Standard Test Method for Determination of Volatile Organic Compounds Emitted from Carpet using a Specific Sorbent Tube and Thermal Desorption / Gas Chromatography

SIGNIFICANCE AND USE
5.1 Manufacturers of carpet need to monitor emissions of VOCs to assess the environmental impact of their products indoors. These results are also used to demonstrate compliance with VOC emission limits for individual VOCs.  
5.2 These data are also used to understand which VOCs are emitted from a product or material and to measure the magnitude of those emissions.  
5.3 Emission data may be used to compare different lots of carpet of the same materials of construction, or carpets composed of different materials of construction, in order to develop products with lower emissions and lower potential environmental impact.  
5.4 This test method should be used in conjunction with practices/guidelines for emissions testing such as Guide D5116, Practice D7143, Practice D7706, ISO 16000-9, and ISO 16000-10. These detail how to select and prepare samples and how and when to carry out emissions tests such that the concentration and profile of vapors in the exhaust air of the emission chamber/cell are representative of the product under test. This test method covers the sampling and analysis of volatile organic compounds in the exhaust gas from the chamber/cell using thermal desorption—compatible sorbent tubes and will provide the necessary analytical consistency to ensure that reproducible data is obtained for the analysis of identical vapor samples by different laboratories.
SCOPE
1.1 This test method describes an analytical procedure for identifying and quantifying the masses of individual volatile organic compounds (individual VOCs or IVOCs) that are emitted into a flow of air from carpet specimens and collected on sorbent sampling tubes during emissions testing.  
1.2 This test method will be used in conjunction with a standard practice for sampling and preparing carpet specimens for emissions testing. If a specific chamber practice is not available for the carpet specimens, this test method should be used in conjunction with approved standard practices for emissions testing and sample preparation.  
1.3 When used in conjunction with standard practices for carpet specimen preparation and collection of vapor-phase emissions , this test method will provide a standardized means of determining the levels of IVOC in the exhaust stream of the emissions test chamber/cell. If this test method is used with a reliable practice for emissions testing, these IVOC levels can be used to determine the emission rate from a unit quantity (usually surface area) of the sample material under test.  
1.4 VOCs in the exhaust stream of an emissions test device are collected on thermal desorption tubes packed with a specific combination of sorbents using active (pumped) sampling. (See Practice D6196 for a more general description of vapor collection using pumped sampling onto sorbent tubes.) The samples are analyzed by thermal desorption (TD) with gas chromatography and mass spectrometry detection (GC/MS) and/or flame ionization detection (FID) depending upon the requirements of the specific materials emissions testing/certification protocol.  
1.5 This test method can be used for the measurement of most GC-compatible organic vapors ranging from the approximate volatility from n-hexane to n-hexadecane (that is, compounds with vapor pressures ranging from 16 kPa to 4 × 10-4 kPa at 25°C). Properties other than a compound’s vapor pressure such as affinity for the sorbent may need to be taken into account. Compounds with vapor pressures outside this range may or may not be quantifiable by this test method. However, qualitative data concerning the identity of a compound(s), outside the stated volatility range for quantitation, may still be useful to the user. This test method can be applied to analytes over a wide concentration range—typically 1 μg/m3 to 1 mg/m3 concentration of vapor in the exhaust air from the emission cell or chamber.  
1.6 This test method is not capable of quantifying all compounds which are emit...

General Information

Status
Published
Publication Date
28-Feb-2018
Technical Committee
D22 - Air Quality
Drafting Committee
D22.05 - Indoor Air

Relations

Effective Date
01-Mar-2018
Effective Date
01-Sep-2023
Effective Date
01-Sep-2020
Effective Date
15-Mar-2020
Effective Date
01-Mar-2020
Effective Date
01-Sep-2018
Effective Date
01-Nov-2017
Effective Date
01-May-2017
Effective Date
01-Mar-2017
Effective Date
01-Oct-2016
Effective Date
01-Nov-2015
Effective Date
15-Oct-2015
Effective Date
01-Jul-2015
Effective Date
01-Dec-2014
Effective Date
01-May-2014

Overview

ASTM D7339-18 is a standard test method developed by ASTM International for the determination of volatile organic compounds (VOCs) emitted from carpet materials. Using a specific sorbent tube and thermal desorption/gas chromatography (TD-GC), this method enables precise identification and quantification of VOCs released from carpets into indoor environments. The standard is primarily used to monitor emissions that may impact indoor air quality and guide manufacturers in achieving compliance with emissions regulations and best practices for product safety.

Key Topics

  • Scope of the Method:
    The test method provides procedures for sampling and analyzing individual VOCs and intermediate VOCs (IVOCs) emitted from carpet specimens. The compounds of interest are collected on sorbent tubes using active (pumped) air sampling, followed by thermal desorption, gas chromatography, and either mass spectrometry (GC/MS) or flame ionization detection (GC/FID).

  • Analytical Consistency:
    By specifying sample collection and analysis protocols, ASTM D7339-18 ensures reproducible results, enabling consistent data across different laboratories.

  • Emissions Data Usage:

    • Supports environmental impact assessments of carpet products.
    • Facilitates compliance with VOC emission limits and indoor air quality standards.
    • Enables comparison between carpet lots or materials to optimize product design for lower emissions.
  • Integration with Other Practices:
    The method is used alongside other emissions testing standards (e.g., ASTM D5116, D7143, and ISO 16000 series) that detail sample selection, preparation, and test conditions to ensure representative and meaningful results.

Applications

  • Product Certification and Compliance:
    The method provides a scientific foundation for manufacturers to verify their carpets meet regulatory VOC emission thresholds and green building requirements.

  • Research and Development:
    Enables carpet engineers and product developers to assess the impact of material changes on VOC emissions and to improve indoor environmental quality by designing low-emitting products.

  • Quality Control:
    Used by quality assurance teams to monitor variations in VOC emissions across production lots, supporting continuous improvement efforts and proactive issue identification.

  • Environmental and Indoor Air Quality Testing:
    Widely adopted by independent testing laboratories to provide third-party certification, support product claims, and aid in risk assessments for building occupants.

Related Standards

  • ASTM D5116: Guide for Small-Scale Environmental Chamber Determinations of Organic Emissions from Indoor Materials/Products
  • ASTM D6196: Practice for Choosing Sorbents, Sampling Parameters and Thermal Desorption Analytical Conditions for Monitoring Volatile Organic Chemicals in Air
  • ASTM D7143: Practice for Emission Cells for the Determination of VOC Emissions from Indoor Materials/Products
  • ASTM D7706: Practice for Rapid Screening of VOC Emissions from Products Using Micro-Scale Chambers
  • ISO 16000-9: Determination of the emission of volatile organic compounds from building products - Emission test chamber method
  • ISO 16000-10: Determination of the emission of VOCs from building products - Emission test cell method
  • USEPA TO-17: Determination of VOCs in Ambient Air Using Active Sampling onto Sorbent Tubes

Practical Value

Adhering to ASTM D7339-18 empowers manufacturers, certification bodies, and laboratory professionals to reliably measure and control VOC emissions from carpets. This not only helps achieve regulatory compliance but also contributes significantly to occupant health and indoor air quality goals in residential and commercial spaces. The standard’s robust, reproducible methodology supports transparent reporting, facilitates innovation in carpet manufacturing, and aligns with global standards for indoor environmental testing.

