oSIST prEN ISO 9360-1:2026
(Main)Heat and moisture exchangers, and breathing system filters for respired gases - Part 1: Common requirements (ISO/DIS 9360-1:2026)
- Status
- Not Published
- Public Enquiry End Date
- 30-Oct-2026
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 4020 - Public enquire (PE) (Adopted Project)
- Start Date
- 04-Sep-2026
- Due Date
- 22-Jan-2027
- Completion Date
- 02-Nov-2026
oSIST prEN ISO 9360-1:2026 is the Slovenian draft standard based on ISO/DIS 9360-1:2026, published by SIST. It specifies common requirements for heat and moisture exchangers (HME), breathing system filters (BSF), and combined HMEF devices for anaesthetic and respiratory use. Manufacturers, test laboratories, and clinical users use it to select connectors, verify performance, and mark the device clearly.
What does oSIST prEN ISO 9360-1:2026 specify?
oSIST prEN ISO 9360-1:2026 specifies the common requirements for HME, BSF and HMEF devices intended for anaesthetic and respiratory use. It covers materials, design, performance, devices supplied sterile, marking, and information to be supplied by the manufacturer.
The document applies to devices with or without a machine port for connection to a ventilator. It excludes filters for vacuum sources, gas sample lines, compressed gases, physiological respiratory test equipment, resuscitation bag inlets, nebuliser ports, and filters used to filter anaesthetic gases from respiratory gas.
The document is organized into clauses on scope, normative references, terms and definitions, general requirements, materials, design requirements, cleaning/disinfection/sterilization, and manufacturer information. It also includes two informative annexes.
| Annex | What it covers |
|---|---|
| Annex A | Rationale for selected requirements, especially housing visibility, colour coding and marking |
| Annex B | Test method for assessing visibility or legibility of markings |
What are the key requirements of oSIST prEN ISO 9360-1:2026?
oSIST prEN ISO 9360-1:2026 uses ISO 18190:2025 as the general baseline for several core clauses, so many general airway-device requirements come from that related standard. In practice, that means this part adds the HME, BSF and HMEF-specific points rather than redefining every general safety rule.
Test conditions and material-related requirements
Clause 4.2 sets the ambient conditions for testing. The tests are done under defined temperature, humidity and pressure conditions, so results are comparable between laboratories.
Clause 5 applies the material requirements of ISO 18190:2025, and Clause 5.2 adds a housing-specific visibility rule. For devices with a machine port intended for mechanical ventilation, liquid water and secretions shall be visible in the housing. This matters because condensate can change pressure drop, reduce humidification performance, and create a risk of inhaled water.
Identification and connector design
Clause 6.2.1 uses colour coding to distinguish device types. BSF housings or markings use yellow, orange or red, while HME and HMEF housings or markings use blue or green. The colour must remain visible in use, which helps clinical users avoid placing the wrong device in a breathing circuit.
Clause 6.3 requires conical connection ports that conform to ISO 5356-1, with port configurations defined for the patient side and the machine side. For HME and HMEF without a machine port, the patient connection port is a 15 mm conical socket. This matters because connector compatibility is a primary control against misconnections.
Accessory and oxygen connections
Clause 6.3.3 covers accessory ports, when provided. An accessory port has to be compatible with a Luer lock connector conforming to ISO 80369-7, have a tethered sealing cap or equivalent retention feature, and carry a warning about the hazard of inadvertent Luer connections. In practice, this is aimed at common sampling ports and reduces the risk of connecting the wrong line.
Clause 6.3.4 covers supplementary oxygen connections on HME or HMEF devices without a machine port for spontaneously breathing patients. Such a connection shall either conform to the R2 socket connector of ISO 80369-2 or have a permanently connected oxygen tube. This helps control how oxygen is added to the breathing path.
