IEC 61526:2005
(Main)Radiation protection instrumentation - Measurement of personal dose equivalents Hp(10) and Hp(0,07) for X, gamma, neutron and beta radiations - Direct reading personal dose equivalent meters and monitors
Radiation protection instrumentation - Measurement of personal dose equivalents Hp(10) and Hp(0,07) for X, gamma, neutron and beta radiations - Direct reading personal dose equivalent meters and monitors
Specifies requirements for the dosemeter and, if supplied, for its associated readout system. Specifies general characteristics, general test procedures, radiation characteristics as well as electrical, mechanical, safety and environmental characteristics. Also specifies classes of personal dose equivalent meters in relation to retention of stored information.
Instrumentation pour la radioprotection - Mesure des équivalents de dose individuels Hp(10) et Hp(0,07) pour les rayonnements X, gamma, neutron et bêta - Appareils de mesure à lecture directe et moniteurs de l'équivalent de dose individuel
Précise les exigences pour les dosimètres et, s'ils sont fournis, pour leurs systèmes de lecture associés. Spécifie leurs caractéristiques générales, leurs procédures générales d'essai, leurs caractéristiques sous rayonnement ainsi que leurs caractéristiques électriques, mécaniques, de sécurité et environnementales. Spécifie aussi les classes des appareils de mesure de l'équivalent de dose individuel en fonction de la conservation des informations.
General Information
Relations
Standards Content (Sample)
NORME CEI
INTERNATIONALE
IEC
INTERNATIONAL
Deuxième édition
STANDARD
Second edition
2005-02
Instrumentation pour la radioprotection –
Mesure des équivalents de dose individuels
H (10) et H (0,07) pour les rayonnements
p p
X, gamma, neutron et bêta –
Appareils de mesure à lecture directe et
moniteurs de l'équivalent de dose individuel
Radiation protection instrumentation –
Measurement of personal dose equivalents
H (10) and H (0,07) for X, gamma, neutron
p p
and beta radiations –
Direct reading personal dose equivalent
meters and monitors
Numéro de référence
Reference number
CEI/IEC 61526:2005
Numérotation des publications Publication numbering
Depuis le 1er janvier 1997, les publications de la CEI As from 1 January 1997 all IEC publications are
sont numérotées à partir de 60000. Ainsi, la CEI 34-1 issued with a designation in the 60000 series. For
devient la CEI 60034-1. example, IEC 34-1 is now referred to as IEC 60034-1.
Editions consolidées Consolidated editions
Les versions consolidées de certaines publications de la The IEC is now publishing consolidated versions of its
CEI incorporant les amendements sont disponibles. Par publications. For example, edition numbers 1.0, 1.1
exemple, les numéros d’édition 1.0, 1.1 et 1.2 indiquent and 1.2 refer, respectively, to the base publication,
respectivement la publication de base, la publication de the base publication incorporating amendment 1 and
base incorporant l’amendement 1, et la publication de the base publication incorporating amendments 1
base incorporant les amendements 1 et 2. and 2.
Informations supplémentaires Further information on IEC publications
sur les publications de la CEI
Le contenu technique des publications de la CEI est The technical content of IEC publications is kept
constamment revu par la CEI afin qu'il reflète l'état under constant review by the IEC, thus ensuring that
actuel de la technique. Des renseignements relatifs à the content reflects current technology. Information
cette publication, y compris sa validité, sont dispo- relating to this publication, including its validity, is
nibles dans le Catalogue des publications de la CEI available in the IEC Catalogue of publications
(voir ci-dessous) en plus des nouvelles éditions, (see below) in addition to new editions, amendments
amendements et corrigenda. Des informations sur les and corrigenda. Information on the subjects under
sujets à l’étude et l’avancement des travaux entrepris consideration and work in progress undertaken by the
par le comité d’études qui a élaboré cette publication, technical committee which has prepared this
ainsi que la liste des publications parues, sont publication, as well as the list of publications issued,
également disponibles par l’intermédiaire de: is also available from the following:
• Site web de la CEI (www.iec.ch) • IEC Web Site (www.iec.ch)
• Catalogue des publications de la CEI • Catalogue of IEC publications
Le catalogue en ligne sur le site web de la CEI The on-line catalogue on the IEC web site
(www.iec.ch/searchpub) vous permet de faire des (www.iec.ch/searchpub) enables you to search by a
recherches en utilisant de nombreux critères, variety of criteria including text searches,
comprenant des recherches textuelles, par comité technical committees and date of publication. On-
d’études ou date de publication. Des informations en line information is also available on recently
ligne sont également disponibles sur les nouvelles issued publications, withdrawn and replaced
publications, les publications remplacées ou retirées, publications, as well as corrigenda.
ainsi que sur les corrigenda.
