SIST EN 62494-1:2009
(Main)Medical electrical equipment - Exposure index of digital X-ray imaging systems - Part 1: Definitions and requirements for general radiography (IEC 62494-1:2008 )
Medical electrical equipment - Exposure index of digital X-ray imaging systems - Part 1: Definitions and requirements for general radiography (IEC 62494-1:2008 )
IEC 62494-1:2008 specifies definitions and requirements for the exposure index of images acquired with digital X-ray imaging systems. IEC 62494-1:2008 is applicable to digital X-ray imaging systems used in general radiography for producing projection X-ray images for general applications, such as, but not exclusively: - computed radiography (CR) systems based on stimulable phosphors; - flat-panel detector based systems; - charge-coupled device (CCD) based systems. Image intensifier based systems and systems for mammographic or dental application are not covered in this first edition. IEC 62494-1:2008 defines the exposure index only for images generated with a single irradiation event. Images generated from multiple irradiations (e.g., tomosynthetic or dual-energy images, multiple views on a single CR plate) are not covered.
Medizinische elektrische Geräte - Dosisindikator digitaler Röntgenbildsysteme - Teil 1: Definitionen und Anforderungen für die allgemeine Radiographie (IEC 62494-1:2008 )
Appareils électro-médicaux - Indice d'exposition des systèmes d’imagerie numérique à rayonnement X - Partie 1: Définitions et exigences pour la radiographie générale (CEI 62494-1:2008 )
La CEI 62494-1:2008 spécifie les définitions et les exigences pour l'indice d'exposition des images acquises avec des systèmes d'imagerie numérique à rayonnement X. La CEI 62494-1:2008 est applicable aux systèmes d'imagerie numérique à rayonnement X utilisés en radiographie générale pour produire des images radiologiques de projection pour les applications générales, comme par exemple, mais pas exclusivement: - les systèmes de radiographie assistée par ordinateur (CR, computed radiography) basés sur des luminophores stimulables; - les systèmes basés sur des détecteurs à écran plat; - les systèmes basés sur un dispositif à transfert de charge (CCD, charge-coupled device). Les systèmes basés sur un intensificateur d'image et les systèmes utilisés en mammographie ou pour les applications dentaires ne sont pas traités dans cette première édition. La CEI 62494-1:2008 définit l'indice d'exposition uniquement pour les images produites avec un seul événement d'irradiation. Les images produites à partir d'irradiations multiples (par exemple, images tomosynthétiques ou biénergétiques, vues multiples sur une seule plaque de radiographie assistée par ordinateur) ne sont pas traitées.
Medicinska električna oprema - Indeks izpostavljenosti pri digitalnih sistemih za rentgensko slikanje - 1. del: Definicije in zahteve za splošno radiografijo (IEC 62494-1:2008 )
General Information
- Status
- Published
- Publication Date
- 03-Dec-2008
- Technical Committee
- IEMO - Electrical equipment in medical practice
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 28-Nov-2008
- Due Date
- 02-Feb-2009
- Completion Date
- 04-Dec-2008
Overview
EN 62494-1:2008 (identical to IEC 62494-1:2008) defines the concept, terminology and general requirements for the Exposure Index used with digital X‑ray imaging systems in general radiography. The standard standardizes how an exposure indicator (a numeric value derived from original image data) is defined, calibrated and reported so operators can assess whether detector exposure was appropriate for the intended image quality. It applies to systems such as computed radiography (CR), flat‑panel detectors, and CCD‑based detectors, but excludes image‑intensifier systems and dedicated mammography or dental systems. Multi‑irradiation images (e.g., tomosynthesis, dual‑energy or multiple views on one CR plate) are also excluded.
Key topics and requirements
- Definitions: formal terms for EXPOSURE INDEX, DEVIATION INDEX, ORIGINAL DATA, RELEVANT IMAGE REGION and related concepts.
- Requirements for creation of original data and selection of the relevant image region from which the exposure index is derived.
- Exposure Index (EI): requirements for scale, meaning and storage (e.g., embedding in image metadata such as DICOM).
- Deviation Index (DI): a standardized way to express deviation of measured EI from a TARGET EXPOSURE INDEX to indicate under‑ or over‑exposure relative to task‑specific targets.
- Calibration: requirements and procedures for calibration of the exposure index, including determination of calibration and inverse calibration functions and beam conditions for calibration.
