SIST EN 62359:2007
(Main)Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
This International Standard is applicable to medical diagnostic ultrasound fields. This standard establishes - parameters related to thermal and non-thermal aspects of diagnostic ultrasonic fields; - methods for the determination of an exposure parameter relating to temperature rise in theoretical tissue-equivalent models, resulting from absorption of ultrasound; - methods for the determination of an exposure parameter appropriate to certain non-thermal effects.
Ultraschall - Charakterisierung von Feldern - Prüfverfahren für die Ermittlung des thermischen und des mechanischen Indexes bezogen auf medizinisch-diagnostische Ultraschallfelder
Ultrasons - Caractérisation du champ - Essais pour la détermination des indices d'échauffement et mécaniques des champs d'ultrasons utilisés pour le diagnostic médical
Is applicable to medical diagnostic ultrasound fields. This standard establishes - parameters related to thermal and non-thermal aspects of diagnostic ultrasonic fields; - methods for the determination of an exposure parameter relating to temperature rise in theoretical tissue-equivalent models, resulting from absorption of ultrasound; - methods for the determination of an exposure parameter appropriate to certain non-thermal effects.
Ultrazvok - Karakterizacija polj - Preskusne metode za ugotavljanje termičnih in mehanskih znakov glede medicinskih diagnostičnih ultrazvočnih polj (IEC 62359:2005)
General Information
Relations
Standards Content (Sample)
6,67(16/29(16.,MDQXDU67$1'$5'LVWRYHWHQ(18OWUD]YRN.DUDNWHUL]DFLMDSROM3UHVNXVQHPHWRGH]DXJRWDYOMDQMHWHUPLþQLKLQPHKDQVNLK]QDNRYJOHGHPHGLFLQVNLKGLDJQRVWLþQLKXOWUD]YRþQLKSROM,(&8OWUDVRQLFV)LHOGFKDUDFWHUL]DWLRQ7HVWPHWKRGVIRUWKHGHWHUPLQDWLRQRIWKHUPDODQGPHFKDQLFDOLQGLFHVUHODWHGWRPHGLFDOGLDJQRVWLFXOWUDVRQLFILHOGV,(& !"#$%&'( )&!*- . 5HIHUHQþQDãWHYLOND6,67(1HQ,&6
EUROPEAN STANDARD
EN 62359 NORME EUROPÉENNE EUROPÄISCHE NORM
October 2005 CENELEC European Committee for Electrotechnical Standardization Comité Européen de Normalisation Electrotechnique Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: rue de Stassart 35, B - 1050 Brussels
© 2005 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 62359:2005 E
ICS 17.140.50
English version
Ultrasonics –
Field characterization –
Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields (IEC 62359:2005)
Ultrasons –
Caractérisation du champ –
Essais pour la détermination des indices d'échauffement et mécaniques
des champs d'ultrasons utilisés
pour le diagnostic médical (CEI 62359:2005)
Ultraschall –
Charakterisierung von Feldern - Prüfverfahren für die Ermittlung des thermischen und des mechanischen Indexes bezogen auf medizinisch-diagnostische Ultraschallfelder (IEC 62359:2005)
This European Standard was approved by CENELEC on 2005-09-13. CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the Central Secretariat or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to the Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EN 62359:2005 - 2 - Foreword The text of document 87/300/FDIS, future edition 1 of IEC 62359, prepared by IEC TC 87, Ultrasonics, was submitted to the IEC-CENELEC parallel vote and was approved by CENELEC as EN 62359 on 2005-09-13. The following dates were fixed: – latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement
(dop)
2006-07-01 – latest date by which the national standards conflicting
with the EN have to be withdrawn
(dow)
2008-10-01 Annex ZA has been added by CENELEC. __________ Endorsement notice The text of the International Standard IEC 62359:2005 was approved by CENELEC as a European Standard without any modification. __________
- 3 - EN 62359:2005
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. NOTE Where an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD applies. Publication Year Title EN/HD Year IEC 60601-2-37 - 1) Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment
EN 60601-2-37 2001 2) IEC 61102 1991 Measurement and characterisation of ultrasonic fields using hydrophones in the frequency range 0,5 MHz to 15 MHz
EN 61102 1993 IEC 61157 1992 Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment
EN 61157 1994 IEC 61161 1992 Ultrasonic power measurement in liquids in the frequency range 0,5 MHz to
25 MHz EN 61161 1994 A1 1998
A1 1998
1) Undated reference. 2) Valid edition at date of issue.
