Inhalational anaesthesia systems - Draw-over anaesthetic systems

ISO 18835:2015 specifies basic safety and essential performance requirements for anaesthetic systems utilizing the draw-over method to provide inhalational anaesthesia. Requirements are included to allow the use of these systems with both non-flammable and flammable anaesthetic agents. ISO 18835:2015 also includes requirements for a bellows-type manual ventilator. NOTE 1 Requirements for automatic anaesthetic ventilators are covered by ISO 80601‑2-13. NOTE 2 Requirements for operator-powered self-inflating bags are covered by ISO 10651‑4. ISO 18835:2015 does not specify requirements for monitoring of the equipment or the patient.

Systèmes d'anesthésie par inhalation — Alimentation en vapeur et équipements annexes

General Information

Status
Published
Publication Date
01-Apr-2015
Current Stage
9093 - International Standard confirmed
Start Date
07-Jul-2021
Completion Date
13-Dec-2025
Ref Project

Relations

Overview

ISO 18835:2015 - "Inhalational anaesthesia systems - Draw-over anaesthetic systems" specifies basic safety and essential performance requirements for anaesthetic systems that use the draw-over method to deliver inhalational anaesthesia. The standard covers systems designed to operate without compressed gas or electricity and includes provisions for use with both non‑flammable and flammable anaesthetic agents. It also contains requirements for a bellows‑type manual ventilator. Note that monitoring of patient parameters and automatic anaesthetic ventilators are outside its scope.

Key topics and technical requirements

  • Scope and risk management
    • Manufacturers must follow a risk management process in accordance with ISO 14971 and mitigate unacceptable risks by design, protection, monitoring/alarms, or labelling.
  • Construction and components
    • Defines required components: draw‑over vaporizer, breathing system, optional reservoir and bellows‑type manual ventilator for oxygen supplementation.
    • Addresses ports, connectors, operator‑detachable parts and marking of operator‑assembled components.
  • Performance
    • Sets performance limits for breathing resistance (example limits included: inspiration ≤ 0.6 kPa, expiration ≤ 0.2 kPa under specified conditions).
    • Specifies requirements for materials, mechanical hazards, particulate protection and environmental resistance.
  • Compatibility with agents
    • Requirements permit the use of both flammable and non‑flammable anaesthetic agents, with specific construction and materials guidance to ensure safety.
  • Documentation and marking
    • Details manufacturer information, labelling, instructions for use and packaging requirements.
  • Supporting annexes
    • Annex A: rationale for selected requirements; Annex B: normative test methods; Annex C: environmental aspects; Annex D: reference to essential principles.

Practical applications - who uses ISO 18835:2015

  • Medical device manufacturers designing or certifying draw‑over vaporizers, breathing systems, reservoirs and manual ventilators.
  • Regulatory bodies and notified bodies assessing conformity and safety claims.
  • Procurement teams and NGOs specifying equipment for low‑resource or field settings where compressed gas and electricity may be unavailable.
  • Biomedical engineers and clinical engineers responsible for maintenance, inspection and integration of draw‑over systems.
  • Clinicians and anesthetists seeking to understand equipment limits and safe use scenarios (note: patient monitoring standards are not included).

Related standards

  • ISO 14971 - Risk management for medical devices
  • ISO 80601‑2‑13 - Automatic anaesthetic ventilators (different scope)
  • ISO 10651‑4 - Operator‑powered self‑inflating bags
  • ISO 4135, ISO 5356‑1/2, ISO 5360, ISO 5367, ISO 23328 series - Connectors, vaporizers, breathing sets and filters
  • IEC 60601‑1, EN 13544‑2 - General electrical and respiratory equipment safety

Keywords: ISO 18835:2015, draw‑over anaesthetic systems, inhalational anaesthesia, draw‑over vaporizer, bellows‑type manual ventilator, low‑resource settings, medical device safety.

