Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1 (ISO/TS 17665-2:2009)

ISO/TS 17665-2:2009 provides general guidance on the development, validation and routine control of moist heat sterilization processes and is intended to explain the requirements set forth in ISO 17665-1. The guidance given in this Technical Specification is provided to promote good practice related to moist heat sterilization processes and to assist those developing and validating a moist heat sterilization process according to ISO 17665-1.

Sterilisation von Produkten für die Gesundheitsfürsorge - Feuchte Hitze - Teil 2: Leitfaden für die Anwendung von ISO 17665-1 (ISO/TS 17665-2:2009)

Diese Technische Spezifikation liefert einen allgemeinen Leitfaden zur Entwicklung, Validierung und Lenkung der Anwendung von Sterilisationsverfahren mit feuchter Hitze und ist vorgesehen, die in ISO 17665-1 dargelegten Anforderungen zu erläutern. Der in der vorliegenden Technischen Spezifikation vorgelegte Leitfaden ist vorgesehen, bewährte Praktiken in Bezug auf Sterilisationsverfahren mit feuchter Hitze zu fördern und diejenigen zu unterstützen, die ein Sterilisationsverfahren mit feuchter Hitze nach ISO 17665-1 entwickeln und validieren. ANMERKUNG 1 Die Gliederung des Hauptteils der vorliegenden Technischen Spezifikation (Abschnitt 1 bis Abschnitt 12) entspricht der Gliederung von ISO 17665-1, daher passt der unter einem bestimmten Abschnitt von diesem Teil von ISO 17665 gegebene Leitfaden zu den im entsprechenden Abschnitt von ISO 17765-1 aufgeführten Anforderungen. Z. B. ist ein für 5.2 von ISO 17665-1:2006 gegebener Leitfaden in ISO/TS 17665-2 in 5.2 zu finden. Dieser Leitfaden wird zusätzlich zu dem in Anhang A von ISO 17665-1:2006 gegebenen Leitfaden vorgelegt. Siehe auch Anhang E. ANMERKUNG 2 In Anhang D werden besondere Hinweise gegeben, die speziell für Sterilisationsverfahren zutreffen, die in der Gesundheitsfürsorge durchgeführt werden.

Stérilisation des produits de santé - Chaleur humide - Partie 2: Directives relatives à l'application de l'ISO 17665-1 (ISO/TS 17665-2:2009)

L'ISO/TS 17665-2:2009 fournit des directives générales sur le développement, la validation et le contrôle de routine de procédés de stérilisation par chaleur humide, et vise à expliquer les exigences présentées dans l'ISO 17665-1. Les directives fournies dans l'ISO/TS 17665-2:2009 sont destinées à promouvoir les bonnes pratiques relatives aux procédés de stérilisation par chaleur humide et à faciliter le développement et la validation d'un procédé de stérilisation par chaleur humide conforme à l'ISO 17665-1.

Sterilizacija izdelkov za zdravstveno nego - Vlažna toplota - 2. del: Navodilo za uporabo ISO 17665-1 (ISO 17665-2:2009)

General Information

Status
Withdrawn
Publication Date
14-Jan-2009
Withdrawal Date
13-Apr-2025
Current Stage
9960 - Withdrawal effective - Withdrawal
Start Date
01-May-2024
Completion Date
14-Apr-2025

Relations

Technical specification
TS CEN ISO/TS 17665-2:2009
English language
55 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Frequently Asked Questions

CEN ISO/TS 17665-2:2009 is a technical specification published by the European Committee for Standardization (CEN). Its full title is "Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1 (ISO/TS 17665-2:2009)". This standard covers: ISO/TS 17665-2:2009 provides general guidance on the development, validation and routine control of moist heat sterilization processes and is intended to explain the requirements set forth in ISO 17665-1. The guidance given in this Technical Specification is provided to promote good practice related to moist heat sterilization processes and to assist those developing and validating a moist heat sterilization process according to ISO 17665-1.

ISO/TS 17665-2:2009 provides general guidance on the development, validation and routine control of moist heat sterilization processes and is intended to explain the requirements set forth in ISO 17665-1. The guidance given in this Technical Specification is provided to promote good practice related to moist heat sterilization processes and to assist those developing and validating a moist heat sterilization process according to ISO 17665-1.

CEN ISO/TS 17665-2:2009 is classified under the following ICS (International Classification for Standards) categories: 11.080.01 - Sterilization and disinfection in general. The ICS classification helps identify the subject area and facilitates finding related standards.

