EN 17346:2020
(Main)Ambient air - Standard method for the determination of the concentration of ammonia using diffusive samplers
Ambient air - Standard method for the determination of the concentration of ammonia using diffusive samplers
This document specifies a method for the sampling and analysis of NH3 in ambient air using diffusive sampling.
It can be used for NH3 measurements at ambient levels, but the concentration range and exposure time are sampler dependent, and the end user is therefore advised to match the sampler type to the measurement requirement and to follow the operating instructions provided by the manufacturer.
Außenluft - Messverfahren zur Bestimmung der Konzentration von Ammoniak mit Passivsammlern
Dieses Dokument legt ein Verfahren zur Probenahme und Analyse von NH3 in Außenluft mit Passivsammlern fest.
Es kann für NH3 Messungen bei Außenluftkonzentrationen verwendet werden; Konzentrationsbereich und Expositionszeit sind aber sammlerabhängig. Dem Anwender wird daher empfohlen, einen den Messanforderungen entsprechenden Sammlertyp zu wählen und sich an die Herstellerangaben zu halten.
Air ambiant - Méthode normalisée pour la détermination de la concentration en ammoniac au moyen d’échantillonneurs par diffusion
Le présent document spécifie une méthode pour l’échantillonnage et l’analyse du NH3 dans l’air ambiant à l’aide d’échantillonneurs par diffusion.
Elle peut être utilisée pour mesurer le NH3 aux niveaux ambiants, mais la gamme de concentrations et la durée d’exposition dépendent de l’échantillonneur. Il est donc conseillé à l’utilisateur final d’adapter le type d’échantillonneur aux exigences de mesure et de se conformer aux instructions d’utilisation fournies par le fabricant.
Kakovost zunanjega zraka - Standardna metoda za določevanje koncentracije amoniaka z difuzijskim vzorčenjem
General Information
Standards Content (Sample)
SLOVENSKI STANDARD
01-julij-2020
Kakovost zunanjega zraka - Standardna metoda za določevanje koncentracije
amoniaka z difuzijskim vzorčenjem
Ambient Air Quality - Standard method for the determination of the concentration of
ammonia by diffusive sampling
Außenluftqualität - Messverfahren zur Bestimmung der Konzentration von Ammoniak mit
Passivsammlern
Air ambiant - Méthode normalisée pour la détermination de la concentration d'ammoniac
au moyen d'échantillonneurs par diffusion
Ta slovenski standard je istoveten z: EN 17346:2020
ICS:
13.040.20 Kakovost okoljskega zraka Ambient atmospheres
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EN 17346
EUROPEAN STANDARD
NORME EUROPÉENNE
May 2020
EUROPÄISCHE NORM
ICS 13.040.20
English Version
Ambient air - Standard method for the determination of
the concentration of ammonia using diffusive samplers
Air ambiant - Méthode normalisée pour la Außenluft - Messverfahren zur Bestimmung der
détermination de la concentration en ammoniac au Konzentration von Ammoniak mit Passivsammlern
moyen d'échantillonneurs par diffusion
This European Standard was approved by CEN on 13 April 2020.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2020 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 17346:2020 E
worldwide for CEN national Members.
Contents Page
European foreword . 4
Introduction . 5
1 Scope . 9
2 Normative references . 9
3 Terms and definitions . 9
4 Description of samplers . 11
4.1 Principle . 11
4.2 Implementation . 11
4.3 Tube-type samplers. 11
4.4 Badge-type samplers . 12
4.5 Radial samplers . 12
5 Calculation of the concentration of NH . 12
5.1 Mass concentration . 12
5.2 Conversion to standard conditions of temperature and pressure . 13
6 Quality control/quality assurance . 13
6.1 Quality control . 13
6.2 Quality assurance . 14
7 Report . 14
8 Performance requirements and measurement uncertainty . 15
Annex A (informative) Tube-type samplers . 16
A.1 Sampler design . 16
A.2 Extraction and analysis . 16
A.3 Application range and conditions . 16
Annex B (informative) Badge-type samplers . 18
B.1 Type 1 badge-type sampler . 18
B.2 Type 2 badge-type sampler . 20
B.3 Type 3 badge-type sampler . 23
B.4 Type 4 badge-type sampler . 25
Annex C (informative) Radial samplers . 29
C.1 Sampler design . 29
C.2 Extraction and analysis . 30
C.3 Application range and conditions . 31
Annex D (informative) Summary of passive diffusive sampling rate data . 32
Annex E (normative) Estimation of the sampling rate of the samplers . 33
Annex F (informative) Measurement uncertainty calculation . 35
F.1 Measurement equation . 35
F.2 Combined standard uncertainty . 35
F.3 Expanded relative uncertainty . 36
F.4 Uncertainty contributions . 36
Bibliography . 41
European foreword
This document (EN 17346:2020) has been prepared by Technical Committee CEN/TC 264 “Air quality”,
the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by November 2020, and conflicting national standards shall
be withdrawn at the latest by November 2020.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
According to the CEN-CENELEC Internal Regulations, the national standards organisations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia,
Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland,
Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of North
Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United
Kingdom.