Performance tests
Clause 6.4.2 requires moisture loss to be measured by the method in ISO 9360-2. Clause 6.4.3 requires filtration performance for BSF and HMEF to be measured by the method in ISO 9360-3. Clause 6.4.4, 6.4.5 and 6.4.6 cover resistance to flow, gas leakage and compliance.
These tests matter because they describe how the device behaves in a breathing system:
- resistance to flow is checked at flowrates for neonatal, paediatric and adult intended use
- gas leakage is limited and measured under internal pressure
- compliance is measured from the change in volume under pressure
The same resistance, leakage and compliance tests are repeated after preconditioning for the recommended maximum time of use, so the device is assessed in a more realistic in-use condition.
Marking, packaging and instructions for use
Clause 8 requires the manufacturer to provide device marking, packaging labelling and instructions for use in line with ISO 18190. The device marking identifies orientation, uses the correct colour coding where applicable, and shows the device type as BSF, HME or HMEF. The packaging label also has to identify the device clearly.
The instructions for use must give the recommended range of tidal volumes and/or flowrates, measured resistance to flow, internal volume, gas leakage, compliance, and, for HME/HMEF, moisture loss. For BSF/HMEF, the instructions also include penetration information, plus warnings about using a BSF in place of an HME, using small-bore connectors, and using nebulizers, inhaled medications or humidifiers when relevant. That information helps buyers, clinicians and trainers match the device to the intended breathing circuit.
What terms does oSIST prEN ISO 9360-1:2026 define?
- Heat and moisture exchanger (HME) - a device that retains some expired heat and moisture and returns them to the patient during inspiration.
- Breathing system filter (BSF) - a device intended to reduce airborne particulates, including microorganisms, in breathing systems.
- HMEF - a heat and moisture exchanger that also incorporates the function of a breathing system filter.
- Active HME - an HME in which water, water vapour or heat is actively added to increase gas humidity.
- Machine port - the port intended for connection to a patient connection port of a breathing set.
- Patient connection port - the port of the breathing system intended for connection to an airway device such as a tracheal tube, tracheostomy tube, face mask or supra-laryngeal airway.
- Internal volume - the compressible volume contained within the device and occupied by breathing gases.
- Pressure drop - the difference between upstream and downstream pressure at a specified constant gas flowrate through the device.
Who uses oSIST prEN ISO 9360-1:2026?
Manufacturers of anaesthetic and respiratory equipment use oSIST prEN ISO 9360-1:2026 to design the housing, ports, markings and performance claims for HME, BSF and HMEF devices. Test laboratories use it to run the moisture loss, filtration, resistance, leakage, compliance and legibility checks.
Clinical engineers, respiratory teams, anaesthesia teams and buyers use it to distinguish the right device type for a breathing circuit and to check compatibility with connectors and accessories. It is especially relevant in hospitals and other healthcare settings that use ventilator breathing systems and anaesthetic breathing systems.
What changed in oSIST prEN ISO 9360-1:2026 from the previous edition?
The foreword says this part technically revises and replaces ISO 9360-1:2000, ISO 9360-2:2001, ISO 23328-1:2008 and ISO 23328-2:2009. It also says the series combines requirements from those earlier standards.
The stated changes include:
- the layout now follows the format of ISO 18190:2025 and uses its clause structure as a reference
- colour coding requirements were added to distinguish between HMEs and BSFs
Which standards are used with oSIST prEN ISO 9360-1:2026?
- ISO 18190:2025 - supplies the general requirements for airway devices and related equipment that this document reuses in several clauses.
- ISO 5356-1 - defines the conical connectors used for the device connection ports.
- ISO 80369-2 - supports the R2 connector option for supplementary oxygen connections.
- ISO 80369-7 - defines the Luer lock connector used for accessory ports.
- IEC 60601-1 - is referenced for the symbols and safety sign used when marking accessory ports.
- ISO 9360-2:202X - gives the method for measuring moisture loss and the preconditioning basis for several tests.