CEN ISO/TS 17665-2:2009 has the following relationships with other standards: It is inter standard links to EN ISO 17665:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase CEN ISO/TS 17665-2:2009 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CEN standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-april-2009
Sterilizacija izdelkov za zdravstveno nego - Vlažna toplota - 2. del: Navodilo za
uporabo ISO 17665-1 (ISO 17665-2:2009)
Sterilization of health care products - Moist heat - Part 2: Guidance on the application of
ISO 17665-1 (ISO 17665-2:2009)
Sterilisation von Produkten für die Gesundheitsfürsorge - Feuchte Hitze - Teil 2:
Leitfaden für die Anwendung von ISO 17665-1 (ISO 17665-2:2009)
Stérilisation des produits de santé - Chaleur humide - Partie 2: Directives relatives à
l'application de l'ISO 17665-1 (ISO 17665-2:2009)
Ta slovenski standard je istoveten z: CEN ISO/TS 17665-2:2009
ICS:
11.080.01 Sterilizacija in dezinfekcija na Sterilization and disinfection
splošno in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

TECHNICAL SPECIFICATION
CEN ISO/TS 17665-2
SPÉCIFICATION TECHNIQUE
TECHNISCHE SPEZIFIKATION
January 2009
ICS 11.080.01
English Version
Sterilization of health care products - Moist heat - Part 2:
Guidance on the application of ISO 17665-1 (ISO 17665-2:2009)
Stérilisation des produits de santé - Chaleur humide - Sterilisation von Produkten für die Gesundheitsfürsorge -
Partie 2: Directives relatives à l'application de l'ISO 17665- Feuchte Hitze - Teil 2: Leitfaden für die Anwendung von
1 (ISO 17665-2:2009) ISO 17665-1 (ISO 17665-2:2009)
This Technical Specification (CEN/TS) was approved by CEN on 23 November 2008 for provisional application.
The period of validity of this CEN/TS is limited initially to three years. After two years the members of CEN will be requested to submit their
comments, particularly on the question whether the CEN/TS can be converted into a European Standard.
CEN members are required to announce the existence of this CEN/TS in the same way as for an EN and to make the CEN/TS available
promptly at national level in an appropriate form. It is permissible to keep conflicting national standards in force (in parallel to the CEN/TS)
until the final decision about the possible conversion of the CEN/TS into an EN is reached.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2009 CEN All rights of exploitation in any form and by any means reserved Ref. No. CEN ISO/TS 17665-2:2009: E
worldwide for CEN national Members.

Contents Page
Foreword .3

Foreword
This document (CEN ISO/TS 17665-2:2009) has been prepared by Technical Committee ISO/TC 198
"Sterilization of health care products" of the International Organization for Standardization (ISO) and has been
taken over as CEN/TS ISO 17665-2:2009 by Technical Committee CEN/TC 204 “Sterilization of medical
devices” the secretariat of which is held by BSI.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to announce this Technical Specification: Austria, Belgium, Bulgaria, Cyprus, Czech
Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.
Endorsement notice
The text of ISO 17665-2:2009 has been approved by CEN as a CEN/TS ISO 17665-2:2009 without any
modification.
TECHNICAL ISO/TS
SPECIFICATION 17665-2
First edition
2009-01-15
Sterilization of health care products —
Moist heat —
Part 2:
Guidance on the application
of ISO 17665-1
Stérilisation des produits de santé — Chaleur humide —
Partie 2: Directives relatives à l'application de l'ISO 17665-1

Reference number
ISO/TS 17665-2:2009(E)
©
ISO 2009
ISO/TS 17665-2:2009(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.

©  ISO 2009
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2009 – All rights reserved

ISO/TS 17665-2:2009(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope . 1
2 Normative references . 1
3 Terms and definitions. 2
4 Quality management system elements .2
5 Sterilizing agent characterization . 2
5.1 Sterilizing agent . 2
5.2 Microbicidal effectiveness . 2
5.3 Material effects. 3
5.4 Environmental considerations . 3
6 Process and equipment characterization .3
6.1 Process . 3
6.2 Equipment . 6
7 Product definition . 7
8 Process definition. 8
9 Validation. 10
9.1 General. 10
9.2 Installation qualification (IQ) . 11
9.3 Operational qualification (OQ). 11
9.4 Performance qualification (PQ) . 13
9.5 Review and approval of the validation . 14
10 Routine monitoring and control . 15
11 Product release from sterilization. 16
12 Maintaining process effectiveness . 17
12.1 Demonstration of continued effectiveness. 17
12.2 Recalibration . 17
12.3 Maintenance of equipment . 17
12.4 Requalification . 17
12.5 Assessment of change. 18
Annex A (informative) Evaluation of a sterilization process primarily based on the measurement
of physical parameters. 19
Annex B (informative) Evaluation of a sterilization process primarily based on biological
inactivation and an accompanying mechanical air removal procedure . 27
Annex C (informative) Temperature and pressure of saturated steam for use in moist heat
sterilization . 30
Annex D (informative) Special considerations for health care settings . 32
Annex E (informative) Index of normative clauses/subclauses of ISO 17665-1 and cited references
or related guidance given in ISO 17665-1 and ISO/TS 17665-2. 41
Bibliography . 44

