ASTM E1319-23
(Guide)Standard Guide for High-Temperature Static Strain Measurement
Standard Guide for High-Temperature Static Strain Measurement
SIGNIFICANCE AND USE
4.1 The use of this guide is voluntary and is intended for use as a procedures guide for selection and application of specific types of strain gages for high-temperature installations. No attempt is made to restrict the type of strain gage types or concepts to be chosen by the user. The provisions of this guide may be invoked in specifications and procedures by specifying those that shall be considered mandatory for the purpose of the specific application. When so invoked, the user shall include in the work statement a notation that provisions of this guide shown as recommendation shall be considered mandatory for the purposes of the specification or procedure concerned, and shall include a statement of any exceptions to or modifications of the affected provisions of this guide.
SCOPE
1.1 This guide covers the selection and application of strain gages for the measurement of static strain up to and including the temperature range from 425 °C to 650 °C (800 °F to 1200 °F). This guide reflects some state-of-the-art techniques in high-temperature strain measurement.
1.2 This guide assumes that the user is familiar with the use of bonded strain gages and associated signal conditioning circuits and instrumentation as discussed in (1) and (2).2 The strain gage systems described are those that have proven effective in the temperature range of interest and were available at the time of issue of this guide. It is not the intent of this guide to limit the user to one of the strain gage types described nor is it the intent to specify the type of strain gage system to be used for a specific application. However, in using any strain gage system including those described, the proposer shall be able to demonstrate the capability of the proposed strain gage system to meet the selection criteria provided in Section 5 and the needs of the specific application.
1.3 The devices and techniques described in this guide can sometimes be applicable at temperatures above and below the range noted, and for making dynamic strain measurements at high temperatures with proper precautions. The strain gage manufacturer should be consulted for recommendations and details of such applications.
1.4 The references are a part of this guide to the extent specified in the text.
1.5 The values stated in metric (SI) units are to be regarded as the standard. The values given in parentheses are for informational purposes only.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
General Information
- Status
- Published
- Publication Date
- 30-Apr-2023
- Technical Committee
- E28 - Mechanical Testing
- Drafting Committee
- E28.01 - Calibration of Mechanical Testing Machines and Apparatus
Relations
- Effective Date
- 15-May-2009
- Effective Date
- 15-May-2009
- Effective Date
- 01-Apr-2009
- Effective Date
- 01-Jan-2009
- Effective Date
- 01-Oct-2008
- Effective Date
- 01-Feb-2008
- Effective Date
- 01-Nov-2007
- Effective Date
- 01-Jun-2007
- Effective Date
- 01-Mar-2007
- Effective Date
- 01-Apr-2006
- Effective Date
- 10-Jun-2003
- Effective Date
- 10-Jun-2003
- Effective Date
- 10-Aug-2002
- Effective Date
- 10-Jul-1999
Overview
ASTM E1319-23, Standard Guide for High-Temperature Static Strain Measurement, is a procedural guide developed by ASTM International to assist users in the selection and application of strain gages for static strain measurement at elevated temperatures. This guide is particularly relevant for applications that require reliable strain measurement in the temperature range of 425 °C to 650 °C (800 °F to 1200 °F). The document outlines criteria for selecting appropriate strain gage systems, recommends installation practices, and highlights considerations specific to high-temperature environments. It provides state-of-the-art techniques, but allows flexibility in the choice of gage types and system concepts based on the needs of the user.
Key Topics
Strain Gage Types
- Resistance strain gages (including bonded weldable, hermetic weldable)
- Capacitive and variable capacitance strain gages
Selection Criteria
- Suitability for temperature range (425 °C to 650 °C)
- Operating environment (including pressure, vibration, radiation, and humidity)
- Expected strain range, resolution, and accuracy
- Thermal compensation and stability during transients
- Duration and nature of the test
Strain Gage System Considerations
- Lead wires, connectors, and environmental protection
- Signal conditioning circuits and bridge completion
- Pre-test and post-test calibration processes
- Space constraints and potential reinforcement effects
Installation and Calibration
- Use of capacitive-discharge welding for gage attachment
- Importance of proper calibration procedures before, during, and after testing
Applications
The ASTM E1319-23 guide is valuable for a broad range of industries and applications requiring precise strain measurement at high temperatures, such as:
- Aerospace and Aviation: Monitoring strain in engine components, turbine blades, and structural parts exposed to extreme heat
- Power Generation: Testing boiler tubes, steam lines, and structural supports in fossil-fuel and nuclear plants
- Materials Research: Evaluating metal and alloy behavior under high-temperature loading, including creep and fatigue
- Automotive and Heavy Industry: Analysis of exhaust systems, high-performance components, and thermal stress evaluation in critical parts
The guide allows engineers, researchers, and quality assurance professionals to confidently specify, install, and interpret data from high-temperature strain gage systems, ensuring reliable and accurate static strain measurements under challenging thermal conditions.
Related Standards
For comprehensive understanding and effective application, users should also consider the following related ASTM standards and references:
- ASTM E6: Terminology Relating to Methods of Mechanical Testing
- Guidelines for calibration and signal conditioning equipment for strain measurements
- Manufacturer datasheets and recommendations for high-temperature strain gages
- International standards for mechanical testing and measurement uncertainty
ASTM E1319-23 is developed in line with internationally recognized principles of standardization, supporting global harmonization of materials testing methodologies. Employing this guide helps ensure compliance with best practices and regulatory requirements for high-temperature static strain measurement.
Keywords: ASTM E1319-23, high-temperature strain measurement, static strain, strain gage selection, resistance strain gage, capacitive strain gage, thermal compensation, strain gage installation, calibration, mechanical testing standards.
Buy Documents
ASTM E1319-23 - Standard Guide for High-Temperature Static Strain Measurement
REDLINE ASTM E1319-23 - Standard Guide for High-Temperature Static Strain Measurement
Get Certified
Connect with accredited certification bodies for this standard
IMP NDT d.o.o.
Non-destructive testing services. Radiography, ultrasonic, magnetic particle, penetrant, visual inspection.
Inštitut za kovinske materiale in tehnologije
Institute of Metals and Technology. Materials testing, metallurgical analysis, NDT.
Q Techna d.o.o.
