ISO 15553:2006
(Main)Water quality — Isolation and identification of Cryptosporidium oocysts and Giardia cysts from water
Water quality — Isolation and identification of Cryptosporidium oocysts and Giardia cysts from water
ISO 15553:2006 specifies a method that is applicable for the detection and enumeration of Cryptosporidium oocysts and Giardia cysts in water. It is applicable for the examination of surface and ground waters, treated waters, mineral waters, swimming pool and recreational waters. This method does not allow identification to species level, the host species of origin or the determination of viability or infectivity of any Cryptosporidium oocyst or Giardia cyst which may be present. These procedures are for use by experienced analysts who have successfully completed competency tests prior to commencing analysis.
Qualité de l'eau — Isolement et identification des oocystes de Cryptosporidium et des kystes de Giardia
Kakovost vode - Izolacija in identifikacija oocist Cryptosporidium in cist Giardia
Ta mednarodni standard določa metodo, ki velja za detekcijo in preštevanje oocist Cryptosporidium in cist Giardia v vodi. Uporablja se pri preučevanju površinskih in podtalnih voda, obdelanih voda, mineralnih voda, voda iz bazenov in rekreacijskih voda. Ta metoda ne omogoča identifikacije na ravni vrste, vrste njenih gostiteljic, ali določevanje sposobnosti za življenje oziroma nalezljivosti katere koli oociste Cryptosporidium ali ciste Giardia, ki so lahko prisotne. Te postopke naj uporabljajo izkušeni analitiki, ki so pred pričetkom analize uspešno opravili preskuse usposobljenosti. Poleg tega bi morali takšni analitiki še naprej dokazovati usposobljenost s preučevanjem zasejanih vzorcev v rednih intervalih in z udeležbo pri zunanjih shemah zagotavljanja kakovosti. OPOMBA: Telesa, ki spominjajo na Cryptosporidium ali Giardia po morfologiji, so lahko prisotna in le-ta se lahko pomotoma zamenjajo z oocisti ali cisti. Rezultate je potrebno razlagati pazljivo. Kadar obstaja dvom glede identitete oocist ali cist ali kadar je dosežen nenavadno visok rezultat, je priporočljivo, da mikroskopske vzorce preučijo strokovnjaki iz drugih laboratorijev, da te ugotovitve potrdijo ali zavrnejo.
General Information
Relations
Buy Standard
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 15553
First edition
2006-11-15
Water quality — Isolation and
identification of Cryptosporidium oocysts
and Giardia cysts from water
Qualité de l'eau — Isolement et identification des oocystes de
Cryptosporidium et des kystes de Giardia
Reference number
ISO 15553:2006(E)
©
ISO 2006
---------------------- Page: 1 ----------------------
ISO 15553:2006(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2006
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2006 – All rights reserved
---------------------- Page: 2 ----------------------
ISO 15553:2006(E)
Contents Page
Foreword. iv
Introduction . v
1 Scope . 1
2 Terms and definitions. 1
3 Principle. 1
4 Reagents. 2
5 Apparatus . 4
6 Sampling and transport. 6
7 Procedure . 7
8 Quality control procedures. 14
9 Reporting of results. 14
Annex A (normative) Preparation of reagents. 16
Annex B (informative) Concentration of oocysts and cysts from small (10 l) volumes of water. 19
Annex C (informative) Calibration of eyepiece graticule. 24
Annex D (informative) Preparation of positive controls and recovery tests. 25
Annex E (informative) Examples of indicative performance data . 28
Annex F (informative) Alternative methods. 30
Annex G (informative) Further information about Cryptosporidium and Giardia. 31
Annex H (informative) Details of manufacturers. 32
Bibliography . 36
© ISO 2006 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO 15553:2006(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 15553 was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 4,
Microbiological methods.
iv © ISO 2006 – All rights reserved
---------------------- Page: 4 ----------------------
ISO 15553:2006(E)
Introduction
Cryptosporidium and Giardia are protozoan parasites that can cause enteric illness in humans. Both
organisms are characterized by an ability to survive in the aquatic environment. Cryptosporidium in particular
is resistant to chlorine at the concentrations used in the treatment of drinking and swimming pool waters.
Consequently the absence of vegetative bacteria as indicators of faecal contamination does not necessarily
indicate the absence of Cryptosporidium oocysts or Giardia cysts. The methods described in this document
may be used to determine whether Cryptosporidium and/or Giardia are present in water supplies. The
techniques have been selected on the basis of method development and peer review publication of the data
thus obtained. They are further selected to give comparable recoveries of the methods or reagents used in the
isolation of the organisms.
© ISO 2006 – All rights reserved v
---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 15553:2006(E)
Water quality — Isolation and identification of Cryptosporidium
oocysts and Giardia cysts from water
1 Scope
This International Standard specifies a method that is applicable for the detection and enumeration of
Cryptosporidium oocysts and Giardia cysts in water. It is applicable for the examination of surface and ground
waters, treated waters, mineral waters, swimming pool and recreational waters.
This method does not allow identification to species level, the host species of origin or the determination of
viability or infectivity of any Cryptosporidium oocyst or Giardia cyst which may be present. These procedures
are for use by experienced analysts who have successfully completed competency tests prior to commencing
analysis. In addition, such analysts should continue to demonstrate competency by examining seeded
samples at regular intervals and taking part in external quality assurance schemes.
NOTE Bodies resembling Cryptosporidium or Giardia in morphology can be present and these may be mistaken for
oocysts or cysts. Results should be interpreted with care. Where there is doubt about the identity of oocysts or cysts or
where an unusually high result is obtained, it is advisable to have the slides examined by experts from other laboratories
to confirm or refute the findings.
2 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
2.1
Cryptosporidium
protozoan parasite, concentrated and selected from water samples with the methods described, which reacts
with specific anti-Cryptosporidium antibodies and exhibits the typical morphological characteristics described
in 7.4 of this International Standard
NOTE A more complete definition of the parasite and the different genotypes and species is given in Annex G.
2.2
Giardia
protozoan parasite, concentrated and selected from water samples with the methods described, which reacts
with specific anti-Giardia antibodies and exhibits the typical morphological characteristics described in 7.4 of
this International Standard
NOTE A more complete definition of the parasite and the different species is given in Annex G.
