IEC 61676:2002/AMD1:2008
(Amendment)Amendment 1 - Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology
Amendment 1 - Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology
General Information
Relations
Standards Content (Sample)
IEC 61676
Edition 1.0 2008-11
INTERNATIONAL
STANDARD
AMENDMENT 1
Medical electrical equipment – Dosimetric instruments used for non-invasive
measurement of X-ray tube voltage in diagnostic radiology
IEC 61676:2002/A1:2008(E)
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester.
If you have any questions about IEC copyright or have an enquiry about obtaining additional rights to this publication,
please contact the address below or your local IEC member National Committee for further information.
IEC Central Office
3, rue de Varembé
CH-1211 Geneva 20
Switzerland
Email: inmail@iec.ch
Web: www.iec.ch
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.
About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigenda or an amendment might have been published.
ƒ Catalogue of IEC publications: www.iec.ch/searchpub
The IEC on-line Catalogue enables you to search by a variety of criteria (reference number, text, technical committee,…).
It also gives information on projects, withdrawn and replaced publications.
ƒ IEC Just Published: www.iec.ch/online_news/justpub
Stay up to date on all new IEC publications. Just Published details twice a month all new publications released. Available
on-line and also by email.
ƒ Electropedia: www.electropedia.org
The world's leading online dictionary of electronic and electrical terms containing more than 20 000 terms and definitions
in English and French, with equivalent terms in additional languages. Also known as the International Electrotechnical
Vocabulary online.
ƒ Customer Service Centre: www.iec.ch/webstore/custserv
If you wish to give us your feedback on this publication or need further assistance, please visit the Customer Service
Centre FAQ or contact us:
Email: csc@iec.ch
Tel.: +41 22 919 02 11
Fax: +41 22 919 03 00
IEC 61676
Edition 1.0 2008-11
INTERNATIONAL
STANDARD
AMENDMENT 1
Medical electrical equipment – Dosimetric instruments used for non-invasive
measurement of X-ray tube voltage in diagnostic radiology
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
PRICE CODE
D
ICS 11.040.50; 11.040.55 ISBN 978-2-88910-551-9
– 2 – 61676 Amend. 1 © IEC:2008
FOREWORD
This amendment has been prepared by subcommittee 62C, Equipment for radiotherapy,
nuclear medicine and radiation dosimetry, of IEC technical committee 62, Electrical equipment
in medical practice.
The text of this amendment is based on the following documents:
Enquiry draft Report on voting
62C/445/CDV 62C/452/RVC
Full information on the voting for the approval of this amendment can be found in the report
on voting indicated in the above table.
NOTE In this amendment, a new influence quantity “Additional tungsten filtration (tube aging)” has been
introduced.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
______________
4.3.5 Limits of variation
Replace the existing Table 2 by the following:
61676 Amend. 1 © IEC:2008 – 3 –
Table 2 – Minimum RATED RANGE OF USE, REFERENCE CONDITIONS, STANDARD TEST CONDITIONS,
LIMITS OF VARIATION (± L) and INTRINSIC ERROR (E) over the EFFECTIVE RANGE of use, for the
pertaining INFLUENCE QUANTITY
INFLUENCE Minimum RATED RANGE of REFERENCE STANDARD TEST ± E ± L Sub-
QUANTITY use CONDITIONS CONDITIONS kV % clause
Voltage waveform
and frequency:
Diagnostic 2,0
Constant potential,
Constant Constant potential, ripple
2-, 6-, 12-pulse and medium
4.4.2
potential less than 4 %
frequency
a
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.