EN IEC 60601-2-28:2019/A1:2026
(Amendment)Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
- Status
- Published
- Publication Date
- 22-Oct-2026
- Technical Committee
- CLC/TC 62 - Electrical equipment in medical practice
- Current Stage
- 6060 - Document made available - Publishing
- Start Date
- 23-Oct-2026
- Due Date
- 29-Oct-2027
- Completion Date
- 29-Oct-2027
EN IEC 60601-2-28:2019/A1:2026 is CLC’s amendment to the IEC 60601-2-28 standard for X-ray tube assemblies used in medical diagnosis. It updates the particular requirements for basic safety and essential performance, mainly for manufacturers, test organizations, and others who specify, mark, document, or verify this equipment. The amendment aligns the text with newer IEC references and clarifies several safety, radiation-protection, and documentation points.
What does EN IEC 60601-2-28:2019/A1:2026 specify?
EN IEC 60601-2-28:2019/A1:2026 specifies changes to the European adoption of Part 2-28 for X-ray tube assemblies for medical diagnosis. It is an amendment, so it revises selected subclauses of the existing part rather than replacing the whole document.
The amendment updates the links to the general IEC 60601 series, the collateral radiation-protection standard, the focal-spot and filtration references, and several product-information requirements. It also revises Annex AA, the bibliography, and the index of defined terms.
What are the key requirements of EN IEC 60601-2-28:2019/A1:2026?
Marking and identification
Clause 201.7.2.101 requires X-RAY TUBES to show the MANUFACTURER name or trademark, a MODEL OR TYPE REFERENCE, and individual identification. The markings must stay clearly legible over the expected service life, which matters because the tube has to be traceable back to its documents during service, repair, and acceptance checking.
Clause 201.7.3.7 says supply terminals must be marked adjacent to the terminals unless the design prevents interchange or interchange would not create an unacceptable RISK. In practice, this tells designers when terminal marking is needed and when the design itself provides the protection.
Clause 201.7.9.3.101 expands the technical description. It requires information on compatible supply equipment and HIGH-VOLTAGE GENERATOR(S), the required high voltage, ELECTRON emitter supply, beam-control supply functions, ANODE rotation supply where relevant, CATHODE emission characteristics, removable HIGH VOLTAGE connectors, auxiliary equipment, and any additional supply interfaces needed for safe operation and captured in the RISK MANAGEMENT FILE.
Safety, temperature, and usability-related approach
Clause 201.7.9.1 replaces the usability note with a risk-based approach. The amendment states that the USABILITY ENGINEERING PROCESS is not applicable for an X-RAY TUBE ASSEMBLY and points instead to the RISK MANAGEMENT PROCESS based on ISO 14971:2019 and ISO/TR 24971:2020. For users, that means usability concerns are handled through risk analysis and intended-use thinking rather than the separate usability process used elsewhere in IEC 60601.
Clause 201.11.1.1 sets a temperature limit for painted metal surfaces that may be touched unintentionally for up to 1 second during NORMAL USE. Clause 201.13.1.2 also uses a 1-second touch assumption when judging enclosure or temperature-related emission requirements. In practice, manufacturers need to classify touchable surfaces and document how they measured them.
Radiation protection and filtration
Clause 203.6.7.3 says the NOMINAL FOCAL SPOT VALUES of the X-RAY TUBE(S) FOCAL SPOT(S) in the equipment shall be stated according to IEC 60336. Clause 203.7.3 requires the X-RAY TUBE ASSEMBLY to be marked with PERMANENT FILTRATION according to IEC 60522-1:2020, or with the relevant material thicknesses and chemical symbols. That means the tube’s radiation output and filtration data must be stated in a way that supports compliance and comparison.
Clause 203.8.3 limits the DIAPHRAGM so the radiation aperture is no larger than needed for the largest X-RAY BEAM across the intended applications, and the material defining the aperture should be placed as close as practicable to the FOCAL SPOT. Compliance is checked by verifying the design data, so the design file matters as much as the physical build.
Clause 201.17 says the MANUFACTURER is not required to check conformity with IEC 60601-1-2 on the stand-alone X-RAY TUBE ASSEMBLY. This narrows the responsibility at the tube-assembly level and helps separate tube assessment from system-level EMC work.
What terms does EN IEC 60601-2-28:2019/A1:2026 define?
- X-RAY TUBE ASSEMBLY - the product covered by the part; the standard sets tube-specific safety and documentation requirements for medical diagnosis.
- HIGH-VOLTAGE GENERATOR - the source that provides the high voltage needed to operate the tube assembly and must be identified in the technical description.
- RISK MANAGEMENT FILE - the file that records the safety justification for additional interfaces or design choices needed for safe operation.
- QUALITY EQUIVALENT FILTRATION - the filtration concept linked to diagnostic X-ray quality, with methods referenced to IEC 60522-1:2020.
