kSIST FprEN 868-9:2026
(Main)Packaging for terminally sterilized medical devices - Part 9: Uncoated nonwoven materials of polyolefines - Requirements and test methods
- Abstract
This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices.
Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.
- Status
- Not Published
- Public Enquiry End Date
- 30-Mar-2026
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 5020 - Formal vote (FV) (Adopted Project)
- Start Date
- 06-Oct-2026
- Due Date
- 24-Nov-2026
- Completion Date
- 06-Oct-2026
kSIST FprEN 868-9:2026 is the SIST version of the draft European Standard prEN 868-9:2026. It covers packaging for terminally sterilized medical devices and specifies test methods and product-specific performance values for uncoated nonwoven materials of polyolefins used in single-use sterile barrier systems and packaging systems. The document is for manufacturers, test laboratories, and users who need to qualify, specify, or verify this material for sterile packaging.
What does kSIST FprEN 868-9:2026 specify?
kSIST FprEN 868-9:2026 specifies the material-specific requirements for uncoated nonwoven polyolefin materials, while the general requirements come from EN ISO 11607-1 and EN ISO 11607-2. It is organized into clauses on scope, normative references, terms and definitions, general requirements, performance requirements and test methods, sterilization compatibility, labelling, and information to be provided, plus one informative annex on environmental aspects.
The document focuses on materials used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. It also notes that these materials can be used in health care facility sterilization processes and, in some cases, in industrial sterilization.
What are the key requirements of kSIST FprEN 868-9:2026?
kSIST FprEN 868-9:2026 sets out the product checks that matter for selecting, testing, and documenting uncoated nonwoven polyolefins. In practice, the user has to show that the material performs as declared, remains suitable after sterilization, and is labelled and documented correctly.
General material and test conditions
Clause 4 requires the general EN ISO 11607-1 requirements to be met and notes that polyethylene and polypropylene non-wovens have different specific performance requirements. The material has to be translucent or opaque, made of continuous polyolefin filaments of high purity, and not release substances in amounts that could create a health risk.
Clause 4 also sets the testing basis for conditioned samples and says that the standard only adds the product-specific requirements for this material family. For users, that means this part cannot be read alone as a full sterile packaging specification.
Performance checks in Clause 5
Clause 5 covers the core performance tests. It checks that colour does not leach from the material, that the declared mass per unit area is met within the allowed tolerance, and that tear resistance is sufficient in both machine and cross directions.
Clause 5 also covers delamination, air permeance, and water penetration. If delamination can be initiated, the material must resist layer separation; air permeance is required for the material as conditioned; and water penetration is checked with a hydrostatic head method, with the test result and conditions documented.
Polymer-specific strength requirements
Clause 5 separates polyethylene-based and polypropylene-based materials because their strength behaviour is different. Polyethylene-based materials have their own tensile and bursting strength requirements, and polypropylene-based materials have a different set of tensile and bursting strength requirements.
For buyers and designers, this means the polymer type is not just a material description - it changes the acceptance criteria and should be matched to the sterilization process and package design.
Sterilization, labelling, and information
Clause 6 requires the sterilization processes to be defined and the compatibility of the packaging material to be evaluated using the EN ISO 11607-1 approach, considering the intended process or processes and number of cycles. Clause 7 requires durable, legible labelling, and for deliveries to health care facilities it names the key label data to provide, including identification, quantity, manufacturer or supplier details, date of manufacture, lot number, nominal mass, size or roll dimensions, and storage conditions where relevant.
Clause 8 covers information to be provided with the material. It includes regulatory and technical information for sterile barrier system use, environmental labels and declarations, end-of-life recommendations, and, where recyclability is assessed, the method used, the product reference, the standards and regulations applied, and the results.
What terms does kSIST FprEN 868-9:2026 define?
Clause 3 uses the terms and definitions from EN ISO 11607-1:2020. Key terms used in the document include:
- sterile barrier system (SBS) - the packaging barrier intended to maintain sterility until point of use.
- packaging system - the wider packaging arrangement used with the sterile barrier system.
- protective packaging - the outer packaging that has to be legibly and durably labelled.