Buy Documents

Standard

ASTM D7339-18 - Standard Test Method for Determination of Volatile Organic Compounds Emitted from Carpet using a Specific Sorbent Tube and Thermal Desorption / Gas Chromatography

English language (10 pages)
sale 15% off
sale 15% off
Standard

REDLINE ASTM D7339-18 - Standard Test Method for Determination of Volatile Organic Compounds Emitted from Carpet using a Specific Sorbent Tube and Thermal Desorption / Gas Chromatography

English language (10 pages)
sale 15% off
sale 15% off

Frequently Asked Questions

ASTM D7339-18 is a standard published by ASTM International. Its full title is "Standard Test Method for Determination of Volatile Organic Compounds Emitted from Carpet using a Specific Sorbent Tube and Thermal Desorption / Gas Chromatography". This standard covers: SIGNIFICANCE AND USE 5.1 Manufacturers of carpet need to monitor emissions of VOCs to assess the environmental impact of their products indoors. These results are also used to demonstrate compliance with VOC emission limits for individual VOCs. 5.2 These data are also used to understand which VOCs are emitted from a product or material and to measure the magnitude of those emissions. 5.3 Emission data may be used to compare different lots of carpet of the same materials of construction, or carpets composed of different materials of construction, in order to develop products with lower emissions and lower potential environmental impact. 5.4 This test method should be used in conjunction with practices/guidelines for emissions testing such as Guide D5116, Practice D7143, Practice D7706, ISO 16000-9, and ISO 16000-10. These detail how to select and prepare samples and how and when to carry out emissions tests such that the concentration and profile of vapors in the exhaust air of the emission chamber/cell are representative of the product under test. This test method covers the sampling and analysis of volatile organic compounds in the exhaust gas from the chamber/cell using thermal desorption—compatible sorbent tubes and will provide the necessary analytical consistency to ensure that reproducible data is obtained for the analysis of identical vapor samples by different laboratories. SCOPE 1.1 This test method describes an analytical procedure for identifying and quantifying the masses of individual volatile organic compounds (individual VOCs or IVOCs) that are emitted into a flow of air from carpet specimens and collected on sorbent sampling tubes during emissions testing. 1.2 This test method will be used in conjunction with a standard practice for sampling and preparing carpet specimens for emissions testing. If a specific chamber practice is not available for the carpet specimens, this test method should be used in conjunction with approved standard practices for emissions testing and sample preparation. 1.3 When used in conjunction with standard practices for carpet specimen preparation and collection of vapor-phase emissions , this test method will provide a standardized means of determining the levels of IVOC in the exhaust stream of the emissions test chamber/cell. If this test method is used with a reliable practice for emissions testing, these IVOC levels can be used to determine the emission rate from a unit quantity (usually surface area) of the sample material under test. 1.4 VOCs in the exhaust stream of an emissions test device are collected on thermal desorption tubes packed with a specific combination of sorbents using active (pumped) sampling. (See Practice D6196 for a more general description of vapor collection using pumped sampling onto sorbent tubes.) The samples are analyzed by thermal desorption (TD) with gas chromatography and mass spectrometry detection (GC/MS) and/or flame ionization detection (FID) depending upon the requirements of the specific materials emissions testing/certification protocol. 1.5 This test method can be used for the measurement of most GC-compatible organic vapors ranging from the approximate volatility from n-hexane to n-hexadecane (that is, compounds with vapor pressures ranging from 16 kPa to 4 × 10-4 kPa at 25°C). Properties other than a compound’s vapor pressure such as affinity for the sorbent may need to be taken into account. Compounds with vapor pressures outside this range may or may not be quantifiable by this test method. However, qualitative data concerning the identity of a compound(s), outside the stated volatility range for quantitation, may still be useful to the user. This test method can be applied to analytes over a wide concentration range—typically 1 μg/m3 to 1 mg/m3 concentration of vapor in the exhaust air from the emission cell or chamber. 1.6 This test method is not capable of quantifying all compounds which are emit...

SIGNIFICANCE AND USE 5.1 Manufacturers of carpet need to monitor emissions of VOCs to assess the environmental impact of their products indoors. These results are also used to demonstrate compliance with VOC emission limits for individual VOCs. 5.2 These data are also used to understand which VOCs are emitted from a product or material and to measure the magnitude of those emissions. 5.3 Emission data may be used to compare different lots of carpet of the same materials of construction, or carpets composed of different materials of construction, in order to develop products with lower emissions and lower potential environmental impact. 5.4 This test method should be used in conjunction with practices/guidelines for emissions testing such as Guide D5116, Practice D7143, Practice D7706, ISO 16000-9, and ISO 16000-10. These detail how to select and prepare samples and how and when to carry out emissions tests such that the concentration and profile of vapors in the exhaust air of the emission chamber/cell are representative of the product under test. This test method covers the sampling and analysis of volatile organic compounds in the exhaust gas from the chamber/cell using thermal desorption—compatible sorbent tubes and will provide the necessary analytical consistency to ensure that reproducible data is obtained for the analysis of identical vapor samples by different laboratories. SCOPE 1.1 This test method describes an analytical procedure for identifying and quantifying the masses of individual volatile organic compounds (individual VOCs or IVOCs) that are emitted into a flow of air from carpet specimens and collected on sorbent sampling tubes during emissions testing. 1.2 This test method will be used in conjunction with a standard practice for sampling and preparing carpet specimens for emissions testing. If a specific chamber practice is not available for the carpet specimens, this test method should be used in conjunction with approved standard practices for emissions testing and sample preparation. 1.3 When used in conjunction with standard practices for carpet specimen preparation and collection of vapor-phase emissions , this test method will provide a standardized means of determining the levels of IVOC in the exhaust stream of the emissions test chamber/cell. If this test method is used with a reliable practice for emissions testing, these IVOC levels can be used to determine the emission rate from a unit quantity (usually surface area) of the sample material under test. 1.4 VOCs in the exhaust stream of an emissions test device are collected on thermal desorption tubes packed with a specific combination of sorbents using active (pumped) sampling. (See Practice D6196 for a more general description of vapor collection using pumped sampling onto sorbent tubes.) The samples are analyzed by thermal desorption (TD) with gas chromatography and mass spectrometry detection (GC/MS) and/or flame ionization detection (FID) depending upon the requirements of the specific materials emissions testing/certification protocol. 1.5 This test method can be used for the measurement of most GC-compatible organic vapors ranging from the approximate volatility from n-hexane to n-hexadecane (that is, compounds with vapor pressures ranging from 16 kPa to 4 × 10-4 kPa at 25°C). Properties other than a compound’s vapor pressure such as affinity for the sorbent may need to be taken into account. Compounds with vapor pressures outside this range may or may not be quantifiable by this test method. However, qualitative data concerning the identity of a compound(s), outside the stated volatility range for quantitation, may still be useful to the user. This test method can be applied to analytes over a wide concentration range—typically 1 μg/m3 to 1 mg/m3 concentration of vapor in the exhaust air from the emission cell or chamber. 1.6 This test method is not capable of quantifying all compounds which are emit...

ASTM D7339-18 is classified under the following ICS (International Classification for Standards) categories: 97.150 - Floor coverings. The ICS classification helps identify the subject area and facilitates finding related standards.

