SIST EN 16181:2018
(Main)Soil, treated biowaste and sludge - Determination of polycyclic aromatic hydrocarbons (PAH) by gas chromatography (GC) and high performance liquid chromatography (HPLC)
Soil, treated biowaste and sludge - Determination of polycyclic aromatic hydrocarbons (PAH) by gas chromatography (GC) and high performance liquid chromatography (HPLC)
This European Standard specifies the quantitative determination of 16 polycyclic aromatic hydrocarbons (PAH) (see Table 2) in sludge, soil and treated biowaste using GC-MS and HPLC-UV-DAD/FLD covering a wide range of PAH contamination levels (see also Annex B).
When using fluorescence detection, acenaphthylene cannot be measured.
(...)
The limit of detection depends on the determinants, the equipment used, the quality of chemicals used for the extraction of the sample and the clean-up of the extract.
Typically, a lower limit of application of 0,01 mg/kg (expressed as dry matter) can be ensured for each individual PAH. This depends on instrument and sample.
Sludge, soil and treated biowaste can differ in properties and also in the expected contamination levels of PAHs and presence of interfering substances. These differences make it impossible to describe one general procedure. This European Standard contains decision tables based on the properties of the sample and the extraction and clean-up procedure to be used. Two general lines are followed, an agitation procedure (shaking) or use of soxhlet/pressurized liquid extraction.
NOTE Other PAH compounds can also be analysed with this method, provided suitability has been proven.
Boden, behandelter Bioabfall und Schlamm - Bestimmung von polycyclischen aromatischen Kohlenwasserstoffen (PAK) mittels Gaschromatographie (GC) und Hochleistungs-Flüssigkeitschromatographie (HPLC)
Diese Europäische Norm legt die quantitative Bestimmung von 16 polycyclischen aromatischen Kohlenwasserstoffen (PAK) (siehe Tabelle 2) in Schlamm und behandeltem Bioabfall unter Anwendung der GC MS und HPLC UV DAD/FLD fest, wobei ein breiter Bereich von PAK Kontaminationsniveaus (siehe auch Anhang B) abgedeckt wird.
Bei Anwendung der Fluoreszenzdetektion kann Acenaphthylen nicht gemessen werden.
Sols, biodéchets traités et boues - Dosage des hydrocarbures aromatiques polycycliques (HAP) par chromatographie en phase gazeuse et chromatographie liquide à haute performance
La présente Norme européenne spécifie le dosage de 16 hydrocarbures aromatiques polycycliques (HAP) (voir Tableau 2) dans les boues, les sols et les biodéchets traités, faisant appel à des méthodes CG-SM et HPLC-UV-DAD/FLD couvrant une large gamme de niveaux de contamination par les HAP (voir également l’Annexe B).
L’utilisation d’une méthode de détection fluorimétrique ne permet pas le mesurage de l’acénaphtylène.
(...)
La limite de détection dépend des composés à analyser, de l’équipement utilisé, de la qualité des réactifs chimiques utilisés pour l’extraction de l’échantillon et la purification de l’extrait.
En général, une limite inférieure d’application de 0,01 mg/kg (exprimée en matière sèche) peut être atteinte pour chacun des HAP. Cela dépend de l’instrument et de l’échantillon.
Les boues, les sols et les biodéchets traités peuvent présenter des propriétés, des niveaux de contamination par les HAP attendus et des substances interférentes différents. En raison de ces différences, il est impossible de décrire un mode opératoire général. La présente Norme européenne contient des tables de décision basées sur les propriétés de l'échantillon ainsi que sur le mode opératoire d'extraction et de purification à utiliser. Deux routes analytiques générales sont suivies : un mode opératoire par agitation ou l’utilisation d’une méthode d’extraction au Soxhlet/par liquide pressurisé.
NOTE D’autres composés HAP peuvent être également analysés avec cette méthode, à condition que l’aptitude à l’emploi soit prouvée.
Tla, obdelani biološki odpadki in blato - Določevanje policikličnih aromatskih ogljikovodikov (PAH) s plinsko kromatografijo (GC) in s tekočinsko kromatografijo visoke ločljivosti (HPLC)
Ta evropski standard določa kvantitativno določevanje 16 policikličnih aromatskih ogljikovodikov (PAH) (glej preglednico 2) v blatu, tleh ter obdelanih bioloških odpadkih z metodama GC-MS in HPLC-UV-DAD/FLD, ki zajemata široko območje ravni onesnaženja PAH (glej tudi dodatek B).
S fluorescenčnim zaznavanjem ni mogoče izmeriti acenaftilena.
(...)
Meja detekcije je odvisna od determinant, uporabljene opreme, kakovosti kemikalij, uporabljenih za ekstrakcijo vzorca in očiščenje izvlečka.
Običajno je mogoče za vsak posamezen policiklični aromatski ogljikovodik zagotoviti spodnjo mejo uporabe 0,01 mg/kg (izraženo kot suha snov). To je odvisno od instrumenta in vzorca.
Blato, tla in obdelani biološki odpadki se lahko razlikujejo glede lastnosti, predvidenih ravni onesnaženja s policikličnimi aromatskimi ogljikovodiki ter prisotnosti motečih snovi. Zaradi teh razlik ni možen opis enotnega splošnega postopka. Ta evropski standard vključuje tabele odločanja, ki temeljijo na lastnostih vzorca, ter postopek za ekstrakcijo in očiščenje, ki ga je treba uporabiti. Upoštevata se dve splošni smernici, in sicer stresanje ali tekočinska ekstrakcija po Soxhletu/pod tlakom.
OPOMBA: S to metodo je mogoče analizirati tudi druge snovi PAH pod pogojem, da je bila dokazana primernost.
