Biological evaluation of medical devices - Part 12: Sample preparation and reference materials - Amendment 1 (ISO 10993 12:2021/DAM 1:2024)

Biologische Beurteilung von Medizinprodukten - Teil 12: Probenvorbereitung und Referenzmaterialien - ÄNDERUNG 1 (ISO 10993 12:2021/DAM 1:2024)

Évaluation biologique des dispositifs médicaux - Partie 12: Préparation des échantillons et matériaux de référence - Amendement 1 (ISO 10993 12:2021/DAM 1:2024)

Biološko ovrednotenje medicinskih pripomočkov - 12. del: Priprava vzorcev in referenčni materiali - Dopolnilo A1 (ISO 10993 12:2021/DAM 1:2024)

General Information

Status
Not Published
Public Enquiry End Date
10-Sep-2024
Technical Committee
Current Stage
4020 - Public enquire (PE) (Adopted Project)
Start Date
01-Jul-2024
Due Date
18-Nov-2024

Relations

Buy Standard

Draft
EN ISO 10993-12:2021/oprA1:2024
English language
8 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-september-2024
Biološko ovrednotenje medicinskih pripomočkov - 12. del: Priprava vzorcev in
referenčni materiali - Dopolnilo A1 (ISO 10993 12:2021/DAM 1:2024)
Biological evaluation of medical devices - Part 12: Sample preparation and reference
materials - Amendment 1 (ISO 10993 12:2021/DAM 1:2024)
Biologische Beurteilung von Medizinprodukten - Teil 12: Probenvorbereitung und
Referenzmaterialien - ÄNDERUNG 1 (ISO 10993 12:2021/DAM 1:2024)
Évaluation biologique des dispositifs médicaux - Partie 12: Préparation des échantillons
et matériaux de référence - Amendement 1 (ISO 10993 12:2021/DAM 1:2024)
Ta slovenski standard je istoveten z: EN ISO 10993-12:2021/prA1
ICS:
11.100.20 Biološko ovrednotenje Biological evaluation of
medicinskih pripomočkov medical devices
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

DRAFT
Amendment
ISO 10993-12:2021/
DAM 1
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 12:
2024-06-19
Sample preparation and reference
Voting terminates on:
materials 2024-09-11
AMENDMENT 1
Évaluation biologique des dispositifs médicaux —
Partie 12: Préparation des échantillons et matériaux de référence
AMENDEMENT 1
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
NATIONAL REGULATIONS.
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Reference number
ISO 10993-12:2021/DAM 1:2024(en)
DRAFT
ISO 10993-12:2021/DAM 1:2024(en)
Amendment
ISO 10993-12:2021/
DAM 1
ISO/TC 194
Biological evaluation of medical
Secretariat: DIN
devices —
Voting begins on:
Part 12:
Sample preparation and reference
Voting terminates on:
materials
AMENDMENT 1
Évaluation biologique des dispositifs médicaux —
Partie 12: Préparation des échantillons et matériaux de référence
AMENDEMENT 1
ICS: 11.100.20
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENTS AND APPROVAL. IT
IS THEREFORE SUBJECT TO CHANGE
AND MAY NOT BE REFERRED TO AS AN
INTERNATIONAL STANDARD UNTIL
PUBLISHED AS SUCH.
This document is circulated as received from the committee secretariat.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
© ISO 2024
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
STANDARDS MAY ON OCCASION HAVE TO
ISO/CEN PARALLEL PROCESSING
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
BE CONSIDERED IN THE LIGHT OF THEIR
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
or ISO’s member body in the country of the requester.
NATIONAL REGULATIONS.
ISO copyright office
RECIPIENTS OF THIS DRAFT ARE INVITED
CP 401 • Ch. de Blandonnet 8
TO SUBMIT, WITH THEIR COMMENTS,
CH-1214 Vernier, Geneva
NOTIFICATION OF ANY RELEVANT PATENT
Phone: +41 22 749 01 11
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION.
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland Reference number
ISO 10993-12:2021/DAM 1:2024(en)
ii
ISO 10993-12:2021/DAM 1:2024(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of
medical devices.
This amendment to ISO 10993-12:2021 refines the following:
— the way in which ISO 14971 (in clause 4), ISO 10993-1 (in clause 7) and ISO 10993-18 and ISO/TS 10993-19
(in 10.1) are referenced and thus the inclusion of these standards in clause 2:
— change of Table 1;
— deletion of a sentence in 10.3.4.
A list of all parts in the ISO 10993 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iii
ISO 10993-12:2021/DAM 1:2024(en)
Biological evaluation of medical devices —
Part 12:
Sample preparation and reference materials
AMENDMENT 1
Clause 2
Change to read:
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 10993-18, Biological evaluation of medical devices — Part 18: Chemical characterization of medical
device materials within a risk management process
ISO/TS 10993-19, Biological evaluation of medical devices — Part 19: Physico-chemical, morphological and
topographical characterization of materials
ISO 14971, Medical devices — Application of risk management to medical devices

Clause 4
Change the wording in the first sentence of the first paragraph from “as described in ISO 14971” to “in
accordance with ISO 14971”.
Clause 7
Change the wording in the first sentence of the first paragraph from “(see ISO 10993-1)” to “in accordance
with ISO 10993-1”.
Clause 10.1
Change the wording in the second sentence of the first paragraph from “(see ISO 10993-18 and
ISO/TS 10993-19)” to “in accordance with ISO 10993-18 and ISO/TS 10993-19”.

10.3.3
Add the following paragraph after the first paragraph:

ISO 10993-12:2021/DAM 1:2024(en)
For medical devices containing absorbent materials, the individual absorption capacity of these
materials should be considered to determine the overall extra
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.