ISO/TS 17665-2:2009(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
In other circumstances, particularly when there is an urgent market requirement for such documents, a
technical committee may decide to publish other types of document:
⎯ an ISO Publicly Available Specification (ISO/PAS) represents an agreement between technical experts in
an ISO working group and is accepted for publication if it is approved by more than 50 % of the members
of the parent committee casting a vote;
⎯ an ISO Technical Specification (ISO/TS) represents an agreement between the members of a technical
committee and is accepted for publication if it is approved by 2/3 of the members of the committee casting
a vote.
An ISO/PAS or ISO/TS is reviewed after three years in order to decide whether it will be confirmed for a
further three years, revised to become an International Standard, or withdrawn. If the ISO/PAS or ISO/TS is
confirmed, it is reviewed again after a further three years, at which time it must either be transformed into an
International Standard or be withdrawn.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO/TS 17665-2 was prepared by Technical Committee ISO/TC 198, Sterilization of health care products.
ISO 17665 consists of the following parts, under the general title Sterilization of health care products — Moist
heat:
⎯ Part 1: Requirements for the development, validation and routine control of a sterilization process for
medical devices
⎯ Part 2: Guidance on the application of ISO 17665-1 [Technical Specification]

iv © ISO 2009 – All rights reserved

ISO/TS 17665-2:2009(E)
Introduction
The guidance given in this Technical Specification is not intended as a checklist for assessing compliance with
ISO 17665-1. This guidance is intended to assist in obtaining a uniform understanding and implementation of
ISO 17665-1 by providing explanations and acceptable methods for achieving compliance with specified
requirements. It highlights important aspects and provides examples. Methods other than those given in this
guidance may be used. However, the use of alternative methods has to be demonstrated to be effective in
achieving compliance with ISO 17665-1.
The main body of this document is applicable to all settings where moist heat sterilization is carried out. The
annexes to this guidance document also specify detailed means of implementing the requirements of
ISO 17665-1 and represent current best practices.
The numbering of the clauses in the main body of this Technical Specification corresponds to that in
ISO 17665-1.
Medical devices reprocessed in health care facilities include a wide variety of product with varying levels of
bioburden. Appropriate and thorough cleaning and, where necessary for safe handling, decontamination
processes are essential prior to presenting product for sterilization. Mixed product loads are common in
healthcare facilities with throughput volumes dictated by historical and predicted demand for sterile product.
Health care facilities do not normally specify sterilization processes for any individual medical device. Also, it
is impractical for health care facilities to determine bioburden on a medical device. It is important that specified
instruments be disassembled before decontamination and thoroughly inspected after completion of the
sterilization process. Reassembly and assessment of functionality are also needed. Therefore, the medical
[23]
device manufacturer's instructions (see ISO 17664 ) should be followed for all aspects of cleaning,
disinfection, packaging and sterilization. Many devices can be fully immersed and can be washed and
[19-22]
disinfected in automated equipment (see ISO 15883 ). For devices that cannot be fully immersed and
that cannot tolerate thermal decontamination, alternative methods of disinfection should be used to ensure
safe handling. Such procedures and policies should be in place to ensure that medical devices undergo
appropriate reprocessing. Particular attention needs to be paid to the drying and storage of sterile medical
[8] [9]
devices. Requirements for packaging of medical devices are covered in ISO 11607-1 and ISO 11607-2 .
If multiple sterilization cycles can lead to degradation and limit the useful life of a medical device, the
manufacturer will specify the number of reprocessing cycles that can normally be tolerated.
When selecting a medical device, priority should be given to properties such as ease of cleaning and
disassembly.
Additional guidance specific to health care is offered in Annex D of this Technical Specification.

TECHNICAL SPECIFICATION ISO/TS 17665-2:2009(E)

