Standard Specification for Thermistor Sensors for General Purpose and Laboratory Temperature Measurements

ABSTRACT
This specification covers the classification, testing, and corresponding requirements for negative-temperature-coefficient thermistor-type sensors intended to be used for clinical laboratory temperature measurements or control, or both, within a specified range. This specification also covers the detailed requirements for ASTM designated sensors.
SCOPE
1.1 This specification covers the general requirements for Negative Temperature Coefficient (NTC) thermistor-type sensors intended to be used for laboratory temperature measurements or control, or both, within the range from −10 °C to 105 °C.  
1.2 This specification also covers the detailed requirements for ASTM designated sensors.  
1.3 This specification also covers the requirements for general purpose, Negative Temperature Coefficient (NTC) thermistor-type sensors intended for use with Digital Contact Thermometers (also known as Digital Thermometers) within the range from –50 °C to +150 °C.  
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
30-Apr-2020
Technical Committee
E20 - Temperature Measurement

Relations

Effective Date
01-Dec-2023
Effective Date
01-Nov-2023
Effective Date
01-Sep-2019
Effective Date
01-Apr-2018
Effective Date
01-Nov-2016
Effective Date
01-May-2016
Effective Date
01-May-2013
Effective Date
01-Jul-2012
Effective Date
01-May-2012
Effective Date
01-May-2011
Effective Date
01-Nov-2010
Effective Date
01-Nov-2010
Effective Date
01-May-2010
Effective Date
01-Nov-2009
Effective Date
01-May-2009

Overview

ASTM E879-20 is the standard specification for thermistor sensors used in general purpose and laboratory temperature measurements. Issued by ASTM International, this standard provides detailed classification, performance, and testing requirements for negative temperature coefficient (NTC) thermistor-type sensors. These sensors are critical for clinical laboratory applications and various general temperature measurement or control tasks. The document outlines specifications for both general purpose thermistor sensors and those designated for use with digital contact thermometers over defined temperature ranges, promoting measurement accuracy and reliability.

Key Topics

  • Sensor Classification
    Thermistor sensors under this standard are comprehensively classified by:

    • Type designation (e.g., silicone coated, epoxy coated, stainless steel housed)
    • Operating temperature range (e.g., −10 °C to 105 °C; –50 °C to +150 °C)
    • Accuracy class (e.g., ±0.02 °C to ±0.5 °C)
    • Calibration type (interchangeable or non-interchangeable)
  • Performance Requirements

    • Zero-power resistance versus temperature relationship
      Sensors must maintain specific resistance values at various calibration points, ensuring accuracy within defined tolerances.
    • Thermal behavior
      Specifications address dissipation constant and response time to guarantee sensor stability and reliability under typical operating conditions.
    • Environmental durability
      Sensors must function across a range of temperatures and relative humidity, and withstand storage as well as thermal shocks.
  • Testing and Quality Assurance

    • Procedures for verifying zero-power resistance and environmental performance
    • Qualification and routine manufacturing screening required for conformity
    • Measurement uncertainties must be traceable and controlled to maintain measurement system integrity

Applications

The ASTM E879-20 standard serves a broad array of applications, underpinning the accuracy and reliability of temperature measurement in both clinical and industrial environments:

  • General Laboratory Measurements
    Used in research laboratories to ensure precise monitoring and control of experimental conditions.
  • Clinical Diagnostic Equipment
    Essential for temperature monitoring in devices like blood analyzers and incubators, supporting patient safety and process integrity.
  • Digital Contact Thermometers
    Specifies requirements for thermistor sensors used in field and portable digital thermometry, benefitting quality control, HVAC, and process monitoring.
  • Calibration Baths and Reference Thermometers
    Enables traceable calibration of instrumentation, forming a foundation for ISO, NIST, or internationally traceable measurements.

Related Standards

ASTM E879-20 references and aligns with several other recognized standards to ensure consistency and interoperability:

  • ASTM E344 - Terminology relating to thermometry and hydrometry
  • ASTM E563 - Preparation and use of ice-point bath as reference temperature
  • ASTM E1502 & E1750 - Use of fixed-point cells for reference temperatures
  • ASTM E2488 - Preparation and evaluation of liquid baths for temperature calibration
  • ASTM F29 - Dumet wire in glass-to-metal seal applications
  • ANSI/NCSL Z540.3 - Calibration of measuring and test equipment
  • NIST Monograph 126 & Special Publications - Platinum resistance thermometry and calibration

These related standards support measurement traceability, calibration consistency, and international compatibility.


Keywords: ASTM E879-20, thermistor sensors, laboratory temperature measurement, NTC thermistor, digital contact thermometer, temperature calibration, sensor classification, measurement traceability, clinical laboratory, standard specification.

Buy Documents

Technical specification

ASTM E879-20 - Standard Specification for Thermistor Sensors for General Purpose and Laboratory Temperature Measurements

English language (21 pages)
sale 15% off
sale 15% off
Technical specification

REDLINE ASTM E879-20 - Standard Specification for Thermistor Sensors for General Purpose and Laboratory Temperature Measurements

English language (21 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

BSMI (Bureau of Standards, Metrology and Inspection)

Taiwan's standards and inspection authority.