3 Principle
3.1 Concentration from water
The isolation of Cryptosporidium and Giardia from water requires the use of a procedure which allows the
volume of the sample to be reduced whilst retaining any oocysts and cysts. The concentration procedure used
however, is dependent upon the water type which is to be analysed, the volume of sample and the amount of
particulate material in the sample. This document describes the use of two concentration techniques for
varying volumes of water using cartridge filtration and elution followed by low speed centrifugation (7.1).
Additional methods for the recovery of oocysts and cysts from small volumes of water or very turbid waters
are given in Annex B. Some examples of recovery data for these techniques are given in Annex E.
© ISO 2006 – All rights reserved 1
---------------------- Page: 6 ----------------------
ISO 15553:2006(E)
Table 1 — Membrane filters/filtration systems used for the concentration of parasites
from water samples
Membrane filter/filtration system Application
Concentration of 10-litre to 200-litre (or more) samples of
TM a
Pall Envirochek STD
water
TM
Pall Envirochek HV Concentration of 10-litre to 1 000-litre samples of water
®
IDEXX Filta-Max Concentration of 10-litre to 1 000-litre samples of water
a
It has been shown by some laboratories that this technique may be used successfully for larger volumes of
water although the manufacturers’ instructions may only include volumes up to 200 litres.
3.2 Purification and further concentration
After concentration of particulate material from filter eluates, oocysts and cysts are isolated using
immunomagnetic separation (IMS) (7.2). Oocysts and cysts are attached to para-magnetic beads coated with
specific antibody, the beads are separated from the unwanted particulate material using a magnet and then
the oocysts and cysts are dissociated from the beads using acid and neutralized using alkali before
immunostaining.
3.3 Detection of Cryptosporidium and Giardia
After IMS, organisms are labelled with monoclonal antibody (mAb) conjugated to a fluorochrome, usually
fluoroscein isothiocyanate (FITC). In addition, any nuclear material is labelled with a nucleic acid stain to aid
identification (7.3). Each sample is then examined for the presence of labelled Cryptosporidium oocysts and
Giardia cysts using epifluorescence and differential interference contrast (DIC) microscopy (7.4).
4 Reagents
TM 1)
4.1 Reagents required for eluting Pall Envirochek STD capsule filters
4.1.1 Deionized water, 0,2 µm filtered at the point of use.
4.1.2 Laureth 12 detergent.
4.1.3 Tris buffer, pH 7,4 (A.1.1).
4.1.4 EDTA solution, 0,5 mol/l, pH 8,0 (A.1.2).
4.1.5 Antifoam A.
4.1.6 Elution buffer (A.1.3).
TM 1)
4.2 Reagents required for eluting Pall Envirochek HV capsule filters
4.2.1 Deionized water, 0,2 µm filtered at point of use.
4.2.2 Pre-treatment buffer (A.1.4).
4.2.3 Laureth 12 detergent
1) All products and reagents are examples of suitable products available commercially. This information is given for the
convenience of users of this International Standard and does not constitute an endorsement by ISO of these products.
2 © ISO 2006 – All rights reserved
---------------------- Page: 7 ----------------------
ISO 15553:2006(E)
4.2.4 Tris buffer, pH 7,4 (A.1.1).
4.2.5 EDTA solution, 0,5 mol/l, pH 8,0 (A.1.2).
4.2.6 Antifoam A.
4.2.7 Elution buffer (A.1.3).
® 1)
4.3 Reagents required for eluting IDEXX Filta-Max filters
4.3.1 Phosphate buffered saline (PBS) (A.2.1).
4.3.2 Polyoxyethylene(20)sorbitan monolaurate (Tween 20).
Store at room temperature (20 ± 5) °C. Expiry date one year.
4.3.3 Elution buffer (A.2.2).
4.4 Concentration and detection reagents
4.4.1 Methanol, analytical grade.
4.4.2 Magnetic beads, for the detection of Cryptosporidium and Giardia.
Expiry date printed by the manufacturer.
NOTE See Annex H for a list of suitable suppliers.
4.4.3 Fluorescently labelled monoclonal antibodies (mAbs) against Cryptosporidium and Giardia.
Store at (5 ± 3) °C. Expiry date as stated by the manufacturer. When stains are prepared from concentrated
material using a diluent supplied by the manufacturer, the prepared solution is stored at (5 ± 3) °C for no
longer than 6 months.
NOTE See Annex H for a list of suitable suppliers.
4.4.4 Immunofluorescence mounting medium (A.3.1).
NOTE See Annex H for a list of suitable suppliers.
4.4.5 4′,6′-Diamidino-2-phenylindole dihydrochloride dihydrate (DAPI) freeze dried reagent.
Store according to the manufacturer's instructions.
Expiry date printed by the manufacturer on each vial.
4.4.6 DAPI stock solution (A.3.2).
4.4.7 DAPI working solution (A.3.3).
4.4.8 Phosphate buffered saline (PBS) (A.2.1).
4.4.9 Non-fluorescing immersion oil.
Store at room temperature (20 ± 5) °C.
4.4.10 Stock suspensions of Cryptosporidium parvum oocysts and Giardia lamblia cysts.
© ISO 2006 – All rights reserved 3
---------------------- Page: 8 ----------------------
ISO 15553:2006(E)
Store at (5 ± 3) °C, never allow the suspension to freeze and check quality regularly. Ideally, suspensions of
oocysts and cysts should be no more than 3 months old. Stock suspensions should be checked
microscopically to confirm that they are monodispersed and discarded if clumps or aggregates are detected.
In addition, if mAb and DAPI staining become weak and oocysts become deformed, they should also be
discarded.
4.4.11 Parasite storage medium (A.3.4).
5 Apparatus
Use usual laboratory equipment and, in particular, the following.
TM 2)
5.1 Scientific apparatus, required for concentration using Pall Envirochek STD or HV.
TM
5.1.1 Sampling capsule, Envirochek STD or HV (Pall).
5.1.2 Peristaltic pump, capable of a flow rate of 2 l/min.
5.1.3 Silicon tubing, for use with the peristaltic pump.
5.1.4 Seeding container, 10 l, if seeding filters is required.
TM
5.1.5 Wrist-action shaker, with arms for the agitation of the Envirochek STD or HV sample capsules.
5.1.6 Centrifuge, capable of a minimum of 1 100 g.
5.1.7 Centrifuge tubes, conical, plastic, screwtop, 250 ml capacity.
5.1.8 Centrifuge tubes, conical, plastic, screwtop, 50 ml capacity.
NOTE A flow meter and flow restrictor are required for taking water samples with the filter.
