Standard Practice for Platelet Leukocyte Count&#x2014;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials

SIGNIFICANCE AND USE
4.1 The purpose of this practice is to determine if thrombus formation has occurred by comparing platelet and leukocyte counts in the blood exposed to the test material relative to the blood cell counts in the control blood that has not been exposed to the test material. A large number of platelets and leukocytes becoming entrapped/incorporated in thrombi adhering to the material will be reflected by a decrease in their counts in blood. Thrombogenic materials should not be used for cardiovascular medical devices, unless the purpose of the device is to promote thrombosis.
SCOPE
1.1 This practice assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as in accordance with ANSI/AAMI/ISO 10993–4. See also Test Method F2382.  
1.2 All safety policies and practices shall be observed during the performance of this practice. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled with the contents, and disposed of by appropriate means.  
1.3 The human blood should be handled at Biosafety Level 2 (BSL-2) as recommended in the Centers for Disease Control/National Institutes of Health publication, Biosafety in Microbiological and Biomedical Laboratories (BMBL). The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood and handled/manipulated using standard precautions.
Note 1: The results of this in-vitro test may not correspond to actual human response.  
1.4 The values stated in SI (International System of Units) units are to be regarded as standard. No other units of measurement are included in this standard.  
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 8 on Hazards.  
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Jan-2019

Relations

Effective Date
01-Feb-2019
Effective Date
01-Oct-2018
Effective Date
01-Sep-2017
Effective Date
01-Sep-2017
Effective Date
01-Jun-2010
Effective Date
01-May-2004
Effective Date
01-May-2004
Effective Date
01-Feb-2019

Overview

ASTM F2888-19 is a recognized standard practice established by ASTM International for assessing the hemocompatibility of cardiovascular materials. Specifically, it guides the in-vitro measurement of platelet and leukocyte counts to evaluate the potential of cardiovascular device materials to induce thrombus (blood clot) formation. This standard is critical in the development and validation of medical devices, ensuring that materials intended for blood contact do not pose an unnecessary risk of thrombosis.

Thrombus formation, marked by a reduction in platelets and leukocytes after exposure to device materials, is a significant safety and performance consideration for cardiovascular medical devices. The ASTM F2888-19 standard helps manufacturers and researchers identify thrombogenic properties of biomaterials before clinical use.

Key Topics

  • Hemocompatibility Assessment: Evaluates materials for their tendency to activate platelets and leukocytes, leading to thrombus formation.
  • In-Vitro Platelet and Leukocyte Counting: Uses blood analyzers to measure changes in blood cell counts after exposure to test materials compared to controls.
  • Material Preparation and Testing Conditions: Specifies methods for sample handling, including the use of fresh human whole blood, appropriate anticoagulants, and controlled incubation environments.
  • Control Materials: Involves positive and negative reference controls to validate assay performance and interpret results.
  • Safety and Biosafety Compliance: Includes guidelines for handling human blood at Biosafety Level 2 (BSL-2) and ensuring proper disposal of biohazardous materials.
  • Result Interpretation: Outlines criteria for determining if a material passes or fails hemocompatibility screening, contributing to device safety assessments.

Applications

ASTM F2888-19 is widely applicable in the cardiovascular device industry and biomedical research, including:

  • Cardiovascular Device Development: Used during the design and prototyping phases to screen materials for blood-contacting devices, such as stents, catheters, and heart valves, to minimize thrombogenicity.
  • Regulatory Submission Support: Provides critical data for demonstrating compliance with regulatory requirements for hemocompatibility as referenced in standards like ANSI/AAMI/ISO 10993-4.
  • Material Screening and Selection: Assists researchers and manufacturers in comparing new biomaterial options against known controls and marketed devices.
  • Risk Assessment and Quality Control: Forms part of comprehensive risk management and quality control activities in clinical and preclinical settings.
  • Research and Development: Enables scientific studies on the blood-material interface, supporting innovations in biocompatible material technologies.

Related Standards

  • ASTM F2382 - Assessment of Circulating Blood-Contacting Medical Device Materials on Partial Thromboplastin Time (PTT)
  • ANSI/AAMI/ISO 10993-4 - Biological Evaluation of Medical Devices-Selection of Tests for Interactions with Blood
  • CDC/NIH BMBL - Biosafety in Microbiological and Biomedical Laboratories, for handling human blood samples

Practical Value

By following ASTM F2888-19, device manufacturers can:

  • Reduce Thrombosis Risk: Identify and avoid materials with high thrombogenic potential early in product development.
  • Meet Regulatory Expectations: Align documentation and testing with global regulatory requirements for cardiovascular devices.
  • Promote Patient Safety: Ensure that only hemocompatible materials are used in devices intended for direct blood contact, reducing the risk of adverse clinical events.

