Cores made of soft magnetic materials - Measuring methods - Part 3: Magnetic properties at high excitation level

IEC 62044-3:2023 specifies measuring methods for power loss and amplitude permeability of magnetic cores forming the closed magnetic circuits intended for use at high excitation levels in inductors, chokes, transformers and similar devices for power electronics applications. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: addition of Annex F and Annex G.

Noyaux en matériaux magnétiques doux - Méthodes de mesure - Partie 3: Propriétés magnétiques à niveau élevé d'excitation

L'IEC 62044-3:2023 spécifie les méthodes de mesure de la perte de puissance et de la perméabilité d'amplitude des noyaux magnétiques qui forment les circuits magnétiques fermés destinés à être utilisés à des niveaux élevés d'excitation dans les bobines d'inductance, les bobines d'arrêt, les transformateurs et les dispositifs similaires pour les applications d'électronique de puissance. Cette deuxième édition annule et remplace la première édition parue en 2000. Cette édition constitue une révision technique. Cette édition inclut les modifications techniques majeures suivantes par rapport à l'édition précédente: ajout de l'Annexe F et de l'Annexe G.

General Information

Status
Published
Publication Date
06-Jul-2023
Current Stage
PPUB - Publication issued
Start Date
07-Jul-2023
Completion Date
03-Aug-2023
Ref Project

Relations

Overview

IEC 62044-3:2023 - "Cores made of soft magnetic materials – Measuring methods – Part 3: Magnetic properties at high excitation level" - specifies standardized measuring methods for power loss and amplitude permeability of magnetic cores used in closed magnetic circuits operating at high excitation levels. It applies to cores for inductors, chokes, transformers and similar devices used in power electronics. The 2023 edition is a technical revision of the 2000 edition and adds new material (Annex F - normative pulse methods, and Annex G - pulse circuit examples).

Keywords: IEC 62044-3, soft magnetic materials, magnetic cores, power loss measurement, amplitude permeability, high excitation level, pulse measurements, power electronics.

Key topics and technical requirements

  • Scope and definitions: Terms, symbols and specimen requirements for repeatable high-excitation measurements.
  • General precautions: Reproducibility of magnetic state, core mounting, and coil winding practices for accurate results.
  • Measuring amplitude permeability: Purpose, measurement principles, circuits and calculation methods for effective amplitude permeability under large excitation.
  • Power loss measurement methods: Multiple methods and principles are covered, including:
    • Root-mean-square (RMS) based methods
    • Multiplying (V-A-W) methods and impedance analyzer techniques
    • Reflection methods
    • Calorimetric techniques (thermal measurements)
  • Pulse and high‑rate excitation: Annex F (normative) defines measurement methods and procedures for magnetic properties under pulse conditions (e.g., pulse inductance factor, magnetizing current non-linearity). Annex G provides example circuits for pulse measurements.
  • Measurement accuracy and reporting: Required test-report contents, information to be stated, and guidance on equipment, coils and specimen handling.

Practical applications

  • Validating core material performance under high flux or pulsed excitation typical in modern power converters and switching supplies.
  • Comparing power loss and effective permeability across core materials (ferrites, powders, etc.) for high-stress applications.
  • Supporting design decisions for inductors, chokes and transformers in DC-DC converters, inverters and motor drives.
  • Establishing quality control and production testing procedures for manufacturers of magnetic cores and components.
  • Enabling R&D and materials characterization for improved efficiency and thermal management in power electronics.

Who should use this standard

  • Magnetic-core manufacturers and material suppliers
  • Power electronics designers and component engineers
  • Test laboratories and certification bodies
  • R&D and materials characterization teams
  • Quality assurance and standards compliance engineers

Related standards

  • Other parts of the IEC 62044 series and IEC standards on magnetic materials and magnetic component testing (consult IEC resources for latest cross-references).

For procurement, compliance or lab implementation, reference IEC 62044-3:2023 directly to ensure correct test circuits, procedures and reporting requirements.

Standard
IEC 62044-3:2023 RLV - Cores made of soft magnetic materials - Measuring methods - Part 3: Magnetic properties at high excitation level Released:7/7/2023 Isbn:9782832272244
English language
143 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
IEC 62044-3:2023 - Cores made of soft magnetic materials - Measuring methods - Part 3: Magnetic properties at high excitation level Released:7/7/2023
English and French language
93 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


IEC 62044-3 ®
Edition 2.0 2023-07
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Cores made of soft magnetic materials – Measuring methods –
Part 3: Magnetic properties at high excitation level

All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.

IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews. With a subscription you will always have
committee, …). It also gives information on projects, replaced access to up to date content tailored to your needs.
and withdrawn publications.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
The world's leading online dictionary on electrotechnology,
Stay up to date on all new IEC publications. Just Published
containing more than 22 300 terminological entries in English
details all new publications released. Available online and once
and French, with equivalent terms in 19 additional languages.
a month by email.
Also known as the International Electrotechnical Vocabulary

(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc

If you wish to give us your feedback on this publication or need
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
IEC 62044-3 ®
Edition 2.0 2023-07
REDLINE VERSION
INTERNATIONAL
STANDARD
colour
inside
Cores made of soft magnetic materials – Measuring methods –
Part 3: Magnetic properties at high excitation level
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
ICS 29.030, 29.100.10 ISBN 978-2-8322-7224-4

– 2 – IEC 62044-3:2023 RLV © IEC 2023
CONTENTS
FOREWORD . 5
INTRODUCTION . 2
1 Scope . 8
2 Normative references . 8
3 Terms, definitions and symbols. 8
3.1 Terms and definitions . 9
3.2 Symbols . 9
4 General precautions requirements for measurements at high excitation level . 13
4.1 General statements . 13
4.1.1 Relation to practice . 13
4.1.2 Core effective parameters and material properties . 13
4.1.3 Reproducibility of the magnetic state . 14
4.2 Measuring coil . 14
4.2.1 General . 14
4.2.2 Number of turns . 14
4.2.3 Single winding and double winding . 15
4.3 Mounting of cores consisting of more than one part . 15
4.4 Measuring equipment . 16
5 Specimens . 18
6 Measuring procedures . 18
6.1 General procedure . 18
6.2 Measuring method for the (effective) amplitude permeability . 20
6.2.1 Purpose . 20
6.2.2 Principle of the measurement . 20
6.2.3 Circuit and equipment . 20
6.2.4 Measuring procedure . 20
6.2.5 Calculation . 20
6.3 Measuring methods for the power loss . 21
6.3.1 Purpose . 21
6.3.2 Methods and principles of the measurements . 21
7 Information to be stated . 24
8 Test report . 25
Annex A (informative) Basic circuits and related equipment for the measurement of
amplitude permeability . 26
Annex B (informative) Root-mean-square method for the measurement of power loss –
Example of a circuit and related procedure . 29
B.1 Method of measurement . 29
B.2 Measuring coil . 29
B.3 Measuring equipment . 30
B.4 Measuring procedure . 30
B.5 Pulse measurement and accuracy . 31
Annex C (informative) Multiplying methods for the measurement of power loss – Basic
circuits and related measurement procedures . 32
C.1 Basic circuits . 32
C.2 Requirements . 34
C.3 Measuring coil . 34