Standard
ISO 18835:2015 - Inhalational anaesthesia systems -- Draw-over anaesthetic systems
English language
22 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


INTERNATIONAL ISO
STANDARD 18835
First edition
2015-04-01
Inhalational anaesthesia systems —
Draw-over anaesthetic systems
Systèmes d’anesthésie par inhalation — Alimentation en vapeur et
équipements annexes
Reference number
©
ISO 2015
© ISO 2015
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2015 – All rights reserved

Contents Page
Foreword .iv
Introduction .v
1 * Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 General requirements . 2
4.1 Risk management . 2
4.2 Construction . 3
4.3 Performance . 4
4.4 * Components for use with flammable anaesthetic agents . 4
4.5 Materials . 4
4.6 Mechanical hazards . 4
4.7 Particulate matter. 4
4.8 Environmental requirements . 4
5 Draw-over vaporizer . 5
5.1 Construction . 5
5.2 Performance . 5
5.3 * Ports and connectors . 6
6 Breathing system. 6
6.1 Construction . 6
6.2 Performance . 7
6.3 Ports and connectors . 7
7 Reservoir . 7
7.1 Construction . 7
7.2 Performance . 7
7.3 Ports and connectors . 8
8 Bellows-type manual ventilator . 8
8.1 * Construction . 8
8.2 Performance . 8
8.3 Ports and connectors . 9
9 Marking of operator-assembled components . 9
9.1 General . 9
9.2 Marking of the draw-over vaporizer . 9
9.3 Marking of breathing-system attachments .10
9.4 Marking of reservoir .10
9.5 Marking of bellows-type manual ventilator .10
9.6 Marking of packages .11
10 Information supplied by the manufacturer .11
Annex A (informative) Rationale .13
Annex B (normative) Test methods .15
Annex C (informative) Environmental aspects .17
Annex D (informative) Reference to the essential principles .18
Bibliography .22
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any
patent rights identified during the development of the document will be in the Introduction and/or on
the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical Barriers
to Trade (TBT), see the following URL: Foreword — Supplementary information.
This International Standard cancels and replaces ISO/TS 18835:2004, which has been technically revised.
iv © ISO 2015 – All rights reserved

