Abstract

The purpose is to give some indications of the way in which accuracy data can be used in various practical situations by: giving a standard method of calculating the repeatability limit, the reproducibility limit and other limits, providing a way of checking the acceptability of test results obtained under repeatability or reproducibility conditions, describing how to assess the stability of results within a laboratory over a period of time, describing how to assess whether a given laboratory is able to use a given standard measurement method in a satisfactory way, describing how to compare alternative measurement methods.

Status
Not Published
Current Stage
6000 - International Standard under publication - 60 Publication
Start Date
12-Nov-2026

Buy Documents

Draft

ISO/PRF 5725-6 - Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values

Release Date:14-Oct-2026
English language (65 pages)
sale 15% off
sale 15% off
Draft

REDLINE ISO/PRF 5725-6 - Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values

Release Date:14-Oct-2026
English language (65 pages)
sale 15% off
sale 15% off

ISO 5725-6 is an ISO standard in the series on accuracy (trueness and precision) of measurement methods and results. It explains how to use established accuracy data in practice, especially for repeatability and reproducibility limits, checking test results, monitoring laboratory stability, assessing whether a laboratory can use a standard method, and comparing alternative methods.

What does ISO 5725-6 specify?

ISO 5725-6 specifies practical uses of accuracy data for measurement methods that produce a single numerical result on a continuous scale. It is organized into clauses on limits, acceptability of test results, laboratory stability, laboratory assessment, and comparison of alternative methods, plus two informative annexes that derive table values.

AnnexWhat it covers
Annex ADerivation of the values in Table 1
Annex BDerivation of the values in Table 2

The scope says the method should already have trueness and precision estimates established in accordance with ISO 5725-1 to ISO 5725-5. That matters in practice because ISO 5725-6 is not the place to establish the accuracy data from scratch - it shows how to use them once they are known.

What are the key requirements of ISO 5725-6?

Repeatability and reproducibility limits

ISO 5725-6 turns precision data into decision limits for comparing results. For two results obtained under repeatability or reproducibility conditions, the standard uses the repeatability limit (r) or the reproducibility limit (R), with a rounded factor of 2.8 times the relevant standard deviation.

In practice, that means a lab does not compare two results to a standard deviation alone. It compares the observed difference to a critical difference that reflects normal variation, which helps separate ordinary spread from a difference that deserves investigation.

Checking the acceptability of test results

ISO 5725-6 gives methods for checking whether test results are acceptable when the measurement method is standardized and the repeatability and reproducibility standard deviations are known. If the range of results exceeds the relevant limit, the document treats one or more results as aberrant and recommends technical investigation.

The clause covers single results, two results, and more than two results under repeatability conditions, and also the case where results come from two laboratories. It also allows the final quoted result to be determined from accepted results, including cases where the median is used.

Stability of results within a laboratory

ISO 5725-6 uses control charts to check whether results remain stable over time within one laboratory. It refers to Shewhart control charts and cumulative sum control (CUSUM) charts, and it notes that CUSUM is more effective for trends or gradual shifts while Shewhart charts are better for sudden changes.

This matters because routine work often changes with time, operator, and equipment. The standard therefore relies on intermediate precision measures rather than true repeatability conditions when checking long-term stability.

Assessing laboratories

ISO 5725-6 uses repeatability and reproducibility standard deviations to assess whether a laboratory can use a standard measurement method satisfactorily. It covers both a laboratory that has not previously been assessed and continued assessment of laboratories that were already approved.

The clause distinguishes between methods where reference materials (RM) exist and methods where they do not. In practice, that lets an assessor set up control experiments and judge whether routine performance still matches the method’s known precision and trueness.

Comparing alternative measurement methods

ISO 5725-6 describes how to compare an international standard method, or a prime candidate for one, with an alternative method. It sets out precision experiments and accuracy experiments, and it also covers the case where method B is a candidate for a routine method.

The practical point is that the comparison shows whether the two methods differ in precision and/or trueness. It does not by itself decide which method is better for a specific use, because cost and equipment availability still matter.

What terms does ISO 5725-6 define?

  • Trueness - The closeness of agreement between the arithmetic mean of many test results and the true or accepted reference value.
  • Precision - The closeness of agreement between test results, showing how much results vary when measurements are repeated.
  • Repeatability limit (r) - A limit used when comparing results obtained under repeatability conditions, so the user can judge whether the difference is larger than expected from normal within-lab variation.
  • Reproducibility limit (R) - A limit used when comparing results obtained under reproducibility conditions, typically between laboratories.
  • Reference material (RM) - A material used as a point of reference for checking trueness or stability when a true value or certified value is available.
  • Laboratory bias - The systematic difference associated with one laboratory, used when assessing trueness and laboratory performance.
  • CUSUM - Cumulative sum control, a charting approach used to detect trends or small shifts over time.
  • Shewhart control chart - A control chart used to watch variation around a centre line and detect sudden departures from stable behaviour.

Who uses ISO 5725-6?

ISO 5725-6 is used by testing laboratories, quality managers, metrology staff, and technical committees that need to turn precision data into practical decisions. It is also relevant to assessors and buyers who need to judge whether a laboratory can use a standardized measurement method reliably.

The standard is aimed at work where results are used in commercial transactions, routine quality control, or method comparison. Its examples and procedures are especially relevant in analytical and industrial testing settings where laboratories monitor long-term performance and compare methods.

What changed in ISO 5725-6 from the previous edition?

  • ISO 5725-6 is the second edition and replaces ISO 5725-6:1994, together with ISO 5725-6:1994/Cor 1:2001.
  • The whole document was revised to align with the latest ISO 5725 series.
  • The descriptions related to the control chart were revised to align with the ISO 7870 series.
  • The calculation processes were made more detailed.
  • The annexes now include information for computing order statistics.

Which standards are used with ISO 5725-6?

  • ISO 3534-1, ISO 3534-2, ISO 3534-3 - provide the statistical vocabulary and symbols used throughout the document.
  • ISO 5725-1 - gives the general principles and definitions for accuracy, trueness, and precision.
  • ISO 5725-2 - provides the basic method for determining repeatability and reproducibility, which ISO 5725-6 uses as the basis for practical decision rules.
  • ISO 5725-3 - is relevant when intermediate conditions apply, especially in stability checking.
  • ISO 5725-4 - is cited in the bibliography as a companion part on trueness.
  • ISO 7870-2 and ISO 7870-4 - support the Shewhart and CUSUM control chart methods used in the stability clauses.
  • ISO/IEC 17025 - is listed in the bibliography as the general competence framework for testing and calibration laboratories.