- Scope limitations: EI is defined for single‑irradiation projection images only; the standard intentionally excludes specific software algorithms for calculating EI to allow technical innovation.
- Non‑substitution for patient dose: the EI refers to detector exposure (signal) and is not a direct patient dose metric; it should not be used to determine compliance with diagnostic reference levels.
Practical applications
- Establishing consistent exposure indicators across vendors within a radiology department to simplify QA and exposure optimization.
- Enabling radiographers and technologists to quickly check if exposures are appropriate for intended image quality using the EI and DI.
- Supporting medical physicists and QA teams in routine monitoring, calibration and documentation of detector exposure levels.
- Facilitating integration of EI/DI into PACS and DICOM workflows for clinical documentation and trending.
Who should use this standard
- Medical device manufacturers (CR, flat‑panel, CCD vendors)
- Hospital radiology departments, medical physicists and QA personnel
- Regulators and accreditation bodies involved in imaging quality and radiation safety
- PACS/DICOM software developers and system integrators
Related standards
- IEC 62494-1:2008 (international edition - same content)
- Normative/referenced documents noted in the standard: IEC/TR 60788 (glossary); harmonized standards referenced in the endorsement (e.g., EN/IEC 60601-1, EN/IEC 62220-1, EN/IEC 62220-1-2).
Frequently Asked Questions
SIST EN 62494-1:2009 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Medical electrical equipment - Exposure index of digital X-ray imaging systems - Part 1: Definitions and requirements for general radiography (IEC 62494-1:2008 )". This standard covers: IEC 62494-1:2008 specifies definitions and requirements for the exposure index of images acquired with digital X-ray imaging systems. IEC 62494-1:2008 is applicable to digital X-ray imaging systems used in general radiography for producing projection X-ray images for general applications, such as, but not exclusively: - computed radiography (CR) systems based on stimulable phosphors; - flat-panel detector based systems; - charge-coupled device (CCD) based systems. Image intensifier based systems and systems for mammographic or dental application are not covered in this first edition. IEC 62494-1:2008 defines the exposure index only for images generated with a single irradiation event. Images generated from multiple irradiations (e.g., tomosynthetic or dual-energy images, multiple views on a single CR plate) are not covered.
IEC 62494-1:2008 specifies definitions and requirements for the exposure index of images acquired with digital X-ray imaging systems. IEC 62494-1:2008 is applicable to digital X-ray imaging systems used in general radiography for producing projection X-ray images for general applications, such as, but not exclusively: - computed radiography (CR) systems based on stimulable phosphors; - flat-panel detector based systems; - charge-coupled device (CCD) based systems. Image intensifier based systems and systems for mammographic or dental application are not covered in this first edition. IEC 62494-1:2008 defines the exposure index only for images generated with a single irradiation event. Images generated from multiple irradiations (e.g., tomosynthetic or dual-energy images, multiple views on a single CR plate) are not covered.
SIST EN 62494-1:2009 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN 62494-1:2009 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-januar-2009
0HGLFLQVNDHOHNWULþQDRSUHPD,QGHNVL]SRVWDYOMHQRVWLSULGLJLWDOQLKVLVWHPLK]D
UHQWJHQVNRVOLNDQMHGHO'HILQLFLMHLQ]DKWHYH]DVSORãQRUDGLRJUDILMR,(&
Medical electrical equipment - Exposure index of digital X-ray imaging systems - Part 1:
Definitions and requirements for general radiography (IEC 62494-1:2008 )
Medizinische elektrische Geräte - Dosisindikator digitaler Röntgenbildsysteme - Teil 1:
Definitionen und Anforderungen für die allgemeine Radiographie (IEC 62494-1:2008 )
Appareils électro-médicaux - Indice d'exposition des systèmes d’imagerie numérique à
rayonnement X - Partie 1: Définitions et exigences pour la radiographie générale (CEI
62494-1:2008 )
Ta slovenski standard je istoveten z: EN 62494-1:2008
ICS:
11.040.50 Radiografska oprema Radiographic equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN 62494-1
NORME EUROPÉENNE
November 2008
EUROPÄISCHE NORM
ICS 11.040.50
English version
Medical electrical equipment -
Exposure index of digital X-ray imaging systems -
Part 1: Definitions and requirements for general radiography
(IEC 62494-1:2008)
Appareils électromédicaux - Medizinische elektrische Geräte -
Indice d'exposition des systèmes Dosisindikator digitaler
d’imagerie numérique à rayonnement X - Röntgenbildsysteme -
Partie 1: Définitions et exigences Teil 1: Definitionen und Anforderungen
pour la radiographie générale für die allgemeine Radiographie
(CEI 62494-1:2008) (IEC 62494-1:2008)
This European Standard was approved by CENELEC on 2008-10-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.
CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B - 1050 Brussels
© 2008 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 62494-1:2008 E
Foreword
The text of document 62B/680/CDV, future edition 1 of IEC 62494-1, prepared by SC 62B, Diagnostic
imaging equipment, of IEC TC 62, Electrical equipment in medical practice, was submitted to the
IEC-CENELEC parallel vote and was approved by CENELEC as EN 62494-1 on 2008-10-01.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2009-07-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2011-10-01
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, notes, general statements, exceptions and references: in smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THIS STANDARD, IN EN 60601-1 OR IN IEC/TR 60788, AS REFERENCED IN
THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 62494-1:2008 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
IEC 60601-1 NOTE Harmonized as EN 60601-1:2006 (not modified).
IEC 60601-2-43 NOTE Harmonized as EN 60601-2-43:2000 (not modified).
IEC 62220-1 NOTE Harmonized as EN 62220-1:2004 (not modified).
IEC 62220-1-2 NOTE Harmonized as EN 62220-1-2:2007 (not modified).
__________
– 3 – EN 62494-1:2008
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
NOTE When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Publication Year Title EN/HD Year
IEC/TR 60788 2004 Medical electrical equipment - Glossary of - -
defined terms
IEC 62494-1
Edition 1.0 2008-08
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment – Exposure index of digital X-ray imaging
systems –
Part 1: Definitions and requirements for general radiography
Appareils électromédicaux – Indice d'exposition des systèmes d’imagerie
numérique à rayonnement X –
Partie 1: Définitions et exigences pour la radiographie générale
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
R
CODE PRIX
ICS 11.040.50 ISBN 2-8318-9944-3
– 2 – 62494-1 © IEC:2008
CONTENTS
FOREWORD.3
INTRODUCTION.5
1 Scope.7
2 Normative references .7
3 Terms and definitions .7
4 Requirements .9
4.1 Creation of ORIGINAL DATA.9
4.2 Determination of the RELEVANT IMAGE REGION and the VALUE OF INTEREST .9
4.3 Requirements for the EXPOSURE INDEX .10
4.4 Calibration of the EXPOSURE INDEX .10
4.5 Determination of the CALIBRATION FUNCTION .11
4.6 Determination of the INVERSE CALIBRATION FUNCTION .11
4.7 Requirements for the DEVIATION INDEX.11
Annex A (informative) Details on the rationale, properties and use of the EXPOSURE
INDEX .13
Annex B (informative) Details on the rationale, properties and use of the DEVIATION
INDEX .17
Annex C (normative) Beam conditions to be used for calibration.18
Bibliography.19
Terminology – Index of defined terms .20
Figure A.1 – Example of an ORIGINAL DATA radiograph with an example of the RELEVANT
IMAGE REGION outlined .14
Figure A.2 – Histogram of the ORIGINAL DATA for the radiograph shown in Figure A.1 .15
Figure A.3 – Relative IMAGE RECEPTOR AIR KERMA required to produce a fixed detector
response for the four x-ray beam qualities defined in ISO 9236-1 .16
62494-1 © IEC:2008 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
EXPOSURE INDEX OF DIGITAL X-RAY IMAGING SYSTEMS –
Part 1: Definitions and requirements for general radiography
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62494-1 has been prepared by subcommittee 62B: Diagnostic
imaging equipment, of IEC technical committee 62: Electrical equipment in medical practice.
The text of this standard is based on the following documents:
Enquiry draft Report on voting
62B/680/CDV 62B/703/RVC
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
– 4 – 62494-1 © IEC:2008
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, notes, general statements, exceptions and references: in smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THIS STANDARD, IN IEC 60601-1 OR IN IEC 60788, AS
REFERENCED IN THE INDEX OF DEFINED TERMS: SMALL CAPITALS.
A list of all parts of the IEC 62494 series, published under the general title Medical electrical
equipment – Exposure index of digital X-ray imaging systems, can be found on the IEC
website.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
62494-1 © IEC:2008 – 5 –
INTRODUCTION
The direct connection between the level of detector exposure and optical density is well
established in film-screen radiology. This is not the case in digital radiography, where almost
always a constant image characteristic is achieved using automatic image processing.