NDT and quality assurance specialist. 30+ years experience. NDT personnel certification per ISO 9712, nuclear and thermal power plant inspections, QA/
Sponsored listings
Frequently Asked Questions
ASTM E1319-23 is a guide published by ASTM International. Its full title is "Standard Guide for High-Temperature Static Strain Measurement". This standard covers: SIGNIFICANCE AND USE 4.1 The use of this guide is voluntary and is intended for use as a procedures guide for selection and application of specific types of strain gages for high-temperature installations. No attempt is made to restrict the type of strain gage types or concepts to be chosen by the user. The provisions of this guide may be invoked in specifications and procedures by specifying those that shall be considered mandatory for the purpose of the specific application. When so invoked, the user shall include in the work statement a notation that provisions of this guide shown as recommendation shall be considered mandatory for the purposes of the specification or procedure concerned, and shall include a statement of any exceptions to or modifications of the affected provisions of this guide. SCOPE 1.1 This guide covers the selection and application of strain gages for the measurement of static strain up to and including the temperature range from 425 °C to 650 °C (800 °F to 1200 °F). This guide reflects some state-of-the-art techniques in high-temperature strain measurement. 1.2 This guide assumes that the user is familiar with the use of bonded strain gages and associated signal conditioning circuits and instrumentation as discussed in (1) and (2).2 The strain gage systems described are those that have proven effective in the temperature range of interest and were available at the time of issue of this guide. It is not the intent of this guide to limit the user to one of the strain gage types described nor is it the intent to specify the type of strain gage system to be used for a specific application. However, in using any strain gage system including those described, the proposer shall be able to demonstrate the capability of the proposed strain gage system to meet the selection criteria provided in Section 5 and the needs of the specific application. 1.3 The devices and techniques described in this guide can sometimes be applicable at temperatures above and below the range noted, and for making dynamic strain measurements at high temperatures with proper precautions. The strain gage manufacturer should be consulted for recommendations and details of such applications. 1.4 The references are a part of this guide to the extent specified in the text. 1.5 The values stated in metric (SI) units are to be regarded as the standard. The values given in parentheses are for informational purposes only. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
SIGNIFICANCE AND USE 4.1 The use of this guide is voluntary and is intended for use as a procedures guide for selection and application of specific types of strain gages for high-temperature installations. No attempt is made to restrict the type of strain gage types or concepts to be chosen by the user. The provisions of this guide may be invoked in specifications and procedures by specifying those that shall be considered mandatory for the purpose of the specific application. When so invoked, the user shall include in the work statement a notation that provisions of this guide shown as recommendation shall be considered mandatory for the purposes of the specification or procedure concerned, and shall include a statement of any exceptions to or modifications of the affected provisions of this guide. SCOPE 1.1 This guide covers the selection and application of strain gages for the measurement of static strain up to and including the temperature range from 425 °C to 650 °C (800 °F to 1200 °F). This guide reflects some state-of-the-art techniques in high-temperature strain measurement. 1.2 This guide assumes that the user is familiar with the use of bonded strain gages and associated signal conditioning circuits and instrumentation as discussed in (1) and (2).2 The strain gage systems described are those that have proven effective in the temperature range of interest and were available at the time of issue of this guide. It is not the intent of this guide to limit the user to one of the strain gage types described nor is it the intent to specify the type of strain gage system to be used for a specific application. However, in using any strain gage system including those described, the proposer shall be able to demonstrate the capability of the proposed strain gage system to meet the selection criteria provided in Section 5 and the needs of the specific application. 1.3 The devices and techniques described in this guide can sometimes be applicable at temperatures above and below the range noted, and for making dynamic strain measurements at high temperatures with proper precautions. The strain gage manufacturer should be consulted for recommendations and details of such applications. 1.4 The references are a part of this guide to the extent specified in the text. 1.5 The values stated in metric (SI) units are to be regarded as the standard. The values given in parentheses are for informational purposes only. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM E1319-23 is classified under the following ICS (International Classification for Standards) categories: 19.060 - Mechanical testing. The ICS classification helps identify the subject area and facilitates finding related standards.
ASTM E1319-23 has the following relationships with other standards: It is inter standard links to ASTM E6-09b, ASTM E6-09be1, ASTM E6-09a, ASTM E6-09, ASTM E6-08a, ASTM E6-08, ASTM E6-07b, ASTM E6-07a, ASTM E6-07, ASTM E6-06, ASTM E6-03, ASTM E6-03e1, ASTM E6-02a, ASTM E6-99e1. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ASTM E1319-23 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1319 − 23
Standard Guide for
High-Temperature Static Strain Measurement
This standard is issued under the fixed designation E1319; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope* priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.1 This guide covers the selection and application of strain
1.7 This international standard was developed in accor-
gages for the measurement of static strain up to and including
dance with internationally recognized principles on standard-
the temperature range from 425 °C to 650 °C (800 °F to
ization established in the Decision on Principles for the
1200 °F). This guide reflects some state-of-the-art techniques
Development of International Standards, Guides and Recom-
in high-temperature strain measurement.
mendations issued by the World Trade Organization Technical
1.2 This guide assumes that the user is familiar with the use
Barriers to Trade (TBT) Committee.
of bonded strain gages and associated signal conditioning
2. Referenced Documents
circuits and instrumentation as discussed in (1) and (2). The
strain gage systems described are those that have proven
2.1 ASTM Standards:
effective in the temperature range of interest and were available
E6 Terminology Relating to Methods of Mechanical Testing
at the time of issue of this guide. It is not the intent of this guide
3. Terminology
to limit the user to one of the strain gage types described nor
is it the intent to specify the type of strain gage system to be
3.1 Definitions of Terms Common to Mechanical Testing:
used for a specific application. However, in using any strain
3.1.1 The terms calibration, elastic limit, error, gage factor,
gage system including those described, the proposer shall be
lead wire, modulus of elasticity, Poisson’s ratio, and Young’s
able to demonstrate the capability of the proposed strain gage
modulus are used as defined in Terminology E6. Some impor-
system to meet the selection criteria provided in Section 5 and
tant terms from E6 are reprinted here.
the needs of the specific application.
3.1.2 gage factor—for strain gages, the ratio between the
unit change of strain gage resistance due to strain and the
1.3 The devices and techniques described in this guide can
causing strain.
sometimes be applicable at temperatures above and below the
3.1.2.1 Discussion—The gage factor is dimensionless and is
range noted, and for making dynamic strain measurements at
expressed as follows:
high temperatures with proper precautions. The strain gage
manufacturer should be consulted for recommendations and
R 2 R
details of such applications.
R 1 ΔR
K 5 5 (1)
L 2 L ε R
1.4 The references are a part of this guide to the extent 0 0
L
specified in the text.
where:
1.5 The values stated in metric (SI) units are to be regarded
as the standard. The values given in parentheses are for K = gage factor,
R = strain gage resistance at test strain,
informational purposes only.
R = strain gage resistance at zero or reference strain,
1.6 This standard does not purport to address all of the
L = test structure length under the strain gage at test strain,
safety concerns, if any, associated with its use. It is the
L = test structure length under the strain gage at zero or
responsibility of the user of this standard to establish appro-
reference strain,
ΔR = change in strain gage resistance when strain is changed
from zero (or reference strain) to test strain, and
L2L
= 0
This guide is under the jurisdiction of ASTM Committee E28 on Mechanical
ε
the mechanical strain
Testing and is the direct responsibility of Subcommittee E28.01 on Calibration of L
Mechanical Testing Machines and Apparatus.
