prEN ISO 9360-1
(Main)Heat and moisture exchangers, and breathing system filters for respired gases - Part 1: Common requirements (ISO/DIS 9360-1:2026)
- Status
- Not Published
- Publication Date
- 07-Feb-2028
- Technical Committee
- CEN/TC 215 - Respiratory and anaesthetic equipment
- Drafting Committee
- CEN/TC 215/WG 4 - Airway devices, related equipment and suction devices
- Current Stage
- 4060 - Closure of enquiry - Enquiry
- Start Date
- 12-Nov-2026
- Due Date
- 12-Nov-2026
- Completion Date
- 12-Nov-2026
prEN ISO 9360-1 is a CEN draft aligned with ISO/DIS 9360-1:2026. It specifies the common requirements for heat and moisture exchangers (HME), breathing system filters (BSF), and combined devices (HMEF) for anaesthetic and respiratory use, including materials, design, performance, marking, and manufacturer information. It is written for manufacturers, test laboratories, and users who need to select or verify these devices in breathing systems.
What does prEN ISO 9360-1 specify?
prEN ISO 9360-1 defines the common requirements for HME, BSF, and HMEF devices used in anaesthetic and respiratory applications. It covers devices with and without a machine port, and it sets requirements for materials, housing visibility, connector design, performance testing, cleaning and reprocessing information, and user information.
The document also states what is outside its scope, so buyers and engineers can separate these devices from other filters used elsewhere in healthcare or equipment testing.
| Annex | What it covers |
|---|---|
| Annex A | Rationale for selected requirements, including housing visibility and device identification |
| Annex B | Test method for assessing visibility and legibility |
The document is organized into clauses on scope, normative references, terms and definitions, general requirements, materials, design requirements, cleaning/disinfection/sterilization, and information supplied by the manufacturer. It also includes a bibliography and informative annexes.
What are the key requirements of prEN ISO 9360-1?
prEN ISO 9360-1 combines design, performance, and labeling requirements so that HME, BSF, and HMEF devices can be recognized, connected correctly, and assessed consistently. In practice, the standard helps avoid wrong-device use, poor humidification, and connection errors.
General conditions and materials
Clause 4 applies the general requirements of ISO 18190 and sets ambient test conditions. Clause 5 also applies ISO 18190 for materials, then adds a specific housing requirement for devices with a machine port intended for mechanical ventilation: liquid water and secretions shall be visible within the housing. This matters because condensation can affect performance and, in the worst case, expose the patient to liquid water.
Device identification and color coding
Clause 6.2 requires color coding for the housing or marking. BSF devices use one color set, while HME and HMEF devices use another set, and the marking shall remain visible during use. This helps clinical staff distinguish devices quickly at the bedside and reduces the risk of placing the wrong device in the breathing system.
Connection ports and accessory ports
Clause 6.3 requires device connection ports to be conical connectors conforming to ISO 5356-1, with specific port configurations depending on whether the device is nearest the patient or the machine. HMEs and HMEFs without a machine port shall have a 15 mm conical socket on the patient side.
If an accessory port is provided, it shall be compatible with a Luer lock connector, have a tethered sealing cap or equivalent retention, and be marked with the required warning symbol or safety sign. The instructions for use shall also warn about the risk of inadvertent connection to other Luer-based medical devices. This is important because accessory ports are commonly used for gas sampling.
If a supplementary oxygen connection is provided on an HME or HMEF without a machine port for a spontaneously breathing patient, it shall conform to the specified respiratory connector requirement or have a permanently connected oxygen tube.
Performance testing
Clause 6.4 requires testing appropriate to the device type:
- Moisture loss is measured by the method in ISO 9360-2.
- Filtration performance is measured by the method in ISO 9360-3 for BSF and HMEF.
- Resistance to flow is measured at intended-use flowrates for neonate, paediatric, and adult applications, and again after preconditioning for maximum recommended use.
- Gas leakage is measured under a defined pressure test, and the device shall meet a low leakage limit.
- Compliance is measured with a defined pressure test and is also repeated after preconditioning.
These tests matter because they describe how the device will behave in real breathing systems - how much humidity it returns, how much it resists breathing flow, whether it leaks gas, and how much it expands under pressure.
Cleaning, disinfection, sterilization, and manufacturer information
Clause 7 applies the cleaning, disinfection, and sterilization requirements of ISO 18190. Clause 8 applies the manufacturer-information requirements of ISO 18190 and adds device marking, packaging labeling, and instructions for use.
The instructions for use shall include the recommended tidal volume and/or flowrate range, measured resistance to flow, internal volume, gas leakage, compliance, and moisture loss for HME/HMEF. For BSF/HMEF, the penetration value shall also be provided. The standard also requires warnings about using a BSF instead of an HME, use with nebulizers or humidifiers where relevant, and risks from small-bore connector misuse.
