ISO/TS 21726:2019
(Main)Biological evaluation of medical devices - Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents
Biological evaluation of medical devices - Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents
This document describes the basis for, selection of, and general applicability of a threshold of toxicological concern (TTC) value for a constituent present in/on a medical device or released from a medical device. The TTC values in this document can be used for: - comparing to a maximum concentration of an identified or unidentified constituent in an extract (see ISO 10993-18); - supporting toxicological equivalence; - comparing to a maximum exposure dose estimate of an identified constituent (see ISO 10993-17). NOTE Constituent is defined in 3.1. ISO 10993-18 specifies how to convert TTC (µg/d) into a concentration (µg/ml). TTC is not applicable to constituents with adequate toxicity data for deriving a tolerable intake (TI) value (see ISO 10993-17). The TTC values established in this document are protective for carcinogens, systemic toxicants, and reproductive toxicants (see Clause 5). This document does not include TTC values for other biological endpoints assessed as part of the biological evaluation of a medical device, per ISO 10993-1, for example: - cytotoxicity; - irritation; - sensitization; - hemocompatibility; - material mediated pyrogenicity; - local effects that occur in tissues at the site of contact between a medical device and the body (e.g. the observations from implantation studies). The TTC values in this document do not apply to potential exposure via gas pathways of medical devices. For application of TTC for constituents present/released from these devices, see the ISO 18562 series. The TTC values presented in this document are not applicable for the safety assessment of cohort of concern (see 5.3).
Évaluation biologique des dispositifs médicaux — Application du seuil de préoccupation toxicologique (TTC) pour évaluer la biocompatibilité des substances extractibles des dispositifs médicaux
General Information
- Status
- Published
- Publication Date
- 31-Jan-2019
- Technical Committee
- ISO/TC 194 - Biological and clinical evaluation of medical devices
- Drafting Committee
- ISO/TC 194/WG 11 - Allowable limits for leachable substances
- Current Stage
- 9092 - International Standard to be revised
- Start Date
- 10-Oct-2025
- Completion Date
- 13-Dec-2025
Overview
ISO/TS 21726:2019 provides guidance on applying the Threshold of Toxicological Concern (TTC) to the biological evaluation and biocompatibility assessment of medical device constituents. The technical specification explains the basis, selection and general applicability of TTC values for constituents present in, on, or released from a medical device. TTC in this document is intended to be protective for carcinogens, systemic toxicants, and reproductive toxicants, and is used when specific toxicity data are not available to derive a tolerable intake (TI).
Key topics and technical requirements
- TTC values and selection: Recommended TTC intake limits (µg/day) are provided according to device contact duration. Typical values include 120 µg/d, 20 µg/d, 10 µg/d, and 1.5 µg/d for increasing exposure durations and long-term contact categories.
- Device contact categories: Selection of the appropriate TTC is based on medical device contact duration (limited, prolonged, long‑term) and must be justified by use conditions.
- Applicability and exclusions:
- TTC applies to individual constituents (identified or unidentified) in extracts or leachables for comparison against concentration or estimated exposure.
- Not applicable where adequate toxicity data exist (see ISO 10993-17).
- Not appropriate for cohorts of concern (highly potent toxicants) such as aflatoxin‑like compounds, N‑nitroso compounds, polyhalogenated dioxins/furans, heavy metals, steroids, organophosphorus compounds, strained heterocycles, etc.
- Excluded substance types: high molecular weight polymers, particles (including nanoparticles), proteins, ceramics, radioactive constituents.
- Mixtures and interactions: TTC values can be applied to individual constituents within complex mixtures; the specification states additional aggregate adjustments generally are not required at very low exposure levels.
- Supporting evidence: Use of negative mutagenicity data or orthogonal in silico models may justify assignment to a Cramer class and selection of the corresponding TTC.
Applications and users
ISO/TS 21726 is practical for:
- Medical device manufacturers conducting biocompatibility and extractables & leachables (E&L) risk assessments.
- Toxicologists and regulatory affairs specialists preparing safety justifications or equivalence arguments.
- Contract testing laboratories interpreting analytical extract data against toxicological thresholds.
- Use cases include:
- Comparing analytical extract concentrations to TTC (per ISO 10993-18 conversion guidance).
- Supporting toxicological equivalence between materials.
- Comparing TTC to estimated exposure doses (per ISO 10993-17) when specific TI values are not available.