C.4 Accuracy . 34
C.5 V-A-W (volt-ampere-watt) meter method . 34
C.6 Impedance analyzer method . 35
C.7 Digitizing method . 35
C.8 Vector spectrum method . 35
C.9 Cross-power method . 36
Annex D (informative) Reflection method for the measurement of power loss – Basic
circuit and related measurement procedures . 37
D.1 Basic circuit . 37
D.2 Requirements . 37
D.3 Measuring coil . 37
D.4 Measuring procedure and accuracy . 38
Annex E (informative) Calorimetric measurement methods for the measurement of
power loss . 39
E.1 Basic circuit . 39
E.2 Requirements . 40
E.3 Measuring coil . 40
E.4 Accuracy . 40
E.5 Measurements at thermal equilibrium . 40
E.5.1 General . 40
E.5.2 Measurement across calibrated thermal resistance . 40
E.5.3 Measurement by matching the temperature rise in the core and resistor . 41
E.6 Measurements at non-thermal equilibrium . 41
Annex F (normative) Magnetic properties under pulse condition . 42
F.1 Object . 42
F.2 Measurement methods . 42
F.3 Principle of the methods . 42
F.4 Specimens . 42
F.5 Measuring coil . 42
F.6 Measuring equipment . 43
F.7 Measuring procedure . 44
F.7.1 General . 44
F.7.2 Measurement of pulse inductance factor and magnetizing current . 45
F.7.3 Measurement of the non-linearity of the magnetizing current . 46
F.8 Calculation . 47
Annex G (informative) Examples of circuits for pulse measurements . 49
Bibliography . 50

Figure 1 – Pulse excitation without biasing field . 10
Figure 2 – Pulse excitation with biasing field . 11
Figure A.1 – Basic circuits for the measurement of amplitude permeability . 28
Figure B.1 – Example of a measuring circuit for the RMS method . 29
Figure C.1 – Basic circuits for multiplying methods . 33
Figure D.1 – Basic circuit . 37
Figure E.1 – Basic circuit and related measurement procedures – Measurement set-up . 39
Figure F.1 – Voltage pulse parameters . 45
Figure F.2 – Typical measuring waveforms . 46

– 4 – IEC 62044-3:2023 RLV © IEC 2023
Figure F.3 – Non-linearity of magnetizing current . 47
Figure G.1 – Measurement without bias and with single pulses . 49
Figure G.2 – Measurement with bias and with repeated pulses . 49

Table 1 – Some multiplying methods and related domains of excitation waveforms,
acquisition, processing . 22

INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
CORES MADE OF SOFT MAGNETIC MATERIALS –
MEASURING METHODS –
Part 3: Magnetic properties at high excitation level

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
This redline version of the official IEC Standard allows the user to identify the changes
made to the previous edition IEC 62044-3:200. A vertical bar appears in the margin
wherever a change has been made. Additions are in green text, deletions are in
strikethrough red text.
– 6 – IEC 62044-3:2023 RLV © IEC 2023
IEC 62044-3 has been prepared by IEC technical committee 51: Magnetic components, ferrite
and magnetic powder materials. It is an International Standard.
This second edition cancels and replaces the first edition published in 2000. This edition
constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) addition of Annex F and Annex G.
The text of this International Standard is based on the following documents:
Draft Report on voting
51/1426/CDV 51/1439/RVC
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
A list of all parts in the IEC 62044 series, published under the general title Cores made of soft
magnetic materials – Measuring methods, can be found on the IEC website.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
IMPORTANT – The "colour inside" logo on the cover page of this document indicates that it
contains colours which are considered to be useful for the correct understanding of its
contents. Users should therefore print this document using a colour printer.

INTRODUCTION
IEC 62044, under the general title Cores made of soft magnetic materials – Measuring methods,
includes the following parts:
IEC 62044-1: Generic specification
IEC 62044-2: Magnetic properties at low excitation level
IEC 62044-3: Magnetic properties at high excitation level

– 8 – IEC 62044-3:2023 RLV © IEC 2023
CORES MADE OF SOFT MAGNETIC MATERIALS –
MEASURING METHODS –
Part 3: Magnetic properties at high excitation level

1 Scope
This part of IEC 62044 provides specifies measuring methods for power loss and amplitude
permeability of magnetic cores forming the closed magnetic circuits intended for use at high
excitation levels in inductors, chokes, transformers and similar devices for power electronics
applications.
The methods given in this document can cover the measurement of magnetic properties for
frequencies ranging practically from direct current to 10 MHz, and even possibly higher, for the
calorimetric and reflection methods. The applicability of the individual methods to specific
frequency ranges is dependent on the level of accuracy that is to be obtained.
The methods in this document are basically the most suitable for sine-wave excitations. Other
periodic waveforms can also be used; however, adequate accuracy can only be obtained if the
measuring circuitry and instruments used are able to handle and process the amplitudes and
phases of the signals involved within the frequency spectrum corresponding to the given
induction magnetic flux density and field strength waveforms with only slightly degraded
accuracy.
NOTE It may can be necessary for some magnetically soft metallic materials to follow specific general principles,
customary for these materials, related to the preparation of specimens and prescribed specified calculations. These
principles are formulated in IEC 60404-8-6.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies.
For undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 60050(221):1990, International Electrotechnical Vocabulary (IEV) – Chapter 221: Magnetic
materials and components
Amendment 1 (1993)
Amendment 2 (1999)
IEC 60205:1966, Calculation of the effective parameters of magnetic piece parts
IEC 60367-1:1982, Cores for inductors and transformers for telecommunications – Part 1:
Measuring methods
IEC 60401:1993, Ferrite materials – Guide on the format of data appearing in manufacturers’
catalogues of transformer and inductor cores
IEC 60404-8-6:1999, Magnetic materials – Part 8-6: Specifications for individual materials –
Soft magnetic metallic materials
IEC 61332:1995, Soft ferrite material classification
IEC 62044-1:2002, Cores made of soft magnetic materials – Measuring methods – Part 1:
Generic specification
3 Terms, definitions and symbols
3.1 Terms and definitions
For the purposes of this document, the following terms and definitions apply in addition to those
of IEC 60050(221).
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp
3.1.1
(effective) amplitude permeability (symbols: amplitude permeability: µ ,
a
effective amplitude permeability: µ )
ea
effective amplitude permeability
μ
ea
magnetic permeability obtained from the peak value of the effective magnetic induction flux
ˆ
ˆ
density, , and the peak value of the effective magnetic field strength, H , at the stated value
B
e e
of either, when the magnetic induction flux density and magnetic field vary periodically with time
and with an average of zero, and the material is initially in a specified neutralized demagnetized
state
NOTE 1 This definition differs from that of IEC 60050 [221-03-07].
NOTE 2 Two amplitude permeabilities are in common use, namely:
– that in which the peak values apply to the actual waveforms of the induction and field strength,
– that in which the peak values apply to the fundamental components of waveforms of the induction
and the field strength.
NOTE 3 The induction and the field strength and, consequently, the amplitude permeability may even be quasi-static
quantities, provided the core is cyclically magnetized and no excursion of the B-H curve appears.
3.1.2
maximum (effective) amplitude permeability
μ
ea max
ˆ
maximum value of the (effective) amplitude permeability when the amplitude of excitation ( B
e
ˆ
H
or ) is varied
e
NOTE This definition differs from that of IEC 60050 [221-03-10].
3.1.3
excitation
either induction magnetic flux density or field strength for which the waveform and amplitude
both remain within the specified tolerance
Note 1 to entry: When the induction magnetic flux density (field strength) mode of excitation is chosen, the resultant
waveform of field strength (induction magnetic flux density) may can be distorted with respect to the excitation
waveform due to the non-linear behaviour of the magnetic material.
3.1.4
high excitation level
excitation at which the permeability depends on excitation amplitude (particularly at low
frequencies) and/or at which the power loss results in a noticeable temperature rise (particularly
at high frequencies), or both
– 10 – IEC 62044-3:2023 RLV © IEC 2023
3.1.5
sinusoidal excitation
excitation of harmonic content of less than 1 %
3.1.5
exciting winding
winding of measuring coil to which the exciting voltage is applied or through which the exciting
current is flowing
3.1.6
voltage sensing winding
unloaded winding of a measuring coil across which the electromotive force induced by the
excitation may can be determined
3.1.7
measuring winding
winding, usually secondary, loaded or unloaded, which can be used for measurement apart
from the exciting and/or voltage sensing winding, or both
3.1.8
power loss
power absorbed by the core
3.1.9
pulse excitation without biasing field
excitation in which a core is energized by a voltage pulse, from a remanent flux density to a
higher value of flux density in the same direction, and in which the core recovers to the same
remanent flux density
Note 1 to entry: The excursion in the B-H plane associated with such a pulse is shown in Figure 1.