What does the ISO 5725-6 document contain?

ISO 5725-6 contains methods, decision rules, and worked examples for using accuracy data in day-to-day laboratory practice. Clause 5 gives the basis for limits, Clause 6 covers acceptability of results and final quoted results, Clause 7 covers stability checks, Clause 8 covers laboratory assessment, and Clause 9 covers comparison of alternative methods.

The document also includes a symbol list, a table of c(n) values for comparing medians with arithmetic means, and informative annexes that derive table values. In practice, these parts support reporting, quality control, method verification, and method comparison without reproducing the full statistical development each time.

Relations

Effective Date
10-Sep-2022
Effective Date
10-Sep-2022

Buy Documents

Draft

ISO/PRF 5725-6 - Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values

Release Date:14-Oct-2026
English language (65 pages)
sale 15% off
sale 15% off
Draft

REDLINE ISO/PRF 5725-6 - Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values

Release Date:14-Oct-2026
English language (65 pages)
sale 15% off
sale 15% off

Get Certified

Connect with accredited certification bodies for this standard

BSMI (Bureau of Standards, Metrology and Inspection)

Taiwan's standards and inspection authority.

TAF Taiwan Verified

Sponsored listings

Frequently Asked Questions

ISO 5725-6 is a draft published by the International Organization for Standardization (ISO). Its full title is "Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values". This standard covers: The purpose is to give some indications of the way in which accuracy data can be used in various practical situations by: giving a standard method of calculating the repeatability limit, the reproducibility limit and other limits, providing a way of checking the acceptability of test results obtained under repeatability or reproducibility conditions, describing how to assess the stability of results within a laboratory over a period of time, describing how to assess whether a given laboratory is able to use a given standard measurement method in a satisfactory way, describing how to compare alternative measurement methods.

The purpose is to give some indications of the way in which accuracy data can be used in various practical situations by: giving a standard method of calculating the repeatability limit, the reproducibility limit and other limits, providing a way of checking the acceptability of test results obtained under repeatability or reproducibility conditions, describing how to assess the stability of results within a laboratory over a period of time, describing how to assess whether a given laboratory is able to use a given standard measurement method in a satisfactory way, describing how to compare alternative measurement methods.

ISO 5725-6 is classified under the following ICS (International Classification for Standards) categories: 17.020 - Metrology and measurement in general. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 5725-6 has the following relationships with other standards: It is inter standard links to ISO 5725-6:1994/Cor 1:2001, ISO 5725-6:1994. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ISO 5725-6 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Symbols
  • 0 Given reference value
  • 5 Determination of limits
  • 5.1 Repeatability and reproducibility limits
  • 5.2 Comparisons based on more than or equal to two values
  • 6 Methods for checking the acceptability of test results and determining the final
  • 6.1 General
  • 6.2 Methods for checking the acceptability of test results obtained under repeatability
  • 6.3 Methods for checking the acceptability of test results obtained under both repeatability
  • 7 Methods for checking the stability of test results within a laboratory
  • 7.1 Background
  • 7.2 Methods for checking stability
  • 2 ITO
  • 1 ITO
  • 2 ITO
  • 8 Use of repeatability and reproducibility standard deviations in assessing
  • 8.1 Assessment method
  • 8.2 Evaluation of the use of a measurement method by a laboratory not previously assessed
  • 8.3 Continued assessment of previously approved laboratories
  • 9 Comparison of alternative measurement methods
  • 9.1 Origin of alternative measurement methods
  • 9.2 Purpose of comparing measurement methods
  • 9.3 Method B is a candidate for an alternative standard method ("Standardization
  • 9.4 Accuracy experiment
  • 9.5 Method B is a candidate for a routine method
  • Annex A
  • Annex B
  • Bibliography

Introduction

0.1 ISO 5725 series uses two terms, "trueness" and “precision", to describe the accuracy of a measurement method. "Trueness" refers to the closeness of agreement between the arithmetic mean of a large number of test results and the true or accepted reference value. "Precision" refers to the closeness of agreement between test results.

0.2 "Precision" must be considered because tests performed on presumably identical materials in presumably identical circumstances do not generally yield identical results. This is attributed to unavoidable random errors inherent in every measurement procedure; the factors that influence the outcome of a measurement cannot all be completely controlled. In the practical interpretation of measurement data, this variability has to be considered. For instance, the difference between a test result and some specified value may be within the scope of unavoidable random errors, in which case a real deviation from such a specific value has not been established. Similarly, comparing test results from two batches of material will not indicate a fundamental quality difference if the difference between them can be attributed to the inherent variation in the measurement procedure.

0.3 This document, as for the other parts, considers that measurement error can be divided into a component of error attributed to random causes and a component of error which can be attributed to systematic causes. It is understood that some causes of error can lead to a constant error during a particular set of observations, appearing to be systematic for that set of observations, but can lead to a different error for another set of observations. For example, in an interlaboratory study, the laboratory component of bias is a systematic error for each laboratory, but varies randomly when considered across a collection of laboratories. Similarly, within a laboratory, a set of observations over a limited time can be subject to a systematic error from a given cause (such as choice of equipment or operator) which can in principle vary randomly over sets of observations at other times. It is accordingly important to consider possible changes in systematic effects when comparing results taken under different conditions, such as in different laboratories or over longer periods of time within a laboratory.

0.4 ISO 5725-1 to ISO 5725-5 discuss the background and give methods regarding the assessment of the precision (in terms of the repeatability standard deviation and the reproducibility standard deviation) and the trueness (in terms of the various components of bias) of measurements produced by a standard measurement method. Such an assessment would, however, be pointless if there were no practical uses for the results.

0.5 Given that the accuracy of a measurement method has been established, this document applies that knowledge in practical situations to facilitate commercial transactions and to monitor and improve the operational performance of laboratories.

