Aerosol drug delivery device design verification - Requirements and test methods (ISO 20072:2009)

This International Standard applies to the design, labelling, instructions for use and testing requirements for hand-held single- and multi-use aerosol drug delivery devices (ADDDs) intended to deliver a metered or premetered aerosolized medication to or by means of the human respiratory tract (including nasal, oral, tracheal, bronchial and alveolar sites). This International Standard applies to both refillable and disposable devices intended for personal use. This International Standard is intended for device design verification and not for drug product quality assessment. The objective of this International Standard is to verify, by laboratory (in-vitro) testing, that the ADDD design consistently meets the manufacturer's design specification by satisfying a device functionality profile and system verification test both of which are determined from a risk assessment and evaluated in accordance with the instructions for use. This International Standard excludes continuous or semi-continuous aerosolization devices covered in ISO 27427, aerosolization devices which do not emit active pharmaceutical ingredient (API), general purpose aerosolization devices (for use with ventilators) and atomizers. This International Standard does not apply to manufacturers of single parts or components of the ADDDs [e.g. (spray) pumps, valves, containers, etc.].

Ausführungsverifizierung von Inhalationsgeräten - Anforderungen und Prüfverfahren (ISO 20072:2009)

Die vorliegende Internationale Norm gilt für die Anforderungen an Ausführung, Kennzeichnung, Gebrauchs-anweisung und Prüfung von Hand gehaltenen Inhalationsgeräten (ADDD) für den Einmal  und den Mehrfachgebrauch, die dazu dienen, ein dosiertes oder vordosiertes Arzneimittel in aerosolierter Form in den Atemtrakt (einschließlich Nase, Mund, Luftröhre, Bronchien und Alveolen) eines Menschen oder über diesen zu verabreichen. Diese Internationale Norm gilt sowohl für nachfüllbare als auch Einmalgeräte für den Eigengebrauch.
Diese Internationale Norm ist für die Ausführungsverifizierung der Geräte, nicht jedoch die Qualitäts-beurteilung von Arzneimittelprodukten vorgesehen. Das Ziel dieser Internationalen Norm besteht darin, durch Labor (In vitro )prüfung zu verifizieren, dass die Ausführung des Inhalationsgeräts dauerhaft die Ausführungs-spezifikation des Herstellers erfüllt, indem sie sowohl ein Geräte Funktionsprofil als auch eine System-verifizierungsprüfung besteht, die beide auf einer Risikobeurteilung beruhen und in Übereinstimmung mit der Gebrauchsanweisung bewertet werden.
Die vorliegende Internationale Norm schließt Folgendes aus: Die von ISO 27427 abgedeckten kontinuierlich und halbkontinuierlich arbeitenden Aerosolvernebler sowie Aerosolvernebler, die keine arzneilich wirksamen Bestandteile (API, en: active pharmaceutical ingredients) abgeben, Universal Vernebler (für die Anwendung mit Gebläsen) und Zerstäuber.
Diese Internationale Norm gilt nicht für die Hersteller von Einzelteilen der Inhalationsgeräte [z. B. (Sprüh )Pumpen, Ventile, Behälter usw.].
ANMERKUNG   In einigen Fällen fällt ein Gerät möglicherweise in den Anwendungsbereich sowohl dieser Internationalen Norm als auch von ISO 27427. Der Ausschuss geht davon aus, dass der Hersteller auf der Grundlage der bestimmungsgemäßen Anwendung des Produkts und der Risikobeurteilung des Gerätes ableiten kann, welche der Internationalen Normen für die Ausführungsverifizierung des Inhalationsgeräts anzuwenden ist. In dieser Internationalen Norm wird der Prozess beschrieben, nach dem die Ausführungsverifizierung des Inhalationsgeräts durchzuführen ist, verbunden mit einem risikobasierten Geräte Funktionsprofil des Inhalationsgeräts, bei dem entweder das Arzneimittel, ein Placebo oder ein repräsentatives Arzneimittel zum Einsatz kommt. In ISO 27427 wird der Prozess beschrieben, nach dem die Charakterisierung der aerodynamischen Aerosolleistung eines Verneblersystems zur Anwendung mit einer unspezifischen Klasse arzneilich wirksamer Bestandteile durchgeführt wird.

