FprEN IEC 61675-2:2026
(Main)Radionuclide imaging devices - Characteristics and test conditions - Part 2: Gamma cameras for planar, wholebody, and SPECT imaging
- Status
- Not Published
- Publication Date
- 17-Dec-2026
- Technical Committee
- CLC/TC 62 - Electrical equipment in medical practice
- Current Stage
- 5060 - Voting results sent to TC, SR - Formal Approval
- Start Date
- 18-Sep-2026
- Due Date
- 18-Sep-2026
- Completion Date
- 18-Sep-2026
FprEN IEC 61675-2:2026 is an IEC/CENELEC Committee Draft for Vote on radionuclide imaging devices, specifically gamma cameras for planar, wholebody, and SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT) imaging. It specifies terminology, test methods, and the accompanying documents used to describe camera performance on a common basis for manufacturers and test organizations.
What does FprEN IEC 61675-2:2026 specify?
FprEN IEC 61675-2:2026 specifies how to describe the characteristics of GAMMA CAMERAS equipped with PARALLEL HOLE COLLIMATORS that can perform planar imaging. It adds test methods for systems used for PLANAR WHOLEBODY IMAGING EQUIPMENT, SPECT, and wholebody SPECT, and it also allows SPECT systems to be equipped with a CT system for hybrid imaging.
The standard is organized into clause 1 Scope, clause 2 Normative references, clause 3 Terms and definitions, clause 4 Test methods, and clause 5 Accompanying documents, followed by a bibliography and an index of defined terms. Clause 4 is split into planar imaging, wholebody imaging, and tomographic imaging sections, so a reader can move straight to the measurement family they need.
The scope excludes novel camera designs with multiple DETECTOR HEADS that are not capable of planar acquisition. In practice, that means the document is aimed at conventional gamma camera architectures and at systems that are intended to support planar, wholebody, or SPECT performance claims.
What are the key requirements of FprEN IEC 61675-2:2026?
General test conditions
FprEN IEC 61675-2:2026 requires measurements to use the PULSE AMPLITUDE ANALYSER WINDOW settings specified for performance testing, and it expects the system to be adjusted in the same way as a normally installed unit. If a test cannot be performed exactly as stated, the reason and the actual conditions used must be reported. This matters because the results are meant to be comparable between systems, not tuned to favor one test setup.
The standard also says that, unless otherwise specified, each DETECTOR HEAD in the system is characterized with a full data set. Planar imaging results are to be reported with the same corrections used in clinical practice, and SPECT results are to use the corrections applied in routine clinical acquisitions. That keeps the measurements relevant to day-to-day use.
Planar imaging performance
Clause 4.2 starts with SYSTEM SENSITIVITY, measured with a calibrated activity assay in a cylindrical phantom. The result is reported together with the COLLIMATOR and the RADIONUCLIDE used. In practice, this gives buyers and users a way to compare how efficiently different camera and collimator combinations detect activity.
Clause 4.2.2 measures SPATIAL RESOLUTION using line sources and FWHM of the LINE SPREAD FUNCTION. The standard distinguishes intrinsic resolution, measured without the COLLIMATOR, from system resolution, measured with scatter material in place. That helps separate detector performance from the effect of the imaging setup.
Clauses 4.2.3 to 4.2.9 cover SPATIAL NON-LINEARITY, NON-UNIFORMITY OF RESPONSE, BAD PIXEL reporting, ENERGY RESOLUTION, MULTIPLE WINDOW SPATIAL REGISTRATION, COUNT RATE performance, and shield leakage. These checks show whether the image remains geometrically stable, whether the response is uniform across the field of view, and how the camera behaves at higher count rates.
Wholebody imaging performance
Clause 4.3 addresses wholebody imaging with tests for scanning constancy and SPATIAL RESOLUTION without scatter. The scan constancy test looks for variations in counts along the scan length, which is important when the detector or patient table is moving. The resolution test checks that motion does not distort the final wholebody image.
SPECT geometry, sensitivity, and image quality
Clause 4.4.1 checks PROJECTION geometry by testing CENTRE OF ROTATION (COR), DETECTOR HEAD TILT, and COLLIMATOR hole misalignment. The method uses point sources over many projection angles and fits sine functions to the measured centroids. This is the geometric foundation for accurate reconstruction.
Clause 4.4.2 measures DETECTOR POSITIONING TIME and NORMALIZED VOLUME SENSITIVITY. These tests separate useful acquisition time from idle time and relate SPECT sensitivity to the AXIAL FIELD OF VIEW. In practice, they help compare step-and-shoot performance and whole-volume efficiency.