General Information
Relations
Standards Content (Sample)
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Boden, behandelter Bioabfall und Schlamm - Bestimmung von polycyclischen aromatischen Kohlenwasserstoffen (PAK) mittels Gaschromatographie (GC) und Hochleistungs-Flüssigkeitschromatographie (HPLC)Sols, biodéchets traités et boues - Dosage des hydrocarbures aromatiques polycycliques (HAP) par chromatographie en phase gazeuse et chromatographie liquide à haute performanceSoil, treated biowaste and sludge - Determination of polycyclic aromatic hydrocarbons (PAH) by gas chromatography (GC) and high performance liquid chromatography (HPLC)71.040.50Fizikalnokemijske analitske metodePhysicochemical methods of analysis13.080.10Chemical characteristics of soils13.030.20Liquid wastes. SludgeICS:Ta slovenski standard je istoveten z:EN 16181:2018SIST EN 16181:2018en,fr,de01-december-2018SIST EN 16181:2018SLOVENSKI
STANDARDSIST-TS CEN/TS 16181:20131DGRPHãþD
SIST EN 16181:2018
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 16181
June
t r s z English Version
Soilá treated biowaste and sludge æ Determination of Solsá biodéchets traités et boues æ Dosage des chromatographie en phase gazeuse et chromatographie liquide à haute performance
Bodená behandelter Bioabfall und Schlamm æ Bestimmung von polycyclischen aromatischen This European Standard was approved by CEN on
v April
t r s zä
egulations which stipulate the conditions for giving this European Standard the status of a national standard without any alterationä Upætoædate lists and bibliographical references concerning such national standards may be obtained on application to the CENæCENELEC Management Centre or to any CEN memberä
translation under the responsibility of a CEN member into its own language and notified to the CENæCENELEC Management Centre has the same status as the official versionsä
CEN members are the national standards bodies of Austriaá Belgiumá Bulgariaá Croatiaá Cyprusá Czech Republicá Denmarká Estoniaá Finlandá Former Yugoslav Republic of Macedoniaá Franceá Germanyá Greeceá Hungaryá Icelandá Irelandá Italyá Latviaá Lithuaniaá Luxembourgá Maltaá Netherlandsá Norwayá Polandá Portugalá Romaniaá Serbiaá Slovakiaá Sloveniaá Spainá Swedená Switzerlandá Turkey and United Kingdomä
EUROPEAN COMMITTEE FOR STANDARDIZATION COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre:
Rue de la Science 23,
B-1040 Brussels
9
t r s z CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Membersä Refä Noä EN
s x s z sã t r s z ESIST EN 16181:2018
EN 16181:2018 (E) 2 Contents Page European foreword . 4 Introduction . 5 1 Scope . 6 2 Normative references . 7 3 Terms and definitions . 7 4 Principle . 8 5 Interferences . 9 5.1 Interference with sampling and extraction. 9 5.2 Interference with GC-MS . 9 5.3 Interferences with the HPLC . 9 6 Safety remarks . 10 7 Reagents . 10 8 Apparatus . 15 9 Sample storage and preservation . 17 9.1 Sample storage . 17 9.2 Sample pretreatment . 17 10 Procedure . 18 10.1 Blank test . 18 10.2 Extraction . 18 10.2.1 General. 18 10.2.2 Extraction procedure 1: acetone/hexane-like solvent and agitation . 19 10.2.3 Extraction procedure 2: Soxhlet extraction (dry samples) . 19 10.2.4 Extraction procedure 3: acetone/petroleum ether/sodium chloride and agitation . 20 10.3 Concentration or dilution . 20 10.3.1 General. 20 10.3.2 For HPLC analysis . 21 10.4 Clean-up of the extract . 21 10.4.1 General. 21 10.4.2 Clean-up A – Aluminium oxide . 22 10.4.3 Clean-up B – Silica gel . 22 10.4.4 Clean-up C – Gel permeation chromatography (styrene divinylbenzene resin). 22 10.4.5 Clean-up D – DMF/cyclohexane partitioning for aliphatic hydrocarbons removal . 23 10.5 Addition of the injection standard . 23 10.6 Gas chromatographic analysis (GC) . 23 10.6.1 Gas chromatographic analysis with mass spectrometric detection . 23 10.6.2 Calibration of the method using an internal standard . 24 10.6.3 Measurement . 25 10.6.4 Identification . 25 10.6.5 Check on method performance . 25 10.6.6 Calculation . 26 10.7 High-performance liquid chromatographic analysis (HPLC) . 27 10.7.1 General. 27 10.7.2 Chromatographic separation. 27 SIST EN 16181:2018
EN 16181:2018 (E) 3 10.7.3 Detection . 27 10.7.4 Calibration. 28 10.7.5 Measurement of samples . 29 10.7.6 Calculation . 29 11 Performance characteristics . 29 12 Precision . 30 13 Test report . 30 Annex A (informative)
Repeatability and reproducibility data . 31 A.1 Materials used in the interlaboratory comparison study . 31 A.2 Interlaboratory comparison results . 31 Annex B (informative)
Examples of instrumental conditions and chromatograms . 34 B.1 Measurement of PAH with GC-MS . 34 B.2 Measurement of PAH with HPLC fluorescence . 39 Bibliography . 46
SIST EN 16181:2018
EN 16181:2018 (E) 4 European foreword This document (EN 16181:2018) has been prepared by Technical Committee CEN/TC 444 “Test methods for environmental characterization of solid matrices”, the secretariat of which is held by NEN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by December 2018, and conflicting national standards shall be withdrawn at the latest by December 2018. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN not be held responsible for identifying any or all such patent rights. This document supersedes CEN/TS 16181:2013. This document has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association. The preparation of this document by CEN is based on a mandate by the European Commission (Mandate M/330), which assigned the development of standards on sampling and analytical methods for hygienic and biological parameters as well as inorganic and organic determinants, aiming to make these standards applicable to sludge, treated biowaste and soil as far as this is technically feasible. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. SIST EN 16181:2018
