Medical electrical equipment - Part 2-78: Particular requirements for basic safety and essential performance of medical robots for rehabilitation, assessment, compensation or alleviation

IEC 80601-2-78:2019 applies to the general requirements for BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL ROBOTS that physically interact with a PATIENT with an IMPAIRMENT to support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the PATIENT’S MOVEMENT FUNCTIONS, as intended by the MANUFACTURER. IEC 80601-2-78:2019 does not apply to • external limb prosthetic devices (use ISO 22523), • electric wheelchairs (use ISO 7176 (all parts)), • diagnostic imaging equipment (e.g. MRI, use IEC 60601-2-33), and • personal care ROBOTS (use ISO 13482).

Medizinische elektrische Geräte - Teil 2-78: Besondere Festlegungen an die Sicherheit, einschließlich der wesentlichen Leistungsmerkmale von medizinischen Robotern zur Rehabilitation, Beurteilung, Kompensation oder Linderung

Appareils électromédicaux - Partie 2-78: Exigences particulières pour la sécurité de base et les performances essentielles des robots médicaux dédiés à la rééducation, l'évaluation, la compensation ou l'atténuation

IEC 80601-2-78:2019 s'applique aux exigences générales pour la SECURITE DE BASE et les PERFORMANCES ESSENTIELLES des ROBOTS MEDICAUX avec interaction physique avec un PATIENT souffrant d’une DEFICIENCE de manière à permettre ou à effectuer une REEDUCATION, une EVALUATION, une COMPENSATION ou une ATTENUATION associée aux FONCTIONS LIEES AU MOUVEMENT DU PATIENT, comme cela est prévu par le FABRICANT. IEC 80601-2-78:2019 ne s'applique pas aux • prothèses de membres externes (utiliser l'ISO 22523), • fauteuils roulants électriques (utiliser l'ISO 7176 (toutes les parties)), • appareils d'imagerie de diagnostic (par exemple, imagerie par résonance magnétique, utiliser l'IEC 60601-2-33), et • ROBOTS d'assistance à la personne (utiliser l'ISO 13482).

Medicinska električna oprema - 2-78. del: Posebne zahteve za osnovno varnost in bistvene lastnosti medicinskih robotov za rehabilitacijo, ocenjevanje, nadomestitev funkcij ali lajšanje simptomov (IEC 80601-2-78:2019)

Standard EN-IEC 62386-105 se uporablja za krmilno opremo in krmilne naprave. Običajno enota vodila v skladu s standardom IEC 62386 (vsi deli) vključuje vdelano programsko opremo. V določenih okoliščinah bo po zaključku proizvodnje ali pošiljanju izdelka morda treba spremeniti vdelano programsko opremo (npr. če enota vodila ne deluje, kot je predvideno). V tem primeru se priporoča posodobitev vdelane programske opreme enote vodila prek vmesnika. Ta postopek posodobitve vdelane programske opreme je namenjen predvsem za odpravljanje napak in ne za razširitve funkcij. Postopek posodobitve vdelane programske opreme je kljub temu mogoče uporabiti za razširitve funkcij, vendar je pomembno pri tem upoštevati tveganje negativnih učinkov na celoten sistem.

General Information

Status
Published
Publication Date
02-Apr-2020
Current Stage
6060 - Document made available - Publishing
Start Date
03-Apr-2020
Completion Date
03-Apr-2020

Relations

Overview

EN IEC 80601-2-78:2020 (IEC 80601-2-78:2019) defines particular requirements for the basic safety and essential performance of medical robots that physically interact with patients to provide rehabilitation, assessment, compensation or alleviation of movement-related impairments. It is a part of the IEC 80601 series addressing medical electrical equipment and applies to a wide range of robot types (exoskeletons, arm/leg robot arms, cane-type devices, body-weight support systems, etc.). The standard explicitly excludes external limb prostheses (ISO 22523), electric wheelchairs (ISO 7176), diagnostic imaging devices (e.g., MRI - IEC 60601-2-33) and personal care robots (ISO 13482).

Key Topics and Requirements

  • Scope and classification of medical robots, intended use and essential performance listings.
  • Risk management integration: aligns with ISO 14971 and requires hazard analysis for patient-robot interaction, including mechanical and electrical hazards.
  • Mechanical safety: protection against moving parts, strength and stopping procedures, and tests for mechanical hazards.
  • Electrical safety & EMC: requirements referencing IEC 60601-1 collateral standards (EMC - IEC 60601-1-2).
  • Programmable systems and controls: requirements for Programmable Electrical Medical Systems (PEMS), closed-loop controllers (see IEC 60601-1-10) and alarm provisions.
  • Usability and situation awareness: usability engineering per IEC 62366-1 and informative guidance on operator/patient interaction, shared-control strategies and loss of situation awareness.
  • Documentation, marking and testing: identification, instructions for use, and test procedures to demonstrate essential performance and safety.
  • Informative annexes: extended rationale, examples (system diagrams, shared control strategies, exoskeleton/walking robot illustrations) and guidance on applying usability and risk approaches.

