prEN IEC 61788-29:2026
(Main)Electromechanical properties measurement - Electromechanical test of practical rebco and bscco composite superconductors at liquid nitrogen temperature
- Status
- Not Published
- Publication Date
- 14-Dec-2027
- Technical Committee
- CLC/SR 90 - Superconductivity
- Current Stage
- 4060 - Enquiry results established and sent to TC, SR, BTTF - Enquiry
- Start Date
- 02-Oct-2026
- Due Date
- 02-Oct-2026
- Completion Date
- 02-Oct-2026
prEN IEC 61788-29:2026 is an IEC/CENELEC draft standard in the superconductivity series. It specifies a stress-based tensile test method for measuring the electromechanical properties of practical REBCO and BSCCO composite superconductor tapes at liquid nitrogen temperature.
The document is for laboratories, manufacturers, and users who need a repeatable way to relate tensile stress to critical current loss. Its main output is the irreversible stress limit for critical current degradation.
What does prEN IEC 61788-29:2026 specify?
prEN IEC 61788-29:2026 specifies a uniaxial tensile test in which critical current is measured while the tape is loaded at liquid nitrogen temperature, without using extensometers. The method is intended to identify the stress level where critical current degradation becomes irreversible.
The document is organized into clauses on scope, references, definitions, principles, apparatus, specimen preparation, test conditions, calculation of results, uncertainty, and reporting. It also includes two informative annexes with practical guidance and uncertainty evaluation.
| Annex | What it covers |
|---|---|
| Annex A | Additional information on gripping, cross-sectional area evaluation, test setup, and the effect of recovery condition |
| Annex B | Relative standard uncertainty, analysis of variance, Type-B uncertainty, and uncertainty correlation |
What are the key requirements of prEN IEC 61788-29:2026?
prEN IEC 61788-29:2026 ties together mechanical loading and critical current measurement. In practice, the test result depends on how the specimen is mounted, cooled, loaded, and measured, so the document is very specific about setup and reporting.
Clauses 4 and 5 - principles and apparatus
Clause 4 combines a uniaxial tensile test with critical current measurement at liquid nitrogen temperature. The specimen is mounted on a loading frame with a critical current measuring system, and the irreversible stress limit may be determined either by the 99 % Ic0 recovery criterion or the 95 % Ic0 retention criterion. For a user, this means the test is not just a mechanical pull test - it is a coupled electromechanical measurement.
Clause 5 requires the apparatus to conform to ISO 7500-1 and the calibration to comply with ISO 376. The tensile machine must hold a constant crosshead speed and the cryostat must keep the specimen temperature stable and homogeneous. The grips, load transfer parts, and current terminals have to work at liquid nitrogen temperature, which matters because slippage, poor thermal design, or weak materials can invalidate the result.
Clauses 6 and 7 - specimen preparation and test conditions
Clause 6 requires careful handling to avoid bending, pre-loading, and delamination. The specimen length is defined from the grip and voltage-tap geometry, and the cross-sectional area is measured after removing insulation, using multiple measurements to account for non-uniform thickness. For the reader, this means the specimen geometry is part of the result, not just a sample preparation detail.
Clause 7 requires the specimen to be mounted at room temperature, cooled in a controlled way, and tested only after thermal equilibrium is reached. The cool-down procedure has to be documented, and any pre-loading caused by thermal contraction has to be reported. This matters because thermal stress can change the tape before the tensile test even starts.
Clauses 7.3 to 7.5 - how the test is run
Clause 7.3 uses displacement control and a stop-and-hold approach during critical current measurement. The document recommends a crosshead speed between 0.05 mm/min and 2 mm/min, and the loading rate may be changed to improve efficiency or stability. Clause 7.4 says the U-I characteristic is measured in accordance with IEC 61788-26.