ASTM D7339-18 has the following relationships with other standards: It is inter standard links to ASTM D7339-12, ASTM D7706-17(2023), ASTM D1356-20a, ASTM D1356-20, ASTM D3686-20, ASTM D6670-18, ASTM D5116-17, ASTM D7706-17, ASTM D7143-17, ASTM D7143-11(2016), ASTM D6196-15, ASTM D1356-15a, ASTM D1356-15, ASTM D1356-14b, ASTM D1356-14a. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ASTM D7339-18 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:D7339 −18
Standard Test Method for
Determination of Volatile Organic Compounds Emitted from
Carpet using a Specific Sorbent Tube and Thermal
Desorption / Gas Chromatography
This standard is issued under the fixed designation D7339; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope kPa at 25°C). Properties other than a compound’s vapor
pressure such as affinity for the sorbent may need to be taken
1.1 This test method describes an analytical procedure for
into account. Compounds with vapor pressures outside this
identifying and quantifying the masses of individual volatile
range may or may not be quantifiable by this test method.
organic compounds (individual VOCs or IVOCs) that are
However, qualitative data concerning the identity of a
emitted into a flow of air from carpet specimens and collected
compound(s), outside the stated volatility range for
on sorbent sampling tubes during emissions testing.
quantitation, may still be useful to the user. This test method
1.2 This test method will be used in conjunction with a
can be applied to analytes over a wide concentration range—
standard practice for sampling and preparing carpet specimens 3 3
typically 1 µg/m to 1 mg/m concentration of vapor in the
for emissions testing. If a specific chamber practice is not
exhaust air from the emission cell or chamber.
available for the carpet specimens, this test method should be
1.6 This test method is not capable of quantifying all
used in conjunction with approved standard practices for
compounds which are emitted from carpets. See the appropri-
emissions testing and sample preparation.
ate test practices/methods for determining other compounds
1.3 When used in conjunction with standard practices for
that are not amenable to analysis by gas chromatography (that
carpet specimen preparation and collection of vapor-phase
is, Test Method D5197 for the determination of aldehydes).
emissions , this test method will provide a standardized means
1.7 Units—The values stated in SI units are to be regarded
of determining the levels of IVOC in the exhaust stream of the
asstandard.Nootherunitsofmeasurementareincludedinthis
emissions test chamber/cell. If this test method is used with a
standard.
reliable practice for emissions testing, these IVOC levels can
be used to determine the emission rate from a unit quantity
1.8 This standard does not purport to address all of the
(usually surface area) of the sample material under test.
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
1.4 VOCs in the exhaust stream of an emissions test device
priate safety, health, and environmental practices and deter-
are collected on thermal desorption tubes packed with a
mine the applicability of regulatory limitations prior to use.
specific combination of sorbents using active (pumped) sam-
1.9 This international standard was developed in accor-
pling. (See Practice D6196 for a more general description of
vapor collection using pumped sampling onto sorbent tubes.) dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Thesamplesareanalyzedbythermaldesorption(TD)withgas
Development of International Standards, Guides and Recom-
chromatography and mass spectrometry detection (GC/MS)
mendations issued by the World Trade Organization Technical
and/or flame ionization detection (FID) depending upon the
Barriers to Trade (TBT) Committee.
requirements of the specific materials emissions testing/
certification protocol.
2. Referenced Documents
1.5 This test method can be used for the measurement of
most GC-compatible organic vapors ranging from the approxi- 2.1 ASTM Standards:
mate volatility from n-hexane to n-hexadecane (that is, com-
D1356Terminology Relating to Sampling and Analysis of
-4
pounds with vapor pressures ranging from 16 kPa to4×10 Atmospheres
This test method is under the jurisdiction of ASTM Committee D22 on Air
Quality and is the direct responsibility of Subcommittee D22.05 on Indoor Air. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved March 1, 2018. Published March 2018. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2007. Last previous edition approved in 2012 as D7339 – 12. Standards volume information, refer to the standard’s Document Summary page on
DOI:10.1520/D7339-18. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7339−18
D3686Practice for Sampling Atmospheres to Collect Or- 3. Terminology
ganic Compound Vapors (Activated Charcoal Tube Ad-
3.1 Definitions—Refer to Terminology D1356 and Practice
sorption Method)
E355 for definitions of terms used in this test method.
D5116Guide for Small-Scale Environmental Chamber De-
4. Summary of Test Method
terminationsofOrganicEmissionsfromIndoorMaterials/
Products
4.1 A sample of the VOCs emitted from a carpet specimen
D5197TestMethodforDeterminationofFormaldehydeand
iscollectedfollowingthepreparationandcollectionguidelines
OtherCarbonylCompoundsinAir(ActiveSamplerMeth-
provided in ISO 10580 or the appropriate chamber/emission
odology)
cell practices/guides. See, for example, Guide D5116 (small
D5337Practice for Flow RateAdjustment of Personal Sam-
chamber), Practice D6670 (full-scale chamber), Practice
pling Pumps
D7706 (micro-scale chamber), Practice D7143 (emission
D6196Practice for Choosing Sorbents, Sampling Param- cells), ISO16000-9 (small chambers), ISO16000-10 (emis-
eters and Thermal Desorption Analytical Conditions for sion cells), and ISO 16000-11 (sample preparation).
Monitoring Volatile Organic Chemicals in Air
4.2 Organicvaporsintheexhauststreamofanemissiontest
D6670Practice for Full-Scale Chamber Determination of
chamber or cell are pumped onto standard thermal desorption
Volatile Organic Emissions from Indoor Materials/
tubes (see Practice D6196) containing ~200 mg of a polyphe-
Products
nyleneoxideresin-basedW-PP(weakporouspolymersorbent)
D7143Practice for Emission Cells for the Determination of
with a short bed (1–2 mm) of quartz wool, a glass frit, or
Volatile Organic Emissions from Indoor Materials/
stainless steel screen (singly or combined) at each end of the
Products
~200 mg of W-PP sorbent. The pump flow rate and sampling
D7706Practice for Rapid Screening of VOC Emissions
time must be controlled (see Practice D6196). The sorbent
from Products Using Micro-Scale Chambers
tubes are then thermally desorbed, in a reverse flow of carrier
E355PracticeforGasChromatographyTermsandRelation-
gas, using an appropriate two-stage desorption apparatus (see
ships
Practice D6196), such that volatile organic compounds are
3 transferred (injected) efficiently into the capillary GC column
2.2 ISO Standards:
for analysis.
ISO 10580Resilient, textile and laminate floor coverings—
Test method for volatile organic compound (VOC) emis- 4.3 GC-compatible organic compounds which are retained
sions bytheW-PPsorbentorquartz/W-PPsorbenttubeduringvapor
collectionandwhichelutebetween n-C and n-C ona100%,
ISO16000-6Determination of volatile organic compounds
6 16
polydimethylsiloxane (PDMS) fused silica capillary column
in indoor and test chamber air by active sampling on
are identified and quantified by gas chromatography/mass
TenaxTAsorbent, thermal desorption and gas chromatog-
spectrometry (see Section 11). Selective ion monitoring, ion
raphy using MS/FID
extraction or spectral de-convolution shall be used to quantify
ISO16000-9IndoorAir—Part9:Determinationoftheemis-
specificvolatileorganiccompounds.Individualcomponentsof
sion of volatile organic compounds from building prod-
interestarequantifiedusingauthenticstandardsofthatparticu-
ucts and furnishings—Emission test chamber method
lar compound. Other compounds are quantified using toluene
ISO16000-10Indoor Air—Part 10: Determination of the
as the surrogate standard reference material (see 11.7.2).
emission of volatile organic compounds from building
NOTE 1—The procedure is similar to that outlined in ISO16000-6.
products and furnishings—Emission test cell method
ISO16000-11Indoor Air—Part 11: Determination of the
5. Significance and Use
emission of volatile organic compounds from building
5.1 Manufacturers of carpet need to monitor emissions of
products and furnishings—Sampling, storage of samples
VOCs to assess the environmental impact of their products