Related standards
- ISO 10993-1: Biological evaluation framework
- ISO 10993-17: Establishment of allowable limits for leachable substances
- ISO 10993-18: Chemical characterization and converting TTC (µg/d) to concentration (µg/mL)
- ISO 18562 series: TTC considerations for gas pathway exposures
- ICH M7(R1): Basis for mutagenic impurity TTC values
Keywords: ISO/TS 21726, Threshold of Toxicological Concern, TTC, biocompatibility, medical devices, extractables and leachables, ISO 10993, cohort of concern, toxicological risk assessment.
Frequently Asked Questions
ISO/TS 21726:2019 is a technical specification published by the International Organization for Standardization (ISO). Its full title is "Biological evaluation of medical devices - Application of the threshold of toxicological concern (TTC) for assessing biocompatibility of medical device constituents". This standard covers: This document describes the basis for, selection of, and general applicability of a threshold of toxicological concern (TTC) value for a constituent present in/on a medical device or released from a medical device. The TTC values in this document can be used for: - comparing to a maximum concentration of an identified or unidentified constituent in an extract (see ISO 10993-18); - supporting toxicological equivalence; - comparing to a maximum exposure dose estimate of an identified constituent (see ISO 10993-17). NOTE Constituent is defined in 3.1. ISO 10993-18 specifies how to convert TTC (µg/d) into a concentration (µg/ml). TTC is not applicable to constituents with adequate toxicity data for deriving a tolerable intake (TI) value (see ISO 10993-17). The TTC values established in this document are protective for carcinogens, systemic toxicants, and reproductive toxicants (see Clause 5). This document does not include TTC values for other biological endpoints assessed as part of the biological evaluation of a medical device, per ISO 10993-1, for example: - cytotoxicity; - irritation; - sensitization; - hemocompatibility; - material mediated pyrogenicity; - local effects that occur in tissues at the site of contact between a medical device and the body (e.g. the observations from implantation studies). The TTC values in this document do not apply to potential exposure via gas pathways of medical devices. For application of TTC for constituents present/released from these devices, see the ISO 18562 series. The TTC values presented in this document are not applicable for the safety assessment of cohort of concern (see 5.3).
This document describes the basis for, selection of, and general applicability of a threshold of toxicological concern (TTC) value for a constituent present in/on a medical device or released from a medical device. The TTC values in this document can be used for: - comparing to a maximum concentration of an identified or unidentified constituent in an extract (see ISO 10993-18); - supporting toxicological equivalence; - comparing to a maximum exposure dose estimate of an identified constituent (see ISO 10993-17). NOTE Constituent is defined in 3.1. ISO 10993-18 specifies how to convert TTC (µg/d) into a concentration (µg/ml). TTC is not applicable to constituents with adequate toxicity data for deriving a tolerable intake (TI) value (see ISO 10993-17). The TTC values established in this document are protective for carcinogens, systemic toxicants, and reproductive toxicants (see Clause 5). This document does not include TTC values for other biological endpoints assessed as part of the biological evaluation of a medical device, per ISO 10993-1, for example: - cytotoxicity; - irritation; - sensitization; - hemocompatibility; - material mediated pyrogenicity; - local effects that occur in tissues at the site of contact between a medical device and the body (e.g. the observations from implantation studies). The TTC values in this document do not apply to potential exposure via gas pathways of medical devices. For application of TTC for constituents present/released from these devices, see the ISO 18562 series. The TTC values presented in this document are not applicable for the safety assessment of cohort of concern (see 5.3).
ISO/TS 21726:2019 is classified under the following ICS (International Classification for Standards) categories: 11.100.20 - Biological evaluation of medical devices. The ICS classification helps identify the subject area and facilitates finding related standards.
You can purchase ISO/TS 21726:2019 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of ISO standards.
Standards Content (Sample)
TECHNICAL ISO/TS
SPECIFICATION 21726
First edition
2019-02
Biological evaluation of medical
devices — Application of the threshold
of toxicological concern (TTC) for
assessing biocompatibility of medical
device constituents
Évaluation biologique des dispositifs médicaux — Application
du seuil de préoccupation toxicologique (TTC) pour évaluer la
biocompatibilité des substances extractibles des dispositifs médicaux
Reference number
©
ISO 2019
© ISO 2019
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Fax: +41 22 749 09 47
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2019 – All rights reserved
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 2
4 Background . 2
4.1 General . 2
4.2 Protectiveness of TTC values . 3
5 Applicability of TTC to medical device constituents . 3
5.1 General . 3
5.2 Selection of TTC value based on duration of body contact . 3
5.3 Cohort of concern constituents . 4
5.3.1 General. 4
5.3.2 Identification of cohort of concern constituents . 5
5.4 Applicability to mixtures . 5
Bibliography . 6
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT) see www .iso
.org/iso/foreword .html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of
medical devices.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/members .html.