Clauses 4.4.3 and 4.4.4 cover SCATTER FRACTION and SPECT SYSTEM SPATIAL RESOLUTION. The standard uses line sources, point sources, and a cylindrical phantom to report radial, tangential, and axial FWHM values. These measurements tell the reader how much scatter is present and how sharply the reconstructed slices can show detail.
Clauses 4.4.5 and 4.4.6 extend the method to Whole Body SPECT and to tomographic image quality and accuracy. They cover overlap-region testing, hot-sphere contrast recovery, background noise, attenuation and scatter correction accuracy, SPECT and CT image registration, and quantitative activity recovery. That is the part to review when the system is used for quantitative SPECT or SPECT/CT work.
Reporting and accompanying documents
Clause 5 requires an accompanying document for GAMMA CAMERAS, PLANAR WHOLEBODY IMAGING EQUIPMENT, and SPECT systems. It brings together the performance data that matter most in practice, so the manufacturer or user can see the tested collimators, sensitivity, resolution, count-rate behavior, uniformity, geometric checks, and quantitative image quality results in one place.
What terms does FprEN IEC 61675-2:2026 define?
- ADDRESS PILE UP - a false address calculation created when two or more events combine and still pass the ENERGY WINDOW.
- CENTRE OF ROTATION (COR) - the origin of the coordinate system used to describe the projections of a transverse slice.
- COUNT LOSS - the difference between measured COUNT RATE and TRUE COUNT RATE caused by the finite resolving time of the instrument.
- DETECTOR FIELD OF VIEW (FOV) - the detector region for which events are included in the display image and for which performance specifications are provided.
- ENERGY WINDOW - the accepted range of signal energies that the device passes for further processing.
- NORMALIZED VOLUME SENSITIVITY - VOLUME SENSITIVITY divided by the AXIAL FIELD OF VIEW of the tomograph or the phantom length, whichever is smaller.
- SPECT - SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY, meaning EMISSION COMPUTED TOMOGRAPHY using single photon detection of gamma-ray emitting RADIONUCLIDES.
Who uses FprEN IEC 61675-2:2026?
FprEN IEC 61675-2:2026 is intended mainly for manufacturers, because the introduction says the test methods are selected so manufacturers can describe system characteristics on a common basis. It is also used by test laboratories, service teams, and quality assurance teams that need to measure or verify camera performance.
In practice, the document supports engineering, acceptance testing, and performance reporting for GAMMA CAMERAS, PLANAR WHOLEBODY IMAGING EQUIPMENT, and SPECT systems. It is also useful to clinical physics and procurement teams that compare sensitivity, resolution, count-rate behavior, geometric accuracy, and quantitative imaging capability across different systems.
What changed in FprEN IEC 61675-2:2026 from the previous edition?
FprEN IEC 61675-2:2026 is the third edition and cancels and replaces the second edition published in 2015. The foreword says it combines IEC 60789:1992, IEC 61675-2:1998, IEC 61675-2:1998/AMD1:2004, and IEC 61675-3:1998.
The stated technical changes are:
- it was modified to address systems based on pixelated semi-conductor detectors;
- tests for wholebody SPECT performance were added;
- tests for quantitative SPECT image accuracy were added;
- the document was reformatted and updated;
- it was partly aligned with NEMA NU 1-2018;
- tests for SPECT systems operated in coincidence mode were removed.
Which standards are used with FprEN IEC 61675-2:2026?
- IEC TR 60788:2004 - provides the glossary of defined terms used throughout the document.
- IEC 61675-1:2022 - provides shared radionuclide imaging terminology and related definitions used for count rate, count loss, PSF, sinogram, and other common concepts.
- IEC 63073-1:2020 - is cited as a related radionuclide imaging device standard and supports some of the shared terminology used in clause 3.
What does the FprEN IEC 61675-2:2026 document contain?
The document contains detailed test setups for planar imaging, wholebody imaging, and SPECT performance, including phantom geometries, source arrangements, scan conditions, and analysis methods. It uses figures for cylindrical phantoms, slit phantoms, point-source geometry, scatter inserts, and wholebody SPECT overlap imaging, so the user can reproduce the measurement setup.
It also contains a table that links RADIONUCLIDES to the ENERGY WINDOW settings used for performance measurements. The methods cover activity assay, decay correction, background correction, centroid fitting, line spread function analysis, and reconstruction-based evaluation.
Clause 4.4.6 adds image quality and accuracy methods that compare hot spheres, background regions, a cold lung insert, and SPECT/CT registration. Clause 5 then turns those results into a structured set of accompanying documents for reporting camera characteristics and test outcomes.