EN 16181:2018 (E) 5 Introduction Polycyclic aromatic hydrocarbons (PAH) are ubiquitous because they are released in appreciable quantities every year into the environment through the combustion of organic matters such as coal, fuel oils, petrol, wood, refuse and plant materials. Since some of these PAH compounds are carcinogenic or mutagenic, their presence in the environment (air, water, soil, sediment and waste) is regularly monitored and controlled. At present determination of PAH is carried out in these matrices in most of the routine laboratories following the preceding steps for sampling, pretreatment, extraction, clean-up by measurement of specific PAH by means of gas chromatography in combination with mass spectrometric detection (GC-MS) or by high performance liquid chromatography (HPLC) in combination with UV-DAD- or fluorescence-detection (HPLC-UV-DAD/FLD). Both the GC-MS and the HPLC methods are included in this horizontal standard. The use of internal and injection standards is described in order to have an internal check on execution of the extraction and clean-up procedure. The method is as far as possible in agreement with the method described for PCBs (see EN 16167). This document is the result of a desk study “Horizontal European Standard for determination of PAH in sludge, soil and biowaste” in the project “Horizontal” and aims at evaluating the latest developments in assessing PAH in sludge, soil, treated biowaste and neighbouring fields. After an evaluation study, in which the ruggedness of the method was studied, a European-wide validation of the draft standard has taken place. The results of the desk studies as well as the evaluation and validation studies have been subject to discussions with all parties concerned in CEN. This European Standard is applicable and validated for several types of matrices as indicated in Table 1 (see also Annex A for the results of the validation). Table 1 — Matrices for which this European Standard is applicable and validated Matrix Materials used for validation Sludge Municipal sludge Biowaste Fresh compost Soil Sandy soil WARNING — Persons using this European Standard should be familiar with usual laboratory practice. This European Standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions. IMPORTANT — It is absolutely essential that tests conducted according to this European Standard be carried out by suitably trained staff. SIST EN 16181:2018
EN 16181:2018 (E) 6 1 Scope This European Standard specifies the quantitative determination of 16 polycyclic aromatic hydrocarbons (PAH) (see Table 2) in sludge, soil and treated biowaste using GC-MS and HPLC-UV-DAD/FLD covering a wide range of PAH contamination levels (see also Annex B). When using fluorescence detection, acenaphthylene cannot be measured. Table 2 — Polycyclic aromatic hydrocarbons which can be analysed using this European Standard Target analyte CAS-RNa Naphthalene 91–20–3 Acenaphthene 83–32–9 Acenaphthylene 208–96–8 Fluorene 86–73–7 Anthracene 120–12–7 Phenanthrene 85–01–8 Fluoranthene 206–44–0 Pyrene 129–00–0 Benz[a]anthracene 56–55–3 Chrysene 218–01–9 Benzo[b]fluoranthene 205–99–2 Benzo[k]fluoranthene 207–08–9 Benzo[a]pyrene 50–32–8 Indeno[1,2,3-cd]pyrene 193–39–5 Dibenz[a,h]anthracene 53–70–3 Benzo[ghi]perylene 191–24–2 a CAS-RN Chemical Abstracts Service Registry Number. The limit of detection depends on the determinants, the equipment used, the quality of chemicals used for the extraction of the sample and the clean-up of the extract. Typically, a lower limit of application of 0,01 mg/kg (expressed as dry matter) can be ensured for each individual PAH. This depends on instrument and sample. Sludge, soil and treated biowaste can differ in properties and also in the expected contamination levels of PAHs and presence of interfering substances. These differences make it impossible to describe one general procedure. This European Standard contains decision tables based on the properties of the sample and the extraction and clean-up procedure to be used. Two general lines are followed, an agitation procedure (shaking) or use of soxhlet/pressurized liquid extraction. NOTE Other PAH compounds can also be analysed with this method, provided suitability has been proven. SIST EN 16181:2018
EN 16181:2018 (E) 7 2 Normative references The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN 15934, Sludge, treated biowaste, soil and waste — Calculation of dry matter fraction after determination of dry residue or water content EN 16179, Sludge, treated biowaste and soil — Guidance for sample pretreatment EN ISO 5667-15, Water quality — Sampling — Part 15: Guidance on the preservation and handling of sludge and sediment samples (ISO 5667-15) EN ISO 16720, Soil quality — Pretreatment of samples by freeze-drying for subsequent analysis (ISO 16720) EN ISO 22892, Soil quality — Guidelines for the identification of target compounds by gas chromatography and mass spectrometry (ISO 22892) ISO 8466-1, Water quality — Calibration and evaluation of analytical methods and estimation of performance characteristics — Part 1: Statistical evaluation of the linear calibration function ISO 18512, Soil quality — Guidance on long and short term storage of soil samples 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 critical pair pair of PAH that will be separated to a predefined degree (e.g. R = 0,5) to ensure chromatographic separation meets minimum quality criteria [SOURCE: EN 15308:2016, 3.6] Note 1 to entry: See Figure 1. SIST EN 16181:2018
EN 16181:2018 (E) 8
Key