- ISO 9360-3:202X - gives the method for measuring filtration performance.
- ISO TS 24172 - provides an alternative test method for heat and moisture exchangers.
- ISO 80601-2-74 - is named in the scope for active HMEs.
- ISO 4135:2022 - supplies the terms and definitions used in Clause 3.
What does the oSIST prEN ISO 9360-1:2026 document contain?
The document contains a set of product requirements, test methods and user-information requirements for HME, BSF and HMEF devices. Its main technical content is in the tables, figures and test procedures that define connector forms, colour coding, resistance test flowrates, leakage limits and compliance measurement.
It includes a housing visibility test, connector configuration guidance, a resistance-to-flow test setup, a gas-leakage method and a compliance method. It also includes a visibility/legibility test in Annex B for checking whether markings can be read in use. For documentation and lab work, the most useful items are Table 1 on colours, Table 2 on connection ports, Table 3 on test requirements, Table 4 on flowrates, Table 5 on device labels, Figure 1 on housing transparency and Figure 3 on the resistance test arrangement.
Relations
- Effective Date
- 19-Feb-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
oSIST prEN ISO 9360-1:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Heat and moisture exchangers, and breathing system filters for respired gases - Part 1: Common requirements (ISO/DIS 9360-1:2026)". This standard covers: La présente partie de l'ISO 9360 spécifie certaines exigences relatives aux échangeurs de chaleur et d'humidité (ECH), y compris ceux équipés de filtres pour systèmes respiratoires, essentiellement destinés à humidifier les gaz respirés par les patients ayant un volume courant supérieur ou égal à 250 ml, et comportant au moins un orifice côté appareil, et décrit des méthodes d'essai permettant de les évaluer.
La présente partie de l'ISO 9360 spécifie certaines exigences relatives aux échangeurs de chaleur et d'humidité (ECH), y compris ceux équipés de filtres pour systèmes respiratoires, essentiellement destinés à humidifier les gaz respirés par les patients ayant un volume courant supérieur ou égal à 250 ml, et comportant au moins un orifice côté appareil, et décrit des méthodes d'essai permettant de les évaluer.
oSIST prEN ISO 9360-1:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
oSIST prEN ISO 9360-1:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 9360-1:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
oSIST prEN ISO 9360-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 accessory port
- 3.2 active HME
- 3.3 breathing set
- 3.4 breathing system filter
- 3.5 compliance
- 3.6 Heat and moisture exchanger, noun
- 3.7 HMEF, FHME, noun
- 3.8 internal volume, noun
- 3.9 machine port
- 3.10 moisture loss, noun
- 3.11 patient connection port
- 3.12 pressure drop, noun
- 4 General requirements
- 4.1 General
- 4.2 Ambient conditions of test
- 5 Materials
- 5.1 General
- 5.2 Housing
- 6 Design Requirements
- 6.1 General
- 6.2 Identification of the device
- 6.3 Device connections
- 6.4 Performance Requirements
- 4 HME/HMEF/BSF under test
- 7 Cleaning, disinfection and sterilization
- 8 Information to be supplied by the manufacturer
- 8.1 Marking on the device
- 8.2 Labelling on the packaging
- 8.3 Instructions for use
- Annex A
- Annex B
- Bibliography
Introduction
The gases generally available for medical use lack sufficient moisture to be physiologically acceptable to the respiratory tract of patients. Heat and moisture exchangers (HME) are used to raise the water content and the temperature of the gas delivered to the respiratory tract of spontaneously and non-spontaneously breathing patients. HMEs are intended to form part of a ventilator breathing system. HMEs are primarily placed between the patient interface (e.g. Tracheal or Tracheostomy Tube) and the anaesthesia or ventilator breathing set. HMEs are also intended for use with tracheostomy tube where the HME is not designed to connected to a anaesthesia or ventilator breathing set.