Keywords: ASTM F2888-19, hemocompatibility assessment, platelet leukocyte count, cardiovascular materials, thrombogenicity, blood-contacting medical devices, biosafety, in-vitro testing, medical device standards.

Buy Documents

Standard

ASTM F2888-19 - Standard Practice for Platelet Leukocyte Count&#x2014;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials

English language (5 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM F2888-19 - Standard Practice for Platelet Leukocyte Count&#x2014;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials

English language (5 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

ASTM F2888-19 is a standard published by ASTM International. Its full title is "Standard Practice for Platelet Leukocyte Count&#x2014;An <emph type="bdit">In-Vitro</emph > Measure for Hemocompatibility Assessment of Cardiovascular Materials". This standard covers: SIGNIFICANCE AND USE 4.1 The purpose of this practice is to determine if thrombus formation has occurred by comparing platelet and leukocyte counts in the blood exposed to the test material relative to the blood cell counts in the control blood that has not been exposed to the test material. A large number of platelets and leukocytes becoming entrapped/incorporated in thrombi adhering to the material will be reflected by a decrease in their counts in blood. Thrombogenic materials should not be used for cardiovascular medical devices, unless the purpose of the device is to promote thrombosis. SCOPE 1.1 This practice assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as in accordance with ANSI/AAMI/ISO 10993–4. See also Test Method F2382. 1.2 All safety policies and practices shall be observed during the performance of this practice. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled with the contents, and disposed of by appropriate means. 1.3 The human blood should be handled at Biosafety Level 2 (BSL-2) as recommended in the Centers for Disease Control/National Institutes of Health publication, Biosafety in Microbiological and Biomedical Laboratories (BMBL). The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood and handled/manipulated using standard precautions. Note 1: The results of this in-vitro test may not correspond to actual human response. 1.4 The values stated in SI (International System of Units) units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 8 on Hazards. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