® 2)
5.2 Specific apparatus, required for concentration using IDEXX Filta-Max .
®
5.2.1 Sampling housing, Idexx Filta-Max .
®
5.2.2 Sampling module, Idexx Filta-Max .
®
5.2.3 Filter membranes, Idexx Filta-Max .
5.2.4 Laboratory pump, capable of supplying 500 kPa (5 bar) pressure.
5.2.5 Peristaltic pump, capable of flow rate of 4 l/min.
5.2.6 Silicon tubing, for use with peristaltic pump.
5.2.7 Seeding container, 10 l, if seeding filters is required.
®
5.2.8 Wash station, automatic or manual, and wash station clamp set, Idexx Filta-Max .
5.2.9 Vacuum set, includes plastic hand pump, waste bottle, tubing and magnetic stirring bar. Idexx
®
Filta-Max .
2) All apparatus are examples of suitable products available commercially. This information is given for the convenience
of users of this International Standard and does not constitute an endorsement by ISO of these products.
4 © ISO 2006 – All rights reserved
---------------------- Page: 9 ----------------------
ISO 15553:2006(E)
5.2.10 Tubing set, includes elution tube, and middle section, concentrator tube and base, with line tap and
®
steel rod Idexx Filta-Max .
5.2.11 Membrane, for tubing set.
5.2.12 Plastic bag, for washing membrane.
5.2.13 Centrifuge, capable of 1 100 g.
5.2.14 Centrifuge tubes, conical, plastic, 50 ml capacity.
5.2.15 Forceps.
NOTE A flow meter and flow restrictor are required for taking water samples with the filter.
2)
5.3 General apparatus .
5.3.1 Incubator, at (36 ± 2) °C.
5.3.2 Refrigerator, at (5 ± 3) °C.
5.3.3 Magnetic stirrer, and magnetic stirring bars.
5.3.4 Vortex mixer.
5.3.5 Wash bottles, polypropylene, 250 ml.
5.3.6 Calibrated micropipettes, adjustable: 1 µl to 10 µl with 1 µl to 10 µl tips; 20 µl to 200 µl with 10 µl to
200 µl tips; 200 µl to 1 000 µl with 100 µl to 1 000 µl tips.
5.3.7 pH meter.
5.3.8 Magnetic particle concentrators, with suitable tubes.
5.3.9 Well microscope slides, with special hydrophobic coating and coverslips.
5.3.10 Epifluorescence microscope, with a UV filter (350 nm excitation, 450 nm emission), FITC filter
TM
(480 nm excitation, 520 nm emission) filters, differential interference contrast (DIC) optics and an eye piece
graticule. Total magnification 1 000 ×.
5.3.11 Microscope stage micrometer, 1 mm, ruled in 100 units.
5.3.12 Eyepiece graticule, ruled in 100 units.
5.3.13 Humidity chamber, e.g. consisting of a tightly sealed plastic container containing damp paper towels
on which the slides are placed.
5.3.14 10 l containers, graduated in 1 l.
5.3.15 Neubauer haemocytometer slide.
© ISO 2006 – All rights reserved 5
---------------------- Page: 10 ----------------------
ISO 15553:2006(E)
6 Sampling and transport
The size of the sample is dependent on the type of water being sampled, the purpose of the analysis, the
sensitivity to be achieved and the speed with which the result is required. Small volume samples (10 l to 100 l)
can be collected in the field and transported and analysed quickly whereas large volume samples (1 000 l)
require on-site filtration. Filtration may take up to 24 h because of a restricted flow rate through the filter. Small
volume samples (10 l) will give an indication of water quality at the time of sampling whereas large volumes
(1 000 l) will give an indication of water quality over an extended period. Since the concentration of
Cryptosporidium and Giardia is usually very low, large test volumes (10 l to 1 000 l) are required. The volume
to be examined may be dictated by regulatory limits.
For large volume filtration, connect the device in-line with the water supply, making sure that the flow through
the filter is in the direction indicated on the housing by the manufacturer. A flow meter should be included with
the filter and this should be read before and after sampling. If the filter is to be transported to the laboratory, it
should be sealed, after sampling, with end caps provided by the manufacturer. Due care shall be taken to
ensure that the flow rates do not exceed those recommended by the manufacturers of the filtration devices.
Take small volume grab samples and transport them to the laboratory in the dark at ambient temperature.
Once at the laboratory, samples should be stored at (5 ± 3) °C unless they are to be analysed immediately.
Samples should be analysed preferably within 24 h of collection and no longer than 4 d.
If the samples are filtered in the field, transport the filters in the dark at ambient temperature. Once at the
laboratory, samples should be stored at (5 ± 3) °C unless they are to be analysed immediately. Samples
should be analysed preferably within 24 h of collection and no longer than 4 d. If filters are stored at (5 ± 3) °C,
they shall be allowed to warm to room temperature before elution starts.
No information is available to date on the behaviour of Cryptosporidium and Giardia during sample or filter
storage. It is therefore advisable to examine the samples as soon after sampling as possible.
A pre-treatment step using sodium polyphosphate before the elution buffer was introduced to improve the
removal of particulate material bound to the filter.
TM
NOTE 1 The Envirochek STD filter consists of a pleated polyether sulfone membrane sealed in a polycarbonate
shell. The filter is supported on a loose polypropylene support. It is supplied packaged with two end caps which can be
used to seal the filter. The filter can be connected to a water supply by connecting to a ribbed inlet and the direction of flow
through the filter is clearly marked. The flow through the filter should not exceed 2,3 l/min and the differential pressure
across the filter should not exceed 210 kPa (2,1 bar).
TM
NOTE 2 The Envirochek HV capsule is comprised of 1 µm pore size polyester track-etched membrane permanently
enclosed in a polycarbonate housing. The polyester membrane is directly laminated to a polypropylene support which
TM
offers a significant strength improvement over the standard Envirochek STD. The capsule housing burst strength
exceeds 1 000 kPa (10 bar) and the differential pressure across the filter membrane is rated to 410 kPa (4,1 bar). Each
TM
Envirochek HV capsule is 100 % integrity tested after assembly to ensure product performance. The effective filtration
TM 2
area of the Envirochek HV is 1 300 cm . The filter is supplied with two end caps which can be used to seal the filter for
transport to the laboratory. The filter can be supplied with a tamper evident label containing a unique identification number.
