ISO 81060-2:2013
(Main)Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type
Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type
ISO 81060-2:2013 specifies the requirements and methods for the clinical investigation of me equipment used for the intermittent non-invasive automated estimation of the arterial blood pressure by utilizing a cuff. ISO 81060-2:2013 is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation. ISO 81060-2:2013 covers sphygmomanometers intended for use in all patient populations and all conditions of use. ISO 81060-2:2013 specifies additional disclosure requirements for the accompanying documents of sphygmomanometers that have undergone clinical investigation according to ISO 81060-2:2013.
Sphygmomanomètres non invasifs — Partie 2: Validation clinique pour type à mesurage automatique
L'ISO 81060-2:2013 précise les exigences et les méthodes d'investigation clinique des appareils électromédicaux utilisés pour estimer ponctuellement, de manière non invasive et automatique, la pression artérielle au moyen d'un brassard. L'ISO 81060-2:2013 est applicable à tous les sphygmomanomètres qui captent ou affichent des pulsations, des flux ou des sons pour l'estimation, l'affichage ou l'enregistrement de la pression artérielle. Il n'est pas nécessaire que ces sphygmomanomètres aient un dispositif de gonflage automatique du brassard. L'ISO 81060-2:2013 couvre les sphygmomanomètres utilisables pour toutes les populations de patients et dans toutes les conditions d'emploi. L'ISO 81060-2:2013 spécifie des exigences supplémentaires de divulgation d'informations pour les documents d'accompagnement des sphygmomanomètres ayant fait l'objet d'une investigation clinique conformément à la'ISO 81060-2:2013.
General Information
Relations
Frequently Asked Questions
ISO 81060-2:2013 is a standard published by the International Organization for Standardization (ISO). Its full title is "Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type". This standard covers: ISO 81060-2:2013 specifies the requirements and methods for the clinical investigation of me equipment used for the intermittent non-invasive automated estimation of the arterial blood pressure by utilizing a cuff. ISO 81060-2:2013 is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation. ISO 81060-2:2013 covers sphygmomanometers intended for use in all patient populations and all conditions of use. ISO 81060-2:2013 specifies additional disclosure requirements for the accompanying documents of sphygmomanometers that have undergone clinical investigation according to ISO 81060-2:2013.
ISO 81060-2:2013 specifies the requirements and methods for the clinical investigation of me equipment used for the intermittent non-invasive automated estimation of the arterial blood pressure by utilizing a cuff. ISO 81060-2:2013 is applicable to all sphygmomanometers that sense or display pulsations, flow or sounds for the estimation, display or recording of blood pressure. These sphygmomanometers need not have automatic cuff inflation. ISO 81060-2:2013 covers sphygmomanometers intended for use in all patient populations and all conditions of use. ISO 81060-2:2013 specifies additional disclosure requirements for the accompanying documents of sphygmomanometers that have undergone clinical investigation according to ISO 81060-2:2013.
ISO 81060-2:2013 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 81060-2:2013 has the following relationships with other standards: It is inter standard links to ISO 81060-2:2018, ISO 81060-2:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
You can purchase ISO 81060-2:2013 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.
Standards Content (Sample)
INTERNATIONAL ISO
STANDARD 81060-2
Second edition
2013-05-01
Non-invasive sphygmomanometers —
Part 2:
Clinical investigation of automated
measurement type
Sphygmomanomètres non invasifs —
Partie 2: Validation clinique pour type à mesurage automatique
Reference number
©
ISO 2013
© ISO 2013
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2013 – All rights reserved
Contents Page
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 General requirements for CLINICAL INVESTIGATIONS . 2
5 CLINICAL INVESTIGATION with an auscultatory REFERENCE SPHYGMOMANOMETER . 3
6 CLINICAL INVESTIGATION with REFERENCE INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT . 17
7 * Pregnant (including pre-eclamptic) PATIENT populations . 22
Annex A (informative) Rationale and guidance . 23
Annex B (normative) Target heart rates for exercise stress testing . 35
Annex C (informative) Reference to the essential principles . 36
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
This second edition cancels and replaces the first edition (ISO 81060-2:2009), subclauses 5.2.4.3.1 and 6.2.4
of which have been technically revised. Numerous clarifications have been added and kPa equivalent values
for the mmHg values have been included in the standard, including the Criterion 2 requirements of 5.2.4.1.2. It
also incorporates the Technical Corrigendum ISO 81060-2:2009/Cor 1:2011.
ISO 81060-2 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment,
Subcommittee SC 3, Lung ventilators and related equipment, in collaboration with Technical Committee
IEC/TC 62, Electrical equipment in medical practice, Subcommittee 62D, Electromedical equipment, in
accordance with ISO/IEC mode of cooperation 5.
ISO 81060 consists of the following parts, under the general title Non-invasive sphygmomanometers:
Part 1: Requirements and test methods for non-automated measurement type
Part 2: Clinical investigation of automated measurement type
In this document, the following print types are used:
requirements, compliance with which can be verified, and definitions: roman type;
notes and examples: smaller roman type;
test methods: italic type;
terms defined in this document: SMALL CAPITALS TYPE.
Throughout this document, text for which a rationale is provided in Annex A is indicated by an asterisk (*).
The attention of Member Bodies and National Committees is drawn to the fact that equipment manufacturers
and testing organizations may need a transitional period following publication of a new, amended or revised
ISO or IEC publication in which to make products in accordance with the new requirements and to equip
themselves for conducting new or revised tests. It is the recommendation of ISO/TC 121 and IEC/TC 62 that
the content of this part of ISO 81060 not be adopted for mandatory implementation nationally earlier than 3
years from the date of publication for equipment newly designed, and not earlier than 5 years from the date of
publication for equipment already in production.
iv © ISO 2013 – All rights reserved
Introduction
Determination of BLOOD PRESSURE is an important procedure that is clinically used to assess the status of a
PATIENT.
Frequent determination of BLOOD PRESSURE is routine during anaesthesia. BLOOD PRESSURE serves to aid in
drug titration and fluid management and to provide warning of conditions that could affect PATIENT morbidity
and mortality.
INTERNATIONAL STANDARD ISO 81060-2:2013(E)
Non-invasive sphygmomanometers —
Part 2:
Clinical investigation of the automated measurement type
1 Scope
This part of ISO 81060 specifies the requirements and methods for the CLINICAL INVESTIGATION of
ME EQUIPMENT used for the intermittent non-invasive automated estimation of the arterial BLOOD PRESSURE by
utilizing a CUFF.
This part of ISO 81060 is applicable to all SPHYGMOMANOMETERS that sense or display pulsations, flow or
sounds for the estimation, display or recording of BLOOD PRESSURE. These SPHYGMOMANOMETERS need not
have automatic CUFF inflation.
This part of ISO 81060 covers SPHYGMOMANOMETERS intended for use in all PATIENT populations (e.g. all age
and weight ranges), and all conditions of use (e.g. ambulatory BLOOD PRESSURE monitoring, stress testing
BLOOD PRESSURE monitoring and BLOOD PRESSURE monitors for the HOME HEALTHCARE ENVIRONMENT for self-
measurement as well as use in a professional healthcare facility).
EXAMPLE AUTOMATED SPHYGMOMANOMETER as given in IEC 80601-2-30 undergoing CLINICAL INVESTIGATION according to
this part of ISO 81060.
This part of ISO 81060 specifies additional disclosure requirements for the ACCOMPANYING DOCUMENTS of
SPHYGMOMANOMETERS that have undergone CLINICAL INVESTIGATION according to this part of ISO 81060.
This part of ISO 81060 is not applicable to CLINICAL INVESTIGATIONS of NON-AUTOMATED SPHYGMOMANOMETERS
as given in ISO 81060-1 or INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT as given in IEC 60601-2-34.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 14155:2011, Clinical investigation of medical devices for human subjects — Good clinical practice
ISO 81060-1, Non-invasive sphygmomanometers — Part 1: Requirements and test methods for non-
automated measurement type
IEC 80601-2-30:2009, Medical electrical equipment — Part 2-30: Particular requirements for basic safety and
essential performance of automated non-invasive sphygmomanometers
IEC 60601-1:2005, Medical electrical equipment — Part 1: General requirements for basic safety and
essential performance
Amendment 1:2012
IEC 60601-1-11:2010, Medical electrical equipment — Part 1-11: General requirements for basic safety and
essential performance — Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in home care applications
IEC 60601-2-34:2011, Medical electrical equipment — Part 2-34: Particular requirements for the basic safety
and essential performance of invasive blood pressure monitoring equipment
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 14155, IEC 80601-2-30,
IEC 60601-1, IEC 60601-1-11, IEC 60601-2-34 and the following apply.
NOTE For convenience, an alphabetized index of defined terms is found beginning on page 40.
3.1
REFERENCE, adj
established accuracy used for the CLINICAL INVESTIGATION of other instruments
3.2
SPHYGMOMANOMETER
ME EQUIPMENT for non-invasive estimation of systemic arterial BLOOD PRESSURE
3.3
SPHYGMOMANOMETER-UNDER-TEST
SPHYGMOMANOMETER undergoing CLINICAL INVESTIGATION
4 General requirements for CLINICAL INVESTIGATIONS
4.1 CLINICAL INVESTIGATION methods
SPHYGMOMANOMETERS other than NON-AUTOMATED SPHYGMOMANOMETERS shall undergo CLINICAL INVESTIGATION
either by using a non-invasive (auscultatory) REFERENCE SPHYGMOMANOMETER or by using REFERENCE INVASIVE
BLOOD PRESSURE MONITORING EQUIPMENT according to this part of ISO 81060 in each mode of operation.
EXAMPLE 1 Adult and neonatal modes.
EXAMPLE 2 Slow and fast CUFF deflation rate modes.
A CLINICAL INVESTIGATION shall be considered a TYPE TEST.
Consider compliance with the requirements of this subclause to exist when the criteria of the relevant
inspections and tests in this part of ISO 81060 are met.
4.2 Good clinical practice
All CLINICAL INVESTIGATIONS shall comply with the requirements of ISO 14155. CLINICAL INVESTIGATION with
REFERENCE INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT should not be used for PATIENTS or subjects
solely for the purpose of investigating SPHYGMOMANOMETER performance.
NOTE Some authorities having jurisdiction have additional requirements.
The requirements of this International Standard, which are more specific than the corresponding requirements
of ISO 14155, shall prevail.
Check compliance by application of the requirements of ISO 14155.
2 © ISO 2013 – All rights reserved
5 CLINICAL INVESTIGATION with an auscultatory REFERENCE SPHYGMOMANOMETER
5.1 Subject requirements
5.1.1 * Number
An auscultatory REFERENCE SPHYGMOMANOMETER CLINICAL INVESTIGATION shall consist of a minimum of 85
subjects. If not otherwise specified, at least three valid BLOOD PRESSURE DETERMINATIONS shall be taken for
each subject. There shall be a minimum of 255 valid paired BLOOD PRESSURE DETERMINATIONS.
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
5.1.2 * Gender distribution
At least 30 % of the subjects shall be male and at least 30 % of the subjects shall be female.
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
5.1.3 * Age distribution
For a SPHYGMOMANOMETER intended for use on adults and/or adolescent PATIENTS, the age of every subject
included in the CLINICAL INVESTIGATION shall be greater than 12 years.
NOTE 1 Minimum total of 85 subjects.
For a SPHYGMOMANOMETER additionally intended for use in children, 35 child subjects aged between 3 years
and 12 years shall be included in the CLINICAL INVESTIGATION.
NOTE 2 Minimum total of 85 subjects.
If the SPHYGMOMANOMETER has a special mode for children, in that mode, children shall be considered a
special PATIENT population (see 5.1.6). In such a study, children are exempt from the BLOOD PRESSURE
distribution requirements of 5.1.5.
Children aged less than 3 years shall not be included in a CLINICAL INVESTIGATION utilizing auscultatory
DETERMINATIONS by observers with a REFERENCE SPHYGMOMANOMETER.
Check compliance by inspection of the ACCOMPANYING DOCUMENT and the CLINICAL INVESTIGATION REPORT.
5.1.4 * Limb size distribution
For a SPHYGMOMANOMETER intended for use with a single CUFF size:
at least 40 % of the subjects shall have a limb circumference which lies within the upper half of the
specified range of use of the CUFF and at least 40 % shall have a limb circumference within the lower half;
and
at least 20 % of the subjects shall have a limb circumference which lies within the upper quarter of the
specified range of use of the CUFF and at least 20 % shall have a limb circumference within the lower
quarter.
For a SPHYGMOMANOMETER intended for use with multiple CUFF sizes, each CUFF size shall be tested on at
least of the subjects, where n is the number of CUFF sizes.
2 n
Check compliance by inspection of the ACCOMPANYING DOCUMENT and the CLINICAL INVESTIGATION REPORT.
5.1.5 * BLOOD PRESSURE distribution
At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE ≤ 100 mmHg
(13,33 kPa).
At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE ≥ 160 mmHg
(21,33 kPa).
At least 20 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE
≥ 140 mmHg (18,66 kPa).
At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE ≤ 60 mmHg
(8,0 kPa).
At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE
≥ 100 mmHg (13,33 kPa).
At least 20 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE
≥ 85 mmHg (11,33 kPa).
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
5.1.6 * Special PATIENT populations
A SPHYGMOMANOMETER that is intended for use in special PATIENT populations where there is OBJECTIVE
EVIDENCE that the accuracy of the SPHYGMOMANOMETER might be problematic in those PATIENT populations,
shall undergo CLINICAL INVESTIGATION in those PATIENT populations.
NOTE Clause 7 has a specific example of a special PATIENT population with specific requirements.
If the SPHYGMOMANOMETER has undergone CLINICAL INVESTIGATION according to the requirements of 5.1.1
and 5.2, it shall then undergo CLINICAL INVESTIGATION in at least an additional 35 special population subjects. If
the SPHYGMOMANOMETER has not previously undergone CLINICAL INVESTIGATION according to the requirements
of 5.1.1 and 5.2, the CLINICAL INVESTIGATION in accordance with the requirements of 5.1.1 and 5.2 shall consist
only of subjects from the special PATIENT population.
The special PATIENT population shall be defined in clear terms and address the following attributes: gender
(see 5.1.2), age (see 5.1.3), limb size (see 5.1.4) and BLOOD PRESSURE (see 5.1.5). A summary of this
information shall be disclosed in the instructions for use.
Check compliance by inspection of the instructions for use and the CLINICAL INVESTIGATION REPORT.
5.2 CLINICAL INVESTIGATION method with a REFERENCE SPHYGMOMANOMETER
5.2.1 * Subject preparation
Unless otherwise indicated by the instructions for use of the SPHYGMOMANOMETER-UNDER-TEST, position the
subject such that the subject:
is comfortable;
EXAMPLE Comfortably seated with legs uncrossed and feet flat on the floor.
has the back, elbow and forearm supported;
has the middle of the CUFF at the level of the right atrium of the heart.
4 © ISO 2013 – All rights reserved
Recommend that the subject be as relaxed as possible and that the subject avoid talking during the entire
procedure. Before the first reading is taken, 5 min should elapse.
NOTE Additional details can be found in Reference [32].
5.2.2 * Observer preparation
Observers should be trained in using a proper methodology for performing a resting BLOOD PRESSURE
DETERMINATION by utilizing an accepted clinical protocol for BLOOD PRESSURE measurement.
References [8], [28], [29], [32] and [45] contain additional information. Observers should have sufficient
practice in performing BLOOD PRESSURE DETERMINATIONS.
Each observer's recording of observations of the REFERENCE SPHYGMOMANOMETER shall not be visible to the
other observer. The readings of the SPHYGMOMANOMETER-UNDER-TEST shall not be visible to either of these
observers.
EXAMPLE 1 Utilizing a third observer for recording the readings of the SPHYGMOMANOMETER-UNDER-TEST.
EXAMPLE 2 Utilizing an electronic means for recording the readings of the SPHYGMOMANOMETER-UNDER-TEST.
Instruct the observers to determine the DIASTOLIC BLOOD PRESSURE as the last audible Korotkoff sound (fifth
phase or K5), except when Korotkoff sounds are still audible with the CUFF deflated or in children between 3
years and 12 years of age, where the fourth phase (K4) is used. If K4 is not audible in a child, either K5 is
used or the subject is excluded.
NOTE Other than for children, K4 should be reserved for subjects in whom there is a large discrepancy between muffling
and disappearance (with the latter at times approaching zero mmHg).
Instruct the observers to record which Korotkoff sound has been used for the DETERMINATION of DIASTOLIC
BLOOD PRESSURE.
The Korotkoff sound used for DETERMINATION of DIASTOLIC BLOOD PRESSURE in the CLINICAL INVESTIGATION shall
be disclosed in the instructions for use of a SPHYGMOMANOMETER.
EXAMPLE 3 K5 was used on 65 subjects and K4 was used on 20 subjects.
5.2.3 * REFERENCE DETERMINATION
Two observers shall make simultaneous BLOOD PRESSURE DETERMINATIONS on each subject using a double
stethoscope.
Unless the SPHYGMOMANOMETER-UNDER-TEST is intended for use during significantly irregular heart rhythm and
if either observer detects significantly irregular heart rhythm, that DETERMINATION shall be excluded.
EXAMPLES Bigeminy, trigeminy, isolated ventricular premature beat (VPB), atrial fibrillation.
NOTE 1 Although CLINICAL INVESTIGATION of BLOOD PRESSURE in PATIENTS with atrial fibrillation is clinically important,
there are currently no generally accepted guidelines for determining the BLOOD PRESSURE in such individuals.
Any pair of observers' DETERMINATIONS with a difference greater than 4 mmHg (0,53 kPa) shall be excluded.
The observers' individual values of each DETERMINATION shall be averaged according to Formula (1) to create
the REFERENCE BLOOD PRESSURE DETERMINATION.
pp
ref ref
ii,1 ,2
p (1)
ref
i
where
th
p is the BLOOD PRESSURE determined by observer 1 for the i DETERMINATION;
ref
i ,1
th
p is the BLOOD PRESSURE determined by observer 2 for the i DETERMINATION;
ref
i,2
th
p is the REFERENCE BLOOD PRESSURE for the i DETERMINATION.
ref
i
The observer-to-observer differences shall be reviewed after completing a set of pairs of test-REFERENCE
DETERMINATIONS. If any DETERMINATIONS are excluded, additional pair(s) of DETERMINATIONS shall be taken to
ensure that the required number of valid test-REFERENCE pairs are available. A maximum of eight pairs of
DETERMINATIONS should be taken.
Use a REFERENCE SPHYGMOMANOMETER that complies with the requirements of ISO 81060-1, except that the
maximum permissible error shall be 1 mmHg (0,13 kPa). Reading of the values on the REFERENCE
SPHYGMOMANOMETER should be as accurate as possible. When reading the value on the REFERENCE
SPHYGMOMANOMETER, the observers should avoid parallax errors and rounding.
NOTE 2 Rounding has a negative effect on the results of the CLINICAL INVESTIGATION.
NOTE 3 For the purposes of this part of ISO 81060, the CUFF is considered part of the REFERENCE SPHYGMOMANOMETER.
A CUFF that does not comply with ISO 81060-1 cannot be used.
5.2.4 CLINICAL INVESTIGATION methods
5.2.4.1 Same arm simultaneous method
5.2.4.1.1 * Procedure
This method shall only be used with a SPHYGMOMANOMETER-UNDER-TEST:
that has a CUFF compliant with ISO 81060-1;
that is designed for use on the upper arm; and
where
the continuous linear deflation rate is between 2 mmHg/s (0,27 kPa/s) and 3 mmHg/s (0,40 kPa/s) or
for a SPHYGMOMANOMETER-UNDER-TEST that controls the deflation as a function of the pulse rate, the
deflation rate is between 2 mmHg/pulse (0,27 kPa/pulse) and 3 mmHg/pulse (0,40 kPa/pulse).
Either arm may be utilized.
The SPHYGMOMANOMETER-UNDER-TEST shall not deflate prior to the detection of the REFERENCE DIASTOLIC
BLOOD PRESSURE. The SPHYGMOMANOMETER-UNDER-TEST may be modified to meet this criterion.
NOTE Valid same arm simultaneous DETERMINATIONS require the SPHYGMOMANOMETER-UNDER-TEST to inflate the CUFF
to at least 20 mmHg (2,67 kPa) higher than the actual SYSTOLIC BLOOD PRESSURE, as determined by the REFERENCE
SPHYGMOMANOMETER, and to at least 20 mmHg (2,67 kPa) below the actual DIASTOLIC BLOOD PRESSURE, as determined by
the REFERENCE SPHYGMOMANOMETER.
Perform the following:
a) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
simultaneously determine the subject's BLOOD PRESSURE utilizing the same CUFF and inflation/deflation
cycle (see Figure 1). These data points are not used in the calculation of accuracy of the
SPHYGMOMANOMETER-UNDER-TEST.
b) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and then wait at least
60 s.
6 © ISO 2013 – All rights reserved
EXAMPLES Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
reset command are methods to clear the memory of the previous DETERMINATION.
c) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
simultaneously determine the subject's BLOOD PRESSURE utilizing the same CUFF and inflation/deflation
cycle.
d) Wait at least 60 s between DETERMINATIONS.
e) Repeat c) and d) until the required number of valid DETERMINATIONS has been performed.
All data from a subject shall be excluded if any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS
differ by more than 12 mmHg (1,60 kPa) or if any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS
differ by more than 8 mmHg (1,07 kPa).
Key
1 double stethoscope
2 REFERENCE SPHYGMOMANOMETER display
3 SPHYGMOMANOMETER-UNDER-TEST
Figure 1 — Illustration of same arm simultaneous method
5.2.4.1.2 * Data analysis
The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
a) Criterion 1
For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, x ,
n
of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , no greater than 8,0 mmHg (1,07 kPa) when
n
calculated according to Formula (2) and Formula (3):
n
xp p (2)
n sut ref
ii
n
i1
n
s (3)
xx
n n
i
n1
i1
where
is the mean value of the differences;
x
n
pp = x difference between a paired BLOOD PRESSURE DETERMINATION [DETERMINATION of
sut ref
i
i i
the SPHYGMOMANOMETER-UNDER-TEST REFERENCE BLOOD PRESSURE as calculated according to
Formula (1)];
i is the index for the individual element;
n is the number of DETERMINATIONS;
and s shall be calculated and expressed to 0,1 mmHg (0,01 kPa).
x
n
n
EXAMPLE 1 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an
85 subject study).
