SIST EN 60601-1-12:2015
(Main)Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment
IEC 60601-1-12:2014 constitutes a collateral standard to IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance hereafter referred to as the general standard. Medical practice is increasingly using medical electrical equipment and medical electrical systems for monitoring, treatment or diagnosis of patients in the emergency medical services environment. The safety of medical electrical equipment in this uncontrolled, rough environment is a cause for concern. This collateral standard was developed with contributions from clinicians, engineers and regulators. The terminology, requirements, general recommendations and guidance of this collateral standard are intended to be useful for manufacturers of medical electrical equipment and medical electrical systems and for technical committees responsible for the development of particular standards. This International Standard applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems, hereafter referred to as ME equipment and ME systems, which are intended, as indicated in the instructions for use by their manufacturer, for use in the EMS environment (Emergency Medical Services environment). The object of this collateral standard is to provide general requirements for ME equipment and ME systems carried to the scene of an emergency and used there, as well as in transport, in situations where the ambient conditions differ from indoor conditions. The object of this collateral standard is to specify general requirements that are in addition to those of the general standard and to serve as the basis for particular standards.
Medizinische elektrische Geräte - Teil 1-12: Allgemeine Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Anforderungen an medizinische elektrische Geräte und medizinische elektrische Systeme in der Umgebung für den Notfalleinsatz
Appareils électromédicaux - Partie 1-12: Exigences générales pour la sécurité de base et les performances essentielles - Norme collatérale: Exigences pour les appareils électromédicaux et les systèmes électromédicaux destinés à être utilisés dans l'environnement des services médicaux d'urgence
L'IEC 60601-1-12:2014 constitue une norme collatérale de l'IEC 60601-1: Appareils électromédicaux - Partie 1: Exigences générales pour la sécurité de base et les performances essentielles désignée ci-après par norme générale. La pratique médicale utilise de plus en plus des appareils électromédicaux et des systèmes électromédicaux pour la surveillance, le traitement ou le diagnostic des patients dans l'environnement des services médicaux d'urgence. La sécurité des appareils électromédicaux dans cet environnement particulièrement difficile et non contrôlé, est source de préoccupation. La présente norme collatérale a été élaborée grâce aux contributions de médecins cliniciens, d'ingénieurs et d'autorités de tutelle. La terminologie, les exigences, les recommandations générales et les lignes directrices contenues dans la présente norme collatérale ont pour objectif d'être utiles aux fabricants d'appareils électromédicaux et de systèmes électromédicaux et aux comités techniques responsables de l'élaboration des normes particulières. La présente Norme internationale s'applique à la sécurité de base et aux performances essentielles des appareils électromédicaux et des systèmes électromédicaux, désignés ci-après par appareils EM et systèmes EM, destinés, comme indiqué dans les instructions d'utilisation, par leur fabricant à être utilisés dans l'environnement SMU (environnement des Services Médicaux d'Urgence). L'objet de la présente norme collatérale est de fournir des exigences générales pour les appareils EM et les systèmes EM acheminés et utilisés sur les lieux d'une intervention en urgence, ainsi que pour leur transport, dans des situations où les conditions ambiantes sont différentes des conditions intérieures. La présente norme collatérale est destinée à spécifier des exigences générales qui viennent s'ajouter à celles de la norme générale et à servir de base pour les normes particulières.
Medicinska električna oprema - 1-12. del: Splošne zahteve za osnovno varnost in bistvene lastnosti - Spremljevalni standard: Zahteve za elektromedicinsko opremo in elektromedicinske sisteme, namenjene za uporabo v okolju nujne medicinske pomoči
Ta mednarodni standard velja za OSNOVNO VARNOST in BISTVENE LASTNOSTI MEDICINSKE ELEKTRIČNE OPREME in MEDICINSKIH ELEKTRIČNIH SISTEMOV (v nadaljevanju tudi: ELEKTROMEDICINSKA OPREMA in ELEKTROMEDICINSKI SISTEM), ki so namenjeni za uporabo v OKOLJU NUJNE MEDICINSKE POMOČI (OKOLJE EMS), kot je PROIZVAJALEC navedel v navodilih za uporabo in
kot je določeno v poglavju 3.1. OPOMBA 1 Za namene tega standarda je namen PROIZVAJALCA naveden v navodilih za uporabo. ODGOVORNA ORGANIZACIJA in IZVAJALEC morata upoštevati, da se lahko zaradi uporabe, ki ni skladna z NAMERAVANO UPORABO PROIZVAJALCA, PACIENT znajde v NEVARNEM STANJU. OKOLJE NUJNE MEDICINSKE POMOČI vključuje:
– odzivanje na in oživljanje PACIENTA na mestu intervencije, ki je poškodovan ali pod vplivom bolezni v predbolnišničnem okolju, ter prevoz PACIENTA do primerne strokovne zdravstvene ustanove za nadaljnje zdravljenje, med katerim se nadaljuje z oživljanjem.