Vérification de la conception d'un dispositif d'administration de médicament sous forme d'aérosol - Exigences et méthodes d'essai (ISO 20072:2009)

L'ISO 20072:2009 s'applique aux exigences de conception, d'étiquetage, de notice d'utilisation et d'essai relatives aux dispositifs d'administration de médicament sous forme d'aérosol ou aérosols-doseurs (ADDD) utilisés une ou plusieurs fois et tenus à la main, destinés à l'administration de médicaments sous forme d'aérosols dosés ou prédosés au système respiratoire humain ou par son intermédiaire (incluant les voies nasale, orale, trachéale, bronchiale et alvéolaire). La présente Norme internationale s'applique aux dispositifs non réutilisables et rechargeables destinés à un usage personnel.
L'ISO 20072:2009 est destinée à la vérification de la conception du dispositif et non pas à l'évaluation de la qualité du médicament. L'objectif de la présente Norme internationale est de vérifier, par des essais (in vitro) de laboratoire, que la conception d'un ADDD satisfait invariablement aux spécifications de conception du fabricant en se conformant au profil de performance et à l'essai de vérification du système, lesquels sont déterminés à partir d'une appréciation du risque et évalués conformément à la notice d'utilisation.

Inhalatorji za namensko uporabo v medicini - Zahteve in preskusne metode (ISO 20072:2009)

Ta mednarodni standard se uporablja za zasnovo, označevanje, navodila za uporabo in preskusne zahteve za ročne inhalatorje (ADDD) za enkratno ali večkratno uporabo, ki so namenjeni za dostavo odmerjenega ali predhodno odmerjenega zdravila v obliki aerosola v dihalni trakt človeka ali čez dihalni trakt človeka (vključno z nazalnimi, oralnimi, trahealnimi, bronhialnimi in alveolarnimi mesti). Ta mednarodni standard se uporablja za pripomočke za osebno uporabo, ki jih je mogoče ponovno napolniti, in za pripomočke, ki se po uporabi zavržejo. Ta mednarodni standard je namenjen za preverjanje zasnove pripomočka in ne za oceno kakovosti zdravila. Cilj tega mednarodnega standarda je z laboratorijskimi (in-vitro) preskusi preveriti, da je zasnova inhalatorjev dosledno v skladu s specifikacijo proizvajalca,tako da upošteva profil funkcionalnosti pripomočka in preskus preverjanja sistema, pri čemer se oboje določi na podlagi ocene tveganja in oceni v skladu z navodili za uporabo. Ta mednarodni standard izključuje kontinuirane ali polkontinuirane pripomočke za aerosolizacijo, ki so zajeti v standardu ISO 27427, pripomočke za aerosolizacijo, ki ne oddajajo aktivne farmacevtske sestavine (API), pripomočke za aerosolizacijo za splošne namene (za uporabo z respiratornimi pripomočki) in razprševalce. Ta mednarodni standard se ne uporablja za proizvajalce posameznih delov ali sestavnih delov inhalatorjev [npr. (pršilnih) črpalk, valjev, posod itd.].

General Information

Status
Published
Publication Date
15-Apr-2013
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
05-Mar-2013
Due Date
10-May-2013
Completion Date
16-Apr-2013

Relations

Buy Standard

Standard
EN ISO 20072:2013
English language
53 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN ISO 20072:2013
01-maj-2013
1DGRPHãþD
SIST EN ISO 20072:2011
Inhalatorji za namensko uporabo v medicini - Zahteve in preskusne metode (ISO
20072:2009)
Aerosol drug delivery device design verification - Requirements and test methods (ISO
20072:2009)
Ausführungsverifizierung von Inhalationsgeräten - Anforderungen und Prüfverfahren
(ISO 20072:2009)
Vérification de la conception d'un dispositif d'administration de médicament sous forme
d'aérosol - Exigences et méthodes d'essai (ISO 20072:2009)
Ta slovenski standard je istoveten z: EN ISO 20072:2013
ICS:
11.040.10 Anestezijska, respiratorna in Anaesthetic, respiratory and
reanimacijska oprema reanimation equipment
SIST EN ISO 20072:2013 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN ISO 20072:2013

