Transfusion equipment for medical use - Part 5: Transfusion sets for single use with pressure infusion apparatus (ISO 1135-5:2015)

ISO 1135-5:2015 specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures up to 200 kPa (2 bar). This International Standard ensures compatibility with containers for blood and blood components as well as intravenous equipment.
Secondary aims of ISO 1135-5:2015 are to provide guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components.
Platelet components should not be transfused under pressure using these sets.
In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-5:2015.

Transfusionsgeräte zur medizinischen Verwendung - Teil 5: Transfusionsgeräte zur einmaligen Verwendung mit Druckinfusionsapparaten (ISO 1135-5:2015)

Dieser Teil von ISO 1135 legt Anforderungen an zur einmaligen Verwendung bestimmte Transfusionsgeräte fest, die mit Druckinfusionsapparaten mit erzeugten Drücken bis zu 200 kPa (2 bar) arbeiten. Diese Internationale Norm stellt die Kompatibilität mit Behältern für Blut und Blutbestandteile und Geräten zur intravenösen Punktion sicher.
Weitere Ziele dieses Teils von ISO 1135 sind die Bereitstellung einer Anleitung zu Spezifikationen der Qualität und Leistungsfähigkeit von Werkstoffen, die für Transfusionsgeräte verwendet werden, die Angabe von Benennungen für Einzelteile von Transfusionsgeräten und die Sicherstellung der Kompatibilität der Geräte mit roten Blutkörperchen und Blutplasmabestandteilen.
Thrombozyten-Bestandteile sollten mit diesen Geräten nicht unter Druck übertragen werden.
In einigen Ländern sind das nationale Arzneibuch oder andere nationale Bestimmungen rechtlich bindend und haben Vorrang vor diesem Teil von ISO 1135.

Matériel de transfusion à usage médical - Partie 5: Appareils de transfusion non réutilisables avec des appareils de perfusion sous pression (ISO 1135-5:2015)

ISO 1135-5:2015 spécifie les exigences relatives aux appareils de transfusion non réutilisables, couplés à des appareils de perfusion sous pression capables de générer des pressions pouvant atteindre 200 kPa (2 bar). La présente Norme internationale assure la compatibilité avec les conteneurs de sang ou de composants sanguins, ainsi qu'avec le matériel d'injection intraveineuse.
Les objectifs secondaires de l'ISO 1135-5:2015 consistent à fournir des préconisations concernant les spécifications de qualité et de performance des matériaux utilisés dans les appareils de transfusion, de proposer des désignations pour les composants de ces appareils de transfusion, et d'assurer la compatibilité de ces appareils avec les produits hématiques et plasmiques du sang.
Il convient de ne pas transfuser de plaquettes sous pression avec ces appareils.
Dans certains pays, la pharmacopée ou d'autres règlements nationaux sont légalement obligatoires et ont donc préséance sur l'ISO 1135-5:2015.

Transfuzijska oprema za uporabo v medicini - 5. del: Transfuzijske garniture za enkratno uporabo s tlačno črpalko (ISO 1135-5:2015)

Ta del standarda ISO 1135 opredeljuje zahteve za transfuzijske garniture za enkratno uporabo s tlačno črpalko, ki lahko ustvari tlak do 200 kPa (2 bara). Ta mednarodni standard zagotavlja združljivost z vsebniki za kri in krvne sestavine ter z intravenozno opremo.
Druga cilja tega dela standarda ISO 1135 sta zagotoviti navodila za specifikacije v zvezi s kakovostjo in učinkovitostjo materialov, uporabljenih za transfuzijske garniture, predstaviti oznake za sestavne dele transfuzijskih garnitur, ter zagotoviti združljivost garnitur z naborom sestavin celic rdeče vrste in plazemskih krvnih sestavin. V teh garniturah naj ne bi izvajali transfuzije sestavin trombocitov pod tlakom. V nekaterih državah je nacionalna farmakopeja ali druga nacionalna zakonodaja pravno zavezujoča in ima prednost pred tem delom standarda ISO 1135.

