Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO/DIS 10993-2:2024)

This document specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices. It is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or that of the medical devices themselves.
This document makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests.
This document applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices.
This document does not apply to tests performed on invertebrate animals and other lower forms; nor (other than with respect to provisions relating to species, source, health status, and care and accommodation) does it apply to testing performed on isolated tissues and organs taken from vertebrate animals that have been euthanized.

Biologische Beurteilung von Medizinprodukten - Teil 2:Tierschutzbestimmungen (ISO/DIS 10993‑2:2024)

Évaluation biologique des dispositifs médicaux - Partie 2: Exigences relatives à la protection des animaux (ISO/DIS 10993-2:2024)

Le présent document spécifie les exigences minimales à satisfaire afin d'assurer et de démontrer que des dispositions appropriées ont été prises pour la protection des animaux utilisés dans le cadre d'expérimentations animales visant à évaluer la biocompatibilité des matériaux utilisés dans des dispositifs médicaux. Il s'adresse aux personnes qui commandent, conçoivent et réalisent des essais ou qui évaluent les données provenant d'expérimentations animales entreprises pour évaluer la biocompatibilité des matériaux destinés à être utilisés dans des dispositifs médicaux, ou des dispositifs médicaux eux-mêmes.
Le présent document donne des recommandations et préconisations visant à réduire davantage à l'avenir le nombre total d'animaux utilisés, à affiner les méthodes d'essai en vue de réduire ou d'éliminer la douleur ou la détresse des animaux, ainsi qu'à remplacer les expérimentations animales par d'autres moyens scientifiquement validées, ne nécessitant pas d'expérimentation animale.
Le présent document est applicable à l'expérimentation animale effectuée sur des vertébrés vivants, à l'exclusion de l'homme, afin d'établir la biocompatibilité des matériaux ou des dispositifs médicaux.
Le présent document n'est pas applicable aux expérimentations effectuées sur des animaux invertébrés et autres formes de vie inférieures ; il ne s'applique pas non plus (sauf en ce qui concerne les dispositions relatives à l'espèce, à l'origine, à l'état de santé, aux soins et à l'hébergement) aux expérimentations réalisées sur des tissus isolés et des organes prélevés à partir d'animaux vertébrés ayant été euthanasiés.

Biološko ovrednotenje medicinskih pripomočkov - 2. del: Zahteve za varstvo živali (ISO/DIS 10993-2:2024)

General Information

Status
Not Published
Publication Date
02-Nov-2025
Current Stage
4599 - Dispatch of FV draft to CMC - Finalization for Vote
Start Date
28-Feb-2025
Completion Date
28-Feb-2025

Relations

Overview

The prEN ISO 10993-2:2024 standard, titled Biological Evaluation of Medical Devices - Part 2: Animal Welfare Requirements, issued by CEN, establishes minimum requirements to ensure the ethical treatment and welfare of animals used in biocompatibility testing of medical device materials. This document targets professionals who commission, design, perform, or evaluate animal tests for assessing the biological safety of medical devices or their materials.

Focusing on living vertebrate animals (excluding humans), prEN ISO 10993-2 provides guidance to reduce animal use, refine testing methods to minimize pain and distress, and promote alternatives to animal testing when scientifically valid. The standard does not cover tests on invertebrates or tissues from euthanized animals except in relation to species care considerations.

Key Topics

  • Animal Welfare Requirements
    Specifies provisions for animal care, accommodation, and ethical treatment during biocompatibility tests including proper restraint, surgical procedures, and humane endpoints.

  • Justification for Animal Testing
    Outlines criteria for the scientific necessity of animal tests, promoting alternatives where possible to reduce reliance on animals.

  • Personnel Competence
    Emphasizes the importance of skilled and trained personnel conducting animal tests to ensure proper welfare and test integrity.

  • Test Planning and Performance
    Details on protocol preparation, use of re-use of animals, and sequencing of in vitro and in vivo tests to optimize humane and scientific outcomes.

  • Humane Endpoints and Euthanasia
    Defines early termination criteria to alleviate animal suffering and prescribes minimally distressful euthanasia methods if necessary.

  • Study Documentation and Data Validity
    Requires thorough documentation to validate test results and support mutual acceptance of data across regulatory bodies.

  • 3Rs Principle: Reduce, Refine, Replace
    Strongly promotes the 3Rs philosophical framework to minimize animal use, refine techniques, and replace animal tests with alternative methods when feasible.