TAF Taiwan Verified

Sponsored listings

Frequently Asked Questions

ASTM E879-20 is a technical specification published by ASTM International. Its full title is "Standard Specification for Thermistor Sensors for General Purpose and Laboratory Temperature Measurements". This standard covers: ABSTRACT This specification covers the classification, testing, and corresponding requirements for negative-temperature-coefficient thermistor-type sensors intended to be used for clinical laboratory temperature measurements or control, or both, within a specified range. This specification also covers the detailed requirements for ASTM designated sensors. SCOPE 1.1 This specification covers the general requirements for Negative Temperature Coefficient (NTC) thermistor-type sensors intended to be used for laboratory temperature measurements or control, or both, within the range from −10 °C to 105 °C. 1.2 This specification also covers the detailed requirements for ASTM designated sensors. 1.3 This specification also covers the requirements for general purpose, Negative Temperature Coefficient (NTC) thermistor-type sensors intended for use with Digital Contact Thermometers (also known as Digital Thermometers) within the range from –50 °C to +150 °C. 1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ABSTRACT This specification covers the classification, testing, and corresponding requirements for negative-temperature-coefficient thermistor-type sensors intended to be used for clinical laboratory temperature measurements or control, or both, within a specified range. This specification also covers the detailed requirements for ASTM designated sensors. SCOPE 1.1 This specification covers the general requirements for Negative Temperature Coefficient (NTC) thermistor-type sensors intended to be used for laboratory temperature measurements or control, or both, within the range from −10 °C to 105 °C. 1.2 This specification also covers the detailed requirements for ASTM designated sensors. 1.3 This specification also covers the requirements for general purpose, Negative Temperature Coefficient (NTC) thermistor-type sensors intended for use with Digital Contact Thermometers (also known as Digital Thermometers) within the range from –50 °C to +150 °C. 1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM E879-20 is classified under the following ICS (International Classification for Standards) categories: 17.200.20 - Temperature-measuring instruments. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM E879-20 has the following relationships with other standards: It is inter standard links to ASTM E344-23, ASTM E1750-23, ASTM E344-19, ASTM E344-18, ASTM E344-16, ASTM E1750-10(2016), ASTM E344-13, ASTM F29-97(2012), ASTM E344-12, ASTM E563-11, ASTM E344-10, ASTM E1502-10, ASTM E1750-10, ASTM E2488-09, ASTM F29-97(2009). Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM E879-20 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:E879 −20
Standard Specification for
Thermistor Sensors for General Purpose and Laboratory
Temperature Measurements
This standard is issued under the fixed designation E879; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope E563Practice for Preparation and Use of an Ice-Point Bath
as a Reference Temperature
1.1 This specification covers the general requirements for
E1502Guide for Use of Fixed-Point Cells for Reference
Negative Temperature Coefficient (NTC) thermistor-type sen-
Temperatures
sors intended to be used for laboratory temperature measure-
E1750Guide for Use of Water Triple Point Cells
ments or control, or both, within the range from−10 °C to
E2488Guide for the Preparation and Evaluation of Liquid
105 °C.
Baths Used for Temperature Calibration by Comparison
1.2 This specification also covers the detailed requirements
F29Specification for Dumet Wire for Glass-to-Metal Seal
for ASTM designated sensors.
Applications
1.3 This specification also covers the requirements for 2.2 NIST Documents:
general purpose, Negative Temperature Coefficient (NTC)
NIST Monograph 126Platinum Resistance Thermometry
thermistor-type sensors intended for use with Digital Contact NIST Special Publication 250-22Platinum Resistance Ther-
Thermometers (also known as Digital Thermometers) within
mometer Calibration
the range from –50 °C to +150 °C. NIST Technical Note 1265Guidelines for Realizing the
International Temperature Scale of 1990
1.4 The values stated in SI units are to be regarded as the
2.3 Other Documents:
standard. The values given in parentheses are for information
JCGM100:2008Evaluationofmeasurementdata–Guideto
only.
the expression of uncertainty in measurement
1.5 This standard does not purport to address all of the
ANSI/NCSL Z540.3-2006 (R2013)Requirements for the
safety concerns, if any, associated with its use. It is the
Calibration of Measuring and Test Equipment
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
3. Terminology
mine the applicability of regulatory limitations prior to use.
3.1 Definitions—The definitions given in Terminology
1.6 This international standard was developed in accor-
E344 shall apply to this specification.
dance with internationally recognized principles on standard-
3.2 Definitions of Terms Specific to This Standard:
ization established in the Decision on Principles for the
3.2.1 dissipation constant, δ,n—the ratio of the change in
Development of International Standards, Guides and Recom-
energy dissipated per unit time (power) in a thermistor,
mendations issued by the World Trade Organization Technical
∆P= P − P , to the resultant temperature change of the
2 1
Barriers to Trade (TBT) Committee.
thermistor, ∆T= T − T .
2 1
2. Referenced Documents
∆P
2 δ 5 (1)
2.1 ASTM Standards: ∆T
E344Terminology Relating to Thermometry and Hydrom-
The dimensions of the dissipation constant are watts per
etry
kelvin (W/K). Note that many NTC thermistors are very
small devices, and as such, the dissipation constant dimen-
This specification is under the jurisdiction of ASTM Committee E20 on
sions are frequently expressed in terms of milliwatts per kel-
Temperature Measurement and is the direct responsibility of Subcommittee E20.03
vin (mW/K).
on Resistance Thermometers.
CurrenteditionapprovedMay1,2020.PublishedJuly2020.Originallyapproved
in 1982. Last previous edition approved in 2012 as E879–12. DOI: 10.1520/
E0879-20. Available from National Institute of Standards and Technology (NIST), 100
For referenced ASTM standards, visit the ASTM website, www.astm.org, or Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Available from the BIPM, Sevres, France, http:// www.bipm.org.
Standards volume information, refer to the standard’s Document Summary page on Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
the ASTM website. 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright ©ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA19428-2959. United States
E879−20
For this specification, T is in the range from 20 °C to 38 °C 4.2.1 Type Designation—Thetypedesignationcodeshallbe
and ∆T=10 °C. a letter symbol to indicate the design and construction of the
thermistor sensor followed by a number to indicate the subset
3.2.2 dumet, n—round,copper-coated42%nickel-ironwire
of specific features as shown on the detailed specification
intended primarily for sealing to soft glass. Also known as
drawingwhenevertherearedesignoptions.(SeeFig.2through
CuNiFe in some communities; see Specification F29 for
Fig. 10).
additional information.
4.2.1.1 Type S—Silicone rubber-coated glass probe with
3.2.3 insulation resistance, dc, n—the resistance at a speci-
tinned Dumet extension leads (see Fig. 2).
fied direct-current voltage between the insulated leads of a
4.2.1.2 Type E—Epoxy-coated glass probe with silver-
thermistor sensor and the metallic enclosure of the sensor, if
plated copper extension leads (see Fig. 3).
such an enclosure is present, or else between the sensor leads
4.2.1.3 Type G—General purpose four-wire sensor in stain-
and a conductive medium in which the sensor is immersed.
less steel housing (see Fig. 4).
3.2.4 qualification test, n—aseriesoftestsconductedbythe
4.2.1.4 Type H—General purpose two-wire sensor in stain-
procuring agency or an agent thereof to determine confor-
less steel housing (see Fig. 5).
mance of thermistor sensors to the requirements of a
4.2.1.5 Type V—Interchangeable sensor enclosed in 1.2 mm
specification, normally for the development of a qualified
vinyl tube (see Fig. 6).
products list under the specification.
4.2.1.6 Type W—Non-interchangeable sensor enclosed in
3.2.5 response time, n—the time required for a sensor to 0.9 mm vinyl tube (see Fig. 7).
change a specified percentage of the total difference between 4.2.1.7 Type P—Interchangeable sensor with flexible cable
its initial and final temperatures as determined from zero- and sealed plastic tip (see Fig. 8).
power resistances when the sensor is subjected to a step 4.2.1.8 Type J—Interchangeablesensorenclosedinstainless
function change in temperature. steel housing (see Fig. 9).
4.2.1.9 Type K—Interchangeable sensor enclosed in stain-
3.2.6 timeconstant,n—the63.2%responsetimeofasensor
less steel housing with threaded pipe fitting (see Fig. 10).
that exhibits a single-exponential response.
4.3 Operating Temperature Range—Theoperatingtempera-
3.2.7 zero-power resistance, n—the dc resistance of a
ture range shall be designated by a letter symbol (see Table 4).
device, at a specified temperature, calculated for zero-power.
3.2.7.1 Discussion—Accurate zero-power resistance is ob-
4.4 Accuracy Class—Theaccuracyclassshallbedesignated
tained by extrapolating to zero-power the resistance values
by a single-digit number (see Table 3).
obtained from measurements at three or more levels of power
4.5 Calibration Type—The type of calibration required for
with the sensor immersed in a constant temperature medium.
each unit shall be designated by a letter symbol. The letter I
For the purpose of this specification, this is obtained from
shall be used to denote units that are interchangeable with
measurements at a single power level that is adjusted or
respect to a single resistance-temperature relationship. The
otherwise determined to be not greater than one-fifth the
letter N shall be used to denote non-interchangeable units for
product of the dissipation constant specified in Table 1 (see
which resistance-temperature information must be furnished
3.2.1 and 7.3) and the appropriate tolerance requirement of
for each unit. For Calibration Type N sensors, serial number
Table 3. Accordingly, the applied power levels required to
identification must be provided.
satisfythezero-powerresistancecriterionasdefinedabovecan
4.6 Zero-power Resistance at 25 °C—Eachoftheindividual
beaslittleas2µW(foraTypeWsensorwithaccuracyclass1)
thermistor sensor types shall have nominal zero-power resis-
to as much as 600 µW (for Types J, K or P sensors with
tance values at 25 °C as specified in Table 1 and as shown on
accuracy class 6). The known excitation current or voltage of
the individual detailed specification drawings. Each nominal
themeasurementsystemalongwiththeresistanceofthesensor
resistance value specified shall also have an associated resis-
at the selected temperature (from the nominal resistance and
tanceratio-versus-temperaturevaluethatischaracteristicofthe
ratiovaluesgiveninTable2)canbeusedtocomputetheactual
thermistor material formulation employed in the sensor. The
applied power to the unit under test (UUT) and to determine if
resistanceratio-versus-temperaturedatafortheselectedtherm-
this power level meets the zero-power criterion.When making
istor material formulations are shown in Table 2. Other
repeat measurements over a specified time period to determine
nominal resistance values at 25 °C, or at another reference
the stability of a sensor, the power level shall be kept constant.
temperature, may be agreed upon by the manufacturer and end
4. Classification
user provided that this information is clearly communicated in
4.1 Thermistorsensorscoveredbythisspecificationshallbe all appropriate documentation; and, that the unique identifica-
classified with a unique identification code that includes the tion code (see Fig. 4.1) is not used under such circumstances.
ASTMdetailedspecificationnumberfollowedbyadescriptive
4.7 Terminations—The detailed specification drawings (see
typedesignationcode,anoperatingtemperaturerangecode,an
Fig. 2 through Fig. 10) show the sensors with the most
accuracy class code, and a calibration type code. (See Fig. 1.)
common type of termination for that type designation. All
