OSIST prEN ISO 10083:2005
(Main)Oxygen concentrator supply systems for use with medical gas pipeline systems (ISO/FDIS 10083:2006)
Oxygen concentrator supply systems for use with medical gas pipeline systems (ISO/FDIS 10083:2006)
Message from D Upstone (Secretary of CEN/TC 215) - 2006-05-31:
Dear Ashok,
I attach document CEN/TC215 N 1092 giving the results of the parallel Formal Vote and a covering letter from the Secretariat of CEN/TC215. You will see that the document passed in ISO but failed in CEN. According to the Guidelines on the Vienna Agreement, these results should be discussed by CEN and ISO including representatives of the TCs concerned (which in the case of ISO is ISO/TC121/SC6 – Chairman Dr David Peel, Secretary Dr Susan Dorsch, ISO programme manager Mr Tim Hancox).
The recommendation of the Chairman and Secretary of CEN/TC215 is that the CEN project should be cancelled and the document should be published as an ISO-only standard. I trust N1092 gives sufficient information and justification to proceed along these lines. Please let me know if you require anything further from me. As far as I can see, none of the pre-formatted resolutions cover this case, i.e. mandated item, not covered by or
Versorgungssysteme mit Sauerstoff-Konzentratoren zur Verwendung mit Rohrleitungssystemen für medizinische Gase (ISO/FDIS 10083:2006)
1.1 Diese Internationale Norm legt Anforderungen für die Gestaltung und Installation eines Versorgungssystems mit Sauerstoff Konzentratoren fest, das zusammen mit einem Rohrverteilersystem für medizinische Gase nach ISO 7396 1 verwendet wird.
1.2 Sie gilt nur für Versorgungssysteme mit Sauerstoff Konzentratoren, die mit Sauerstoff angereicherte Luft mit einer Sauerstoffkonzentration nicht unter 90 % herstellen (siehe 4.5.1).
1.3 Sauerstoff Konzentratoren für den häuslichen Gebrauch liegen nicht im Anwendungsbereich dieser Internationalen Norm.
ANMERKUNG Anforderungen an Sauerstoff Konzentratoren für den häuslichen Gebrauch sind in ISO 8359 festgelegt.
Systemes d'approvisionnement concentrateur d'oxygene pour utilisation dans des réseaux de distribution de gaz médicaux (ISO/FDIS 10083:2006)
Sistemi za koncentriranje kisika za uporabo s sistemi napeljav za medicinske pline (ISO/DIS 10083:2004)
General Information
- Status
- Not Published
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 98 - Abandoned project (Adopted Project)
- Start Date
- 03-Aug-2012
- Due Date
- 08-Aug-2012
- Completion Date
- 03-Aug-2012
Overview
The OSIST prEN ISO 10083:2005 standard, titled Oxygen Concentrator Supply Systems for Use with Medical Gas Pipeline Systems (ISO/FDIS 10083:2006), is a critical guideline developed by CEN/TC 215 in collaboration with ISO/TC 121/SC 6. This standard defines the safety, technical, and operational requirements for oxygen concentrator systems connected to medical gas pipeline installations in healthcare settings. The document serves as a comprehensive resource ensuring the reliable and safe supply of oxygen directly from concentrators to medical gas networks, supporting patient care and clinical outcomes.
The standard addresses the design, installation, testing, and maintenance of oxygen concentrator supply systems, aligning with essential EU directives such as Directive 93/42/EEC on medical devices. Compliance with OSIST prEN ISO 10083:2005 provides presumption of conformity with relevant regulatory frameworks and promotes harmonization in medical gas supply systems within European and international markets.
Key Topics
Key topics covered in OSIST prEN ISO 10083:2005 include:
- System Design and Safety – Requirements for robust construction, use of appropriate materials, and design specifications to ensure continuous, safe oxygen supply.
- Supply Sources – Guidelines for primary, secondary, and reserve oxygen sources, including the integration of cylinder reserves and system location best practices.
- Component Specifications – Standards for crucial components like oxygen concentrator units, pressure-relief valves, shut-off valves, oxygen monitors, and sample ports to guarantee system integrity.
- Monitoring and Alarms – Detailed requirements for monitoring systems, alarm signals, and operating alerts to detect faults promptly and maintain patient safety.
- Installation Procedures – Installation standards including electrical system considerations and compliances to facilitate safe and effective setup.
- Testing and Commissioning – Procedures for systematic testing, commissioning, certification, and documentation to verify system compliance and operational readiness.
- Maintenance and Documentation – Recommendations for preventive maintenance, retrials, and comprehensive documentation, including installation instructions, operating manuals, and certification forms.
