Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy

This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use.
The method is applicable to the determination of the mass con¬centration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for CO for the type approval test.
NOTE 1   Other ranges may be used depending on the levels present in ambient air.
NOTE 2   When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply.
The method covers the determination of ambient air concentrations of carbon monoxide in areas classified as rural, urban-background and traffic-orientated and those influenced by industrial sources.  
The results are expressed in mg/m3 (at 20 °C and 101,3 kPa).
NOTE 3   100 mg/m3 of CO corresponds to 86 mol/mol of CO.
This standard contains information for different groups of users.
Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment.
Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about:
-   Type-approval test conditions, test procedures and test requirements;
-   Analyser performance requirements;
-   Evaluation of the type-approval test results;
-   Evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type-approval test results.
Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about:
-   Initial installation of the analyser in the monitoring network and acceptance testing;
-   Ongoing quality assurance/quality control;
-   Calculation and reporting of measurement results;
-   Evaluation of the uncertainty of measurement results under practical monitoring conditions.

Außenluft - Messverfahren zur Bestimmung der Konzentration von Kohlenmonoxid mit nicht-dispersiver Infrarot-Photometrie

Dieses Dokument legt ein kontinuierliches Messverfahren nach dem nicht-dispersiven Infrarot-Messprinzip zur Bestimmung der Konzentration von Kohlenmonoxid in Außenluft fest. Es nennt die Leistungs¬kenngrößen und legt die Mindestanforderungen für die Eignungsprüfung von Infrarot-Photometern fest. Es beinhaltet auch die Bewertung der Eignung eines Messgeräts für den Einsatz an einem bestimmten ortsfesten Messstandort bezüglich der Anforderungen an die Datenqualität nach Anhang I der Richtlinie 2008/50/EG [1] und bezüglich der Anforderungen für Probennahme, Kalibrierung und Qualitätssicherung.
Das Verfahren ist anwendbar zur Bestimmung der Massenkonzentration von Kohlenmonoxid in Außenluft bis 100 mg/m3. Dieser Konzentrationsbereich entspricht dem Zertifizierungsbereich für die Eignungsprüfung.
ANMERKUNG 1      Je nach Konzentrationsniveau in der Luft können auch andere Bereiche gewählt werden.
ANMERKUNG 2      Wird diese Norm für andere Zwecke als für Messungen nach Richtlinie 2008/50/EG eingesetzt, dann brauchen die Anforderungen hinsichtlich Anwendungsbereich und Unsicherheit nicht zur Geltung kommen.
Das Verfahren umfasst die Bestimmung der Konzentration von Kohlenmonoxid an Standorten, die als ländlich, städtischer Hintergrund, verkehrsbezogen und durch industrielle Quellen beeinflusst klassifiziert sind.
Die Ergebnisse werden in µg/m3 (bei 20 °C und 101,3 kPa) angegeben.
ANMERKUNG 3   100 mg/m3 Kohlenmonoxid entsprechen 86 µmol/mol Kohlenmonoxid.
Dieses Dokument enthält Informationen für verschiedene Nutzergruppen.
Die Abschnitte 5 bis 7 und die Anhänge B bis D enthalten allgemeine Informationen über die Grundlagen der Kohlenmonoxid-Messung mit einem nicht-dispersiven Infrarot-Photometer sowie über die Probennahme-einrichtung.
Abschnitt 8 und Anhang E richten sich speziell an Prüfinstitute und -laboratorien, die Eignungsprüfungen von Kohlenmonoxid-Messgeräten durchführen. Diese Abschnitte enthalten Informationen über
-   Prüfbedingungen, -verfahren und -anforderungen bei der Eignungsprüfung;
-   Anforderungen an die Leistungsfähigkeit des Messgeräts;
-   die Bewertung der Ergebnisse der Eignungsprüfung;
-   die Bewertung der Unsicherheit der Messergebnisse des Kohlenmonoxid-Messgeräts anhand der Eignungsprüfungsergebnisse.
Die Abschnitte 9 bis 11 und die Anhang F richten sich an die Betreiber von Messnetzen, mit denen Kohlenmonoxid-Messungen in Außenluft durchgeführt werden. Diese Abschnitte enthalten Informationen über
-   die Erstinstallation des Messgeräts im Messnetz und Eignungsbewertung;
-   die laufende Qualitätssicherung und -kontrolle;
-   die Berechnung und Übermittlung der Messergebnisse;
-   die Bewertung der Unsicherheit der Messergebnisse unter Messbedingungen.