difference in retention times of the two peaks a and b in seconds (s) Ya peak width at the base of peak a in seconds (s) Yb peak width at the base of peak b in seconds (s) Figure 1 — Example of a chromatogram of a critical pair 3.2 internal standard isotopically labelled PAH or PAH unlikely to be present in the sample, added to samples prior to extraction, against which the concentrations of native substances are calculated 3.3 injection standard PAH that is unlikely to be present in samples added to the sample extract before injection into the gas chromatograph, to monitor variability of instrument response and the recovery of the internal standards [SOURCE: EN 15527:2008, 3.5, modified – the wording of the definition was revised.] 3.4 Surrogate PAH that is unlikely to be present in samples added to the sample prior to analysis and used to correct for losses of the PAH analytes during sample extraction or clean-up 4 Principle Due to the multi matrix (horizontal) character of this European Standard, different procedures for different steps (modules) are allowed. Which modules should be used depends on the sample. A recommendation is given in this European Standard. Performance criteria are described and it is the responsibility of the laboratories applying this European Standard to show that these criteria are met. Using of spiking standards (internal standards) allows an overall check on the efficiency of a specific combination of modules for a specific sample. But it does not necessarily give the information upon the extensive extraction efficiency of the native PAH bonded to the matrix. SIST EN 16181:2018
EN 16181:2018 (E) 9 After pretreatment to reduce the moisture content and to increase the homogeneity (see 9.2), the test sample is extracted with a solvent. The extract is concentrated and interfering compounds are removed by a clean-up method suitable for the specific matrix. The eluate is concentrated. For HPLC analysis, the concentrated eluate is taken up in an appropriate less volatile water miscible polar solvent and the non-polar eluate residue is removed. The extract is analysed by GC-MS using a capillary column with a stationary phase of low polarity or by HPLC-UV-DAD/FLD with an appropriate reversed phase column. PAH are identified and quantified with GC-MS by comparison of relative retention times and relative peak heights (or peak areas) with respect to internal standards added, and with HPLC by using the corresponding variables of the external standard solutions. The efficiency of the procedure depends on the composition of the matrix that is investigated. 5 Interferences 5.1 Interference with sampling and extraction Use sampling containers of materials (preferably of steel, aluminium or glass) that do not change the sample during the contact time. Avoid plastics and other organic materials during sampling, sample storage or extraction. Keep the samples from direct sunlight and prolonged exposure to light. During storage of the samples, losses of PAH may occur due to adsorption on the walls of the containers. The extent of the losses depends on the storage time. 5.2 Interference with GC-MS Substances that co-elute with the target PAH may interfere with the determination. These interferences may lead to incomplete resolved signals and may, depending on their magnitude, affect accuracy and precision of the analytical results. Peak overlap does not allow an interpretation of the result. Unsymmetrical peaks and peaks broader than the corresponding peaks of the reference substance suggest interferences. Chromatographic separation between dibenzo[a,h]anthracene and indeno[1,2,3-cd]pyrene are mostly critical. Due to their molecular mass differences, quantification can be made by mass selective detection. When incomplete resolution is encountered, peak integration shall be checked and, if necessary, corrected. A chromatographic resolution of R > 0,8 is required for benzo[b]fluoranthene, benzo[k]fluoranthene and benzo[j]fluoranthene (m/z = 252). If the criterion is not met for benzo[j]fluoranthene, the result shall be reported as the sum of the coeluted PAH. 5.3 Interferences with the HPLC Substances that show either fluorescence or quenching and co-elute with the PAHs to be determined may interfere with the determination. These interferences may lead to incompletely resolved signals and may, depending on their magnitude, affect accuracy and precision of the analytical results. Peak overlap does not allow an interpretation of the result. Asymmetrical peaks and peaks being broader than the corresponding peaks of the reference substance suggest interferences. This problem may arise for naphthalene and phenanthrene depending on the selectivity of the phases used. Incomplete removal of the solvents used for sample extraction and clean-up may lead to poor reproducibility of the retention times and wider peaks or double peaks especially for the 2- and 3-ring PAHs. Extracts shall be diluted sufficiently with acetonitrile for the HPLC-analysis, otherwise the detection of naphthalene and 3-ring PAH can be interfered with by a broad toluene peak. Usually perylene is incompletely resolved from benzo[b]fluoranthene, but by choosing a selective wavelength, the perylene peak can be suppressed. SIST EN 16181:2018
EN 16181:2018 (E) 10 6 Safety remarks Certain PAH are highly carcinogenic and shall be handled with extreme care. Avoid contact with solid materials, solvent extracts and solutions of standard PAH. PAH may co-distil with solvent and become deposited outside of stoppered bottles. All containers containing solutions of PAH in solvent shall therefore always be handled using gloves which are solvent resistant and preferably disposable. PAH contamination of vessels can be detected by irradiation with 366 nm UV-light. Vessels containing PAH solutions should be stored standing in beakers to contain any spillage in the case of breakage. Solid PAH are the most dangerous and give rise to a dust hazard due to their crystals becoming electrostatically charged. These materials shall only be handled where proper facilities are available (e.g. adequate fume hoods, protective clothing, dust masks). It is strongly advised that standard solutions are prepared centrally in suitably equipped laboratories or are purchased from suppliers specialized in their preparation. Solvent solutions containing PAH shall be disposed of in a manner approved for disposal of toxic wastes. National regulations shall be followed with respect to all hazards associated with this method. 