- Implementation and Use of Oxygen 90+ – Specific considerations relating to the acceptance, calibration, timing, and mixing constraints for oxygen with 90+% purity used in these systems.
Applications
The OSIST prEN ISO 10083:2005 standard is essential for:
- Healthcare Facilities – Hospitals and clinics deploying oxygen concentrator systems integrated with medical gas pipeline systems to ensure a safe and effective oxygen supply.
- Medical Gas Suppliers – Companies involved in the design, manufacture, installation, and maintenance of medical gas systems utilizing oxygen concentrators.
- Regulatory Bodies – Authorities assessing compliance with EU medical device directives and international safety standards.
- Maintenance Teams – Professionals carrying out preventive maintenance and certification on oxygen supply systems within healthcare infrastructures.
- Contractors and Engineers – Specialists undertaking installation and commissioning work on medical gas pipeline systems involving oxygen concentrators.
By following this standard, healthcare providers ensure consistent oxygen delivery, reduce risk of supply interruptions, and maintain compliance with international medical gas safety regulations.
Related Standards
The OSIST prEN ISO 10083:2005 standard is related to and should be used in conjunction with the following standards for comprehensive medical gas system management:
- ISO 7396-1 – Medical gas pipeline systems - Part 1: Pipeline systems for compressed medical gases and vacuum, which provides overarching pipeline system design and safety standards.
- ISO 10083:1992 – The original edition of oxygen concentrator supply systems, which OSIST prEN ISO 10083:2005 revises and updates to reflect current technology and safety practices.
- EU Directive 93/42/EEC – Relevant medical devices directive establishing essential requirements for safety and performance.
- ISO 13485 – Quality management systems for medical devices, covering manufacturing quality and control relevant to oxygen concentrators.
Healthcare providers and system designers should leverage these associated standards to ensure integrated, compliant, and effective medical gas supply infrastructures.
Keywords: oxygen concentrator supply systems, medical gas pipeline systems, ISO 10083, medical oxygen supply, healthcare gas safety, oxygen purity 90+, medical gas installation, system commissioning, preventive maintenance medical gases, CEN/TC 215 standard, EU medical device directives.
Frequently Asked Questions
OSIST prEN ISO 10083:2005 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Oxygen concentrator supply systems for use with medical gas pipeline systems (ISO/FDIS 10083:2006)". This standard covers: Message from D Upstone (Secretary of CEN/TC 215) - 2006-05-31: Dear Ashok, I attach document CEN/TC215 N 1092 giving the results of the parallel Formal Vote and a covering letter from the Secretariat of CEN/TC215. You will see that the document passed in ISO but failed in CEN. According to the Guidelines on the Vienna Agreement, these results should be discussed by CEN and ISO including representatives of the TCs concerned (which in the case of ISO is ISO/TC121/SC6 – Chairman Dr David Peel, Secretary Dr Susan Dorsch, ISO programme manager Mr Tim Hancox). The recommendation of the Chairman and Secretary of CEN/TC215 is that the CEN project should be cancelled and the document should be published as an ISO-only standard. I trust N1092 gives sufficient information and justification to proceed along these lines. Please let me know if you require anything further from me. As far as I can see, none of the pre-formatted resolutions cover this case, i.e. mandated item, not covered by or
Message from D Upstone (Secretary of CEN/TC 215) - 2006-05-31: Dear Ashok, I attach document CEN/TC215 N 1092 giving the results of the parallel Formal Vote and a covering letter from the Secretariat of CEN/TC215. You will see that the document passed in ISO but failed in CEN. According to the Guidelines on the Vienna Agreement, these results should be discussed by CEN and ISO including representatives of the TCs concerned (which in the case of ISO is ISO/TC121/SC6 – Chairman Dr David Peel, Secretary Dr Susan Dorsch, ISO programme manager Mr Tim Hancox). The recommendation of the Chairman and Secretary of CEN/TC215 is that the CEN project should be cancelled and the document should be published as an ISO-only standard. I trust N1092 gives sufficient information and justification to proceed along these lines. Please let me know if you require anything further from me. As far as I can see, none of the pre-formatted resolutions cover this case, i.e. mandated item, not covered by or
OSIST prEN ISO 10083:2005 is classified under the following ICS (International Classification for Standards) categories: 11.040.10 - Anaesthetic, respiratory and reanimation equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
OSIST prEN ISO 10083:2005 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/023. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
OSIST prEN ISO 10083:2005 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI OSIST prEN ISO 10083:2005
PREDSTANDARD
januar 2005
Sistemi za koncentriranje kisika za uporabo s sistemi napeljav za medicinske
pline (ISO/DIS 10083:2004)
(istoveten prEN ISO 10083:2004)
Oxygen concentrator supply systems for use with medical gas pipeline systems
(ISO/DIS 10083:2004)
ICS 11.040.10 Referenčna številka
© Standard je založil in izdal Slovenski inštitut za standardizacijo. Razmnoževanje ali kopiranje celote ali delov tega dokumenta ni dovoljeno
EUROPEAN STANDARD
DRAFT
NORME EUROPÉENNE
EUROPÄISCHE NORM
October 2004
ICS
English version
Oxygen concentrator supply systems for use with medical gas
pipeline systems (ISO/DIS 10083:2004)
Systèmes d'approvisionnement concentrateurs d'oxygène
pour utilisation dans des réseaux de distribution de gaz
médicaux (ISO/DIS 10083:2004)
This draft European Standard is submitted to CEN members for parallel enquiry. It has been drawn up by the Technical Committee
CEN/TC 215.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other language
made by translation under the responsibility of a CEN member into its own language and notified to the Management Centre has the same
status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France,
Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia,
Slovenia, Spain, Sweden, Switzerland and United Kingdom.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2004 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN ISO 10083:2004: E
worldwide for CEN national Members.