Air ambiant - Méthode normalisée de mesurage de la concentration en monoxyde de carbone par spectroscopie à rayonnement infrarouge non dispersif

La présente Norme européenne spécifie une méthode de mesurage en continu pour la détermination de la concentration en monoxyde de carbone dans l’air ambiant, selon le principe de mesure par spectroscopie à rayonnement infrarouge non dispersif. La présente norme décrit les caractéristiques de performance et fixe les critères minimaux requis pour sélectionner un analyseur par spectroscopie à rayonnement infrarouge non dispersif à l’aide d’essais d’approbation de type. Elle présente également l’évaluation de l’aptitude à l’emploi d’un analyseur sur un site fixe spécifique de manière à répondre aux exigences de qualité des données spécifiées dans l’Annexe I de la Directive 2008/50/CE [1] ainsi qu’aux exigences à observer au cours du prélèvement, de l’étalonnage et de l’assurance qualité.
La méthode s’applique à la détermination de la concentration massique en monoxyde de carbone présent dans l’air ambiant jusqu’à 100 mg/m3. Cette plage de concentration représente l’étendue de mesure certifiée pour l’essai d’approbation de type.
NOTE 1   D’autres plages peuvent être utilisées selon les niveaux présents dans l’air ambiant.
NOTE 2   Lorsque la norme est utilisée à d’autres fins que la Directive 2008/50/CE, les exigences relatives à la plage et à l’incertitude peuvent ne pas s’appliquer.
La méthode couvre la détermination des concentrations en monoxyde de carbone présent dans l’air ambiant dans les zones classées comme zones rurales, urbaines de fond, proches d’axes de circulation automobile et influencées par des sources industrielles.  
Les résultats sont exprimés en mg/m3 (à 20 °C et 101,3 kPa).
NOTE 3   100 mg/m3 de CO correspondent à 86 µmol/mol de CO.
La présente norme contient des informations pour différents groupes d’utilisateurs.
Les Articles 5 à 7 et les Annexes B, C et D contiennent des informations générales relatives aux principes de mesurage du monoxyde de carbone à l’aide d’un analyseur par spectroscopie à rayonnement infrarouge non dispersif et d’un équipement de prélèvement.
L’Article 8 et l’Annexe E concernent tout particulièrement les entreprises et les laboratoires d’essai qui effectuent des essais d’approbation de type des analyseurs de monoxyde de carbone. Ces sections contiennent des informations concernant :
-   les conditions de l’essai d’approbation de type, les modes opératoires d’essai et les exigences d’essai ;
-   les exigences de performance de l’analyseur ;
-   l’évaluation des résultats de l’essai d’approbation de type ;
-   l’évaluation de l’incertitude des résultats de mesurage de l’analyseur de monoxyde de carbone d’après les résultats de l’essai d’approbation de type.
Les Articles 9 à 11 et les Annexes F à H concernent les réseaux de surveillance effectuant les mesurages pratiques du monoxyde de carbone dans l’air ambiant. Ces sections contiennent des informations concernant :
-   l’installation initiale de l’analyseur dans le réseau de surveillance et l’essai de réception ;
-   l’assurance qualité / le contrôle qualité en continu ;
-   le calcul et le compte-rendu des résultats de mesurage ;
-   l’évaluation de l’incertitude des résultats de mesurage sous des conditions de surveillance pratiques.

Zunanji zrak - Standardna metoda za določevanje koncentracije ogljikovega monoksida z nedisperzivno infrardečo spektroskopijo