General Information

Status
Withdrawn
Public Enquiry End Date
30-Dec-2013
Publication Date
01-Feb-2016
Withdrawal Date
21-Dec-2025
Technical Committee
VAZ - Healthcare
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
18-Dec-2025
Due Date
10-Jan-2026
Completion Date
22-Dec-2025

Relations

Effective Date
01-Mar-2016
Effective Date
01-Feb-2026
Standard

SIST EN ISO 1135-5:2016

English language
30 pages
Preview
Preview
e-Library read for
1 day

Frequently Asked Questions

SIST EN ISO 1135-5:2016 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Transfusion equipment for medical use - Part 5: Transfusion sets for single use with pressure infusion apparatus (ISO 1135-5:2015)". This standard covers: ISO 1135-5:2015 specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures up to 200 kPa (2 bar). This International Standard ensures compatibility with containers for blood and blood components as well as intravenous equipment. Secondary aims of ISO 1135-5:2015 are to provide guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components. Platelet components should not be transfused under pressure using these sets. In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-5:2015.

ISO 1135-5:2015 specifies requirements for single use transfusion sets for use with pressure infusion equipment capable of generating pressures up to 200 kPa (2 bar). This International Standard ensures compatibility with containers for blood and blood components as well as intravenous equipment. Secondary aims of ISO 1135-5:2015 are to provide guidance on specifications relating to the quality and performance of materials used in transfusion sets, to present designations for transfusion set components, and to ensure the compatibility of sets with red cell and plasma blood components. Platelet components should not be transfused under pressure using these sets. In some countries, the national pharmacopoeia or other national regulations are legally binding and take precedence over ISO 1135-5:2015.

SIST EN ISO 1135-5:2016 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

SIST EN ISO 1135-5:2016 has the following relationships with other standards: It is inter standard links to SIST EN ISO 1135-4:2012, SIST EN ISO 1135-5:2026. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

SIST EN ISO 1135-5:2016 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/023. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

You can purchase SIST EN ISO 1135-5:2016 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of SIST standards.

Standards Content (Sample)


SLOVENSKI STANDARD
01-marec-2016
1DGRPHãþD
SIST EN ISO 1135-4:2012
7UDQVIX]LMVNDRSUHPD]DXSRUDERYPHGLFLQLGHO7UDQVIX]LMVNHJDUQLWXUH]D
HQNUDWQRXSRUDERVWODþQRþUSDONR ,62
Transfusion equipment for medical use - Part 5: Transfusion sets for single use with
pressure infusion apparatus (ISO 1135-5:2015)
Transfusionsgeräte zur medizinischen Verwendung - Teil 5: Transfusionsgeräte zur
einmaligen Verwendung mit Druckinfusionsapparaten (ISO 1135-5:2015)
Matériel de transfusion à usage médical - Partie 5: Appareils de transfusion non
réutilisables avec des appareils de perfusion sous pression (ISO 1135-5:2015)
Ta slovenski standard je istoveten z: EN ISO 1135-5:2015
ICS:
11.040.20 Transfuzijska, infuzijska in Transfusion, infusion and
injekcijska oprema injection equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EN ISO 1135-5
EUROPEAN STANDARD
NORME EUROPÉENNE
December 2015
EUROPÄISCHE NORM
ICS 11.040.20 Supersedes EN ISO 1135-4:2012
English Version
Transfusion equipment for medical use - Part 5:
Transfusion sets for single use with pressure infusion
apparatus (ISO 1135-5:2015)
Matériel de transfusion à usage médical - Partie 5: Transfusionsgeräte zur medizinischen Verwendung -
Appareils de transfusion non réutilisables avec les Teil 5: Transfusionsgeräte zur einmaligen Verwendung
appareils de perfusion sous pression (ISO 1135- mit Druckinfusionsapparaten (ISO 1135-5:2015)
5:2015)
This European Standard was approved by CEN on 24 July 2015.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 1135-5:2015 E
worldwide for CEN national Members.

Contents Page
European foreword . 3
Annex ZA (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on Medical devices . 5
European foreword
This document (EN ISO 1135-5:2015) has been prepared by Technical Committee ISO/TC 76
“Transfusion, infusion and injection, and blood processing equipment for medical and pharmaceutical
use” in collaboration with Technical Committee CEN/TC 205 “Non-active medical devices” the
secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by June 2016, and conflicting national standards shall be
withdrawn at the latest by June 2016.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent
rights.
Together with EN ISO 1135-4:2015 this document supersedes EN ISO 1135-4:2012.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this
document.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta,
Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
Endorsement notice
The text of ISO 1135-5:2015 has been approved by CEN as EN ISO 1135-5:2015 without any
modification.
The following referenced documents are indispensable for the application of this document. For
undated references, the latest edition of the referenced document (including any amendments) applies.
For dated references, only the edition cited applies. However, for any use of this standard ‘within the
meaning of Annex ZA’, the user should always check that any referenced document has not been
superseded and that its relevant contents can still be considered the generally acknowledged state-of-
art.
When an IEC or ISO standard is referred to in the ISO standard text, this shall be understood as a
normative reference to the corresponding EN standard, if available, and otherwise to the dated ISO or
IEC standard, as listed in Table 1.
NOTE The way in which these references documents are cited in normative requirements determines the
extent (in whole or in part) to which they apply.