– nadzorovanje, zdravljenje ali diagnosticiranje med prevozom med strokovnimi zdravstvenimi ustanovami.
Ta mednarodni standard se ne uporablja za ELEKTROMEDICINSKO OPREMO in ELEKTROMEDICINSKE SISTEME, namenjene izključno za uporabo v OKOLJU DOMAČE ZDRAVSTVENE OSKRBE ki so zajeti v standardu IEC 60601-1-11, ali namenjene izključno za uporabo v strokovnih zdravstvenih ustanovah, ki so zajeti v standardu IEC 60601-1 brez dodatkov standarda IEC 60601-1-11 ali tega spremljevalnega standarda. ELEKTROMEDICINSKA OPREMA in ELEKTROMEDICINSKI SISTEMI so pogosto namenjeni uporabi v različnih okoljih. Taka ELEKTROMEDICINSKA OPREMA in ELEKTROMEDICISNKI SISTEMI so lahko namenjeni uporabi v različnih okoljih in so tako v področju uporabe tega standarda tudi, če so med drugim namenjeni za uporabo v OKOLJU NUJNE MEDICINSKE POMOČI.
PRIMER: ELEKTROMEDICINSKA OPREMA ali ELEKTROMEDICINSKI SISTEMI namenjeni tako za uporabo v OKOLJU NUJNE MEDICINSKE POMOČI in okolju strokovnih zdravstvenih ustanov.
OPOMBA 2: ELEKTROMEDICINSKA OPREMA in ELEKTROMEDICINSKI SISTEMI, namenjeni uporabi v OKOLJU NUJNE MEDICINSKE POMOČI, se lahko uporabljajo na mestih z nezanesljivimi električnimi viri in v zunanjih okoljskih pogojih.
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-september-2015
0HGLFLQVNDHOHNWULþQDRSUHPDGHO6SORãQH]DKWHYH]DRVQRYQRYDUQRVWLQ
ELVWYHQHODVWQRVWL6SUHPOMHYDOQLVWDQGDUG=DKWHYH]DHOHNWURPHGLFLQVNRRSUHPR
LQHOHNWURPHGLFLQVNHVLVWHPHQDPHQMHQH]DXSRUDERYRNROMXQXMQHPHGLFLQVNH
SRPRþL
Medical electrical equipment - Part 1-12: General requirements for basic safety and
essential performance - Collateral standard: Requirements for medical electrical
equipment and medical electrical systems intended for use in the emergency medical
services environment
Medizinische elektrische Geräte - Teil 1-12: Allgemeine Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale - Ergänzungsnorm: Anforderungen
an medizinische elektrische Geräte und medizinische elektrische Systeme in der
Umgebung für den Notfalleinsatz
Appareils électromédicaux - Partie 1-12: Exigences générales pour la sécurité de base et
les performances essentielles - Norme collatérale: Exigences pour les appareils
électromédicaux et les systèmes électromédicaux destinés à être utilisés dans
l'environnement des services médicaux d'urgence
Ta slovenski standard je istoveten z: EN 60601-1-12:2015
ICS:
11.040.01 Medicinska oprema na Medical equipment in general
splošno
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN 60601-1-12
NORME EUROPÉENNE
EUROPÄISCHE NORM
May 2015
ICS 11.040
English Version
Medical electrical equipment - Part 1-12: General requirements
for basic safety and essential performance - Collateral Standard:
Requirements for medical electrical equipment and medical
electrical systems intended for use in the emergency medical
services environment
(IEC 60601-1-12:2014)
Appareils électromédicaux - Partie 1-12: Exigences Medizinische elektrische Geräte - Teil 1-12: Allgemeine
générales pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles - Norme collatérale: Exigences pour les wesentlichen Leistungsmerkmale - Ergänzungsnorm:
appareils électromédicaux et les systèmes électromédicaux Anforderungen an medizinische elektrische Geräte und
destinés à être utilisés dans l'environnement des services medizinische elektrische Systeme in der Umgebung für den
médicaux d'urgence Notfalleinsatz
(IEC 60601-1-12:2014) (IEC 60601-1-12:2014)
This European Standard was approved by CENELEC on 2014-07-24. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN 60601-1-12:2015 E
Foreword
The text of document 62A/932/FDIS, future edition 1 of IEC 60601-1-12, prepared by SC 62A
"Common aspects of electrical equipment used in medical practice", of IEC/TC 62 "Electrical
equipment in medical practice", was submitted to the IEC-CENELEC parallel vote and approved by
CENELEC as EN 60601-1-12:2015.