SIGNIFICANCE AND USE 4.1 The purpose of this practice is to determine if thrombus formation has occurred by comparing platelet and leukocyte counts in the blood exposed to the test material relative to the blood cell counts in the control blood that has not been exposed to the test material. A large number of platelets and leukocytes becoming entrapped/incorporated in thrombi adhering to the material will be reflected by a decrease in their counts in blood. Thrombogenic materials should not be used for cardiovascular medical devices, unless the purpose of the device is to promote thrombosis. SCOPE 1.1 This practice assists in the evaluation of cardiovascular device materials for their ability to induce thrombus formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and materials contacting human blood, as in accordance with ANSI/AAMI/ISO 10993–4. See also Test Method F2382. 1.2 All safety policies and practices shall be observed during the performance of this practice. All human blood and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled with the contents, and disposed of by appropriate means. 1.3 The human blood should be handled at Biosafety Level 2 (BSL-2) as recommended in the Centers for Disease Control/National Institutes of Health publication, Biosafety in Microbiological and Biomedical Laboratories (BMBL). The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses. The blood should be treated like any patient blood and handled/manipulated using standard precautions. Note 1: The results of this in-vitro test may not correspond to actual human response. 1.4 The values stated in SI (International System of Units) units are to be regarded as standard. No other units of measurement are included in this standard. 1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 8 on Hazards. 1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM F2888-19 is classified under the following ICS (International Classification for Standards) categories: 11.100.30 - Analysis of blood and urine. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM F2888-19 has the following relationships with other standards: It is inter standard links to ASTM F2888-13, ASTM F2382-18, ASTM F2382-17e1, ASTM F2382-17, ASTM F2382-04(2010), ASTM F2382-04e1, ASTM F2382-04, ASTM E3219-20. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM F2888-19 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: F2888 − 19
Standard Practice for
Platelet Leukocyte Count—An In-Vitro Measure for
Hemocompatibility Assessment of Cardiovascular Materials
This standard is issued under the fixed designation F2888; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.6 This international standard was developed in accor-
dance with internationally recognized principles on standard-
1.1 This practice assists in the evaluation of cardiovascular
ization established in the Decision on Principles for the
devicematerialsfortheirabilitytoinducethrombusformation.
Development of International Standards, Guides and Recom-
Thrombus formation is assessed by means of a reduction in
mendations issued by the World Trade Organization Technical
human platelets and leukocytes when consumed by thrombus
Barriers to Trade (TBT) Committee.
after activation on the material surface.This assay may be part
of the hemocompatibility evaluation for devices and materials
2. Referenced Documents
contacting human blood, as in accordance with ANSI/AAMI/
2.1 ASTM Standards:
ISO 10993–4. See also Test Method F2382.
F2382Test Method for Assessment of Circulating Blood-
1.2 All safety policies and practices shall be observed
Contacting Medical Device Materials on Partial Throm-
during the performance of this practice. All human blood and
boplastin Time (PTT)
any materials that had contact with human blood shall be
2.2 Other Standards:
bagged in a biohazard bag, properly labeled with the contents,
ANSI/AAMI/ISO 10993–4Biological Evaluation of Medi-
and disposed of by appropriate means.
cal Devices—Part 4: Selection of Tests for Interactions
with Blood
1.3 The human blood should be handled at Biosafety Level
BMBL Biosafety in Microbiological and Biomedical
2(BSL-2)asrecommendedintheCentersforDiseaseControl/
Laboratories, 5th ed., 2009
National Institutes of Health publication, Biosafety in Micro-
biological and Biomedical Laboratories (BMBL). The human
3. Summary of Practice
blood donor must have tested negative for Hepatitis B (HBV)
3.1 This practice identifies materials which are capable of
and Human Immunodeficiency (HIV) viruses. The blood
activatingbloodplateletsandleukocytesontheirsurfacewhen
should be treated like any patient blood and handled/
exposed to freshly drawn human blood and causing the
manipulated using standard precautions.
formation of thrombi on the material surface. A significant
NOTE 1—The results of this in-vitro test may not correspond to actual
decrease in the number of platelets and leukocytes when
human response.
counted by a blood analyzer is an indication of these cells
1.4 The values stated in SI (International System of Units)
being entrapped in thrombi.The materials are exposed to fresh
units are to be regarded as standard. No other units of
whole blood that is drawn with anticoagulant.Another antico-
measurement are included in this standard.
agulant is added at the appropriate time (one hour) to stop
further thrombus formation. Different blood analyzers may be
1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the used.
responsibility of the user of this standard to establish appro-
4. Significance and Use
priate safety, health, and environmental practices and deter-
4.1 The purpose of this practice is to determine if thrombus