EXAMPLE 2 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and12 years (an 85 subject study).
EXAMPLE 3 n = 105 for a SPHYGMOMANOMETER intended for a special INTENDED USE (a 35 subject study), when the
SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according to 5.1.1
and 5.2.
b) Criterion 2
For the SYSTOLIC and DIASTOLIC BLOOD PRESSURES for each of the m subjects, the standard deviation, s , of
m
the averaged paired DETERMINATIONS per subject of the SPHYGMOMANOMETER-UNDER-TEST and of the
observers' DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER shall meet the criteria listed in Table 1 or
Table 2 when calculated according to Formula (4):
m
s (4)
xx
m n
j
m1
j1
where
is the mean value of the differences as calculated according to Formula (2);
x
n
m is the number of subjects;
x is the index for the individual element;
j
m is calculated according to Formula (5).
d
xp()p (5)
j sut ref
kk
d
k1
where
d is the number of DETERMINATIONS per subject;
k
is the index for the individual element.
8 © ISO 2013 – All rights reserved
EXAMPLE 4 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
study).
EXAMPLE 5 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and 12 years (an 85 subject study).
EXAMPLE 6 m = 35 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
when the SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according
to 5.1.1 and 5.2.
EXAMPLE 7 m = 85 for a SPHYGMOMANOMETER intended only for a special INTENDED USE (an 85 subject study).
Table 1 — Averaged subject data acceptance (criterion 2) in mmHg
Maximum permissible standard deviation, s , as function of,
x
m
n
mmHg
x
n
0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9
0, 6,95 6,95 6,95 6,95 6,93 6,92 6,91 6,90 6,89 6,88
1, 6,87 6,86 6,84 6,82 6,80 6,78 6,76 6,73 6,71 6,68
2, 6,65 6,62 6,58 6,55 6,51 6,47 6,43 6,39 6,34 6,30
3, 6,25 6,20 6,14 6,09 6,03 5,97 5,89 5,83 5,77 5,70
4, 5,64 5,56 5,49 5,41 5,33 5,25 5,16 5,08 5,01 4,90
5, 4,79 — — — — — — — — —
EXAMPLE For mean of 4,2 mmHg, the maximum permissible standard deviation is 5,49 mmHg.
Table 2 — Averaged subject data acceptance (criterion 2) in kPa
Maximum permissible standard deviation, s , as function of,
x
m n
kPa
x
n
0,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070 0,080 0,090
0,0 0,9266 0,9266 0,9266 0,9266 0,9266 0,9246 0,9233 0,9223 0,9213 0,9203
0,1 0,9193 0,9183 0,9173 0,9163 0,9152 0,9138 0,9119 0,9099 0,9079 0,9059
0,2 0,9039 0,9007 0,8989 0,8970 0,8946 0,8906 0,8878 0,8855 0,8826 0,8785
0,3 0,8756 0,8723 0,8679 0,8641 0,8601 0,8562 0,8519 0,8471 0,8414 0,8374
0,4 0,8333 0,8283 0,8226 0,8169 0,8119 0,8059 0,7999 0,7933 0,7853 0,7793
0,5 0,7739 0,7669 0,7599 0,7531 0,7463 0,7388 0,7319 0,7237 0,7157 0,7077
0,6 0,6999 0,6891 0,6802 0,6723 0,6670 0,6595 0,6488 0,6386 — —
EXAMPLE For mean of 0,520 kPa, the maximum permissible standard deviation is 0,7599 kPa.
5.2.4.2 * Same arm sequential method
5.2.4.2.1 * Procedure
Either arm may be utilized.
Perform the following:
a) Using the REFERENCE SPHYGMOMANOMETER, have the observers determine the subject's BLOOD PRESSURE
(see Figure 2).
b) Wait at least 60 s.
c) Using the SPHYGMOMANOMETER-UNDER-TEST, have the observers determine the subject's BLOOD PRESSURE.
d) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and then wait at least
60 s.
EXAMPLES Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
DETERMINATION.
reset command are methods to clear the memory of the previous
Key
1 double stethoscope
2 REFERENCE SPHYGMOMANOMETER display
3 REFERENCE SPHYGMOMANOMETER hand pump
4 SPHYGMOMANOMETER-UNDER-TEST
NOTE Only one CUFF is connected to the SPHYGMOMANOMETER-UNDER-TEST.
Figure 2 — Illustration of the same arm sequential method
e) Do not use the data points obtained in a) and c) in the DETERMINATION of accuracy.
f) Using the REFERENCE SPHYGMOMANOMETER, have the observers determine the subject's BLOOD PRESSURE.
g) Have the observers use the SPHYGMOMANOMETER-UNDER-TEST and the REFERENCE SPHYGMOMANOMETER to
determine the subject's BLOOD PRESSURE sequentially.
h) Wait at least 60 s between each DETERMINATION.
i) Repeat g) and h) until the required number of valid DETERMINATIONS has been performed.
All data from a subject shall be excluded if any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS
differ by more than 12 mmHg (1,60 kPa) or if any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS
10 © ISO 2013 – All rights reserved
differ by more than 8 mmHg (1,07 kPa). Notwithstanding this requirement, if the REFERENCE BLOOD PRESSURE
of an individual subject does not meet these criteria during the period of the test, two compliant consecutive
DETERMINATION pairs that meet these criteria may be used. In this case, additional subjects shall be used to
complete the minimum number of DETERMINATIONS. No more than 10 % of the subjects shall have fewer than
three valid DETERMINATION pairs.
5.2.4.2.2 Data analysis
The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
a) Criterion 1
For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, ,
x
n
of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , no greater than 8,0 mmHg (1,07 kPa) when
n
calculated according to Formula (6) and Formula (7):
n
xp()p (6)
n sut refsq
ii
n
i1
n
s (7)
xx
n n
i
n1
i1
where
is the mean value of the differences;
x
n
th
pp x difference between the i paired BLOOD PRESSURE DETERMINATIONS (DETERMINATION
sut refsq i
ii
of the SPHYGMOMANOMETER-UNDER-TEST REFERENCE BLOOD PRESSURE);
i is the index for the individual element;
n is the number of DETERMINATIONS;
th
p is the REFERENCE BLOOD PRESSURE for the i DETERMINATION as calculated according to
refsq
i
Formula (8);
and s shall be calculated and expressed to at least 0,1 mmHg (0,01 kPa).
x
n
n
The REFERENCE BLOOD PRESSURE, p , (the observers' DETERMINATIONS with the REFERENCE
refsq
i
SPHYGMOMANOMETER) shall be the average of the preceding and following REFERENCE BLOOD PRESSURES.
pp( p p p) (8)
refsq ref ref ref ref
ii,1 i,2 i1,1 i 1,2
where
th
p is the BLOOD PRESSURE determined by observer 1 for the i DETERMINATION;
ref
i ,1
th
p is the BLOOD PRESSURE determined by observer 2 for the i DETERMINATION.
ref
i,2
EXAMPLE 1 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
study).
EXAMPLE 2 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and 12 years (an 85 subject study).
EXAMPLE 3 n = 105 for a SPHYGMOMANOMETER intended for a special INTENDED USE (a 35 subject study), when the
SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according to 5.1.1
and 5.2.
b) Criterion 2
For the SYSTOLIC and DIASTOLIC BLOOD PRESSURES for each of the m subjects, the standard deviation, s , of
m
the averaged paired DETERMINATIONS per subject of the SPHYGMOMANOMETER-UNDER-TEST and of the
observers' DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER shall meet the criteria listed in Table 1 or
Table 2 when calculated according to Formula (9).
m
sx()x (9)
mjn
m1
j1
where
is the mean value of the differences as calculated according to Formula (6);
x
n
m is the number of subjects;
j is the index for the individual element;
x is SPHYGMOMANOMETER-UNDER-TEST error as calculated according to Formula (10);
j
d
xp( p) (10)
j sut refsq
jk,,j k
d
k1
where
d is the number of DETERMINATIONS per subject;
k is the index for the individual element;
p is the REFERENCE BLOOD PRESSURE calculated according to Formula (11).
refsq
j ,k
pp( p p p) (11)
refsq ref ref ref ref
j,,k jk,1 j,k,2 jk,1,1 jk,1,2
EXAMPLE 4 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
study).
EXAMPLE 5 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and 12 years (an 85 subject study).
EXAMPLE 6 m = 35 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
when the SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according
to 5.1.1 and 5.2.
EXAMPLE 7 m = 85 for a SPHYGMOMANOMETER intended only for a special INTENDED USE (an 85 subject study).
12 © ISO 2013 – All rights reserved
5.2.4.3 * Opposite limb simultaneous method
5.2.4.3.1 Procedure
The starting arm side of the SPHYGMOMANOMETER-UNDER-TEST and the REFERENCE SPHYGMOMANOMETER
DETERMINATIONS shall be alternated between subjects.
Perform the following:
a) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
simultaneously determine the subject's BLOOD PRESSURE in opposite arms (see Figure 3).
b) These data points are not used in the calculation of accuracy.
c) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and wait at least 60 s.
EXAMPLES Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
reset command are methods to clear the memory of the previous DETERMINATION.
d) Interchange arm sides of the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST.
e) Wait at least 60 s from the completion of the previous determination.
f) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
simultaneously determine the subject's BLOOD PRESSURE in opposite arms.
g) Repeat d) to f) until six paired determinations have been performed.
h) If the DETERMINATION by the observers with the REFERENCE SPHYGMOMANOMETER or the
SPHYGMOMANOMETER-UNDER-TEST is not successfully completed, repeat the DETERMINATION on the same
arm (i.e. without interchanging sides).
i) Repeat a) to h) until the required number of valid subjects and DETERMINATIONS has undergone CLINICAL
INVESTIGATION.
All data from a subject shall be excluded if
any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS on the same arm differ by more than
12 mmHg (1,60 kPa) or any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS on the same arm
differ by more than 8 mmHg (1,07 kPa).
the lateral difference of the REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS is more than 15 mmHg
(2,00 kPa) or the lateral difference of the REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS is more
than 10 mmHg (1,33 kPa).
Key
1 double stethoscope
2 REFERENCE SPHYGMOMANOMETER display
3 REFERENCE SPHYGMOMANOMETER hand pump
4 SPHYGMOMANOMETER-UNDER-TEST
NOTE Only one CUFF is connected to the SPHYGMOMANOMETER-UNDER-TEST.
Figure 3 — Illustration of opposite arm simultaneous method
The lateral difference, LD, is calculated separately for SYSTOLIC and DIASTOLIC BLOOD PRESSURES, according to
Formula (12).
LD p p (12)
ref_R ref_L
ij
ij11
where
p and p are REFERENCE BLOOD PRESSURES in the right (R) arm and left (L) arm, respectively.
ref_R ref_L
i j
5.2.4.3.2 * Data analysis
The SPHYGMOMANOMETER-UNDER-TEST error, x, is calculated by taking the difference between the
SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURE and the REFERENCE SPHYGMOMANOMETER BLOOD PRESSURE
and by adding the lateral difference, LD, as calculated according to Formula (13) if the SPHYGMOMANOMETER-
UNDER-TEST BLOOD PRESSURE was taken in the left arm or by subtracting the lateral difference, LD, as
calculated according to Formula (14) if the SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURE was taken in
the right arm.
xppLD (13)
sut_L ref_R
14 © ISO 2013 – All rights reserved
xppLD (14)
sut_R ref_L
where p and p are SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURES in right (R) arm and left (L)
sut_R sut_L
arm, respectively.
The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
a) Criterion 1
For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, x ,
n
of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , not greater than 8,0 mmHg (1,07 kPa) when
n
calculated according to Formula (15) and Formula (16).
n
x x (15)
ni
n
i1
n
sx x (16)
nin
n1
i1
where
x is the SPHYGMOMANOMETER-UNDER-TEST error as calculated according to Formula (13) and
i
Formula (14);
n is the number of DETERMINATIONS;
i is the index for the individual element.
EXAMPLE 1 n = 510 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
study).
EXAMPLE 2 n = 510 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and 12 years (an 85 subject study).
EXAMPLE 3 n = 210 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
when the SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according
to 5.1.1 and 5.2.
b) Criterion 2
For the average of the SYSTOLIC BLOOD PRESSURE and the average of the DIASTOLIC BLOOD PRESSURE for each
subject, the standard deviation, s , of the m averaged paired DETERMINATIONS of the SPHYGMOMANOMETER-
m
UNDER-TEST and of the observers' DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER, per subject, shall
meet the criteria listed in Table 1 or Table 2 when calculated according to Formula (17).
m 2
s (17)
x x
m n
j
m1
j1
where
is the mean value as calculated according to Formula (15);
x
n
m is the number of subjects;
j is the index for the individual element;
x is the mean per subject calculated according to Formula (18).
j
x x (18)
j k
k1
where
x is the SPHYGMOMANOMETER-UNDER-TEST error for each subject as calculated according to
k
Formula (13) and Formula (14);
k is the index for the individual element.
EXAMPLE 4 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
study).
EXAMPLE 5 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
between 3 years and 12 years (an 85 subject study).
EXAMPLE 6 m = 35 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
when the SPHYGMOMANOMETER has previously undergone CLINICAL INVESTIGATION in a separate 85 subject study according
to 5.1.1 and 5.2.
EXAMPLE 7 m = 85 for a SPHYGMOMANOMETER intended only for a special INTENDED USE (an 85 subject study).
5.2.5 * Additional requirements for a SPHYGMOMANOMETER intended for use in exercise stress testing
environments
For a SPHYGMOMANOMETER intended for use in exercise stress testing, an additional CLINICAL INVESTIGATION
shall be performed. During this CLINICAL INVESTIGATION, the subjects shall be stressed by dynamic (aerobic)
exercise on a bicycle ergometer or a treadmill so as to increase their heart rate from their resting heart rate to
a target heart rate of 50 % to 70 % of their average predicted maximum heart rate (see Annex B). The
physical load setting of the ergometer and target heart rate shall be recorded. The arm used for a
DETERMINATION shall be supported. The CUFF shall be at the level of the right atrium during the DETERMINATION
of BLOOD PRESSURE.
The same arm sequential method of 5.2.4.2 shall not be used. The CLINICAL INVESTIGATION shall consist of a
minimum of 35 subjects. A stress testing study shall be exempt from the BLOOD PRESSURE distribution
requirements of 5.1.5. At least 10 % of the subjects shall have a resting SYSTOLIC BLOOD PRESSURE
≥ 160 mmHg (21,33 kPa). At least 10 % of the subjects shall have a resting DIASTOLIC BLOOD PRESSURE
≥ 100 mmHg (13,33 kPa).
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
5.2.6 Additional requirements for a SPHYGMOMANOMETER intended for use in ambulatory monitoring
For a SPHYGMOMANOMETER intended for use in ambulatory monitoring, an additional CLINICAL INVESTIGATION
shall be performed. During this CLINICAL INVESTIGATION, the subjects shall be stressed by dynamic (aerobic)
exercise on a bicycle ergometer or treadmill so as to increase their heart rate to 10 % to 20 % above their
resting heart rate. The physical load setting of the ergometer and heart rate shall be recorded. The arm used
for a DETERMINATION shall be supported. The CUFF shall be at the level of the right atrium during the
DETERMINATION of BLOOD PRESSURE.
The same arm sequential method of 5.2.4.2 shall not be used. The CLINICAL INVESTIGATION shall consist of a
minimum of 35 subjects. An ambulatory monitoring study shall be exempt from the BLOOD PRESSURE
distribution requirements of 5.1.5. At least 10 % of the subjects shall have a resting SYSTOLIC BLOOD PRESSURE
16 © ISO 2013 – All rights reserved
160 mmHg (21,33 kPa). At least 10 % of the subjects shall have a resting DIASTOLIC BLOOD PRESSURE
100 mmHg (13,33 kPa).
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
6 CLINICAL INVESTIGATION with REFERENCE INVASIVE BLOOD PRESSURE MONITORING
EQUIPMENT
6.1 PATIENT requirements
6.1.1 Number
A REFERENCE INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT CLINICAL INVESTIGATION shall consist of a
minimum of 15 PATIENTS. For each PATIENT, no more than 10 valid BLOOD PRESSURE measurements shall be
taken. There shall be a minimum of 150 valid BLOOD PRESSURE measurements in the CLINICAL INVESTIGATION.
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
6.1.2 * Gender distribution
At least 30 % of the measurements shall be on male PATIENTS and at least 30 % of the measurements shall be
on female PATIENTS.
Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
6.1.3 * Age distribution
6.1.3.1 SPHYGMOMANOMETERS intended for use in adults, adolescents or children
For a SPHYGMOMANOMETER intended for use in adult and/or adolescent PATIENTS, the age of every PATIENT
included in the CLINICAL INVESTIGATION shall be greater than 12 years.
NOTE 1 Minimum total of 15 subjects.
For a SPHYGMOMANOMETER additionally intended for use in children, at least an additional 5 children aged
between 3 years and 12 years shall be included in the CLINICAL INVESTIGATION.
NOTE 2 Minimum total of 20 subjects.
For a SPHYGMOMANOMETER additionally intended for use in children, the data analysis (see 6.1.1) of adults,
adolescents and children shall be pooled. Children are exempt from the BLOOD PRESSURE dist
...
DRAFT INTERNATIONAL STANDARD ISO/DIS 81060-2
ISO/TC 121/SC 3 Secretariat: ANSI
Voting begins on Voting terminates on
2011-12-02 2012-05-02
INTERNATIONAL ORGANIZATION FOR STANDARDIZATION • МЕЖДУНАРОДНАЯ ОРГАНИЗАЦИЯ ПО СТАНДАРТИЗАЦИИ • ORGANISATION INTERNATIONALE DE NORMALISATION
INTERNATIONAL ELECTROTECHNICAL COMMISSION • МЕЖДУНАРОДНАЯ ЭЛЕКТРОТЕХНИЧЕСКАЯ КОММИСИЯ • COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
Non-invasive sphygmomanometers —
Part 2:
Clinical investigation of automated measurement type
Sphygmomanomètres non invasifs —
Partie 2: Validation clinique pour type à mesurage automatique
[Revision of first edition (ISO 81060-2:2009)]
ICS 11.040.10
This draft is submitted to a parallel enquiry in ISO and a CDV vote in the IEC.
In accordance with the provisions of Council Resolution 15/1993 this document is circulated
in the English language only.
Conformément aux dispositions de la Résolution du Conseil 15/1993, ce document est
distribué en version anglaise seulement.
To expedite distribution, this document is circulated as received from the committee
secretariat. ISO Central Secretariat work of editing and text composition will be undertaken at
publication stage.
Pour accélérer la distribution, le présent document est distribué tel qu'il est parvenu du
secrétariat du comité. Le travail de rédaction et de composition de texte sera effectué au
Secrétariat central de l'ISO au stade de publication.
THIS DOCUMENT IS A DRAFT CIRCULATED FOR COMMENT AND APPROVAL. IT IS THEREFORE SUBJECT TO CHANGE AND MAY NOT BE REFERRED TO
AS AN INTERNATIONAL STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS BEING ACCEPTABLE FOR INDUSTRIAL, TECHNOLOGICAL, COMMERCIAL AND USER PURPOSES, DRAFT
INTERNATIONAL STANDARDS MAY ON OCCASION HAVE TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL TO BECOME STANDARDS TO WHICH
REFERENCE MAY BE MADE IN NATIONAL REGULATIONS.
RECIPIENTS OF THIS DRAFT ARE INVITED TO SUBMIT, WITH THEIR COMMENTS, NOTIFICATION OF ANY RELEVANT PATENT RIGHTS OF WHICH THEY
ARE AWARE AND TO PROVIDE SUPPORTING DOCUMENTATION.
© International Organization for Standardization, 2011
ISO/DIS 81060-2
Copyright notice
This ISO document is a Draft International Standard and is copyright-protected by ISO. Except as permitted
under the applicable laws of the user’s country, neither this ISO draft nor any extract from it may be reproduced,
stored in a retrieval system or transmitted in any form or by any means, electronic, photocopying, recording or
otherwise, without prior written permission being secured.
Requests for permission to reproduce should be addressed to either ISO at the address below or ISO’s member
body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Reproduction may be subject to royalty payments or a licensing agreement.
Violators may be prosecuted.
ii © ISO 2011 – All rights reserved
ISO/DIS 81060-2
1 Contents Page
2 Foreword .iv
3 Introduction . v
4 1 Scope .2
5 2 Normative references .2
6 3 Terms and definitions .3
7 4 General requirements for clinical investigations .3
8 4.1 Clinical investigation methods .3
9 4.2 Good clinical practice .3
10 5 Clinical investigation with auscultatory REFERENCE SPHYGMOMANOMETER .4
11 5.1 Subject requirements .4
12 5.2 Clinical investigation method with REFERENCE SPHYGMOMANOMETER .5
13 6 Clinical investigation with REFERENCE INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT . 18
14 6.1 PATIENT requirements . 18
15 6.2 Clinical investigation methods with REFERENCE INVASIVE BLOOD PRESSURE MONITORING
16 EQUIPMENT . 20
17 7 Pregnant (including pre-eclamptic) PATIENT populations . 23
18 Annex A (informative) Rationale and guidance . 25
19 Annex B (normative) Target heart rates for exercise stress testing . 37
20 Annex C (informative) Reference to the essential principles . 38
21 Annex ZA (informative) Relationship between this International Standard and the Essential
22 Requirements of EU Directive 93/42/EEC . 39
23 Bibliography . 40
ISO/DIS 81060-2
25 Foreword
26 ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
27 (ISO member bodies). The work of preparing International Standards is normally carried out through ISO
28 technical committees. Each member body interested in a subject for which a technical committee has been
29 established has the right to be represented on that committee. International organizations, governmental and
30 non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
31 International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
32 International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
33 The main task of technical committees is to prepare International Standards. Draft International Standards
34 adopted by the technical committees are circulated to the member bodies for voting. Publication as an
35 International Standard requires approval by at least 75 % of the member bodies casting a vote.
36 Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
37 rights. ISO shall not be held responsible for identifying any or all such patent rights.