Introduction
Atmospheric ammonia (NH ) is a pollutant of major environmental concern with adverse effects on
forests, species composition of semi-natural ecosystems and soils [1-4]. Emission and deposition of NH
can contribute significantly to total nitrogen deposition to the environment, contributing to
eutrophication (nutrient enrichment) and acidification (oxidation of NH to nitrate resulting in release of
+
H ions) of land and freshwaters, leading to a reduction in both soil and water quality, loss of biodiversity
and ecosystem change [5-10].
In addition to these effects, NH3 is the major precursor for neutralization of atmospheric acids, affecting
the long-range transport distance of both SO and NO and leading to the formation of secondary particles
2 x
(primarily ammonium sulphate and ammonium nitrate) [11-13]. These particles have multiple impacts
including effects on atmospheric visibility, radiative scattering (and the greenhouse effect) and on human
health.
The recognition of NH as an important air pollutant led to its inclusion in international agreements to
reduce air pollutant emissions, first under the 1999 UNECE Gothenburg Protocol and then the National
Emissions Ceilings Directive (NECD) (2001/81/EC) of the EU. The target of both these agreements is that
NH emissions should not exceed emission ceilings set for EU member states, with a particular focus on
reducing the extent of critical loads exceedance for acidification and eutrophication effects. Revision of
the Gothenburg Protocol (2012) and the NEC Directive (2016) include new, more stringent emission
ceilings for 2020 that seek more environmental protection and improvement in air quality than has so
far been committed, including the introduction of an emissions ceiling for particulate matter (PM). Under
the 2012 UNECE Gothenburg Protocol, EU member states have to jointly cut their emissions of NH by
6 % and particles by 22 % between 2005 and 2020. As a precursor of PM, controlling NH is important to
reducing particle emissions of PM and PM . A recent study employing three chemical transport models
2,5 10
found that the models underestimated the formation of ammonium particles and concluded that the role
of NH on PM is larger than originally thought [14]. Thus the implementation of 2020 targets detailed
above may not be enough to deliver compliance with proposed particle limit values, and further local
measures may be required to be compliant.
Other legislations to abate NH emissions include the Industrial Emissions Directive (IED) (2010/75/EU)
which requires pig and poultry farms (above stated size thresholds) to reduce emissions using Best
Available Techniques. For the protection of vegetation and ecosystems, new revised “Critical Levels” (CL)
3 3
of NH concentrations were adopted in 2007 (see Table 1), of 1 µg/m and 3 µg/m annual mean for the
protection of lichens/bryophytes and higher plants under field conditions, respectively, which replaced
3 3
the previous CL annual mean value of 8 µg/m . A monthly critical level of 23 µg /m was retained as a
provisional value in order to deal with the possibility of high peak emissions during periods of manure
application (e.g. in spring) ([15]). In Germany, the recommended exposure limit for the protection of
ecosystems is 10 µg/m (TA Luft, Annex 1, [16]).
Table 1 — Summary of upper limits of NH concentrations for protection of ecosystems under
field conditions
Concentration Specification Types of locality
(µg/m )
1 UNECE Critical Level (annual mean) for Sensitive ecosystems in
lower plants (lichens, bryophytes) which the lichens and
bryophytes are important
components, e.g. designated
sites for nature
conservation and protection
of sensitive species, e.g.