Breathing system filters (BSF) are used to reduce particulates, including microorganisms, in gases delivered to and exhaled from patients. BSFs are intended to form part of a ventilator breathing system. BSFs may be placed between the patient interface and the anaesthesia or ventilator breathing set or placed between the anaesthesia or ventilator breathing set and the anaesthesia or ventilator machine. BSF are exposed to various levels of humidity during clinical use.
Many devices combine the characteristics of both an HME and a BSF in one combined device, called an HMEF. HMEs and HMEFs are only intended for use between the patient interface and the anaesthesia or ventilator breathing set.
oSIST prEN ISO 9360-1:2026
oSIST prEN ISO 9360-1:2026 DRAFT International Standard ISO/DIS 9360-1:2026(en) Heat and moisture exchangers, and breathing system filters for respired gases — Part 1:
Common requirements
1 Scope
This document specifies the common requirements for heat and moisture exchangers (HME), breathing system filters (BSF) and combined devices (HMEF) intended for anaesthetic and respiratory use. It addresses materials, design, performance, requirements for devices supplied sterile, marking, and information to be provided by the manufacturer.
NOTE 1 Wherever the term “device” is used in this document, it refers to an HME or, HMEF or BSF. This document also applies to HMEs or HMEFs that do not have a machine port for connecting to a ventilator. [1] Requirements for active HMEs are located in ISO 80601-2-74 .
This document is not applicable to other types of filters:
- those designed to protect vacuum sources or gas sample lines;
- to filter compressed gases;
- to protect test equipment for physiological respiratory measurements;
- on inlet of resuscitation bags;
- port of nebulisers.
NOTE 2 A method for assessing moisture loss from an HME/HMEF is given in ISO 9360-2: 202X. NOTE 3 A method for assessing filtration performance of BSF/HMEF is given in ISO 9360-3:202X. Filters used for filtering anaesthetic gases from the respiratory gas are not within scope of this document. [1] NOTE 4 Requirements for Active HMEs are included in ISO 80601-2-74 .
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 5356-1:2015, Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets ISO 5356-1, Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets ISO 18190:2025, Anaesthetic and respiratory equipment — General requirements for airway devices and related equipment ISO 80369-2, Small-bore connectors for liquids and gases in healthcare applications — Part 2: Connectors for respiratory applications
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en) ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications IEC 60601-1:2005+AMD1: 2012+ AMD2: 2020 Medical electrical equipment — Part 1: General requirements for basic safety and essential performance ISO 9360-2: 202X Heat and moisture exchangers, and breathing system filters for respired gases — Part 2:
Measurement of moisture loss ISO 9360-3: 202X Heat and moisture exchangers, and breathing system filters for respired gases — Part 3:
Measurement of filtration performance ISO TS 24172 Alternative test method for heat and moisture exchangers
3 Terms and definitions
[2] For the purposes of this part of ISO 9360, the terms and definitions given in ISO 4135:2022 and the following apply.
3.1 accessory port
port of the device intended for connection to an accessory EXAMPLE Connection to a respiratory gas monitor.
3.2 active HME
device where water, water vapour or heat is actively added to the HME to increase the humidity level of the gas delivered from the HME to the patient [2] [SOURCE: ISO 4135:2022 , 3.7.2.3]
3.3 breathing set
assembly of breathing tubes, connectors and components that form the inspiratory and expiratory limbs of the gas pathways of an anaesthetic breathing system or ventilator breathing system between the ventilator and the patient's airway device [2] [SOURCE: ISO 4135:2022 , 3.6.1.6]
3.4 breathing system filter
BSF device intended to reduce transmission of airborne particulates, including microorganisms, in breathing systems Note 1 to entry: A BSF may be incorporated into an HME.
