Ultrasonics - Hydrophones -- Part 2: Calibration for ultrasonic fields up to 40 MHz

This part of IEC 62127 specifies: - absolute hydrophone calibration methods; - relative (comparative) hydrophone calibration methods. This standard is applicable to - hydrophones used for measurements made in water and in the ultrasonic frequency range up to 40 MHz; - hydrophones employing circular piezoelectric sensor elements, designed to measure the pulsed wave and continuous wave ultrasonic fields generated by ultrasonic equipment; - hydrophones with or without a hydrophone pre-amplifier. IEC 62127-1, IEC 62127-2 and IEC 62127-3 are being published simultaneously. Together these cancel and replace IEC 60866:1987, IEC 61101:1991, IEC 61102:1991, IEC 61220:1993 and IEC 62092:2001. The contents of the corrigendum of August 2008 have been included in this copy.

Ultraschall - Hydrophone - Teil 2: Kalibrierung für Ultraschallfelder bis zu 40 MHz

Ultrasons - Hydrophones -- Partie 2: Etalonnage pour les champs ultrasonores jusqu'à 40 Mhz

La CEI 62127-2:2007 spécifie les méthodes d'étalonnage absolues de l'hydrophone et les méthodes d'étalonnage (comparatives) relatives de l'hydrophone. Cette norme s'applique:
- aux hydrophones utilisés pour procéder à des mesures dans l'eau dans la plage de fréquences ultrasonores pouvant atteindre 40 MHz;
- aux hydrophones utilisant des capteurs piézoélectriques circulaires, conçus pour mesurer les champs ultrasoniques à ondes pulsées et entretenues générés par les appareils à ultrasons;
- et aux hydrophones avec ou sans préamplificateur d'hydrophone. La CEI 62127-1, la CEI 62127-2 et la CEI 62127-3 ont été publiées conjointement. Ensemble, elles annulent et remplacent la CEI 60866:1987, la CEI 61101:1991, la CEI 61102:1991, la CEI 61220:1993 et la CEI 62092:2001. Le contenu du corrigendum (août 2008, disponible en anglais seulement) a été inclus dans cette norme. La version française n'a pas été soumise au vote.

Ultrazvok - Hidrofoni - 2. del: Kalibracija za ultrazvočna polja do 40 MHz (IEC 62127-2:2007)

Ta del standarda IEC 62127 določa: absolutne metode kalibracije hidrofonov; relativne (primerjalne) metode kalibracije hidrofonov. Priporočila in sklici na sprejeto literaturo so podani za številne relativne in absolutne metode kalibracije v frekvenčnem območju, ki ga zajema ta standard. Ta standard se uporablja za hidrofone, ki se uporabljajo za meritve, opravljene v vodi in ultrazvočnem frekvenčnem območju do 40 MHz; hidrofone, ki uporabljajo krožne piezoelektrične senzorne elemente in so zasnovani za merjenje ultrazvočnih polj z impulznimi in neprekinjenimi valovi, ki jih ustvari ultrazvočna oprema; hidrofone s hidrofonskim predojačevalnikom ali brez njega.

General Information

Status
Published
Publication Date
05-Nov-2007
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
24-Oct-2007
Due Date
29-Dec-2007
Completion Date
06-Nov-2007

Relations

Standard
SIST EN 62127-2:2008
English language
86 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-januar-2008
1DGRPHãþD
SIST EN 61101:2002
SIST EN 61102:2002
SIST EN 61102:2002/A1:2002
SIST EN 61220:2002
SIST EN 62092:2002
8OWUD]YRN+LGURIRQLGHO.DOLEUDFLMD]DXOWUD]YRþQDSROMDGR0+] ,(&