Ta evropski standard določa metodo stalnega merjenja za določitev koncentracije ogljikovega monoksida v zunanjem zraku na podlagi principa nedisperzne infrardeče spektrometrije. Ta standard opisuje lastnosti delovanja in vzpostavlja ustrezna minimalna merila, potrebna za izbiro ustreznega nedisperznega infrardečega spektroskopskega analizatorja s homologacijskimi preizkusi. Vključuje tudi oceno ustreznosti analizatorja za uporabo na specifičnem fiksnem mestu, da se izpolnijo zahteve o kakovosti podatkov, kot je opredeljeno v Prilogi I Direktive 2008/50/ES, ter zahteve med vzorčenjem, kalibriranjem in zagotavljanjem kakovosti za uporabo. Ta metoda velja za določitev masne koncentracije ogljikovega monoksida v zunanjem zraku do vrednosti 100 mg/m3 ogljikovega monoksida. Ta razpon koncentracij predstavlja razpon certificiranja za homologacijski preizkus. Ta metoda pokriva določitev koncentracij ogljikovega monoksida v zunanjem zraku na področjih, ki so uvrščena med podeželska območja, območja v urbanem okolju, prometna območja in lokacije, na katere vplivajo industrijski viri. Rezultati so izraženi v mg/m3 (pri 20 °C in 101,3 kPa). Ta standard vsebuje informacije za različne skupine uporabnikov. Določbe od 5 do 7 ter dodatki B, C in D vsebujejo splošne informacije o načelih merjenja ogljikovega monoksida z nedisperznim infrardečim spektroskopskim analizatorjem in opremo za vzorčenje. Osrednji predmet obravnave določbe 8 in dodatka E so laboratoriji, ki izvajajo homologacijske preizkuse analizatorjev ogljikovega monoksida. Ti razdelki vsebujejo informacije o naslednjem: – pogoji homologacijskih preizkusov, preizkusni postopki in zahteve; – zahteve glede učinkovitosti analizatorja; ocena rezultatov homologacijskih preizkusov; ocena negotovosti rezultatov merjenja analizatorja ogljikovega monoksida na podlagi rezultatov homologacijskih preizkusov. Osrednji predmet obravnave določb od 9 do 11 in dodatka F so omrežja za spremljanje stanja, ki izvajajo praktično merjenje ogljikovega monoksida v zunanjem zraku. Ti razdelki vsebujejo informacije o naslednjem: – začetna namestitev analizatorja v omrežju za spremljanje stanja in preizkušanje ustreznosti; – tekoče zagotavljanje kakovosti/nadzor kakovosti; izračun in poročanje rezultatov merjenja; ocena negotovosti rezultatov merjenja pod pogoji praktičnega spremljanja.

General Information

Status
Withdrawn
Publication Date
21-Aug-2012
Withdrawal Date
13-Apr-2025
Technical Committee
CEN/TC 264 - Air quality
Current Stage
9960 - Withdrawal effective - Withdrawal
Start Date
18-Dec-2024
Completion Date
14-Apr-2025

Relations

Effective Date
22-Dec-2008
Effective Date
13-Jun-2018

Frequently Asked Questions

EN 14626:2012 is a standard published by the European Committee for Standardization (CEN). Its full title is "Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy". This standard covers: This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass con¬centration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for CO for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in areas classified as rural, urban-background and traffic-orientated and those influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 mol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: - Type-approval test conditions, test procedures and test requirements; - Analyser performance requirements; - Evaluation of the type-approval test results; - Evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type-approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: - Initial installation of the analyser in the monitoring network and acceptance testing; - Ongoing quality assurance/quality control; - Calculation and reporting of measurement results; - Evaluation of the uncertainty of measurement results under practical monitoring conditions.

This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass con¬centration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for CO for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in areas classified as rural, urban-background and traffic-orientated and those influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 mol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: - Type-approval test conditions, test procedures and test requirements; - Analyser performance requirements; - Evaluation of the type-approval test results; - Evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type-approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: - Initial installation of the analyser in the monitoring network and acceptance testing; - Ongoing quality assurance/quality control; - Calculation and reporting of measurement results; - Evaluation of the uncertainty of measurement results under practical monitoring conditions.

EN 14626:2012 is classified under the following ICS (International Classification for Standards) categories: 13.040.20 - Ambient atmospheres. The ICS classification helps identify the subject area and facilitates finding related standards.

EN 14626:2012 has the following relationships with other standards: It is inter standard links to EN 14626:2005, EN 14626:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN 14626:2012 is associated with the following European legislation: EU Directives/Regulations: 2000/69/EC. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

You can purchase EN 14626:2012 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CEN standards.

Standards Content (Sample)


2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.MLNRYHJDLuftqualität - Messverfahren zur Bestimmung der Konzentration von Kohlenmonoxid mit nicht-dispersiver Infrarot-PhotometrieAir ambiant - Méthode normalisée de mesurage de la concentration en monoxyde de carbone par spectroscopie à rayonnement infrarouge non dispersifAmbient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy13.040.20Kakovost okoljskega zrakaAmbient atmospheresICS:Ta slovenski standard je istoveten z:EN 14626:2012SIST EN 14626:2012en,fr,de01-december-2012SIST EN 14626:2012SLOVENSKI
STANDARDSIST EN 14626:20051DGRPHãþD

EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 14626
August 2012 ICS 13.040.20 Supersedes EN 14626:2005English Version
Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy
Air ambiant - Méthode normalisée de mesurage de la concentration en monoxyde de carbone par spectroscopie à rayonnement infrarouge non dispersif
Luftqualität - Messverfahren zur Bestimmung der Konzentration von Kohlenmonoxid mit nicht-dispersiver Infrarot-Photometrie This European Standard was approved by CEN on 10 May 2012.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2012 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 14626:2012: ESIST EN 14626:2012

Test of lack of fit . 48Annex B (informative)
Sampling equipment. 50Annex C (informative)
Schematics of non-dispersive infrared spectrometer . 52Annex D (informative)
Manifold testing . 54Annex E (normative)
Type approval . 56Annex F (informative)
Calculation of uncertainty in field operation at the 8-hour limit value . 75Annex G (informative)
Significant technical changes . 83Bibliography . 84 SIST EN 14626:2012

According to the CEN/CENELEC Internal Regulations, the national standards organisations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. SIST EN 14626:2012

NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources.
The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 µmol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about:  type-approval test conditions, test procedures and test requirements;  analyser performance requirements;  evaluation of the type-approval test results;  evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type-approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about:  initial installation of the analyser in the monitoring network and acceptance testing;  ongoing quality assurance/quality control;  calculation and reporting of measurement results;  evaluation of the uncertainty of measurement results under practical monitoring conditions.
(ISO 6144) EN ISO 6145-6, Gas analysis — Preparation of calibration gas mixtures using dynamic volumetric methods — Part 6: Critical orifices (ISO 6145-6) EN ISO 6145-7, Gas analysis — Preparation of calibration gas mixtures using dynamic volumetric methods — Part 7: Thermal mass-flow controllers (ISO 6145-7) EN ISO 14956, Air quality — Evaluation of the suitability of a measurement procedure by comparison with a required measurement uncertainty (ISO 14956) EN ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories
(ISO/IEC 17025) ENV 13005:1999, Guide to the expression of uncertainty in measurement
3 Terms and definitions For the purpose of this document, the following terms and definitions apply. 3.1 adjustment set of operations carried out on a measuring system so that it provides prescribed indications corresponding to given values of a quantity to be measured Note 1 to entry:
Types of adjustment of a measuring system include zero adjustment of a measuring system, offset adjustment, and span adjustment (sometimes called gain adjustment). Note 2 to entry:
Adjustment of a measuring system should not be confused with calibration, which is a prerequisite for adjustment. [SOURCE: JCGM 200:2012 (VIM) [2]] Note 3 to entry:
In the context of this standard, adjustment is performed on measurement data rather than on the analyser. 3.2 ambient air outdoor air in the troposphere, excluding workplaces as defined by Directive 89/654/EEC, where provisions concerning health and safety at work apply and to which members of the public do not have regular access [SOURCE: 2008/50/EC[1]] SIST EN 14626:2012

3.5 calibration operation that, under specified conditions, in a first step, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties and, in a second step, uses this information to establish a relation for obtaining a measurement result from an indication Note 1 to entry: A calibration may be expressed by a statement, calibration function, calibration diagram, calibration curve, or calibration table. In some cases, it may consist of an additive or multiplicative correction of the indication with associated measurement uncertainty. Note 2 to entry: Calibration should not be confused with adjustment of a measuring system, often mistakenly called “self-calibration”, nor with verification of a calibration. Note 3 to entry:
Often, the first step alone in the above definition is perceived as being calibration.
[SOURCE: JCGM 200:2012 (VIM) [2]] Note 4 to entry:
In the context of this standard, calibration is a comparison of the analyser response to a known gas concentration with a known uncertainty when the information obtained from the comparison is used for the successive adjustment (if needed) of the analyser. 3.6 certification range concentration range for which the analyser is type-approved 3.7 check verification that the analyser is still operating within specified performance limits
3.8 combined standard uncertainty standard uncertainty of the result of a measurement when that result is obtained from the values of a number of other quantities, equal to the positive square root of a sum of terms, the terms being the variances or co-variances of these other quantities weighted according to how the measurement result varies with changes in these quantities [SOURCE: ENV 13005:1999] 3.9 coverage factor numerical factor used as a multiplier of the combined standard uncertainty in order to obtain an expanded uncertainty [SOURCE: ENV 13005:1999] 3.10 designated body body which has been designated for a specific task (type approval tests and/or QA/QC activities in the field) by the competent authority in the Member States SIST EN 14626:2012