7 Reagents All reagents shall be of recognized analytical grade. The purity of the reagents used shall be checked by running a blank test as described in 10.1. The blank shall be less than 50 % of the lowest reporting limit. 7.1 Reagents for extraction 7.1.1 Propanone (Acetone), C3H6O. 7.1.2 Hexane-like solvent, boiling range 40 °C to 60 °C. NOTE Hexane-like solvents with a boiling range between 40 °C and 89 °C are allowed. 7.1.3 Toluene, C7H8. 7.1.4 Anhydrous sodium sulphate, Na2SO4. The anhydrous sodium sulphate shall be kept carefully sealed. 7.1.5 Distilled water or water of equivalent quality, H2O. 7.1.6 Sodium chloride, NaCl, anhydrous. 7.2 Reagents for clean-up 7.2.1 Clean-up using aluminium oxide 7.2.1.1 Aluminium oxide, Al2O3. Basic or neutral, specific surface 200 m2/g, activity Super I according to Brockmann. SIST EN 16181:2018
EN 16181:2018 (E) 11 NOTE Brockman Activity Scale is a measure of the percentage of water added to the adsorbent based upon weight/weight relationships between water and the adsorbent. Grade I corresponds to 0 % water added. See [15], [16]. 7.2.1.2 Deactivated aluminium oxide. Deactivated with approximately 10 % water. Add approximately 10 g of water (7.1.5) to 90 g of aluminium oxide (7.2.1.1). Shake until all lumps have disappeared. Allow the aluminium oxide to condition before use for some 16 h, sealed from the air, use it for maximum of two weeks. NOTE The activity depends on the water content. It might be necessary to adjust the water content. 7.2.2 Clean-up using silica gel 60 for column chromatography 7.2.2.1 Silica gel 60, particle size 63 µm to 200 µm. 7.2.2.2 Silica gel 60, water content: mass fraction w(H2O) = 10 %. Silica gel 60, heated for at least 3 h at 450 °C, cooled down and stored in a desiccator containing magnesium perchlorate or a suitable drying agent. Before use, heat at least for 5 h at 130 °C in a drying oven. Then allow cooling in a desiccator and add 10 % water (mass fraction) in a flask. Shake for 5 min intensively until all lumps have disappeared and then for 2 h in a shaking device (8.1.2). Store the deactivated silica gel in the absence of air; use it for a maximum of two weeks. Silica gel 60 is stable for at most 2 weeks. 7.2.3 Clean-up using gel permeation chromatography (GPC) 7.2.3.1 Bio-Beads® S-X31). 7.2.3.2 Ethyl acetate, C4H8O2. 7.2.3.3 Cyclohexane, C6H12. 7.2.3.4 Spherical, porous styrene divinylbenzene resin Preparation of GPC, for example: Put 50 g Bio-Beads® S-X3 (7.2.3.1) into a 500 ml Erlenmeyer flask and add 300 ml elution mixture made up of cyclohexane (7.2.3.3) and ethyl acetate (7.2.3.2) 1:1 (volume) in order to allow the beads to swell. After swirling for a short time until no lumps are left, maintain the flask closed for 24 h. Drain the slurry into the chromatography tube for GPC. After approximately three days, push in the plungers of the column so that a filling level of approximately 35 cm is obtained. To further compress the gel, pump approximately 2 l of elution mixture through the column at a flow rate of 5 ml · min1 and push in the plungers to obtain a filling level of approximately 33 cm. 7.2.4 Clean-up using liquid-liquid partition/DMF/cyclohexane 7.2.4.1 Dimethylformamide (DMF), C3H7NO. 7.2.4.2 Dimethylformamide:water, 9:1.
1) Bio-Beads® is an example of a suitable product available commercially. This information is given for the convenience of this document and does not constitute an endorsement by CEN of this product. Equivalent products may be used if they can be shown to lead to the same results. SIST EN 16181:2018
EN 16181:2018 (E) 12 7.3 Reagents for chromatographic analysis 7.3.1 GC-Analysis 7.3.1.1 Carrier gas for GC-MS: helium or hydrogen of high purity and in accordance with the manufacturer’s specifications. 7.3.2 HPLC-analysis 7.3.2.1 Mobile phase 7.3.2.2 Acetonitrile, CH3CN or methanol, CH3OH, HPLC purity grade. 7.3.2.3 Ultra-pure water, HPLC purity grade. 7.3.2.4 Helium, He, of suitable purity for degasification of solvents. 7.4 Standards 7.4.1 Reference substanc
...
SLOVENSKI STANDARD
oSIST prEN 16181:2017
01-december-2017
%ODWRREGHODQLELRORãNLRGSDGNLLQWOD'RORþHYDQMHSROLFLNOLþQLKDURPDWVNLK
RJOMLNRYRGLNRY3$+VSOLQVNRNURPDWRJUDILMR*&LQVWHNRþLQVNRNURPDWRJUDILMR
YLVRNHORþOMLYRVWL+3/&
Soil, treated biowaste and sludge - Determination of polycyclic aromatic hydrocarbons
(PAH) by gas chromatography (GC) and high performance liquid chromatography
(HPLC)
Boden, behandelter Bioabfall und Schlamm - Bestimmung von polycyclischen
aromatischen Kohlenwasserstoffen (PAK) mittels Gaschromatographie (GC) und
Hochleistungs-Flüssigkeitschromatographie (HPLC)
Sols, biodéchets traités et boues - Dosage des hydrocarbures aromatiques polycycliques
(HAP) par chromatographie en phase gazeuse et chromatographie liquide à haute
performance
Ta slovenski standard je istoveten z: prEN 16181
ICS:
13.030.20 7HNRþLRGSDGNL%ODWR Liquid wastes. Sludge
13.080.10 .HPLMVNH]QDþLOQRVWLWDO Chemical characteristics of
soils
oSIST prEN 16181:2017 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------
oSIST prEN 16181:2017
---------------------- Page: 2 ----------------------
oSIST prEN 16181:2017
DRAFT
EUROPEAN STANDARD
prEN 16181
NORME EUROPÉENNE
EUROPÄISCHE NORM
October 2017
ICS 13.030.01; 13.080.10 Will supersede CEN/TS 16181:2013
English Version
Soil, treated biowaste and sludge - Determination of
polycyclic aromatic hydrocarbons (PAH) by gas
chromatography (GC) and high performance liquid
chromatography (HPLC)
Schlamm, behandelter Bioabfall und Boden -
Bestimmung von polycyclischen aromatischen
Kohlenwasserstoffen (PAK) mittels
Gaschromatographie (GC) und Hochleistungs-
Flüssigkeitschromatographie (HPLC)
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 444.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2017 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 16181:2017 E
worldwide for CEN national Members.