Relations
- Effective Date
- 23-May-2023
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
FprEN IEC 61675-2:2026 is a draft published by CLC. Its full title is "Radionuclide imaging devices - Characteristics and test conditions - Part 2: Gamma cameras for planar, wholebody, and SPECT imaging". This standard covers: Radionuclide imaging devices - Characteristics and test conditions - Part 2: Gamma cameras for planar, wholebody, and SPECT imaging
Radionuclide imaging devices - Characteristics and test conditions - Part 2: Gamma cameras for planar, wholebody, and SPECT imaging
FprEN IEC 61675-2:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
FprEN IEC 61675-2:2026 has the following relationships with other standards: It is inter standard links to EN 61675-2:2015. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
FprEN IEC 61675-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 3 FOREWORD
- 4 INTRODUCTION
- 5 1 Scope
- 6 2 Normative references
- 7 3 Terms and definitions
- 8 4 Test methods
- 9 4.1 General
- 10 4.2 Planar imaging
- 11 4.2.1 SYSTEM SENSITIVITY
- 12 4.2.2 SPATIAL RESOLUTION
- 13 4.2.3 SPATIAL NON-LINEARITY
- 14 4.2.4 NON-UNIFORMITY OF RESPONSE
- 15 4.2.5 BAD PIXEL reporting
- 16 4.2.6 ENERGY RESOLUTION
- 17 4.2.7 Intrinsic MULTIPLE WINDOW SPATIAL REGISTRATION
- 18 4.2.8 COUNT RATE performance
- 19 4.2.9 Shield leakage test
- 20 4.3 Wholebody imaging
- 21 4.3.1 Scanning constancy
- 22 4.3.2 SPATIAL RESOLUTION without scatter
- 23 4.4 Tomographic imaging (SPECT)
- 24 4.4.1 Test of PROJECTION geometry
- 25 4.4.2 Measurement of SPECT SYSTEM SENSITIVITY
- 26 4.4.3 Scatter measurement
- 27 4.4.4 SPECT SYSTEM SPATIAL RESOLUTION
- 28 4.4.5 Whole Body SPECT
- 29 4.4.6 Tomographic image quality and accuracy
- 30 5 ACCOMPANYING DOCUMENTS
- 31 5.1 General
- 32 5.2 General parameters for GAMMA CAMERAS
- 33 5.2.1 COLLIMATORS
- 34 5.2.2 Shield leakage values
- 35 5.2.3 Pre-set PULSE AMPLITUDE ANALYSER WINDOWS
- 36 5.2.4 INTRINSIC ENERGY RESOLUTION
- 37 5.2.5 COLLIMATOR dependent quantities
- 38 5.2.6 COUNT RATE CHARACTERISTICS
- 39 5.2.7 Measured COUNT RATE that is 80 % of the corresponding TRUE COUNT 40 RATE
- 41 5.2.8 Dimensions of the DETECTOR FIELD OF VIEW
- 42 5.2.9 Non-uniformity characteristics
- 43 5.2.10 INTRINSIC SPATIAL RESOLUTION (FWHM) of the DETECTOR HEAD without 44 COLLIMATOR
- 45 5.2.11 INTRINSIC SPATIAL NON-LINEARITY
- 46 5.2.12 Intrinsic MULTIPLE WINDOW SPATIAL REGISTRATION
- oSIST prEN IEC 61675-2:2024 IEC CDV 61675-2 © IEC 2024 – 3 – 62C/912/CDV 47 5.3 GAMMA CAMERA based wholebody imaging system
- 48 5.3.1 Scanning constancy
- 49 5.3.2 SPATIAL RESOLUTION
- 50 5.4 SPECT
- 51 5.4.1 Calibration measurements of COR
- 52 5.4.2 Measurement of head tilt
- 53 5.4.3 Measurement of COLLIMATOR hole misalignment
- 54 5.4.4 TRANSVERSE RESOLUTION (radial and tangential)
- 55 5.4.5 AXIAL RESOLUTION
- 56 5.4.6 Axial PIXEL size
- 57 5.4.7 Transaxial PIXEL size
- 58 5.4.8 DETECTOR POSITIONING TIME
- 59 5.4.9 NORMALIZED VOLUME SENSITIVITY
- 60 5.4.10 SCATTER FRACTIONS SF and SF
- 61 5.4.11 Wholebody SPECT spatial resolution and performance
- 62 5.4.12 Image Quality Scan set up and phantom ACTIVITY concentration
- 63 5.4.13 Image quality
- 64 5.4.14 Accuracy of ATTENUATION correction and scatter correction
- 65 5.4.15 Accuracy of SPECT and CT image registration
- 66 5.4.16 Accuracy of quantitation
- 67 Bibliography
- 68 Index of defined terms
- 70 Figure 1 – Geometry of PROJECTIONS
- 71 Figure 2 – Cylindrical phantom
- 72 Figure 3 – Container
- 73 Figure 4 – Slit phantom
- 74 Figure 5 – Source arrangement for intrinsic measurements
- 75 Figure 6 – Calculation of FWHM
- 76 Figure 7 – Uniform source
- 77 Figure 8 – Small shielded liquid source
- 78 Figure 9 – Source positions for scanning constancy for wholebody imagingError! Bookmark not defined. 79 Figure 10 – Cylindrical phantom