mine the applicability of regulatory limitations prior to use.
formation has occurred by comparing platelet and leukocyte
Some specific hazards statements are given in Section 8 on
Hazards.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
ThispracticeisunderthejurisdictionofASTMCommitteeF04onMedicaland the ASTM website.
Surgical Materials and Devices and is the direct responsibility of Subcommittee Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
F04.16 on Biocompatibility Test Methods. 4th Floor, New York, NY 10036, http://www.ansi.org.
Current edition approved Feb. 1, 2019. Published February 2019. Originally Available from U.S. Department of Health and Human Services, Centers for
approved in 2013. Last previous edition approved in 2013 as F2888 – 13. DOI: DiseaseControlandPrevention(CDC),1600CliftonRd.,Atlanta,GA30329-4027,
10.1520/F2888–19. http://www.cdc.gov.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2888 − 19
counts in the blood exposed to the test material relative to the blood only, shall be prepared based on a ratio of 12 cm of
bloodcellcountsinthecontrolbloodthathasnotbeenexposed materialto1mLofbloodandplacedintopolypropylenetubes.
to the test material.Alarge number of platelets and leukocytes Also, three empty test tubes shall be prepared for the negative
becoming entrapped/incorporated in thrombi adhering to the control (blood in test tubes without test materials).
materialwillbereflectedbyadecreaseintheircountsinblood.
9.2 Thirty-six square centimetres of the test sample and
Thrombogenic materials should not be used for cardiovascular
each of the controls are divided into three 12 cm samples, cut
medicaldevices,unlessthepurposeofthedeviceistopromote
to maximize blood exposure, for triplicate testing.
thrombosis.
NOTE 2—If this surface area cannot be achieved with the device, or
other volumes of blood are used, the ratio of total surface area to blood
5. Interferences
volume should remain at 12 cm :1 mL.
5.1 There is potential for interference if the materials of the
10. Preparation of Apparatus
test tubes used are thrombogenic (for example, glass tubes).
10.1 Initialize the hematology analyzer and allow it to
Therefore, polyethylene or polypropylene tubes should be
perform internal self-checks. If no errors are noted, the
used.
analyzer is ready for use.
6. Apparatus
10.2 To verify the analyzer is functioning properly, prior to
6.1 Hematology analyzer capable of determination of a analyzing samples, cell count controls shall be run to confirm
complete blood count. that the results fall within the allowable ranges.
10.3 Fresh human blood from at least 3 donors should be
6.2 Polypropylene test tubes with caps.
usedforthetesttoaccountforbloodvariabilityamongdonors.
6.3 Commercial blood collection tubes containing 3.2%,
Testing with blood from each donor separately or testing with
0.105 M sodium citrate.
pooled blood is acceptable. Blood should be from donors who
6.4 Agitating water bath/incubator, 37 6 2ºC.
have not taken nonsteroidal, anti-inflammatory drugs
(NSAIDs; for example, ibuprofen, aspirin, naproxen),
6.5 Pipettes and tips (non-glass).
acetaminophen, or antithrombotic drugs (for example, heparin,
7. Reagents and Materials
warfarin) for ten days prior to blood collection. Blood should
becollectedinatubecontaining3.2%,0.105Msodiumcitrate
7.1 Cell count controls suitable for hematology analyzer.
(at a ratio of 9:1 v/v blood to sodium citrate ), gently mixed by
7.2 Fresh human whole blood.
inversion, and stored at room temperature until use.
7.3 EDTA (ethylenediaminetetraacetic acid), 500 mM.
Alternatively, blood should be collected in a tube containing a
low concentration of heparin (final concentration approxi-
7.4 Saline, optional.
mately 1 unit/mL). Other blood collection tubes/anticoagulant
7.5 Positive control material (for example, black rubber,
concentrationsmaybeusedifvalidatedforuseinthispractice,
natural rubber latex, and glass).
using two or more moderately thrombogenic positive control
7.6 Negative reference control material (for example, high materials (for example, black rubber, natural rubber latex, and
glass) for the validation.
density polyethylene (HDPE)).
7.7 Marketed comparator device (a legally marketed device NOTE 3—Blood from an individual donor could be used directly to
perform the test, as long as the test is repeated separately with blood
that has similar blood contact and clinical use as the material/
collected from 2 additional donors. Alternatively, blood from 3 donors
device being investigated), optional.
maybepooledforanalyses,butinthiscase,bloodtypemismatchbetween
donors should be avoided. The test report should indicate the type and
8. Hazards
concentration of anticoagulant used, and whether the blood samples were
pooled or not.
8.1 Human blood should be handled according to standard
10.4 For each analysis, a single donor’s blood (or pooled
microbiological practices and techniques (for example, use of
personal protective equipment) and specific required precau- blood, if applicable) will be exposed to the test sample,
negative control reference material, positive control material,
tions (for example, OSHA Bloodborne Pathogen Standard ).
and marketed comparator device (when applicable) to evaluate
9. Sampling, Test Specimens, and Test Units
their impact on platelet and leukocyte counts. It is recom-
mendedtopre-screenthebloodtoensurethebloodparameters
9.1 Each test sample, negative reference control material,
fall within the normal expected range (normal leukocyte count
positive control material, and marketed comparator device (if
3.4 to 8.37 × 10 /µL, normal platelet count 116 to 329 ×