4.2 ASTM Specification Number—The ASTM specification sensors,regardlessoftypedesignation,maybeterminatedwith
number specifies uniquely the design and construction of the any suitable electrical connector (as agreed upon by manufac-
sensor including the type designation if more than one type turer and end user) such that it allows the sensor to interface
appears in the same specification. withaspecificthermometersystem.Itisrecommendedthatthe
E879−20
TABLE 1 Specification for ASTM Laboratory Thermistor Sensors
E879 Type S E879 Type E E879 Type G E879 Type H E879 Type V E879 Type W E879 Type P E879 Type J E879 Type K
Non-
4-Wire Non- 2-Wire Non- Interchangeable Interchangeable Interchangeable
Silicone Rubber interchangeable Interchangeable
Epoxy Coated interchangeable interchangeable Sensor Enclosed Sensor with flex- Sensor in S.S.
Description Coated Glass Sensor Enclosedin Sensor in S.S.
Glass Probe Sensor in S.S. Sensor in S.S. in 1.17 mm Plastic ible cable and Housing with
Probe 0.92 mm Plastic Housing
Housing Housing Tube plastic tip 1/8-27 NPT body
Tube
Design and
Fig. 2 Fig. 3 Fig. 4 Fig. 5 Fig. 6 Fig. 7 Fig. 8 Fig. 9 Fig. 10
Construction
General Purpose General Purpose General Purpose General Purpose General Purpose General Purpose General Purpose
Major Laboratory Laboratory Laboratory Laboratory Cuvette Cuvette Laboratory Laboratory Laboratory
Applications Temperature Temperature Temperature Temperature Thermometry Thermometry Temperature Temperature Temperature
Measurement Measurement Measurement Measurement Measurement Measurement Measurement
Design Exposure
Temperature
Limits
minimum –50 °C –20 °C –20 °C –20 °C –20 °C –20 °C –20 °C –50 °C –50 °C
maximum 125 °C 105 °C 105 °C 105 °C 70 °C 70 °C 105 °C 105 °C 150 °C
1 (±0.01 °C) 1 (±0.01 °C) 1 (±0.01 °C) 1 (±0.01 °C) . . . . . . . . . . . . . . .
2 (±0.02 °C) 2 (±0.02 °C) 2 (±0.02 °C) 2 (±0.02 °C) . . . . . . . . . . . . . . .
Available
3 (±0.05 °C) 3 (±0.05 °C) 3 (±0.05 °C) 3 (±0.05 °C) . . . . . . . . . . . . . . .
Accuracy
4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C)
Classes
. . . . . . . . . . . . 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C)
... ... ... ... ... ... ... ... 6(±0.50°C)
A A A A A A A
... ...
(–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (-10 °C to 105 °C)
Available B (–10 °C to B (–10 °C to 60 B (–10 °C to 60 B (–10 °C to 60 B (–10 °C to 60
B (–10 °C to 60 °C) B (–10 °C to 60 °C) B (–10 °C to 60 °C) B (–10 °C to 60 °C)
Operating 60 °C) °C) °C) °C) °C)
Temperature C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C) C(0°Cto60°C)
Ranges D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C) D(0°Cto70°C)
E(0°Cto100°C) E(0°Cto100°C) E(0°Cto100°C) E(0°Cto100°C) . . E(0°Cto100°C) E(0°Cto100°C) E(0°Cto100°C)
... ... ... ... ... ... ... F(-50°Cto50°C) F(-50°Cto50°C)
Accuracies for
Other Tempera- Class 3 (±0.05 °C) Class 3 (±0.05 °C)
tures Within n/a n/a n/a n/a Over 24 °C to 45 Over 24 °C to 45 n/a n/a n/a
Specified Tem- °C °C
perature Range
Non- Non-
Non- Non- Non- Non-
Calibration Type Interchangeable Interchangeable Interchangeable
interchangeable interchangeable interchangeable interchangeable interchangeable nterchangeable
all fluids compat- all fluids compat-
all fluids compatible all fluids compatible
Type of Immersion water, air water, oil, air water water water, oil, air ible with Type 304 ible with Type 304
with Type 304 S.S. with Type 304 S.S.
S.S. S.S.
Nominal
Resistance at 2500 (19.86) 2500 (19.86) 5000 (19.86) 5000 (19.86) 11 000 (20.37) 10 000 (19.86) 2252 (29.25) 2252 (29.25) 2252 (29.25)
25 °C in Ohms
with applicable
Curve (Ratio 25/ 10 000 (22.73) 10 000 (22.73) 10 000 (22.06) 10 000 (22.06) 44 000 (20.37) 22 000 (20.37) 10 000 (29.25) 10 000 (29.25) 10 000 (29.25)
125)
Dissipation
3.5 ± 0.9 mW/K 5.0 ± 1.2 mW/K 4.8 ± 1.2 mW/K 4.8 ± 1.2 mW/K 1.1 ± 0.3 mW/K 0.8 ± 0.2 mW/K 6.0 ± 1.5 mW/K 6.0 ± 1.5 mW/K 6.0 ± 1.5 mW/K
Constant
63.2% Response
0.55±0.16s 0.45±0.11 4.5±1.1s 4.5±1.1s 0.5±0.12s 0.26±0.06s 4.5±1.1s 4.5±1.1s 4.5±1.1s
Time
E879−20
TABLE 2 Resistance-temperature Data for ASTM Laboratory Thermistor Sensors
Curve Designation (19.86) (20.37) (22.06) (22.73) (29.25)
11 000
Applicable Nominal Resistance at 4000 5000 10 000 2252
22 000
25 °C (Ohms) 5000 10 000 15 000 10 000
44 000
10 000
Resistance ratio 25 °C to 125 °C 19.86±10% 20.37±10% 22.06±10% 2.73±10% 29.25±10%
Resistance ratio 0 °C to 50 °C 7.04±6% 7.08±6% 7.44±6% 7.58±6% 9.06±6%
Resistance ratio 0 °C to 70 °C 13.33±8% 13.45±8% 14.37±8% 14.75±8% 18.65±8%
Nominal ratio data:
–50 °C 40.15 40.07 44.97 46.74 67.01
–40 °C 22.06 22.07 24.16 24.96 33.65
–30 °C 12.59 12.60 13.53 13.89 17.70
–20 °C 7.433 7.448 7.863 8.025 9.707
–10 °C 4.534 4.543 4.728 4.800 5.532
0°C 2.849 2.853 2.932 2.962 3.265
10 °C 1.840 1.841 1.870 1.881 1.990
20 °C 1.840 1.219 1.224 1.227 1.249
25 °C 1.0000 1.0000 1.0000 1.0000 1.0000
30 °C 0.8262 0.8253 0.8215 0.8197 0.8057
40 °C 0.5725 0.5711 0.5633 0.5600 0.5327
50 °C 0.4048 0.4032 0.3942 0.3906 0.3603
60 °C 0.2917 0.2899 0.2811 0.2776 0.2487
70 °C 0.2138 0.2121 0.2040 0.2008 0.1751
80 °C 0.1593 0.1576 0.1504 0.1477 0.1255
90 °C 0.1206 0.1189 0.1126 0.1102 0.09156
100 °C 0.09245 0.0909 0.08547 0.08346 0.06784
105 °C 0.08138 0.0799 0.07484 0.07298 0.05876
110 °C 0.07186 0.0704 0.06573 0.06402 0.05107
120 °C 0.05654 0.0552 0.05117 0.04971 0.03896
125 °C 0.05036 0.0491 0.04534 0.04400 0.03419
130 °C 0.04487 0.0438 0.04029 0.03905 0.03010
140 °C 0.03619 0.0351 0.03209 0.03100 0.02352
150 °C 0.02953 0.0284 0.02577 0.02485 0.01859
TABLE 3 Equivalent Temperature Tolerances for Different Class
When tested in accordance with 7.2, the zero-power resistance
Sensors (See 4.1 and 4.4)
versus temperature relationship for interchangeable parts shall
Accuracy Temperature
comply to within the tolerance specified in Table 3. The
Class Tolerance, °C
manufacturerofthesensorshall,fornon-interchangeableparts,
1 ±0.02
supply this relationship with each part shipped. This relation-
2 ±0.03
shipmaybesuppliedintheformofdatatables,orbyproviding
3 ±0.05
4 ±0.1
the constants for one or more of the following thermistor
5 ±0.2
equations, or both.
6 ±0.5
Equations representing this relationship use the standard
curve fitting technique of considering the natural logarithm of
the zero-power resistance to be a polynomial in reciprocal
zero-power resistance measurements be performed only after
temperature, or vice versa. Although the polynomial equation
any such modifications have taken place. Examples of such
can be of any higher order, historically it has been shown that
terminations shall include (but are not limited to) phone plugs;
athirdorderpolynomial(Eq2and3)providesexcellentresults
stereo plugs; multiple pin DIN type plugs; multiple pin
over the relatively narrow temperature ranges of thermistor
“Deutsches Institut für Normung” (DIN) circular connectors;
sensor applications. Furthermore, Steinhart and Hart have
or, gold-plated terminals.
proven in oceanographic research studies over the range of
–2 °C to +30 °C that satisfactory results are obtained when the
5. Requirements
squared term is omitted (Eq 4 and 5).
5.1 Specifications—Sensors shall comply with the general
Eq4and5havethreeunknownconstantsthatarederivedby
requirements specified herein as well as with the applicable
the simultaneous solution of three equations based upon three
detailed specifications of Table 1, Table 2, and Fig. 2 through
calibration data points. Eq 2 and 3 have four unknown
Fig. 10. In the event of conflict between this requirement
constants and thus a minimum of four calibration data points
subsection and the detailed specification of Table 1, Table 2,
are required to solve for the constants. Frequently, more than
and Fig. 2 through Fig. 10, the latter shall govern.
four calibration data points are obtained and a polynomial
regression analysis is performed to derive the constants for
5.2 Zero-Power Resistance Versus Temperature
equations Eq 2 and 3. This has the benefit of statistically
Relationship—The zero-power resistance versus temperature
improving the accuracy of the data.
relationship shall be presented in a form such that any
temperature within the specified operating temperature range
can be obtained from that relationship and have an uncertainty
Steinhart, J.S., and Hart, S.R., “Calibration Curves forThermistors,” Deep Sea
no greater than one-tenth the specified tolerance in Table 3. Research, Vol 15, No. 497, 1968.
E879−20
FIG. 1 Thermistor Sensor Type Designation Code
FIG. 2Silicone Rubber Coated Glass Probe (Type S)
2 3
lnR 5 A 1 A ⁄ T 1 A ⁄ T 1 A ⁄ T (2) 5.3 Thermal Requirements:
@ #
T 0 1 2 3
2 3
5.3.1 Dissipation Constant—When tested in accordance
1⁄T 5 a 1a ln R 1a ln R 1a ln R (3)
~ ! ~ ! ~ !
0 1 T 2 T 3 T
with 7.3, the dissipation constant shall be as specified in the
lnR 5 @B 1 B ⁄ T 1 B ⁄ T # (4)
T 0 1 3
detailed specification.
1⁄T 5 b 1b ln R 1b ln R (5)
~ ! ~ !
0 1 T 3 T
5.3.2 Response Time—When tested in accordance with 7.4,
theresponsetimeortimeconstant,orboth,shallbeasspecified
where:
in the detailed specification.
A,a = unique constants derived from calibrations per-
n n
formed at four or more specified data points,
5.4 Environmental Requirements:
B,b = unique constants derived from calibrations per-
n n 5.4.1 Operating Temperature Range—The operating tem-
formed at three specified data points,
peraturerangeshallbeasspecifiedinthetypedesignationcode
T = is the absolute temperature in kelvins, and
(see 4.1 and 4.3).
R = is the zero-power resistance at temperature T.
T
5.4.2 StorageTemperatureRange—Sensorsshallbecapable
5.2.1 Accuracy—The resistance-temperature relationship, of meeting all requirements specified herein as well as those
provided in Table 2, or with the sensor, or both, shall not differ listed in the applicable detailed specification after storage at
from that obtained from measurements made in accordance any temperature (or combination thereof) in the range
with7.2bymorethanthetolerancesspecifiedinTable3forthe from–50 °C minimum (or the design limit for minimum
applicable intervals specified in Table 1. When additional exposure temperature, whichever is higher) up to 60 °C
temperature intervals are specified within the operating tem- maximum (or the design limit for maximum exposure
perature range, the accuracy class for each interval shall be temperature,whicheverislower)foraperiodof1year.Sensors
clearly indicated by the manufacturer. subjected to operation or storage at temperatures that exceed
E879−20
FIG. 3Epoxy Coated Glass Probe (Type E)
FIG. 4General Purpose Four-wire Sensor in Stainless Steel Housing (Type G)
their design limits may experience mechanical damage, de- greater than 10% of the tolerance shown in Table 3 for the
graded stability, or both, as well as unreliable performance. accuracy class specified.
5.4.3 Humidity Requirement—Sensors shall be capable of
5.5.2 Long-term Stability (120 days)—When tested in ac-
being operated or stored at relative humidity from 0 to 95%
cordance with 7.5.2, the equivalent temperature shift shall be
without condensation.
no greater than 25% of the tolerance shown in Table 3 for the
5.5 Stability:
accuracy class specified.
5.5.1 Short-term Stability (10 days)—When tested in accor-
dance with 7.5.1, the equivalent temperature shift shall be no
E879−20
FIG. 5General Purpose Two-wire Sensor in Stainless Steel Housing (Type H)
FIG. 6Interchangeable Sensor Enclosed in 1.2 mm Vinyl Tube (Type V)
5.6 Low-temperature Storage—When tested in accordance 5.8 Insulation Resistance:
with7.6,thereshallbenoevidenceofmechanicaldamageand
5.8.1 Dry Test—Thisrequirementshallapplytosensorsthat
the sensor shall comply with the accuracy requirements of 5.2.
have exposed metallic surfaces, but are not designed for
immersion in conductive fluids. When tested in accordance
5.7 Thermal Shock—When tested in accordance with 7.7,
with 7.8.1, there shall be no evidence of mechanical damage
there shall be no evidence of mechanical damage and the
sensor shall comply with the accuracy requirements of 5.2. and the insulation resistance shall be sufficiently high that its
E879−20
FIG. 7Non-Interchangeable Sensor Enclosed in 0.9 mm Vinyl Tube
FIG. 8Interchangeable Sensor with Flexible Cable and Sealed Plastic Tip (Type P)
shunting effect will not prevent the unit from complying with 6. Quality Assurance Provisions
the accuracy requirement of Table 3. In no case shall the
6.1 General—The methods of examination and tests con-
insulation resistance be less than 10 Ω.
tainedinSection7aretobeusedtodeterminetheconformance
5.8.2 Wet Test—This requirement shall apply to sensors that
of sensors to the requirements of this specification. Each
