Packaging for terminally sterilized medical devices - Part 9: Uncoated nonwoven materials of polyolefines - Requirements and test methods

This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices.
Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.

Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 9: Unbeschichtete Faservliesmaterialien aus Polyolefinen - Anforderungen und Prüfverfahren

Dieses Dokument legt Prüfverfahren und Werte für unbeschichtete Faservliesmaterialien aus Polyolefinen fest, die in Sterilbarrieresystemen zum einmaligen Gebrauch und/oder Verpackungssystemen für in der Endverpackung sterilisierte Medizinprodukte verwendet werden.
Über die allgemeinen Anforderungen hinaus, welche durch EN ISO 11607 1 und EN ISO 11607 2 festgelegt sind, legt dieser Teil der Normenreihe EN 868 Materialien, Prüfverfahren und Werte fest, die spezifisch für die in diesem Dokument betrachteten Produkte sind.

Emballages des dispositifs médicaux stérilisés au stade terminal - Partie 9 : Matériaux non tissés à base de polyoléfines, non enduits - Exigences et méthodes d'essai

Le présent document spécifie des méthodes d’essai et des valeurs relatives aux matériaux non tissés à base de polyoléfines, non enduits, destinés à un usage unique dans des systèmes de barrière stérile et/ou des systèmes d’emballage pour les dispositifs médicaux stérilisés au stade terminal.
Outre les exigences générales telles que spécifiées dans l’EN ISO 11607-1 et l’EN ISO 11607-2, la présente partie de la série EN 868 spécifie les matériaux, les méthodes d’essai et les valeurs spécifiques aux produits traités dans le présent document.

Embalaža za končno sterilizirane medicinske pripomočke - 9. del: Površinsko neobdelani netkani materiali iz poliolefinov - Zahteve in preskusne metode

General Information

Status
Not Published
Public Enquiry End Date
30-Mar-2026
Technical Committee
VAZ - Healthcare
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
23-Jan-2026
Due Date
12-Jun-2026

Relations

Effective Date
13-Nov-2024

Overview

oSIST prEN 868-9:2026: Packaging for Terminally Sterilized Medical Devices - Part 9: Uncoated Nonwoven Materials of Polyolefines – Requirements and Test Methods is a draft European Standard developed by CEN/TC 102. This document defines performance requirements and test methods for uncoated, nonwoven polyolefin materials used in sterile barrier systems and packaging for single-use, terminally sterilized medical devices. It specifies criteria beyond EN ISO 11607-1 and EN ISO 11607-2, focusing on material-specific requirements relevant to polyolefins-such as polyethylene and polypropylene-used in healthcare sterilization processes.

Key Topics

  • Material Specifications: Details the required physical and chemical properties for uncoated, nonwoven polyolefin materials including purity, opacity, and lack of leachable substances that could pose health risks.
  • Performance Testing: Outlines test methods and minimum performance values for characteristics like tensile strength, tearing resistance, bursting strength, air permeance, and water penetration resistance.
  • Sterilization Compatibility: Ensures that materials can maintain sterility post-sterilization and withstand the effects of various sterilization processes.
  • Labelling and Information: Specifies durable labelling requirements to ensure traceability and user information, including manufacturer details, product reference, and instructions for storage or handling.
  • Environmental Aspects: Addresses eco-design, recyclability, environmental labelling (per EN ISO 14021 and EN ISO 14025), and recommendations for minimizing environmental impact throughout the product lifecycle.

Applications

Uncoated nonwoven polyolefin materials as defined in oSIST prEN 868-9:2026 are primarily used for:

  • Sterile Barrier Systems: Providing packaging layers that maintain the sterility of medical devices up to the point of patient use.
  • Healthcare Facility Sterilization: Suitable for various sterilization methods commonly used in hospitals, including steam, ethylene oxide, or irradiation, provided compatibility is validated.
  • Industry Packaging: Applicable in manufacturing and industrial packaging processes to ensure finished medical devices remain sterile during storage, transportation, and handling.
  • Risk Management: Materials conforming to this standard support compliance with healthcare regulations around device safety, sterility assurance, and patient health protection.

