ISO/FDIS 15193
(Main)In vitro diagnostic medical devices — Requirements for reference measurement procedures
In vitro diagnostic medical devices — Requirements for reference measurement procedures
Dispositifs médicaux de diagnostic in vitro — Exigences relatives aux procédures de mesure de référence
General Information
- Status
- Not Published
- Current Stage
- 5000 - FDIS registered for formal approval
- Start Date
- 12-Jul-2024
- Completion Date
- 09-Jul-2024
Relations
- Effective Date
- 12-Feb-2026
- Consolidates
ISO/IEC 19566-4:2020 - Information technologies — JPEG systems — Part 4: Privacy and security - Effective Date
- 06-Jun-2022
- Effective Date
- 06-Jun-2022
Overview
ISO/FDIS 15193: In vitro diagnostic medical devices - Requirements for reference measurement procedures provides internationally recognized requirements for reference measurement procedures in the field of laboratory medicine. Developed by ISO Technical Committee 212, this standard is central to ensuring accuracy, reliability, and consistency in the results produced by in vitro diagnostic (IVD) medical devices. Reference measurement procedures enable the metrological traceability of laboratory results, building confidence in both clinical decision-making and regulatory compliance.
As healthcare systems and regulatory authorities increasingly rely on harmonized laboratory data, adopting standards such as ISO/FDIS 15193 becomes essential. This document is applicable to developers, manufacturers, and institutions involved in creating or verifying reference measurement procedures for clinical laboratory testing.
Key Topics
ISO/FDIS 15193 addresses essential topics and core requirements, including:
- Elements of Reference Measurement Procedures
Specifying technical descriptions for reagents, calibrators, apparatus, and sample preparation. - Validation and Quality Assurance
Defining clear procedures for validation, including analytical sensitivity, trueness, precision, limit of detection, and measurement uncertainty. - Metrological Traceability
Ensuring measurement results are traceable through a documented, unbroken chain of calibration. - Calibration and Reference Materials
Utilizing certified reference materials and calibrators for assigning accurate quantity values. - Data Processing and Reporting
Outlining calculations, validation steps, and standard reporting of laboratory measurement results. - Requirements for Documentation
Providing comprehensive technical documentation to enable reproducibility and transparency. - Interlaboratory Comparison
Detailing processes for comparison among laboratories to assess consistency of reference measurement procedures.
Applications
The practical value of ISO/FDIS 15193 spans a variety of settings within laboratory medicine and the IVD industry:
- Calibration of Diagnostic Devices
Using reference measurement procedures to calibrate and verify the performance of clinical diagnostic equipment. - Regulatory Compliance
Demonstrating conformity with regulatory bodies that require internationally accepted reference measurement procedures for IVD products. - Assignment of Reference Values
Assigning accurate values to calibrators and control materials, which in turn ensures comparability of patient test results across different platforms and laboratories. - Quality Management Systems
Integrating into laboratory quality assurance processes for ongoing system accreditation and proficiency assessment. - Research and Development
Supporting the development, validation, and comparison of new medial analyses and device technologies.
Implementing ISO/FDIS 15193 supports patient safety by reducing potential diagnostic errors caused by inconsistent laboratory results. Laboratories, manufacturers, and oversight bodies alike benefit from the strengthened foundation this standard provides for harmonized clinical measurements.
Related Standards
Organizations involved in laboratory medicine or in vitro diagnostic device manufacturing may also reference or implement these related standards for comprehensive quality and traceability:
- ISO 17511: In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples.
- ISO/IEC Guide 98-3: Uncertainty of measurement - Guide to the expression of uncertainty in measurement (GUM).
- ISO/IEC Guide 99: International vocabulary of metrology - Basic and general concepts and associated terms (VIM).
- ISO 15194: In vitro diagnostic medical devices - Requirements for certified reference materials and the content of supporting documentation.
- ISO 15195: Laboratory medicine - Requirements for reference measurement laboratories.
By adopting ISO/FDIS 15193 and related standards, laboratories and manufacturers contribute to a robust, interoperable, and internationally recognized framework for clinical laboratory measurement quality.
Buy Documents
ISO/FDIS 15193 - In vitro diagnostic medical devices — Requirements for reference measurement procedures/31/2023
ISO/FDIS 15193 - Dispositifs médicaux de diagnostic in vitro — Exigences relatives aux procédures de mesure de référence/5/2024
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
ISO/FDIS 15193 is a draft published by the International Organization for Standardization (ISO). Its full title is "In vitro diagnostic medical devices — Requirements for reference measurement procedures". This standard covers: In vitro diagnostic medical devices — Requirements for reference measurement procedures
In vitro diagnostic medical devices — Requirements for reference measurement procedures
ISO/FDIS 15193 is classified under the following ICS (International Classification for Standards) categories: 11.100.10 - In vitro diagnostic test systems. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO/FDIS 15193 has the following relationships with other standards: It is inter standard links to prEN ISO 15193, ISO/IEC 19566-4:2020, ISO 15193:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO/FDIS 15193 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
DRAFT INTERNATIONAL STANDARD
ISO/DIS 15193
ISO/TC 212 Secretariat: ANSI
Voting begins on: Voting terminates on:
2023-12-26 2024-03-19
In vitro diagnostic medical devices — Requirements for
reference measurement procedures
ICS: 11.100.10
This document is circulated as received from the committee secretariat.
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENT AND APPROVAL. IT IS
ISO/CEN PARALLEL PROCESSING
THEREFORE SUBJECT TO CHANGE AND MAY
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
ISO/DIS 15193:2023(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. © ISO 2023
ISO/DIS 15193:2023(E)
DRAFT INTERNATIONAL STANDARD
ISO/DIS 15193
ISO/TC 212 Secretariat: ANSI
Voting begins on: Voting terminates on:
In vitro diagnostic medical devices — Requirements for
reference measurement procedures
ICS: 11.100.10
This document is circulated as received from the committee secretariat.
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENT AND APPROVAL. IT IS
© ISO 2023
ISO/CEN PARALLEL PROCESSING
THEREFORE SUBJECT TO CHANGE AND MAY
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
NOT BE REFERRED TO AS AN INTERNATIONAL
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on STANDARD UNTIL PUBLISHED AS SUCH.
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
IN ADDITION TO THEIR EVALUATION AS
or ISO’s member body in the country of the requester. BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
ISO copyright office
USER PURPOSES, DRAFT INTERNATIONAL
CP 401 • Ch. de Blandonnet 8
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
CH-1214 Vernier, Geneva
POTENTIAL TO BECOME STANDARDS TO
Phone: +41 22 749 01 11
WHICH REFERENCE MAY BE MADE IN
Reference number
Email: copyright@iso.org
NATIONAL REGULATIONS.