Ultrasonics - Hydrophones - Part 2: Calibration for ultrasonic fields up to 40 MHz (IEC
62127-2:2007)
Ultraschall - Hydrophone - Teil 2: Kalibrierung für Ultraschallfelder bis zu 40 MHz (IEC
62127-2:2007)
Ultrasons - Hydrophones - Partie 2: Etalonnage pour les champs ultrasonores jusqu'a 40
Mhz (IEC 62127-2:2007)
Ta slovenski standard je istoveten z: EN 62127-2:2007
ICS:
11.040.01 Medicinska oprema na Medical equipment in general
splošno
17.140.50 Elektroakustika Electroacoustics
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD
EN 62127-2
NORME EUROPÉENNE
October 2007
EUROPÄISCHE NORM
ICS 17.140.50 Partially supersedes EN 61101:1993, EN 61102:1993 + A1:1994, EN 61220:1995 and EN 62092:2001

English version
Ultrasonics -
Hydrophones -
Part 2: Calibration for ultrasonic fields up to 40 MHz
(IEC 62127-2:2007)
Ultrasons -  Ultraschall -
Hydrophones - Hydrophone -
Partie 2: Etalonnage pour les champs Teil 2: Kalibrierung für Ultraschallfelder
ultrasonores jusqu'à 40 Mhz bis zu 40 MHz
(CEI 62127-2:2007) (IEC 62127-2:2007)

This European Standard was approved by CENELEC on 2007-09-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels

© 2007 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 62127-2:2007 E
Foreword
The text of document 87/353/CDV, future edition 1 of IEC 62127-2, prepared by IEC TC 87, Ultrasonics,
was submitted to the IEC-CENELEC parallel Unique Acceptance Procedure and was approved by
CENELEC as EN 62127-2 on 2007-09-01.
EN 62127-1, EN 62127-2 and EN 62127-3 are being published simultaneously. Together these European
Standards cancel and replace EN 61101:1993, EN 61102:1993 + A1:1994, EN 61220:1995 and
EN 62092:2001.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2008-06-01
– latest date by which the national standards conflicting
(dow) 2010-09-01
with the EN have to be withdrawn
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 62127-2:2007 was approved by CENELEC as a European
Standard without any modification.
__________
- 3 - EN 62127-2:2007
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.
Publication Year Title EN/HD Year

IEC 60050-801 1994 International Electrotechnical Vocabulary - -
(IEV) -
Chapter 801: Acoustics and electroacoustics

1) 2)
IEC 60565 - Underwater acoustics - Hydrophones - EN 60565 2007
Calibration in the frequency range
0,01 Hz to 1 MHz
IEC 61161 2006 Ultrasonics - Power measurement - EN 61161 2007
Radiation force balances and performance
requirements
IEC 61828 2001 Ultrasonics - Focusing transducers - EN 61828 2001
Definitions and measurement methods
for the transmitted fields
1) 2)
IEC 62127-1 - Ultrasonics - Hydrophones - EN 62127-1 2007
Part 1: Measurement and characterization
of medical ultrasonic fields up to 40 MHz

1) 2)
IEC 62127-3 - Ultrasonics - Hydrophones - EN 62127-3 2007
Part 3: Properties of hydrophones for
ultrasonic fields up to 40 MHz

1)
Undated reference.
2)
Valid edition at date of issue.

IEC 62127-2
Edition 1.0 2007-08
INTERNATIONAL
STANDARD
Ultrasonics – Hydrophones –
Part 2: Calibration for ultrasonic fields up to 40 MHz

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
PRICE CODE
XC
ICS 17.140.50 ISBN 2-8318-9277-5