Natura 2000 sites
3 UNECE Critical Level (annual mean) for Sensitive ecosystems in
higher plants which the higher plants are
important components, e.g.
designated sites for nature
conservation and protection
of sensitive species, e.g.
Natura 2000 sites
10 German First General Administrative Near installations
Regulation Pertaining the Federal
Immission Control Act Maximum near
installations where ecological
monitoring undertaken.
23 UNECE critical level (monthly mean) – for In close proximity to
peak emission periods such as in months emission sources
where slurry spreading takes place.
Improving knowledge on levels of NH in the ambient air and near sources is therefore important for the
assessment of:
— environmental effects on ecosystems (Contribution to eutrophication and acidification processes);
— contributions to the formation of PM and PM ;
10 2,5
— effectiveness of current and future abatement measures to reduce NH emissions.
The simplest to the latest state-of–the-art techniques for measurement of atmospheric NH are presented
in Table 2.
Table 2 — Measurement methods suitable for determination of atmospheric NH gas and
ammonium particle concentrations
Monitoring Methods Time resolution References
Integrative methods: passive
Passive diffusion samplers daily to monthly [17]
[18]
[19]
[20]
Integrative methods: active
Simple denuder systems with offline chemical analysis daily to monthly [17]
[19]
[21]
Annular denuder systems (ADS) with offline chemical hourly to daily [22]
analysis
Conditional sampling with denuders at different heights weekly to monthly [23]
(COTAG)
Continuous: wet chemistry methods
Annular Denuder Systems with online analysis hourly or better [24]
depending on set-
Membrane stripping with online analysis
up
Steam Jet Aerosol Collector Systems for gas and aerosol hourly or better [25]
depending on set-
[26]
up
Continuous: optical methods
Differential Optical Absorption Spectrometry (DOAS) hourly or better [27]
depending on set-
up
Tunable Diode Laser Absorption Spectrometry and hourly or better [28]
Quantum Cascade Laser (TDL and QCL AS, respectively) depending on set-
up
Photoacoustic spectrometry hourly or better [29]
depending on set-
up
Chemiluminescence with catalytic conversion hourly or better [30]
depending on set-
up
Integrative atmospheric sampling methods such as passive diffusion samplers and active samplers
provide measurement of concentrations of NH averaged over the chosen sampling time. The diffusive
samplers used include those that are available commercially and those that have been developed in-
house by organisations to meet specific research requirements. A full validation of diffusive sampling
methods for NH in accordance with the European Standard (EN 13528-2 [31]) would be costly and
would also require specialist facilities only available at well-equipped large metrological institutes.
Validation of the quantitative measurement of NH through comparison with “reference” methods is
problematic for NH as there is no currently accepted and defined reference method. Automatic
continuous analysers for NH , employing spectroscopic or other techniques (Table 2) are available
commercially, but there is a lack of robust published calibration data and procedures for reliable field
measurements under ambient concentrations and conditions [32].
1 Scope
This document specifies a method for the sampling and analysis of NH in ambient air using diffusive
sampling.
It can be used for NH measurements at ambient levels, but the concentration range and exposure time
are sampler dependent, and the end user is therefore advised to match the sampler type to the
measurement requirement and to follow the operating instructions provided by the manufacturer.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at https://www.iso.org/obp/ui
3.1
combined standard measurement uncertainty
combined standard uncertainty
standard measurement uncertainty that is obtained using the individual standard measurement
uncertainties associated with the input quantities in a measurement model
[SOURCE: JCGM 200:2012, 2.31] [33]
3.2
extraction efficiency
ratio of the mass of analyte extracted from a sampling device to that applied
3.3
diffusive sampler
device which is capable of taking samples of gases or vapours from the atmosphere at a rate controlled
by a physical process such as gaseous diffusion through a static air layer or a porous material and/or
permeation through a membrane, but which does not involve the active movement of air through the
device
Note 1 to entry: Active normally refers to the pumped movement of air.
[SOURCE: EN 13528-2:2002, 3.6] [31]
3.4
diffusive sampling rate
diffusive uptake rate
rate at which the diffusive sampler collects a particular gas or vapour from the atmosphere
3 3
Note 1 to entry: The sampling rate is usually expressed in units of (m /h), (ml/min) or (cm /min).
