IEC 60601-2-2:2017
(Main)Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories.
HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements.
The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
- refinement and additions to the defined terms;
- additional separation of the requirements for HF surgical equipment and HF surgical accessories;
- a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes;
- new requirements for devices that have or use a high current mode.
Appareils électromédicaux - Partie 2-2: Exigences particulières pour la sécurité de base et les performances essentielles des appareils d'électrochirurgie à courant haute fréquence et des accessoires d'électrochirurgie à courant haute fréquence
L'IEC 60601-2-2:2017 s’applique à la sécurité de base et aux performances essentielles des appareils d’électrochirurgie HF et des accessoires d'électrochirurgie HF. Les appareils d’électrochirurgie HF dont la puissance de sortie assignée est inférieure ou égale à 50 W (destinés, par exemple, à la micro coagulation, à l’ophtalmologie ou à l’usage dentaire) sont exemptés de certaines exigences de la présente norme particulière. Ces exemptions sont indiquées dans les exigences correspondantes.
La présente norme particulière a pour objet d’établir des exigences particulières relatives à la sécurité de base et aux performances essentielles des appareils d’électrochirurgie HF et des accessoires d’électrochirurgie HF.
Cette sixième édition annule et remplace la cinquième édition parue en 2009. Cette édition constitue une révision technique. Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente:
- des précisions et des ajouts aux termes définis;
- une séparation supplémentaire des exigences relatives aux appareils d'électrochirurgie à courant haute fréquence (HF) et aux accessoires d'électrochirurgie à courant haute fréquence (HF);
- une nouvelle exigence concernant les électrodes neutres adultes devant servir d'électrodes neutres de surveillance de la qualité du contact;
- de nouvelles exigences relatives aux appareils ayant ou utilisant un mode de courant élevé.
General Information
- Status
- Published
- Publication Date
- 20-Feb-2023
- Technical Committee
- SC 62D - Particular medical equipment, software, and systems
- Drafting Committee
- MT 17 - TC 62/SC 62D/MT 17
- Current Stage
- PPUB - Publication issued
- Start Date
- 31-Mar-2017
- Completion Date
- 10-Mar-2017
Relations
- Effective Date
- 05-Sep-2023
- Effective Date
- 26-Oct-2025
- Effective Date
- 05-Sep-2023
- Effective Date
- 05-Sep-2023
Overview
IEC 60601-2-2:2017 - "Medical electrical equipment - Part 2-2" specifies the particular basic safety and essential performance requirements for high frequency (HF) surgical equipment and HF surgical accessories. This sixth edition (2017) replaces the 2009 edition and constitutes a technical revision. It applies to HF surgical equipment and accessories used for electrosurgery (including monopolar and bipolar techniques), with limited exemptions for equipment rated ≤ 50 W (e.g., micro-coagulation, dentistry, ophthalmology) noted in the standard.
Key topics and requirements
- Scope and object: Establishes particular safety and essential performance criteria that supplement IEC 60601-1 (the general standard for medical electrical equipment).
- Defined terms and classification: Refinements to definitions and clearer separation of requirements for HF equipment vs HF accessories.
- Electrical safety: Protection against electrical hazards, leakage current measurement methods, earthing/isolated patient circuit considerations (illustrated test circuits are included).
- Neutral electrode requirements: New requirement that adult neutral electrodes be contact quality monitoring neutral electrodes.
- High current mode: Additional requirements for devices that provide or use a high current mode.
- Performance and outputs: Measurement of HF output power (monopolar and bipolar outputs) and protection against hazardous outputs.
- Mechanical, thermal and radiation safety: Protection against mechanical hazards, excessive temperatures and unwanted radiation.
- Programmable systems & alarms: Requirements for PEMS (programmable electrical medical systems) and alarm systems used in HF surgical devices.
- Electromagnetic compatibility (EMC): EMC requirements and emissions testing specific to HF surgical equipment (Annex BB addresses emissions).
- Documentation and markings: Identification, labelling, and user information required for safe operation.
- Testing: Mandatory test specifications and informative guidance (Annex AA) to support compliance.
Applications and users
IEC 60601-2-2:2017 is intended for:
- Manufacturers and designers of electrosurgical units (ESUs), HF generators, and surgical accessories (neutral/active electrodes, cords).
- Regulatory and conformity assessment bodies assessing product safety and essential performance.
- Test laboratories and certification bodies performing electrical safety, EMC and performance testing.
- Hospital biomedical engineers and clinical procurement teams ensuring equipment meets safety standards and selecting compatible accessories.
Practical uses include product design validation, pre-market compliance testing, clinical risk mitigation, and supplier quality requirements for electrosurgical systems.
Related standards
- IEC 60601-1 (General requirements for basic safety and essential performance) - mandatory companion standard.
- Other parts of the IEC 60601 series covering alarms, EMC, and specific device types.
Keywords: IEC 60601-2-2:2017, high frequency surgical equipment, HF surgical accessories, electrosurgery, medical electrical equipment, neutral electrode, high current mode, safety, essential performance, EMC.
IEC 60601-2-2:2017 RLV - Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories Released:3/31/2017 Isbn:9782832241851
IEC 60601-2-2:2017 - Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
IEC 60601-2-2:2017+AMD1:2023 CSV - Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories Released:2/21/2023
Frequently Asked Questions
IEC 60601-2-2:2017 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories". This standard covers: IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories. HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements. The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: - refinement and additions to the defined terms; - additional separation of the requirements for HF surgical equipment and HF surgical accessories; - a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes; - new requirements for devices that have or use a high current mode.
IEC 60601-2-2:2017 applies to the basic safety and essential performance of HF surgical equipment and HF surgical accessories. HF surgical equipment having a rated output power not exceeding 50 W (for example for micro-coagulation, or for use in dentistry or ophthalmology) is exempt from certain of the requirements of this particular standard. These exemptions are indicated in the relevant requirements. The object of this particular standard is to establish particular basic safety and essential performance requirements for HF surgical equipment and HF surgical accessories. This sixth edition cancels and replaces the fifth edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: - refinement and additions to the defined terms; - additional separation of the requirements for HF surgical equipment and HF surgical accessories; - a new requirement for adult neutral electrodes to be contact quality monitoring neutral electrodes; - new requirements for devices that have or use a high current mode.
IEC 60601-2-2:2017 is classified under the following ICS (International Classification for Standards) categories: 11.040.30 - Surgical instruments and materials. The ICS classification helps identify the subject area and facilitates finding related standards.
IEC 60601-2-2:2017 has the following relationships with other standards: It is inter standard links to IEC 60601-2-2:2017/AMD1:2023, IEC 60601-2-2:2017/AMD1:2023/ISH1:2025, IEC 60601-2-2:2009, IEC 60601-2-2:2009/COR1:2014. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
You can purchase IEC 60601-2-2:2017 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of IEC standards.
Standards Content (Sample)
IEC 60601-2-2 ®
Edition 6.0 2017-03
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
IEC 60601-2-2 ®
Edition 6.0 2017-03
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 11.040.30 ISBN 978-2-8322-4185-1
– 2 – IEC 60601-2-2:2017 RLV © IEC 2017
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 9
201.3 Terms and definitions . 10
201.4 General requirements . 14
201.5 General requirements for testing of ME EQUIPMENT . 15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 16
201.7 ME EQUIPMENT identification, marking and documents . 16
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 21
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 36
201.10 Protection against unwanted and excessive radiation HAZARDS . 36
201.11 Protection against excessive temperatures and other HAZARDS . 37
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 38
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 45
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 45
201.15 Construction of ME EQUIPMENT . 45
201.16 ME SYSTEMS . 50
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 50
202 * ELECTROMAGNETIC compatibility DISTURBANCES – Requirements and tests . 50
208 General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems . 51
Annexes . 53
Annex AA (informative) Particular guidance and rationale . 54
Annex BB (informative) ELECTROMAGNETIC DISTURBANCES created by HF SURGICAL
EQUIPMENT . 81
Bibliography . 90
Index of defined terms used in this particular standard . 92
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT . 16
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT . 16
Figure 201.103 – Circuit suitable for testing compliance to 201.8.4.101 . 22
Figure 201.104 – Measurement of HF LEAKAGE CURRENT with NEUTRAL ELECTRODE
referenced to earth for EARTH REFERENCED PATIENT CIRCUITS and load between
electrodes . 25
Figure 201.105 – Measurement of HF LEAKAGE CURRENT WITH NEUTRAL ELECTRODE
REFERENCED TO EARTH for EARTH REFERENCED PATIENT CIRCUITS and a load resistance
from ACTIVE ELECTRODE to earth . 26
Figure 201.106 – Measurement of HF LEAKAGE CURRENT with NEUTRAL ELECTRODE
isolated from earth at high frequency for HF ISOLATED PATIENT CIRCUITS . 27
Figure 201.107 – Measurement of HF LEAKAGE CURRENT from a BIPOLAR ELECTRODE
ACCESSORY . 28
Figure 201.108 – Test apparatus for anchorages of cords of ACTIVE ACCESSORY . 35
Figure 201.109 – Measurement of output power – MONOPOLAR output . 40
Figure 201.110 – Measurement of output power – BIPOLAR output . 41
Figure 201.111 – Method of testing feedback from one active output to another in
simultaneous activation . 44
Figure AA.1 – Examples of various parts of an HF surgical ME SYSTEM . 56
Figure AA.2 – Example of MONOPOLAR method of HF surgery using a NEUTRAL
ELECTRODE . 56
Figure AA.3 – Example of BIPOLAR method of HF surgery . 57
Figure AA.4 – CREST FACTOR vs. peak voltage . 62
Figure AA.5 – Example of PATIENT circuit with NEUTRAL ELECTRODE referenced to earth
at operating frequencies . 66
Figure BB.1 – E-FIELD EMISSIONS test setup . 84
Figure BB.2 – H-FIELD EMISSIONS test setup . 85
Figure BB.3 – Conducted EMISSIONS test setup . 86
Figure BB.4 – Unit ad hoc test . 88
Figure BB.5 – Power cord ad hoc test . 89
Figure BB.6 – ACCESSORY cord ad hoc test . 89
Table 201.101 – Colours of indicator lights and their meaning for HF SURGICAL
EQUIPMENT . 17
Table 201.102 – Maximum output powers in SINGLE FAULT CONDITIONS . 43
Table 201.103 – Test currents by weight range . 47
Table AA.1 – Summary of measured current and durations for 25 TUR procedures . 75
Table AA.2 – Summary of measured currents and durations for general surgical
procedures. 76
Table BB.1 – Worst case EMISSIONS of spark gap type HF SURGICAL EQUIPMENT . 86
Table BB.2 – Worst case EMISSIONS of non-spark gap (modern) HF SURGICAL EQUIPMENT . 86
– 4 – IEC 60601-2-2:2017 RLV IEC 2017
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This redline version of the official IEC Standard allows the user to identify the changes
made to the previous edition. A vertical bar appears in the margin wherever a change
has been made. Additions are in green text, deletions are in strikethrough red text.