Website: www.iso.org ISO/DIS 15193:2023(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
Published in Switzerland
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
ii
PROVIDE SUPPORTING DOCUMENTATION. © ISO 2023
ISO/DIS 15193:2023(E)
Contents Page
Foreword .v
Introduction . vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 Requirements for a reference measurement procedure .10
4.1 General . 10
4.2 Elements of a reference measurement procedure . 10
4.3 Warning and safety precautions . 11
4.4 Introduction . 11
4.5 Scope . .12
4.6 Terms, definitions, symbols, and abbreviated terms .12
4.6.1 Concepts .12
4.6.2 Nomenclature . 12
4.6.3 Trivial names .12
4.7 Measurement principle and measurement method .12
4.8 Reagents and materials .13
4.8.1 General .13
4.8.2 Descriptive items .13
4.8.3 Influence quantities . 14
4.8.4 Expression of concentration . 14
4.8.5 Expression of dilution . 14
4.8.6 Reference to patented items . 15
4.9 Apparatus and equipment .15
4.9.1 Description .15
4.9.2 Auxiliary equipment .15
4.10 Consumables . . . 15
4.11 Sampling and sample .15
4.11.1 General .15
4.11.2 Samples .15
4.12 Preparation of measuring system and analytical portion . 16
4.12.1 General . 16
4.12.2 Preparation of apparatus . 16
4.12.3 Calibration . . . 16
4.12.4 Types of analytical sample . 16
4.12.5 Design of analytical series . 17
4.12.6 Analytical portion . 17
4.12.7 Analytical solution . 17
4.13 Operation of measuring system . 17
4.13.1 Sequence of measurement steps . 17
4.13.2 Blank . 17
4.13.3 Validation of measurement conditions . 17
4.14 Data processing. 18
4.14.1 Calculation of measurement results . 18
4.14.2 Conversion equations . 18
4.15 Validation . 18
4.15.1 Concepts, values, and their use . . 18
4.15.2 Analytical calibration function . 19
4.15.3 Analytical measurement function . 19
4.15.4 Analytical sensitivity . 19
4.15.5 Analytical influence quantities . 19
4.15.6 Blank measurement . 19
4.15.7 Recovery studies . 19
iii
ISO/DIS 15193:2023(E)
4.15.8 Measurement trueness .20
4.15.9 Measurement precision .20
4.15.10 Repeatability standard deviation, s . 20
r
4.15.11 Intermediate precision standard deviation . 20
4.15.12 Reproducibility standard deviation, s . 20
R
4.15.13 Limit of detection . 20
4.15.14 Measurement interval . 20
4.15.15 Validation by using reference material . 21
4.15.16 Interlaboratory comparison . 21
4.15.17 Measurement uncertainty . 21
4.16 Special cases . 21
4.17 Reporting of measurement results. 22
4.18 Quality assurance . 22
4.19 Bibliography . 22
4.20 Dates of authorization and revision . 22
Annex A (informative) Reference procedures for properties other than differential and
rational quantities .23
Annex ZA (informative) Relationship between this European Standard the General Safety
and Performance Requirements of Regulation (EU) 2017/746 aimed to be covered .25
Bibliography .28
iv
ISO/DIS 15193:2023(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use
of (a) patent(s). ISO takes no position concerning the evidence, validity, or applicability of any claimed
patent rights in respect thereof. As of the date of publication of this document, ISO had not received
notice of (a) patent(s) which may be required to implement this document. However, implementers are
cautioned that this may not represent the latest information, which may be obtained from the patent
database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all
such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 212, Clinical laboratory testing and in
vitro diagnostic test systems.
This third edition cancels and replaces the second edition (ISO 15193:2009), which has been technically
revised.
The main changes are as follows:
— title and chapter headings have been changed to better reflect the objective of the document;
— requirements, concepts and definitions have been incorporated for consistency with ISO 17511:2020,
ISO 15194:XXXX, and ISO 15195:2018;
— content has been adapted to make the document applicable to all types of measurands;
— the subchapters of Chapter 4 "Requirements for a reference measurement procedure" have been
revised and specified in order to present the requirements more transparently;
— subclause 4.1 has been added to emphasize quality requirements for a reference measurement
procedure; and
— aspects of validation have been updated and summarized in Chapter 4.15.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
v
ISO/DIS 15193:2023(E)
Introduction
Reference measurement systems are needed to enable the results produced by end user measurement
procedures to be metrologically traceable to measurement standards and/or measurement procedures
of the highest metrological level. Such systems exist within a traceability chain/calibration hierarchy as
described in ISO 17511:2020. In the context of in vitro diagnostic (IVD) medical devices, they mitigate
the risk of harm to patients by avoiding inconsistent results from different measuring systems.
Reference measurement procedures play a crucial role in this metrological traceability system, because
they can be used for the following:
a) assessing performance properties of measuring systems – comprising measuring instruments,
auxiliary equipment as well as reagents,
b) assessing whether there is a functional interchangeability of different end user measurement
procedures purporting to measure the same quantity,
c) assigning quantity values to reference materials that are then used for purposes of calibration or
measurement trueness control of end user measurement procedures, and
d) detecting analytical influence quantities in biological samples measured using end user
measurement procedures.
For medical laboratory measurements, in particular, it is vitally important to both patient care and
health screening that the measurement results reported by different end user systems be equivalent,
comparable over time, reproducible and accurate. Establishing metrological traceability of an end user
measuring system to a reference measurement procedure enables equivalent results to be reported. It
is advisable that a reference measurement procedure be specified, especially when:
— it is required by e.g. standards, technical specifications, or technical regulations,
— quantity values are to be stated by the manufacturer, and
— technical requirements have a direct relationship to the performance of a product or process.
The advantages of having a standard for reference measurement procedures are listed in the ISO/IEC
Directives, Part 1, 2021, (Annex SM).
In this document, defined concepts are indicated by italicized text.
vi
DRAFT INTERNATIONAL STANDARD ISO/DIS 15193:2023(E)
In vitro diagnostic medical devices — Requirements for
reference measurement procedures
1 Scope
This document specifies requirements for reference measurement procedures for measurands used in
laboratory medicine.
This document applies to
a) reference measurement procedures providing values of differential or rational quantities. Annex A
provides information on nominal properties and ordinal quantities.
b) any person, body or institution developing reference measurement procedures for measurands
used in laboratory medicine.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO/IEC Guide 98-3:2008, Uncertainty of measurement — Part 3: Guide to the expression of uncertainty in
me a s ur ement (GUM: 1995)
ISO/IEC Guide 99:2007, International vocabulary of metrology — Basic and general concepts and
associated terms (VIM)
ISO 17511:2020, In vitro diagnostic medical devices — Requirements for establishing metrological
traceability of values assigned to calibrators, trueness control materials and human samples
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO/IEC Guide 99 and ISO 17511
and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
3.1
analyte
component represented in the name of a quantity (3.29)
EXAMPLE In the type of quantity (3.29) "mass of protein in 24-hour urine", "protein" is the analyte. In
“amount of substance of glucose in plasma”, “glucose” is the analyte. In both cases the full phrase describes the
measurand (3.18).
[SOURCE: ISO 17511:2020, 3.1]
3.2
analytical interference
systematic effect on a measurement caused by an influence quantity (3.14) which does not by itself
produce an indication, but which causes an enhancement or depression of the indication
ISO/DIS 15193:2023(E)
3.3
analytical portion
portion of material taken from the analytical sample (3.4) and on which the measurement is actually
carried out, either directly or following dissolution
Note 1 to entry: The analytical portion is taken directly from the primary sample (3.28) or laboratory sample
(3.16) if no preparation of these is required. The analytical portion is sometimes dissolved to give an analytical
solution (3.7) before being exposed to the measuring device.
3.4
analytical sample
sample prepared from the laboratory sample (3.16) and from which analytical portions (3.3) can be
taken
Note 1 to entry: The analytical sample can be subjected to various treatments before an analytical portion (3.3)
is taken.
3.5
analytical selectivity
selectivity of a measuring system
selectivity
property of a measuring system (3.24), used with a specified measurement procedure (3.21), whereby it
provides measured quantity (3.29) values for one or more measurands (3.18) such that the values of each
measurand (3.18) are independent of other measurands (3.18) or other quantities in the phenomenon,
body, or substance being investigated
Note 1 to entry: Adapted from ISO 17511:2020 and ISO/IEC Guide 99:2007.
3.6
analytical sensitivity
quotient of the change in an indication of a measuring system (3.24) and the corresponding change in
the value of a quantity (3.29) being measured
Note 1 to entry: The term analytical sensitivity is not intended to be used as a synonym for detection limit (3.12).
Note 2 to entry: ISO/IEC Guide 99:2007, 4.12 uses the term “sensitivity of a measuring system”.
3.7
analytical solution
solution prepared prior to measurement by dissolving an analytical portion (3.3) in a liquid material or
dispersing it in a solid material, with or without reaction
3.8
blank
lack of the analyte (3.1) or another component necessary to produce an indication of a measuring system
(3.24) that is specific to the analyte (3.1)
3.9
calibrator
measurement standard (3.22) used in calibration of a measuring system (3.24) according to a specified
measurement procedure (3.21)
Note 1 to entry: Adapted from ISO/IEC Guide 99:2007, 5.12 and ISO 17511:2020.
Note 2 to entry: The measurement standard (3.22) should be a certified reference material (3.10), if available.
ISO/DIS 15193:2023(E)
3.10
certified reference material
CRM
reference material (3.31), accompanied by documentation issued by an authoritative body, that provides
one or more specified property values with associated uncertainty (3.41) and traceability using valid
procedures
Note 1 to entry: Documentation is given in the form of a reference material (3.31) certificate and a certification
report.
Note 2 to entry: Requirements for the production and characterization of CRMs are given in ISO 17034 and
ISO 15194 ; guidance on characterization and value assignment is provided in ISO Guide 35.
Note 3 to entry: In this definition, “uncertainty” covers both measurement uncertainty (3.41) and uncertainty
associated with the value of a nominal property, such as for identity or sequence. Traceability covers both
metrological traceability (3.25) of a quantity (3.29) and traceability of a nominal property.