3 3 –8
Note 2 to entry: cm /min may be converted to SI units of m /s by factor 1,67 × 10 .
3.5
expanded measurement uncertainty
product of a combined standard measurement uncertainty and a factor larger than the number one
Note 1 to entry: The factor depends upon the type of probability distribution of the output quantity in a
measurement model and on the selected coverage probability.
Note 2 to entry: The term “factor” in this definition refers to a coverage factor.
[SOURCE: JCGM 200:2012, 2.35]
3.6
field blank
unused sampler, taken from the same batch used for NH monitoring, handled in the same way as a
sampler that is used for NH monitoring, except it is not used for collecting a sample
Note 1 to entry: Adapted from EN 14902:2005, 3.1.6.
Note 2 to entry: The results from the analysis of field blanks are used to identify contamination of the sample
arising from handling in the field and during transport.
[SOURCE: EN 1540:2011, 3.3.8] [34]
Note 3 to entry: A transport blank is considered to be a special case of a field blank. A transport blank is taken to
the exposure site, left unopened and returned to the laboratory immediately after placement or collection of the
samplers. Transport blanks may be used when regular field blanks reveal an unacceptable level of ammonium to
investigate the possibility of contamination of samplers during transport. This blank is only used for quality control
purposes.
3.7
laboratory blank
sealed sampler drawn from the same batch as the samplers being used for NH monitoring which is stored
for the duration of the sampling period and is analysed at the same time as the exposed samplers
3.8
measurement uncertainty
uncertainty of measurement
non-negative parameter characterizing the dispersion of the quantity values being attributed to a
measurand, based on the information used
Note 1 to entry: For notes to the definition the reader is referred to the parent document JCGM 200:2012.
[SOURCE: JCGM 200:2012, 2.26]
3.9
standard measurement uncertainty
standard uncertainty
measurement uncertainty expressed as a standard deviation
[SOURCE: JCGM 200:2012, 2.30]
4 Description of samplers
4.1 Principle
The diffusive sampler is exposed in air for a measured time period. NH migrates through the sampler
along a diffusion path of defined dimensions and is collected by reaction onto an acid sorbent.
Determining the sampling rate is essential when deploying diffusive NH samplers in the field, either by:
— calculation based on Fick’s first law of diffusion (see EN 13528-3 [35] and Annex E),
— calibration by exposure to standard atmospheres, or
— co-located calibration studies against another well characterized NH measurement method in the
field.
NOTE Denuders can be used as a cost effective surrogate reference method until there are improvements in
the continuous optical methods.
Details of these approaches shall be documented.
Samplers can be provided with manufacturer measured sampling rates. Samplers in networks often have
on-going measurements of sampling rates. Users can calculate a locally derived sampling rate. Sampling
rates are also documented in literature [see [36], and Annex D].
The sampling rate in the field is a function of local meteorology. Samplers can be deployed with protective
shelters to minimize meteorological influences. When doing so, the user shall apply a suitable protocol to
ensure a consistent approach for all samplers. Ideally, the effect of the shelter on the sampler
performance should be characterized.
4.2 Implementation
Samplers shall be sealed and stored under cool conditions, for example at temperatures between 0 °C and
4 °C, in the dark, in order to minimize any undesired reactions before and after deployment. After
deployment, samplers shall be analysed as soon as possible, according to manufacturer’s specifications.
Disposable gloves shall be worn at all times, including during deployment in the field. This serves to
protect the samples from contamination by contact with the skin. It is also advised to avoid breathing
directly on the samples, as exhaled breath contains NH .
Since there are different sampler designs, each common sampler type is briefly described below.
4.3 Tube-type samplers
The tube-type samplers are hollow cylindrical tubes oriented vertically. A cap at the top end holds in
place either a cellulose filter paper, glass fibre filter or stainless steel grid, which is coated with a sorbent
that collects the gas of interest. This type of sampler is characterized by a high length to cross sectional
area ratio [15, 37]. To collect NH , sorbents used include citric, phosphoric, phosphorous, sulphuric and
tartaric acid [38]. The analysis is carried out using various methods including ion chromatography, flow
injection analysis with detection of conductivity and spectrophotometry.
