Ambient air - Determination of lung deposited surface area (LDSA) concentration using aerosol monitors based on diffusion charging

This document specifies a process for the electrical diffusion charging of aerosols with subsequent measurement of particle charge. With the aid of this method, it is possible to determine the lung-deposited surface area (LDSA) concentration of particles in ambient air. Depending on the design of the electrical diffusion charger, the LDSA of particles in the size range of approximately 20 nm to approximately 300 nm is measurable.
Furthermore, this document specifies design criteria for LDSA measuring aerosol monitors as well as performance criteria and the associated test procedures. The performance criteria depend on the application and they are more stringent when the instrument is operated in an air quality monitoring station.
In the determination of the LDSA concentration, the share of geometric particle surface area concentration is determined that can be deposited in the alveolar region of the human lung. Typical particle surface area concentrations with alveolar deposition measured in urban areas range from 5 µm2/cm3 to 50 µm2/cm3.
Instruments based on this measurement principle can be designed to be very compact with a low power consumption. This makes them ideally suited for handheld measurements, other forms of mobile application or to measure personal exposure. On the other hand, they can be easily adapted to serve as a stationary instrument in air quality monitoring stations.

Außenluft - Bestimmung der lungendeponierbaren Oberflächenkonzentration (LDSA) mit Aerosolmonitoren auf Basis der Diffusionsaufladung

Dieses Dokument legt ein Verfahren zur elektrischen Diffusionsaufladung von Aerosolen mit anschließender Messung der Partikelladung fest. Mit diesem Verfahren kann die lungendeponierbare Oberflächen¬konzentration (LDSA, en: lung-deposited surface area concentration) von Partikeln in Außenluft gemessen werden. Je nach Ausführungsform des elektrischen Diffusionsaufladers ist die LDSA von Partikeln im Größenbereich von etwa 20 nm bis etwa 300 nm messbar.
Darüber hinaus legt dieses Dokument Konstruktionskriterien für Aerosolmonitore zur LDSA-Messung sowie Leistungskriterien und die damit verbundenen Prüfverfahren fest. Die Leistungskriterien hängen von der Anwendung ab und sind stringenter, wenn das Instrument in einer Luftgütemessstation betrieben wird.
Bei der Bestimmung der LDSA-Konzentration wird der Anteil der geometrischen Partikeloberflächen-konzentration, der im alveolaren Bereich der menschlichen Lunge abgeschieden werden kann, ermittelt. Typische in städtischen Gebieten gemessene alveolar deponierbare Partikeloberflächenkonzentrationen liegen zwischen 5 µm2/cm3 und 50 µm2/cm3.
Auf diesem Messprinzip basierende Geräte können sehr kompakt und mit geringem Stromverbrauch ausgeführt werden. Hierdurch sind sie einerseits sehr gut für Handmessungen, andere mobile Anwendungen und die Messung der persönlichen Exposition geeignet. Andererseits können sie leicht für den stationären Betrieb in Luftgütemessstationen angepasst werden.

Air ambiant - Détermination de la concentration en surface spécifique des particules pouvant se déposer dans les poumons (LDSA) à l’aide de moniteurs d’aérosols basés sur la charge par diffusion

Zunanji zrak - Določanje koncentracije delcev. ki se lahko usedajo na površino pljuč (LDSA) z uporabo monitorjev aerosolov na podlagi difuzije in naboja

General Information

Status
Not Published
Public Enquiry End Date
01-Jul-2024
Technical Committee
Current Stage
5520 - Unique Acceptance Procedure (UAP) (Adopted Project)
Start Date
22-Apr-2024
Due Date
09-Sep-2024

Buy Standard

Draft
kTS FprCEN/TS 18073:2024 - BARVE
English language
32 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
kSIST-TS FprCEN/TS 18073:2024
01-junij-2024
Zunanji zrak - Določanje koncentracije delcev. ki se lahko usedajo na površino
pljuč (LDSA) z uporabo monitorjev aerosolov na podlagi difuzije in naboja
Ambient air - Determination of lung deposited surface area (LDSA) concentration using
aerosol monitors based on diffusion charging
Außenluft - Bestimmung der lungendeponierbaren Oberflächenkonzentration (LDSA) mit
Aerosolmonitoren auf Basis der Diffusionsaufladung
Air ambiant - Détermination de la concentration en surface spécifique des particules
pouvant se déposer dans les poumons (LDSA) à l’aide de moniteurs d’aérosols basés
sur la charge par diffusion
Ta slovenski standard je istoveten z: FprCEN/TS 18073
ICS:
13.040.20 Kakovost okoljskega zraka Ambient atmospheres
kSIST-TS FprCEN/TS 18073:2024 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
kSIST-TS FprCEN/TS 18073:2024

---------------------- Page: 2 ----------------------
kSIST-TS FprCEN/TS 18073:2024


FINAL DRAFT
TECHNICAL SPECIFICATION
FprCEN/TS 18073
SPÉCIFICATION TECHNIQUE

TECHNISCHE SPEZIFIKATION

April 2024
ICS 13.040.20
English Version

Ambient air - Determination of lung deposited surface area
(LDSA) concentration using aerosol monitors based on
diffusion charging
Air ambiant - Détermination de la concentration Außenluft - Bestimmung der lungendeponierbaren
surfacique des particules d'aérosol atmosphérique à Oberflächenkonzentration (LDSA) mit
l'aide de moniteurs d'aérosols électriques basés sur la Aerosolmonitoren auf Basis der Diffusionsaufladung
charge par diffusion


This draft Technical Specification is submitted to CEN members for Vote. It has been drawn up by the Technical Committee
CEN/TC 264.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a Technical Specification. It is distributed for review and comments. It is subject to change
without notice and shall not be referred to as a Technical Specification.


EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2024 CEN All rights of exploitation in any form and by any means reserved Ref. No. FprCEN/TS 18073:2024 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------
kSIST-TS FprCEN/TS 18073:2024
FprCEN/TS 18073:2024 (E)
Contents Page
European foreword . 4
Introduction . 5
1 Scope . 6
2 Normative references . 6
3 Terms and definitions . 6
3.1 Aerosol properties. 6
3.2 Particle size metrics . 7
3.3 Particle concentration metrics . 7
4 Principle . 8
4.1 Physical principles . 8
4.2 Physiological principles . 8
5 Function .
...