– 2 – 62127-2 © IEC:2007(E)
CONTENTS
FOREWORD.5
INTRODUCTION.7

1 Scope.8
2 Normative references.8
3 Terms, definitions and symbols .9
4 List of symbols.13
5 Overview of calibration procedures .15
5.1 Principles .15
5.2 Summary of calibration procedures .16
5.3 Reporting of results .17
5.4 Recommended calibration periods .18
6 Generic requirements of a hydrophone calibration system .19
6.1 Mechanical positioning.19
6.1.1 General .19
6.1.2 Accuracy of the axial hydrophone position .19
6.1.3 Accuracy of the lateral hydrophone position.19
6.2 Temperature measurements and temperature stability .19
6.3 Hydrophone size.20
6.4 Measurement vessel and water properties .20
6.5 Measurement of output voltage .20
7 Electrical considerations .21
7.1 Signal type .21
7.2 Earthing.21
7.3 Measurement of hydrophone output voltage .21
7.3.1 General .21
7.3.2 Electrical loading by measuring instrument.21
7.3.3 Electrical loading by extension cables .22
7.3.4 Noise.22
7.3.5 Cross-talk (radio-frequency rf pick-up) and acoustic interference .22
7.3.6 Integral hydrophone pre-amplifiers .23
8 Preparation of hydrophones .23
8.1 General .23
8.2 Wetting.23
8.3 Hydrophone support .23
8.4 Influence of cable .23
9 Free field reciprocity calibration .23
9.1 General .23
9.2 Object .23
9.3 General principles.24
9.3.1 General .24
9.3.2 Three-transducer reciprocity calibration method .24
9.3.3 Self-reciprocity calibration method .24
9.3.4 Two-transducer reciprocity calibration method.24
9.4 Two-transducer reciprocity calibration method.24
9.4.1 Apparatus .24

62127-2 © IEC:2007(E) – 3 –
9.4.2 Procedure.25
10 Free field calibration by planar scanning .25
10.1 General .25
10.2 Object .25
10.3 General principle .25
10.4 Procedural requirements.27
10.4.1 Hydrophone scanning .27
10.5 Procedure.27
10.5.1 Power measurement .27
10.5.2 Transducer mounting .27
10.5.3 Measurement conditions .27
10.5.4 Measurements .28
10.6 Corrections and sources of uncertainty .28
11 Free field calibration by optical interferometry .28
11.1 General .28
11.2 Principle .28
12 Calibration by comparison using a standard hydrophone .28
12.1 General .28
12.2 Object .28
12.3 Principle .29
12.4 Procedural requirements.29
12.4.1 Source transducer .29
12.4.2 Source transducer drive signal .29
12.4.3 Measurement system .29
12.5 Procedure.30
12.5.1 Measurements (Type I): determination of the directional response of a
hydrophone.30
12.5.2 Measurements (Type II): calibration by comparison using a standard
hydrophone.30
12.6 Maximum hydrophone size.31

Annex A (informative) Assessment of uncertainty in free field calibration of
hydrophones .32
Annex B (informative) Behaviour of PVDF polymer sensors in high intensity ultrasonic
fields.34
Annex C (informative) Electrical loading corrections .37
Annex D (informative)  Absolute calibration of hydrophones using the planar scanning
technique .38
Annex E (informative) Properties of water.46
Annex F (informative)  The absolute calibration of hydrophones by optical
interferometry up to 40 MHz .48
Annex G (informative)  Waveform concepts.58
Annex H (informative) Time delay spectrometry – requirements and a brief review of
the technique .68
Annex I (informative) Determination of the phase response of hydrophones .71
Annex J (informative)  Maximum size considerations for the active element of a
hydrophone.77

– 4 – 62127-2 © IEC:2007(E)
Bibliography .79

Figure F.1 – Experimental set-up of the interferometric foil technique .51
Figure F.2 – End-of-cable open-circuit sensitivity, M , of a coplanar membrane
c
hydrophone.53
Figure F.3 – Hydrophone waveform generated by a 9 μm coplanar membrane
hydrophone positioned at the focus of a 5 MHz transducer (focal length 51 mm) .54
Figure F.4 – Interferometer (displacement) waveform generated with the pellicle
positioned at the focus of the 5 MHz transducer (focal position 51 mm) .55
Figure F.5 – Frequency spectrum of the displacement waveform (lower curve) and the
differentiated displacement waveform (upper curve) .55
Figure F.6 – Sensitivity of a 0,2 mm active element diameter of a 9 μm bilaminar
membrane hydrophone determined at 5 MHz intervals over the frequency range 5 MHz
to 60 MHz .56
Figure G.1 – Coordinates of a field point, P, in the near field of a plane- circular source
transducer of radius, a .65
t
Figure I.1 – Phase of end-of-cable open-circuit sensitivity for two membrane
hydrophones .73
Figure I.2 – Phase of end-of-cable open-circuit sensitivity for a ∅0,2 mm needle
hydrophone.75