38 This second edition cancels and replaces the first edition (ISO 81060-2:2009), subclauses 5.2.4.3.1 and 6.2.4
39 of which have been technically revised. Numerous clarifications have been added and kPa equivalent values
40 for the mmHg values have been included in the standard including the Criterion 2 requirements of 5.2.4.1.2.
41 ISO 81060-2 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment,
42 Subcommittee SC 3, Lung ventilators and related equipment and Technical Committee IEC/TC 62, Electrical
43 Equipment in Medical Practice, Subcommittee 62D, Electromedical Equipment.
44 ISO 81060 consists of the following parts, under the general title Non-invasive sphygmomanometers:
45 ⎯ Part 1: Requirements and test methods for non-automated measurement type
46 ⎯ Part 2: Clinical investigation of automated measurement type
47 In this document, the following print types are used:
48 ⎯ requirements, compliance with which can be verified, and definitions: roman type;
49 ⎯ notes and examples: smaller roman type;
50 ⎯ test methods: italic type;
51 ⎯ terms defined in this document: SMALL CAPITALS TYPE.
52 Throughout this document, text for which a rationale is provided in Annex A is indicated by an asterisk (*).
iv © ISO 2011 – All rights reserved
ISO/DIS 81060-2
54 Introduction
55 Determination of BLOOD PRESSURE is an important procedure that is clinically used to assess the health of the
56 PATIENT.
57 Frequent determination of BLOOD PRESSURE is routine during anaesthesia. BLOOD PRESSURE serves to aid in
58 drug titration and fluid management and to provide warning of conditions that could affect PATIENT morbidity
59 and mortality.
DRAFT INTERNATIONAL STANDARD ISO/DIS 60-2
Non-invasive sphygmomanometers —
Part 2:
Clinical investigation of automated measurement type
ISO/DIS 81060-2
63 1 Scope
64 This part of ISO 81060 specifies the requirements and methods for the clinical investigation of ME EQUIPMENT
65 used for the intermittent non-invasive automated estimation of the arterial BLOOD PRESSURE by utilizing a CUFF.
66 This part of ISO 81060 is applicable to all SPHYGMOMANOMETERS that sense or display pulsations, flow or
67 sounds for the estimation, display or recording of BLOOD PRESSURE. These SPHYGMOMANOMETERS need not
68 have automatic CUFF inflation. This part of ISO 81060 covers SPHYGMOMANOMETERS intended for use in all
69 PATIENT populations (e.g. all age and weight ranges), and all conditions of use (e.g. ambulatory BLOOD
70 PRESSURE monitoring, stress testing BLOOD PRESSURE monitoring and BLOOD PRESSURE monitors for the HOME
71 HEALTHCARE ENVIRONMENT or self-measurement).
72 EXAMPLE AUTOMATED SPHYGMOMANOMETER as given in IEC 80601-2-30 clinically investigated by this part of ISO 81060.
73 This part of ISO 81060 specifies additional disclosure requirements for the ACCOMPANYING DOCUMENTS of
74 SPHYGMOMANOMETERS clinically investigated according to this part of ISO 81060.
75 This part of ISO 81060 is not applicable to clinical investigations of NON-AUTOMATED SPHYGMOMANOMETERS as
76 given in ISO 81060-1 or INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT as given in IEC 60601-2-34.
77 2 Normative references
78 The following referenced documents are indispensable for the application of this document. For dated
79 references, only the edition cited applies. For undated references, the latest edition of the referenced
80 document (including any amendments) applies.
81 ISO 14155:2011, Clinical investigation of medical devices for human subjects — Good clinical practice
82 ISO 81060-1:2007, Non-invasive sphygmomanometers — Part 1: Requirements and test methods for non-
83 automated measurement type
84 IEC 80601-2-30:2009, Medical electrical equipment — Part 2-30: Particular requirements for basic safety and
85 essential performance of automated non-invasive sphygmomanometers
86 IEC 60601-1:2005, Medical electrical equipment — Part 1: General requirements for basic safety and
87 essential performance
88 IEC 60601-1-11:2010, Medical electrical equipment — Part 1-11: General requirements for basic safety and
89 essential performance — Collateral standard: Requirements for medical electrical equipment and medical
90 electrical systems used in home care applications
91 IEC 60601-2-34:2000, Medical electrical equipment — Part 2-34: Particular requirements for the safety,
92 including essential performance, of invasive blood pressure monitoring equipment
ISO/DIS 81060-2
93 3 Terms and definitions
94 For the purposes of this document, the terms and definitions given in ISO 14155:2011, IEC 80601-2-30:2009
95 IEC 60601-1:2005, IEC 60601-1-11:2010, IEC 60601-2-34:2000 and the following apply. For convenience, an
96 alphabetized index of defined terms is found beginning on page 43.
97 3.1
98 REFERENCE, adj
99 established accuracy used for clinical investigation of other instruments
100 3.2
101 SPHYGMOMANOMETER
102 ME EQUIPMENT for non-invasive estimation of systemic arterial BLOOD PRESSURE
103 3.3
104 SPHYGMOMANOMETER-UNDER-TEST
105 SPHYGMOMANOMETER being clinically investigated
106 4 General requirements for clinical investigations
107 4.1 Clinical investigation methods
108 SPHYGMOMANOMETERS other than NON-AUTOMATED SPHYGMOMANOMETERS shall be clinically investigated either
109 by using a non-invasive (auscultatory) REFERENCE SPHYGMOMANOMETER or by using REFERENCE INVASIVE
110 BLOOD PRESSURE MONITORING EQUIPMENT according to this part of ISO 81060 in each mode of operation.
111 EXAMPLE 1 Adult and neonatal mode.
112 EXAMPLE 2 Slow and fast CUFF deflation rate mode.
113 A clinical investigation shall be considered a TYPE TEST.
114 Consider compliance with the requirements of this subclause to exist when the criteria of the relevant
115 inspections and tests in this part of ISO 81060 are met.
116 4.2 Good clinical practice
117 All clinical investigations shall comply with the requirements of ISO 14155:2011. Clinical investigation with
118 REFERENCE INVASIVE BLOOD PRESSURE MONITORING EQUIPMENT should not be used for PATIENTS or subjects
119 solely for the purpose of investigating SPHYGMOMANOMETER performance.
120 NOTE Some authorities with jurisdiction have additional requirements.
121 The following requirements that are more specific to a SPHYGMOMANOMETER prevail on the corresponding
122 requirements of ISO 14155:2011.
123 Check compliance by application of the requirements of ISO 14155:2011.
ISO/DIS 81060-2
124 5 Clinical investigation with auscultatory REFERENCE SPHYGMOMANOMETER
125 5.1 Subject requirements
126 5.1.1 * Number
127 An auscultatory REFERENCE SPHYGMOMANOMETER clinical investigation shall consist of a minimum of 85
128 subjects. If not otherwise specified, at least three valid BLOOD PRESSURE DETERMINATIONS shall be taken for
129 each subject. There shall be a minimum of 255 valid paired BLOOD PRESSURE DETERMINATIONS.
130 Check compliance by inspection of the CLINICAL INVESTIGATION report.
131 5.1.2 * Gender distribution
132 At least 30 % of the subjects shall be male and at least 30 % of the subjects shall be female.
133 Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
134 5.1.3 * Age distribution
135 For a SPHYGMOMANOMETER intended for use on adults and/or adolescent PATIENTS, the age of every subject
136 included in the clinical investigation shall be greater than 12 years.
137 NOTE 1 Minimum total of 85 subjects.
138 For a SPHYGMOMANOMETER additionally intended for use in children, 35 child subjects aged between 3 years
139 and 12 years shall be included in the clinical investigation.
140 NOTE 2 Minimum total of 85 subjects.
141 If the SPHYGMOMANOMETER has a special mode for children, in that mode, children shall be considered a
142 special PATIENT population (see 5.1.6). In such a study, children are exempt from the BLOOD PRESSURE
143 distribution requirements of 5.1.5.
144 Children aged less than 3 years shall not be included in a clinical investigation utilizing auscultatory
145 DETERMINATIONS by observers with a REFERENCE SPHYGMOMANOMETER.
146 Check compliance by inspection of the ACCOMPANYING DOCUMENT and the CLINICAL INVESTIGATION REPORT.
147 5.1.4 * Limb size distribution
148 For a SPHYGMOMANOMETER intended for use with a single CUFF size:
149 ⎯ at least 40 % of the subjects shall have a limb circumference which lies within the upper half of the
150 specified range of use of the CUFF and at least 40 % shall have a limb circumference within the lower half;
151 and
152 ⎯ at least 20 % of the subjects shall have a limb circumference which lies within the upper quarter of the
153 specified range of use of the CUFF and at least 20 % shall have a limb circumference within the lower
154 quarter.
155 For a SPHYGMOMANOMETER intended for use with multiple CUFF sizes, each CUFF size shall be tested on at
156 least 1/(2 × n) of the subjects, where n is the number of CUFF sizes.
ISO/DIS 81060-2
157 Check compliance by inspection of the ACCOMPANYING DOCUMENT and the CLINICAL INVESTIGATION REPORT.
158 5.1.5 * BLOOD PRESSURE distribution
159 At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE ≤ 100 mmHg
160 (13,33 kPa).
161 At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE ≥ 160 mmHg
162 (21,33 kPa).
163 At least 20 % of the REFERENCE BLOOD PRESSURE readings shall have a SYSTOLIC BLOOD PRESSURE ≥
164 140 mmHg (18,66 kPa).
165 At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE ≤ 60 mmHg
166 (8,0 kPa).
167 At least 5 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE ≥
168 100 mmHg (13,33 kPa).
169 At least 20 % of the REFERENCE BLOOD PRESSURE readings shall have a DIASTOLIC BLOOD PRESSURE ≥
170 85 mmHg (11,33 kPa).
171 Check compliance by inspection of the CLINICAL INVESTIGATION REPORT.
172 5.1.6 * Special PATIENT populations
173 A SPHYGMOMANOMETER that is intended for use in special PATIENT populations where there is OBJECTIVE
174 EVIDENCE that the accuracy of the SPHYGMOMANOMETER might be problematic in those PATIENT populations,
175 shall be clinically investigated in those PATIENT populations.
176 NOTE Clause 7 has a specific example of a special PATIENT population with specific requirements.
177 If the SPHYGMOMANOMETER has been clinically investigated according to the requirements of 5.1.1 and 5.2, it
178 shall then be clinically investigated in at least an additional 35 special population subjects. If the
179 SPHYGMOMANOMETER has not been previously clinically investigated according to the requirements of 5.1.1
180 and 5.2, the clinically investigation in accordance with the requirements of 5.1.1 and 5.2 shall only consist of
181 subjects from the special PATIENT population.
182 The special PATIENT population shall be defined in clear terms and address the following attributes: gender
183 (see 5.1.2), age (see 5.1.3), limb size (see 5.1.4) and BLOOD PRESSURE (see 5.1.5). A summary of this
184 information shall be disclosed in the instructions for use.
185 Check compliance by inspection of the instructions for use and the CLINICAL INVESTIGATION REPORT.
186 5.2 Clinical investigation method with REFERENCE SPHYGMOMANOMETER
187 5.2.1 * Subject preparation
188 Unless otherwise indicated by the instructions for use of the SPHYGMOMANOMETER-UNDER-TEST, position the
189 subject such that the subject:
190 ⎯ is comfortable;
191 EXAMPLE Comfortably seated with legs uncrossed and feet flat on the floor.
192 ⎯ has the back, elbow and forearm supported;
193 ⎯ has the middle of CUFF at the level of the right atrium of the heart.
ISO/DIS 81060-2
194 Recommend that the subject be as relaxed as possible and that the subject avoid talking during the entire
195 procedure. Before the first reading is taken, 5 min should elapse.
196 NOTE Additional details can be found in Reference [32].
197 5.2.2 * Observer preparation
198 Observers should be trained in using a proper methodology for performing a resting BLOOD PRESSURE
199 DETERMINATION by utilizing an accepted clinical protocol for BLOOD PRESSURE measurement. References [8],
200 [28], [29], [32] and [45] contain additional information. Observers should have sufficient practice in performing
201 BLOOD PRESSURE DETERMINATIONS.
202 Each observer's recording of observations of the REFERENCE SPHYGMOMANOMETER shall not be visible to the
203 other observer. The readings of the SPHYGMOMANOMETER-UNDER-TEST shall not be visible to either of these
204 observers.
205 EXAMPLE 1 Utilizing a third observer for recording the readings of the SPHYGMOMANOMETER-UNDER-TEST.
206 EXAMPLE 2 Utilizing an electronic means for recording the readings of the SPHYGMOMANOMETER-UNDER-TEST.
207 Instruct the observers to determine DIASTOLIC BLOOD PRESSURE as the last audible Korotkoff sound (fifth phase
208 or K5), except when Korotkoff sounds are still audible with the CUFF deflated or in children between 3 years
209 and 12 years of age, where the fourth phase (K4) is used. If K4 is not audible in a child, either K5 is used or
210 the subject is excluded.
211 NOTE Other than for children, K4 should be reserved for subjects in whom there is a large discrepancy between muffling
212 and disappearance (with the latter at times approaching zero mmHg).
213 Instruct the observers to record which Korotkoff sound has been used for the DETERMINATION of DIASTOLIC
214 BLOOD PRESSURE.
215 The Korotkoff sound used for DETERMINATION of DIASTOLIC BLOOD PRESSURE in the clinical investigation shall be
216 disclosed in the instructions for use of a SPHYGMOMANOMETER.
217 EXAMPLE 3 K5 was used on 65 subjects and K4 was used on 20 subjects.
218 5.2.3 * REFERENCE DETERMINATION
219 Two observers shall make simultaneous BLOOD PRESSURE DETERMINATIONS on each subject using a double
220 stethoscope.
221 Unless the SPHYGMOMANOMETER-UNDER-TEST is intended for use during significantly irregular heart rhythm and
222 if either observer detects significantly irregular heart rhythm, that DETERMINATION shall be excluded.
223 EXAMPLES Bigeminy, trigeminy, isolated VPB, atrial fibrillation.
224 NOTE 1 Although clinical investigation of BLOOD PRESSURE in PATIENTS with atrial fibrillation is clinically important, there
225 are currently no generally accepted guidelines for determining the BLOOD PRESSURE in such individuals.
226 Any pair of observers' DETERMINATIONS with a difference greater than 4 mmHg (0,53 kPa) shall be excluded.
227 The observers' individual values of each DETERMINATION shall be averaged according to Equation (1) to create
228 the REFERENCE BLOOD PRESSURE DETERMINATION.
p + p
ref ref
i,1 i,2
229 p = (1)
ref
i
ISO/DIS 81060-2
230 where
th
231 p is the BLOOD PRESSURE determined by observer 1 for the i DETERMINATION
ref
i,1
th
232 p is the BLOOD PRESSURE determined by observer 2 for the i DETERMINATION
ref
i,2
th
233 p is the REFERENCE BLOOD PRESSURE for the i DETERMINATION
ref
i
234 The observer-to-observer differences shall be reviewed after completing a set of pairs of test-REFERENCE
235 DETERMINATIONS. If any DETERMINATIONS are excluded, additional pair(s) of DETERMINATIONS shall be taken to
236 ensure that the required number of valid test-REFERENCE pairs are available. A maximum of eight pairs of
237 DETERMINATIONS should be taken.
238 Use a REFERENCE SPHYGMOMANOMETER that complies with the requirements of ISO 81060-1, except that the
239 maximum permissible error shall be ± 1 mmHg (0,13 kPa). Reading of the values on the REFERENCE
240 SPHYGMOMANOMETER should be as accurate as possible. When reading the value on the REFERENCE
241 SPHYGMOMANOMETER, the observers should avoid parallax errors. Rounding has a negative effect on the
242 results of the clinical investigation.
243 NOTE 2 For the purposes of this standard, the CUFF is considered part of the REFERENCE SPHYGMOMANOMETER. A CUFF
244 that does not comply with ISO 81060-1 cannot be used.
245 5.2.4 Clinical investigation methods
246 5.2.4.1 Same arm simultaneous method
247 5.2.4.1.1 * Procedure
248 This method shall only be used with a SPHYGMOMANOMETER-UNDER-TEST:
249 ⎯ that has a CUFF compliant with ISO 81060-1;
250 ⎯ that is designed for use on the upper arm; and
251 ⎯ where
252 ⎯ the continuous linear deflation rate is between 2 mmHg/s (0,27 kPa/s) and 3 mmHg/s (0,40 kPa/s) or
253 ⎯ for a SPHYGMOMANOMETER-UNDER-TEST that controls the deflation as a function of the pulse rate, the
254 deflation rate is between 2 mmHg/pulse (0,27 kPa/pulse) and 3 mmHg/pulse (0,40 kPa/pulse).
255 Either arm may be utilized.
256 The SPHYGMOMANOMETER-UNDER-TEST shall not deflate prior to the detection of the REFERENCE DIASTOLIC
257 BLOOD PRESSURE. The SPHYGMOMANOMETER-UNDER-TEST may be modified to meet this criterion.
258 NOTE Valid same arm simultaneous DETERMINATIONS require the SPHYGMOMANOMETER-UNDER-TEST to inflate the CUFF
259 to at least 20 mmHg (2,67 kPa) higher than the actual SYSTOLIC BLOOD PRESSURE, as determined by the REFERENCE
260 SPHYGMOMANOMETER, and to at least 20 mmHg (2,67 kPa) below the actual DIASTOLIC BLOOD PRESSURE, as determined by
261 the REFERENCE SPHYGMOMANOMETER.
262 Perform the following:
263 a) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
264 simultaneously determine the subject's BLOOD PRESSURE utilizing the same CUFF and inflation/deflation
ISO/DIS 81060-2
265 cycle (see Figure 1). These data points are not used in the calculation of accuracy of the
266 SPHYGMOMANOMETER-UNDER-TEST.
267 b) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and then wait at least
268 60 s.
269 EXAMPLE Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
270 reset command are methods to clear the memory of the previous DETERMINATION.
271 c) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
272 simultaneously determine the subject's BLOOD PRESSURE utilizing the same CUFF and inflation/deflation
273 cycle.
274 d) Wait at least 60 s between DETERMINATIONS.
275 e) Repeat c) and d) until the required number of valid DETERMINATIONS have been performed.
276 All data from a subject shall be excluded if any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS
277 differ by more than 12 mmHg (1,60 kPa) or if any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS
278 differ by more than 8 mmHg (1,07 kPa).
280 Key
281 1 double stethoscope
282 2 REFERENCE SPHYGMOMANOMETER display
283 3 SPHYGMOMANOMETER-UNDER-TEST
284 Figure 1 — Illustration of same arm simultaneous method
285 5.2.4.1.2 * Data analysis
286 The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
287 a) Criterion 1
288 For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, x ,
n
289 of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
290 DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
ISO/DIS 81060-2
291 ± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , no greater than 8,0 mmHg (1,07 kPa) when calculated
n
292 according to Equation (2) and Equation (3):
n
293 xp=× −p (2)
n ()sut ref
∑
ii
n
i=1
n
294 s=× (3)
xx− n
n ∑()
i
n−1
i=1
295 where
296 is the mean value of the differences;
x
n
297 p − p = x difference between a paired BLOOD PRESSURE DETERMINATION (DETERMINATION of the
sut ref i
ii
298 SPHYGMOMANOMETER-UNDER-TEST − REFERENCE BLOOD PRESSURE as calculated according to
299 Equation (1) );
300 i is the index for the individual element;
301 n is the number of DETERMINATIONS;
302 and s shall be calculated and expressed to 0,1 mmHg (0,01 kPa).
x
n
n
303 EXAMPLE 1 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85
304 subject study).
305 EXAMPLE 2 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
306 between 3 years and12 years (an 85 subject study).
307 EXAMPLE 3 n = 105 for a SPHYGMOMANOMETER intended for a special INTENDED USE (a 35 subject study), when the
308 SPHYGMOMANOMETER has previously been clinically investigated in a separate 85 subject study according to 5.1.1 and 5.2.
309 b) Criterion 2
310 For the SYSTOLIC and DIASTOLIC BLOOD PRESSURES for each of the m subjects, the standard deviation, s , of the
m
311 averaged paired DETERMINATIONS per subject of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
312 DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER shall meet the criteria listed in Table 1 or Table 2
313 when calculated according to Equation (4):
m 2
314 s=× (4)
xx−
m ∑ n
()
j
m−1
j=1
315 where
316 is the mean value of the differences as calculated according to Equation (2);
x
n
317 m is the number of subjects;
318 j is the index for the individual element;
319 x is calculated according to Equation (5).
j
ISO/DIS 81060-2
320 Table 1 — Averaged subject data acceptance (criterion 2) in mmHg
Maximum permissible standard deviation, s , as function of,
x
m
n
mmHg
x
n
0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9
± 0, 6,95 6,95 6,95 6,95 6,93 6,92 6,91 6,90 6,89 6,88
± 1, 6,87 6,86 6,84 6,82 6,80 6,78 6,76 6,73 6,71 6,68
± 2, 6,65 6,62 6,58 6,55 6,51 6,47 6,43 6,39 6,34 6,30
± 3, 6,25 6,20 6,14 6,09 6,03 5,97 5,89 5,83 5,77 5,70
± 4, 5,64 5,56 5,49 5,41 5,33 5,25 5,16 5,08 5,01 4,90
± 5, 4,79 — — — — — — — — —
EXAMPLE For mean of ± 4,2 mmHg, the maximum permissible standard deviation is 5,49 mmHg.
322 Table 2 — Averaged subject data acceptance (criterion 2) in kPa
Maximum permissible standard deviation, s , as function of,
x
m
n
kPa
x
n
0,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070 0,080 0,090
± 0,0 0,9266 0,9266 0,9266 0,9266 0,9266 0,9246 0,9233 0,9223 0,9213 0,9203
± 0,1 0,9193 0,9183 0,9173 0,9163 0,9152 0,9138 0,9119 0,9099 0,9079 0,9059
± 0,2 0,9039 0,9007 0,8989 0,8970 0,8946 0,8906 0,8878 0,8855 0,8826 0,8785
± 0,3 0,8756 0,8723 0,8679 0,8641 0,8601 0,8562 0,8519 0,8471 0,8414 0,8374
± 0,4 0,8333 0,8283 0,8226 0,8169 0,8119 0,8059 0,7999 0,7933 0,7853 0,7793
± 0,5 0,7739 0,7669 0,7599 0,7531 0,7463 0,7388 0,7319 0,7237 0,7157 0,7077
± 0,6 0,6999 0,6891 0,6802 0,6723 0,6670 0,6595 0,6488 0,6386 — —
EXAMPLE For mean of ± 0,520 kPa, the maximum permissible standard deviation is 0,7599 kPa.
d
324 xp=×()−p (5)
j sut ref
∑
kk
d
k=1
325 where
326 d is the number of DETERMINATIONS per subject;
327 k is the index for the individual element.