There is one commonly used design of tube type samplers, the 3,5 cm short membrane diffusion tube.
For more information, see Annex A.
4.4 Badge-type samplers
The badge-type samplers have a lower length to area ratio of the sample body, with enhanced sensitivity
over the tube-type samplers [15, 37]. There are many badge-type samplers in use with different
geometry. Due to the short diffusion path length, they have a gas permeable barrier at the inlet to prevent
wind-induced turbulent diffusion (wind shortening effect on sampling rate). The sorbents used to collect
ammonia are the same as employed with tube-type samplers, and the following analytical assessment is
performed by using the same techniques.
For more information, see Annex B.
4.5 Radial samplers
The radial-type sampler has a cylindrical outer surface that acts as a permeation barrier through which
NH diffuses [39]. NH molecules move axially parallel towards an absorbent bed that is also cylindrical
3 3
and coaxial to the diffusive surface. This type of sampler uses phosphoric acid as a sorbent. Exposed
samplers are analysed using various methods, including spectrophotometry, and ion chromatography.
For more information, see Annex C.
5 Calculation of the concentration of NH
5.1 Mass concentration
The concentration of NH in ambient air under actual conditions of sampling is calculated using
Formula (1):
M
mm−
NH
3 sb
c= .
M e⋅⋅υ t
+
NH
(1)
where
c is the concentration of NH at ambient conditions in µg/m ;
is the molar mass of NH in g/mol;
M 3
NH
+
is the molar mass of NH in g/mol;
M 4
+
NH
m is the mass of ammonium found in the sample in µg;
s
m is the mass of ammonium found in the mean laboratory blank in µg;
b
NOTE 1 In normal operations the transport and field blanks are expected to record similar masses of
ammonium compared to the laboratory blank. In cases where the transport and/or field blank is higher than
the laboratory blank the transport or field blank can be used for the subtraction. This information needs to be
clearly documented.
υ 3
is the sampling rate at ambient conditions during sampling in m /h;
e is the efficiency of extraction of ammonium;
NOTE 2 It is not necessary to include the efficiency of extraction of ammonium if this efficiency is shown not
be significantly different from 100 % or if it is already included into the estimation of the sampling rate.
t is the sampling time in h.
NOTE 3 The sampling rate can be in µg/(nmol/mol)/min, in which case c is expressed in units of nmol/mol.
5.2 Conversion to standard conditions of temperature and pressure
The mass concentration of NH in air is calculated at the ambient temperature and pressure during
exposure using Formula (1). This mass concentration shall be referred to at standard conditions of
temperature and pressure, as required in Directive 2008/50/EC [40] and defined e.g. by EN 16339 [41],
using Formula (2):
T 101,3
cc=⋅⋅
STP
293 P
(2)
where
c is the concentration of NH at standard temperature and pressure in µg/m ;
STP 3
c is the concentration of NH at ambient conditions in µg/m ;
T is the average temperature during exposure in K;
P is the average pressure during exposure in kPa.
NOTE Temperature and pressure data can be obtained from nearby meteorological stations.
6 Quality control/quality assurance
6.1 Quality control
For each series of analyses, the following control checks shall be performed and recorded:
a) inspection of each sampler before and after exposure, reject those with visible damage or
contamination and record this information in the report;
b) analysis for each batch of samplers, field blanks, transport blanks and/or laboratory blanks to detect
contamination of samplers during transport, in the field and during subsequent storage and
handling;
c) analysis of calibration solutions to determine instrument drift and appropriate re-calibration at
regular intervals, e.g. at the start of each day . If a check of the calibration response is outside the
expected performance criteria of the instrument, then a further investigation is required to
demonstrate that it is functioning correctly.
NOTE For a linear calibration curve, a check can be carried out using at least 3 points (zero, 50 % of calibration
range and full scale).
At regular intervals, the following control checks shall be performed and registered:
d) analysis of reagent solutions to determine variations of reagent blank levels;
e) determination of extraction efficiency by spiking coated samplers with known masses of ammonium
followed by the determination of their recoveries;
f) duplicate analysis of sample extracts to check analytical repeatability;
g) analysis of samples taken in parallel at one field site to check reproducibility;
h) checking the sensitivity of the analytical method used (e.g. the slope of the calibration curve). For
routine NH monitoring with diffusive samplers, it is necessary to complement with other quality
control checks (e.g. instrument drift) that will ensure proper analysis over time;
i) for new samplers and any re-used components of diffusive samplers, employment of a protocol to
check for the absence of ammonium.