Consequently, deviations from the intended exposure, i.e., over- and underexposure, are not
noticeable by a corresponding deviation in image brightness. While considerable
underexposure results in an increased level of noise, the more alarming aspect (from a
radiation protection point of view) is that overexposure cannot be recognized easily in the
displayed image.
Therefore, various manufacturers of digital radiography systems have introduced so-called
exposure indicators for their equipment. These are numbers, determined from the original
image data of each image taken, which allow conclusions about the level of the exposure at
the image receptor. However, the exposure indicators are manufacturer or system specific,
i.e. they differ for the systems of different manufacturers in their definition and scaling. A
unified EXPOSURE INDEX for all digital radiography systems is needed to simplify its usage, e.g.
for the establishment of exposure guidelines, particularly when systems of different
manufacturers are used within the same department.
This standard defines such a concept of the EXPOSURE INDEX. What is laid down here refers to
the definition, the scale and the general requirements for the EXPOSURE INDEX. The process of
its calculation in detail (software algorithm) is excluded from this standard as to not obstruct
technical progress.
The EXPOSURE INDEX allows the OPERATOR to judge if an image was taken at a detector
exposure level suitable for the intended level of image quality. It is important to note that the
EXPOSURE INDEX, as defined in this standard, is derived from the image signal, which in turn is
usually related to the energy absorbed in the detector, i.e. the detector dose, but not directly
to the air kerma at the image receptor. The relation to IMAGE RECEPTOR AIR KERMA (air kerma
at the detector surface) is introduced only at one radiation quality through calibration.
However, this definition is appropriate as the image quality in digital radiography is
determined mainly by the signal-to-noise level, which in turn is determined by the absorbed
energy. Annex A provides more details on the rationale, properties and use of the EXPOSURE
INDEX.
The level of detector exposure needed to obtain a suitable level of image quality may vary
depending on body part, view, or the x-ray imaging system used, as may the appropriate
EXPOSURE INDEX. This standard introduces a second parameter, called DEVIATION INDEX, which
quantifies the deviation of an actual EXPOSURE INDEX from the appropriate EXPOSURE INDEX
(called TARGET EXPOSURE INDEX). While this parameter does not relate to the image receptor
dose on an absolute scale, it allows the operator an easy check whether the exposure is
considered acceptable for the specific imaging task. Annex B provides more details on the
rationale, properties and use of the DEVIATION INDEX.
The storage of the EXPOSURE INDEX (and the DEVIATION INDEX) together with the image data,
e.g., in a DICOM tag field, allows the documentation and communication of the image
receptor dose level in clinical practice.
The EXPOSURE INDEX does not obviate the use of dose parameters that describe the patient’s
REFERENCE AIR KERMA or the kerma-area
exposure to radiation, such as, for example, the
product. Because the relation between patient exposure and detector exposure is influenced
by a number of factors that are generally not known under clinical conditions, the EXPOSURE
INDEX should not be used to calculate or estimate patient dose.
– 6 – 62494-1 © IEC:2008
The EXPOSURE INDEX cannot be used to control the compliance with diagnostic reference
1)
levels, which refer to patient dose [1] .
___________
1)
Figures in square brackets refer to the Bibliography.
62494-1 © IEC:2008 – 7 –
MEDICAL ELECTRICAL EQUIPMENT –
EXPOSURE INDEX OF DIGITAL X-RAY IMAGING SYSTEMS –
Part 1: Definitions and requirements for general radiography
1 Scope
This part of IEC 62494 specifies definitions and requirements for the EXPOSURE INDEX of
images acquired with DIGITAL X-RAY IMAGING SYSTEMS.
This standard is applicable to DIGITAL X-RAY IMAGING SYSTEMS used in general radiography for
producing PROJECTION X-ray images for general applications, such as, but not exclusively:
• computed radiography (CR) systems based on stimulable phosphors;
• flat-panel detector based systems;
• charge-coupled device (CCD) based systems.
Image intensifier based systems and systems for mammographic or dental application are not
covered in this first edition.
This standard defines the EXPOSURE INDEX only for images generated with a single
IRRADIATION event. Images generated from multiple IRRADIATIONS (e.g., tomosynthetic or dual-
energy images, multiple views on a single CR plate) are not covered.