Applications and Users

EN IEC 80601-2-78 is intended for:

  • Manufacturers and design engineers designing medical robots (rehabilitation robots, assessment and assistive movement devices).
  • Test laboratories and conformity assessors performing safety and performance testing and CE-marking support.
  • Regulatory and procurement teams, clinical engineers and healthcare providers assessing device compliance and safe clinical deployment.
  • Usability and risk management professionals integrating human factors into design and validation.

Practical uses include product development, safety verification, risk documentation, clinical deployment planning, and regulatory submissions.

Related Standards

  • IEC 60601-1 series (basic safety and essential performance)
  • IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-1-10 (closed-loop controllers), IEC 60601-1-11 (home healthcare)
  • ISO 14971 (risk management)
  • IEC 62366-1 (usability engineering)
  • ISO 22523 (external prostheses), ISO 13482 (personal care robots), ISO 7176 (wheelchairs)

Keywords: EN IEC 80601-2-78:2020, IEC 80601-2-78, medical robot standard, rehabilitation robot safety, essential performance, medical electrical equipment.

Standard
EN IEC 80601-2-78:2020 - BARVE
English language
79 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Frequently Asked Questions

EN IEC 80601-2-78:2020 is a standard published by CLC. Its full title is "Medical electrical equipment - Part 2-78: Particular requirements for basic safety and essential performance of medical robots for rehabilitation, assessment, compensation or alleviation". This standard covers: IEC 80601-2-78:2019 applies to the general requirements for BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL ROBOTS that physically interact with a PATIENT with an IMPAIRMENT to support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the PATIENT’S MOVEMENT FUNCTIONS, as intended by the MANUFACTURER. IEC 80601-2-78:2019 does not apply to • external limb prosthetic devices (use ISO 22523), • electric wheelchairs (use ISO 7176 (all parts)), • diagnostic imaging equipment (e.g. MRI, use IEC 60601-2-33), and • personal care ROBOTS (use ISO 13482).

IEC 80601-2-78:2019 applies to the general requirements for BASIC SAFETY and ESSENTIAL PERFORMANCE of MEDICAL ROBOTS that physically interact with a PATIENT with an IMPAIRMENT to support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the PATIENT’S MOVEMENT FUNCTIONS, as intended by the MANUFACTURER. IEC 80601-2-78:2019 does not apply to • external limb prosthetic devices (use ISO 22523), • electric wheelchairs (use ISO 7176 (all parts)), • diagnostic imaging equipment (e.g. MRI, use IEC 60601-2-33), and • personal care ROBOTS (use ISO 13482).

EN IEC 80601-2-78:2020 is classified under the following ICS (International Classification for Standards) categories: 11.040.01 - Medical equipment in general. The ICS classification helps identify the subject area and facilitates finding related standards.

EN IEC 80601-2-78:2020 has the following relationships with other standards: It is inter standard links to EN IEC 80601-2-78:2020/A1:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN IEC 80601-2-78:2020 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/295. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

You can purchase EN IEC 80601-2-78:2020 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of CLC standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-julij-2020
Medicinska električna oprema - 2-78. del: Posebne zahteve za osnovno varnost in
bistvene lastnosti medicinskih robotov za rehabilitacijo, ocenjevanje,
nadomestitev funkcij ali lajšanje simptomov (IEC 80601-2-78:2019)
Medical Electrical Equipment - Part 2-78: Particular requirements for basic safety and
essential performance of medical robots for rehabilitation, assessment, compensation or
alleviation (IEC 80601-2-78:2019)
Medizinische elektrische Geräte - Teil 2-78: Besondere Festlegungen an die Sicherheit,
einschließlich der wesentlichen Leistungsmerkmale von medizinischen Robotern zur
Rehabilitation, Beurteilung, Kompensation oder Linderung (IEC 80601-2-78:2019)
Appareils électromédicaux - Partie 2-78: Exigences particulières pour la sécurité de base
et les performances essentielles des robots médicaux dédiés à la rééducation,
l'évaluation, la compensation ou l'atténuation (IEC 80601-2-78:2019)
Ta slovenski standard je istoveten z: EN IEC 80601-2-78:2020
ICS:
11.040.60 Terapevtska oprema Therapy equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EUROPEAN STANDARD EN IEC 80601-2-78

NORME EUROPÉENNE
EUROPÄISCHE NORM
April 2020
ICS 11.040.01
English Version
Medical electrical equipment - Part 2-78: Particular requirements
for basic safety and essential performance of medical robots for
rehabilitation, assessment, compensation or alleviation
(IEC 80601-2-78:2019)
Appareils électromédicaux - Partie 2-78: Exigences Medizinische elektrische Geräte - Teil 2-78: Besondere
particulières pour la sécurité de base et les performances Festlegungen an die Sicherheit, einschließlich der
essentielles des robots médicaux dédiés à la rééducation, wesentlichen Leistungsmerkmale von medizinischen
l'évaluation, la compensation ou l'atténuation Robotern zur Rehabilitation, Beurteilung, Kompensation
(IEC 80601-2-78:2019) oder Linderung
(IEC 80601-2-78:2019)
This European Standard was approved by CENELEC on 2019-08-13. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.