Foreword
This document (prEN ISO 10083:2004) has been prepared by Technical Committee ISO/TC
121 "Anaesthetic and respiratory equipment" in collaboration with Technical Committee
CEN/TC 215 "Respiratory and anaesthetic equipment", the secretariat of which is held by
BSI.
This document is currently submitted to the parallel Enquiry.
This document has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association, and supports essential requirements
of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of
this document.
Endorsement notice
The text of ISO 10083:2004 has been approved by CEN as prEN ISO 10083:2004 without
any modifications.
ANNEX ZA
(informative)
Relationship between this European Standard and the
Essential Requirements of EU Directive 93/42 EEC
This European Standard has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association to provide a means of conforming to
Essential Requirements of the New Approach Directive 93/42 EEC.
Once this standard is cited in the Official Journal of the European Communities under that
Directive and has been implemented as a national standard in at least one Member State,
compliance with the normative clauses of this standard given in Table ZA. confers, within the
limits of the scope of this standard, a presumption of conformity with the corresponding
Essential Requirements of that Directive and associated EFTA regulations.
WARNING: Other requirements and other EU Directives may be applicable to the product(s)
falling within the scope of this standard.
DRAFT INTERNATIONAL STANDARD ISO/DIS 10083
ISO/TC 121/SC 6 Secretariat: ANSI
Voting begins on: Voting terminates on:
2004-10-07 2005-03-07
INTERNATIONAL ORGANIZATION FOR STANDARDIZATION • МЕЖДУНАРОДНАЯ ОРГАНИЗАЦИЯ ПО СТАНДАРТИЗАЦИИ • ORGANISATION INTERNATIONALE DE NORMALISATION
Oxygen concentrator supply systems for use with medical gas
pipeline systems
Systèmes d'approvisionnement concentrateur d'oxygène pour utilisation dans des réseaux de distribution de
gaz médicaux
[Revision of first edition (ISO 10083:1992)]
ICS 11.040.10
ISO/CEN PARALLEL ENQUIRY
The CEN Secretary-General has advised the ISO Secretary-General that this ISO/DIS covers a subject
of interest to European standardization. In accordance with the ISO-lead mode of collaboration as
defined in the Vienna Agreement, consultation on this ISO/DIS has the same effect for CEN
members as would a CEN enquiry on a draft European Standard. Should this draft be accepted, a
final draft, established on the basis of comments received, will be submitted to a parallel two-month FDIS
vote in ISO and formal vote in CEN.
In accordance with the provisions of Council Resolution 15/1993 this document is circulated in
the English language only.
Conformément aux dispositions de la Résolution du Conseil 15/1993, ce document est distribué
en version anglaise seulement.
To expedite distribution, this document is circulated as received from the committee secretariat.
ISO Central Secretariat work of editing and text composition will be undertaken at publication
stage.
Pour accélérer la distribution, le présent document est distribué tel qu'il est parvenu du
secrétariat du comité. Le travail de rédaction et de composition de texte sera effectué au
Secrétariat central de l'ISO au stade de publication.
THIS DOCUMENT IS A DRAFT CIRCULATED FOR COMMENT AND APPROVAL. IT IS THEREFORE SUBJECT TO CHANGE AND MAY NOT BE
REFERRED TO AS AN INTERNATIONAL STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS BEING ACCEPTABLE FOR INDUSTRIAL, TECHNOLOGICAL, COMMERCIAL AND USER PURPOSES, DRAFT
INTERNATIONAL STANDARDS MAY ON OCCASION HAVE TO BE CONSIDERED IN THE LIGHT OF THEIR POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN NATIONAL REGULATIONS.