Applications

  • Medical Device Development and Safety Testing
    Ensures that biocompatibility assessment of new materials or devices complies with internationally recognized animal welfare standards.

  • Regulatory Compliance
    Assists manufacturers and testing bodies in meeting legal and ethical requirements around animal testing in medical device evaluation.

  • Laboratory Animal Management
    Provides guidance for veterinarians and animal care staff in maintaining high welfare standards through competent care and procedural training.

  • Test Method Optimization
    Supports scientists in designing test strategies that balance humane treatment with scientific rigor, including the integration of in vitro alternatives.

  • Ethical Review Processes
    Acts as a reference for institutional animal care and use committees to evaluate proposed animal studies and ensure justification and humane conduct.

Related Standards

  • ISO 10993-1: Biological Evaluation of Medical Devices Part 1 - Evaluation and Testing Within a Risk Management Process
  • ISO 10993 Series: Comprehensive guidelines addressing various aspects of biological evaluation including cytotoxicity, sensitization, and systemic toxicity tests.
  • National and International Animal Welfare Regulations: Complementary laws and directives governing animal testing ethics and welfare.
  • OECD Guidelines for Testing of Chemicals: Provides standardized animal testing protocols relevant to biocompatibility assessment.

Keywords: prEN ISO 10993-2, biological evaluation, medical devices, animal welfare, biocompatibility testing, 3Rs principle, humane endpoints, animal testing guidelines, in vivo testing, veterinary care, animal ethics, test method refinement, reduction and replacement of animal use.

Draft
prEN ISO 10993-2:2025
English language
20 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Frequently Asked Questions

prEN ISO 10993-2 is a draft published by the European Committee for Standardization (CEN). Its full title is "Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO/DIS 10993-2:2024)". This standard covers: This document specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices. It is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or that of the medical devices themselves. This document makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests. This document applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices. This document does not apply to tests performed on invertebrate animals and other lower forms; nor (other than with respect to provisions relating to species, source, health status, and care and accommodation) does it apply to testing performed on isolated tissues and organs taken from vertebrate animals that have been euthanized.

This document specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices. It is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or that of the medical devices themselves. This document makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests. This document applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices. This document does not apply to tests performed on invertebrate animals and other lower forms; nor (other than with respect to provisions relating to species, source, health status, and care and accommodation) does it apply to testing performed on isolated tissues and organs taken from vertebrate animals that have been euthanized.

prEN ISO 10993-2 is classified under the following ICS (International Classification for Standards) categories: 11.100.20 - Biological evaluation of medical devices. The ICS classification helps identify the subject area and facilitates finding related standards.

prEN ISO 10993-2 has the following relationships with other standards: It is inter standard links to EN ISO 10993-2:2022. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase prEN ISO 10993-2 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CEN standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-januar-2025
Biološko ovrednotenje medicinskih pripomočkov - 2. del: Zahteve za varstvo živali
(ISO/DIS 10993-2:2024)
Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO/DIS
10993-2:2024)
Biologische Beurteilung von Medizinprodukten - Teil 2:Tierschutzbestimmungen
(ISO/DIS 10993-2:2024)
Évaluation biologique des dispositifs médicaux - Partie 2: Exigences relatives à la
protection des animaux (ISO/DIS 10993-2:2024)
Ta slovenski standard je istoveten z: prEN ISO 10993-2
ICS:
11.100.20 Biološko ovrednotenje Biological evaluation of
medicinskih pripomočkov medical devices
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

DRAFT
International
Standard
ISO/DIS 10993-2
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 2: 2024-11-11
Animal welfare requirements
Voting terminates on:
2025-02-03
Évaluation biologique des dispositifs médicaux —
Partie 2: Exigences relatives à la protection des animaux
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS.
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Reference number
ISO/DIS 10993-2:2024(en)
DRAFT
ISO/DIS 10993-2:2024(en)
International
Standard
ISO/DIS 10993-2
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 2:
Animal welfare requirements
Voting terminates on:
Évaluation biologique des dispositifs médicaux —
Partie 2: Exigences relatives à la protection des animaux
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
© ISO 2024
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
BE CONSIDERED IN THE LIGHT OF THEIR
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
or ISO’s member body in the country of the requester.
NATIONAL REGULATIONS.
ISO copyright office
RECIPIENTS OF THIS DRAFT ARE INVITED
CP 401 • Ch. de Blandonnet 8
TO SUBMIT, WITH THEIR COMMENTS,
CH-1214 Vernier, Geneva
NOTIFICATION OF ANY RELEVANT PATENT
Phone: +41 22 749 01 11
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland Reference number
ISO/DIS 10993-2:2024(en)
ii
ISO/DIS 10993-2:2024(en)
Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Requirements . 3
4.1 General .3
4.2 Justification for animal tests .4
4.3 Competence of personnel .5
4.4 Planning and performance of animal tests .5
4.4.1 General .5
4.4.2 Re-use .5
4.5 Test strategy — Sequence of in vitro and in vivo tests .6
4.6 Animal care and accommodation .6
4.6.1 General .6
4.6.2 Restraint .6
4.6.3 Surgical procedures . .6
4.7 Humane end points .7
4.7.1 General .7
4.7.2 Euthanasia .7
4.8 Study documentation . .7
4.9 Validity of test results and mutual acceptance of data .8
Annex A (informative) Rationale for the development of this document . 9
Annex B (informative) Further suggestions for replacing, reducing and refining animal tests .13
Bibliography . 14