European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2020 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 80601-2-78:2020 E

European foreword
The text of document 62D/1676/FDIS, future edition 1 of IEC 80601-2-78, prepared by SC 62D
"Electromedical equipment" of IEC/TC 62 "Electrical equipment in medical practice" was submitted to
the IEC-CENELEC parallel vote and approved by CENELEC as EN IEC 80601-2-78:2020.
The following dates are fixed:
• latest date by which the document has to be implemented at national (dop) 2020-10-03
level by publication of an identical national standard or by endorsement
• latest date by which the national standards conflicting with the (dow) 2023-04-03
document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association, and supports essential requirements of EU
Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
Endorsement notice
The text of the International Standard IEC 80601-2-78:2019 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards
indicated:
ISO 13482:2014 NOTE Harmonized as EN ISO 13482:2014 (not modified)
ISO 9999:2016 NOTE Harmonized as EN ISO 9999:2016 (not modified)
ISO 10535:2006 NOTE Harmonized as EN ISO 10535:2006 (not modified)
IEC 60601-2-33 NOTE Harmonized as EN 60601-2-33
ISO 10218-1:2011 NOTE Harmonized as EN ISO 10218-1:2011 (not modified)
IEC 60601-1-9:2007 NOTE Harmonized as EN 60601-1-9:2008 (not modified)
IEC 61924-2:2012 NOTE Harmonized as EN 61924-2:2013 (not modified)
ISO 11064-7:2006 NOTE Harmonized as EN ISO 11064-7:2006 (not modified)

Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments)
applies.
NOTE 1  Where an International Publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.
NOTE 2  Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu.
The Annex ZA of EN 60601-1:2006 is applicable, except as follows:
Publication Year Title EN/HD Year
Replacement
IEC 60601-1-2 2014 Medical electrical equipment - Part 1-2: EN 60601-1-2 2015
General requirements for basic safety and
essential performance - Collateral
Standard: Electromagnetic disturbances -
Requirements and tests
IEC 60601-1-6 2010 Medical electrical equipment - Part 1-6: EN 60601-1-6 2010
General requirements for basic safety and
essential performance - Collateral
standard: Usability
IEC 60601-1-8 2006 Medical electrical equipment_- Part_1-8: - -
General requirements for basic safety and
essential performance_- Collateral
standard: General requirements, tests and
guidance for alarm systems in medical
electrical equipment and medical electrical
systems
ISO 14971 2007 Medical devices_- Application of risk - -
management to medical devices
Addition:
IEC 60601-1-10 2007 Medical electrical equipment - Part 1-10: EN 60601-1-10 2008
General requirements for basic safety and
essential performance - Collateral
Standard: Requirements for the
development of physiologic closed-loop
controllers
IEC 60601-1-11 2015 Medical electrical equipment_- Part_1-11: - -
General requirements for basic safety and
essential performance_- Collateral
standard: Requirements for medical
electrical equipment and medical electrical
systems used in the home healthcare
environment
Publication Year Title EN/HD Year
IEC 62366-1 2015 Medical devices - Part 1: Application of EN 62366-1 2015
usability engineering to medical devices
+AC 2015
ISO 22523  External limb prostheses and external EN ISO 22523 2006
orthoses – Requirements and test methods

IEC 80601-2-78
Edition 1.0 2019-07
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –

Part 2-78: Particular requirements for basic safety and essential performance of

medical robots for rehabilitation, assessment, compensation or alleviation

Appareils électromédicaux –
Partie 2-78: Exigences particulières pour la sécurité de base et les performances

essentielles des robots médicaux dédiés à la rééducation, l'évaluation, la

compensation ou l'atténuation
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
INTERNATIONALE
ICS 11.040.01 ISBN 978-2-8322-7000-4

