CEN/TS 15413:2006
(Main)Solid recovered fuels - Methods for the preparation of the test sample from the laboratory sample
Solid recovered fuels - Methods for the preparation of the test sample from the laboratory sample
This Technical Specification specifies the correct sequence of operations to ensure the representativity of the test portions that has been taken according to the sampling plan, prior to physical and/or chemical analysis (e.g. extractions, digestion and/or analytical determinations) of solid samples.
This Technical Specification specifies the correct sequence of operations and treatments to be applied to the laboratory sample in order to obtain suitable test portions in compliance with the specific requirements defined in the corresponding analytical procedures.
Feste Sekundärbrennstoffe - Verfahren zur Herstellung der Versuchsprobe aus der Laboratoriumsprobe
Combustibles solides de récupération - Méthodes pour la préparation d'échantillons pour essais à partir d'échantillons de laboratoire
Trdno alternativno gorivo - Metode za pripravo preskusnega vzorca iz laboratorijskega vzorca
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-marec-2007
Trdno alternativno gorivo - Metode za pripravo preskusnega vzorca iz
laboratorijskega vzorca
Solid recovered fuels - Methods for the preparation of the test sample from the laboratory
sample
Feste Sekundärbrennstoffe - Verfahren zur Herstellung der Versuchsprobe aus der
Laboratoriumsprobe
Combustibles solides de récupération - Méthodes pour la préparation d'échantillons pour
essais a partir d'échantillons de laboratoire
Ta slovenski standard je istoveten z: CEN/TS 15413:2006
ICS:
75.160.10 Trda goriva Solid fuels
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
TECHNICAL SPECIFICATION
CEN/TS 15413
SPÉCIFICATION TECHNIQUE
TECHNISCHE SPEZIFIKATION
September 2006
ICS 75.160.10
English Version
Solid recovered fuels - Methods for the preparation of the test
sample from the laboratory sample
Combustibles solides de récupération - Méthodes pour la Feste Sekundärbrennstoffe - Verfahren zur zur Herstellung
préparation d'échantillons pour essais à partir d'échantillons einer Prüfprobe aus einer Laborprobe
de laboratoire
This Technical Specification (CEN/TS) was approved by CEN on 25 March 2006 for provisional application.
The period of validity of this CEN/TS is limited initially to three years. After two years the members of CEN will be requested to submit their
comments, particularly on the question whether the CEN/TS can be converted into a European Standard.
CEN members are required to announce the existence of this CEN/TS in the same way as for an EN and to make the CEN/TS available
promptly at national level in an appropriate form. It is permissible to keep conflicting national standards in force (in parallel to the CEN/TS)
until the final decision about the possible conversion of the CEN/TS into an EN is reached.
CEN members are the national standards bodies of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France,
Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania,
Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2006 CEN All rights of exploitation in any form and by any means reserved Ref. No. CEN/TS 15413:2006: E
worldwide for CEN national Members.
Contents Page
Foreword.3
Introduction .4
1 Scope .5
2 Normative references .5
3 Terms and definitions .5
4 Safety remarks .6
5 Principle.6
6 Apparatus .6
7 Interferences and sources of error .6
8 Procedure .7
9 Quality control.10
10 Performance characteristics .10
11 Test report .10
Annex A (normative) Guideline for choosing sample treatment techniques .11
Annex B (informative) Relationship between minimum amount of sample and particle size.20
Annex C (informative) Sample treatment equipment.23
Annex D (normative) Guidelines - Characteristics of the laboratory sample for chemical analysis
of SRF .24
Bibliography .26
Foreword
This document (CEN/TS 15413:2006) has been prepared by Technical Committee CEN/TC 343 “Solid
Recovered Fuels”, the secretariat of which is held by SFS.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to announce this Technical Specification: Austria, Belgium, Cyprus, Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden,
Switzerland and United Kingdom.
Introduction
In laboratory praxis different analytical procedures need to be applied very often to the laboratory sample that
has been taken according to the sampling plan. For this purpose sub-sampling is applied in a way, that the
different test portions are representative for the original laboratory sample with respect to the compounds of
interest and the specific analytical procedures. The representativity of the laboratory sample and of the test
portions is of major importance to guarantee the quality and accuracy of analytical results. The representativity
of the laboratory sample is specified by the sampling plan.
This Technical Specification is largely based on the work already done by CEN/TC 292 "Characterization of
waste", and in particular on latest drafts of just published EN 15002; actually, some experts who developed
EN 15002 actively participated to the preparation of this Technical Specification as well.
EN 15002 was developed for the majority of waste samples, and most of its concepts and specifications are
actually applicable to SRF samples as well, but there would be a number of major problems:
several points of Annex A (normative) of EN 15002:2006 ("Guideline for choosing sample treatment
techniques") are simply not applicable for SRF samples, due to the very particular nature of these
samples, and in some cases this could be actually misleading.
the main peculiarity that makes SRF samples significantly different from other kind of waste is that very
often SRFs are solid, but neither "granular" nor monolithic; it often happens that SRF samples are fibrous-
like materials, so the statistical formula for sampling (Annex B normative of EN 15002:2006, that links the
minimum amount of sample depending on the particle size and other parameters), that is one of the
foundations of EN 15002, is not applicable "as it is": one more term in the statistical equation is needed,
namely the "shape factor" (s).
all examples contained in Annex E of EN 15002:2006 are just not applicable for SRF samples, which may
lead users who need to analyze SRF samples to misunderstandings.