Clause 7.5 defines two ways to find the irreversible stress limit. A preliminary test uses a broad stress interval to locate the transition stress RT, and the regular test then uses a narrower interval around RT. For Rrev, the specimen is unloaded after each step and the retained critical current is checked against the 99 % Ic0 recovery criterion. For Rret, the critical current is measured continuously during loading and the 95 % Ic0 retention criterion is used. This gives laboratories two practical routes, depending on whether they can or want to unload between stress steps.
Clause 9 and Clause 10 - uncertainty and reporting
Clause 9 sets the uncertainty framework. The liquid nitrogen bath is maintained between 76.8 K and 77.7 K unless otherwise specified, the force measuring cell is expected to have a relative standard uncertainty below 0.1 %, and the critical current measurement follows the IEC 61788-26 uncertainty limit. The document also states that the relative standard uncertainty for Rrev and Rret should be below 3 %.
Clause 10 lists what a report has to identify and what results it should include. In practice, this is what allows a buyer, quality team, or test lab to compare results from different specimens and batches.
What terms does prEN IEC 61788-29:2026 define?
- Tensile stress (R) - The tensile force divided by the original cross-sectional area of the specimen at any moment during the test.
- Stress interval (ΔR) - The stress range used to observe how critical current changes under tensile loading.
- Transition stress (RT) - The stress where the stress interval changes from broad to narrow while the critical current is being measured.
- Critical current (Ic) - The maximum direct current that can flow in the HTS tape specimen practically without resistance, under a specified electric field criterion and temperature.
- Irreversible stress limit for Ic degradation (Rrev or Rret) - Rrev is based on 99 % Ic0 recovery after unloading, while Rret is based on 95 % Ic0 retention during loading.
- Constant sweep rate method - A U-I measurement method where current is swept at a constant rate and the voltage-current data are recorded continuously or frequently.
- Ramp-and-hold method - A U-I measurement method where current is increased in stages and held at each stage while the voltage-current data are recorded.
Who uses prEN IEC 61788-29:2026?
prEN IEC 61788-29:2026 is used by test laboratories, superconducting tape manufacturers, and organizations that design or buy REBCO and BSCCO products. It is also relevant to makers of superconducting magnets and power devices, where winding stress and electromagnetic force can change tape performance.
Quality managers use it to define a consistent test method and a reportable stress limit. Buyers and technical reviewers use it to compare batches, check supplier claims, and interpret test reports from different laboratories.
What changed in prEN IEC 61788-29:2026 from the previous edition?
prEN IEC 61788-29:2026 supersedes 90/551/CD and 90/557/CC.
Which standards are used with prEN IEC 61788-29:2026?
- IEC 60050-815:2024 - Provides superconductivity vocabulary used in the definitions.
- IEC 61788-26 - Gives the DC critical current measurement method used for the U-I characteristic.
- IEC 61788-28 - Gives the tensile test procedure for practical REBCO and BSCCO composite superconductors at cryogenic temperatures and supports the specimen mounting approach.
- ISO 376:2011 - Covers calibration of force-proving instruments used with the testing machine.
- ISO 7500-1:2018 - Covers calibration and verification of static uniaxial testing machines.
What does the prEN IEC 61788-29:2026 document contain?
The document includes the specimen-length equation, the cross-sectional-area procedure, and the test sequence that links stress steps to critical current readings. Figure 1 shows the specimen layout with voltage taps, and Figure 2 shows how Ic/Ic0 and Icr/Ic0 are used to identify Rrev and Rret.
Annex A adds practical guidance on gripping, area measurement, the electromechanical test setup, and how recovery condition affects the result. Annex B collects round-robin data, laboratory-condition tables, statistical summaries, F-test analysis, Type-B uncertainty equations, and plots that show how uncertainty in Ic relates to uncertainty in the electromechanical properties.