---------------------- Page: 3 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
Contents Page
European foreword . 4
Introduction . 5
1 Scope . 6
2 Normative references . 7
3 Terms and definitions . 7
4 Principle . 8
5 Interferences . 9
5.1 Interference with sampling and extraction. 9
5.2 Interference with GC-MS . 9
5.3 Interferences with the HPLC . 9
6 Safety remarks . 10
7 Reagents . 10
8 Apparatus . 15
9 Sample storage and preservation . 17
9.1 Sample storage . 17
9.2 Sample pretreatment . 17
10 Procedure . 18
10.1 Blank test . 18
10.2 Extraction . 18
10.2.1 General. 18
10.2.2 Extraction procedure 1: acetone/hexane-like solvent and agitation . 19
10.2.3 Extraction procedure 2: Soxhlet extraction (dry samples) . 19
10.2.4 Extraction procedure 3: acetone/petroleum ether/sodium chloride and agitation . 20
10.3 Concentration or dilution . 20
10.3.1 General. 20
10.3.2 For HPLC analysis . 21
10.4 Clean-up of the extract . 21
10.4.1 General. 21
10.4.2 Clean-up A – Aluminium oxide . 22
10.4.3 Clean-up B – Silica gel . 22
10.4.4 Clean-up C – Gel permeation chromatography (styrene divinylbenzene resin). 22
10.4.5 Clean-up D – DMF/cyclohexane partitioning for aliphatic hydrocarbons removal . 22
10.5 Addition of the injection standard . 23
10.6 Gas chromatographic analysis (GC) . 23
10.6.1 Gas chromatographic analysis with mass spectrometric detection . 23
10.6.2 Calibration of the method using an internal standard . 24
10.6.3 Measurement . 25
10.6.4 Identification . 25
10.6.5 Check on method performance . 25
10.6.6 Calculation . 26
10.7 High-performance liquid chromatographic analysis (HPLC) . 27
10.7.1 General. 27
10.7.2 Chromatographic separation. 27
2
---------------------- Page: 4 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
10.7.3 Detection . 27
10.7.4 Calibration. 28
10.7.5 Measurement of samples . 29
10.7.6 Calculation . 29
11 Performance characteristics . 29
12 Precision . 30
13 Test report . 30
Annex A (informative) Repeatability and reproducibility data . 31
A.1 Materials used in the interlaboratory comparison study . 31
A.2 Interlaboratory comparison results . 31
Annex B (informative) Examples of instrumental conditions and chromatograms . 34
B.1 Measurement of PAH with GC-MS . 34
B.2 Measurement of PAH with HPLC fluorescence . 40
Bibliography . 46
3
---------------------- Page: 5 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
European foreword
This document (prEN 16181:2017) has been prepared by Technical Committee CEN/TC 444 “Test
methods for environmental characterization of solid matrices”, the secretariat of which is held by NEN.
This document is currently submitted to the CEN Enquiry.
This document will supersede CEN/TS 16181:2013.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association.
The preparation of this document by CEN is based on a mandate by the European Commission
(Mandate M/330), which assigned the development of standards on sampling and analytical methods
for hygienic and biological parameters as well as inorganic and organic determinants, aiming to make
these standards applicable to sludge, treated biowaste and soil as far as this is technically feasible.
4
---------------------- Page: 6 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
Introduction
Polycyclic aromatic hydrocarbons (PAH) are ubiquitous because they are released in appreciable
quantities every year into the environment through the combustion of organic matters such as coal, fuel
oils, petrol, wood, refuse and plant materials. Since some of these PAH compounds are carcinogenic or
mutagenic, their presence in the environment (air, water, soil, sediment and waste) is regularly
monitored and controlled. At present determination of PAH is carried out in these matrices in most of
the routine laboratories following the preceding steps for sampling, pretreatment, extraction, clean-up
by measurement of specific PAH by means of gas chromatography in combination with mass
spectrometric detection (GC-MS) or by high performance liquid chromatography (HPLC) in
combination with UV-DAD- or fluorescence-detection (HPLC-UV-DAD/FLD). Both the GC-MS and the
HPLC methods are included in this horizontal standard.
It is to be underlined that the target contamination level of PAH can lie in the range of about 0,01 mg/kg
per individual PAH (agricultural soil and sediment) to about 200 mg/kg and higher (e.g. contaminated
soil at coking plant sites or waste). The use of internal and injection standards is described in order to
have an internal check on execution of the extraction and clean-up procedure. The method is as far as
possible in agreement with the method described for PCBs (see EN 16167).
This document is the result of a desk study “Horizontal European Standard for determination of PAH in
sludge, soil and biowaste” in the project “Horizontal” and aims at evaluating the latest developments in
assessing PAH in sludge, soil, treated biowaste and neighbouring fields. After an evaluation study, in
which the ruggedness of the method was studied, a European-wide validation of the draft standard has
taken place. The results of the desk studies as well as the evaluation and validation studies have been
subject to discussions with all parties concerned in CEN.
This European Standard is applicable and validated for several types of matrices as indicated in Table 1
(see also Annex A for the results of the validation).
Table 1 — Matrices for which this European Standard is applicable and validated
Matrix Materials used for validation
Sludge Municipal sludge
Biowaste Fresh compost
Soil Sandy soil
WARNING — Persons using this European Standard should be familiar with usual laboratory practice.
This European Standard does not purport to address all of the safety problems, if any, associated with
its use. It is the responsibility of the user to establish appropriate safety and health practices and to
ensure compliance with any national regulatory conditions.
IMPORTANT — It is absolutely essential that tests conducted according to this European
Standard be carried out by suitably trained staff.
5
---------------------- Page: 7 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
1 Scope
This European Standard specifies the quantitative determination of 16 polycyclic aromatic
hydrocarbons (PAH) (see Table 2) in sludge, soil and treated biowaste using GC-MS and HPLC-UV-
DAD/FLD covering a wide range of PAH contamination levels (see also Annex B).