iii
ISO/DIS 10993-2:2024(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any patent
rights identified during the development of the document will be in the Introduction and/or on the ISO list of
patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of medical
devices, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/
TC 206, Biological and clinical evaluation of medical devices, in accordance with the Agreement on technical
cooperation between ISO and CEN (Vienna Agreement).
This fourth edition cancels and replaces the third edition (ISO 10993-2:2022).
The main change is the update of the bibliography entries.
A list of all parts in the ISO 10993 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iv
ISO/DIS 10993-2:2024(en)
Introduction
The goal of the ISO 10993 series is the protection of humans in the context of the use of medical devices.
This document supports the goal of the ISO 10993 series by promoting good science through paying proper
regard to maximizing the use of scientifically sound non-animal tests and by ensuring that those animal
tests performed to evaluate the biological properties of materials used in medical devices are conducted
humanely according to recognized ethical and scientific principles.
The application of such humane experimental techniques, including high standards of animal care and
accommodation, both help to ensure the scientific validity of safety testing and enhance the welfare of the
animals used.
v
DRAFT International Standard ISO/DIS 10993-2:2024(en)
Biological evaluation of medical devices —
Part 2:
Animal welfare requirements
1 Scope
This document specifies the minimum requirements to be satisfied to ensure and demonstrate that proper
provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of
materials used in medical devices. It is aimed at those who commission, design and perform tests or evaluate
data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical
devices, or that of the medical devices themselves.
This document makes recommendations and offers guidance intended to facilitate future further reductions
in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in
animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests.
This document applies to tests performed on living vertebrate animals, other than man, to establish the
biocompatibility of materials or medical devices.
This document does not apply to tests performed on invertebrate animals and other lower forms; nor (other
than with respect to provisions relating to species, source, health status, and care and accommodation) does
it apply to testing performed on isolated tissues and organs taken from vertebrate animals that have been
euthanized.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 10993-1 and the following apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at https:// www .electropedia .org/
— ISO Online browsing platform: available at https:// www .iso .org/ obp
3.1
alternative method
test method that replaces an animal test (3.3), reduces the numbers of animals (3.2) used or refines the
procedures applied
3.2
animal
live non-human vertebrate, excluding immature forms during the first half of gestation or incubation

ISO/DIS 10993-2:2024(en)
3.3
animal test
use of an animal (3.2) for scientific purposes
Note 1 to entry: The definition of an animal test excludes acts of recognized veterinary practice applied for the benefit
of an animal or the group of animals of which it is part; recognized husbandry practices to manage or conserve the
animal or the group of which it is part; marking by methods which cause no more than momentary pain or distress;
and euthanasia (3.5).
Note 2 to entry: The prevention of pain, suffering, distress or lasting harm by the effective use of anaesthesia or
analgesia or other methods of rendering the animal insentient to pain (e.g. decerebration) does not place animal tests
outside the scope of this definition. The administration of anaesthetics, analgesics or other methods of rendering the
animal insentient to pain are considered to constitute an integral part of the animal test.
3.4
competent authority
body designated or recognized by a national government to take responsibility for overseeing, supervising
or regulating animal tests (3.3), or the breeding and supply of purpose-bred animals (3.10) for use on such
tests, within the scope of this document
3.5
euthanasia
humane killing of an animal (3.2) by a method causing minimal physical and mental suffering
3.6
humane endpoint
pre-determined, specific criteria and measures to be implemented to minimize or terminate pain, suffering
or distress caused by animal tests (3.3) as soon as
— the scientific objectives have been met, or
— when it is realized they cannot be met, or
— when the animal (3.2) welfare problems being encountered are greater than can be justified by the
importance, potential benefits, objectives and nature of the study
3.7
laboratory animal veterinarian
qualified laboratory animal veterinarian
person responsible for the health and well-being of all laboratory animals (3.2) used at the institution
Note 1 to entry: Often called “attending veterinarian” who is certified or has training or experience in laboratory
animal science and medicine or is otherwise qualified in the care of the species being used.
Note 2 to entry: It is recommended that a laboratory animal veterinarian appropriately qualified by the relevant
competent authority should be used as an attending veterinarian.
Note 3 to entry: There is an International Association of Colleges of Laboratory Animal Medicine (IACLAM) that is an
association of associations, specifically the member Colleges of laboratory animal medicine. Each college has members
that, in addition to their demonstrated proficiency in laboratory animal medicine, also possess subspecialization in a
variety of areas that have direct bearing on the care, use and welfare of laboratory animals.
3.8
procedural training
prior training and acclimatizing of animals (3.2) to the interventions to be performed during an animal test,
with a view to minimizing stress to the animal when animal tests are conducted
3.9
protocol
documentation prepared in advance of animal tests (3.3) being undertaken setting out the justification,
rationale and test method [including scientific and humane endpoints (3.6)] for the animal tests