Figure 1 – Pulse excitation without biasing field
Note 2 to entry: If the back-e.m.f. during recovery is limited only by the time constant of the test circuit, then the
magnetizing current decays exponentially.
Alternatively the back-e.m.f. can be limited to a constant value, for example by returning the energy via a secondary
winding to a voltage source; then the magnetizing current decay is approximately linear. The latter method can
prevent excessively high back-e.m.f. and high rates of change of flux. The distinction is mainly relevant to loss
measurements.
3.1.10
pulse excitation with biasing field
excitation in which a core is energized by a voltage pulse, from a value of the flux density
determined by a biasing field to a flux density in the opposite direction, and in which the core
recovers to the same value determined by the biasing field
Note 1 to entry: The excursion in the B-H plane associated with such a pulse is shown in Figure 2.

Figure 2 – Pulse excitation with biasing field
Note 2 to entry: See Note to entry 2 of 3.1.9.
3.1.11
pulse permeability
μ
p
relative permeability obtained from the change of flux density and the corresponding change of
the field strength when either quantity is varying in an arbitrary form between stated limits:
1 ΔB
μ ⋅
p
μ ΔH
SEE: Figure 1 and Figure 2.
Note 1 to entry: The value of the pulse permeability depends strongly on the limits of the flux density or field strength
excursions; it is not necessary for these limits to be symmetrical with respect to zero.
Note 2 to entry: Often pulse permeability refers to the special case of square voltage pulses being applied to an
exciting winding; the flux density waveform is then approximately triangular.
3.1.12
pulse amplitude
U
m
maximum instantaneous value which an ideal voltage pulse would have with respect to the
steady value of the voltage between pulses
Note 1 to entry: An ideal pulse is derived from an actual voltage pulse by ignoring unwanted or non-pertinent
phenomena such as overshoot (see Figure 1).
=
– 12 – IEC 62044-3:2023 RLV © IEC 2023
3.1.13
pulse duration
t
d
time interval during which the instantaneous value of the pulse exceeds 50 % of the pulse
amplitude
U
m
L =
p
Δi t
md
where
L
p is the pulse inductance
Δi is the total change in i during the pulse
m m
SEE: Figure F.1.
Note 1 to entry: For unidirectional drive pulses Δi = î .
m m
3.1.14
pulse inductance factor
A
LP
pulse inductance divided by the square of the number of turns of the test coil
L
p
A =
LP
N
3.1.15
voltage-time product limit
(U·t)
lim
specified limit of the product of the amplitude of a voltage pulse and the time elapsed from the
start of the pulse
Note 1 to entry: Within this limit the non-linearity of the magnetizing current through the measuring coil placed on
the core should not exceed a specified value.
3.1.16
non-linearity (with time)
ratio of the actual instantaneous value of a characteristic at a time t to the value reached by the
extrapolated linear portion of its graph versus time, at the same instant
SEE: Figure F.3.
3.1.17
pulse repetition rate
frequency of recurrence of the pulses in a periodic sequence of pulses
3.2 Symbols
All the formulae in this document use basic SI units. When multiples or sub-multiples are used,
the appropriate power of 10 shall be introduced.
A effective cross-sectional area of the core
e
ˆ
B peak value of the effective induction magnetic flux density in the core
e
f frequency
ˆ
H peak value of the effective magnetic field strength in the core
e
l effective magnetic path length of the core
e
L inductance
i instantaneous value of the current
I current
N number of turns of winding of the measuring coil
P power loss in the core
Q Q quality factor of the core for a given frequency
c
R resistance
t time
T temperature
u instantaneous value of the voltage
U voltage
effective volume of the core
V
e
δ relative error, deviation, etc.
∆ absolute error, deviation, etc.
µ (effective) amplitude permeability
ea
−7
µ magnetic constant permeability of vacuum: approximately 4π ×10 H/m
π the number 3,14159.
ϕ phase shift
ω angular frequency = 2πf
NOTE 1 The additional subscript, upper script, etc., gives a more specific meaning to the given symbol.
NOTE 2 Symbols which are used sporadically are defined in the place where they appear in the text.
NOTE 3 Effective parameters, such as effective magnetic path length, l , effective cross-sectional area, A , and
e e
effective volume of the core, V , are calculated in accordance with IEC 60205.
e
NOTE 4 In the further text of this standard, the terms induction and field strength stand for the shortened terms
magnetic induction and magnetic field strength. In the text of this document, the term flux density stands for the
shortened term of magnetic flux density.
4 General precautions requirements for measurements at high excitation level
4.1 General statements
4.1.1 Relation to practice
The measuring conditions, methods and procedures shall be chosen in such a way that the
measured results are suitable for predicting the performance of the core under practical
circumstances. This does not imply that all these stipulations, especially those related to the
excitation waveforms, have to correspond to terms encountered in practice.
4.1.2 Core effective parameters and material properties
Since the core is in general of non-uniform cross-section and generally has non-uniformly
distributed windings along the core path, the measurement does not yield the amplitude
permeability and the power loss of the material, but the effective values of these parameters
ˆ
appropriate to the effective induction magnetic flux density B and the effective field strength
e
ˆ
in the core.
H
e
For the measurement of the amplitude permeability and the power loss of the material, the core
shall have a ring or toroidal shape in which the ratio of the outer to the inner diameter should