iv © ISO 2019 – All rights reserved
TECHNICAL SPECIFICATION ISO/TS 21726:2019(E)
Biological evaluation of medical devices — Application of
the threshold of toxicological concern (TTC) for assessing
biocompatibility of medical device constituents
1 Scope
This document describes the basis for, selection of, and general applicability of a threshold of
toxicological concern (TTC) value for a constituent present in/on a medical device or released from a
medical device. The TTC values in this document can be used for:
— comparing to a maximum concentration of an identified or unidentified constituent in an extract
(see ISO 10993-18);
— supporting toxicological equivalence;
— comparing to a maximum exposure dose estimate of an identified constituent (see ISO 10993-17).
NOTE Constituent is defined in 3.1.
ISO 10993-18 specifies how to convert TTC (µg/d) into a concentration (µg/ml).
TTC is not applicable to constituents with adequate toxicity data for deriving a tolerable intake (TI)
value (see ISO 10993-17).
The TTC values established in this document are protective for carcinogens, systemic toxicants, and
reproductive toxicants (see Clause 5). This document does not include TTC values for other biological
endpoints assessed as part of the biological evaluation of a medical device, per ISO 10993-1, for example:
— cytotoxicity;
— irritation;
— sensitization;
— hemocompatibility;
— material mediated pyrogenicity;
— local effects that occur in tissues at the site of contact between a medical device and the body (e.g.
the observations from implantation studies).
The TTC values in this document do not apply to potential exposure via gas pathways of medical devices.
For application of TTC for constituents present/released from these devices, see the ISO 18562 series.
The TTC values presented in this document are not applicable for the safety assessment of cohort of
concern (see 5.3).
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 10993-1:2018, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management process
ISO 10993-17, Biological evaluation of medical devices — Part 17: Establishment of allowable limits for
leachable substances
ISO 10993-18, Biological evaluation of medical devices — Part 18: Chemical characterization of materials
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https: //www .iso .org/obp
— IEC Electropedia: available at http: //www .electropedia .org/
3.1
constituent
chemical or compound present in or on a finished medical device or material(s) of construction
Note 1 to entry: Constituents may be intentionally present (e.g. an additive such as an antioxidant) or
unintentionally present (e.g. an impurity).
Note 2 to entry: When applying TTC to an extractable or leachable, the identity of the extractable/leachable
represents a constituent to which individual(s) are potentially exposed due to medical device use.
3.2
extractable
constituent released when the medical device or material of construction is extracted using laboratory
extraction conditions and vehicles
Note 1 to entry: When applying TTC to an extractable, the extracted amount is assumed to potentially contact
the individual(s) to whom the medical device contacts during clinical use, see ISO 10993-17.
3.3
identified constituent
constituent assigned a full chemical structure
3.4
leachable
constituent released from a medical device and potentially contacts the individual(s) during its
clinical use
Note 1 to entry: When applying TTC to a leachable, the leached amount is assumed to potentially contact the
user(s) of the medical device during its clinical use, see ISO 10993-17.
3.5
threshold of toxicological concern
TTC
[1]
level of exposure for constituents, below which there would be no appreciable risk to human health
4 Background
4.1 General
The Threshold of Toxicological Concern (TTC) was originally developed for evaluating the toxicological
[2]
risk of impurities present at low levels when impurity toxicity data are not available . The concept was
developed to address impurities present in food contact materials, and was then adapted for impurities
[3]to[16][20][21]
in pharmaceutical products . The TTC concept can be used to evaluate constituents
present, or released, at low amounts from a medical device.
2 © ISO 2019 – All rights reserved
4.2 Protectiveness of TTC values
Separate threshold values have been developed to be protective for cancer-based and non-cancer
effects. Threshold values for non-cancer effects are stratified into Cramer Class categories and
[2]
identified constituents can be assigned to a specific Cramer Class based on their chemical structure
[16]
. Although some evaluation schemes that employ the TTC concept use both non-cancer- and cancer-
based threshold values, only one set of TTC values is described in Table 1 to simplify the application of
TTC values for the safety assessment of constituents that may be released from medical devices.
The TTC values defined in this document are based on the Acceptable Daily Intake values derived in
[9][17]
ICH M7 (R1) (2017) for individual mutagenic impurities in pharmaceuticals . The threshold value
derived by ICH for mutagenic impurities in the longest exposure duration category (1,5 µg/d) is lower
than the threshold value assigned to the most protective Cramer Class non-carcinogenic TTC value
[2][9]
(90 µg/d), which is intended to be protective for chronic exposure to non-carcinogens .