The following dates are fixed:
– latest date by which the document has to be implemented at (dop) 2015-11-22
national level by publication of an identical national
standard or by endorsement
– latest date by which the national standards conflicting with (dow) 2018-12-31
the document have to be withdrawn
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such
patent rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directives 93/42/EEC and 90/385/EEC, see informative Annexes ZZA
and ZZB, which are integral parts of this document.
Endorsement notice
The text of the International Standard IEC 60601-1-12:2014 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
IEC 60038:2009 NOTE Harmonized as EN 60038:2011 (modified).
IEC 60065 NOTE Harmonized as EN 60065.
IEC 60335-1:2010 NOTE Harmonized as EN 60335-1:2012 (modified).
IEC 60364 NOTE Harmonized in HD 384 / HD 60364 series (partly modified).
IEC 60721-3-7:1995 NOTE Harmonized as EN 60721-3-7:1995 (not modified)
+ A1:1996 + A1:1997 (not modified).
IEC 60950-1:2005 NOTE Harmonized as EN 60950-1:2006 (modified).
IEC 61032:1997 NOTE Harmonized as EN 61032:1998 (not modified).
ISO 10651-2:2004 NOTE Harmonized as EN ISO 10651-2:2009 (not modified).
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
NOTE 1 When an International Publication has been modified by common modifications, indicated by (mod),
the relevant EN/HD applies.
NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is
available here: www.cenelec.eu.
Publication Year Title EN/HD Year
IEC 60068-2-27 2008 Environmental testing - EN 60068-2-27 2009
Part 2-27: Tests - Test Ea and guidance:
Shock
IEC 60068-2-31 2008 Environmental testing - EN 60068-2-31 2008
Part 2-31: Tests - Test Ec: Rough handling
shocks, primarily for equipment-type
specimens
IEC 60068-2-64 2008 Environmental testing - EN 60068-2-64 2008
Part 2-64: Tests - Test Fh: Vibration,
broadband random and guidance
IEC 60529 1989 Degrees of protection provided by EN 60529 1991
- - enclosures (IP Code) + corrigendum May 1993
+ A1 1999 + A1 2000
IEC 60601-1 2005 Medical electrical equipment - EN 60601-1 2006
- - Part 1: General requirements for basic + corrigendum Mar. 2010
+ A1 2012 safety and essential performance + A1 2013
- - + A1/AC 2014
IEC 60601-1-2 2014 Medical electrical equipment - EN 60601-1-2 2014
Part 1-2: General requirements for basic
safety and essential performance -
Collateral standard: Electromagnetic
disturbances - Requirements and tests
IEC 60601-1-6 2010 Medical electrical equipment - EN 60601-1-6 2010
+ A1 2013 Part 1-6: General requirements for basic + A1 2015
safety and essential performance -
Collateral standard: Usability
IEC 60601-1-8 2006 Medical electrical equipment - EN 60601-1-8 2007
- - Part 1-8: General requirements for basic + corrigendum Mar. 2010
+ A1 2012 safety and essential performance - + A1 2013
- - Collateral Standard: General requirements, + A1/AC 2014
tests and guidance for alarm systems in
medical electrical equipment and medical
electrical systems
Publication Year Title EN/HD Year
IEC 60601-1-11 2015 Medical electrical equipment - EN 60601-1-11 2015
Part 1-11: General requirements for basic
safety and essential performance -
Collateral Standard: Requirements for
medical electrical equipment and medical
electrical systems used in the home
healthcare environment
CISPR 11 (mod) 2009 Industrial, scientific and medical EN 55011 2009
equipment - Radio-frequency disturbance
characteristics - Limits and methods of
measurement
ISO 7000 2014 Graphical symbols for use on equipment - - -
Registered symbols
ISO 7010 2011 Graphical symbols - Safety colours and EN ISO 7010 2012
+ A1 2012 safety signs - Registered safety signs + A1 2014
+ A2 2012 + A2 2014
+ A3 2012 + A3 2014
+ A4 2013 + A4 2014
+ A5 2014 + A5 2015
ISO 15223-1 2012 Medical devices - Symbols to be used with EN ISO 15223-1 2012
medical device labels, labelling and
information to be supplied -
Part 1: General requirements
Annex ZZA
(informative)
Coverage of Essential Requirements of EU Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association, and within its scope the Standard covers all
relevant essential requirements given in Annex I of EU Directive 93/42/EEC of 14 June 1993
concerning medical devices.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING: Other requirements and other EU Directives can be applied to the products falling within
the scope of this standard.
