by Amendment 1, with the following
new reference:
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic
safety and essential performance - Collateral standard: Electromagnetic disturbances -
Requirements and tests
IEC 60601-1-2:2014/AMD1:2020
Replace the existing reference to IEC 60601-1, modified by Amendment 1, with the following
new reference:
IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety
and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1:2005/AMD2:2020
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
Replace the existing reference to ISO 15223-1, modified by Amendment 1, with the following
new reference:
ISO 15223-1:2021, Medical devices - Symbols to be used with information to be supplied by
the manufacturer - Part 1: General requirements
Add the following normative references:
IEC 60601-1-8:2006, Medical electrical equipment - Part 1-8: General requirements for basic
safety and essential performance - Collateral standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-8:2006/AMD1:2012
IEC 60601-1-8:2006/AMD2:2020
IEC TR 61000-2-5, Electromagnetic compatibility (EMC) - Part 2-5: Environment - Description
and classification of electromagnetic environments
IEC 61000-4-3, Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement
techniques - Radiated, radio-frequency, electromagnetic field immunity test
IEC 61000-4-4, Electromagnetic compatibility (EMC) - Part 4-4: Testing and measurement
techniques - Electrical fast transient/burst immunity test
IEC 61000-4-5:2014, Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement
techniques - Surge immunity test
IEC 61000-4-8, Electromagnetic compatibility (EMC) - Part 4-8: Testing and measurement
techniques - Power frequency magnetic field immunity test
201.3 Terms and definitions
Replace the existing text in the first paragraph with the following updated text:
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020, IEC 60601-1-2:2014, and
IEC 60601-1-2:2014/AMD1:2020 apply, except as follows:
201.7.2.103 Disposable defibrillator electrodes
Replace, in the existing item a), modified by Amendment 1, "ISO 15223-1:2007" with
"ISO 15223-1:2021".
201.8.5.3 * MAXIMUM MAINS VOLTAGE
Replace, in the existing last paragraph, modified by Amendment 1, "250 V" with "240 V".
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
201.8.5.5.1 * Defibrillation protection
Replace the existing second sentence that reads “Replacement of the third paragraph ("After
the operation of S, the peak.”) by the following:” with the following:
Replacement of the second-to-last paragraph (“After the operation of S, the peak….”) by the
following:
Add, at the end of the existing Subclause 201.12.2.102, the following new Subclause 201.12.3
and text:
201.12.3 * ALARM SYSTEMS
Replacement:
RISK MANAGEMENT is used to determine RISK CONTROL measures required for reducing the RISK(S)
due to unsatisfactory PATIENT physiological states or unsatisfactory functional states of the
MEDICAL ELECTRICAL EQUIPMENT or MEDICAL ELECTRICAL SYSTEM, to an acceptable level.
Requirements of IEC 60601-1-8:2006, IEC 60601-1-8:2006/AMD1:2012 and
IEC 60601-1-8:2006/AMD2:2020 apply when RISK MANAGEMENT has determined a requirement
to indicate to the OPERATOR of unsatisfactory physiological PATIENT states or unsatisfactory
functional states of the MEDICAL ELECTRICAL EQUIPMENT or MEDICAL ELECTRICAL SYSTEM, or to
warn the OPERATOR of HAZARDS to the PATIENT or OPERATOR due to the MEDICAL ELECTRICAL
EQUIPMENT or MEDICAL ELECTRICAL SYSTEM.
RISK CONTROL measures shall be appropriate for the environment of use and level of OPERATOR
training.
Deviations from the alarm standard shall be validated (e.g. usability).
201.101 * Charging time
201.101.1 Requirements for FREQUENT USE, MANUAL DEFIBRILLATORS
Replace the entire existing text of this subclause, including items a) and b), modified by
Amendment 1, with the following new text:
a) The time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 15 s under the following conditions:
1) when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the MANUAL DEFIBRILLATOR is operated with batteries depleted by whichever of the
following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which the
battery is specified as still useable by the MANUFACTURER.
b) The time from initially switching power on, or from within any OPERATOR programming mode,
to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not exceed
25 s under the following conditions:
1) when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the MANUAL DEFIBRILLATOR is operated with batteries depleted by whichever of the
following conditions comes last:
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which the
battery is specified as still useable by the MANUFACTURER.