What terms does prEN ISO 9360-1 define?
- Heat and moisture exchanger (HME) - A device that retains part of the patient’s expired heat and moisture and returns it during inspiration.
- Breathing system filter (BSF) - A device intended to reduce airborne particulates, including microorganisms, in breathing systems.
- HMEF - A heat and moisture exchanger that also includes the function of a breathing system filter.
- Accessory port - A port intended for connection to an accessory, such as a respiratory gas monitor.
- Machine port - The port intended for connection to the patient connection port of a breathing set.
- Patient connection port - The end of the breathing system intended for connection to an airway device.
- Moisture loss - The amount of water lost during the specified test method.
- Compliance - The change in gas volume per unit pressure change in an enclosed space.
Who uses prEN ISO 9360-1?
prEN ISO 9360-1 is used by manufacturers of anaesthetic and respiratory equipment, including HME, BSF, and HMEF devices. It is also used by test laboratories and quality teams that verify connector fit, performance, labeling, and user information.
Clinicians, procurement teams, and hospital technical staff may use it when choosing devices for ventilator breathing systems or checking whether a device is suitable for the patient-side or machine-side position. It is especially relevant where wrong-device selection or connector mix-ups could affect patient safety.
What changed in prEN ISO 9360-1 from the previous edition?
prEN ISO 9360-1 technically revises and replaces ISO 9360-1:2000, ISO 9360-2:2001, ISO 23328-1:2008, and ISO 23328-2:2009. The document says that Part 1 combines requirements from ISO 9360-1:2000 and ISO 23328-2:2009.
The stated changes include:
- the layout now follows the format of ISO 18190:202x and references clauses from ISO 18190;
- colour coding requirements were added to distinguish between HMEs and BSFs.
Which standards are used with prEN ISO 9360-1?
- ISO 18190:2025 - supplies general requirements for airway devices and related equipment; prEN ISO 9360-1 applies its clauses for general requirements, materials, cleaning, and manufacturer information.
- ISO 5356-1 - defines the conical connectors used for device connection ports.
- ISO 80369-2 - gives the respiratory connector used when a supplementary oxygen connection is provided.
- ISO 80369-7 - defines the Luer lock connector used for accessory ports.
- IEC 60601-1 - is cited for symbol and safety-sign references used in accessory-port marking.
- ISO 9360-2 - provides the moisture-loss test method for HME and HMEF devices.
- ISO 9360-3 - provides the filtration-performance test method for BSF and HMEF devices.
- ISO TS 24172 - gives an alternative test method for assessing HME efficiency.
- ISO 80601-2-74 - covers active HMEs.
- ISO 4135:2022 - provides the vocabulary used for the defined terms.
What does the prEN ISO 9360-1 document contain?
The document contains tables for color coding, connector configurations, test requirements by device type, resistance-to-flow test conditions, and device-identification symbols. It also includes figures for housing transparency and resistance-to-flow test setup, which help laboratories and manufacturers set up the correct checks.
Clause 5.2 gives a practical housing visibility test, Clause 6.4 gives test methods for moisture loss, filtration performance, resistance to flow, gas leakage, and compliance, and Annex B gives a legibility test based on user viewpoint and lighting. Annex A explains why the color, visibility, and labeling requirements were chosen, which helps users understand the safety intent behind the requirements.
Relations
- Effective Date
- 19-Feb-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
prEN ISO 9360-1 is a draft published by the European Committee for Standardization (CEN). Its full title is "Heat and moisture exchangers, and breathing system filters for respired gases - Part 1: Common requirements (ISO/DIS 9360-1:2026)". This standard covers: La présente partie de l'ISO 9360 spécifie certaines exigences relatives aux échangeurs de chaleur et d'humidité (ECH), y compris ceux équipés de filtres pour systèmes respiratoires, essentiellement destinés à humidifier les gaz respirés par les patients ayant un volume courant supérieur ou égal à 250 ml, et comportant au moins un orifice côté appareil, et décrit des méthodes d'essai permettant de les évaluer.