– 2 – IEC 80601-2-78:2019 © IEC 2019
CONTENTS
FOREWORD . 4
INTRODUCTION . 6
201.1 Scope, object and related standards . 7
201.2 Normative references . 9
201.3 Terms and definitions . 9
201.4 General requirements . 13
201.5 General requirements for testing of ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 14
201.7 ME EQUIPMENT identification, marking and documents . 14
201.8 Protection against electrical HAZARDS from ME EQUIPMENT. 15
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 15
201.10 Protection against unwanted and excessive radiation HAZARDS . 21
201.11 Protection against excessive temperatures and other HAZARDS . 21
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 21
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 22
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 22
201.15 Construction of ME EQUIPMENT . 22
201.16 ME SYSTEMS . 24
201.17 ELECTROMAGNETIC COMPATIBILITY of ME EQUIPMENT and ME SYSTEMS . 24
202 ELECTROMAGNETIC DISTURBANCES – Requirements and tests . 24
206 USABILITY . 24
208 * General requirements, tests and guidance for alarm systems in MEDICAL
ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS . 25
210 * Process requirements for the development of PHYSIOLOGIC CLOSED-LOOP
CONTROLLERS . 25
211 * Requirements for MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL
SYSTEMS used in the HOME HEALTHCARE ENVIRONMENT . 26
Annexes . 27
Annex A (informative) General guidance and rationale . 27
Annex AA (informative) Particular guidance and rationale . 28
Annex BB (informative) Guidance and examples of SITUATION AWARENESS . 57
Bibliography . 71
Index of defined terms used in this particular standard . 74

Figure AA.1 – Relationship of the terms used to describe equipment, accessories or
equipment parts . 29
Figure AA.2 – Endsley's model of SITUATION AWARENESS (based on [10], drawn by Dr.
Peter Lankton, May 2007) . 33
Figure AA.3 – Model of user-medical device interaction . 34
Figure AA.4 – RACA ROBOT shared control system block diagram: control by PATIENT
and RACA ROBOT . 38
Figure AA.5 – RACA ROBOT shared control system block diagram: control by PATIENT,
OPERATOR and RACA ROBOT . 39

IEC 80601-2-78:2019 © IEC 2019 – 3 –
Figure AA.6 – RACA ROBOT shared control system block diagram: control by PATIENT
and RACA ROBOT, and control modulation by OPERATOR . 40
Figure AA.7 – WALKING RACA ROBOT using motion-related biosignal input . 41
Figure AA.8 – System block diagram of a WALKING RACA ROBOT using a motion-related
biosignal as input . 41
Figure AA.9 – RACA ROBOT that is an arm exoskeleton for REHABILITATION that applies a
PATIENT-cooperative shared control strategy . 42
Figure AA.10 – System block diagram of an arm exoskeleton for REHABILITATIOn that
applies a PATIENT-cooperative shared control strategy . 43
Figure AA.11 – Cane-type RACA ROBOT for REHABILITATION of walking . 44
Figure AA.12 – System block diagram of a cane-type RACA ROBOT . 44
Figure AA.13 – Example of ROBOT arm type RACA ROBOT for lower extremities . 45
Figure AA.14 – Example of ROBOT arm type RACA ROBOT for upper extremities . 46
Figure AA.15 – Example of exoskeleton type RACA ROBOT for upper extremities . 47
Figure AA.16 – Example of exoskeleton type RACA ROBOT for knee joint . 49
Figure AA.17 – Example of soft artificial muscle-type RACA ROBOT for knee joint . 50
Figure AA.18 – Example of exoskeleton-type WALKING RACA ROBOT . 51
Figure AA.19 – Example of RACA ROBOT for balance control . 52
Figure AA.20 – Example of a body-weight support-type RACA ROBOT with gait following

function . 54
Figure BB.1 – All the proximate causes of loss of SITUATION AWARENESS [19] . 58
Figure BB.2 – Relationship between SITUATION AWARENESS, the RISK MANAGEMENT
PROCESS (ISO 14971:2007) and the USABILITY ENGINEERING PROCESS (IEC 62366-
1:2015) . 60
Figure BB.3 – Relationship between GDTA and RISK MANAGEMENT and USABILITY
........................................................................................................ 63
ENGINEERING PROCESSES
Figure BB.4 – WALKING exoskeleton RACA ROBOT . 69

Table 201.101 – List of potential ESSENTIAL PERFORMANCEs . 14
Table 19 – MECHANICAL HAZARDS covered by this clause . 16
Table 201.102 – Overview of different stopping procedures . 16
Table 28 – Mechanical strength test applicability . 23
Table 1 – Mechanical strength test applicability, non-TRANSIT-OPERABLE . 26
Table 2 – Mechanical strength test applicability, TRANSIT-OPERABLE . 26
Table AA.1 – Correlation mapping between Figure AA.2 and Figure AA.3 . 35
Table BB.1 – Example of using GDTA in BB.5.2 . 64
Table BB.2 – Example of using GDTA in BB.5.3 . 66
Table BB.3 – Example of using GDTA in BB.5.4 . 68
Table BB.4 – Example of using GDTA in BB.5.5 . 70