Check compliance with a) and b) above by measurement. In the case of INTERNALLY POWERED
ME EQUIPMENT, the test is to start with a new and fully charged battery. In the case of such
ME EQUIPMENT also capable of charging the ENERGY STORAGE DEVICE when connected to the
SUPPLY MAINS or to a separate battery charger, compliance is also checked when the
ME EQUIPMENT is connected to the SUPPLY MAINS or to the battery charger. In cases with a
discharged or missing battery, verify performance is consistent with the markings provided as
required by 201.7.2.102.
201.101.2 Requirements for INFREQUENT USE, MANUAL DEFIBRILLATORS
Replace the entire existing text of this subclause, including items a) and b), with the following
new text:
a) For the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy, the
following charge time requirements apply:
1) the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall
not exceed 20 s when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS
VOLTAGE;
2) the time from the MANUAL DEFIBRILLATOR discharging or disarming its ENERGY STORAGE
DEVICE to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall
not exceed 25 s when the MANUAL DEFIBRILLATOR is operated with batteries depleted by
whichever of the following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
b) For the time from initially switching power on, or from within any OPERATOR programming
mode, to the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy, the
following charge time requirements apply:
1) the time from switching power on, or from within any OPERATOR programming mode, to
the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 30 s when the MANUAL DEFIBRILLATOR is operated on 90 % of the RATED MAINS
VOLTAGE;
2) the time from switching power on, or from within any OPERATOR programming mode, to
the MANUAL DEFIBRILLATOR being ready for discharge at maximum energy shall not
exceed 35 s when the MANUAL DEFIBRILLATOR is operated with batteries depleted by
whichever of the following conditions comes last:
i) the delivery of charge and discharge cycles until the MANUAL DEFIBRILLATOR indicates
when non-rechargeable batteries require replacement or rechargeable batteries
require recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
IEC 60601-2-4:2010/AMD2:2026 © IEC 2026
201.101.3 * Requirements for FREQUENT USE, AUTOMATED EXTERNAL DEFIBRILLATORS
Replace the existing text of this subclause with the following new text:
a) The time from activation of the RHYTHM RECOGNITION DETECTOR to the AUTOMATED EXTERNAL
DEFIBRILLATORS (AED) being ready for discharge at maximum energy shall not exceed 30 s
under the following conditions:
1) when the AED is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the AED is operated with batteries depleted by whichever of the following conditions
comes last:
i) the delivery of charge and discharge cycles until the AED indicates when non-
rechargeable batteries require replacement or rechargeable batteries require
recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
b) * The time from initially switching power on, or from within any OPERATOR programming
mode, to the AED being ready for discharge at maximum energy shall not exceed 40 s under
the following conditions:
1) when the AED is operated on 90 % of the RATED MAINS VOLTAGE;
2) when the AED is operated with batteries depleted by whichever of the following conditions
comes last:
i) the delivery of charge and discharge cycles until the AED indicates when non-
rechargeable batteries require replacement or rechargeable batteries require
recharging;
ii) the delivery of the maximum number of charge and discharge cycles after which is
specified as still useable by the MANUFACTURER.
Check compliance with a) and b) above by the following test:
A shockable simulated patient rhythm signal as defined by the manufacturer is applied to the
separate monitoring electrodes or the defibrillator electrodes. The visual or audible instructions
given by the defibrillator are followed. The charge time is measured from RRD activation [for
201.101.3 a)] or initial power on or exiting an operator programming mode [for 201.101.3 b)] to
the AED being ready for discharge at maximum energy.
In the case of internally powered ME EQUIPMENT, the test is to start with a new and fully charged
battery. In the case of such ME EQUIPMENT also capable of charging the energy storage device
when connected to the supply mains or to a separate battery charger, compliance is also
checked when the ME EQUIPMENT is connected to the SUPPLY MAINS or to the battery charger. In
cases with a discharged or missing battery, verify the performance is consistent with the
markings provided as required by 201.7.2.102.
201.101.4 * Requirements for INFREQUEN
...