© International Organization for Standardization, 2004
ISO/DIS 10083
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but shall
not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the
unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
Copyright notice
This ISO document is a Draft International Standard and is copyright-protected by ISO. Except as permitted
under the applicable laws of the user's country, neither this ISO draft nor any extract from it may be
reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, photocopying,
recording or otherwise, without prior written permission being secured.
Requests for permission to reproduce should be addressed to either ISO at the address below or ISO's
member body in the country of the requester.
ISO copyright office
Case postale 56 CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Reproduction may be subject to royalty payments or a licensing agreement.
Violators may be prosecuted.
©
ii ISO 2004 – All rights reserved
ISO/DIS 10083
Contents Page
Foreword.v
Introduction.vi
1 Scope .1
2 Normative references.1
3 Terms and definitions .2
4 General Requirements.4
4.1 Safety .4
4.2 R Alternative construction .5
4.3 Materials.5
4.4 System design.6
4.5 Specifications for oxygen 90+ .6
4.6 Cylinder filling.6
5 Sources of supply .6
5.1 General.7
5.2 Primary source of supply .7
5.3 Secondary source of supply .7
5.4 Reserve source of supply.8
5.5 Sources of supply with cylinders .8
5.6 Location of oxygen concentrator supply systems .8
6 Requirements for components.8
6.1 Oxygen concentrator unit.8
6.2 Oxygen Monitors.9
6.3 Pressure-relief valves. 10
6.4 Shut-off valves. 10
6.5 Sample port. 10
7 Monitoring and alarm systems. 10
7.1 General. 10
7.2 Monitoring and alarm signals. 10
7.3 Operating alarms. 11
7.4 Information signals . 11
8 Marking . 11
9 Installation. 11
9.1 General. 11
9.2 Electrical Systems. 12
10 Testing, commissioning and certification . 12
10.1 General. 12
10.2 Tests and procedures. 12
10.3 Specific tests. 12
10.4 Commissioning and certification . 14
11 Information to be supplied by the manufacturer . 14
11.1 Instructions for installation . 14
11.2 Instructions for use . 14
11.3 Instructions for preventive maintenance. 15
11.4 ‘As installed’ drawings. 15
11.5 Electrical schematics. 15
11.6 Certificates of inspection, performance testing or qualification. 15
ISO/DIS 10083
12 Implementation of use of oxygen 90+ . 15
12.1 Acceptance of oxygen 90+ . 15
12.2 Timing. 15
12.3 Mixing of oxygen 90+ and oxygen. 16
12.4 Calibration of medical equipment . 16
12.5 Labelling of terminal units . 16
12.6 Compliance with ISO 7396-1 . 16
Annex A (informative) Schematic representations of oxygen concentrator supply systems. 17
Annex B (informative) General guidelines for locations of supply systems. 25
Annex C (informative) Guidelines for emergency procedures . 27
C.1 General . 27
C.2 Communication. 27
C.3 Conservation of gas supplies . 27
C.4 Remedial actions . 28
C.5 Training. 28
C.6 Additional cylinder reserves. 28
Annex D (informative) Procedure for testing and commissioning . 29
D.1 Introduction . 29
D.2 General . 29
D.3 Procedure . 29
Annex E (informative) Typical forms for certification of an oxygen concentrator supply system. 30
Annex F (informative) Recommended minimum requirements for preventive maintenance. 34
F.1 General . 34
F.2 Organization . 34
F.3 Retesting. 35
F.4 Documentation. 35
Annex G (informative) Recommendations for installation . 36
Annex H (informative) Risk and risk management . 37
Annex I (informative) Recommendations for sizing and capacity . 38
Annex J (informative) Recommendations for filling cylinders with oxygen 90+. 39
J.1 General . 39
J.2 Filling system control . 39
J.3 Manifolds . 39
J.4 Connectors . 39
J.5 Termination of filling. 39
Annex K (informative) Rationale . 40
Bibliography . 41
Annex ZA (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC Medical Devices . 42
iv © ISO 2004 – All rights reserved
ISO/DIS 10083
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 10083 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment,
Subcommittee SC 6, Medical gas systems
This second edition cancels and replaces the first edition (1992), which has been technically revised.
Annexes A, B, C, D, E, F, G, H, I, J, K, L and M are given for information only.