Current edition approved May 1, 2023. Published May 2023. Originally
approved in 1989. Last previous edition approved in 2021 as E1319 - 21. DOI: For referenced ASTM standards, visit the ASTM website, www.astm.org, or
10.1520/E1319-23. contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
The boldface numbers in parentheses refer to the list of references at the end of Standards volume information, refer to the standard’s Document Summary page on
this guide. the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1319 − 23
3.1.3 lead wire—for strain gages, an electrical conductor
used to connect a strain gage to its instrumentation.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 capacitive strain gage—a strain gage whose response
to strain is a change in electrical capacitance which is predict-
ably related to that strain.
3.2.1.1 Discussion—Capacitive strain gages are also com-
monly referred to as capacitance strain gages.
3.2.2 compensating strain gage—a strain gage element that
is subject to the same environment as the active strain gage
element, and that is placed in the adjacent leg of a Wheatstone
bridge to provide thermal, pressure, or other compensation in
the strain gage system.
3.2.3 electrical simulation—a method of calibration
whereby a known voltage is generated at the input of an
amplifier, equivalent to the voltage produced by a specific
FIG. 1 Relationship Between Static and Dynamic Strain
amount of strain.
3.2.4 free-filament strain gage—a resistance strain gage
3.2.11 test article—an item to which a strain gage system is
made from a continuous wire or foil filament that is fixed to the
installed for the purpose of measuring strain in that item.
test article along the entire length of the strain gage, and that is
3.2.12 thermal compensation—the process by which the
supplied without a permanent matrix.
thermal output of a strain gage system is counteracted through
3.2.4.1 Discussion—The matrix is an electrically noncon-
the use of one or more supplementary devices, such as a
ductive layer of material used to support a strain gage grid.
thermocouple or compensating strain gage.
3.2.5 integral lead wire—a lead wire or portion of a lead
3.2.12.1 Discussion—The counteraction may be integral to
wire that is furnished by a strain gage manufacturer as part of
the strain gage system or may be accomplished by data
the strain gage assembly.
processing methods, or both.
3.2.6 resistance strain gage—a strain gage whose response
3.2.13 thermal output—the reversible part of the
to strain is a change in electrical resistance that is predictably
temperature-induced indicated strain of a strain gage installed
related to that strain.
on an unrestrained test article when exposed to a change in
temperature.
3.2.7 shunt calibration—a method of calibration whereby a
resistor or capacitor of known value is placed electrically in
4. Significance and Use
parallel with another resistor or capacitor in a circuit, causing
a calculable change in the total resistance or capacitance that is 4.1 The use of this guide is voluntary and is intended for use
predictably related to a specific amount of strain. as a procedures guide for selection and application of specific
types of strain gages for high-temperature installations. No
3.2.8 signal conditioning circuit—a circuit or instrument
attempt is made to restrict the type of strain gage types or
subsystem that applies excitation to a strain gage, detects an
concepts to be chosen by the user. The provisions of this guide
electrical change in the strain gage, and converts this change to
may be invoked in specifications and procedures by specifying
an output that is related to strain in the test article.
those that shall be considered mandatory for the purpose of the
3.2.8.1 Discussion—The signal conditioning circuit may
specific application. When so invoked, the user shall include in
include one or more of the following: bridge completion
the work statement a notation that provisions of this guide
circuit, signal amplification, zero adjustment, excitation
shown as recommendation shall be considered mandatory for
adjustment, calibration, and gain (span) adjustment.
the purposes of the specification or procedure concerned, and
3.2.9 strain gage system—the sum total of all components
shall include a statement of any exceptions to or modifications
used to obtain a strain measurement.
of the affected provisions of this guide.
3.2.9.1 Discussion—The strain gage system may include a
strain gage; a means of attaching the strain gage to the test
5. Strain Gage Selection Criteria
articles; lead wires; splices; lead-wire attachments; signal
5.1 The factors listed in this section shall be considered
conditioning circuit and read-out instrumentation; data-logging
when selecting a strain gage system for use in the temperature
system; calibration and control system; environmental protec-
range specified in 1.1. The risk of compromising certain test
tion; or any combination of these and other elements required
objectives shall be evaluated.
for the tests.
NOTE 1—It is possible that no strain gage has all of the desired
3.2.10 static strain—a strain that is measured relative to a
capabilities to meet all requirements of a particular test. Some test
constant reference value, as opposed to dynamic strain, which
objectives will have to be modified to match the capabilities of the
is the peak-to-peak value of a cyclic phenomenon, without
available strain gage selected. Guidelines for this evaluation are provided
reference to a constant zero or reference value (Fig. 1). in Section 9.
E1319 − 23
5.2 Operating Temperature: be demonstrated; if multiple tests are required on the same test
5.2.1 Isothermal Tests—Stability of the reference value with article, the capability and effect of strain gage replacement
respect to time is essential when tests are to be made at shall also be established.
constant temperature. The stability of the candidate strain gage
5.4 Strain Rate—The time response of the candidate strain
system at the specified temperature shall be such that any shift
gage system shall be adequate to meet test requirements if
that occurs in the reference value is tolerable for the duration
rapid changes of load are anticipated.
of the test.
NOTE 6—Limiting the loading rate of the test is one solution to
5.2.2 Thermal Compensation and Transients—The ad-
accommodate limitations of the selected strain gage system.
equacy of the thermal compensation shall be considered when
the measurement of strain during a thermal transient is re-
5.5 Environment—Some strain gages are limited to specific
quired.
operating environments and therefore, the strain gage system
selected shall be capable of withstanding the environment in
NOTE 2—Thermal output is a function of temperature, thus its value at
which it will operate. Such limitations shall be carefully
a temperature depends not only on temperature, but on the temperature
history followed in reaching that temperature. If significant hysteresis in
considered when selecting the strain gage system to be used.
the thermal response is present, large errors or uncertainties can result.
5.5.1 Factors such as pressure, vibration, radiation, mag-
This is especially true when the calibration procedure used to characterize
netic fields, and humidity shall be considered.
the thermal output does not accurately reflect the temperature sequence to
which the strain gages will be exposed during testing.
5.5.2 The ambient and test environments of the elements of
5.2.2.1 If the response time of the thermal compensation is the strain gage system shall be considered in the selection of
exceeded, the resulting uncertainty shall be considered. lead wires, connectors, instrumentation, and seals (when re-
5.2.2.2 The ability of the strain gage system to withstand the
quired).
transient without a detrimental shift of the reference value shall
5.6 Strain Range:
be verified. This is true whether or not strain is measured
5.6.1 Total Strain Range—The maximum strain ranges of
during the transient.
the candidate strain gage types shall be defined and shall be
5.2.2.3 Any gage factor change as a function of temperature
adequate for the test. Mechanical strain attenuators, when
change shall also be considered.
permissible, may be added to extend the strain range of a given
5.2.3 Pre-test Calibration:
strain gage system, subject to the limitation of 5.6.2.