- machine direction - the main direction of the material used for tensile and tear testing.
- cross direction - the direction across the material width used for tensile and tear testing.
- delamination strength - resistance to layers separating when delamination can be initiated.
- environmental declaration - a claim or declaration made under EN ISO 14021 or EN ISO 14025.
Who uses kSIST FprEN 868-9:2026?
kSIST FprEN 868-9:2026 is used by manufacturers of sterile packaging materials, designers of sterile barrier systems, testing laboratories, and quality or regulatory teams. It is also relevant to buyers and suppliers who need to agree material specifications, label content, and documentation for delivery to health care facilities or industrial users.
What changed in kSIST FprEN 868-9:2026 from the previous edition?
kSIST FprEN 868-9:2026 will supersede EN 868-9:2018. The foreword states that Clause 4 was revised for clarity and alignment with the EN 868 series, Clause 8 on environmental declarations was added, the former annex with significant technical changes was deleted, and a new Annex A on environmental aspects was added.
Which standards are used with kSIST FprEN 868-9:2026?
| Standard | What it contributes |
|---|---|
| EN ISO 11607-1 | General requirements for materials, sterile barrier systems, and packaging systems |
| EN ISO 11607-2 | Validation requirements for forming, sealing, and assembly processes |
| EN ISO 536 | Grammage determination |
| EN ISO 811 | Water penetration resistance by hydrostatic pressure |
| EN ISO 1924-2 | Tensile properties |
| EN ISO 1974 | Tearing resistance |
| EN ISO 2758 | Bursting strength |
| EN ISO 14021 | Self-declared environmental claims |
| EN ISO 14025 | Type III environmental declarations |
| ISO 5636-3 | Air permeance |
| ISO 6588-2 | Hot extraction method used for the colour-leach check |
| ISO 8601-1 | Date format for labelling |
| ASTM D2724 | Delamination strength test |
The bibliography also points to EN ISO 10993-1, ISO 22095, and CEN Guide 4.
What does the kSIST FprEN 868-9:2026 document contain?
kSIST FprEN 868-9:2026 brings together test methods for colour leaching, mass per area, tear resistance, delamination, air permeance, water penetration, tensile strength, and bursting strength. It also sets out the sterilization compatibility approach, the required package labelling data, and the environmental information that should accompany the material.
Annex A gives guidance on environmental aspects across the lifecycle, including material use, production, transport packaging, end-of-life treatment, and transportation. The bibliography links the document to related guidance on biocompatibility, chain of custody, and environmental issues.
Relations
- Effective Date
- 13-Nov-2024
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
kSIST FprEN 868-9:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Packaging for terminally sterilized medical devices - Part 9: Uncoated nonwoven materials of polyolefines - Requirements and test methods". This standard covers: This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.
This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.
kSIST FprEN 868-9:2026 is classified under the following ICS (International Classification for Standards) categories: 11.080.30 - Sterilized packaging. The ICS classification helps identify the subject area and facilitates finding related standards.
kSIST FprEN 868-9:2026 has the following relationships with other standards: It is inter standard links to SIST EN 868-9:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
kSIST FprEN 868-9:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 General requirements
- 4.1 General
- 4.2 Materials
- 5 Performance requirements and test methods
- 5.1 General
- 5.2 Specific performance requirements for polyethylene based materials
- 5.3 Specific perform ace requirements for polypropylene based materials
- 6 Sterilization compatibility
- 7 Labelling of the protective packaging
- 8 Information to be provided
- 8.1 Regulatory and technical information
- 8.2 Environmental declarations
- Annex A
- Bibliography
Introduction
The EN ISO 11607 series of standards consists of two parts under the general title “Packaging for terminally sterilized medical devices”. EN ISO 11607-1 specifies general requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. EN ISO 11607-2 specifies validation requirements for forming, sealing and assembly processes. General requirements for all types of sterile barrier systems are provided by EN ISO 11607-1 and EN ISO 11607-2 [2].