ISO/DIS 10083
Introduction
This purpose of this International Standard is to specify minimum safety and performance requirements for
oxygen concentrator supply systems used to deliver oxygen 90+ to a pipeline distribution system. Oxygen 90+
is a generic term, developed by the subcommittee and intended to include the range of oxygen concentrations
produced by oxygen concentrator supply systems. National, regional or local regulations may stipulate the
minimum concentration of oxygen to be produced by an oxygen concentrator supply system, or the range of
concentrations which the supply system shall produce.
Oxygen concentrators can be used to deliver oxygen-enriched air to a medical gas pipeline system as a
substitute for medical oxygen. Oxygen concentrators may be combined with sources of supply containing
100% medical oxygen (i.e. in cylinders or cryogenic vessels).
Oxygen concentrators can supply a product gas with an oxygen concentration variable within a specified
range (e.g. 93 ± 3%) depending on the characteristic of the concentrator and the flow supplied. The oxygen
concentration supplied will therefore vary from 90 to 100% in normal operating conditions.
The selection of oxygen 90+ is a decision of the healthcare facility and outside the scope of this standard. This
standard should not be construed as an endorsement or recommendation of one concentration of oxygen over
another. The use of supply systems with oxygen concentrators may require the approval of regional or
national authorities.
A supply system with oxygen concentrators can be installed at the time of the installation of the pipeline
distribution system or as a replacement or augmentation of an existing supply system. A supply system with
oxygen concentrators can be supplied as a package and may be installed by a third party. In this case the
manufacturer of the oxygen concentrator supply system must provide the installer with appropriate information
for installation and testing before connecting the supply system to the pipeline distribution system and before
use.
Objectives of this standard are to ensure the following:
- appropriate introduction of an oxygen concentrator supply system into a health care facility;
- acceptable quality of the oxygen 90+ delivered by the system;
- continuous supply of oxygen 90+;
- use of suitable materials;
- cleanliness of components;
- correct installation;
- provision of appropriate control, monitoring and alarm systems for the supply system;
- testing, commissioning and certification.
It is intended for use by those persons involved in the design, construction, inspection or operation of health
care facilities. Those persons involved in the design, manufacture, calibration or testing of equipment intended
to be connected to a pipeline system supplied by an oxygen concentrator supply system should also be aware
of the contents of this document.
Annex K contains rationale statements for some of the requirements of ISO 10083. It is included to provide
additional insight into the reasoning that led to the requirements and recommendations that have been
incorporated in this International Standard. The clauses and subclauses marked with R after their number
have corresponding rationale contained in annex K. It is considered that knowledge of the reasons for the
requirements will not only facilitate the proper application of this International Standard, but will expedite any
subsequent revisions.
vi © ISO 2004 – All rights reserved
DRAFT INTERNATIONAL STANDARD ISO/DIS 10083
Oxygen concentrator supply systems for use with medical gas
pipeline systems
1 Scope
1.1 This standard specifies requirements for design and installation of an oxygen concentrator supply
system for use with a pipeline distribution system.
1.2 Oxygen concentrators for domiciliary use are excluded from the scope of this standard.
NOTE Requirements for oxygen concentrators for domiciliary use are specified in ISO 8359.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 407:1991, Small medical gas cylinders - Pin-index yoke-type valve connections.
ISO 3746:1995, Acoustics – Determination of sound power levels of noise sources using sound pressure –
Survey method using an enveloping measurement surface over a reflecting plane.
ISO 5145:1990, Cylinder valve outlets for gases and gas mixtures - - Selection and dimensioning.
ISO 5359:2000, Low-pressure hose assemblies for use with medical gases.
ISO 7396-1:2002, Medical gas pipeline systems – Part 1: Pipelines for compressed medical gases and
vacuum.
ISO/TR 7470:1998, Valve outlets for gas cylinders -- List of provisions which are either standardized or in use.
ISO 9703-1:1992, Anaesthesia and respiratory care alarm signals – Part 1: Visual alarm signals.
ISO 14971:2000, Medical devices – Risk management -- Part 1: Application of risk analysis.
ISO 15001:2003, Anaesthetic and respiratory equipment---compatibility with oxygen.
ISO 21969, High-pressure flexible connections for use with medical gases.
EN 286-1:2002, Simple unfired pressure vessels designed to contain air or nitrogen – Part 1: Pressure
vessels for general purposes.
ISO/DIS 10083
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1
commissioning
proof of function to verify that the agreed system specification is met and is accepted by the user or his
representative
3.2
control equipment
those items necessary to maintain the oxygen 90+ supply within the specified operating parameters
Examples are pressure regulators, pressure-relief valves, alarms, sensors and oxygen monitors.