Table 1 — Correlations between undated normative references and dated EN and ISO standards
Normative references as listed in Equivalent dated standard
Clause 2 of the ISO standard
EN ISO or IEC
ISO 594-1 --- ISO 594-1:1986
ISO 594-2 --- ISO 594-2:1998
ISO 3696 EN ISO 3696:1995 ISO 3696:1987
ISO 3826-1:2013 EN ISO 3826-1:2013 ISO 3826-1:2013
ISO 3826-2 EN ISO 3826-2:2008 ISO 3826-2:2008
ISO 10993-1 EN ISO 10993-1:2009 ISO 10993-1:2009
ISO 10993-4 EN ISO 10993-4:2009 ISO 10993-4:2002 plus ISO 10993-4 AMD
1:2006
ISO 14644-1 EN ISO 14644-1:1999 ISO 14644-1:1999
ISO 15223-1 EN ISO 15223-1:2012 ISO 15223-1:2012

Annex ZA
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on Medical devices
This European Standard has been prepared under a mandate given to CEN by the European
Commission and the European Free Trade Association to provide a means of conforming to Essential
Requirements of the New Approach Directive 93/42/EEC, Medical devices
Once this standard is cited in the Official Journal of the European Union under that Directive and has
been implemented as a national standard in at least one Member State, compliance with the clauses of
this standard given in Table ZA.1 confers, within the limits of the scope of this standard, a presumption
of conformity with the corresponding Essential Requirements of that Directive and associated EFTA
regulations.
NOTE 1 Where a reference from a clause of this standard to the risk management process is made, the risk
management process needs to be in compliance with Directive 93/42/EEC / Directive 90/385/EEC, as amended
by 2007/47/EC. This means that risks have to be reduced ‘as far as possible’, ‘to a minimum’, ‘to the lowest
possible level’, ‘minimized’ or ‘removed’, according to the wording of the corresponding essential requirement.
NOTE 2 The manufacturer’s policy for determining acceptable risk must be in compliance with essential
requirements 1, 2, 5, 6, 7, 8, 9, 11 and 12 of the Directive.
NOTE 3 This Annex ZA is based on Normative References according to Table of References, replacing the
references in the core text.
NOTE 4 When an Essential Requirement does not appear in Table ZA.1, it means that it is not addressed by this
European Standard.
Table ZA.1 — Correspondence between this European Standard and Directive 93/42/EEC,
Medical devices
Clause(s)/sub-clause(s) of this EN Essential Requirements (ERs) Qualifying remarks/Notes
of Directive 93/42/EEC
4.2, 6.1, 6.2, 6.3, Clause 7, Clause 8 7.2 The part of ER 7.2 relating to
packaging is not addressed (→ for
packaging see Clause 10 of this
standard).
Clause 5, 6.1, 6.2, 6.3, Clause 7, 7.3 ER covered by biological evaluation
Clause 8
6.2, 6.3, 6.10, 9.2, 9.3, A.2, A.4 7.5 Only the first paragraph is covered.
6.1, 6.3 7.6
4.2, Clause 6 8.1
6.12, Clause 9, Clause 10 8.3 Maintenance of sterility in storage is
covered.
8.2 8.4 Sterilization process is covered.
6.1, A.1 8.5
9.2, 9.3 8.7
Clause(s)/sub-clause(s) of this EN Essential Requirements (ERs) Qualifying remarks/Notes
of Directive 93/42/EEC
6.4 9.1 The second sentence of ER 9.1 is not
addressed.
6.4 refers to ISO 3826-1.
Clauses 4, 5, 6, 7, 8 9.2
6.2, 6.3, A.2 12.7 Only 12.7.1 is addressed.
Only tensile strength is addressed.
Clause 9 13.1
9.2, 9.3 13.2 ISO 15223-1 and ISO 3826-2 are
addressed when using symbols.
9.2 a), b), c), d), e), f), g), i), j), k), 9.3 13.3 The part of 13.3a) relating to the
a), b), c), d), e), f), g) authorized representative is not
addressed. Presumption of
conformity to the rest of 13.3a) is
only provided if the name and
address of the manufacturer are
given.
13.3b) is addressed in Clause 3.1 and
4.3.
13.3d) is only covered if the batch
number is preceded by the word
'LOT'.
13.3f) Requirement „indication of
single use must be consistent across
the Community“ is not addressed in
the standard.
13.3g) and h) are not addressed in
the standard.
9.2, 9.3 13.4 13.4 is addressed regarding to the
label.
WARNING — Other requirements and other EU Directives may be applicable to the product(s)
falling within the scope of this standard.