INTERNATIONAL STANDARD IEC62359 First edition2005-04 Ultrasonics – Field characterization – Test methods for the determination of
thermal and mechanical indices related
to medical diagnostic ultrasonic fields
IEC 2005
Copyright - all rights reserved No part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from the publisher. International Electrotechnical Commission,
3, rue de Varembé, PO Box 131, CH-1211 Geneva 20, SwitzerlandTelephone: +41 22 919 02 11 Telefax: +41 22 919 03 00 E-mail: inmail@iec.ch
Web: www.iec.ch X For price, see current cataloguePRICE CODE
Commission Electrotechnique InternationaleInternational Electrotechnical Commission
– 2 – 62359 IEC:2005(E) CONTENTS FOREWORD.3 INTRODUCTION.5
1 Scope.6 2 Normative references.6 3 Terms and definitions.6 4 List of symbols.15 5 Test methods for determining the mechanical index and the thermal index.16 5.1 General.16 5.2 Determination of mechanical index.17 5.3 Determination of thermal index – general.17 5.4 Determination of thermal index in non-scanning mode.17 5.5 Determination of thermal index in scanning mode.19 5.6 Calculations for combined-operating mode.19 5.7 Summary of measured quantities for index determination.20
Annex A (informative)
Relationships with other standards.22 Annex B (informative)
Guidance notes for measurement of output power
in scanning mode.23 Annex C (informative)
Rationale and derivation of index models.27 Annex D (informative)
Guidance on the interpretation of TI and MI.40
Bibliography.41
Figure B.1 – Suggested 1 cm-wide aperture mask.25 Figure B.2 – Suggested orientation of transducer, mask slit and RFB target.25 Figure B.3 – Suggested orientation of transducer and 1 cm RFB target.25 Figure C.1 – Focused transducer with a large aperture.35 Figure C.2 – Focused transducer with smaller aperture (≥1 cm2).35 Figure C.3 – Focused transducer with a weak focus (Aeq > 1 cm2).36 Figure C.4 – Weakly focused transducer.36
Table 1 – Summary of combination formulae for each of the THERMAL INDEX categories.20 Table 2 – Summary of the acoustic quantities required for the determination of the indices.21 Table C.1 – THERMAL INDEX categories and models.29 Table C.2 – Thermal index formulae.33
62359 IEC:2005(E) – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION ____________
ULTRASONICS –
FIELD CHARACTERIZATION –
TEST METHODS FOR THE DETERMINATION OF THERMAL
AND MECHANICAL INDICES RELATED TO
MEDICAL DIAGNOSTIC ULTRASONIC FIELDS
FOREWORD 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with may participate in this preparatory work. International, governmental and non-governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for Standardization (ISO) in accordance with conditions determined by agreement between the two organizations. 2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international consensus of opinion on the relevant subjects since each technical committee has representation from all interested IEC National Committees.
3) IEC Publications have the form of requirements for international use and are accepted by IEC National Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any misinterpretation by any end user. 4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications transparently to the maximum extent possible in their national and regional publications. Any divergence between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter. 5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any equipment declared to be in conformity with an IEC Publication. 6) All users should ensure that they have the latest edition of this publication. 7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and members of its technical committees and IEC National Committees for any personal injury, property damage or other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is indispensable for the correct application of this publication. 9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of patent rights. IEC shall not be held responsible for identifying any or all such patent rights. International Standard IEC 62359 has been prepared by IEC technical committee 87: Ultrasonics The text of this standard is based on the following documents: FDIS Report on voting 87/300/FDIS 87/305/RVD
Full information on the voting for the approval of this standard can be found in the report on voting indicated in the above table. This publication has been drafted in accordance with the ISO/IEC Directives, Part 2. This standard may be used to support the requirements of IEC 60601-2-37.