PROOF/ÉPREUVE

International Standard ISO 5725-6:2026(en) Accuracy (trueness and precision) of measurement methods and results — Part 6:

Use in practice of accuracy values

1 Scope

1.1 This document gives some indications of how accuracy data, including trueness and precision data, can be used in various practical situations by a) giving a standard method of calculating the repeatability limit, the reproducibility limit, and other limits to be used in examining the test results obtained by a standard measurement method, b) providing a way of checking the acceptability of test results obtained under repeatability or reproducibility conditions, c) providing a method of quality control of the operations over time within that laboratory, d) describing how to assess whether a given laboratory can use a given standard measurement method satisfactorily, and e) describing how to compare alternative measurement methods.

1.2 This document is concerned exclusively with measurement methods which yield measurements on a continuous scale and give a single numerical figure as a result. This single figure may be the outcome of a calculation from a set of observations.

NOTE 1 It is assumed that the estimates of trueness and precision for the method have been obtained in accordance with ISO 5725-1 to ISO 5725-5.

NOTE 2 Any additional information regarding the field of application will be given at the beginning of each particular application.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 3534-1, Statistics — Vocabulary and symbols — Part 1: General statistical terms and terms used in probability ISO 3534-2, Statistics — Vocabulary and symbols — Part 2: Applied statistics ISO 3534-3, Statistics — Vocabulary and symbols — Part 3: Design of experiments ISO 5725-1, Accuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definitions ISO 5725-2, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method PROOF/ÉPREUVE

ISO 5725-6:2026(en)

3 Terms and definitions

For the purposes of this document, the definitions given in ISO 3534-1, ISO 3534-2, ISO 3534-3, and ISO 5725-1 apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https:// www .iso .org/ obp
  • IEC Electropedia: available at https:// www .electropedia .org/

4 Symbols

A Factor used to calculate the control limits of X chart Ratio of the standard deviations of the median to the arithmetic mean of n test results cn

C Critical difference d Critical range at the 95 % probability level (for n test results) CC, n

RR,,0950,,95 CL Centre line Cov Covariance (see Annex B) CUSUM Cumulative sum control Factor used to calculate the centre line of a Range chart d

Factor used to calculate the lower control limit of a Range chart when process standard D

deviation is pre specified Factor used to calculate the upper control limit of a Range chart when process standard D

deviation is pre-specified E Expectation (see Annex B) f Critical range factor Probability density function of a range (see Annex A) F , γ -quantile of the F distribution with degrees of freedom ν (numerator) and ν

 AB A B (denominator), where 01 .

g Density function of order statistics (see Annex B) G Grubbs’ test statistic h Standardized decision limit of a CUSUM chart H Observed reference shift of a CUSUM chart k Standardized reference value of a CUSUM chart Reference value for the upper side of a CUSUM chart K

Reference value for the lower side of a CUSUM chart K

L Lower control limit CL m Arbitrary integer (see Annex B) PROOF/ÉPREUVE

ISO 5725-6:2026(en) n Number of test results obtained in one laboratory at one level (i.e. per cell) Number of samples (see Annex A and Annex B) Number of test results obtained in laboratory i at one level (i.e. per cell) n

n Arithmetic mean of the number of each laboratory's test results n .

p Number of laboratories participating in the interlaboratory experiment P Probability (see Annex A) q Number of levels of the test property in the interlaboratory experiment r Repeatability limit Arbitrary integer (see Annex B) R Reproducibility limit R Moving range m RM Reference material s Estimate of a standard deviation S CUSUM statistic U Upper control limit CL Var Variance (see Annex B) w Range w Arithmetic mean of (moving) ranges x Test results (see Clause 7) Function argument (see Annex A and Annex B)

Average of test results (see Clause 7) X Individual value (see Clause 7) Random variable following a normal distribution (see Annex A and Annex B) X Sample (arithmetic) mean of random variables XX,,…,X (see Annex B) 12 n

X Sample median of random variables XX,,…,X (see Annex B) 12 n y Test results (Except Clause 7) Function argument (see Annex B) y Arithmetic mean of each laboratory's test results (Except Clause 7) y Grand mean of test results (Except Clause 7) z Probability density function of the standard normal distribution (see Annex A and Annex B) α Maximum probability of rejecting the null hypothesis when in fact it is true, also referred to as ‘level of significance’ β Probability associated with a risk of failing to detect PROOF/ÉPREUVE

ISO 5725-6:2026(en) δ Laboratory bias ˆ Estimate of a laboratory bias δ λ Detectable difference between two laboratory biases µ True value or accepted reference value of a test property ν Degree of freedom ρ Detectable ratio between precision measures of the two methods when the precision is expressed as the repeatability standard deviation σ True value of a standard deviation

Between-laboratory variance σ L

Repeatability variance σ r

Reproducibility variance σ R φ Detectable ratio between precision measures of the two methods when the precision is expressed as the square root of the between-laboratory mean squares ϕ Inverse function of the cumulative distribution function of the standard normal distribution (see Annex B) Φ Cumulative distribution function of the standard normal distribution (see Annex A and Annex B) 2 2   γ -quantile of the χ distribution with ν degrees of freedom 01 .

Subscripts A, B Numbering for measurement methods i Identifier for a particular laboratory Identifier for a particular sample (see Annex B) Index for summation (see Annex B) Identifier for a particular trial when the number of levels = 1 j Identifier for a particular level when the number of levels ≥ 2 Identifier for a particular sample (see Annex B) Identifier for a trial when the number of levels ≥ 2 k L Between-laboratory (interlaboratory) Lower Lower side P Probability M Between-test-sample min Minimum value max Maximum value n Number of test results obtained in one laboratory at one level (i.e. per cell) PROOF/ÉPREUVE

ISO 5725-6:2026(en) r Repeatability R Reproducibility t Identifier for time Upper Upper side W Within-laboratory (intralaboratory)

0 Given reference value

1, 2, 3, ... For test results, numbering in the order of obtaining them; for other cases (laboratories), as arbitrary identifiers st nd rd (1), (2), (3), ... For test results, (1), (2), (3), … denote the 1 , 2 , 3 ,… etc. order statistic, that is, the st nd rd 1 , 2 , 3 ,… etc. value numbered in the order of increasing magnitude (1|n), (2|n), (3|n)... Same as (1), (2), (3),... etc., but emphasize the sample size is n (see Annex B)

5 Determination of limits

5.1 Repeatability and reproducibility limits

5.1.1 In ISO 5725-2, attention has been focused on estimating the standard deviations associated

with operations under repeatability or reproducibility conditions. However, normal laboratory practice requires examination of the difference(s) observed between two (or more) test results, either within a single laboratory or between two different laboratories. For this purpose, some measure akin to a critical difference is required rather than a standard deviation.