– 14 – IEC 62044-3:2023 RLV © IEC 2023
not be greater than 1,4 and should have windings distributed uniformly, close to the core, of
inductive coupling coefficient practically equal to unity.
4.1.3 Reproducibility of the magnetic state
To obliterate various remanence and time effects in the core material, the measurement shall
be made at a well-defined and reproducible magnetic state.
Any measurement under specified excitation, unless otherwise stated, is to be made at the time
t = t + ∆t after the magnetic conditioning start; t is the time period within which the magnetic
m c c
conditioning is completed and, whereupon, the specified excitation is set; ∆t is the time period
during which the core is kept stable under the excitation being set.
4.2 Measuring coil
4.2.1 General
Normally, a measuring coil will be used, but in principle any coaxial line, cavity or other suitable
device providing the necessary interaction between the magnetic material and the
electromagnetic signal, may also be used.
For measurement on toroid using coils, the turns of the measuring coil shall be distributed in
such a way as to keep both the stray capacitance and the stray field as low as necessary for
sufficiently accurate measurement.
For measurements made on cores assembled around a coil, the shape of the measuring coil
shall correspond to that of the coils used for normal application of the core and its influence on
the variation of the inductance to be measured shall be negligible.
Unless otherwise specified, the test coil complete with coil former or encapsulation, or both,
shall be positioned in a coaxial way to the limb which it embraces, and the side of the coil at
which the start of the winding is located shall be lightly pressed into contact with the core at
one end of this limb as follows:
• for a symmetrical core, the coil assembly shall contact the core at either one end or the
other;
• for a core with an air-gap that is asymmetrical because the gap has been made by grinding
away material on the leg of one core half and not the other, the coil assembly shall make
end contact to the half of the core that has not been ground away to create the gap.
One of the coil faces shall be marked so as to define its orientation. The coil shall be kept in
the defined position during the whole measurement in order to obtain the maximum
reproducibility of the measurement.
4.2.2 Number of turns
The number of turns shall be specified for each winding in relation to the measuring conditions,
the equipment used and the accuracy to be obtained. The windings shall be wound as close to
the core as possible, to make the coupling (magnetic flux linkage) coefficients between the
measuring coil windings and the core and between the windings of measuring coil, as close to
100 % as possible.
The resistance, self-capacitance and inter-winding capacitance of windings should be as low
as possible to make the related errors negligible.
In the case of ring or toroidal cores, the turns shall be distributed evenly around the core
circumference.
The connectors, primarily of exciting winding, should consist of insulated strands, if this is
necessary for measurements at high frequencies.
NOTE When winding a sharp-edge core, care should be taken to ensure that the wire insulation is not ruptured
and, in the case of stranded wire, strands are not broken.
When winding a sharp-edge core, the wire insulation should not rupture and, in the case of
stranded wire, the strands should not break.
4.2.3 Single winding and double winding
The use of a single winding both for excitation and voltage sensing is recommended if
– the coupling between the exciting winding and the voltage sensing winding is so reduced
that it results in a non-negligible error in the determination of the measuring induction
magnetic flux density B in the core;
– the inter-winding capacitance is too high;
– there is no measuring circuitry contra-indication against the direct connection of the exciting
winding to the input(s) of measuring instruments.
NOTE When a single winding is used, it is recommended that its resistance be made as low
as possible to make the winding ohmic power loss negligible compared to the power loss in the
core.
The use of separate exciting and voltage sensing windings (double winding) is recommended
if, for whatever reason, the exciting winding should be is galvanically separated from the voltage
and the current measuring instruments, for example, to avoid a floating or DC connection to
their inputs.
NOTE 1 When the exciting and voltage sensing windings are used, it is critical to make their
magnetic coupling coefficient shall be as close to 100 % as possible.
NOTE 2 When the voltage needed for calculation of the induction in the core is measured across the voltage sensing
winding then only the power loss in the core is determined with the exclusion of the ohmic power loss in the current-
carrying (exciting) winding.
The voltage necessary for calculation of the magnetic flux density in the core is typically
measured across a voltage sensing winding that is separate from the current-carrying (exciting)
winding. When measuring core losses, ohmic losses in the voltage sensing winding do not affect
the calculation, but ohmic losses in the current-carrying (exciting) winding shall be excluded
from the core loss calculation.
NOTE 3 The use of two windings is recommended at more than 200 kHz.
4.3 Mounting of cores consisting of more than one part
A ferrite core set consisting of more than one part and which is to be assembled around the
measuring coil, shall be held together with glue, tape or a clamping device throughout the
measurement.
Whichever method is used to join the core parts together, it shall have the following
characteristics:
• distribution of the joining force uniformly over the mating surfaces, without the introduction
of bending stresses in the core;
• holding of all the core parts rigidly and without changing the position with regard to each
other;
– 16 – IEC 62044-3:2023 RLV © IEC 2023
• when a specified clamping method is used, an initial over-force of about 10 % shall be
applied when the core is closed, in order to break down fine irregularities between the
cleaned mating surfaces. Next, the specified clamping force ±5 % shall be applied;
• keeping the joining force constant within ±1 % during all measuring operations within all
measuring conditions, including the full specified temperature range.
The mounting of such cores shall be carried out in accordance with the following instructions.
The mating surface shall be inspected for damage and cleanness. Damaged cores shall not be
used. The mating surface shall be cleaned by non-abrasive means, for example, by rubbing
gently on a dry washing-leather. Next, the mating surfaces shall be degreased if they have to
be glued. Dust particles shall be blown off with clean dry compressed air. The mating surfaces
shall never be touched with bare fingers. The core parts shall then be assembled around the
measuring coil, the latter being locked in position with respect to the core by suitable means,
for example, a foam-washer. The core parts are centered and glued or placed in a clamping
device. The glue, if used, shall be spread evenly on the mating surface to form a film as thin as
possible and then properly hardened.
In the case where the clamping device is used, the clamping force specified in the relevant
specification shall be applied. The glued, taped or clamped cores shall relax under the specified
conditions (see IEC 62044-1:2002, Clause 3 of IEC 60367-1) for a long enough time sufficient
to allow any variation of stress effects, due to clamping, gluing or taping, to become negligible.
4.4 Measuring equipment
4.4.1 Any suitable measuring equipment may be used. Examples of appropriate circuits are
given in Annex A to Annex E.
In addition to any requirement specified for the particular method and/or measuring circuit, or
both, used, the following general requirements shall be met.
4.4.2 To ensure the induction magnetic flux density (field strength) mode of excitation, the
output impedance of the exciting source shall be low (high) compared with the series impedance
of the exciting winding of the measuring coil assembled with the core under test and the current
sensing resistor.
4.4.3 When the sinusoidal waveform of excitation is specified, the total harmonic content of
the excitation source shall be less than 1 %. When square pulses are specified, the relevant
requirements of clause 16 of IEC 60367-1 Annex F shall be met.
4.4.4 During the period of measurement, the excitation amplitude variations shall not exceed
±0,05 % and the frequency stability shall be adequate for the measuring method and the
equipment used.
4.4.5 The frequency range of voltmeters and other voltage sensing instruments shall include
all harmonics of the measured voltage having amplitudes of 1 % or more of their fundamentals.
This frequency range shall be specified in the relevant instrument specification.
4.4.6 The voltmeters and other voltage sensing instruments used shall be high-impedance
instruments, the connection of which will have only a negligible effect on the measuring circuit,
especially at high frequencies. The probes of a high-input resistance and a low-input capa-
citance can reduce the load effects.
4.4.7 The accuracy of the voltmeters and/or voltage sensing instruments, determined for
the calibrating sinusoidal waveform, shall be within ±0,5 % for RMS and average values and
±1 % for peak values, provided that the peak factor of waveforms to be measured is within limits
imposed by the instrument.
If inaccuracies exceed the above limits, only a sine-wave excitation of total harmonic content
less than 1 % is recommended and
• to determine the RMS, average and peak values of sinusoidal waveforms, a true RMS
sensing voltmeter of accuracy within ±1 % is recommended. The average and peak values
are obtained by multiplying the indicated RMS values by the following factors: average value
= 0,900 × RMS value, peak value = 1,414 × RMS value;
• to determine the RMS, average and peak values of non-sinusoidal waveforms, a digital
storage and processing oscilloscope or appropriate acquisition and processing instrument
shall be used. It shall be capable
...