TM
The flow through the Envirochek HV should not exceed 3,4 l/min.
®
NOTE 3 The Filta-Max filter consists of a foam filter module comprising 60 open cell reticulated foam discs with an
external diameter of 55 mm and an internal diameter of 15 mm. The discs are sandwiched between two retaining plates
and compressed by tightening a retaining bolt to give a nominal porosity of 1 µm. The filter module fits into a filter housing
which has a screw top and seal. The filter housing has stainless steel barbed inlet and outlet ports. The sample enters
through the lid of the housing and exits through the base. Water flows into the housing, through the compressed foam
rings into the centre of the module and through the outlet port. Removal of the retaining bolt during the elution stage allows
the filter to expand during washing. Filter housings are supplied with two tools for the removal of the top and two rubber
®
bungs to seal water in the sample. After sampling, Filta-Max should be kept wet during storage and transportation. If
stored or transported in the filter housing, the inlet and outlet should be securely plugged with the rubber stoppers
provided. During transportation or storage, the filter module may be removed from the housing and aseptically placed in
an airtight container along with several milliliters of additional deionized water.
6 © ISO 2006 – All rights reserved
---------------------- Page: 11 ----------------------
ISO 15553:2006(E)
7 Procedure
7.1 Concentration
TM
7.1.1 Pall Envirochek STD Filtration
Support the filter vertically with the white bleed valve uppermost. Remove the two end caps and allow any
water in the sample to drain out through the filter. Replace the bottom end cap, fill the cartridge with elution
buffer (4.1.6) through the inlet fitting until it covers the filter by approximately 1 cm. Replace the upper end cap
and secure the cartridge horizontally into the wrist shaker (5.1.5) with the white bleed valve in the 12 o’clock
position. Shake at 600 cycles per minute (cpm) ± 25 cpm for 5 min ± 30 s.
Remove the upper end cap and pour the washings into a 250 ml conical centrifuge tube (5.1.7). Add a further
aliquot of elution buffer into the capsule, replace the upper end cap and shake for a further 5 min ± 30 s.
Ensure that the white bleed valve is in the 3 o’clock or 9 o’clock position.
After 5 min of shaking, remove the upper end cap and decant the washings into the 250 ml centrifuge tube
and centrifuge at 1 100 × g for 15 min without braking during the deceleration phase. Record the pellet volume
(volume of solids) immediately after centrifugation.
A second centrifugation step may be required in a 50 ml centrifuge tube in order to measure the volume.
Alternatively, 50 ml centrifuge tubes may be used to concentrate the particulate material eluted from the filter.
Using a pipette and a vacuum source of less than 20 kPa (0,2 bar), carefully aspirate off the supernatant
leaving 2 ml to 5 ml above the pellet. If no pellet is visible, extra care shall taken to avoid aspirating oocysts
and cysts during this step.
Add deionized water to the centrifuge tube to bring the total volume to 9 ml. Vortex the tube for 10 s to 15 s to
resuspend the pellet and either store the sample at (5 ± 3) °C for future IMS or proceed directly to 7.2.
If the pellet volume exceeds that recommended by the manufacturer of the IMS test kit, centrifuge the sample
a second time in a tube that permits the pellet volume to be measured accurately. Subdivide the sample into
aliquots for IMS such that each aliquot represents the maximum pellet volume recommended by the
manufacturer and make up each aliquot to 9 ml with deionized water.
TM
7.1.2 Pall Envirochek HV Filtration
Support the filter vertically with the white bleed valve uppermost. Remove the two end caps and allow any
water in the sample to drain out through the filter. Replace the bottom end cap, fill the cartridge with
pre-treatment buffer (4.2.2) through the inlet fitting until it covers the filter by approximately 1 cm. Replace the
upper end cap and secure the cartridge horizontally into the wrist shaker (5.1.5) with the white bleed valve in
the vertical position. Shake at 600 cycles per minute (cpm) ± 25 cpm for 5 min ± 30 s.
Secure the filter vertically with the white bleed valve uppermost, remove the end caps and allow the
pre-treatment buffer to drain out through the filter. Replace the bottom end cap and fill the cartridge as above
with deionized water (4.2.1). Replace the upper end cap and rinse the membrane by gently rotating the filter
for 30 s. Secure the filter vertically, remove the end caps and allow the deionized water to drain out through
the filter.
Replace the bottom end cap, fill the cartridge with elution buffer (4.2.7) through the inlet fitting until it covers
the filter by approximately 1 cm. Replace the upper end cap and secure the cartridge into the wrist shaker
(5.1.5) with the white bleed valve in the 12 o’clock position. Shake at 600 cpm ± 25 cpm for 5 min ± 30 s.
Remove the upper end cap and pour the washings into a 250 ml conical centrifuge tube (5.1.7). Add a further
aliquot of elution buffer into the capsule, replace the upper end cap and shake for a further 5 min ± 30 s.
Ensure that the white bleed valve is in the 4 o’clock position. After 5 min of shaking, turn the filter such that the
white valve is in the 8 o’clock position and shake for a further 5 min.
© ISO 2006 – All rights reserved 7
---------------------- Page: 12 ----------------------
ISO 15553:2006(E)
Remove the upper end cap and decant the washings into the 250 ml centrifuge tube and centrifuge at 1 100 g
for 15 min without braking during the deceleration phase. Record the pellet volume (volume of solids)
immediately after centrifugation.
A second centrifugation step may be required in a 50 ml centrifuge tube in order to measure the volume.
Alternatively, 50 ml centrifuge tubes may be used to concentrate the particulate material eluted from the filter.
Using a pipette and a vacuum source of less than 20 kPa (0,2 bar), carefully aspirate off the supernatant
leaving 2 ml to 5 ml above the pellet. If no pellet is visible, extra care shall be taken to avoid aspirating oocysts
and cysts during this step.
Add deionized water to the centrifuge tube to bring the total volume to 9 ml. Vortex the tube for 10 s to 15 s to
re-suspend the pellet and either store the sample at (5 ± 3) °C for IMS or proceed directly to 7.2.