3.3 cylinder bundle
pack or pallet of cylinders linked together with a single connector for filling and emptying
3.4
gas-specific
having characteristics which prevent connections between different gas services
3.5
manifold
device for connecting the outlet(s) of one or more cylinders or cylinder bundles of the same medical gas to the
pipeline system
3.6
manufacturer
the natural or legal person with responsibility for the design, manufacture, packaging and labelling of a device
before it is placed on the market under his own name, regardless of whether these operations are carried out
by that person himself or on his behalf by a third party
3.7
medical gas pipeline system
complete system which comprises a supply system, a monitoring and alarm system and a pipeline
distribution system with terminal units at the points where medical gases or vacuum may be
required
3.8
nominal distribution pressure
pressure which the medical gas pipeline system is intended to deliver at the terminal units
3.9
non-return valve
valve which permits flow in one direction only
3.10
operating alarm
alarm to indicate to technical staff that it is necessary to replenish the supply or to correct a malfunction
3.11
oxygen concentrator
device which produces oxygen 90+ from ambient air by extraction of nitrogen
3.12
oxygen concentrator supply system
supply system containing one or more oxygen concentrators
2 © ISO 2004 – All rights reserved
ISO/DIS 10083
3.13
oxygen concentrator unit
source of supply for oxygen 90+ comprising compressor(s), sieve bed(s) and at least one storage vessel
3.14
oxygen 90+ storage vessel
pressurized container to store oxygen 90+
3.15
oxygen 90+
medical gas produced by an oxygen concentrator and meeting the specifications in section 4.5 of this
International Standard
3.16
peak demand
maximum anticipated oxygen flowrate required by a health care facility
NOTE This is commonly expressed in litres per minute.
3.17
pipeline distribution system
that part of a medical gas pipeline system linking the supply system to the terminal units
3.18
pressure regulator
device which reduces the inlet pressure of a gas and maintains its set outlet pressure within specified limits
3.19
pressure-relief valve
valve which opens to atmosphere at a preset pressure and which is intended to prevent excess positive
pressure
3.20
primary source of supply
that portion of the supply system that normally delivers the medical gas to the pipeline distribution system
3.21
reserve source of supply
that portion of the supply system which delivers the medical gas to the pipeline distribution system in the event
of failure or exhaustion of both the primary and secondary sources of supply
3.23
secondary source of supply
that portion of the supply system that delivers the medical gas to the pipeline distribution system in the event
of failure or exhaustion of the primary source of supply
3.24
shut-off valve
valve which prevents flow in both directions when closed
3.25
single fault condition
condition in which a single means for protection against a safety hazard in equipment is defective or a single
external abnormal condition is present
3.26
source of supply
that portion of the supply system with associated control equipment which supplies the pipeline distribution
system
ISO/DIS 10083
3.27
supply system
assembly which delivers a medical gas to the pipeline distribution system and which includes all sources of
supply
3.28
system design flow
flow calculated from the maximum flow requirement of the health care facility and corrected by the diversity
factor(s)
3.29
terminal unit
outlet assembly (inlet for vacuum) in a medical gas pipeline system at which the operator makes connections
and disconnections
4 General Requirements
4.1 Safety
4.1.1 Oxygen concentrator supply systems shall, when installed, commissioned, operated in normal use
and maintained according to the instructions of the manufacturer, cause no safety hazard which could
reasonably be foreseen using risk analysis procedures in accordance with ISO 14971 and which is connected
with their intended application, in normal condition and in single fault condition.
NOTE 1 Loss of mains electrical power or water supply is considered a single fault condition.
NOTE 2 Some risks and risk management measures are given in annex H.
NOTE 3 Risk management may require that critical components [e.g. air compressor(s)] be rated for continuous
service.
NOTE 4 It may be necessary to connect some oxygen concentrators to the emergency electrical power supply. In
circumstances where an emergency power system is not available or cannot accommodate an oxygen concentrator, the
risk management procedure shall determine the appropriate capacity of the secondary and reserve sources of supply.
NOTE 5 See ISO/TR 16142 for information.
4.1.2 Control equipment shall be designed so that components such as pressure regulators can be
maintained without interrupting the gas supply to the pipeline distribution system.
4.1.3 The system shall be designed so that maintenance or failure of any component shall not require the
isolation of two sources of supply at the same time.
4.1.4 An oxygen concentrator supply system shall cause no interruption of supply in normal condition and in
single fault condition.
NOTE Loss of mains electrical power or water supply is a single fault condition.
4.1.5 Shutting off or failure of an oxygen concentrator unit shall not affect the delivery of gas from the
oxygen concentrator supply system to the pipeline distribution system.