– 4 – 62359 IEC:2005(E) The committee has decided that the contents of this publication will remain unchanged until the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in the data related to the specific publication. At this date, the publication will be
• reconfirmed; • withdrawn; • replaced by a revised edition, or • amended.
A bilingual version of this standard may be issued at a later date.
62359 IEC:2005(E) – 5 –
INTRODUCTION Medical diagnostic ultrasonic equipment is widely used in clinical practice for imaging and monitoring purposes. Equipment normally operates at frequencies in the low megahertz frequency range and comprises an ultrasonic transducer acoustically coupled to the patient and associated electronics. There is an extremely wide range of different types of systems in current clinical practice.
The ultrasound entering the patient interacts with the patient's tissue and this interaction can be considered in terms of both thermal and non-thermal effects. The purpose of this International Standard is to specify methods of determining thermal and non-thermal exposure indices which can be used to help in assessing the hazard caused by exposure to a particular ultrasonic field used for medical diagnosis or monitoring. It is recognised that these indices have limitations and a knowledge of the indices at the time of an examination is not sufficient in itself to make an informed clinical risk assessment. It is intended that these limitations will be addressed in future revisions of this standard and as scientific understanding increases. Under certain conditions specified in IEC 60601-2-37 these indices are displayed on medical ultrasonic equipment intended for these purposes.
– 6 – 62359 IEC:2005(E) ULTRASONICS –
FIELD CHARACTERIZATION –
TEST METHODS FOR THE DETERMINATION OF THERMAL
AND MECHANICAL INDICES RELATED TO
MEDICAL DIAGNOSTIC ULTRASONIC FIELDS
1 Scope This International Standard is applicable to medical diagnostic ultrasound fields. This standard establishes – parameters related to thermal and non-thermal aspects of diagnostic ultrasonic fields; – methods for the determination of an exposure parameter relating to temperature rise in theoretical tissue-equivalent models, resulting from absorption of ultrasound; – methods for the determination of an exposure parameter appropriate to certain non-thermal effects. NOTE In this standard where multiples or submultiples of SI units are used this is clearly stated and the usage is self-consistent. 2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. IEC 60601-2-37, Medical electrical equipment – Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment IEC 61102:1991, Measurement and characterisation of ultrasonic fields using hydrophones in the frequency range 0,5 MHz to 15 MHz IEC 61157:1992, Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment IEC 61161:1992, Ultrasonic power measurement in liquids in the frequency range 0,5 MHz to 25 MHz 1)
Amendment 1 (1998) 3 Terms and definitions
For the purposes of this International standard, the terms and definitions given in IEC 61102:1991, IEC 61157:1992 and IEC 61161:1998 (several of which are repeated below for convenience) and the following apply. 3.1
acoustic attenuation coefficient coefficient intended to account for ultrasonic attenuation of tissue between the source and a specified point Symbol: α Unit: decibels per centimetre per megahertz, dB cm–1 MHz–1 ——————— 1) A consolidated edition (1.1) exists, including IEC 61161:1992 and its Amendment 1 (1998).
62359 IEC:2005(E) – 7 –
3.2
acoustic working frequency arithmetic mean of the most widely separated frequencies f1 and f2 at which the amplitude of the pressure spectrum of the acoustic signal is 3 dB lower than the peak amplitude [3.4.2 of IEC 61102:1991, modified] Symbol: fawf Unit: megahertz, MHz 3.3
attenuated output power value of the acoustic output power after attenuation and at a specified distance from the transducer, and given by dB) /10-(awf10 fz PPαα= where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest; fawf is the acoustic working frequency; P is the output power measured in water.