– 4 – IEC 80601-2-78:2019 © IEC 2019
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________
MEDICAL ELECTRICAL EQUIPMENT –

Part 2-78: Particular requirements for basic safety and essential
performance of medical robots for rehabilitation, assessment,
compensation or alleviation
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as "IEC
Publication(s)"). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 80601-2-78 has been prepared by IEC subcommittee 62D:
Electromedical equipment, of IEC Technical Committee 62: Electrical equipment in medical
practice, and ISO Technical Committee 299: Robotics.
This publication is published as a double logo standard.
The text of this International Standard is based on the following documents:
FDIS Report on voting
62D/1676/FDIS 62D/1688/RVD
Full information on the voting for the approval of this International Standard can be found in
the report on voting indicated in the above table.
This document has been drafted in accordance with the ISO/IEC Directives, Part 2.

IEC 80601-2-78:2019 © IEC 2019 – 5 –
In this document, the following print types are used:
– requirements and definitions: roman type;
– test specifications: italic type;
informative material appearing outside of tables, such as notes, examples and references in smaller type.

Normative text of tables is also in a smaller type;
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this document, the term
– "clause" means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this document are preceded by the term "Clause" followed by the
clause number. References to subclauses within this particular standard are by number only.
In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any
combination of the conditions is true.
The verbal forms used in this document conform to usage described in Clause 7 of the
ISO/IEC Directives, Part 2. For the purposes of this document, the auxiliary verb:
– "shall" means that compliance with a requirement or a test is mandatory for compliance
with this document;
– "should" means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this document;
– "may" is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 80601 and IEC 60601 International Standard, published under the
general title Medical electrical equipment, can be found on the IEC website.
The committee has decided that the contents of this document will remain unchanged until the
stability date indicated on the IEC website under "http://webstore.iec.ch" in the data related to
the specific document. At this date, the document will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of users of this document is drawn to the fact that equipment manufacturers and testing
organizations may need a transitional period following publication of a new, amended or revised IEC publication in
which to make products in accordance with the new requirements and to equip themselves for conducting new or
revised tests. It is the recommendation of the committees that the content of this publication be adopted for
implementation nationally not earlier than 3 years from the date of publication.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.
– 6 – IEC 80601-2-78:2019 © IEC 2019
INTRODUCTION
This part of IEC 80601 International Standard was written at a time when technical evolution
of MEDICAL ROBOTS was in rapid progress and the scientific foundation of safe use was still
being expanded.
This document is the result of work that began in ISO/TC 184/SC 2/WG 7 in October 2006 on
personal care ROBOTS, to address an emerging type of MEDICAL ROBOT that was used outside
of an industrial environment. That group was working on a new standard, ISO 13482, which
was published as an International Standard (IS) in 2014. While initially focused on non-
medical applications, WG 7 recognized that work was needed on medical devices utilizing
robotic technology. In October 2009, ISO/TC 184/SC 2 established a WG 7, Study Group (SG)
on Medical care robots, comprised of experts from Canada, France, Germany, Japan, Korea,
Romania, Switzerland, UK and USA.

The work of ISO/TC 184/SC 2/WG 7 SG cumulated in a proposal to form a Joint Working
Group (JWG 9) with IEC/TC 62/SC 62A focusing on MEDICAL ELECTRICAL EQUIPMENT using
robotic technology. This JWG began developing a technical report (IEC TR 60601-4-1)
dealing with degree of autonomy. While developing this document, a particular standard was
deemed required for REHABILITATION type ROBOTS. This led to the creation of a Joint Working
Group 36 (MEDICAL ROBOTS for REHABILITATION) in April, 2015 within IEC/TC 62/SC 62D to
develop particular requirements of SAFETY of MEDICAL ELECTRICAL EQUIPMENT and MEDICAL
ELECTRICAL SYSTEMS for REHABILITATION type ROBOTS. ISO/TC 184/SC 2 has since been
promoted to ISO/TC 299, and JWG 9 has merged with JWG35 and 36 to form JWG 5
(MEDICAL ROBOT Safety) on the ISO side. This proposal was approved from both IEC and
ISO and work began.
The minimum safety requirements specified in this particular standard are presented to
provide for an acceptable degree of BASIC SAFETY and ESSENTIAL PERFORMANCE for MEDICAL
ROBOTS that physically interact with a PATIENT with an IMPAIRMENT, to support or perform
REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the PATIENT’S MOVEMENT
FUNCTIONS .
The requirements are followed by particular specifications for the relevant tests.