3.12 expanded uncertainty quantity defining an interval about the result of a measurement that may be expected to encompass a large fraction of the distribution of values that could reasonably be attributed to the measurand
Note 1 to entry:
The fraction may be viewed as the coverage probability or level of confidence of the interval. Note 2 to entry:
To associate a specific level of confidence with the interval defined by the expanded uncertainty requires explicit or implicit assumptions regarding the probability distribution characterised by the measurement result and its combined standard uncertainty. The level of confidence that may be attributed to this interval can be known only to the extent to which such assumptions may be justified. [SOURCE: ENV 13005:1999] Note 3 to entry: For the purpose of this standard the expanded uncertainty is the combined standard uncertainty multiplied by a coverage factor k=2 resulting in an interval with a level of confidence of 95 %. 3.13 fall time difference between the response time (fall) and the lag time (fall) 3.14 independent measurement individual measurement that is not influenced by a previous individual measurement by separating two individual measurements by at least four response times Note 1 to entry: The largest value of response time (rise) and response time (fall) are intended. 3.15 individual measurement measurement averaged over a time period equal to the response time of the analyser Note 1 to entry: The largest value of response time (rise) and response time (fall) are intended. Note 2 to entry: This definition differs from the meaning of the concept “individual measurement” in Directive 2008/50/EC [1]. 3.16 influence quantity quantity that is not the measurand but that affects the result of the measurement
[SOURCE: ENV 13005:1999] 3.17 interferent component of the air sample, excluding the measured constituent, that affects the output signal 3.18 lack of fit maximum deviation from the linear regression line of the average of a series of measurement results at the same concentration
3.23 parallel measurements measurements from different analysers, sampling from one and the same sampling manifold, starting at the same time and ending at the same time 3.24 performance characteristic one of the parameters assigned to equipment in order to define its performance 3.25 performance criterion limiting quantitative numerical value assigned to a performance characteristic, to which conformance is tested 3.26 period of unattended operation time period over which the drift complies with the performance criterion for long term drift 3.27 repeatability (of results of measurement) closeness of the agreement between the results of successive individual measurements of carbon monoxide carried out under the same conditions of measurement Note 1 to entry: These conditions include: a) the same measurement procedure; b) the same observer; c) the same analyser, used under the same conditions; d) at the same location; e) repetition over a short period of time. 3.28 reproducibility under field conditions closeness of the agreement between the results of simultaneous measurements with two analysers in ambient air carried out under the same conditions of measurement Note 1 to entry: These conditions are called field reproducibility conditions and include: a) the same measurement procedure; b) two identical analysers, used under the same conditions; SIST EN 14626:2012

Note 1 to entry: The term ‘gas’ may refer to a test gas used in type-approval testing or to ambient air transferred to the analyser.
3.32 sampling system the assembly of components needed to transfer ambient air to the analyser 3.33 short-term drift difference between zero or span readings at the beginning and end of a 12 h period 3.34 standard uncertainty uncertainty of the result of a measurement expressed as a standard deviation
[SOURCE: ENV 13005:1999] 3.35 surrounding temperature temperature of the air directly surrounding the analyser 3.36 type approval decision taken by a designated body that the pattern of an analyser conforms to specified requirements
3.37 type approval test examination of two or more analysers of the same pattern which are submitted by a manufacturer to a designated body including the tests necessary for approval of the pattern 3.38 uncertainty (of measurement) parameter associated with the result of a measurement that characterises the dispersion of the values that could reasonably be attributed to the measurand [SOURCE: ENV 13005:1999] 4 Abbreviated terms
FEP
perfluoro-ethylene-propylene MFC mass flow controller SIST EN 14626:2012

6.1 General Depending on the installation of the infrared analyser at a monitoring station, a single sampling line for the analyser may be chosen. Alternatively, sampling can take place from a sampling system consisting of a common sampling inlet with a sampling manifold to which other analysers and equipment may be attached. Conditions and layout of the sampling system will contribute to the uncertainty of the measurement; to minimise this contribution to the expanded uncertainty, requirements for the sampling equipment are given in the following sub-clauses. NOTE
In Annex B, different arrangements of the sampling equipment are schematically presented. 6.2 Sampling location
The location where the ambient air shall be sampled and analysed is not specified as this depends strongly on the category of a monitoring station (such as measurements taken in e.g. a rural area or background area). Criteria on sampling points on a micro scale are given in Annex III of Directive 2008/50/EC [1]. 6.3 Sampling system 6.3.1 Construction The sampling system shall include a sampling inlet and may include the following components:  a sampling line or manifold;  a particle filter placed between the sampling line or manifold and the analyser;  a sampling pump in case a sampling manifold is used. The sample inlet shall be constructed in such a way that ingress of rainwater into the sampling line or manifold is prevented. The sampling line or manifold shall be as short as practical to minimise the residence time. In the case where a sampling manifold is used, an additional pump is necessary with sufficient capacity to fulfil the sampling requirements stated in the previous sub-clauses (see also 6.5 and Annex D).
The material of the sample inlet as well as the sampling line or manifold can influence the composition of the sample. In practice, the best materials, such as polytetrafluoroethylene (PTFE), perfluoro-ethylene-propylene SIST EN 14626:2012