Frequently Asked Questions
prEN IEC 61788-29:2026 is a draft published by CLC. Its full title is "Electromechanical properties measurement - Electromechanical test of practical rebco and bscco composite superconductors at liquid nitrogen temperature". This standard covers: Electromechanical properties measurement - Electromechanical test of practical rebco and bscco composite superconductors at liquid nitrogen temperature
Electromechanical properties measurement - Electromechanical test of practical rebco and bscco composite superconductors at liquid nitrogen temperature
prEN IEC 61788-29:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 3 FOREWORD
- 4 INTRODUCTION
- 5 1 Scope
- 6 2 Normative references
- 7 3 Terms and definitions
- 8 3.1 tensile stress R
- 9 3.2 stress interval ΔR
- 10 3.3 transition stress RT
- 11 3.4 critical current I
- 12 3.5 reversible stress limit for Ic degradation Rrev or Rret
- 13 3.6 constant sweep rate method
- 14 3.7 ramp-and-hold method
- 15 4 Principles
- 16 5 Apparatus
- 17 5.1 General
- 18 5.2 Testing machine
- 19 5.3 Critical current measuring system
- 20 6 Specimen preparation
- 21 6.1 General
- 22 6.2 Specimen length
- 23 6.3 Determination of cross-sectional area (S0)
- 24 7 Testing conditions
- 25 7.1 Specimen gripping
- 26 7.2 Cooling procedure
- 27 7.3 Testing speed
- 28 7.4 Critical current measurement
- 29 7.5 Test
- 30 8 Calculation of results
- 31 8.1 Irreversible stress limit for I degradation R or R
- c rev ret 32 9 Uncertainty of measurand
- 33 10 Test report
- 34 10.1 Identification of specimen
- 35 10.2 Reporting of test conditions
- 36 10.3 Results
- 37 Annex A
- 38 A.1 Scope
- 39 A.2 Gripping of the specimen
- 40 A.3 Evaluation of the cross-sectional area of HTS tapes
- 41 A.4 Electromechanical properties evaluation setup
- 42 A.5 Effect of recovery condition in defining the electromechanical properties of HTS 43 tapes
- 44 Annex B
162 INTRODUCTION 163 Several types of composite superconductors have now been commercialized. The rare-earth-based 164 oxide superconductor (SC) with the chemical formula RE1Ba2Cu3O7-δ is used in practical SC tapes, where 165 the rare-earth element RE is Y, Dy, Gd, Eu, Nd, Ho, and Sm or a mixture of them. This type of practical SC 166 tape is usually called a REBCO coated conductor. A typical architecture consists of a substrate of Hastelloy, 167 Ni-W alloy, or stainless steel, a buffer layer consisting of plural oxides, an SC layer, protection layers of Ag, 168 and electroplated copper layers. The substrate and buffer layers act as templates to facilitate the well- 169 oriented crystal grains of the SC layer. To resist the strong electromagnetic force, the tapes are in some 170 cases externally reinforced by laminating thin foils of stainless steel, copper, or a copper alloy. 171 Two types of bismuth–based oxide superconductors with the chemical formula (Bi, Pb)2Sr2Can-1Cun 172 O where n = 2 or 3, have been developed as practical SC tapes/wires. In either case, superconducting 2n+4 173 filaments are embedded in a pure silver matrix. This type of practical SC tape/wire is usually called BSCCO 174 tape/wire. In the Bi-2223 tapes/wires, a silver alloy outer sheath surrounds the matrix to increase mechanical 175 strength. They are made exclusively in a flat tape-shaped form. Tapes are commercially available in which 176 thin metallic sheets of copper alloy, stainless steel, or nickel alloy are soldered on both sides of the SC tapes 177 to improve the stress dependence of critical current and the mechanical properties.
178 Practical composite superconductors have a high current density and a small cross-sectional area. 179 The major application of composite superconductors is to build electrical power devices and 180 superconducting magnets. While the magnet is being manufactured, complex stresses/strains are applied 181 to its windings, and when it is energized, a large electromagnetic force is applied to the superconducting 182 tapes due to their high current density. It is therefore indispensable to determine the electromechanical 183 properties of the practical SC tapes at cryogenic temperatures.
