Standard Practice for 99 %/95 % Interlaboratory Detection Estimate (IDE) for Analytical Methods with Negligible Calibration Error

SIGNIFICANCE AND USE
Appropriate application of this practice should result in an IDE achievable by most laboratories properly using the test method studied. This IDE provides the basis for any prospective use of the test method by qualified laboratories for reliable detection of low-level concentrations of the same analyte as the one studied in this practice and same media (matrix).
The IDE values may be used to compare the detection power of different methods for analysis of the same analyte in the same matrix.
The IDE provides high probability (approximately 95 %) that result values of the method studied which exceed the IDE represent presence of analyte in the sample and high probability (approximately 99 %) that blank samples will not result in a detection.
The IDE procedure should be used to establish the interlaboratory detection capability for any application of a method where interlaboratory detection is important to data use. The intent of IDE is not to set reporting limits.
SCOPE
1.1 This practice establishes a standard for computing a 99 %/95 % Interlaboratory Detection Estimate (IDE) and provides guidance concerning the appropriate use and application. The calculations involved in this practice can be performed with DQCALC, Microsoft Excel-based software available from ASTM.
1.2 The IDE is computed to be the lowest concentration at which there is 90 % confidence that a single measurement from a laboratory selected from the population of qualified laboratories represented in an interlaboratory study will have a true detection probability of at least 95 % and a true nondetection probability of at least 99 % (when measuring a blank sample).
1.3 The fundamental assumption of the collaborative study is that the media tested, the concentrations tested, and the protocol followed in the study provide a representative and fair evaluation of the scope and applicability of the test method as written. When properly applied, the IDE procedure ensures that the 99 %/95 % IDE has the following properties:
1.3.1 Routinely Achievable IDE Value—Most laboratories are able to attain the IDE detection performance in routine analyses, using a standard measurement system, at reasonable cost. This property is needed for a detection limit to be practically feasible. Representative laboratories must be included in the data to calculate the IDE.
1.3.2 Routine Sources of Error Accounted for—The IDE should realistically include sources of bias and variation which are common to the measurement process. These sources include, but are not limited to: intrinsic instrument noise, some typical amount of carryover error, plus differences in laboratories, analysts, sample preparation, and instruments.
1.3.3 Avoidable Sources of Error Excluded—The IDE should realistically exclude avoidable sources of bias and variation, that is, those which can reasonably be avoided in routine field measurements. Avoidable sources would include, but are not limited to: modifications to the sample, measurement procedure, or measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there was a way to detect and either correct or eliminate them).
1.3.4 Low Probability of False Detection—The IDE is a true concentration consistent with a measured concentration threshold (critical measured value) that will provide a high probability, 99 %, of true nondetection (a low probability of false detection, α = 1 %). Thus, when measuring a blank sample, the probability of not detecting the analyte would be 99 %. To be useful, this must be demonstrated for the particular matrix being used, and not just for reagent water.
1.3.5 Low Probability of False Nondetection—The IDE should be a true concentration at which there is a high probability, at least 95 %, of true detection (a low probability of false nondetection, β = 5 %, at the IDE), with a simultaneous low probability of false detection (see 1.3.4). Thu...

General Information

Status
Historical
Publication Date
28-Feb-2007
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D6091-07 - Standard Practice for 99 %/95 % Interlaboratory Detection Estimate (IDE) for Analytical Methods with Negligible Calibration Error
English language
13 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D6091 − 07 AnAmerican National Standard
Standard Practice for
99 %/95 % Interlaboratory Detection Estimate (IDE) for
1
Analytical Methods with Negligible Calibration Error
This standard is issued under the fixed designation D6091; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope routine field measurements. Avoidable sources would include,
but are not limited to: modifications to the sample, measure-
1.1 This practice establishes a standard for computing a
ment procedure, or measurement equipment of the validated
99%/95% Interlaboratory Detection Estimate (IDE) and pro-
method, and gross and easily discernible transcription errors
videsguidanceconcerningtheappropriateuseandapplication.
(provided there was a way to detect and either correct or
The calculations involved in this practice can be performed
eliminate them).
with DQCALC, Microsoft Excel-based software available
2
1.3.4 Low Probability of False Detection —The IDE is a
from ASTM.
true concentration consistent with a measured concentration
1.2 The IDE is computed to be the lowest concentration at
threshold (critical measured value) that will provide a high
whichthereis90%confidencethatasinglemeasurementfrom
probability, 99%, of true nondetection (a low probability of
a laboratory selected from the population of qualified labora-
false detection, α=1%). Thus, when measuring a blank
tories represented in an interlaboratory study will have a true
sample, the probability of not detecting the analyte would be
detection probability of at least 95% and a true nondetection
99%.Tobeuseful,thismustbedemonstratedfortheparticular
probability of at least 99% (when measuring a blank sample).
matrix being used, and not just for reagent water.
1.3 The fundamental assumption of the collaborative study
1.3.5 Low Probability of False Nondetection—The IDE
is that the media tested, the concentrations tested, and the
should be a true concentration at which there is a high
protocolfollowedinthestudyprovidearepresentativeandfair
probability, at least 95%, of true detection (a low probability
evaluation of the scope and applicability of the test method as of false nondetection, β=5%, at the IDE), with a simultane-
written.Whenproperlyapplied,theIDEprocedureensuresthat
ous low probability of false detection (see 1.3.4). Thus, when
the 99%/95% IDE has the following properties: measuring a sample at the IDE, the probability of detection
1.3.1 Routinely Achievable IDE Value —Most laboratories
wouldbeatleast95%.Tobeuseful,thismustbedemonstrated
are able to attain the IDE detection performance in routine
for the particular matrix being used, and not just for reagent
analyses, using a standard measurement system, at reasonable
water.
cost. This property is needed for a detection limit to be
NOTE 1—The referenced probabilities, α and β, are key parameters for
practically feasible. Representative laboratories must be in-
risk-based assessment of a detection limit.
cluded in the data to calculate the IDE.
1.4 The IDE applies to measurement methods for which
1.3.2 Routine Sources of Error Accounted for—The IDE
calibration error is minor relative to other sources, such as
shouldrealisticallyincludesourcesofbiasandvariationwhich
when the dominant source of variation is one of the following
are common to the measurement process. These sources
(with comment):
include,butarenotlimitedto:intrinsicinstrumentnoise,some
1.4.1 Sample Preparation, and calibration standards do not
typical amount of carryover error, plus differences in
have to go through sample preparation.
laboratories, analysts, sample preparation, and instruments.
1.4.2 Differences in Analysts,andanalystshavelittleoppor-
1.3.3 Avoidable Sources of Error Excluded—The IDE
tunity to affect calibration results (such as with automated
should realistically exclude avoidable sources of bias and
calibration).
variation, that is, those which can reasonably be avoided in
1.4.3 Differences in Laboratories , for whatever reasons,
perhaps difficult to identify and eliminate.
1
This practice is under the jurisdiction ofASTM Committee D19 on Water and
1.4.4 Differences in Instruments (measurement equipment),
is the direct responsibility of Subcommittee D19.02 on Quality Systems,
which could take the form of differences in manufacturer,
Specification, and Statistics.
Current edition approved March 1, 2007. Published April 2007. Originally
model,hardware,electronics,samplingrate,chemicalprocess-
approved in 1997. Last previous edition approved in 2003 as D6091–03. DOI:
ing rate, integration time, software algorithms, interna
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.