International standard IEC 60601-2-2 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This sixth edition cancels and replaces the fifth edition published in 2009. This edition
constitutes a technical revision. This edition includes the following significant technical
changes with respect to the previous edition:
– refinement and additions to the defined terms;
– additional separation of the requirements for HF surgical equipment and HF surgical
accessories;
– a new requirement for adult neutral electrodes to be contact quality monitoring neutral
electrodes;
– new requirements for devices that have or use a high current mode.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/1427/FDIS 62D/1442/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– 6 – IEC 60601-2-2:2017 RLV IEC 2017
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC website under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The “colour inside” logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct understanding
of its contents. Users should therefore print this publication using a colour printer.
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HIGH FREQUENCY SURGICAL
EQUIPMENT.
This particular standard amends and supplements IEC 60601-1:2005 and Amendment 1:2012,
Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance, hereinafter referred to as the general standard (see 201.1.4).
The requirements are followed by specifications for the relevant tests.
A "Particular guidance and rationale" section giving some explanatory notes, where
appropriate, about the more important requirements is included in Annex AA.
Clauses or subclauses for which there are explanatory notes in Annex AA are marked with an
asterisk (*).
It is considered that a knowledge of the reasons for these requirements will not only facilitate
the proper application of the standard but will, in due course, expedite any revision neces-
sitated by changes in clinical practice or as a result of developments in technology. However,
this annex does not form part of the requirements of this document.
– 8 – IEC 60601-2-2:2017 RLV IEC 2017
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HF
SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as defined in 201.3.224 and 201.3.223.
HF SURGICAL EQUIPMENT having a RATED OUTPUT POWER not exceeding 50 W (for example for
micro-COAGULATION, or for use in dentistry or ophthalmology) is exempt from certain of the
requirements of this particular standard. These exemptions are indicated in the relevant
requirements.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for HF SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as
defined in 201.3.224 and 201.3.223.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2:2014 and IEC 60601-1-8:2006 apply as modified in Clauses 202 and 208
respectively. IEC 60601-1-3, IEC 60601-1-10 and IEC 60601-1-11 do not apply. All other
published collateral standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
______________
The general standard is IEC 60601-1:2005/AMD1:2012, Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.
IEC 60601-1-11, Medical electrical equipment – Part 1-11: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment (in preparation).
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this document addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 87.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2007 2014, Medical electrical equipment – Part 1-2: General requirements for
basic safety and essential performance – Collateral Standard: Electromagnetic compatibility
disturbances – Requirements and tests
– 10 – IEC 60601-2-2:2017 RLV IEC 2017
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral Standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
Addition:
CISPR 11:2003 2015, Industrial, scientific and medical equipment – Radio-frequency
equipment – Electromagnetic disturbance characteristics – Limits and methods of
measurement
IEC 61000-4-3:2006, Electromagnetic compatibility (EMC) – Part 4-3: Testing and
measurement techniques – Radiated, radio-frequency electromagnetic field immunity test
IEC 61000-4-6:2003 2013, Electromagnetic compatibility (EMC) – Part 4-6: Testing and
measurement techniques – Immunity to conducted disturbances, induced by radio-frequency
fields
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
the following apply, except as follows:
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
Replace NOTE 1 with the following:
NOTE 1 Where the terms “voltage” and “current” are used in this document, they mean the RMS values of an
alternating, direct or composite voltage or current averaged over 1 s unless stated otherwise.
Addition:
201.3.201
ACTIVE ACCESSORY
HF SURGICAL ACCESSORY intended for manipulation by the OPERATOR to produce surgical an
effect by electrical conduction adjacent to the ACTIVE ELECTRODE at the intended site on the
PATIENT, generally comprising an ACTIVE HANDLE, the cord of an ACTIVE ACCESSORY, ACTIVE
CONNECTOR and ACTIVE ELECTRODE
201.3.202
ACTIVE CONNECTOR
part of an ACTIVE ACCESSORY intended for connection to an ACTIVE OUTPUT TERMINAL, which
may include additional terminals for connection of a FINGERSWITCH to a SWITCH SENSOR
201.3.203
ACTIVE ELECTRODE
part of an ACTIVE ACCESSORY extending from the ACTIVE HANDLE to the surgical site and
intended to pass HF current into body tissue
201.3.204
ACTIVE ELECTRODE INSULATION
electrical insulation material affixed to part of an ACTIVE ELECTRODE intended to prevent
unintended injury to adjacent PATIENT tissue or the OPERATOR
201.3.205
ACTIVE HANDLE
part of an ACTIVE ACCESSORY intended to be held by the OPERATOR
201.3.206
ACTIVE OUTPUT TERMINAL
part of HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an ACTIVE
ACCESSORY and for delivery of HF current thereto
Note 1 to entry: An ACTIVE CONNECTOR is that which plugs into an ACTIVE OUTPUT TERMINAL.
Note 2 to entry: See Figure AA.1.
201.3.207
*ASSOCIATED EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT other than HF SURGICAL EQUIPMENT that may be electrically
connected to the PATIENT circuit and not intended for independent use
201.3.208
*BIPOLAR
method of applying HF output current to a PATIENT via multiple-pole between two or more
ACTIVE ELECTRODES without the need for a separately connected NEUTRAL ELECTRODE (or the
need to use the PATIENT’S body capacitance to earth) in which an effect is intended in tissue
near one or more ACTIVE ELECTRODES
Note 1 to entry: The BIPOLAR method includes devices energizing pairs of ACTIVE ELECTRODES as well as devices
energizing groups of ACTIVE ELECTRODES where the HF current source and return may have different numbers of
electrodes.
Note 2 to entry: See Figure AA.1 and Figure AA.3.
201.3.209
BIPOLAR ELECTRODE ACCESSORY
assembly of ACTIVE ACCESSORY comprising two or more ACTIVE ELECTRODES on the same
support, so constructed that, when energized, the HF current flows mainly amongst these
electrodes
201.3.210
COAGULATION
use of HF current to elevate the temperature of tissue, e.g. to reduce or terminate undesired
bleeding induce a thermal effect, e.g. to control or prevent bleeding, induce tissue destruction,
or induce tissue shrinkage
Note 1 to entry: COAGULATION may take the form of contact or non-contact COAGULATION.
Note 2 to entry: FULGURATION, desiccation, spray, forced, swift, soft and argon beam (plasma) COAGULATION are
all names of COAGULATION types.
201.3.211
CONTACT QUALITY MONITOR
CQM
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to a
MONITORING NE providing an alarm in the event that NEUTRAL ELECTRODE (NE) contact with the
PATIENT becomes insufficient
Note 1 to entry: CONTACT QUALITY MONITOR is functional only when used with a MONITORING NE.
201.3.212
CONTINUITY MONITOR
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an NE,
except MONITORING NE, providing an alarm in the event of electrical discontinuity in the NE
cable or its connections
– 12 – IEC 60601-2-2:2017 RLV IEC 2017
201.3.213
*CREST FACTOR
dimensionless value equal to the peak output voltage divided by the RMS voltage as
measured at the output of HF SURGICAL EQUIPMENT in an open circuit condition
Note 1 to entry: Specific information on the correct way to make the measurements needed to calculate this value
may be found in Annex AA.
201.3.214
*CUTTING
resection or dissection division of body tissue caused by the passage of HIGH FREQUENCY
current of high current density at the ACTIVE ELECTRODE (S)
201.3.215
*EARTH REFERENCED PATIENT CIRCUIT
PATIENT circuit which includes components, such as capacitors, installed to provide a low-
impedance path to earth for HF currents
201.3.216
FINGERSWITCH
device generally included with an ACTIVE ACCESSORY which, when manipulated by the
OPERATOR, enables HF output to be produced and, when released disables HF output
Note 1 to entry: Requirements for similar switches intended to perform functions other than activation of HF output
are under consideration.
201.3.217
*FULGURATION
form of COAGULATION using long (0,5 mm or more) electrical sparks to heat tissue surfaces
superficially, with no intentional mechanical contact between the ACTIVE ELECTRODE and the
tissue the use of HF current to produce an effect on a tissue surface by electrical sparks from
an ACTIVE ELECTRODE that is not in physical contact with the tissue
201.3.218
*HEATING FACTOR
a value equal to I × t where I is the MONOPOLAR current in amperes and t is the duration of
the current flow in s
Note 1 to entry: The HEATING FACTOR is expressed as A s (amperes squared seconds).
Note 2 to entry: See subclause 201.15.101.5 in Annex AA for additional information.