328 EXAMPLE 4 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
329 study).
330 EXAMPLE 5 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
331 between 3 years and 12 years (an 85 subject study).
ISO/DIS 81060-2
332 EXAMPLE 6 m = 35 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
333 when the SPHYGMOMANOMETER has previously been clinically investigated in a separate 85 subject study according to 5.1.1
334 and 5.2.
335 EXAMPLE 7 m = 85 for a SPHYGMOMANOMETER intended only for a special INTENDED USE (an 85 subject study).
336 5.2.4.2 * Same arm sequential method
337 5.2.4.2.1 * Procedure
338 Either arm may be utilized.
339 Perform the following:
340 a) Have the observers using the REFERENCE SPHYGMOMANOMETER determine the subject's BLOOD PRESSURE
341 (see Figure 2).
342 b) Wait at least 60 s.
343 c) Have the observers using the SPHYGMOMANOMETER-UNDER-TEST determine the subject's BLOOD PRESSURE.
344 d) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and then wait at least
345 60 s.
346 EXAMPLES Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
347 reset command are methods to clear the memory of the previous DETERMINATION.
349 Key
350 1 double stethoscope
351 2 REFERENCE SPHYGMOMANOMETER display
352 3 REFERENCE SPHYGMOMANOMETER hand pump
353 4 SPHYGMOMANOMETER-UNDER-TEST
354 NOTE Only one CUFF is connected to the SPHYGMOMANOMETER-UNDER-TEST.
355 Figure 2 — Illustration of same arm sequential method
356 e) Do not use the data points obtained in a) and c) in the DETERMINATION of accuracy.
ISO/DIS 81060-2
357 f) Have the observers use the REFERENCE SPHYGMOMANOMETER to determine the subject's BLOOD PRESSURE.
358 g) Have the observers use the SPHYGMOMANOMETER-UNDER-TEST and the REFERENCE SPHYGMOMANOMETER to
359 determine the subject's BLOOD PRESSURE sequentially.
360 h) Wait at least 60 s between each DETERMINATION.
361 i) Repeat g) and h) until the required number of valid DETERMINATIONS has been performed.
362 j) Have the observers use the REFERENCE SPHYGMOMANOMETER to determine the subject's BLOOD PRESSURE
363 after the last DETERMINATION with the SPHYGMOMANOMETER-UNDER-TEST.
364 All data from a subject shall be excluded if any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS
365 differ by more than 12 mmHg (1,60 kPa) or if any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS
366 differ by more than 8 mmHg (1,07 kPa). Notwithstanding this requirement, if the REFERENCE BLOOD PRESSURE
367 of an individual subject does not meet these criteria during the period of the test, two compliant consecutive
368 DETERMINATION pairs that meet these criteria may be used. In this case, additional subjects shall be used to
369 complete the minimum number of DETERMINATIONS. No more than 10 % of the subjects shall have fewer than
370 three valid DETERMINATION pairs.
371 5.2.4.2.2 Data analysis
372 The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
373 a) Criterion 1
374 For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, x ,
n
375 of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
376 DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
377 ± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , no greater than 8,0 mmHg (1,07 kPa) when calculated
n
378 according to Equation (6) and Equation (7):
n
379 x = × ( p − p ) (6)
n ∑ sut ref−sq
n
i i
i=1
n
380 s=× (7)
xx−
n n
∑()
i
n−1
i=1
381 where
382 is the mean value of the differences;
x
n
th
383 p − p = x difference between the i paired BLOOD PRESSURE DETERMINATIONS (DETERMINATION
sut ref−sq i
i i
384 of the SPHYGMOMANOMETER-UNDER-TEST − REFERENCE BLOOD PRESSURE);
385 i is the index for the individual element;
386 n is the number of DETERMINATIONS.
th
387 p is the REFERENCE BLOOD PRESSURE for the i DETERMINATION as calculated according to
ref−sq
i
388 Equation (8)
ISO/DIS 81060-2
389 x and s shall be calculated and expressed to at least 0,1 mmHg (0,01 kPa).
n n
390 The REFERENCE BLOOD PRESSURE, p , (the observers' DETERMINATIONS with the REFERENCE
ref-sq
i
391 SPHYGMOMANOMETER) shall be the average of the preceding and following REFERENCE BLOOD PRESSURES.
392 p = ×( p + p + p + p ) (8)
ref−sq ref ref ref ref
i i,1 i,2 i+1,1 i+1,2
393 where
th
394 p is the BLOOD PRESSURE determined by observer 1 for the i DETERMINATION
ref
i,1
th
395 p is the BLOOD PRESSURE determined by observer 2 for the i DETERMINATION
ref
i,2
396 EXAMPLE 1 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
397 study).
398 EXAMPLE 2 n = 255 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
399 between 3 years and 12 years (an 85 subject study).
400 EXAMPLE 3 n = 105 for a SPHYGMOMANOMETER intended for a special INTENDED USE (a 35 subject study), when the
401 SPHYGMOMANOMETER has previously been clinically investigated in a separate 85 subject study according to 5.1.1 and 5.2.
402 b) Criterion 2
403 For the SYSTOLIC and DIASTOLIC BLOOD PRESSURES for each of the m subjects, the standard deviation, s , of the
m
404 averaged paired DETERMINATIONS per subject of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
405 DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER shall meet the criteria listed in Table 1 or Table 2
406 when calculated according to Equation (9).
m 2
407 s=× (9)
xx−
m ∑ n
()
j
m−1
j=1
408 where
409 is the mean value of the differences as calculated according to Equation (6);
x
n
410 m is the number of subjects;
411 j is the index for the individual element;
412 x is SPHYGMOMANOMETER-UNDER-TEST error as calculated according to Equation (10);
j
d
413 xj= × ( p − p ) (10)
∑ sut ref−sq
j,k j,k
d
k=1
414 where
415 d is the number of DETERMINATIONS per subject;
416 k is the index for the individual element;
417 p the REFERENCE BLOOD PRESSURE is calculated according to Equation (11).
ref−sq
j,k
ISO/DIS 81060-2
418 p = ×( p + p + p + p ) (11)
ref−sq ref ref ref ref
j,k j,k,1 j,k,2 j,k+1,1 j,k+1,2
419 EXAMPLE 4 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
420 study).
421 EXAMPLE 5 m = 85 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
422 between 3 years and 12 years (an 85 subject study).
423 EXAMPLE 6 m = 35 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
424 when the SPHYGMOMANOMETER has previously been clinically investigated in a separate 85 subject study according to 5.1.1
425 and 5.2.
426 EXAMPLE 7 m = 85 for a SPHYGMOMANOMETER intended only for a special INTENDED USE (an 85 subject study).
428 5.2.4.3 * Opposite limb simultaneous method
429 5.2.4.3.1 Procedure
430 The starting limb side of the SPHYGMOMANOMETER-UNDER-TEST and the REFERENCE SPHYGMOMANOMETER
431 DETERMINATIONS shall be alternated between subjects.
432 Perform the following:
433 a) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
434 simultaneously determine the subject's BLOOD PRESSURE in opposite arms (see Figure 3).
435 b) These data points are not used in the calculation of accuracy.
436 c) Clear the SPHYGMOMANOMETER-UNDER-TEST memory of the previous DETERMINATION and wait at least 60 s.
437 EXAMPLES Switching the power off and on, removing the BLOOD PRESSURE module and reinstalling it or issuing a
438 reset command are methods to clear the memory of the previous DETERMINATION.
439 d) Interchange arm sides of the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST.
440 e) Wait at least 60 s from the completion of the previous determination.
441 f) Have the observers using the REFERENCE SPHYGMOMANOMETER and the SPHYGMOMANOMETER-UNDER-TEST
442 simultaneously determine the subject's BLOOD PRESSURE in opposite arms.
443 g) Repeat d) to f) until six paired determinations have been performed.
444 h) If the DETERMINATION by the observers' with the REFERENCE SPHYGMOMANOMETER or the
445 SPHYGMOMANOMETER-UNDER-TEST is not successfully completed, repeat the DETERMINATION on the same
446 arm, i.e., without interchanging limb sides.
447 i) Repeat a) to h) until the required number of valid subjects and DETERMINATIONS has been clinically
448 investigated.
449 All data from a subject shall be excluded if
ISO/DIS 81060-2
450 ⎯ any two REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS on the same arm differ by more than
451 12 mmHg (1,60 kPa) or any two REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS on the same arm
452 differ by more than 8 mmHg (1,07 kPa).
453 ⎯ the lateral difference of the REFERENCE SYSTOLIC BLOOD PRESSURE DETERMINATIONS is more than 15 mmHg
454 (2,00 kPa) or the lateral difference of the REFERENCE DIASTOLIC BLOOD PRESSURE DETERMINATIONS is more
455 than 10 mmHg (1,33 kPa).
457 Key
458 1 double stethoscope
459 2 REFERENCE SPHYGMOMANOMETER display
460 3 REFERENCE SPHYGMOMANOMETER hand pump
461 4 SPHYGMOMANOMETER-UNDER-TEST
462 NOTE Only one CUFF is connected to the SPHYGMOMANOMETER-UNDER-TEST.
463 Figure 3 — Illustration of opposite arm simultaneous method
464 The lateral difference, LD, is calculated separately for SYSTOLIC and DIASTOLIC BLOOD PRESSURES, according to
465 Equation (12).
⎛⎞
⎜⎟
466 LD=× − (12)
∑∑
PP
ref_Rijref_L
3⎜⎟
ij==11
⎝⎠
467 where
468 P and P are REFERENCE BLOOD PRESSURE in right (R) limb and left (L) limb, respectively.
ref_Ri ref_Lj
469 5.2.4.3.2 * Data analysis
470 The SPHYGMOMANOMETER-UNDER-TEST error, x, is calculated by taking the difference between the
471 SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURE and REFERENCE SPHYGMOMANOMETER BLOOD PRESSURE and
ISO/DIS 81060-2
472 by adding the lateral difference, LD, as calculated according to Equation (13) if the SPHYGMOMANOMETER-
473 UNDER-TEST BLOOD PRESSURE was taken in the left arm or by subtracting the lateral difference, LD, as
474 calculated according to Equation (14) if the SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURE was taken in
475 the right arm.
476 x = P − P + LD (13)
sut_L ref_R
477 x = P − P − LD (14)
sut_R ref_L
478 where P and P are SPHYGMOMANOMETER-UNDER-TEST BLOOD PRESSURES in right (R) arm and left (L)
sut_R sut_L
479 arm, respectively.
480 The SPHYGMOMANOMETER-UNDER-TEST shall meet the following two criteria.
481 a) Criterion 1
482 For SYSTOLIC and DIASTOLIC BLOOD PRESSURES, the mean value of the differences of the DETERMINATIONS, x ,
n
483 of the n individual paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and of the observers'
484 DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER for all subjects shall be within or equal to
485 ± 5,0 mmHg (± 0,67 kPa), with a standard deviation, s , not greater than 8,0 mmHg (1,07 kPa) when
n
486 calculated according to Equation (15) and Equation (16).
n
487 x=× x (15)
n ∑ i
n
i=1
n
488 s=× (16)
xx−
n ∑()n
i
n−1
i=1
489 where
490 x is the SPHYGMOMANOMETER-UNDER-TEST error as calculated according to Equation (13) and
i
491 Equation (14);
492 n is the number of DETERMINATIONS;
493 i is the index for the individual element.
494 EXAMPLE 1 n = 510 for a SPHYGMOMANOMETER intended for use in adults and/or adolescent PATIENTS (an 85 subject
495 study).
496 EXAMPLE 2 n = 510 for a SPHYGMOMANOMETER intended for use in adults and/or adolescents and children aged
497 between 3 years and 12 years (an 85 subject study).
498 EXAMPLE 3 n = 210 for a SPHYGMOMANOMETER intended for (an additional) special INTENDED USE (a 35 subject study),
499 when the SPHYGMOMANOMETER has previously been clinically investigated in a separate 85 subject study according to 5.1.1
500 and 5.2.
501 b) Criterion 2
502 For the average of the SYSTOLIC and the average of the DIASTOLIC BLOOD PRESSURES for each subject, the
503 standard deviation, s , of the m averaged paired DETERMINATIONS of the SPHYGMOMANOMETER-UNDER-TEST and
m
504 of the observers' DETERMINATIONS with the REFERENCE SPHYGMOMANOMETER, per subject, shall meet the criteria
505 listed in Table 1 or Table 2 when calculated according to Equation (17).
ISO/DIS 81060-2
m
506 s=× (17)
x − x
m n
∑()
j
m−1
j=1
507 where
508 is the mean value as calculated according to Equation (15);
x
n
509 m is the number of subjects;
510 j is the index for the individual element;
511 x is the mean per subject calculated
...
NORME ISO
INTERNATIONALE 81060-2
Deuxième édition
2013-05-01
Sphygmomanomètres non invasifs —
Partie 2:
Validation clinique pour type à mesurage
automatique
Non-invasive sphygmomanometers—
Part 2: Clinical investigation of automated measurement type
Numéro de référence
©
ISO/CEI 2013
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO/CEI 2013
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur l’internet ou sur un
Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à l’adresse ci-après ou au comité
membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Version française parue en 2014
Publié en Suisse
ii © ISO/CEI 2013 – Tous droits réservés
Sommaire Page
Avant-propos . iv
Introduction . vi
1 Domaine d'application . 1
2 Références normatives . 1
3 Termes et définitions . 2
4 Exigences générales relatives aux INVESTIGATIONS CLINIQUES . 2
5 INVESTIGATION CLINIQUE avec un SPHYGMOMANOMETRE DE REFERENCE auscultatoire . 3
6 INVESTIGATION CLINIQUE au moyen d’un APPAREIL DE SURVEILLANCE DE LA PRESSION SANGUINE
PRELEVEE DIRECTEMENT . 18
7 * Populations de PATIENTES enceintes (y compris patientes pré-éclamptiques) . 23
Annexe A (informative) Justification et lignes directrices . 25
Annexe B (normative) Fréquences cardiaques cibles pour les épreuves d'effort . 38
Annexe C (informative) Renvoi aux principes essentiels . 39
Bibliographie . 40
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
La tâche principale des comités techniques est d'élaborer les Normes internationales. Les projets de Normes
internationales adoptés par les comités techniques sont soumis aux comités membres pour vote. Leur
publication comme Normes internationales requiert l'approbation de 75 % au moins des comités membres
votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de propriété et averti de leur existence.
Cette deuxième édition annule et remplace la première édition (ISO 81060-2:2009), dont les paragraphes
5.2.4.3.1 et 6.2.4 ont fait l'objet d'une révision technique. De nombreux éclaircissements ont été ajoutés et les
valeurs en kPa équivalentes aux valeurs en mmHg ont été incluses dans la norme, y compris les exigences
du critère 2 du 5.2.4.1.2. Elle incorpore également le Rectificatif technique ISO 81060-2:2009/Cor 1:2011.
L'ISO 81060-2 a été élaborée par le comité technique ISO/TC 121, Matériel d'anesthésie et de réanimation
respiratoire, sous-comité SC 3, Ventilateurs pulmonaires et équipements connexes, en collaboration avec le
comité technique CEI/TC 62, Équipements électriques dans la pratique médicale, sous-comité 62D, Appareils
électromédicaux, conformément au mode de collaboration 5 de l'ISO/CEI.
L’ISO 81060 comprend les parties suivantes, présentées sous le titre général Sphygmomanomètres non
invasifs:
Partie 1: Exigences et méthodes d’essai pour type à mesurage non automatique;
Partie 2: Investigation clinique pour type à mesurage automatique.
Dans le présent document, les caractères suivants sont employés:
exigences dont la conformité peut être vérifiée et définitions: caractères romains;
notes et exemples: petits caractères romains;
méthodes d'essai: italiques;
termes définis dans le présent document: PETITES MAJUSCULES.
Tout au long du présent document, le texte pour lequel une justification est fournie dans l’Annexe A est
précédé d’un astérisque (*).
iv © ISO 2013 – Tous droits réservés
L’attention des comités membres et des comités nationaux est attirée sur le fait que les fabricants
d'équipements et les organismes d'essai peuvent avoir besoin d'une période de transition après la publication
d'une norme ISO ou CEI nouvelle, amendée ou révisée, pour mettre leurs produits en conformité avec les
nouvelles exigences et s'équiper afin de réaliser les nouveaux essais ou les essais révisés. L'ISO/TC 121 et
le CEI/TC 62 recommandent que le contenu de la présente partie de l'ISO 81060 ne soit adopté en vue de sa
mise en œuvre obligatoire à l'échelle nationale que trois ans au plus tôt après sa date de publication pour les
équipements de conception nouvelle, et cinq ans au plus tôt après sa date de publication pour les
équipements déjà en production.
Introduction
La détermination de la PRESSION ARTERIELLE est une procédure importante utilisée en pratique clinique pour
évaluer l'état d'un PATIENT.
En cours d’anesthésie, il est systématiquement procédé à une surveillance fréquente de la PRESSION
ARTERIELLE. La PRESSION ARTERIELLE aide au dosage de l’anesthésique et permet une meilleure gestion des
fluides; elle permet également de prévenir des états qui pourraient affecter la morbidité et la mortalité du
PATIENT.
vi © ISO 2013 – Tous droits réservés
Error! Reference source not found. ISO 81060-2:2013(F)
Sphygmomanomètres non invasifs — Partie 2: Investigation
clinique pour type à mesurage automatique
1 Domaine d'application
La présente partie de l’ISO 81060 précise les exigences et les méthodes d'INVESTIGATION CLINIQUE des
APPAREILS ELECTROMEDICAUX utilisés pour estimer ponctuellement, de manière non invasive et automatique, la
PRESSION ARTERIELLE au moyen d’un BRASSARD.
La présente partie de l’ISO 81060 est applicable à tous les SPHYGMOMANOMETRES qui captent ou affichent des
pulsations, des flux ou des sons pour l’estimation, l’affichage ou l’enregistrement de la PRESSION ARTERIELLE. Il
n’est pas nécessaire que ces SPHYGMOMANOMETRES aient un dispositif de gonflage automatique du BRASSARD.
La présente partie de l’ISO 81060 couvre les SPHYGMOMANOMETRES utilisables pour toutes les populations de
PATIENTS (par exemple: toutes les tranches d’âge et gammes de poids) et dans toutes les conditions d’emploi
(par exemple: contrôle ambulatoire de la PRESSION ARTERIELLE, contrôle de la PRESSION ARTERIELLE lors
d’épreuves d’effort, auto-contrôle de la PRESSION ARTERIELLE en ENVIRONNEMENT DE SOINS A DOMICILE et
contrôle de la pression artérielle dans un établissement de soins de santé).
EXEMPLE SPHYGMOMANOMETRE AUTOMATIQUE tel que défini dans la CEI 80601-2-30 faisant l'objet d'une INVESTIGATION
CLINIQUE conformément à la présente partie de l’ISO 81060.
La présente partie de l’ISO 81060 spécifie des exigences supplémentaires de divulgation d’informations pour
les DOCUMENTS D’ACCOMPAGNEMENT des SPHYGMOMANOMETRES ayant fait l'objet d'une INVESTIGATION CLINIQUE
conformément à la présente partie de l’ISO 81060.
La présente partie de l’ISO 81060 n’est pas applicable à l'INVESTIGATION CLINIQUE des SPHYGMOMANOMETRES
NON AUTOMATIQUES tels que définis dans l’ISO 81060-1 ou des APPAREILS DE SURVEILLANCE DE LA PRESSION
SANGUINE PRELEVEE DIRECTEMENT tels que définis dans la CEI 60601-2-34.
2 Références normatives
Les documents ci-après, dans leur intégralité ou non, sont des références normatives indispensables à
l’application du présent document. Pour les références datées, seule l'édition citée s'applique. Pour les
références non datées, la dernière édition du document de référence s'applique (y compris les éventuels
amendements).
ISO 14155:2011, Investigation clinique des dispositifs médicaux pour sujets humains — Bonnes pratiques
cliniques
ISO 81060-1, Sphygmomanomètres non invasifs — Partie 1: Exigences et méthodes d'essai pour type à
mesurage non automatique
CEI 80601-2-30:2009, Appareils électromédicaux — Partie 2-30: Exigences particulières pour la sécurité de
base et les performances essentielles des sphygmomanomètres non invasifs automatiques
CEI 60601-1:2005, Appareils électromédicaux — Partie 1: Exigences générales pour la sécurité de base et
les performances essentielles
Amendement 1:2012
CEI 60601-1-11:2010, Appareils électromédicaux – Partie 1-11: Exigences générales pour la sécurité de base
et les performances essentielles – Norme Collatérale: Exigences pour les appareils électromédicaux et les
systèmes électromédicaux utilisés dans l’environnement des soins à domicile
CEI 60601-2-34:2011, Appareils électromédicaux — Partie 2-34: Exigences particulières pour la sécurité de
base et les performances essentielles des appareils de surveillance de la pression sanguine prélevée
directement
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions donnés dans l’ISO 14155, la CEI 80601-2-30,
la CEI 60601-1, la CEI 60601-1-11 et la CEI 60601-2-34 ainsi que les suivants s’appliquent.
NOTE Pour des raisons pratiques, un index alphabétique des termes définis est donné à partir de la page 43.