IEC 62044-3 ®
Edition 2.0 2023-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Cores made of soft magnetic materials – Measuring methods –
Part 3: Magnetic properties at high excitation level

Noyaux en matériaux magnétiques doux – Méthodes de mesure –
Partie 3: Propriétés magnétiques à niveau élevé d'excitation
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form
or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from
either IEC or IEC's member National Committee in the country of the requester. If you have any questions about IEC
copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or
your local IEC member National Committee for further information.

Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni
utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et
les microfilms, sans l'accord écrit de l'IEC ou du Comité national de l'IEC du pays du demandeur. Si vous avez des
questions sur le copyright de l'IEC ou si vous désirez obtenir des droits supplémentaires sur cette publication, utilisez
les coordonnées ci-après ou contactez le Comité national de l'IEC de votre pays de résidence.

IEC Secretariat Tel.: +41 22 919 02 11
3, rue de Varembé info@iec.ch
CH-1211 Geneva 20 www.iec.ch
Switzerland
About the IEC
The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes
International Standards for all electrical, electronic and related technologies.

About IEC publications
The technical content of IEC publications is kept under constant review by the IEC. Please make sure that you have the
latest edition, a corrigendum or an amendment might have been published.

IEC publications search - webstore.iec.ch/advsearchform IEC Products & Services Portal - products.iec.ch
The advanced search enables to find IEC publications by a Discover our powerful search engine and read freely all the
variety of criteria (reference number, text, technical publications previews. With a subscription you will always have
committee, …). It also gives information on projects, replaced access to up to date content tailored to your needs.
and withdrawn publications.
Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
The world's leading online dictionary on electrotechnology,
Stay up to date on all new IEC publications. Just Published
containing more than 22 300 terminological entries in English
details all new publications released. Available online and once
and French, with equivalent terms in 19 additional languages.
a month by email.
Also known as the International Electrotechnical Vocabulary

(IEV) online.
IEC Customer Service Centre - webstore.iec.ch/csc

If you wish to give us your feedback on this publication or need
further assistance, please contact the Customer Service
Centre: sales@iec.ch.
A propos de l'IEC
La Commission Electrotechnique Internationale (IEC) est la première organisation mondiale qui élabore et publie des
Normes internationales pour tout ce qui a trait à l'électricité, à l'électronique et aux technologies apparentées.

A propos des publications IEC
Le contenu technique des publications IEC est constamment revu. Veuillez vous assurer que vous possédez l’édition la
plus récente, un corrigendum ou amendement peut avoir été publié.

Recherche de publications IEC -  IEC Products & Services Portal - products.iec.ch
webstore.iec.ch/advsearchform Découvrez notre puissant moteur de recherche et consultez
La recherche avancée permet de trouver des publications IEC gratuitement tous les aperçus des publications. Avec un
en utilisant différents critères (numéro de référence, texte, abonnement, vous aurez toujours accès à un contenu à jour
comité d’études, …). Elle donne aussi des informations sur les adapté à vos besoins.
projets et les publications remplacées ou retirées.

Electropedia - www.electropedia.org
IEC Just Published - webstore.iec.ch/justpublished
Le premier dictionnaire d'électrotechnologie en ligne au monde,
Restez informé sur les nouvelles publications IEC. Just
avec plus de 22 300 articles terminologiques en anglais et en
Published détaille les nouvelles publications parues.
français, ainsi que les termes équivalents dans 19 langues
Disponible en ligne et une fois par mois par email.
additionnelles. Egalement appelé Vocabulaire

Electrotechnique International (IEV) en ligne.
Service Clients - webstore.iec.ch/csc

Si vous désirez nous donner des commentaires sur cette
publication ou si vous avez des questions contactez-nous:
sales@iec.ch.
IEC 62044-3 ®
Edition 2.0 2023-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Cores made of soft magnetic materials – Measuring methods –

Part 3: Magnetic properties at high excitation level

Noyaux en matériaux magnétiques doux – Méthodes de mesure –

Partie 3: Propriétés magnétiques à niveau élevé d'excitation

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 29.030, 29.100.10 ISBN 978-2-8322-7172-8

– 2 – IEC 62044-3:2023 © IEC 2023
CONTENTS
FOREWORD . 5
INTRODUCTION . 7
1 Scope . 8
2 Normative references . 8
3 Terms, definitions and symbols. 8
3.1 Terms and definitions . 8
3.2 Symbols . 12
4 General requirements for measurements at high excitation level . 13
4.1 General statements . 13
4.1.1 Relation to practice . 13
4.1.2 Core effective parameters and material properties . 13
4.1.3 Reproducibility of the magnetic state . 13
4.2 Measuring coil . 13
4.2.1 General . 13
4.2.2 Number of turns . 14
4.2.3 Single winding and double winding . 14
4.3 Mounting of cores consisting of more than one part . 15
4.4 Measuring equipment . 15
5 Specimens . 18
6 Measuring procedures . 18
6.1 General procedure . 18
6.2 Measuring method for the effective amplitude permeability . 19
6.2.1 Purpose . 19
6.2.2 Principle of the measurement . 19
6.2.3 Circuit and equipment . 19
6.2.4 Measuring procedure . 19
6.2.5 Calculation . 20
6.3 Measuring methods for the power loss . 20
6.3.1 Purpose . 20
6.3.2 Methods and principles of the measurements . 20
7 Information to be stated . 23
8 Test report . 24
Annex A (informative) Basic circuits and related equipment for the measurement of
amplitude permeability . 25
Annex B (informative) Root-mean-square method for the measurement of power loss –
Example of a circuit and related procedure . 27
B.1 Method of measurement . 27
B.2 Measuring coil . 27
B.3 Measuring equipment . 28
B.4 Measuring procedure . 28
B.5 Pulse measurement and accuracy . 29
Annex C (informative) Multiplying methods for the measurement of power loss – Basic
circuits and related measurement procedures . 30
C.1 Basic circuits . 30
C.2 Requirements . 31
C.3 Measuring coil . 31