Symbol: Pα Unit: milliwatts, mW 3.4
attenuated peak-rarefactional acoustic pressure value of the peak-rarefactional acoustic pressure after attenuation and at a specified point, and given by dB)/20(-rr,awf10 )()(fz .zpzpα= where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest; fawf is the acoustic working frequency; pr(z) is the peak-rarefactional acoustic pressure measured in water. Symbol: pr,α Unit: megapascals, MPa 3.5
attenuated pulse-average intensity value of the acoustic pulse-average intensity after attenuation and at a specified point, and given by dB)/10-(papa,awf10 )(fz zIIαα=
– 8 – 62359 IEC:2005(E) where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest; fawf is the acoustic working frequency; Ipa(z) is the pulse-average intensity measured in water. Symbol: Ipa,α Unit: watts per centimetre squared, W cm–2 3.6
attenuated pulse-intensity integral value of the pulse-intensity integral after attenuation and at a specified point, and given by dB)/10-(pipi,awf10 fz IIαα= where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest; fawf is the acoustic working frequency; Ipi is the pulse-intensity integral measured in water. Symbol: Ipi,α Unit: millijoules per centimetre squared, mJ cm–2 3.7
attenuated spatial-peak temporal-average intensity value of the spatial-peak temporal-average intensity after attenuation and at a specified distance z, and given by dB) /10-(zptazpta,awf10 )()(fz zIzIαα= where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest; fawf is the acoustic working frequency; Izpta(z) is the spatial-peak temporal-average intensity, at a specified distance z measured in water. Symbol: Izpta,α(z) Unit: milliwatts per centimetre squared, mW cm–2 3.8
attenuated temporal-average intensity value of the temporal-average intensity after attenuation and at a specified point, and given by dB)/10(-tata,awf10 )()(fz .zIzIα= where α is the acoustic attenuation coefficient; z is the distance from the source to the point of interest;
62359 IEC:2005(E) – 9 –
fawf is the acoustic working frequency; Ita(z) is the temporal-average intensity measured in water. Symbol: Ita,α(z) Unit: milliwatts per centimetre squared, mW cm–2 3.9
beam area area in a specified plane perpendicular to the beam-alignment axis consisting of all points at which the pulse-intensity integral is greater than a specified fraction of the maximum pulse-intensity integral in that plane [3.6 of IEC 61102:1991, modified] NOTE For measurement purposes the pulse intensity integral can be taken as being proportional to the pulse pressure-squared integral 3.10
beam alignment axis straight line joining the points of maximum pulse intensity integral measured at several different distances in the far field. For the purposes of alignment, this line may be projected to the face of the ultrasonic transducer [3.5 of IEC 61102:1991, modified] 3.11
bone thermal index thermal index for applications, such as foetal (second and third trimester) or neonatal cephalic (through the fontanelle), in which the ultrasound beam passes through soft tissue and a focal region is in the immediate vicinity of bone Symbol: TIB Unit: None NOTE See 5.4.2 and 5.5.2 for methods of determining the bone thermal index. 3.12
bounded output power output power emitted in scanning mode from a region of the active area of the transducer whose width in the scan plane is limited to 1 cm Symbol: P1
Unit: milliwatts, mW 3.13
break-point depth value equal to 1,5 times the equivalent aperture diameter, and given by zbp = 1,5 Deq where Deq is the equivalent aperture diameter. Symbol: zbp Unit: centimetres, cm
– 10 – 62359 IEC:2005(E) 3.14
combined-operating mode mode of operation of an equipment which combines more than one discrete-operating mode [3.6 of IEC 61157:1992, modified] 3.15
cranial-bone thermal index thermal index for applications, such as paediatric and adult cranial applications, in which the ultrasound beam passes through bone near the beam entrance into the body Symbol: TIC Unit: None NOTE See 5.4.3 and 5.5.3 for methods of determining the cranial bone thermal index. 3.16
default setting specific state of control, the ultrasonic diagnostic equipment will enter upon power-up, new patient select or change from non-foetal to foetal applications 3.17