3.1
REFERENCE, adj
exactitude établie utilisée pour l'INVESTIGATION CLINIQUE d’autres instruments
3.2
SPHYGMOMANOMETRE
APPAREIL ELECTROMEDICAL d’estimation non invasive de la PRESSION ARTERIELLE systémique
3.3
SPHYGMOMANOMETRE EN ESSAI
SPHYGMOMANOMETRE faisant l'objet d'une INVESTIGATION CLINIQUE
4 Exigences générales relatives aux INVESTIGATIONS CLINIQUES
4.1 Méthodes d'INVESTIGATION CLINIQUE
Les SPHYGMOMANOMETRES autres que les SPHYGMOMANOMETRES NON AUTOMATIQUES doivent faire l'objet d'une
INVESTIGATION CLINIQUE, soit en utilisant un SPHYGMOMANOMETRE DE REFERENCE non invasif (auscultatoire), soit
au moyen d’un APPAREIL DE SURVEILLANCE DE LA PRESSION SANGUINE PRELEVEE DIRECTEMENT, conformément à
la présente partie de l’ISO 81060 pour chaque mode de fonctionnement.
EXEMPLE 1 Modes adulte et nouveau-né.
EXEMPLE 2 Modes de vitesse de dégonflage lente et rapide du BRASSARD.
Une INVESTIGATION CLINIQUE doit être considérée comme un ESSAI DE TYPE.
Les exigences du présent paragraphe sont considérées satisfaites lorsque les critères des examens et essais
applicables de la présente partie de l’ISO 81060 sont remplis.
4.2 Bonnes pratiques cliniques
Toutes les INVESTIGATIONS CLINIQUES doivent satisfaire aux exigences de l'ISO 14155. Il convient de ne pas
effectuer d'INVESTIGATION CLINIQUE au moyen d'APPAREILS DE SURVEILLANCE DE LA PRESSION SANGUINE PRELEVEE
DIRECTEMENT sur les PATIENTS ou des sujets dans le seul but d'étudier les performances de
SPHYGMOMANOMETRES.
NOTE Certaines autorités compétentes ont des exigences supplémentaires.
Les exigences de la présente Norme internationale, qui sont plus spécifiques que les exigences
correspondantes de l'ISO 14155, doivent prévaloir.
La conformité est vérifiée selon les exigences de l’ISO 14155.
5 INVESTIGATION CLINIQUE avec un SPHYGMOMANOMETRE DE REFERENCE auscultatoire
5.1 Exigences relatives aux sujets
5.1.1 * Nombre de sujets
L’INVESTIGATION CLINIQUE au moyen d’un SPHYGMOMANOMETRE DE REFERENCE auscultatoire doit comprendre au
minimum 85 sujets. Sauf spécification contraire, chaque sujet doit faire l’objet d’au moins trois
DETERMINATIONS valides de la PRESSION ARTERIELLE. Il doit y avoir au minimum 255 paires de DETERMINATIONS
valides de la PRESSION ARTERIELLE.
La conformité est vérifiée par examen du RAPPORT d'INVESTIGATION CLINIQUE.
5.1.2 * Distribution par sexe
Au moins 30 % des sujets doivent être de sexe masculin et 30 % de sexe féminin.
La conformité est vérifiée par examen du RAPPORT D'INVESTIGATION CLINIQUE.
5.1.3 * Distribution par âge
Pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents, chaque sujet inclus dans
l’INVESTIGATION CLINIQUE doit être âgé de plus de 12 ans.
NOTE 1 Nombre total minimal: 85 sujets.
Pour un SPHYGMOMANOMETRE qui est en outre destiné à des enfants, 35 sujets âgés de 3 ans à 12 ans
doivent être inclus dans l’INVESTIGATION CLINIQUE.
NOTE 2 Nombre total minimal: 85 sujets.
Si le SPHYGMOMANOMETRE comporte un mode spécial pour enfants, les enfants doivent être considérés
comme une population spéciale de PATIENTS lorsque l’appareil est utilisé dans ce mode (voir 5.1.6). Dans une
telle étude, la population d’enfants est exemptée des exigences de distribution de la PRESSION ARTERIELLE
décrites en 5.1.5.
Les enfants âgés de moins de 3 ans ne doivent pas être inclus dans une INVESTIGATION CLINIQUE utilisant des
DETERMINATIONS auscultatoires réalisées par des observateurs avec un SPHYGMOMANOMETRE DE REFERENCE.
La conformité est vérifiée par examen du DOCUMENT D'ACCOMPAGNEMENT et du RAPPORT D'INVESTIGATION
CLINIQUE.
5.1.4 * Distribution des tailles de bras
Lorsque le SPHYGMOMANOMETRE est destiné à être utilisé avec une seule taille de BRASSARD:
au moins 40 % des sujets doivent avoir une circonférence du bras s’inscrivant dans la moitié supérieure
de la plage spécifiée d’utilisation du BRASSARD et au moins 40 % doivent avoir une circonférence du bras
dans la moitié inférieure de cette même plage; et
au moins 20 % des sujets doivent avoir une circonférence du bras s’inscrivant dans le quart supérieur de
la plage spécifiée d’utilisation du BRASSARD et au moins 20 % doivent avoir une circonférence du bras
dans le quart inférieur de cette même plage.
Lorsque le SPHYGMOMANOMETRE est destiné à être utilisé avec plusieurs tailles de BRASSARD, chaque taille de
BRASSARD doit être soumise à essai sur au moins des sujets, où n est le nombre de tailles de
2 n
BRASSARD.
La conformité est vérifiée par examen du DOCUMENT D'ACCOMPAGNEMENT et du RAPPORT D'INVESTIGATION
CLINIQUE.
5.1.5 * Distribution de la PRESSION ARTERIELLE
Au moins 5 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
SYSTOLIQUE ≤ 100 mmHg (13,33 kPa).
Au moins 5 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
SYSTOLIQUE ≥ 160 mmHg (21,33 kPa).
Au moins 20 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
SYSTOLIQUE ≥ 140 mmHg (18,66 kPa).
Au moins 5 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
DIASTOLIQUE ≤ 60 mmHg (8,0 kPa).
Au moins 5 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
DIASTOLIQUE ≥ 100 mmHg (13,33 kPa).
Au moins 20 % des lectures de PRESSION ARTERIELLE DE REFERENCE doivent avoir une PRESSION ARTERIELLE
DIASTOLIQUE ≥ 85 mmHg (11,33 kPa).
La conformité est vérifiée par examen du RAPPORT D'INVESTIGATION CLINIQUE.
5.1.6 * Populations spéciales de PATIENTS
Lorsqu’il existe une PREUVE OBJECTIVE que l’exactitude d’un SPHYGMOMANOMETRE conçu pour des populations
spéciales de PATIENTS pourrait poser problème, le SPHYGMOMANOMETRE doit faire l’objet d’une INVESTIGATION
CLINIQUE sur ladite population de PATIENTS.
NOTE L'Article 7 contient un exemple spécifique d'une population spéciale de PATIENTS ayant des exigences
spécifiques.
Si le SPHYGMOMANOMETRE a fait l'objet d'une INVESTIGATION CLINIQUE conformément aux exigences du 5.1.1 et
du 5.2, il doit ensuite faire l'objet d'une INVESTIGATION CLINIQUE sur au moins 35 sujets supplémentaires de la
population spéciale. Si le SPHYGMOMANOMETRE n'a pas fait préalablement l'objet d'une INVESTIGATION CLINIQUE
conformément aux exigences du 5.1.1 et du 5.2, l'INVESTIGATION CLINIQUE conformément aux exigences
du 5.1.1 et du 5.2 doit être réalisée uniquement sur des sujets issus de la population spéciale de PATIENTS.
La population spéciale de PATIENTS doit être définie en termes clairs et faire état des attributs suivants: sexe
(voir 5.1.2), âge (voir 5.1.3), taille de bras (voir 5.1.4) et PRESSION ARTERIELLE (voir 5.1.5). Une synthèse de
ces informations doit être contenue dans les instructions d’utilisation.
La conformité est vérifiée par examen des instructions d'utilisation et du RAPPORT D'INVESTIGATION CLINIQUE.
5.2 Méthode d'INVESTIGATION CLINIQUE avec un SPHYGMOMANOMETRE DE REFERENCE
5.2.1 * Préparation des sujets
Sauf spécification contraire dans les instructions d’utilisation du SPHYGMOMANOMETRE EN ESSAI, positionner le
sujet de façon à ce:
qu’il soit confortablement installé;
EXEMPLE Confortablement assis avec les jambes décroisées et les pieds à plat sur le sol.
que son dos, ses coudes et ses avant-bras soient soutenus;
que le milieu du BRASSARD soit situé au niveau de l’oreillette droite.
Il est recommandé que le sujet soit le plus détendu possible et qu’il évite de parler pendant toute la procédure.
Il convient d’attendre 5 min avant d’effectuer la première lecture.
NOTE Des détails supplémentaires sont donnés dans la référence [32].
5.2.2 * Préparation des observateurs
Il convient que les observateurs soient formés à l’application d’une méthodologie correcte de DETERMINATION
de la PRESSION ARTERIELLE au repos fondée sur un protocole clinique accepté de mesure de la PRESSION
ARTERIELLE. Les références [8], [28], [29], [32] et [45] contiennent des informations supplémentaires. Il
convient que les observateurs aient une pratique suffisante en matière de DETERMINATION de la PRESSION
ARTERIELLE.
L’enregistrement des mesures du SPHYGMOMANOMETRE DE REFERENCE effectuées par chaque observateur ne
doit pas être visible pour l’autre observateur. Les lectures du SPHYGMOMANOMETRE EN ESSAI ne doivent être
visibles pour aucun de ces observateurs.
EXEMPLE 1 Utilisation d’un troisième observateur pour enregistrer les lectures du SPHYGMOMANOMETRE EN ESSAI.
EXEMPLE 2 Utilisation d’un dispositif électronique pour enregistrer les lectures du SPHYGMOMANOMETRE EN ESSAI.
Instruire les observateurs pour que la détermination de la PRESSION ARTERIELLE DIASTOLIQUE soit effectuée au
dernier bruit audible de Korotkoff (cinquième phase ou K5), sauf lorsque les bruits de Korotkoff sont encore
audibles avec le BRASSARD dégonflé ou pour les enfants âgés de 3 ans à 12 ans, auquel cas la quatrième
phase (K4) est utilisée. Si K4 n'est pas audible chez un enfant, utiliser K5 ou exclure le sujet.
NOTE Outre les enfants, il convient de réserver K4 aux sujets présentant un important écart entre assourdissement et
disparition du bruit (quelquefois proche de zéro mmHg).
Instruire les observateurs pour qu’ils indiquent dans leurs enregistrements quel bruit de Korotkoff a été utilisé
pour la DETERMINATION de la PRESSION ARTERIELLE DIASTOLIQUE.
Le bruit de Korotkoff utilisé pour la DETERMINATION de la PRESSION ARTERIELLE DIASTOLIQUE dans le cadre de
l’INVESTIGATION CLINIQUE doit être mentionné dans les instructions d’utilisation d’un SPHYGMOMANOMETRE.
EXEMPLE 3 K5 a été utilisé sur 65 sujets et K4 a été utilisé sur 20 sujets.
5.2.3 * DETERMINATION DE REFERENCE
Deux observateurs doivent effectuer des DETERMINATIONS simultanées de la PRESSION ARTERIELLE de chaque
sujet au moyen d’un stéthoscope double.
Cette DETERMINATION doit être exclue à moins que le SPHYGMOMANOMETRE EN ESSAI ne soit conçu pour des
mesures impliquant des rythmes cardiaques substantiellement irréguliers et si les deux observateurs
détectent un rythme cardiaque substantiellement irrégulier.
EXEMPLES Bigéminisme, trigéminisme, extrasystole ventriculaire (ESV) isolée, fibrillation auriculaire.
NOTE 1 Bien que l’INVESTIGATION CLINIQUE de la PRESSION ARTERIELLE chez des PATIENTS présentant une fibrillation
auriculaire soit importante en pratique clinique, il n’existe pas actuellement de règles généralement admises pour la
détermination de la PRESSION ARTERIELLE de ces personnes.
Toute paire de DETERMINATIONS d’observateurs présentant une différence supérieure à 4 mmHg (0,53 kPa)
doit être exclue. La moyenne des valeurs individuelles de chaque DETERMINATION effectuée par les
observateurs doit être calculée selon la Formule (1) pour obtenir la DETERMINATION de la PRESSION ARTERIELLE
DE REFERENCE.
p p
ref ref
i,1 i,2
p (1)
ref
i
où:
ème
p est la PRESSION ARTERIELLE déterminée par l'observateur 1 pour la i DETERMINATION;
ref
i,1
ème
p est la PRESSION ARTERIELLE déterminée par l'observateur 2 pour la i DETERMINATION;
ref
i,2
ème
p est la PRESSION ARTERIELLE DE REFERENCE pour la i DETERMINATION.
ref
i
Les différences de résultats entre observateurs doivent être examinées une fois obtenu un ensemble de
paires de DETERMINATIONS de REFERENCE d’essai. Si d’éventuelles DETERMINATIONS sont exclues, une (des)
paire(s) de DETERMINATIONS supplémentaires doi(vent) être effectuée(s) pour disposer du nombre requis de
paires valides de REFERENCE d’essai. Il convient d’effectuer au maximum huit paires de DETERMINATIONS.
Utiliser un SPHYGMOMANOMETRE DE REFERENCE conforme aux exigences de l’ISO 81060-1, à l’exception du fait
que l’erreur maximale tolérée doit être de 1 mmHg (0,13 kPa). Il est souhaitable que la lecture des valeurs
du SPHYGMOMANOMETRE DE REFERENCE soit aussi exacte que possible. Il convient en l’occurrence que les
observateurs évitent les erreurs de parallaxe et d'arrondissage.
NOTE 2 L’arrondissage a un effet préjudiciable sur les résultats de l'INVESTIGATION CLINIQUE.
NOTE 3 Pour les besoins de la présente partie de l'ISO 81060, le BRASSARD est considéré comme faisant partie du
SPHYGMOMANOMETRE DE REFERENCE. Il n’est pas possible d’utiliser un BRASSARD non conforme à l’ISO 81060-1.
5.2.4 Méthodes d'INVESTIGATION CLINIQUE
5.2.4.1 Méthode de détermination simultanée sur un même bras
5.2.4.1.1 * Mode opératoire
Cette méthode ne doit être utilisée qu’avec un SPHYGMOMANOMETRE EN ESSAI:
ayant un BRASSARD conforme à l'ISO 81060-1;
conçu pour être utilisé sur la partie supérieure du bras; et
dont
la vitesse de dégonflage linéaire continue est de 2 mmHg/s (0,27 kPa/s) à 3 mmHg/s (0,40 kPa/s) ou
si le dégonflage est contrôlé par un SPHYGMOMANOMETRE EN ESSAI en fonction de la fréquence du
pouls, la vitesse de dégonflage est comprise entre 2 mmHg/pulsation (0,27 kPa/pulsation) et
3 mmHg/pulsation (0,40 kPa/pulsation).
N'importe quel bras peut être utilisé.
Le SPHYGMOMANOMETRE EN ESSAI ne doit pas se dégonfler avant détection de la PRESSION ARTERIELLE
DIASTOLIQUE DE REFERENCE. Le SPHYGMOMANOMETRE EN ESSAI peut être modifié pour remplir ce critère.
NOTE Pour qu’elles soient valides, les DETERMINATIONS simultanées sur un même bras exigent que le
SPHYGMOMANOMETRE EN ESSAI puisse gonfler le BRASSARD à une pression supérieure d’au moins 20 mmHg (2,67 kPa) à la
PRESSION ARTERIELLE SYSTOLIQUE réelle, telle que déterminée par le SPHYGMOMANOMETRE DE REFERENCE, et inférieure d’au
moins 20 mmHg (2,67 kPa) à la PRESSION ARTERIELLE DIASTOLIQUE réelle, telle que déterminée par le SPHYGMOMANOMETRE
DE REFERENCE.
La procédure est la suivante:
a) Les observateurs utilisant le SPHYGMOMANOMETRE DE REFERENCE et le SPHYGMOMANOMETRE EN ESSAI
déterminent simultanément la PRESSION ARTERIELLE du sujet au moyen du même BRASSARD et selon le
même cycle de gonflage/dégonflage (voir Figure 1). Ces points de données ne sont pas utilisés pour le
calcul de l'exactitude du SPHYGMOMANOMETRE EN ESSAI.
b) Effacer de la mémoire du SPHYGMOMANOMETRE EN ESSAI la DETERMINATION précédente et attendre ensuite
au moins 60 s.
EXEMPLES La mise hors tension puis sous tension, le retrait du module de détermination de la PRESSION ARTERIELLE
et sa réinstallation ou une commande de réinitialisation sont des méthodes permettant d'effacer la DETERMINATION
précédente de la mémoire.
c) Les observateurs utilisant le SPHYGMOMANOMETRE DE REFERENCE et le SPHYGMOMANOMETRE EN ESSAI
déterminent simultanément la PRESSION ARTERIELLE du sujet au moyen du même BRASSARD et selon le
même cycle de gonflage/dégonflage.
d) Attendre au moins 60 s entre les DETERMINATIONS.
e) Répéter les points c) et d) jusqu'à ce que le nombre requis de DETERMINATIONS valides ait été effectué.
L’ensemble des données obtenues pour un sujet donné doit être exclu si deux DETERMINATIONS quelconques
de la PRESSION ARTERIELLE SYSTOLIQUE DE REFERENCE présentent une différence de plus de 12 mmHg
(1,60 kPa) ou si deux DETERMINATIONS quelconques de la PRESSION ARTERIELLE DIASTOLIQUE DE REFERENCE
présentent une différence de plus de 8 mmHg (1,07 kPa).
Légende
1 stéthoscope double
2 affichage du SPHYGMOMANOMETRE DE REFERENCE
3 SPHYGMOMANOMETRE EN ESSAI
Figure 1 — Illustration de la méthode de détermination simultanée sur un même bras
5.2.4.1.2 * Analyse des données
Le SPHYGMOMANOMETRE EN ESSAI doit satisfaire aux deux critères suivants:
a) Critère 1
Pour les PRESSIONS ARTERIELLES SYSTOLIQUE et DIASTOLIQUE, la valeur moyenne des différences entre les
DETERMINATIONS, x , des n DETERMINATIONS individuelles appariées réalisées avec le SPHYGMOMANOMETRE EN
n
ESSAI et des DETERMINATIONS réalisées par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE pour
tous les sujets, doivent être inférieures ou égales à ± 5,0 mmHg (± 0,67 kPa), avec un écart-type, s , ne
n
dépassant pas 8,0 mmHg (1,07 kPa), le calcul étant effectué selon la Formule (2) et la Formule (3):
n
x p p (2)
n sut ref
i i
n
i1
n
s x x (3)
n i n
n1
i1
où:
x est la valeur moyenne des différences;
n
p p = x différence entre une paire de DETERMINATIONS de la PRESSION ARTERIELLE
sut ref i
i i
[DETERMINATION réalisée avec le SPHYGMOMANOMETRE EN ESSAI PRESSION ARTERIELLE DE
REFERENCE, calculée selon la Formule (1)];
i est l’indice de l’élément individuel;
n est le nombre de DETERMINATIONS;
x et s doivent être calculés et exprimés à 0,1 mmHg (0,01 kPa) près.
n
n
EXEMPLE 1 n = 255 pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents (étude portant
sur 85 sujets).
EXEMPLE 2 n = 255 pour un SPHYGMOMANOMETRE destiné à des adultes et/ou des adolescents ainsi qu’à des
enfants âgés de 3 ans à 12 ans (étude portant sur 85 sujets).
EXEMPLE 3 n = 105 pour un SPHYGMOMANOMETRE destiné à un USAGE particulier (étude portant sur 35 sujets), lorsque
le SPHYGMOMANOMETRE a préalablement fait l'objet d'une INVESTIGATION CLINIQUE lors d'une étude séparée portant sur
85 sujets conformément à 5.1.1 et 5.2.
b) Critère 2
Pour les PRESSIONS ARTERIELLES SYSTOLIQUE et DIASTOLIQUE de chacun des m sujets, l'écart-type, s , de la
m
moyenne des DETERMINATIONS appariées, pour chaque sujet, réalisées avec le SPHYGMOMANOMETRE EN ESSAI
et des DETERMINATIONS réalisées par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE, doit
remplir les critères du Tableau 1 ou du Tableau 2, le calcul étant effectué selon la Formule (4):
m
s x x (4)
m j n
m 1
j1
où:
x est la valeur moyenne des différences, calculée selon la Formule (2);
n
m est le nombre de sujets;
est l’indice de l’élément individuel;
x
j
m est calculé selon la Formule (5).
d
x p p (5)
j sut ref
k k
d
k1
où:
d est le nombre de DETERMINATIONS par sujet;
k est l’indice de l’élément individuel.
EXEMPLE 4 m = 85 pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents (étude portant sur
85 sujets).
EXEMPLE 5 m = 85 pour un SPHYGMOMANOMETRE destiné à des adultes et/ou des adolescents ainsi qu’à des enfants
âgés de 3 ans à 12 ans (étude portant sur 85 sujets).
EXEMPLE 6 m = 35 pour un SPHYGMOMANOMETRE destiné à un USAGE (supplémentaire) particulier (étude portant sur
35 sujets), lorsque le SPHYGMOMANOMETRE a préalablement fait l'objet d'une INVESTIGATION CLINIQUE lors d'une étude
séparée portant sur 85 sujets conformément à 5.1.1 et 5.2.
EXEMPLE 7 m = 85 pour un SPHYGMOMANOMETRE destiné uniquement à un USAGE particulier (étude portant sur
85 sujets).
Tableau 1 — Acceptation des données moyennées des sujets (critère 2) en mmHg
Écart-type maximal admissible, s , en fonction de x
m n
mmHg
x
n
0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9
0, 6,95 6,95 6,95 6,95 6,93 6,92 6,91 6,90 6,89 6,88
1, 6,87 6,86 6,84 6,82 6,80 6,78 6,76 6,73 6,71 6,68
2, 6,65 6,62 6,58 6,55 6,51 6,47 6,43 6,39 6,34 6,30
3, 6,25 6,20 6,14 6,09 6,03 5,97 5,89 5,83 5,77 5,70
4, 5,64 5,56 5,49 5,41 5,33 5,25 5,16 5,08 5,01 4,90
5, 4,79 — — — — — — — — —
EXEMPLE Pour une moyenne de 4,2 mmHg, l’écart-type maximal admissible est de 5,49 mmHg.