5.2.3.1 If thermal output calibration on the test article is not
possible, strain gages shall be precalibrated on a similar
5.6.2 Resolution—The ability of the candidate strain gage
material.
system to measure small increments of strain within the total
–6 strain range should be compared with the incremental strain
NOTE 3—Variations of up to 0.3 × 10 /°C are possible within a
measurement requirements of the test. When mechanical strain
material. Often, rolling direction will influence thermal expansion coeffi-
cient.
attenuators are used, the resulting loss of resolution shall be
considered.
5.2.3.2 Precalibrate resistance or capacitive strain gages
using a calibration fixture made from material similar to the
5.7 Strain Gradient—The length of the candidate strain
test article. The calibration fixture shall be made to precisely fit
gage establishes the length over which the unit strain is
the strain gage, especially if curvature is involved. Mating parts
averaged. This factor shall be considered.
shall be lapped together to provide uniform clamping pressure
around the periphery of the strain gage weld area. 5.8 Uncertainty Factor—Uncertainty information that is
5.2.3.3 The calibration test should be repeated to ensure available from the manufacturer shall be considered, in con-
precise duplication of the calibration. Zero return should also junction with conditions that are unique to the test, in order to
repeat exactly.
estimate the total uncertainty.
NOTE 4—If calibration data does not repeat, either the calibration setup
5.9 Space Requirements—Working space for installation of
or the strain gages are faulty.
the strain gage system shall be considered. Space adjacent to
5.2.4 Post-Test Calibration—If a more precise thermal out-
the installed strain gage should be provided for installation of
put calibration is needed, a post-test calibration should be
room-temperature strain gages required for making in-place
conducted.
calibrations.
5.2.4.1 Remove the test strain gage (cut it out of the test
NOTE 7—If space on or adjacent to the test article is limited, the space
article) and run a precision test on the test strain gage still
requirements for the complete strain gage system can be a critical
consideration in determining the suitability of a particular strain gage
attached to the test article material.
system.
NOTE 5—The test article is relieved of all induced stresses (thermal,
5.10 Effects of the Strain Gage on the Test Article—If a
mechanical, residual) and is free to expand freely with temperature.
weldable strain gage is to be used on thin sections, an
5.2.4.2 The strain gage integral lead wire should be exposed
evaluation of the reinforcing effect should be made.
to thermal gradients similar to those that occurred during the
NOTE 8—In most cases the reinforcing effect of the strain gage on the
test program.
test article is negligible, particularly in the case of capacitive strain gages
5.3 Duration of Test—The ability of all parts of the strain
where the spring rate is extremely low. Technical data concerning this
gage system to function for the specified duration of test should effect can be obtained from a strain gage manufacturer.
E1319 − 23
6. Characteristics of Available Strain Gages alloys are available: (1) self temperature compensated nickel
chrome alloy sensors usable to 340 °C in quarter bridge (single
6.1 The two basic types of strain gages used for high
element) configuration, (2) platinum tungsten and palladium
temperature static strain measurements are resistance strain
chrome alloy sensors (dual element) compensated with plati-
gages and capacitive strain gages.
num elements in half-bridge configuration, and (3) iron chrome
6.1.1 Resistance strain gages are usually small, low-profile
aluminum alloys having low temperature coefficient are avail-
units superbly suited for dynamic strain measurements and
able in half- or full-bridge configuration for applications where
relatively short-term static measurements. Because high tem-
active-dummy combinations (slow temperature changes) are
peratures cause metallurgical instability, oxidation, relaxation,
usable.
and phase change of the strain-sensing materials, all of which
6.2.2 Except for long-term stability, the bonded strain gage
affect resistance change, resistance gages are generally not
has excellent performance with minimal hysteresis, small zero
used for long-term measurements.
shift, long fatigue life, and accurate gage factor among its
6.1.2 Capacitive strain gages are devices that measure
salient features. An integral weldable terminal and integral
changes in geometry and are unaffected by temperature or
high temperature cable are usually supplied with these units,
temperature changes, oxidation, relaxation, creep, grain
especially when the strain gages are supplied precalibrated for
growth, or phase change. They are best suited for measuring
apparent strain.
creep strains, or for very long-term tests on applications where
6.2.3 The thermal output of the dual-element strain gages
a relatively large strain gage can be used, and when the strain
can be adjusted to produce a zero output at any two selected
gage will not be subjected to high vibration, gravity, or
temperatures. The thermal output of the platinum tungsten
acceleration forces, shock loading, or an electrically conduc-
strain gage is usually well within 6200 μm/m between 20 °C
tive atmosphere.
and 500 °C. The shape of the thermal output curve is influ-
6.1.3 When selecting a specific strain gage for a given
enced by the thermal expansion characteristics of the test
application, the strain gage system shall be qualified for the
material. Fig. 3 shows the bridge completion circuit for the
specific conditions under which it will be required to operate
dual-element half-bridge strain gage. There are two methods of
and for the characteristics it shall exhibit under service
thermal compensation: (1) NASA method (3) and (2) the wire
conditions. This section describes some of the capability of
method (4).
currently available strain gages, suitable for use in the specified
6.2.3.1 With the NASA method, the strain gage is manufac-
temperature range, to meet the selection criteria of Section 5.
tured to fit a specific type of material with the platinum
NOTE 9—Wire and foil free-filament strain gages are usable to
thermometer element resistance value selected to provide an
approximately 400 °C (750 °F) under static conditions, and to approxi-
almost perfectly balanced bridge. This permits a three-wire
mately 1250 °C (2280 °F) for certain dynamic applications. However, the
cable to be used without sacrificing inherent lead wire com-
bonding methods used (ceramic cement, flame spray) are cumbersome and
pensation. The five-wire system employs a thermometer ele-
difficult to employ on large structures, particularly under field conditions.
Ceramic cements require heat-curing and are generally unsuitable for
ment sufficiently high in resistance to compensate on virtually
large structures such as nuclear or fossil-fuel power-generating equipment.
any material. This makes for universal thermal compensation
Flame spray is also difficult to use in the field. Free-filament strain gages,
on any material. The drawback of this universal system is that
although useful for strain measurement on small articles under laboratory
five wires are required. A shunt calibration resistor R placed
conditions, are, therefore, not included in this guide. This does not g
across the thermometer element only shunts the output of the
preclude the use of these strain gages for specific tests based on the
selection criteria of Section 5.
compensator. If the shunt calibration resistor were placed
across the thermometer and lead wire, inherent lead compen-
6.1.4 The strain gages described in this section have been
sation would be sacrificed.
used at high temperature for sufficient time and with sufficient
6.2.4 The user may precalibrate the strain gage and cable
success to warrant consideration in this guide. Each type has
system and determine the bridge completion resistor values
unique features, advantages, and limitations that shall be
using a set of equations provided with the strain gage, or the
carefully evaluated relative to the selection factors of
strain gage may be precalibrated at the factory (5) and supplied
Section 5.
precalibrated with the bridge completion resistors included in a
6.2 Bonded Weldable Resistance Strain Gage:
circuit attached to the cool end of the cable. The user needs
6.2.1 This strain gage, shown in Fig. 2, consists of a
free-filament strain gage ceramic bonded to a shim. While it is
not usually sealed or intended for underwater use, some
hermetically sealed strain gages are bonded to the shim with
ceramic cements or flame sprayed ceramics. The following
FIG. 2 Bonded Weldable Resistance Strain Gage FIG. 3 Five-Wire Circuit
E1319 − 23
only to connect the strain gage as a full bridge transducer and
insert the calibration curve into the data acquisition system.