- 80 Figure 11 – Phantom insert with holders for the scatter source
- 81 Figure 12 – Evaluation of scatter fraction
- 82 Figure 13 – Reporting transverse resolution
- 83 Figure 14 – Cross-section of body phantom
- 84 Figure 15 – Phantom insert with hollow spheres
- 85 Figure 16: Image quality phantom position for data acquisition with masses
- 86 Figure 17 – Placement of ROIs in the phantom background
- 87 Figure 18: Placement of quantitative ROIs in phantom background
- 89 Table 1 – RADIONUCLIDES and ENERGY WINDOWS to be used for performance 90 measurements
170 INTRODUCTION 171 The test methods specified in this part of IEC 61675 have been selected to reflect as much as 172 possible the clinical use of GAMMA CAMERAS for planar imaging, PLANAR WHOLEBODY IMAGING 173 EQUIPMENT, and SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT). It is intended that 174 the test methods are carried out by manufacturers thereby enabling them to describe the 175 characteristics of the systems on a common basis.
177 RADIONUCLIDE IMAGING DEVICES – 178 CHARACTERISTICS AND TEST CONDITIONS – 179 180 Part 2: Gamma cameras for planar, wholebody, and SPECT imaging 181 182 183 1 Scope 184 This part of IEC 61675 specifies terminology and test methods for describing the 185 characteristics of GAMMA CAMERAS equipped with PARALLEL HOLE COLLIMATORS that are capable 186 of planar imaging. Additional tests are specified for those GAMMA CAMERAS that are capable of 187 SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY (SPECT) or planar wholebody imaging 188 (PLANAR WHOLEBODY IMAGING EQUIPMENT) or SPECT wholebody imaging. SPECT systems may 189 also be equipped with a CT system for hybrid imaging.
190 These GAMMA CAMERAS consist of a gantry, single or multiple DETECTOR HEADS, and a 191 computer for data acquisition, processing, storage, and display. The DETECTOR HEADS may 192 contain single or multiple scintillation crystals or solid state detectors. 193 Novel camera designs with multiple DETECTOR HEADS that are not capable of planar acquisition 194 are not included in the scope of this standard.
195 2 Normative references 196 The following documents are referred to in the text in such a way that some or all of their 197 content constitutes requirements of this document. For dated references, only the edition 198 cited applies. For undated references, the latest edition of the referenced document (including 199 any amendments) applies.
200 IEC TR 60788:2004, Medical electrical equipment – Glossary of defined terms 201 IEC 61675-1:2022, Radionuclide imaging devices – Characteristics and test conditions – 202 Part 1: Positron emission tomographs 203 IEC 63073-1:2020, Dedicated radionuclide imaging devices – Characteristics and test 204 conditions – Part 1: Cardiac SPECT 205 3 Terms and definitions 206 For the purposes of this document the terms and definitions given in IEC TR 60788:2004, 207 IEC 63071-1:2020 and IEC 61675-1:2022 (some of which are repeated here for convenience), 208 and the following terms and definitions apply.
209 ISO and IEC maintain terminology databases for use in standardization at the following 210 addresses:
211 • IEC Electropedia: available at https://www.electropedia.org/ 212 • ISO Online browsing platform: available at https://www.iso.org/obp 213 3.1 214 ADDRESS PILE UP 215 <GAMMA CAMERA> false address calculation of an artificial event which passes the ENERGY 216 WINDOW, but is formed from two or more events by the PILE UP EFFECT 218 AXIAL FIELD OF VIEW 219 dimensions of a slice through the TOMOGRAPHIC VOLUME parallel to and including the SYSTEM 220 AXIS 221 Note 1 to entry: In practice it is specified only by its axial dimension given by the distance between the centres of 222 the outermost defined IMAGE PLANES plus the average of the measured AXIAL SLICE WIDTH measured as EQUIVALENT 223 WIDTH (EW).