INTERNATIONAL ISO
STANDARD 1135-5
First edition
2015-12-01
Transfusion equipment for medical
use —
Part 5:
Transfusion sets for single use with
pressure infusion apparatus
Matériel de transfusion à usage médical —
Partie 5: Appareils de transfusion non réutilisables avec les appareils
de perfusion sous pression
Reference number
ISO 1135-5:2015(E)
©
ISO 2015
ISO 1135-5:2015(E)
© ISO 2015, Published in Switzerland
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Ch. de Blandonnet 8 • CP 401
CH-1214 Vernier, Geneva, Switzerland
Tel. +41 22 749 01 11
Fax +41 22 749 09 47
copyright@iso.org
www.iso.org
ii © ISO 2015 – All rights reserved

ISO 1135-5:2015(E)
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 General requirements . 2
4.1 Nomenclature for components of the transfusion set . 2
4.2 Maintenance of sterility. 3
5 Materials . 3
6 Physical requirements . 4
6.1 Particulate contamination . 4
6.2 Leakage . 4
6.3 Tensile strength . 4
6.4 Closure-piercing device . 4
6.5 Tubing . 5
6.6 Filter for blood and blood components . 5
6.7 Drip chamber and drip tube . 5
6.8 Flow regulator . 5
6.9 Flow rate of blood and blood components . 5
6.10 Injection site . 6
6.11 Male conical fitting . 6
6.12 Protective caps . 6
6.13 Storage volume . 6
7 Chemical requirements . 6
7.1 Reducing (oxidizable) matter . 6
7.2 Metal ions . 6
7.3 Titration acidity or alkalinity. 6
7.4 Residue on evaporation . 6
7.5 UV absorption of extract solution . 7
8 Biological requirements . 7
8.1 General . 7
8.2 Sterility . 7
8.3 Pyrogenicity . 7
8.4 Haemolysis . 7
8.5 Toxicity . 7
8.6 Assessment of blood component depletion . 7
8.7 Assessment of damage to blood components . 7
9 Labelling . 8
9.1 General . 8
9.2 Unit container . 8
9.3 Shelf or multi-unit container . 9
10 Packaging . 9
11 Disposal . 9
Annex A (normative) Physical tests .10
Annex B (normative) Chemical tests .14
Annex C (normative) Biological tests .16
Annex D (normative) Storage volume .17
Bibliography .20
ISO 1135-5:2015(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation on the meaning of ISO specific terms and expressions related to conformity
assessment, as well as information about ISO’s adherence to the WTO principles in the Technical
Barriers to Trade (TBT) see the following URL: Foreword - Supplementary information
The committee responsible for this document is ISO/TC 76, Transfusion, infusion and injection, and blood
processing equipment for medical and pharmaceutical use.
This first edition of ISO 1135-5, together with ISO 1135-4, cancels and replaces ISO 1135-4:2012, which
has been technically revised with the following changes:
— the scope of ISO 1135-4 has been restricted to gravity feed applications, whereby, ISO 1135-5 is
focused on pressure infusion applications;
— a new Annex D on ‘Storage volume’ has been added.
ISO 1135 consists of the following parts, under the general title Transfusion equipment for medical use:
— Part 3: Blood-taking sets for single use
— Part 4: Transfusion sets for single use, gravity feed
— Part 5: Transfusion sets for single use with pressure infusion apparatus
iv © ISO 2015 – All rights reserved