201.3.219
*HIGH CURRENT MODE
MONOPOLAR output mode whose INTENDED USE (MAXIMUM OUTPUT CURRENT and maximum DUTY
CYCLE) results in a HEATING FACTOR of greater than 30 A s in any 60 s period
201.3.220
*HIGH FREQUENCY
HF
frequencies less than 5 MHz and generally greater than 200 kHz
201.3.221
HF ISOLATED PATIENT CIRCUIT
HF PATIENT CIRCUIT where there are no components installed to provide a low-impedance path
to earth for HF currents
201.3.222
HF PATIENT CIRCUIT
any electrical circuit which contains one or more PATIENT CONNECTIONS including all conductive
parts of the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT circuits through which HF
current is intended to flow between the ME EQUIPMENT and the PATIENT in NORMAL CONDITION or
SINGLE FAULT CONDITION
201.3.223
HF SURGICAL ACCESSORY
ACCESSORY intended to conduct, supplement or monitor HF energy applied to the PATIENT from
HF SURGICAL EQUIPMENT
Note 1 to entry: HF SURGICAL ACCESSORIES include HF surgical application electrodes ACTIVE ACCESSORIES,
including cords and connectors for attachment to HF SURGICAL EQUIPMENT, NEUTRAL ELECTRODES, as well as other
ASSOCIATED EQUIPMENT intended for connection to the HF surgical PATIENT circuit. See Figure AA.1.
Note 2 to entry: Not all accessories used with HF surgical equipment are HF surgical accessories.
201.3.224
HF SURGICAL EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT including associated ACCESSORIES which generates HIGH
FREQUENCY currents intended for the performance of surgical operations tasks, such as the
CUTTING or COAGULATION of biological tissue by means of these HIGH FREQUENCY currents
Note 1 to entry: HF SURGICAL EQUIPMENT is also variously known as surgical diathermy, electrosurgical equipment,
electrosurgical generator, RF generator or HF generator.
Note 2 to entry: A footswitch is an example of an associated ACCESSORY that is part of HF SURGICAL EQUIPMENT.
See Figure AA.1.
201.3.225
*HF SURGICAL MODE
any of a number of OPERATOR selectable HF output characteristics intended to provide a
specific indicated surgical effect at a connected ACTIVE ACCESSORY, such as CUTTING,
COAGULATION and the like
Note 1 to entry: Each available HF SURGICAL MODE may be provided with an OPERATOR-adjustable output control to
set the desired intensity or speed of the surgical effect.
201.3.226
*MAXIMUM OUTPUT CURRENT
for each available HF SURGICAL MODE, the magnitude of the maximum possible HF output
current during INTENDED USE
201.3.227
*MAXIMUM OUTPUT VOLTAGE
for each available HF SURGICAL MODE, the magnitude of the maximum possible peak HF output
voltage appearing between PATIENT circuit connections
201.3.228
*MONITORING NE
NE intended for use with a CONTACT QUALITY MONITOR
Note 1 to entry: A MONITORING NEUTRAL ELECTRODE is also known as a split plate, dual plate, dual foil electrode or
CQM electrode.
201.3.229
*MONOPOLAR
method of applying HF output current to a PATIENT via an ACTIVE ELECTRODE and returning via a
separately PATIENT-connected NEUTRAL ELECTRODE (or via the PATIENT’S body capacitance to
earth) in which an effect is intended only in tissue at or near the ACTIVE ELECTRODE
Note 1 to entry: See Figures AA.1 and AA.2.
– 14 – IEC 60601-2-2:2017 RLV IEC 2017
201.3.230
NEUTRAL ELECTRODE
NE
electrode of a relatively large area for connection to the body of the PATIENT, intended to
provide an electrical return path for the MONOPOLAR application of HIGH FREQUENCY current
with such a low current density in the body PATIENT’S tissue that physical effects such as
excessive rise in temperature or unwanted burns are avoided
Note 1 to entry: The NEUTRAL ELECTRODE is also known as plate, plate electrode, electrosurgical pad, passive,
return or dispersive electrode.
Note 2 to entry: To keep the current density low enough to prevent unwanted heating, the NEUTRAL ELECTRODE
needs to have a large enough area.
Note 3 to entry: A NEUTRAL ELECTRODE is usually in contact with the PATIENT at a location that is separate from the
MONOPOLAR ACTIVE ELECTRODE.
Note 4 to entry: See Figures AA.1 and AA.2.
201.3.231.1
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be
applied with respect to an NE connected to the PATIENT
201.3.231.2
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be applied to
pairs of opposite polarity
201.3.232
RATED LOAD
value of non-reactive load resistance which, when connected results in the maximum HF
output power from each HF SURGICAL MODE of the HF SURGICAL EQUIPMENT
201.3.233
RATED OUTPUT POWER
for each HF SURGICAL MODE set at its maximum output setting, the power in watts produced
when all ACTIVE OUTPUT TERMINALS which can be activated simultaneously are connected to
their respective RATED LOADS
201.3.234
SWITCH SENSOR
part of HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT which controls activation of HF output
in response to operation of a connected FINGERSWITCH or footswitch
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
Additional subclauses:
201.4.1.101 * Additional conditions for application
The compliance of HF SURGICAL EQUIPMENT to this document and the compliance of HF
SURGICAL ACCESSORIES to this document shall be independent of each other, except where
specifically required by conformance tests or by the MANUFACTURER.
201.4.2.3.101 * Evaluating RISK MANAGEMENT PROCESS for ME EQUIPMENT or ME SYSTEMS
Addition:
MANUFACTURERS shall include, within their RISK ANALYSIS, the potential for their HF SURGICAL
EQUIPMENT and/or HF SURGICAL ACCESSORIES to be used in MONOPOLAR HIGH CURRENT MODE
situations and the impact this would have on the heating under the NEUTRAL ELECTRODE (for
example, see 201.7.9.2.2.101 f)).
201.4.3 * ESSENTIAL PERFORMANCE
Addition:
The requirements listed in the third hyphen of 201.8.4.101 and in 201.12.4.101 shall be
considered ESSENTIAL PERFORMANCE requirements.
NOTE 101 Please refer to Annex AA.
201.4.7 SINGLE FAULT CONDITION for ME EQUIPMENT
Addition:
Additional SINGLE FAULT CONDITIONS subclause:
201.4.7.101 Specific SINGLE FAULT CONDITIONS
The following SINGLE FAULT CONDITIONS are the subject of specific requirements and tests in
this document:
a) failure in the CONTINUITY MONITOR or CONTACT QUALITY MONITOR which might cause a
RISK (see 201.8.4.101);
unacceptable
b) a defect in the output switching circuit resulting in an excessive low-frequency PATIENT
LEAKAGE CURRENT (see 201.8.10.4.101.1);
c) any defect which results in the unwanted energization of the PATIENT circuit (see
201.12.4.2.101);
d) any defect which results in a significant increase in output power relative to the output
setting (see 201.12.4.4.101).
201.4.11 Power input
Amendment:
Replacement of first dash in compliance tests:
– The HF SURGICAL EQUIPMENT shall be operated in the output mode and using the load
which creates the greatest steady state input current. Input current is measured and
compared with markings and the contents of the technical description.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.4 * Other conditions
Addition:
aa) Particular care shall be taken to ensure accuracy and safety during measurement of HF
output. See Annex AA for guidance.
– 16 – IEC 60601-2-2:2017 RLV IEC 2017
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2.8.2 Other power sources
Amendment:
Subclause 7.2.8.2 of the general standard does not apply to ACTIVE OUTPUT TERMINALS or NE
terminals.
201.7.2.10 APPLIED PARTS
Addition:
The relevant symbols required for marking DEFIBRILLATION-PROOF APPLIED PARTS shall be
attached to the front panel, but are not required on the APPLIED PARTS.
Connections on the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT for the connection of
NE leads shall be marked with the symbols given in Figures 201.101 and 201.102 as follows:
IEC
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT
IEC
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT
Additional subclause:
201.7.2.10.101 * HF SURGICAL ACCESSORIES
HF SURGICAL ACCESSORIES (excluding HF ASSOCIATED EQUIPMENT) shall not be required to
display the TYPE BF or TYPE CF mark on the ACCESSORY itself, the ACCOMPANYING DOCUMENTS,
or on the packaging unless the RISK MANAGEMENT FILE identifies an unacceptable RISK
associated with this exclusion.
201.7.4.2 * Control devices
Addition:
The output control shall have a scale and/or associated indicator showing the relative units of
HIGH FREQUENCY output. The indication shall not be marked in watts unless the indicated
power is delivered with an accuracy of ± 20 % over the total load resistance range specified in
201.7.9.3.1.
The numeral "0" shall not be used unless no HF power in excess of 10 mW is delivered from
an ACTIVE ELECTRODE or BIPOLAR ELECTRODE ACCESSORY in this position.
NOTE The compliance test is the application of subclause 201.12.1.102.
201.7.8.1 * Colours of indicator lights
Replace Table 2 in the general standard with the following Table 201.101:
Table 201.101 – Colours of indicator lights and their meaning
for HF SURGICAL EQUIPMENT
Colour Meaning
Red Warning – immediate response by the
OPERATOR is required, for example, a fault in
the PATIENT circuit
Yellow CUTTING mode
Blue COAGULATION mode
Green Ready for use
Any other colour Meaning other than that of red, yellow, blue or
green
201.7.8.2 * Colours of controls
Addition:
Where operating controls, output terminals, indicator lights, pedals (see 201.12.2) and
pushbuttons of FINGERSWITCHES (see 201.12.2) are associated with a particular HF SURGICAL
MODE, they shall be identified by a consistent, unique colour not in conflict with Table 201.101.
Compliance is checked by inspection.
201.7.9.2.2 Warning and safety notices
Additional subclause:
201.7.9.2.2.101 Additional information in instructions for use
a) * Notes on the application of HF SURGICAL EQUIPMENT. These notes shall draw the attention
of the OPERATOR to certain precautions which are necessary in order to reduce the RISK of
accidental burns. In particular, advice, when appropriate, shall be given on the following:
NEUTRAL ELECTRODE should be reliably attached to a suitably
1) The entire area of the
prepared and appropriate area of the PATIENT's body as defined by the
MANUFACTURER.