---------------------- Page: 2 ----------------------

SIST EN ISO 20072:2013


EUROPEAN STANDARD
EN ISO 20072

NORME EUROPÉENNE

EUROPÄISCHE NORM
February 2013
ICS 11.040.10 Supersedes EN ISO 20072:2010
English Version
Aerosol drug delivery device design verification - Requirements
and test methods (ISO 20072:2009)
Vérification de la conception d'un dispositif d'administration Ausführungsverifizierung von Inhalationsgeräten -
de médicament sous forme d'aérosol - Exigences et Anforderungen und Prüfverfahren (ISO 20072:2009)
méthodes d'essai (ISO 20072:2009)
This European Standard was approved by CEN on 8 January 2013.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same
status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United
Kingdom.





EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2013 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 20072:2013: E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------

SIST EN ISO 20072:2013
EN ISO 20072:2013 (E)
Contents
Page
Foreword . 3
Annex ZA (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on Medical Devices . 4


2

---------------------- Page: 4 ----------------------

SIST EN ISO 20072:2013
EN ISO 20072:2013 (E)
Foreword
The text of ISO 20072:2009 has been prepared by Technical Committee ISO/TC 84 “Devices for
administration of medicinal products and intravascular catheters” of the International Organization for
Standardization (ISO) and has been taken over as EN ISO 20072:2013 by Technical Committee CEN/TC 205
“Non-active medical devices” the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an identical
text or by endorsement, at the latest by August 2013, and conflicting national standards shall be withdrawn at
the latest by August 2013.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN ISO 20072:2010.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directive.
For relationship with EU Directive, see informative Annex ZA, which is an integral part of this document.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech
Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.
Endorsement notice
The text of ISO 20072:2009 has been approved by CEN as EN ISO 20072:2013 without any modification.
3

---------------------- Page: 5 ----------------------

SIST EN ISO 20072:2013
EN ISO 20072:2013 (E)
Annex ZA
(informative)

Relationship between this European Standard and the Essential Requirements of
EU Directive 93/42/EEC on Medical Devices
This European Standard has been prepared under a mandate given to CEN by the European Commission
and the European Free Trade Association to provide a means of conforming to Essential Requirements of the
New Approach Directive 93/42/EEC on Medical Devices.
Once this standard is cited in the Official Journal of the European Union under that Directive and has been
implemented as a national standard in at least one Member State, compliance with the clauses of this
standard given in Table ZA.1 confers, within the limits of the scope of this standard, a presumption of
conformity with the corresponding Essential Requirements of that Directive and associated EFTA regulations.
Table ZA.1 — Correspondence between this European Standard and Directive 93/42/EEC on
Medical Devices
Clause(s)/subclause(s) of this Essential Requirements (ERs) Qualifying remarks/Notes
EN of Directive 93/42/EEC
5.1, parts h, i, j, l and 5.2 7
5.1, parts h, i, j, l and 5.2 8
5.1 part d and 5.6 9
5.1, 5.5, 6.4.2, 6.4.3, 6.4.4, 8.2 10
5.1, parts k, m, n and 5.6.8 12
8 (all parts) 13 The parts of ER 13.3 a) relating
to the address of manufacturer
and to the authorized
representative are not
addressed.
ERs 13.3 f) and 13.6 h) relating
to single-use are not addressed.
ER 13.6 q) is not addressed.
WARNING — Other requirements and other EU Directives may be applicable to the product(s) falling
within the scope of this standard.
4

---------------------- Page: 6 ----------------------

SIST EN ISO 20072:2013

INTERNATIONAL ISO
STANDARD 20072
First edition
2009-08-01

Aerosol drug delivery device design
verification — Requirements and test
methods
Vérification de la conception d'un dispositif d'administration de
médicament sous forme d'aérosol — Exigences et méthodes d'essai




Reference number
ISO 20072:2009(E)
©
ISO 2009

---------------------- Page: 7 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


COPYRIGHT PROTECTED DOCUMENT


©  ISO 2009
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2009 – All rights reserved