and preparation of test specimens
indoors.Theseresultsarealsousedtodemonstratecompliance
2.3 US EPA Methods:
with VOC emission limits for individual VOCs.
TO-15 Determination of Volatile Organic Compounds
5.2 These data are also used to understand whichVOCs are
(VOCs) in Air Collected in Specially-Prepared Canisters
emitted from a product or material and to measure the
and Analyzed by Gas Chromatography/Mass Spectrom-
magnitude of those emissions.
etry (GC/MS)
5.3 Emission data may be used to compare different lots of
TO-17—Determination of Volatile Organic Compounds in
carpet of the same materials of construction, or carpets
Ambient Air Using Active Sampling onto Sorbent Tubes
composed of different materials of construction, in order to
develop products with lower emissions and lower potential
environmental impact.
Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
5.4 This test method should be used in conjunction with
4th Floor, New York, NY 10036, http://www.ansi.org.
Found in “Compendium of Methods for the Determination of Toxic Organic
practices/guidelines for emissions testing such as Guide
Compounds in Ambient Air,” 2nd Ed., 1999, US. Environmental Protection
D5116, Practice D7143, Practice D7706, ISO16000-9, and
Agency/625/R-96/010b. Available from United States Environmental Protection
ISO16000-10.These detail how to select and prepare samples
Agency (EPA), William Jefferson Clinton Bldg., 1200 Pennsylvania Ave., NW,
Washington, DC 20460, http://www.epa.gov. and how and when to carry out emissions tests such that the
D7339−18
NOTE 3—The use of a secondary back-up tube can serve as a useful
concentration and profile of vapors in the exhaust air of the
check on the breakthrough volume of the primary W-PP or quartz/W-PP
emission chamber/cell are representative of the product under
tubes. Breakthrough should be determined using two sorbent tubes
test. This test method covers the sampling and analysis of
containing the same sorbent and placed in series. Tube performance
volatile organic compounds in the exhaust gas from the
should be addressed by individual laboratory QC programs, see EPA
chamber/cell using thermal desorption—compatible sorbent
Method TO-17 for guidance.
tubes and will provide the necessary analytical consistency to
7.2 Sorbent Tube End Caps for Long-Term Storage—Blank
ensure that reproducible data is obtained for the analysis of
and sampled tubes should be sealed with metal screw-cap
identical vapor samples by different laboratories.
fittings with combined (one-piece) PTFE ferrules for storage
and transportation. If alternate fittings are used, the laboratory
6. Interferences
should determine that they meet storage and transportation
6.1 Organic compounds that have the same or nearly the
stability requirements.
same retention times as the analyte of interest can interfere
duringgaschromatographicanalysis.Highresolutioncapillary NOTE 4—As a quick test that long term storage caps have been fitted
correctly, check the length of the capped tube to make sure the seals are
columnsarerequiredtominimizetheseissues.Artifactscanbe
seated as far down the tube as possible and check that the caps cannot be
generated during sampling and analysis. Interferences can be
pulled off the tubes by hand using reasonable force.
minimized by proper selection of gas chromatographic col-
7.3 Syringes—A precision 10-µL liquid syringe readable to
umnsandconditions,andbystringentconditioningofboththe
0.1 µL.
sorbent tubes and the analytical system before use. Artifacts
may be formed during storage of blank sorbent tubes. This is
7.4 Soap Bubble Meter—A soap bubble flow meter or
minimized by correctly sealing and storing blank and sampled
another suitable calibrated device is required for calibrating
tubes(see7.3and10.1).Theeffectivenessofthesemethodsfor
pump, desorption, and split flows. Follow the manufacturer’s
controlling the potential interferences can be demonstrated by
recommended procedure and or the participating laboratory
properqualityassuranceproceduresincludingtheuseofblanks
QC program. See Practice D3686 for further guidance.
and spiked sampling tubes.
7.5 Thermal Desorption Apparatus—Atwo-stage apparatus
NOTE 2—The inherent artifact levels will vary from sorbent to sorbent
is required for thermally desorbing VOCs retained on the
but are generally at sub-nanogram levels for quartz wool, W-PP and for
sorbent tubes and transferring/injecting them into a gas chro-
MS-GCB (medium to strong graphitized carbon black) type sorbent (see
matograph (GC) in a flow of inert carrier gas. A typical
Practice D6196).
apparatus contains a mechanism for holding the tubes to be
6.2 Compounds of interest that co-elute
desorbedwhiletheyareheatedandpurgedsimultaneouslywith
chromatographically, are not distinguishable when using an
inert carrier gas. The sample flow path through the thermal
FID. Identification and quantification shall be done using a
desorbermustbeconstructedentirelyofinertmaterials(thatis,
mass spectrometer in the selected ion monitoring (SIM) mode,
quartz, fused silica, silica-coated steel, PTFE, etc.), including
or in SCAN mode in combination with post-run processing
all valve components which may come into contact with
using spectral deconvolution, or ion extraction, or both.
samplevapors.Thedesorptiontemperatureandtimeshouldbe
6.3 Even if mass spectroscopy is employed, it may not be
adjustable, as should the carrier gas flow rate. Air must be
possible to uniquely identify individual compounds when
purged from the sample tube and analytical system before heat
similar compounds co-elute exactly (co-maximize) under the
isappliedtopreventsorbentandanalyteoxidation.Noneofthe
analytical conditions selected.
purged air should be allowed to reach the GC column or
6.4 This test method is suitable for sampling and analyzing detector.The apparatus should incorporate a stringent leak test
of every sample (see Note 5) to check flow path integrity
vapor samples ranging up to 95% relative humidity for all
hydrophobic sorbents such as quartz wool, W-PP and MS- before sample desorption/analysis. The secondary focusing
(cold) trap should contain appropriate sorbents and be of
GCB. When less hydrophobic, strong sorbents such as CMS
(carbonized molecular sieves) are used in a secondary (back- sufficient internal diameter to prevent ice from blocking the
flow path during the focusing of humid samples. It should be
up) tube (see Note 3 and Note 7), care should be taken to
reduce the mass of water retained from humid samples (see desorbed in back-flush direction (that is, with carrier gas
flowinginthereversedirectiontothatusedduringthetrapping
Practice D6196).
stage)toensurecompatibilitywithcomponentsoverthewidest
7. Apparatus
possiblevolatilityrange.Theoptionforsamplesplittingshould
7.1 Sorbent Tubes for Pumped Sampling—Sample tubes beavailableduringprimary(tube)desorption,secondary(trap)
(see Practice D6196) packed with 200 mg of W-PP sorbent or desorption or both. Tubes on automated thermal desorption
with a combination of 1 or 2 mm of loosely packed quartz systems must be sealed, both before and after desorption, to
wool, glass frit, or stainless steel screens bracketing 200 mg of protect the integrity of sampled and desorbed (blank) tubes.
W-PP sorbent should be used for collection of the volatile The sample should be desorbed rapidly from the secondary
organic vapors in the exhaust gas from the emission chamber/ focusing trap to ensure efficient transfer to the capillary GC
cell.Analyses of glass or stainless steel spiked tubes indicates column and optimum sensitivity. The sample is routed to the
that similar results are obtained using either glass wool, or gas chromatograph by way of a heated capillary-lined, or
stainless steel frits, as long as the sorbent is in the heated zone silica-lined stainless steel, transfer line. The option of internal
of the thermal desorber (see 12.2). standard addition, whereby gas phase internal standard is
D7339−18
automatically introduced onto the sampling end of sorbent 7.8 Use ordinary laboratory apparatus (for example, volu-
tubes post leak test and before analytical desorption, should metric flasks for preparing standard solutions) as needed.
also be considered as a tool for checking system stability/
8. Reagents and Materials
performance over time (see 11.7.3).
8.1 Unless otherwise stated, all reagents shall conform to
NOTE 5—Leak testing should be carried out under no-flow conditions,
the specifications of the committee on Analytical Reagents of
at low temperature, and at carrier gas column head pressure such that it is