C.4 Accuracy . 31
C.5 V-A-W (volt-ampere-watt) meter method . 32
C.6 Impedance analyzer method . 32
C.7 Digitizing method . 32
C.8 Vector spectrum method . 33
C.9 Cross-power method . 33
Annex D (informative) Reflection method for the measurement of power loss – Basic
circuit and related measurement procedures . 34
D.1 Basic circuit . 34
D.2 Requirements . 34
D.3 Measuring coil . 34
D.4 Measuring procedure and accuracy . 35
Annex E (informative) Calorimetric measurement methods for the measurement of
power loss . 36
E.1 Basic circuit . 36
E.2 Requirements . 37
E.3 Measuring coil . 37
E.4 Accuracy . 37
E.5 Measurements at thermal equilibrium . 37
E.5.1 General . 37
E.5.2 Measurement across calibrated thermal resistance . 37
E.5.3 Measurement by matching the temperature rise in the core and resistor . 38
E.6 Measurements at non-thermal equilibrium . 38
Annex F (normative) Magnetic properties under pulse condition . 39
F.1 Object . 39
F.2 Measurement methods . 39
F.3 Principle of the methods . 39
F.4 Specimens . 39
F.5 Measuring coil . 39
F.6 Measuring equipment . 40
F.7 Measuring procedure . 41
F.7.1 General . 41
F.7.2 Measurement of pulse inductance factor and magnetizing current . 42
F.7.3 Measurement of the non-linearity of the magnetizing current . 43
F.8 Calculation . 44
Annex G (informative) Examples of circuits for pulse measurements . 46
Bibliography . 47

Figure 1 – Pulse excitation without biasing field . 10
Figure 2 – Pulse excitation with biasing field . 10
Figure A.1 – Basic circuits for the measurement of amplitude permeability . 26
Figure B.1 – Example of a measuring circuit for the RMS method . 27
Figure C.1 – Basic circuits for multiplying methods . 31
Figure D.1 – Basic circuit . 34
Figure E.1 – Basic circuit and related measurement procedures – Measurement set-up . 36
Figure F.1 – Voltage pulse parameters . 42
Figure F.2 – Typical measuring waveforms . 43

– 4 – IEC 62044-3:2023 © IEC 2023
Figure F.3 – Non-linearity of magnetizing current . 44
Figure G.1 – Measurement without bias and with single pulses . 46
Figure G.2 – Measurement with bias and with repeated pulses . 46

Table 1 – Some multiplying methods and related domains of excitation waveforms,
acquisition, processing . 21

INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
CORES MADE OF SOFT MAGNETIC MATERIALS –
MEASURING METHODS –
Part 3: Magnetic properties at high excitation level

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international
co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and
in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports,
Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their
preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with
may participate in this preparatory work. International, governmental and non-governmental organizations liaising
with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for
Standardization (ISO) in accordance with conditions determined by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence between
any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) IEC draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). IEC takes no position concerning the evidence, validity or applicability of any claimed patent rights in
respect thereof. As of the date of publication of this document, IEC had not received notice of (a) patent(s), which
may be required to implement this document. However, implementers are cautioned that this may not represent
the latest information, which may be obtained from the patent database available at https://patents.iec.ch. IEC
shall not be held responsible for identifying any or all such patent rights.
IEC 62044-3 has been prepared by IEC technical committee 51: Magnetic components, ferrite
and magnetic powder materials. It is an International Standard.
This second edition cancels and replaces the first edition published in 2000. This edition
constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous
edition:
a) addition of Annex F and Annex G.

– 6 – IEC 62044-3:2023 © IEC 2023
The text of this International Standard is based on the following documents:
Draft Report on voting
51/1426/CDV 51/1439/RVC
Full information on the voting for its approval can be found in the report on voting indicated in
the above table.
The language used for the development of this International Standard is English.
This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in
accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available
at www.iec.ch/members_experts/refdocs. The main document types developed by IEC are
described in greater detail at www.iec.ch/publications.
A list of all parts in the IEC 62044 series, published under the general title Cores made of soft
magnetic materials – Measuring methods, can be found on the IEC website.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under webstore.iec.ch in the data related to the
specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
INTRODUCTION
IEC 62044, under the general title Cores made of soft magnetic materials – Measuring methods,
includes the following parts:
IEC 62044-1: Generic specification
IEC 62044-2: Magnetic properties at low excitation level
IEC 62044-3: Magnetic properties at high excitation level

– 8 – IEC 62044-3:2023 © IEC 2023
CORES MADE OF SOFT MAGNETIC MATERIALS –
MEASURING METHODS –
Part 3: Magnetic properties at high excitation level

1 Scope
This part of IEC 62044 specifies measuring methods for power loss and amplitude permeability
of magnetic cores forming the closed magnetic circuits intended for use at high excitation levels
in inductors, chokes, transformers and similar devices for power electronics applications.
The methods given in this document can cover the measurement of magnetic properties for
frequencies ranging practically from direct current to 10 MHz, and even possibly higher, for the
calorimetric and reflection methods. The applicability of the individual methods to specific
frequency ranges is dependent on the level of accuracy that is to be obtained.
The methods in this document are basically the most suitable for sine-wave excitations. Other
periodic waveforms can also be used; however, adequate accuracy can only be obtained if the
measuring circuitry and instruments used are able to handle and process the amplitudes and
phases of the signals involved within the frequency spectrum corresponding to the given
magnetic flux density and field strength waveforms with only slightly degraded accuracy.
NOTE It can be necessary for some magnetically soft metallic materials to follow specific general principles,
customary for these materials, related to the preparation of specimens and specified calculations. These principles
are formulated in IEC 60404-8-6.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies.
For undated references, the latest edition of the referenced document (including any
amendments) applies.
IEC 62044-1:2002, Cores made of soft magnetic materials – Measuring methods – Part 1:
Generic specification
3 Terms, definitions and symbols
3.1 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at https://www.electropedia.org/
• ISO Online browsing platform: available at https://www.iso.org/obp

3.1.1
effective amplitude permeability
μ
ea
ˆ
magnetic permeability obtained from the peak value of the effective magnetic flux density, B ,
e
ˆ
H
and the peak value of the effective magnetic field strength, , at the stated value of either,
e
when the magnetic flux density and magnetic field vary periodically with time and with an
average of zero, and the material is initially in a specified demagnetized state
3.1.2
maximum effective amplitude permeability
μ
ea max
ˆ
maximum value of the effective amplitude permeability when the amplitude of excitation ( B or
e
ˆ
H ) is varied
e
3.1.3
excitation
either magnetic flux density or field strength for which the waveform and amplitude both remain
within the specified tolerance
Note 1 to entry: When the magnetic flux density (field strength) mode of excitation is chosen, the resultant waveform
of field strength (magnetic flux density) can be distorted with respect to the excitation waveform due to the non-linear
behaviour of the magnetic material.
3.1.4
high excitation level
excitation at which the permeability depends on excitation amplitude (particularly at low
frequencies) or at which the power loss results in a noticeable temperature rise (particularly at
high frequencies), or both
3.1.5
exciting winding
winding of measuring coil to which the exciting voltage is applied or through which the exciting
current is flowing
3.1.6
voltage sensing winding
unloaded winding of a measuring coil across which the electromotive force induced by the
excitation can be determined
3.1.7
measuring winding
winding, usually secondary, loaded or unloaded, which can be used for measurement apart
from the exciting or voltage sensing winding, or both
3.1.8
power loss
power absorbed by the core
3.1.9
pulse excitation without biasing field
excitation in which a core is energized by a voltage pulse, from a remanent flux density to a
higher value of flux density in the same direction, and in which the core recovers to the same
remanent flux density
Note 1 to entry: The excursion in the B-H plane associated with such a pulse is shown in Figure 1.