• IEC Just Published • IEC Just Published
Ce résumé des dernières publications parues This summary of recently issued publications
(www.iec.ch/online_news/justpub) est aussi dispo- (www.iec.ch/online_news/justpub) is also available
nible par courrier électronique. Veuillez prendre by email. Please contact the Customer Service
contact avec le Service client (voir ci-dessous) Centre (see below) for further information.
pour plus d’informations.
• Service clients • Customer Service Centre
Si vous avez des questions au sujet de cette If you have any questions regarding this
publication ou avez besoin de renseignements publication or need further assistance, please
supplémentaires, prenez contact avec le Service contact the Customer Service Centre:
clients:
Email: custserv@iec.ch Email: custserv@iec.ch
Tél: +41 22 919 02 11 Tel: +41 22 919 02 11
Fax: +41 22 919 03 00 Fax: +41 22 919 03 00
.
NORME CEI
INTERNATIONALE
IEC
INTERNATIONAL
Deuxième édition
STANDARD
Second edition
2005-02
Instrumentation pour la radioprotection –
Mesure des équivalents de dose individuels
H (10) et H (0,07) pour les rayonnements
p p
X, gamma, neutron et bêta –
Appareils de mesure à lecture directe et
moniteurs de l'équivalent de dose individuel
Radiation protection instrumentation –
Measurement of personal dose equivalents
H (10) and H (0,07) for X, gamma, neutron
p p
and beta radiations –
Direct reading personal dose equivalent
meters and monitors
© IEC 2005 Droits de reproduction réservés ⎯ Copyright - all rights reserved
Aucune partie de cette publication ne peut être reproduite ni No part of this publication may be reproduced or utilized in any
utilisée sous quelque forme que ce soit et par aucun procédé, form or by any means, electronic or mechanical, including
électronique ou mécanique, y compris la photocopie et les photocopying and microfilm, without permission in writing from
microfilms, sans l'accord écrit de l'éditeur. the publisher.
International Electrotechnical Commission, 3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, Switzerland
Telephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch Web: www.iec.ch
CODE PRIX
XA
PRICE CODE
Commission Electrotechnique Internationale
International Electrotechnical Commission
МеждународнаяЭлектротехническаяКомиссия
Pour prix, voir catalogue en vigueur
For price, see current catalogue
– 2 – 61526 © CEI:2005
SOMMAIRE
AVANT-PROPOS.8
INTRODUCTION.12
1 Domaine d'application et objet.14
2 Références normatives.16
3 Termes et définitions .20
4 Unités et liste des symboles .32
4.1 Unités .32
4.2 Liste des symboles.32
5 Caractéristiques mécaniques.36
5.1 Taille .36
5.2 Masse .36
5.3 Boîtier .38
5.4 Boutons de commutation .38
6 Caractéristiques générales .38
6.1 Classification des appareils de mesure de l'équivalent de dose individuel .38
6.2 Indication .38
6.3 Marquages sur le dosimètre .38
6.4 Décontamination .38
6.5 Domaines de mesure de l'équivalent de dose et du débit d'équivalent de
dose.38
6.6 Etendue de mesure .38
6.7 Domaine de variation d’une grandeur d’influence .40
6.8 Utilisation de plus d'un dosimètre .40
6.9 Dérive de zéro et indication due au rayonnement naturel .40
6.10 Alarmes préréglables en dose ou en débit de dose.40
6.11 Indication de dysfonctionnement .42
7 Procédures générales d'essai.42
7.1 Nature des essais .42
7.2 Conditions de référence et conditions normales d'essai.42
7.3 Essais concernant les grandeurs d'influence de type F.42
7.4 Essais concernant les grandeurs d'influence de type S .42
7.5 Choix du fantôme utilisé pour les essais.44
7.6 Position du dosimètre au cours des essais .44
7.7 Position du dosimètre pendant son utilisation .44
7.8 Domaine minimal de variation d'une grandeur d'influence.44
7.9 Débits d'équivalent de dose faibles .44
7.10 Fluctuations statistiques .44
7.11 Production de rayonnement de référence .46
8 Additivité des valeurs indiquées .46
8.1 Exigences .46
8.2 Méthode d’essai .46
8.3 Interprétation des résultats.46
61526 © IEC:2005 – 3 –
CONTENTS
FOREWORD.9
INTRODUCTION.13
1 Scope and object.15
2 Normative references .17
3 Terms and definitions .21
4 Units and list of symbols.33
4.1 Units .33
4.2 List of symbols .33
5 Mechanical characteristics.37
5.1 Size.37
5.2 Mass .37
5.3 Case .39
5.4 Switches .39
6 General characteristics.39
6.1 Personal dose equivalent meter classification .39
6.2 Indication .39
6.3 Dosemeter markings .39
6.4 Retention of radioactive contamination .39
6.5 Ranges for dose equivalent and dose equivalent rate.39
6.6 Effective range of measurement .39
6.7 Rated range of an influence quantity .41