- RADIATION APERTURE - the opening that shapes the beam; its size and placement are controlled by Clause 203.8.3.
- X-RAY BEAM - the emitted beam whose size and application range determine the diaphragm limit.
- USABILITY - a defined term carried over from the general IEC 60601 framework, but the amendment routes usability considerations through risk management for this product.
Who uses EN IEC 60601-2-28:2019/A1:2026?
Manufacturers of X-RAY TUBE ASSEMBLIES use this standard to define markings, technical data, filtration information, and safety-related design choices. Test organizations use it when checking temperature, identification, radiation-protection, and documentation-related points.
Regulatory and quality staff use it to review conformity files, risk management records, and product descriptions. System integrators and medical imaging equipment designers use it when selecting compatible supply equipment, HIGH-VOLTAGE GENERATOR(S), and auxiliary equipment. Buyers and technical evaluators use it to compare tube assemblies by documented performance and safety information.
What changed in EN IEC 60601-2-28:2019/A1:2026 from the previous edition?
This amendment to IEC 60601-2-28:2017 updates references to IEC 60601-1-3, IEC 60336, and IEC 60522-1:2020, and aligns the text with IEC 60601-1:2005/AMD2:2020. It revises marking, technical-description, temperature, EMC, and radiation-protection wording, and it updates Annex AA terminology from filament-based wording to ELECTRON emitter wording. The foreword says it was prepared to provide a complete set of safety requirements for X-RAY TUBE ASSEMBLIES and to correct references to actual standards.
Which standards are used with EN IEC 60601-2-28:2019/A1:2026?
- IEC 60601-1-3:2008 / AMD1:2013 / AMD2:2021 - provides the collateral radiation-protection requirements used throughout the amendment.
- IEC 60336 - gives the basis for focal spot dimensions and related characteristics, including nominal focal spot values.
- IEC 60522-1:2020 - provides methods for quality equivalent filtration and permanent filtration.
- ISO 14971:2019 and ISO/TR 24971:2020 - support the risk-management approach used for usability-related considerations.
- IEC 60806:2022 - is the reference for stating the maximum symmetrical radiation field when that value is given.
- IEC 60601-1-2 and IEC 60601-1-6 - are named in the text to clarify stand-alone EMC and usability handling for X-RAY TUBE ASSEMBLIES.
What does the EN IEC 60601-2-28:2019/A1:2026 document contain?
The amendment contains replacement text for several clauses, not a complete standalone test method. Its content focuses on updated references, marking rules, technical-description items, temperature notes, and radiation-protection requirements.
It also revises Annex AA general considerations, especially terminology around ELECTRON emitter supply. The bibliography is updated with the standards and guidance documents used by the amended requirements.
The document also includes an updated index of defined terms for items such as ELECTRON, RADIATION APERTURE, RADIATION FIELD, X-RAY BEAM, and QUALITY EQUIVALENT FILTRATION.
Relations
- Effective Date
- 13-Oct-2026
- Effective Date
- 13-Oct-2026
- Effective Date
- 13-Oct-2026
- Effective Date
- 13-Oct-2026
- Effective Date
- 13-Oct-2026
- Effective Date
- 13-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 27-Oct-2026
- Effective Date
- 23-Sep-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN IEC 60601-2-28:2019/A1:2026 is a amendment published by CLC. Its full title is "Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis". This standard covers: Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
EN IEC 60601-2-28:2019/A1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.55 - Diagnostic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
EN IEC 60601-2-28:2019/A1:2026 has the following relationships with other standards: It is inter standard links to EN IEC 60522-1:2021, EN 60601-1-3:2008, EN 60601-1-3:2008/A2:2021, EN 60601-1-3:2008/A1:2013, EN IEC 60336:2021, EN IEC 60806:2023, EN 13718-1:2008, EN 1789:2007+A1:2010, EN 1789:2007+A2:2014, EN IEC 60601-2-28:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN IEC 60601-2-28:2019/A1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
INTRODUCTION This first amendment to the third edition of IEC 60601-2-28 has been prepared to provide a complete set of safety requirements for X-RAY TUBE ASSEMBLIES, based on the second amendment (2020) to IEC 60601-1:2005 and associated collateral standards. It corrects references to actual standards (e.g. IEC 60336, IEC 60522-1, and the IEC 60601 series) and clarifies single requirements within the standard.
Users of this document should note that when constructing the dated references to specific elements in a standard, such as definitions, amendments are only referenced if they modified the text being cited. For example, if a reference is made to a definition that has not been modified by an amendment, then the reference to the amendment is not included in the dated reference.
