Standard Test Method for Ethylene, Other Hydrocarbons, and Carbon Dioxide in High-Purity Ethylene by Gas Chromatography

SIGNIFICANCE AND USE
High-purity ethylene is required as a feedstock for some manufacturing processes, and the presence of trace amounts of carbon dioxide and some hydrocarbons can have deleterious effects. This method is suitable for setting specifications, for use as an internal quality control tool and for use in development or research work.
SCOPE
1.1 This test method covers the determination of carbon dioxide, methane, ethane, acetylene, and other hydrocarbons in high-purity ethylene. Hydrogen, nitrogen, oxygen, and carbon monoxide are determined in accordance with Test Method D2504. The percent ethylene is obtained by subtracting the sum of the percentages of the hydrocarbon and nonhydrocarbon impurities from 100. The method is applicable over the range of impurities from 1 to 500 parts per million volume (ppmV).
1.2 This standard may involve hazardous materials, operations, and equipment. This standard does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For some specific hazard statements, see Section 6.
1.3 The values stated in acceptable metric units are to be regarded as the standard. The values in parentheses are for information only.

General Information

Status
Historical
Publication Date
31-Oct-2004
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D2505-88(2004)e1 - Standard Test Method for Ethylene, Other Hydrocarbons, and Carbon Dioxide in High-Purity Ethylene by Gas Chromatography
English language
7 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information.
´1
Designation:D2505–88 (Reapproved 2004)
Standard Test Method for
Ethylene, Other Hydrocarbons, and Carbon Dioxide in High-
Purity Ethylene by Gas Chromatography
This standard is issued under the fixed designation D2505; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
´ NOTE—Warning notes were moved into text editorially in November 2004.
1. Scope 3. Summary of Test Method
1.1 This test method covers the determination of carbon 3.1 The sample is separated in a gas chromatograph system
dioxide, methane, ethane, acetylene, and other hydrocarbons in utilizing four different packed chromatographic columns with
high-purity ethylene. Hydrogen, nitrogen, oxygen, and carbon helium as the carrier gas. Methane and ethane are determined
monoxide are determined in accordance with Test Method by using a silica gel column. Propylene and heavier hydrocar-
D2504. The percent ethylene is obtained by subtracting the bons are determined using a hexamethylphosphoramide
sum of the percentages of the hydrocarbon and nonhydrocar- (HMPA) column. Acetylene is determined by using, in series,
bon impurities from 100. The method is applicable over the a hexadecane column and a squalane column. Carbon dioxide
range of impurities from 1 to 500 parts per million volume is determined using a column packed with activated charcoal
(ppmV). impregnated with a solution of silver nitrate in b,b8-
1.2 This standard does not purport to address all of the oxydipropionitrile. Columns other than those mentioned above
safety concerns, if any, associated with its use. It is the may be satisfactory (see 5.3). Calibration data are obtained
responsibility of the user of this standard to establish appro- using standard samples containing the impurities, carbon
priate safety and health practices and determine the applica- dioxide, methane, and ethane in the range expected to be
bility of regulatory limitations prior to use. For some specific encountered. Calibration data for acetylene are obtained as-
hazard statements, see Section 6. suming that acetylene has the same peak area response on a
1.3 The values stated in acceptable metric units are to be weight basis as methane. The acetylene content in a sample is
regarded as the standard. The values in parentheses are for calculatedonthebasisoftheratioofpeakareaoftheacetylene
information only. peak to the peak area of a known amount of methane.
Calculations for carbon dioxide, methane, and ethane are
2. Referenced Documents
carried out by the peak-height measurement method.
2.1 ASTM Standards:
4. Significance and Use
D2504 Test Method for Noncondensable Gases in C and
Lighter Hydrocarbon Products by Gas Chromatography 4.1 High-purity ethylene is required as a feedstock for some
D4051 Practice for Preparation of Low-Pressure Gas manufacturing processes, and the presence of trace amounts of
Blends carbon dioxide and some hydrocarbons can have deleterious
E260 Practice for Packed Column Gas Chromatography effects. This method is suitable for setting specifications, for
F307 Practice for Sampling Pressurized Gas for GasAnaly- use as an internal quality control tool and for use in develop-
sis ment or research work.
5. Apparatus
This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee 5.1 Any chromatographic instrument with an overall sensi-
D02.D0 on Hydrocarbons for Chemical and Special Uses.
tivity sufficient to detect 2 ppmV or less of the compounds
Current edition approved Nov. 1, 2004. Published November 2004. Originally
listed with a peak height of at least 2 mm without loss of
approved in 1966. Last previous edition approved in 1998 as D2505 – 88 (1998).
resolution.
DOI: 10.1520/D2505-88R04E01.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
5.2 Detectors—Thermal Conductivity—If a methanation re-
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
actor is used, a flame ionization detector is also required. To
Standards volume information, refer to the standard’s Document Summary page on
determine carbon dioxide with a flame ionization detector, a
the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
´1
D2505–88 (2004)
methanation reactor must be inserted between the column and
the detector and hydrogen added as a reduction gas (see Test
Method D2504, Appendix X1, Preparation of Methanation
Reactor).