Tableau 2 — Acceptation des données moyennées des sujets (critère 2) en kPa
Écart-type maximal admissible, s , en fonction de x
m n
kPa
x
n
0,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070 0,080 0,090
0,0 0,9266 0,9266 0,9266 0,9266 0,9266 0,9246 0,9233 0,9223 0,9213 0,9203
0,1 0,9193 0,9183 0,9173 0,9163 0,9152 0,9138 0,9119 0,9099 0,9079 0,9059
0,2 0,9039 0,9007 0,8989 0,8970 0,8946 0,8906 0,8878 0,8855 0,8826 0,8785
0,3 0,8756 0,8723 0,8679 0,8641 0,8601 0,8562 0,8519 0,8471 0,8414 0,8374
0,4 0,8333 0,8283 0,8226 0,8169 0,8119 0,8059 0,7999 0,7933 0,7853 0,7793
0,5 0,7739 0,7669 0,7599 0,7531 0,7463 0,7388 0,7319 0,7237 0,7157 0,7077
0,6 0,6999 0,6891 0,6802 0,6723 0,6670 0,6595 0,6488 0,6386 — —
EXEMPLE Pour une moyenne de 0 520 kPa, l’écart-type maximal admissible est de 0,7599 kPa.
5.2.4.2 * Méthode de détermination séquentielle sur un même bras
5.2.4.2.1 * Mode opératoire
N'importe quel bras peut être utilisé.
La procédure est la suivante:
a) A l'aide du SPHYGMOMANOMETRE DE REFERENCE, les observateurs déterminent la PRESSION ARTERIELLE du
sujet (voir Figure 2).
b) Attendre au moins 60 s.
c) A l'aide du SPHYGMOMANOMETRE EN ESSAI, les observateurs déterminent la PRESSION ARTERIELLE du sujet.
d) Effacer de la mémoire du SPHYGMOMANOMETRE EN ESSAI la DETERMINATION précédente et attendre ensuite
au moins 60 s.
EXEMPLES La mise hors tension puis sous tension, le retrait du module de détermination de la PRESSION ARTERIELLE
et sa réinstallation ou une commande de réinitialisation sont des méthodes permettant d'effacer la DETERMINATION
précédente de la mémoire.
Légende
1 stéthoscope double
2 affichage du SPHYGMOMANOMETRE DE REFERENCE
3 pompe à main du SPHYGMOMANOMETRE DE REFERENCE
4 SPHYGMOMANOMETRE EN ESSAI
NOTE Un seul BRASSARD est raccordé au SPHYGMOMANOMETRE EN ESSAI.
Figure 2 — Illustration de la méthode de détermination séquentielle sur un même bras
e) Ne pas utiliser les points de données obtenus en a) et c) pour la DETERMINATION de l'exactitude.
f) A l'aide du SPHYGMOMANOMETRE DE REFERENCE, les observateurs déterminent la PRESSION ARTERIELLE du
sujet.
g) Les observateurs utilisent le SPHYGMOMANOMETRE EN ESSAI et le SPHYGMOMANOMETRE DE REFERENCE pour
déterminer séquentiellement la PRESSION ARTERIELLE du sujet.
h) Attendre au moins 60 s entre chaque DETERMINATION.
i) Répéter les points g) et h) jusqu'à ce que le nombre requis de DETERMINATIONS valides ait été effectué.
L’ensemble des données obtenues pour un sujet donné doit être exclu si deux DETERMINATIONS quelconques
de la PRESSION ARTERIELLE SYSTOLIQUE DE REFERENCE présentent une différence de plus de 12 mmHg
(1,60 kPa) ou si deux DETERMINATIONS quelconques de la PRESSION ARTERIELLE DIASTOLIQUE DE REFERENCE
présentent une différence de plus de 8 mmHg (1,07 kPa). Nonobstant cette exigence, si la PRESSION
ARTERIELLE DE REFERENCE d'un sujet individuel ne remplit pas ces critères pendant la durée de l'essai, il est
possible d'utiliser deux paires de DETERMINATIONS consécutives conformes répondant à ces critères. Dans ce
cas, des sujets supplémentaires doivent être utilisés pour atteindre le nombre minimal de DETERMINATIONS. Le
pourcentage de sujets ayant fait l’objet de moins de trois paires de DETERMINATIONS valides ne doit pas
dépasser 10 %.
5.2.4.2.2 Analyse des données
Le SPHYGMOMANOMETRE EN ESSAI doit satisfaire aux deux critères suivants:
a) Critère 1
Pour les PRESSIONS ARTERIELLES SYSTOLIQUE et DIASTOLIQUE, la valeur moyenne des différences entre les
DETERMINATIONS, x , des n DETERMINATIONS individuelles appariées réalisées avec le SPHYGMOMANOMETRE EN
n
ESSAI et des DETERMINATIONS réalisées par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE pour
tous les sujets, doit être inférieure ou égale à ± 5,0 mmHg (± 0,67 kPa), avec un écart-type, s , ne dépassant
n
pas 8,0 mmHg (1,07 kPa), le calcul étant effectué selon la Formule (6) et la Formule (7):
n
(6)
x p p
n sut ref -sq
i
i
n
i1
n
s x x (7)
n i n
n 1
i1
où:
x est la valeur moyenne des différences;
n
ème
p p x différence entre les i DETERMINATIONS appariées de la PRESSION ARTERIELLE
sut ref -sq i
i i
(DETERMINATION réalisée avec le SPHYGMOMANOMETRE EN ESSAI PRESSION ARTERIELLE DE
REFERENCE);
i est l’indice de l’élément individuel;
n est le nombre de DETERMINATIONS;
ème
p est la PRESSION ARTERIELLE DE REFERENCE pour la i DETERMINATION, calculée selon la
ref-sq
i
Formule (8);
x et s doivent être calculés et exprimés à au moins 0,1 mmHg (0,01 kPa) près.
n n
La PRESSION ARTERIELLE DE REFERENCE, p , (DETERMINATIONS par les observateurs avec le
ref-sq
i
SPHYGMOMANOMETRE DE REFERENCE) doit être la moyenne des PRESSIONS ARTERIELLES DE REFERENCE
précédente et suivante.
p p p p p (8)
ref -sq ref ref ref ref
i,1 i,2 i1,1 i1,2
i
où:
ème
p est la PRESSION ARTERIELLE déterminée par l'observateur 1 pour la i DETERMINATION;
ref
i,1
ème
p est la PRESSION ARTERIELLE déterminée par l'observateur 2 pour la i DETERMINATION.
ref
i,2
EXEMPLE 1 n = 255 pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents (étude portant sur
85 sujets).
EXEMPLE 2 n = 255 pour un SPHYGMOMANOMETRE destiné à des adultes et/ou des adolescents ainsi qu’à des enfants
âgés de 3 ans à 12 ans (étude portant sur 85 sujets).
EXEMPLE 3 n = 105 pour un SPHYGMOMANOMETRE destiné à un USAGE particulier (étude portant sur 35 sujets), lorsque
le SPHYGMOMANOMETRE a préalablement fait l'objet d'une INVESTIGATION CLINIQUE lors d'une étude séparée portant sur
85 sujets conformément à 5.1.1 et 5.2.
b) Critère 2
Pour les PRESSIONS ARTERIELLES SYSTOLIQUE et DIASTOLIQUE de chacun des m sujets, l'écart-type, s , de la
m
moyenne des DETERMINATIONS appariées, pour chaque sujet, réalisées avec le SPHYGMOMANOMETRE EN ESSAI
et des DETERMINATIONS réalisées par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE, doit
remplir les critères du Tableau 1 ou du Tableau 2, le calcul étant effectué selon la Formule (9):
m
s x x (9)
m j n
m 1
j1
où:
x est la valeur moyenne des différences, calculée selon la Formule (6);
n
m est le nombre de sujets;
j est l’indice de l’élément individuel;
x est l'erreur du SPHYGMOMANOMETRE EN ESSAI, calculée selon la Formule (10);
j
d
x p p (10)
j sut ref-sq
j,k
j,k
d
k1
où:
d est le nombre de DETERMINATIONS par sujet;
k est l’indice de l’élément individuel;
p est la PRESSION ARTERIELLE DE REFERENCE calculée selon la Formule (11).
ref-sq
j,k
p p p p p (11)
ref-sq ref ref ref ref
j,k,1 j,k,2 j,k1,1 j,k1,2
j,k
EXEMPLE 4 m = 85 pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents (étude portant sur
85 sujets).
EXEMPLE 5 m = 85 pour un SPHYGMOMANOMETRE destiné à des adultes et/ou des adolescents ainsi qu’à des enfants
âgés de 3 ans à 12 ans (étude portant sur 85 sujets).
EXEMPLE 6 m = 35 pour un SPHYGMOMANOMETRE destiné à un USAGE (supplémentaire) particulier (étude portant sur
35 sujets), lorsque le SPHYGMOMANOMETRE a préalablement fait l'objet d'une INVESTIGATION CLINIQUE lors d'une étude
séparée portant sur 85 sujets conformément à 5.1.1 et 5.2.
EXEMPLE 7 m = 85 pour un SPHYGMOMANOMETRE destiné uniquement à un USAGE particulier (étude portant sur
85 sujets).
5.2.4.3 * Méthode de détermination simultanée sur bras opposés
5.2.4.3.1 Mode opératoire
Le côté du bras de départ des DETERMINATIONS réalisées avec le SPHYGMOMANOMETRE EN ESSAI et le
SPHYGMOMANOMETRE DE REFERENCE doit alterner entre sujets.
La procédure est la suivante:
a) Les observateurs utilisant le SPHYGMOMANOMETRE DE REFERENCE et le SPHYGMOMANOMETRE EN ESSAI
déterminent simultanément la PRESSION ARTERIELLE du sujet sur les bras opposés (voir Figure 3).
b) Ces points de données ne sont pas utilisés pour le calcul de l’exactitude.
c) Effacer de la mémoire du SPHYGMOMANOMETRE EN ESSAI la DETERMINATION précédente et attendre au
moins 60 s.
EXEMPLES La mise hors tension puis sous tension, le retrait du module de détermination de la PRESSION ARTERIELLE
et sa réinstallation ou une commande de réinitialisation sont des méthodes permettant d'effacer la DETERMINATION
précédente de la mémoire.
d) Permuter le SPHYGMOMANOMETRE DE REFERENCE et le SPHYGMOMANOMETRE EN ESSAI entre les deux bras.
e) Attendre au moins 60 s après achèvement de la détermination précédente.
f) Les observateurs utilisant le SPHYGMOMANOMETRE DE REFERENCE et le SPHYGMOMANOMETRE EN ESSAI
déterminent simultanément la PRESSION ARTERIELLE du sujet sur les bras opposés.
g) Répéter les points d) à f) jusqu’à ce que six déterminations appariées aient été effectuées.
h) Si la DETERMINATION réalisée par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE ou le
SPHYGMOMANOMETRE EN ESSAI n'est pas réussie, répéter la DETERMINATION sur le même bras (c'est-à-dire
sans permutation).
i) Répéter les points a) à h) jusqu'à ce que le nombre requis de sujets et de DETERMINATIONS valides aient
fait l'objet d'une INVESTIGATION CLINIQUE.
L’ensemble des données obtenues pour un sujet donné doivent être exclues si
deux DETERMINATIONS quelconques de la PRESSION ARTERIELLE SYSTOLIQUE DE REFERENCE réalisées sur le
même bras présentent une différence de plus de 12 mmHg (1,60 kPa) ou si deux DETERMINATIONS
quelconques de la PRESSION ARTERIELLE DIASTOLIQUE DE REFERENCE réalisées sur le même bras
présentent une différence de plus de 8 mmHg (1,07 kPa);
la différence de latéralité des DETERMINATIONS de la PRESSION ARTERIELLE SYSTOLIQUE DE REFERENCE est
supérieure à 15 mmHg (2,00 kPa) ou si la différence de latéralité des DETERMINATIONS de la PRESSION
ARTERIELLE DIASTOLIQUE DE REFERENCE est supérieure à 10 mmHg (1,33 kPa).
Légende
1 stéthoscope double
2 affichage du SPHYGMOMANOMETRE DE REFERENCE
3 pompe à main du SPHYGMOMANOMETRE DE REFERENCE
4 SPHYGMOMANOMETRE EN ESSAI
NOTE Un seul BRASSARD est raccordé au SPHYGMOMANOMETRE EN ESSAI.
Figure 3 — Illustration de la méthode de détermination simultanée sur bras opposés
La différence de latéralité, LD, est calculée séparément pour les PRESSIONS ARTERIELLES SYSTOLIQUE et
DIASTOLIQUE selon la Formule (12).
3 3
LD p p (12)
ref _ R ref _ L
i j
ij1 1
où:
p et p sont respectivement les PRESSIONS ARTERIELLES DE REFERENCE déterminées sur le
ref _R ref _L
i j
bras droit (R) et sur le bras gauche (L).
5.2.4.3.2 * Analyse des données
L'erreur du SPHYGMOMANOMETRE EN ESSAI, x, est calculée en prenant la différence entre la PRESSION
ARTERIELLE DU SPHYGMOMANOMETRE EN ESSAI et la PRESSION ARTERIELLE DU SPHYGMOMANOMETRE DE
REFERENCE et en ajoutant la différence de latéralité, LD, calculée selon la Formule (13), si la PRESSION
ARTERIELLE DU SPHYGMOMANOMETRE EN ESSAI a été déterminée sur le bras gauche ou en soustrayant la
différence de latéralité, LD, calculée selon la Formule (14), si la PRESSION ARTERIELLE DU SPHYGMOMANOMETRE
EN ESSAI a été déterminée sur le bras droit.
x p p LD (13)
sut _ L ref _ R
x p p LD (14)
sut _ R ref _ L
où p et p sont respectivement les PRESSIONS ARTERIELLES DU SPHYGMOMANOMETRE EN ESSAI
sut _ R sut _ L
déterminées sur le bras droit (R) et sur le bras gauche (L).
Le SPHYGMOMANOMETRE EN ESSAI doit satisfaire aux deux critères suivants:
a) Critère 1
Pour les PRESSIONS ARTERIELLES SYSTOLIQUE et DIASTOLIQUE, la valeur moyenne des différences entre les
DETERMINATIONS, x , des n DETERMINATIONS individuelles appariées réalisées avec le SPHYGMOMANOMETRE EN
n
ESSAI et des DETERMINATIONS réalisées par les observateurs avec le SPHYGMOMANOMETRE DE REFERENCE pour
tous les sujets, doit être inférieure ou égale à ± 5,0 mmHg (± 0,67 kPa), avec un écart-type, s , ne dépassant
n
pas 8,0 mmHg (1,07 kPa), le calcul étant effectué selon la Formule (15) et la Formule (16):
n
x x (15)
n i
n
i1
n
s x x (16)
n i n
n 1
i1
où:
x est l'erreur du SPHYGMOMANOMETRE EN ESSAI, calculée selon la Formule (13) et la Formule (14);
i
n est le nombre de DETERMINATIONS;
i est l’indice de l’élément individuel.
EXEMPLE 1 n = 510 pour un SPHYGMOMANOMETRE destiné à des PATIENTS adultes et/ou adolescents (étude portant sur
85 sujets).
EXEMPLE 2 n = 510 pour un SPHYGMOMANOMETRE destiné à des adultes et/ou des adolescents ainsi qu’à des enfants
âgés de 3 ans à 12 ans (étude portant sur 85 sujets).
EXEMPLE 3 n = 210 pour un SPHYGMOMANOMETRE destiné à un USAGE (supplémentaire) particulier (étude portant sur
35 su
...
МЕЖДУНАРОДНЫЙ ISO
СТАНДАРТ 81060-2
Второе издание
2013-05-01
Неинвазивные сфигмоманометры.
Часть 2.
Клинические исследования моделей с
автоматическим типом измерения
Non-invasive sphygmomanometers —
Part 2: Clinical investigation of automated measurement type
Ответственность за подготовку русской версии несёт GOST R
(Российская Федерация) в соответствии со статьёй 18.1 Устава ISO
Ссылочный номер
©
ISO 2013
Отказ от ответственности при работе в PDF
Настоящий файл PDF может содержать интегрированные шрифты. В соответствии с условиями лицензирования, принятыми
фирмой Adobe, этот файл можно распечатать или смотреть на экране, но его нельзя изменить, пока не будет получена
лицензия на интегрированные шрифты и они не будут установлены на компьютере, на котором ведется редактирование. В
случае загрузки настоящего файла заинтересованные стороны принимают на себя ответственность за соблюдение
лицензионных условий фирмы Adobe. Центральный секретариат ISO не несет никакой ответственности в этом отношении.
Adobe - торговый знак фирмы Adobe Systems Incorporated.
Подробности, относящиеся к программным продуктам, использованные для создания настоящего файла PDF, можно найти в
рубрике General Info файла; параметры создания PDF были оптимизированы для печати. Были приняты во внимание все
меры предосторожности с тем, чтобы обеспечить пригодность настоящего файла для использования комитетами-членами
ISO. В редких случаях возникновения проблемы, связанной со сказанным выше, просьба проинформировать Центральный
секретариат по адресу, приведенному ниже.
ДОКУМЕНТ ЗАЩИЩЕН АВТОРСКИМ ПРАВОМ
Все права сохраняются. Если не указано иное, никакую часть настоящей публикации нельзя копировать или использовать в
какой-либо форме или каким-либо электронным или механическим способом, включая фотокопии и микрофильмы, без
предварительного письменного согласия ISO по адресу ниже или представительства ISO в соответствующей стране.
Бюро авторского права ISO
Почтовый ящик 56 • CH-1211 Женева 20
Тел. + 41 22 749 01 11
Факс + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Опубликовано в Швейцарии
ii © ISO 2013– Все права сохраняются
Содержание Страница
1 Область применения .1
2 Нормативные ссылки .1
3 Термины и определения .2
4 Общие требования к КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ.2
5 КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ с аускультативным ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ .3
6 КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ с ЭТАЛОННЫМ инвазивным ОБОРУДОВАНИЕМ ДЛЯ КОНТРОЛЯ
КРОВЯНОГО ДАВЛЕНИЯ .18
7 * Беременные ПАЦИЕНТЫ (включая ПАЦИЕНТОВ с преэклампсией).23
Приложение A (информативное) Обоснование и руководство .25
Приложение B (нормативное) Целевая частота пульса при испытаниях с физической
нагрузкой.37
Приложение C (информативное) Ссылки на основные принципы .38
Предисловие
Международная организация по стандартизации (ISO) является всемирной федерацией национальных
организаций по стандартизации (комитетов-членов ISO). Разработка международных стандартов
обычно осуществляется техническими комитетами ISO. Каждый комитет-член, заинтересованный в
деятельности, для которой был создан технический комитет, имеет право быть представленным в этом
комитете. Международные правительственные и неправительственные организации, имеющие связи с
ISO, также принимают участие в работах. Что касается стандартизации в области электротехники, то
ISO работает в тесном сотрудничестве с Международной электротехнической комиссией (IEC).
Проекты международных стандартов разрабатываются в соответствии с правилами Директив ISO/IEC,
Часть 2.
Основная задача технических комитетов заключается в подготовке международных стандартов.
Проекты международных стандартов, принятые техническими комитетами, рассылаются комитетам-
членам на голосование. Их опубликование в качестве международных стандартов требует одобрения
не менее 75 % комитетов-членов, принимающих участие в голосовании.
Следует иметь в виду, что некоторые элементы настоящего международного стандарта могут быть
объектом патентных прав. ISO не может нести ответственность за идентификацию какого-либо одного
или всех патентных прав.
Данное второе издание отменяет и заменяет первое издание (ISO 81060-2:20092009), подразделы 5.2.4.3.1 и
6.2.4, которого были технически пересмотрены. Были добавлены многочисленные пояснения, а также в
данный стандарт были добавлены значения в кПа, эквивалентные значениям в мм рт. ст., включая критерий
2 требований 5.2.4.1.2. Оно также включает Техническую поправку ISO 81060-2:2009/Cor 1:2011.
ISO 81060-2 был подготовлен Техническим комитетом ISO/TC 121, Оборудование для анестезии и
искусственной вентиляции легких, Подкомитетом SC 3, Вентиляторы легких и сопутствующее
оборудование, совместно с техническим комитетом IEC/TC 62, Электрооборудование в медицинской
практике, Подкомитетом SC 62D, Электромедицинское оборудование в соответствии с 5 режимом
сотрудничества ISO/IEC.
ISO 81060 состоит из следующих частей под общим заголовком Неинвазивные сфигмоманометры:
⎯ Часть 1. Требования и методы испытания моделей с неавтоматическим типом измерения
⎯ Часть 2. Клинические исследования моделей с автоматическим типом измерения
В данном документе используются следующие шрифты:
⎯ требования, соответствие которым может быть проверено, и определения: прямой шрифт.
⎯ примечание и примеры: уменьшенный шрифт.
⎯ метод испытания: курсив.
⎯ термины, определенные в данном документе: МАЛЫМИ ПРОПИСНЫМИ.
В данном документе текст, обоснование которого приведено в Приложении А, отмечен звездочкой (*).
Обращаем внимание комитетов-участников и национальных комитетов на факт того, что
производители оборудования и испытательные организации могут нуждаться в переходном периоде
после публикации новой, дополненной или пересмотренной, публикации ISO или IEC, в течение
которого они приведут продукты в соответствие с новыми требованиями и оснастятся для проведения
новых или пересмотренных испытаний. ISO/TC 121 и IEC/TC 62 рекомендуют, чтобы содержание
данной части ISO 81060 не адаптировалось для обязательного национального применения раньше,
чем через 3 года после публикации для впервые разрабатываемого оборудования, и не ранее, чем
через 5 лет после публикации для уже производящегося оборудования.
iv © ISO 2013– Все права сохраняются
Введение
Определение КРОВЯНОГО ДАВЛЕНИЯ является важной процедурой, которая клинически используется для
оценки здоровья ПАЦИЕНТА.
Частое определение КРОВЯНОГО ДАВЛЕНИЯ является рутинной процедурой во время анестезии.
КРОВЯНОЕ ДАВЛЕНИЕ является вспомогательным параметром при титрации лекарств и инфузионной
терапии и может предупреждать об условиях, которые могут влиять на заболеваемость и смертность
ПАЦИЕНТОВ.
МЕЖДУНАРОДНЫЙ СТАНДАРТ ISO 81060-2:2013(R)
Неинвазивные сфигмоманометры.