INTERNATIONAL STANDARD ISO 1135-5:2015(E)
Transfusion equipment for medical use —
Part 5:
Transfusion sets for single use with pressure infusion
apparatus
1 Scope
This part of ISO 1135 specifies requirements for single use transfusion sets for use with pressure infusion
equipment capable of generating pressures up to 200 kPa (2 bar). This International Standard ensures
compatibility with containers for blood and blood components as well as intravenous equipment.
Secondary aims of this part of ISO 1135 are to provide guidance on specifications relating to the quality
and performance of materials used in transfusion sets, to present designations for transfusion set
components, and to ensure the compatibility of sets with red cell and plasma blood components.
Platelet components should not be transfused under pressure using these sets.
In some countries, the national pharmacopoeia or other national regulations are legally binding and
take precedence over this part of ISO 1135.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
1)
ISO 594-1 , Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical
equipment — Part 1: General requirements
1)
ISO 594-2 , Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical
equipment — Part 2: Lock fittings
ISO 3696, Water for analytical laboratory use — Specification and test methods
ISO 3826-1:2013, Plastics collapsible containers for human blood and blood components — Part 1:
Conventional containers
ISO 3826-2, Plastics collapsible containers for human blood and blood components — Part 2: Graphical
symbols for use on labels and instruction leaflets
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 10993-4, Biological evaluation of medical devices — Part 4: Selection of tests for interactions with blood
ISO 14644-1, Cleanrooms and associated controlled environments — Part 1: Classification of air cleanliness
ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to
be supplied — Part 1: General requirements

1) To be replaced by ISO 80369-7.
ISO 1135-5:2015(E)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
NOTE These terms and definitions are specifically applicable to Annex D.
3.1
filling volume
V
F
volume of tube during “pressure less” filling, respectively filling by gravity
Note 1 to entry: The tube remains unstressed.
Note 2 to entry: The filling volume is to be equated with the calculated volume of the tube.
3.2
storage volume
V
S
tube volume during pressurization equal to filling volume, V , plus bolus volume, V : V = V +V
F S S F B
3.3
bolus volume
V
B
increased tube volume during pressurization (storage volume, V ) in comparison with the unstressed
S
tube (filling volume, V )
F
Note 1 to entry: For illustration of the bolus volume, see Figure 1.
Key
1 patient 4 bolus volume
2 occlusion 5 syringe pump
3 tube
Figure 1 — Bolus volume
4 General requirements
4.1 Nomenclature for components of the transfusion set
The nomenclature for components of transfusion sets is given in Figure 2.
2 © ISO 2015 – All rights reserved

ISO 1135-5:2015(E)
Key
1 protective cap of the closure-piercing device 8 flow regulator
2 closure-piercing device 9 injection site
3 fluid channel 10 male conical fitting
4 drip tube 11 protective cap of the male conical fitting
a
5 drip chamber Indicates alternative locations of the filter for blood
and blood components. Other designs are acceptable,
if the same safety aspects are ensured.
b
6 filter for blood and blood components Injection site is optional.
7 tubing
Figure 2 — Example of a transfusion set
4.2 Maintenance of sterility
The transfusion set shall be provided with protective caps to maintain sterility of the internal parts of
the set until the set is used.
5 Materials
The materials from which the transfusion sets given in Clause 4 are manufactured shall comply with the
requirements specified in Clause 6. If components of the transfusion set come into contact with blood
and blood components, they shall additionally comply with the requirements specified in Clauses 7 and 8.
ISO 1135-5:2015(E)
6 Physical requirements
6.1 Particulate contamination
The transfusion sets shall be manufactured under conditions that minimize particulate contamination.
All parts shall be smooth and clean at the fluid pathway surfaces. When tested as specified in A.1, the
number of particles detected shall not exceed the contamination index limit.
6.2 Leakage
The transfusion set, when tested in accordance with A.2, shall show no signs of air leakage.
6.3 Tensile strength
Any connections between the components of the transfusion set, excluding protective caps, shall
withstand a static tensile force of not less than 15 N for 15 s.
6.4 Closure-piercing device
6.4.1 The dimensions of the closure-piercing device shall conform to the
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...