NOTE An oxygen concentrator takes a certain time to achieve the specified concentration of oxygen after a
prolonged shutdown.
4 © ISO 2004 – All rights reserved
ISO/DIS 10083
4.2 R Alternative construction
Installations and components or parts thereof, using materials or having forms of construction different from
those detailed in this International Standard shall be accepted if it can be demonstrated that an equivalent
degree of safety is obtained. Evidence of an equivalent degree of safety shall be provided by the manufacturer.
4.3 Materials
4.3.1 Compatibility with oxygen
4.3.1.1 All components of an oxygen concentrator supply system that are liable to come into contact with
a medical gas (compressed air, oxygen or oxygen 90+) shall be compatible with oxygen under the operating
conditions specified by the manufacturer, taking into account the requirements of 4.1.1.
NOTE Compatibility with oxygen or oxygen 90+ involves both combustibility and ease of ignition. Materials which
burn in air will burn violently in pure oxygen or oxygen 90+. Many materials which do not burn in air will do so in pure
oxygen or oxygen 90+, particularly under pressure. Similarly, materials which can be ignited in air require less energy to
ignite in oxygen or oxygen 90+. Many such materials may be ignited by friction at a valve seat or by adiabatic
compression produced when oxygen or oxygen 90+ at high pressure is rapidly introduced into a system initially at low
pressure.
4.3.1.2 If lubricants are used, they shall be compatible with oxygen at the operating conditions of the
supply system.
Evidence of compliance with 4.3.1.1 and 4.3.1.2 shall be made available by the manufacturer upon request.
4.3.2 Cleanliness
All components of an oxygen concentrator supply system that are liable to come into contact with a medical
gas (compressed air, oxygen or oxygen 90+) shall meet the cleanliness requirements of ISO 15001. Such
components shall be protected from contamination prior to and during installation.
Evidence of compliance shall be made available by the manufacturer upon request.
NOTE Examples of cleaning procedures are given in ISO 15001.
4.3.3 Resistance to corrosion
The manufacturer shall disclose, upon request, evidence of the corrosion resistance of the materials used for
pipes and fittings.
NOTE Corrosion resistance includes resistance against the influence of moisture and the surrounding materials.
4.3.4 Exposure to cylinder pressure
Components which may be exposed to cylinder pressure in normal or single fault condition shall function
according to their specifications after being exposed to a pressure of 1,5 times the cylinder working pressure
for 5 min.
Evidence shall be provided by the manufacturer upon request.
ISO/DIS 10083
4.4 System design
4.4.1 Sizing
An oxygen concentrator supply system shall be designed to deliver the oxygen flow requirements specified by
the health care facility. See annex I for information on sizing.
4.4.2 Pipeline distribution pressure control
The oxygen concentrator supply system shall be capable of maintaining the nominal distribution pressure as
required in ISO 7396-1.
4.5 Specifications for oxygen 90+
4.5.1 Oxygen 90+ produced by an oxygen concentrator supply system shall comply with national or regional
regulations. Where such regulations do not exist, the oxygen 90+ shall comply with the following:
a) minimum oxygen concentration 90% volume fraction of oxygen
b) maximum carbon monoxide concentration 5 ml /m
c) maximum carbon dioxide concentration 300 ml/m
d) 3
maximum particulate concentration 0,5 mg/m
e) maximum oil concentration 0,5 mg/m measured at ambient pressure
f) maximum water content 60 mg/m .
NOTE 1 The balance of the gas comprises predominantly argon and nitrogen.
NOTE 2 Other terms may be used by national or regional regulations.
4.5.2 Particulate filters shall be installed immediately downstream of the oxygen concentrator(s) to ensure
that the oxygen 90+ meets the particulate requirements of 4.5.1. Means shall be provided to indicate the
status of filter elements, e.g. by measuring the pressure drop across the filter.
4.6 Cylinder filling
If an oxygen concentrator unit is used to fill cylinders with oxygen 90+ the following conditions shall be met:
+
a) Means shall be provided to ensure that cylinder filling does not affect delivery of oxygen 90 to the
pipeline distribution system.
b) A sample port with a shut-off valve shall be provided adjacent to the filling system.
c) The cylinder filling system shall comply with the requirements of national or regional regulations.
NOTE Recommendations for filling cylinders with oxygen 90+ are provided in annex J.
5 Sources of supply
Schematic representations of oxygen concentrator supply systems are shown in annex A.
6 © ISO 2004 – All rights reserved
ISO/DIS 10083
5.1 General
5.1.1 An oxygen concentrator supply system shall be designed for automatic operation, and shall contain
the following sources of supply (see annex A):
a) a primary supply;
b) a secondary supply and
c) a reserve supply.