depth for bone thermal index distance from the plane where the –12 dB output beam dimensions are determined along the beam alignment axis to the plane where the product of attenuated output power and attenuated pulse-intensity integral is maximum Symbol: zb
Unit: centimetres, cm 3.18
depth for soft-tissue thermal index distance from the plane where the –12 dB output beam dimensions are determined along the beam alignment axis to the plane at which the lower value of the attenuated output power and the product of the attenuated spatial-peak temporal-average intensity and 1 cm2 is maximized over the distance range equal to, or more than, 1,5 times the equivalent aperture diameter Symbol: zs
Unit: centimetres, cm NOTE In this standard, the restricted definition of spatial-peak temporal-average intensity from 3.49 of IEC 61102:1991 relating to a specified plane is used where spatial-peak temporal-average intensity is replaced by attenuated spatial-peak temporal-average intensity. 3.19
discrete-operating mode mode of operation of ultrasonic diagnostic equipment in which the purpose of the excitation of the ultrasonic transducer or ultrasonic transducer element group is to utilize only one diagnostic methodology [3.7 of IEC 61157:1992]
62359 IEC:2005(E) – 11 –
3.20
equivalent aperture diameter diameter of a circle whose area is the –12 dB output beam area and given by
4aprteqADπ≡ where Aaprt is the –12 dB output beam area. Symbol: Deq Unit: centimetres, cm NOTE This formula gives the diameter of a circle whose area is the –12 dB output beam area. It is used in the calculation of the cranial-bone thermal index and the soft tissue thermal index. 3.21
equivalent beam area value of the area of the acoustic beam at the distance z in terms of power and intensity, and given by ())()(zptazpta,zIPzIzP)z(Aeq=≡αα where Pα(z) is the attenuated output power, at the distance z; Izpta,α(z) is the attenuated spatial-peak temporal-average intensity, at the distance z; P is the output power; Izpta(z)
is the spatial-peak temporal-average intensity, at the distance z; and z is the distance from the source to the specified point. Symbol: Aeq(z) Unit: centimetres squared, cm2 3.22
equivalent beam diameter value of the diameter of the acoustic beam at the distance z in terms of the equivalent beam area, and given by
)(4)(eqeqzAzdπ= where Aeq(z) is the equivalent beam area; z is the distance from the source to the specified point. Symbol: deq(z) Unit: centimetres, cm
– 12 – 62359 IEC:2005(E) 3.23
mechanical index mechanical index is given by
1/2awf,rMICfpMI−=α
where CMI = 1 MPa MHz–1/2;
pr,α is the attenuated peak-rarefactional acoustic pressure; fawf is the acoustic-working frequency.
Symbol: MI Unit: None 3.24
non-scanning mode mode of operation of ultrasonic diagnostic equipment that involves a sequence of ultra-sonic pulses which give rise to ultrasonic scan lines that follow the same acoustic path [3.12 of IEC 61157:1992, modified] 3.25
–12 dB output beam area area of the ultrasonic beam derived from the –12 dB output beam dimensions [3.13 of IEC 61157:1992, modified] Symbol: Aaprt Unit: centimetre squared, cm2 3.26
–12 dB output beam dimensions
dimensions of the ultrasonic beam (–12 dB pulse beam width) in specified directions normal to the beam alignment axis and at the transducer output face [3.14 of IEC 61157:1992, modified] NOTE 1 For reasons of measurement accuracy, the –12 dB output beam dimensions can be derived from measurements at a distance chosen to be as close as possible to the face of the transducer, and if possible no more than 1 mm from the face. NOTE 2 For contact transducers, these dimensions can be taken as the dimensions of the radiating element. Symbol: X, Y Unit: centimetres, cm 3.27
output power time-average ultrasonic power radiated by an ultrasonic transducer into an approximately free field under specified conditions in a specified medium, preferably water [3.5 of IEC 61161:1998] Symbol: P Unit: milliwatts, mW
62359 IEC:2005(E) – 13 –
3.28
peak-rarefactional acoustic pressure maximum of the modulus of the negative instantaneous acoustic pressure in an acoustic field during an acoustic repetition period [3.34 of IEC 61157:1992, modified] Symbol: pr Unit: megapascals, MPa 3.29