are designed for use in conductive solutions. When tested in
manufacturer or distributor who represents his products as
accordance with 7.8.2, there shall be no evidence of mechani-
conforming to this specification may, as agreed upon between
cal damage and the insulation resistance shall be sufficiently
the purchaser and seller, use statistically based sampling plans
high that its shunting effect will not prevent the unit from that are appropriate for each inspection lot. Records shall be
complying with the accuracy requirement of Table 3.Inno
kept as necessary to document the claim that all of the
case shall the insulation resistance be less than 10 Ω. requirementsofthisspecificationaremet.Thetestsspecifiedin
E879−20
FIG. 9Interchangeable Sensor In Stainless Steel Housing (Type J)
FIG. 10Interchangeable Sensor In Stainless Steel Housing With Pipe Fitting (Type K)
TABLE 4 Letter Symbol Designation of Operating Temperature
Ranges (See 4.1 and 4.3)
Operating Temperature
Letter Symbol
Range
−10 °C to 105 °C A
−10 °C to 60 °C B
0°Cto60°C C
0°Cto70°C D
0°Cto100°C E
–50 °C to 50 °C F
E879−20
TABLE 6 Manufacturing Screening Tests
this section are intended as minimum requirements.Additional
sampling and testing of the product, as may be agreed upon Examination Requirement Test Method
or Test Section Section
between the purchaser and the seller, are not precluded by this
Visual and mechanical 5.1 7.1
section.
Zero-power resistance versus 5.2 7.2
temperature relationship
6.2 Classification of Inspection:
Insulation resistance 5.8 7.8
6.2.1 Qualification Tests—Qualification tests shall be per-
formed for each basic design manufactured in accordance with
this specification. The sample size required for the tests
conducted shall be in accordance with Table 5. In order for a
Standards and Technology (NIST) using suitable reference
design to qualify, there shall be no failures resulting from any
standards with documentation.
of the tests.
7.2.1.1 Measurement System Uncertainties—Uncertainties
6.2.2 Responsibility for Qualification Testing—The manu-
existinallmeasurementsandthesemustbeproperlyevaluated
facturer shall perform qualification testing, at least once, for
in order to have confidence in the measurement results. An
eachbasicdesignforwhichthisspecificationapplies.Ifabasic
uncertainty budget is an analysis tool that should be used
designincorporatesmorethanoneresistancevalueofaspecific
before the zero-power measurements are taken in order to
materialformulationoraparticularstyleofthermistor,orboth,
determine if the measurement system is capable of achieving
different resistance values may be combined for the qualifica-
the desired results. The uncertainty budget will identify the
tion sample. The highest and lowest resistance values for a
sources of uncertainty and their individual contributions to the
specified thermistor design (type, material formulation, and
overall uncertainty of the measurement system. The result of
geometry) must be included in the qualification sample. Quali-
theanalysisisacombinedexpandeduncertainty,symbolU,for
fication testing, by the manufacturer, must be repeated when-
the measurement system. Typically, U is given at a coverage
ever a design change which may affect the performance of the
factor of 2, approximating to a 95 % confidence level. For the
sensor with regard to Section 5 of this specification is intro-
purpose of performing the zero-power resistance-versus-
duced.
temperature measurements herein, it is recommended that the
6.2.3 Manufacturing Screening Tests—During manufacture,
resulting combined expanded uncertainty for the measurement
all parts produced in accordance with this specification shall
system in use should not exceed 25 % of the allowable
receive 100% testing for compliance with the requirements of
tolerance for the accuracy class of the sensor to be tested.This
Table 6.
provides a desired test uncertainty ratio, TUR, of 4:1 and thus
there is a reasonable degree of confidence in the measurement
7. Methods of Examination and Test
data. Refer to Appendix X2 for a more in-depth discussion of
7.1 Visual and Mechanical Examination—Examine sensors
the concepts and sources of measurement uncertainty, as well
to verify that their design, construction, physical dimensions,
as examples. Also refer to ANSI/NCSL Z540.3 for additional
markings, and workmanship comply with the detailed specifi-
guidance on requirements for calibration of measuring and test
cation.
equipment.
7.2 Zero-power Resistance versus Temperature
7.2.2 Temperature-controlled Medium:
7,8,9
Relationship:
7.2.2.1 Comparison Baths—Make all zero-power resistance
7.2.1 Traceability—All measurements shall be traceable to
versus temperature relationship measurements in a
theInternationalSystemofUnits(SI)throughaNational4,5,6.
temperature-controlled liquid bath (such as a water bath). The
Metrology Institute (NMI) such as the National Institute of
volume of the liquid should be at least 1000 times the volume
ofthesensor(s)undertest,butshallnotbelessthan1L.Baths
having volumes as large as 100 L have been found to be
Mangum,B.W.,“PlatinumResistanceThermometerCalibration,” NBS Special
convenient to use and to be satisfactory with respect to
Publication 250-22, 1987.
temperature control. Ensure that the bath medium is suffi-
Mangum, B. W., and Furukawa, G. T., “Guidelines for Realizing the Interna-
ciently well-stirred so that temperature gradients are small
tional Temperature Scale of 1990 (ITS-90),” NIST Technical Note 1265, 1990.
Riddle, J. L., Furukawa, G. T., and Plumb, H. H., “Platinum Resistance
compared with the temperature accuracy required. Survey the
Thermometry,” NBS Monograph 126, 1973.
bath with a thermometer to ensure that its temperature is
uniform and stable to the extent necessary to perform the tests.
TABLE 5 Qualification Tests Required for Each Basic Design
For further information refer to Guide E2488 which describes
Requirement Method
methods to define the working space of a bath and to evaluate
Examination or Test Sample Size
Section(s) Section(s)
the temperature uniformity and stability within this space.
Visual and mechanical 5.1 7.1 10
Ideally, the working space will be as close as possible to
Zero-power resistance versus 5.2 7.2 10
temperature relationship isothermal.Ifneeded,athermalintegratingblock(equalization
Dissipation constant 5.3.1 7.3 5
block) can be utilized to enhance performance and thus
Response time 5.3.2 7.4 5
minimize the uncertainty contributed by the comparison bath.
Short-term stability 5.5.1 7.5.1 10
Long-term stability 5.5.2 7.5.2 10 7.2.2.2 Fixed-point temperatures—Iftheoperatingtempera-
Low-temperature storage 5.6 7.6 10
ture range of the thermistor sensor includes the ice-point
Thermal shock 5.7 7.7 10
temperature, the water triple-point temperature, or the gallium
Insulation resistance 5.8 7.8 10
melting-point temperature, then an ice-point bath, a water
E879−20
triple-pointcell,oragalliummelting-pointcellmaybeusedas to this type of sensor must not exceed 14.4 µW (one fifth the
the temperature-controlled medium at that respective tempera- product of the dissipation constant and the temperature toler-
ture. Refer to Practice E563 and Guides E1502 and E1750 for ance requirement of Table 3). If the measurement system uses
additional guidance and information on the use of fixed-point an instrument that applies a constant current, then the worst
temperatures. case condition will occur when the thermistor resistance is at
its maximum (29 320 Ω at 0 °C). In this case, the maximum
7.2.3 Temperature Monitoring and Control—While per-
value of the applied current may not exceed 2.2 µA. If the
forming the zero-power resistance versus temperature
measurementsystemusesaninstrumentthatappliesaconstant
measurements,observethetemperaturefluctuationsofthebath
voltage, then the worst case condition will occur when the
withacheckstandardthermometerhavingaresponsetimethat
thermistor resistance is at its minimum value (2811 Ω at
is shorter than or equal to that of the unit under test. The total
60 °C). In this case, the maximum value of the applied voltage
uncertainty resulting from the combined uncertainties of the
may not exceed 200 mV.
check standard thermometer and the bath temperature (due to
temperaturefluctuationsandbathgradientswithintheworking
7.3 Dissipation Constant—Determine the dissipation con-
volume) shall comply with the desired TUR of 4:1.
stant in water unless another fluid is specified. As determined
7.2.4 Resistance Measurement: Evaluate the contributions
here, the dissipation constant is for the specific environment
to overall system uncertainty for all test instruments used to
described in 7.3.1. Measurements made with the sensor in air,
performthezero-powerresistancemeasurementstoensurethat
oil, still water, etc. will yield different values.
the total uncertainty complies with the desired TUR of 4:1 for
7.3.1 Mount the sensor in a fluid bath that is controlled at
the specific accuracy class as specified in Table 3.
some temperature, T, in the range from 24 °C to 38 °C. The
i
7.2.5 Test Procedure: fluid specified for the bath shall have a velocity of no less than
7.2.5.1 Temperature Stabilization—After inserting the sen- 1 m/s and its volume shall be no less than 1000 times the
volume of the sensor. Determine the zero-power resistance, R,
sor into the bath, allow enough time for the sensor and bath to
i
come to equilibrium (see 7.2.3). from measurements made in accordance with 7.2.
7.3.2 Increase the measuring current (or voltage) until the
7.2.5.2 Immersion—Bestresultswillbeobtainedwhenmea-
sensor indicates a resistance R , equivalent to that at a
surements are made with the sensor totally immersed. The i+10
temperature of T , a temperature which is 10 °C higher
manufacturer shall specify the minimum immersion length
i+10
than that of the initial temperature T.
required to obtain the specified tolerance within the tempera- i
7.3.3 Measure the sensor current (or voltage) to within an
ture range permitted. (See Table 3.)
uncertainty of 61% and compute the dissipation constant
7.2.5.3 Zero-power Resistance:
from Eq 6:
(a) Sensors Designed for Operating Temperature Ranges
2 2
With Code Letter Symbols B, C, or D—Determine the zero-
δ 5 ∆P⁄∆T 5 I · R ⁄10 5 E ⁄ 10 · R (6)
~ i 1 10 ! ~ i 1 10 !
~ ! ~ !
power resistance of the sensor at 0 °C 6 0.3 °C, 30 °C 6 0.3
where:
°C, and 60 °C 6 0.5 °C.
δ = dissipation constant,
(b) Sensors Designed or Operating Temperature Ranges
∆ = change in applied power,
With Code Letter Symbols A or E—Determine the zero-power
∆ = resulting change in thermistor temperature,
resistance of the sensor at 0 °C 6 0.3 °C, 30 °C 6 0.3 °C,
I = measured current,
60 °C 6 0.5 °C, and 105 °C 6 1.0 °C.
E = measured voltage, and
(c) Sensors Designed for Operating Temperature Range
R = resistance of thermistor at a temperature 10 °C
(i+ 10)
With Code Letter Symbol F—Determine the zero-power resis-
above the initial temperature, T.
I
tance of the sensor at –20 °C 6 1.0 °C, 0 °C 6 0.3 °C, 30 °C
7.4 Response Time—Determine the response time in water
6 0.3 °C, and 50 °C 6 0.5 °C.
unless another fluid is specified. As determined here, the
7.2.5.3.1 Discussion—While performing the zero-power re-
response time is for the specific environment described in
sistance measurements at any of the specified calibration
7.4.2. Measurements made with the sensor in air, oil, still
temperatures above, verify that the applied power level to the
water, etc. will yield different values.
unit under test (UUT) does not exceed the zero-power resis-
7.4.1 Connect the sensor to an instrument that continuously
tance criterion as described in 3.2.7 (Discussion).
records the sensor output signal. It is desirable that the
Example—A thermistor sensor with ASTM specification
recorded signal be linearly related to temperature. See Appen-
numberE879G2B2Nistobemeasuredpertherequirementsof
dix X3 for information on the design of a thermistor voltage
7.2.5.3(a). This thermistor sensor is defined by type designa-
divider circuit that provides a linear output signal.
tion “G2” (10 000 Ω nominal at 25 °C with a curve ratio value
7.4.2 Mount the sensor in a plunger-type fixture above a
of 22.06); an operating temperature range “B” (–10 °C to 60
°C); an accuracy class “2” (60.02 °C), and a minimum fluid bath having a minimum volume of 1000 times the sensor
volume and a temperature somewhere in the range from
dissipation constant of 3.6 mW/K. Using the nominal resis-
tance as well as the ratio values of Table 2, we can determine 0.01 °C to 5 °C that is constant during the time of measure-