When using fluorescence detection, acenaphthylene cannot be measured.
Table 2 — Polycyclic aromatic hydrocarbons which can be analysed using this European
Standard
Target analyte a
CAS-RN
Naphthalene 91–20–3
Acenaphthene 83–32–9
Acenaphthylene 208–96–8
Fluorene 86–73–7
Anthracene 120–12–7
Phenanthrene 85–01–8
Fluoranthene 206–44–0
Pyrene 129–00–0
Benz[a]anthracene 56–55–3
Chrysene 218–01–9
Benzo[b]fluoranthene 205–99–2
Benzo[k]fluoranthene 207–08–9
Benzo[a]pyrene 50–32–8
Indeno[1,2,3-cd]pyrene 193–39–5
Dibenz[a,h]anthracene 53–70–3
Benzo[ghi]perylene 191–24–2
a
CAS-RN Chemical Abstracts Service Registry Number.
The limit of detection depends on the determinants, the equipment used, the quality of chemicals used
for the extraction of the sample and the clean-up of the extract.
Typically, a lower limit of application of 0,01 mg/kg (expressed as dry matter) can be ensured for each
individual PAH. This depends on instrument and sample.
Sludge, soil and treated biowaste can differ in properties and also in the expected contamination levels
of PAHs and presence of interfering substances. These differences make it impossible to describe one
general procedure. This European Standard contains decision tables based on the properties of the
sample and the extraction and clean-up procedure to be used. Two general lines are followed, an
agitation procedure (shaking) or use of soxhlet/pressurized liquid extraction.
NOTE Other PAH compounds can also be analysed with this method, provided suitability has been proven.
6
---------------------- Page: 8 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN 15934, Sludge, treated biowaste, soil and waste — Calculation of dry matter fraction after
determination of dry residue or water content
EN 16179, Sludge, treated biowaste and soil — Guidance for sample pretreatment
EN ISO 5667-15, Water quality — Sampling — Part 15: Guidance on the preservation and handling of
sludge and sediment samples (ISO 5667-15)
EN ISO 16720, Soil quality — Pretreatment of samples by freeze-drying for subsequent analysis
(ISO 16720)
EN ISO 22892, Soil quality — Guidelines for the identification of target compounds by gas
chromatography and mass spectrometry (ISO 22892)
ISO 8466-1, Water quality — Calibration and evaluation of analytical methods and estimation of
performance characteristics — Part 1: Statistical evaluation of the linear calibration function
ISO 18512, Soil quality — Guidance on long and short term storage of soil samples
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
critical pair
pair of PAH that will be separated to a predefined degree (e.g. R = 0,5) to ensure chromatographic
separation meets minimum quality criteria
[SOURCE: EN 15308:2016, 3.6]
Note 1 to entry: See Figure 1.
7
---------------------- Page: 9 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
Key
Δt difference in retention times of the two peaks a and b in seconds (s)
Y peak width at the base of peak a in seconds (s)
a
Y peak width at the base of peak b in seconds (s)
b
Figure 1 — Example of a chromatogram of a critical pair
3.2
internal standard
isotopically labelled PAH or PAH unlikely to be present in the sample, added to samples prior to
extraction, against which the concentrations of native substances are calculated
3.3
injection standard
PAH that is unlikely to be present in samples added to the sample extract before injection into the gas
chromatograph, to monitor variability of instrument response and the recovery of the internal
standards
[SOURCE: EN 15527:2008, 3.5, modified – the wording of the definition was revised.]
3.4
Surrogate
PAH that is unlikely to be present in samples added to the sample prior to analysis and used to correct
for losses of the PAH analytes during sample extraction or clean-up
4 Principle
Due to the multi matrix (horizontal) character of this European Standard, different procedures for
different steps (modules) are allowed. Which modules should be used depends on the sample. A
recommendation is given in this European Standard. Performance criteria are described and it is the
responsibility of the laboratories applying this European Standard to show that these criteria are met.
Using of spiking standards (internal standards) allows an overall check on the efficiency of a specific
combination of modules for a specific sample. But it does not necessarily give the information upon the
extensive extraction efficiency of the native PAH bonded to the matrix.
8
---------------------- Page: 10 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
After pretreatment to reduce the moisture content and to increase the homogeneity (see 9.2), the test
sample is extracted with a solvent.
The extract is concentrated and interfering compounds are removed by a clean-up method suitable for
the specific matrix. The eluate is concentrated. For HPLC analysis, the concentrated eluate is taken up in
an appropriate less volatile water miscible polar solvent and the non-polar eluate residue is removed.
The extract is analysed by GC-MS using a capillary column with a stationary phase of low polarity or by
HPLC-UV-DAD/FLD with an appropriate reversed phase column.
PAH are identified and quantified with GC-MS by comparison of relative retention times and relative
peak heights (or peak areas) with respect to internal standards added, and with HPLC by using the
corresponding variables of the external standard solutions. The efficiency of the procedure depends on
the composition of the matrix that is investigated.
5 Interferences
5.1 Interference with sampling and extraction
Use sampling containers of materials (preferably of steel, aluminium or glass) that do not change the
sample during the contact time. Avoid plastics and other organic materials during sampling, sample
storage or extraction. Keep the samples from direct sunlight and prolonged exposure to light.
During storage of the samples, losses of PAH may occur due to adsorption on the walls of the containers.
The extent of the losses depends on the storage time.
5.2 Interference with GC-MS
Substances that co-elute with the target PAH may interfere with the determination. These interferences
may lead to incomplete resolved signals and may, depending on their magnitude, affect accuracy and
precision of the analytical results. Peak overlap does not allow an interpretation of the result.
Unsymmetrical peaks and peaks broader than the corresponding peaks of the reference substance
suggest interferences.