201.1 Scope, object and related standards Replace, in the existing footnote 1, "IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012", with "IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020". 201.1.3 Collateral standards Replace, in the existing second paragraph of the Addition, "IEC 60601-1-3:2008 and IEC 60601-1-3:2008/AMD1:2013 applies as modified in Clause 203.", with "IEC 60601-1-3:2008, IEC 60601-1-3:2008/AMD1:2013 and IEC 60601-1-3:2008/AMD2:2021 apply as modified in Clause 203.".
SIST EN IEC 60601-2-28:2020/A1:2026
201.1.4 Particular standards Replace, in the existing third paragraph of the Replacement, "IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012", with "IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020".
Replace, in the existing eighth paragraph of the Replacement, "3.147" with "3.154". 201.2 Normative references Replace the existing reference of the Replacement "IEC 60601-1-3:2008" with:
IEC 60601-1-3:2008, Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment IEC 60601-1-3:2008/AMD1:2013 IEC 60601-1-3:2008/AMD2:2021 Replace the existing reference of the Addition "IEC 60336" with:
IEC 60336, Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics Replace the existing reference of the Addition "IEC 60522" with:
IEC 60522-1, Medical electrical equipment - Diagnostic X-rays - Part 1: Determination of quality equivalent filtration and permanent filtration Add the following reference to the Addition:
IEC 60806, Determination of the maximum symmetrical radiation field of X-ray tube assemblies and X-ray source assemblies for medical diagnosis 201.3 Terms and definitions In the second paragraph, replace the existing reference to "IEC 60522" with "IEC 60522-1". 201.7.2.2 Identification Replace the third dash item by:
- individual identification (e.g. serial number); and 201.7.2.11 Mode of operation Delete the existing NOTE.
201.7.2.101 Marking of X-RAY TUBES Replace the existing subclause by:
X-RAY TUBES shall be provided with the following markings:
- name or trademark of the MANUFACTURER;
- MODEL OR TYPE REFERENCE; and
- individual identification (e.g. serial number).
SIST EN IEC 60601-2-28:2020/A1:2026
The markings shall enable individual products, series or types to be correlated with their ACCOMPANYING DOCUMENTS.
The markings on the X-RAY TUBE shall remain CLEARLY LEGIBLE during the EXPECTED SERVICE LIFE.
The above markings may be given in the form of a combined designation explained in the ACCOMPANYING DOCUMENTS.
Add, after the existing subclause 201.7.3.2, the following new subclause 201.7.3.7:
201.7.3.7 Supply terminals Replacement of the first paragraph by:
If terminals for supply conductors are non-interchangeable by design, or, if otherwise, there is no unacceptable RISK if the terminals are interchanged, then marking of the supply terminals or marking adjacent to the supply terminals is not required. Otherwise, supply terminals shall be marked adjacent to the terminals.
NOTE For this modification of the general standard the term supply terminal according to IEC 60050-845:2020, 845-28-064, has been considered.
201.7.9 ACCOMPANYING DOCUMENTS 201.7.9.1 General Replace the existing NOTE 104 with:
NOTE 104 Instead of the USABILITY ENGINEERING PROCESS (IEC 60601-1-6 is not applicable for an X-RAY TUBE ASSEMBLY), the RISK MANAGEMENT PROCESS (based on ISO 14971:2019 and its accompanying TR – ISO/TR 24971:2020) is the carrier of the considerations. ISO 14971:2019, 5.2 (INTENDED USE and reasonable foreseeable misuse) and ISO/TR 24971:2020 A.2.31 (USABILITY of the user interface) cover the USABILITY aspects of X-RAY TUBE ASSEMBLIES sufficiently.
201.7.9.3.101 Technical description of X-RAY TUBE ASSEMBLIES Replace, in the existing list item e), the reference to "IEC 60522" with "IEC 60522-1". Replace the existing list items h) to k) with:
h) type-designation of compatible electrical supply equipment required to power the X-RAY TUBE ASSEMBLY, including HIGH-VOLTAGE GENERATOR(S), or specifications concerning:
1) the HIGH VOLTAGE required from the HIGH-VOLTAGE GENERATOR, 2) electrical supply of the ELECTRON emitter(s), 3) electrical supply for electron beam control, e.g. focusing, pulsing, deflecting, etc. (as applicable) 4) electrical supply for rotating the ANODE (as applicable), 5) CATHODE EMISSION CHARACTERISTICS.
i) type-designation or specification of HIGH VOLTAGE connectors, if removable in NORMAL USE; j) type-designation of compatible auxiliary equipment (such as for cooling purposes), or requirements for such auxiliary functions; k) any supply interface requirement(s) in addition to items h) through j) above which are necessary for the safe operation of the X-RAY TUBE ASSEMBLY as documented by the MANUFACTURER in the RISK MANAGEMENT FILE; Delete the existing NOTE 102.
SIST EN IEC 60601-2-28:2020/A1:2026
...