Annex ZZB
(informative)
Coverage of Essential Requirements of EU Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association, and within its scope the Standard covers all
relevant essential requirements given in Annex 1 of EU Directive 90/385/EEC of 20 June 1990 relating
to active implantable medical devices.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING: Other requirements and other EU Directives can be applied to the products falling within
the scope of this standard.
IEC 60601-1-12
Edition 1.0 2014-06
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
colour
inside
Medical electrical equipment –
Part 1-12: General requirements for basic safety and essential performance –
Collateral Standard: Requirements for medical electrical equipment and medical
electrical systems intended for use in the emergency medical services
environment
Appareils électromédicaux –
Partie 1-12: Exigences générales pour la sécurité de base et les performances
essentielles – Norme collatérale: Exigences pour les appareils électromédicaux
et les systèmes électromédicaux destinés à être utilisés dans l'environnement
des services médicaux d'urgence
INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
CODE PRIX XA
ICS 11.040 ISBN 978-2-8322-1654-5
– 2 – IEC 60601-1-12:2014 © IEC 2014
CONTENTS
FOREWORD . 4
INTRODUCTION . 7
1 Scope, object and related standards . 8
1.1 * Scope . 8
1.2 * Object . 8
1.3 Related standards . 9
1.3.1 IEC 60601-1 . 9
1.3.2 Particular standards . 9
2 Normative references . 9
3 Terms and definitions . 10
4 General requirements . 11
4.1 * Additional requirements for SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS . 11
4.2 * Environmental conditions for ME EQUIPMENT . 11
4.2.1 * Environmental conditions of transport and storage between uses . 12
4.2.2 * Environmental operating conditions . 13
5 * Classification of ME EQUIPMENT and ME SYSTEMS . 15
6 ME EQUIPMENT identification, marking and documents . 16
6.1 * Additional requirements for legibility of markings . 16
6.2 * Additional requirements for marking of IP classification . 16
6.3 * Instructions for use . 16
6.3.1 Additional general requirements . 16
6.3.2 * Additional requirements for an electrical power source . 17
6.3.3 Additional requirements for ME EQUIPMENT start-up PROCEDURE . 17
6.3.4 * Additional requirements for operating instructions . 18
6.3.5 Additional requirements for ME EQUIPMENT messages . 18
6.4 Technical description – FIXED or PERMANENTLY INSTALLED CLASS I
ME EQUIPMENT . 18
7 * Protection against electrical HAZARDS from ME EQUIPMENT . 18
8 Protection against excessive temperatures and other HAZARDS . 19
8.1 Additional requirements for ingress of water or particulate matter into
ME EQUIPMENT and ME SYSTEMS . 19
8.1.1 * Ingress of water or particulate matter into ME EQUIPMENT . 19
8.1.2 * Ingress of water or particulate matter into ME SYSTEMS . 19
8.2 Additional requirements for interruption of the power supply to ME EQUIPMENT
and ME SYSTEM . 19
8.3 * Additional requirements for INTERNAL ELECTRICAL POWER SOURCE for
ME EQUIPMENT . 20
9 * Accuracy of controls and instruments and protection against hazardous outputs . 21
10 Construction of ME EQUIPMENT . 21
10.1 * Additional requirements for mechanical strength of ME EQUIPMENT intended
for the EMS ENVIRONMENT . 21
10.1.1 General requirements for mechanical strength . 21
10.1.2 * Requirements for mechanical strength for FIXED or PERMANENTLY
INSTALLED ME EQUIPMENT intended for use in a road ambulance . 22
10.1.3 * Requirements for mechanical strength for TRANSPORTABLE
ME EQUIPMENT . 23
IEC 60601-1-12:2014 © IEC 2014 – 3 –
10.1.4 * Requirements for mechanical strength for ME EQUIPMENT intended for
airborne use . 24
10.2 Requirements for mounting of ME EQUIPMENT . 25
11 Additional requirements for electromagnetic compatibility of ME EQUIPMENT and
ME SYSTEMS . 25
Annex A (informative) General guidance and rationale . 26
A.1 General guidance . 26
A.2 Rationale for particular clauses and subclauses . 28
Annex B (informative) Guide to marking and labelling requirements for ME EQUIPMENT
and ME SYSTEMS . 42
B.1 Marking on the outside of ME EQUIPMENT, ME SYSTEMS or their parts . 42
B.2 ACCOMPANYING DOCUMENTS, instructions for use . 42
B.3 ACCOMPANYING DOCUMENTS, technical description . 43
Annex C (informative) Symbols on marking . 44
Bibliography . 46
Index of defined terms used in this collateral standard . 48
Figure A.1 – Saturation water vapour pressure as function of temperature . 31
Table 1 – Mechanical strength test applicability . 22
Table A.1 – Saturation water vapour pressure as function of temperature . 32
Table B.1 – Marking on the outside of ME EQUIPMENT, ME SYSTEMS or their parts . 42
Table B.2 – ACCOMPANYING DOCUMENTS, instructions for use . 42