NOTE This value can be achieved when the quality assurance and quality control requirements (see Clause 9) are followed.
The sampling line or manifold may be moderately heated to avoid condensation. Condensation may occur in the case of high ambient temperature and/or humidity. The influence on the measured concentrations of the pressure drop along the sampling inlet and line or manifold and the particle filter shall be ≤ 1,0 %. 6.3.2 Particle filter A particle filter shall be placed between the sampling line or manifold and the inlet of the analyser. The filter shall retain all particles likely to alter the performance of the analyser. It shall be made of PTFE. The material of the filter housing shall be chemically inert to carbon monoxide. The filter may be internal to the analyser (see 7.7) or external. In case the analyser contains a built-in filter, an external filter is not necessary. NOTE 1 A pore size of the filter of 5 µm usually fulfils this requirement. NOTE 2 Suitable materials for the filter housing are for example PTFE, stainless steel, or borosilicate glass.
The particle filter shall be conditioned before used in measurements. The filter shall be changed periodically depending on the dust loading at the sampling site (as indicated in 9.7). The filter housing shall be cleaned at least every six months. Overloading of the filter may cause loss of carbon monoxide by adsorption on the particle matter and may increase the pressure drop in the sampling line. 6.3.3 Loss of carbon monoxide Depending on the location of the particle filter, the sampling system can be contaminated by deposition of dust. This can induce losses of carbon monoxide. The sampling system shall be cleaned (as stated in 9.4.1) with a frequency which is dependent on the site-specific conditions. 6.3.4 Conditioning The sampling system and the particle filter shall be conditioned (at initial installation and after each cleaning) to avoid temporary decreases in the measured carbon monoxide concentrations by sampling ambient air for a period of at least 30 min at the nominal sample flow rate. Conditioning may also be done in the laboratory before installation. These conditioning periods shall not be included in the calculation of the availability of the analyser during the type approval test (see 8.5.7). NOTE Conditioning during field operation is considered a part of normal maintenance. Consequently, the concentrations measured during conditioning need not be included in the calculation of data capture, and eight-hourly averages. 6.4 Control and regulation of sample flow rate The sample flow rate into the analyser shall be maintained within the specifications of the manufacturer of the analyser.
NOTE The flow rate into the infrared analyser is usually controlled by means of restrictors. SIST EN 14626:2012

The influence of the pressure drop induced by the manifold sampling pump on the measured concentration shall be ≤ 1,0 %. 7 Analyser equipment 7.1 General In Annex C, schematic diagrams are given of two types of non-dispersive infrared analysers. The non-dispersive infrared analyser consists of the principal components that are described in 7.3 to 7.6. 7.2 Interferents 7.2.1 General As various gases absorb infrared radiation, interference from these gases can occur when their infrared absorption bands coincide or overlap the CO infrared absorption bands. The degree of interference varies among individual NDIR analysers. In general, gas correlation spectrophotometers are less sensitive to the influence of interferences (see 5.2). 7.2.2 Water vapour The primary interferent is water vapour.
NOTE Water vapour interference can be minimised by using one or more of the following measures: a) drying the sampled air with use of a semi permeable membrane or a similar drying agent; b) maintaining a constant humidity in the sample and calibration gases by refrigeration or saturation and making a volume correction for the constant humidity; c) using narrow-band optical filters. 7.2.3 Carbon dioxide Interference may be caused by carbon dioxide (CO2).
NOTE The effect of CO2 interference at concentrations normally present in ambient air is minimal; that is, 340 µmol/mol volume fraction may give a response equivalent to 0,2 µmol/mol volume fraction [4]. If necessary, CO2 may be scrubbed with soda lime. 7.2.4 Hydrocarbons Hydrocarbons at concentrations normally found in the ambient air do not ordinarily interfere. NOTE 500 µmol/mol volume fraction of methane may give a response equivalent to 0,5 µmol/mol volume fraction [4]. 7.3 Details about analyser equipment
The main parts of the analyser consist of an infrared source, one or two absorption cells, an infrared detector and a chopper wheel for modulation of the infrared radiation to the detector. SIST EN 14626:2012