185 SUPERCONDUCTIVITY – 186 Part 29: Electromechanical properties measurement – 187 Electromechanical test under uniaxial tensile load of REBCO and BSCCO 188 composite superconductor tapes at liquid nitrogen temperature 189 1 Scope 190 This part of IEC 61788 specifies a stress-based tensile test method for determining the electromechanical 191 properties of practical REBCO and BSCCO composite superconductor tapes at liquid nitrogen temperature 192 by measuring the critical current (Ic) under uniaxial tensile loading without the use of extensometers. This 193 stress-based test method is intended to determine the irreversible stress limits for I degradation under c 194 tensile loading.
195 When REBCO tapes are tested using this procedure, they are a rectangular cross-section with an area of 2 2 196 0,06 mm to 5,0 mm (corresponding to tapes 2,0 mm to 12,5 mm in width and 0,03 mm to 0,4 mm in 2 197 thickness). When BSCCO tapes are tested, they are a rectangular cross-section with an area of 0,3 mm to 2 198 2,5 mm (corresponding to tape-shaped ones 2,0 mm to 5,0 mm in width and 0,15 mm to 0,5 mm in 199 thickness).
200 2 Normative references 201 The following documents are referred to in the text in such a way that some or all of their content 202 constitutes requirements of this document. For dated references, only the edition cited applies. For 203 undated references, the latest edition of the referenced document (including any amendments) applies. 204 IEC 60050-815:2024, International Electrotechnical Vocabulary (IEV) – Part 815: Superconductivity 205 IEC 61788-26, Ed. 1: Superconductivity - Part 26: Critical current measurement – DC critical current of 206 RE-Ba-Cu-O composite superconductors 207 IEC 61788-28, CDV: Superconductivity – Part 28: Mechanical properties measurement – Tensile test of 208 practical REBCO and BSCCO composite superconductors at cryogenic temperatures 209 ISO 376:2011, Metallic materials – Calibration of force-proving instruments used for the verification of 210 uniaxial testing machines 211 ISO 7500-1:2018, Metallic materials – Calibration and verification of static uniaxial testing machines – Part 212 1: Tension/compression testing machines – Calibration and verification of the force-measuring system 213 3 Terms and definitions 214 For the purposes of this document, the terms and definitions given in IEC 60050-815 and ISO 6892, as 215 well as the following, apply.
216 ISO and IEC maintain terminology databases for use in standardization at the following addresses:
217 • IEC Electropedia: available at https://www.electropedia.org/ 218 • ISO Online browsing platform: available at https://www.iso.org/obp 219 3.1 220 tensile stress 221 R 222 Tensile force divided by the original cross-sectional area (So) of the test piece at any moment during the 223 test.
224 3.2 225 stress interval 226 ΔR 227 Stress range set to measure the behavior of the critical current (I ) under the tensile stress applied to the c 228 superconducting tape specimen during a stress-based tensile test.
229 3.3 230 transition stress 231 R
T 232 Stress at the point where ΔR changes from a broad to a narrow range while measuring the Ic under tensile 233 stress (see Figure 2).
234 Note: R is used to establish the irreversible stress limit for Ic degradation accurately. It typically corresponds to the stress level at T 235 which I decreases to approximately 95 % of its initial critical current (I ). Alternatively, if applicable, the transition stress may be set to c c0 236 80% of the specimen's yield stress [1]. This alternative may be used as guidance for the preliminary test described in Clause 7.5. 237 3.4 238 critical current 239 I c 240 Maximum direct current that can be regarded as flowing through an HTS tape specimen practically without 241 resistance, defined by a specified electric field criterion (E ) at a specified temperature. c 242 Note: I measured at the state before the load or deformation is applied is regarded as the initial I , I . I in the unloaded state is c c c0 c 243 referred to as the retained or residual I (I ).
c cr 244 3.5 245 irreversible stress limit for I degradation c 246 R or R rev ret 247 Irreversible stress limit for Ic degradation may be identified using two criteria: Rrev is based on the 99 % Ic0 248 recovery criterion (see Figure 2) and corresponds to the stress level just before irreversible I degradation c 249 begins during loading-unloading under tensile stress R. At this point, after unloading, the measured critical 250 current in the unloaded state (I ), which is also known as the recovered or retained I , does not recover to cr c 251 a level exceeding 99 % of Ic0 in subsequent loading-unloading cycles.