6.8 Use of more than one dosemeter.41
6.9 Zero effect and indication due to natural environmental radiation .41
6.10 Resettable dose or dose rate alarms .41
6.11 Indication of malfunction.43
7 General test procedures .43
7.1 Nature of tests.43
7.2 Reference conditions and standard test conditions .43
7.3 Tests for influence quantities of type F .43
7.4 Tests for influence quantities of type S.43
7.5 Choice of phantom for testing.45
7.6 Position of dosemeter for the purpose of testing.45
7.7 Position of dosemeter during use .45
7.8 Minimum rated range of influence quantity .45
7.9 Low dose equivalent rates.45
7.10 Statistical fluctuations .45
7.11 Production of reference radiation.47
8 Additivity of indicated value .47
8.1 Requirements.47
8.2 Method of test .47
8.3 Interpretation of the results .47
– 4 – 61526 © CEI:2005
9 Exigences de performances sous rayonnement et essais .48
9.1 Généralités.48
9.2 Considération relative à l’incertitude de la valeur conventionnellement vraie .48
9.3 Linéarité de la réponse.48
9.4 Variation de la réponse en dose en fonction du débit de dose .50
9.5 Variation de la réponse en fonction de l’énergie des photons et de l’angle
d'incidence .52
9.6 Variation de la réponse en fonction de l’énergie des neutrons et de l’angle
d'incidence .56
9.7 Variation de la réponse en fonction de l’énergie des bêtas et de l’angle
d'incidence .58
9.8 Conservation de la lecture en équivalent de dose.60
9.9 Caractéristiques de surcharge.62
9.10 Alarmes.62
9.11 Fonction modèle.66
10 Caractéristiques électriques et d'environnement, exigences et essais .68
10.1 Généralités.68
10.2 Alimentation électrique – Fonctionnement sur piles ou batteries.68
10.3 Température ambiante .72
10.4 Humidité relative .74
10.5 Pression atmosphérique .74
10.6 Etanchéité .76
10.7 Stockage .76
11 Performances électromagnétiques, exigences et essais .76
11.1 Généralités.76
11.2 Décharge électrostatique.76
11.3 Champs électromagnétiques radiatifs .78
11.4 Champs électromagnétiques émis par les téléphones portables ou réseaux
sans fils.78
11.5 Perturbations induites dans les conducteurs par des transitoires rapides ou
des impulsions .80
11.6 Perturbations induites dans les conducteurs par des surtensions .80
11.7 Perturbations induites dans les conducteurs par des radiofréquences .82
11.8 Champ magnétique à 50 Hz/60 Hz .82
11.9 Chutes de tension et microcoupures.82
12 Performances mécaniques, exigences et essais .84
12.1 Généralités.84
12.2 Essai de chute.84
12.3 Essai de vibration.84
12.4 Essai microphonique .86
13 Incertitude .86
14 Documentation .86
14.1 Rapport d'essai de type.86
14.2 Certificat .88
15 Manuel d'utilisation et de maintenance .88
61526 © IEC:2005 – 5 –
9 Radiation performance requirements and tests .49
9.1 General .49
9.2 Consideration of the uncertainty of the conventional true value .49
9.3 Linearity of the response .49
9.4 Variation of the response due to dose rate dependence of dose
measurements.51
9.5 Variation of the response due to photon radiation energy and angle of
incidence.53
9.6 Variation of the response due to neutron radiation energy and angle of
incidence.57
9.7 Variation of the response due to beta radiation energy and angle of
incidence.59
9.8 Retention of dose equivalent reading .61
9.9 Overload characteristics .63
9.10 Alarm .63
9.11 Model function.67
10 Electrical and environmental performance requirements and tests.69
10.1 General .69
10.2 Power supplies – Battery operation .69
10.3 Ambient temperature .73
10.4 Relative humidity.75
10.5 Atmospheric pressure.75
10.6 Sealing.77
10.7 Storage .77
11 Electromagnetic performance requirements and tests.77
11.1 General .77
11.2 Electrostatic discharge .77
11.3 Radiated electromagnetic fields.79
11.4 Radiated electromagnetic fields of mobile phones or wireless LAN.79
11.5 Conducted disturbances induced by fast transients or bursts.81
11.6 Conducted disturbances induced by surges.81
11.7 Conducted disturbances induced by radio-frequencies .83
11.8 50 Hz/60 Hz magnetic field.83
11.9 Voltage dips and short interruptions .83
12 Mechanical performance, requirements and tests .85
12.1 General .85
12.2 Drop test .85
12.3 Vibration test.85
12.4 Microphonics test .87
13 Uncertainty.87
14 Documentation .87
14.1 Type test report.87
14.2 Certificate .89
15 Operation and maintenance manual .89
– 6 – 61526 © CEI:2005
Annexe A (normative) Fluctuations statistiques.104
Annexe B (informative) Procédure pour déterminer la variation de la réponse relative
en fonction de l'énergie et de l'angle d'incidence des rayonnements .108
Error! No table of figures entries found.