Sterilization of health care products — Moist heat —
Part 2:
Guidance on the application of ISO 17665-1
1 Scope
This Technical Specification provides general guidance on the development, validation and routine control of
moist heat sterilization processes and is intended to explain the requirements set forth in ISO 17665-1. The
guidance given in this Technical Specification is provided to promote good practice related to moist heat
sterilization processes and to assist those developing and validating a moist heat sterilization process
according to ISO 17665-1.
NOTE 1 The structure of the main body of this ISO Technical Specification (Clauses 1 to 12) corresponds to the
structure of ISO 17665-1, so that the guidance given under a particular clause or subclause of this part of ISO 17665
applies to the requirements given in the corresponding clause or subclause of ISO 17665-1. For example, guidance for
subclause 5.2 of ISO 17665-1:2006 is given in 5.2. This guidance is provided in addition to the guidance given in
ISO 17665-1:2006, Annex A. See also Annexe E.
NOTE 2 Special considerations specific to sterilization processes performed in health care facilities are given in
Annex D.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 17665-1:2006, Sterilization of health care products — Moist heat — Part 1: Requirements for the
development, validation and routine control of a sterilization process for medical devices
NOTE The normative references in ISO 17665-1 refer to published standards, the content of which should be used to
assist in demonstrating compliance to the clause in which they are cited. Some are required mainly for moist heat
sterilization in industry or for manufacturers of moist heat sterilizers and could go beyond typical practice for those
performing sterilization in health care facilities.
ISO 17665-1 specifies a number of methods and procedures that can be used to monitor sterilization
processes. The equipment required will normally be commercially available. A number of the normative
references cited describe the specification and test methods used by commercial suppliers to qualify their
products. The user of such products should ensure that purchased products comply with these standards, but
will not normally need to refer to the standards.
ISO 17665-1 specifies the use of packaging complying with ISO 11607-1 and ISO 11607-2. Healthcare
facilities should purchase packaging complying with these International Standards.
One method of process validation specified in ISO 17665-1 is based on the determination of bioburden. The
[6],[7]
ISO 11737 series specifies a number of microbiological methods used during this process. Health care
facilities would not normally utilize this approach for process validation.
ISO/TS 17665-2:2009(E)
3 Terms and definitions
For the purposes of this Technical Specification, the terms and definitions given in ISO 17665-1 and the
following apply.
3.1
tests for sterility
technical operation defined in pharmacopoeia performed on product following exposure to a sterilization
process
4 Quality management system elements
The guidance offered in Annex A of ISO 17665-1:2006 applies.
NOTE For additional considerations specific to health care facilities, see Clause D.2.
5 Sterilizing agent characterization
5.1 Sterilizing agent
5.1.1 Moist heat is water at elevated temperatures. Moist heat may be provided as saturated steam or can
be generated in situ by applying thermal energy to water already present in the product. Moisture acts as the
medium for transferring thermal energy to microorganisms.
5.1.2 Contaminants suspended in the sterilizing agent can be both toxic and corrosive and may generate a
barrier between the microorganism and the sterilizing agent. They originate from water, that is heated or
evaporated into steam or from contact between materials and the sterilizing agent during generation and
transport to the sterilizer (see Clause 6, Clause 7 and Annex A). If the level of contaminants in the sterilizing
agent can be affected by the quality of the feed water to the steam generation system, the feed water quality
should be specified.
5.2 Microbicidal effectiveness
The microbicidal activity of moist heat is based on the temperature and the duration of contact between water
molecules and microorganisms.
For the purpose of moist heat sterilization there are a number of acceptable time and temperature
combinations recognised by some pharmacopoeias. These combinations include but are not limited to those
listed in Table 1. All combinations listed are based on the concept of overkill with a safety factor that has been
established for saturated steam or water in contact with the microorganism. Superheated steam behaves
more like a dry gas and has a low microbicidal effectiveness compared with saturated steam. Superheated
steam can result from pressure reduction and/or thermodynamic compression of saturated steam. It can also
occur from the rehydration of parts of the sterilization load, particularly those parts containing natural fibres.
Superheated steam conditions can be minimized by engineering of the steam supply system, for example by:
a) having a series of pressure reduction stages from the supply pipe to the sterilizer chamber and ensuring
the pressure reduction ratio for each stage does not exceed 2:1;
b) ensuring steam velocity does not exceed 25 m/s;
c) ensuring materials made from natural fibres are pre-conditioned to a humidity greater than 40 % RH prior
to sterilization.
2 © ISO 2009 – All rights reserved