5.1.2 When a quantity is based on sums or differences of n independent estimates, and each has a

standard deviation σ, that resultant quantity will have a standard deviation σ n. Because the reproducibility limit (R) or repeatability limit (r) is for differences between two test results, the associated standard deviation is σ 2 . In normal statistical practice, the critical difference for examining the difference between these two values is f times this standard deviation, that is, fσ 2 . The value of f (the critical range factor) depends on the probability level to be associated with the critical difference and on the shape of the underlying distribution. For the reproducibility and repeatability limits, the probability level is specified as 95 %, and throughout the analysis in ISO 5725 (series), the underlying distribution is assumed to be approximately normal. For a normal distribution at the 95 % probability level, f is 1,96 and then f 2 is 2,77. As this document aims to give some simple "rule of thumb" to be applied by non-statisticians when examining the test results, it seems reasonable to use a rounded value of 2,8, instead of 2,77. 5.1.3 As stated above, the process of estimating precision leads to estimates of the true standard deviations, because the true standard deviations remain unknown. Therefore, in statistical practice, they should be denoted by s rather than σ . However, if the procedures in ISO 5725-1 and ISO 5725-2 are followed, then these estimates will be based on an appreciable number of test results and provide the best information on the true standard deviations. In the applications described later in this document, the symbol s (estimate of a standard deviation) is used for estimates of these standard deviations based on more limited data. Therefore, it is considered best to use the symbol σ to denote the values obtained from a full precision experiment and treat them as the true standard deviations to be used for comparison with other estimates (s).

5.1.4 In view of 5.1.1 to 5.1.3, when examining two single test results obtained under repeatability or reproducibility conditions, the comparison shall be made with the repeatability limit r 28,  or the r reproducibility limit R28,  .

R PROOF/ÉPREUVE

ISO 5725-6:2026(en)

5.2 Comparisons based on more than or equal to two values

5.2.1 Two groups of measurements in one laboratory

It is assumed that, in one laboratory under repeatability conditions, two groups of measurements are performed with the first and second groups of n test results giving an arithmetic mean of y �and n test 1 1 2 results giving an arithmetic mean of y , respectively. Then, the standard deviation of yy is given by

2 12 Formula (1):

 (1)

r

nn

and the critical difference for yy− is given by Formula (2):

1 1 C f28,  (2) d r 2nn2

at the 95 % probability level.

NOTE If nn 1 , then C reduces to r 28,  , as expected.

d 12 r

NOTE To derive Formula (2), the relationship ff  2  is used.

5.2.2 Two groups of measurements in two laboratories

It is assumed that the first laboratory obtains n test results to give an arithmetic mean of y , while the 1 1 second laboratory obtains n test results to give an arithmetic mean of y , in each case under repeatability 2 2 conditions. Then the standard deviation of yy is given in Formula (3):

2 22 2

Lr Lr

nn

1 2

2 

Lr

nn  12 

1 1 22 2 22  1 

Lr r

2nn2

1 1

22 1 (3)

Rr

2nn2  12  and the critical difference for yy− is as shown in Formula (4):

C  f d

1 1

28,  1

Rr

2nn2

PROOF/ÉPREUVE

ISO 5725-6:2026(en)

22 1 1  28,, 28 1 (4)

Rr

2nn2

at the 95 % probability level.

NOTE If nn 1 , then C reduces to R28,  , as expected.

d 12 R

5.2.3 Comparison with a reference value for one laboratory

It is assumed that n test results are obtained under repeatability conditions within one laboratory, which gives an arithmetic mean of y . If there is no specific knowledge of the laboratory component of bias, the comparison with a given reference value µ shall be made by using a standard deviation of y ,

0 0

Lr n 1 1

22 2 22  1

Lr r n

n1

 , (5)

Rr n

Then, the critical difference for y is

C  f d 1 n1

28, 

Rr n

1 22 n1  28,, 28 (6)

Rr n 2   at the 95 % probability level.

5.2.4 Comparison with a reference value for more than two laboratories

It is assumed that p laboratories have obtained n test results with the arithmetic means of y (in each case i i under repeatability conditions) and the grand mean y is computed by p

y y (7)

p i1 If this grand mean is to be compared with a reference value µ , then the standard deviation of y is

0 0 p 11 1 2 2

Lr

p n p

i1 p

2 1 1 22 2 r 22 

Lr r

pn 2p

i1 PROOF/ÉPREUVE

ISO 5725-6:2026(en) p

1 11

 1 (8) Rr

pn p

i1

and the critical difference for y is

C  f d p

1 11

28,  1 Rr

pn 2p

i1

p

1 22 11

 28,, 28 1 (9)

Rr

pn 2p

 i1  at the 95 % probability level.

5.2.5 Summarising the outcome of a comparison

If the absolute difference exceeds the appropriate limit as given in the preceding clauses, then the difference shall be considered significant at the 95 % level of confidence.

6 Methods for checking the acceptability of test results and determining the final

quoted result

6.1 General

6.1.1 The checking method described in this clause should be applied only to the case where the

measurement was carried out according to a measurement method which has been standardized and whose standard deviations, σ and σ , are known. Therefore, when the range of n test results exceeds the r R appropriate limit, as given in Clause 5, it is considered that some or all of the n test results are aberrant. It is recommended that the cause of the aberrant result(s) should be investigated from a technical point of view. However, it may be necessary for commercial reasons to obtain some acceptable value, and in such cases, the test results shall be treated according to the stipulations of this clause.

6.1.2 This clause has been prepared on the assumptions that the test results were obtained under

repeatability and reproducibility conditions and that the probability level to be used is 95 %. If intermediate conditions (see ISO 5725-3) were in force, then it is necessary to replace σ by the appropriate intermediate r measure.

6.1.3 In some cases where the procedures described in 6.2 lead to the median being quoted as the final result, it might be better to abandon the data.

6.2 Methods for checking the acceptability of test results obtained under repeatability

conditions

6.2.1 Single test result

It is not common in commercial practice to obtain only one test result. When only one test result is obtained, it is impossible to make an immediate statistical test of the acceptability of that test result with respect to the given repeatability measure. If there is any suspicion that the test result may not be correct, then a second test result should be obtained. The availability of two test results leads to the more common practice described below.