The EN 868 series of standards have been developed mainly for materials and sterile barrier systems used in health care facilities sterilization processes. Materials complying with part 9 of the EN 868 series can also be used for industrial sterilization. The EN 868 series of standards can be used to demonstrate compliance with one or more of the requirements specified in EN ISO 11607-1. Considering CEN guide 4 [4] and the CEN environmental checklists, this revision has been complemented with a new annex with guidance to encourage users to also include environmental aspects when applying the EN 868 series of standards with the objective to minimize the environmental impact. Environmental aspects have also been included into the description of test methods with the same objective.
oSIST prEN 868-9:2026 prEN 868-9:2026 (E)
1 Scope
This document specifies test methods and values for uncoated nonwoven materials of polyolefins used for single-use sterile barrier systems and/or packaging systems for terminally sterilized medical devices. Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2, this part of EN 868 series specifies materials, test methods and values that are specific to the products covered by this document.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN ISO 536, Paper and board - Determination of grammage (ISO 536) EN ISO 811, Textiles - Determination of resistance to water penetration - Hydrostatic pressure test (ISO 811) EN ISO 1924-2, Paper and board - Determination of tensile properties - Part 2: Constant rate of elongation method (20 mm/min) (ISO 1924-2) EN ISO 1974, Paper - Determination of tearing resistance - Elmendorf method (ISO 1974) EN ISO 2758, Paper - Determination of bursting strength (ISO 2758)
EN ISO 11607-1:2020, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems (ISO 11607-1:2019) EN ISO 14021, Environmental labels and declarations - Self-declared environmental claims (Type II environmental labelling) (ISO 14021) EN ISO 14025, Environmental labels and declarations - Type III environmental declarations - Principles and procedures (ISO 14025) ISO 5636-3, Paper and board — Determination of air permeance (medium range) — Part 3: Bendtsen method ISO 6588-2:2021, Paper, board and pulps — Determination of pH of aqueous extracts — Part 2: Hot extraction ISO 8601-1, Date and time — Representations for information interchange — Part 1: Basic rules ASTM D2724, Standard Test Methods for Bonded, Fused, and Laminated Apparel Fabrics
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2020 apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
As impacted by EN ISO 11607-1:2020/A1:2023 and EN ISO 11607-1:2020/A11:2022.
oSIST prEN 868-9:2026 prEN 868-9:2026 (E) - IEC Electropedia: available at http://www.electropedia.org/ - ISO Online browsing platform: available at https://www.iso.org/obp
4 General requirements
4.1 General
4.1.1 For any material, preformed sterile barrier system or sterile barrier system, the requirements of EN ISO 11607-1 shall apply.
4.1.2 Polyethylene and polypropylene non-wovens have different specific performance requirements
and this shall be considered during design and use of sterile barrier systems for specific applications. NOTE When additional materials are used inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) then other requirements, including the determination of the acceptability of these materials during validation activities, can apply. 4.1.3 This part of EN 868 series only introduces performance requirements and test methods that are specific to the products covered by this part of EN 868 series but does not add or modify the general requirements specified in EN ISO 11607-1.
4.1.4 As such, the particular requirements in Clause 5 can be used to demonstrate compliance with one or more but not all of the requirements of EN ISO 11607-1.
NOTE Conformity with this document does not automatically mean conformity with EN ISO 11607-1.
4.1.5 A confirmation of compliance to this document shall contain a statement of whether
EN ISO 11607-1 is covered.
4.2 Materials
4.2.1 The uncoated material shall be translucent or opaque and made of continuous filaments of
polyolefins of a high level of purity and shall not release any substances in such quantities as could constitute a health risk.
NOTE Attention is drawn to EN ISO 10993-1.
4.2.2 Unless otherwise specified in the test methods, test samples shall be conditioned at (23 ± 1) °C and (50 ± 2) % relative humidity for a minimum of 24 h. The conditioning time may be reduced for polyolefin materials if demonstrated as appropriate.
5 Performance requirements and test methods
5.1 General
5.1.1 No colour shall leach out of the material. Compliance shall be tested by visual examination of a hot extract prepared in accordance with the method given in ISO 6588-2:2021 modified to test temperature of (60 ± 5) °C.
5.1.2 The average mass of 1 m of the conditioned material when tested in accordance with EN ISO 536 shall be within ± 7 % of the nominal value stated by the manufacturer.
...