applicable) shall be prepared in triplicate.All test samples and
10 /µL). If the donor blood parameters fall outside of the
controls, with the exception of, the negative control which is
normal expected ranges, blood from another donor should be
used.
U.S. Department of Health and Human Services, Centers for Disease Control
11. Calibration and Standardization
and Prevention, and National Institutes of Health, “Section III—Principles of
Biosafety,” Biosafety in Microbiological and Biomedical Laboratories (BMBL), L.
11.1 Perform daily calibration procedures as in accordance
C. Chosewood and D. E. Wilson, eds. 2009.
with instrument instructions. (Typically the instrument self-
U.S.DepartmentofLabor,“Oc
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: F2888 − 13 F2888 − 19
Standard Test Method Practice for
Platelet Leukocyte Count—An In-Vitro Measure for
Hemocompatibility Assessment of Cardiovascular Materials
This standard is issued under the fixed designation F2888; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method practice assists in the evaluation of cardiovascular device materials for their ability to induce thrombus
formation. Thrombus formation is assessed by means of a reduction in human platelets and leukocytes when consumed by
thrombus after activation on the material surface. This assay may be part of the hemocompatibility evaluation for devices and
materials contacting human blood, as per in accordance with ANSI/AAMI/ISO 10993–4. See also Test Method F2382.
1.2 All safety policies and practices shall be observed during the performance of this test method. practice. All human blood
and any materials that had contact with human blood shall be bagged in a biohazard bag, properly labeled aswith the contents, and
disposed of by appropriate means.
1.3 The human blood should be handled at Biosafety Level 2 (BSL-2) as recommended in the Centers for Disease
Control/National Institutes of Health Manualpublication, Biosafety in Microbiological Laboratories. and Biomedical Laboratories
(BMBL). The human blood donor must have tested negative for Hepatitis B (HBV) and Human Immunodeficiency (HIV) viruses.
The blood should be treated like any patient blood in and handled/manipulated using universalstandard precautions.
NOTE 1—The results of this in-vitro test may not correspond to actual human response.
1.4 The values stated in SI (International System of Units) units are to be regarded as standard. No other units of measurement
are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 78 on Hazards.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
F2382 Test Method for Assessment of Circulating Blood-Contacting Medical Device Materials on Partial Thromboplastin Time
(PTT)
2.2 Other Standards:
ANSI/AAMI/ISO 10993–4 Biological Evaluation of Medical Devices—Part 4: Selection of Tests for Interactions with Blood
Centers for Disease ControlBMBL /National Institutes of Health Manual Biosafety in Microbiological Laboratories, 1993and
Biomedical Laboratories, 5th ed., 2009
3. Summary of Test MethodPractice
3.1 This test method practice identifies materials which are capable of activating blood platelets and leukocytes on their surface
when exposed to freshly drawn human blood and causing the formation of thrombi on the material surface. A significant decrease
This test method practice is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of
Subcommittee F04.16 on Biocompatibility Test Methods.
Current edition approved Feb. 15, 2013Feb. 1, 2019. Published March 2013February 2019. Originally approved in 2013. Last previous edition approved in 2013 as F2888
– 13. DOI: 10.1520/F2888–13.10.1520/F2888–19.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Available from U.S. Department of Health and Human Services, Centers for Disease Control and Prevention (CDC), 1600 Clifton Rd., Atlanta, GA 30329-4027,
http://www.cdc.gov.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
F2888 − 19
in the number of platelets and leukocytes when counted by a blood analyzer is an indication of these cells being entrapped in
thrombi. The materials are exposed to blood immediately after the blood fresh whole blood that is drawn with anticoagulant.
Another anticoagulant is added at the appropriate time (one hour) to stop the reaction. further thrombus formation. Different blood
analyzers may be used.
4. Significance and Use
4.1 The purpose of this test method practice is to determine if medical materials exposed to human whole blood will adversely
affect the thrombus formation has occurred by comparing platelet and leukocyte counts in whole blood. the blood exposed to the
test material relative to the blood cell counts in the control blood that has not been exposed to the test material. A large number
of platelets and leukocytes as part ofbecoming entrapped/incorporated in thrombi adhering to the material will be reflected by a
decrease in their counts in blood. Thrombogenic materials should not be used for cardiovascular medical devices, unless the
purpose of the device is to promote thrombosis.
5. Interferences
5.1 There is potential for interference if the materials of the test tubes used are thrombogenic (for example, glass tubes).
Therefore, polyethylene or polypropylene tubes should be used.
6. Apparatus
6.1 Hematology analyzer capable of determination of a complete blood count.
6.2 Polypropylene test tubes with caps.
6.3 Commercial blood collection tubes containing 3.2 %, 0.105 M sodium citrate.
6.4 Agitating water bath/incubator, 37 6 2ºC.
6.5 Pipettes and tips (non-glass).
7. Reagents and Materials
7.1 Cell count controls suitable for hematology analyzer.
7.2 Fresh whole human whole blood.
7.3 EDTA (ethylenediaminetetraacetic acid), 500 mM.