ment. The fluid specified for the bath shall have a velocity of
the UUT resistance will be approximately 29 320, 8215, and
2811 Ωatthespecifiedcalibrationtemperaturesof0°C,30°C, no less than 1 m/s.
and 60 °C respectively. To meet the zero-power resistance 7.4.3 Allowthesensortocometoequilibriuminairatroom
criterion for these measurements the maximum applied power temperature in the range of 20 °C to 25 °C.
E879−20
7.4.4 Plunge the sensor into the bath to the immersion point 7.5.2.4 Compute the fractional change in zero-power
specified in 7.2.5.2. The transit time between the start of the resistance, ∆R /R .
tm tm
plunge and the submerged rest position of the sensor shall be
7.5.2.5 Compute the equivalent temperature shift in accor-
determined to be less than 3% of the 95% thermal response dance with 7.5.3.
time obtained in 7.4.5.
7.5.3 Computation of Equivalent Temperature Shift—
Although it may not always be valid for evaluation purposes,
7.4.5 Observetherecordinganddeterminethetimerequired
for the sensor to change from the initial to the final sensor theassumptionismadethatthestabilityofathermistormaybe
characterized by a fractional change in its zero-power
temperature. Determine the 95% and 63.2% response times
and calculate their ratio. If the ratio lies between 3.0 and 3.7, resistance, which is dependent on time and maximum storage
thenthesensormaybeassumedtoexhibitasingleexponential temperature but independent of the temperature at which the
change is evaluated.
responseandthe63.2%responsetimemaybeconsideredtobe
the time constant of the sensor. If the ratio is greater than 3.7,
7.5.3.1 Sensors That Include 37 °C Within Their Operating
the63.2%responsetimeshallnotbeusedasthetimeconstant Ranges—Compute the equivalent temperature shift at 37 °C
and the total response curve should be considered.
from Eq 7:
∆T 5 @R ~∆ R ⁄ R !#⁄5~R 2 R ! (7)
7.5 Stability: 37 Tm Tm 36.9 37.1
7.5.1 Short-term Stability:
7.5.3.2 Sensors That Do Not Include 37 °C Within Their
7.5.1.1 Class 1 and Class 2 Sensors—Measurements in a
Operating Ranges—Compute the equivalent temperature shift
Triple Point of Water Cell (as described in NBS Monograph from Eq 8 at a temperature, ts, which is the operating
126)oraGalliumMeltingPointStandard(NationalInstituteof
temperature closest to 37 °C.
Standards and Technology SRM 1968 or commercially avail-
∆T 5 R R ⁄ R ⁄5 R 2 R (8)
@ ~ ! ~ !#
Ts Tm Tm Ts20.1 Ts10.1
ableequivalent)arerequiredfortestingtheshort-termstability
7.6 Low-temperature Storage:
of Class 1 and Class 2 sensors. The use of either of these
7.6.1 Determine the zero-power resistance versus tempera-
fixed-point cells is optional for Class 3, 4, 5, and 6 sensors.
ture relationship in accordance with 7.2. If the results of a
ReferenceGuidesE1502andE1750foradditionalinformation
previous set of measurements made within 500 h of the
on the use of these fixed-point cells.
low-temperature exposure exist, then this step may be elimi-
(a)Determine the zero-power resistance, R , in one of the
tm
nated.
above mentioned cells at a measurement temperature, tm.
7.6.2 Place the sensor in a chamber (whose volume and
(b)Store the sensor, with no power applied, at its maxi-
massareatleast1000timesthevolumeandmassofthesensor)
mum rated temperature for a minimum period of ten days.
at room temperature.
(c)Repeat Step (a) at the same temperature.
7.6.3 Reduce the chamber temperature until the chamber is
(d)Compute ∆R /R .
tm tm
controlled at−65 6 5 °C. Allow the sensor to remain at this
(e)Computetheequivalenttemperatureshiftinaccordance
temperature for a period of 6 h 6 15 min.
with 7.5.3.
7.6.4 Remove the sensor from the chamber and allow it to
7.5.1.2 Class3,4,5,and6Sensors—Theuseofeitheranice
stabilize at room temperature for not less than one hour.
bath (see Practice E563 for preparation of ice bath) or a
7.6.5 Determine the zero-power resistance versus tempera-
temperature-controlledbathisoptionalfortestingClass3,4,5,
ture relationship in accordance with 7.2 and verify that it
and 6 sensors.
complies with the accuracy requirement of Table 1.
(a)Determine the zero-power resistance of the sensor in
accordancewith7.2attheicepointorsometemperatureinthe 7.6.6 Examine the sensor for evidence of mechanical dam-
range from 23 °C to 38 °C. age.
(b)Store the sensor, with no power applied, at its maxi-
7.7 Thermal Shock:
mum rated temperature for a minimum period of ten days.
7.7.1 Determine the zero-power resistance versus tempera-
(c)Repeat step (a) at the same temperature.
ture relationship in accordance with 7.2 to ascertain if the
(d)Compute ∆R /R .
tm tm
sensor complies with 5.2.1. This step may be omitted if the
(e)Computetheequivalenttemperatureshiftinaccordance
results of a previous set of measurements were made within
with 7.5.3.
500 h of the thermal shock exposure.
7.5.2 Long-term Stability:
7.7.2 Plungethesensorintoanicebathandallowittocome
7.5.2.1 Determinethezero-powerresistanceofthesensorin
to equilibrium. Leave the sensor in the bath for a period of not
accordance with 7.2 at a measurement temperature, tm, corre-
less than ten times the thermal time constant.
spondingtotheicepoint,triplepointofwater,galliummelting
7.7.3 Remove the sensor from the bath and allow at least
point (see 7.5.1), or some temperature in the range from 23 °C
15 min for it to come to equilibrium at room temperature.
to 38 °C.
7.7.4 Plunge the sensor into a water bath set to control at
7.5.2.2 Store the sensor, with no power applied, at its
55 °C to 60 °C for sensors with operating temperature range
maximum rated temperature for a minimum period of
code letter symbol B, C, D, or F; or set to control at 95 °C to
120 days.
100°Cforsensorswithoperatingtemperaturerangecodeletter
7.5.2.3 Repeat the step outlined in 7.5.2.1 at the same symbol A or E. Allow a period of not less than ten times the
temperature. thermal time constant for the sensor to reach equilibrium.
E879−20
7.7.5 Remove the sensor from the bath and allow at least (c)DeterminethevalueofT ,correspondingtothevalueof
s
15 min for it to come to equilibrium at room temperature. R computedfromEq9,usingthezero-powerresistanceversus
s
temperature relationship for the sensor.
7.7.6 Repeat Steps 7.7.2 – 7.7.5 for a total of five cycles.
(d)Verify that the absolute value of T − T does not
s L
7.7.7 Examine the sensor for evidence of mechanical dam-
exceed the tolerance specified in Table 3 for the accuracy class
age.
specified in Table 1.
7.7.8 Determine the zero-power resistance versus tempera-
7.8.2 Wet Test—Perform this test on sensors that are de-
ture relationship in accordance with 7.2 and verify that it
signed for immersion in a conductive solution.
complies with the accuracy requirement of Table 1.
7.8.2.1 Immerse the sensor in a saturated water solution of
7.8 Insulation Resistance:
sodium chloride for a period of not less than 24 h. The
7.8.1 Dry Test—Perform this test on sensors that have
immersion depth shall be the same as that used in 7.2.
exposed metallic surfaces but are not designed for immersion
7.8.2.2 While the sensor is immersed, connect its leads
in a conductive solution.
together and measure the insulation resistance between the
7.8.1.1 The insulation resistance shall be measured by sensor leads and the solution with 100 V dc applied, unless
applying 100 V dc between the insulated leads connected otherwise specified in the detail specification.
7.8.2.3 Repeat Step 7.8.2.2 with the polarity reversed.
together and the exposed metallic surface of the sensor.
7.8.2.4 Examine the sensors for evidence of mechanical
7.8.1.2 Repeat Step 7.8.1.1 with the polarity reversed.
damage.
7.8.1.3 Examine the sensor for evidence of mechanical
7.8.2.5 Verify that the measured value of insulation resis-
damage.
tance is greater than 10 Ω.
7.8.1.4 Verify that the measured value of insulation resis-
7.8.2.6 Using the zero-power resistance versus temperature
tance is greater than 10 Ω.
relationship specified or provided with the sensor,
7.8.1.5 Using the zero power resistance versus temperature
(a)Determine the zero-power resistance at the lowest
relationship specified or provided with the sensor,
temperature specified in Table 1.
(a)Determine the zero-power resistance at the lowest
(b)Determine the value of zero-power resistance resulting
temperature specified in Table 1.
from the shunting effect of the insulation resistanc
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E879 − 12 E879 − 20
Standard Specification for
Thermistor Sensors for General Purpose and Laboratory
Temperature Measurements
This standard is issued under the fixed designation E879; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This specification covers the general requirements for Negative Temperature Coefficient (NTC) thermistor-type sensors
intended to be used for laboratory temperature measurements or control, or both, within the range from −10from −10 °C to
105°C.105 °C.
1.2 This specification also covers the detailed requirements for ASTM designated sensors.
1.3 This specification also covers the requirements for general purpose, Negative Temperature Coefficient (NTC) thermistor-
type sensors intended for use with Digital Contact Thermometers (also known as Digital Thermometers) within the range from
–50–50 °C to +150°C.+150 °C.
1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
E344 Terminology Relating to Thermometry and Hydrometry
E563 Practice for Preparation and Use of an Ice-Point Bath as a Reference Temperature
E1502 Guide for Use of Fixed-Point Cells for Reference Temperatures
E1750 Guide for Use of Water Triple Point Cells
E2488 Guide for the Preparation and Evaluation of Liquid Baths Used for Temperature Calibration by Comparison
F29 Specification for Dumet Wire for Glass-to-Metal Seal Applications
2.2 NIST Documents:
NIST Monograph 126 Platinum Resistance Thermometry
NIST Special Publication 250-22 Platinum Resistance Thermometer Calibration
NIST Technical Note 1265 Guidelines for Realizing the International Temperature Scale of 1990
2.3 Other Documents:
JCGM 100:2008 Evaluation of measurement data – Guide to the expression of uncertainty in measurement
ANSI/NCSL Z540.3-2006 (R2013) Requirements for the Calibration of Measuring and Test Equipment
This specification is under the jurisdiction of ASTM Committee E20 on Temperature Measurement and is the direct responsibility of Subcommittee E20.03 on Resistance
Thermometers.
Current edition approved May 1, 2012May 1, 2020. Published June 2012July 2020. Originally approved in 1982. Last previous edition approved in 20072012 as E879 – 01
(2007).E879 – 12. DOI: 10.1520/E0879-01R07.10.1520/E0879-20.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from National Institute of Standards and Technology (NIST), 100 Bureau Dr., Stop 1070, Gaithersburg, MD 20899-1070, http://www.nist.gov.
Available from the BIPM, Sevres, France, http:// www.bipm.org.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E879 − 20
3. Terminology
3.1 Definitions—The definitions given in Terminology E344 shall apply to this specification.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 dissipation constant, δ, n—the ratio of the change in energy dissipated per unit time (power) in a thermistor, ΔP = P −
P , to the resultant temperature change of the thermistor, ΔT = T − T .
1 2 1
ΔP
δ5 (1)
ΔT
The dimensions of the dissipation constant are W/K.watts per kelvin (W/K). Note that many NTC thermistors are very
small devices, and as such, the dissipation constant dimensions are frequently expressed in terms of milliwatts per kelvin
(mW/K).
For this specification, T is in the range from 2020 °C to 38°C38 °C and ΔT = 10°C. = 10 °C.
3.2.2 dumet, n—round, copper-coated 42 % nickel-iron wire intended primarily for sealing to soft glass. Also known as CuNiFe
in some communities.communities; see Specification F29 for additional information.
3.2.3 insulation resistance, dc, n—the resistance at a specified direct-current voltage between the insulated leads of a thermistor
sensor and the metallic enclosure of the sensor, if such an enclosure is present, or else between the sensor leads and a conductive
medium in which the sensor is immersed.
3.2.4 qualification test, n—a series of tests conducted by the procuring agency or an agent thereof to determine conformance
of thermistor sensors to the requirements of a specification, normally for the development of a qualified products list under the
specification.
3.2.5 response time, n—the time required for a sensor to change a specified percentage of the total difference between its initial
and final temperatures as determined from zero-power resistances when the sensor is subjected to a step function change in