---------------------- Page: 8 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope .1
2 Normative references .1
3 Terms and definitions .2
4 Symbols and abbreviated terms .6
5 Requirements.7
5.1 General.7
5.2 Risk assessment requirements.8
5.3 Device functionality profile.8
5.4 System verification test .8
5.5 Uncertainty of measurements and conformance with specification.9
5.6 Test requirements.9
6 Test methods.11
6.1 General.11
6.2 Test procedures.12
6.3 Test conditions .16
6.4 Test evaluations.17
7 Test report .19
8 Information supplied by the manufacturer .19
8.1 General.19
8.2 Marking .19
8.3 Instructions for use .20
Annex A (informative) Rationale for requirements .22
Annex B (informative) Further guidance and clarification of the device functionality profile .24
Annex C (informative) Rationale for test methods.26
Annex D (informative) Two-sided tolerance limit factors (k).29
Annex E (informative) Alternative acceptance criteria for the device functionality profile
evaluation .35
Bibliography .42

© ISO 2009 – All rights reserved iii

---------------------- Page: 9 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 20072 was prepared by Technical Committee ISO/TC 84, Devices for administration of medicinal
products and intravascular catheters.
iv © ISO 2009 – All rights reserved

---------------------- Page: 10 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
Introduction
This International Standard applies to hand-held aerosol drug delivery devices (ADDD) intended to administer
medication to humans. To avoid unnecessarily restricting innovation, given the broad variation in device
designs, this International Standard addresses the more general design/labelling requirements rather than
specific physical and prescriptive design requirements. However, this International Standard does require the
elaboration of a device functionality profile (DFP) specific to the ADDD in question. This International Standard
also addresses ADDD design requirements from both the user interface and safety perspectives.
An ADDD is used as part of a system consisting of the ADDD, the container, the medication and the labelling,
including the instructions for use. Therefore, design verification of the ADDD includes a final system
verification test conducted in accordance with the instructions for use.
From a regulatory perspective, the ADDD system may be reviewed and approved as part of a drug product
(combination of ADDD and medication) or as a device by itself. For the purposes of this International
Standard, such regulatory distinctions do not alter the intent of the design verification process described
herein. As an example, in the European Union (EU), if an ADDD is placed on the market in such a way that
the ADDD and the medication form a single integral product (i.e. the system) that is intended exclusively for
use in the given combination and which is not refillable, that single product shall be governed by Directive
2001/83/EEC. However, the relevant essential requirements of Annex I of the Medical Device Directive
(93/42/EEC) shall apply as far as safety and performance-related ADDD features are concerned, which is the
specific objective of this design verification standard.
Regardless of the distinctions (“drug” or “device,” pre-filled or refillable), it is recognised that ADDD design
verification is an important component of the overall validation process. Moreover, design verification is
iterative, to be conducted at various phases throughout the ADDD’s development and subsequent ADDD
post-approval modifications. In all instances, design verification is conducted using the phase-appropriate
instructions for use. It is understood that in the early phases of ADDD development an appropriate subset of
the requirements contained herein might apply, but that all of the requirements will be satisfied as part of the
final design verification exercise. Furthermore, design verification should be considered a minimum
requirement for the safe and effective use of the ADDD, and in many instances additional testing may be
appropriate as indicated by a risk assessment that shall also be conducted.
This International Standard introduces the requirement for developers and/or manufacturers to create a device
functionality profile (DFP) for a given ADDD based on the ISO Standard for device risk assessment (as a part
of ISO 14971). The device functionality profile defines the parameters and tolerance intervals used to verify
the ADDD's ability to meet the manufacturer's design specifications during in-use conditions and following
environmental and electromechanical extreme use conditions. This International Standard also includes a
system verification test conducted at standard atmosphere and nominal flow rate as a simple bridge between
the device design and the patient interface.
The purpose of this International Standard is to ensure a method and guide for independent testing of the
repeatability and reproducibility of ADDD functionality that verifies compliance with its design specification.
The design verification process may include use of applicable regulatory agency requirements and/or test
methods. The sampling plans for this International Standard are intended to verify the design at a high
confidence level. They do not replace the more general manufacturing quality systems, including lot release,
which appear in standards on quality systems (e.g. the ISO 9000 series or ISO 13485).
Figure 1 illustrates the process this International Standard advises to use in order to assess and verify
whether a design meets the determined DFP.
© ISO 2009 – All rights reserved v