Introduction
The continuous-flow anaesthetic workstations described in ISO 80601-2-13 rely upon an uninterrupted
supply of compressed medical gases and electricity. These in turn depend upon a highly developed
infrastructure of transport, power generation, and technical services.
The World Health Organization (WHO) and the World Federation of Societies of Anaesthesiologists
(WFSA) have requested ISO ensure that the needs for safe anaesthesia for people in populous and low
to middle income countries of the world are also addressed in ISO standards for anaesthetic equipment.
In accordance with this request, ISO/TC 121/SC 1 has developed a standard for anaesthetic systems
(ISO 8835-7) that can give a safe inhalation anaesthetic without relying on electricity or compressed gas.
To achieve this, it is recognized that the draw-over anaesthetic system is an essential part of this
system. A technical specification for draw-over vaporizers and associated equipment, ISO/TS 18835
has been in publication since 2004 and forms the basis of this International Standard.
Throughout this International Standard, text for which a rationale is provided in Annex A is indicated
by an asterisk (*).
In this International Standard, the following print types are used:
— requirements, compliance with which can be verified, and definitions: roman type;
— notes and examples: smaller roman type;
— test methods: italic type;
— terms defined in this document: small caps.
INTERNATIONAL STANDARD ISO 18835:2015(E)
Inhalational anaesthesia systems — Draw-over
anaesthetic systems
1 * Scope
This International Standard specifies basic safety and essential performance requirements for
anaesthetic systems utilizing the draw-over method to provide inhalational anaesthesia.
Requirements are included to allow the use of these systems with both non-flammable and flammable
anaesthetic agents.
This International Standard also includes requirements for a bellows-type manual ventilator.
NOTE 1 Requirements for automatic anaesthetic ventilators are covered by ISO 80601-2-13.
NOTE 2 Requirements for operator-powered self-inflating bags are covered by ISO 10651-4.
This International Standard does not specify requirements for monitoring of the equipment or the patient.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
ISO 4135:2001, Anaesthetic and respiratory equipment — Vocabulary
ISO 5356-1, Anaesthetic and respiratory equipment — Conical connectors — Part 1: Cones and sockets
ISO 5356-2, Anaesthetic and respiratory equipment — Conical connectors — Part 2: Screw-threaded weight-
bearing connectors
ISO 5360, Anaesthetic vaporizers — Agent-specific filling systems
ISO 5367, Anaesthetic and respiratory equipment — Breathing sets and connectors
ISO 14971, Medical devices — Application of risk management to medical devices
ISO 23328-1, Breathing system filters for anaesthetic and respiratory use — Part 1: Salt test method to
assess filtration performance
ISO 23328-2, Breathing system filters for anaesthetic and respiratory use — Part 2: Non-filtration aspects
1)
ISO 80369-7 , Small bore connectors for liquids and gases in healthcare applications — Part 7: Connectors
with 6% (Luer) taper for intravascular or hypodermic applications
IEC 60601-1:2005+A1:2012, Medical electrical equipment — Part 1: General requirements for basic safety
and essential performance
EN 13544-2:2002+A1:2009, Respiratory therapy equipment — Part 2: Tubing and connectors
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 4135:2001 and the following apply.
1) This reference will be replaced by ISO 80369-2 once this International Standard has been published.
3.1
carrier gas
respirable gas that carries the anaesthetic agent to the patient
Note 1 to entry: A common example of a carrier gas is entrained ambient air supplemented with oxygen.
3.2
draw-over anaesthetic system
low-resistance system for administering inhalational anaesthesia that can be used in the absence of
compressed gas or electricity
3.3
draw-over valve system
a valve, or combination of valves, that controls unidirectional flow to the patient during inspiration
and unidirectional flow from the patient during exhalation under both spontaneous ventilation and
intermittent positive pressure ventilation (IPPV)
3.4
draw-over vaporizer
vaporizer from which a sufficient flow of gas vapour mixture is produced by lowering the pressure at
the outlet of the vaporizer below that at its inlet by a patient’s inspiratory effort or by a ventilator
[SOURCE: ISO 4135:2001, 4.1.8, modified — the breathing system and the vaporizer are treated as
separate devices in this International Standard, and thus, the phrase “either in the breathing system
and the vaporizer,” as used in ISO 4135, is inappropriate in this International Standard]
3.5
exhaust port
port through which the patient exhales to the atmosphere or into the inlet port of an anaesthetic gas
scavenging transfer system
[SOURCE: ISO 4135:2001, 4.2.1.6, modified — in order to be more specific as the exhaust port is an
integral part of the draw-over valve system]
3.6
operator-detachable
detachable without the use of a tool
3.7
patient connection port
opening at the patient end of a breathing system intended for connection of an airway device
[SOURCE: ISO 4135:2001, 4.2.1.2, modified — the examples that were not considered necessary for this
International Standard have been deleted]
3.8
reservoir
container where the carrier gas mixes with supplementary oxygen
4 General requirements
4.1 Risk management
The manufacturer of a draw-over anaesthetic system or parts intended for use in a draw-over
anaesthetic system shall follow a risk management process in accordance with ISO 14971. Any
unacceptable risk shall be mitigated by in this order:
a) design features which prevent the hazard;
b) inclusion of a means of protection;
2 © ISO 2015 – All rights reserved