suitably stringent, but does not compromise sample integrity. Tubes that
theAmerican Chemical Society, where such specifications are
fail the leak test should not be analyzed but resealed to await user
available. Other grades may be used, provided that it is
intervention.
ascertainedthatuseofthereagentdoesnotlessentheaccuracy
7.6 Gas Chromatographic (GC) Apparatus:
of the practice.
7.6.1 Gas Chromatograph—Fitted with a mass spectromet-
8.2 Reagents:
ric (MS) detector and with a flame ionization detector (FID) if
8.2.1 Volatile Organic Compounds—To be prepared as liq-
the latter is required (see 4.3). The gas chromatograph should
uidstandardsforcalibration.Thesewouldideallybewithinthe
be capable of split/splitless injections.The MS detector should
volatility range of n-hexane to n-hexadecane (see 1.5) and
be capable of scanning between mass ion 25 and 450.
should reflect the compounds of interest. Toluene must be
7.6.2 Gas Chromatographic Column—A100%polydimeth-
included.
ylsiloxane (PDMS) fused silica capillary column should be
8.2.2 Methanol Dilution Solvent—To be used as the solvent
used. Typical dimensions: 50–60-m long with 0.32-mm I.D.
for preparation of liquid standards of toluene and the volatile
and a 1.0-µm film thickness. Higher speed alternatives with
organiccompoundsofinterest.Themethanolusedshouldbeof
similar phase ratio (narrower I.D. and thinner film) can also be
chromatographicquality,freefromcompoundsco-elutingwith
used. If a more polar column (for example, cyanopropyl-
thecompoundorcompoundsofinterest(see8.2.1).Alternative
phenyl-polymethylsiloxane) is used for these analyses, the
dilutionsolvents,forexample,ethylacetateorcyclohexanecan
testinglaboratorymustdemonstratethattheyobtainrecoveries
be used, particularly if there is a possibility of reaction or
and precision that meet the guidelines outlined in 12.1 and
chromatographic co-elution.
12.2.
8.3 Sorbents—Sorbent tubes, pre-packed with weighed
7.6.3 Effluent Splitting—If both FID and MS detection are
amounts of sorbent, are available commercially. Empty
required by the relevant emissions test protocol, the outlet of
industry-standardsizedtubescanalsobepackedbytheuser.In
the capillary column should be connected to both the FID and
thiscase200mgofW-PPsorbentranginginparticle-sizefrom
MS detector using a conventional, zero or low-dead-volume,
35 to 80 mesh should be weighed into standard sample tubes
capillary effluent splitting device. Connections from each
that are either empty or contain 1-2 mm of loosely packed,
respective detector to the capillary effluent splitter should
clean and non-friable quartz wool, glass frits, or stainless steel
ensure that sufficient effluent from the capillary analytical
screens at the sampling end. New commercial or self-packed
columnisdirectedtotheFIDandtotheMSdetectortoachieve
W-PP sorbent or quartz-W-PP sorbent tubes should be strin-
thedesireddetectionlimits,takingintoaccountthefactthatthe
gently conditioned before initial use. Example conditions for
MS detector operates at vacuum.
cleaning the sorbent tubes are as follows: 320°C under a flow
7.7 Injection Facility for Preparing Standards Purpose-
of >100 mL/min of pure inert gas (helium or nitrogen) for >2
Built—Injectionportsareavailableforintroducingstandardsto
hours, followed by a further 30 minutes at 335°C (see Practice
the sampling end of sorbent tubes in the vapor-phase in a
D6196). Cleaning procedures specified by the tube supplier or
stream of carrier gas. A conventional gas chromatographic
manufacturer should always be followed.
injection port may also be used. Injection port temperatures
between 75°C and 140°C have been shown to give reliable NOTE 6—Less stringent conditions are required for cleaning used tubes
(for example 10 minutes at a temperature ~20°C higher than that to be
resultsinpreparingsamples(see12.2andTable1).Thiscanbe
used for analysis), provided this doesn’t exceed the maximum safe
leftin-situ,oritcanbemountedseparately.Thecarriergasline
temperature for W-PP sorbent. Practically, the analytical desorption
totheinjectorshouldberetained.Thebackoftheinjectionport
temperatureshouldnotbegreaterthat300°Cbecausehighertemperatures
should be adapted if necessary to fit the sampling end of a
may lead to the production of trace amounts of benzene. Tubes used for
tracelevelmonitoring(individualVOClevelsbelow20ng/L)canoftenbe
sorbent tube. This can be done conveniently by means of a
re-used immediately after analysis (that is, without further cleaning).
⁄4-in.compressioncouplingwithaPTFEorgraphitizedvespel
ferrule.Alternatively,commerciallypreparedstandardsmaybe 8.4 Calibration Solutions—Following guidance given in
used. Practice D6196, prepare standard solutions of toluene and
TABLE 1 Average Recoveries for GC/MS Analysis of Six IVOCs spiked into Glass and Stainless Steel Thermal Desorption Tubes
Average % Average % Average % Average % Average % Average %
Recovery Recovery Recovery Recovery Recovery Recovery
Toluene Benzene 4-PCH Styrene 2-EHA Caprolactam
Glass Tube (Glass Wool) 102 97 98 99 120 70
Glass Tube (Glass Frit) 101 94 99 97 119 72
A
Stainless Steel (1) 94 92 89 86 109 66
A
Stainless Steel (2) 98 98 97 93 127 73
A
(1) and (2) refer to 75°C and 140°C preparation temperatures for the spiked tubes. The glass tubes were prepared at 75°C.
D7339−18
other VOCs of interest (for example; styrene, caprolactam and example of an MS calibration curve (area response versus ng
4-phenylcyclohexene) in methanol such thata1to4µL toluene loaded on a thermal desorption tube). Prepare loaded
injection, via the calibration solution loading apparatus (see
tubes by injecting aliquots of standard solutions onto clean
7.7), introduces 20 to 2500 ng of each compound, or an
W-PPsorbent or quartz-W-PPsorbent tubes as follows: Fit the
alternativemassrangeifmoreappropriatetothesamplesbeing
sampling end of the clean sorbent tube into the injection unit
tested. In any event, the lowest and highest concentration
(see 7.7) through which inert purge gas is typically passing at
standards must be prepared such that the mass of the analyte
50-100 mL/min and introducea1to4µL aliquot of an
introducedinthehighestlevelstandardisatleastafactorof20
appropriate standard solution injected through the septum.
higher than that introduced in the lowest standard.
After 1 to 5 minutes, disconnect the tube and seal it using long
termstoragecaps(see7.3).Thismulti-levelcalibrationshallbe
9. Tuning and Calibration of the TD-GC/MS(FID)
carried out when the calibration check (see 9.4 and 9.5) falls
Analytical System
outside the specified range. The MS response factors shall
9.1 Tuning and mass standardization is performed in accor-
agree within 20% across all calibration levels or demonstrate
dance with the manufacturer’s instructions, generally using
an R value of >0.995.
perfluorotributylamine (FC-43) commonly known as PFTBA.
9.3 The use of a linear regression of the standards for
This process may vary among instruments. Consult the tuning
quantification is recommended. Use of a linear regression of
instructions for the specific instrument being used for the
the standard responses does not eliminate the possibility of
appropriate tune conditions. The FC-43 is introduced directly
into the ion source through a molecular leak. Instrument large systematic errors if the intercept value is large relative to
parameters are adjusted to give acceptable relative ion abun- the quantity being measured. The analyst needs to be aware of
dances. An example of the ranges for the relative ion abun-
unusuallylargeinterceptswhichmaybiastheresults.Typically
dances based on the NIST02 mass spectral database are given
observedinterceptvaluesfortoluene,benzene,styrene,4PCH,
below.
caprolactam, and 4-ethylhexanoic acid have ranged from 5 to
50 nanograms.
Mass % Relative Abundance
69 100
131 25–55 9.4 Split or splitless injections can be used for these
219 45–75
analyses. In either case, care must be taken to ensure that the
502 3–7
calibration standards do not exceed the capacity of the sorbent
The mass range scanned is 35 amu to 450 amu with an
material, column, or the detector. Appendix X1 gives sug-
allowable scan rate >0.5 Hz.The mass spectrometer should be
gestedsplitconditions,howevertheanalystmayneedtoadjust
tuned prior to an initial calibration, after vacuum is broken
the split flow (or lack thereof) to accommodate the sensitivity
(column change, new source installed, etc.), and if the calibra-
needed,thebreakthroughvolumeofthecompoundsofinterest,
tion check has failed.
or the capacity of the column and/or detector.
9.2 StandardcalibrationcurveswillberequiredforMS(and
9.5 If
...