224 3.3 225 AXIAL RESOLUTION 226 for tomographs with sufficiently fine axial sampling fulfilling the sampling theorem, SPATIAL 227 RESOLUTION along a line parallel to the SYSTEM AXIS 228 3.4 229 BAD PIXEL 230 detector pixel within the DETECTOR FIELD OF VIEW of the detector that has been physically or 231 electronically turned off such that gamma rays which interact in that pixel are not recorded by 232 the camera 233 [SOURCE: IEC 63073-1:2020, 3.2] 234 3.5 235 CENTRE OF ROTATION 236 COR 237 origin of that coordinate system, which describes the PROJECTIONS of a transverse slice with 238 respect to their orientation in space 239 Note 1 to entry: The CENTRE OF ROTATION of a transverse slice is given by the intersection of the SYSTEM AXIS with 240 the mid-plane of the corresponding OBJECT SLICE.
241 Note 2 to entry: The second note to entry concerns the French text only.
242 3.6 243 COLLIMATOR AXIS 244 straight line which passes through the geometrical centre of the exit field and entrance field of 245 the COLLIMATOR 246 3.7 247 COLLIMATOR FRONT FACE 248 surface of the COLLIMATOR which is closest to the object being imaged 249 3.8 250 COORDINATE SYSTEM OF PROJECTION 251 Cartesian system of the IMAGE MATRIX of each two-dimensional PROJECTION with axes X and p 252 Y p 253 Note 1 to entry: Axes X and Y are defined by the axes of the IMAGE MATRIX. p p 254 Note 2 to entry: The Y axis and the PROJECTION of the SYSTEM AXIS onto the COLLIMATOR FRONT FACE have to be p 255 in parallel.
256 Note 3 to entry: The origin of the COORDINATE SYSTEM OF PROJECTION may be the centre of the IMAGE MATRIX (see 257 Figure 1).
Y
Z
X
Y p
X p
X p
IEC 151/98 258 259 NOTE The FIXED COORDINATE SYSTEM X, Y, Z has its origin at the centre of the TOMOGRAPHIC VOLUME (shown as a 260 cylinder), the Z-axis being the SYSTEM AXIS. The COORDINATE SYSTEM OF PROJECTION X , Y is shown for a p p 261 PROJECTION ANGLE . For each , the one-dimensional PROJECTION of the marked OBJECT SLICE has the address 262 range shown (hatched). Within this range the CENTRE OF ROTATION is projected onto the address X (offset). p 263 Figure 1 – Geometry of PROJECTIONS 264 3.9 265 COUNT LOSS 266 difference between measured COUNT RATE and TRUE COUNT RATE, which is caused by the finite 267 RESOLVING TIME of the instrument 268 [SOURCE: IEC 61675-1:2022, 3.8.1] 269 3.10 270 COUNT RATE 271 number of counts per unit of time 272 [SOURCE: IEC 61675-1:2022, 3.8.2] 274 COUNT RATE CHARACTERISTIC 275 function giving the relationship between observed COUNT RATE and TRUE COUNT RATE 276 [SOURCE: IEC TR 60788:2004, rm-34-21] 277 3.12 278 DETECTOR FIELD OF VIEW 279 FOV 280 region of the detector within which events are included in the display image, and for which all 281 performance specifications are provided 282 Note 1 to entry: The note to entry regarding the abbreviation concerns the French text only. 283 3.13 284 DETECTOR HEAD TILT 285 deviation of the COLLIMATOR AXIS from orthogonality with the SYSTEM AXIS 286 3.14 287 DETECTOR POSITIONING TIME 288 fraction of the total time spent on an acquisition which is not used in collecting data 289 3.15 290 DETECTOR PIXEL 291 smallest discrete unit of a pixelated DETECTOR HEAD that is able to provide distinct energy, 292 spatial, and timing information about detected photons 293 3.16 294 EMISSION COMPUTED TOMOGRAPHY 295 ECT 296 imaging method for the representation of the spatial distribution of RADIONUCLIDES in selected 297 two-dimensional slices through the object 298 3.17 299 ENERGY WINDOW 300 range defining the energy of the signals accepted by the device for further processing 301 3.18 302 EQUIVALENT WIDTH 303 EW 304 width of that rectangle having the same area and the same height as the response function, 305 e.g., the POINT SPREAD FUNCTION 306 [SOURCE: IEC TR 60788:2004, rm-34-45] 307 3.19 308 FIXED COORDINATE SYSTEM 309 Cartesian system with axes X, Y, and Z 310 Note 1 to entry: Z being the SYSTEM AXIS.
311 Note 2 to entry: The origin of the FIXED COORDINATE SYSTEM is defined by the centre of the TOMOGRAPHIC VOLUME 312 (see Figure 1).