Table 1 – List of typical uncertainty values obtained by the calibration methods
specified in this standard and for the frequency range listed here .17
Table E.1 – Speed of sound c and specific acoustic impedance, ρc, as a function of
temperature, for propagation in water .46
Table G.1 – Temporal waveform and hydrophone position concepts described in this
Annex .
Table I.1. – Example of uncertainties (where a coverage factor, k = 2, is used) for a
HTDS phase calibration of a needle hydrophone with a diameter of 0,2 mm, expressed
at a confidence level of 95 %.73

62127-2 © IEC:2007(E) – 5 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
ULTRASONICS – HYDROPHONES –
Part 2: Calibration for ultrasonic fields up to 40 MHz

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 62127-2 has been prepared by IEC technical committee 87:
Ultrasonics.
IEC 62127-1, IEC 62127-2 and IEC 62127-3 are being published simultaneously. Together
these cancel and replace IEC 60866:1987, IEC 61101:1991, IEC 61102:1991, IEC 61220:1993
and IEC 62092:2001.
The text of this standard is based on the following documents:
Enquiry draft Report on voting
87/353/CDV 87/372/RVC
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.

– 6 – 62127-2 © IEC:2007(E)
A list of all parts of IEC 62127 series, published under the general title Ultrasonics –
Hydrophones, can be found on the IEC website.
NOTE Words in bold in the text are defined in Clause 3.
The committee has decided that the contents of this publication will remain unchanged until the
maintenance result date indicated on the IEC website under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition; or
• amended.
A bilingual version of this publication may be issued at a later date.

62127-2 © IEC:2007(E) – 7 –
INTRODUCTION
The spatial and temporal distribution of acoustic pressure in an ultrasonic field in a liquid
medium is commonly determined using miniature ultrasonic hydrophones. These devices are
not absolute measurement instruments and require calibration. The purpose of this part of
IEC 62127 is to specify those calibration methods to be used in determining the response of a
hydrophone in the ultrasonic range, i.e. above 20 kHz up to a frequency of 40 MHz. The main
hydrophone application in this context lies in the measurement of ultrasonic fields emitted by
medical diagnostic equipment in water. Hydrophone behaviour over this wide frequency band
is required in order to reliably characterize the acoustic parameters of the applied acoustic
field. In particular, the frequency range above 15 MHz is important to fully characterize this
equipment, primarily due to the increased appearance of high-frequency components in the
ultrasonic signals, caused by non-linear propagation. In addition, the number of medical
ultrasonic systems that use frequencies above 15 MHz, particularly intra-operative probes, is
growing. It has turned out in recent years that the hydrophone response below 0,5 MHz is also
required to reliably determine the peak-negative (rarefactional) acoustic pressure.
While the term "hydrophone" can be used in a wider sense, it is understood here as referring
to miniature piezoelectric hydrophones. It is this instrument type that is used today in various
areas of medical ultrasonics and, in particular, to characterize quantitatively the field structure
of medical diagnostic instruments. With regard to other pressure sensor types, such as those
based on fibre optics, some of the requirements of this standard are applicable to these as well
but others are not. If in the future these other "hydrophone" types gain more importance in
field measurement practice, their characteristics and calibration will have to be dealt with in a
revised version of this standard or in a separate one.
NOTE This standard covers the ultrasonic frequency range, from 20 kHz to an upper frequency of 40 MHz.
Standards dealing with hydrophone properties (IEC 62127-3) and hydrophone use (IEC 62127-1) are being
developed in parallel as part of a programme of maintenance activities aimed at restructuring and merging, where
possible, all existing ultrasonic hydrophone standards. This will eventually lead to unified standards covering the
whole field of practical hydrophone application.