This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: D7339 − 12 D7339 − 18
Standard Test Method for
Determination of Volatile Organic Compounds Emitted from
Carpet using a Specific Sorbent Tube and Thermal
Desorption / Gas Chromatography
This standard is issued under the fixed designation D7339; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method describes an analytical procedure for identifying and quantifying the masses of individual volatile organic
compounds (individual VOCs or IVOCs) that are emitted into a flow of air from carpet specimens and collected on sorbent
sampling tubes during emissions testing.
1.2 This test method will be used in conjunction with a standard practice for sampling and preparing carpet specimens for
emissions testing. If a specific chamber practice is not available for the carpet specimens, this standard test method should be used
in conjunction with approved standard practices for emissions testing and sample preparation.
1.3 When used in conjunction with standard practices for carpet specimen preparation and collection of vapor-phase emissions
, this test method will provide a standardized means of determining the levels of IVOC in the exhaust stream of the emissions test
chamber/cell. If this test method is used with a reliable practice for emissions testing, these IVOC levels can be used to determine
the emission rate from a unit quantity (usually surface area) of the sample material under test.
1.4 VOCs in the exhaust stream of an emissions test device are collected on thermal desorption tubes packed with a specific
combination of sorbents using active (pumped) sampling. (See Practice D6196 for a more general description of vapor collection
using pumped sampling onto sorbent tubes).tubes.) The samples are analyzed by thermal desorption (TD) with gas chromatography
and mass spectrometry detection (GC/MS) and/or flame ionization detection (FID) depending upon the requirements of the specific
materials emissions testing/certification protocol.
1.5 This test method can be used for the measurement of most GC-compatible organic vapors ranging from the approximate
-4
volatility from n-hexane to n-hexadecane (that is, compounds with vapor pressures ranging from 16 kPa to 4 × 10 kPa at 25°C).
Properties other than a compound’s vapor pressure such as affinity for the sorbent may need to be taken into account. Compounds
with vapor pressures outside this range may or may not be quantifiable by this test method. However, qualitative data concerning
the identity of a compound(s), outside the stated volatility range for quantitation, may still be useful to the user. The This test
3 3
method can be applied to analytes over a wide concentration range—typically 1 μg/m to 1 mg/m concentration of vapor in the
exhaust air from the emission cell or chamber.
1.6 This test method is not capable of quantifying all compounds which are emitted from carpets. See the appropriate test
practices/methods for determining other compounds that are not amenable to analysis by gas chromatography (that is, Test Method
D5197 for the determination of aldehydes).
1.7 Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this
standard.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.9 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
This test method is under the jurisdiction of ASTM Committee D22 on Air Quality and is the direct responsibility of Subcommittee D22.05 on Indoor Air.
Current edition approved Oct. 15, 2012March 1, 2018. Published November 2012March 2018. Originally approved in 2007. Last previous edition approved in 20072012
as D7339 – 07. DOI:10.1520/D7339-12.12. DOI:10.1520/D7339-18.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D7339 − 18
2. Referenced Documents
2.1 ASTM Standards:
D1356 Terminology Relating to Sampling and Analysis of Atmospheres
D3686 Practice for Sampling Atmospheres to Collect Organic Compound Vapors (Activated Charcoal Tube Adsorption Method)
D5116 Guide for Small-Scale Environmental Chamber Determinations of Organic Emissions from Indoor Materials/Products
D5197 Test Method for Determination of Formaldehyde and Other Carbonyl Compounds in Air (Active Sampler Methodology)
D5337 Practice for Flow Rate Adjustment of Personal Sampling Pumps
D6196 Practice for Choosing Sorbents, Sampling Parameters and Thermal Desorption Analytical Conditions for Monitoring
Volatile Organic Chemicals in Air
D6670 Practice for Full-Scale Chamber Determination of Volatile Organic Emissions from Indoor Materials/Products
D7143 Practice for Emission Cells for the Determination of Volatile Organic Emissions from Indoor Materials/Products
D7706 Practice for Rapid Screening of VOC Emissions from Products Using Micro-Scale Chambers
E355 Practice for Gas Chromatography Terms and Relationships
2.2 ISO Standards:
ISO 10580 Resilient, textile and laminate floor coverings—Test method for volatile organic compound (VOC) emissions
ISO 16000-6 Determination of volatile organic compounds in indoor and test chamber air by active sampling on Tenax TA
sorbent, thermal desorption and gas chromatography using MS/FID
ISO 16000-9 Indoor Air—Part 9: Determination of the emission of volatile organic compounds from building products and
furnishings—Emission test chamber method
ISO 16000-10 Indoor Air—Part 10: Determination of the emission of volatile organic compounds from building products and
furnishings—Emission test cell method
ISO 16000-11 Indoor Air—Part 11: Determination of the emission of volatile organic compounds from building products and
furnishings—Sampling, storage of samples and preparation of test specimens
2.3 US EPA Methods:
TO-15 Determination of Volatile Organic Compounds (VOCs) in Air Collected in Specially-Prepared Canisters and Analyzed
by Gas Chromatography/Mass Spectrometry (GC/MS)
TO-17 —Determination of Volatile Organic Compounds in Ambient Air Using Active Sampling onto Sorbent Tubes
3. Terminology
3.1 Definitions—Refer to Terminology D1356 and Practice E355 for definitions of terms used in this test method.
4. Summary of Test Method
4.1 A sample of the VOCs emitted from a carpet specimen is collected following the preparation and collection guidelines
provided in ISO 10580 or the appropriate chamber/emission cell practices/guides. See, for example, Guide D5116 (small chamber),
Practice D6670 (full-scale chamber), Practice D7706 (micro-scale chamber), Practice D7143 (emission cells), ISO 16000-9 (small
chambers), ISO 16000-10 (emission cells), and ISO 16000-11 (sample preparation).
4.2 Organic vapors in the exhaust stream of an emission test chamber or cell are pumped onto standard thermal desorption tubes
(see Practice D6196) containing ~200 mg of a polyphenylene oxide resin-based (PPOR-B) sorbent W-PP (weak porous polymer
sorbent) with a short bed (1-2(1–2 mm) of quartz wool, a glass frit, or stainless steel screen (singly or combined) at each end of
the ~200 mg of PPOR-BW-PP sorbent. The pump flow rate and sampling time must be controlled (see Practice D6196). The
sorbent tubes are then thermally desorbed, in a reverse flow of carrier gas, using an appropriate two-stage desorption apparatus,
(Seeapparatus (see Practice D6196)), such that volatile organic compounds are transferred (injected) efficiently into the capillary
GC column for analysis.
4.3 GC-compatible organic compounds which are retained by the PPOR-BW-PP sorbent or quartz/PPOR-Bquartz/W-PP sorbent
tube during vapor collection and which elute between n-C and n-C on a 100 %, polydimethylsiloxane (PDMS) fused silica
6 16
capillary column are identified and quantified by gas chromatography/mass spectrometry (see Section 11). Selective ion
monitoring, ion extraction or spectral de-convolution shall be used to quantify specific volatile organic compounds. Individual
components of interest are quantified using authentic standards of that particular compound. Other compounds are quantified using
toluene as the surrogate standard reference material (see 11.7.2).
NOTE 1—The procedure is similar to that outlined in ISO 16000-6.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New York, NY 10036, http://www.ansi.org.
Found in “Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air,” 2nd Ed., 1999, US. Environmental Protection
Agency/625/R-96/010b. Available from United States Environmental Protection Association (EPA), Ariel Rios Agency (EPA), William Jefferson Clinton Bldg., 1200
Pennsylvania Ave., NW, Washington, DC 20460, http://www.epa.gov.
D7339 − 18
5. Significance and Use
5.1 Manufacturers of carpet need to monitor emissions of VOCs to assess the environmental impact of their products indoors.
These results are also used to demonstrate compliance with VOC emission limits for individual VOCs.
5.2 These data are also used to understand which VOCs are emitted from a product or material and to measure the magnitude
of those emissions.