IEC 80601-2-78:2019 © IEC 2019 – 7 –
MEDICAL ELECTRICAL EQUIPMENT –

Part 2-78: Particular requirements for basic safety and essential
performance of medical robots for rehabilitation, assessment,
compensation or alleviation
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This part of IEC 80601 applies to the general requirements for BASIC SAFETY and ESSENTIAL
PERFORMANCE of MEDICAL ROBOTS that physically interact with a PATIENT with an IMPAIRMENT to
support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION related to the
PATIENT’S MOVEMENT FUNCTIONS, as intended by the MANUFACTURER.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
NOTE See also 4.2 of the general standard.
This particular standard does not apply to
• external limb prosthetic devices (use ISO 22523),
• electric wheelchairs (use ISO 7176 (all parts)),
• diagnostic imaging equipment (e.g. MRI, use IEC 60601-2-33), and
• personal care ROBOTS (use ISO 13482).
201.1.2 Object
Replacement:
BASIC SAFETY and ESSENTIAL
The object of this particular standard is to establish particular
PERFORMANCE requirements for MEDICAL ROBOTS that physically interact with a PATIENT with an
IMPAIRMENT, to support or perform REHABILITATION, ASSESSMENT, COMPENSATION or ALLEVIATION
related to the PATIENT’S MOVEMENT FUNCTIONS.
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
____________
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance.

– 8 – IEC 80601-2-78:2019 © IEC 2019
IEC 60601-1-2:2014, IEC 60601-1-6:2010 and IEC 60601-1-6:2010/AMD1:2013, IEC 60601-1-
8:2006 and IEC 60601-1-8:2006/AMD1:2012, IEC 60601-1-10:2007 and IEC 60601-1-
10:2007/AMD1:2013, and IEC 60601-1-11:2015 apply as modified in Clauses 202, 206, 208,
210 and 211 respectively. IEC 60601-1-3 and IEC 60601-1-12 do not apply. All other
published collateral standards in the IEC 60601-1 series apply as published.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012 are referred to in this
particular standard as the general standard. Collateral standards are referred to by their
document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix "201" (e.g. 201.1 in this document addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix "20x"
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the 60601-1-2 collateral standard,
203.4 in this particular standard addresses the content of Clause 4 of the IEC 60601-1-3
collateral standard, etc.). The changes to the text of the general standard are specified by the
use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.147, additional definitions in this document are
numbered beginning from 201.3.201. Additional Annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where "x" is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this document" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.

IEC 80601-2-78:2019 © IEC 2019 – 9 –
201.2 Normative references
NOTE Informative references are listed in the Bibliography.
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic disturbances –
Requirements and tests
IEC 60601-1-6:2010, Medical electrical equipment – Part 1-6: General requirements for basic
safety and essential performance – Collateral standard: Usability
IEC 60601-1-6:2010/AMD1:2013
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for safety
and essential performance – Collateral Standard: General requirements, tests and guidance
for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-8:2006/AMD1:2012
ISO 14971:2007, Medical devices – Application of risk management to medical devices
Addition:
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for the
development of physiologic closed-loop controllers
IEC 60601-1-10:2007/AMD1:2013
IEC 60601-1-11:2015, Medical electrical equipment – Part 1-11: General requirements for
basic safety and essential performance – Collateral Standard: Requirements for medical
electrical equipment and medical electrical systems used in the home healthcare environment
IEC 62366-1:2015, Medical devices – Part 1: Application of usability engineering to medical
devices
ISO 22523:2006, External limb prostheses and external orthoses – Requirements and test
methods
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012 apply.
ISO and IEC maintain terminological databases for use in standardization at the following
addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
NOTE An Index of defined terms is found beginning on page 74.
Amendment:
– 10 – IEC 80601-2-78:2019 © IEC 2019
201.3.63
* MEDICAL ELECTRICAL EQUIPMENT
ME EQUIPMENT
Addition:
Note 1 to entry: See Figure AA.1.
201.3.65
MOBILE
Addition:
Note 1 to entry: MOBILE also includes equipment intended to support the movement of a PATIENT from one location
to another.
201.3.144
* BODY-WORN
Replacement:
PORTABLE equipment whose INTENDED USE includes operation while being worn by a PATIENT or
attached to a PATIENT’S clothing
Addition:
201.3.201
*ACTUATED APPLIED PART
subcategory of APPLIED PART that is intended to provide actively controlled physical
interactions with the PATIENT that are related to the PATIENT’s MOVEMENT FUNCTIONS, to perform
a CLINICAL FUNCTION of a RACA ROBOT
Note 1 to entry: "Actively controlled" as used above is intended to mean controlled by the RACA ROBOT, including
shared control with the PATIENT or OPERATOR.
Note 2 to entry: Actively controlled physical interactions include position control, force control, impedance control,
admittance control, or any other controls that regulate the interaction between a RACA ROBOT and the PATIENT.
Note 3 to entry: Each ACTUATED APPLIED PART is part of an actuation system according to 9.8.1 of the general
standard.
201.3.202
ALLEVIATION
treatment to ease symptoms due to an IMPAIRMENT of a PATIENT
Note 1 to entry: An example of ALLEVIATION is physical therapy performed by a RACA ROBOT to reduce pain or
other secondary effects of an IMPAIRMENT.
201.3.203
* ASSESSMENT
procedure to quantify or to aid in the qualification of the level of IMPAIRMENT of a PATIENT.
Note 1 to entry: The term "ASSESSMENT" should not be confused with the term "RISK ASSESSMENT’"
Note 2 to entry: ASSESSMENT in this definition focuses on the level of IMPAIRMENT of body functions and structures
and not of activity limitations or participation restrictions ([1] , Chapter 4).
Note 3 to entry: ASSESSMENT can be distinguished from both measurement and diagnosis. Measurement typically
refers to the physiological parameter being measured (e.g. PATIENT hand position by a RACA ROBOT) and is typically
related to, or is the direct output of a transducer. Those measurements are then used to quantify an IMPAIRMENT
(e.g., proprioception, planning of a movement, etc). Because IMPAIRMENTs are not the same as the underlying
pathology, but are the manifestations of an underlying pathology, an ASSESSMENT that quantifies the IMPAIRMENT
____________
Numbers in square brackets refer to the Bibliography.