7.4 Saline, optional.
7.5 Positive reference control material (for example, black rubber, natural rubber latex, black rubber), optional.and glass).
7.6 Positive control material (for example, black rubber, natural rubber latex).
7.6 Negative reference control material (for example, high density polyethylene (HDPE)).
7.7 Marketed comparator device (a legally marketed device that has similar blood contact and clinical use as the material/device
being investigated), optional.
8. Hazards
8.1 Human blood should be treated with the proper universal precautions, including eye wear and laboratory gloves.handled
according to standard microbiological practices and techniques (for example, use of personal protective equipment) and specific
required precautions (for example, OSHA Bloodborne Pathogen Standard ).
9. Sampling, Test Specimens, and Test Units
9.1 Prepare each Each test sample, negative reference material, negative control, and positive control control material, positive
control material, and marketed comparator device (if applicable) shall be prepared in triplicate. All material samples are test
samples and controls, with the exception of, the negative control which is blood only, shall be prepared based on a ratio of 12 cm
of material to 1 mL of blood and placed into polypropylene tubes. Also, three empty test tubes shall be prepared for the negative
control (blood in test tubes without test materials).
9.2 Thirty-six square centimeterscentimetres of the test sample, positive and reference sample and each of the controls are
divided into three 12 cm samples, cut to maximize blood exposure, for triplicate testing.
NOTE 2—If this surface area cannot be achieved with the device, or other volumes of blood are used, the ratio of total surface area to blood volume
should remain at 12:1.12 cm :1 mL.
U.S. Department of Health and Human Services, Centers for Disease Control and Prevention, and National Institutes of Health, “Section III—Principles of Biosafety,”
Biosafety in Microbiological and Biomedical Laboratories (BMBL), L. C. Chosewood and D. E. Wilson, eds. 2009.
U.S. Department of Labor, “Occupational Exposure to Bloodborne Pathogens,” Final Rule, 29 CFR, 1910.1030, 1991.
F2888 − 19
9.3 For each test sample, the percentage of test value (platelet count or leukocyte count) per negative control (blood not exposed
to a material) is calculated as follows:
A⁄ B 3100 5 C (1)
where:
A = average count (platelets or leukocytes),
B = average count (platelets or leukocytes) of negative control, and
C = percentage (%) of negative control.
When applicable, a comparison article should be used in the formula in place of the negative control.
10. Preparation of Apparatus
10.1 Initialize the hematology analyzer and allow it to perform internal self-checks. If no errors are noted, the analyzer is ready
for use.
10.2 To verify the analyzer is functioning properly, prior to analyzing samples, cell count controls shall be run to
conformconfirm that the results fall within the allowable ranges.
10.3 Fresh human blood is drawn from at least 3 donors should be used for the test system. to account for blood variability
among donors. Testing with blood from each donor separately or testing with pooled blood is acceptable. Blood should be from
donors who have not taken aspirin, acetaminophen, naproxen, warfarin, heparin, or ibuprofen for ten days. nonsteroidal,
anti-inflammatory drugs (NSAIDs; for example, ibuprofen, aspirin, naproxen), acetaminophen, or antithrombotic drugs (for
example, heparin, warfarin) for ten days prior to blood collection. Blood should be collected in a tube containing 3.2 %, 0.105 M
sodium citrate (at a ratio of 9:1 v/v blood to sodium citrate as per commercial blood collection tubes), pooled, ), gently mixed by
inversion, and stored on ice until use.at room temperature until use. Alternatively, blood should be collected in a tube containing
a low concentration of heparin (final concentration approximately 1 unit/mL). Other blood collection tubes/anticoagulant
concentrations may be used if validated for use in this practice, using two or more moderately thrombogenic positive control
materials (for example, black rubber, natural rubber latex, and glass) for the validation.
NOTE 3—Blood from an individual donor could be used directly to perform the test, as long as the test is repeated separately with blood collected from
2 additional donors. Alternatively, blood from 3 donors may be pooled for analyses, but in this case, blood type mismatch between donors should be
avoided. The test report should indicate the type and concentration of anticoagulant used, and whether the blood samples were pooled or not.
10.4 For each analysis, a single donor’s blood (or pooled blood, if applicable) will be exposed to the test sample, reference
material, negative control, and positive control negative control reference material, positive control material, and marketed
comparator device (when applicable) to provide a consistent test system to evaluate all reference and test materials.evaluate their
impact on platelet and leukocyte counts. It is recommended to pre-screen the blood to ensure the blood parameters fall within the
3 3
normal expected range (normal leukocyte count 3.4 to 8.37 × 10 /uL,/μL, normal platelet count 116 to 329 × 10 /uL)./μL). If the
donor blood parameters fall outside of the normal expected ranges, blood from another donor should be used.
11. Calibration and Standardization
11.1 Perform daily calibration procedures as per in accordance with instrument instructions. (Typically the instrument
self-calib
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...