[19]
Carcinogens elicit cancer via genotoxic or non-genotoxic mechanisms . The TTC values in Table 1 are
intended to be protective for exposure to carcinogens that exert their effect via either mechanism.
5 Applicability of TTC to medical device constituents
5.1 General
The applicability of the threshold values established by ICH for mutagenic
...
記事のタイトル:ISO/TS 21726:2019 - 医療機器の生物学的評価 - 医療機器成分の生物学的適合性評価における毒性閾値(TTC)の適用 記事の内容:この文書は、医療機器の成分に関して、その医療機器内またはから放出される成分に対する毒性閾値(TTC)の値の選択および一般的な適用について説明しています。この文書のTTC値は、次の目的で使用することができます:- 抽出物中の特定または未特定の成分の最大濃度と比較するために(ISO 10993-18を参照)- 毒性の等価性をサポートするために- 特定の成分の最大暴露量の推定と比較するために(ISO 10993-17を参照) ただし、TTCは十分な毒性データがある成分には適用できません。また、この文書では発がん物質、全身毒性物質、生殖毒性物質に対して保護が提供されるTTC値が設定されています。ただし、この文書には細胞毒性、刺激性、感作性、血液適合性、材料介在性発熱性、医療機器と体との接触部位で起こる局所作用など、医療機器の生物学的評価の一環として評価される他の生物学的評価点に対するTTC値は含まれていません。また、この文書のTTC値は医療機器のガス経路を通じた潜在的な曝露には適用されません。これらの場合には、ISO 18562シリーズを参照してください。また、この文書で提示されるTTC値は関心のある集団の安全性評価には適用されません。
The article discusses ISO/TS 21726:2019, which pertains to the evaluation of biological compatibility of medical devices. It introduces the concept of a threshold of toxicological concern (TTC) for assessing the presence and release of constituents in or on a medical device. The TTC values provided in the document can be used for various purposes such as comparing with maximum concentration levels, supporting toxicological equivalence, and estimating maximum exposure doses. However, it is important to note that TTC is not applicable for constituents with sufficient toxicity data to establish a tolerable intake value. The document specifically addresses the protection against carcinogens, systemic toxicants, and reproductive toxicants. It does not include TTC values for other biological endpoints or potential gas exposure pathways. For these cases, readers are referred to other ISO standards. The TTC values in this document are also not suitable for assessing the safety assessment of a cohort of concern.
제목: ISO/TS 21726:2019 - 의료기기의 생물학적 평가 - 의료기기 성분의 생물적 합성성 평가를 위한 독성 임계치 적용 내용: 이 문서는 의료기기에 존재하는 성분 또는 의료기기로부터 방출되는 성분에 대한 독성 임계치 (TTC) 값을 선택하고 적용하기 위한 기초를 설명한다. 이 문서의 TTC 값은 다음과 같은 용도로 사용될 수 있다: - 추출물 내에 식별 또는 미식별 성분의 최대 농도와 비교하기 위해 (ISO 10993-18 참조) - 독성 동등성을 지원하기 위해 - 식별된 성분의 최대 노출 용량 추정치와 비교하기 위해 (ISO 10993-17 참조) 참고로, 성분은 3.1에서 정의되어 있다. ISO 10993-18은 TTC (µg/d)를 농도 (µg/ml)로 변환하는 방법을 기재하고 있다. TTC는 용해 가능한 요구량 (TI) 값을 도출하기 위한 충분한 독성 데이터가 있는 성분에는 적용할 수 없다 (ISO 10993-17 참조). 이 문서에서 설정된 TTC 값은 암원, 전신 독성물질 및 생식 독성물질에 대해서 보호적이다 (제5절 참조). 이 문서에는 의료기기의 생물학적 평가의 일환으로 평가되는 다른 생물학적 종점에 대한 TTC 값이 포함되어 있지 않다. 예를 들어, 세포독성, 자극성, 감각성, 혈액 적합성, 물질 매개된 발열성, 의료기기와 체액의 접촉 부위에서 발생하는 국소 작용 등이 포함된다. 또한, 이 문서의 TTC 값은 의료기기의 가스 경로를 통한 잠재적 노출에는 적용되지 않는다. 이러한 경우에는 ISO 18562 시리즈를 참조하라. 이 문서에서 제시된 TTC 값은 관심 집단의 안전성 평가에 적용되지 않는다.










Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...