313 Note 3 to entry: The SYSTEM AXIS is orthogonal to all transverse slices.
314 3.20 315 IMAGE MATRIX 316 arrangement of MATRIX ELEMENTS in a preferentially Cartesian coordinate system 318 IMAGE PLANE 319 plane assigned to a plane in the OBJECT SLICE 320 Note 1 to entry: Usually the IMAGE PLANE is the mid-plane of the corresponding OBJECT SLICE. 321 3.22 322 INTRINSIC ENERGY RESOLUTION 323 FULL WIDTH AT HALF MAXIMUM of the full energy absorption peak in the INTRINSIC ENERGY 324 SPECTRUM for a specified RADIONUCLIDE 325 3.23 326 INTRINSIC ENERGY SPECTRUM 327 measured histogram of pulse heights for the DETECTOR HEAD without COLLIMATOR 328 Note 1 to entry: The pulse height should be expressed as corresponding energy. 329 3.24 330 INTRINSIC NON-UNIFORMITY OF RESPONSE 331 NON-UNIFORMITY OF RESPONSE of the DETECTOR HEAD without COLLIMATOR 332 3.25 333 INTRINSIC SPATIAL NON-LINEARITY 334 SPATIAL NON-LINEARITY of the DETECTOR HEAD without COLLIMATOR 335 3.26 336 INTRINSIC SPATIAL RESOLUTION 337 <GAMMA CAMERA> SPATIAL RESOLUTION in air for a specified RADIONUCLIDE measured without 338 the COLLIMATOR 339 3.27 340 LINE SOURCE 341 straight RADIOACTIVE SOURCE approximating a -function in two dimensions and being constant 342 (uniform) in the third dimension 343 3.28 344 LOW-ENERGY-TAIL RATIO 345 ratio of the counts measured in an energy window of width 2*EFWHM centred at energy Epeak - 346 2* E divided by the counts measured in an energy window of width 2*E centred at an FWHM FWHM 347 energy of E , where E is the peak energy of the RADIOISOTOPE being measured and peak peak 348 EFWHM is the energy resolution of the detector 349 [SOURCE: IEC 63073-1:2020, 3.12] 350 3.29 351 MATRIX ELEMENT 352 smallest unit of an IMAGE MATRIX, which is assigned in location and size to a certain volume 353 element of the object (VOXEL) 354 3.30 355 MULTIPLE WINDOW SPATIAL REGISTRATION 356 measured position of a source as a function of the ENERGY WINDOW setting 357 3.31 358 NORMALIZED VOLUME SENSITIVITY 359 VOLUME SENSITIVITY divided by the AXIAL FIELD OF VIEW of the tomograph or the phantom 360 length, whichever is the smaller 362 OBJECT SLICE 363 slice in the object 364 Note 1 to entry: The physical property of this slice that determines the measured information is displayed in the 365 tomographic image.
366 3.33 367 OFFSET 368 deviation of the position of the PROJECTION of the COR (X' ) from X = 0 (see Figure 1) p p 369 3.34 370 PARALLEL HOLE COLLIMATOR 371 COLLIMATOR with a number of apertures, the axes of which are parallel 372 3.35 373 PILE UP EFFECT 374 false measurement of the pulse amplitude, due to the absorption of two or more gamma rays, 375 reaching the same radiation detector within the RESOLVING TIME 376 3.36 377 PIXEL 378 MATRIX ELEMENT in a two-dimensional IMAGE MATRIX 379 3.37 380 PLANAR WHOLEBODY IMAGING EQUIPMENT 381 <GAMMA CAMERA> GAMMA CAMERA, with one or two DETECTOR HEAD(S), in which the image of an 382 extended object is formed by moving the DETECTOR HEAD(S) or the object in the axial direction 383 relative to each other 384 3.38 385 POINT SOURCE 386 RADIOACTIVE SOURCE approximating a -function in all three dimensions 387 3.39 388 POINT SPREAD FUNCTION 389 PSF 390 scintigraphic image of a POINT SOURCE 391 3.40 392 PROJECTION 393 transformation of a three-dimensional object into its two-dimensional image or of a two- 394 dimensional object into its one-dimensional image, by integrating the physical property which 395 determines the image along the direction of the PROJECTION BEAM 396 Note 1 to entry: This process is mathematically described by line integrals in the direction of PROJECTION and 397 called the Radon-transform.
398 3.41 399 PROJECTION ANGLE 400 angle at which the PROJECTION is measured or acquired 401 Note 1 to entry: See Figure 1.
402 3.42 403 PROJECTION BEAM 404 determines the smallest possible volume in which the physical property which determines the 405 image is integrated during the measurement process 406 Note 1 to entry: Its shape is limited by the SPATIAL RESOLUTION in all three dimensions and may vary with photon 407 energy and COLLIMATOR design.