Table B.3 – ACCOMPANYING DOCUMENTS, technical description . 43
Table C.1 – General symbols . 44
– 4 – IEC 60601-1-12:2014 © IEC 2014
INTERNATIONAL ELECTROTECHNICAL COMMISSION
_______________
MEDICAL ELECTRICAL EQUIPMENT –
Part 1-12: General requirements for basic
safety and essential performance –
Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems intended for use
in the emergency medical services environment
FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60601-1-12 has been prepared by a joint working group of
IEC subcommittee 62A: Common aspects of electrical equipment used in medical practice, of
IEC technical committee 62: Electrical equipment in medical practice, and ISO subcommittee
SC3: Lung ventilators and related devices, of ISO technical committee 121: Anaesthetic and
respiratory equipment.
This first edition constitutes a collateral standard to IEC 60601-1 (third edition): Medical
electrical equipment – Part 1: General requirements for basic safety and essential
performance hereafter referred to as the general standard.
IEC 60601-1-12:2014 © IEC 2014 – 5 –
The text of this collateral standard is based on the following documents:
FDIS Report on voting
62A/932/FDIS 62A/938/RVD
Full information on the voting for the approval of this collateral standard can be found in the
report on voting indicated in the above table. In ISO, this International Standard has been
approved by 18 P-members out of 19 having cast a vote.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In the 60601 series of publications, collateral standards specify general requirements for
safety applicable to:
– a subgroup of MEDICAL ELECTRICAL EQUIPMENT (e.g. radiological equipment); or
– a specific characteristic of all MEDICAL ELECTRICAL EQUIPMENT, not fully addressed in the
general standard (e.g. ALARM SYSTEMS).
In this collateral standard, the following print types are used:
– requirements and definitions: roman type.
– test specifications: italic type.
informative material appearing outside of tables, such as notes, examples and references: in smaller type.
–
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS COLLATERAL STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.3.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this collateral standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
Clauses, subclauses and definitions for which a rationale is provided in informative Annex A
are marked with an asterisk (*).
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website.
– 6 – IEC 60601-1-12:2014 © IEC 2014
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.
NOTE The attention of Member Bodies and National Committees is drawn to the fact that equipment
manufacturers and testing organizations may need a transitional period following publication of a new, amended or
revised ISO or IEC publication in which to make products in accordance with the new requirements and to equip
themselves for conducting new or revised tests. It is the recommendation of the committee that the content of this
publication be adopted for implementation nationally not earlier than 3 years from the date of publication.
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
re print this document using a
understanding of its contents. Users should therefo
colour printer.
IEC 60601-1-12:2014 © IEC 2014 – 7 –
INTRODUCTION
Medical practice is increasingly using MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL
SYSTEMS for monitoring, treatment or diagnosis of PATIENTS in the EMERGENCY MEDICAL
SERVICES ENVIRONMENT (see 3.1). The safety of MEDICAL ELECTRICAL EQUIPMENT in this
uncontrolled, rough environment is a cause for concern.
This collateral standard was developed with contributions from clinicians, engineers and
regulators. The terminology, requirements, general recommendations and guidance of this
collateral standard are intended to be useful for MANUFACTURERS of MEDICAL ELECTRICAL
EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS and for technical committees responsible for the
development of particular standards.
– 8 – IEC 60601-1-12:2014 © IEC 2014
MEDICAL ELECTRICAL EQUIPMENT –
Part 1-12: General requirements for basic
safety and essential performance –
Collateral Standard: Requirements for medical electrical
equipment and medical electrical systems intended for use
in the emergency medical services environment
1 Scope, object and related standards
1.1 * Scope
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS, hereafter referred to as
ME EQUIPMENT and ME SYSTEMS, which are intended, as indicated in the instructions for use by
MANUFACTURER, for use in the EMS ENVIRONMENT (EMERGENCY MEDICAL SERVICES
their
ENVIRONMENT), as defined in 3.1.