temperature.
3.2.6 time constant, n—the 63.2 % response time of a sensor that exhibits a single-exponential response.
3.2.7 zero-power resistance, n—the dc resistance of a device, at a specified temperature, calculated for zero-power.
3.2.7.1 Discussion—
Accurate zero-power resistance is obtained by extrapolating to zero-power the resistance values obtained from measurements at
three or more levels of power with the sensor immersed in a constant temperature medium. For the purpose of this specification,
this is obtained from measurements at a single power level adjusted such that the power isthat is adjusted or otherwise determined
to be not greater than one-fifth the product of the dissipation constant specified in Table 1 (see 3.2.13.2.1 and 7.3 and 7.3) and the
appropriate tolerance requirement of Table 3. Accordingly, the applied power levels required to satisfy the zero-power resistance
criterion as defined above can be as little as 2 μW (for a Type W sensor with accuracy class 1) to as much as 600 μW (for Types
J, K or P sensors with accuracy class 6). The known excitation current or voltage of the measurement system along with the
resistance of the sensor at the selected temperature (from the nominal resistance and ratio values given in Table 2. When making
stability measurements, the power ) can be used to compute the actual applied power to the unit under test (UUT) and to determine
if this power level meets the zero-power criterion. When making repeat measurements over a specified time period to determine
the stability of a sensor, the power level shall be kept constant.
4. Classification
4.1 Thermistor sensors covered by this specification shall be classified with a type designationunique identification code that
includes the ASTM detailed specification number followed by a descriptive code. Seetype designation code, an operating
temperature range code, an accuracy class code, and a calibration type code. (See Fig. 1.)
4.2 ASTM Specification Number—The ASTM specification number specifies uniquely the design and construction of the sensor
including the type designation if more than one type appears in the same specification.
4.2.1 Type Designation—The type designation code shall be a letter symbol to indicate the design and construction of the
thermistor sensor.sensor followed by a number to indicate the subset of specific features as shown on the detailed specification
drawing whenever there are design options. (See Fig. 2 through Fig. 10).
4.2.1.1 Type S—Silicone rubber-coated glass probe with tinned Dumet extension leads (see Fig. 2).
4.2.1.2 Type E—Epoxy-coated glass probe with silver-plated copper extension leads (see Fig. 3).
4.2.1.3 Type G—General purpose four wire four-wire sensor in stainless steel housing (see Fig. 4).
4.2.1.4 Type H—General purpose two-wire sensor in stainless steel housing (see Fig. 5).
4.2.1.5 Type V—Interchangeable sensor enclosed in 1.2-mm1.2 mm vinyl tube (see Fig. 6).
4.2.1.6 Type W—Non-interchangeable sensor enclosed in 0.9-mm0.9 mm vinyl tube (see Fig. 7).
4.2.1.7 Type P—Interchangeable sensor with flexible cable and sealed plastic tip (see Fig. 8).
E879 − 20
TABLE 3 Letter Symbol Designation of Operating Temperature Ranges (See 4.1 and 4.3)
Operating
Temperature Letter Symbol
Range
−10
A
to 105°C
−10 to 60°C B
TABLE 1 Specification for ASTM Laboratory Thermistor Sensors
E879 Type S E879 Type E E879 Type G E879 Type H E879 Type V E879 Type W E879 Type P E879 Type J E879 Type K
Non-
4-Wire Non- 2-Wire Non- Interchangeable Interchangeable Interchangeable
Silicone Rubber interchangeable Interchangeable
Epoxy Coated interchangeable interchangeable Sensor Enclosed Sensor with flex- Sensor in S.S.
Description Coated Glass Sensor Enclosedin Sensor in S.S.
Glass Probe Sensor in S.S. Sensor in S.S. in 1.17 mm Plastic ible cable and Housing with
Probe 0.92 mm Plastic Housing
Housing Housing Tube plastic tip 1/8-27 NPT body
Tube
Design and
Fig. 2 Fig. 3 Fig. 4 Fig. 5 Fig. 6 Fig. 7 Fig. 8 Fig. 9 Fig. 10
Construction
General Purpose General Purpose General Purpose General Purpose General Purpose General Purpose General Purpose
Major Laboratory Laboratory Laboratory Laboratory Cuvette Cuvette Laboratory Laboratory Laboratory
Applications Temperature Temperature Temperature Temperature Thermometry Thermometry Temperature Temperature Temperature
Measurement Measurement Measurement Measurement Measurement Measurement Measurement
Design Exposure
Temperature
Limits
minimum –50 °C –20 °C –20 °C –20 °C –20 °C –20 °C –20 °C –50 °C –50 °C
maximum 125 °C 105 °C 105 °C 105 °C 70 °C 70 °C 105 °C 105 °C 150 °C
1 (±0.01 °C) 1 (±0.01 °C) 1 (±0.01 °C) 1 (±0.01 °C) . . . . . . . . . . . . . . .
2 (±0.02 °C) 2 (±0.02 °C) 2 (±0.02 °C) 2 (±0.02 °C) . . . . . . . . . . . . . . .
Available
3 (±0.05 °C) 3 (±0.05 °C) 3 (±0.05 °C) 3 (±0.05 °C) . . . . . . . . . . . . . . .
Accuracy
4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C) 4 (±0.10 °C)
Classes
. . . . . . . . . . . . 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C) 5 (±0.20 °C)
. . . . . . . . . . . . . . . . . . . . . . . . 6 (±0.50 °C)
A A A A A A A
. . . . . .
(–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (–10 °C to 105 °C) (-10 °C to 105 °C)
B (–10 °C to B (–10 °C to 60 B (–10 °C to 60 B (–10 °C to 60 B (–10 °C to 60
Available
B (–10 °C to 60 °C) B (–10 °C to 60 °C) B (–10 °C to 60 °C) B (–10 °C to 60 °C)
Operating 60 °C) °C) °C) °C) °C)
C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C) C (0 °C to 60 °C)
Temperature
Ranges D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C) D (0 °C to 70 °C)
E (0 °C to 100 °C) E (0 °C to 100 °C) E (0 °C to 100 °C) E (0 °C to 100 °C) . . . . . . E (0 °C to 100 °C) E (0 °C to 100 °C) E (0 °C to 100 °C)
. . . . . . . . . . . . . . . . . . . . . F (-50 °C to 50 °C) F (-50 °C to 50 °C)
Accuracies for
Other Tempera- Class 3 (±0.05 °C) Class 3 (±0.05 °C)
tures Within n/a n/a n/a n/a Over 24 °C to 45 Over 24 °C to 45 n/a n/a n/a
Specified Tem- °C °C
perature Range
Non- Non- Non- Non- Non- Non-
Calibration Type Interchangeable Interchangeable Interchangeable
interchangeable interchangeable interchangeable interchangeable interchangeable nterchangeable
all fluids compat- all fluids compat-
all fluids compatible all fluids compatible
Type of Immersion water, air water, oil, air water water water, oil, air ible with Type 304 ible with Type 304
with Type 304 S.S. with Type 304 S.S.
S.S. S.S.
Nominal
Resistance at 2500 (19.86) 2500 (19.86) 5000 (19.86) 5000 (19.86) 11 000 (20.37) 10 000 (19.86) 2252 (29.25) 2252 (29.25) 2252 (29.25)
25 °C in Ohms
with applicable
Curve (Ratio 25/ 10 000 (22.73) 10 000 (22.73) 10 000 (22.06) 10 000 (22.06) 44 000 (20.37) 22 000 (20.37) 10 000 (29.25) 10 000 (29.25) 10 000 (29.25)
125)
Dissipation
3.5 ± 0.9 mW/K 5.0 ± 1.2 mW/K 4.8 ± 1.2 mW/K 4.8 ± 1.2 mW/K 1.1 ± 0.3 mW/K 0.8 ± 0.2 mW/K 6.0 ± 1.5 mW/K 6.0 ± 1.5 mW/K 6.0 ± 1.5 mW/K
Constant
E879 − 20
63.2% Response
0.55 ± 0.16 s 0.45 ± 0.11 4.5 ± 1.1 s 4.5 ± 1.1 s 0.5 ± 0.12 s 0.26 ± 0.06 s 4.5 ± 1.1 s 4.5 ± 1.1 s 4.5 ± 1.1 s
Time
E879 − 20
TABLE 56 Manufacturing Screening Tests
Examination Requirement Test Method
or Test Section Section
Visual and mechanical 5.1 7.1
Zero-power resistance versus 5.2 7.2
temperature relationship
Insulation resistance 5.8 7.8
4.2.1.8 Type J—Interchangeable sensor enclosed in stainless steel housing (see Fig. 9).
4.2.1.9 Type K—Interchangeable sensor enclosed in stainless steel housing with threaded pipe fitting (see Fig. 10).
4.3 Operating Temperature Range—The operating temperature range shall be designated by a letter symbol (see Table 34).
4.4 Accuracy Class—The accuracy class shall be designated by a single-digit number (see Table 23).
4.5 Calibration Type—The type of calibration required for each unit shall be designated by a letter symbol. The letter I shall
be used to denote units that are interchangeable with respect to a single resistance-temperature relationship. The letter N shall be
used to denote non-interchangeable units for which resistance-temperature information must be furnished for each unit. For
Calibration Type N sensors, serial number identification must be provided.
4.6 Zero-power Resistance at 25 °C—Each of the individual thermistor sensor types shall have nominal zero-power resistance
values at 25 °C as specified in Table 1 and as shown on the individual detailed specification drawings. Each nominal resistance
value specified shall also have an associated resistance ratio-versus-temperature value that is characteristic of the thermistor
material formulation employed in the sensor. The resistance ratio-versus-temperature data for the selected thermistor material
formulations are shown in Table 2. Other nominal resistance values at 25 °C, or at another reference temperature, may be agreed
upon by the manufacturer and end user provided that this information is clearly communicated in all appropriate documentation;
and, that the unique identification code (see Fig. 4.1) is not used under such circumstances.
4.7 Terminations—The detailed specification drawings (see Fig. 2 through Fig. 10) show the sensors with the most common type
of termination for that type designation. All sensors, regardless of type designation, may be terminated with any suitable electrical
connector (as agreed upon by manufacturer and end user) such that it allows the sensor to interface with a specific thermometer
system. It is recommended that the zero-power resistance measurements be performed only after any such modifications have taken
place. Examples of such terminations shall include (but are not limited to) phone plugs; stereo plugs; multiple pin DIN type plugs;
multiple pin “Deutsches Institut für Normung” (DIN) circular connectors; or, gold-plated terminals.
5. Requirements
5.1 Specifications—Sensors shall comply with the general requirements specified herein as well as with the applicable detailed
specifications of Table 1, Table 2, and Fig. 2 through Fig. 10. In the event of conflict between this requirement paragraphsubsection
and the detailed specification of Table 1, Figs. 2-7Table 2, and Fig. 2 through Fig. 10, the latter shall govern.
5.2 Zero-Power Resistance versusVersus Temperature Relationship—The zero-power resistance versus temperature relationship
shall be presented in a form such that any temperature within the specified operating temperature range can be obtained from that
relationship and have an uncertainty no greater than one-tenth the specified tolerance in Table 23. When tested in accordance with
7.2, the zero-power resistance versus temperature relationship for interchangeable parts shall comply to within the tolerance
specified in Table 23. The manufacturer of the sensor shall, for non-interchangeable parts, supply this relationship with each part
shipped. This relationship may be supplied in the form of data tables, or by providing the constants for one or more of the following
thermistor equations, or both.
Equations representing this relationship use the standard curve fitting technique of considering the natural logarithm of the
zero-power resistance to be a polynomial in reciprocal temperature, or vice versa. Although the polynomial equation can be of any
higher order, historically it has been shown that a third order polynomial (Eq 2 and 3) provides excellent results over the relatively
narrow temperature ranges of thermistor sensor applications. Furthermore, Steinhart and Hart have proven in oceanographic
research studies over the range of –2 °C to +30 °C that satisfactory results are obtained when the squared term is omitted (Eq 4
and 5).
Eq 4 and 5 have three unknown constants that are derived by the simultaneous solution of three equations based upon three
calibration data points. Eq 2 and 3 have four unknown constants and thus a minimum of four calibration data points are required
to solve for the constants. Frequently, more than four calibration data points are obtained and a polynomial regression analysis is
performed to derive the constants for equations Eq 2 and 3. This has the benefit of statistically improving the accuracy of the data.
2 3
lnR 5 @A 1 A ⁄ T 1 A ⁄ T 1 A ⁄ T # (2)
T 0 1 2 3
2 3
1⁄T 5 a 1a ln R 1a ln R 1a ln R (3)
~ ! ~ ! ~ !
0 1 T 2 T 3 T
Steinhart, J.S., and Hart, S.R., “Calibration Curves for Thermistors,” Deep Sea Research, Vol 15, No. 497, 1968.
E879 − 20
TABLE 2 Resistance-temperature Data for ASTM Laboratory Thermistor Sensors
Curve Designation (19.86) (20.37) (22.06) (22.73) (29.25)
11 000
Applicable Nominal Resistance at 4000 5000 10 000 2252
22 000