Because of these reasons, a significant revision of just-published EN 15002 would have been necessary in
order to fulfil all requirements for SRF samples, which presumably had better to be carried out jointly by CEN/
TC 292 and TC 343. Moreover, even other CEN/TC 292 standards and TSs on sampling of waste would have
become inconsistent and would have had to be revised in order to include the "shape factor" in the statistical
formula. However, all of this work would probably have caused unacceptable delays for both CEN TCs. So,
CEN TC 343 decided to proceed with the development of a new Technical Specification.
1 Scope
This Technical Specification specifies the correct sequence of operations to ensure the representativity of the
test portions that has been taken according to the sampling plan, prior to physical and/or chemical analysis
(e.g. extractions, digestion and/or analytical determinations) of solid samples.
This Technical Specification specifies the correct sequence of operations and treatments to be applied to the
laboratory sample in order to obtain suitable test portions in compliance with the specific requirements defined
in the corresponding analytical procedures.
2 Normative references
The following referenced documents are indispensable for the application of this Technical Specification. For
dated references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
CEN/TS 15357:2006, Solid recovered fuels — Terminology, definitions and descriptions
prCEN/TS 15443, Solid recovered fuels — Methods for laboratory sample preparation
3 Terms and definitions
For the purposes of this Technical Specification, the terms and definitions given in CEN/TS 15357:2006 and
the following apply.
3.1
drying
process of removing water from a sample
NOTE For the purpose of test portion preparation, it may be useful to remove just the amount of water that could
interfere with other processes involved (e.g. during crushing or milling). In order to minimise the alteration of the sample
during test portion preparation, removing the total amount of water present in the sample is not necessarily needed.
3.2
fraction separation
process of dividing components, particles or layers if homogenisation of the sample is practically not
applicable and/or the analyses of different fractions or phases are appropriate
3.3
homogenisation
process of combining of components, particles or layers into a more homogeneous state of the original
samples (in the case of composite samples) or pre-treated fractions of samples in order to ensure equal
distribution of substances in and properties of the sample
3.4
sub-sampling
process of selecting one or more sub-samples from a sample
3.5
test portion; analytical portion
quantity of material of proper size, for measurement of the concentration or other properties of interest,
removed from the test sample
NOTE The test portion may be taken from the laboratory sample directly if no preparation of sample is required (e.g.
with liquids or samples of proper homogeneity, size and fineness), but usually it is taken from the prepared test sample.
3.6
test sample; analytical sample
sample, prepared from the laboratory sample, from which test portions are removed for testing or analysis
NOTE 1 When the laboratory sample is further prepared (reduced) by subdividing, mixing, grinding, or by combinations
of these operations, the result is the test sample. When no preparation of the laboratory sample is required, the laboratory
sample is the test sample. A test portion is removed from the test sample for the performance of the test or for analysis.
NOTE 2 The laboratory sample is the final sample from the point of view of sample collection but it is the initial sample
from the point of view of the laboratory.
NOTE 3 Several laboratory samples may be prepared and sent to different laboratories or to the same laboratory for
different purposes. When sent to the same laboratory, the set is generally considered as a single laboratory sample and is
documented as a single sample.
4 Safety remarks
The safety in handling of potentially hazardous materials is dealt with relevant national and European
regulations, which every laboratory should refer to.
In addition the following information is given:
the apparatus for grinding, cutting, milling, and homogenisation may result harmful for the users. They
have to be operated by skilled personnel strictly according to the manufacturer instructions;
all procedures have to be performed in a hood or in closed force-ventilated equipment, due to the
possibility of generation of fine powders.
5 Principle
The laboratory sample is reduced in particle size and mass using different apparatus and procedures
depending on the type of sample and the type of analysis to which the sample will be submitted.
6 Apparatus
For the purpose of preparation of test portions from the laboratory samples appropriate equipment has to be
chosen depending on the procedures selected according to Annex A.
In the selection of the type of treatment techniques, one should keep in mind that each of them has some
potential impact on analytical results, because it can introduce contamination or alter the physical-chemical
properties of the sample.
All glassware and devices that come in contact with the sample shall be made out of a suitable material,
chemically compatible with the sample, selected in order to minimize contamination of samples. Care shall be
taken to ensure a good cleaning, in order to avoid cross-contamination of samples.
An informative list of appropriate equipment for the sample treatment procedures is given in Annex C.
7 Interferences and sources of error
The (sub)-sample shall be re-homogenised after any operation that may have resulted in segregation of
different sized particles.
Care should be taken to avoid loss of material and contamination of the sample via the air, by dust, by the use
of the apparatus (e.g. from the ambient laboratory atmosphere or between samples stored or processed close
to one another).
Three types of contamination could occur from the apparatus:
abrasion;
cross-contaminati
...