ISO/DIS 10993-2:2024(en)
3.10
purpose-bred animal
animal (3.2) bred with the intention that it be used in animal tests (3.3) or for other experimental or scientific
purposes
3.11
reduction
decrease to the essential minimum the number of animals (3.2) used in an animal test (3.3) to meet a defined
scientific objective
3.12
refinement
total of the measures taken to safeguard the welfare of the test animals (3.14) by minimizing any resulting
pain, suffering, distress or lasting harm to the animals (3.2) that are used
3.13
replacement
scientifically valid and reasonably and practically available test method that either completely or partially
replaces the use of living vertebrate animals (3.2) with test methods that have not the potential to cause
pain or distress to animals
3.14
test animal
animal (3.2) used in in vivo animal tests (3.3), or used to provide tissue for ex vivo or in vitro tests
3.15
validation
formal process by which the reliability and relevance of a test method is established for a particular purpose
3.16
veterinary care
responsibility for promoting an animal’s (3.2) health and welfare before, during and after animal procedures
and providing advice and guidance based on best practice
Note 1 to entry: Veterinary care includes attention to the physical and behavioural status of the animal.
Note 2 to entry: The laboratory animal veterinarian (3.7) shall have authority and responsibility for making judgements
concerning animal welfare.
Note 3 to entry: Veterinary advice and care shall be available at all times.
4 Requirements
4.1 General
This document sets forth essential requirements when animal tests are being considered, planned or
performed for the biological evaluation of materials used in medical devices.
It has been developed to protect the welfare of animals used in the biological evaluation of materials used
in medical devices without compromising, indeed to help to ensure, the scientific validity of the test results
and the risk assessments that shall subsequently be performed.
This document focuses on the need to demonstrate that animal welfare is properly considered when expert
judgement has to be exercised in relation to the biological evaluation of medical device materials, and that
the principles of humane experimental technique are demonstrably applied to the design and conduct of
animal tests.
This document requires that the need to perform animal tests is justified, and any pain, suffering, distress
or lasting harm that is caused during essential animal tests is minimized.