– 10 – IEC 62044-3:2023 © IEC 2023

Figure 1 – Pulse excitation without biasing field
Note 2 to entry: If the back-e.m.f. during recovery is limited only by the time constant of the test circuit, then the
magnetizing current decays exponentially.
Alternatively the back-e.m.f. can be limited to a constant value, for example by returning the energy via a secondary
winding to a voltage source; then the magnetizing current decay is approximately linear. The latter method can
prevent excessively high back-e.m.f. and high rates of change of flux. The distinction is mainly relevant to loss
measurements.
3.1.10
pulse excitation with biasing field
excitation in which a core is energized by a voltage pulse, from a value of the flux density
determined by a biasing field to a flux density in the opposite direction, and in which the core
recovers to the same value determined by the biasing field
Note 1 to entry: The excursion in the B-H plane associated with such a pulse is shown in Figure 2.

Figure 2 – Pulse excitation with biasing field
Note 2 to entry: See Note to entry 2 of 3.1.9.

3.1.11
pulse permeability
μ
p
relative permeability obtained from the change of flux density and the corresponding change of
the field strength when either quantity is varying in an arbitrary form between stated limits:
1 ΔB
μ ⋅
p
μ ΔH
SEE: Figure 1 and Figure 2.
Note 1 to entry: The value of the pulse permeability depends strongly on the limits of the flux density or field strength
excursions; it is not necessary for these limits to be symmetrical with respect to zero.
Note 2 to entry: Often pulse permeability refers to the special case of square voltage pulses being applied to an
exciting winding; the flux density waveform is then approximately triangular.
3.1.12
pulse amplitude
U
m
maximum instantaneous value which an ideal voltage pulse would have with respect to the
steady value of the voltage between pulses
Note 1 to entry: An ideal pulse is derived from an actual voltage pulse by ignoring unwanted or non-pertinent
phenomena such as overshoot (see Figure 1).
3.1.13
pulse duration
t
d
time interval during which the instantaneous value of the pulse exceeds 50 % of the pulse
amplitude
U
m
L =
p
Δi t
md
where
L
p is the pulse inductance
Δi is the total change in i during the pulse
m m
SEE: Figure F.1.
Note 1 to entry: For unidirectional drive pulses Δi = î .
m m
3.1.14
pulse inductance factor
A
LP
pulse inductance divided by the square of the number of turns of the test coil
L
p
A =
LP
N
=
– 12 – IEC 62044-3:2023 © IEC 2023
3.1.15
voltage-time product limit
(U·t)
lim
specified limit of the product of the amplitude of a voltage pulse and the time elapsed from the
start of the pulse
Note 1 to entry: Within this limit the non-linearity of the magnetizing current through the measuring coil placed on
the core should not exceed a specified value.
3.1.16
non-linearity (with time)
ratio of the actual instantaneous value of a characteristic at a time t to the value reached by the
extrapolated linear portion of its graph versus time, at the same instant
SEE: Figure F.3.
3.1.17
pulse repetition rate
frequency of recurrence of the pulses in a periodic sequence of pulses
3.2 Symbols
All the formulae in this document use basic SI units. When multiples or sub-multiples are used,
the appropriate power of 10 shall be introduced.
A effective cross-sectional area of the core
e
ˆ
peak value of the effective magnetic flux density in the core
B
e
f frequency
ˆ
H
peak value of the effective magnetic field strength in the core
e
l effective magnetic path length of the core
e
L inductance
i instantaneous value of the current
I current
N number of turns of winding of the measuring coil
P power loss in the core
Q quality factor of the core for a given frequency
R resistance
t time
T temperature
u instantaneous value of the voltage
U voltage
V effective volume of the core
e
δ relative error, deviation, etc.
∆ absolute error, deviation, etc.
µ effective amplitude permeability
ea
−7
µ permeability of vacuum: approximately 4π ×10 H/m
ϕ phase shift
ω angular frequency = 2πf
NOTE 1 The additional subscript, upper script, etc., gives a more specific meaning to the given symbol.

NOTE 2 Symbols which are used sporadically are defined in the place where they appear in the text.
NOTE 3 Effective parameters, such as effective magnetic path length, l , effective cross-sectional area, A , and
e e
effective volume of the core, V , are calculated in accordance with IEC 60205.
e
NOTE 4 In the text of this document, the term flux density stands for the shortened term of magnetic flux density.
4 General requirements for measurements at high excitation level
4.1 General statements
4.1.1 Relation to practice
The measuring conditions, methods and procedures shall be chosen in such a way that the
measured results are suitable for predicting the performance of the core under practical
circumstances. This does not imply that all these stipulations, especially those related to the
excitation waveforms, have to correspond to terms encountered in practice.
4.1.2 Core effective parameters and material properties
Since the core is in general of non-uniform cross-section and generally has non-uniformly
distributed windings along the core path, the measurement does not yield the amplitude
permeability and the power loss of the material, but the effective values of these parameters
ˆ ˆ
appropriate to the effective magnetic flux density B and the effective field strength H in the
e e
core.
For the measurement of the amplitude permeability and the power loss of the material, the core
shall have a ring or toroidal shape in which the ratio of the outer to the inner diameter should
not be greater than 1,4 and should have windings distributed uniformly, close to the core, of
inductive coupling coefficient practically equal to unity.
4.1.3 Reproducibility of the magnetic state
To obliterate various remanence and time effects in the core material, the measurement shall
be made at a well-defined and reproducible magnetic state.
Any measurement under specified excitation, unless otherwise stated, is to be made at the time
t = t + ∆t after the magnetic conditioning start; t is the time period within which the magnetic
m c c
conditioning is completed and, whereupon, the specified excitation is set; ∆t is the time period
during which the core is kept stable under the excitation being set.
4.2 Measuring coil
4.2.1 General
Normally, a measuring coil will be used, but in principle any coaxial line, cavity or other suitable
device providing the necessary interaction between the magnetic material and the
electromagnetic signal, may also be used.
For measurement on toroid using coils, the turns of the measuring coil shall be distributed in
such a way as to keep both the stray capacitance and the stray field as low as necessary for
sufficiently accurate measurement.
For measurements made on cores assembled around a coil, the shape of the measuring coil
shall correspond to that of the coils used for normal application of the core and its influence on
the variation of the inductance to be measured shall be negligible.