2) The PATIENT should not come into contact with metal parts which are earthed or
which have an appreciable capacitance to earth (for example operating table
supports, etc.).
3) Skin-to-skin contact (for example between the arms and body of the PATIENT) should
be avoided, for example by insertion of dry gauze.
4) When HF surgical equipment and physiological monitoring equipment are used
simultaneously on the same PATIENT, any monitoring electrodes should be placed as
far as possible from the surgical electrodes. Needle monitoring electrodes are not
recommended.
In all cases, monitoring systems incorporating HIGH FREQUENCY current limiting
devices are recommended.
– 18 – IEC 60601-2-2:2017 RLV IEC 2017
5) The PATIENT leads should be positioned in such a way that contact with the PATIENT
or other leads is avoided.
Temporarily unused ACTIVE ELECTRODES should be stored in a location that is isolated
from the PATIENT.
6) For surgical procedures where the HF current could flo
...
IEC 60601-2-2 ®
Edition 6.0 2017-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite
ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie
et les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Central Office Tel.: +41 22 919 02 11
3, rue de Varembé Fax: +41 22 919 03 00
CH-1211 Geneva 20 info@iec.ch
Switzerland www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
IEC Catalogue - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
The stand-alone application for consulting the entire The world's leading online dictionary of electronic and
bibliographical information on IEC International Standards, electrical terms containing 20 000 terms and definitions in
Technical Specifications, Technical Reports and other English and French, with equivalent terms in 16 additional
documents. Available for PC, Mac OS, Android Tablets and languages. Also known as the International Electrotechnical
iPad. Vocabulary (IEV) online.
IEC publications search - www.iec.ch/searchpub IEC Glossary - std.iec.ch/glossary
The advanced search enables to find IEC publications by a 65 000 electrotechnical terminology entries in English and
variety of criteria (reference number, text, technical French extracted from the Terms and Definitions clause of
committee,…). It also gives information on projects, replaced IEC publications issued since 2002. Some entries have been
and withdrawn publications. collected from earlier publications of IEC TC 37, 77, 86 and
CISPR.
IEC Just Published - webstore.iec.ch/justpublished
Stay up to date on all new IEC publications. Just Published IEC Customer Service Centre - webstore.iec.ch/csc
details all new publications released. Available online and If you wish to give us your feedback on this publication or
also once a month by email. need further assistance, please contact the Customer Service
Centre: csc@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Catalogue IEC - webstore.iec.ch/catalogue Electropedia - www.electropedia.org
Application autonome pour consulter tous les renseignements
Le premier dictionnaire en ligne de termes électroniques et
bibliographiques sur les Normes internationales,
électriques. Il contient 20 000 termes et définitions en anglais
Spécifications techniques, Rapports techniques et autres
et en français, ainsi que les termes équivalents dans 16
documents de l'IEC. Disponible pour PC, Mac OS, tablettes
langues additionnelles. Egalement appelé Vocabulaire
Android et iPad.
Electrotechnique International (IEV) en ligne.
Recherche de publications IEC - www.iec.ch/searchpub
Glossaire IEC - std.iec.ch/glossary
65 000 entrées terminologiques électrotechniques, en anglais
La recherche avancée permet de trouver des publications IEC
en utilisant différents critères (numéro de référence, texte, et en français, extraites des articles Termes et Définitions des
comité d’études,…). Elle donne aussi des informations sur les publications IEC parues depuis 2002. Plus certaines entrées
projets et les publications remplacées ou retirées. antérieures extraites des publications des CE 37, 77, 86 et
CISPR de l'IEC.
IEC Just Published - webstore.iec.ch/justpublished
Service Clients - webstore.iec.ch/csc
Restez informé sur les nouvelles publications IEC. Just
Published détaille les nouvelles publications parues. Si vous désirez nous donner des commentaires sur cette
Disponible en ligne et aussi une fois par mois par email. publication ou si vous avez des questions contactez-nous:
csc@iec.ch.
IEC 60601-2-2 ®
Edition 6.0 2017-03
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.30 ISBN 978-2-8322-4008-3
– 2 – IEC 60601-2-2:2017 IEC 2017
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 9
201.3 Terms and definitions . 10
201.4 General requirements . 14
201.5 General requirements for testing of ME EQUIPMENT . 15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 15
201.7 ME EQUIPMENT identification, marking and documents . 15
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 20
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 36
201.10 Protection against unwanted and excessive radiation HAZARDS . 36
201.11 Protection against excessive temperatures and other HAZARDS . 36
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 38
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 43
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 44
201.15 Construction of ME EQUIPMENT . 44
201.16 ME SYSTEMS . 49
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 49
202 * ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 49
208 General requirements, tests and guidance for alarm systems in medical
electrical equipment and medical electrical systems . 50
Annexes . 51
Annex AA (informative) Particular guidance and rationale . 52
Annex BB (informative) ELECTROMAGNETIC DISTURBANCES created by HF SURGICAL
EQUIPMENT . 78
Bibliography . 87
Index of defined terms used in this particular standard . 89
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT . 16
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT . 16
Figure 201.103 – Circuit suitable for testing compliance to 201.8.4.101 . 22
Figure 201.104 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENT
CIRCUITS and load between electrodes . 25
Figure 201.105 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENT
CIRCUITS and a load resistance from ACTIVE ELECTRODE to earth . 26
Figure 201.106 – Measurement of HF LEAKAGE CURRENT for HF ISOLATED PATIENT
CIRCUITS . 27
Figure 201.107 – Measurement of HF LEAKAGE CURRENT from a BIPOLAR ACCESSORY . 28
Figure 201.108 – Test apparatus for anchorages of cords of ACTIVE ACCESSORY . 34
Figure 201.109 – Measurement of output power – MONOPOLAR output . 39
Figure 201.110 – Measurement of output power – BIPOLAR output . 40
Figure 201.111 – Method of testing feedback from one active output to another in
simultaneous activation . 43
Figure AA.1 – Examples of various parts of an HF surgical ME SYSTEM . 54
Figure AA.2 – Example of MONOPOLAR method of HF surgery using a NEUTRAL
ELECTRODE . 54
Figure AA.3 – Example of BIPOLAR method of HF surgery . 55
Figure AA.4 – CREST FACTOR vs. peak voltage . 60
Figure AA.5 – Example of PATIENT circuit with NEUTRAL ELECTRODE referenced to earth
at operating frequencies . 64
Figure BB.1 – E-FIELD EMISSIONS test setup . 81
Figure BB.2 – H-FIELD EMISSIONS test setup . 82
Figure BB.3 – Conducted EMISSIONS test setup . 83
Figure BB.4 – Unit ad hoc test . 85
Figure BB.5 – Power cord ad hoc test . 86
Figure BB.6 – ACCESSORY cord ad hoc test . 86
Table 201.101 – Colours of indicator lights and their meaning for HF SURGICAL
EQUIPMENT . 16
Table 201.102 – Maximum output powers in SINGLE FAULT CONDITIONS . 42
Table 201.103 – Test currents by weight range . 46
Table AA.1 – Summary of measured current and durations for 25 TUR procedures . 73
Table AA.2 – Summary of measured currents and durations for general surgical
procedures. 74
Table BB.1 – Worst case EMISSIONS of spark gap type HF SURGICAL EQUIPMENT . 84
Table BB.2 – Worst case EMISSIONS of non-spark gap (modern) HF SURGICAL EQUIPMENT . 84
– 4 – IEC 60601-2-2:2017 IEC 2017
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-2-2 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This sixth edition cancels and replaces the fifth edition published in 2009. This edition
constitutes a technical revision. This edition includes the following significant technical
changes with respect to the previous edition:
– refinement and additions to the defined terms;
– additional separation of the requirements for HF surgical equipment and HF surgical
accessories;
– a new requirement for adult neutral electrodes to be contact quality monitoring neutral
electrodes;
– new requirements for devices that have or use a high current mode.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/1427/FDIS 62D/1442/RVD
Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-2:2017 IEC 2017
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC website under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HIGH FREQUENCY SURGICAL
EQUIPMENT.
This particular standard amends and supplements IEC 60601-1:2005 and Amendment 1:2012,
Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance, hereinafter referred to as the general standard (see 201.1.4).
The requirements are followed by specifications for the relevant tests.
A "Particular guidance and rationale" section giving some explanatory notes, where
appropriate, about the more important requirements is included in Annex AA.
Clauses or subclauses for which there are explanatory notes in Annex AA are marked with an
asterisk (*).
It is considered that a knowledge of the reasons for these requirements will not only facilitate
the proper application of the standard but will, in due course, expedite any revision neces-
sitated by changes in clinical practice or as a result of developments in technology. However,
this annex does not form part of the requirements of this document.
– 8 – IEC 60601-2-2:2017 IEC 2017
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HF
SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as defined in 201.3.224 and 201.3.223.
HF SURGICAL EQUIPMENT having a RATED OUTPUT POWER not exceeding 50 W (for example for
micro-COAGULATION, or for use in dentistry or ophthalmology) is exempt from certain of the
requirements of this particular standard. These exemptions are indicated in the relevant
requirements.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for HF SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as
defined in 201.3.224 and 201.3.223.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2:2014 and IEC 60601-1-8:2006 apply as modified in Clauses 202 and 208
respectively. IEC 60601-1-3, IEC 60601-1-10 and IEC 60601-1-11 do not apply. All other
published collateral standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
______________
The general standard is IEC 60601-1:2005/AMD1:2012, Medical electrical equipment – Part 1: General
requirements for basic safety and essential performance.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this document addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 87.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and tests
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral Standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
– 10 – IEC 60601-2-2:2017 IEC 2017
Addition:
CISPR 11:2015, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurement
IEC 61000-4-3:2006, Electromagnetic compatibility (EMC) – Part 4-3: Testing and
measurement techniques – Radiated, radio-frequency electromagnetic field immunity test
IEC 61000-4-6:2013, Electromagnetic compatibility (EMC) – Part 4-6: Testing and
measurement techniques – Immunity to conducted disturbances, induced by radio-frequency
fields
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1 and the
following apply.