– 8 – 62127-2 © IEC:2007(E)
ULTRASONICS — HYDROPHONES —
Part 2: Calibration for ultrasonic fields up to 40 MHz

1 Scope
This part of IEC 62127 specifies:
• absolute hydrophone calibration methods;
• relative (comparative) hydrophone calibration methods.
Recommendations and references to accepted literature are made for the various relative and
absolute calibration methods in the frequency range covered by this standard.
This standard is applicable to
• hydrophones used for measurements made in water and in the ultrasonic frequency range
up to 40 MHz;
NOTE 1 Although some physiotherapy medical applications of medical ultrasound are developing which
operate in the frequency range 40 kHz to 100 kHz, the primary frequency range of diagnostic imaging remains
above 2 MHz. It has recently been established that, even in the latter case, the hydrophone response at
substantially lower frequencies can influence measurements made of key acoustic parameters [1].
• hydrophones employing circular piezoelectric sensor elements, designed to measure the
pulsed wave and continuous wave ultrasonic fields generated by ultrasonic equipment;
NOTE 2 Some hydrophones can have non-circular active elements, arising from slight deviations from a
circular structure caused, for example by electrode structure, or conversely, the active elements can actually
be squares. The clauses within this standard remain valid, although, in these cases, special attention should
be paid to the directional response and to the effective radii of the active element through various axes of
rotation.
• hydrophones with or without a hydrophone pre-amplifier.
2 Normative references
The following referenced documents are indispensable for the application of this document. For
dated references, only the edition cited applies. For undated references, the latest edition of
the referenced document (including any amendments) applies.
IEC 60050-801:1994, International Electrotechnical Vocabulary – Chapter 801: Acoustics
and electro-acoustics
IEC 60565, Underwater acoustics – Hydrophones – Calibration in the frequency range 0,01 Hz
to 1 MHz
IEC 61161:2006, Ultrasonics – Power measurement – Radiation force balances and performance
requirements
IEC 61828:2006, Ultrasonics – Focusing transducers – Definitions and measurement methods
for the transmitted fields
IEC 62127-1, Ultrasonics – Hydrophones – Part 1: Measurement and characterization of
medical ultrasonic fields up to 40 MHz
IEC 62127-3, Ultrasonics – Hydrophones – Part 3: Properties of hydrophones for ultrasonic
fields up to 40 MHz
62127-2 © IEC:2007(E) – 9 –
3 Terms, definitions and symbols
For the purposes of this document, the terms and definitions given in IEC 62127-1 and the
following apply.
3.1
acoustic centre
point on or near a transducer from which the spherically divergent sound waves emitted by the
transducer, and observable at remote points, appear to diverge
3.2
beam axis
straight line that passes through the beam centrepoints of two planes perpendicular to the line
which connects the point of maximal pulse-pressure-squared integral with the centre of the
external transducer aperture
NOTE 1 The location of the first plane is the location of the plane containing the maximum pulse-pressure-
squared integral or, alternatively, is one containing a single main lobe which is in the focal Fraunhofer zone. The
location of the second plane is as far as is practicable from the first plane and parallel to the first with the same two
orthogonal scan lines (x and y axes) used for the first plane.

NOTE 2 In a number of cases, the term pulse-pressure-squared integral is replaced in the above definition by
any linearly related quantity, for example
a) in the case of a continuous wave signal the term pulse-pressure-squared integral is replaced by mean square
acoustic pressure as defined in IEC 61689,
b) in cases where signal synchronisation with the scanframe is not available the term pulse-pressure-squared
integral may be replaced by temporal average intensity.
NOTE 3 See Figure 1 of IEC 62127-1.
NOTE 4 Definition adopted from IEC 62127-1.
3.3
beam centrepoint
position determined by the intersection of two lines passing through the beamwidth midpoints
of two orthogonal planes, xz and yz