The signal shall be corrected also for external pressure and temperature fluctuations. Significant pressure variations are due to synoptic meteorological changes (up to + 3 %) or due to the altitude of the measurement site (about 10 % decrease in pressure for an 800 m rise in altitude). 7.5 Flow rate indicator A flow rate indicator shall be included in the analyser. 7.6 Sampling pump for the analyser The sampling pump is situated at the outlet of the analyser, and draws the sample through the analyser. It can be separate or part of the analyser. In any case it shall be capable of operating within the specified flow requirements of the manufacturer of the analyser and pressure conditions required for the NDIR absorption cell. 7.7 Particle filter An analyser will generally contain an internal particle filter (see 6.3.2). This filter shall be considered an integral part of the analyser in type-approval testing. 8 Type approval of carbon monoxide analysers 8.1 General
The determination of the concentration of carbon monoxide in ambient air shall fulfil the requirement of a maximum uncertainty in the measured values, which is prescribed by Annex I of Directive 2008/50/EC (15 % for fixed measurements or 25 % for indicative measurements). In order to achieve an uncertainty less than (or equal to) this required uncertainty, the infrared spectrometric analyser shall fulfil all the criteria for a number of performance characteristics which are given in this standard. The values of the selected performance characteristics shall be evaluated by means of laboratory tests and field tests. By combining the values of the selected performance characteristics in the expanded uncertainty calculation, a judgement shall be made whether or not the infrared spectrometric analyser meets the criterion of maximum uncertainty prescribed by Annex I of Directive 2008/50/EC (15 % for fixed measurements or 25 % for indicative measurements). This process of assessment (type approval test) of the values of the performance characteristics comprises laboratory tests and field tests and the calculation of the expanded uncertainty. At least two analysers of the same type shall be tested in the laboratory. Two of these analysers shall be tested during the field test. All analysers tested are required to pass all tests. The type approval procedure shall fulfil the certification requirements laid down in EN 15267-1 and EN 15267-2. A designated body shall perform the type approval tests. The designated body for the type approval test shall be SIST EN 14626:2012

a) performance testing of an AMS; b) initial assessment of the AMS manufacturer’s quality management system; c) certification; d) post-certification product-surveillance. EN 15267-2 covers the supplementary requirements for an AMS manufacturer’s management system to
EN ISO 9001 [6] for the control of design and manufacturing of AMS. This European Standard also serves as a reference document for auditing the AMS manufacturer’s management system.
8.2 Relevant performance characteristics and performance criteria The performance characteristics which shall be determined during a laboratory and field test, and their related performance criteria, are given in Table 1. Table 1 applies to those ranges that are specified as normative in the scope of this standard. When type-approval tests are performed on analysers with other certification ranges, the performance criteria stated in absolute units such as µmol/mol/K may need to be modified in order to meet the uncertainty criteria in Annex I from Directive 2008/50/EC. All criteria expressed in relative units (e.g. %) shall remain the same. The determination of the value of the performance characteristics stated in Table 1 shall be performed by a designated body during the laboratory test and field test according to the procedures described in 8.4, 8.5 and Annex A. Table 1 — Relevant performance characteristics and criteria No. Performance characteristic Symbol SectionLab. t
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

This article discusses EN 14626:2012, which is a European Standard for measuring the concentration of carbon monoxide in ambient air using non-dispersive infrared spectroscopy. The standard sets performance criteria for selecting a suitable analyzer and includes requirements for sampling, calibration, and quality assurance. The method is applicable for carbon monoxide concentrations up to 100 mg/m3. The standard provides information for different user groups, including general principles of measurement and specific instructions for test houses and laboratories conducting type-approval testing, as well as monitoring networks performing practical measurements. The results of the measurement are expressed in mg/m3 at specific temperature and pressure conditions. The standard also mentions that the ranges and uncertainty requirements may not apply when used for purposes other than those specified in Directive 2008/50/EC.