Tableau 1 – Symboles (et abréviations) .32
Tableau 2 – Conditions de référence et conditions normales d'essai .90
Tableau 3 – Caractéristiques sous rayonnement des dosimètres mesurant H (0,07)
p
pour les rayonnements X, gamma et bêta .92
Tableau 4 – Caractéristiques sous rayonnement des dosimètres mesurant H (10) pour
p
les rayonnements X, gamma et bêta .94
Tableau 5 – Caractéristiques sous rayonnement des dosimètres mesurant H (10) pour
p
les neutrons.96
Tableau 6 – Caractéristiques électriques et environnementales des dosimètres.98
Tableau 7 – Caractéristiques des dosimètres aux perturbations électromagnétiques .100
Tableau 8 – caractéristiques des dosimètres aux perturbations mécaniques.102
Tableau A.1 – Nombre de lectures requises pour démontrer une réelle différence
(niveau de confiance 95 %) entre deux séries de lectures effectuées sur le même
instrument .106
61526 © IEC:2005 – 7 –
Annex A (normative) Statistical fluctuations .105
Annex B (informative) Procedure to determine the variation of the relative response
due to radiation energy and angle of radiation incidence.109
Table 1 – Symbols (and abbreviated terms) .33
Table 2 – Reference conditions and standard test conditions .91
Table 3 – Radiation characteristics of H (0,07) dosemeters for X, gamma and beta
p
radiation .93
Table 4 – Radiation characteristics of H (10) dosemeters for X and gamma radiation .95
p
Table 5 – Radiation characteristics of H (10) dosemeters for neutron radiation .97
p
Table 6 – Electrical and environmental characteristics of dosemeters .99
Table 7 – Electromagnetic disturbance characteristics of dosemeters .101
Table 8 – Mechanical disturbances characteristics of dosemeters.103
Table A.1 – Number of instrument readings required to detect true differences (95 %
confidence level) between two sets of instrument readings on the same instrumentTT.107
– 8 – 61526 © CEI:2005
COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
____________
INSTRUMENTATION POUR LA RADIOPROTECTION –
MESURE DES ÉQUIVALENTS DE DOSE INDIVIDUELS H (10) ET H (0,07)
P P
POUR LES RAYONNEMENTS X, GAMMA, NEUTRON ET BETA –
APPAREILS DE MESURE À LECTURE DIRECTE ET MONITEURS
DE L’ÉQUIVALENT DE DOSE INDIVIDUEL
AVANT-PROPOS
1) La Commission Electrotechnique Internationale (CEI) est une organisation mondiale de normalisation
composée de l'ensemble des comités électrotechniques nationaux (Comités nationaux de la CEI). La CEI a
pour objet de favoriser la coopération internationale pour toutes les questions de normalisation dans les
domaines de l'électricité et de l'électronique. A cet effet, la CEI – entre autres activités – publie des Normes
internationales, des Spécifications techniques, des Rapports techniques, des Spécifications accessibles au
public (PAS) et des Guides (ci-après dénommés "Publication(s) de la CEI"). Leur élaboration est confiée à des
comités d'études, aux travaux desquels tout Comité national intéressé par le sujet traité peut participer. Les
organisations internationales, gouvernementales et non gouvernementales, en liaison avec la CEI, participent
également aux travaux. La CEI collabore étroitement avec l'Organisation Internationale de Normalisation (ISO),
selon des conditions fixées par accord entre les deux organisations.