PROOF/ÉPREUVE

ISO 5725-6:2026(en)

6.2.2 Two test results to start with
6.2.2.1 General

Two test results should be obtained under repeatability conditions. The absolute difference between the two test results should then be compared with the repeatability limit r 28,  . r NOTE In 6.2.2.2 and 6.2.2.3, reference made to expensive or inexpensive measurements is interpreted not only in financial terms but also in terms of whether the measurement is complex, troublesome, and time-consuming.

6.2.2.2 Case where obtaining test results is inexpensive

If the absolute difference between the two test results does not exceed r , then both are considered acceptable, and the final quoted result should be the arithmetic mean of the two test results. If the absolute difference does exceed r , then the laboratory should obtain two further test results. If the range yy of the four test results is less than or equal to the critical range at the 95 %

maxmin probability level for n= 4 , C 4 , then the arithmetic mean of the four test results should be reported

R,09, 5 as the final quoted result. The critical range is computed by C nf n  , (10)

R,09, 5 r where fn is a critical range factor. Table 1 lists the values of fn for n = 2 to 20. If the range of the four

...

Contents

  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 4 Symbols
  • 0 Given reference value
  • 5 Determination of limits
  • 5.1 Repeatability and reproducibility limits
  • 5.2 Comparisons based on more than or equal to two values
  • 6 Methods for checking the acceptability of test results and determining the final
  • 6.1 General
  • 6.2 Methods for checking the acceptability of test results obtained under repeatability
  • 6.3 Methods for checking the acceptability of test results obtained under both repeatability
  • 7 Methods for checking the stability of test results within a laboratory
  • 7.1 Background
  • 7.2 Methods for checking stability
  • 8 Use of repeatability and reproducibility standard deviations in assessing
  • 8.1 Assessment method
  • 8.2 Evaluation of the use of a measurement method by a laboratory not previously assessed
  • 8.3 Continued assessment of previously approved laboratories
  • 9 Comparison of alternative measurement methods
  • 9.1 Origin of alternative measurement methods
  • 9.2 Purpose of comparing measurement methods
  • 9.3 Method B is a candidate for an alternative standard method ("Standardization
  • 9.4 Accuracy experiment
  • 9.5 Method B is a candidate for a routine method
  • Annex A
  • Annex B
  • Bibliography

Introduction

0.1 0.1 ISO 5725 series uses two terms, "trueness" and “precision", to describe the accuracy of a measurement method. "Trueness" refers to the closeness of agreement between the arithmetic mean of a large number of test results and the true or accepted reference value. "Precision" refers to the closeness of agreement between test results.

0.2 0.2 "Precision" must be considered because tests performed on presumably identical materials in presumably identical circumstances do not generally yield identical results. This is attributed to unavoidable random errors inherent in every measurement procedure; the factors that influence the outcome of a measurement cannot all be completely controlled. In the practical interpretation of measurement data, this variability has to be considered. For instance, the difference between a test result and some specified value may be within the scope of unavoidable random errors, in which case a real deviation from such a specific value has not been established. Similarly, comparing test results from two batches of material will not indicate a fundamental quality difference if the difference between them can be attributed to the inherent variation in the measurement procedure.

0.3 0.3 This document, as for the other parts, considers that measurement error can be divided into a component of error attributed to random causes and a component of error which can be attributed to systematic causes. It is understood that some causes of error can lead to a constant error during a particular set of observations, appearing to be systematic for that set of observations, but can lead to a different error for another set of observations. For example, in an interlaboratory study, the laboratory component of bias is a systematic error for each laboratory, but varies randomly when considered across a collection of laboratories. Similarly, within a laboratory, a set of observations over a limited time can be subject to a systematic error from a given cause (such as choice of equipment or operator) which can in principle vary randomly over sets of observations at other times. It is accordingly important to consider possible changes in systematic effects when comparing results taken under different conditions, such as in different laboratories or over longer periods of time within a laboratory.

0.4 0.4 ISO 5725-1 to ISO 5725-5 discuss the background and give methods regarding the assessment of the precision (in terms of the repeatability standard deviation and the reproducibility standard deviation) and the trueness (in terms of the various components of bias) of measurements produced by a standard measurement method. Such an assessment would, however, be pointless if there were no practical uses for the results.

0.5 0.5 Given that the accuracy of a measurement method has been established, this document applies that knowledge in practical situations to facilitate commercial transactions and to monitor and improve the operational performance of laboratories.

ISO/PRF 5725-6:2026(en) Accuracy (trueness and precision) of measurement methods and results — Part 6: Use in practice of accuracy values — Part 6:

Use in practice of accuracy values

1 Scope

1.1 1.1 This document gives some indications of how accuracy data, including trueness and precision data, can be used in various practical situations by a) a) giving a standard method of calculating the repeatability limit, the reproducibility limit, and other limits to be used in examining the test results obtained by a standard measurement method, b) b) providing a way of checking the acceptability of test results obtained under repeatability or reproducibility conditions, c) c) providing a method of quality control of the operations over time within that laboratory, d) d) describing how to assess whether a given laboratory can use a given standard measurement method satisfactorily, and e) e) describing how to compare alternative measurement methods. 1.2 1.2 This document is concerned exclusively with measurement methods which yield measurements on a continuous scale and give a single numerical figure as a result. This single figure may be the outcome of a calculation from a set of observations.

NOTE 1 It is assumed that the estimates of trueness and precision for the method have been obtained in accordance with ISO 5725-1 to ISO 5725-5.

NOTE 2 Any additional information regarding the field of application will be given at the beginning of each particular application.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 3534--1, Statistics — Vocabulary and symbols — Part 1: General statistical terms and terms used in probability ISO 3534--2, Statistics — Vocabulary and symbols — Part 2: Applied statistics ISO 3534-3, Statistics — Vocabulary and symbols — Part 3: Design of experiments ISO 5725--1, Accuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definitions ISO 5725--2, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method

ISO/PRF 5725-6:2026(en)

3 Terms and definitions

For the purposes of this document, the definitions given in ISO 3534-1, ISO 3534-2, ISO 3534-3, and ISO 5725- 1 apply.

