IEC 61674:1997/AMD1:2002
(Amendment)Amendment 1 - Medical electrical equipment - Dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging
Amendment 1 - Medical electrical equipment - Dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging
Contains a new subclause on surges and a modified subclause on voltage dips, short interruptions and voltage variations (electromagnetic compatibility).
General Information
- Status
- Published
- Publication Date
- 04-Jun-2002
- Drafting Committee
- WG 3 - TC 62/SC 62C/WG 3
- Current Stage
- DELPUB - Deleted Publication
- Start Date
- 29-Nov-2012
- Completion Date
- 13-Feb-2026
Relations
- Effective Date
- 05-Sep-2023
- Effective Date
- 05-Sep-2023
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
IEC 61674:1997/AMD1:2002 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Amendment 1 - Medical electrical equipment - Dosimeters with ionization chambers and/or semi-conductor detectors as used in X-ray diagnostic imaging". This standard covers: Contains a new subclause on surges and a modified subclause on voltage dips, short interruptions and voltage variations (electromagnetic compatibility).
Contains a new subclause on surges and a modified subclause on voltage dips, short interruptions and voltage variations (electromagnetic compatibility).
IEC 61674:1997/AMD1:2002 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 61674:1997/AMD1:2002 has the following relationships with other standards: It is inter standard links to IEC 61674:1997, IEC 61674:2012. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
IEC 61674:1997/AMD1:2002 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
INTERNATIONAL IEC
STANDARD
AMENDMENT 1
2002-06
Amendment 1
Medical electrical equipment –
Dosimeters with ionization chambers
and/or semi-conductor detectors
as used in X-ray diagnostic imaging
Amendement 1
Appareils électromédicaux –
Dosimètres à chambres d'ionisation et/ou
à détecteurs à semi-conducteurs utilisés
en imagerie de diagnostic à rayonnement X
IEC 2002 Droits de reproduction réservés Copyright - all rights reserved
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
PRICE CODE
Commission Electrotechnique Internationale
C
International Electrotechnical Commission
Международная Электротехническая Комиссия
For price, see current catalogue
– 2 – 61674 Amend. 1 IEC:2002(E)
FOREWORD
This amendment has been prepared by subcommittee SC 62C: Equipment for radiotherapy,
nuclear medicine and radiation dosimetry, of IEC Technical Committee 62: Electrical equip-
ment in medical practice.
The text of this amendment is based on the following documents:
FDIS Report on voting
62C/333/FDIS 62C/339/RVD
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.
The committee has decided that the contents of the base publication and its amendments will
remain unchanged until 2006. At this date, in accordance with the committee’s decision, the
publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
Page 3
CONTENTS
Add, on page 5, after 6.8.3, the title of the new subclause 6.8.4, and renumber the existing
6.8.4 as 6.8.5:
6.8.4 Surges
6.8.5 Voltage dips, short interruptions and voltage variations
Page 51
Add the following new subclause 6.8.4:
6.8.4 Surges
The maximum spurious indications (both transient and permanent) of the display or data
output due to surges shall be less than the limits given in table 7). The test is not to be
performed on the connection lines between the detector and the measuring assembly.
For mains-operated instruments compliance shall be checked by observing and recording the
indication
...