Corrective actions shall be taken in case of exceedance of internal quality control limits set by the user, in
order to ensure the quality of the NH measurement results, and in accordance with the QA/QC
procedures of the laboratory.
6.2 Quality assurance
A competent body (or bodies) shall perform the method. The competent body (or bodies) performing the
method shall be able to demonstrate that work is in conformity with the requirements of internationally
accepted standards for test laboratories.
NOTE 1 EN ISO/IEC 17025 [42] is the harmonized internationally accepted standard that applies.
NOTE 2 A formal accreditation by a member body of the European Accreditation Organization to
EN ISO/IEC 17025 is a demonstration of conformity.
7 Report
The test report shall contain at least the following information:
— the type of sampler used;
— the complete identification of the sampler(s);
— the sampling location(s), and sampling period(s);
— the average barometric pressure and average temperature during sampling, if relevant;
NOTE 1 The average barometric pressure and temperature can be obtained from available data from nearby
meteorological stations.
— any unusual features noted during the measurements;
— the value(s) of the NH sampling rate(s) used with uncertainty statement and reference;
— the method of analysis used;
— the analytical results;
— the NH concentrations determined accompanied by an uncertainty statement;
— any deviations from procedures described in this European Standard;
— the responsible person checking the analysis, the calculation of the test results and their
uncertainties shall be identified.
NOTE 2 Any other metadata such as personnel involved in deployment and analysis can be included.
Information about the following items needs not to be included in the report but shall be made available
upon request:
— the date of preparation and method of preparation of the sampler; or, if commercial samplers are
used, the identification code of the series of samplers;
— results for field/transport blanks;
— sampling height.
8 Performance requirements and measurement uncertainty
The parameters given in Table 3 have been identified to contribute to the uncertainty of concentrations
of NH measured by diffusive sampling and subsequent analysis.
Table 3 — Uncertainty parameters and minimum requirements for a single exposure period for
diffusive samplers
Uncertainty source Symbol Criterion
Volume of air sampled
Sampling rate υ Relative uncertainty ≤ 10 %
Sampling time t Relative uncertainty ≤ 0,2 %
Average air temperature during T Relative uncertainty ≤ 2 %
exposure for conversion to
standard temperature and
pressure
Average air pressure during P Relative uncertainty ≤ 2 %
exposure for conversion to
standard temperature and
pressure
Mass of ammonium determined
Extraction efficiency E Is included in the effective sampling rate, or
when calculated separately the relative
uncertainty shall be ≤ 5 %
Concentration in calibration m Relative uncertainty ≤ 3 %, excluding zero
CS
standards concentration
Lack-of-fit of calibration function δ Relative residuals over the calibration
i
range ≤ 2 %, excluding zero
Response drift between D ≤ 5 %
calibrations
Analytical repeatability R ≤ 5 %
A recent laboratory study showed that the expanded uncertainty of tube-type and badge-type samplers
is within the range of 10-12 %, whereas the expanded uncertainty of radial-type samplers is within the
range of 15-23 %, for 1-23 µg/m . The tube-type and badge-type samplers were exposed for 4 weeks,
while the radial-type samplers were exposed for 2 weeks [36].
For the expanded uncertainty of Ferm-type samplers, see VDI 3869-4 [43].
Annex A
(informative)
Tube-type samplers
A.1 Sampler design
There is one commonly used design of tube-type samplers, the 3,5 cm short membrane diffusion tube,
see Figure A.1.
Figure A.1 — Tube-type sampler
The tube-type sampler is a hollow cylindrical tube oriented vertically. A cap at the top end holds in place
either a filter paper or stainless steel grid, which is coated with an absorbent that collects the gas of
interest. For NH sampling, a number of acids have been used, including citric, phosphoric, phosphorous,
sulphuric and tartaric acid [44]. The lower end of the tube is capped with a membrane [45].