If the pellet volume exceeds that recommended by the manufacturer of the IMS test kit, centrifuge the sample
a second time in a tube that permits the pellet volume to be measured accurately. Subdivide the sample into
aliquots for IMS such that each aliquot represents the maximum pellet volume recommended by the
manufacturer and make up each aliquot to 9 ml with deionized water.
NOTE Warming all the elution solutions to 37 °C improves the removal of particulate material. Elution is also helped
by increasing the shaking speed to 900 cpm.
®
7.1.3 Idexx Filta-Max Filtration
The elution apparatus consists of an upper and lower wash tube (5.2.10), a wash station (5.2.8) and a vacuum
set (5.2.9) designed to
...
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Kakovost vode - Izolacija in identifikacija oocist Cryptosporidium in cist GiardiaQualité de l'eau - Isolement et identification des oocystes de Cryptosporidium et des kystes de GiardiaWater quality - Isolation and identification of Cryptosporidium oocysts and Giardia cysts from water07.100.20Mikrobiologija vodeMicrobiology of waterICS:Ta slovenski standard je istoveten z:ISO 15553:2006SIST ISO 15553:2010en01-september-2010SIST ISO 15553:2010SLOVENSKI
STANDARD
SIST ISO 15553:2010
Reference numberISO 15553:2006(E)© ISO 2006
INTERNATIONAL STANDARD ISO15553First edition2006-11-15Water quality — Isolation and identification of Cryptosporidium oocysts and Giardia cysts from water Qualité de l'eau — Isolement et identification des oocystes de Cryptosporidium et des kystes de Giardia SIST ISO 15553:2010
ISO 15553:2006(E) PDF disclaimer This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat accepts no liability in this area. Adobe is a trademark of Adobe Systems Incorporated. Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
©
ISO 2006 All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO's member body in the country of the requester. ISO copyright office Case postale 56 • CH-1211 Geneva 20 Tel.
+ 41 22 749 01 11 Fax
+ 41 22 749 09 47 E-mail
copyright@iso.org Web
www.iso.org Published in Switzerland
ii © ISO 2006 – All rights reserved
SIST ISO 15553:2010
ISO 15553:2006(E) © ISO 2006 – All rights reserved iiiContents Page Foreword.iv Introduction.v 1 Scope.1 2 Terms and definitions.1 3 Principle.1 4 Reagents.2 5 Apparatus.4 6 Sampling and transport.6 7 Procedure.7 8 Quality control procedures.14 9 Reporting of results.14 Annex A (normative)
Preparation of reagents.16 Annex B (informative)
Concentration of oocysts and cysts from small (10 l) volumes of water.19 Annex C (informative)
Calibration of eyepiece graticule.24 Annex D (informative)
Preparation of positive controls and recovery tests.25 Annex E (informative)
Examples of indicative performance data.28 Annex F (informative)
Alternative methods.30 Annex G (informative)
Further information about Cryptosporidium and Giardia.31 Annex H (informative)
Details of manufacturers.32 Bibliography.36
SIST ISO 15553:2010
ISO 15553:2006(E) iv © ISO 2006 – All rights reserved Foreword ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2. The main task of technical committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to the member bodies for voting. Publication as an International Standard requires approval by at least 75 % of the member bodies casting a vote. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent rights. ISO 15553 was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 4, Microbiological methods. SIST ISO 15553:2010
ISO 15553:2006(E) © ISO 2006 – All rights reserved vIntroduction Cryptosporidium and Giardia are protozoan parasites that can cause enteric illness in humans. Both organisms are characterized by an ability to survive in the aquatic environment. Cryptosporidium in particular is resistant to chlorine at the concentrations used in the treatment of drinking and swimming pool waters. Consequently the absence of vegetative bacteria as indicators of faecal contamination does not necessarily indicate the absence of Cryptosporidium oocysts or Giardia cysts. The methods described in this document may be used to determine whether Cryptosporidium and/or Giardia are present in water supplies. The techniques have been selected on the basis of method development and peer review publication of the data thus obtained. They are further selected to give comparable recoveries of the methods or reagents used in the isolation of the organisms.
SIST ISO 15553:2010
SIST ISO 15553:2010
INTERNATIONAL STANDARD ISO 15553:2006(E) © ISO 2006 – All rights reserved 1Water quality — Isolation and identification of Cryptosporidium oocysts and Giardia cysts from water 1 Scope This International Standard specifies a method that is applicable for the detection and enumeration of Cryptosporidium oocysts and Giardia cysts in water. It is applicable for the examination of surface and ground waters, treated waters, mineral waters, swimming pool and recreational waters. This method does not allow identification to species level, the host species of origin or the determination of viability or infectivity of any Cryptosporidium oocyst or Giardia cyst which may be present. These procedures are for use by experienced analysts who have successfully completed competency tests prior to commencing analysis. In addition, such analysts should continue to demonstrate competency by examining seeded samples at regular intervals and taking part in external quality assurance schemes. NOTE Bodies resembling Cryptosporidium or Giardia in morphology can be present and these may be mistaken for oocysts or cysts. Results should be interpreted with care. Where there is doubt about the identity of oocysts or cysts or where an unusually high result is obtained, it is advisable to have the slides examined by experts from other laboratories to confirm or refute the findings. 2 Terms and definitions For the purposes of this document, the following terms and definitions apply. 2.1 Cryptosporidium protozoan parasite, concentrated and selected from water samples with the methods described, which reacts with specific anti-Cryptosporidium antibodies and exhibits the typical morphological characteristics described in 7.4 of this International Standard NOTE A more complete definition of the parasite and the different genotypes and species is given in Annex G. 2.2 Giardia protozoan parasite, concentrated and selected from water samples with the methods described, which reacts with specific anti-Giardia antibodies and exhibits the typical morphological characteristics described in 7.4 of this International Standard NOTE A more complete definition of the parasite and the different species is given in Annex G. 3 Principle 3.1 Concentration from water The isolation of Cryptosporidium and Giardia from water requires the use of a procedure which allows the volume of the sample to be reduced whilst retaining any oocysts and cysts. The concentration procedure used however, is dependent upon the water type which is to be analysed, the volume of sample and the amount of particulate material in the sample. This document describes the use of two concentration techniques for varying volumes of water using cartridge filtration and elution followed by low speed centrifugation (7.1). Additional methods for the recovery of oocysts and cysts from small volumes of water or very turbid waters are given in Annex B. Some examples of recovery data for these techniques are given in Annex E. SIST ISO 15553:2010
ISO 15553:2006(E) 2 © ISO 2006 – All rights reserved Table 1 — Membrane filters/filtration systems used for the concentration of parasites from water samples Membrane filter/filtration system Application Pall EnvirochekTM STD a Concentration of 10-litre to 200-litre (or more) samples of water Pall EnvirochekTM HV Concentration of 10-litre to 1 000-litre samples of water IDEXX Filta-Max® Concentration of 10-litre to 1 000-litre samples of water a It has been shown by some laboratories that this technique may be used successfully for larger volumes of water although the manufacturers’ instructions may only include volumes up to 200 litres.