Часть 2.
Клинические исследования моделей с автоматическим
типом измерения
1 Область применения
Данная часть ISO 81060 определяет требования и методы КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ
СМЭ ОБОРУДОВАНИЯ, используемого для периодической неинвазивной автоматической оценки
артериального КРОВЯНОГО ДАВЛЕНИЯ, используя МАНЖЕТУ.
Данная часть ISO 81060 применима ко всем СФИГМОМАНОМЕТРАМ, которые чувствуют или отображают
пульсации, поток или звуки для оценки, отображения или записи КРОВЯНОГО ДАВЛЕНИЕ. Эти
СФИГМОМАНОМЕТРЫ не должны иметь автоматической накачки МАНЖЕТЫ.
Данная часть ISO 81060 покрывает СФИГМОМАНОМЕТРЫ, предназначенные для использования у всех
групп ПАЦИЕНТОВ (например, всех возрастов и весовых категорий), и при всех условиях использования
(например, амбулаторный контроль КРОВЯНОГО ДАВЛЕНИЯ, контроль КРОВЯНОГО ДАВЛЕНИЯ с нагрузкой и
контроль КРОВЯНОГО ДАВЛЕНИЯ при МЕДИЦИНСКОЙ ПОМОЩИ НА ДОМУ и самоконтроль).
ПРИМЕР Автоматические СФИГМОМАНОМЕТРЫ, определенные IEC 80601-2-30, подвергаются КЛИНИЧЕСКИМ
ИССЛЕДОВАНИЯМ согласно данной части ISO 81060.
Данная часть ISO 81060 определяет дополнительные требования к СОПРОВОДИТЕЛЬНОЙ ДОКУМЕНТАЦИИ
СФИГМОМАНОМЕТРОВ, которые подвергались КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в соответствии с данной
частью ISO 81060.
Данная часть ISO 81060 не применима к КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ НЕАВТОМАТИЧЕСКИХ
СФИГМОМАНОМЕТРОВ, определенных в ISO 81060-1 или ИНВАЗИВНОГО ОБОРУДОВАНИЯ ДЛЯ КОНТРОЛЯ
КРОВЯНОГО ДАВЛЕНИЯ, определенного в IEC 60601-2-34.
2 Нормативные ссылки
Ссылка на следующие документы обязательна при использовании данного документа. Для жестких
ссылок применяются только указанное по тексту издание. Для плавающих ссылок необходимо
использовать самое последнее издание нормативного ссылочного документа (включая любые
изменения).
ISO 14155:2011, Клинические исследования медицинских изделий для человека. Приемлемая
клиническая практика
ISO 81060-1:2007, Неинвазивные сфигмоманометры. Часть 1. Требования и методы испытания
неавтоматического типа измерения
IEC 80601-2-30:2009, Медицинское электрооборудование. Часть 2-30. Специальные требования к
основам безопасности и основным рабочим характеристикам автоматических неинвазивных
сфигмоманометров
IEC 60601-1:2005, Медицинское электрооборудование. Часть 1. Общие требования к основам
безопасности и основным рабочим характеристикам
Поправка 1:2012
IEC 60601-1-11, Медицинское электрооборудование. Часть 1-11. . Общие требования к основам
безопасности и основным рабочим характеристикам. Связанный стандарт. Требования к
медицинскому электрооборудованию и медицинским электрическим системам, используемым при
домашнем применении
IEC 60601-2-34:2000, Медицинское электрооборудование. Часть 2-34. . Специальные требования к
безопасности, включая основные рабочие характеристики, к инвазивному оборудованию для
контроля КРОВЯНОГО ДАВЛЕНИЯ
3 Термины и определения
В рамках данного документа используются термины и определения, приведенные в ISO 14155,
IEC 80601-2-30, IEC 60601-1, IEC 60601-1-11, IEC 60601-2-34 и следующие.
ПРИМЕЧАНИЕ Для удобства, начиная со страницы 40, можно найти алфавитный список определенных
терминов.
3.1
ЭТАЛОННЫЙ, прил.
REFERENCE, adj
с установленной точностью, используемый для КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ других инструментов
3.2
СФИГМОМАНОМЕТР
SPHYGMOMANOMETER
МЭ ОБОРУДОВАНИЕ для неинвазивной оценки системного артериального КРОВЯНОГО ДАВЛЕНИЯ
3.3
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР
SPHYGMOMANOMETER-UNDER-TEST
СФИГМОМАНОМЕТР, подвергающийся КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ
4 Общие требования к КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ
4.1 Методы валидации
СФИГМОМАНОМЕТРЫ, отличные от НЕАВТОМАТИЧЕСКИХ СФИГМОМАНОМЕТРОВ должны подвергаться
КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ либо с использованием неинвазивного (аускультативного) ЭТАЛОННОГО
СФИГМОМАНОМЕТРА, либо используя ЭТАЛОННОЕ ИНВАЗИВНОЕ ОБОРУДОВАНИЕ ДЛЯ КОНТРОЛЯ КРОВЯНОГО
ДАВЛЕНИЯ в соответствии с данной частью ISO 81060 в каждом режиме работы.
ПРИМЕР 1 Режим взрослых и новорожденных.
ПРИМЕР 2 Режим с быстрой и медленной скоростью спуска МАНЖЕТЫ.
КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ следует считать ТИПОВЫМ ИСПЫТАНИЕМ.
Считается, что соответствие требованиям данного подраздела достигнуто, если выполнены
критерии соответствующих инспекций и испытаний в данной части ISO 81060.
2 © ISO 2013– Все права сохраняются
4.2 Приемлемая клиническая практика
Все КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ должны соответствовать требованиям ISO 14155. КЛИНИЧЕСКИЕ
ИССЛЕДОВАНИЯ с ЭТАЛОННЫМ ИНВАЗИВНЫМ ОБОРУДОВАНИЕМ ДЛЯ КОНТРОЛЯ КРОВЯНОГО ДАВЛЕНИЯ не должны
проводиться на ПАЦИЕНТАХ или лицах только в целях исследования рабочих характеристик
СФИГМОМАНОМЕТРА.
ПРИМЕЧАНИЕ Некоторые компетентные органы имеют дополнительные требования.
Проверьте соответствие применением требований ISO 14155.
5 КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ с аускультативным ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ
5.1 Требования к субъекту
5.1.1 * Количество
КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ с аускультативным ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ должны состоять из,
как минимум, 85 субъектов. Если не определено иное, для каждого субъекта должно быть проведено,
по крайней мере, три валидных ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ. Должно быть, как минимум, 255
валидных парных определения КРОВЯНОГО ДАВЛЕНИЯ.
Проверьте соответствие инспекцией ОТЧЕТА О КЛИНИЧЕСКИХ ИССЛЕДОВАНИЯХ.
5.1.2 * Распределение по полам
По крайней мере, 30 % субъектов должно быть мужчинами и, по крайней мере, 30 % субъектов должно
быть женщинами.
Проверьте соответствие инспекцией ОТЧЕТА О КЛИНИЧЕСКИХ ИССЛЕДОВАНИЯХ.
5.1.3 * Распределение по возрастам
Для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или ПАЦИЕНТОВ
подросткового возраста, возраст субъектов, включенных в КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ должен быть
больше 12 лет.
ПРИМЕЧАНИЕ 1 Всего минимум 85 субъектов.
Для СФИГМОМАНОМЕТРОВ, дополнительно предназначенных для использования у детей в КЛИНИЧЕСКИЕ
ИССЛЕДОВАНИЯ должны быть включены 35 субъектов-детей в возрасте между 3 годами и 12 годами.
ПРИМЕЧАНИЕ 2 Всего минимум 85 субъектов.
Если СФИГМОМАНОМЕТР имеет специальный режим для детей, в этом режиме дети считаются
специальной выборкой ПАЦИЕНТОВ (см. 5.1.6). В этих исследованиях дети освобождаются от
требований 5.1.5 к распределению КРОВЯНОГО ДАВЛЕНИЯ.
Детей возрастом менее 3 лет не следует включать в КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ, использующим
аускультативное ОПРЕДЕЛЕНИЕ за счет снятия показаний с ЭТАЛОННОГО СФИГМОМАНОМЕТРА.
Проверьте соответствие инспекцией СОПРОВОДИТЕЛЬНОЙ ДОКУМЕНТАЦИИ и ОТЧЕТА О КЛИНИЧЕСКИХ
ИССЛЕДОВАНИЯХ.
5.1.4 * Распределение размеров конечностей
Для СФИГМОМАНОМЕТРОВ, предназначенных для использования с МАНЖЕТОЙ одного размера:
⎯ по крайней мере, 40 % субъектов должны иметь обхват руки, который лежит в пределах верхней
половины определенного диапазона использования МАНЖЕТЫ и, по крайней мере, 40 % должны
иметь обхват руки, который лежит в пределах нижней половины, и
⎯ по крайней мере, 20 % субъектов должны иметь обхват руки, который лежит в пределах верхней
четверти определенного диапазона использования МАНЖЕТЫ и, по крайней мере, 20 % должны
иметь обхват руки, который лежит в пределах нижней четверти.
Для СФИГМОМАНОМЕТРОВ, предназначенных для использования с несколькими размерами МАНЖЕТЫ,
каждый размер МАНЖЕТЫ должен испытываться на, по крайней мере, от всех субъектов, где
2× n
n - число размеров МАНЖЕТЫ.
Проверьте соответствие инспекцией СОПРОВОДИТЕЛЬНОЙ ДОКУМЕНТАЦИИ и ОТЧЕТА О КЛИНИЧЕСКИХ
ИССЛЕДОВАНИЯХ.
5.1.5 * Распределение КРОВЯНОГО ДАВЛЕНИЯ
По крайней мере, 5 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ u 100 мм рт. ст. (13, 33 кПа).
По крайней мере, 5 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ W 160 мм рт. ст. (21, 33 кПа).
По крайней мере, 20 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ W 140 мм рт. ст. (18,66 кПа).
По крайней мере, 5 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ u 60 мм рт. ст. (8 кПа).
По крайней мере, 5 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ W 100 мм рт. ст. (13, 33 кПа).
По крайней мере, 20 % показаний ЭТАЛОННОГО КРОВЯНОГО ДАВЛЕНИЯ должны иметь СИСТОЛИЧЕСКОЕ
КРОВЯНОЕ ДАВЛЕНИЕ W 85 мм рт. ст. (11, 33 кПа).
Проверьте соответствие инспекцией ОТЧЕТА О КЛИНИЧЕСКИХ ИССЛЕДОВАНИЯХ.
5.1.6 * Специальные выборки ПАЦИЕНТОВ
СФИГМОМАНОМЕТРЫ, предназначенные для использования в специальных выборках ПАЦИЕНТОВ, если
существуют ОБЪЕКТИВНЫЕ ДОКАЗАТЕЛЬСТВА, что точность СФИГМОМАНОМЕТРА может быть
проблематичной в этих выборках ПАЦИЕНТОВ, должны подвергаться КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в
этих выборках ПАЦИЕНТОВ. См. Также Раздел 0.
ПРИМЕЧАНИЕ В Разделе 7 определены примеры специальных популяций ПАЦИЕНТОВ со специальными
требованиями.
Если СФИГМОМАНОМЕТРЫ подвергаются КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в соответствии с требованиями
5.1.1 и 5.2, они должны подвергаться КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ на, по крайней мере, 35
дополнительных субъектах специальной выборки. Если СФИГМОМАНОМЕТРЫ не подвергаются
предварительным КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в соответствии с требованиями 5.1.1 и 5.2,
4 © ISO 2013– Все права сохраняются
КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ в соответствии с требованиями 5.1.1 и 5.2 должны состоять только из
субъектов из специальной выборки ПАЦИЕНТОВ.
Специальная выборка ПАЦИЕНТОВ должна быть определена в ясных терминах и рассматривать
следующие параметры: пол (см. 5.1.2), возраст (см. 5.1.3), размер руки (см. 5.1.4) и КРОВЯНОЕ ДАВЛЕНИЕ
(см. 5.1.5). Сводка по данной информации должна быть приведена в инструкции по эксплуатации.
Проверьте соответствие инспекцией СОПРОВОДИТЕЛЬНОЙ ДОКУМЕНТАЦИИ и ОТЧЕТА О КЛИНИЧЕСКИХ
ИССЛЕДОВАНИЯХ.
5.2 Метод КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ с ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ
5.2.1 * Подготовка субъекта
Если в инструкции по эксплуатации ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА не указано иное, расположите
субъект так, чтобы субъекту:
⎯ было комфортно;
ПРИМЕР Комфортно усадите так, чтобы ноги не перекрещивались, а ступни были на полу.
⎯ имел спинку, подлокотники и поддержку предплечья;
⎯ середина МАНЖЕТЫ была на уровне правого предсердья сердца.
Рекомендуется, чтобы субъект максимально расслабился и не разговаривал во время всей процедуры.
До считывания первых показаний должно пройти 5 мин.
ПРИМЕЧАНИЕ Дополнительные подробности можно найти по Ссылке [32].
5.2.2 * Подготовка наблюдателя
Наблюдатели должны быть обучены использованию соответствующих методик для проведения
ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ покоя, используя приемлемые клинические протоколы измерения
КРОВЯНОГО ДАВЛЕНИЯ. См. Ссылки [8], [28], [29], [32] и [45], содержащие дополнительную информацию.
Наблюдатели должны иметь достаточный опыт проведения ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ.
Все наблюдатели, записывающие показания ЭТАЛОННОГО СФИГМОМАНОМЕТРА не должны быть видимы
другим наблюдателям. Показания ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА не должны быть видимы любому
из этих наблюдателей.
ПРИМЕР 1 Использование третьего наблюдателя для записи показаний ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА.
ПРИМЕР 2 Использование электронных средств записи показаний ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА.
Проинструктируйте наблюдателей, чтобы они определяли ДИАСТОЛИЧЕСКОЕ КРОВЯНОЕ ДАВЛЕНИЕ как
последний слышимый тон Короткова (пятая фаза или K5), за исключением случаев, когда тоны
Короткова остаются слышимыми при спущенной МАНЖЕТЕ, и детей возрастом между 3 годами и
12 годами, у которых используется четвертая фаза (K4). Если у детей не прослушивается К4, либо
используется К5, либо субъект исключается.
ПРИМЕЧАНИЕ Кроме детей, К4 должен также использоваться у субъектов, у которых большая разница между
приглушением и исчезанием тонов (при этом второе порой стремиться к нулю мм рт. ст.).
Проинструктируйте наблюдателей, чтобы они записывали, какой тон Короткова использовался для
ОПРЕДЕЛЕНИЯ ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ.
Тоны Короткова, используемые для ОПРЕДЕЛЕНИЯ ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ в КЛИНИЧЕСКИХ
ИССЛЕДОВАНИЯХ, должны быть указаны в инструкции по эксплуатации СФИГМОМАНОМЕТРА.
ПРИМЕР 3 K5 использовался у 65 субъектов и K4 использовался у 20 субъектов.
5.2.3 * ЭТАЛОННОЕ ОПРЕДЕЛЕНИЕ
Два наблюдателя должны проводить одновременное ОПРЕДЕЛЕНИЕ КРОВЯНОГО ДАВЛЕНИЯ на каждом
субъекте, используя двойной стетоскоп.
Если испытываемый СФИГМОМАНОМЕТР не предназначен для использования при существенно
нерегулярном ритме сердце и если любой из наблюдателей определяет существенно нерегулярный
ритм сердца, это ОПРЕДЕЛЕНИЕ должно быть исключено.
ПРИМЕРЫ Бигеминия, тригеминия, изолированный VPB, мерцательная аритмия.
ПРИМЕЧАНИЕ 1 Хотя КЛИНИЧЕСКОЕ ИССЛЕДОВАНИЕ КРОВЯНОГО ДАВЛЕНИЯ у ПАЦИЕНТОВ с мерцательной аритмией
клинически важно, в настоящее время отсутствует общепринятое руководство по определению КРОВЯНОГО
ДАВЛЕНИЯ у таких лиц.
Любая пара показаний наблюдателей с разницей более 4 мм рт. ст. (0,53 кПа) должна быть исключена.
Отдельные значения наблюдателей для каждого определения должны быть приведены к среднему в
соответствии с Формулой (1) для получения ЭТАЛОННОГО ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ.
p + p
этал этал
i ,1 i ,2
p = (1)
этал
i
где
p КРОВЯНОЕ ДАВЛЕНИЕ, определенное наблюдателем 1 для i-ого ОПРЕДЕЛЕНИЯ;
этал
i ,1
p КРОВЯНОЕ ДАВЛЕНИЕ, определенное наблюдателем 2 для i-ого ОПРЕДЕЛЕНИЯ;
этал
i ,2
p ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ для i-ого ОПРЕДЕЛЕНИЯ.
этал
i
Погрешности между наблюдателями должны быть просмотрены после завершения ряда парных
ОПРЕДЕЛЕНИЙ для ЭТАЛОННОГО ОПРЕДЕЛЕНИЯ и испытываемого прибора. Если какие-то ОПРЕДЕЛЕНИЯ
исключены, должна быть получена дополнительная пара (пары) ОПРЕДЕЛЕНИЙ, чтобы гарантировать,
что доступно необходимое число валидных пар для ЭТАЛОННОГО ОПРЕДЕЛЕНИЯ и испытываемого
прибора. Должно быть проведено максимум восемь пар ОПРЕДЕЛЕНИЙ.
Используйте ЭТАЛОННЫЙ СФИГМОМАНОМЕТР, соответствующий требованиям ISO 81060-1, за
исключением того, что допустимая ошибка должна быть ± 1 мм рт. ст. (0,13 кПа). Считывание значений
с ЭТАЛОННОГО СФИГМОМАНОМЕТРА должно быть настолько точным, насколько это возможно. При
считывании значений с ЭТАЛОННОГО СФИГМОМАНОМЕТРА, наблюдатели должны избегать ошибок
вследствие параллакса и округления.
ПРИМЕЧАНИЕ 2 Округление оказывает негативное влияние на результаты КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ.
ПРИМЕЧАНИЕ 3 В рамках данной части ISO 81060 МАНЖЕТА считается частью ЭТАЛОННОГО СФИГМОМАНОМЕТРА. Не
может использоваться МАНЖЕТА, не соответствующая ISO 81060-1.
6 © ISO 2013– Все права сохраняются
5.2.4 Методы КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ
5.2.4.1 Метод одновременного ОПРЕДЕЛЕНИЯ на одной руке
5.2.4.1.1 * Процедура
Этот метод должен использоваться только с ИСПЫТЫВАЕМЫМ СФИГМОМАНОМЕТРОМ:
⎯ который имеет МАНЖЕТУ, соответствующую ISO 81060-1;
⎯ который разработан для использования на предплечье; и
⎯ у которого:
⎯ постоянная линейная скорость спуска от 2 мм рт. ст./с (0,27 кПа/с) до 3 мм рт. ст./с (0,40 кПа/с)
или
⎯ для испытываемого СФИГМОМАНОМЕТРА, который контролирует спуск как функцию пульса,
скорость спуска от 2 мм рт. ст./удар (0,27 кПа/удар) до 3 мм рт. ст./удар (0,40 кПа/удар).
Может использоваться любая рука.
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР не должен спускать до ОПРЕДЕЛЕНИЯ ЭТАЛОННОГО ДИАСТОЛИЧЕСКОГО
КРОВЯНОГО ДАВЛЕНИЯ. ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР может быть изменен, чтобы соответствовать
данному критерию.
ПРИМЕЧАНИЕ Корректное одновременное определение на одной руке требует, чтобы испытываемый
СФИГМОМАНОМЕТР надувал МАНЖЕТУ до уровня, по крайней мере, на 20 мм рт. ст. (2,67 кПа) выше, чем реальное
СИСТОЛИЧЕСКОЕ КРОВЯНОЕ ДАВЛЕНИЕ, определенное ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ, и, по крайней мере, на
20 мм рт. ст. (2,67 кПа) ниже, чем реальное ДИАСТОЛИЧЕСКОЕ КРОВЯНОЕ ДАВЛЕНИЕ, определенное ЭТАЛОННЫМ
СФИГМОМАНОМЕТРОМ.
Выполните следующее:
a) Проведите наблюдение, используя одновременное ОПРЕДЕЛЕНИЕ КРОВЯНОГО ДАВЛЕНИЯ субъекта,
используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР и ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР, используя одну и ту
же МАНЖЕТУ и один и тот же цикл накачки/спуска (см. Рисунок 1). Эти данные не используются
для вычисления точности ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА.
b) Очистите память ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА от предыдущего ОПРЕДЕЛЕНИЯ и затем
выждите, по крайней мере, 60 с.
ПРИМЕРЫ Методы очистки памяти от предыдущего ОПРЕДЕЛЕНИЯ: выключение и включение питания,
удаление модуля КРОВЯНОГО ДАВЛЕНИЯ и повторная его установка или запуск команды перезагрузки.
c) Проведите наблюдение, используя одновременное ОПРЕДЕЛЕНИЕ КРОВЯНОГО ДАВЛЕНИЯ субъекта,
используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР и ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР, используя одну и ту
же МАНЖЕТУ и один и тот же цикл накачки/спуска.
d) Выждите, по крайней мере, 60 с между ОПРЕДЕЛЕНИЯМИ.
e) Повторите с) и d) пока не будет проведено необходимое количество ОПРЕДЕЛЕНИЙ.
Все данные от субъекта должны быть исключены, если любые два ЭТАЛОННЫХ ОПРЕДЕЛЕНИЯ
СИСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ отличаются более чем на 12 мм рт. ст. (1,60 кПа) или любые два
эталонных ОПРЕДЕЛЕНИЯ ДИАСТОЛИЧЕСКОГО кровяного давление отличаются более чем на 8 мм рт. ст.
(1,07 кПа).