6.3 Hermetic Weldable Resistance Strain Gage—This strain
gage, which is shown several times the actual size in Fig. 4, is
hermetically sealed and furnished with integral lead wires, and
may be used in a variety of severe environments at high
temperature. The strain tube is welded to a thin mounting
FIG. 5 Bridge Completion Circuit and Power Supply
flange, which is welded to the surface of the test article, thus
providing transfer of strain from the test article to the strain
gage. Although Fig. 4 shows stainless steel strain tube, mount-
6.3.2.2 To qualify the strain gage for thermal shock, labo-
ing flange, and cladding of the integral lead wire, other
ratory tests should be made to determine the stability charac-
materials are available to meet the requirements of specific
teristics and the limits of thermal compensation.
applications; consult the manufacturer for available materials.
6.3.3 Electrical Requirements—Bridge completion, as
Within limits, the thermal output of the strain gage due to
shown in Fig. 5, shall be made. While there are several
temperature can be adjusted to produce a zero output at any
standard strain gage systems with bridge completion capability
two selected temperatures by inserting a thermal compensation
available, for static strain measurement in the temperature
resistor, R in Fig. 5 in series with either the active or
tc
range of this guide, an individual signal conditioning circuit
compensating strain gage element; the proper resistor is
should have the following features for each strain gage.
furnished by the manufacturer. Because of the added resistance
6.3.3.1 Excitation Power Supply—A power supply for pro-
in series with one of the strain gage elements, the bridge-
viding constant DC voltage, continuously variable from 1 V to
completion resistors shall also be adjusted for balance by
15 V across a 120 Ω external load, shall be used. If more than
adding a balancing resistor (R in Fig. 5) in the opposite half
bal
one bridge circuit is excited by the same power supply, the
of the bridge. This resistor is also furnished by the strain gage
electrical configuration shall provide electrical isolation of
manufacturer. The value of R is based on the use of 120 Ω
bal
each circuit to protect it in the event of a direct short of the
bridge-completion resistors to produce a balanced bridge when
excitation of any of the adjacent circuits.
the strain gage is connected.
6.3.1 Operating Temperature and Thermal Stability:
NOTE 12—Constant current excitation cannot be used with some of the
6.3.1.1 The platinum tungsten element is essentially stable
thermal compensation techniques generally used with the hermetic weld-
able resistance strain gage.
for short-term testing to 500 °C (days) with shorter excursions
up to 580 °C (hours) without damage to the strain gage. Longer
6.3.3.2 Balance Control—Means shall be provided for bal-
tests (weeks) can be run to up to 425 °C. Beyond these limits
ancing the bridge with a T balance resistor circuit across the
a capacitive strain gage should be used.
completion half of the bridge. This means may be omitted if the
6.3.2 Thermal Compensation and Transients—Thermal out-
data acquisition equipment automatically compensates for
put characteristics shall be considered for operation at varying
initial bridge unbalance.
temperatures.
6.3.3.3 Shunt Calibration—Shunt calibration capability
should be provided on the completion half of the bridge. The
NOTE 10—Information about the thermal output characteristics is
active or compensating legs of the bridge should not be
furnished by the manufacturer for use of the strain gage on the material
specified by the user.
shunted, because of changes in the resistances of the lead wires
with temperature. Multiple shunt calibration should be done.
6.3.2.1 For precise evaluation, the strain gage system shall
6.3.3.4 Signal Conditioning Circuit—The signal condition-
be calibrated, with thermal output determined at the tempera-
ing circuit shall be capable of handling a half-bridge circuit
tures of interest. Temperatures should be measured by a
with precision completion resistors, configured to permit the
thermocouple(s) mounted immediately adjacent to the strain
addition of series balance resistors to either leg. The signal
gage.
conditioning circuit should be able to accommodate a half-
NOTE 11—Thermal output and hysteresis of a test are usually repeatable
bridge, five-wire configuration, with two additional leads, for
under identical test conditions; however, even the slightest change in test
remote sensing of the excitation voltage.
conditions can result in a change of thermal output, hysteresis, or both.
6.3.3.5 Means shall be provided for continuous monitoring
of bridge excitations and bridge output. An amplifier may be
used or omitted, depending on the input capability of the strain
gage system used. Amplifier requirements are not covered in
this guide; however, a good quality, stable amplifier with true
differential input, and input impedance of not less than 10 MΩ
and shunted by 750 pF when DC-coupled should be used.
6.4 Differential Capacitance Strain Gage—Fig. 6 identifies
major elements of the strain gage and shows principal dimen-
sions. Fig. 7 shows an isometric view. The compensating rod
(1) is usually made of the same material as the test article
FIG. 4 Hermetic Weldable Resistance Strain Gage (specified by user). The cylindrical excitation plates (2) are
E1319 − 23
FIG. 6 Differential Capacitance Strain Gage
NOTE 1—Overall strain gage dimensions are 3.175 cm (1.25 in.) by 1.524 cm (0.60 in.)
FIG. 7 Isometric View, Differential Capacitance Strain Gage
mounted coaxially on, but are electrically insulated from, the instrumented with thermocouples to obtain data for computing
compensating rod. The sensing ring (3) is mounted coaxially the corrections required if there is a temperature difference
with the excitation plates but is separated from them by an air betw
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1319 − 21 E1319 − 23
Standard Guide for
High-Temperature Static Strain Measurement
This standard is issued under the fixed designation E1319; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This guide covers the selection and application of strain gages for the measurement of static strain up to and including the
temperature range from 425 °C to 650 °C (800 °F to 1200 °F). This guide reflects some state-of-the-art techniques in
high-temperature strain measurement.
1.2 This guide assumes that the user is familiar with the use of bonded strain gages and associated signal conditioning circuits and
instrumentation as discussed in (1) and (2). The strain gage systems described are those that have proven effective in the
temperature range of interest and were available at the time of issue of this guide. It is not the intent of this guide to limit the user
to one of the strain gage types described nor is it the intent to specify the type of strain gage system to be used for a specific
application. However, in using any strain gage system including those described, the proposer shall be able to demonstrate the
capability of the proposed strain gage system to meet the selection criteria provided in Section 5 and the needs of the specific
application.
1.3 The devices and techniques described in this guide can sometimes be applicable at temperatures above and below the range
noted, and for making dynamic strain measurements at high temperatures with proper precautions. The strain gage manufacturer
should be consulted for recommendations and details of such applications.