Related Standards

oSIST prEN 868-9:2026 belongs to a broader series of packaging standards for terminally sterilized medical devices. Key related standards include:

  • EN ISO 11607-1: General requirements for packaging systems intended to maintain the sterility of terminally sterilized medical devices.
  • EN ISO 11607-2: Validation requirements for forming, sealing, and assembly processes.
  • EN 868 Series: Covers additional packaging material standards:
    • Part 2: Sterilization wrap
    • Part 3: Paper for bag and pouch manufacture
    • Part 4: Paper bags
    • Part 5: Sealable pouches and reels
    • Part 6: Low temperature process paper
    • Part 7: Adhesive coated paper for low temperature sterilization
    • Part 8: Re-usable sterilization containers
    • Part 10: Adhesive coated nonwoven polyolefins

Practical Value

  • Consistent Quality: Ensuring polyolefin packaging materials meet industry-accepted benchmarks for strength and durability protects the sterility of medical devices during their entire lifecycle.
  • Regulatory Compliance: Conformity with oSIST prEN 868-9:2026 and related standards helps organizations demonstrate compliance with EU and international requirements, supporting market access and patient safety.
  • Environmental Considerations: The standard promotes environmentally responsible choices in material selection, use, and disposal-encouraging recyclable products and reduced material use without compromising sterility or safety.
  • Streamlined Procurement: Clear, harmonized requirements simplify sourcing decisions and supplier audits for manufacturers and healthcare facilities.

Adopting oSIST prEN 868-9:2026 ensures that packaging for terminally sterilized medical devices performs as needed in demanding healthcare environments, while also addressing sustainability and regulatory obligations.

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

oSIST prEN 868-9:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Packaging for terminally sterilized medical devices - Part 9: Uncoated nonwoven materials of polyolefines - Requirements and test methods". This standard covers: This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.

This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.

oSIST prEN 868-9:2026 is classified under the following ICS (International Classification for Standards) categories: 11.080.30 - Sterilized packaging. The ICS classification helps identify the subject area and facilitates finding related standards.

oSIST prEN 868-9:2026 has the following relationships with other standards: It is inter standard links to SIST EN 868-9:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

oSIST prEN 868-9:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-marec-2026
Embalaža za končno sterilizirane medicinske pripomočke - 9. del: Površinsko
neobdelani netkani materiali iz poliolefinov - Zahteve in preskusne metode
Packaging for terminally sterilized medical devices - Part 9: Uncoated nonwoven
materials of polyolefines - Requirements and test methods
Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 9:
Unbeschichtete Faservliesmaterialien aus Polyolefinen - Anforderungen und
Prüfverfahren
Emballages des dispositifs médicaux stérilisés au stade terminal - Partie 9 : Matériaux
non tissés à base de polyoléfines, non enduits - Exigences et méthodes d'essai
Ta slovenski standard je istoveten z: prEN 868-9
ICS:
11.080.30 Sterilizirana embalaža Sterilized packaging
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

DRAFT
EUROPEAN STANDARD
NORME EUROPÉENNE
EUROPÄISCHE NORM
January 2026
ICS 11.080.30 Will supersede EN 868-9:2018
English Version
Packaging for terminally sterilized medical devices - Part
9: Uncoated nonwoven materials of polyolefines -
Requirements and test methods
Emballages des dispositifs médicaux stérilisés au stade Verpackungen für in der Endverpackung zu
terminal - Partie 9 : Matériaux non tissés à base de sterilisierende Medizinprodukte - Teil 9:
polyoléfines, non enduits - Exigences et méthodes Unbeschichtete Faservliesmaterialien aus Polyolefinen
d'essai - Anforderungen und Prüfverfahren
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 102.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.

EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 868-9:2026 E
worldwide for CEN national Members.