3.2 Purification and further concentration After concentration of particulate material from filter eluates, oocysts and cysts are isolated using immunomagnetic separation (IMS) (7.2). Oocysts and cysts are attached to para-magnetic beads coated with specific antibody, the beads are separated from the unwanted particulate material using a magnet and then the oocysts and cysts are dissociated from the beads using acid and neutralized using alkali before immunostaining. 3.3 Detection of Cryptosporidium and Giardia After IMS, organisms are labelled with monoclonal antibody (mAb) conjugated to a fluorochrome, usually fluoroscein isothiocyanate (FITC). In addition, any nuclear material is labelled with a nucleic acid stain to aid identification (7.3). Each sample is then examined for the presence of labelled Cryptosporidium oocysts and Giardia cysts using epifluorescence and differential interference contrast (DIC) microscopy (7.4). 4 Reagents 4.1 Reagents required for eluting Pall EnvirochekTM STD capsule filters 1) 4.1.1 Deionized water, 0,2 µm filtered at the point of use. 4.1.2 Laureth 12 detergent. 4.1.3 Tris buffer, pH 7,4 (A.1.1). 4.1.4 EDTA solution, 0,5 mol/l, pH 8,0 (A.1.2). 4.1.5 Antifoam A. 4.1.6 Elution buffer (A.1.3). 4.2 Reagents required for eluting Pall EnvirochekTM HV capsule filters 1) 4.2.1 Deionized water, 0,2 µm filtered at point of use. 4.2.2 Pre-treatment buffer (A.1.4). 4.2.3 Laureth 12 detergent
1) All products and reagents are examples of suitable products available commercially. This information is given for the convenience of users of this International Standard and does not constitute an endorsement by ISO of these products. SIST ISO 15553:2010
ISO 15553:2006(E) © ISO 2006 – All rights reserved 34.2.4 Tris buffer, pH 7,4 (A.1.1). 4.2.5 EDTA solution, 0,5 mol/l, pH 8,0 (A.1.2). 4.2.6 Antifoam A. 4.2.7 Elution buffer (A.1.3). 4.3 Reagents required for eluting IDEXX Filta-Max® filters 1) 4.3.1 Phosphate buffered saline (PBS) (A.2.1). 4.3.2 Polyoxyethylene(20)sorbitan monolaurate (Tween 20). Store at room temperature (20 ± 5) °C. Expiry date one year. 4.3.3 Elution buffer (A.2.2). 4.4 Concentration and detection reagents 4.4.1 Methanol, analytical grade. 4.4.2 Magnetic beads, for the detection of Cryptosporidium and Giardia. Expiry date printed by the manufacturer. NOTE See Annex H for a list of suitable suppliers. 4.4.3 Fluorescently labelled monoclonal antibodies (mAbs) against Cryptosporidium and Giardia. Store at (5 ± 3) °C. Expiry date as stated by the manufacturer. When stains are prepared from concentrated material using a diluent supplied by the manufacturer, the prepared solution is stored at (5 ± 3) °C for no longer than 6 months. NOTE See Annex H for a list of suitable suppliers. 4.4.4 Immunofluorescence mounting medium (A.3.1). NOTE See Annex H for a list of suitable suppliers. 4.4.5 4,6-Diamidino-2-phenylindole dihydrochloride dihydrate (DAPI) freeze dried reagent. Store according to the manufacturer's instructions. Expiry date printed by the manufacturer on each vial. 4.4.6 DAPI stock solution (A.3.2). 4.4.7 DAPI working solution (A.3.3). 4.4.8 Phosphate buffered saline (PBS) (A.2.1). 4.4.9 Non-fluorescing immersion oil. Store at room temperature (20 ± 5) °C. 4.4.10 Stock suspensions of Cryptosporidium parvum oocysts and Giardia lamblia cysts. SIST ISO 15553:2010
ISO 15553:2006(E) 4 © ISO 2006 – All rights reserved Store at (5 ± 3) °C, never allow the suspension to freeze and check quality regularly. Ideally, suspensions of oocysts and cysts should be no more than 3 months old. Stock suspensions should be checked microscopically to confirm that they are monodispersed and discarded if clumps or aggregates are detected. In addition, if mAb and DAPI staining become weak and oocysts become deformed, they should also be discarded. 4.4.11 Parasite storage medium (A.3.4). 5 Apparatus Use usual laboratory equipment and, in particular, the following. 5.1 Scientific apparatus, required for concentration using Pall EnvirochekTM STD or HV. 2) 5.1.1 Sampling capsule, EnvirochekTM STD or HV (Pall). 5.1.2 Peristaltic pump, capable of a flow rate of 2 l/min. 5.1.3 Silicon tubing, for use with the peristaltic pump. 5.1.4 Seeding container, 10 l, if seeding filters is required. 5.1.5 Wrist-action shaker, with arms for the agitation of the EnvirochekTM STD or HV sample capsules. 5.1.6 Centrifuge, capable of a minimum of 1 100 g. 5.1.7 Centrifuge tubes, conical, plastic, screwtop, 250 ml capacity. 5.1.8 Centrifuge tubes, conical, plastic, screwtop, 50 ml capacity. NOTE A flow meter and flow restrictor are required for taking water samples with the filter. 5.2 Specific apparatus, required for concentration using IDEXX Filta-Max®. 2) 5.2.1 Sampling housing, Idexx Filta-Max®. 5.2.2 Sampling module, Idexx Filta-Max®. 5.2.3 Filter membranes, Idexx Filta-Max®. 5.2.4 Laboratory pump, capable of supplying 500 kPa (5 bar) pressure. 5.2.5 Peristaltic pump, capable of flow rate of 4 l/min. 5.2.6 Silicon tubing, for use with peristaltic pump. 5.2.7 Seeding container, 10 l, if seeding filters is required. 5.2.8 Wash station, automatic or manual, and wash station clamp set, Idexx Filta-Max®. 5.2.9 Vacuum set, includes plastic hand pump, waste bottle, tubing and magnetic stirring bar. Idexx Filta-Max®.