NOTE 1 For the purposes of this standard, the intent of the MANUFACTURER is indicated in the instructions for use.
The RESPONSIBLE ORGANIZATION and the OPERATOR need to be aware that any other use outside the
MANUFACTURER’S INTENDED USE can result in a HAZARDOUS SITUATION for the PATIENT.
The EMS ENVIRONMENT includes
– responding to and providing life support at the scene of an emergency to a PATIENT
reported as experiencing injury or illness in a pre-hospital setting, and transporting the
PATIENT, while continuing such life support care, to an appropriate professional healthcare
facility for further care.
– providing monitoring, treatment or diagnosis during transport between professional
healthcare facilities.
This International Standard does not apply to ME EQUIPMENT and ME SYSTEMS intended solely
for use in the HOME HEALTHCARE ENVIRONMENT covered by IEC 60601-1-11 or solely for use in
professional healthcare facilities covered by IEC 60601-1 without the additions of
IEC 60601-1-11 or this collateral standard. ME EQUIPMENT and ME SYSTEMS are often not solely
intended for one environment. Such ME EQUIPMENT or ME SYSTEM can be intended for multiple
use environments, and as such, if also intended for use in the EMS ENVIRONMENT, are within
the scope of this standard.
EXAMPLE ME EQUIPMENT or ME SYSTEM intended for both the EMS ENVIRONMENT and the professional healthcare
facility environment.
NOTE 2 EMS ENVIRONMENT ME EQUIPMENT and ME SYSTEMS can be used in locations with unreliable electrical
sources and outdoor environmental conditions.
1.2 * Object
The object of this collateral standard is to provide general requirements for ME EQUIPMENT and
ME SYSTEMS carried to the scene of an emergency and used there, as well as in transport, in
situations where the ambient conditions differ from indoor conditions.
The object of this collateral standard is to specify general requirements that are in addition to
those of the general standard and to serve as the basis for particular standards.
IEC 60601-1-12:2014 © IEC 2014 – 9 –
1.3 Related standards
1.3.1 IEC 60601-1
For ME EQUIPMENT and ME SYSTEMS, this collateral standard complements IEC 60601-1,
hereafter referred to as the general standard.
When referring to IEC 60601-1 or to this collateral standard, either individually or in
combination, the following conventions are used:
– "the general standard" designates IEC 60601-1 alone;
– "this collateral standard" designates IEC 60601-1-12 alone;
– "this standard" designates the combination of the general standard and this collateral
standard.
1.3.2 Particular standards
A requirement in a particular standard takes priority over the corresponding requirement in
this collateral standard.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and
are indispensable for its application. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any
amendments) applies.
NOTE 1 The way in which these referenced documents are cited in normative requirements determines the extent
(in whole or in part) to which they apply.
NOTE 2 Informative references are listed in the bibliography on page 46.
IEC 60068-2-27:2008, Environmental testing – Part 2-27: Tests – Test Ea and guidance:
Shock
IEC 60068-2-31:2008, Environmental testing – Part 2-31: Tests – Test Ec: Rough handling
shocks, primarily for equipment-type specimens
IEC 60068-2-64:2008, Environmental testing – Part 2-64: Tests – Test Fh: Vibration,
broadband random and guidance
IEC 60529:1989, Degrees of protection provided by enclosures (IP Code)
IEC 60529:1989/AMD1:1999
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral standard: Electromagnetic disturbances –
Requirements and tests
______________
There exists a consolidated edition 2.1(2001) including IEC 60529:1989 and its Amendment 1:1999.
There exists a consolidated edition 3.1(2012) including IEC 60601-1:2005 and its Amendment 1:2012.