1.4 The references are a part of this guide to the extent specified in the text.
1.5 The values stated in metric (SI) units are to be regarded as the standard. The values given in parentheses are for informational
purposes only.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of
regulatory limitations prior to use.
1.7 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
This guide is under the jurisdiction of ASTM Committee E28 on Mechanical Testing and is the direct responsibility of Subcommittee E28.01 on Calibration of Mechanical
Testing Machines and Apparatus.
Current edition approved Feb. 1, 2021May 1, 2023. Published April 2021May 2023. Originally approved in 1989. Last previous edition approved in 20142021 as
E1319 - 98 (2014).E1319 - 21. DOI: 10.1520/E1319-21.10.1520/E1319-23.
The boldface numbers in parentheses refer to the list of references at the end of this guide.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1319 − 23
2. Referenced Documents
2.1 ASTM Standards:
E6 Terminology Relating to Methods of Mechanical Testing
3. Terminology
3.1 Definitions of Terms Common to Mechanical Testing:
3.1.1 The terms calibration, elastic limit, error, gage factor, lead wire, modulus of elasticity, Poisson’s ratio, and Young’s modulus
are used as defined in Terminology E6. Some important terms from E6 are reprinted here.
3.1.2 gage factor—for strain gages, the ratio between the unit change of strain gage resistance due to strain and the causing strain.
3.1.2.1 Discussion—
The gage factor is dimensionless and is expressed as follows:
R 2 R
R 1ΔR
K 5 5 (1)
L 2 L ε R
0 0
L
where:
K = gage factor,
R = strain gage resistance at test strain,
R = strain gage resistance at zero or reference strain,
L = test structure length under the strain gage at test strain,
L = test structure length under the strain gage at zero or reference strain,
ΔR = change in strain gage resistance when strain is changed from zero (or reference strain) to test strain, and
L2L
= 0
ε
the mechanical strain
L
3.1.3 lead wire—for strain gages, an electrical conductor used to connect a strain gage to its instrumentation.
3.2 Definitions of Terms Specific to This Standard:
3.2.1 capacitive strain gage—a strain gage whose response to strain is a change in electrical capacitance which 1sis predictably
related to that strain.
3.2.1.1 Discussion—
Capacitive strain gages are also commonly referred to as capacitance strain gages.
3.2.2 compensating strain gage—a strain gage element that is subject to the same environment as the active strain gage element,
and that is placed in the adjacent leg of a Wheatstone bridge to provide thermal, pressure, or other compensation in the strain gage
system.
3.2.3 electrical simulation—a method of calibration whereby a known voltage is generated at the input of an amplifier, equivalent
to the voltage produced by a specific amount of strain.
3.2.4 free-filament strain gage—a resistance strain gage made from a continuous wire or foil filament that is fixed to the test article
along the entire length of the strain gage, and that is supplied without a permanent matrix.
3.2.4.1 Discussion—
The matrix is an electrically nonconductive layer of material used to support a strain gage grid.
3.2.5 integral lead wire—a lead wire or portion of a lead wire that is furnished by a strain gage manufacturer as part of the strain
gage assembly.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
E1319 − 23
3.2.6 resistance strain gage—a strain gage whose response to strain is a change in electrical resistance that is predictably related
to that strain.
3.2.7 shunt calibration—a method of calibration whereby a resistor or capacitor of known value is placed electrically in parallel
with another resistor or capacitor in a circuit, causing a calculable change in the total resistance or capacitance that is predictably
related to a specific amount of strain.
3.2.8 signal conditioning circuit—a circuit or instrument subsystem that applies excitation to a strain gage, detects an electrical
change in the strain gage, and converts this change to an output that is related to strain in the test article.
3.2.8.1 Discussion—
The signal conditioning circuit may include one or more of the following: bridge completion circuit, signal amplification, zero
adjustment, excitation adjustment, calibration, and gain (span) adjustment.
3.2.9 strain gage system—the sum total of all components used to obtain a strain measurement.
3.2.9.1 Discussion—
The strain gage system may include a strain gage; a means of attaching the strain gage to the test articles; lead wires; splices;
lead-wire attachments; signal conditioning circuit and read-out instrumentation; data-logging system; calibration and control
system; environmental protection; or any combination of these and other elements required for the tests.
3.2.10 static strain—a strain that is measured relative to a constant reference value, as opposed to dynamic strain, which is the
peak-to-peak value of a cyclic phenomenon, without reference to a constant zero or reference value (Fig. 1).
3.2.11 test article—an item to which a strain gage system is installed for the purpose of measuring strain in that item.
3.2.12 thermal compensation—the process by which the thermal output of a strain gage system is counteracted through the use
of one or more supplementary devices, such as a thermocouple or compensating strain gage.
3.2.12.1 Discussion—
The counteraction may be integral to the strain gage system or may be accomplished by data processing methods, or both.
3.2.13 thermal output—the reversible part of the temperature-induced indicated strain of a strain gage installed on an unrestrained
test article when exposed to a change in temperature.
4. Significance and Use
4.1 The use of this guide is voluntary and is intended for use as a procedures guide for selection and application of specific types
of strain gages for high-temperature installations. No attempt is made to restrict the type of strain gage types or concepts to be
chosen by the user. The provisions of this guide may be invoked in specifications and procedures by specifying those that shall
FIG. 1 Relationship Between Static and Dynamic Strain
E1319 − 23
be considered mandatory for the purpose of the specific application. When so invoked, the user shall include in the work statement
a notation that provisions of this guide shown as recommendation shall be considered mandatory for the purposes of the
specification or procedure concerned, and shall include a statement of any exceptions to or modifications of the affected provisions
of this guide.
5. Strain Gage Selection Criteria
5.1 The factors listed in this section shall be considered when selecting a strain gage system for use in the temperature range
specified in 1.1. The risk of compromising certain test objectives shall be evaluated.
NOTE 1—It is possible that no strain gage has all of the desired capabilities to meet all requirements of a particular test. Some test objectives will have
to be modified to match the capabilities of the available strain gage selected. Guidelines for this evaluation are provided in Section 9.
E1319 − 23
5.2 Operating Temperature:
5.2.1 Isothermal Tests—Stability of the reference value with respect to time is essential when tests are to be made at constant
temperature. The stability of the candidate strain gage system at the specified temperature shall be such that any shift that occurs
in the reference value is tolerable for the duration of the test.
5.2.2 Thermal Compensation and Transients—The adequacy of the thermal compensation shall be considered when the
measurement of strain during a thermal transient is required.
NOTE 2—Thermal output is a function of temperature, thus its value at a temperature depends not only on temperature, but on the temperature history
followed in reaching that temperature. If significant hysteresis in the thermal response is present, large errors or uncertainties can result. This is especially
true when the calibration procedure used to characterize the thermal output does not accurately reflect the temperature sequence to which the strain gages
will be exposed during testing.