ISO/TS 17665-2:2009(E)
Table 1 — Examples of minimum temperatures and times established
for adequate levels of microbial lethality in sterilization processes
Temperature Time
°C min
121 15
126 10
134 3
5.3 Material effects
Material effects are generally limited to deformation and fracture caused by the temperatures and pressures of
the sterilizing agent.
5.4 Environmental considerations
Principles of an environmental management system can be applied to a moist heat sterilization process.
[11] [12]
ISO 14001 provides a specification for an environmental management system. ISO 14040 provides
guidance on designing a life cycle assessment study. The presence of noxious substances in the exhausts
[15]
from the sterilizer should be considered. Further guidance on this clause is given in E.3 of ISO 14937:— .
6 Process and equipment characterization
NOTE The purpose of this activity is to characterize the entire sterilization process and the equipment necessary to
deliver the sterilization process safely and reproducibly.
6.1 Process
6.1.1 General
A sterilization process should be specified for each product family and/or load configuration presented for
sterilization.
Process parameters should apply to the equipment used. They should be optimised to ensure that for defined
product families specified exposure conditions will be routinely obtained throughout the sterilizer chamber, and
the maximum temperatures and rates of change of process variable (e.g. temperature and pressure) will not
cause damage or degradation to the product.
The sterilization process specification should include all the process parameters that define the exposure
profile throughout the operating cycle. It should also include the ones used to verify reproducibility. The portion
of the operating cycle over which lethality is established should be identified, and the upper and lower limits of
each process parameter that can affect both this lethality and the performance of the medical device should
be defined.
Provision should be made to record data for judging the effectiveness and suitability of a routine sterilization
process. The accuracy of measurement should be related to the tolerances of the process parameters.
If it is proposed to use an existing sterilization process to treat a new medical device, the existing sterilization
process should be detailed and contain information and data sufficient to enable process definition
(see Clause 8) to be carried out for the proposed new medical device(s) or loading configuration. The
challenge identified for the new medical device or loading condition should be less than or equal to the
challenge from the existing sterilization load(s). For some product families, assurance that defined exposure
conditions will be reproduced might only be possible if the size of the sterilization load and the load
configuration have been clearly defined.
ISO/TS 17665-2:2009(E)
If biological indicators and chemical indicators are to be used, they should not replace routine monitoring,
measurement of process variables and any periodic tests.
Compatibility of a new medical device to the least favourable sterilization process conditions should be
assessed. Such assessment should include process parameter tolerances, uncertainties of measurement
associated with process parameters and the quality of the services (see Annex A).
Any restrictions on the size and mass of the sterilization load and its configuration should be identified and
included in the operating instructions.
The relationship between the temperature measured at the reference measuring point and the temperature
measured in the sterilization load should be known for each product family.
The performance of a medical device can be affected by contaminants on its surface. The contaminants and
maximum acceptable concentration(s) contained in each fluid coming into contact with the medical device
should be specified and included in the sterilization process specification. Some of the contaminants and their
maximum levels which need to be considered are identified in Annex A.
6.1.2 Saturated steam processes
Steam may be generated in, or admitted to a sterilizer chamber from an external source. Air in the sterilizer
chamber will be gradually removed by gravity displacement, active flow or by forced evacuation. The presence
of saturated steam will be obtained at the measurement location, e.g. the chamber discharge, when the
measured temperature is coincident with the temperature of saturated steam calculated from the pressure
(see Annex C). Both temperature and pressure are process variables, and the point of temperature
measurement is defined as the reference measuring point.
If variations in process parameters and/or the amount of non-condensable gas remaining in the sterilizer
chamber at the end of air removal can result in an ineffective process, the sterilizer manufacturer or
designated person (see A.4.2 in ISO 17665-1:2006) should provide adequate information to the user and
should include:
⎯ the upper and lower limits for each process parameter, and the method used for air removal;
⎯ sources of non-condensable gas;
⎯ test methods, test frequency and acceptance criteria for sterilization process evaluation.
The removal of air from the sterilizer chamber by either active flow or by gravity displacement is only
predictable for simple solid medical devices. Air removal is unpredictable for medical devices such as
instruments containing lumens, heavy solid masses and instruments and textiles contained within their
primary packaging. For such medical devices, an operating cycle that employs forced or dynamic air removal
should be used. An example is one that employs a number of vacuum and/or steam pulses to serially dilute
the air from the sterilizer chamber and medical device(s). During each pulse, steam will move into and out of
the medical device and the condensing steam will re-evaporate and cause a dynamic ‘scouring’ of the residual
air contained in packages, crevices and lumens. The number of pulses, the upper and lower pressures
associated with each pulse, the rate of change of pressure and temperature, and the interval of time between
each change, are process variables and will play a part in effecting air removal. When assigning the suitability
of a product family to a sterilization process, the combination of these pressures and temperature changes,
the rates of change, and the duration of each change should be considered.
Whenever the measured temperature exceeds the theoretical temperature calculated from measured
pressure as described in Annex C, superheated steam may be present. The presence of superheated steam
may be detrimental to the medical device and or its packaging and may compromise the sterilization process.
Effective air removal from lumens, porous loads and other complex designs incorporating enclosed spaces is
difficult. The physical conditions required for effective air removal are influenced by length, width and shape of
lumen, wall thickness, material of the product, mass, density, the packaging system and other items in the
same packaging system. A sterilization process that removes air from the sterilizer chamber to a low level
4 © ISO 2009 – All rights reserved

ISO/TS 17665-2:2009(E)
may fail to remove sufficient air from a lumen to permit steam penetration. Dalton's law states that the total
pressure in an enclosed space is equal to the sum of the partial pressures of the individual gases present. In
theory the temperature in a sterilizer chamber containing a mixture of steam and residual air will be lower than
the calculated temperature derived from the pressure in accordance with steam table values (see Annex C).
However there is evidence to show that an amount of residual air sufficient to cause a process failure in a
sterilization load may only depress steam temperature by as little as 0,01 °C. As a consequence the
differences between the temperature measured at the temperature measurement point and the temperature
calculated from the sterilizer chamber pressure using steam table values (see Annex C) may not be adequate
to detect the small volumes of air which could concentrate in lumens or enclosed spaces and prevent steam
penetration. Under such circumstances adequate air removal and steam penetration should be predicted from
data obtained from a steam penetration test and/or a process monitoring device.
A steam penetration test is designed for a specified product family(ies) and is used to check that the amount
of non-condensable gas remaining in the sterilizer chamber at the commencement of the plateau period will
not obstruct the presence of saturated steam on the surfaces of the medical device for the duration of the
holding time. The efficiency of the air removal system, air leakage into the sterilizer chamber and non-
condensable gas carried by the steam contribute to this amount. Air leakage into the sterilizer chamber and
non condensable gases carried by the steam can be checked by tests (see for example, Annex A and EN
[25]
285 ). The total presence of non-condensable gas is monitored by the steam penetration test.
A steam penetration test may be based upon temperature measurement, biological indicators or chemical
indicators, as applicable. The test system should provide a challenge representative of the product family(ies)
it represents. A number of steam penetration and air removal test devices are available. Performance
[55] [56] [57]
requirements for chemical indicators can be found in ISO 11140-3 , ISO 11140-4 , ISO 11140-5 ,
1) [25]
ISO 11140-6 and EN 285 . Guidance on the selection and use of chemical indicators is given in
[18] [4]
ISO 15882 . Requirements for biological indicators are found in ISO 11138-3 . Guidance on the selection
[13]
and use of biological indicators is found in ISO 14161 .
A reference load can consist of a single medical device type, medical devices from different product families
or medical devices assigned to different product families but assembled into a single package. For any
reference product or medical device, difficulty in air removal and the challenge to the sterilization process
should not be less than that for any medical device in the product family(ies) assigned to the sterilization
process (see also Annex A and Annex B).
If it is proposed to use a process challenge device (e.g. an air detector or other monitoring device) to
represent a product family(ies), then the validity of the device when exposed to the sterilization process should
be established by the process challenge device manufacturer, sterilizer manufacturer or designated person
(see A.4.2 of ISO 17665-1:2006).
6.1.3 Contained product processes
A product may be heated in a water immersion cycle, a water spray cycle, a cycle with an air and steam
mixture, a cycle with steam and gravity displacement, or a cycle with forced air removal. Air and steam
mixtures are often used to prevent distortion or fracture of the sterilized container caused by the internal
pressure generated from heating both the water-based solution and air in any sealed container.
The energy required to heat up a sterilization load to the defined sterilization temperature depends on the
product family, the size of a sterilization load and its initial temperature. Heat transfer will depend on the
heating medium, its contact with the product container, the material of the container and container support
system, and the temperature difference at the heat transfer site. The type of product family and the load
configuration will have a major influence on temperature differences between containers. These differences
may be minimized by increasing the flow and distribution of the heating medium by forced circulation. Mass
flow and homogeneity of the heat transfer medium throughout the sterilizer chamber can be verified by
process variables such as fan speed, circulation pressure and flow. Temperature of the heat transfer medium
at the outlet should be identified as a process variable. If steam is used, the temperature of the steam
environment should also be handled as a process variable. Consideration may need to be given to ensure the