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
Replace NOTE 1 with the following:
NOTE 1 Where the terms “voltage” and “current” are used in this document, they mean the RMS values of an
alternating, direct or composite voltage or current averaged over 1 s unless stated otherwise.
Addition:
201.3.201
ACTIVE ACCESSORY
HF SURGICAL ACCESSORY intended for manipulation by the OPERATOR to produce an effect by
electrical conduction adjacent to the ACTIVE ELECTRODE at the intended site on the PATIENT,
generally comprising an ACTIVE HANDLE, the cord of an ACTIVE ACCESSORY, ACTIVE CONNECTOR
and ACTIVE ELECTRODE
201.3.202
ACTIVE CONNECTOR
part of an ACTIVE ACCESSORY intended for connection to an ACTIVE OUTPUT TERMINAL, which
may include additional terminals for connection of a FINGERSWITCH to a SWITCH SENSOR
201.3.203
ACTIVE ELECTRODE
part of an ACTIVE ACCESSORY extending from the ACTIVE HANDLE to the surgical site and
intended to pass HF current into body tissue
201.3.204
ACTIVE ELECTRODE INSULATION
electrical insulation material affixed to part of an ACTIVE ELECTRODE intended to prevent
unintended injury to PATIENT tissue or the OPERATOR
201.3.205
ACTIVE HANDLE
part of an ACTIVE ACCESSORY intended to be held by the OPERATOR
201.3.206
ACTIVE OUTPUT TERMINAL
part of HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an ACTIVE
ACCESSORY and for delivery of HF current thereto
Note 1 to entry: An ACTIVE CONNECTOR is that which plugs into an ACTIVE OUTPUT TERMINAL.
Note 2 to entry: See Figure AA.1.
201.3.207
*ASSOCIATED EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT other than HF SURGICAL EQUIPMENT that may be electrically
connected to the PATIENT circuit
201.3.208
*BIPOLAR
method of applying HF current to a PATIENT between two or more ACTIVE ELECTRODES without
the need for a separately connected NEUTRAL ELECTRODE (or the need to use the PATIENT’S
body capacitance to earth) in which an effect is intended in tissue near one or more ACTIVE
ELECTRODES
Note 1 to entry: The BIPOLAR method includes devices energizing pairs of ACTIVE ELECTRODES as well as devices
energizing groups of ACTIVE ELECTRODES where the HF current source and return may have different numbers of
electrodes.
Note 2 to entry: See Figure AA.1 and Figure AA.3.
201.3.209
BIPOLAR ACCESSORY
ACTIVE ACCESSORY comprising two or more ACTIVE ELECTRODES on the same support, so
constructed that, when energized, the HF current flows mainly amongst these electrodes
201.3.210
COAGULATION
use of HF current to induce a thermal effect, e.g. to control or prevent bleeding, induce tissue
destruction, or induce tissue shrinkage
Note 1 to entry: COAGULATION may take the form of contact or non-contact COAGULATION.
Note 2 to entry: FULGURATION, desiccation, spray, forced, swift, soft and argon beam (plasma) COAGULATION are
all names of COAGULATION types.
201.3.211
CONTACT QUALITY MONITOR
CQM
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to a
MONITORING NE providing an alarm in the event that NEUTRAL ELECTRODE (NE) contact with the
PATIENT becomes insufficient
Note 1 to entry: CONTACT QUALITY MONITOR is functional only when used with a MONITORING NE.
201.3.212
CONTINUITY MONITOR
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an NE
providing an alarm in the event of electrical discontinuity in the NE cable or its connections
201.3.213
*CREST FACTOR
dimensionless value equal to the peak output voltage divided by the RMS voltage as
HF SURGICAL EQUIPMENT in an open circuit condition
measured at the output of
Note 1 to entry: Specific information on the correct way to make the measurements needed to calculate this value
may be found in Annex AA.
– 12 – IEC 60601-2-2:2017 IEC 2017
201.3.214
*CUTTING
division of body tissue caused by the passage of HIGH FREQUENCY current of high current
density at the ACTIVE ELECTRODE (S)
201.3.215
*EARTH REFERENCED PATIENT CIRCUIT
PATIENT circuit which includes components, such as capacitors, installed to provide a low-
impedance path to earth for HF currents
201.3.216
FINGERSWITCH
device generally included with an ACTIVE ACCESSORY which, when manipulated by the
OPERATOR, enables HF output to be produced and, when released disables HF output
Note 1 to entry: Requirements for similar switches intended to perform functions other than activation of HF output
are under consideration.
201.3.217
*FULGURATION
the use of HF current to produce an effect on a tissue surface by electrical sparks from an
ACTIVE ELECTRODE that is not in physical contact with the tissue
201.3.218
*HEATING FACTOR
a value equal to I × t where I is the MONOPOLAR current in amperes and t is the duration of
the current flow in s
Note 1 to entry: The HEATING FACTOR is expressed as A s (amperes squared seconds).
Note 2 to entry: See subclause 201.15.101.5 in Annex AA for additional information.
201.3.219
*HIGH CURRENT MODE
MONOPOLAR output mode whose INTENDED USE (MAXIMUM OUTPUT CURRENT and maximum DUTY
CYCLE) results in a HEATING FACTOR of greater than 30 A s in any 60 s period
201.3.220
*HIGH FREQUENCY
HF
frequencies less than 5 MHz and generally greater than 200 kHz
201.3.221
HF ISOLATED PATIENT CIRCUIT
HF PATIENT CIRCUIT where there are no components installed to provide a low-impedance path
to earth for HF currents
201.3.222
HF PATIENT CIRCUIT
any electrical circuit which contains one or more PATIENT CONNECTIONS including all conductive
parts of the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT circuits through which HF
current is intended to flow between the ME EQUIPMENT and the PATIENT in NORMAL CONDITION or
SINGLE FAULT CONDITION
201.3.223
HF SURGICAL ACCESSORY
ACCESSORY intended to conduct, supplement or monitor HF energy applied to the PATIENT from
HF SURGICAL EQUIPMENT
Note 1 to entry: HF SURGICAL ACCESSORIES include ACTIVE ACCESSORIES, including cords and connectors for
attachment to HF SURGICAL EQUIPMENT, NEUTRAL ELECTRODES, as well as other ASSOCIATED EQUIPMENT intended for
connection to the HF surgical PATIENT circuit. See Figure AA.1.
Note 2 to entry: Not all accessories used with HF surgical equipment are HF surgical accessories.
201.3.224
HF SURGICAL EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT which generates HIGH FREQUENCY currents intended for the
performance of surgical tasks, such as the CUTTING or COAGULATION of biological tissue by
means of these HIGH FREQUENCY currents
Note 1 to entry: HF SURGICAL EQUIPMENT is also variously known as surgical diathermy, electrosurgical equipment,
electrosurgical generator, RF generator or HF generator.
Note 2 to entry: A footswitch is an example of an associated ACCESSORY that is part of HF SURGICAL EQUIPMENT.
See Figure AA.1.
201.3.225
*HF SURGICAL MODE
any of a number of OPERATOR selectable HF output characteristics intended to provide a
specific effect at a connected ACTIVE ACCESSORY, such as CUTTING, COAGULATION and the like
Note 1 to entry: Each available HF SURGICAL MODE may be provided with an OPERATOR-adjustable output control to
set the desired intensity or speed of the effect.
201.3.226
*MAXIMUM OUTPUT CURRENT
for each available HF SURGICAL MODE, the magnitude of the maximum possible HF output
current during INTENDED USE
201.3.227
*MAXIMUM OUTPUT VOLTAGE
for each available HF SURGICAL MODE, the magnitude of the maximum possible peak HF output
voltage appearing between PATIENT circuit connections
201.3.228
*MONITORING NE
NE intended for use with a CONTACT QUALITY MONITOR
Note 1 to entry: A MONITORING NEUTRAL ELECTRODE is also known as a split plate, dual plate, dual foil electrode or
CQM electrode.
201.3.229
*MONOPOLAR
method of applying HF output current to a PATIENT via an ACTIVE ELECTRODE and returning via a
separate PATIENT-connected NEUTRAL ELECTRODE (or via the PATIENT’S body capacitance to
earth) in which an effect is intended only in tissue at or near the ACTIVE ELECTRODE
Note 1 to entry: See Figures AA.1 and AA.2.
201.3.230
NEUTRAL ELECTRODE
NE
electrode intended to provide an electrical return path for the MONOPOLAR application of HIGH
FREQUENCY current with such a low current density in the PATIENT’S tissue that effects such as
excessive rise in temperature or unwanted burns are avoided
Note 1 to entry: The NEUTRAL ELECTRODE is also known as plate, plate electrode, electrosurgical pad, passive,
return or dispersive electrode.
Note 2 to entry: To keep the current density low enough to prevent unwanted heating, the NEUTRAL ELECTRODE
needs to have a large enough area.
– 14 – IEC 60601-2-2:2017 IEC 2017
Note 3 to entry: A NEUTRAL ELECTRODE is usually in contact with the PATIENT at a location that is separate from the
MONOPOLAR ACTIVE ELECTRODE.
Note 4 to entry: See Figures AA.1 and AA.2.