5.2.2.1 If the response time of the thermal compensation is exceeded, the resulting uncertainty shall be considered.
5.2.2.2 The ability of the strain gage system to withstand the transient without a detrimental shift of the reference value shall be
verified. This is true whether or not strain is measured during the transient.
5.2.2.3 Any gage factor change as a function of temperature change shall also be considered.
5.2.3 Pre-test Calibration:
5.2.3.1 If thermal output calibration on the test article is not possible, strain gages shall be precalibrated on a similar material.
–6
NOTE 3—Variations of up to 0.3 × 10 /°C are possible within a material. Often, rolling direction will influence thermal expansion coefficient.
5.2.3.2 Precalibrate resistance or capacitive strain gages using a calibration fixture made from material similar to the test article.
The calibration fixture shall be made to precisely fit the strain gage, especially if curvature is involved. Mating parts shall be lapped
together to provide uniform clamping pressure around the periphery of the strain gage weld area.
5.2.3.3 The calibration test should be repeated to ensure precise duplication of the calibration. Zero return should also repeat
exactly.
NOTE 4—If calibration data does not repeat, either the calibration setup or the strain gages are faulty.
5.2.4 Post-Test Calibration—If a more precise thermal output calibration is needed, a post-test calibration should be conducted.
5.2.4.1 Remove the test strain gage (cut it out of the test article) and run a precision test on the test strain gage still attached to
the test article material.
NOTE 5—The test article is relieved of all induced stresses (thermal, mechanical, residual) and is free to expand freely with temperature.
5.2.4.2 The strain gage integral lead wire should be exposed to thermal gradients similar to those that occurred during the test
program.
5.3 Duration of Test—The ability of all parts of the strain gage system to function for the specified duration of test should be
demonstrated; if multiple tests are required on the same test article, the capability and effect of strain gage replacement shall also
be established.
5.4 Strain Rate—The time response of the candidate strain gage system shall be adequate to meet test requirements if rapid
changes of load are anticipated.
NOTE 6—Limiting the loading rate of the test is one solution to accommodate limitations of the selected strain gage system.
E1319 − 23
5.5 Environment—Some strain gages are limited to specific operating environments and therefore, the strain gage system selected
shall be capable of withstanding the environment in which it will operate. Such limitations shall be carefully considered when
selecting the strain gage system to be used.
5.5.1 Factors such as pressure, vibration, radiation, magnetic fields, and humidity shall be considered.
5.5.2 The ambient and test environments of the elements of the strain gage system shall be considered in the selection of lead
wires, connectors, instrumentation, and seals (when required).
5.6 Strain Range:
5.6.1 Total Strain Range—The maximum strain ranges of the candidate strain gage types shall be defined and shall be adequate
for the test. Mechanical strain attenuators, when permissible, may be added to extend the strain range of a given strain gage system,
subject to the limitation of 5.6.2.
5.6.2 Resolution—The ability of the candidate strain gage system to measure small increments of strain within the total strain
range should be compared with the incremental strain measurement requirements of the test. When mechanical strain attenuators
are used, the resulting loss of resolution shall be considered.
5.7 Strain Gradient—The length of the candidate strain gage establishes the length over which the unit strain is averaged. This
factor shall be considered.
5.8 Uncertainty Factor—Uncertainty information that is available from the manufacturer shall be considered, in conjunction with
conditions that are unique to the test, in order to estimate the total uncertainty.
5.9 Space Requirements—Working space for installation of the strain gage system shall be considered. Space adjacent to the
installed strain gage should be provided for installation of room-temperature strain gages required for making in-place calibrations.
NOTE 7—If space on or adjacent to the test article is limited, the space requirements for the complete strain gage system can be a critical consideration
in determining the suitability of a particular strain gage system.
5.10 Effects of the Strain Gage on the Test Article—If a weldable strain gage is to be used on thin sections, an evaluation of the
reinforcing effect should be made.
NOTE 8—In most cases the reinforcing effect of the strain gage on the test article is negligible, particularly in the case of capacitive strain gages where
the spring rate is extremely low. Technical data concerning this effect can be obtained from a strain gage manufacturer.
6. Characteristics of Available Strain Gages
6.1 The two basic types of strain gages used for high temperature static strain measurements are resistance strain gages and
capacitive strain gages.
6.1.1 Resistance strain gages are usually small, low-profile units superbly suited for dynamic strain measurements and relatively
short-term static measurements. Because high temperatures cause metallurgical instability, oxidation, relaxation, and phase change
of the strain sensing strain-sensing materials, all of which affect resistance change, resistance gages are generally not used for
long-term measurements.
6.1.2 Capacitive strain gages are devices that measure changes in geometry and are unaffected by temperature or temperature
changes, oxidation, relaxation, creep, grain growth, or phase change. They are best suited for measuring creep strains, or for very
long-term tests on applications where a relatively large strain gage can be used, and when the strain gage will not be subjected
to high vibration, gravity, or acceleration forces, shock loading, or an electrically conductive atmosphere.
6.1.3 When selecting a specific strain gage for a given application, the strain gage system shall be qualified for the specific
conditions under which it will be required to operate and for the characteristics it shall exhibit under service conditions. This
section describes some of the capability of currently available strain gages, suitable for use in the specified temperature range, to
meet the selection criteria of Section 5.
E1319 − 23
NOTE 9—Wire and foil free-filament strain gages are usable to approximately 400 °C (750 °F) under static conditions, and to approximately 1250 °C
(2280 °F) for certain dynamic applications. However, the bonding methods used (ceramic cement, flame spray) are cumbersome and difficult to employ
on large structures, particularly under field conditions. Ceramic cements require heat-curing and are generally unsuitable for large structures such as
nuclear or fossil-fuel power-generating equipment. Flame spray is also difficult to use in the field. Free-filament strain gages, although useful for strain
measurement on small articles under laboratory conditions, are, therefore, not included in this guide. This does not preclude the use of these strain gages
for specific tests based on the selection criteria of Section 5.
6.1.4 The strain gages described in this section have been used at high temperature for sufficient time and with sufficient success
to warrant consideration in this guide. Each type has unique features, advantages, and limitations that shall be carefully evaluated
relative to the selection factors of Section 5.
6.2 Bonded Weldable Resistance Strain Gage:
6.2.1 This strain gage, shown in Fig. 2, consists of a free-filament strain gage ceramic bonded to a shim. While it is not usually
sealed or intended for underwater use, some hermetically sealed strain gages are bonded to the shim with ceramic cements or flame
sprayed ceramics. The following alloys are available: (1) self temperature compensated nickel chrome alloy sensors usable to
340 °C in quarter bridge (single element) configuration, (2) platinum tungsten and palladium chrome alloy sensors (dual element)
compensated with platinum elements in half-bridge configuration, and (3) iron chrome aluminum alloys having low temperature
coefficient are available in half- or full-bridge configuration for applications where active-dummy combinations (slow temperature
changes) are usable.