La présente partie de l'ISO 9360 spécifie certaines exigences relatives aux échangeurs de chaleur et d'humidité (ECH), y compris ceux équipés de filtres pour systèmes respiratoires, essentiellement destinés à humidifier les gaz respirés par les patients ayant un volume courant supérieur ou égal à 250 ml, et comportant au moins un orifice côté appareil, et décrit des méthodes d'essai permettant de les évaluer.
prEN ISO 9360-1 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
prEN ISO 9360-1 has the following relationships with other standards: It is inter standard links to EN ISO 9360-1:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
prEN ISO 9360-1 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 accessory port
- 3.2 active HME
- 3.3 breathing set
- 3.4 breathing system filter
- 3.5 compliance
- 3.6 Heat and moisture exchanger, noun
- 3.7 HMEF, FHME, noun
- 3.8 internal volume, noun
- 3.9 machine port
- 3.10 moisture loss, noun
- 3.11 patient connection port
- 3.12 pressure drop, noun
- 4 General requirements
- 4.1 General
- 4.2 Ambient conditions of test
- 5 Materials
- 5.1 General
- 5.2 Housing
- 6 Design Requirements
- 6.1 General
- 6.2 Identification of the device
- 6.3 Device connections
- 6.4 Performance Requirements
- 4 HME/HMEF/BSF under test
- 7 Cleaning, disinfection and sterilization
- 8 Information to be supplied by the manufacturer
- 8.1 Marking on the device
- 8.2 Labelling on the packaging
- 8.3 Instructions for use
- Annex A
- Annex B
- Bibliography
Introduction
The gases generally available for medical use lack sufficient moisture to be physiologically acceptable to the respiratory tract of patients. Heat and moisture exchangers (HME) are used to raise the water content and the temperature of the gas delivered to the respiratory tract of spontaneously and non-spontaneously breathing patients. HMEs are intended to form part of a ventilator breathing system. HMEs are primarily placed between the patient interface (e.g. Tracheal or Tracheostomy Tube) and the anaesthesia or ventilator breathing set. HMEs are also intended for use with tracheostomy tube where the HME is not designed to connected to a anaesthesia or ventilator breathing set.
Breathing system filters (BSF) are used to reduce particulates, including microorganisms, in gases delivered to and exhaled from patients. BSFs are intended to form part of a ventilator breathing system. BSFs may be placed between the patient interface and the anaesthesia or ventilator breathing set or placed between the anaesthesia or ventilator breathing set and the anaesthesia or ventilator machine. BSF are exposed to various levels of humidity during clinical use.
Many devices combine the characteristics of both an HME and a BSF in one combined device, called an HMEF. HMEs and HMEFs are only intended for use between the patient interface and the anaesthesia or ventilator breathing set.
oSIST prEN ISO 9360-1:2026
oSIST prEN ISO 9360-1:2026 DRAFT International Standard ISO/DIS 9360-1:2026(en) Heat and moisture exchangers, and breathing system filters for respired gases — Part 1:
Common requirements
1 Scope
This document specifies the common requirements for heat and moisture exchangers (HME), breathing system filters (BSF) and combined devices (HMEF) intended for anaesthetic and respiratory use. It addresses materials, design, performance, requirements for devices supplied sterile, marking, and information to be provided by the manufacturer.
NOTE 1 Wherever the term “device” is used in this document, it refers to an HME or, HMEF or BSF. This document also applies to HMEs or HMEFs that do not have a machine port for connecting to a ventilator. [1] Requirements for active HMEs are located in ISO 80601-2-74 .
This document is not applicable to other types of filters:
- those designed to protect vacuum sources or gas sample lines;
- to filter compressed gases;
- to protect test equipment for physiological respiratory measurements;
- on inlet of resuscitation bags;
- port of nebulisers.
NOTE 2 A method for assessing moisture loss from an HME/HMEF is given in ISO 9360-2: 202X. NOTE 3 A method for assessing filtration performance of BSF/HMEF is given in ISO 9360-3:202X. Filters used for filtering anaesthetic gases from the respiratory gas are not within scope of this document. [1] NOTE 4 Requirements for Active HMEs are included in ISO 80601-2-74 .
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 5356-1:2015, Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets ISO 5356-1, Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets ISO 18190:2025, Anaesthetic and respiratory equipment — General requirements for airway devices and related equipment ISO 80369-2, Small-bore connectors for liquids and gases in healthcare applications — Part 2: Connectors for respiratory applications
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en) ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications IEC 60601-1:2005+AMD1: 2012+ AMD2: 2020 Medical electrical equipment — Part 1: General requirements for basic safety and essential performance ISO 9360-2: 202X Heat and moisture exchangers, and breathing system filters for respired gases — Part 2:
Measurement of moisture loss ISO 9360-3: 202X Heat and moisture exchangers, and breathing system filters for respired gases — Part 3:
Measurement of filtration performance ISO TS 24172 Alternative test method for heat and moisture exchangers
3 Terms and definitions
[2] For the purposes of this part of ISO 9360, the terms and definitions given in ISO 4135:2022 and the following apply.
3.1 accessory port
port of the device intended for connection to an accessory EXAMPLE Connection to a respiratory gas monitor.