Note 4 to entry: Specified quantity (3.29) values of CRMs require metrological traceability (3.25) with associated
measurement uncertainty (3.41).
Note 5 to entry: ISO 17034 has an analogous definition.
Note 6 to entry: For a specified material, a calibration certificate provided by an accredited calibration laboratory
is not by itself sufficient to confer the status of CRM on these types of materials.
EXAMPLE Human serum with an assigned quantity (3.29) value and associated measurement uncertainty
(3.41) for concentration (amount of substance per unit volume) of cholesterol inherently present in the serum
and used as a calibrator (3.9) or as a trueness control material (3.11) is an example of a CRM.
[SOURCE: ISO 17511: 2020, 3.9]
3.11
control material
substance, material or artefact intended by its producer to be used to verify the performance of a
measurement procedure (3.21)
Note 1 to entry: The control material should be a reference material (3.31), if available.
3.12
detection limit
limit of detection
measured quantity (3.29) value, obtained by a given measurement procedure (3.21), for which the
probability of falsely claiming the absence of a component in a material is β, given a probability α of
falsely claiming its presence
Note 1 to entry: IUPAC recommends default values for α and β equal to 0.05.
Note 2 to entry: The abbreviation LOD is sometimes used.
Note 3 to entry: The term “sensitivity” is discouraged for this concept.
Note 4 to entry: Adapted from ISO/IEC Guide 99:2007, 4.18.
3.13
examination
set of operations having the objective of determining the numerical value, text value or characteristics
of a property
Note 1 to entry: An examination may be the total of a number of activities, observations or measurements
required to determine a value or characteristic.
Note 2 to entry: Laboratory examinations that determine a numerical value of a property are called “quantitative
examinations”; those that determine the characteristics of a property are called “qualitative examinations”.
ISO/DIS 15193:2023(E)
[SOURCE: ISO 15189: 2022, 3.8]
3.14
influence quantity
quantity (3.29) that, in a direct measurement, does not affect a quantity (3.29) that is actually measured,
but affects the relation between the indication and the measurement result
Note 1 to entry: Adapted from ISO/IEC Guide 99:2007, 2.52.
3.15
intermediate precision standard deviation
closeness of agreement between independent test results obtained under stipulated conditions
(ISO 5725-1)
Note 1 to entry: Stipulations can be measurements made at different times, different operators, different
equipment, calibrations carried out between measurements or any combination of those stipulations (from
Table 1 ISO 5725-3:1994)
3.16
laboratory sample
subsample of the primary sample (3.28), as prepared for sending to or as received by the laboratory and
intended for measurement
Note 1 to entry: In cases where there is no subsample of the primary sample (3.28), the primary sample (3.28)
could also be used as the laboratory sample.
3.17
matrix
system matrix
components of a material, except the analyte (3.1)
Note 1 to entry: The biological system excluding the analyte (3.1) is the matrix of the material.
EXAMPLE In the measurand (3.18) “mass concentration of protein in urine”, “urine” is the matrix. In the
measurand (3.18) “amount of substance of glucose in plasma”, “plasma” is the matrix.
[SOURCE: ISO 17511:2020, 3.24]
3.18
measurand
quantity (3.29) intended to be measured
Note 1 to entry: Specification of a measurand requires knowledge of the kind of quantity (3.29), description of the
state of the phenomenon, body, or substance carrying the quantity (3.29), including any relevant component, and
the chemical entities involved.
Note 2 to entry: The measurement, including the measuring system (3.24) and the conditions under which the
measurement is carried out, could change the phenomenon, body, or substance such that the quantity being
measured can differ from the measurand as defined. In this case, adequate correction is necessary.
EXAMPLE The length of a steel rod in equilibrium at ambient Celsius temperature of 23 °C will be different
from the length at the specified temperature of 20 °C, which is the measurand. In this case, a correction is
necessary.
Note 3 to entry: “Analyte” (3.1), or the name of a substance or compound, are terms sometimes used for
“measurand”. This usage is erroneous because these terms do not refer to quantities (3.29).
Note 4 to entry: In laboratory medicine, the description of the measurand includes the name of the quantity
(3.29) (e.g. amount of substance concentration), the component/analyte (3.1) (e.g. β-D-glucose), and the biological
system in which it is found (e.g. blood plasma).
[SOURCE: ISO 17511:2020, 3.26]
ISO/DIS 15193:2023(E)
3.19
measurement interval
set of values of quantities (3.29) of the same kind that can be measured by a given measuring instrument
or measuring system (3.24) with specified instrumental uncertainty (3.41), under defined conditions
Note 1 to entry: The lower limit of a measurement interval should not be confused with limit of detection (3.12).
3.20
measurement method
method of measurement
generic description of a logical organization of operations used in a measurement
[SOURCE: ISO 17511:2020, 3.30]
3.21
measurement procedure
MP
description of a measurement according to one or more measurement principles and to a given
measurement method (3.20), based on a measurement model and including any calculation to obtain a
measurement result. A measurement procedure refers to a written specification for how a measurement
is performed, including a technical description of reagents, calibrators (3.9), equipment, instrument,
and other details necessary to create and operate a measuring system (3.24) that implements those
specifications
Note 1 to entry: A MP is usually documented in sufficient detail to enable an operator to perform a measurement.
Note 2 to entry: A MP can include a statement concerning a target measurement uncertainty (3.41).
Note 3 to entry: A MP is sometimes called a standard operating procedure, abbreviated SOP.
Note 4 to entry: In medical laboratory measurements, a specific instance of a measurement procedure, an IVD
medical device, is used to make a measurement on a clinical sample (3.26) to produce a measurement result
which is used to inform medical decisions for a patient.
[SOURCE: ISO 17511:2020, 3.30 – Note 4 added]
3.22
measurement standard
realization of the definition of a given quantity (3.29), with a stated value and associated measurement
uncertainty (3.41), used as a reference
[SOURCE: ISO 17511:2020, 3.28]
3.23
measurement trueness
trueness of measurement
trueness
closeness of agreement between the average of an infinite number of replicate measured quantity (3.29)
values and a reference quantity (3.29) value
Note 1 to entry: Measurement trueness is not a quantity (3.29) and thus cannot be expressed numerically, but
measures for closeness of agreement are given in ISO 5725.
Note 2 to entry: Measurement trueness is inversely related to systematic measurement error but is not related to
random measurement error.
Note 3 to entry: “Measurement accuracy” should not be used for “measurement trueness”.
[SOURCE: VIM3: 2012, 2.14]
ISO/DIS 15193:2023(E)
3.24
measuring system
set of one or more measuring instruments and often other devices, including any reagent and supply,
assembled and adapted to give information used to generate measured quantity (3.29) values within
specified intervals for quantities (3.29) of specified kinds
Note 1 to entry: A measuring system may consist of only one measuring instrument.
[S OU RC E: V I M3: 2012 , 3.2]
3.25
metrological traceability
property of a measurement result whereby the result can be related to a reference through a
documented unbroken chain of calibrations, each contributing to the measurement uncertainty (3.41)
Note 1 to entry: For this definition, a ‘reference’ can be a definition of a measurement unit through its practical
realization, or a measurement procedure (3.21) including the measurement unit for a quantity (3.29) value, or a
measurement standard (3.22).
Note 2 to entry: Metrological traceability requires an established calibration hierarchy.
Note 3 to entry: The abbreviated term “traceability” is commonly used to refer to other concepts, such as ‘sample
traceability’ or ‘document traceability’ or ‘instrument traceability’ or ‘material traceability’, where the history
(“trace”) of an item is meant. Therefore, the full term of “metrological traceability” is preferred if there is any risk
of confusion.
[SOURCE: ISO 17511:2020, 3.30 –Notes 1, 2 and 9 (here note 3) only are retained from the source]
3.26
precision of measurement
measurement precision
closeness of agreement between indications or measured quantity (3.29) values obtained by replicate
measurements on the same or similar objects under specified conditions
Note 1 to entry: Measurement precision is usually expressed numerically by measures of imprecision, such as
standard deviation, variance, or coefficient of variation under the specified conditions of measurement.
Note 2 to entry: The ‘specified conditions’ can be, for example, repeatability conditions of measurement,
intermediate precision conditions of measurement, or reproducibility conditions (3.35) of measurement (see
ISO 5725-1:1994).
Note 3 to entry: Measurement precision is used to define measurement repeatability, intermediate measurement
precision, and measurement reproducibility.
Note 4 to entry: Sometimes “measurement precision” is erroneously used to mean measurement accuracy.