c) inclusion of a monitoring and/or an alarm system;
d) safety and handling advice by marking or labelling.
If the inclusion of such risk mitigation measures is not feasible, the instructions for use shall contain
— a statement recommending that such risk mitigation measures be added prior to the use of the
draw-over anaesthetic system,
— sufficient specification of such risk mitigation measures, and
— a description of any residual risk to the patient or user/bystander.
Check compliance by inspection of the risk management file and, if applicable, the instructions for use.
4.2 Construction
A draw-over anaesthetic system shall comprise a draw-over vaporizer and a breathing system. It
can also include a ventilator and a reservoir if oxygen is to be added to the carrier gas.
NOTE An example of a draw-over anaesthetic system is shown in Figure 1.
Key
1 ambient air inlet 8 ventilator
2 means of particle protection 9 part of draw-over valve system
3 supplementary oxygen inlet 10 flow to and from patient
4 reservoir 11 exhaust port
5 draw-over vaporizer 12 direction of flow
6 breathing system tubing 13 pressure relief valve
7 part of draw-over valve system
Figure 1 — Schematic representation of the components of a draw-over anaesthetic system
4.3 Performance
The resistance to inspiration for the draw-over anaesthetic system shall not exceed 0,6 kPa and the
resistance to expiration shall not exceed 0,2 kPa under the following conditions:
a) a tidal volume of (600 ± 60) ml;
b) a frequency of 12 cycles per minute;
c) using a sinusoidal wave form or at an I:E ratio of 1:1.
Check compliance by the test given in B.3.
4.4 * Components for use with flammable anaesthetic agents
All components intended to be used with flammable anaesthetic agents shall comply with the applicable
requirements of IEC 60601-1:2005+A1:2012, 11.4.
Check compliance by inspection of the technical file.
4.5 Materials
The materials from which all components are made shall be selected to take into account the chemical
and physical properties of any substances with which the manufacturer declares that they may come
into contact with during use.
The selection procedures used for materials shall be documented and retained by the manufacturer.
NOTE draw-over anaesthetic systems have been used in areas where “normal” cleaning materials are not
available and many corrosive and abrasive materials have been substituted as cleaners. Water has also been used
to flush out draw-over vaporizers after use.
Check compliance by inspection of the technical file.
4.6 Mechanical hazards
Draw-over anaesthetic systems shall comply with IEC 60601-1:2005+A1:2012, Clause 9, where applicable.
Check compliance by inspection of the technical file and the tests given in IEC 60601-1:2005+A1:2012.
4.7 Particulate matter
Means shall be provided to reduce the risk of particulate matter from entering the draw-over
anaesthetic system and being inhaled by the patient.
NOTE Particulate matter can be taken to mean any foreign bodies including small creatures.
If a breathing system filter is used, it shall comply with ISO 23328-1 and ISO 23328-2.
Check compliance by visual inspection.
4.8 Environmental requirements
Draw over anaesthetic systems and components thereof shall not be adversely affected when
transported and stored under the following environmental conditions:
a) ambient temperatures between −20 °C and +70°C;
b) relative humidity between 30 % and 100 %;
c) atmospheric pressures between 500 hPa and 1060 hPa.
4 © ISO 2015 – All rights reserved

Check compliance by functional testing.
5 Draw-over vaporizer
5.1 Construction
5.1.1 A visual indication of the level of liquid anaesthetic agent contained within the draw-over
vaporizer shall be provided [see 9.2.1 f)].
Check compliance by visual inspection.
5.1.2 A control shall be provided to adjust the vapour concentration (volume fraction) calibrated for
each intended anaesthetic agent and
a) under normal operating conditions, it shall not be possible to set the control above the calibrated range,
b) the control can have a separate “OFF” position in addition to a “0” or “zero” position, and
c) a means shall be provided to reduce the risk of unintended change of the control from its set position.
Check compliance by visual inspection and functional testing.
5.1.3 When operated in accordance with the manufacturer’s instructions, it shall not be possible to
overfill the draw-over vaporizer such that
a) its performance is affected, or
b) the fluid level is no longer visible or indicated.
Check compliance by visual inspection and functional testing.
5.1.4 The draw-over vaporizer shall either
a) be provided with mounting fittings suitable to enable it to be rigidly supported, or
b) have a base designed to provide stability when freestanding.
Check compliance by visual inspection and functional testing.
5.2 Performance
5.2.1 The output in the “0”, “OFF”, or “zero” position shall be less than 0,1 %.
Check compliance by the test given in B.2.
5.2.2 * The accuracy of output shall be within ±20 % of set value for concentrations (volume fraction)
greater than 1 % and ±50 % of set value for concentrations of 1 % or below under the following conditions:
a) throughout the temperature range of 20 °C to 30 °C;
b) through the range of minute volumes from 2 l to 8 l;
c) at a frequency of 12 breaths per minute;
d) using a sinusoidal wave form or at an I:E ratio of 1:1.
The manufacturer shall disclose, in his technical file, the test method used to confirm the accuracy
requirements of this International Standard.
Check compliance by inspection of the technical file.
5.2.3 If the draw-over vaporizer is designed to operate without being rigidly attached to a mounting
rail, the draw-over vaporizer output shall remain within the manufacturer’s stated performance if the
draw-over vaporizer is tilted up to an angle of 30° from the vertical.
Check compliance by functional testing.
5.3 * Ports and connectors
5.3.1 The inlet port, if operator-detachable, shall be a 22 mm socket complying with ISO 5356-1
or ISO 5356-2.
Check compliance by visual inspection and functional testing.
5.3.2 The outlet port, if operator-detachable, shall be a 22 mm cone complying with ISO 5356-1
or ISO 5356-2.
Check compliance by visual inspection and functional testing.
5.3.3 If the draw-over vaporizer is fitted with an agent-specific filler port, it shall comply with the
requirements of ISO 5360.
Check compliance by visual inspection and the tests given in ISO 5360.
6 Breathing system
6.1 Construction
6.1.1 The breathing system shall include a draw-over valve system. The draw-over valve system
shall ensure that
a) the patient can inhale and exhale during both spontaneous and positive pressure ventilation,
b) the risk of jamming is minimized,
c) the rebreathing of expired gas is reduced to a minimum (see 6.2.1),
d) the unintentional entrainment of ambient air into the system is reduced to a minimum (see 6.2.2), and
e) the operation of the valve can be seen by the operator.
Check compliance by functional testing.
6.1.2 The breathing system shall be designed so that
a) it is operator-detachable, and
b) the risk of incorrect assembly is reduced.
Check compliance by visual inspection, functional testing, and inspection of the risk management file.
6.1.3 Breathing system tubes shall comply with ISO 5367.
Check compliance by inspection of the technical file.
6.1.4 Breathing systems should be designed to reduce the risk of cross contamination.
6 © ISO 2015 – All rights reserved