2) All apparatus are examples of suitable products available commercially. This information is given for the convenience of users of this International Standard and does not constitute an endorsement by ISO of these products. SIST ISO 15553:2010
ISO 15553:2006(E) © ISO 2006 – All rights reserved 55.2.10 Tubing set, includes elution tube, and middle section, concentrator tube and base, with line tap and steel rod Idexx Filta-Max®. 5.2.11 Membrane, for tubing set. 5.2.12 Plastic bag, for washing membrane. 5.2.13 Centrifuge, capable of 1 100 g. 5.2.14 Centrifuge tubes, conical, plastic, 50 ml capacity. 5.2.15 Forceps. NOTE A flow meter and flow restrictor are required for taking water samples with the filter. 5.3 General apparatus 2). 5.3.1 Incubator, at (36 ± 2) °C. 5.3.2 Refrigerator, at (5 ± 3) °C. 5.3.3 Magnetic stirrer, and magnetic stirring bars. 5.3.4 Vortex mixer. 5.3.5 Wash bottles, polypropylene, 250 ml. 5.3.6 Calibrated micropipettes, adjustable: 1 µl to 10 µl with 1 µl to 10 µl tips; 20 µl to 200 µl with 10 µl to 200 µl tips; 200 µl to 1 000 µl with 100 µl to 1 000 µl tips. 5.3.7 pH meter. 5.3.8 Magnetic particle concentrators, with suitable tubes. 5.3.9 Well microscope slides, with special hydrophobic coating and coverslips. 5.3.10 Epifluorescence microscope, with a UV filter (350 nm excitation, 450 nm emission), FITC filter (480 nm excitation, 520 nm emission) filters,TM differential interference contrast (DIC) optics and an eye piece graticule. Total magnification 1 000 ×. 5.3.11 Microscope stage micrometer, 1 mm, ruled in 100 units. 5.3.12 Eyepiece graticule, ruled in 100 units. 5.3.13 Humidity chamber, e.g. consisting of a tightly sealed plastic container containing damp paper towels on which the slides are placed. 5.3.14 10 l containers, graduated in 1 l. 5.3.15 Neubauer haemocytometer slide. SIST ISO 15553:2010
ISO 15553:2006(E) 6 © ISO 2006 – All rights reserved 6 Sampling and transport The size of the sample is dependent on the type of water being sampled, the purpose of the analysis, the sensitivity to be achieved and the speed with which the result is required. Small volume samples (10 l to 100 l) can be collected in the field and transported and analysed quickly whereas large volume samples (1 000 l) require on-site filtration. Filtration may take up to 24 h because of a restricted flow rate through the filter. Small volume samples (10 l) will give an indication of water quality at the time of sampling whereas large volumes (1 000 l) will give an indication of water quality over an extended period. Since the concentration of Cryptosporidium and Giardia is usually very low, large test volumes (10 l to 1 000 l) are required. The volume to be examined may be dictated by regulatory limits. For large volume filtration, connect the device in-line with the water supply, making sure that the flow through the filter is in the direction indicated on the housing by the manufacturer. A flow meter should be included with the filter and this should be read before and after sampling. If the filter is to be transported to the laboratory, it should be sealed, after sampling, with end caps provided by the manufacturer. Due care shall be taken to ensure that the flow rates do not exceed those recommended by the manufacturers of the filtration devices. Take small volume grab samples and transport them to the laboratory in the dark at ambient temperature. Once at the laboratory, samples should be stored at (5 ± 3) °C unless they are to be analysed immediately. Samples should be analysed preferably within 24 h of collection and no longer than 4 d. If the samples are filtered in the field, transport the filters in the dark at ambient temperature. Once at the laboratory, samples should be stored at (5 ± 3) °C unless they are to be analysed immediately. Samples should be analysed preferably within 24 h of collection and no longer than 4 d. If filters are stored at (5 ± 3) °C, they shall be allowed to warm to room temperature before elution starts. No information is available to date on the behaviour of Cryptosporidium and Giardia during sample or filter storage. It is therefore advisable to examine the samples as soon after sampling as possible. A pre-treatment step using sodium polyphosphate before the elution buffer was introduced to improve the removal of particulate material bound to the filter. NOTE 1 The EnvirochekTM STD filter consists of a pleated polyether sulfone membrane sealed in a polycarbonate shell. The filter is supported on a loose polypropylene support. It is supplied packaged with two end caps which can be used to seal the filter. The filter can be connected to a water supply by connecting to a ribbed inlet and the direction of flow through the filter is clearly marked. The flow through the filter should not exceed 2,3 l/min and the differential pressure across the filter should not exceed 210 kPa (2,1 bar). NOTE 2 The EnvirochekTM HV capsule is comprised of 1 µm pore size polyester track-etched membrane permanently enclosed in a polycarbonate housing. The polyester membrane is directly laminated to a polypropylene support which offers a significant strength improvement over the standard EnvirochekTM STD. The capsule housing burst strength exceeds 1 000 kPa (10 bar) and the differential pressure across the filter membrane is rated to 410 kPa (4,1 bar). Each EnvirochekTM HV capsule is 100 % integrity tested after assembly to ensure product performance. The effective filtration area of the EnvirochekTM HV is 1 300 cm2. The filter is supplied with two end caps which can be used to seal the filter for transport to the laboratory. The filter can be supplied with a tamper evident label containing a unique identification number. The flow through the EnvirochekTM HV should not exceed 3,4 l/min. NOTE 3 The Filta-Max® filter consists of a foam filter module comprising 60 open cell reticulated foam discs with an external diameter of 55 mm and an internal diameter of 15 mm. The discs are sandwiched between two retaining plates and compressed by tightening a retaining bolt to give a nominal porosity of 1 µm. The filter module fits into a filter housing which has a screw top and seal. The filter housing has stainless steel barbed inlet and outlet ports. The sample enters through the lid of the housing and exits through the base. Water flows into the housing, through the compressed foam rings into the centre of the module and through the outlet port. Removal of the retaining bolt during the elution stage allows the filter to expand during washing. Filter housings are supplied with two tools for the removal of the top and two rubber bungs to seal water in the sample. After sampling, Filta-Max® should be kept wet during storage and transportation. If stored or transported in the filter housing, the inlet and outlet should be securely plugged with the rubber stoppers provided. During transportation or storage, the filter module may be removed from the housing and aseptically placed in an airtight container along with several milliliters of additional deionized water. SIST ISO 15553:2010