252 R is based on the 95 % I retention criterion (see Figure 2) during loading without engaging unloading and ret c0 253 corresponds to the stress at which Ic drops to 95 % of Ic0. This provides an alternative practical criterion to 254 assess the electromechanical response of HTS tapes under tensile loading.
255 3.6 256 constant sweep rate method 257 Voltage-current (U-I) data acquisition method where a current is swept at a constant rate from zero to a 258 current above I , and where the U-I data are acquired continuously or frequently.
c 259 3.7 260 ramp-and-hold method 261 U-I data acquisition method where a current is swept in stages from zero to a current above Ic, where the 262 current is held for an appropriate period at each stage, during which the U-I data are acquired 263 continuously or frequently.
264 4 Principles 265 This stress-based tensile test method combines the Ic measurement method with a uniaxial tensile test on 266 HTS tapes at liquid nitrogen temperature to determine electromechanical properties, such as the irreversible 267 stress limit for Ic degradation. The specimen shall be mounted on the loading frame of a testing machine 268 equipped with a critical current measuring system (see Figure A.1).
269 The electromechanical properties shall be evaluated by measuring Ic under controlled tensile stress 270 conditions during loading and, where applicable, unloading. The irreversible stress limits for I degradation c 271 may be determined using either: a) the 99 % Ic0 recovery criterion (R ev); or b) the 95 % Ic0 retention criterion r 272 during loading (R ).
ret 273 The U-I characteristic shall be measured in accordance with IEC 61788-26.
274 It is the responsibility of the user of this document to establish appropriate safety and health practices and 275 to determine the applicability of regulatory limitations prior to use. Specific precautionary statements are 276 given below.
277 5 Apparatus 278 5.1 General 279 The testing apparatus shall conform to the specifications outlined in ISO 7500-1. Calibration procedures 280 shall comply with ISO 376. Furthermore, this standard imposes specific additional requirements, detailed 281 as follows:
282 5.2 Testing machine 283 The testing machine shall include a tensile machine control system capable of maintaining a constant 284 crosshead speed and a cryostat system capable of maintaining a constant, homogeneous specimen 285 temperature during testing.
286 The grips shall have sufficient structure and strength appropriate for the test specimen and shall provide a 287 firm connection to the tensile machine. The grip faces may be filed or knurled, or otherwise roughened to 288 prevent specimen slippage during testing. Gripping may be a screw type, depending on the cryogenic test 289 system.
290 Structural materials used for grips and load-transferring components shall maintain sufficient strength and 291 toughness at liquid nitrogen temperature. Materials with low thermal conductivity shall be used to reduce 292 heat flow into the cryostat system. Care shall be taken to ensure that the specimen is not allowed to be 293 slack and that compressive stresses or shocks do not occur. Abrupt removal of the load can cause specimen 294 buckling, which could invalidate the test.
295 The load cell capacity shall be selected according to the expected tensile load range of the specimen. 296 Hazards exist in measurements at cryogenic conditions. Commonly, cryogenic liquids (e.g., liquid nitrogen 297 or liquid helium) are used to cool the specimen directly by bath cooling or gas exchange. Other cryogenic 298 systems may also be applied. Direct contact of the skin with cold liquid or gas transfer lines, storage Dewars, 299 or other components can cause immediate skin damage. Safety precautions for handling cryogenic devices 300 shall be observed.
301 5.3 Critical current measuring system 302 The apparatus used for measuring the U-I characteristic shall consist of a specimen mounted on the upper 303 and lower grip supports using Cu blocks for the current terminals, a Dewar with an open bath, and a U-I 304 measuring system.