6.2 Performance
6.2.1 The reverse flow of gas from the patient shall be less than 1 % of the minute volume over a range
of 2 l to 8 l at 12 breaths per minute using a sinusoidal waveform.
Check compliance by functional testing.
6.2.2 The unintentional entrainment of air into the breathing system shall be less than 1 % of the
minute volume over a range of 2 l to 8 l at 12 breaths per minute using a sinusoidal waveform.
Check compliance by functional testing.
6.3 Ports and connectors
6.3.1 The inlet port of the breathing system, if operator-detachable, shall be a 22 mm socket
complying with ISO 5356-1.
Check compliance by visual inspection and functional testing.
6.3.2 The patient connection port shall be a 22/15 mm conical coaxial cone/socket complying
with ISO 5356-1.
Check compliance by visual inspection and functional testing.
6.3.3 * The exhaust port of the breathing system shall be a 30 mm cone complying with ISO 5356-1.
Check compliance by visual inspection and functional testing.
6.3.4 Other operator-detachable conical cones and sockets used in the breathing system shall comply
with ISO 5356-1 or ISO 5356-2.
Check compliance by visual inspection and functional testing.
7 Reservoir
7.1 Construction
7.1.1 The reservoir shall allow entry of ambient air and be fitted with a supplementary oxygen
inlet connector.
Check compliance by visual inspection and functional testing.
7.1.2 * The supplementary oxygen inlet connector shall be placed as close as possible to the ou
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

Frequently Asked Questions

ISO 18835:2015 is a standard published by the International Organization for Standardization (ISO). Its full title is "Inhalational anaesthesia systems - Draw-over anaesthetic systems". This standard covers: ISO 18835:2015 specifies basic safety and essential performance requirements for anaesthetic systems utilizing the draw-over method to provide inhalational anaesthesia. Requirements are included to allow the use of these systems with both non-flammable and flammable anaesthetic agents. ISO 18835:2015 also includes requirements for a bellows-type manual ventilator. NOTE 1 Requirements for automatic anaesthetic ventilators are covered by ISO 80601‑2-13. NOTE 2 Requirements for operator-powered self-inflating bags are covered by ISO 10651‑4. ISO 18835:2015 does not specify requirements for monitoring of the equipment or the patient.

ISO 18835:2015 specifies basic safety and essential performance requirements for anaesthetic systems utilizing the draw-over method to provide inhalational anaesthesia. Requirements are included to allow the use of these systems with both non-flammable and flammable anaesthetic agents. ISO 18835:2015 also includes requirements for a bellows-type manual ventilator. NOTE 1 Requirements for automatic anaesthetic ventilators are covered by ISO 80601‑2-13. NOTE 2 Requirements for operator-powered self-inflating bags are covered by ISO 10651‑4. ISO 18835:2015 does not specify requirements for monitoring of the equipment or the patient.