– 10 – IEC 60601-1-12:2014 © IEC 2014
IEC 60601-1-6:2010, Medical electrical equipment – Part 1-6: General requirements for basic
safety and essential performance – Collateral standard: Usability
IEC 60601-1-6:2010/AMD1:2013
IEC 60601-1-8:2006, Medical electrical equipment – Part 1-8: General requirements for basic
safety and essential performance – Collateral standard: General requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-8:2006/AMD1:2012
IEC 60601-1-11:— , Medical electrical equipment – Part 1-11: General requirements for basic
safety and essential performance – Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment
CISPR 11:2009, Industrial, scientific and medical equipment – Radio-frequency disturbance
characteristics – Limits and methods of measurement
ISO 7000:2014, Graphical symbols for use on equipment – Registered symbols
ISO 7010:2011, Graphical symbols – Safety colours and safety signs – Registered safety
signs
Amendment 1:2012
Amendment 2:2012
Amendment 3:2012
Amendment 4:2013
Amendment 5:2014
ISO 15223-1:2012, Medical devices – Symbols to be used with medical device labels,
labelling and information to be supplied – Part 1: General requirements
EUROCAE ED-14G, Environmental conditions and test procedures for airborne equipment
RTCA DO-160G, Environmental Conditions and Test Procedures for Airborne Equipment
3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005 and
IEC 60601-1:2005/AMD1:2012, IEC 60601-1-6:2006 and IEC 60601-1-6:2006/AMD1:2013,
IEC 60601-1-8:2006 and IEC 60601-1-8:2006/AMD1:2012, IEC 60601-1-11:— and the
following definitions apply.
NOTE 1 Where the terms “voltage” and “current” are used in this document, they mean the r.m.s. values of an
alternating, direct or composite voltage or current unless stated otherwise.
NOTE 2 The term “electrical equipment” is used to mean ME EQUIPMENT or other electrical equipment. This
standard also uses the term “equipment” to mean ME EQUIPMENT or other electrical or non-electrical equipment in
the context of an ME SYSTEM.
NOTE 3 An index of defined terms used in this collateral standard is found beginning on page 48.
______________
There exists a consolidated edition 2.1(2012) including IEC 60601-1-8:2006 and Amendment 1:2012.
Second edition, to be published.
5 EUROCAE (European Organization for Civil Aviation Electronics), 102 rue Etienne Dolet, 92240 Malakoff,
France.
6 RTCA (Radio Technical Commission for Aeronautics), 1150 18th St, NW., Suite 910, Washington, DC 20036,
USA.
IEC 60601-1-12:2014 © IEC 2014 – 11 –
3.1
* EMS ENVIRONMENT
EMERGENCY MEDICAL SERVICES ENVIRONMENT
actual conditions and settings, in which OPERATORS interact with the ME EQUIPMENT or
ME SYSTEM, in and around the scene of an emergency outside of a professional healthcare
facility where a PATIENT can be given medical care, basic or advanced life support as well as
during professional transport to a professional healthcare facility or between professional
healthcare facilities
EXAMPLE 1 Responding to and providing life support at the scene of an emergency to a PATIENT reported as
experiencing injury or illness in a pre-hospital setting, and transporting the PATIENT, while continuing such life
support care, to an appropriate professional healthcare facility for further care.
EXAMPLE 2 Providing monitoring, treatment or diagnosis during transport between professional healthcare
facilities.
Note 1 to entry: For the purposes of this standard, use of equipment intended for the EMS ENVIRONMENT and
temporarily used in the HOME HEALTHCARE ENVIRONMENT by emergency medical personnel is considered use in the
EMS ENVIRONMENT.
Note 2 to entry: For the purposes of this standard, the OPERATORS of equipment intended for the
EMS ENVIRONMENT are presumed to be professional medical personnel or personnel with relevant specialized
training.
Note 3 to entry: Professional healthcare facilities include hospitals, physician offices, freestanding surgical
centres, dental offices, freestanding birthing centres, limited care facilities, first aid rooms or rescue rooms and
multiple treatment facilities.
Note 4 to entry: Emergency medical services are known by various names in different countries and regions.
Note 5 to entry: For the purposes of this standard, transport includes road, rotary and fixed-wing ambulances.
4 General requirements
4.1 * Additional requirements for SUPPLY MAINS for ME EQUIPMENT and ME SYSTEMS
For ME EQUIPMENT or ME SYSTEMS intended for the EMS ENVIRONMENT, the characteristics of the
SUPPLY MAINS specified in 4.10.2 of the general standard apply, with the following additions.
SUPPLY MAINS in the EMS ENVIRONMENT shall be assumed to have the following characteristics
of no voltage in excess of 110 % or lower than 85 % of the NOMINAL voltage between any of
the conductors of the system or between any of these conductors and earth.
The RATED range of NOMINAL voltage of the ME EQUIPMENT in the EMS ENVIRONMENT shall
include at least 12,4 V to 15,1 V for operation from 12 V d.c. SUPPLY MAINS and at least 24,8 V
to 30,3 V for operation from 24 V d.c. SUPPLY MAINS.