[SOURCE: ISO/IEC Guide 99:2007, 2.15]
3.27
primary calibrator
primary calibration material
calibrator (3.9) established using a primary reference measurement procedure (3.32), or created as
an artefact, chosen by convention. A primary calibrator serves as the anchor point for a calibration
hierarchy
EXAMPLE 1 Primary calibrator of amount of substance concentration prepared by dissolving a known amount
of substance of a primary reference material (3.31) for a chemical component into a known volume of solution.
EXAMPLE 2 Primary calibrator for isotope amount of substance ratio measurements, prepared by mixing
known amounts-of-substance of specified isotopes.
Note 1 to entry: ISO 17511:2020 further explains the role of a primary calibrator, defined as position m.2 within
a calibration hierarchy.
ISO/DIS 15193:2023(E)
[SOURCE: ISO 17511:2020, 3.37 – with “measurement standard” replaced by “calibrator”, “Note” added
and Examples 1 to 5 in the source text replaced by two examples relevant to laboratory medicine.]
3.28
primary sample
specimen
discrete portion of a body fluid or tissue or other sample associated with the human body taken for
examination (3.13), study or analysis of one or more quantities (3.29) or characteristics to determine
the character of the whole
[SOURCE: ISO 15189:2022, 3.25]
3.29
quantity
property of a phenomenon, body, or substance, where the property has a magnitude that can be
expressed as a number and a reference
+
EXAMPLE 1 “Plasma (Blood) — Na ion; amount of substance concentration of 143 mmol/l in a given person at
a given time”.
EXAMPLE 2 Number concentration of erythrocytes in a blood sample (whole blood) of 5 × 10 /μl in a given
person at a given time.
Note 1 to entry: Quantity is not to be confused with analyte (3.1).
Note 2 to entry: A measurement procedure (3.21) for which the result is expressed in a qualitative manner (e.g.
nucleotide sequence for a DNA sample or “present” or “not present” against a ratio or counting scale with a pre-
determined decision threshold), are consistent with this definition of the term quantity.
[SOURCE: ISO 17511:2020, 3.38]
3.30
recovery
fraction of analyte (3.1) measured as a proportion of the amount of analyte (3.1) present in or added to
the analytical portion (3.3) of the test material, which is presented for measurement after preparation
of the sample
[14]
Note 1 to entry: Cited from and modified.
3.31
reference material
RM
material sufficiently homogeneous and stable with reference to specified properties, which has been
established to be fit for its intended use in measurement or in examination (3.13) of nominal properties
EXAMPLE 1 Examples of RMs embodying a quantity (3.29):
a) water of stated purity, the dynamic viscosity of which is used to calibrate viscometers;
b) human serum without an assigned quantity (3.29) value for the amount of substance concentration
of cholesterol inherently present in the serum and used as a control material (3.11).
EXAMPLE 2 Example of RMs embodying nominal properties:
DNA compound containing a specified nucleotide sequence.
EXAMPLE 3 Spheres of uniform size mounted on a microscope slide.
Note 1 to entry: RMs with or without assigned quantity (3.29) values can be used as control materials (3.11)
whereas only CRMs (3.10) with traceable quantity (3.29) values can be used as calibrators (3.9) or as trueness
control materials (3.11).
Note 2 to entry: RMs normally embody quantity (3.29) values but can also embody nominal properties.
ISO/DIS 15193:2023(E)
Note 3 to entry: RMs can have assigned quantity (3.29) values that are traceable to a measurement unit outside
the International System of units (SI).
Note 4 to entry: An RM accompanied by documentation issued by an authoritative body and referring to valid
measurement procedures (3.21) used to assign a quantity (3.29) value with associated measurement uncertainty
(3.41) and metrological traceability (3.25) is referred to as a CRM (3.10).
[SOURCE: ISO 17511: 2020, 3.39 with modifications]
3.32
reference measurement procedure
RMP
measurement procedure (3.21) accepted as providing measurement results fit for their intended use
in assessing measurement trueness (3.23) of measured quantity (3.29) values obtained from other
measurement procedures (3.21) for a quantity (3.29) of the same kind, in calibration, or in assigning a
quantity (3.29) value to a CRM (3.10)
Note 1 to entry: Definitions of reference materials (3.31) are described in ISO 15194.
Note 2 to entry: Explanation of the role of a primary RMP within a calibration hierarchy can be found in ISO 17511.
Note 3 to entry: When several reference measurement procedures exist for a given measurable quantity (3.29), it
can be possible to arrange them in a hierarchy according to size of measurement uncertainty (3.41).
Note 4 to entry: This document refers to reference measurement procedures in general and includes higher order
reference measurement procedures as defined in ISO 17511:2020.
[SOURCE: ISO 17511:2020, 3.40 – modified]
3.33
reference measurement system
measuring system (3.24) accepted as fit for its intended purpose in assessing or establishing measurement
trueness (3.23) for quantity (3.29) values obtained from other measurement procedures (3.21) for the
measurand (3.18); comprised of (1) a unit of measurement, (2) a definition of the measurand (3.18), (3)
RMP(s), (4) RM(s) and (5) one or more laboratories providing reference measurement services
3.34
repeatability standard deviation
the standard deviation of test results obtained with the same method on identical test items in the
same laboratory by the same operator using the same equipment within short intervals of time
[SOURCE: ISO 5725-1:1994]
3.35
reproducibility condition
(of measurement)
condition of measurement, out of a set of conditions that includes different locations, operators,
measuring systems (3.24), and replicate measurements on the same or similar objects
[SOURCE: VIM3: 2012, 2.24]
3.36
reproducibility standard deviation
the standard deviation of test results obtained with the same method on identical test items in different
laboratories with different operators using different equipment
[SOURCE: ISO 5725-1:1994]
ISO/DIS 15193:2023(E)
3.37
secondary calibrator
secondary calibration material
calibrator (3.9) established through calibration with respect to a primary calibrator (3.27) or by a RMP
(3.25) that defines the measurand (3.18) or by a consensus of results obtained using measurement
procedures (3.21)
Note 1 to entry: Also called “commutable reference material” or “secondary reference material”.
Note 2 to entry: Calibration may be obtained directly between a primary calibrator (3.27) and a secondary
calibrator or involve an intermediate measurement system (3.16) calibrated by the primary calibrator (3.27) and
used to assign a measurement result to the secondary calibrator.
Note 3 to entry: ISO 17511 further explains the role of a secondary calibrator, defined as position m.3 within a
calibration hierarchy.
[SO
...
PROJET
Norme
internationale
ISO/DIS 15193
ISO/TC 212
Dispositifs médicaux de diagnostic
Secrétariat: ANSI
in vitro — Exigences relatives aux
Début de vote:
procédures de mesure de référence
2023-12-26
In vitro diagnostic medical devices — Requirements for reference
Vote clos le:
measurement procedures
2024-03-19
ICS: 11.100.10
CE DOCUMENT EST UN PROJET DIFFUSÉ
POUR OBSERVATIONS ET APPROBATION. IL
EST DONC SUSCEPTIBLE DE MODIFICATION
ET NE PEUT ÊTRE CITÉ COMME NORME
INTERNATIONALE AVANT SA PUBLICATION EN
TANT QUE TELLE.
Le présent document est distribué tel qu’il est parvenu du secrétariat
du comité. OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES
FINS INDUSTRIELLES, TECHNOLOGIQUES ET
COMMERCIALES, AINSI QUE DU POINT DE VUE
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
TRAITEMENT PARALLÈLE ISO/CEN
CONSIDÉRÉS DU POINT DE VUE DE LEUR
POSSIBILITÉ DE DEVENIR DES NORMES
POUVANT SERVIR DE RÉFÉRENCE DANS LA
RÉGLEMENTATION NATIONALE.
LES DESTINATAIRES DU PRÉSENT PROJET
SONT INVITÉS À PRÉSENTER, AVEC LEURS
OBSERVATIONS, NOTIFICATION DES DROITS
DE PROPRIÉTÉ DONT ILS AURAIENT
ÉVENTUELLEMENT CONNAISSANCE
ET À FOURNIR UNE DOCUMENTATION
EXPLICATIVE.