5.3 Column—Any column or set of columns can be used
that separates carbon dioxide, methane, acetylene and C and
heavier compounds. There may be tailing of the ethylene peak
but do not use any condition such that the depth of the valleys
ahead of the trace peak is less than 50 % of the trace peak
height. (See Fig. 1 for example.)
5.4 Recorder—A recorder with a full-scale response of 2 s
or less and a maximum rate of noise of 60.3 % of full scale.
5.5 Gas-BlendingApparatus—Atypical gas-blending appa-
ratus is shown in Fig. 2. A high-pressure manifold equipped
withagagecapableofaccuratelymeasuringethylenepressures
up to 3.4 MN/m gage (500 psig) is required. Other types of
gas-blending equipment, such as described in Practice D4051,
can be used.
FIG. 2 Gas-Blending Manifold
NOTE 1— Practice E260 contains information that will be helpful to
those using this method.
6.3 Active Solids—Activated carbon, 30 to 40-mesh, silica
6. Reagents and Materials
gel, 100 to 200-mesh. Other sizes may be satisfactory.
6.1 CopperorAluminum,orStainlessSteelTubing, 6.4-mm 6
6.4 Liquid Phases—Hexamethylphosphoramide (HMPA ),
1 1
( ⁄4-in. ) outside diameter, and nylon tubing, 3.2-mm ( ⁄8-in.) 6 6
hexadecane. Squalene, silver nitrate, and b,b8-
outside diameter. 7
oxydipropionitrile. Other liquid phases may be satisfactory.
6.2 Solid Supports—Crushed firebrick or calcined diatoma-
(Warning—Combustible solvents. See A1.7.) (Warning—
ceous earth, such as Chromosorb P, 35 to 80-mesh and 80 to
HMPAmaybeharmfulifinhaled.Causesirritation.Apotential
100-mesh. Other supporting materials or mesh sieves can be
carcinogen (lungs). See A1.5.)
satisfactory.
6.5 Helium.(Warning—Compressed Gas, Hazardous Pres-
sure. See A1.2.)
6.6 Hydrogen.(Warning—Flammable Gas, Hazardous
Pressure. See A1.6.)
6.7 Acetone.(Warning—Extremely Flammable. SeeA1.1.)
6.8 Gases for Calibration—Pure or research grade carbon
dioxide, methane, ethane, acetylene, ethylene, propane, and
propylene. Certified calibration blends are commercially avail-
able from numerous sources and may be used. (Warning—
Flammable Gases, Hazardous Pressure. See A1.2 and A1.3.)
6.9 Methanol.(Warning—Flammable. Vapor Harmful. See
A1.4.)
A fraction sieved in the laboratory to 30 to 40 mesh from medium activity
charcoal, 20 to 60 mesh, sold by Central Scientific Co., 1700 Irving Park Road,
Chicago, IL60613, has been found satisfactory for this purpose. If you are aware of
alternative suppliers, please provide this information to ASTM International
Headquarters.Your comments will receive careful consideration at a meeting of the
responsible technical committee , which you may attend.
The sole source of supply of the apparatus known to the committee at this time
is Silica gel Code 923 available from the Davison Chemical Co., Baltimore,
Md.21203.Ifyouareawareofalternativesuppliers,pleaseprovidethisinformation
to ASTM International Headquarters. Your comments will receive careful consid-
erationatameetingoftheresponsibletechnicalcommittee ,whichyoumayattend.
FIG. 1 Typical Chromatogram for Propylene
The sole source of supply of the apparatus known to the committee at this time
is available from the Fisher Scientific Co., St. Louis, MO. If you are aware of
alternative suppliers, please provide this information to ASTM International
Headquarters.Your comments will receive careful consideration at a meeting of the
responsible technical committee , which you may attend.
3 7
ThesolesourceofsupplyoftheapparatusisavailablefromtheCeliteDivision, b,b8-oxydipropionitrile, sold by Distillation Products Industries, Division of
Johns Mansville Co., NewYork, NY. f you are aware of alternative suppliers, please Eastman Kodak Co., Rochester, NY, has been found to be satisfactory. If you are
provide this information toASTM International Headquarters. Your comments will aware of alternative suppliers, please provide this information to ASTM Interna-
receive careful consideration at a meeting of the responsible technical committee , tional Headquarters.Your comments will receive careful consideration at a meeting
which you may attend. of the responsible technical committee , which you may attend.
´1
D2505–88 (2004)
NOTE 2—The use of copper tubing is not recommended with samples
packing prepared in the same manner as the hexadecane
containing acetylene as this could lead to the formation of potentially
packing. Plug the open end of the tubing with glass wool and
explosive copper acetylide.
shape the column to fit into the chromatograph with the
hexadecane portion of the column at the front end of the
7. Sampling
column. The column shall be purged under test conditions (no
7.1 Samples should be supplied to the laboratory in high
sample added) until a constant baseline is obtained.
pressure sample cylinders, obtained using the procedures
NOTE 3—Columns made with liquid phases listed above were used
described in Practice F307, or similar methods.
satisfactorily in cooperative work. Other columns may be used (see 5.3).