The SIST EN ISO 1135-5:2016 standard is a comprehensive guideline for the design and use of transfusion sets intended for single use with pressure infusion apparatus. It succinctly outlines the critical specifications and requirements necessary to assure safety and efficacy in transfusion practices. One of the key strengths of ISO 1135-5:2015 lies in its clear delineation of requirements for equipment capable of generating pressures up to 200 kPa (2 bar). This aspect is vital for healthcare providers as it ensures that transfusion sets are designed to withstand the rigors of high-pressure infusion without compromising their structural integrity or function. Moreover, the standard emphasizes compatibility with containers for blood and blood components, as well as with intravenous equipment, which is crucial for maintaining the integrity of blood products during administration. This compatibility reduces the risk of contamination and ensures that clinical outcomes are optimized, thereby reinforcing patient safety. Another notable feature of the standard is its guidance on material quality and performance. By specifying the types of materials that should be used in the construction of transfusion sets, it aims to enhance the overall reliability and durability of these medical devices. Additionally, the standard provides designations for individual components within transfusion sets, facilitating better understanding and usage among healthcare professionals. It is also essential to note the standard's clear directive regarding the transfusion of platelet components, which should not be conducted under pressure using these sets. This consideration reflects an acute awareness of the nuances involved in blood component therapy and serves to further mitigate potential complications. Lastly, while ISO 1135-5:2015 sets a high benchmark for transfusion equipment, it acknowledges that in certain jurisdictions, national pharmacopoeias or other regulations may supersede its guidelines. This recognition underscores the importance of remaining compliant with local legislation, thereby fortifying the standard's relevance in a global context. Overall, SIST EN ISO 1135-5:2016 serves as a vital resource, ensuring that transfusion sets meet consistent quality and safety standards while addressing the specific requirements for their use with pressure infusion apparatus. Its comprehensive approach makes it an essential reference for manufacturers, healthcare providers, and regulatory bodies alike.

SIST EN ISO 1135-5:2016 표준은 압력 주입 장치와 함께 사용할 수 있도록 설계된 일회용 수혈 세트의 요구사항을 규정하고 있습니다. 이 표준은 최대 200 kPa(2 bar)의 압력을 생성할 수 있는 장비와의 호환성을 보장하여, 수혈 세트가 혈액 및 혈액 구성 요소, 그리고 정맥 주입 장비와 효과적으로 연계될 수 있도록 합니다. 이 표준의 강점 중 하나는 수혈 세트에 사용되는 재료의 품질 및 성능과 관련된 사양을 제공하는 것입니다. 이는 최상의 환자 안전과 치료 결과를 보장하기 위한 필수적인 요소입니다. 또한, 수혈 세트 구성 요소에 대한 명칭을 명확히 하여, 제조업체와 의료 제공자 간의 의사소통을 원활하게 하는 데 기여하고 있습니다. SIST EN ISO 1135-5:2016은 적혈구 및 플라즈마 혈액 구성 요소와의 호환성을 보장하는 데 중점을 두고 있으며, 이를 통해 혈액 수혈 과정에서 발생할 수 있는 안전 문제를 최소화하고자 합니다. 그러나, 이 표준은 혈소판 성분을 압력 하에 수혈할 수 없도록 안내하고 있어서, 의료 환경에서의 안전성을 더욱 강화합니다. 또한, 몇몇 국가에서는 국가 약전이나 기타 법적 규정이 ISO 1135-5:2015보다 우선할 수 있으므로, 관련 법률을 준수하는 것이 매우 중요합니다. 이러한 점에서 SIST EN ISO 1135-5:2016은 의료 기기 사용자와 제조업체가 준수해야 할 중요한 기준을 제공하는 표준으로서 그 필요성과 관련성을 충분히 갖추고 있습니다.

SIST EN ISO 1135-5:2016は、医療用の輸血機器に関する国際標準の一部であり、圧力注入装置用の使い捨て輸血セットについての要求事項を定めています。この標準は、圧力を200 kPa(2 bar)まで生成できる圧力注入機器と互換性を持つことを保証しており、輸血セットが血液および血液成分の容器、ならびに静脈内機器と適合するよう設計されています。 この標準の大きな強みは、輸血セットに使用される材料の品質および性能に関する仕様の指針を提供している点です。輸血セットの各コンポーネントの名称が明確にされており、セットの赤血球や血漿血液成分との互換性が確保されています。この明確性は、医療現場における適切な機器の選択と使用を促進し、患者の安全性を高めるために重要です。 また、ISO 1135-5:2015では、血小板成分を圧力をかけて輸血することが推奨されていないため、使用上の注意点も明示されています。これにより、医療従事者は特定の条件下で使用することのリスクについて理解し、安全な輸血を実施するための情報を得ることができます。 さらに、特定の国々では、国内の薬典または他の国の規制が法的に拘束力を持ち、ISO 1135-5:2015よりも優先される場合があることが明記されているため、国際標準と地域的な規制の両方に対応するための注意も促されています。 このように、SIST EN ISO 1135-5:2016は、圧力注入装置用の使い捨て輸血セットに関する包括的かつ実用的な基準を提供しており、その適用範囲と関連性は非常に高いと言えます。