2 Normative references
The following referenced documents are indispensable for the application of this document.
For dated references, only the edition cited applies. For undated references, the latest edition
of the referenced document (including any amendments) applies.
IEC/TR 60788:2004, Medical electrical equipment – Glossary of defined terms
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC TR 60788 and the
following apply.
3.1
CALIBRATION CONDITIONS
set of conditions under which EXPOSURE INDEX calibration is done
3.2
CALIBRATION FUNCTION
function expressing the VALUE OF INTEREST as a function of the IMAGE RECEPTOR AIR KERMA that
is valid under CALIBRATION CONDITIONS
3.3
DETECTOR SURFACE
accessible area which is closest to the IMAGE RECEPTOR PLANE
NOTE After removal of all parts (including the ANTI-SCATTER GRID and components for AUTOMATIC EXPOSURE
CONTROL, if applicable) that can be safely removed from the RADIATION BEAM without damaging the digital X-ray
detector.
[IEC 62220-1-2:2007, definition 3.3]
– 8 – 62494-1 © IEC:2008
3.4
DEVIATION INDEX
DI
number quantifying the deviation of the actual EXPOSURE INDEX from a TARGET EXPOSURE INDEX
3.5
DIGITAL X-RAY IMAGING DEVICE
device consisting of a digital X-ray detector including the protective layers installed for use in
practice, the amplifying and digitizing electronics, and a computer providing the ORIGINAL DATA
(DN) of the image
[IEC 62220-1:2003, definition 3.5]
NOTE This may include protecting parts, such as anti-scatter grids or AEC components
3.6
DIGITAL X-RAY IMAGING SYSTEM
X-ray equipment using a DIGITAL X-RAY IMAGING DEVICE, providing PROJECTION images in digital
format, comprising subsystems allowing to process, display, print or store the images
3.7
EXPOSURE INDEX
EI
measure of the detector response to radiation in the RELEVANT IMAGE REGION of an image
acquired with a DIGITAL X-RAY IMAGING SYSTEM
NOTE For a fixed RADIATION QUALITY, the signal generated in the detector is proportional to the IMAGE RECEPTOR
AIR KERMA (or exposure).
3.8
IMAGE RECEPTOR AIR KERMA
K
AIR KERMA at the position of the DETECTOR SURFACE, free-in-air (excluding backscatter)
3.9
INVERSE CALIBRATION FUNCTION
function expressing the IMAGE RECEPTOR AIR KERMA as a function of the VALUE OF INTEREST that
CALIBRATION CONDITIONS
is valid under
3.10
ORIGINAL DATA
DN
RAW DATA to which the corrections allowed in this standard have been applied
[IEC 62220-1:2003, definition 3.12]
NOTE The relation of the ORIGINAL DATA to the IMAGE RECEPTOR AIR KERMA may include a non-linear, e.g.,
logarithmic or square-root characteristic.
3.11
RAW DATA
pixel values read directly after the analogue-digital-conversion from the DIGITAL X-RAY IMAGING
DEVICE without any software corrections
[IEC 62220-1:2003, definition 3.14]]
3.12
RELEVANT IMAGE REGION
examination-specific sub-area or sub-areas of the image containing the diagnostically relevant
information
NOTE This is typically the region for which the exposure parameters should be optimized.
62494-1 © IEC:2008 – 9 –
3.13
TARGET EXPOSURE INDEX
EI
T
expected value of the EXPOSURE INDEX when exposing the X-RAY IMAGE RECEPTOR properly
NOTE The TARGET EXPOSURE INDEX may depend on the type of detector, on the type of examination, on the
diagnostic question and on other parameters.
3.14
VALUE OF INTEREST
V
central tendency of the original data in the relevant image region
NOTE Central tendency is a statistical term depicting generally the centre of a distribution. It may refer to a
variety of measures such as the mean, the median or the mode.
4 Requirements
4.1 Creation of ORIGINAL DATA
The following image-independent corrections of the RAW DATA are allowed for the creation of
ORIGINAL DATA in advance of the processing of the data for the determination of the
CALIBRATION FUNCTION and the EXPOSURE INDEX.
All the following corrections if used shall be made as in normal clinical use:
– replacement of the RAW DATA of bad or defective pixels by appropriate data;
– a flat-field correction comprising for example:
• correction of the non-uniformity of the RADIATION FIELD;
• correction for the offset o
...










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...