---------------------- Page: 11 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)

Figure 1 — ADDD design verification process
This International Standard specifically addresses the most basic elements regarding the safe and effective
use of ADDD in humans. It does not define the pharmaceutical or clinical performance of an ADDD. Any
labelling indicating ADDD use to deliver medication to specific regions of the respiratory tract falls under the
authority of national governments or regional agencies regulating the manufacture and marketing of medical
devices and pharmaceutical products. In some countries national regulations exist, and their requirements can
supersede or complement this International Standard.
For a given manufacturer, existing marketed products and those currently under development might not fulfil
some of the requirements. However, manufacturers should comply with this International Standard when
improving the functional design of existing ADDDs or developing new ADDDs to obtain an even higher level of
quality.
Annex A describes the reasoning for establishing the various requirements in this International Standard.
vi © ISO 2009 – All rights reserved

---------------------- Page: 12 ----------------------

SIST EN ISO 20072:2013
INTERNATIONAL STANDARD ISO 20072:2009(E)

Aerosol drug delivery device design verification —
Requirements and test methods
1 Scope
This International Standard applies to the design, labelling, instructions for use and testing requirements for
hand-held single- and multi-use aerosol drug delivery devices (ADDDs) intended to deliver a metered or pre-
metered aerosolized medication to or by means of the human respiratory tract (including nasal, oral, tracheal,
bronchial and alveolar sites). This International Standard applies to both refillable and disposable devices
intended for personal use.
This International Standard is intended for device design verification and not for drug product quality
assessment. The objective of this International Standard is to verify, by laboratory (in-vitro) testing, that the
ADDD design consistently meets the manufacturer's design specification by satisfying a device functionality
profile and system verification test both of which are determined from a risk assessment and evaluated in
accordance with the instructions for use.
This International Standard excludes continuous or semi-continuous aerosolization devices covered in
ISO 27427, aerosolization devices which do not emit active pharmaceutical ingredient (API), general purpose
aerosolization devices (for use with ventilators) and atomizers.
This International Standard does not apply to manufacturers of single parts or components of the ADDDs [e.g.
(spray) pumps, valves, containers, etc.].
NOTE There might be times when a device falls under the scope of this International Standard and that of
ISO 27427. The committee envisions that the intended use of the product and the risk assessment of the device will derive
which International Standard the manufacturer chooses for design verification of the ADDD. This International Standard
outlines the process by which ADDD design verification is to be performed in conjunction with a risk-based device
functionality profile of the ADDD with either the medication, a placebo or a representative medication. ISO 27427 outlines
the process by which the characterization of the aerodynamic aerosol performance of a nebulizing system for use with a
non-specific class of active pharmaceutical ingredient(s) is made.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 14971:2007, Medical devices — Application of risk management to medical devices
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing
ISO 11135-1, Sterilization of health care products — Ethylene oxide — Part 1: Requirements for development,
validation and routine control of a sterilization process for medical devices
ISO 11137 (all parts), Sterilization of health care products — Radiation
ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to
be supplied — Part 1: General requirements
© ISO 2009 – All rights reserved 1