8. Preparation of Apparatus
9. Calibration
8.1 Silica Gel Column—Dry the silica gel in an oven at
9.1 Preparation of Standard Mixtures:
204°C (400°F) for 3 h, cool in a desiccator, and store in
9.1.1 PreparationofConcentrate—Prepare a concentrate of
screw-cap bottles. Pour the activated silica gel into a 0.9-m
the impurities expected to be encountered. A certified calibra-
(3-ft) length of 6.4-mm ( ⁄4-in.) outside diameter copper or
tion blend containing the expected impurities can be obtained
aluminum tubing plugged with glass wool at one end. Tap or
and used as the concentrate. An example of a satisfactory
vibrate the tube while adding the silica gel to ensure uniform
concentrate is given in Table 1. The concentrate can be
packing and plug the top end with glass wool. Shape the
prepared using the gas blending manifold as shown in Fig. 2 or
column to fit into the chromatograph.
using a similar apparatus as follows: Evacuate the apparatus
8.2 Silver Nitrate—b,b8-Oxydipropionitrile—Activated
and add the components in the order of increasing vapor
CarbonColumn—Weigh10gof b,b8-oxydipropionitrileintoa
pressure; that is, propylene, carbon dioxide, ethane and meth-
brown 125-mL(4-oz) bottle.Add5gof silver nitrate (AgNO )
ane. Record the increase in pressure on the manometer as each
crystals. With occasional shaking, dissolve as muchAgNO as
component is added. Close the reservoir and evacuate the
possible, and allow the bottle to stand overnight to ensure
manometer after each addition.
saturation. Prepare this solution fresh, as required. Without
9.1.2 Dilution of Concentrate—Dilute the concentrate with
disturbing the crystals at the bottom of the bottle, weigh 2.5 g
high-purity ethylene in a ratio of approximately 1:4000. This
of supernatant AgNO solution into a 250-mL beaker and add
canbedonebyaddingthecalculatedamountoftheconcentrate
50 mL of methanol. While stirring this mixture, slowly add
and high purity ethylene to an evacuated cyclinder using the
22.5 g of activated carbon. Place the beaker on a steam bath to
gas-blending apparatus (Fig. 2). Use a source of high-pressure,
evaporate the methanol. When the impregnated activated
high-purity ethylene equipped with a needle valve and a
carbon appears to be dry, remove the beaker from the steam
pressure gage capable of accurately measuring the pressure of
bath and finish drying in an oven at 100 to 110°C for 2 h. Plug
theblendastheethyleneisaddedtothecylindercontainingthe
one end of a 4-ft (1.2-m) length of 6.4-mm ( ⁄4-in.) outside
concentrate.Add the calculated amount of ethylene; warm one
diameter aluminum or stainless steel tubing with glass wool.
end of the cylinder to ensure mixing of the blend. Allow the
Hold the tubing vertically with the plugged end down and pour
temperature to reach equilibrium before recording the final
freshly dried column packing into it, vibrating the column
pressure on the cylinder. Prepare at least three calibration
during filling to ensure uniform packing. Plug the top end with
samples containing the compounds to be determined over the
glass wool and shape the tubing so that it may be mounted
range of concentration desired in the products to be analyzed.
conveniently in the chromatograph.
9.2 Calculation of Composition of Standard Mixtures—
8.3 Hexamethylphosphoramide Column (HMPA)—Dry the
Calculate the exact ratio of the concentrate dilution with
35 to 80-mesh inert support at 204°C (400°F). Weigh 75 g into
ethylene by correcting the pressure of the ethylene added for
awide-mouth500-mL(16-oz)bottle.Add15gofHMPAtothe
the compressibility of ethylene (Table 2). Multiply the dilution
inert support and shake and roll the mixture until the support
ratio or factor by the percentage of each component present in
appears to be uniformly wet with the HMPA. Pour the packing
the original concentrate (Table 1). These calculations give the
into a 6-m (20-ft) length of 6.4-mm ( ⁄4-in.) outside diameter
amount of each component that has been added to the
copper of aluminum tubing plugged at one end with glass
high-purityethyleneblendstock.Theactualcompositionofthe
wool. Vibrate the tubing while filling to ensure more uniform
final blend must be ascertained by making corrections for the
packing. Plug the top end of the column with glass wool and
impurities present in the high-purity ethylene used for the
shape the column to fit into the chromatograph.
blendstock.Theamountofcorrectionisdeterminedbymaking
8.4 Hexadecane-SqualaneColumn—Dissolve 30 g of hexa-
chromatograph runs on the high-purity ethylene and measuring
decane into approximately 100 mL of acetone.Add 70 g of 80
the peak heights of the impurities. These peak heights will be
to 100-mesh inert support. Mix thoroughly and pour the
mixture into an open pan for drying. The slurry should be
TABLE 1 Suggested Composition of a Concentrate of Impurities
stirred during drying to ensure uniform distribution. When the
Used in
...