6.2.2 Except for long-term stability, the bonded strain gage has excellent performance with minimal hysteresis, small zero shift,
long fatigue life, and accurate gage factor among its salient features. An integral weldable terminal and integral high temperature
cable are usually supplied with these units, especially when the strain gages are supplied precalibrated for apparent strain.
6.2.3 The thermal output of the dual element dual-element strain gages can be adjusted to produce a zero output at any two
selected temperatures. The thermal output of the platinum tungsten strain gage is usually well within 6200 μm/m between 20 °C
and 500 °C. The shape of the thermal output curve is influenced by the thermal expansion characteristics of the test material. Fig.
3 shows the bridge completion circuit for the dual element dual-element half-bridge strain gage. There are two methods of thermal
compensation: (1) NASA method (3) and (2) the wire method (4).
6.2.3.1 With the NASA method, the strain gage is manufactured to fit a specific type of material with the platinum thermometer
element resistance value selected to provide an almost perfectly balanced bridge. This permits a three-wire cable to be used without
sacrificing inherent lead wire compensation. The five-wire system employs a thermometer element sufficiently high in resistance
to compensate on virtually any material. This makes for universal thermal compensation on any material. The drawback of this
universal system is that five wires are required. A shunt calibration resistor R placed across the thermometer element only shunts
g
the output of the compensator. If the shunt calibration resistor were placed across the thermometer and lead wire, inherent lead
compensation would be sacrificed.
6.2.4 The user may precalibrate the strain gage and cable system and determine the bridge completion resistor values using a set
of equations provided with the strain gage, or the strain gage may be precalibrated at the factory (5) and supplied precalibrated
with the bridge completion resistors included in a circuit attached to the cool end of the cable. The user needs only to connect the
strain gage as a full bridge transducer and insert the calibration curve into the data acquisition system.
6.3 Hermetic Weldable Resistance Strain Gage—This strain gage, which is shown several times the actual size in Fig. 4, is
hermetically sealed and furnished with integral lead wires, and may be used in a variety of severe environments at high
temperature. The strain tube is welded to a thin mounting flange, which is welded to the surface of the test article, thus providing
transfer of strain from the test article to the strain gage. Although Fig. 4 shows stainless steel strain tube, mounting flange, and
cladding of the integral lead wire, other materials are available to meet the requirements of specific applications; consult the
FIG. 2 Bonded Weldable Resistance Strain Gage
E1319 − 23
FIG. 3 Five-Wire Circuit
FIG. 4 Hermetic Weldable Resistance Strain Gage
manufacturer for available materials. Within limits, the thermal output of the strain gage due to temperature can be adjusted to
produce a zero output at any two selected temperatures by inserting a thermal compensation resistor, R in Fig. 5 in series with
tc
either the active or compensating strain gage element; the proper resistor is furnished by the manufacturer. Because of the added
resistance in series with one of the strain gage elements, the bridge-completion resistors shall also be adjusted for balance by
adding a balancing resistor (R in Fig. 5) in the opposite half of the bridge. This resistor is also furnished by the strain gage
bal
manufacturer. The value of R is based on the use of 120 Ω bridge-completion resistors to produce a balanced bridge when the
bal
strain gage is connected.
6.3.1 Operating Temperature and Thermal Stability:
6.3.1.1 The platinum tungsten element is essentially stable for short-term testing to 500 °C (days) with shorter excursions up to
580 °C (hours) without damage to the strain gage. Longer tests (weeks) can be run to up to 425 °C. Beyond these limits a capacitive
strain gage should be used.
6.3.2 Thermal Compensation and Transients—Thermal output characteristics shall be considered for operation at varying
temperatures.
NOTE 10—Information about the thermal output characteristics is furnished by the manufacturer for use of the strain gage on the material specified by
the user.
6.3.2.1 For precise evaluation, the strain gage system shall be calibrated, with thermal output determined at the temperatures of
interest. Temperatures should be measured by a thermocouple(s) mounted immediately adjacent to the strain gage.
NOTE 11—Thermal output and hysteresis of a test are usually repeatable under identical test conditions; however, even the slightest change in test
conditions can result in a change of thermal output, hysteresis, or both.
FIG. 5 Bridge Completion Circuit and Power Supply
E1319 − 23
6.3.2.2 To qualify the strain gage for thermal shock, laboratory tests should be made to determine the stability characteristics and
the limits of thermal compensation.
6.3.3 Electrical Requirements—Bridge completion, as shown in Fig. 5, shall be made. While there are several standard strain gage
systems with bridge completion capability available, for static strain measurement in the temperature range of this guide, an
individual signal conditioning circuit should have the following features for each strain gage.
6.3.3.1 Excitation Power Supply—A power supply for providing constant DC voltage, continuously variable from 1 V to 15 V
across a 120 Ω external load, shall be used. If more than one bridge circuit is excited by the same power supply, the electrical
configuration shall provide electrical isolation of each circuit to protect it in the event of a direct short of the excitation of any of
the adjacent circuits.
NOTE 12—Constant current excitation cannot be used with some of the thermal compensation techniques generally used with the hermetic weldable
resistance strain gage.
6.3.3.2 Balance Control—Means shall be provided for balancing the bridge with a T balance resistor circuit across the completion
half of the bridge. This means may be omitted if the data acquisition equipment automatically compensates for initial bridge
unbalance.
6.3.3.3 Shunt Calibration—Shunt calibration capability should be provided on the completion half of the bridge. The active or
compensating legs of the bridge should not be shunted, because of changes in the resistances of the lead wires with temperature.
Multiple shunt calibration should be done.
6.3.3.4 Signal Conditioning Circuit—The signal conditioning circuit shall be capable of handling a half-bridge circuit with
precision completion resistors, configured to permit the addition of series balance resistors to either leg. The signal conditioning
circuit should be able to accommodate a half-bridge, five-wire configuration, with two additional leads, for remote sensing of the
excitation voltage.
6.3.3.5 Means shall be provided for continuous monitoring of bridge excitations and bridge output. An amplifier may be used or
omitted, depending on the input capability of the strain gage system used. Amplifier requirements are not covered in this guide;
however, a good quality, stable amplifier with true differential input, and input impedance of not less than 10 MΩ and shunted by
750 pF when DC-coupled should be used.
6.4 Differential Capacitance Strain Gage—Fig. 86 identifies major elements of the strain gage and shows principal dimensions.
Fig. 97 shows an isometric view. The compensating rod (1) is usually made of the same material as the test article (specified by
user). The cylindrical excitation plates (2) are mounted coaxially on, but are electrically insulated from, the compensating rod. The
sensing ring (3) is mounted coaxially with the excitation plates but is separated from them by an air gap. The attachment ribbons
(4) (see the isometric view, Fig. 97) provide means for welding the strain gage to the test article. The alignment flexures (5) (see
the isometric view in Fig. 97), maintain the coaxial alignment of the sensing ring relative to the excitation plates and compensating
rod. Leads from
...