NOTE Definition adopted from IEC 61828:2001.
3.4
beamwidth
w , w , w
6 12 20
greatest distance between two points on a specified axis perpendicular to the beam axis where
the pulse-pressure-squared integral falls below its maximum on the specified axis by a
specified amount
NOTE 1 In a number of cases, the term pulse-pressure-squared integral is replaced in the above definition by
any linearly related quantity, for example
a) in the case of a continuous wave signal the term pulse-pressure-squared integral is replaced by mean square
acoustic pressure as defined in IEC 61689,
b) in cases where signal synchronisation with the scanframe is not available the term pulse-pressure-squared
integral may be replaced by temporal average intensity.
NOTE 2 Commonly used beamwidths are specified at – 6 dB, –12 dB and –20 dB levels below the maximum. The
decibel calculation implies taking 10 times the logarithm of the ratios of the integrals.
NOTE 3 Beamwidth is expressed in metres (m).
NOTE 4 Definition adopted from IEC 62127-1.

– 10 – 62127-2 © IEC:2007(E)
3.5
beamwidth midpoint
linear average of the location of the centres of beamwidths in a plane
NOTE 1 The average is taken over 20 beamwidth levels corresponding to intervals in the −0,1 dB to −26 dB range
(see IEC 61828, Clause B.2).
NOTE 2 Definition adopted from IEC 61828:2001.
3.6
beam centre
point in a plane in the far field, usually perpendicular to the beam axis, at which the spatial-
peak temporal-peak acoustic pressure occurs
3.7
diametrical beam scan
set of measurements of the hydrophone output voltage made while moving the hydrophone in a
straight line passing through a point on the beam axis and in a direction normal to the beam
axis
NOTE 1 The diametrical beam scan may extend to different distances on either side of the beam axis.
NOTE 2 Definition adopted from IEC 62127-1.
3.8
directional response
description, generally presented graphically, of the response of a hydrophone, as a function of
direction of propagation of the incident plane sound wave, in a specified plane through the
reference centre and at a specified frequency
NOTE Definition adopted from IEC 60565.
3.9
effective radius of a non-focused ultrasonic transducer
a
t
radius of a perfect disc piston-like ultrasonic source transducer that has a predicted axial
acoustic pressure distribution approximately equivalent to the observed axial acoustic pressure
distribution over an axial distance until at least the last axial maximum has passed

NOTE 1 The effective radius of an non-focused ultrasonic transducer is expressed in metres (m).
NOTE 2 Definition adopted from IEC 62127-1.
3.10
effective hydrophone radius
a , a , a
h h3 h6
radius of a stiff disc receiver hydrophone that has a predicted directional response function
with an angular width equal to the observed angular width
NOTE 1 The angular width is determined at a specified level below the peak of the directional response function.
For the specified levels of 3 dB and 6 dB, the radii are denoted by a and a respectively.
h3 h6
NOTE 2 The effective hydrophone radius is expressed in metres (m).
NOTE 3 The radius is usually a function of frequency. For representative experimental data, see [2].
NOTE 4 Definition adopted from IEC 62127-3.

62127-2 © IEC:2007(E) – 11 –
3.11
electric load impedance
Z
L
complex electric input impedance (consisting of a real and an imaginary part) to which the
hydrophone assembly output cable is connected or is to be connected
NOTE 1 The electric load impedance is expressed in ohms (Ω).
NOTE 2 Definition adopted from IEC 62127-3.
3.12
end-of-cable loaded sensitivity
end-of-cable loaded sensitivity of a hydrophone (or hydrophone-assembly)
M (f)
L
ratio of the instantaneous voltage at the end of any integral cable or output connector of a
hydrophone or hydrophone-assembly, when connected to a specified electric load
impedance, to the instantaneous acoustic pressure in the undisturbed free field of a plane
wave in the position of the reference centre of the hydrophone if the hydrophone were
removed
NOTE 1 End-of-cable loaded sensitivity is expressed in volts per pascal (V/Pa).
NOTE 2 Definition adopted from IEC 62127-3.
3.13
end-of-cable open-circuit sensitivity
end-of-cable open-circuit sensitivity of a hydrophone
M (f)
c
ratio of the instantaneous open-circuit voltage at the end of any integral cable or output
connector of a hydrophone to the instantaneous acoustic pressure in the undisturbed free
field of a plane wave in the position of the reference centre of the hydrophone if the
hydrophone were removed
NOTE 1 End-of-cable open-circuit sensitivity is expressed in volts per pascal (V/Pa).
NOTE 2 Definition adopted from IEC 62127-3.
3.14
external transducer aperture
part of the surface of the ultrasonic transducer or ultrasonic transducer element group
assembly that emits ultrasonic radiation into the propagation medium.