이 기사는 EN 14626:2012라는 유럽 표준에 대해 설명하고 있다. 이 표준은 비산란 적외선 분광법을 이용하여 대기 중 일산화탄소 농도를 측정하는 지속적인 측정 방법을 규정한다. 이 표준은 성능 특성을 설명하며, 적합한 비산란 적외선 분광분석기를 선택하기 위해 유형 승인 테스트를 통해 필요한 최소 기준을 설정한다. 또한, 특정 고정 사이트에서 사용하기 위해 분석기의 적합성 검토와 Directive 2008/50/EC의 첨부 I에서 명시된 데이터 품질 요구 사항 및 측정, 교정, 품질 보증 기간의 요구 사항도 포함한다. 이 방법은 최대 100 mg/m3까지의 대기 중 일산화탄소 질량 농도를 측정하는 데 적용될 수 있다. 이 농도 범위는 CO의 유형 승인 테스트를 위한 인증 범위를 나타낸다. 참고로, 대기 중 존재하는 레벨에 따라 다른 범위를 사용할 수 있다(NOTE 1). Directive 2008/50/EC에서 요구하는 측정 외의 목적으로 표준을 사용할 때는 범위와 불확실성 요구 사항이 적용되지 않을 수 있다(NOTE 2). 이 방법은 시골, 도시 배경 및 교통 지향 지역 및 산업원인 영향을 받는 지역에서 대기 중 일산화탄소 농도를 측정한다. 측정 결과는 20°C와 101.3 kPa에서 mg/m3로 표시된다(NOTE 3. 100 mg/m3의 CO는 86 μmol/mol의 CO에 해당한다). 본 표준은 다른 사용자 그룹을 위한 정보도 포함하고 있다. 5에서 7항과 B, C, D부는 비산란 적외선 분광 분석기와 샘플링 장비를 통해 일산화탄소 측정의 원리에 대한 일반적인 정보를 포함하고 있다. 8항과 E조는 특히 타입 승인 테스트를 수행하는 시험소와 실험실에 대해 특정 지침을 제시한다. 이 섹션에는 다음과 같은 정보가 포함되어 있다: - 타입 승인 테스트 조건, 테스트 절차 및 테스트 요구 사항 - 분석기 성능 요구 사항 - 타입 승인 테스트 결과 평가 - 타입 승인 테스트 결과를 기반으로 한 일산화탄소 분석기 측정 결과의 불확실성 평가 9에서 11항과 F조는 대기 중 일산화탄소를 실제로 측정하는 모니터링 네트워크에 해당된다. 이들 섹션은 다음과 같은 정보를 제공한다: - 모니터링 네트워크에 대한 분석기의 초기 설치 및 수락 테스트 - 지속적인 품질 보증/품질 통제 - 측정 결과의 계산 및 보고 - 실제 모니터링 조건에서의 측정 결과의 불확실성 평가

この記事では、EN 14626:2012について説明されています。このヨーロッパ規格は、非分散赤外分光法を用いて大気中の一酸化炭素濃度を測定するための連続測定法を指定しています。この規格では、適切な非分散赤外分光分析装置を選択するための性能基準を定め、タイプ承認試験によって必要な最小基準を設定しています。また、指令2008/50/ECの付録Iで指定されたデータ品質要件を満たし、サンプリング、キャリブレーション、品質保証の要件も含まれています。 この方法は、大気中の一酸化炭素濃度を100 mg/m3まで測定することができます。この濃度範囲は、COのタイプ承認試験の認証範囲を表しています。 注1: 大気中の存在するレベルに応じて他の範囲を使用することもあります。 注2: 指令2008/50/ECで要求される測定以外の目的でこの規格を使用する場合、範囲と不確かさの要件は適用されない場合があります。 この方法は、田舎、都市の背景、交通が集中する地域、および産業源の影響を受ける地域における大気中の一酸化炭素濃度の測定をカバーしています。 結果は、20°Cおよび101.3 kPaでmg/m3で表されます。 注3: COの100 mg/m3は、COの86 μmol/molに相当します。 この規格には、異なるユーザーグループ向けの情報も含まれています。5~7条と付録B、C、Dには、非分散赤外分光分析装置とサンプリング装置による一酸化炭素測定の原理についての一般的な情報が含まれています。8条と付録Eは、一酸化炭素分析装置のタイプ承認試験を行う試験所と研究室を対象としており、次の情報が含まれています。 - タイプ承認試験の条件、手順、要件 - 分析装置の性能要件 - タイプ承認試験結果の評価 - タイプ承認試験結果に基づく一酸化炭素分析装置の測定結果の不確かさの評価 9~11条および付録Fは、実地で一酸化炭素を測定するモニタリングネットワークを対象としており、次の情報が含まれています。 - 分析装置の初期設置と受け入れテスト - 継続的な品質保証/品質管理 - 測定結果の計算と報告 - 実地モニタリング条件での測定結果の不確かさの評価