Numéro de référence
ISO/DIS 15193:2023(fr)
̹ʹͲʹ͵–‘—•†”‘‹–•”±•‡”˜±•
ISO/DIS 15193:2023(fr)
ISO/TC 212
ƒ–‡ǣʹͲʹͶǦͲ͵Ǧͳͻ
ISO/DIS 15193:2023(F)
ȀʹͳʹȀ
ʹ
‡…”±–ƒ”‹ƒ–ǣ
Dispositifs médicaux de diagnostic in vitro — Exigences relatives
aux procédures de mesure de référence
In vitro diagnostic medical devices — Requirements for reference measurement procedures
Avertissement
‡†‘…—‡–’‡•–’ƒ•—‡‘”‡‹–‡”ƒ–‹‘ƒŽ‡†‡Ž’Ǥއ•–†‹•–”‹„—±’‘—”‡šƒ‡‡–‘„•‡”˜ƒ–‹‘•Ǥ
އ•–•—•…‡’–‹„އ†‡‘†‹ˆ‹…ƒ–‹‘•ƒ•’”±ƒ˜‹•‡–‡’‡—–²–”‡…‹–±…‘‡‘”‡‹–‡”ƒ–‹‘ƒŽ‡Ǥ
‡•†‡•–‹ƒ–ƒ‹”‡•†—’”±•‡–’”‘Œ‡–•‘–‹˜‹–±•’”±•‡–‡”ǡƒ˜‡…އ—”•‘„•‡”˜ƒ–‹‘•ǡ‘–‹ˆ‹…ƒ–‹‘†‡•
†”‘‹–•†‡’”‘’”‹±–±†‘–‹Ž•ƒ—”ƒ‹‡–±˜‡–—‡ŽŽ‡‡–…‘ƒ‹••ƒ…‡‡–ˆ‘—”‹”—‡†‘…—‡–ƒ–‹‘
‡š’Ž‹…ƒ–‹˜‡Ǥ
ǣͳͳǤͳͲͲǤͳͲ
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2023
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org ›’‡†—†‘…—‡–ǣ‘”‡‹–‡”ƒ–‹‘ƒŽ‡
Web: www.iso.org
‘—•Ǧ–›’‡†—†‘…—‡–ǣ
Publié en Suisse –ƒ†‡†—†‘…—‡–ǣȋͶͲȌ“—²–‡
ƒ‰—‡†—†‘…—‡–ǣ
ii
ISO/DIS 15193:2023(F)
Sommaire Page
Avant-propos . v
Introduction . vii
1 Domaine d’application .1
2 Références normatives .1
3 Termes et définitions .1
4 Exigences relatives aux procédures de mesure de référence . 12
4.1 Généralités . 12
4.2 Éléments d’une procédure de mesure de référence . 12
4.3 Avertissement et précautions de sécurité . 13
4.4 Introduction. 13
4.5 Domaine d’application . 13
4.6 Termes, définitions, symboles et abréviations . 14
4.6.1 Concepts . 14
4.6.2 Nomenclature . 14
4.6.3 Noms communs . 14
4.7 Principe de mesure et méthode de mesure . 14
4.8 Réactifs et matériaux . 14
4.8.1 Généralités . 14
4.8.2 Éléments descriptifs . 15
4.8.3 Grandeurs d’influence . 16
4.8.4 Expression de la concentration . 16
4.8.5 Expression de la dilution . 16
4.8.6 Référence à des éléments brevetés . 17
4.9 Appareillage et équipement . 17
4.9.1 Description . 17
4.9.2 Équipement auxiliaire . 17
4.10 Consommables . 17
4.11 Échantillonnage et échantillon . 17
4.11.1 Généralités . 17
4.11.2 Échantillons . 17
4.12 Préparation du système de mesure et de la prise d’essai pour analyse . 18
4.12.1 Généralités . 18
4.12.2 Préparation de l’appareillage . 18
4.12.3 Étalonnage . 18
4.12.4 Types d’échantillon pour analyse . 19
4.12.5 Conception des séries analytiques . 19
4.12.6 Prise d’essai pour analyse . 19
4.12.7 Solution pour analyse . 19
4.13 Utilisation du système de mesure. 19
4.13.1 Déroulement des étapes de mesure. 19
4.13.2 Blanc . 20
4.13.3 Validation des conditions de mesure . 20
4.14 Traitement des données . 20
4.14.1 Calcul des résultats de mesure . 20
4.14.2 Équations de conversion . 21
4.15 Validation . 21
4.15.1 Concepts, valeurs et leur utilisation . 21
iii
ISO/DIS 15193:2023(F)
4.15.2 Fonction d’étalonnage analytique . 21
4.15.3 Fonction de mesure analytique . 21
4.15.4 Sensibilité analytique . 21
4.15.5 Grandeurs d’influence analytique . 22
4.15.6 Mesurage du blanc . 22
4.15.7 Études de récupération . 22
4.15.8 Justesse de mesure . 22
4.15.9 Fidélité de mesure . 22
4.15.10 Écart-type de répétabilité, s . 22
r
4.15.11 Écart-type de fidélité intermédiaire . 23
4.15.12 Écart-type de reproductibilité, s . 23
R
4.15.13 Limite de détection . 23
4.15.14 Intervalle de mesure . 23
4.15.15 Validation à l’aide d’un matériau de référence . 23
4.15.16 Comparaison interlaboratoires. 24
4.15.17 Incertitude de mesure . 24
4.16 Cas particuliers . 24
4.17 Rapport des résultats de mesure . 25
4.18 Assurance qualité . 25
4.19 Bibliographie . 25
4.20 Dates d’autorisation et de révision . 25
(informative) Procédures de référence pour les propriétés autres que les
grandeurs différentielles et proportionnelles . 26
Annexe ZA (informative) Relation entre la présente Norme européenne et les exigences
générales en matière de sécurité et de performances concernées du Règlement (UE)
2017/746 . 28
Bibliographie . 31
iv
ISO/DIS 15193:2023(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude a le
droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (IEC) en ce qui
concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L’ISO attire l’attention sur le fait que la mise en application du présent document peut entraîner
l’utilisation d’un ou de plusieurs brevets. L’ISO ne prend pas position quant à la preuve, à la validité et à
l’applicabilité de tout droit de propriété revendiqué à cet égard. À la date de publication du présent
document, l’ISO n’avait pas reçu notification qu’un ou plusieurs brevets pouvaient être nécessaires à sa
mise en application. Toutefois, il y a lieu d’avertir les responsables de la mise en application du présent
document que des informations plus récentes sont susceptibles de figurer dans la base de données de
brevets, disponible à l’adresse www.iso.org/brevets. L’ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de brevets.
Les appellations commerciales éventuellement mentionnées dans le présent document sont données
pour information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l’ISO liés à l’évaluation de la conformité, ou pour toute information au sujet de l’adhésion
de l’ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles
techniques au commerce (OTC), voir www.iso.org/iso/foreword.html.
Le présent document a été élaboré par le comité technique ISO/TC 212, Laboratoires d’analyses de
biologie médicale et systèmes de diagnostic in vitro.
Cette troisième édition annule et remplace la deuxième édition (ISO 15193:2009), qui a fait l’objet d’une
révision technique.
Les principales modifications sont les suivantes :
— modification du titre et des intitulés des articles/paragraphes afin de mieux refléter l’objectif du
document ;
— introduction d’exigences, de concepts et de définitions dans un but d’harmonisation avec
l’ISO 17511:2020, l’ISO 15194:XXXX et l’ISO 15195:2018 ;
— adaptation du contenu pour rendre le document applicable à tous les types de mesurandes ;
— révision et clarification des paragraphes de l’Article 4, Présentation d’une procédure de mesure de
référence, afin de présenter les exigences de manière plus transparente ;
v
ISO/DIS 15193:2023(F)
— ajout du paragraphe 4.1 afin de souligner les exigences de qualité d’une procédure de mesure de
référence ;
— mise à jour et résumé des aspects concernant la validation dans le paragraphe 4.15.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes
est disponible à l’adresse www.iso.org/members.html.
vi
ISO/DIS 15193:2023(F)
Introduction
Les systèmes de mesure de référence doivent permettre la traçabilité métrologique des résultats
produits par les procédures de mesure de l’utilisateur final à des étalons et/ou des procédures de
mesure du plus haut niveau métrologique. De tels systèmes existent au sein d’une chaîne de
traçabilité/hiérarchie d’étalonnage comme décrit dans l’ISO 17511:2020. Dans le contexte des
dispositifs médicaux de diagnostic in vitro (DIV), ils atténuent le risque de préjudice pour les patients en
évitant les résultats incohérents obtenus avec des systèmes de mesure différents.