[27]
1) ISO 11140-6 is under development and is based on EN 867-5 .
ISO/TS 17665-2:2009(E)
heat transfer medium is pyrogen free and free of chemical impurities that may cause spotting on packaging. In
addition, the heat transfer medium may need to be sterile during cooling and the period of the operating cycle
for which lethality is claimed.
The temperature distribution within the product container will depend on the shape of the container, viscosity
of the product, conduction through the container wall and product, and convection within the product. Large
product containers will need longer times to heat up and cool down, which could restrict the size of container
that can be used for products sensitive to prolonged exposure.
During the sterilization process, the locations of the product containers exhibiting the highest and lowest
temperatures during the heating phase and the highest and lowest temperatures during the cooling phase in
the sterilization load should be identified. The temperatures measured in these locations should be treated as
process variables; however, if either location cannot be reproduced, a statistical approach may need to be
used to ensure the specified lethality is consistently attained while maintaining product integrity.
6.2 Equipment
NOTE For additional considerations specific to health care facilities, see D.3.2.
6.2.1 Regional and national standards for sterilizing equipment have been published (e.g. EN 285), which
recommend materials that can be used in the construction of a sterilizer. Materials used by a manufacturer for
the construction of a sterilizer can be based on the sterilization process delivered by the sterilizer and the
product family(ies) that will be sterilized. The materials chosen should minimize corrosion and any
contaminant that can be released during routine operation. Steam, heat transfer, fluids or air used to
pressurize the sterilizer chamber can carry corrosive and toxic agents. These should be identified and
maximum permissible levels specified (see Annex A). Protection of materials by filming amines such as
hydrazine should not be used as an alternative to the correct choice of material and the control of corrosive
contaminants.
It is preferable that sterilization records be established independent from the automatic controller and
indicating instruments. A system that combines recording, control and indication may lead to an ineffective
sterilization process being interpreted as effective. Independent recorders are characterized by separate
measurement, data processing and printing of values. Interchange of informative data between the recorder
and the controller for other purposes is not excluded.
An air detector may be fitted to a sterilizer that uses vacuum and steam pulsing to remove air during the air
removal stage of a saturated steam sterilization process. It is used to predict whether non-condensable gas
remaining in the sterilizer chamber at the commencement of the plateau period could accumulate in parts of
the sterilization load (e.g. lumens) and cause a failure of the sterilization process in these parts. The setting of
the air detector is based on the defined process parameters and the product family(ies) that the sterilization
process is designed to process. Non-condensable gas identified by an air detector may also contain gas
released when a product or its packaging is heated. Air detector tests are specified in Annex A and EN 285.
6.2.2 The specification for the equipment should include sufficient information to perform a process
definition for a new product or loading configuration. (See Clause 8).
6.2.3 A sterilization process delivered in accordance with its specification is dependant upon the quality of
the services provided. During maximum demand, pressure measured at the connection to the sterilizer for
each fluid, gas or steam service should not fall below the minimum specified by the sterilizer manufacturer.
For example, the efficiency of a water ring vacuum pump and a heat exchanger deteriorates with falling water
pressure and rising water temperature. Microbial contamination can occur if air entering the sterilizer chamber
contains particles greater than 0,2 µm. If services are provided by another party, recommendations from the
sterilizer manufacturer should be followed and conformity confirmed.
Local regulations for environmental considerations could govern the discharge of high temperature effluent
from the sterilizer chamber into the public sewer system, the leakage of the materials used to generate the
sterilizing agent, the particulates released from the product and/or packaging during sterilization, and the
volume of water used during the process.
6 © ISO 2009 – All rights reserved