Обозначение
1 двойной стетоскоп
2 дисплей ЭТАЛОННОГО СФИГМОМАНОМЕТРА
3 ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР
Рисунок 1 — Иллюстрация метода одновременного ОПРЕДЕЛЕНИЯ на одной и той же руке
5.2.4.1.2 * Анализ данных
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР должен соответствовать следующим двум критериям.
a) Критерий 1
Для СИСТОЛИЧЕСКОГО и ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ средняя погрешность ОПРЕДЕЛЕНИЯ, x ,
n
для n отдельных парных ОПРЕДЕЛЕНИЙ ИСПЫТЫВАЕМЫМ СФИГМОМАНОМЕТРОМ и ЭТАЛОННЫМ
СФИГМОМАНОМЕТРОМ для всех субъектов должна быть в пределах или равна ± 5,0 мм рт. ст. (± 0,67 кПа),
со средним квадратным отклонением, s , не более 8,0 мм рт. ст. (1,07 кПа) при вычислении в
n
соответствии с Формулой (2) и Формулой (3):
n
x = ×()p − p (2)
n ∑ исп этал
i i
n
i=1
n
s=× (3)
xx−
n ∑()n
i
n−1
i=1
где
x
n
средняя погрешность;
p − p = x разница между парным ОПРЕДЕЛЕНИЕМ КРОВЯНОГО ДАВЛЕНИЯ [ОПРЕДЕЛЕНИЕ на
исп этал i
i i
ИСПЫТЫВАЕМОМ СФИГМОМАНОМТРЕ − ЭТАЛОННЫОЕ КРОВЯНОЕ ДАВЛЕНИЕ, вычисленное в
соответствии с Формулой (1)];
i индекс отдельного элемента;
n число ОПРЕДЕЛЕНИЙ.
8 © ISO 2013– Все права сохраняются
и s должны быть вычислены и выражены с точностью до 0,1 мм рт. ст. (0,01 кПа)
x
n
n
ПРИМЕР 1 n = 255 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
ПРИМЕР 2 n = 255 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 3 n = 105 для СФИГМОМАНОМЕТРОВ, предназначенных для специального ПРЕДПОЛАГАЕМОГО
ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно подвергался КЛИНИЧЕСКИМ
ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
b) Критерий 2
Для СИСТОЛИЧЕСКОГО и ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ для каждого из m субъектов среднее
квадратичное отклонение, s , усредненных парных ОПРЕДЕЛЕНИЙ для субъекта на ИСПЫТЫВАЕМОМ
m
СФИГМОМАНОМЕТРЕ и ЭТАЛОННОМ СФИГМОМАНОМЕТРЕ должно удовлетворять критериям, перечисленным
в Таблице 1 или Таблице 2 при вычислении в соответствии с Формулой (4):
m
s=× (4)
xx−
m ∑ n
()
j
m−1
j=1
где
среднее отклонение, вычисленное в соответствии с Формулой (2);
x
n
m число субъектов;
j индекс отдельного элемента;
x вычисленное в соответствии с Формулой (5).
j
d
x = ×()p − p (5)
∑
j исп этал
ki ki
d
k=1
где
d число ОПРЕДЕЛЕНИЙ на субъект;
k индекс отдельного элемента.
ПРИМЕР 4 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
ПРИМЕР 5 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 6 m = 35 для СФИГМОМАНОМЕТРОВ, предназначенных для (дополнительного) специального
ПРЕДПОЛАГАЕМОГО ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно
подвергался КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
ПРИМЕР 7 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных только для специального предполагаемого
использование (исследование на 85 субъектах).
Таблица 1 — Приемлемость усредненных данных субъекта (критерий 2) в мм. рт. ст.
Максимально допустимое среднее квадратичное отклонение, s , как функция средней ошибки,
x
m
n
мм рт. ст.
x
n
0,0 0,1 0,2 0,3 0,4 0,5 0,6 0,7 0,8 0,9
± 0, 6,95 6,95 6,95 6,95 6,93 6,92 6,91 6,90 6,89 6,88
± 1, 6,87 6,86 6,84 6,82 6,80 6,78 6,76 6,73 6,71 6,68
± 2, 6,65 6,62 6,58 6,55 6,51 6,47 6,43 6,39 6,34 6,30
± 3, 6,25 6,20 6,14 6,09 6,03 5,97 5,89 5,83 5,77 5,70
± 4, 5,64 5,56 5,49 5,41 5,33 5,25 5,16 5,08 5,01 4,90
± 5, 4,79 — — — — — — — — —
ПРИМЕР Для средней ошибки ± 4,2 мм рт. ст. максимально допустимое среднее квадратичное отклонение 5,49 мм рт. ст.
Таблица 2 — Приемлемость усредненных данных субъекта (критерий 2) в кПа
Максимально допустимое среднее квадратичное отклонение, s , как функция средней ошибки,
x
m
n
мм рт. ст.
x
n
0,000 0,010 0,020 0,030 0,040 0,050 0,060 0,070 0,080 0,090
± 0,0 0,9266 0,9266 0,9266 0,9266 0,9266 0,9246 0,9233 0,9223 0,9213 0,9203
± 0,1 0,9193 0,9183 0,9173 0,9163 0,9152 0,9138 0,9119 0,9099 0,9079 0,9059
± 0,2 0,9039 0,9007 0,8989 0,8970 0,8946 0,8906 0,8878 0,8855 0,8826 0,8785
± 0,3 0,8756 0,8723 0,8679 0,8641 0,8601 0,8562 0,8519 0,8471 0,8414 0,8374
± 0,4 0,8333 0,8283 0,8226 0,8169 0,8119 0,8059 0,7999 0,7933 0,7853 0,7793
± 0,5 0,7739 0,7669 0,7599 0,7531 0,7463 0,7388 0,7319 0,7237 0,7157 0,7077
ПРИМЕР Для средней ошибки ± 0,520 кПа. максимально допустимое среднее квадратичное отклонение 0,7599 кПа.
5.2.4.2 * Метод последовательного ОПРЕДЕЛЕНИЯ на одной руке
5.2.4.2.1 *Процедура
Может использоваться любая рука.
Выполните следующее:
a) Выполните наблюдение, используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР для ОПРЕДЕЛЕНИЯ КРОВЯНОГО
ДАВЛЕНИЯ субъекта (см. Рисунок 2).
b) Выждите, по крайней мере, 60 с.
c) Выполните наблюдение, используя ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР для ОПРЕДЕЛЕНИЯ КРОВЯНОГО
ДАВЛЕНИЯ субъекта.
d) Очистите память ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА от предыдущих ОПРЕДЕЛЕНИЙ и затем
выждите, по крайней мере, 60 с.
ПРИМЕРЫ Методы очистки памяти от предыдущего ОПРЕДЕЛЕНИЯ: выключение и включение питания,
удаление модуля КРОВЯНОГО ДАВЛЕНИЯ и повторная его установка или запуск команды перезагрузки.
10 © ISO 2013– Все права сохраняются
Обозначение
1 двойной стетоскоп
2 дисплей ЭТАЛОННОГО СФИГМОМАНОМЕТРА
3 ручной насос ЭТАЛОННОГО СФИГМОМАНОМЕТРА
4 ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР
ПРИМЕЧАНИЕ К ИСПЫТЫВАЕМОМУ СФИГМОМАНОМЕТРУ подсоединена только одна МАНЖЕТА.
Рисунок 2 — Иллюстрация метода последовательного ОПРЕДЕЛЕНИЯ на одной и той же руке
e) Данные, полученные в а) и с) не должны использоваться при ОПРЕДЕЛЕНИИ точности.
f) Используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР, выполните определение КРОВЯНОГО ДАВЛЕНИЯ субъекта.
g) Выполните наблюдения, последовательно используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР и
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР для ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ субъектов.
h) Выжидайте, по крайней мере, 60 с между каждыми ОПРЕДЕЛЕНИЯМИ.
i) Повторяйте g) и h) пока не будет проведено необходимое число валидных ОПРЕДЕЛЕНИЙ.
Все данные от субъекта должны быть исключены, если любые два ЭТАЛОННЫХ ОПРЕДЕЛЕНИЯ
СИСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ отличаются более чем на 12 мм рт. ст. (1,60 кПа) или любые два
ЭТАЛОННЫХ ОПРЕДЕЛЕНИЯ ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ отличаются более чем на 8 мм рт. ст.
(1,07 кПа). Несмотря на данное требование, если ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ отдельного субъекта
не удовлетворяет данному критерию во время испытания, могу использоваться две соответствующие
последовательные пары ОПРЕДЕЛЕНИЙ, удовлетворяющие критерию. В данном случае, должны
использоваться дополнительные субъекты для получения минимального числа ОПРЕДЕЛЕНИЙ. Не более
10 % субъектов должны иметь менее трех валидных пар ОПРЕДЕЛЕНИЙ.
5.2.4.2.2 Анализ данных
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР должен соответствовать следующим двум критериям.
a) Критерий 1
Для СИСТОЛИЧЕСКОГО и ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ средняя погрешность ОПРЕДЕЛЕНИЯ, x ,
n
для n отдельных парных ОПРЕДЕЛЕНИЙ ИСПЫТЫВАЕМЫМ СФИГМОМАНОМЕТРОМ и ЭТАЛОННЫМ
СФИГМОМАНОМЕТРОМ для всех субъектов должна быть в пределах или равна ± 5,0 мм рт. ст. (± 0,67 кПа),
со средним квадратным отклонением, s , не более 8,0 мм рт. ст. (1,07 кПа) при вычислении в
n
соответствии с Формулой (6) и Формулой (7):
n
x = ×()p − p (6)
n ∑ исп этал−ср
i i
n
i=1
n
s=× (7)
xx−
n ∑()n
i
n−1
i=1
где
x
n
средняя погрешность;
p − p = x разница между i-ыми парными ОПРЕДЕЛЕНИЕМ КРОВЯНОГО ДАВЛЕНИЯМИ
исп этал i
i i
[ОПРЕДЕЛЕНИЕ на ИСПЫТЫВАЕМОМ СФИГМОМАНОМТРЕ − ЭТАЛОННЫОЕ КРОВЯНОЕ ДАВЛЕНИЕ];
i индекс отдельного элемента;
n число ОПРЕДЕЛЕНИЙ;
p ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ i-ого ОПРЕДЕЛЕНИЯ, вычисленное в соответствии с
этал−ср
i
Формулой (8);
и s должны быть вычислены и выражены с точностью до 0,1 мм рт. ст. (0,01 кПа)
x
n
n
ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ p (определенное наблюдателем с помощью ЭТАЛОННОГО
этал−ср
i
СФИГМОМАНОМЕТРА) должно быть средним предшествующего и последующего ЭТАЛОННОГО КРОВЯНОГО
ДАВЛЕНИЯ).
(8)
p = × ( p + p + p + p )
этал−ср этал этал этал этал
i i ,1 i, 2 +1i ,1 +1i, 2
где
p КРОВЯНОЕ ДАВЛЕНИЕ, определенное наблюдателем 1 для i-ого ОПРЕДЕЛЕНИЯ;
этал
i ,1
p КРОВЯНОЕ ДАВЛЕНИЕ, определенное наблюдателем 2 для i-ого ОПРЕДЕЛЕНИЯ.
этал2
ПРИМЕР 1 n = 255 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
ПРИМЕР 2 n = 255 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 3 n = 105 для СФИГМОМАНОМЕТРОВ, предназначенных для специального ПРЕДПОЛАГАЕМОГО
ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно подвергался КЛИНИЧЕСКИМ
ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
b) Критерий 2
Для СИСТОЛИЧЕСКОГО и ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ для каждого из m субъектов среднее
квадратичное отклонение, s , усредненных парных ОПРЕДЕЛЕНИЙ для субъекта на ИСПЫТЫВАЕМОМ
m
12 © ISO 2013– Все права сохраняются
СФИГМОМАНОМЕТРЕ и ЭТАЛОННОМ СФИГМОМАНОМЕТРЕ должно удовлетворять критериям, перечисленным
в Таблице 1 или Таблице 2 при вычислении в соответствии с Формулой (9):
m
s=× (9)
xx−
m ∑ n
()
j
m−1
j=1
где
среднее отклонение, вычисленное в соответствии с Формулой (6);
x
n
m число субъектов;
j индекс отдельного элемента;
x ошибка ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА, вычисленная в соответствии с Формулой (10).
j
d
x = ×()p − p (10)
∑
j исп этал−ср
ki ki
d
k=1
где
d число ОПРЕДЕЛЕНИЙ на субъект;
k индекс отдельного элемента;
ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ i-ого ОПРЕДЕЛЕНИЯ, вычисленное в соответствии с
p
этал−ср
i
Формулой (11);
p = × ( p + p + p + p ) (11)
этал−ср этал этал этал этал
i i ,1 i, 2 +1i ,1 +1i, 2
ПРИМЕР 4 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
ПРИМЕР 5 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 6 m = 35 для СФИГМОМАНОМЕТРОВ, предназначенных для (дополнительного) специального
ПРЕДПОЛАГАЕМОГО ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно
подвергался КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
ПРИМЕР 7 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных только для специального предполагаемого
использование (исследование на 85 субъектах).
5.2.4.3 Метод одновременного ОПРЕДЕЛЕНИЯ на противоположных руках
5.2.4.3.1 * Процедура
Исходная рука при определении ИСПЫТЫВАЕМЫМ СФИГМОМАНОМЕТРОМ и ЭТАЛОННЫМ СФИГМОМАНОМЕТРОМ
должна отличаться у разных субъектов.
Выполните следующее:
a) Выполните наблюдение, используя одновременно ЭТАЛОННЫЙ СФИГМОМАНОМЕТР и ИСПЫТЫВАЕМЫЙ
СФИГМОМАНОМЕТР для ОПРЕДЕЛЕНИЯ КРОВЯНОГО ДАВЛЕНИЯ субъекта на противоположных руках
(см. Рисунок 3).
b) Эти данные не должны использоваться для вычисления точности.
c) Очистите память ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА от предыдущих ОПРЕДЕЛЕНИЙ и затем
выждите, по крайней мере, 60 с.
ПРИМЕРЫ Методы очистки памяти от предыдущего ОПРЕДЕЛЕНИЯ: выключение и включение питания,
удаление модуля КРОВЯНОГО ДАВЛЕНИЯ и повторная его установка или запуск команды перезагрузки.
d) Поменяйте руки для ЭТАЛОННОГО СФИГМОМАНОМЕТРА и ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА.
e) Выждите, по крайней мере, 60 с со времени завершения предыдущего ОПРЕДЕЛЕНИЯ.
f) Проведите наблюдение, используя одновременное ОПРЕДЕЛЕНИЕ КРОВЯНОГО ДАВЛЕНИЯ субъекта,
используя ЭТАЛОННЫЙ СФИГМОМАНОМЕТР и ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР на разных руках.
g) Повторите d) – f) пока не будет выполнено шесть парных ОПРЕДЕЛЕНИЙ.
h) Если определение наблюдателем с использованием ЭТАЛОННОГО СФИГМОМАНОМЕТРА или
ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА не выполнено успешно, повторите ОПРЕДЕЛЕНИЕ на той же
руке (т.е. без изменения руки).
i) Повторите a) – h) пока необходимое число валидных субъектов и ОПРЕДЕЛЕНИЙ не будет
подвергнуто КЛИНИЧСЕКИМ ИССЛЕДОВАНИЯМ.
Обозначение
1 двойной стетоскоп
2 дисплей ЭТАЛОННОГО СФИГМОМАНОМЕТРА
3 ручной насос ЭТАЛОННОГО СФИГМОМАНОМЕТРА
4 ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР
ПРИМЕЧАНИЕ К ИСПЫТЫВАЕМОМУ СФИГМОМАНОМЕТРУ подсоединена только одна МАНЖЕТА.
Рисунок 3 — Иллюстрация метода одновременного ОПРЕДЕЛЕНИЯ на противоположных руках
14 © ISO 2013– Все права сохраняются
Все данные от субъекта должны быть исключены, если
⎯ любые два ЭТАЛОННЫХ ОПРЕДЕЛЕНИЯ СИСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ на одной и той же руке
отличаются более чем на 12 мм рт. ст. (1,60 кПа) или любые два эталонных ОПРЕДЕЛЕНИЯ
ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ на одной и той же руке отличаются более чем на
8 мм рт. ст. (1,07 кПа).
⎯ Боковое отклонение ЭТАЛОННОГО ОПРЕДЕЛЕНИЯ СИСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ более
15 мм рт. ст. (2,00 кПа) или боковое отклонение ЭТАЛОННОГО ОПРЕДЕЛЕНИЯ ДИАСТОЛИЧЕСКОГО
КРОВЯНОГО ДАВЛЕНИЯ более 10 мм рт. ст. (1,33 кПа).
Боковое отклонение, (lateral difference, LD), вычисляется отдельно для СИСТОЛИЧЕСКОГО и
ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ, в соответствии с Формулой (12).
3 3
⎛ ⎞
⎜ ⎟
LD= P − P (12)
∑ этал _Пi ∑ этал _Лj
⎜ ⎟
i=1 j=1
⎝ ⎠
где
P и P – это ЭТАЛОННОЕ КРОВЯНОЕ ДАВЛЕНИЕ в правой (П) и левой (Л) руке
'этал_Пi этал_Лj
соответственно.
5.2.4.3.2 * Анализ данных
Ошибка ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА, x, вычисляется, взяв разницу между КРОВЯНЫМ ДАВЛЕНИЕМ
ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА и КРОВЯНЫМ ДАВЛЕНИЕМ ЭТАЛОННОГО СФИГМОМАНОМЕТРА и добавив
боковое отклонение, LD, вычисленное в соответствии с Формулой (13), если КРОВЯНОЕ ДАВЛЕНИЕ
ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА было получено на левой руке и вычтя боковое отклонение, LD,
вычисленное в соответствии с Формулой (13), если КРОВЯНОЕ ДАВЛЕНИЕ ИСПЫТЫВАЕМОГО
СФИГМОМАНОМЕТРА было получено на правой руке.
x = P − P + LD (13)
исп_Л этал_П
x = P − P − LD (14)
исп_П этал_Л
где P и P – КРОВЯНОЕ ДАВЛЕНИЕ ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА на правой (П) руке и левой
исп_П исп_Л
(Л) руке, соответственно.
ИСПЫТЫВАЕМЫЙ СФИГМОМАНОМЕТР должен удовлетворять следующим двум критериям.
a) Критерий 1
Для СИСТОЛИЧЕСКОГО и ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ средняя погрешность ОПРЕДЕЛЕНИЯ, x ,
n
для n отдельных парных ОПРЕДЕЛЕНИЙ ИСПЫТЫВАЕМЫМ СФИГМОМАНОМЕТРОМ и ЭТАЛОННЫМ
СФИГМОМАНОМЕТРОМ для всех субъектов должна быть в пределах или равна ± 5,0 мм рт. ст. (± 0,67 кПа),
со средним квадратным отклонением, s , не более 8,0 мм рт. ст. (1,07 кПа) при вычислении в
n
соответствии с Формулой (15) и Формулой (16):
n
x=× x (15)
i
n ∑
n
i=1
n
s=× (16)
xx−
n ∑()n
i
n−1
i=1
где
x ошибка ИСПЫТЫВАЕМОГО СФИГМОМАНОМЕТРА, вычисленная в соответствии с Формулой (13) и
i
Формулой 14.;
n число определений;
i индекс отдельного элемента.
ПРИМЕР 1 n = 510 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
ПРИМЕР 2 n = 510 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 3 n = 210 для СФИГМОМАНОМЕТРОВ, предназначенных для специального ПРЕДПОЛАГАЕМОГО
ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно подвергался КЛИНИЧЕСКИМ
ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
b) Критерий 2
Для среднего СИСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ и среднего ДИАСТОЛИЧЕСКОГО КРОВЯНОГО ДАВЛЕНИЯ
для каждого субъекта среднее квадратичное отклонение, s , усредненных парных ОПРЕДЕЛЕНИЙ для
m
субъекта на ИСПЫТЫВАЕМОМ СФИГМОМАНОМЕТРЕ и ЭТАЛОННОМ СФИГМОМАНОМЕТРЕ должно удовлетворять
критериям, перечисленным в Таблице 1 или Таблице 2 при вычислении в соответствии с
Формулой (17).
m
s=× (17)
xx−
m n
∑()
j
m−1
j=1
где
среднее отклонение, вычисленное в соответствии с Формулой (15);
x
n
m число субъектов;
j индекс отдельного элемента;
x среднее значение на субъект, вычисленное в соответствии с Формулой (18).
j
x=× x (18)
j ∑ k
k=1
где
x ошибка испытываемого сфигмоманометра для каждого субъекта, вычисленная в соответствии
k
с Формулой (13) и Формулой (14);
k индекс отдельного элемента.
ПРИМЕР 4 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков (исследование на 85 субъектах).
16 © ISO 2013– Все права сохраняются
ПРИМЕР 5 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных для использования у взрослых ПАЦИЕНТОВ и/или
подростков и детей в возрасте от 3 до 12 лет (исследование на 85 субъектах).
ПРИМЕР 6 m = 35 для СФИГМОМАНОМЕТРОВ, предназначенных для (дополнительного) специального
ПРЕДПОЛАГАЕМОГО ИСПОЛЬЗОВАНИЯ (исследование на 35 субъектах), если СФИГМОМАНОМЕТР предварительно
подвергался КЛИНИЧЕСКИМ ИССЛЕДОВАНИЯМ в отдельных исследованиях на 85 субъектах в соответствии с 5.1.1 и 5.2.
ПРИМЕР 7 m = 85 для СФИГМОМАНОМЕТРОВ, предназначенных только для специального предполагаемого
использование (исследование на 85 субъектах).
5.2.5 * Дополнительные требования к СФИГМОМАНОМЕТРАМ, предназначенным для
использования в условиях испытаний с нагрузкой
Для СФИГМОМАНОМЕТРОВ, предназначенных для использования в испытаниях с нагрузкой, должны быть
проведены дополнительные КЛИНИЧЕСКИЕ ИССЛЕДОВАНИЯ. Во время КЛИНИЧЕСКИХ ИССЛЕДОВАНИЙ
субъекты должны подвергаться динамической (аэробной) нагрузке на велоэргометре или беговой
дорожке так, чтобы увеличить частоту сердечных сокращений с их частоты сердечных сокращений
покоя до целевой частоты сердечных сокращений величиной от 50 % до 70 % от средней
максимальной частоты сердечных сокращений (см. Приложение B). Должны записываться параметры
физической нагрузки велоэргометра и целевая частота сердечных сокращений. Рука, используемая
для ОПРЕДЕЛЕНИЯ должна поддерживаться на уровне сердца во время
...














Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...