Les procédures de mesure de référence jouent un rôle essentiel dans ce système de traçabilité
métrologique du fait qu’elles peuvent être utilisées pour :
a) évaluer les propriétés de performances des systèmes de mesure — comprenant les instruments de
mesure, les équipements auxiliaires et les réactifs ;
b) évaluer s’il y a interchangeabilité fonctionnelle des différentes procédures de mesure de
l’utilisateur final destinées à mesurer la même grandeur ;
c) assignation des valeurs de grandeur aux matériaux de référence, qui sont ensuite utilisés pour
l’étalonnage ou le contrôle de la justesse de mesure des procédures de mesure de l’utilisateur final ;
d) détecter les grandeurs d’influence analytique dans les échantillons biologiques mesurées en
utilisant les procédures de mesure de l’utilisateur final.
En ce qui concerne les mesurages de laboratoires d’analyses de biologie médicale, en particulier, il est
fondamental, tant pour les soins aux patients que pour la surveillance sanitaire, que les résultats de
mesure fournis par les différents systèmes d’utilisateur final soient équivalents, comparables dans le
temps, reproductibles et exacts. L’établissement de la traçabilité métrologique d’un système de mesure
d’utilisateur final à une procédure de mesure de référence permet de rapporter des résultats
équivalents. Il est conseillé de spécifier une procédure de mesure de référence, en particulier lorsque :
— elle est exigée, par exemple, par les normes, les spécifications techniques ou les règlements
techniques ;
— les valeurs d’une grandeur doivent être déclarées par le fabricant ;
— les exigences techniques sont directement liées aux performances d’un produit ou d’un procédé.
Les avantages du fait de disposer d’une norme pour les procédures de mesure de référence sont
énumérés dans les Directives ISO/IEC, Partie 1, 2021, (Annexe SM).
Dans le présent document, les concepts définis sont indiqués par du texte en italique.
vii
PROJET DE NORME INTERNATIONALE ISO/DIS 15193:2023(F)
Dispositifs médicaux de diagnostic in vitro — Exigences
relatives aux procédures de mesure de référence
1 Domaine d’application
Le présent document spécifie les exigences relatives aux procédures de mesure de référence pour les
mesurandes utilisés en biologie médicale.
Le présent document s’applique :
a) à des procédures de mesure de référence permettant d’obtenir des valeurs de grandeurs
différentielles ou proportionnelles. L’Annexe A fournit des informations sur les propriétés
nominales et les grandeurs ordinales ;
b) à toute personne, tout organisme ou toute institution développant des procédures de mesure de
référence pour les mesurandes utilisés en biologie médicale.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique.
Pour les références non datées, la dernière édition du document de référence s’applique (y compris les
éventuels amendements).
ISO/IEC Guide 98-3:2008, Incertitude de mesure — Partie 3 : Guide pour l’expression de l’incertitude de
mesure (GUM:1995).
ISO/IEC Guide 99:2007, Vocabulaire international de métrologie — Concepts fondamentaux et généraux
et termes associés (VIM).
ISO 17511:2020, Dispositifs médicaux de diagnostic in vitro — Exigences pour l’établissement d’une
traçabilité métrologique des valeurs attribuées aux étalons, aux matériaux de contrôle de la justesse et aux
échantillons humains.
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions du Guide ISO/IEC 99 et de l’ISO 17511
et les suivants s’appliquent.
L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en
normalisation, consultables aux adresses suivantes :
— ISO Online browsing platform : disponible à l’adresse https://www.iso.org/obp
— IEC Electropedia : disponible à https://www.electropedia.org/
ISO/DIS 15193:2023(F)
3.1
analyte
composant représenté au nom d’une grandeur (3.29)
EXEMPLE Dans le type de grandeur (3.29) « masse de protéines dans l’urine de 24 h », « protéine » est
l’analyte. Dans l’expression « quantité de matière du glucose dans le plasma », « glucose » est l’analyte. Dans les
deux cas, l’expression complète représente le mesurande (3.18).
[SOURCE : ISO 17511:2020, 3.1]
3.2
interférence analytique
effet systématique causé par une grandeur d’influence (3.14) qui ne produit pas par lui-même une
indication, mais qui provoque une augmentation ou une diminution de l’indication
3.3
prise d’essai pour analyse
portion de matériel prélevée dans l’échantillon pour analyse (3.4) et sur laquelle des opérations de
mesure sont effectuées, directement ou après dissolution
Note 1 à l’article : Lorsqu’aucune préparation n’est requise, la prise d’essai pour analyse est directement prélevée
dans l’échantillon primaire (3.28) ou dans l’échantillon pour laboratoire (3.16). La prise d’essai pour analyse est
parfois dissoute de manière à obtenir une solution pour analyse (3.7) avant d’être exposée au dispositif de mesure.
3.4
échantillon pour analyse
échantillon préparé à partir de l’échantillon pour laboratoire (3.16) et duquel des prises d’essai pour
analyse (3.3) peuvent être prélevées
Note 1 à l’article : L’échantillon pour analyse peut être soumis à divers traitements avant le prélèvement de la
prise d’essai pour analyse (3.3).
3.5
sélectivité analytique
sélectivité d’un système de mesure
sélectivité
propriété d’un système de mesure (3.24), utilisé avec une procédure de mesure (3.21) spécifiée,
selon laquelle le système fournit des valeurs mesurées de grandeur (3.29) pour un ou plusieurs
mesurandes (3.18), de sorte que les valeurs de chaque mesurande (3.18) sont indépendantes d’autres
mesurandes (3.18) ou d’autres grandeurs dans le phénomène, le corps ou la substance en cours
d’examen
Note 1 à l’article : Adapté d’après l’ISO 17511:2020 et le Guide ISO/IEC 99:2007.
3.6
sensibilité analytique
quotient de la variation d’une indication d’un système de mesure (3.24) par la variation correspondante
de la valeur de la grandeur (3.29) mesurée
Note 1 à l’article : Le terme sensibilité analytique n’est pas censé être utilisé comme synonyme de limite de
détection (3.12).
Note 2 à l’article : Le Guide ISO/IEC 99:2007, 4.12 utilise le terme « sensibilité d’un système de mesure ».
ISO/DIS 15193:2023(F)
3.7
solution pour analyse
solution préparée avant le mesurage par dissolution d’une prise d’essai pour analyse (3.3) dans un
matériau liquide ou sa dispersion dans un matériau solide, avec ou sans réaction
3.8
blanc
absence de l’analyte (3.1) ou d’un autre composant nécessaire pour produire une indication d’un
système de mesure (3.24) qui est spécifique de l’analyte (3.1)
3.9
étalon
étalon de mesure (3.22) utilisé pour l’étalonnage d’un système de mesure (3.24) conformément à une
procédure de mesure (3.21) spécifiée
Note 1 à l’article : Adapté d’après le Guide ISO/IEC 99:2007, 5.12 et l’ISO 17511:2020.
Note 2 à l’article : Il convient que l’étalon de mesure (3.22) soit un matériau de référence certifié (3.10), s’il est
disponible.
3.10
matériau de référence certifié
MRC
matériau de référence (3.31) accompagné d’une documentation délivrée par un organisme faisant
autorité et fournissant une ou plusieurs valeurs de propriétés spécifiées avec l’incertitude (3.41) et la
traçabilité associées, en utilisant des procédures valables
Note 1 à l’article : La documentation est délivrée sous la forme d’un certificat de matériau de référence (3.31) et
d’un rapport de certification.
Note 2 à l’article : Des exigences pour la production et la caractérisation des MRC sont fournies dans l’ISO 17034 et
l’ISO 15194 ; des recommandations pour la caractérisation et l’assignation de valeur sont fournies dans le
Guide ISO 35.
Note 3 à l’article : Dans la définition, le terme « incertitude » peut désigner soit une « incertitude de
mesure » (3.41), soit l’incertitude associée à la valeur d’une propriété nominale, telle que l’identité ou la séquence.
La traçabilité englobe à la fois la traçabilité métrologique (3.25) d’une grandeur (3.29) et la traçabilité d’une valeur
nominale.
Note 4 à l’article : Les valeurs de grandeur (3.29) spécifiées des MRC exigent une traçabilité métrologique (3.25)
avec une incertitude de mesure (3.41) associée.
Note 5 à l’article : L’ISO 17034 possède une définition analogue.
Note 6 à l’article : Tout certificat d’étalonnage fourni par un laboratoire d’étalonnage agréé pour un matériau
précis ne saurait conférer le statut de MRC à ce type de matériaux.