ME EQUIPMENT and ME SYSTEMS in the EMS ENVIRONMENT shall maintain BASIC SAFETY and
ESSENTIAL PERFORMANCE during and following a 30 s dip to 10 V from a 12 V d.c. SUPPLY MAINS
and during and following a 30 s dip to 20 V for operation from a 24 V d.c. SUPPLY MAINS.
For ME EQUIPMENT or ME SYSTEMS intended to be powered from an aircraft, the SUPPLY MAINS
shall comply with Section 16 of either EUROCAE ED-14G or RTCA DO-160G.
4.2 * Environmental conditions for ME EQUIPMENT
NOTE In IEC 60601-1:2005, the MANUFACTURER specifies the permissible environmental conditions of use,
including conditions for transport and storage in the technical description (see 7.9.3.1, second dash). These
conditions are referenced in requirements for testing throughout the general standard, (e.g. 5.3 and 11.1.1).
– 12 – IEC 60601-1-12:2014 © IEC 2014
4.2.1 * Environmental conditions of transport and storage between uses
The instructions for use shall indicate the permissible environmental conditions of transport
and storage of ME EQUIPMENT after the ME EQUIPMENT has been removed from its protective
packaging and subsequently between uses.
Unless otherwise indicated in the instructions for use, the ME EQUIPMENT shall comply with this
standard and shall remain operational in NORMAL USE within its specification after transport or
storage in the following environmental range:
– - 40 °C to + 5 °C without relative humidity control;
– + 5 °C to + 35 °C at a relative humidity up to 90 %, non-condensing;
– > 35 °C to 70 °C at a water vapour pressure up to 50 hPa;
after having been removed from its protective packaging and subsequently between uses.
NOTE 1 This represents class 7K4, as described in IEC TR 60721-4-7:2001 [6] .
If the instructions for use state a more restricted range of environmental transport and storage
conditions between uses, these environmental conditions shall be:
– justified in the RISK MANAGEMENT FILE;
– marked on the ME EQUIPMENT, unless such marking is not practicable, in which case the
more restricted range need only be disclosed in the instructions for use; and
– marked on the carrying case, if the instructions for use indicate that the ME EQUIPMENT is
intended to be transported or stored in a carrying case between uses.
Symbols 5.3.5 (ISO 7000-0534), 5.3.6 (ISO 7000-0533) or 5.3.7 (ISO 7000-0632) of
ISO 15223-1:2012 may be used to mark the temperature range (see Table C.1, symbols 2, 3
and 4). Symbol 5.3.8 (ISO 7000-2620) of ISO 15223-1:2012 may be used to mark the
humidity range (see Table C.1, symbol 5) and symbol 5.3.9 (ISO 7000-2621) of
ISO 15223-1:2012 may be used to mark the atmospheric pressure range (see Table C.1,
symbol 6). Where ME EQUIPMENT has different markings for conditions of transport and storage
between uses, continuous operating conditions (see 4.2.2.1) and transient operating
conditions (see 4.2.2.2), those markings shall be accompanied by supplementary marking
(e.g. appropriate wording) except where the respective applicability would be obvious (e.g.
limits for transport and storage between uses on the carrying case and limits for operation on
the ME EQUIPMENT itself).
Compliance is checked by the following test and, when a more restricted range is stated in the
instructions for use, inspection of the RISK MANAGEMENT FILE.
a) Prepare the ME EQUIPMENT for transport or storage according to instructions for use.
EXAMPLES Removal of batteries, emptying fluid reservoirs
b) Expose the ME EQUIPMENT at its lowest specified environmental transport and storage
conditions (temperature °C) for:
−4
− at least 16 h; or
− ensure that the ME EQUIPMENT reaches THERMAL STABILITY for at least 2 h.
c) Then expose the ME EQUIPMENT to 34 °C ± 4 °C and 93 % ± 3 % relative humidity until the
test chamber reaches equilibrium. The transition from low to high conditions should be
made slowly enough to provide a non-condensing environment. Hold for at least 2 h.
______________
Numbers in square brackets refer to the Bibliography.
IEC 60601-1-12:2014 © IEC 2014 – 13 –
d) Then expose the ME EQUIPMENT at its highest specified environmental transport and
storage conditions, but not requiring a water vapour partial pressure greater than 50 hPa,
+4
(temperature °C) for:
− at least 16 h; or
− ensure that the ME EQUIPMENT reaches THERMAL STABILITY for at least 2 h.
NOTE 2 The intent of specifying a minimum duration of the exposure to both the low and high temp
...








Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...