ISO/DIS 10993-2:2024(en)
This document sets out essential requirements that safeguard animal welfare by minimizing the pain and
distress caused when animal tests are considered or undertaken by:
— establishing a framework that reflects the relevant ethical and, in many jurisdictions, the legal
considerations relating to the use of animals for experimental or other scientific purposes;
— minimizing the number of animal tests by the appropriate use of literature searches, data-sharing,
validated replacement alternatives, and appropriate testing strategies and study designs;
— minimizing any pain, suffering, distress and lasting harm caused to animals used in tests to evaluate the
biocompatibility of materials used in medical devices by requiring appropriate use of relevant reduction
and refinement alternatives;
— promoting consistent, high standards of accommodation and care to safeguard both the welfare of the
animals used and the scientific validity and the reproducibility of the data generated;
— appropriate veterinary care program overseen by a qualified laboratory animal veterinarian is
implemented.
To these ends the design and conduct of animal tests to evaluate the biocompatibility of materials used in
medical devices shall be formed by, and incorporate, relevant strategies for the replacement, reduction and
refinement of animal tests.
Commissioning animal tests without seeking and obtaining this information, exercising these judgements
and implementing these measures does not comply with the essential requirements of this document.
NOTE These principles, and the essential requirements of this document, can also be relevant to animal use for
medical device training and development.
4.2 Justification for animal tests
When required to make proper provision to ensure human safety, animal testing to enable the proper
biological characterization of materials used in medical devices is acceptable.
For the purposes of the ISO 10993 series, animal tests shall only be deemed to be justified when:
— the resulting data are not otherwise available, but are essential to properly characterize the test material
in the context in which it is to be used;
— when no suitable scientifically validated test method not involving the use of living animals is reasonably
and practically available;
— when relevant reduction and refinement strategies have been identified and implemented including, if
appropriate, obtaining test data from manufacturers and suppliers, and literature searches for toxicity
and biocompatibility data.
To avoid unnecessary duplication, before animal tests to evaluate the biocompatibility of materials used
in medical devices are undertaken, a review of available, relevant information on the properties of the test
material shall be undertaken and documented. This shall include taking reasonable steps to enable data
sharing.
Animal tests are deemed to be justified only when:
— they have been shown to be relevant and reliable for the purposes for which they are undertaken;
— the resulting data are essential to properly characterize and evaluate the test material in the context in
which it is to be used in medical devices;
— no scientifically valid test method not requiring the use of living animals is reasonably and practically
available;
— other relevant and appropriate strategies to minimize the pain, suffering, distress and lasting harm
caused to the animals that are used have been identified and implemented.

ISO/DIS 10993-2:2024(en)
4.3 Competence of personnel
Animal tests shall be designed, conducted and interpreted by persons competent to discharge the
responsibilities assigned to them.
Animal tests shall be designed and conducted with the involvement of personnel with expertise in veterinary
science, laboratory animal science and medicine, and animal husbandry and care.
Details of how staff are equipped by experience, qualification and training (including continued professional
development) to satisfy these requirements shall be documented.
NOTE 1 Although this document does not provide an objective specification, it is considered important that those
involved in the conduct of animal tests demonstrate a caring and respectful attitude to the animals used, i.e. that they
have an appropriate “culture of care”.
NOTE 2 For further information on assurance of training and competency, see 7.8.5 of Reference [1].
4.4 Planning and performance of animal tests
4.4.1 General
The selection and design of animal tests shall be appropriate to meet the specific scientific objectives of
the study while minimizing the pain, suffering, distress or lasting harm that can be produced to the test
animals.
As specified in 4.2, animal testing shall only be undertaken when the information required is essential to
characterize the test material, is not otherwise available and when no suitable scientifically validated test
method not involving the use of living animals is reasonably and practically available.
Following consideration of relevant and reasonably available potential replacement, reduction and
refinement strategies, and before animal tests are undertaken, principal investigators and/or sponsors shall
attest and document that no other replacement, reduction or refinement strategies are required to minimize
the animal welfare costs of the studies.
NOTE In some instances pilot studies can be required to optimize study design before definitive studies can be
designed and performed.
Where the provisions of the ISO 10993 series require or permit that an informed choice be made from a
range of species, stages of development or animal numbers for an animal test, the decisions taken shall both
safeguard the scientific validity of the test and minimize any pain, suffering, distress or lasting harm to the
animals used. The rationale for the decisions taken shall be documented.
4.4.2 Re-use
The need to avoid undue cumulative welfare costs to the individual animals used shall be balanced against
the need to minimize the number of animals used.
In general, an animal should not be used for more than one test.
Animals that have experienced pain and distress in the course of an animal test, or whose previous use can
influence the outcome of further tests, shall not be re-used.
Re-use shall be consistent with the scientific objective and shall not impose unreasonable cumulative
welfare costs on the individual animal.
Any re-use shall be documented, giving summary details of the earlier use and confirming that the
requirements set out in this subclause are considered and met.

ISO/DIS 10993-2:2024(en)
4.5 Test strategy — Sequence of in vitro and in vivo tests
Testing strategies shall, as appropriate, adopt a tiered or hierarchical approach to minimize both the amount
of animal testing required and any pain or distress that can be caused when animal tests are justified and
undertaken. Specifically, unnecessary animal tests shall not be performed before appropriate, scientifically
valid, and reasonably and practically available preliminary in vitro tests have been carried out, and the
results evaluated.
Animal tests shall not be performed if the available data (e.g. from literature and/or database searches,
results from previous screening tests, validated in vitro tests, previous animal tests or any other available
relevant evidence) provide sufficient information on the biocompa
...