EXEMPLE Un sérum humain avec une valeur de grandeur (3.29) assignée et une incertitude de mesure (3.41)
associée pour la concentration (quantité de matière par unité de volume) de cholestérol présent de manière
inhérente et qui sert d’étalon (3.9) ou de matériau de contrôle (3.11) de la justesse est un exemple de MRC.
[SOURCE : ISO 17511:2020, 3.9]
3.11
matériau de contrôle
substance, matériau ou objet conçu par son producteur pour vérifier les performances d’une procédure
de mesure (3.21)
Note 1 à l’article : Il convient que le matériau de contrôle soit un matériau de référence (3.31), s’il est disponible.
ISO/DIS 15193:2023(F)
3.12
limite de détection
valeur de grandeur (3.29) mesurée, obtenue selon une procédure de mesure (3.21) donnée,
pour laquelle la probabilité de conclure faussement à l’absence d’un constituant dans un matériau est β,
avec une probabilité α de conclure faussement à sa présence
Note 1 à l’article : L’IUPAC recommande une valeur par défaut de 0,05 pour α et β.
Note 2 à l’article : L’abréviation LOD est parfois utilisée.
Note 3 à l’article : Le terme « sensibilité » est déconseillé pour ce concept.
Note 4 à l’article : Adapté d’après le Guide ISO/IEC 99:2007, 4.18.
3.13
analyse
examen
ensemble des opérations destinées à déterminer la valeur numérique, la valeur textuelle ou les
caractéristiques d’une propriété
Note 1 à l’article : Un examen peut correspondre à la totalité des activités, des observations ou des mesures
effectuées, nécessaires pour déterminer une valeur ou une caractéristique.
Note 2 à l’article : Les analyses de laboratoire qui déterminent une valeur numérique d’une propriété sont
nommées « analyses quantitatives » ; celles qui déterminent les caractéristiques d’une propriété sont nommées
« analyses qualitatives ».
[SOURCE : ISO 15189:2022, 3.8]
3.14
grandeur d’influence
grandeur (3.29) qui, lors d’un mesurage direct, n’a pas d’effet sur la grandeur (3.29) effectivement
mesurée, mais a un effet sur la relation entre l’indication et le résultat de mesure
Note 1 à l’article : Adapté d’après le Guide ISO/IEC 99:2007, 2.52.
3.15
écart-type de fidélité intermédiaire
étroitesse de l’accord entre des résultats d’essai indépendants obtenus dans des conditions stipulées
(ISO 5725-1)
Note 1 à l’article : Les stipulations peuvent être des mesures réalisées à différents moments, par des opérateurs
différents, en utilisant différents équipements, des étalonnages réalisés entre les mesures ou toute combinaison
de ces stipulations (d’après le Tableau 1 ISO 5725-3:1994)
3.16
échantillon pour laboratoire
sous-échantillon de l’échantillon primaire (3.28) préparé pour être envoyé au laboratoire ou reçu par ce
dernier à des fins de mesurage
Note 1 à l’article : Dans les cas où il n’existe pas de sous-échantillon de l’échantillon primaire (3.28), l’échantillon
primaire (3.28) peut également être utilisé comme échantillon pour laboratoire.
ISO/DIS 15193:2023(F)
3.17
matrice
matrice d’un système
composants d’un système de matériau à l’exception de l’analyte (3.1)
Note 1 à l’article : Le système biologique à l’exclusion de l’analyte (3.1) est la matrice du matériau.
EXEMPLE Dans le mesurande (3.18) « concentration en masse de protéines dans l’urine », « urine » est la
matrice. Dans le mesurande (3.18) « quantité de matière du glucose dans le plasma », « plasma » est la matrice.
[SOURCE : ISO 17511:2020, 3.24]
3.18
mesurande
grandeur (3.29) que l’on veut mesurer
Note 1 à l’article : La spécification d’un mesurande nécessite la connaissance de la nature de la grandeur (3.29) et
la description de l’état du phénomène, du corps ou de la substance dont la grandeur (3.29) est une propriété,
incluant tout constituant pertinent, et les entités chimiques en jeu.
Note 2 à l’article : Il se peut que le mesurage, incluant le système de mesure (3.24) et les conditions dans lesquelles
le mesurage est effectué, modifie le phénomène, le corps ou la substance de sorte que la grandeur mesurée peut
différer du mesurande tel qu’il est défini. Dans ce cas, une correction appropriée est nécessaire.
EXEMPLE La longueur d’une tige d’acier en équilibre à la température ambiante de 23 °C sera différente de la
longueur à la température spécifiée de 20 °C, qui est le mesurande. Dans ce cas, une correction est nécessaire.
Note 3 à l’article : Le terme « analyte » (3.1) ou le nom d’une substance ou d’un composé sont quelquefois utilisés à
la place de « mesurande ». Cet usage est erroné puisque ces termes ne désignent pas des grandeurs (3.29).
Note 4 à l’article : En biologie médicale, la description du mesurande comprend le nom de la grandeur (3.29)
(par exemple la concentration en quantité de matière), le composant/l’analyte (3.1) (par exemple β-D-glucose) et
le système biologique dans lequel il se trouve (par exemple, le plasma sanguin).
[SOURCE : ISO 17511:2020, 3.26]
3.19
intervalle de mesure
ensemble de valeurs de grandeurs (3.29) du même type pouvant être mesurées par un instrument de
mesure ou système de mesure (3.24) donné avec une incertitude (3.41) instrumentale spécifiée, dans des
conditions définies
Note 1 à l’article : Il convient de ne pas confondre la limite inférieure de l’intervalle de mesure avec la limite de
détection (3.12).
3.20
méthode de mesure
description générique de l’organisation logique des opérations mises en œuvre dans un mesurage
[SOURCE : ISO 17511:2020, 3.30]
ISO/DIS 15193:2023(F)
3.21
procédure de mesure
PM
description d’un mesurage conformément à un ou plusieurs principes de mesure et à une méthode de
mesure (3.20) donnée, fondée sur un modèle de mesure et incluant tout calcul destiné à obtenir un
résultat de mesure. Une procédure de mesure fait référence à une spécification écrite de la manière
dont une mesure est réalisée, incluant une description technique des réactifs, des étalons (3.9),
de l’équipement, de l’instrument et d’autres détails nécessaires pour créer et utiliser un système de
mesure (3.24) intégrant ces spécifications
Note 1 à l’article : Une PM est habituellement documentée avec assez de détails pour permettre à un opérateur
d’effectuer un mesurage.
Note 2 à l’article : Une PM peut inclure une assertion concernant une incertitude de mesure (3.41) cible.
Note 3 à l’article : Une PM est quelquefois appelée en anglais standard operating procedure, abrégé en SOP.
Note 4 à l’article : En biologie médicale, un type spécifique de procédure de mesure, un dispositif médical de DIV,
est utilisé pour effectuer une mesure sur un échantillon (3.26) clinique pour produire un résultat de mesure qui
est utilisé pour éclairer des décisions médicales pour un patient.
[SOURCE : ISO 17511:2020, 3.30 – Note 4 ajoutée]
3.22
étalon de mesure
réalisation de la définition d’une grandeur (3.29) donnée, avec une valeur déterminée et une incertitude
de mesure (3.41) associée, utilisée comme référence
[SOURCE : ISO 17511:2020, 3.28]
3.23
justesse de mesure
justesse
étroitesse de l’accord entre la moyenne d’un nombre infini de valeurs de grandeur (3.29) mesurées
répétées et une grandeur (3.29) de référence
Note 1 à l’article : La justesse de mesure n’est pas une grandeur (3.29) et ne peut donc pas s’exprimer
numériquement, mais l’ISO 5725 donne des caractéristiques pour l’étroitesse de l’accord.
Note 2 à l’article : La justesse de mesure varie en sens inverse de l’erreur systématique, mais n’est pas liée à
l’erreur de mesure aléatoire.
Note 3 à l’article : Il convient de ne pas utiliser « exactitude de mesure » pour la « justesse de mesure ».
[SOURCE : VIM3:2012, 2.14]
ISO/DIS 15193:2023(F)
3.24
système de mesure
ensemble d’un ou plusieurs instruments de mesure et souvent d’autres dispositifs, comprenant
si nécessaire réactifs et alimentations, assemblés et adaptés pour fournir des informations destinées à
obtenir des valeurs mesurées dans des intervalles spécifiés pour des grandeurs (3.29) de natures
spécifiées
Note 1 à l’article : Un système de mesure peut consister en un seul instrument de mesure.