ISO/TS 17665-2:2009(E)
[24]
Safety is part of equipment design and operation. Reference should be made to IEC 61010-2-040 and
national regulations.
6.2.4 Systems such as containers, shelving, racks and carriers designed to support and/or contain the
medical device should not unduly restrict uniform steam distribution, circulation of heat transfer fluid, removal
of residual air, drainage of condensate or drainage of water. The system should also prevent damage to the
medical device and/or its packaging and retain the integrity of the sterilization load.
6.2.5 No guidance offered.
6.2.6 Software design should be structured. Guidance is given in Good Automated Manufacturing Practice,
[39]
Guide For Validation Of Automated Systems In Pharmaceutical Manufacturing (GAMP 4) .
7 Product definition
NOTE For additional considerations specific to health care facilities, see Clause D.4.
7.1 During product design, consideration should be given to the procedures for disassembly (if appropriate),
cleaning, disinfection, inspection and sterilization.
Guidance and methods for the cleaning and disinfection of medical devices prior to sterilization are addressed
[19-22]
in the ISO 15883 series of standards . Information to be provided by a medical device manufacturer for
[25]
the reprocessing of a medical device is given in ISO 17664 .
7.2 The major function of a package is to ensure that the medical device remains sterile until opened for
use. Packaging should withstand the stresses that occur during a sterilization process, remain secure, and
should not have a negative effect on the quality of the medical device (for example, by generating particles).
[8],[9]
Packaging for a medical device sterilized by saturated steam should meet the requirements of ISO 11607 .
For non-permeable packaging (e.g. vials, ampoules, flexible pouches), the material and design should permit
heat transfer to the product and, if a closure is fitted, it should remain secure and sealed.
Secondary packaging should protect the product during customary handling and distribution. If secondary
packaging is exposed to the sterilization process it should retain its ability to protect the product and should
not be adversely affected by the sterilization process.
If, at the end of a sterilizat
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

CEN ISO/TS 17665-2:2009는 의료 제품의 멸균을 위한 지침으로서, 습열 멸균 과정에 대한 일반적인 안내를 제공한다. 이 표준은 ISO 17665-1에서 설정된 요구사항을 설명하기 위해 고안되었으며, 멸균 과정의 개발, 검증 및 일상 관리를 위한 방향성을 제시한다. 이 문서의 주요 강점은 습열 멸균 프로세스를 개발하고 검증하는 데 도움이 되는 최선의 관행을 촉진한다는 점이다. 사용자는 ISO 17665-1에 따라 멸균 프로세스를 효과적으로 구축하고 운영할 수 있는 프레임워크를 제공받는다. 또한, CEN ISO/TS 17665-2:2009는 멸균 과정의 일관성을 보장하기 위한 검증 방법에 대한 명확하고 실용적인 조언을 포함하고 있어, 관련 분야의 전문가들이 따라야 할 기준을 제시한다. 이 표준은 의료 분야에서 멸균의 깊은 이해와 엄격한 실행을 필요로 하는 모든 전문가들에게 필수적이며, 현대의 건강 관리 제품을 안전하게 제공하기 위한 기초 자료로서 그 중요성이 크다. 습열 멸균의 복잡성을 이해하면, 의료 제품의 안전성과 효과성을 극대화하는 데 기여할 수 있으며, 따라서 CEN ISO/TS 17665-2:2009는 그 적용 가능성 면에서 매우 중요한 자료라고 할 수 있다.

Die Norm CEN ISO/TS 17665-2:2009 bietet umfassende Leitlinien zur Entwicklung, Validierung und routinemäßigen Kontrolle von Prozessen zur Sterilisation durch feuchte Hitze. Sie ist eine wertvolle Ergänzung zur ISO 17665-1, da sie die darin enthaltenen Anforderungen präzisiert und erklärt. Ein klarer Vorteil dieser Norm liegt in ihrem breiten Anwendungsbereich. Sie richtet sich nicht nur an Hersteller von Gesundheitsprodukten, sondern auch an Einrichtungen, die Sterilisationsverfahren durchführen. Dies fördert ein einheitliches Verständnis von den Prozessen und deren Validierung, was letztlich die Qualität und Sicherheit der medizinischen Produkte erheblich verbessert. Die Stärke der CEN ISO/TS 17665-2:2009 liegt in der Förderung bewährter Praktiken im Bereich der feuchten Hitze-Sterilisation. Durch die Bereitstellung präziser Leitlinien leistet die Norm einen wichtigen Beitrag zur Standardisierung innerhalb der Branche. Diese Standardisierung ist entscheidend, um die Konsistenz und Zuverlässigkeit der Sterilisationsprozesse zu gewährleisten. Darüber hinaus ist die Norm relevant, da sie auf die notwendigen Anforderungen eingeht, die für die Entwicklung und Validierung eines feuchten Hitze-Sterilisationsprozesses erforderlich sind. Dies unterstützt nicht nur die Qualitätskontrolle, sondern trägt auch dazu bei, die regulatorischen Anforderungen zu erfüllen, die in vielen Ländern für die Sterilisation von Gesundheitsprodukten gelten. Zusammengefasst bietet die CEN ISO/TS 17665-2:2009 eine wertvolle Ressource für alle, die in der Medizinprodukteindustrie tätig sind. Ihre klaren und detaillierten Leitlinien zur feuchten Hitze-Sterilisation sind ein unverzichtbares Instrument für die Gewährleistung von Qualität und Sicherheit in der Herstellung und Anwendung von Gesundheitsprodukten.