IEC 80601-2-78:2019 © IEC 2019 – 11 –
can then be used by a clinician to support a diagnosis that identifies the pathology, or to monitor the progression of
a pathology or recovery from an injury. See Annex AA for an example.
201.3.204
CLINICAL FUNCTION
clinically significant operation that the ME EQUIPMENT or ME SYSTEM is intended to perform, as
specified by the MANUFACTURER
Note 1 to entry: CLINICAL FUNCTION is generally a subset of the INTENDED USE of the ME EQUIPMENT or ME SYSTEM
and is related to the PATIENT.
Note 2 to entry: In the context of this document, a clinically significant operation refers to:
• support of the diagnosis of a PATIENT with an IMPAIRMENT,
• treatment, mitigation, or monitoring of an IMPAIRMENT of a PATIENT.
[SOURCE: IEC TR 60601-4-1:2017, 3.6, modified – In the definition, the words "medical
operation" have been replaced by "clinically significant operation", and Note 2 to entry has
been added.]
201.3.205
COMPENSATION
mitigation of IMPAIRMENT of a PATIENT through support of body structures or through support or
replacement of body functions
Note 1 to entry: COMPENSATION could be provided by, for example, externally powered orthoses. COMPENSATION
does not include the improvement of MOVEMENT FUNCTIONS related to an IMPAIRMENT, which is defined as
REHABILITATION.
201.3.206
EMERGENCY STOP
manually initiated interruption of operation intended to stop the RACA ROBOT to prevent HARM
Note 1 to entry: EMERGENCY STOP is meant as a last resort when no other means to reduce RISK are available.
Note 2 to entry: EMERGENCY STOP does not allow an automatic restart, in contrast to a PROTECTIVE STOP.
Note 3 to entry: The activation of an EMERGENCY STOP does not necessarily shut down all the power in a RACA
ROBOT. Refer to 9.2.4 of the general standard for further clarification.
201.3.207
* IMPAIRMENT
problem in body function or structure, such as significant deviation or loss
Note 1 to entry: "Problems" here is used to refer to a (negative) deviation from generally accepted population
normality levels in the biomedical status of the body and its functions. Notification of the constituents of IMPAIRMENT
is undertaken primarily by those qualified to judge functioning according to the generally accepted normality levels.
An IMPAIRMENT usually follows a disease or accident, but also includes birth defects.
Note 2 to entry: This document only concerns ME EQUIPMENT or ME SYSTEM, meaning that it addresses IMPAIRMENT
of a PATIENT and not normal age-related IMPAIRMENTS.
[SOURCE: ISO 9999:2016, 2.11, modified – The term is here in singular instead of plural. The
notes to entry have been added.]
201.3.208
MEDICAL ROBOT
ROBOT intended to be used as ME EQUIPMENT or ME SYSTEM
[SOURCE: IEC TR 60601-4-1:2017, 3.20]