Chromatographic separation between dibenzo[a,h]anthracene and indeno[1,2,3-cd]pyrene are mostly
critical. Due to their molecular mass differences, quantification can be made by mass selective
detection. When incomplete resolution is encountered, peak integration shall be checked and, if
necessary, corrected.
A chromatographic resolution of R > 0,8 is required for benzo[b]fluoranthene, benzo[k]fluoranthene
and benzo[k]fluoranthene (m/z = 252). If the criterion is not met for benzo[j]fluoranthene, the result
shall be reported as the sum of the coeluted PAH.
5.3 Interferences with the HPLC
Substances that show either fluorescence or quenching and co-elute with the PAHs to be determined
may interfere with the determination. These interferences may lead to incompletely resolved signals
and may, depending on their magnitude, affect accuracy and precision of the analytical results. Peak
overlap does not allow an interpretation of the result. Asymmetrical peaks and peaks being broader
than the corresponding peaks of the reference substance suggest interferences. This problem may arise
for naphthalene and phenanthrene depending on the selectivity of the phases used.
Incomplete removal of the solvents used for sample extraction and clean-up may lead to poor
reproducibility of the retention times and wider peaks or double peaks especially for the 2- and 3-ring
PAHs. Extracts shall be diluted sufficiently with acetonitrile for the HPLC-analysis, otherwise the
detection of naphthalene and 3-ring PAH can be interfered with by a broad toluene peak.
Usually perylene is incompletely resolved from benzo[b]fluoranthene, but by choosing a selective
wavelength, the perylene peak can be suppressed.
9
---------------------- Page: 11 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
6 Safety remarks
Certain PAH are highly carcinogenic and shall be handled with extreme care. Avoid contact with solid
materials, solvent extracts and solutions of standard PAH.
PAH may co-distil with solvent and become deposited outside of stoppered bottles. All containers
containing solutions of PAH in solvent shall therefore always be handled using gloves which are solvent
resistant and preferably disposable.
PAH contamination of vessels may be detected by irradiation with 366 nm UV-light.
Vessels containing PAH solutions should be stored standing in beakers to contain any spillage in the
case of breakage.
Solid PAH are the most dangerous and give rise to a dust hazard due to their crystals becoming
electrostatically charged. These materials shall only be handled where proper facilities are available
(e.g. adequate fume hoods, protective clothing, dust masks). It is strongly advised that standard
solutions are prepared centrally in suitably equipped laboratories or are purchased from suppliers
specialized in their preparation.
Solvent solutions containing PAH shall be disposed of in a manner approved for disposal of toxic
wastes.
National regulations shall be followed with respect to all hazards associated with this method.
7 Reagents
All reagents shall be of recognized analytical grade. The purity of the reagents used shall be checked by
running a blank test as described in 10.1. The blank shall be less than 50 % of the lowest reporting limit.
7.1 Reagents for extraction:
7.1.1 Acetone, C H O.
3 6
7.1.2 Hexane-like solvent, boiling range 40 °C to 60 °C.
NOTE Hexane-like solvents with a boiling range between 40 °C and 89 °C are allowed.
7.1.3 Toluene, C H .
7 8
7.1.4 Anhydrous sodium sulfate, Na SO .
2 4
The anhydrous sodium sulfate shall be kept carefully sealed.
7.1.5 Distilled water or water of equivalent quality, H O.
2
7.1.6 Sodium chloride, NaCl, anhydrous.
7.2 Reagents for clean-up:
7.2.1 Clean-up using aluminium oxide:
7.2.1.1 Aluminium oxide, Al O .
2 3
2
Basic or neutral, specific surface 200 m /g, activity Super I according to Brockmann.
10
---------------------- Page: 12 ----------------------
oSIST prEN 16181:2017
prEN 16181:2017 (E)
NOTE Brockman Activity Scale is a measure of the percentage of water added to the adsorbent based upon
weight/weight relationships between water and the adsorbent. Grade I corresponds to 0 % water added. See [15],
[16].
7.2.1.2 Deactivated aluminium oxide.
Deactivated with approximately 10 % water.
Add approximately 10 g of water (7.1.5) to 90 g of aluminium oxide (7.2.1.1). Shake until all lumps have
disappeared. Allow the aluminium oxide to condition before use for some 16 h, sealed from the air, use
it for maximum of two weeks.
NOTE The activity depends on the water content. It might be necessary to adjust the water content.
7.2.2 Clean-up using silica gel 60 for column chromatography:
7.2.2.1 Silica gel 60, particle size 63 µm to 200 µm.
7.2.2.2 Silica gel 60, water content: mass fraction w(H O) = 10 %.
2
Silica gel 60, heated for at least 3 h at 450 °C, cooled down in a desiccator and stored containing
magnesium perchlorate or a suitable drying agent. Before use, heat at least for 5 h at 130 °C in a drying
oven. Then allow cooling in a desiccator and add 10 % water (mass fraction) in a flask. Shake for 5 min
intensively until all lumps have disappeared and then for 2 h in a shaking device (8.1.2). Store the
deactivated silica gel in the absence of air; use it for a maximum of two weeks.
Silica gel 60 is stable for at most 2 weeks.
7.2.3 Clean-up using gel permeation chromatography (GPC):
® 1)
7.2.3.1 Bio-Beads S-X3 .
7.2.3.2 Ethyl acetate, C H O .
4 8 2
7.2.3.3 Cyclohexane, C H .
6 12
7.2.3.4 Spherical, porous styrene divinylbenzene resin:
®
Preparation of GPC, for example: Put 50 g Bio-Beads S-X3 (7.2.3.1) into a 500 ml Erlenmeyer flask and
add 300 ml elution mixture made up of cyclohexane (7.2.3.3) and ethyl acetate (7.2.3.2) 1:1 (volume) in
order to allow the beads to swell. After swirling for a short time until no lumps are left, maintain the
f
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.