2) Les décisions ou accords officiels de la CEI concernant les questions techniques représentent, dans la mesure
du possible, un accord international sur les sujets étudiés, étant donné que les Comités nationaux de la CEI
intéressés sont représentés dans chaque comité d’études.
3) Les Publications de la CEI se présentent sous la forme de recommandations internationales et sont agréées
comme telles par les Comités nationaux de la CEI. Tous les efforts raisonnables sont entrepris afin que la CEI
s'assure de l'exactitude du contenu technique de ses publications; la CEI ne peut pas être tenue responsable
de l'éventuelle mauvaise utilisation ou interprétation qui en est faite par un quelconque utilisateur final.
4) Dans le but d'encourager l'uniformité internationale, les Comités nationaux de la CEI s'engagent, dans toute la
mesure possible, à appliquer de façon transparente les Publications de la CEI dans leurs publications
nationales et régionales. Toutes divergences entre toutes Publications de la CEI et toutes publications
nationales ou régionales correspondantes doivent être indiquées en termes clairs dans ces dernières.
5) La CEI n’a prévu aucune procédure de marquage valant indication d’approbation et n'engage pas sa
responsabilité pour les équipements déclarés conformes à une de ses Publications.
6) Tous les utilisateurs doivent s'assurer qu'ils sont en possession de la dernière édition de cette publication.
7) Aucune responsabilité ne doit être imputée à la CEI, à ses administrateurs, employés, auxiliaires ou
mandataires, y compris ses experts particuliers et les membres de ses comités d'études et des Comités
nationaux de la CEI, pour tout préjudice causé en cas de dommages corporels et matériels, ou de tout autre
dommage de quelque nature que ce soit, directe ou indirecte, ou pour supporter les coûts (y compris les frais
de justice) et les dépenses découlant de la publication ou de l'utilisation de cette Publication de la CEI ou de
toute autre Publication de la CEI, ou au crédit qui lui est accordé.
8) L'attention est attirée sur les références normatives citées dans cette publication. L'utilisation de publications
référencées est obligatoire pour une application correcte de la présente publication.
9) L’attention est attirée sur le fait que certains des éléments de la présente Publication de la CEI peuvent faire
l’objet de droits de propriété intellectuelle ou de droits analogues. La CEI ne saurait être tenue pour
responsable de ne pas avoir identifié de tels droits de propriété et de ne pas avoir signalé leur existence.
La Norme internationale CEI 61526 a été établie par le sous-comité 45B: Instrumentation pour
la radioprotection, du comité d'études 45 de la CEI: Instrumentation nucléaire.
Cette seconde édition annule et remplace la première édition publiée en 1998, de même que
la CEI 61283 (1995), la CEI 61323 (1995) et la CEI 61525 (1996). Cette édition constitue une
révision technique. Par rapport à la première édition, celle-ci comprend les principales
modifications techniques suivantes:
– donner des informations nécessaires pour le calcul de l'incertitude de la valeur de la dose
mesurée;
– prendre en considération les normes ISO pertinentes concernant les rayonnements de
référence et l'étalonnage, par exemple la série ISO 4037 pour les photons, l’ISO 6980
pour les rayonnements bêta et la série ISO 8529 pour les neutrons.
61526 © IEC:2005 – 9 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
RADIATION PROTECTION INSTRUMENTATION –
MEASUREMENT OF PERSONAL DOSE EQUIVALENTS HB(10) BP
AND HB(0,07) FOR X, GAMMA, NEUTRON AND BETA RADIATIONS – BP
DIRECT READING PERSONAL DOSE EQUIVALENT METERS
AND MONITORS
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61526 has been prepared by subcommittee 45B: Radiation
protection instrumentation, of IEC technical committee 45: Nuclear instrumentation.
This second edition cancels and replaces the first edition published in 1998, as well as
IEC 61283 (1995), IEC 61323 (1995) and IEC 61525 (1996). This edition constitutes a
technical revision. This edition includes the following significant technical changes from the
previous edition:
– inclusion of the determination of the uncertainty of the measured dose value;
– consideration of the relevant ISO standards on reference radiation and calibration, for
example, the ISO 4037 series for photons, ISO 6980 for beta rays and the TISO 8529
series for neutrons.