305 The U-I measuring system shall include a DC source, a data acquisition system, preamplifiers, filters, 306 voltmeters, or a combination thereof. A computer-assisted data acquisition system shall be used. 307 6 Specimen preparation 308 6.1 General 309 Care shall be taken to prevent bending or pre-loading during specimen handling. The specimen shall be 310 used in its original shape, and care shall be taken when cutting specimens to avoid delamination, particularly 311 for REBCO tapes.
312 6.2 Specimen length 313 The length (L) of the test specimen shall be determined as the combined measurement of the distance 314 between the inward points of the grips and the respective lengths of the grips themselves. This definition 315 is specified as follows.
316 L = L1 + 2 × L2 + 2 × L3 + L4 (1) 317 318 Figure 1 – Schematic view of a specimen with voltage taps installed for critical current 319 measurement 320 321 L1, L2, L3 >> W (2) 322 The individual components of the L calculation are defined as follows:
323 • L is the distance between voltage taps; 1 324 • L2 is the length of the current contact; 325 • L3 is the shortest distance from a current contact to a voltage tap; 326 • L is the length of a voltage tap; 4 327 • W is the width of the specimen to be measured.
328 329 The schematic of the specimen length is shown in Figure 1.
330 The dimension L2 may be increased as the specimen's current-carrying capacity increases. This 331 increase becomes more significant when dealing with specimens that have stainless steel substrates 332 or other high-resistivity materials serving as backings or reinforcements. When higher voltage 333 sensitivity is required, L1 may be increased.
334 6.3 Determination of cross-sectional area (S ) 0 335 A micrometer or other dimension-measuring apparatus shall be used to determine the cross-sectional area 336 (S0) of the specimen after removal of the insulation coating. Multiple measurements shall be performed to 337 account for dimensional non-uniformity.
338 The thickness shall be measured using a suitable micrometer equipped with a blade or spherical anvil. This 339 choice of equipment helps reduce errors.
340 7 Testing conditions 341 7.1 Specimen gripping 342 The specimen mounting procedure shall follow IEC 61788-28.
343 The specimen shall be mounted on the upper and lower grip supports at room temperature, then cooled to 344 liquid nitrogen temperature. The specimen shall be firmly clamped using Cu blocks serving as current 345 terminals. Care shall be taken to avoid damage to the superconducting layer of the specimen (see Clause 346 A.2).
347 When the specimen is mounted on the grips, the specimen and tensile loading axis shall be aligned along 348 a straight line. Sandpaper may be inserted as a cushioning material between the specimen and the grip 349 support to prevent the gripped surfaces of the specimen from slipping or fracturing. In this case, the 350 specimen and each grip support shall be electrically insulated using suitable insulating materials such as 351 GFRP (G10) spacers. During specimen mounting, bending or deformation shall be prevented (see Clause 352 A.2).
353 7.2 Cooling procedure 354 The system shall be cooled at a controlled rate to prevent damage from thermal stresses. The cool-down 355 procedure shall be documented in the test report (setup, cool-down rate).
356 Pre-loading caused by thermal contraction of the specimen or gripping device shall be avoided. Feedback- 357 controlled fixtures or systems may be used to minimize unintended loading during cool-down. In the case 358 of pre-loading during cool-down, the exerted stress shall be reported.
359 Testing shall commence after the system has reached thermal equilibrium at liquid nitrogen temperature 360 (active boiling around the parts immersed in liquid nitrogen has ceased). The temperature within the cryostat 361 shall be monitored during testing. Add liquid nitrogen if the level drops to near the upper grip of the specimen. 362 7.3 Testing speed 363 After the specimen has been cooled to liquid nitrogen temperature, tensile loading shall be applied in 364 displacement control mode. The tensile load corresponding to the desired stress level shall be maintained 365 constant during I measurement using a stop-and-hold method.
c 366 A crosshead speed between 0,05 mm/min and 2 mm/min is recommended. Different loading rates (e.g., 2 367 and 0,05 mm/min for I measurements below and above the R ) may be used to improve measurement c T 368 efficiency and stability.