ISO 15553:2006(E) © ISO 2006 – All rights reserved 77 Procedure 7.1 Concentration 7.1.1 Pall EnvirochekTM STD Filtration Support the filter vertically with the white bleed valve uppermost. Remove the two end caps and allow any water in the sample to drain out through the filter. Replace the bottom end cap, fill the cartridge with elution buffer (4.1.6) through the inlet fitting until it covers the filter by approximately 1 cm. Replace the upper end cap and secure the cartridge horizontally into the wrist shaker (5.1.5) with the white bleed valve in the 12 o’clock position. Shake at 600 cycles per minute (cpm) ± 25 cpm for 5 min ± 30 s. Remove the upper end cap and pour the washings into a 250 ml conical centrifuge tube (5.1.7). Add a further aliquot of elution buffer into the capsule, replace the upper end cap and shake for a further 5 min ± 30 s. Ensure that the white bleed valve is in the 3 o’clock or 9 o’clock position. After 5 min of shaking, remove the upper end cap and decant the washings into the 250 ml centrifuge tube and centrifuge at 1 100 × g for 15 min without braking during the deceleration phase. Record the pellet volume (volume of solids) immediately after centrifugation. A second centrifugation step may be required in a 50 ml centrifuge tube in order to measure the volume. Alternatively, 50 ml centrifuge tubes may be used to concentrate the particulate material eluted from the filter. Using a pipette and a vacuum source of less than 20 kPa (0,2 bar), carefully aspirate off the supernatant leaving 2 ml to 5 ml above the pellet. If no pellet is visible, extra care shall taken to avoid aspirating oocysts and cysts during this step. Add deionized water to the centrifuge tube to bring the total volume to 9 ml. Vortex the tube for 10 s to 15 s to resuspend the pellet and either store the sample at (5 ± 3) °C for future IMS or proceed directly to 7.2. If the pellet volume exceeds that recommended by the manufacturer of the IMS test kit, centrifuge the sample a second time in a tube that permits the pellet volume to be measured accurately. Subdivide the sample into aliquots for IMS such that each aliquot represents the maximum pellet volume recommended by the manufacturer and make up each aliquot to 9 ml with deionized water. 7.1.2 Pall EnvirochekTM HV Filtration Support the filter vertically with the white bleed valve uppermost. Remove the two end caps and allow any water in the sample to drain out through the filter. Replace the bottom end cap, fill the cartridge with pre-treatment buffer (4.2.2) through the inlet fitting until it covers the filter by approximately 1 cm. Replace the upper end cap and secure the cartridge horizontally into the wrist shaker (5.1.5) with the white bleed valve in the vertical position. Shake at 600 cycles per minute (cpm) ± 25 cpm for 5 min ± 30 s. Secure the filter vertically with the white bleed valve uppermost, remove the end caps and allow the pre-treatment buffer to drain out through the filter. Replace the bottom end cap and fill the cartridge as above with deionized water (4.2.1). Replace the upper end cap and rinse the membrane by gently rotating the filter for 30 s. Secure the filter vertically, remove the end caps and allow the deionized water to drain out through the filter. Replace the bottom end cap, fill the cartridge with elution buffer (4.2.7) through the inlet fitting until it covers the filter by approximately 1 cm. Replace the upper end cap and secure the cartridge into the wrist shaker (5.1.5) with the white bleed valve in the 12 o’clock position. Shake at 600 cpm ± 25 cpm for 5 min ± 30 s. Remove the upper end cap and pour the washings into a 250 ml conical centrifuge tube (5.1.7). Add a further aliquot of elution buffer into the capsule, replace the upper end cap and shake for a further 5 min ± 30 s. Ensure that the white bleed valve is in the 4 o’clock position. After 5 min of shaking, turn the filter such that the white valve is in the 8 o’clock position and shake for a further 5 min. SIST ISO 15553:2010
ISO 15553:2006(E) 8 © ISO 2006 – All rights reserved Remove the upper end cap and decant the washings into the 250 ml centrifuge tube and centrifuge at 1 100 g for 15 min without braking during the deceleration phase. Record the pellet volume (volume of solids) immediately after centrifugation. A second centrifugation step may be required in a 50 ml centrifuge tube in order to measure the volume. Alternatively, 50 ml centrifuge tubes may be used to concentrate the particulate material eluted from the filter. Using a pipette and a vacuum source of less than 20 kPa (0,2 bar), carefully aspirate off the supernatant leaving 2 ml to 5 ml above the pellet. If no pellet is visible, extra care shall be taken to avoid aspirating oocysts and cysts during this step. Add deionized water to the centrifuge tube to bring the total volume to 9 ml. Vortex the tube for 10 s to 15 s to re-suspend the pellet and either store the sample at (5 ± 3) °C for IMS or proceed directly to 7.2. If the pellet volume exceeds that recommended by the manufacturer of the IMS test kit, centrifuge the sample a second time in a tube that permits the pellet volume to be measured accurately. Subdivide the sample into aliquots for IMS such that each aliquot represents the maximum pellet volume recommended by the manufacturer and make up each aliquot to 9 ml with deionized water. NOTE
Warming all the elution solutions to 37 °C improves the removal of particulate material. Elution is also helped by increasing the shaking speed to 900 cpm. 7.1.3 Idexx Filta-Max® Filtration The elution apparatus consists of an upper and lower wash tube (5.2.10), a wash station (5.2.8) and a vacuum set (5.2.9) designed to reduce the volume of the eluate to 50 ml through a membrane (5.2.11). The elution procedure is as follows: Place a membrane filter (rough surface uppermost) in the base of the lower wash tube and put the tube into the base. Make sure that the membrane is held securely and that the tap on the base is closed. Unscrew the housing top using the tools provided, rem
...