ISO and IEC maintain terminologicalterminology databases for use in standardization at the following addresses:

  • — ISO Online browsing platform: available at https://www.iso.org/obp
  • — IEC Electropedia: available at https://www.electropedia.org/

4 Symbols

A Factor used to calculate the control limits of 𝑋𝑋 chart (𝑛𝑛) Ratio of the standard deviations of the median to the arithmetic mean of n test results CDC Critical difference d 𝐶𝐶 ,𝐶𝐶 (𝑛𝑛) Critical range at the 95 % probability level (for n test results) R ,0,95 R ,0,95 CL Centre line Cov Covariance (see Annex BAnnex B)) CUSUM Cumulative sum control 𝑑𝑑 Factor used to calculate the centre line of a Range chart

𝐷𝐷 Factor used to calculate the lower control limit of a Range chart when process standard

deviation is pre specified 𝐷𝐷 Factor used to calculate the upper control limit of a Range chart when process standard

deviation is pre-specified E Expectation (see Annex BAnnex B)) f Critical range factor Probability density function of a range (see Annex AAnnex A)) 𝐹𝐹 (𝜈𝜈 ,𝜈𝜈 ) -𝛾𝛾-quantile of the F distribution with degrees of freedom 𝜈𝜈 (numerator) and 𝜈𝜈 𝛾𝛾 𝐴𝐴 𝐵𝐵 𝐴𝐴 𝐵𝐵 (denominator), where .0≤𝛾𝛾≤ 1.

g Density function of order statistics (see Annex BAnnex B)) G Grubbs’ test statistic h Standardized decision limit of a CUSUM chart H Observed reference shift of a CUSUM chart k Standardized reference value of a CUSUM chart 𝐾𝐾 Reference value for the upper side of a CUSUM chart

𝐾𝐾 Reference value for the lower side of a CUSUM chart

L Lower control limit CL m Arbitrary integer (see Annex BAnnex B)) n Number of test results obtained in one laboratory at one level (i.e. per cell) Number of samples (see Annex A and Annex BAnnex A, Annex B)) 𝑛𝑛 Number of test results obtained in laboratory i at one level (i.e. per cell) 𝑖𝑖

ISO/PRF 5725-6:2026(en) 𝑛𝑛¯ Arithmetic mean of the number of each laboratory's test results 𝑛𝑛 . 𝑖𝑖 p Number of laboratories participating in the interlaboratory experiment P Probability (see Annex AAnnex A)) q Number of levels of the test property in the interlaboratory experiment r Repeatability limit Arbitrary integer (see Annex BAnnex B)) R Reproducibility limit R Moving range m RM Reference material s Estimate of a standard deviation S CUSUM statistic U Upper control limit CL Var Variance (see Annex BAnnex B)) w Range 𝑤𝑤¯ Arithmetic mean of (moving) ranges x Test results (see Clause 7Clause 7)) Function argument (see Annex A and Annex BAnnex A, Annex B)) 𝑥𝑥¯ Average of test results (see Clause 7Clause 7)) X Individual value (see Clause 7Clause 7)) Random variable following a normal distribution (see Annex A and Annex BAnnex A, Annex B))

𝑋𝑋 Sample (arithmetic) mean of random variables 𝑋𝑋 ,𝑋𝑋 , … ,𝑋𝑋 (see Annex B (Annex B)) 1 2 𝑛𝑛

𝑋𝑋 Sample median of random variables 𝑋𝑋 ,𝑋𝑋 , … ,𝑋𝑋 (see Annex B (Annex B)) 1 2 𝑛𝑛 y Test results (Except Clause 7Clause 7)) Function argument (see Annex BAnnex B)) 𝑦𝑦¯ Arithmetic mean of each laboratory's test results (Except Clause 7Clause 7))

𝑦𝑦¯ Grand mean of test results (Except Clause 7Clause 7)) 𝑧𝑧 Probability density function of the standard normal distribution (see Annex A and Annex BAnnex A, Annex B)) 𝛼𝛼 Maximum probability of rejecting the null hypothesis when in fact it is true, also referred to as ‘level of significance’ 𝛽𝛽 Probability associated with a risk of failing to detect 𝛿𝛿 Laboratory bias

Estimate of a laboratory bias 𝛿𝛿 𝜆𝜆 Detectable difference between two laboratory biases 𝜇𝜇 True value or accepted reference value of a test property 𝜈𝜈 Degree of freedom 𝜌𝜌 Detectable ratio between precision measures of the two methods when the precision is expressed as the repeatability standard deviation

ISO/PRF 5725-6:2026(en) 𝜎𝜎 True value of a standard deviation

𝜎𝜎 Between-laboratory variance 𝐿𝐿

𝜎𝜎 Repeatability variance 𝑟𝑟

Reproducibility variance 𝜎𝜎 𝑅𝑅 𝜙𝜙 Detectable ratio between precision measures of the two methods when the precision is expressed as the square root of the between-laboratory mean squares 𝜑𝜑 Inverse function of the cumulative distribution function of the standard normal distribution (see Annex BAnnex B)) 𝛷𝛷 Cumulative distribution function of the standard normal distribution (see Annex A and Annex BAnnex A, Annex B)) 2 2 𝜒𝜒 (𝜈𝜈) -𝛾𝛾-quantile of the 𝜒𝜒 distribution with 𝜈𝜈 degrees of freedom .(0≤𝛾𝛾≤ 1). 𝛾𝛾 Subscripts A, B Numbering for measurement methods i Identifier for a particular laboratory Identifier for a particular sample (see Annex BAnnex B)) Index for summation (see Annex BAnnex)) j Identifier for a particular trial when the number of levels = 1 Identifier for a particular level when the number of levels ≥ 2 Identifier for a particular sample (see Annex BAnnex B)) k Identifier for a trial when the number of levels ≥ 2 L Between-laboratory (interlaboratory) Lower Lower side P Probability M Between-test-sample min Minimum value max Maximum value n Number of test results obtained in one laboratory at one level (i.e. per cell) r Repeatability R Reproducibility t Identifier for time Upper Upper side W Within-laboratory (intralaboratory)

0 Given reference value

1, 2, 3, ... For test results, numbering in the order of obtaining them; for other cases (laboratories), as arbitrary identifiers st nd rd (1), (2), (3), ... For test results, (1), (2), (3), … denote the 1 , 2 , 3 ,… etc. order statistic, that is, the st nd rd 1 , 2 , 3 ,… etc. value numbered in the order of increasing magnitude (1|n), (2|n), (3|n)... Same as (1), (2), (3),... etc., but emphasize the sample size is n (see Annex BAnnex B))

ISO/PRF 5725-6:2026(en)

5 Determination of limits

5.1 Repeatability and reproducibility limits

5.1.1 5.1.1 In ISO 5725-2, attention has been focused on estimating the standard deviations associated with operations under repeatability or reproducibility conditions. However, normal laboratory practice requires examination of the difference(s) observed between two (or more) test results, either within a single laboratory or between two different laboratories. For this purpose, some measure akin to a critical difference is required rather than a standard deviation.