25 °C (Ohms) 5000 10 000 15 000 10 000
44 000
10 000
Resistance ratio 25 °C to 125 °C 19.86 ± 10 % 20.37 ± 10 % 22.06 ± 10 % 2.73 ± 10 % 29.25 ± 10 %
Resistance ratio 0 °C to 50 °C 7.04 ± 6 % 7.08 ± 6 % 7.44 ± 6 % 7.58 ± 6 % 9.06 ± 6 %
Resistance ratio 0 °C to 70 °C 13.33 ± 8 % 13.45 ± 8 % 14.37 ± 8 % 14.75 ± 8 % 18.65 ± 8 %
Nominal ratio data:
–50 °C 40.15 40.07 44.97 46.74 67.01
–40 °C 22.06 22.07 24.16 24.96 33.65
–30 °C 12.59 12.60 13.53 13.89 17.70
–20 °C 7.433 7.448 7.863 8.025 9.707
–10 °C 4.534 4.543 4.728 4.800 5.532
0 °C 2.849 2.853 2.932 2.962 3.265
10 °C 1.840 1.841 1.870 1.881 1.990
20 °C 1.840 1.219 1.224 1.227 1.249
25 °C 1.0000 1.0000 1.0000 1.0000 1.0000
30 °C 0.8262 0.8253 0.8215 0.8197 0.8057
40 °C 0.5725 0.5711 0.5633 0.5600 0.5327
50 °C 0.4048 0.4032 0.3942 0.3906 0.3603
60 °C 0.2917 0.2899 0.2811 0.2776 0.2487
70 °C 0.2138 0.2121 0.2040 0.2008 0.1751
80 °C 0.1593 0.1576 0.1504 0.1477 0.1255
90 °C 0.1206 0.1189 0.1126 0.1102 0.09156
100 °C 0.09245 0.0909 0.08547 0.08346 0.06784
105 °C 0.08138 0.0799 0.07484 0.07298 0.05876
110 °C 0.07186 0.0704 0.06573 0.06402 0.05107
120 °C 0.05654 0.0552 0.05117 0.04971 0.03896
125 °C 0.05036 0.0491 0.04534 0.04400 0.03419
130 °C 0.04487 0.0438 0.04029 0.03905 0.03010
140 °C 0.03619 0.0351 0.03209 0.03100 0.02352
150 °C 0.02953 0.0284 0.02577 0.02485 0.01859
TABLE 23 Equivalent Temperature Tolerances for Different Class
Sensors (See 4.1 and 4.4)
Accuracy Temperature
Class Tolerance, °C
1 ±0.02
2 ±0.03
3 ±0.05
4 ±0.1
5 ±0.2
6 ±0.5
lnR 5 @B 1 B ⁄ T 1 B ⁄ T # (4)
T 0 1 3
1⁄T 5 b 1b ln~R !1b ln~R ! (5)
0 1 T 3 T
where:
A , a = unique constants derived from calibrations performed at four or more specified data points,
n n
B , b = unique constants derived from calibrations performed at three specified data points,
n n
T = is the absolute temperature in kelvins, and
R = is the zero-power resistance at temperature T.
T
5.2.1 Accuracy—The resistance-temperature relationship, provided in Table 12, or with the sensor, or both, shall not differ from
that obtained from measurements made in accordance with 7.2 by more than the tolerances specified in Table 23 for the applicable
intervals specified in Table 1. When additional temperature intervals are specified within the operating temperature range, the
accuracy class for each interval shall be clearly indicated by the manufacturer.
5.3 Thermal Requirements:
5.3.1 Dissipation Constant—When tested in accordance with 7.3, the dissipation constant shall be as specified in the detailed
specification.
5.3.2 Response Time—When tested in accordance with 7.4, the response time or time constant, or both, shall be as specified in
the detailed specification.
5.4 Environmental Requirements:
5.4.1 Operating Temperature Range—The operating temperature range shall be as specified in the type designation code (see
4.1 and 4.3).
E879 − 20
FIG. 1 Classification Thermistor Sensor Type Designation Code
FIG. 2 Silicone Rubber Coated Glass Probe (Type S)
5.4.2 Storage Temperature Range—Sensors shall be capable of meeting all requirements specified herein as well as those listed
in the applicable detailed specification after storage at any temperature (or combination thereof) in the range from −40 to 60°C for
from –50 °C minimum (or the design limit for minimum exposure temperature, whichever is higher) up to 60 °C maximum (or
the design limit for maximum exposure temperature, whichever is lower) for a period of 1 year. Sensors subjected to operation
or storage at temperatures that exceed their design limits may experience mechanical damage, degraded stability, or both, as well
as unreliable performance.
5.4.3 Humidity Requirement—Sensors shall be capable of being operated or stored at relative humidity from 0 to 95 % without
condensation.
5.5 Stability:
5.5.1 Short-TermShort-term Stability (10 days)—When tested in accordance with 7.5.1, the equivalent temperature shift shall
be no greater than 10 % of the tolerance shown in Table 23 for the accuracy class specified.
5.5.2 Long-TermLong-term Stability (120 days)—When tested in accordance with 7.5.2, the equivalent temperature shift shall
be no greater than 25 % of the tolerance shown in Table 23 for the accuracy class specified.
5.6 Low-TemperatureLow-temperature Storage—When tested in accordance with 7.6, there shall be no evidence of mechanical
damage and the sensor shall comply with the accuracy requirements of 5.2.
5.7 Thermal Shock—When tested in accordance with 7.7, there shall be no evidence of mechanical damage and the sensor shall
comply with the accuracy requirements of 5.2.
5.8 Insulation Resistance:
E879 − 20
FIG. 3 Epoxy Coated Glass Probe (Type E)
FIG. 4 General Purpose Four-WireFour-wire Sensor in Stainless Steel Housing (Type G)
5.8.1 Dry Test—This requirement shall apply to sensors that have exposed metallic surfaces, but are not designed for immersion
in conductive fluids. When tested in accordance with 7.8.1, there shall be no evidence of mechanical damage and the insulation
resistance shall be sufficiently high that its shunting effect will not prevent the unit from complying with the accuracy requirement
of Table 23. In no case shall the insulation resistance be less than 10 ohms. Ω.
5.8.2 Wet Test—This requirement shall apply to sensors that are designed for use in conductive solutions. When tested in
accordance with 7.8.2, there shall be no evidence of mechanical damage and the insulation resistance shall be sufficiently high that
its shunting effect will not prevent the unit from complying with the accuracy requirement of Table 23. In no case shall the
insulation resistance be less than 10 ohms. Ω.
E879 − 20
FIG. 5 General Purpose Two-WireTwo-wire Sensor in Stainless Steel Housing (Type H)
FIG. 6 Interchangeable Sensor Enclosed in 1.2 mm Vinyl Tube (Type V)
6. Quality Assurance Provisions
6.1 General—The methods of examination and tests con-tained in Section 7 are to be used to determine the conformance of
sensors to the requirements of this specification. Each manufacturer or distributor who represents his products as conforming to
this specification may, as agreed upon between the purchaser and seller, use statistically based sampling plans that are appropriate
for each inspection lot. Records shall be kept as necessary to document the claim that all of the requirements of this specification
are met. The tests specified in this section are intended as minimum requirements. Additional sampling and testing of the product,
as may be agreed upon between the purchaser and the seller, are not precluded by this section.
6.2 Classification of Inspection:
E879 − 20
FIG. 7 Non-Interchangeable Sensor Enclosed in 0.9 mm Vinyl Tube
FIG. 8 Interchangeable Sensor with Flexible Cable and Sealed Plastic Tip (Type P)
6.2.1 Qualification Tests—Qualification tests shall be performed for each basic design manufactured in accordance with this
specification. The sample size required for the tests conducted shall be in accordance with Table 45. In order for a design to qualify,
there shall be no failures resulting from any of the tests.
6.2.2 Responsibility for Qualification Testing—The manufacturer shall perform qualification testing, at least once, for each basic
design for which this specification applies. If a basic design incorporates more than one resistance value of a specific material
formulation or a particular style of thermistor, or both, different resistance values may be combined for the qualification sample.
The highest and lowest resistance values for a specified thermistor design (type, material formulation, and geometry) must be
included in the qualification sample. Qualification testing, by the manufacturer, must be repeated whenever a design change is
introduced which may affect the performance of the sensor with regard to Section 5 of this specification.specification is introduced.
6.2.3 Manufacturing Screening Tests—During manufacture, all parts produced in accordance with this specification shall receive
100 % testing for compliance with the requirements of Table 56.
7. Methods of Examination and Test
7.1 Visual and Mechanical Examination—Examine sensors to verify that their design, construction, physical dimensions,
markings, and workmanship comply with the detailed specification.
E879 − 20
FIG. 9 Interchangeable Sensor In Stainless Steel Housing (Type J)
FIG. 10 Interchangeable Sensor In Stainless Steel Housing With Pipe Fitting (Type K)
7,8,9
7.2 Zero-PowerZero-power Resistance versus Temperature RelationshipRelationship: :
7.2.1 Traceability—All measurements shall be traceable to the National Institute International System of Units (SI) through a
National 4,5,6. Metrology Institute (NMI) such as the National Institute of Standards and Technology (NIST) through the use of
using suitable reference standards with documentation.
7.2.1.1 Measurement System Uncertainties—Uncertainties exist in all measurements and these must be properly evaluated in
order to have confidence in the measurement results. An uncertainty budget is an analysis tool that should be used before the
zero-power measurements are taken in order to determine if the measurement system is capable of achieving the desired results.
The uncertainty budget will identify the sources of uncertainty and their individual contributions to the overall uncertainty of the
Mangum, B. W., “Platinum Resistance Thermometer Calibration,” NBS Special Publication 250-22250-22, (1987).1987.
Mangum, B. W., and Furukawa, G. T., “Guidelines for Realizing the International Temperature Scale of 1990 (ITS-90)(ITS-90),” NIST Technical Note 1265, (1990).1990.
Riddle, J. L., Furukawa, G. T., and Plumb, H. H., “Platinum Resistance Thermometry,” NBS Monograph 126, (1973).1973.
E879 − 20
TABLE 4 Letter Symbol Designation of Operating Temperature
Ranges (See 4.1 and 4.3)
Operating Temperature
Letter Symbol
Range
−10 °C to 105 °C A
−10 °C to 60 °C B
0 °C to 60 °C C
0 °C to 70 °C D
0 °C to 100 °C E
–50 °C to 50 °C F
TABLE 45 Qualification Tests Required for Each Basic Design
Requirement Method
Examination or Test Sample Size
Section(s) Section(s)
Visual and mechanical 5.1 7.1 10
Zero-power resistance versus 5.2 7.2 10
temperature relationship
Dissipation constant 5.3.1 7.3 5
Response time 5.3.2 7.4 5
Short-term stability 5.5.1 7.5.1 10
Long-term stability 5.5.2 7.5.2 10
Low-temperature storage 5.6 7.6 10
Thermal shock 5.7 7.7 10
Insulation resistance 5.8 7.8 10
measurement system. The result of the analysis is a combined expanded uncertainty, symbol U, for the measurement system.
Typically, U is given at a coverage factor of 2, approximating to a 95 % confidence level. For the purpose of performing the
zero-power resistance-versus-temperature measurements herein, it is recommended that the resulting combined expanded
uncertainty for the measurement system in use should not exceed 25 % of the allowable tolerance for the accuracy class of the
sensor to be tested. This provides a desired test uncertainty ratio, TUR, of 4:1 and thus there is a reasonable degree of confidence
in the measurement data. Refer to Appendix X2 for a more in-depth discussion of the concepts and sources of measurement
uncertainty, as well as examples. Also refer to ANSI/NCSL Z540.3 for additional guidance on requirements for calibration of
measuring and test equipment.
7.2.2 Temperature-Controlled Medium—Temperature-controlled Medium: Make all measurements in a temperature-controlled
liquid bath (such as a water bath). The volume of the liquid should be at least 1000 times the volume of the sensor(s) under test,
but shall not be less than 1 L. Baths having volumes as large as 100 L have been found to be convenient to use and to be satisfactory
with respect to temperature control. Ensure that the bath medium is sufficiently well-stirred that temperature gradients are small
compared with the temperature accuracy required. Survey the bath with a thermometer to ensure that its temperature is uniform
to the extent necessary to perform the tests. If the operating temperature range of the thermistor sensor includes the ice-point
temperature, the water triple-point temperature, or the gallium melting-point temperature, then an ice-point bath, a water
triple-point cell, or a gallium melting-point cell may be used as the temperature-controlled medium at that respective temperature.
7.2.2.1 Comparison Baths—Make all zero-power resistance versus temperature relationship measurements in a temperature-
controlled liquid bath (such as a water bath). The volume of the liquid should be at least 1000 times the volume of the sensor(s)