ISO 18835:2015 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 18835:2015 has the following relationships with other standards: It is inter standard links to ISO/TS 18835:2004. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase ISO 18835:2015 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.

The article discusses ISO 18835:2015, which outlines safety and performance requirements for inhalational anaesthesia systems that use the draw-over method. The standard covers systems that can be used with both non-flammable and flammable anaesthetic agents, as well as the requirements for a bellows-type manual ventilator. However, it does not include requirements for equipment or patient monitoring. Automatic anaesthetic ventilators and operator-powered self-inflating bags are covered by other ISO standards.

The article discusses ISO 18835:2015, which sets safety and performance requirements for draw-over anaesthetic systems used for inhalational anaesthesia. These systems can be used with both flammable and non-flammable anaesthetic agents. The standard also includes requirements for a bellows-type manual ventilator, but does not cover monitoring requirements for the equipment or the patient.

記事タイトル:ISO 18835:2015 - 吸入麻酔システム - ドローオーバー麻酔システム 記事内容:ISO 18835:2015は、吸入麻酔を提供するためにドローオーバー方式を利用する麻酔システムの基本的な安全性と必要な性能要件を指定しています。このシステムは可燃性および非可燃性の麻酔薬とともに使用するための要件が含まれています。ISO 18835:2015にはベローズ型マニュアルベンチレーターの要件も含まれています。注意事項1:自動麻酔ベンチレーターの要件はISO 80601-2-13でカバーされています。注意事項2:操作者駆動型の自己膨張袋の要件はISO 10651-4でカバーされています。ISO 18835:2015は、機器や患者のモニタリングの要件を指定していません。

記事のタイトル:ISO 18835:2015 - 吸入麻酔システム-ドローオーバー麻酔システム 記事の内容:ISO 18835:2015は、吸入麻酔を提供するためにドローオーバー法を使用する麻酔システムの基本的な安全性と必須のパフォーマンス要件を定めています。これらのシステムを非可燃性および可燃性の麻酔薬と併用するための要件も含まれています。ISO 18835:2015にはベローズ型のマニュアル換気器の要件も含まれています。注意1:自動麻酔換気器の要件はISO 80601-2-13でカバーされています。注意2:操作者駆動の自己膨張バッグの要件はISO 10651-4でカバーされています。 ISO 18835:2015は、機器または患者の監視の要件を特定していません。

기사 제목: ISO 18835:2015 - 흡입 마취 시스템 - 드로오버 마취 시스템 기사 내용: ISO 18835:2015는 드로오버 방법을 사용하여 흡입 마취를 제공하는 마취 시스템에 대한 기본 안전 및 필수 성능 요구 사항을 명시합니다. 이러한 시스템을 비불꽃성 및 불꽃성 마취제와 함께 사용할 수 있도록 요구 사항이 포함되어 있습니다. ISO 18835:2015에는 풍선형 수동 환기장치에 대한 요구 사항도 포함되어 있습니다. 참고 1: 자동 마취 환기 장치에 대한 요구 사항은 ISO 80601‑2-13에 다루어져 있습니다. 참고 2: 작동자가 동작시키는 자가 팽창 가방에 대한 요구 사항은 ISO 10651‑4에서 다루어집니다. ISO 18835:2015는 장비 또는 환자 모니터링에 대한 요구 사항을 명시하지 않습니다.

기사 제목: ISO 18835:2015 - 흡입 마취 시스템 - Draw-over 마취 시스템 기사 내용: ISO 18835:2015는 흡입 마취를 제공하기 위해 Draw-over 방법을 사용하는 마취 시스템에 대한 기본적인 안전 및 필수 성능 요구 사항을 명시합니다. 이러한 시스템을 불연 및 연소성 마취제와 함께 사용할 수 있도록 요구 사항이 포함되어 있습니다. ISO 18835:2015에는 벨로우 유형의 수동 환기 장치에 대한 요구 사항도 포함되어 있습니다. 주의 1: 자동 마취 환기 장치에 대한 요구 사항은 ISO 80601-2-13에서 다루고 있습니다. 주의 2: 작업자가 작동하는 자체 팽창 가방에 대한 요구 사항은 ISO 10651-4에서 다루고 있습니다. ISO 18835:2015는 장비나 환자의 모니터링 요구 사항을 명시하지 않습니다.