Le document de normalisation SIST-TS CEN ISO/TS 17665-2:2009 offre une approche claire et structurée pour la stérilisation des produits de santé par chaleur humide, se concentrant particulièrement sur les exigences énoncées dans la norme ISO 17665-1. Ce texte est fondamental pour les professionnels du secteur, car il fournit des lignes directrices générales relatives à la conception, la validation et le contrôle régulier des processus de stérilisation par chaleur humide. L'une des forces majeures de cette norme est sa capacité à encourager de bonnes pratiques dans le domaine de la stérilisation. En détaillant les principes fondamentaux et en offrant des recommandations pratiques, la norme joue un rôle essentiel dans l'harmonisation des procédures de stérilisation, améliorant ainsi la sécurité et l'efficacité des produits de santé traités. De plus, l'application des directives données dans l'ISO/TS 17665-2:2009 permet aux établissements de santé de renforcer la conformité avec les exigences légales et réglementaires. La pertinence de cette norme ne peut être sous-estimée, surtout dans un contexte où la stérilisation appropriée est vitale pour prévenir les infections nosocomiales et assurer la sécurité des patients. En fournissant un cadre de référence solide, le SIST-TS CEN ISO/TS 17665-2:2009 répond à un besoin essentiel d'information et de soutien pour ceux qui développent et valident des processus de stérilisation par chaleur humide, facilitant ainsi leur mise en œuvre et leur contrôle. En résumé, cette norme se positionne comme un outil indispensable pour toutes les entités engagées dans la stérilisation des produits de santé, rendant ainsi la stérilisation par chaleur humide non seulement plus efficace mais également plus conforme aux standards de qualité.

The standard CEN ISO/TS 17665-2:2009 is a crucial document within the field of sterilization of health care products, focusing specifically on the application and guidance of moist heat sterilization processes as detailed in ISO 17665-1. Its relevance is underscored by the increasing need for robust sterilization methodologies in healthcare settings, where the safety and efficacy of medical devices and products are paramount. One of the significant strengths of ISO/TS 17665-2:2009 lies in its comprehensive guidance on the development, validation, and routine control of moist heat sterilization processes. It provides clear frameworks that practitioners can rely on to ensure their sterilization processes meet rigorous safety standards. By outlining best practices, the standard assists healthcare professionals in navigating the complexities of sterilization, thereby promoting an elevated level of quality and consistency in healthcare delivery. Furthermore, this standard emphasizes the importance of validating sterilization processes-an essential step to confirm that the intended lethality is achieved. The guidance on routine control contained within the document further supports organizations in maintaining compliance with regulatory requirements and industry standards. The systematic approach encouraged by ISO/TS 17665-2:2009 fosters a culture of continuous improvement and adherence to the best practices in moist heat sterilization. In terms of scope, the specification is designed to serve a broad audience, including those responsible for developing, implementing, and auditing sterilization protocols. By simplifying the application of the requirements set forth in ISO 17665-1, it aligns the efforts of various stakeholders in the healthcare sector towards a unified approach to moist heat sterilization. In conclusion, CEN ISO/TS 17665-2:2009 provides essential guidance that enhances the field of health care product sterilization. Its strengths in promoting good practices, ensuring validation, and aiding in the routine control of moist heat sterilization processes make it an invaluable resource for improving health care product safety.

CEN ISO/TS 17665-2:2009の標準は、医療機器の滅菌に関するガイダンスを提供するもので、特に湿熱滅菌プロセスの開発、バリデーション、日常管理に焦点を当てています。この標準は、ISO 17665-1に示された要件の理解を助け、湿熱滅菌プロセスに関して良好な実践を促進することを目的としています。 この文書の重要な強みは、湿熱滅菌の具体的な手法やプロセスについての詳細なガイダンスが含まれている点です。特に、ISO 17665-1の要件に従うことを前提とした内容で構成されており、そのため、開発者やバリデーションを行う専門家にとって非常に有益です。このガイダンスは、湿熱滅菌プロセスの整備と継続的な管理に必要な基準を具体的に示しており、実践的なアプローチを提供します。 CEN ISO/TS 17665-2:2009は、医療製品の安全性を確保するための不可欠なリソースであり、特に湿熱滅菌に関する最新の技術や知識を求める業界関係者にとって、その関連性は非常に高いと言えます。また、この標準は、国際的に認められた基準であるため、各国の規制当局にも対応しやすく、国際的な品質保証の向上にも寄与するでしょう。 結果として、CEN ISO/TS 17665-2:2009の標準は、湿熱滅菌プロセスにおける信頼性、整合性、効率性を高めるための重要なツールといえます。その内容は、医療機関や滅菌施設が最高の衛生基準を維持するための強力なサポートを提供することに貢献しています。