408 Note 2 to entry: In SPECT the PROJECTION BEAM usually has the shape of a long thin diverging cone.
410 RADIAL RESOLUTION 411 TRANSVERSE RESOLUTION along a line passing through the position of the source and the 412 SYSTEM AXIS 413 [SOURCE: IEC 61675-1:2022, 3.4.1.1] 414 3.44 415 RADIOACTIVE SOURCE 416 quantity of radioactive material having both an ACTIVITY and a specific ACTIVITY above specific 417 levels 418 [SOURCE: IEC TR 60788:2004, rm-20-02] 419 3.45 420 RADIUS OF ROTATION 421 distance between the SYSTEM AXIS and the COLLIMATOR FRONT FACE 422 3.46 423 SCATTER FRACTION 424 SF 425 <GAMMA CAMERA> ratio between the number of scattered photons and the sum of scattered 426 plus unscattered photons for a given experimental set-up which includes a specific ENERGY 427 WINDOW size and position and a specific scattering medium 428 3.47 429 SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY 430 SPECT 431 EMISSION COMPUTED TOMOGRAPHY utilizing single photon detection of gamma-ray emitting 432 RADIONUCLIDES 433 Note 1 to entry: The note to entry regarding the abbreviation concerns the French version only. 434 3.48 435 SINOGRAM 436 two-dimensional display of all one-dimensional PROJECTIONS of an OBJECT SLICE, as a function 437 of the PROJECTION ANGLE 438 Note 1 to entry: The PROJECTION ANGLE is displayed on the ordinate, the linear PROJECTION coordinate is 439 displayed on the abscissa.
440 [SOURCE: IEC 61675-1:2022, 3.1.1.4] 441 3.49 442 SLICE SENSITIVITY 443 ratio of COUNT RATE as measured on the SINOGRAM to the ACTIVITY concentration in the 444 phantom 445 Note 1 to entry: In SPECT the measured counts are not numerically corrected for scatter by subtracting the 446 SCATTER FRACTION.
447 [SOURCE: IEC 61675-1:2022, 3.6, modified – The note to entry has been changed] 449 SPATIAL NON-LINEARITY 450 deviations of the image of a straight LINE SOURCE from a straight line 452 SPATIAL RESOLUTION 453 <nuclear medicine> ability to concentrate the count density distribution in the image of a POINT 454 SOURCE to a point 455 [SOURCE: IEC 61675-1:2022, 3.4] 456 3.52 457 SYSTEM AXIS 458 axis of symmetry characterized by geometrical and physical properties of the arrangement of 459 the system 460 Note 1 to entry: The SYSTEM AXIS of a GAMMA CAMERA with rotating detectors is the axis of rotation. 461 [SOURCE: IEC 61675-1:2022, 3.1.1.7, modified – The note to entry has been changed] 462 3.53 463 SYSTEM NON-UNIFORMITY OF RESPONSE 464 NON-UNIFORMITY OF RESPONSE of the DETECTOR HEAD with COLLIMATOR 465 3.54 466 SYSTEM SENSITIVITY 467 <GAMMA CAMERA> with a specified COLLIMATOR and ENERGY WINDOW, the ratio of the COUNT 468 RATE of the DETECTOR HEAD to the ACTIVITY of a plane source of specific dimensions and 469 containing a specified RADIONUCLIDE placed perpendicular to and centred on the COLLIMATOR 470 AXIS under specified conditions 471 Note 1 to entry: See also Figure 2.
300 mm 170 mm
Source
COLLIMATOR FRONT FACE (GAMMA CAMERA)
Material: polymethylmethacrylate
IEC 1707/05 472 473 Figure 2 – Cylindrical phantom showing a source container within a scattering media 474 used in tests for planar system sensitivity and count rate performance. The hashed area 475 indicates the scattering material.
10 mm 20 mm 80 mm d 477 SYSTEM SPATIAL RESOLUTION 478 <GAMMA CAMERA> SPATIAL RESOLUTION in a scattering medium for a specified COLLIMATOR, or a 479 specified RADIONUCLIDE, and at a specified distance from the COLLIMATOR FRONT FACE 480 3.56 481 TANGENTIAL RESOLUTION 482 TRANSVERSE RESOLUTION in the direction orthogonal to the direction of RADIAL RESOLUTION 483 [SOURCE: IEC 61675-1:2022, 3.4.1.2] 484 3.57 485 TOMOGRAPHIC VOLUME 486 juxtaposition of all volume elements which contribute to the measured PROJECTIONS for all 487 PROJECTION ANGLES 488 Note 1 to entry: For a rotating GAMMA CAMERA with a circular field of view the TOMOGRAPHIC VOLUME is a sphere 489 provided that the RADIUS OF ROTATION is larger than the radius of the field of view. For a rectangular field of view, 490 the TOMOGRAPHIC VOLUME is a cylinder.