under test, but shall not be less than 1 L. Baths having volumes as large as 100 L have been found to be convenient to use and
to be satisfactory with respect to temperature control. Ensure that the bath medium is sufficiently well-stirred so that temperature
gradients are small compared with the temperature accuracy required. Survey the bath with a thermometer to ensure that its
temperature is uniform and stable to the extent necessary to perform the tests. For further information refer to Guide E2488 which
describes methods to define the working space of a bath and to evaluate the temperature uniformity and stability within this space.
Ideally, the working space will be as close as possible to isothermal. If needed, a thermal integrating block (equalization block)
can be utilized to enhance performance and thus minimize the uncertainty contributed by the comparison bath.
7.2.2.2 Fixed-point temperatures—If the operating temperature range of the thermistor sensor includes the ice-point
temperature, the water triple-point temperature, or the gallium melting-point temperature, then an ice-point bath, a water
triple-point cell, or a gallium melting-point cell may be used as the temperature-controlled medium at that respective temperature.
Refer to Practice E563 and Guides E1502 and E1750 for additional guidance and information on the use of fixed-point
temperatures.
7.2.3 Temperature Monitoring and Control—Determine the temperature While performing the zero-power resistance versus
temperature measurements, observe the temperature fluctuations of the bath with a check standard thermometer having a response
time that is shorter than or equal to that of the unit under test. The thermometer used to monitor the bath shall have a maximum
uncertainty of one-third of the tolerance specified in Table 2. The total uncertainty resulting from the combined uncertainties of
the monitor check standard thermometer and the bath temperature (due to temperature fluctuations and bath gradients within the
E879 − 20
working volume) shall not be greater than one half of the tolerance specified in comply with the desired TUR Table 2. When
stability measurements are made in which the difference between two measurements must be considered, the total uncertainty shall
not be greater than one third of the maximum difference allowed (see of 4:1.5.5).
7.2.4 Resistance Measurement:
7.2.4.1 Accuracy:
(a) Class 1 and Class 2 Sensors—The uncertainty of the resistance measurement shall be less than 60.01 % for zero-power
resistance determinations (see section 4.2.6) and less than 60.005 % for stability determinations.
(b) Class 3 and Class 4 Sensors—The uncertainty of the resistance measurement shall be less than 60.03 % for zero-power
resistance determinations and less than 60.01 % for stability determinations.Evaluate the contributions to overall system
uncertainty for all test instruments used to perform the zero-power resistance measurements to ensure that the total uncertainty
complies with the desired TUR of 4:1 for the specific accuracy class as specified in Table 3.
7.2.5 Test Procedure:
7.2.5.1 Temperature Stabilization—After inserting the sensor into the bath, allow enough time for the sensor and bath to come
to equilibrium (see 7.2.3).
7.2.5.2 Immersion—Best results will be obtained when measurements are made with the sensor totally immersed. The
manufacturer shall specify the minimum immersion length required to obtain the specified tolerance within the temperature range
permitted. (See Table 23.)
7.2.5.3 Zero-PowerZero-power Resistance:
(a) Sensors Designed for Operating Temperature Ranges With Code Letter Symbols B, C, or D—Determine the zero-power
resistance of the sensor at 0 °C 6 0.3 °C, 30 °C 6 0.3 °C, and 60 °C 6 0.5 °C.
(b) Sensors Designed or Operating Temperature Ranges With Code Letter Symbols A or E—Determine the zero-power
resistance of the sensor at 0 °C 6 0.3 °C, 30 °C 6 0.3 °C, 60 °C 6 0.5 °C, and 105 °C 6 1.0 °C.
(c) Sensors Designed for Operating Temperature Range With Code Letter Symbol F—Determine the zero-power resistance of
the sensor at –20 °C 6 1.0 °C, 0 °C 6 0.3 °C, 30 °C 6 0.3 °C, and 50 °C 6 0.5 °C. (a) Sensors Designed to Operate in the Range
from −10 to 60°C—Determine the zero-power resistance of the sensor at 0 6 0.3°C, 30 6 0.3°C, and 60 6 0.5°C.
7.2.5.3.1 Discussion—
While performing the zero-power resistance measurements at any of the specified calibration temperatures above, verify that the
applied power level to the unit under test (UUT) does not exceed the zero-power resistance criterion as described in 3.2.7
(Discussion).
Example—A thermistor sensor with ASTM specification number E879G2B2N is to be measured per the requirements of
7.2.5.3(a). This thermistor sensor is defined by type designation “G2” (10 000 Ω nominal at 25 °C with a curve ratio value of
22.06); an operating temperature range “B” (–10 °C to 60 °C); an accuracy class “2” (60.02 °C), and a minimum dissipation
constant of 3.6 mW/K. Using the nominal resistance as well as the ratio values of Table 2, we can determine the UUT resistance
will be approximately 29 320, 8215, and 2811 Ω at the specified calibration temperatures of 0 °C, 30 °C, and 60 °C respectively.
To meet the zero-power resistance criterion for these measurements the maximum applied power to this type of sensor must not
exceed 14.4 μW (one fifth the product of the dissipation constant and the temperature tolerance requirement of Table 3). If the
measurement system uses an instrument that applies a constant current, then the worst case condition will occur when the
thermistor resistance is at its maximum (29 320 Ω at 0 °C). In this case, the maximum value of the applied current may not exceed
2.2 μA. If the measurement system uses an instrument that applies a constant voltage, then the worst case condition will occur when
the thermistor resistance is at its minimum value (2811 Ω at 60 °C). In this case, the maximum value of the applied voltage may
not exceed 200 mV.
(b) Sensors Designed to Operate in the Range from −10 to 105°C—Determine the zero-power resistance of the sensor at 0 6
0.3°C, 30 6 0.3°C, 60 6 0.5°C, and 105 6 1.0°C.
7.3 Dissipation Constant—Determine the dissipation constant in water unless another fluid is specified. As determined here, the
dissipation constant is for the specific environment described in 7.3.1. Measurements made with the sensor in air, oil, still water,
etc. will yield different values.
7.3.1 Mount the sensor in a fluid bath that is controlled at some temperature, T , in the range from 2424 °C to 38°C. 38 °C. The
i
fluid specified for the bath shall have a velocity of no less than 1 m/s and its volume shall be no less than 1000 times the volume
of the sensor. Determine the zero-power resistance, R , from measurements made in accordance with 7.2.
i
7.3.2 Increase the measuring current (or voltage) until the sensor indicates a resistance R , equivalent to that at a temperature
i + 10
of T , a temperature which is 10°C 10 °C higher than that of the initial temperature T .
i + 10 i
7.3.3 Measure the sensor current (or voltage) to within an uncertainty of 61 % and compute the dissipation constant from Eq
26:
ΔP
2 2
δ 5 5 I R /10 5 E /10R (6)
i110 i110
ΔT
E879 − 20
2 2
δ 5 ΔP⁄ΔT 5 I · R ⁄10 5 E ⁄ 10 · R (6)
~ i 1 10 ! ~ i 1 10 !
~ ! ~ !
where:
δ = dissipation constant,
Δ = change in applied power,
Δ = resulting change in thermistor temperature,
I = measured current,
E = measured voltage, and
R = resistance of thermistor at a temperature 10 °C above the initial temperature, T .
(i + 10) I
7.4 Response Time—Determine the response time in water unless another fluid is specified. As determined here, the response
time is for the specific environment described in 7.4.2. Measurements made with the sensor in air, oil, still water, etc. will yield
different values.
7.4.1 Connect the sensor to an instrument that continuously records the sensor output signal. It is desirable that the recorded
signal be linearly related to temperature. See Appendix X3 for information on the design of a thermistor voltage divider circuit
that provides a linear output signal.
7.4.2 Mount the sensor in a plunger-type fixture above a fluid bath having a minimum volume of 1000 times the sensor volume
and a temperature somewhere in the range from 0.010.01 °C to 5°C 5 °C that is constant during the time of measurement. The
fluid specified for the bath shall have a velocity of no less than 1 m/s.
7.4.3 Allow the sensor to come to equilibrium in air at room temperature.temperature in the range of 20 °C to 25 °C.
7.4.4 Plunge the sensor into the bath to the immersion point specified in 7.2.5.2. (See Table 1 and Table 3.) The transit time
between the start of the plunge and the submerged rest position of the sensor shall be determined to be less than 3 % of the
90 %95 % thermal response time.time obtained in 7.4.5.
7.4.5 Observe the recording and determine the time required for the sensor to change from the initial to the final sensor
temperature. Determine the 95 % and 63.2 % response times and calculate their ratio. If the ratio lies between 3.0 and 3.7, then
the sensor may be assumed to exhibit a single exponential response and the 63.2 % response time may be considered to be the time
constant of the sensor. If the ratio is greater than 3.7, the 63.2 % response time shall not be used as the time constant and the total
response curve should be considered.
7.5 Stability:
7.5.1 Short-TermShort-term Stability:
7.5.1.1 Class 1 and Class 2 Sensors—Measurements in a Triple Point of Water Cell (as described in NBS Monograph 126) or
a Gallium Melting Point Standard (National Institute of Standards and Technology SRM 1968 or commercially available
equivalent) are required for testing the short-term stability of Class 1 and Class 2 sensors. Its use The use of either of these
fixed-point cells is optional for Class 3 and Class 4 sensors.3, 4, 5, and 6 sensors. Reference Guides E1502 and E1750 for
additional information on the use of these fixed-point cells.
(a) Determine the zero-power resistance, R , in one of the above mentioned cells at a measurement temperature, tm.
tm
(a) Determine the zero-power resistance, R , in one of the above mentioned cells at a measurement temperature, tm.
tm
(b) Store the sensor, with no power applied, at its maximum rated temperature for a minimum period of ten days.
(c) Repeat Step (a) at the same temperature.
(d) Compute ΔR /R .
tm tm
(e) Compute the equivalent temperature shift in accordance with 7.5.3.
(b) Store the sensor, with no power applied, at its maximum rated temperature for a minimum period of 10 days.
(c) Repeat Step (a) at the same temperature.
(d) Compute ΔR /R .
tm tm
(e) Compute the equivalent temperature shift in accordance with 7.5.3.
7.5.1.2 Class 3 and Class 4 3, 4, 5, and 6 Sensors—The use of either an ice bath (See(see Practice E563 for preparation of ice
bath) or a temperature-controlled bath is optional for testing Class 3 and Class 4 3, 4, 5, and 6 sensors.
(a) Determine the zero-power resistance of the sensor in accordance with 7.2 at the ice point or some temperature in the range
from 23 to 38°C.
(a) Determine the zero-power resistance of the sensor in accordance with 7.2 at the ice point or some temperature in the range
from 23 °C to 38 °C.
(b) Store the sensor, with no power applied, at its maximum rated temperature for a minimum period of ten days.
(c) Repeat step (a) at the same temperature.
(d) Compute ΔR /R .
tm tm
(e) Compute the equivalent temperature shift in accordance with 7.5.3.
(b) Store the sensor, with no power applied, at its maximum rated temperature for a minimum period of 10 days.
(c) Repeat step (a) at the same temperature.
(d) Compute ΔR /R .
tm tm
(e) Compute the equivalent temperature shift in accordance with 7.5.3.
E879 − 20
7.5.2 Long-TermLong-term Stability:
7.5.2.1 Determine the zero-power resistance of the sensor in accordance with 7.2 at a measurement temperature, tm,
corresponding to the ice point, triple point of water, gallium melting point (see 7.5.1), or some temperature in the range from
2323 °C to 38°C.38 °C.
7.5.2.2 Store the sensor, with no power applied, at its maximum rated temperature for a minimum period of 120 days.120 days.
7.5.2.3 Repeat the step outlined in 7.5.2.1 at the same temperature.
7.5.2.4 Compute the fractional change in zero-power resistance, ΔR /R .
tm tm
7.5.2.5 Compute the equivalent temperature shift in accordance with 7.5.3.
7.5.3 Computation of Equivalent Temperature Shift—Although it may not always be valid for evaluation purposes, the
assumption is made that the
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...