– 12 – IEC 80601-2-78:2019 © IEC 2019
201.3.209
MOVEMENT FUNCTION
function consisting of one or more of sensory, neuromusculoskeletal or movement-related
body functions that comprise PATIENT motor control
Note 1 to entry: ”Sensory”, "neuromusculoskeletal" and "movement-related" come from the body functions
classification presented in [1].
Note 2 to entry: A RACA ROBOT does not need to address all of these aspects.
201.3.210
* PROTECTIVE STOP
interruption of operation automatically initiated by the RACA ROBOT, that allows a cessation of
motion for BASIC SAFETY and ESSENTIAL PERFORMANCE purposes, which could allow the
PROGRAMMABLE ELECTRICAL MEDICAL SYSTEM (PEMS)/PROGRAMMABLE ELECTRONIC SUBSYSTEM
(PESS) to facilitate a restart
Note 1 to entry: See also ISO 8373:2012, 5.17.
Note 2 to entry: A PROTECTIVE STOP is defined in contrast to an EMERGENCY STOP or a normal stop; clarification is
provided in Table 201.102.
Note 3 to entry: “Automatically initiated” means that the RACA ROBOT is detecting a potentially HAZARDOUS
SITUATION and as a result of this detection the RACA ROBOT then initiates a stopping procedure
Note 4 to entry: The defined term is related to FALLBACK MODE defined in IEC 60601-1-10:2007, 3.11.
Note 5 to entry: Even during the cessation of motion, the actuators are not necessarily deactivated. In certain
conditions, the actuators may hold the PATIENT or PATIENT’S limbs in a stable position.
201.3.211
* REHABILITATION
treatment to improve MOVEMENT FUNCTIONS related to an IMPAIRMENT of a PATIENT
EXAMPLE Relearning of motor control of the upper or lower extremities, restoration of muscle strength and
endurance.
Note 1 to entry: REHABILITATION can be carried out in situations where a PATIENT has an IMPAIRMENT due to an
accident or disease or congenital conditions (e.g. cerebral palsy) or can be used to slow down the expected loss of
body functions due to neurodegenerative diseases (e.g. Parkinson’s, multiple sclerosis).
Note 2 to entry: This definition differs from the World Health Organization (WHO) definition and covers only a
subset of that definition, in order to agree with the scope of this document.
201.3.212
REHABILITATION, ASSESSMENT, COMPENSATION AND ALLEVIATION ROBOT
RACA ROBOT
MEDICAL ROBOT intended by its MANUFACTURER to perform REHABILITATION, ASSESSMENT,
COMPENSATION or ALLEVIATION comprising an ACTUATED APPLIED PART
201.3.213
*ROBOT
programmed actuated mechanism with a degree of autonomy, moving within its environment,
to perform intended tasks
201.3.214
* SITUATION AWARENESS
OPERATOR’S perception, comprehension and projection of a RACA ROBOT’s behaviour in its
environment
Note 1 to entry: SITUATION AWARENESS allows the OPERATOR of the RACA ROBOT to react appropriately and timely
to the RACA ROBOT's behaviour both under NORMAL USE and in the case of a SINGLE FAULT CONDITION.
Note 2 to entry: SITUATION AWARENESS has a direct link to US
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

The article discusses EN IEC 80601-2-78:2020, which is a standard that applies to medical robots used in rehabilitation, assessment, compensation, or alleviation of patients with impairments. The standard outlines the basic safety and essential performance requirements for these medical robots. However, the standard does not apply to external limb prosthetic devices, electric wheelchairs, diagnostic imaging equipment, or personal care robots, as they have their own specific standards.

기사 제목: EN IEC 80601-2-78:2020 - 의료용 전기 장비 - 파트 2-78: 재활, 평가, 보상 또는 경감을 위한 의료용 로봇의 기본 안전 및 필수 성능에 대한 특정 요구 사항 기사 내용: IEC 80601-2-78:2019은 장애를 가진 환자와 물리적으로 상호작용하는 의료용 로봇의 기본 안전 및 필수 성능에 대한 일반 요구 사항에 적용됩니다. 이러한 로봇은 환자의 움직임 기능과 관련하여 지원하거나 수행하기 위해 제조업체의 의도와 같이 사용됩니다. 그러나 IEC 80601-2-78:2019은 외부 근지질 장치(ISO 22523 사용), 전동 휠체어(ISO 7176 (전체 부분) 사용), 진단 영상 장비(예: MRI, IEC 60601-2-33 사용) 및 개인용 돌봇(ISO 13482 사용)에는 적용되지 않습니다.

記事タイトル:EN IEC 80601-2-78:2020 - 医療電気機器 - 第2-78部:リハビリ、評価、補償または緩和用の医療用ロボットの基本的な安全性と必須性能に関する特定の要件 記事内容:IEC 80601-2-78:2019は、患者と物理的に相互作用する医療用ロボットの基本的な安全性と必須性能に関する一般的な要件に適用されます。これらのロボットは、製造業者の意図に従って、患者の運動機能に関連するリハビリテーション、評価、補償または緩和をサポートまたは実行するために使用されます。ただし、IEC 80601-2-78:2019は、外部義肢装置(ISO 22523を使用)、電動車いす(ISO 7176(全てのパーツ)を使用)、診断用画像装置(例:MRI、IEC 60601-2-33を使用)、個人介護ロボット(ISO 13482を使用)には適用されません。