3.2 active HME
device where water, water vapour or heat is actively added to the HME to increase the humidity level of the gas delivered from the HME to the patient [2] [SOURCE: ISO 4135:2022 , 3.7.2.3]
3.3 breathing set
assembly of breathing tubes, connectors and components that form the inspiratory and expiratory limbs of the gas pathways of an anaesthetic breathing system or ventilator breathing system between the ventilator and the patient's airway device [2] [SOURCE: ISO 4135:2022 , 3.6.1.6]
3.4 breathing system filter
BSF device intended to reduce transmission of airborne particulates, including microorganisms, in breathing systems Note 1 to entry: A BSF may be incorporated into an HME.
[2] [SOURCE: ISO 4135:2022 , 3.6.1.5]
3.5 compliance
change in volume of gas per unit pressure change within an enclosed space Note 1 to entry: It is deprecated to include the measurement conditions into the definition of compliance; in general compliance may be expressed at any specified measurement condition. [3] Note 2 to entry: ISO 19223 provides definitions for pulmonary compliance, respiratory system compliance, static compliance and dynamic compliance, all of which relate to the patient. Various other standards from ISO/TC 121 reference compliance with respect to items of equipment such as breathing tubes and reservoir bags. [2] [SOURCE: ISO 4135:2022 , 3.1.1.12]
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en)
3.6 Heat and moisture exchanger, noun
HME, noun device intended to retain a portion of the patient’s expired moisture and heat, and return it to the patient's respiratory tract during inspiration [2] [SOURCE: ISO 4135:2022 , 3.7.4.1]
3.7 HMEF, FHME, noun
heat and moisture exchanger that incorporates the function of a breathing system filter [2] [SOURCE: ISO 4135:2022 , 3.7.4.1, deleted note]
3.8 internal volume, noun
compressible volume contained within the device Note 1 to entry: The internal volume is intended to refer to the compressible volume occupied by breathing gases. Note 2 to entry: The internal volume is bounded by the internal surfaces of the impermeable material of the device, the ports of the device and excludes the internal volume of socket gas connectors. Note 3 to entry: The internal volume excludes the volume of solid elements within the device, such as exchange media.
3.9 machine port
port of the device which is intended for connection to a patient connection port of a breathing set
3.10 moisture loss, noun
total amount of water lost from the test apparatus when tested as specified in ISO 9360-2: 202X Note 1 to entry: It is expressed in milligrams water per litre of air.
3.11 patient connection port
port of a breathing system intended for connection to an airway device Note 1 to entry: The patient-connection port is the end of the breathing system proximal to the patient. Note 2 to entry: The patient-connection port is typically a connector suitable for connection to an airway device such as a tracheal tube, tracheostomy tube, face mask or supra-laryngeal airway. Note 3 to entry: Current product standards typically specify that the patient-connection port is required to be in the form of specific standardized connectors, for example, a connector conforming to ISO 5356-1:2015. Note 4 to entry: In anaesthetic and respiratory equipment where its function is dependent upon a design feature of the component that connects the equipment to the patient's airway, then there is no patient-connection-port. Examples would include non-invasive ventilation using a face mask with an integrated expiratory port, or oxygen therapy using an open mask.
[2] [SOURCE: ISO 4135:2022 , 3.1.4.41]
3.12 pressure drop, noun
difference between the pressure measured in a gas stream flowing into a device and the pressure measured in the gas stream flowing out of the device, with a specified constant gas flowrate through the device [2] [SOURCE: ISO 4135:2022 , 3.1.4.43]
oSIST prEN ISO 9360-1:2026 ISO/DIS 9360-1:2026(en)
4 General requirements
4.1 General
The applicable requirements of ISO 18190:2025, Clause 4 shall apply.
4.2 Ambient conditions of test
The ambient conditions during the tests shall be:
- temperature: (23 ± 2) °C;
- relative humidity: (60 ± 15) % RH;
- pressure: (96 ± 10) kPa.
5 Materials
5.1 General
The applicable requirements of ISO 18190:2025, Clause 5 shall apply.
5.2 Housing
NOTE 1 There is guidance or rationale for this subclause contained in Annex A. For devices with a machine port intended for mechanical ventilation:
a) The housing, if coloured shall meet the requirements of 6.2.1. Check conformity by inspection.
b) Liquid water and secretions shall be visible within the device housing. Check conformity by the following test:
- Determine the orientation of the device where water can collect within the housing and occlude the media using the minimum amount of water, see Figure 1 b).
NOTE 2 Some devices have eccentric connectors that do not allow water to collect under certain orientations, see Figure 1 a).
- Determine the volume of water the housing can contain before overflowing from the patient end connector or when the pressure drop exceeds 6cmH O at 30LPM.
- Fill the housing with 50% of the volume of water and place the housing in its worst case horizontal orientation.
...