201.3.231.1
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be
applied with respect to an NE connected to the PATIENT
201.3.231.2
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be applied to
pairs of opposite polarity
201.3.232
RATED LOAD
results in the maximum HF
value of non-reactive load resistance which, when connected
output power from each HF SURGICAL MODE of the HF SURGICAL EQUIPMENT
201.3.233
RATED OUTPUT POWER
for each HF SURGICAL MODE set at its maximum output setting, the power in watts produced
when all ACTIVE OUTPUT TERMINALS which can be activated simultaneously are connected to
their respective RATED LOADS
201.3.234
SWITCH SENSOR
part of HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT which controls activation of HF output
in response to operation of a connected FINGERSWITCH or footswitch
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
Additional subclauses:
201.4.1.101 * Additional conditions for application
The compliance of HF SURGICAL EQUIPMENT to this document and the compliance of HF
SURGICAL ACCESSORIES to this document shall be independent of each other, except where
specifically required by conformance tests or by the MANUFACTURER.
201.4.2.3.101 * Evaluating RISK
MANUFACTURERS shall include, within their RISK ANALYSIS, the potential for their HF SURGICAL
EQUIPMENT and/or HF SURGICAL ACCESSORIES to be used in HIGH CURRENT MODE and the impact
this would have on the heating under the NEUTRAL ELECTRODE (for example, see
201.7.9.2.2.101 f)).
201.4.3 * ESSENTIAL PERFORMANCE
Addition:
The requirements listed in the third hyphen of 201.8.4.101 and in 201.12.4.101 shall be
considered ESSENTIAL PERFORMANCE requirements.
NOTE 101 Please refer to Annex AA.
201.4.7 SINGLE FAULT CONDITION for ME EQUIPMENT
Additional subclause:
201.4.7.101 Specific SINGLE FAULT CONDITIONS
The following SINGLE FAULT CONDITIONS are the subject of specific requirements and tests in
this document:
a) failure in the CONTINUITY MONITOR or CONTACT QUALITY MONITOR which might cause a
unacceptable RISK (see 201.8.4.101);
b) a defect in the output switching circuit resulting in an excessive low-frequency PATIENT
LEAKAGE CURRENT (see 201.8.10.4.101.1);
c) any defect which results in the unwanted energization of the PATIENT circuit (see
201.12.4.2.101);
d) any defect which results in a significant increase in output power relative to the output
setting (see 201.12.4.4.101).
201.4.11 Power input
Replacement of first dash in compliance tests:
– The HF SURGICAL EQUIPMENT shall be operated in the output mode and using the load
which creates the greatest steady state input current. Input current is measured and
compared with markings and the contents of the technical description.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.4 * Other conditions
Addition:
aa) Particular care shall be taken to ensure accuracy and safety during measurement of HF
output. See Annex AA for guidance.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2.8.2 Other power sources
Amendment:
Subclause 7.2.8.2 of the general standard does not apply to ACTIVE OUTPUT TERMINALS or NE
terminals.
201.7.2.10 APPLIED PARTS
Addition:
The relevant symbols required for marking DEFIBRILLATION-PROOF APPLIED PARTS shall be
attached to the front panel, but are not required on the APPLIED PARTS.
– 16 – IEC 60601-2-2:2017 IEC 2017
Connections on the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT for the connection of
NE leads shall be marked with the symbols given in Figures 201.101 and 201.102 as follows:
IEC
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT
IEC
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT
Additional subclause:
201.7.2.10.101 * HF SURGICAL ACCESSORIES
HF SURGICAL ACCESSORIES (excluding HF ASSOCIATED EQUIPMENT) shall not be required to
display the TYPE BF or TYPE CF mark on the ACCESSORY itself, the ACCOMPANYING DOCUMENTS,
or on the packaging unless the RISK MANAGEMENT FILE identifies an unacceptable RISK
associated with this exclusion.
201.7.4.2 * Control devices
Addition:
The output control shall have a scale and/or associated indicator showing the relative units of
HIGH FREQUENCY output. The indication shall not be marked in watts unless the indicated
power is delivered with an accuracy of ± 20 % over the total load resistance range specified in
201.7.9.3.1.
The numeral "0" shall not be used unless no HF power in excess of 10 mW is delivered from
an ACTIVE ELECTRODE or BIPOLAR ACCESSORY in this position.
NOTE The compliance test is the application of subclause 201.12.1.102.
201.7.8.1 * Colours of indicator lights
Replace Table 2 in the general standard with the following Table 201.101:
Table 201.101 – Colours of indicator lights and their meaning
for HF SURGICAL EQUIPMENT
Colour Meaning
Red Warning – immediate response by the
OPERATOR is required, for example, a fault in
the PATIENT circuit
Yellow CUTTING mode
Blue COAGULATION mode
Green Ready for use
Any other colour Meaning other than that of red, yellow, blue or
green
201.7.8.2 * Colours of controls
Addition:
Where operating controls, output terminals, indicator lights, pedals (see 201.12.2) and
pushbuttons of FINGERSWITCHES (see 201.12.2) are associated with a particular HF SURGICAL
MODE, they shall be identified by a consistent, unique colour not in conflict with Table 201.101.
Compliance is checked by inspection.
201.7.9.2.2 Warning and safety notices
Additional subclause:
201.7.9.2.2.101 Additional information in instructions for use
a) * Notes on the application of HF SURGICAL EQUIPMENT. These notes shall draw the attention
of the OPERATOR to certain precau
...
IEC 60601-2-2 ®
Edition 6.1 2023-02
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et
les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.
IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.
IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews. With a subscription you will always have
committee, …). It also gives information on projects, replaced access to up to date content tailored to your needs.
and withdrawn publications.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
The world's leading online dictionary on electrotechnology,
Stay up to date on all new IEC publications. Just Published
containing more than 22 300 terminological entries in English
details all new publications released. Available online and once
and French, with equivalent terms in 19 additional languages.
a month by email.
Also known as the International Electrotechnical Vocabulary
(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc
If you wish to give us your feedback on this publication or need
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.
A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.
Recherche de publications IEC - Découvrez notre puissant moteur de recherche et consultez
webstore.iec.ch/advsearchform gratuitement tous les aperçus des publications. Avec un
La recherche avancée permet de trouver des publications IEC abonnement, vous aurez toujours accès à un contenu à jour
en utilisant différents critères (numéro de référence, texte, adapté à vos besoins.
comité d’études, …). Elle donne aussi des informations sur les
projets et les publications remplacées ou retirées. Electropedia - www.electropedia.org
Le premier dictionnaire d'électrotechnologie en ligne au monde,
IEC Just Published - webstore.iec.ch/justpublished
avec plus de 22 300 articles terminologiques en anglais et en
Restez informé sur les nouvelles publications IEC. Just
français, ainsi que les termes équivalents dans 19 langues
Published détaille les nouvelles publications parues.
additionnelles. Egalement appelé Vocabulaire
Disponible en ligne et une fois par mois par email.
Electrotechnique International (IEV) en ligne.
Service Clients - webstore.iec.ch/csc
Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
IEC Products & Services Portal - products.iec.ch
IEC 60601-2-2 ®
Edition 6.1 2023-02
CONSOLIDATED VERSION
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.30 ISBN 978-2-8322-6542-0
IEC 60601-2-2 ®
Edition 6.1 2023-02
CONSOLIDATED VERSION
REDLINE VERSION
VERSION REDLINE
colour
inside
Medical electrical equipment –
Part 2-2: Particular requirements for the basic safety and essential performance
of high frequency surgical equipment and high frequency surgical accessories
Appareils électromédicaux –
Partie 2-2: Exigences particulières pour la sécurité de base et les performances
essentielles des appareils d'électrochirurgie à courant haute fréquence et des
accessoires d'électrochirurgie à courant haute fréquence
– 2 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
INTRODUCTION to Amendment 1 . 7
201.1 Scope, object and related standards . 8
201.2 Normative references . 9
201.3 Terms and definitions . 10
201.4 General requirements . 14
201.5 General requirements for testing of ME EQUIPMENT . 15
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 16
201.7 ME EQUIPMENT identification, marking and documents . 16
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 21
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 37
201.10 Protection against unwanted and excessive radiation HAZARDS . 37
201.11 Protection against excessive temperatures and other HAZARDS . 38
201.12 Accuracy of controls and instruments and protection against hazardous outputs . 39
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 45
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 46
201.15 Construction of ME EQUIPMENT . 46
201.16 ME SYSTEMS . 51
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS. 51
202 * ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 51
208 General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems . 52
Annexes . 53
Annex AA (informative) Particular guidance and rationale . 54
Annex BB (informative) ELECTROMAGNETIC DISTURBANCES created by HF SURGICAL
EQUIPMENT . 80
Bibliography . 89
Index of defined terms used in this particular standard . 91
Figure 201.101 – Symbol used with an EARTH REFERENCED PATIENT CIRCUIT . 16
Figure 201.102 – Symbol used with a HF ISOLATED PATIENT CIRCUIT . 16
Figure 201.103 – Circuit suitable for testing compliance to 201.8.4.101 . 23
Figure 201.104 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENT
CIRCUITS and load between electrodes . 26
Figure 201.105 – Measurement of HF LEAKAGE CURRENT for EARTH REFERENCED PATIENT
CIRCUITS and a load resistance from ACTIVE ELECTRODE to earth . 27
Figure 201.106 – Measurement of HF LEAKAGE CURRENT for HF ISOLATED PATIENT
CIRCUITS . 28
Figure 201.107 – Measurement of HF LEAKAGE CURRENT from a BIPOLAR ACCESSORY . 29
Figure 201.108 – Test apparatus for anchorages of cords of ACTIVE ACCESSORY . 36
Figure 201.109 – Measurement of output power – MONOPOLAR output . 41
Figure 201.110 – Measurement of output power – BIPOLAR output . 42
IEC 2023
Figure 201.111 – Method of testing feedback from one active output to another in
simultaneous activation . 45
Figure AA.1 – Examples of various parts of an HF surgical ME SYSTEM . 56
Figure AA.2 – Example of MONOPOLAR method of HF surgery using a NEUTRAL
ELECTRODE . 56
Figure AA.3 – Example of BIPOLAR method of HF surgery . 57
Figure AA.4 – CREST FACTOR vs. peak voltage . 62
Figure AA.5 – Example of PATIENT circuit with NEUTRAL ELECTRODE referenced to earth
at operating frequencies . 66
Figure BB.1 – E-FIELD EMISSIONS test setup . 83
Figure BB.2 – H-FIELD EMISSIONS test setup . 84
Figure BB.3 – Conducted EMISSIONS test setup . 85
Figure BB.4 – Unit ad hoc test . 87
Figure BB.5 – Power cord ad hoc test . 88
Figure BB.6 – ACCESSORY cord ad hoc test . 88
Table 201.101 – Colours of indicator lights and their meaning for HF SURGICAL
EQUIPMENT . 17
Table 201.102 – Maximum output powers in SINGLE FAULT CONDITIONS . 44
Table 201.103 – Test currents by weight range . 48
Table AA.1 – Summary of measured current and durations for 25 TUR procedures . 75
Table AA.2 – Summary of measured currents and durations for general surgical
procedures. 76
Table BB.1 – Worst case EMISSIONS of spark gap type HF SURGICAL EQUIPMENT . 86
Table BB.2 – Worst case EMISSIONS of non-spark gap (modern) HF SURGICAL EQUIPMENT . 86
– 4 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
This consolidated version of the official IEC Standard and its amendment has been
prepared for user convenience.