5.1.2 5.1.2 When a quantity is based on sums or differences of 𝒏𝒏 independent estimates, and each has a standard deviation ,𝝈𝝈, that resultant quantity will have a standard deviation .𝝈𝝈√𝒏𝒏. Because the reproducibility limit (R) or repeatability limit (r) is for differences between two test results, the associated standard deviation is .𝝈𝝈√𝟐𝟐. In normal statistical practice, the critical difference for examining the difference between these two values is f times this standard deviation, that is, .𝒇𝒇𝝈𝝈 𝟐𝟐. The value of 𝒇𝒇 (the critical range factor) depends on

the probability level to be associated with the critical difference and on the shape of the underlying distribution. For the reproducibility and repeatability limits, the probability level is specified as 95 %, and throughout the analysis in ISO 5725, (series), the underlying distribution is assumed to be approximately normal. For a normal distribution at the 95 % probability level, 𝒇𝒇 is 1,96 and then 𝒇𝒇 𝟐𝟐 is 2,77. As this document

aims to give some simple "rule of thumb" to be applied by non-statisticians when examining the test results, it seems reasonable to use a rounded value of 2,8, instead of 2,77. 5.1.3 5.1.3 As stated above, the process of estimating precision leads to estimates of the true standard deviations, because the true standard deviations remain unknown. Therefore, in statistical practice, they should be denoted by s rather than .𝝈𝝈. However, if the procedures in ISO 5725-1 and ISO 5725-2 are followed, then these estimates will be based on an appreciable number of test results and provide the best information on the true standard deviations. In the applications described later in this document, the symbol s (estimate of a standard deviation) is used for estimates of these standard deviations based on more limited data. Therefore, it is considered best to use the symbol 𝝈𝝈 to denote the values obtained from a full precision experiment and treat them as the true standard deviations to be used for comparison with other estimates (s). 5.1.4 5.1.4 In view of 5.1.15.1.1 to 5.1.35.1.3,, when examining two single test results obtained under repeatability or reproducibility conditions, the comparison shall be made with the repeatability limit 𝒓𝒓 = 𝟐𝟐,𝟖𝟖𝝈𝝈 or the reproducibility limit .𝑹𝑹 =𝟐𝟐,𝟖𝟖𝝈𝝈 .

𝒓𝒓 𝑹𝑹

5.2 Comparisons based on more than or equal to two values

5.2.1 Two groups of measurements in one laboratory

It is assumed that, in one laboratory under repeatability conditions, two groups of measurements are performed with the first and second groups of 𝑛𝑛 test results giving an arithmetic mean of 𝑦𝑦¯ and 𝑛𝑛 test 1 1 2 results giving an arithmetic mean of ,𝑦𝑦¯ , respectively. Then, the standard deviation of (𝑦𝑦¯ −𝑦𝑦¯ ) is given by 2 1 2 Formula (1)Formula (1)::

(1) 1 1

𝜎𝜎 = 𝜎𝜎 ( + ) (1)

𝑟𝑟 𝑛𝑛 𝑛𝑛 1 2 and the critical difference for |𝑦𝑦¯ −𝑦𝑦¯ | is given by Formula (2)Formula (2)::

1 2 (2) 1 1 𝐶𝐶 =𝑓𝑓𝜎𝜎 = 2,8𝜎𝜎� + (2) 𝑑𝑑 𝑟𝑟 2𝑛𝑛 2𝑛𝑛 1 2 at the 95 % probability level.

NOTE If ,𝑛𝑛 =𝑛𝑛 = 1, then C reduces to ,𝑟𝑟 = 2,8𝜎𝜎 , as expected. d 1 2 𝑟𝑟

ISO/PRF 5725-6:2026(en) 𝜎𝜎 NOTE To derive Formula (2) (2),, the relationshiprelationship𝑓𝑓𝜎𝜎 = (𝑓𝑓 2)⋅ ( ) is used.

5.2.2 Two groups of measurements in two laboratories

It is assumed that the first laboratory obtains 𝑛𝑛 test results to give an arithmetic mean of ,𝑦𝑦¯ , while the second 1 1 laboratory obtains 𝑛𝑛 test results to give an arithmetic mean of ,𝑦𝑦¯ , in each case under repeatability conditions. 2 2 Then the standard deviation of (𝑦𝑦¯ −𝑦𝑦¯ ) is given in Formula (3)Formula (3)::

1 2

(3) 1 1 2 2 2 2 𝜎𝜎 =�(𝜎𝜎 + 𝜎𝜎 ) + (𝜎𝜎 + 𝜎𝜎 ) 𝐿𝐿 𝑟𝑟 𝐿𝐿 𝑟𝑟 𝑛𝑛 𝑛𝑛 1 2 1 1 2 2     =�2𝜎𝜎 +𝜎𝜎 ( + ) 𝐿𝐿 𝑟𝑟 𝑛𝑛 𝑛𝑛 1 2 1 1 2 2 2     =�2(𝜎𝜎 +𝜎𝜎 )− 2𝜎𝜎 (1− − ) 𝐿𝐿 𝑟𝑟 𝑟𝑟 2𝑛𝑛 2𝑛𝑛 1 2 1 1 2 2     = 2𝜎𝜎 − 2𝜎𝜎 (1− − ) (3)

𝑅𝑅 𝑟𝑟 2𝑛𝑛 2𝑛𝑛 1 2 and the critical difference for |𝑦𝑦¯ −𝑦𝑦¯ | is as shown in Formula (4)Formula (4)::

1 2

(4) 𝐶𝐶 =𝑓𝑓𝜎𝜎 d 1 1 2 2

= 2,8 𝜎𝜎 −𝜎𝜎 (1− − ) 𝑅𝑅 𝑟𝑟 2𝑛𝑛 2𝑛𝑛 1 2 1 1 2 2      =�(2,8𝜎𝜎 ) − (2,8𝜎𝜎 ) (1− − ) (4) 𝑅𝑅 𝑟𝑟 2𝑛𝑛 2𝑛𝑛 1 2 at the 95 % probability level.