NOTE 1 This surface is either directly in contact with the patient or is in contact with a water or liquid path to the
patient (see Figure 2 of IEC 62127-1).
NOTE 2 Definition adopted from IEC 61828:2001.
3.15
far field
acoustic (sound) field at distances from an ultrasonic transducer where the values of the
instantaneous acoustic pressure and particle velocity are substantially in phase (see also
IEC 60050-801, 801-23-30)
NOTE 1 In the far field, the sound pressure appears to be spherically divergent from a point on or near the
radiating surface. Hence, the pressure produced by the sound source is approximately inversely proportional to the
distance from the source.
NOTE 2 The term "far field" is used in this standard only in connection with non-focusing source transducers. For
focusing transducers a different terminology for the various parts of the transmitted field applies (see IEC 61828).

– 12 – 62127-2 © IEC:2007(E)
3.16
free field
sound field in a homogeneous and isotropic medium in which the effects of boundaries are
negligible
NOTE Definition adopted from IEC 60565: 2006, 3.13.
3.17
hydrophone
transducer that produces electric signals in response to waterborne acoustic signals.
NOTE Definition adopted from IEC 60050-801, 801-32-26.
3.18
hydrophone assembly
combination of hydrophone and hydrophone pre-amplifier

NOTE 2 Definition adopted from IEC 62127-3.
3.19
hydrophone axis
nominal symmetry axis of the hydrophone active element
NOTE 1 Unless stated otherwise (explicitly and quantitatively) by the manufacturer, it is understood for the
purposes of this standard that this is given by the apparent geometrical symmetry axis of the hydrophone.
NOTE 2 Definition adopted from IEC 62127-3.
3.20
hydrophone geometrical radius
geometrical radius of a hydrophone active element
a
g
radius defined by the dimensions of the active element of a hydrophone
NOTE 1 The hydrophone geometrical radius is expressed in metres (m)
NOTE 2 Definition adopted from IEC 62127-3.
3.21
hydrophone pre-amplifier
active electronic device connected to, or to be connected to, a particular hydrophone and
reducing its output impedance
NOTE 1 A hydrophone pre-amplifier requires a supply voltage (or supply voltages).
NOTE 2 The hydrophone pre-amplifier may have a forward voltage transmission factor of less than one, i.e. it
need not necessarily be a voltage amplifier in the strict sense.
NOTE 3 Definition adopted from IEC 62127-3.
3.22
instantaneous acoustic pressure
p(t)
pressure minus the ambient pressure at a particular instant in time and at a particular point in
an acoustic field (see also IEC 60050-801, 801-21-19)
NOTE 1 Instantaneous acoustic pressure is expressed in pascal (Pa).
NOTE 2 Definition adopted from IEC 62127-1.