The tube-type sampler has a gas permeable membrane placed at the inlet to establish a turbulent free
layer of air inside the sampler [46]. The membrane does however impose an additional resistance against
gas diffusion, which should be taken into account when deriving a sampler’s effective sampling rate.
A.2 Extraction and analysis
Acid impregnated grids in diffusion tubes are usually extracted into ultrapure water [47] and analysed
using an ion chromatographic method, and this method of analysis is available from commercial
laboratories. Other laboratories use different techniques, for example, Sutton et al. [18] used flow
injection analysis method with detection of conductivity ([48]).
A.3 Application range and conditions
The application range and conditions recommended for use by manufacturers or by researchers are
summarized below in Table A.1.
Table A.1 — Performance data of tube-type sampler
Value(s)
Sampling rate 2,5 to 3,5 ml/min
Working range 1 – 35 µg/m
Sampling period Weekly – 8 weeks
Detection limit Approximately 1,5 µg/m for 4 week
exposure
External influences: Influence of wind speed < 10 % up to
4,5 m/s using protection shelters
wind speed
T: no influence between 10 – 30 °C
temperature
RH: no influence between 20 – 80 %
relative humidity
Storage Before use: 8 weeks from preparation
Interferences Potential volatile aerosol particles
+
containing NH on the permeable
membrane. Avoided by removing
membrane after sampling and sealing
the inlet with a clean cap
Expanded uncertainty 12 % at 1 μg/m level, laboratory
conditions
More detailed information is available in [49].
Annex B
(informative)
Badge-type samplers
B.1 Type 1 badge-type sampler
B.1.1 Sampler design
The type 1 badge-type sampler collects NH molecules onto an absorbent material impregnated with
phosphoric acid. Thus, NH is trapped as ammonium phosphate on the absorbent material. A schematic
of the sampler is shown in Figure B.1. It consists of a polypropylene housing (sampler case) with an inner
diameter of 23 mm and a length of 20 mm. The opening is equipped with a wind protection (glass fibre
and steel screen) to reduce the effect of wind on the sampling rate. In addition, the samplers should be
installed in a protective shelter to minimize the effects from the environmental factors (e.g. wind, rain,
sunlight, etc.), see Figure B.2.
Key
1 cap
2 glass fibre
3 steel screen
4 sampler case
5 fixed cap
Figure B.1 — Schematic of type 1 badge-type sampler
Figure B.2 — Protective shelter for sampler (left) and mounting sampler with clips (right)
B.1.2 Extraction and analysis
By reaction with phenol and alkaline sodium hypochlorite indophenol is produced, a blue dye. The
reaction is accelerated by the addition of sodium nitroprusside as a catalyst. Two types of solution may
be used:
— solution I: 10 g of phenol, 50 mg of sodium nitroprusside, made up to 1 l with ultrapure water;
— solution II: 5 g sodium hydroxide, 4,2 ml sodium hypochlorite (12 % available chlorine) made up to
1 l with ultrapure water (resistivity greater or equal to 18 MΩ·cm at 298 K).
The glass fibre filter is removed from the sampler and transferred to a 10 ml polypropylene tube. 2,0 ml
of water is added and thoroughly shaken by a vortex. Hereafter, 1,0 ml of reagent I and 1,0 ml of reagent
2 are added. After 30 min of colour development, the sample is centrifuged and the absorption of the
colour is measured spectrophotometrically at 630 nm.
B.1.3 Application range and conditions
Table B.1 — Performance data of type 1 badge-type sampler
Parameter Value(s)
Sampling rate 31,5 ml/min at 20 °C
Working range 0,5 – 30 µg/m
Sampling period 2 – 4 weeks
Detection limit 0,2 µg/m for a one month exposure period
External influences: Influence of wind speed < 10 % up to 4,5 m/s
using protection shelters
Wind speed
No influence between 10 to 30 °C
Temperature
No influence between 20 to 80 %
Humidity
Storage Before use: 6 months
After exposure: 4 months
Interferences Ammonium particles are not collected
Expanded uncertainty 29,3 % at a 1 to 5 μg/m level
B.2 Type 2 badge-type sampler
B.2.1 Sampler design
The type 2 badge-type diffusive sampler [46] (Figure B.3) is made up of a circular polyethylene vial
(26 mm height, 27 mm diameter) with one open end. An internal ridge (G, 17 mm height) supports a filter
(F, 24 mm diameter). The filter is coated with an acid such as citric acid, which serves to capture the NH ,
and is held in place with a polyethylene ring (E, 6 mm height). The open end is capped with a polyethylene
cap with a hole punched out in the centre (D, 23 mm diameter), which holds in place a white PTFE
(polytetrafluoroethylene (PTFE)) membrane (C, 27 mm diameter, 5 µm pore size) allowing gaseous NH
to diffuse through. This end is positioned facing downwards. The membrane-cap end of the sampler is
sealed with a top protective cap (B), which is removed prior to exposure. At the end of exposure, the
membrane cap plus membrane are removed and replaced with a replacement solid cap (A).