369 7.4 Critical current measurement 370 The Ic measurement procedure shall, in principle, be performed in accordance with IEC 61788-26. 371 I shall be measured after the specimen and loading frame have reached liquid nitrogen temperature. c 372 When using the constant-sweep-rate method, the sweep rate shall be selected to avoid influencing the 373 voltage measurement.
374 When using the ramp-and-hold method, the current sweep rate between stages shall be lower than the 375 equivalent of ramping from zero current to I in 3 seconds. Data acquisition at each stage shall start as soon c 376 as the flow or creep voltage generated by the current ramp can be considered negligible. The current drift 377 during each current setpoint shall be less than 1 % of I .
c 378 The U-I characteristic shall be recorded with increasing current at each stress level. 379 After completion of the measurements, the specimen shall be warmed up to room temperature. 380 7.5 Test 381 Following these procedures, the tensile machine shall be started after the testing speed is set to the 382 specified level. The stress applied to the specimen shall be calculated from the output signal of the load 383 cell.
384 The test shall consist of: a) a preliminary test for determining RT; and b) a regular test for determining Rrev 385 or R .
ret 386 Prior to tensile loading, the initial critical current Ic0 of the specimen shall be measured at liquid nitrogen 387 temperature in self-field conditions without mechanical deformation.
388 During the preliminary test step, Ic shall be roughly measured while increasing R using a broad stress interval 389 (ΔR), typically between 50 MPa and 100 MPa, depending on the tape structure. The normalized critical 390 current Ic/Ic0 shall be plotted as a function of the applied R. The transition stress (RT) shall be defined as the 391 stress level at which Ic decreases to approximately 95 % to 97% of Ic0, depending on the tape structure. 392 During the regular test, Ic under each R shall be measured using a new specimen taken from the same 393 batch. Below R , I shall be measured using a broad interval ΔR. When the applied R reaches R , I shall T c T c 394 be measured using a narrow ΔR, typically between 10 MPa and 20 MPa, or 1/5 of the broad ΔR, until 395 electrical failure occurs or irreversible degradation is observed.
396 For the determination of Rrev, unloading shall be performed to approximately 5-10 N after each loading step, 397 and the retained critical current I shall be measured. Following that, as R increases, I measurements shall cr c 398 be repeated using the loading-unloading scheme until Ic degradation becomes significant and irreversible. 399 If it is necessary to obtain I data over the entire stress range, I may be measured through unloading from cr cr 400 the beginning of the test.
401 For the determination of R , I shall be measured continuously during loading without unloading procedures. ret c 402 8 Calculation of results 403 8.1 Irreversible stress limit for I degradation R and R c rev ret 404 Figure 2 shows representative variations of Ic/Ic0 and Icr/Ic0 during loading-unloading under tensile stresses 405 (R) applied to the specimen at liquid nitrogen temperature. The figure also shows the determination of R rev 406 and Rret using the 99 % Ic0 recovery criterion and the 95 % Ic0 retention criterion during loading. 407 One curve (marked with ■) is formed by plotting I /I during loading. When I decreases to 97 % I , ΔR c c0 c c0 408 changes from a broad interval to a narrow interval at RT, and unloading begins. The other curve (marked 409 with □) is formed by plotting I /I measured in the unloaded state against the R just before unloading. cr c0 410 When measuring Ic during the loading-unloading scheme, if Icr after unloading remains above 99 % Ic0, the 411 I degradation is considered reversibly recovered. If the applied stress is further increased by the narrow c 412 ΔR (10 MPa - 20 MPa), then Ic is measured again after repeating the loading-unloading scheme. 413 If I after unloading falls below 99 % I , I is no longer reversibly recovered, and the specimen exhibits cr c0 c 414 irreversible degradation due to damage or cracks induced in the superconducting layer. The tensile stress 415 just before
...