5.1.2 Each source of supply shall be capable of supplying the system design flow which is determined by
the healthcare facility.
5.1.2 A non-return valve and a shut-off valve shall be fitted immediately downstream of each source of
supply.
5.2 Primary source of supply
The primary source of supply shall consist of one of the following:
a) one or more oxygen concentrator unit(s);
b) a combination of one or more oxygen concentrator unit(s) and high-pressure cylinders or cryogenic
vessels;
c) a combination of one or more oxygen concentrator unit(s) and a liquid oxygen supply.
The cylinder(s), or cylinder bundle(s) shall be connected downstream of the oxygen storage vessel shut-off
valve and upstream of the connection of the secondary supply.
5.3 Secondary source of supply
5.3.1 The secondary source of supply shall consist of one of the following:
a) one or more oxygen concentrator units;
b) gas (oxygen or oxygen 90+) in cylinder(s) or cylinder bundle(s);
c) a liquid oxygen supply.
5.3.2 If the secondary source of supply consists of only cylinders or cylinder bundles, the manifold shall:
a) consist of at least two banks of cylinders;
b) be supplied from one bank at a time;
c) on depletion of one bank switch to another bank.
5.3.3 If an emergency electrical power supply is not available, the secondary source of supply shall not
comprise only oxygen concentrator unit(s).
5.3.4 The secondary source shall be connected downstream of the primary supply.
5.3.5 The secondary source of supply shall be connected downstream of the primary source of supply
storage vessel.
ISO/DIS 10083
5.4 Reserve source of supply
5.4.1 The reserve source of supply shall consist of gas (oxygen or oxygen 90+) in cylinders or cylinder
bundles.
5.4.2 The reserve source of supply shall:
a) consist of at least two banks of cylinders;
b) be supplied from one bank of cylinders at a time;
c) on depletion of one bank of cylinders switch automatically to another bank.
NOTE This configuration is needed to facilitate replenishment of the reserve source of supply without interruption of
supply.
5.4.3 The reserve source of supply shall be connected downstream of the supply system shut-off valve.
5.5 Sources of supply with cylinders
NOTE Cylinders or cylinder bundles may be used in sources of supply with cylinders.
5.5.1 Each bank of cylinders shall have its cylinders connected to a manifold with its own pressure regulator.
Vent valves, if fitted on manifolds, should be vented outside of the building.
5.5.2 A filter having a pore size no greater than 100 micrometers shall be provided between the cylinder(s)
and the first pressure regulator.
5.5.3 A non-return valve shall be installed at the manifold end of each flexible connection between a
cylinder and the manifold.
5.5.4 The flexible connections between each cylinder and the manifold shall comply with ISO 21969.
5.5.5 Cylinder valve outlets shall be in accordance with ISO 5145 or the relevant national standard.
5.6 Location of oxygen concentrator supply systems
NOTE Informative guidelines are given in annex B.
5.6.1 The ambient temperature in rooms for the oxygen concentrator supply system shall be in the range of
10 °C to 40 °C.
5.6.2 The noise level from the oxygen concentrator supply system shall comply with local regulations.
NOTE Oxygen concentrator units may generate noise in excess of 70 db.
6 Requirements for components
6.1 Oxygen concentrator unit
6.1.1 An oxygen concentrator unit shall consist of:
a) a compressed air supply with at least one air compressor,
b) at least one sieve bed and switching valves,
c) at least one oxygen 90+ storage vessel and
8 © ISO 2004 – All rights reserved
ISO/DIS 10083
d) a sample port with a shut-off valve immediately downstream of the oxygen 90+ storage vessel.
6.1.2 An air compressor may be connected to a receiver.
6.1.3 A receiver may be served by more than one compressor.
6.1.4 The intake(s) for the air compressors shall be located where there is minimal contamination from
internal combustion engine exhaust (e.g. from motor vehicles), vacuum system exhausts, vents from medical
gas pipeline systems, anaesthetic gas scavenging systems, ventilation system discharges and other sources
of contamination. The intake shall be provided with means to prevent the ingress of insects, debris and water.
Consideration should be given to the potential effects of prevailing winds on the location of intake(s).
NOTE The above requirements are taken from ISO 7396-1.
6.1.5 Receivers shall:
a) comply with EN 286-1 or equivalent national standards;
b) be fitted with shut-off valve(s), an automatic drain, a pressure gauge and a pressure-relief valve.
6.1.6 Each receiver shall be fitted with a means of pressure control, e.g. pressure switch (es) or pressure
transducer(s).
6.1.7 Each
...




Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...