---------------------- Page: 13 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
ISO 17665-1, Sterilization of health care products — Moist heat — Part 1: Requirements for the development,
validation and routine control of a sterilization process for medical devices
IEC 60068-2-27, Environmental testing — Part 2-27: Tests — Test Ea and guidance: Shock
IEC 60068-2-30:2005, Environmental testing — Part 2-30: Tests — Test Db: Damp heat, cyclic (12 h + 12 h
cycle)
IEC 60068-2-32, Environmental testing — Part 2: Tests. Test Ed: Free fall
IEC 60068-2-64, Environmental testing — Part 2-64: Tests — Test Fh: Vibration, broadband random and
guidance
IEC 60601-1, Medical electrical equipment — Part 1: General requirements for basic safety and essential
performance
IEC 60601-1-2, Medical electrical equipment — Part 1-2: General requirements for basic safety and essential
performance — Collateral standard: Electromagnetic compatibility — Requirements and tests
IEC 60601-1-8, Medical electrical equipment — Part 1-8: General requirements for basic safety and essential
performance — Collateral standard: General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems
IEC 60721-3-7, Classification of environmental conditions — Part 3-7: Classification of groups of
environmental parameters and their severities — Portable and non-stationary use
IEC 61000-4-2, Electromagnetic compatibility (EMC) — Part 4-2: Testing and measurement techniques —
Electrostatic discharge immunity test
IEC 61000-4-3, Electromagnetic compatibility (EMC) — Part 4-3: Testing and measurement techniques —
Radiated, radio-frequency, electromagnetic field immunity test
IEC 62304, Medical device software — Software life-cycle processes
IEC 62366, Medical devices — Application of usability engineering to medical devices
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
accessory
add-on device (specifically referenced in the ADDD instructions for use) that may be used in conjunction with
an ADDD to enable or enhance its performance
EXAMPLE Spacers, holding chambers, actuation counters, content indicators, etc.
3.2
active pharmaceutical ingredient
API
molecule(s) responsible for producing the intended therapeutic action
3.3
actuation
operation of the ADDD to release medication that will be aerosolized
NOTE The actuation can consist of the loading and release of the medication or only the release of the medication.
2 © ISO 2009 – All rights reserved

---------------------- Page: 14 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
3.4
actuation counter
dose counter
mechanism numerically counting down the number of actuations
NOTE The actuation counter may either be an accessory or integrated with the ADDD.
3.5
ADDD system
integrated system comprised of the ADDD, patient interface and the medication (i.e. a combination product)
3.6
aerosol
suspension of particles in gas
[ISO 27427:2009, definition 3.1]
NOTE 1 Particles can be liquid and/or solid.
NOTE 2 The gas can be the driving gas or ambient air.
3.7
aerosol drug delivery device
ADDD
device for the delivery of medication in the form of an aerosol
3.8
claimed lifetime
time period, and/or number of actuations, stated by the ADDD manufacturer within which the device
functionality profile of the ADDD will remain within the design specification comprising the timeframe within
which the patient uses the ADDD
NOTE Claimed lifetime is not necessarily the same as the shelf life of the ADDD.
3.9
claimed lifetime testing
performance evaluation simulating the claimed lifetime stated by the ADDD manufacturer within which the
device functionality profile of the ADDD will remain within the design specification
3.10
combination product
ADDD used with a specific medication
3.11
content indicator
visual indicator showing the amount of medication remaining in the ADDD
NOTE A content indicator can be either an accessory or integrated with the ADDD.
3.12
design verification
confirmation by examination and provision of objective evidence that specified design requirements have been
fulfilled
3.13
device functionality profile
DFP
parameters and tolerance intervals used to assess whether the ADDD meets the manufacturer's design
specification
NOTE It must be possible to evaluate the properties using laboratory (in vitro) testing.
© ISO 2009 – All rights reserved 3

---------------------- Page: 15 ----------------------

SIST EN ISO 20072:2013
ISO 20072:2009(E)
3.14
dose
mass of API prescribed to elicit a therapeutic response
NOTE 1 More than one actuation of the ADDD may be required to achieve the specified dose.
NOTE 2 For certain APIs, mass can be replaced by the use of biological equivalent units.
3.15
emitted mass
EM
mass of medication per actuation emitted from the ADDD mouthpiece at the patient interface
3.16
excipient
any substance included with the active pharmaceutical ingredient(s) of the medication
3.17
fixed-dose ADDD
ADDD where the amount of medication delivered per actuation (mass or bioequivalent units), either pre-
metered or from a reservoir, is pre-set by the manufacturer
3.18
hand-held
capable of being held in the hand and moved to the patient's mouth or nose for use
3.19
harm
physical injury or damage to the health of people, or damage to property or the environment
[ISO 14971:2007, definition 2.2]
3.20
holding chamber
accessory comprising a volume between the ADDD and the patient's mouth or nose and designed to contain
the aerosol following an actuation
NOTE A holding chamber has some means for retaining the aerosol after ADDD actuation has occur
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.