5.3 Emission data may be used to compare different lots of carpet of the same materials of construction, or carpets composed
of different materials of construction, in order to develop products with lower emissions and lower potential environmental impact.
5.4 This test method should be used in conjunction with practices/guidelines for emissions testing such as Guide D5116,
Practice D7143, Practice D7706, ISO 16000-9, and ISO 16000-10. These detail how to select and prepare samples and how and
when to carry out emissions tests such that the concentration and profile of vapors in the exhaust air of the emission chamber/cell
are representative of the product under test. This standardtest method covers the sampling and analysis of volatile organic
compounds in the exhaust gas from the chamber/cell using thermal desorption—compatible sorbent tubes and will provide the
necessary analytical consistency to ensure that reproducible data is obtained for the analysis of identical vapor samples by different
laboratories.
6. Interferences
6.1 Organic compounds that have the same or nearly the same retention times as the analyte of interest can interfere during gas
chromatographic analysis. High resolution capillary columns are required to minimize these issues. Artifacts can be generated
during sampling and analysis. Interferences can be minimized by proper selection of gas chromatographic columns and conditions,
and by stringent conditioning of both the sorbent tubes and the analytical system before use. Artifacts may be formed during
storage of blank sorbent tubes. This is minimized by correctly sealing and storing blank and sampled tubes (see 7.3 and 10.1). The
effectiveness of these methods for controlling the potential interferences can be demonstrated by proper quality assurance
procedures including the use of blanks and spiked sampling tubes.
NOTE 2—Note that The inherent artifact levels will vary from sorbent to sorbent but are generally at sub-nanogram levels for quartz wool, PPOR-B
and for carbon-black W-PP and for MS-GCB (medium to strong graphitized carbon black) type sorbent (see Practice D6196).
6.2 Compounds of interest that co-elute chromatographically, are not distinguishable when using an FID. Identification and
quantification shall be done using a mass spectrometer in the selected ion monitoring (SIM) mode, or in SCAN mode in
combination with post-run processing using spectral deconvolution, or ion extraction, or both.
6.3 Even if mass spectroscopy is employed, it may not be possible to uniquely identify individual compounds when similar
compounds co-elute exactly (co-maximize) under the analytical conditions selected.
6.4 The This test method is suitable for sampling and analyzing vapor samples ranging up to 95 % relative humidity for all
hydrophobic sorbents such as quartz wool, PPOR-B and graphitized carbon blacks. W-PP and MS-GCB. When less hydrophobic,
strong sorbents such as carbonized CMS (carbonized molecular sievessieves) are used in a secondary (back-up) tube,tube (see Note
3 and Note 87)), care mustshould be taken to reduce the mass of water retained from humid samples (see Practice D6196).
7. Apparatus
7.1 Sorbent Tubes for Pumped Sampling—Sample tubes (see Practice D6196) packed with 200 mg of PPOR-BW-PP sorbent
or with a combination of 1 or 2 mm of loosely packed quartz wool, glass frit, or stainless steel screens bracketing 200 mg of
PPOR-BW-PP sorbent should be used for collection of the volatile organic vapors in the exhaust gas from the emission
chamber/cell. Analyses of glass or stainless steel spiked tubes indicates that similar results are obtained using either glass wool,
or stainless steel frits, as long as the sorbent is in the heated zone of the thermal desorber (see 12.2).
NOTE 3—Note that The use of a secondary back-up tube can serve as a useful check on the breakthrough volume of the primary PPOR-BW-PP or
quartz/PPOR-Bquartz/W-PP tubes. Breakthrough should be determined using two sorbent tubes containing the same sorbent and placed in series. Tube
performance should be addressed by individual laboratory QC programs, see EPA Method TO-17 for guidance.
7.2 Sorbent Tube End Caps for Long-termLong-Term Storage—Blank and sampled tubes should be sealed with metal screw-cap
fittings with combined (one-piece) PTFE ferrules for storage and transportation. If alternate fittings are used, the laboratory should
determine that they meet storage and transportation stability requirements.
NOTE 4—As a quick test that long term storage caps have been fitted correctly, check the length of the capped tube to make sure the seals are seated
as far down the tube as possible and check that the caps cannot be pulled off the tubes by hand using reasonable force.
7.3 Syringes—A precision 10–μL10-μL liquid syringe readable to 0.1 μL.
7.4 Soap Bubble Meter—A soap bubble flow meter or another suitable calibrated device is required for calibrating pump,
desorption, and split flows. Follow the manufacturer’s recommended procedure and or the participating laboratory QC program.
See Practice D3686 for further guidance.
D7339 − 18
7.5 Thermal Desorption Apparatus—A two-stage apparatus is required for thermally desorbing VOCs retained on the sorbent
tubes and transferring/injecting them into a gas chromatograph (GC) in a flow of inert carrier gas. A typical apparatus contains a
mechanism for holding the tubes to be desorbed while they are heated and purged simultaneously with inert carrier gas. The sample
flow path through the thermal desorber must be constructed entirely of inert materials (that is, quartz, fused silica, silica-coated
steel, PTFE, etc.), including all valve components which may come into contact with sample vapors. The desorption temperature
and time should be adjustable, as should the carrier gas flow rate. Air must be purged from the sample tube and analytical system
before heat is applied to prevent sorbent and analyte oxidation. None of the purged air should be allowed to reach the GC column
or detector. The apparatus should incorporate a stringent leak test of every sample (see Note 5) to check flow path integrity before
sample desorption/analysis. The secondary focusing (cold) trap should contain appropriate sorbents and be of sufficient internal
diameter to prevent ice from blocking the flow path during the focusing of humid samples. It should be desorbed in back-flush
direction (that is, with carrier gas flowing in the reverse direction to that used during the trapping stage) to ensure compatibility
with components over the widest possible volatility range. The option for sample splitting should be available during primary
(tube) desorption, secondary (trap) desorption or both. Tubes on automated thermal desorption systems must be sealed, both before
and after desorption, to protect the integrity of sampled and desorbed (blank) tubes. The sample should be desorbed rapidly from
the secondary focusing trap to ensure efficient transfer to the capillary GC column and optimum sensitivity. The sample is routed
to the gas chromatograph by way of a heated capillary-lined, or silica-lined stainless steel, transfer line. The option of internal
standard addition, whereby gas phase internal standard is automatically introduced onto the sampling end of sorbent tubes post leak
test and before analytical desorption, should also be considered as a tool for checking system stability/performance over time (see
11.7.3).
NOTE 5—Note that leak Leak testing should be carried out under no-flow conditions, at low temperature, and at carrier gas column head pressure such
that it is suitably stringent, but does not compromise sample integrity. Tubes that fail the leak test should not be analyzed but resealed to await user
intervention.
7.6 Gas Chromatographic (GC) Apparatus:
7.6.1 Gas Chromatograph—Fitted with a mass spectrometric (MS) detector and with a flame ionization detector (FID) if the
latter is required (see 4.3). The gas chromatograph should be capable of split/splitless injections. The MS detector should be
capable of scanning between mass ion 25 and 450.
7.6.2 Gas Chromatographic Column—A100 % polydimethylsiloxane (PDMS) fused silica capillary column should be used.
Typical dimensions: 50 – 60-m long with 0.32-mm I.D. and a 1.0-μm film thickness. Higher speed alternatives with similar phase
ratio (narrower I.D. and thinner film) can also be used. If a more polar column (for example, cyanopropyl-phenyl-
polymethylsiloxane) is used for these analyses, the testing laboratory must demonstrate that they obtain recoveries and precision
that meet the guidelines outlined in 12.1 and 12.2.
7.6.3 Effluent Splitting—If both FID and MS detection are required by the relevant emissions test protocol, the outlet of the
capillary column should be connected to both the FID and MS detector using a conventional, zero or low-dead-volume, capillary
effluent splitting device. Connections from each respective detector to the capillary effluent splitter should ensure that sufficient