NOTE If ,𝑛𝑛 =𝑛𝑛 = 1, then CDCd reduces to ,𝑅𝑅 = 2,8𝜎𝜎 , as expected. 1 2 𝑅𝑅

5.2.3 Comparison with a reference value for one laboratory

It is assumed that 𝑛𝑛 test results are obtained under repeatability conditions within one laboratory, which gives an arithmetic mean of .𝑦𝑦¯. If there is no specific knowledge of the laboratory component of bias, the comparison with a given reference value 𝜇𝜇 shall be made by using a standard deviation of ,(𝑦𝑦¯−𝜇𝜇 ), 0 0

ISO/PRF 5725-6:2026(en)

(5)

2 2

𝜎𝜎 = 𝜎𝜎 + 𝜎𝜎 𝐿𝐿 𝑟𝑟 𝑛𝑛 1 1 2 2 2

= 2(𝜎𝜎 +𝜎𝜎 )− 2𝜎𝜎 (1− ) 𝐿𝐿 𝑟𝑟 𝑟𝑟 𝑛𝑛 √2 𝑛𝑛−1 2 2     =�𝜎𝜎 −𝜎𝜎 ( ), (5) 𝑅𝑅 𝑟𝑟 𝑛𝑛 Then, the critical difference for |𝑦𝑦¯−𝜇𝜇 | is

(6) 𝐶𝐶 =𝑓𝑓𝜎𝜎 d 1 𝑛𝑛− 1 2 2

= 2,8 𝜎𝜎 −𝜎𝜎 ( ) 𝑟𝑟 𝑅𝑅 𝑛𝑛 √2 1 𝑛𝑛−1 2 2       = �(2,8𝜎𝜎 ) − (2,8𝜎𝜎 ) ( ) (6) 𝑅𝑅 𝑟𝑟 2 𝑛𝑛

at the 95 % probability level.

5.2.4 Comparison with a reference value for more than two laboratories

It is assumed that p laboratories have obtained 𝑛𝑛 test results with the arithmetic means of 𝑦𝑦¯ (in each case 𝑖𝑖 𝑖𝑖

under repeatability conditions) and the grand mean 𝑦𝑦¯ is computed by

𝑝𝑝

𝑦𝑦¯ = 𝑦𝑦¯ (7) 𝑖𝑖 𝑖𝑖=1 𝑝𝑝

If this grand mean is to be compared with a reference value ,𝜇𝜇 , then the standard deviation of (𝑦𝑦¯−𝜇𝜇 ) is 0 0

𝑝𝑝 1 1

2 2

𝜎𝜎 = 𝜎𝜎 + 𝜎𝜎 � 𝑟𝑟 𝐿𝐿

𝑝𝑝 𝑝𝑝 𝑛𝑛 𝑖𝑖 𝑖𝑖=1

ISO/PRF 5725-6:2026(en) 𝑝𝑝

1 2𝜎𝜎 1 𝑟𝑟 2 2 2

= 2(𝜎𝜎 +𝜎𝜎 )− 2𝜎𝜎 + � 𝑟𝑟 𝑟𝑟 𝐿𝐿 𝑝𝑝 𝑛𝑛 2𝑝𝑝 � 𝑖𝑖 𝑖𝑖=1 1 1 1 𝑝𝑝 2 2     = 𝜎𝜎 −𝜎𝜎 (1− ∑ ) (8)

𝑅𝑅 𝑟𝑟 𝑖𝑖=1 𝑝𝑝 𝑝𝑝 𝑛𝑛 √ 𝑖𝑖

and the critical difference for |𝑦𝑦¯−𝜇𝜇 | is

(9) 𝐶𝐶 =𝑓𝑓𝜎𝜎 d 𝑝𝑝 1 1 1 2 2

= 2,8 𝜎𝜎 −𝜎𝜎 (1− � ) 𝑅𝑅 𝑟𝑟 𝑝𝑝 𝑛𝑛 �2𝑝𝑝 𝑖𝑖 𝑖𝑖=1 1 1 𝑝𝑝 1 2 2     = (2,8𝜎𝜎 ) − (2,8𝜎𝜎 ) (1− ∑ ) (9)

𝑅𝑅 𝑟𝑟 𝑖𝑖=1 �2𝑝𝑝 𝑝𝑝 𝑛𝑛 𝑖𝑖 at the 95 % probability level.

5.2.5 Summarising the outcome of a comparison

If the absolute difference exceeds the appropriate limit as given in the preceding clauses, then the difference shall be considered significant at the 95 % level of confidence.

6 Methods for checking the acceptability of test results and determining the final

quoted result

6.1 General

6.1.1 6.1.1 The checking method described in this clause should be applied only to the case where the measurement was carried out according to a measurement method which has been standardized and whose standard deviations, 𝝈𝝈 and ,𝝈𝝈 , are known. Therefore, when the range of n test results exceeds the 𝒓𝒓 𝑹𝑹 appropriate limit, as given in Clause 5Clause 5,, it is considered that some or all of the n test results are aberrant. It is recommended that the cause of the aberrant result(s) should be investigated from a technical point of view. However, it may be necessary for commercial reasons to obtain some acceptable value, and in such cases, the test results shall be treated according to the stipulations of this clause. 6.1.2 6.1.2 This clause has been prepared on the assumptions that the test results were obtained under repeatability and reproducibility conditions and that the probability level to be used is 95 %. If intermediate conditions (see ISO 5725-3) were in force, then it is necessary to replace 𝝈𝝈 by the appropriate intermediate 𝒓𝒓 measure.

6.1.3 6.1.3 In some cases where the procedures described in 6.26.2 lead to the median being quoted as the final result, it might be better to abandon the data.

ISO/PRF 5725-6:2026(en)

6.2 Methods for checking the acceptability of test results obtained under repeatability

conditions

6.2.1 Single test result

It is not common in commercial practice to obtain only one test result. When only one test result is obtained, it is impossible to make an immediate statistical test of the acceptability of that test result with respect to the given repeatability measure. If there is any suspicion that the test result may not be correct, then a second test result should be obtained. The availability of two test results leads to the more common practice described below.

...