– 10 – 61526 © CEI:2005
Le texte de cette norme est issu des documents suivants:
FDIS Rapport de vote
45B/456/FDIS 45B/461/RVD
Le rapport de vote indiqué dans le tableau ci-dessus donne toute information sur le vote ayant
abouti à l'approbation de cette norme.
Cette publication a été rédigée selon les Directives ISO/CEI, Partie 2.
Le comité a décidé que le contenu de cette publication ne sera pas modifié avant la date de
maintenance indiquée sur le site web de la CEI sous «http://webstore.iec.ch» dans les
données relatives à la publication recherchée. A cette date, la publication sera
• reconduite;
• supprimée;
• remplacée par une édition révisée, ou
• amendée.
61526 © IEC:2005 – 11 –
The text of this standard is based on the following documents:
FDIS Report on voting
45B/456/FDIS 45B/461/RVD
Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.
– 12 – 61526 © CEI:2005
INTRODUCTION
La présente Norme Internationale s'applique aux appareils de mesure ou aux moniteurs
actifs, à lecture directe utilisés pour mesurer l'équivalent de dose individuel H (10) et
p
H (0,07) pour les rayonnements X, gamma, neutron et bêta.
p
Pour l'équivalent de dose individuel H (10) ou pour le débit d'équivalent de dose individuel
p
&
H (10) le domaine d'énergie couvert s'étend, pour les rayonnements X et gamma, d’au
p
moins 20 keV à 1,5 MeV et pour les rayonnements neutrons, depuis au moins les neutrons
thermiques à 15 MeV.
Pour l'équivalent de dose H (0,07) le domaine d'énergie couvert s'étend, pour les
p
rayonnements X et gamma, d’au moins 30 keV à 250 keV et pour les rayonnements bêta, à
tous ceux d'énergies moyennes supérieures à 60 keV (énergie nécessaire pour traverser la
couche morte de la peau de 0,07 mm).
Pour certaines utilisations, par exemple près d’un réacteur nucléaire, où l’on rencontre des
rayonnements photoniques de 6 MeV, il est nécessaire de mesurer l’équivalent de dose
individuel H (10) (ou son débit) jusqu’à des énergies de 10 MeV. Pour d’autres applications, il
p
sera nécessaire de mesurer H (10) jusqu’à des valeurs aussi bases que 10 keV.
p
Pour les appareils de mesure de l'équivalent de dose individuel,sont données les exigences
pour la mesure des doses H (10) et H (0,07) par les appareils de mesure de l'équivalent de
p p
& &
dose et pour la surveillance des débits de dose H (10) et H (0,07). La mesure des valeurs
p p
du débit de dose est une option pour laquelle aucune exigence ne s'applique.
Dans certains pays, les institutions nationales peuvent souhaiter que ce type d'appareil de
mesure de l'équivalent de dose individuel soit utilisé, par les services agréés de surveillance
individuelle, pour l'enregistrement des doses des travailleurs exposés au rayonnement.
61526 © IEC:2005 – 13 –
INTRODUCTION
This International Standard applies to non-passive, direct reading personal dose equivalent
meters and monitors used for measuring the personal dose equivalents H (10) and H (0,07)
p p
for X, gamma, neutron and beta radiations.
&
For the personal dose equivalent H (10) or the personal dose equivalent rate H (10) and for
p
p
X and gamma radiations, the energy range from at least 20 keV to 1,5 MeV is covered and for
neutron radiation, from at least thermal neutrons to 15 MeV.
For the personal dose equivalent H (0,07) and for X and gamma radiations, the energy range
p
from at least 30 keV to 250 keV is covered and for beta radiation, all mean energies above
60 keV (the energy required to penetrate the dead layer of the skin of 0,07 mm).
In some applications, for example at a nuclear reactor installation where 6 MeV photon
radiation is present, measurement of personal dose equivalent (rate) H (10) for photon
p
energies up to 10 MeV is required. In some other applications, measurement of H (10) down
p
to 10 keV is required.
For personal dose equivalent meters, requirements for measuring of the dose quantities
& &
H (10) and H (0,07) and for monitoring of the dose rate quantities H (10) and H (0,07) are
p p p p
given. The measurement of these dose rate quantities is an option for personal dose
equivalent meters, for which no requirements are given.
Establishments in some countries may wish to use this type of personal dose equivalent
meter as the basis of a personal monitoring service approved for dose record-keeping
purposes for radiation workers.
– 14 – 61526 © CEI:2005
INSTRUMENTATION POUR LA RADIOPROTECTION –
M
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...