effluent from the capillary analytical column is directed to the FID and to the MS detector to achieve the desired detection limits,
taking into account the fact that the MS detector operates at vacuum.
7.7 Injection Facility for Preparing Standards Purpose-Built—Injection ports are available for introducing standards to the
sampling end of sorbent tubes in the vapor-phase in a stream of carrier gas. A conventional gas chromatographic injection port may
also be used. Injection port temperatures between 75°C and 140°C have been shown to give reliable results in preparing samples
(see 12.2 and Table 1). This can be left in-situ, or it can be mounted separately. The carrier gas line to the injector should be
retained. The back of the injection port should be adapted if necessary to fit the sampling end of a sorbent tube. This can be done
conveniently by means of a ⁄4-in. compression coupling with a PTFE or graphitized vespel ferrule. Alternatively, commercially
prepared standards may be used.
7.8 Use ordinary laboratory apparatus (for example, volumetric flasks for preparing standard solutions) as needed.
8. Reagents and Materials
8.1 Unless otherwise stated, all reagents shall conform to the specifications of the committee on Analytical Reagents of the
American Chemical Society, where such specifications are available. Other grades may be used, provided that it is ascertained that
use of the reagent does not lessen the accuracy of the practice.
TABLE 1 Average Recoveries for GC/MS Analysis of Six IVOCs spiked into Glass and Stainless Steel Thermal Desorption Tubes
Average % Average % Average % Average % Average % Average %
Recovery Recovery Recovery Recovery Recovery Recovery
Toluene Benzene 4-PCH Styrene 2-EHA Caprolactam
Glass Tube (Glass Wool) 102 97 98 99 120 70
Glass Tube (Glass Frit) 101 94 99 97 119 72
A
Stainless Steel (1) 94 92 89 86 109 66
A
Stainless Steel (2) 98 98 97 93 127 73
A
(1) and (2) refer to 75°C and 140°C preparation temperatures for the spiked tubes. The glass tubes were prepared at 75°C.
D7339 − 18
8.2 Reagents:
8.2.1 Volatile Organic Compounds—To be prepared as liquid standards for calibration. These would ideally be within the
volatility range of n-hexane to n-hexadecane (see 1.5) and should reflect the compounds of interest. Toluene must be included.
8.2.2 Methanol Dilution Solvent—To be used as the solvent for preparation of liquid standards of toluene and the volatile
organic compounds of interest. The methanol used should be of chromatographic quality, free from compounds co-eluting with the
compound or compounds of interest (see 8.2.1). Alternative dilution solvents, for example, ethyl acetate or cyclohexane can be
used, particularly if there is a possibility of reaction or chromatographic co-elution.
8.3 Sorbents—Sorbent tubes, pre-packed with weighed amounts of sorbent, are available commercially. Empty industry-
standard sized tubes can also be packed by the user. In this case 200 mg of PPOR-BW-PP sorbent ranging in particle-size from
35 to 80 mesh should be weighed into standard sample tubes that are either empty or contain 1-2 mm of loosely packed, clean and
non-friable quartz wool, glass frits, or stainless steel screens at the sampling end. New commercial or self-packed PPOR-BW-PP
sorbent or quartz-PPOR-Bquartz-W-PP sorbent tubes should be stringently conditioned before initial use. Example conditions for
cleaning the sorbent tubes are as follows: 320°C under a flow of >100 mL/min of pure inert gas (helium or nitrogen) for >2 hours,
followed by a further 30 minutes at 335°C (see Practice D6196). Cleaning procedures specified by the tube supplier or
manufacturer should always be followed.
NOTE 6—Note that much less Less stringent conditions are required for cleaning used tubes (for example 10 minutes at a temperature ~20°C higher
than that to be used for analysis), provided this doesn’t exceed the maximum safe temperature for PPOR-BW-PP sorbent. Practically, the analytical
desorption temperature should not be greater that 300°C because higher temperatures may lead to the production of trace amounts of benzene. Tubes used
for trace level monitoring (individual VOC levels below 20 ng/L) can often be re-used immediately after analysis,analysis (that is, without further
cleaning).
8.4 Calibration Solutions—Following guidance given in Practice D6196, prepare standard solutions of toluene and other VOCs
of interest (for example; styrene, caprolactam and 4-phenylcyclohexene) in methanol such that a 1 to 4 μL injection, via the
calibration solution loading apparatus (see 7.7), introduces 20 to 2500 ng of each compound, or an alternative mass range if more
appropriate to the samples being tested. In any event, the lowest and highest concentration standards must be prepared such that
the mass of the analyte introduced in the highest level standard is at least a factor of 20 higher than that introduced in the lowest
standard.
9. Tuning and Calibration of the TD-GC/MS(FID) Analytical System
9.1 Tuning and mass standardization is performed in accordance with the manufacturer’s instructions, generally using
perfluorotributylamine (FC-43) commonly known as PFTBA. This process may vary among instruments. Consult the tuning
instructions for the specific instrument being used for the appropriate tune conditions. The FC-43 is introduced directly into the
ion source through a molecular leak. Instrument parameters are adjusted to give acceptable relative ion abundances. An example
of the ranges for the relative ion abundances based on the NIST02 mass spectral database are given below.
Mass % Relative Abundance
69 100
131 25-55
131 25–55
219 45-75
219 45–75
502 3-7
502 3–7
The mass range scanned is 35 amu to 450 amu with an allowable scan rate >0.5 Hz. The mass spectrometer should be tuned
prior to an initial calibration, after vacuum is broken (column change, new source installed, etc.), and if the calibration check has
failed.
9.2 Standard calibration curves will be required for MS (and on FID, if required by the relevant materials emissions testing
protocol) for each compound of interest comprising at least five points within the range of 20 to 2500 ng of each compound. The
standard concentrations distribution should not be greater than five times the concentration of the next lower standard (for example
20, 100, 500, 1750, 2500 ng). See Fig. 1 for a typical example of an MS calibration curve (area response versus ng toluene loaded
on a thermal desorption tube). Prepare loaded tubes by injecting aliquots of standard solutions onto clean PPOR-BW-PP sorbent
or quartz-PPOR-Bquartz-W-PP sorbent tubes as follows: Fit the sampling end of the clean sorbent tube into the injection unit (see
7.7) through which inert purge gas is typically passing at 50-100 mL/min and introduce a 1 to 4 μL aliquot of an appropriate
standard solution injected through the septum. After 1 to 5 minutes, disconnect the tube and seal it using long term storage caps
(see 7.3). This multi-level calibration shall be carried out when the calibration check (see 9.4 and 9.5) falls outside the specified
range. The MS response factors shall agree within 20 % across all calibration levels or demonstrate an R value of >0.995.
9.3 The use of a linear regression of the standards for quantification is recommended. Use of a linear regression of the standard
responses does not eliminate the possibility of large systematic errors if the intercept value is large relative to the quantity being
measured. The analyst needs to be aware of unusually large intercepts which may bias the results. Typically observed intercept
values for toluene, benzene, styrene, 4PCH, caprolactam, and 4-ethylhexanoic acid have ranged from 5 to 50 nanograms.
D7339 − 18
FIG. 1 Toluene MS Calibration Curve
9.4 Split or splitless injections can be used for these analyses. In either case, care must be taken to ensure that the calibration
standards do not exceed the capacity of the sorbent material, column, or the detector. Appendix X1 gives suggested split conditions,
however the analyst may need to adjust the split flow (or lack thereof) to accommodate the sensitivity needed, the breakthrough
volume of the compounds of interest, or the capacity of the column and/or detector.
9.5 If an FID is used, a single, mid-point calibration check of detector response will be carried out before and after the analysis
of a series of sorbent tubes collected during emissions testing of a single sample of carpet. The FID response factor shall agree
within 10 % to the average of that obtained from all five standards during the most recent multi-level calibration. If this is not the
case, the five-level calibration curves must be repeated before proceeding with the analysis.
9.6 A check on MS response for each target VOC will be carried out at the midpoint level of the
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...