62127-2 © IEC:2007(E) – 13 –
3.23
instantaneous intensity
I(t)
acoustic energy transmitted per unit time in the direction of acoustic wave propagation per unit
area normal to this direction at a particular instant in time and at a particular point in an
acoustic field
NOTE 1 Instantaneous intensity is the product of instantaneous acoustic pressure and particle velocity. It is
difficult to measure intensity in the ultrasound frequency range. For the measurement purposes referred to in this
standard, and if it is reasonable to assume far field conditions, the instantaneous intensity, I is approximated as
p(t)
(1)
I(t) =
ρ c
where
p(t) is the instantaneous acoustic pressure;
ρ is the density of the medium;
c is the velocity of sound in the medium.
NOTE 2 Instantaneous intensity is expressed in watts per metre squared (W/m ).
3.24
reference centre
point on or near a hydrophone about which its acoustic receiving sensitivity is defined
NOTE Unless stated otherwise (explicitly and quantitatively) by the manufacturer, it is understood for the purposes
of this standard that this is given by the geometrical centre of the front surface of the hydrophone active element.
(See IEC 60565: 2006, 3.25)
3.25
uncertainty
parameter, associated with the result of a measurement, that characterizes the dispersion of
the values that could reasonably be attributed to the measurand
NOTE See the ISO Guide to the Expression of Uncertainty in Measurement [3], 2.2.3.
4 List of symbols
a effective hydrophone radius (a , a : with special reference to a 3 dB or
h h3 h6
6 dB definition, respectively)
a hydrophone geometrical radius
g
a maximum effective radius for a specific hydrophone application
max
a lateral distance from the beam axis (a , a : maximum values with
P PmaxE PmaxH
respect to avoiding edge wave and head wave interference, respectively)
a effective radius of an non-focused ultrasonic transducer
t
A geometrical area of an ultrasonic transducer
g
B/A Fox-Wallace non-linearity parameter
c speed of sound in a medium (usually water)

C end-of-cable capacitance of the hydrophone including any integral cable and
H
connector
normalized directional response function
D(θ)
e base of natural logarithms
f frequency
f fundamental drive frequency of a signal used to generate non-linear
f
distortion
– 14 – 62127-2 © IEC:2007(E)
f upper frequency limit of the stated frequency band of a hydrophone
u
I magnitude of the instantaneous intensity assuming proportionality with

p
acoustic pressure squared
r
instantaneous intensity vector at the point (x,y,z) at time t

I (x,y,z,t)
I(x,y,z,t)  component of the instantaneous intensity vector in the propagation
direction at the point (x,y,z) at time t
F geometric focal length of a focusing transducer

jmax the number of the farthest scan point in a radial scan

k
circular wave number (= 2π/λ)
M
general symbol for the complex hydrophone sensitivity, M = M being its
modulus and arg(M) being its argument (= phase angle)
M (f) end-of-cable open-circuit sensitivity
c
M (f) end-of-cable loaded sensitivity
L
N number of waveform averages taken to generate a time-averaged voltage
av
waveform
N number of diametrical samples
N harmonic number
harm
p pressure amplitude
p acoustic pressure amplitude at the face of a transducer
p(x,y,z,t) instantaneous acoustic pressure at the point (x,y,z) at time t
P(l) total ultrasonic power passing through a plane of infinite extent in an
acoustic field at a distance l from an ultrasonic transducer
P total ultrasonic power emitted by a transducer
o
rf radio-frequency signals
r distance from the ultrasonic beam centre to a scan point
R
(lλ/πa ) - normalized distance between a transducer and a hydrophone in
terms of the Rayleigh distance (πa /λ)
R ,R distances from the beam centre to the extremes of the ith diametrical
1i 2i
beam scan
s distance from the ultrasonic beam centre to the nearest scan point
t arrival time of the nearest head wave
H
t time available for a free field measurement in time delay spectrometry
TDS
TF acoustic transmission factor
U (x,y,z,t) end-of-cable voltage for a hydrophone with the hydrophone at the
L
reference point (x,y,z) and at time t

end-of-cable voltage for a hydrophone including noise with the
U (x,y,z,t)
L
hydrophone at the reference point (x,y,z) and at time t
U (x,y,z)  root-mean-square (rms) noise level measured with a hydrophone at the
n
reference point (x,y,z)
U voltage at the terminals of an ultrasonic transducer
T
r
Instantaneous particle velocity vector at the point (x,y,z) at time t
v(x,y,z,t)
v speed of a radial wave in a transducer plate
t
w  beamwidth of the fundamental-frequency field component
f
z  axial distance between a hydrophone and an ultrasonic transducer
surface. (z , z , z and z are special distance values according to certain
1 2 3 4
criteria involving edge waves and head waves)
z minimum distance for a finite size hydrophone from a transducer
min
z  distance of the pressure focus from a focusing transducer
pf
distance difference
Δz
Z complex electric output impedance of a hydrophone
h
62127-2 © IEC:2007(E) – 15 –
...