[SOURCE : VIM3:2012, 3.2]
3.25
traçabilité métrologique
propriété d’un résultat de mesure selon laquelle ce résultat peut être relié à une référence par
l’intermédiaire d’une chaîne ininterrompue et documentée d’étalonnages, dont chacun contribue à
l’incertitude de mesure (3.41)
Note 1 à l’article : La « référence » mentionnée dans la définition peut être une définition d’une unité de mesure
sous la forme de sa réalisation pratique, une procédure de mesure (3.21), qui indique l’unité de mesure dans le cas
d’une grandeur (3.29), ou un étalon de mesure (3.22).
Note 2 à l’article : La traçabilité métrologique nécessite l’existence d’une hiérarchie d’étalonnage.
Note 3 à l’article : Le terme abrégé « traçabilité » est fréquemment employé pour désigner d’autres concepts,
tels que la « traçabilité d’un échantillon », la « traçabilité d’un document » ou la « traçabilité d’un matériau »,
où intervient l’historique (la « trace ») d’une entité. Il est donc préférable d’utiliser le terme complet « traçabilité
métrologique » s’il y a risque de confusion.
[SOURCE : ISO 17511:2020, 3.30 — Les Notes 1, 2 et 9 (ici note 3) sont conservées de la source.]
3.26
fidélité de mesure
étroitesse de l’accord entre les indications ou les valeurs de grandeur (3.29) mesurées obtenues par des
mesurages répétés du même objet ou d’objets similaires dans des conditions spécifiées
Note 1 à l’article : La fidélité de mesure est en général exprimée numériquement par des caractéristiques telles
que l’écart-type, la variance ou le coefficient de variation dans les conditions spécifiées de mesure.
Note 2 à l’article : Les « conditions spécifiées » peuvent être, par exemple, des conditions de répétabilité,
des conditions de fidélité intermédiaire ou des conditions de reproductibilité (3.35) de la mesure
(voir l’ISO 5725-1:1994).
Note 3 à l’article : La fidélité de mesure sert à définir la répétabilité de mesure, la fidélité intermédiaire de mesure
et la reproductibilité de mesure.
Note 4 à l’article : Le terme « fidélité de mesure » est quelquefois utilisé improprement pour désigner l’exactitude
de mesure.
[SOURCE : Guide ISO/IEC 99:2007, 2.15]
ISO/DIS 15193:2023(F)
3.27
étalon primaire
étalon de mesure primaire
étalon (3.9) établi à l’aide d’une procédure de mesure de référence primaire (3.32) ou créé comme objet
choisi par convention. Un étalon primaire sert de point d’ancrage pour une hiérarchie d’étalonnage
EXEMPLE 1 Étalon primaire de concentration en quantité de matière préparé en dissolvant une quantité de
matière connue d’un matériau de référence (3.31) primaire pour une substance chimique dans un volume connu
de solution.
EXEMPLE 2 Étalon primaire pour les mesurages du rapport molaire d’isotopes, préparé en mélangeant des
quantités de matière connues d’isotopes spécifiés.
Note 1 à l’article : L’ISO 17511:2020 explique en outre le rôle d’un étalon primaire, défini comme la position m.2
au sein d’une hiérarchie d’étalonnage.
[SOURCE : ISO 17511:2020 3.37 – avec ajout de la « Note » et remplacement des Exemples 1 à 5 dans le
texte source par deux exemples pertinents pour la biologie médicale.]
3.28
échantillon primaire
prélèvement
partie discrète d’un liquide ou d’un tissu corporel ou d’un autre échantillon en lien avec le corps
humain, prélevée en vue de l’examen (3.13), l’étude ou l’analyse d’une ou plusieurs grandeurs (3.29) ou
caractéristiques afin de déterminer le caractère de l’ensemble
[SOURCE : ISO 15189:2022, 3.25]
3.29
grandeur
propriété d’un phénomène, d’un corps ou d’une substance, que l’on peut exprimer quantitativement
sous forme d’un nombre et d’une référence
+
EXEMPLE 1 « Plasma (sang) — ion Na ; concentration en quantité de matière égale à 143 mmol/l chez une
personne donnée à un instant donné ».
EXEMPLE 2 Concentration numérique d’érythrocytes dans l’échantillon sanguin (sang total) égale à 5 × 10 /ul
chez une personne donnée à un instant donné.
Note 1 à l’article : Le terme « grandeur » ne doit pas être confondu avec analyte (3.1).
Note 2 à l’article : Une procédure de mesure (3.21) pour laquelle le résultat est exprimé de manière qualitative
(par exemple, séquence nucléotidique dans un échantillon d’ADN « présente » ou « absente » par rapport à un
rapport ou à une échelle de dénombrement avec un seuil de décision prédéterminé) est conforme à la présente
définition du terme grandeur.
[SOURCE : ISO 17511:2020, 3.38]
3.30
récupération
fraction d’analyte (3.1) mesurée sous la forme d’une proportion de la quantité d’analyte (3.1) présente
dans ou ajoutée à la prise d’essai pour analyse (3.3) du matériau pour essai, présentée pour mesure
après préparation de l’échantillon
[14]
Note 1 à l’article : Cité d’après et modifié.
ISO/DIS 15193:2023(F)
3.31
matériau de référence
MR
matériau suffisamment homogène et stable en ce qui concerne des propriétés spécifiées, qui a été
préparé pour être adapté à son utilisation prévue pour un mesurage ou pour l’examen (3.13) de
propriétés qualitatives
EXEMPLE 1 Exemples de MR supports de grandeurs (3.29) :
a) eau de pureté déterminée, dont la viscosité dynamique est utilisée pour l’étalonnage de viscosimètres ;
b) sérum humain sans valeur de grandeur (3.29) assignée pour la concentration en quantité de matière du
cholestérol présent de manière inhérente dans le sérum et qui sert de matériau de contrôle (3.11).
EXEMPLE 2 Exemple de MR supports de propriétés nominales :
ADN contenant une séquence spécifiée de nucléotides.
EXEMPLE 3 Sphères à granulométrie uniforme montées sur une lame de microscope.
Note 1 à l’article : Des MR avec ou sans valeurs de grandeur (3.29) assignées peuvent servir de matériaux de
contrôle (3.11), tandis que seuls des MRC (3.10) avec des valeurs de grandeur (3.29) traçables peuvent servir
d’étalons (3.9) ou de matériaux de contrôle (3.11) de la justesse.
Note 2 à l’article : Les MR comprennent normalement des valeurs de grandeur (3.29), mais ils peuvent également
inclure des propriétés nominales.
Note 3 à l’article : Les MR peuvent avoir des valeurs de grandeur (3.29) assignées qui sont traçables à une unité de
mesure en dehors du Système international d’unités (SI).
Note 4 à l’article : Un MR accompagné d’une documentation délivrée par un organisme faisant autorité et faisant
référence à des procédures de mesure (3.21) valides utilisées pour assigner une valeur de grandeur (3.29) avec
l’incertitude (3.41) de mesure et la traçabilité métrologique (3.25) associées est désigné par MRC (3.10).
[SOURCE : ISO 17511:2020, 3.39 avec modifications]
3.32
procédure de mesure de référence
PMR
procédure de mesure (3.21) considérée comme fournissant des résultats de mesure adaptés à leur
utilisation prévue pour l’évaluation de la justesse de mesure (3.23) de valeurs de grandeur (3.29)
mesurées obtenues à partir d’autres procédures de mesure (3.21) pour une grandeur (3.29) de la même
nature, pour un étalonnage ou pour l’assignation d’une valeur de grandeur (3.29) à un MRC (3.10)
Note 1 à l’article : Les définitions des matériaux de référence (3.31) sont décrites dans l’ISO 15194.
Note 2 à l’article : Une explication du rôle d’un PMR primaire au sein d’une hiérarchie d’étalonnage peut être
trouvée dans l’ISO 17511.
Note 3 à l’article : Lorsque plusieurs procédures de mesure de référence existent pour une grandeur (3.29)
mesurable donnée, il peut être possible de les organiser en hiérarchie selon la taille de l’incertitude (3.41)
de mesure.
Note 4 à l’article : Le présent document fait référence aux procédures de mesure de référence en général et inclut
des procédures de mesure de référence de niveau supérieur comme défini dans l’ISO 17511:2020.
[SOURCE : ISO 17511:2020, 3.40 – modifié]
--------------------
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...