IEC 60601-2-2 edition 6.1 contains the sixth edition (2017-03) [documents 62D/1427/FDIS
and 62D/1442/RVD] and its amendment 1 (2023-02) [documents 62D/2010/FDIS and
62D/2021/RVD].
In this Redline version, a vertical line in the margin shows where the technical content
is modified by amendment 1. Additions are in green text, deletions are in strikethrough
red text. A separate Final version with all changes accepted is available in this
publication.
IEC 2023
International standard IEC 60601-2-2 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC technical committee 62: Electrical equipment in medical
practice.
This sixth edition constitutes a technical revision. This edition includes the following
significant technical changes with respect to the previous edition:
– refinement and additions to the defined terms;
– additional separation of the requirements for HF surgical equipment and HF surgical
accessories;
– a new requirement for adult neutral electrodes to be contact quality monitoring neutral
electrodes;
– new requirements for devices that have or use a high current mode.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
– informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Clause 7 of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
The committee has decided that the contents of the base publication and its amendment will
remain unchanged until the stability date indicated on the IEC web site under webstore.iec.ch
in the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
IEC 2023
INTRODUCTION
The minimum safety requirements specified in this particular standard are considered to
provide for a practical degree of safety in the operation of HIGH FREQUENCY SURGICAL
EQUIPMENT.
This particular standard amends and supplements IEC 60601-1:2005 and Amendment 1:2012,
Medical electrical equipment – Part 1: General requirements for basic safety and essential
performance, hereinafter referred to as the general standard (see 201.1.4).
The requirements are followed by specifications for the relevant tests.
A "Particular guidance and rationale" section giving some explanatory notes, where
appropriate, about the more important requirements is included in Annex AA.
Clauses or subclauses for which there are explanatory notes in Annex AA are marked with an
asterisk (*).
It is considered that a knowledge of the reasons for these requirements will not only facilitate
the proper application of the standard but will, in due course, expedite any revision neces-
sitated by changes in clinical practice or as a result of developments in technology. However,
this annex does not form part of the requirements of this document.
INTRODUCTION to Amendment 1
th
The 6 Edition of IEC 60601-2-2 was published in 2017. This amendment is intended to align
the standard to IEC 60601-1:2005/AMD2:2020. Additionally, this amendment is intended to
address several issues reported from the national committees, including but not limited to:
• requirement for including the length of an accessory in the instructions for use;
HF LEAKAGE CURRENTS;
• clarification of test setup for
• considering modes with high DUTY CYCLES above 45 % in the risk management;
• including text of the interpretation sheet 62D/1703/INF regarding the HIGH CURRENT
MODE to Annex AA.
– 8 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
MEDICAL ELECTRICAL EQUIPMENT –
Part 2-2: Particular requirements for the basic safety and essential
performance of high frequency surgical equipment and
high frequency surgical accessories
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 * Scope
Replacement:
This part of IEC 60601 applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of HF
SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as defined in 201.3.224 and 201.3.223.
HF SURGICAL EQUIPMENT having a RATED OUTPUT POWER not exceeding 50 W (for example for
micro-COAGULATION, or for use in dentistry or ophthalmology) is exempt from certain of the
requirements of this particular standard. These exemptions are indicated in the relevant
requirements.
201.1.2 Object
Replacement:
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for HF SURGICAL EQUIPMENT and HF SURGICAL ACCESSORIES as
defined in 201.3.224 and 201.3.223.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-2:2014 and IEC 60601-1-2:2014/AMD1:2020 and IEC 60601-1-8:2006,
IEC 60601-1-8:2006/AMD1:2012 and IEC 60601-1-8:2006/AMD2:2020, apply as modified in
Clauses 202 and 208 respectively. IEC 60601-1-3, IEC 60601-1-10, IEC 60601-1-11 and
IEC 60601-1-12 do not apply. All other published collateral standards in the IEC 60601-1
series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
______________
The general standard is IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 and IEC 60601-1:2005/AMD2:2020,
Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
IEC 2023
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this document addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
NOTE Informative references are listed in the bibliography beginning on page 89.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and tests
IEC 60601-1-2:2014/AMD1:2020
– 10 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral Standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-8:2006/AMD1:2012
IEC 60601-1-8:2006/AMD2:2020
Addition:
CISPR 11:2015, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurement
CISPR 11:2015/AMD1:2016
CISPR 11:2015/AMD2:2019
IEC 61000-4-3:2006, Electromagnetic compatibility (EMC) – Part 4-3: Testing and
measurement techniques – Radiated, radio-frequency electromagnetic field immunity test
IEC 61000-4-6:2013, Electromagnetic compatibility (EMC) – Part 4-6: Testing and
measurement techniques – Immunity to conducted disturbances, induced by radio-frequency
fields
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1 and the
following apply.
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
Replace NOTE 1 with the following:
NOTE 1 Where the terms “voltage” and “current” are used in this document, they mean the RMS values of an
alternating, direct or composite voltage or current averaged over 1 s unless stated otherwise.
Addition:
201.3.201
ACTIVE ACCESSORY
HF SURGICAL ACCESSORY intended for manipulation by the OPERATOR to produce an effect by
electrical conduction adjacent to the ACTIVE ELECTRODE at the intended site on the PATIENT,
generally comprising an ACTIVE HANDLE, the cord of an ACTIVE ACCESSORY, ACTIVE CONNECTOR
and ACTIVE ELECTRODE
201.3.202
ACTIVE CONNECTOR
part of an ACTIVE ACCESSORY intended for connection to an ACTIVE OUTPUT TERMINAL, which
may include additional terminals for connection of a FINGERSWITCH to a SWITCH SENSOR
201.3.203
ACTIVE ELECTRODE
part of an ACTIVE ACCESSORY extending from the ACTIVE HANDLE to the surgical site and
intended to pass HF current into body tissue
IEC 2023
201.3.204
ACTIVE ELECTRODE INSULATION
electrical insulation material affixed to part of an ACTIVE ELECTRODE intended to prevent
unintended injury to PATIENT tissue or the OPERATOR
201.3.205
ACTIVE HANDLE
part of an ACTIVE ACCESSORY intended to be held by the OPERATOR
201.3.206
ACTIVE OUTPUT TERMINAL
part of HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an ACTIVE
ACCESSORY and for delivery of HF current thereto
Note 1 to entry: An ACTIVE CONNECTOR is that which plugs into an ACTIVE OUTPUT TERMINAL.
Note 2 to entry: See Figure AA.1.
201.3.207
*ASSOCIATED EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT other than HF SURGICAL EQUIPMENT that may be electrically
connected to the PATIENT circuit
201.3.208
*BIPOLAR
method of applying HF current to a PATIENT between two or more ACTIVE ELECTRODES without
the need for a separately connected NEUTRAL ELECTRODE (or the need to use the PATIENT’S
body capacitance to earth) in which an effect is intended in tissue near one or more ACTIVE
ELECTRODES
Note 1 to entry: The BIPOLAR method includes devices energizing pairs of ACTIVE ELECTRODES as well as devices
energizing groups of ACTIVE ELECTRODES where the HF current source and return may have different numbers of
electrodes.
Note 2 to entry: See Figure AA.1 and Figure AA.3.
201.3.209
BIPOLAR ACCESSORY
ACTIVE ACCESSORY comprising two or more ACTIVE ELECTRODES on the same support, so
constructed that, when energized, the HF current flows mainly amongst these electrodes
201.3.210
COAGULATION
use of HF current to induce a thermal effect, e.g. to control or prevent bleeding, induce tissue
destruction, or induce tissue shrinkage
Note 1 to entry: COAGULATION may take the form of contact or non-contact COAGULATION.
Note 2 to entry: FULGURATION, desiccation, spray, forced, swift, soft and argon beam (plasma) COAGULATION are
all names of COAGULATION types.
201.3.211
CONTACT QUALITY MONITOR
CQM
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to a
MONITORING NE providing an alarm in the event that NEUTRAL ELECTRODE (NE) contact with the
PATIENT becomes insufficient
Note 1 to entry: CONTACT QUALITY MONITOR is functional only when used with a MONITORING NE.
– 12 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
201.3.212
CONTINUITY MONITOR
circuit in HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT intended for connection to an NE
providing an alarm in the event of electrical discontinuity in the NE cable or its connections
201.3.213
CREST FACTOR
*
dimensionless value equal to the peak output voltage divided by the RMS voltage as
measured at the output of HF SURGICAL EQUIPMENT in an open circuit condition
Note 1 to entry: Specific information on the correct way to make the measurements needed to calculate this value
may be found in Annex AA.