[2] [SOURCE: ISO 4135:2022 , 3.6.1.5]
3.5 compliance
change in volume of gas per unit pressure change within an enclosed space Note 1 to entry: It is deprecated to include the measurement conditions into the definition of compliance; in general compliance may be expressed at any specified measurement condition. [3] Note 2 to entry: ISO 19223 provides definitions for pulmonary compliance, respiratory system compliance, static compliance and dynamic compliance, all of which relate to the patient. Various other standards from ISO/TC 121 reference compliance with respect to items of equipment such as breathing tubes and reservoir bags. [2] [SOURCE: ISO 4135:2022 , 3.1.1.12]
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en)
3.6 Heat and moisture exchanger, noun
HME, noun device intended to retain a portion of the patient’s expired moisture and heat, and return it to the patient's respiratory tract during inspiration [2] [SOURCE: ISO 4135:2022 , 3.7.4.1]
3.7 HMEF, FHME, noun
heat and moisture exchanger that incorporates the function of a breathing system filter [2] [SOURCE: ISO 4135:2022 , 3.7.4.1, deleted note]
3.8 internal volume, noun
compressible volume contained within the device Note 1 to entry: The internal volume is intended to refer to the compressible volume occupied by breathing gases. Note 2 to entry: The internal volume is bounded by the internal surfaces of the impermeable material of the device, the ports of the device and excludes the internal volume of socket gas connectors. Note 3 to entry: The internal volume excludes the volume of solid elements within the device, such as exchange media.
3.9 machine port
port of the device which is intended for connection to a patient connection port of a breathing set
3.10 moisture loss, noun
total amount of water lost from the test apparatus when tested as specified in ISO 9360-2: 202X Note 1 to entry: It is expressed in milligrams water per litre of air.
3.11 patient connection port
port of a breathing system intended for connection to an airway device Note 1 to entry: The patient-connection port is the end of the breathing system proximal to the patient. Note 2 to entry: The patient-connection port is typically a connector suitable for connection to an airway device such as a tracheal tube, tracheostomy tube, face mask or supra-laryngeal airway. Note 3 to entry: Current product standards typically specify that the patient-connection port is required to be in the form of specific standardized connectors, for example, a connector conforming to ISO 5356-1:2015. Note 4 to entry: In anaesthetic and respiratory equipment where its function is dependent upon a design feature of the component that connects the equipment to the patient's airway, then there is no patient-connection-port. Examples would include non-invasive ventilation using a face mask with an integrated expiratory port, or oxygen therapy using an open mask.
[2] [SOURCE: ISO 4135:2022 , 3.1.4.41]
3.12 pressure drop, noun
difference between the pressure measured in a gas stream flowing into a device and the pressure measured in the gas stream flowing out of the device, with a specified constant gas flowrate through the device [2] [SOURCE: ISO 4135:2022 , 3.1.4.43]
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en)
4 General requirements
4.1 General
The applicable requirements of ISO 18190:2025, Clause 4 shall apply.
4.2 Ambient conditions of test
The ambient conditions during the tests shall be:
- temperature: (23 ± 2) °C;
- relative humidity: (60 ± 15) % RH;
- pressure: (96 ± 10) kPa.
5 Materials
5.1 General
The applicable requirements of ISO 18190:2025, Clause 5 shall apply.
5.2 Housing
NOTE 1 There is guidance or rationale for this subclause contained in Annex A. For devices with a machine port intended for mechanical ventilation:
a) The housing, if coloured shall meet the requirements of 6.2.1. Check conformity by inspection.
b) Liquid water and secretions shall be visible within the device housing. Check conformity by the following test:
- Determine the orientation of the device where water can collect within the housing and occlude the media using the minimum amount of water, see Figure 1 b).
NOTE 2 Some devices have eccentric connectors that do not allow water to collect under certain orientations, see Figure 1 a).
- Determine the volume of water the housing can contain before overflowing from the patient end connector or when the pressure drop exceeds 6cmH O at 30LPM.
- Fill the housing with 50% of the volume of water and place the housing in its worst case horizontal orientation.
...