prudent-use statement affirmation of the principle advising avoidance of primarily high exposure levels and secondarily long exposure times while acquiring necessary clinical information NOTE See Bibliography [1, 2, 3, 4, 5 ]2) 3.30
pulse beam-width distance between two points, on a specified surface in a specified direction passing through the point of maximum pulse-pressure-squared integral (pi) in that surface, at which the pulse-pressure-squared integral is a specified fraction of the maximum value in that surface [3.18 of IEC 61157:1992, modified] Symbol: d–6 (for pulse beam-width defined at –6dB) Unit: centimetres, cm 3.31
pulse duration 1,25 times the interval between the time when the time integral of intensity in an acoustic pulse at a point reaches 10 % and when it reaches 90 % of the pulse intensity integral [3.30 of IEC 61102:1991, modified] Symbol: td Unit: seconds, s 3.32
pulse-intensity integral time integral of the instantaneous intensity at a particular point in an acoustic field integrated over the acoustic pulse waveform [3.31 of IEC 61102:1991] Symbol: Ipi Unit: millijoules per centimetre squared, mJ cm–2 3.33
pulse-pressure-squared integral time integral of the square of the instantaneous acoustic pressure at a particular point in an acoustic field integrated over the acoustic pulse waveform [3.33 of IEC 61102:1991] ——————— 2) Figures in square brackets refer to the Bibliography.
– 14 – 62359 IEC:2005(E) Symbol: pi Unit: Pascal squared seconds, Pa2s 3.34
pulse repetition rate inverse of the time interval between two successive acoustic pulses [3.35 of IEC 61102:1991, modified] Symbol: prr Unit: hertz, Hz 3.35
scan line ultrasonic scan line for automatic scanning systems, the beam alignment axis either for a particular ultrasonic transducer element group, or for a single or multiple excitation of an ultrasonic transducer or of an ultrasonic transducer element group. [3.27 of IEC 61157:1992] 3.36
scanning mode mode of operation of an ultrasonic diagnostic equipment that involves a sequence of ultrasonic pulses which give rise to scan lines that do not follow the same acoustic path [3.21 of IEC 61157:1992, modified] 3.37
soft tissue thermal index thermal index related to soft tissues
Symbol: TIS Unit: None NOTE 1 See 5.4.1 and 5.5.1 and the following for methods of determination of the soft-tissue thermal index. NOTE 2 For the purposes of this document, soft tissue includes all body tissues and fluids but excludes skeletal tissues. 3.38
spatial-peak temporal-average intensity maximum value of the temporal-average intensity in a specified plane at a specified distance z from the transducer
[3.49 of IEC 61102:1991, modified] Symbol: Izpta(z) Unit: milliwatts per centimetre squared, mW cm–2 NOTE In this standard the restricted definition from 3.49 of IEC 61102 relating to a specified plane is used. 3.39
system medical diagnostic ultrasonic equipment combination of the ultrasound instrument console and the transducer assembly making up a complete system [3.11 of IEC 61157:1992]
62359 IEC:2005(E) – 15 –
3.40
temporal-average intensity time-average of the instantaneous intensity at a particular point in an acoustic field [3.53 of IEC 61102:1991, modified] Symbol: Ita(z) Unit: milliwatts per centimetre squared, mW cm–2 3.41
thermal index ratio of attenuated acoustic power at a specified point to the attenuated acoustic power required to raise the temperature at that point in a specific tissue model by 1 °C
Symbol: TI Unit: None 3.42
transducer assembly transducer housing (probe), any associated electronic circuitry and any liquids contained in the housing and the integral cable which connects the transducer probe to an ultrasound console [see 3.22 of IEC 61157:1992] 3.43
transmit pattern combination of a specific set of transducer beam-forming characteristics (determined by the transmit aperture size, apodization shape and relative timing/phase delay pattern across the aperture, resulting in a specific focal length and direction), and an electrical drive waveform of a specific fixed shape but variable amplitude 3.44
ultrasonic diagnostic equipment medical electrical equipment which is intended for in vivo ultrasonic
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.