201.3.214
*CUTTING
division of body tissue caused by the passage of HIGH FREQUENCY current of high current
density at the ACTIVE ELECTRODE (S)
201.3.215
*EARTH REFERENCED PATIENT CIRCUIT
PATIENT circuit which includes components, such as capacitors, installed to provide a low-
impedance path to earth for HF currents
201.3.216
FINGERSWITCH
device generally included with an ACTIVE ACCESSORY which, when manipulated by the
OPERATOR, enables HF output to be produced and, when released disables HF output
Note 1 to entry: Requirements for similar switches intended to perform functions other than activation of HF output
are under consideration.
201.3.217
*FULGURATION
the use of HF current to produce an effect on a tissue surface by electrical sparks from an
ACTIVE ELECTRODE that is not in physical contact with the tissue
201.3.218
*HEATING FACTOR
a value equal to I × t where I is the MONOPOLAR current in amperes and t is the duration of
the current flow in s
Note 1 to entry: The HEATING FACTOR is expressed as A s (amperes squared seconds).
Note 2 to entry: See subclause 201.15.101.5 in Annex AA for additional information.
201.3.219
*HIGH CURRENT MODE
MONOPOLAR output mode whose INTENDED USE (MAXIMUM OUTPUT CURRENT and maximum DUTY
CYCLE) results in a HEATING FACTOR of greater than 30 A s in any 60 s period
201.3.220
*HIGH FREQUENCY
HF
frequencies less than 5 MHz and generally greater than 200 kHz
Note 1 to entry: HIGH FREQUENCY (HF) and radio frequency (RF) are considered as equivalent in the context of this
document as long as the frequency is within the range defined in this definition.
IEC 2023
201.3.221
HF ISOLATED PATIENT CIRCUIT
HF PATIENT CIRCUIT where there are no components installed to provide a low-impedance path
to earth for HF currents
201.3.222
HF PATIENT CIRCUIT
any electrical circuit which contains one or more PATIENT CONNECTIONS including all conductive
parts of the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT circuits through which HF
current is intended to flow between the ME EQUIPMENT and the PATIENT in NORMAL CONDITION or
SINGLE FAULT CONDITION
201.3.223
HF SURGICAL ACCESSORY
ACCESSORY intended to conduct, supplement or monitor HF energy applied to the PATIENT from
HF SURGICAL EQUIPMENT
Note 1 to entry: HF SURGICAL ACCESSORIES include ACTIVE ACCESSORIES, including cords and connectors for
attachment to HF SURGICAL EQUIPMENT, NEUTRAL ELECTRODES, as well as other ASSOCIATED EQUIPMENT intended for
connection to the HF surgical PATIENT circuit. See Figure AA.1.
Note 2 to entry: Not all accessories used with HF surgical equipment are HF surgical accessories.
201.3.224
HF SURGICAL EQUIPMENT
MEDICAL ELECTRICAL EQUIPMENT which generates HIGH FREQUENCY currents intended for the
performance of surgical tasks, such as the CUTTING or COAGULATION of biological tissue by
means of these HIGH FREQUENCY currents
Note 1 to entry: HF SURGICAL EQUIPMENT is also variously known as surgical diathermy, electrosurgical equipment,
electrosurgical generator, RF generator or HF generator.
Note 2 to entry: A footswitch is an example of an associated ACCESSORY that is part of HF SURGICAL EQUIPMENT.
See Figure AA.1.
201.3.225
*HF SURGICAL MODE
any of a number of OPERATOR selectable HF output characteristics intended to provide a
specific effect at a connected ACTIVE ACCESSORY, such as CUTTING, COAGULATION and the like
Note 1 to entry: Each available HF SURGICAL MODE may be provided with an OPERATOR-adjustable output control to
set the desired intensity or speed of the effect.
201.3.226
*MAXIMUM OUTPUT CURRENT
for each available HF SURGICAL MODE, the magnitude of the maximum possible HF output
current during INTENDED USE
201.3.227
*MAXIMUM OUTPUT VOLTAGE
for each available HF SURGICAL MODE, the magnitude of the maximum possible peak HF output
voltage appearing between PATIENT circuit connections
201.3.228
*MONITORING NE
NE intended for use with a CONTACT QUALITY MONITOR
Note 1 to entry: A MONITORING NEUTRAL ELECTRODE is also known as a split plate, dual plate, dual foil electrode or
CQM electrode.
– 14 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
201.3.229
*MONOPOLAR
method of applying HF output current to a PATIENT via an ACTIVE ELECTRODE and returning via a
separate PATIENT-connected NEUTRAL ELECTRODE (or via the PATIENT’S body capacitance to
earth) in which an effect is intended only in tissue at or near the ACTIVE ELECTRODE
Note 1 to entry: See Figures AA.1 and AA.2.
201.3.230
NEUTRAL ELECTRODE
NE
electrode intended to provide an electrical return path for the MONOPOLAR application of HIGH
FREQUENCY current with such a low current density in the PATIENT’S tissue that effects such as
excessive rise in temperature or unwanted burns are avoided
Note 1 to entry: The NEUTRAL ELECTRODE is also known as plate, plate electrode, electrosurgical pad, passive,
return or dispersive electrode.
Note 2 to entry: To keep the current density low enough to prevent unwanted heating, the NEUTRAL ELECTRODE
needs to have a large enough area.
Note 3 to entry: A NEUTRAL ELECTRODE is usually in contact with the PATIENT at a location that is separate from the
MONOPOLAR ACTIVE ELECTRODE.
Note 4 to entry: See Figures AA.1 and AA.2.
201.3.231.1
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be
applied with respect to an NE connected to the PATIENT
201.3.231.2
RATED ACCESSORY VOLTAGE
maximum peak HF output voltage which may be applied to
pairs of opposite polarity
201.3.232
RATED LOAD
value of non-reactive load resistance which, when connected results in the maximum HF
output power from each HF SURGICAL MODE of the HF SURGICAL EQUIPMENT
201.3.233
RATED OUTPUT POWER
for each HF SURGICAL MODE set at its maximum output setting, the power in watts produced
when all ACTIVE OUTPUT TERMINALS which can be activated simultaneously are connected to
their respective RATED LOADS
201.3.234
SWITCH SENSOR
HF SURGICAL EQUIPMENT or ASSOCIATED EQUIPMENT which controls activation of HF output
part of
in response to operation of a connected FINGERSWITCH or footswitch
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
Additional subclauses:
IEC 2023
201.4.1.101 * Additional conditions for application
The compliance of HF SURGICAL EQUIPMENT to this document and the compliance of HF
SURGICAL ACCESSORIES to this document shall be independent of each other, except where
specifically required by conformance tests or by the MANUFACTURER.
201.4.2.3.101 * Evaluating RISK
MANUFACTURERS shall include, within their RISK ANALYSIS, the potential for their HF SURGICAL
EQUIPMENT and/or HF SURGICAL ACCESSORIES to be used in HIGH CURRENT MODE and the impact
this would have on the heating under the NEUTRAL ELECTRODE (for example, see
201.7.9.2.2.101 f)).
Additionally, the impact on the heating under the NEUTRAL ELECTRODE shall be considered
within RISK ANALYSIS for any mode with a duty cycle above 45 % according to its intended use
even if the HEATING FACTOR is below 30 A s in any 60 s interval.
201.4.3 * ESSENTIAL PERFORMANCE
Addition:
The requirements listed in the third hyphen of 201.8.4.101 and in 201.12.4.101 shall be
considered ESSENTIAL PERFORMANCE requirements.
NOTE 101 Please refer to Annex AA.
201.4.7 SINGLE FAULT CONDITION for ME EQUIPMENT
Additional subclause:
201.4.7.101 Specific SINGLE FAULT CONDITIONS
The following SINGLE FAULT CONDITIONS are the subject of specific requirements and tests in
this document:
a) failure in the CONTINUITY MONITOR or CONTACT QUALITY MONITOR which might cause a
unacceptable RISK (see 201.8.4.101);
b) a defect in the output switching circuit resulting in an excessive low-frequency PATIENT
LEAKAGE CURRENT (see 201.8.10.4.101.1);
c) any defect which results in the unwanted energization of the PATIENT circuit (see
201.12.4.2.101);
d) any defect which results in a significant increase in output power relative to the output
setting (see 201.12.4.4.101).
201.4.11 Power input
Replacement of first dash in compliance tests:
– The HF SURGICAL EQUIPMENT shall be operated in the output mode and using the
manner (combination of operating setting, load, etc.) which creates the greatest steady
state input current. Input current is measured and compared with the markings and the
contents of the technical description.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.4 * Other conditions
Addition:
– 16 – IEC 60601-2-2:2017+AMD1:2023 CSV
IEC 2023
aa) Particular care shall be taken to ensure accuracy and safety during measurement of HF
output. See Annex AA for guidance.
201.6 Classification of ME EQUIPMENT and ME SYSTEMS
Clause 6 of the general standard applies.
201.7 ME EQUIPMENT identification, marking and documents
Clause 7 of the general standard applies, except as follows:
201.7.2.8.2 Other power sources
Amendment:
Subclause 7.2.8.2 of the general standard does not apply to ACTIVE OUTPUT TERMINALS or NE
terminals.
201.7.2.10 APPLIED PARTS
Addition:
The relevant symbols required for marking DEFIBRILLATION-PROOF APPLIED PARTS shall be
attached to the front panel, but are not required on the APPLIED PARTS.
Connections on the HF SURGICAL EQUIPMENT and ASSOCIATED EQUIPMENT for the connection of
NE leads shall be marked with the symbols given in Figures 201.101 and 201.102 as follows:
IEC
Figure 201.101 – Symbol used with an EARTH REF
...