– 14 – IEC 62044-3:2023 © IEC 2023
Unless otherwise specified, the test coil complete with coil former or encapsulation, or both,
shall be positioned in a coaxial way to the limb which it embraces, and the side of the coil at
which the start of the winding is located shall be lightly pressed into contact with the core at
one end of this limb as follows:
• for a symmetrical core, the coil assembly shall contact the core at either one end or the
other;
• for a core with an air-gap that is asymmetrical because the gap has been made by grinding
away material on the leg of one core half and not the other, the coil assembly shall make
end contact to the half of the core that has not been ground away to create the gap.
One of the coil faces shall be marked so as to define its orientation. The coil shall be kept in
the defined position during the whole measurement in order to obtain the maximum
reproducibility of the measurement.
4.2.2 Number of turns
The number of turns shall be specified for each winding in relation to the measuring conditions,
the equipment used and the accuracy to be obtained. The windings shall be wound as close to
the core as possible, to make the coupling (magnetic flux linkage) coefficients between the
measuring coil windings and the core and between the windings of measuring coil, as close to
100 % as possible.
The resistance, self-capacitance and inter-winding capacitance of windings should be as low
as possible to make the related errors negligible.
In the case of ring or toroidal cores, the turns shall be distributed evenly around the core
circumference.
The connectors, primarily of exciting winding, should consist of insulated strands, if this is
necessary for measurements at high frequencies.
When winding a sharp-edge core, the wire insulation should not rupture and, in the case of
stranded wire, the strands should not break.
4.2.3 Single winding and double winding
The use of a single winding both for excitation and voltage sensing is recommended if
– the coupling between the exciting winding and the voltage sensing winding is so reduced
that it results in a non-negligible error in the determination of the measuring magnetic flux
density B in the core;
– the inter-winding capacitance is too high;
– there is no measuring circuitry contra-indication against the direct connection of the exciting
winding to the input(s) of measuring instruments.
When a single winding is used, it is recommended that its resistance be made as low as possible
to make the winding ohmic power loss negligible compared to the power loss in the core.
The use of separate exciting and voltage sensing windings (double winding) is recommended
if, for whatever reason, the exciting winding is galvanically separated from the voltage and the
current measuring instruments, for example, to avoid a floating or DC connection to their inputs.
When the exciting and voltage sensing windings are used, their magnetic coupling coefficient
shall be as close to 100 % as possible.

The voltage necessary for calculation of the magnetic flux density in the core is typically
measured across a voltage sensing winding that is separate from the current-carrying (exciting)
winding. When measuring core losses, ohmic losses in the voltage sensing winding do not affect
the calculation, but ohmic losses in the current-carrying (exciting) winding shall be excluded
from the core loss calculation.
The use of two windings is recommended at more than 200 kHz.
4.3 Mounting of cores consisting of more than one part
A ferrite core set consisting of more than one part and which is to be assembled around the
measuring coil, shall be held together with glue, tape or a clamping device throughout the
measurement.
Whichever method is used to join the core parts together, it shall have the following
characteristics:
• distribution of the joining force uniformly over the mating surfaces, without the introduction
of bending stresses in the core;
• holding of all the core parts rigidly and without changing the position with regard to each
other;
• when a specified clamping method is used, an initial over-force of about 10 % shall be
applied when the core is closed, in order to break down fine irregularities between the
cleaned mating surfaces. Next, the specified clamping force ±5 % shall be applied;
• keeping the joining force constant within ±1 % during all measuring operations within all
measuring conditions, including the full specified temperature range.
The mounting of such cores shall be carried out in accordance with the following instructions.
The mating surface shall be inspected for damage and cleanness. Damaged cores shall not be
used. The mating surface shall be cleaned by non-abrasive means, for example, by rubbing
gently on a dry washing-leather. Next, the mating surfaces shall be degreased if they have to
be glued. Dust particles shall be blown off with clean dry compressed air. The mating surfaces
shall never be touched with bare fingers. The core parts shall then be assembled around the
measuring coil, the latter being locked in position with respect to the core by suitable means,
for example, a foam-washer. The core parts are centered and glued or placed in a clamping
device. The glue, if used, shall be spread evenly on the mating surface to form a film as thin as
possible and then properly hardened.
In the case where the clamping device is used, the clamping force specified in the relevant
specification shall be applied. The glued, taped or clamped cores shall relax under the specified
conditions (see IEC 62044-1:2002, Clause 3) for a long enough time to allow any variation of
stress effects, due to clamping, gluing or taping, to become negligible.
4.4 Measuring equipment
4.4.1 Any suitable measuring equipment may be used. Examples of appropriate circuits are
given in Annex A to Annex E.
In addition to any requirement specified for the particular method or measuring circuit, or both,
used, the following general requirements shall be met.
4.4.2 To ensure the magnetic flux density (field strength) mode of excitation, the output
impedance of the exciting source shall be low (high) compared with the series impedance of
the exciting winding of the measuring coil assembled with the core under test and the current
sensing resistor.
– 16 – IEC 62044-3:2023 © IEC 2023
4.4.3 When the sinusoidal waveform of excitation is specified, the total harmonic content of
the excitation source shall be less than 1 %. When square pulses are specified, the relevant
requirements of Annex F shall be met.
4.4.4 During the period of measurement, the excitation amplitude variations shall not exceed
±0,05 % and the frequency stability shall be adequate for the measuring method and the
equipment used.
4.4.5 The frequency range of voltmeters and other voltage sensing instruments shall include
all harmonics of the measured voltage having amplitudes of 1 % or more of their fundamentals.
This frequency range shall be specified in the relevant instrument specification.
4.4.6 The voltmeters and other voltage sensing instruments used shall be high-impedance
instruments, the connection of which will have only a negligible effect on the measuring circuit,
especially at high frequencies. The probes of a high-input resistance and a low-input capa-
citance can reduce the load effects.
4.4.7 The accuracy of the voltmeters or voltage sensing instruments, determined for
the calibrating sinusoidal waveform, shall be within ±0,5 % for RMS and average values and
±1 % for peak values, provided that the peak factor of waveforms to be measured is within limits
imposed by the instrument.
If inaccuracies exceed the above limits, only a sine-wave excitation of total harmonic content
less than 1 % is recommended and
• to determine the RMS, average and peak values of sinusoidal waveforms, a true RMS
sensing voltmeter of accuracy within ±1 % is recommended. The average and peak values
are obtained by multiplying the indicated RMS values by the following factors: average value
= 0,900 × RMS value, peak value = 1,414 × RMS value;
• to determine the RMS, average and peak values of non-sinusoidal waveforms, a digital
storage and processing oscilloscope or appropriate acquisition and processing instrument
shall be used. It shall be capable of capturing and processing the waveform with the
sampling rate not less than 150 samples per waveform period and the resolution not less
than 8 bits.
NOTE The peak factor is the ratio of the peak value to the RMS value of the measured waveform.
4.4.8 The resistance of the in-series
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

Frequently Asked Questions

IEC 62044-3:2023 is a standard published by the International Electrotechnical Commission (IEC). Its full title is "Cores made of soft magnetic materials - Measuring methods - Part 3: Magnetic properties at high excitation level". This standard covers: IEC 62044-3:2023 specifies measuring methods for power loss and amplitude permeability of magnetic cores forming the closed magnetic circuits intended for use at high excitation levels in inductors, chokes, transformers and similar devices for power electronics applications. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: addition of Annex F and Annex G.

IEC 62044-3:2023 specifies measuring methods for power loss and amplitude permeability of magnetic cores forming the closed magnetic circuits intended for use at high excitation levels in inductors, chokes, transformers and similar devices for power electronics applications. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition: addition of Annex F and Annex G.

IEC 62044-3:2023 is classified under the following ICS (International Classification for Standards) categories: 29.030 - Magnetic materials; 29.100.10 - Magnetic components. The ICS classification helps identify the subject area and facilitates finding related standards.

IEC 62044-3:2023 has the following relationships with other standards: It is inter standard links to IEC 62044-3:2000/COR1:2021, IEC 62044-3:2000. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

You can purchase IEC 62044-3:2023 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of IEC standards.