Key
A replacement solid cap F filter paper
B top protective cap G internal ridge to support filter paper
C 5 µm PTFE membrane (27 mm diameter) H extended body for ease of handling
D cap with hole for membrane I Velcro® for attachment to holder
E support ring (6 mm height)
Figure B.3 — Schematic view of type 2 badge-type sampler
The type 2 badge-type sampler geometry permits an optimized sampling rate for long-term sampling (1-
2 months) plus high sensitivity to allow measurements down to low concentrations (<1 µg/m NH ) in
background areas. The membrane inlet is placed directly at the mouth of the sampler, which minimizes
the formation of a boundary layer in front of the membrane, whilst achieving a stable, turbulent-free
diffusion path length behind the membrane.
To make handling easier, the sampler body was extended behind the coated filter. The PTFE membrane
and membrane cap are removed at end of exposure and sampler capped with replacement cap provided.
Velcro® is the trade name or trademark of a product. This information is given for the convenience of users of
this European Standard and does not constitute an endorsement by CEN of the product named. Equivalent
products may be used if they can be shown to lead to the same results.
This serves to avoid potential contamination arising from possible volatilisation of volatile ammonium
salts on the outside of the PTFE membrane during exposure.
B.2.2 Extraction and analysis
B.2.2.1 General
Exposed samples are stored at 4 °C until analysis. In order to avoid potential contamination during the
extraction process, care should be taken in handling the type 2 badge-type samplers and components.
B.2.2.2 Extraction of acid coated filter paper
The following steps are used for analyte extraction:
1) using forceps, lift the plastic retainer ring out of the type 2 badge-type sampler.
2) place the open end of the type 2 badge-type sampler over a 20 ml pot (or other suitable container)
and tap the sampler very gently against the pot until the coated filter paper drops out into the pot. If
the paper does not drop out, then use a pair of clean forceps and remove the coated filter paper very
carefully.
3) add 3 ml deionised water to the filter paper in the pot. Cap the pot with a clean cap, label with sample
name and leave to extract for at least one hour. Repeat this extraction with all of the samples
(including a suitable number of type 2 badge-type sampler blanks), noting the time at which the
extraction process commenced.
4) using a pair of clean forceps (rinse the forceps with deionised water in between different samples),
pinch an edge of the filter paper and dunk it up and down in the extractant a few times to ensure that
all the chemical species of interest enters into the extractant, and that the solution is well mixed.
5) for analysis using flow injection systems, 1,5 ml of extracted solution is decanted into 3 ml
autosampler tubes. Care should be taken to ensure that any bits of loose fibres coming off the filter
paper are not decanted into the autosampler tubes, which may cause blockage.
B.2.2.3 Analysis of ammonium
B.2.2.3.1 Flow Injection Conductivity Ammonium Analysers
The extracted samples are analysed using a system based on selective dialysis of ammonium across a
membrane at high pH with subsequent analysis of conductivity (see Figure B.4). The calibration range
normally used is 0 - 10 mg/l (0,1 mg/l, 1 mg/l and 10 mg/l calibration standards), and the calibration
concentrations used are 0,2 mg/l (low range), 0,9 mg/l (high range), 2 (low range) and 9 mg/l (high
range). It may be necessary to dilute the extracted samples and repeat the analysis if they are outside the
calibration range.
Measuring characteristics:
— dynamic range 5 – 30 000 µg/l;
— resolution 5 µg/l;
— accuracy 5 %.
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.