Contents Page
European foreword . 3
Introduction . 4
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 5
4 General requirements . 6
4.1 General . 6
4.2 Materials . 6
5 Performance requirements and test methods . 6
5.1 General . 6
5.2 Specific performance requirements for polyethylene based materials. 7
5.3 Specific perform ace requirements for polypropylene based materials . 7
6 Sterilization compatibility . 7
7 Labelling of the protective packaging . 7
8 Information to be provided . 8
8.1 Regulatory and technical information . 8
8.2 Environmental declarations . 8
Annex A (informative) Environmental aspects . 9
Bibliography . 12
European foreword
This document (prEN 868-9:2026) has been prepared by Technical Committee CEN/TC 102 “Sterilizers
and associated equipment for processing of medical devices”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This document will supersede EN 868-9:2018.
a) Clause 4 “General requirements” was slightly revised for clarity and aligned with the other parts of
the EN 868 series, and a statement was added clarifying when acceptance criteria apply.
b) Clause 8 “Environmental declarations” was added and aligned with the other parts of EN 868 series.
c) The list of major changes was moved to the foreword, thus the Annex with “Details of significant
technical changes between this European Standard and the previous edition” (former Annex A) was
deleted.
d) New Annex A regarding environmental aspects was added.
EN 868 series consists of the following parts, under the general title Packaging for terminally sterilized
medical devices:
— Part 2: Sterilization wrap — Requirements and test methods;
— Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture
of pouches and reels (specified in EN 868-5) — Requirements and test methods;
— Part 4: Paper bags — Requirements and test methods;
— Part 5: Sealable pouches and reels constructed of porous materials and plastic film— Requirements and
test methods;
— Part 6: Paper for low temperature sterilization processes — Requirements and test methods;
— Part 7: Adhesive coated paper for low temperature sterilization processes — Requirements and test
methods;
— Part 8: Re-usable sterilization containers for large steam sterilizers — Requirements and test methods;
— Part 9: Uncoated nonwoven materials of polyolefins — Requirements and test methods;
— Part 10: Adhesive coated nonwoven materials of polyolefins — Requirements and test methods.
Introduction
The EN ISO 11607 series of standards consists of two parts under the general title “Packaging for
terminally sterilized medical devices”. EN ISO 11607-1 specifies general requirements and test methods
for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are
intended to maintain sterility of terminally sterilized medical devices to the point of use. EN ISO 11607-2
specifies validation requirements for forming, sealing and assembly processes.
General requirements for all types of sterile barrier systems are provided by EN ISO 11607-1 and
EN ISO 11607-2 [2].
The EN 868 series of standards have been developed mainly for materials and sterile barrier systems
used in health care facilities sterilization processes. Materials complying with part 9 of the EN 868 series
can also be used for industrial sterilization. The EN 868 series of standards can be used to demonstrate
compliance with one or more of the requirements specified in EN ISO 11607-1.
Considering CEN guide 4 [4] and the CEN environmental checklists, this revision has been complemented
with a new annex with guidance to encourage users to also include environmental aspects when applying
the EN 868 series of standards with the objective to minimize the environmental impact. Environmental
aspects have also been included into the description of test methods with the same objective.
1 Scope
This document specifies test methods and values for uncoated nonwoven materials of polyolefins used
for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices.
Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of
EN 868 series specifies materials, test methods and values that are specific to the products covered by
this document.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN ISO 536, Paper and board - Determination of grammage (ISO 536)
EN ISO 811, Textiles - Determination of resistance to water penetration - Hydrostatic pressure test (ISO 811)
EN ISO 1924-2, Paper and board - Determination of tensile properties - Part 2: Constant rate of elongation
method (20 mm/min) (ISO 1924-2)
EN ISO 1974, Paper - Determination of tearing resistance - Elmendorf method (ISO 1974)
EN ISO 2758, Paper - Determination of bursting strength (ISO 2758)
EN ISO 11607-1:2020, Packaging for terminally sterilized medical devices - Part 1: Requirements for
materials, sterile barrier systems and packaging systems (ISO 11607-1:2019)
EN ISO 14021, Environmental labels and declarations - Self-declared environmental claims (Type II
environmental labelling) (ISO 14021)
EN ISO 14025, Environmental labels and declarations - Type III environmental declarations - Principles and
procedures (ISO 14025)
ISO 5636-3, Paper and board — Determination of air permeance (medium range) — Part 3: Bendtsen
method
ISO 6588-2:2021, Paper, board and pulps — Determination of pH of aqueous extracts — Part 2: Hot
extraction
ISO 8601-1, Date and time — Representations for information interchange — Part 1: Basic rules
ASTM D2724, Standard Test Methods for Bonded, Fused, and Laminated Apparel Fabrics
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2020 apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:

As impacted by EN ISO 11607-1:2020/A1:2023 and EN ISO 11607-1:2020/A11:2022.
— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at https://www.iso.org/obp
4 General requirements
4.1 General
4.1.1 For any material, preformed sterile barrier system or sterile barrier system, the requirements of
EN ISO 11607-1 shall apply.
4.1.2 Polyethylene and polypropylene non-wovens have different specific performance requirements
and this shall be considered during design and use of sterile barrier systems for specific applications.
NOTE When additional materials are used inside the sterile barrier system in order to ease the organization,
drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer sets, tray liners
or an additional envelope around the medical device) then other requirements, including the determination of the
acceptability of these materials during validation activities, can apply.
4.1.3 This part of EN 868 series only introduces performance requirements and test methods that are
specific to the products covered by this part of EN 868 series but does not add or modify the general
requirements specified in EN ISO 11607-1.
4.1.4 As such, the particular requirements in Clause 5 can be used to demonstrate compliance with one
or more but not all of the requirements of EN ISO 11607-1.
NOTE Conformity with this document does not automatically mean conformity with EN ISO 11607-1.
4.1.5 A confirmation of compliance to this document shall contain a statem
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...