491 [SOURCE: IEC 61675-1:2022, 3.1.1.8, modified – A note to entry has been added.] 492 3.58 493 TRANSVERSE POINT SPREAD FUNCTION 494 reconstructed two-dimensional POINT SPREAD FUNCTION in a tomographic IMAGE PLANE 495 Note 1 to entry: In TOMOGRAPHY, the TRANSVERSE POINT SPREAD FUNCTION can also be obtained from a LINE 496 SOURCE located parallel to the SYSTEM AXIS.
497 [SOURCE: IEC 61675-1:2022, 3.3.3] 498 3.59 499 TRANSVERSE RESOLUTION 500 SPATIAL RESOLUTION in a reconstructed plane perpendicular to the SYSTEM AXIS 501 [SOURCE: IEC 61675-1:2022, 3.4.1] 502 3.60 503 VOLUME SENSITIVITY 504 sum of the individual SLICE SENSITIVITIES 505 [SOURCE: IEC 61675-1:2022, 3.7] 506 3.61 507 VOXEL 508 volume element in the object which is assigned to a MATRIX ELEMENT in a two- or three- 509 dimensional IMAGE MATRIX 510 Note 1 to entry: The dimensions of the VOXEL are determined by the dimensions of the corresponding MATRIX 511 ELEMENT via the appropriate scale factors and by the systems SPATIAL RESOLUTION in all three dimensions. 512 [SOURCE: IEC 61675-1:2022, 3.2.3] 513 4 Test methods 514 4.1 General 515 All measurements shall be performed with the PULSE AMPLITUDE ANALYSER WINDOW set as 516 specified in Table 1. Additional measurements with other settings as specified by the 517 manufacturer can be performed. Before the measurements are performed, the tomographic 518 system shall be adjusted by the procedure normally used by the manufacturer for an installed 519 unit and shall not be adjusted specially for the measurement of specific parameters. If any 520 test cannot be carried out exactly as specified in this document, the reason for the deviation 521 and the exact conditions under which the test was performed shall be stated clearly. 522 Table 1 – RADIONUCLIDES and ENERGY WINDOWS to be used for performance measurements ENERGY WINDOW
RADIONUCLIDE keV 99m
Tc 141 ( 10 %) 131
I 364 ( 10 %) 67
Ga 93, 184, 300 ( 10 %) 57
Co 122 ( 10 %) 177
Lu 113, 208 ( 10 %)
NOTE Because the characteristics of a GAMMA CAMERA may change noticeably 57 99m between 122 keV ( Co) and 141 keV ( Tc), the former is not included as a suitable RADIONUCLIDE. However, it may be useful in some circumstances, e.g., for quality 177 67 control. Lu is included as a possible alternative to Ga for assessing multiple window spatial registration.
523 524 Unless otherwise specified, each DETECTOR HEAD in the system shall be characterized by a full 525 data set.
526 Unless otherwise specified, planar imaging characterization shall be provided for data 527 acquired with those corrections used in clinical practice.
528 Unless otherwise specified, SPECT characterization shall be provided for an acquisition 529 covering the minimal rotation required to obtain a complete set of data (e.g., 120° for a three- 530 headed system). If the tomograph is specified to operate in a non-circular orbiting mode 531 influencing the performance parameters, test results for the non-circular orbiting mode shall 532 be reported in addition. For SPECT characterization, all corrections that are used for clinical 533 acquisitions are applied unless explicitly stated otherwise.
534 Unless otherwise specified, measurements shall be carried out at COUNT RATES not exceeding 535 40 000 counts per second.
536 4.2 Planar imaging 537 4.2.1 SYSTEM SENSITIVITY 538 4.2.1.1 General 539 SYSTEM SENSITIVITY is a parameter that characterizes the effectiveness of a system to identify 540 the radiation emitted from a RADIOACTIVE SOURCE, i.e., the rate at which events are detected in 541 the presence of a RADIOACTIVE SOURCE with low ACTIVITY where COUNT LOSS is negligible. The 542 measured COUNT RATE for a given ACTIVITY and RADIONUCLIDE depends on many factors, 543 including the detector material, its size and thickness, the size and shape of the RADIOACTIVE 544 SOURCE including its absorption and scatter properties, and instrument’s dead time, energy 545 thresholds and COLLIMATOR.
546 4.2.1.2 Purpose 547 The purpose of this measurement is to determine the detected rate of events per unit of 548 ACTIVITY for a standard volume source of given dimensions and a specified COLLIMATOR.
...